PHARM 462
1
2009
European and International regulatory bodies and
their guidelines on different aspects of QA
Body Full name Guidance on
Eurachem Focus for Analytical Chemistry in Europe Method validation
CITAC Cooperation of International Traceability in
Analytical Chemistry
Proficiency testing
Quality Assurance
EA European Cooperation for Accreditation Accreditation
CEN European Committee for Normalization Standardization
IUPAC International Union of Pure & Applied Chem. Method validation
ISO International Standardization Organisation Standardisation
AOAC
ILAC
Association of Official Analytical Chemists
International Laboratory Accreditation Cooperat.
Internal qual. Control
Proficiency testing
Accreditation
FDA US Food and Drug Administration Method validation
USP United States Pharmacopoeia Method validation
ICH International Conference on Harmonization Method validation
2 2009
Method Validation
 Validation of analytical procedures is the process of determining the
suitability of a given methodology for providing useful
analytical data.
J. Guerra, Pharm. Tech. March 1986
 Validation is the formal and systematic proof that a method compiles
with the requirements for testing a product when
observing a defined procedures.
G. Maldener, Chromatographia, July 1989
3
2009
 Method validation is the process of demonstrating that analytical
procedures are suitable for their intended use and that they support
the identity, strength, quality, purity and potency of the
drug substances and drug products
 Method validation is primarily concerned with:
identification of the sources of potential errors
quantification of the potential errors in the method
 An method validation describes in mathematical and quantifiable
terms the performance characteristics of an assay
4
2009
Examples of Methods That Require
Validation Documentation
 Chromatographic Methods - HPLC, GC, TLC, GC/MS, etc.
Pharmaceutical Analysis - In support of CMC.
Bioanalytical Analysis - In support of PK/PD/Clinical Studies.
 Spectrophotometric Methods – UV/VIS, IR, NIR, AA, NMR,
XRD,MS
 Capillary Electrophoresis Methods - Zone, Isoelectric Focusing
 Particle Size Analysis Methods - Laser, Microscopic, Sieving, SEC,
etc.
 Automated Analytical Methods - Robots, Automated Analysis.
5
2009
Considerations Prior to
Method Validation
Suitability of Instrument
 Status of Qualification and Calibration
Suitability of Materials
 Status of Reference Standards, Reagents, Placebo Lots
Suitability of Analyst
 Status of Training and Qualification Records
Suitability of Documentation
 Written analytical procedure and proper approved protocol
with pre-established acceptance criteria
6
2009
Validation Step
 Define the application, purpose and scope of the method.
 Analytes? Concentration? Sample matrices?
 Develop a analytical method.
 Develop a validation protocol.
 Qualification of instrument.
 Qualify/train operator
 Qualification of material.
 Perform pre-validation experiments.
 Adjust method parameters and/or acceptance criteria if necessary.
 Perform full validation experiments.
 Develop SOP for executing the method in routine analysis.
 Document validation experiments and results in the validation report.
7
2009
Purpose of Method Validation
 Identification of Sources and Quantitation of Potential errors
 Determination if Method is Acceptable for Intended Use
 Establish Proof that a Method Can be Used for Decision Making
 Satisfy FDA Requirements
8
2009
What is not Analytical Method
Validation?
 Calibration
The Process of Performing Tests on Individual System
Components to Ensure Proper function
For example) HPLC Detector calibration
 Wavelength Accuracy/ Linear Range/ Noise Level/ Drift
9
2009
 System Suitability
Test to verify the proper functioning of the operating
system, i.e., the electronics, the equipment, the specimens
and the analytical operations.
 Minimum Resolution of 3.0 between the analyte peak and
internal standard peaks
 Relative Standard Deviation of replicate standard injections
of not more than 2.0%
10
2009
11
System Suitability
Sample
Validation
Method
Analyst
Calibration
Pump
Detector
Injector
Data System
2009
Method Life Cycle
12
Validation
Development Optimization
2009
Verification vs. Validation
 Compendial vs. Non-compendial Methods
 Compendial methods-Verification
 Non-compendial methods-Validation requirement
13
2009
Compendial Analytical Procedures
 The Analytical procedures in the USP 25/NF 20 are legally recognized under
section 501(b) of the Federal Food, Drug and Cosmetic Act as the regulatory
analytical procedures for the compendial items. The suitability of these
procedures must be verified under actual conditions of use. When using USP
25/NF 20 analytical procedures, the guidance recommends that information be
provided for the following
characteristics:
 Specificity of the procedure
 Stability of the sample solution
 Intermediate precision
14
2009
Published Validation Guidelines
 1978 Current Good Manufacturing Practices (cGMPs)
 1987 FDA Validation Guideline
 1989 Supplement 9 to USP XXI
 1994 CDER Reviewer Guidance:
Validation of Chromatographic Method
 1995 ICH Validation Definitions:
Q2A, Text on Validation of Analytical procedures
 1997 ICH Validation Methodology:
Q2B, Validation of Analytical Procedures: Methodology
 1999 Supplement 10 to USP 23 <1225>: Validation of Compendial Methods
 1999 CDER “Bioanalytical Method Validation for Human Studies”
 2000 CDER Draft “Analytical Procedures and Method Validation”
15
2009
Regulatory and Compliance
Requirements Review
 Validation of an analytical method is the
process by which it is established, by
laboratory studies, that the performance
characteristics of the method meet the
requirements for the intended analytical
applications
16
USP 23 General
Information <1225>
2009
 The accuracy, sensitivity, specificity, and
reproducibility of test methods employed by the firm
shall be established and documented. Such validation
and documentation may be accomplished in
accordance with 211.194(a)(2).
17
21 CFR PART 211 - CURRENT GOOD MANUFACTURING
PRACTICE FOR FINISHED PHARMACEUTICALS
Subpart I-Laboratory Controls
211.165 Testing and release for distribution (e)
2009
 The objective of validation of an analytical
procedure is to demonstrate that it is suitable
for its intended purpose
18
ICH Guideline for
Industry
Q2A, Text on
Validation of
Analytical
Procedures
March 1995
2009
 In practice, it is usually possible to design the experimental
work such that the appropriate validation characteristics
can be considered simultaneously to provide a sound,
overall knowledge of the capabilities of the analytical
procedure, for instance: Specificity, Linearity, Range,
Accuracy, and
Precision.
19
ICH Guideline for Industry
Q2B, Validation of
Analytical Procedures:
Methodology
2009
Today’s Validation Requirements
20
ICH/USP
GMPs
(legal) FDA
2009
ICH/USP Validation Requirements &
Parameters
 Specificity
 Linearity
 Range
 Accuracy
 Precision
 Repeatability
 Intermediate Precision
 Reproducibility
 Limit of Detection
 Limit of Quantitation
21
ICH
 Specificity
Specificity
 Linearity and Range
Linearity and Range
 Accuracy
Accuracy
 Precision
Precision
 Limit of Detection
Limit of Detection
 Limit of Quantitation
Limit of Quantitation
 Ruggedness
Ruggedness
 Robustness
Robustness
USP
2009
USP Data Elements Required
For Assay Validation
22
Analytical
Performance
Parameter
Assay
Category 1
Assay Category 2
Assay
Category 3
Quantitativ
e
Limit Tests
Accuracy Yes Yes * *
Precision Yes Yes No Yes
Specificity Yes Yes Yes *
LOD No No Yes *
LOQ No Yes No *
Linearity Yes Yes No *
Range Yes Yes * *
Ruggedness Yes Yes Yes Yes
* May be required, depending on the nature of the specific test.
2009
USP Categories
 Category 1: Quantitation of major components or
active ingredients
 Category 2: Determination of impurities or
degradation products
 Category 3: Determination of performance
characteristics
23
2009
ICH Validation Characteristics vs.
Type of Analytical Procedure
24
Type of
Analytical
Procedure
Identificati
on
Impurity testing
Assay
Quantitativ
e
Limit Tests
Accuracy No Yes No Yes
Precision
Repeatability
No Yes No Yes
Interm.
Prec.
No Yes No Yes
Specificity Yes Yes Yes Yes
LOD No No Yes No
LOQ No Yes No No
Linearity No Yes No Yes
2009
Specificity/Selectivity
 Ability of an analytical method to measure the analyte free from
interference due to other components.
 Selectivity describes the ability of an analytical method to differentiate
various substances in a sample
Original term used in USP
Also Preferred by IUPAC and AOAC
Also used to characterize chromatographic columns
 Degree of Bias (Used in USP)
The difference in assay results between the two groups
- the sample containing added impurities, degradation products, related chemical
compounds, placebo ingredients
- the sample without added substances
25
2009
Specificity: Impurities Assay
 Chromatographic Methods
 Demonstrate Resolution
 Impurities/Degradants Available
 Spike with impurities/degradants
 Show resolution and a lack of interference
 Impurities/Degradants Not Available
 Stress Samples
 For assay, Stressed and Unstressed Samples should be
compared.
 For impurity test, impurity profiles should be compared.
26
2009
Forced Degradation Studies
 Temperature (50-60 )
℃
 Humidity (70-80%)
 Acid Hydrolysis (0.1 N HCl)
 Base Hydrolysis (0.1 N NaOH)
 Oxidation (3-30%)
 Light (UV/Vis/Fl)
Intent is to create 10 to 30 % Degradation
27
2009
Linearity
 Ability of an assay to
elicit a direct and
proportional response
to changes in analyte
concentration.
28
2009
Linearity Should be Evaluated
 By Visual Inspection of plot of signals vs. analyte
concentration
 By Appropriate statistical methods
 Linear Regression (y = mx + b)
 Correlation Coefficient, y-intercept (b), slope (m)
 Acceptance criteria: Linear regression r2
> 0.95
Requires a minimum of 5 concentration levels
29
2009
Range
 Acceptable range having linearity, accuracy, precision.
 For Drug Substance & Drug product Assay
 80 to 120% of test Concentration
 For Content Uniformity Assay
 70 to 130% of test Concentration
 For Dissolution Test Method
 +/- 20% over entire Specification Range
 For Impurity Assays
 From Reporting Level to 120% of Impurity Specification for Impurity
Assays
 From Reporting Level to 120% of Assay Specification for Impurity/Assay
Methods
30
2009
Accuracy
 Closeness of the test
results obtained by the
method to the true value.
31
Accuracy
 Should be established across specified range of
analytical procedure.
 Should be assessed using a minimum of 3 concentration
levels, each in triplicate (total of 9 determinations)
 Should be reported as:
 Percent recovery of known amount added or
 The difference between the mean assay result and the accepted
value
32
2009
Accuracy Data Set (1 of 3)
33
Amount
Added (mg)
Amount
Found (mg)
Percent
Recovery
0.0 0.0 ---
50.2 50.4 100.5
79.6 80.1 100.6
99.9 100.7 100.8
120.2 119.8 99.7
150.4 149.7 99.5
2009
Precision
 The closeness of agreement (degree of
scatter) between a series of
measurements obtained from
multiple samplings of the same
homogeneous sample.

Should be investigated using
homogeneous, authentic samples
.
34
2009
Precision… Considered at 3 Levels
 Repeatability
 Intermediate Precision
 Reproducibility
35
2009
Repeatability
 Express the precision
under the same
operating conditions
over a short interval of
time.
 Also referred to as
Intra-assay precision
36
Should be assessed
using minimum of 9
determinations
(3 concentrations/ 3
replicates) or
or
Minimum of 6
determinations at
the 100% level.
2009
Intermediate Precision
37
Express within-
laboratory variations.
Expressed in terms of
standard deviation,
relative standard
deviation (coefficient of
variation) and
confidence interval.
Depends on the
circumstances under
which the procedure is
intended to be used.
Studies should include
varying days, analysts,
equipment, etc.
2009
Repeatability & Intermediate Precision
Day 1 Day 2
100.6 99.5
100.8 99.9
100.1 98.9
100.3 99.2
100.5 99.7
100.4 99.6
38
Grand
Mean = 100.0
RSD = 0.59%
Mean = 100.5
RSD = 0.24%
Mean = 99.5
RSD = 0.36%
2009
Reproducibility
 Definition: Ability reproduce data
within the predefined precision
 Determination: SD, RSD and
confidence interval
 Repeatability test at two different
labs.
Note: Data not required for BLA/NDA
Lab 1 Lab 2 Lab 3
Day
1
Day
2
Day
1
Day
2
Day
1
Day
2
Man
1
Man
2
Man
1
Man
2
Man
1
Man
2
3
Prep
3
Prep
3
Pre
p
3
Prep
3
Pre
p
3
Prep
39
Detection Limit (LOD)/
Quantitation Limit (LOQ)
 LOD
Lowest amount of analyte in a
sample that can be detected
but not necessarily
quantitated.
Estimated by Signal to Noise
Ratio of 3:1.
40
LOQ
Lowest amount of
analyte in a sample that
can be quantified with
suitable accuracy and
precision.
Estimated by Signal to
Noise Ratio of 10:1.
2009
1. Based in Visual Evaluations
- Used for non-instrumental methods
2. Based on Signal-to Noise-Ratio
- 3:1 for Detection Limit
- 10:1 for Quantitation Limit
3. Based on Standard Deviation of the Response and
the Slope
41
LOD and LOQ Estimated by
2009
 S = slope of calibration curve
 s = standard deviation of blank readings or
standard deviation of regression line
Validated by assaying samples at DL or QL
42
DL =
DL =
3.3s
3.3s
QL =
QL =
10s
10s
S
S S
S
LOD and LOQ Estimated by
2009
43
Ybl
LOD LOQ
Statistical estimate of LOD & LOQ
LOD = 3.3 Sbl / b LOQ = 10 Sbl / b
Y = b X + a
2009
 Definition: Capacity to remain unaffected by small but deliberate
variations in method parameters
 Determination: Comparison results under differing conditions with
precision under normal conditions
 Examples of typical variations in LC
 Influence of variations of pH in a mobile phase
 Influence of variations in mobile phase composition
 Different columns (different lots and/or suppliers)
 Temperature
 Flow rate
44
Robustness
2009
Ruggedness
 Degree of reproducibility of test results
under a variety of conditions
 Different Laboratories
 Different Analysts
 Different Instruments
 Different Reagents
 Different Days
 Etc.
 Expressed as %RSD
45
2009
ICH/USP System Suitability
 ICH
 Definition: evaluation of equipment, electronic,
analytical operations and samples as a whole
 Determination: repeatability, tailing factor (T), capacity
factor (k’), resolution (R), and theoretical Plates (N)
46
2009
 USP 23 <621>
 System Suitability Requirements
47
Parameters Recommendations
K’ In general k’ 2.0
≥
R
R > 2, between the peak of interest and the
closest potential interferent (degradant,
internal STD, impurity, excipient, etc…..)
T T 2
≤
N In general N > 2000
Repeatability RSD 2.0% (n 5)
≤ ≥
2009
Re-validation
 When
 Method parameters have been changed
 The scope of the method has been changed
 Synthetic methods have been changed
 Impurity profile has been changed
 What
 Preferably everything. Exceptions should be
scientifically justified
48
2009
How do we Know the expectations of
the FDA
?
 FDA Form 483
 FDA Warning Letters
 Personal Experiences
49
2009
483
Observations
 There was inadequate method validation specificity
data to demonstrate that each method was capable of
distinguishing the active ingredient from its impurities
and degradation products.
 Specificity studies did not include the minimum stress
conditions of acid and base hydrolysis, oxidation,
thermal degradation and photolysis, degradation
schematic for the active ingredient that identifies the
major degradation products
was not included for each product.
50
2009
FDA Waning Letter

On addition to the example of modifying both compendial
methods and customer supplied methods, we also observed
the use of unvalidated in-house methods as well as
unvalidated
modifications to in-house methods
.

A statement indicating that the method has not been
validated in the particular formulation was included in the
certificate of analysis for…use of this statement does not
absolve…from using valid, accurate, and
reproducible methods. (June 2000)
51
2009
FDA Systems Based Inspection:
Laboratory System
52
Method
Validation
13%
Training/Qual.
4%
Stability Program
21%
Inadequate
Records
27%
Controls. General
35%
Feb – July 2002: 212 Inspections (US)
* Reference: Albinus D’ Sa, FDA, CDER Office of Compliance, from AAPS, Nov. 2002
presentation.2009
ICH Update
:
2009 53
A Unique Approach
 International Conference on Harmonisation
(ICH) was created in 1990
 Agreement between the EU, Japan and the
USA to harmonize different regional
requirements for registration of pharmaceutical
drug products
 Unique because joint effort by regulators and
associated pharmaceutical industry trade
associations
2009 54
ICH Objectives
 Identification and elimination of the need to duplicate
studies to meet different regulatory requirements
 More efficient use of resources in the R&D process,
as a consequence
 Quicker access for patients to safe and effective new
medicines
2009 55
Working Groups
SAFETY EFFICACY
QUALITY MULTIDISCIPLINARY
STEERING COMMITTEE
Endorses topics, guidelines and monitors progress
2009 56
Related Site
 www.fda.gov
 www.fda.gov/cder/
 www.waters.com
 www.usp.org
 www.ich.org
 www.aoac.org
 www.pharmweb.net
57
2009

Neutron_Validation-of-analytical-methods_02.ppt

  • 1.
  • 2.
    European and Internationalregulatory bodies and their guidelines on different aspects of QA Body Full name Guidance on Eurachem Focus for Analytical Chemistry in Europe Method validation CITAC Cooperation of International Traceability in Analytical Chemistry Proficiency testing Quality Assurance EA European Cooperation for Accreditation Accreditation CEN European Committee for Normalization Standardization IUPAC International Union of Pure & Applied Chem. Method validation ISO International Standardization Organisation Standardisation AOAC ILAC Association of Official Analytical Chemists International Laboratory Accreditation Cooperat. Internal qual. Control Proficiency testing Accreditation FDA US Food and Drug Administration Method validation USP United States Pharmacopoeia Method validation ICH International Conference on Harmonization Method validation 2 2009
  • 3.
    Method Validation  Validationof analytical procedures is the process of determining the suitability of a given methodology for providing useful analytical data. J. Guerra, Pharm. Tech. March 1986  Validation is the formal and systematic proof that a method compiles with the requirements for testing a product when observing a defined procedures. G. Maldener, Chromatographia, July 1989 3 2009
  • 4.
     Method validationis the process of demonstrating that analytical procedures are suitable for their intended use and that they support the identity, strength, quality, purity and potency of the drug substances and drug products  Method validation is primarily concerned with: identification of the sources of potential errors quantification of the potential errors in the method  An method validation describes in mathematical and quantifiable terms the performance characteristics of an assay 4 2009
  • 5.
    Examples of MethodsThat Require Validation Documentation  Chromatographic Methods - HPLC, GC, TLC, GC/MS, etc. Pharmaceutical Analysis - In support of CMC. Bioanalytical Analysis - In support of PK/PD/Clinical Studies.  Spectrophotometric Methods – UV/VIS, IR, NIR, AA, NMR, XRD,MS  Capillary Electrophoresis Methods - Zone, Isoelectric Focusing  Particle Size Analysis Methods - Laser, Microscopic, Sieving, SEC, etc.  Automated Analytical Methods - Robots, Automated Analysis. 5 2009
  • 6.
    Considerations Prior to MethodValidation Suitability of Instrument  Status of Qualification and Calibration Suitability of Materials  Status of Reference Standards, Reagents, Placebo Lots Suitability of Analyst  Status of Training and Qualification Records Suitability of Documentation  Written analytical procedure and proper approved protocol with pre-established acceptance criteria 6 2009
  • 7.
    Validation Step  Definethe application, purpose and scope of the method.  Analytes? Concentration? Sample matrices?  Develop a analytical method.  Develop a validation protocol.  Qualification of instrument.  Qualify/train operator  Qualification of material.  Perform pre-validation experiments.  Adjust method parameters and/or acceptance criteria if necessary.  Perform full validation experiments.  Develop SOP for executing the method in routine analysis.  Document validation experiments and results in the validation report. 7 2009
  • 8.
    Purpose of MethodValidation  Identification of Sources and Quantitation of Potential errors  Determination if Method is Acceptable for Intended Use  Establish Proof that a Method Can be Used for Decision Making  Satisfy FDA Requirements 8 2009
  • 9.
    What is notAnalytical Method Validation?  Calibration The Process of Performing Tests on Individual System Components to Ensure Proper function For example) HPLC Detector calibration  Wavelength Accuracy/ Linear Range/ Noise Level/ Drift 9 2009
  • 10.
     System Suitability Testto verify the proper functioning of the operating system, i.e., the electronics, the equipment, the specimens and the analytical operations.  Minimum Resolution of 3.0 between the analyte peak and internal standard peaks  Relative Standard Deviation of replicate standard injections of not more than 2.0% 10 2009
  • 11.
  • 12.
  • 13.
    Verification vs. Validation Compendial vs. Non-compendial Methods  Compendial methods-Verification  Non-compendial methods-Validation requirement 13 2009
  • 14.
    Compendial Analytical Procedures The Analytical procedures in the USP 25/NF 20 are legally recognized under section 501(b) of the Federal Food, Drug and Cosmetic Act as the regulatory analytical procedures for the compendial items. The suitability of these procedures must be verified under actual conditions of use. When using USP 25/NF 20 analytical procedures, the guidance recommends that information be provided for the following characteristics:  Specificity of the procedure  Stability of the sample solution  Intermediate precision 14 2009
  • 15.
    Published Validation Guidelines 1978 Current Good Manufacturing Practices (cGMPs)  1987 FDA Validation Guideline  1989 Supplement 9 to USP XXI  1994 CDER Reviewer Guidance: Validation of Chromatographic Method  1995 ICH Validation Definitions: Q2A, Text on Validation of Analytical procedures  1997 ICH Validation Methodology: Q2B, Validation of Analytical Procedures: Methodology  1999 Supplement 10 to USP 23 <1225>: Validation of Compendial Methods  1999 CDER “Bioanalytical Method Validation for Human Studies”  2000 CDER Draft “Analytical Procedures and Method Validation” 15 2009
  • 16.
    Regulatory and Compliance RequirementsReview  Validation of an analytical method is the process by which it is established, by laboratory studies, that the performance characteristics of the method meet the requirements for the intended analytical applications 16 USP 23 General Information <1225> 2009
  • 17.
     The accuracy,sensitivity, specificity, and reproducibility of test methods employed by the firm shall be established and documented. Such validation and documentation may be accomplished in accordance with 211.194(a)(2). 17 21 CFR PART 211 - CURRENT GOOD MANUFACTURING PRACTICE FOR FINISHED PHARMACEUTICALS Subpart I-Laboratory Controls 211.165 Testing and release for distribution (e) 2009
  • 18.
     The objectiveof validation of an analytical procedure is to demonstrate that it is suitable for its intended purpose 18 ICH Guideline for Industry Q2A, Text on Validation of Analytical Procedures March 1995 2009
  • 19.
     In practice,it is usually possible to design the experimental work such that the appropriate validation characteristics can be considered simultaneously to provide a sound, overall knowledge of the capabilities of the analytical procedure, for instance: Specificity, Linearity, Range, Accuracy, and Precision. 19 ICH Guideline for Industry Q2B, Validation of Analytical Procedures: Methodology 2009
  • 20.
  • 21.
    ICH/USP Validation Requirements& Parameters  Specificity  Linearity  Range  Accuracy  Precision  Repeatability  Intermediate Precision  Reproducibility  Limit of Detection  Limit of Quantitation 21 ICH  Specificity Specificity  Linearity and Range Linearity and Range  Accuracy Accuracy  Precision Precision  Limit of Detection Limit of Detection  Limit of Quantitation Limit of Quantitation  Ruggedness Ruggedness  Robustness Robustness USP 2009
  • 22.
    USP Data ElementsRequired For Assay Validation 22 Analytical Performance Parameter Assay Category 1 Assay Category 2 Assay Category 3 Quantitativ e Limit Tests Accuracy Yes Yes * * Precision Yes Yes No Yes Specificity Yes Yes Yes * LOD No No Yes * LOQ No Yes No * Linearity Yes Yes No * Range Yes Yes * * Ruggedness Yes Yes Yes Yes * May be required, depending on the nature of the specific test. 2009
  • 23.
    USP Categories  Category1: Quantitation of major components or active ingredients  Category 2: Determination of impurities or degradation products  Category 3: Determination of performance characteristics 23 2009
  • 24.
    ICH Validation Characteristicsvs. Type of Analytical Procedure 24 Type of Analytical Procedure Identificati on Impurity testing Assay Quantitativ e Limit Tests Accuracy No Yes No Yes Precision Repeatability No Yes No Yes Interm. Prec. No Yes No Yes Specificity Yes Yes Yes Yes LOD No No Yes No LOQ No Yes No No Linearity No Yes No Yes 2009
  • 25.
    Specificity/Selectivity  Ability ofan analytical method to measure the analyte free from interference due to other components.  Selectivity describes the ability of an analytical method to differentiate various substances in a sample Original term used in USP Also Preferred by IUPAC and AOAC Also used to characterize chromatographic columns  Degree of Bias (Used in USP) The difference in assay results between the two groups - the sample containing added impurities, degradation products, related chemical compounds, placebo ingredients - the sample without added substances 25 2009
  • 26.
    Specificity: Impurities Assay Chromatographic Methods  Demonstrate Resolution  Impurities/Degradants Available  Spike with impurities/degradants  Show resolution and a lack of interference  Impurities/Degradants Not Available  Stress Samples  For assay, Stressed and Unstressed Samples should be compared.  For impurity test, impurity profiles should be compared. 26 2009
  • 27.
    Forced Degradation Studies Temperature (50-60 ) ℃  Humidity (70-80%)  Acid Hydrolysis (0.1 N HCl)  Base Hydrolysis (0.1 N NaOH)  Oxidation (3-30%)  Light (UV/Vis/Fl) Intent is to create 10 to 30 % Degradation 27 2009
  • 28.
    Linearity  Ability ofan assay to elicit a direct and proportional response to changes in analyte concentration. 28 2009
  • 29.
    Linearity Should beEvaluated  By Visual Inspection of plot of signals vs. analyte concentration  By Appropriate statistical methods  Linear Regression (y = mx + b)  Correlation Coefficient, y-intercept (b), slope (m)  Acceptance criteria: Linear regression r2 > 0.95 Requires a minimum of 5 concentration levels 29 2009
  • 30.
    Range  Acceptable rangehaving linearity, accuracy, precision.  For Drug Substance & Drug product Assay  80 to 120% of test Concentration  For Content Uniformity Assay  70 to 130% of test Concentration  For Dissolution Test Method  +/- 20% over entire Specification Range  For Impurity Assays  From Reporting Level to 120% of Impurity Specification for Impurity Assays  From Reporting Level to 120% of Assay Specification for Impurity/Assay Methods 30 2009
  • 31.
    Accuracy  Closeness ofthe test results obtained by the method to the true value. 31
  • 32.
    Accuracy  Should beestablished across specified range of analytical procedure.  Should be assessed using a minimum of 3 concentration levels, each in triplicate (total of 9 determinations)  Should be reported as:  Percent recovery of known amount added or  The difference between the mean assay result and the accepted value 32 2009
  • 33.
    Accuracy Data Set(1 of 3) 33 Amount Added (mg) Amount Found (mg) Percent Recovery 0.0 0.0 --- 50.2 50.4 100.5 79.6 80.1 100.6 99.9 100.7 100.8 120.2 119.8 99.7 150.4 149.7 99.5 2009
  • 34.
    Precision  The closenessof agreement (degree of scatter) between a series of measurements obtained from multiple samplings of the same homogeneous sample.  Should be investigated using homogeneous, authentic samples . 34 2009
  • 35.
    Precision… Considered at3 Levels  Repeatability  Intermediate Precision  Reproducibility 35 2009
  • 36.
    Repeatability  Express theprecision under the same operating conditions over a short interval of time.  Also referred to as Intra-assay precision 36 Should be assessed using minimum of 9 determinations (3 concentrations/ 3 replicates) or or Minimum of 6 determinations at the 100% level. 2009
  • 37.
    Intermediate Precision 37 Express within- laboratoryvariations. Expressed in terms of standard deviation, relative standard deviation (coefficient of variation) and confidence interval. Depends on the circumstances under which the procedure is intended to be used. Studies should include varying days, analysts, equipment, etc. 2009
  • 38.
    Repeatability & IntermediatePrecision Day 1 Day 2 100.6 99.5 100.8 99.9 100.1 98.9 100.3 99.2 100.5 99.7 100.4 99.6 38 Grand Mean = 100.0 RSD = 0.59% Mean = 100.5 RSD = 0.24% Mean = 99.5 RSD = 0.36% 2009
  • 39.
    Reproducibility  Definition: Abilityreproduce data within the predefined precision  Determination: SD, RSD and confidence interval  Repeatability test at two different labs. Note: Data not required for BLA/NDA Lab 1 Lab 2 Lab 3 Day 1 Day 2 Day 1 Day 2 Day 1 Day 2 Man 1 Man 2 Man 1 Man 2 Man 1 Man 2 3 Prep 3 Prep 3 Pre p 3 Prep 3 Pre p 3 Prep 39
  • 40.
    Detection Limit (LOD)/ QuantitationLimit (LOQ)  LOD Lowest amount of analyte in a sample that can be detected but not necessarily quantitated. Estimated by Signal to Noise Ratio of 3:1. 40 LOQ Lowest amount of analyte in a sample that can be quantified with suitable accuracy and precision. Estimated by Signal to Noise Ratio of 10:1. 2009
  • 41.
    1. Based inVisual Evaluations - Used for non-instrumental methods 2. Based on Signal-to Noise-Ratio - 3:1 for Detection Limit - 10:1 for Quantitation Limit 3. Based on Standard Deviation of the Response and the Slope 41 LOD and LOQ Estimated by 2009
  • 42.
     S =slope of calibration curve  s = standard deviation of blank readings or standard deviation of regression line Validated by assaying samples at DL or QL 42 DL = DL = 3.3s 3.3s QL = QL = 10s 10s S S S S LOD and LOQ Estimated by 2009
  • 43.
    43 Ybl LOD LOQ Statistical estimateof LOD & LOQ LOD = 3.3 Sbl / b LOQ = 10 Sbl / b Y = b X + a 2009
  • 44.
     Definition: Capacityto remain unaffected by small but deliberate variations in method parameters  Determination: Comparison results under differing conditions with precision under normal conditions  Examples of typical variations in LC  Influence of variations of pH in a mobile phase  Influence of variations in mobile phase composition  Different columns (different lots and/or suppliers)  Temperature  Flow rate 44 Robustness 2009
  • 45.
    Ruggedness  Degree ofreproducibility of test results under a variety of conditions  Different Laboratories  Different Analysts  Different Instruments  Different Reagents  Different Days  Etc.  Expressed as %RSD 45 2009
  • 46.
    ICH/USP System Suitability ICH  Definition: evaluation of equipment, electronic, analytical operations and samples as a whole  Determination: repeatability, tailing factor (T), capacity factor (k’), resolution (R), and theoretical Plates (N) 46 2009
  • 47.
     USP 23<621>  System Suitability Requirements 47 Parameters Recommendations K’ In general k’ 2.0 ≥ R R > 2, between the peak of interest and the closest potential interferent (degradant, internal STD, impurity, excipient, etc…..) T T 2 ≤ N In general N > 2000 Repeatability RSD 2.0% (n 5) ≤ ≥ 2009
  • 48.
    Re-validation  When  Methodparameters have been changed  The scope of the method has been changed  Synthetic methods have been changed  Impurity profile has been changed  What  Preferably everything. Exceptions should be scientifically justified 48 2009
  • 49.
    How do weKnow the expectations of the FDA ?  FDA Form 483  FDA Warning Letters  Personal Experiences 49 2009
  • 50.
    483 Observations  There wasinadequate method validation specificity data to demonstrate that each method was capable of distinguishing the active ingredient from its impurities and degradation products.  Specificity studies did not include the minimum stress conditions of acid and base hydrolysis, oxidation, thermal degradation and photolysis, degradation schematic for the active ingredient that identifies the major degradation products was not included for each product. 50 2009
  • 51.
    FDA Waning Letter  Onaddition to the example of modifying both compendial methods and customer supplied methods, we also observed the use of unvalidated in-house methods as well as unvalidated modifications to in-house methods .  A statement indicating that the method has not been validated in the particular formulation was included in the certificate of analysis for…use of this statement does not absolve…from using valid, accurate, and reproducible methods. (June 2000) 51 2009
  • 52.
    FDA Systems BasedInspection: Laboratory System 52 Method Validation 13% Training/Qual. 4% Stability Program 21% Inadequate Records 27% Controls. General 35% Feb – July 2002: 212 Inspections (US) * Reference: Albinus D’ Sa, FDA, CDER Office of Compliance, from AAPS, Nov. 2002 presentation.2009
  • 53.
  • 54.
    A Unique Approach International Conference on Harmonisation (ICH) was created in 1990  Agreement between the EU, Japan and the USA to harmonize different regional requirements for registration of pharmaceutical drug products  Unique because joint effort by regulators and associated pharmaceutical industry trade associations 2009 54
  • 55.
    ICH Objectives  Identificationand elimination of the need to duplicate studies to meet different regulatory requirements  More efficient use of resources in the R&D process, as a consequence  Quicker access for patients to safe and effective new medicines 2009 55
  • 56.
    Working Groups SAFETY EFFICACY QUALITYMULTIDISCIPLINARY STEERING COMMITTEE Endorses topics, guidelines and monitors progress 2009 56
  • 57.
    Related Site  www.fda.gov www.fda.gov/cder/  www.waters.com  www.usp.org  www.ich.org  www.aoac.org  www.pharmweb.net 57 2009