Frequency of Calibration
1 HIGH - VOLUME SAMPLER
Parts Calibration Frequency
Blower By orifice calibrator Once in six months or
whenever . the blower is
replaced or reminded
Rotameter By soap bubble method Once in six months
Time Totalizer By calibrated clock Once in six months.
STACK MONITORING KIT
Pitot Tube Against STD. Pitot tube
with the help of wind tunnel
After every 10-15
monitoring
Rotameter ( Particulate ) (1) By STD. Rotameter
(2) Water Displacement
o y b i o i 11
(3) Calibrated dry gas
meter
After every 10-15
monitoring
Rotameter ( Gases ) By soap bubble method After every 10-15
monitoring
Pyrometer By Calibrated Pyrometer &
thermocouple
After every 10-15
Monitoring
Stop Watch
i
By Calibrated Clock After every 10-15
Monitoring
I Pressure Gauge By Calibrated Gauge After every 10-15
Monitoring
| Thermometer
|
By Calibrated thermometer After every 10-15
Monitoring
If the parts are not calibrated then the flow measured will not be correct and
subsequently the measured concentration will not be actual representative
concentrations.
i—•jgntrri—r- • - i r - r - T m i

Frequency of calibration

  • 1.
    Frequency of Calibration 1HIGH - VOLUME SAMPLER Parts Calibration Frequency Blower By orifice calibrator Once in six months or whenever . the blower is replaced or reminded Rotameter By soap bubble method Once in six months Time Totalizer By calibrated clock Once in six months. STACK MONITORING KIT Pitot Tube Against STD. Pitot tube with the help of wind tunnel After every 10-15 monitoring Rotameter ( Particulate ) (1) By STD. Rotameter (2) Water Displacement o y b i o i 11 (3) Calibrated dry gas meter After every 10-15 monitoring Rotameter ( Gases ) By soap bubble method After every 10-15 monitoring Pyrometer By Calibrated Pyrometer & thermocouple After every 10-15 Monitoring Stop Watch i By Calibrated Clock After every 10-15 Monitoring I Pressure Gauge By Calibrated Gauge After every 10-15 Monitoring | Thermometer | By Calibrated thermometer After every 10-15 Monitoring If the parts are not calibrated then the flow measured will not be correct and subsequently the measured concentration will not be actual representative concentrations. i—•jgntrri—r- • - i r - r - T m i