5.0 RESULT / DATA ANALYSIS
Table 1: Absorbance value at 516 nm of Carmoisine Standard at different concentration
Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L.mol−1
. cm−1
)
0 (Blank) 0 0
5 0.342 3.4366 × 104
15 0.739 2.4753 × 104
25 1.106 2.2228 × 104
35 1.455 2.0887 × 104
45 1.792 2.0008 × 104
Table 2: Absorbance value at 516 nm of Unknown Samples at different concentration
Unknown Concentration (ppm) Absorbance (516 nm) Molar absorptivity
(L. mol−1
. cm−1
)
A 27.8 1.005 1.8163 × 104
B 36.4 1.317 1.8179 × 104
Figure 1: Standard calibration curve of Carmoisine
y = 0.0362x + 0.1828
R² = 0.9989
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
0 5 10 15 20 25 30 35 40 45 50
Absorbance(516nm)
Concentration (ppm)
Standart Calibration Curve of Carmoisine
6.0 DISCUSSION
The purpose of this experiment were to determine Lambda Max for Carmoisine sample
(wavelength scan). To determine ɛ the molar absorptivity of Carmoisine and to prepare a serial
dilution and generate a standard calibration graph for sample quantitation (photometric scan).
The Carmoisine standard solution samples were run by using Uv-Vis Spectrophotometer.
Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis)
refers to absorption spectroscopy in the ultraviolet-visible spectral region. This means it uses light
in the visible and adjacent (near-UV and near-infrared (NIR)) ranges. UV-Vis spectrophotometer
is used in the quantitative determination of concentrations of the absorber in the solutions of
transition metal ions and highly conjugated organic compounds (IIT Kanpur, 2012). Besides, UV
spectrophotometer principle follows the Beer-Lambert Law. This law states that whenever a beam
of monochromatic light is passed through a solution with an absorbing substance, the decreasing
rate of the radiation intensity along with the thickness of the absorbing solution is actually
proportional to the concentration of the solution and the incident radiation (Choudhary, A. 2017).
The experiment was started with 100 ppm of stock solution of Carmoisine was prepared
to dilute the stock solution into different concentration which were 5 ppm, 15 ppm, 25 ppm, 35
ppm and 45 ppm with distilled water into 50 ml volumetric flask. The volume needed to prepare
the serial dilution from 100 ppm of stock solution were 2.5 ml, 7.5 ml, 12.5 ml, 17.5 ml and
22.5 ml respectively. The volume of serial dilution were calculated by using the
formula M1V1 = M2V2. Moreover, the unknown sample was prepared with the same method
as serial dilution but the volume of stock solution must be between 2.5 ml to 22.5 ml.
After setting up the UV-Vis, the blank solution was placed into instrument and baseline
was click. After the blank solution became zero, the standard solution of 45 ppm was placed to
another cuvette. The maximum wavelength was recorded at 516 nm. The maximum
wavelength was the wavelength with the higher absorbance. For the photometric scan, the
cuvette was filled with standard solution 5 ppm to 45 ppm to determine the absorbance and
prepared for the standard calibration curve.
According to the result obtained, the absorbance reading showed for the standard
solution 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm were 0.342, 0.739, 1.106, 1.455 and
1.792 respectively. Based on graph plotted, the absorbance (516 nm) was directly proportional
to the concentration (ppm). Because of when the concentration was higher, more radiation is
absorbed and the absorbance also become higher. In addition, for the unknown samples the
absorbance obtained were 1.005 and 1.317. The concentration of unknown samples have found
by using Beer’s Lambert law formula A = εbc with the slope obtained was 0.0362 which
indicates the molar absorptivity of Carmoisine. Thus, the concentration of unknown samples
were 27.8 ppm and 36.4 ppm. Furthermore, the graph showed the R2
value that is a statistical
measure of how close the data are to the fitted regression line (Ogee, A. 2013). The correlation
coefficient varies about +1 to -1. The graph showed the correlation coefficient was 0.9989
which the R2
value was nearest to the +1 which indicates the accurate of the result.
As conclusion, there are several possibilities precaution that should be considered.
Firstly, make sure use the pipette to measure the volume of Carmoisine to get the accuracy.
Secondly, make sure wipe and do not touched the clean sides of cuvette and carrying out the
set-up procedure in the correct order.
REFERENCES
1. Ogee, A. (2013). Regression Analysis: How Do I Interpret R-squared and Assess the
Goodness-of-Fit?. Retrieved from https://blog.minitab.com/blog/adventures- in-
statistics-2/regression-analysis-how-do-i-interpret-r-squared-and-assess-the-goodness-
of-fit.
2. Wenzel, T. (2019). Beer’s Law. Retrieved from
https://chem.libretexts.org/Bookshelves/Analytical_Chemistry/Supplemental_Module
s_(Analytical_Chemistry)/Analytical_Sciences_Digital_Library/Active_Learning/In_
Class_Activities.
3. IIT Kanpur. (2012). UV-Vis Spectrophotometer. Retrieved from
https://www.iitk.ac.in/dordold/index.php?option=com_content&view=category&layo
ut=blog&id=219&Itemid=238.
4.
Result & discussion uv vis (exp 3) - for merge

More Related Content

What's hot

BABY CARE PRODUCTS
BABY CARE PRODUCTS BABY CARE PRODUCTS
Hplc
Hplc Hplc
DENTAL PRODUCTS
DENTAL PRODUCTS DENTAL PRODUCTS
Iodine value, Rancidity, Peroxide value.
Iodine value, Rancidity, Peroxide value.Iodine value, Rancidity, Peroxide value.
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
Priyanka Yadav
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)
GH Yeoh
 
Nano LC and UPLC
Nano LC and UPLCNano LC and UPLC
Nano LC and UPLC
Trilochana Margana
 
Chiral hplc intro
Chiral hplc introChiral hplc intro
Chiral hplc intro
sameerr98
 
Q3A(R2) IMPURITIES IN NEW DRUG SUBSTANCES
Q3A(R2)  IMPURITIES IN NEW DRUG SUBSTANCESQ3A(R2)  IMPURITIES IN NEW DRUG SUBSTANCES
Q3A(R2) IMPURITIES IN NEW DRUG SUBSTANCES
Muhamad Abdalkader
 
hplc
 hplc hplc
hplc
hadeia1
 
General Analytical Method for Milk
General Analytical Method for MilkGeneral Analytical Method for Milk
Uplc
UplcUplc
Analytical evaluation of herbal drugs
Analytical evaluation of herbal drugsAnalytical evaluation of herbal drugs
Analytical evaluation of herbal drugs
Dr. Harish Kakrani
 
HPLC method development
HPLC method developmentHPLC method development
HPLC method development
Amy Mehaboob
 
Factors Affecting Resolution In HPLC.
Factors Affecting Resolution In HPLC.Factors Affecting Resolution In HPLC.
Food analysis using hplc-a class seminar
Food analysis using hplc-a class seminarFood analysis using hplc-a class seminar
Food analysis using hplc-a class seminar
ANIL RAUT
 
qualification of Hptlc
qualification of Hptlcqualification of Hptlc
qualification of Hptlc
Ankush Sule
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Method Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and RobustnessMethod Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and Robustness
labgo
 
Advanced pharmaceutical analysis
Advanced pharmaceutical analysisAdvanced pharmaceutical analysis
Advanced pharmaceutical analysis
Neha Suresh
 

What's hot (20)

BABY CARE PRODUCTS
BABY CARE PRODUCTS BABY CARE PRODUCTS
BABY CARE PRODUCTS
 
Hplc
Hplc Hplc
Hplc
 
DENTAL PRODUCTS
DENTAL PRODUCTS DENTAL PRODUCTS
DENTAL PRODUCTS
 
Iodine value, Rancidity, Peroxide value.
Iodine value, Rancidity, Peroxide value.Iodine value, Rancidity, Peroxide value.
Iodine value, Rancidity, Peroxide value.
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)
 
Nano LC and UPLC
Nano LC and UPLCNano LC and UPLC
Nano LC and UPLC
 
Chiral hplc intro
Chiral hplc introChiral hplc intro
Chiral hplc intro
 
Q3A(R2) IMPURITIES IN NEW DRUG SUBSTANCES
Q3A(R2)  IMPURITIES IN NEW DRUG SUBSTANCESQ3A(R2)  IMPURITIES IN NEW DRUG SUBSTANCES
Q3A(R2) IMPURITIES IN NEW DRUG SUBSTANCES
 
hplc
 hplc hplc
hplc
 
General Analytical Method for Milk
General Analytical Method for MilkGeneral Analytical Method for Milk
General Analytical Method for Milk
 
Uplc
UplcUplc
Uplc
 
Analytical evaluation of herbal drugs
Analytical evaluation of herbal drugsAnalytical evaluation of herbal drugs
Analytical evaluation of herbal drugs
 
HPLC method development
HPLC method developmentHPLC method development
HPLC method development
 
Factors Affecting Resolution In HPLC.
Factors Affecting Resolution In HPLC.Factors Affecting Resolution In HPLC.
Factors Affecting Resolution In HPLC.
 
Food analysis using hplc-a class seminar
Food analysis using hplc-a class seminarFood analysis using hplc-a class seminar
Food analysis using hplc-a class seminar
 
qualification of Hptlc
qualification of Hptlcqualification of Hptlc
qualification of Hptlc
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
Method Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and RobustnessMethod Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and Robustness
 
Advanced pharmaceutical analysis
Advanced pharmaceutical analysisAdvanced pharmaceutical analysis
Advanced pharmaceutical analysis
 

Similar to Result & discussion uv vis (exp 3) - for merge

a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...
guo jun chen
 
Reducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesReducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC Analyses
Shimadzu Scientific Instruments
 
exp 2 adsorbtion from solution
exp 2 adsorbtion from solutionexp 2 adsorbtion from solution
exp 2 adsorbtion from solutionRashid Alsuwaidi
 
Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...
SriramNagarajan18
 
Application of UV-Visible spectroscopy
Application of UV-Visible spectroscopyApplication of UV-Visible spectroscopy
Application of UV-Visible spectroscopy
ROHIT
 
Uv and visible appications
Uv and visible appicationsUv and visible appications
Uv and visible appications
suhasini
 
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
ijtsrd
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometryCody Taylor
 
Titrimetric Methods
Titrimetric Methods Titrimetric Methods
Titrimetric Methods
Shoshi Prasad Shil
 
Serial Dilution Vs Accuracy
Serial Dilution Vs AccuracySerial Dilution Vs Accuracy
Serial Dilution Vs Accuracy
bladon
 
Spectrophotometry Lecture
Spectrophotometry LectureSpectrophotometry Lecture
Spectrophotometry Lecture
ouopened
 
AppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity StudyAppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity StudyColin Merrifield
 
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetricallyTo estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
Mithil Fal Desai
 
Estimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric methodEstimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric method
PRAVIN SINGARE
 
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
iosrphr_editor
 
Scanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docxScanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docx
todd331
 
Experiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical MemoExperiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical MemoVanessa Ferrero
 
Analysis of analgesics and antipyretics.indu
Analysis of analgesics and  antipyretics.induAnalysis of analgesics and  antipyretics.indu
Analysis of analgesics and antipyretics.induhdghcfgfgftf
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...Ninad Phadke
 

Similar to Result & discussion uv vis (exp 3) - for merge (20)

a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...
 
Reducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesReducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC Analyses
 
exp 2 adsorbtion from solution
exp 2 adsorbtion from solutionexp 2 adsorbtion from solution
exp 2 adsorbtion from solution
 
Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...
 
Application of UV-Visible spectroscopy
Application of UV-Visible spectroscopyApplication of UV-Visible spectroscopy
Application of UV-Visible spectroscopy
 
Uv and visible appications
Uv and visible appicationsUv and visible appications
Uv and visible appications
 
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometry
 
B37
B37B37
B37
 
Titrimetric Methods
Titrimetric Methods Titrimetric Methods
Titrimetric Methods
 
Serial Dilution Vs Accuracy
Serial Dilution Vs AccuracySerial Dilution Vs Accuracy
Serial Dilution Vs Accuracy
 
Spectrophotometry Lecture
Spectrophotometry LectureSpectrophotometry Lecture
Spectrophotometry Lecture
 
AppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity StudyAppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity Study
 
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetricallyTo estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
 
Estimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric methodEstimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric method
 
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
 
Scanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docxScanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docx
 
Experiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical MemoExperiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical Memo
 
Analysis of analgesics and antipyretics.indu
Analysis of analgesics and  antipyretics.induAnalysis of analgesics and  antipyretics.indu
Analysis of analgesics and antipyretics.indu
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
 

Recently uploaded

The Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdfThe Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
Nettur Technical Training Foundation
 
Railway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdfRailway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdf
TeeVichai
 
Forklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella PartsForklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella Parts
Intella Parts
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
gdsczhcet
 
6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)
ClaraZara1
 
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdfTop 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
Teleport Manpower Consultant
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Dr.Costas Sachpazis
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
AJAYKUMARPUND1
 
Basic Industrial Engineering terms for apparel
Basic Industrial Engineering terms for apparelBasic Industrial Engineering terms for apparel
Basic Industrial Engineering terms for apparel
top1002
 
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
MdTanvirMahtab2
 
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming PipelinesHarnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Christina Lin
 
Building Electrical System Design & Installation
Building Electrical System Design & InstallationBuilding Electrical System Design & Installation
Building Electrical System Design & Installation
symbo111
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
Kamal Acharya
 
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdfGoverning Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
WENKENLI1
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
Osamah Alsalih
 
Fundamentals of Induction Motor Drives.pptx
Fundamentals of Induction Motor Drives.pptxFundamentals of Induction Motor Drives.pptx
Fundamentals of Induction Motor Drives.pptx
manasideore6
 
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABSDESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
itech2017
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
gerogepatton
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
ydteq
 
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
zwunae
 

Recently uploaded (20)

The Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdfThe Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
 
Railway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdfRailway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdf
 
Forklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella PartsForklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella Parts
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
 
6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)
 
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdfTop 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
Top 10 Oil and Gas Projects in Saudi Arabia 2024.pdf
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
 
Basic Industrial Engineering terms for apparel
Basic Industrial Engineering terms for apparelBasic Industrial Engineering terms for apparel
Basic Industrial Engineering terms for apparel
 
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
 
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming PipelinesHarnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
 
Building Electrical System Design & Installation
Building Electrical System Design & InstallationBuilding Electrical System Design & Installation
Building Electrical System Design & Installation
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
 
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdfGoverning Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
 
Fundamentals of Induction Motor Drives.pptx
Fundamentals of Induction Motor Drives.pptxFundamentals of Induction Motor Drives.pptx
Fundamentals of Induction Motor Drives.pptx
 
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABSDESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
DESIGN AND ANALYSIS OF A CAR SHOWROOM USING E TABS
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
 
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
 

Result & discussion uv vis (exp 3) - for merge

  • 1. 5.0 RESULT / DATA ANALYSIS Table 1: Absorbance value at 516 nm of Carmoisine Standard at different concentration Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L.mol−1 . cm−1 ) 0 (Blank) 0 0 5 0.342 3.4366 × 104 15 0.739 2.4753 × 104 25 1.106 2.2228 × 104 35 1.455 2.0887 × 104 45 1.792 2.0008 × 104 Table 2: Absorbance value at 516 nm of Unknown Samples at different concentration Unknown Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L. mol−1 . cm−1 ) A 27.8 1.005 1.8163 × 104 B 36.4 1.317 1.8179 × 104 Figure 1: Standard calibration curve of Carmoisine y = 0.0362x + 0.1828 R² = 0.9989 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 0 5 10 15 20 25 30 35 40 45 50 Absorbance(516nm) Concentration (ppm) Standart Calibration Curve of Carmoisine
  • 2. 6.0 DISCUSSION The purpose of this experiment were to determine Lambda Max for Carmoisine sample (wavelength scan). To determine ɛ the molar absorptivity of Carmoisine and to prepare a serial dilution and generate a standard calibration graph for sample quantitation (photometric scan). The Carmoisine standard solution samples were run by using Uv-Vis Spectrophotometer. Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis) refers to absorption spectroscopy in the ultraviolet-visible spectral region. This means it uses light in the visible and adjacent (near-UV and near-infrared (NIR)) ranges. UV-Vis spectrophotometer is used in the quantitative determination of concentrations of the absorber in the solutions of transition metal ions and highly conjugated organic compounds (IIT Kanpur, 2012). Besides, UV spectrophotometer principle follows the Beer-Lambert Law. This law states that whenever a beam of monochromatic light is passed through a solution with an absorbing substance, the decreasing rate of the radiation intensity along with the thickness of the absorbing solution is actually proportional to the concentration of the solution and the incident radiation (Choudhary, A. 2017). The experiment was started with 100 ppm of stock solution of Carmoisine was prepared to dilute the stock solution into different concentration which were 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm with distilled water into 50 ml volumetric flask. The volume needed to prepare the serial dilution from 100 ppm of stock solution were 2.5 ml, 7.5 ml, 12.5 ml, 17.5 ml and 22.5 ml respectively. The volume of serial dilution were calculated by using the formula M1V1 = M2V2. Moreover, the unknown sample was prepared with the same method as serial dilution but the volume of stock solution must be between 2.5 ml to 22.5 ml. After setting up the UV-Vis, the blank solution was placed into instrument and baseline was click. After the blank solution became zero, the standard solution of 45 ppm was placed to another cuvette. The maximum wavelength was recorded at 516 nm. The maximum wavelength was the wavelength with the higher absorbance. For the photometric scan, the cuvette was filled with standard solution 5 ppm to 45 ppm to determine the absorbance and prepared for the standard calibration curve.
  • 3. According to the result obtained, the absorbance reading showed for the standard solution 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm were 0.342, 0.739, 1.106, 1.455 and 1.792 respectively. Based on graph plotted, the absorbance (516 nm) was directly proportional to the concentration (ppm). Because of when the concentration was higher, more radiation is absorbed and the absorbance also become higher. In addition, for the unknown samples the absorbance obtained were 1.005 and 1.317. The concentration of unknown samples have found by using Beer’s Lambert law formula A = εbc with the slope obtained was 0.0362 which indicates the molar absorptivity of Carmoisine. Thus, the concentration of unknown samples were 27.8 ppm and 36.4 ppm. Furthermore, the graph showed the R2 value that is a statistical measure of how close the data are to the fitted regression line (Ogee, A. 2013). The correlation coefficient varies about +1 to -1. The graph showed the correlation coefficient was 0.9989 which the R2 value was nearest to the +1 which indicates the accurate of the result. As conclusion, there are several possibilities precaution that should be considered. Firstly, make sure use the pipette to measure the volume of Carmoisine to get the accuracy. Secondly, make sure wipe and do not touched the clean sides of cuvette and carrying out the set-up procedure in the correct order. REFERENCES 1. Ogee, A. (2013). Regression Analysis: How Do I Interpret R-squared and Assess the Goodness-of-Fit?. Retrieved from https://blog.minitab.com/blog/adventures- in- statistics-2/regression-analysis-how-do-i-interpret-r-squared-and-assess-the-goodness- of-fit. 2. Wenzel, T. (2019). Beer’s Law. Retrieved from https://chem.libretexts.org/Bookshelves/Analytical_Chemistry/Supplemental_Module s_(Analytical_Chemistry)/Analytical_Sciences_Digital_Library/Active_Learning/In_ Class_Activities. 3. IIT Kanpur. (2012). UV-Vis Spectrophotometer. Retrieved from https://www.iitk.ac.in/dordold/index.php?option=com_content&view=category&layo ut=blog&id=219&Itemid=238. 4.