SlideShare a Scribd company logo
1 of 40
Download to read offline
SOUTHCOASTAIR QUALITYMANAGEMENTDISI'RICI'
DETERMINATIONOFNITROGEN OXIDEEMISSIONS
FROM STATIONARY SOURCES
TECHNICALSUPPORTSERVICES
APPLIED SCIENCEAND TECHNOLOGY
MARCH 1989
D E T E ~ T I O ~OF HITROGEN OXIDE EKISSIOblS
FROM STATZObfaRY SOURCES
TABLE OF CONTENTS
Section
1, Overview
1.1 Principle 

1.2 Applicability 

2 , F i e l d P r o c e d u r e s
2.1 Sampling Apparatus 

2.2 Sampling Reagents 

2.3 Preparation of Sampling Equipment 

2.4 Sampling Procedure 

3, L a b o r a t o r y Procedures
3.1 Apparatus 

3.2 Reagents 

3.3 Pretest Preparation 

3.4 Preparation of Sampling Collection Flask 

3.5 Sample Collection Flask Leak Check 

3.6 Sample Recovery 

3.7 Analysis
3.8 Calculation and Reporting 

3.9 Calibration 

4. Engineering Calculations and Reporting
5. Alternative Laboratory Procedures
5.1 Apparatus 

5.2 Reagents 

5.3 Preparation of Sample Collection Flask 

5.4 Sample Collection Flask Leak Check 

5.5 Recovery 

5.6 Sample Analysis 

5.7 Calibrations 

5.8 Calculations 

5.9 Quality Control
Section 1 of 5 

1. overview' 

1-1 Principle
A grab sample is collected in an evacuated flask 

containing a dilute sulfuric acid-hydrogen 

peroxide absorbing solution, and tpe nitrogen 

oxides except nitrous oxide are measured 

colorimetrically using the phenoldisulfonic acid 

(PDS) procedure or ion chromatograph procedures 

(see Section 5 ) .
1.2 Applicability 

This method measures nitrogen oxides (NOx)
emitted from stationary sources, Its range is 2 

to 400 mg NOx (as NOZ) per dry standard cubic 

meter, without diluting the sample.
DETERMIHATIOX OF OXIDE EMISSIOHS
FROM STATIOHBRP SOURCES
Section 2 of 5
2. Field Procedures
2.1 Sampling Apparatus
Sample collection requires the equipment l i s t e d
below and shown i n Figure 7.1-1.
T
a- Probe
~ o r o s i l i c a t eglass tubing, sufficiently
heated t o prevent water condensation and
equipped with an in-stack or out-stack f i l t e r
t o r&ove particulate matter. (A plug of
glass wool'is satisfactory for t h i s purpose.)
Stainless steel or Teflon tubing also may be
used for the probe. Heating is not necessary
i f the probe remains dry during the purging
-period, or less than 2 percent of the NOx
present is NO2 .
b. Collection Flask 
'
Two-liter borosilicate, round bottom flask, 

with short neck and 24/40 standard taper 

opening, protected against implosion or 

breakage with a T-bore stopcock connected to 

a 24/40 standard taper joint. Alternatively, 

this flask may have the short neck replaced 

with 48 mm (1 1/2 in.) long 12 mm (1/2 in.)
glass tubing sealed with rubber tubing and 

screw clamp. Flasks must be calibrated to 

the nearest 10 m l .
No difference in NO, results has been 

observed between the use of rubbertubing and 

the use of ground joints. Ground glass joints 

tend to have more leakage problems than do 

rubber tubing and screw clamps. When tubing 

is used, the surface area exposed to the 

sample should be minimized. See Section 3.9 

for calibration procedure. 

c. Vacuum Line 

Tubing capable of withstanding a vacuum of 75 

mm (3 in.) Hg absolute pressure, with "Tn
connection and T-bore stopcock o r rubber
tubing and s c r e w clamp.
d. Vacuum Gauge
The collection flask must be equipped with a
vacuum gauge capable of measuring t o 760 mm
(30 in,) Hg with 2.5 mm (0.1 in,) Hg
division,
e. One-way Squeeze Bulb
f, Stopcock and Ground Joint Grease
A high vacuum, high temperature
chlorofluorcarbon grease is required when
ground glass joints are used, Halocarbon 25-
SS is effective,
2 - 2 Sampling Reagents
a. Sulfuric Acid-Hydrogen Peroxide Solution
See Section 3.2.1 f o r preparation procedure.
The solution should be prepared daily and not
exposed t o extreme heat o r direct sunlight-
2.3 Preparation of Sampling Equipment
This may be done in the laboratory prior t o the
test. U s e two-liter round bottam flasks, Add 25
m l of absorbing solution t o each flask- Evacuate
the flask t o the vapor pressure of the solution,
tightly close the s c r e w c l a p and insert a solid
glass plug (optional) into the open end of the
tubing u n t i l ready f o r sampling,
2.4 Sampling Procedure
Assemble the sampling t r a i n a s shown i n Figure
7-1-1 and place the probe a t t h e sampling point-
(Select the sampling point a s i n Method 3-1,
Section 2.2.) Make sure a l l f i t t i n g s are tight
and leak-free, and i f necessary a l l glass joints
have been properly greased with a high-
temperature, chlorfluorocarbon-based stopcock 444,
grease.
Turn the flask valve and the pump to the
"evacuatem1positions. Evacuate the flask to a
pressure approaching the vapor pressure of water
at the existing temperature. Turn the pump valve
to its "ventn position and turn off the pump.
Check for leakage by observing the manometer for
any pressure fluctuation. A variation more than
10 mm (0.4 in.) Hg over a period of one minute is
-
unacceptable. Do not use the flask until the
leakage problem is'wrrected. If flaskk are
evacuated prior to setup, the evacuation steps
are unnecessary.
Turn the flask valve counterclockwise to its
"purgen position. Purge the probe and the vacui
tube using the squeeze bulb. If condensation
occurs in the probe and the flask valve area,
heat the probe and purge until the condensation
disappears. Turn the pump valve to its nventn
position. Turn the flask valve clockwise to its
"evacuaten position and record the difference in
the mercury levels in the manometer. The absolute
internal pressure in the flask (Pi) is equal to -=%
the barometric pressure, minus the manometer 

reading. 

Immediately turn the flask valve to the wsamplen 

position and permit the gas to enter the flask 

until pressure in the flask and sample line (i.e. 

duct stack) are equal. (Vacuum gauge should read 

near zero.) This will usually require about 15 

seconds; a longer period indicates a mplugn in 

the probe, which must be corrected before 

sampl&g is continued. 

After collecting the sample, turn valve to its 

npurgewposition and disconnect the flask 

sampling train. Shake the flask for at least 5 

minutes. Alternatively, open or.closethe screw 

clamps in the appropriate manner to achieve the 

same results, i.e. (1) purge, (2) vacuum check, 

(3) sample, (4) seal flask, (5) shake, 

When an alternate setup is used, perform all the 

above operations by opening and closing the 

appropriate screw clamps, 

Record data on a form similar to Figure 7.1-2.
DETERKINATIOH OF 10ITBOOE# OXIDE EKI88IOEI8
FROM STATIOXARY SOURCES
Section 3 of 4 

3. Laboratory Procedures 

3..1 Apparatus 

. . 3.1 .1 Sampling Apparatus
a. Probe, Collection Flask, vacuum' Line, 

Vacuum Gauge, Squeeze Bulb 

See Section 2.1.
b. Pipet 

Class A, 25 ml 

3.1.2 Sample Recovery 

a. Graduated Cylinder, 50 ml with 1 m l
Divisions
b. Polyethylene Wash Bottle 

c. Glass Stirring Rod 

d. Test Paper for Indicating pH, 

Range 7 to 14 

e. Mercury Manometer, or Calibrated 

Electronic Manometer, with Fitting for 

Sample Bulb Attachment 

f. Thermometer 

Ambient range, OC, traceable to NBS 

thermometer. 

3.1.3 Analysis 

For the analysis, including calibration, 

the following equipment is needed: 

a. volumetric Pipets 

Class A, or a microburet to measure-2, 

4, 6, 8, 10, 12, 16 and 20 ml of 

working calibration solution; one loo 

ml, class A pipet.
b. Porcelain Evaporating Dishes
250 m l capacity with lip for pouring,
one for each sample and each standard.
The Coors NO-45006 (shallow-form, 196
ml) is satisfactory.
c. 'Steam Bath
Low-temperature oven or
thermostatically controlled hot plates
-kept below 70°c ( 1 5 8 ~ ~ )are acceptable
alternatives.
d. Dropping Pipet or Dropper
Three required.
e. Glass Rod
One for each sample, blank, and
standard.
f. Graduated Cylinder
100 ml with 1 m l divisions.
g. volumetric Flasks 

Class A, 1000 ml (four), 

h, Spectrophotometer 

To measure absorbance at 405 nm, using
a 10 mm cell, 

i, Graduated Pipet, 10 ml, 0-1 ml 

divisions 

j, Test Paper for Indicating pH, Range 7 

to 14 pH 

k. Analytical Balance 

To measure to within 0.1 mg. 

, .
1, Centrifuge 

To remove solid contained in sample, 

m. Graduated Centrifuge Tube, 50 ml, 1 ml 

Divisions
3-2 Reagents 

Unless otherwise indicated, it is intended that 

all reagent conform to the specification 

established by the Committee on Analytical 

Reagents of the American Chemical Society, where 

such specifications are available; otherwise, use 

the best available grade. 

3.2.1 Sampling 

To prepare the absorbing solution, 

cautiously add 2.8 ml concentrated H2S04 

-Y
to 1 liter of water. Mix well and add 2 

ml of 30 percent H202. The absorbing 

solution should be prepared daily. Do not 

expose to extreme heat or direct sunlight. 

3.2.2 Sample Recovery 

a. Sodium Hydroxide (NaOH), 1N
Dissolve 40 g NaOH in water and dilute 

to 1 liter.
b. Water 

Deionized, distilled to conform to 

ASTM specification D1193-77, Type 3.
At the option of the chemist, the 

KMn04 test for oxidizable organic 

matter may be omitted when high 

concentrations of organic matters are 

not expected to be present. Reference 

to water throughout this method 

implies deionized, distilled water, 

3.2-3 Analysis 

a. Fuming Sulfuric Acid (H2S04) 

15 to 18 percent, by weight, free SO3. 

HANDLE WITH CAUTION. Used to make 

phenoldisulfonic acid (PDSA) reagent: 

See Section 3.2.3.h. 

b. Phenol 

White solid. Used to make 

phenoldisulfonic acid (PDSA) reagent: 

See Section 3.2.3.h.
c. S u l f u r i c Acid ' (H2S04)
Concentrated, 95 percent minimum
assay. Handle with care.
d. Potassium N i t r a t e (KN03)
D r i e d a t 105 t o 1 1 0 ~ ~(220 t o 2 3 0 ~ ~ )
f o r a minimum of 2 hours j u s t p r i o r t o
preparation of standard solution.
e. Standard KN03 Solution
Dissolve exactly 0.5495 g of d r i e d
KN03 i n w a t e r and d i l u t e t o 1 liter
w i t h water i n a 1000 m l volumetric
flask.' One m i l l i l i t e r of t h i s
standard solution contains the
.equivalent o f 0.250 m g of NO2.
..
f . Working Standard KN03 S o l u t i o n '
Dilute 100 m l of the standard solution
t o 1000 m l with water. One m i l l i l i t e r
of the working standard solution is -'%
equivalent t o 0.025 mg of NO2.
7.1-14
g. Water
See Section 2.2.2.
h. Phenoldisulf onic Acid Solution
Dissolve 25 g of pure white phenol i n
150 m l concentrated H2S04, and heat a t
1 0 0 ~ ~( 2 1 2 ~ ~ )f o r 2 hours. Store i n a
dark stoppered bottle, Protect from
ambient moisture a t a l l t i m e s .
Phenoldisulfonic acid purchased from a
commercial source should be rejected
i f blank absorbance is greater than
0,030, and i f linearity is not
obtained during calibration (see
Section 3,8) ..
3.3 Pretest Preparaticn
A l l equipment; including t h e balance, steam bath,
centrifuge, spectrophotometer, cleaned glassware,
reagents, and safety equipment should be checked
f o r readiness before conducting .the following
procedure,
3.4 Preparation of Sample Collection Flask
Use two-liter round bottom flasks for sampling.
Add 25 m l of absorbing solution to each flask.
Reserve a portion of this reagent for a blank and
for control preparation. Evacuate the flask to
the vapor pressure of the solution, tightly close
the screw clamp, and insert a solid glass plug
(optional) into the open end of the tubing. Pack
the flasks in a sturdy box for transport to the
field, Prepare fresh flasks each day.
NOTE: No difference in NOx results have been
observed with use of rubber tubing as compared to
T
glass joints, In addition, rubber tubing has
been virtually leak-free as compared to glass
joints.
3.5 Sample Collection Flask Leak Check
Check for sample flask leakage during the flask
evacuation step (see Section 3.4). The contents
of the flask will boil when evacuated if no leak
is present. More leak checks may be performed in
the field with a manometer or vacuum gauge,
though these checks may be avoided by sampling in
triplicate or quadruplicate. The analytical
results will indicate which flasks may have 

leaked and should be rejected. 

Check the flask screw clamps (if used) as soon as 

flasks are returned to the laboratory. Record 

the presence of slightly loose clamps for 

consideration in comparing flask analytical 

results. Very loose clamps are clear cause for 

rejection. 

3.6 Sample Recovery 

Shake the sealed collection flasks for 15 minutes 

on a mechanical shaker, and rotate frequently to 

provide a thorough scrubbing action. Allow the 

flasks to stand overnight (or at least 16 hours) 

to ensure complete oxidation and absorption of 

the oxides of nitrogen by the solution. When 

absorption is complete, the oxides of nitrogen 

(except nitrous oxide) will be converted to HN03. 

Record the temperature and absolute gas pressure 

in each flask. Rince the contents into a 200 ml 

evaporating dish, using three 15 ml portions of 

water. Prepare a blank using the same amount of 

absorbing solution and wash water as used for 

each sampling flask.
3.7 Analysis 

Add 1 N sodium hydroxide to.the solution in each 

evaporating dish, including the blank, until it 

is barely alkaline (<pH 9) to litmus paper. 

Evaporate the solution in the evaporating dish to 

. dryness on a steam bath and allow to cool. Add 2
m l phenoldisulfonic acid solution to the dried
residue and titrate thoroughly with a glass rod. 

Make sure the solution contacts all the residue. 

Add 1 ml water and four drops of concentrated 

H2S04. Heat the solution on a steam bath for 3
minutes with occasional stirring. Allow the 

solution to cool, add 20 ml'water, mix well by 

stirring, and add 10 ml of concentrated NHqOH, 

dropwise with constant stirring. Transfer the 

solution to a 50 ml graduated centrifuge tube, 

washing the evaporating dish three times with 5 

m l portions of water. Dilute to the mark with
water, stir thoroughly, and centrifuge. Transfer 

the clear solution to a spectrophotometer cell. 

Measure absorbance of each solution against the 

reagent blank at a wavelength of 405 nm. .If a 

greater dilution of the sample is required, 

dilute the blank to the same volume. Thinner 

cells or dilution blocks may be used. Process
and dilute each sample determination in a manner 

similar to that used in preparing the 

corresponding concentration range of the standard 

curves. In each case, subtract the bldnlr reading 

from the sample reading. The dilution used for 

obtaining the calibration curves in the ranges 

given may have to be varied, depending on the 

make of spectrophotometer used. 

Determine the weight of NOZ, in milligrams, from 

the light absorption reading on the calibration 

curve corresponding to dilution used. See 

Section 3.9 for calibrations. 

3.8 Calculations and Reporting 

Net ODxCalipration Factom515x760xT 

NO2 =
Net Pressure x Net Volume x 288.6 

where:
N02 = NOx as NO2, ppm
Net OD = Sample light absorption - blank
light absorption 

Calibration = pg N02/OD, from calibration
Factor c u m e
T = Temperature in the flask, OK
Net = Measured pressure in the flask,
Pressure minus the water vapor pressure 

at temperature T, mm Hg 

Net Volume = Volume of flask - 25 m l (volume
of absorbing solution in the 

flask), liters
Report NOx as NO2 (ppm), using three significant 

figures. Discard decimal digits. Round the 

third digit to the nearest 5. 

3.9 Calibration 

3.9.1 PDSA 

For a 0-10 ppm calibration curve, use 

pipets or a microburet to transfer 0.0 

(blank), 2.0, 4.0, 6.0, 8.0, 10.0, 12.0, 

16.0, and 20.0 ml of potassium nitrate 

working standard solution into separate 

200 ml evaporating dishes, and add 25 ml
of absorbing solution to each evaporating 

dish. Follow as directed in Section 3.7.
Read the light absorption of each solution 

and subtract the blank reading. If 

absorption is too high, dilute 

appropriately with distilled water. 

Construct the curve for the concentration 

range by plotting the net light absorption 

of the solution as ordinate against 

milligrams of NO2 as abscissa, or use a 

calculator programmed for linear 

regression analysis: 

where:
m = slope
b = intercept
3.9.2 Flask Calibration 

The temperature of the water is measured. 

Volume is calculated according to the 

formula. 

where:
v = Volume of the bulb, m l
W1 = Weight of flask and water, g
W2 = Weight of empty fiask, g
Dt = Density of water, g/ml
at temperature, t
4
DETERHINATION OF ~ T R O G ~OXIDE ZXISSIOHS 

PBOH STAT1Ob;QBY 80uBcE8
Section 4 of 5 

Engineering Calculations and Reporting
When mass.emission rate of NO, is required, follow this 

calculation. 

lb of NOx per hour = A-B.C-D
-- (ppm NO^) (46)(1.58 x (dscfm)
A -- NOX concentration in ppm
B -- Molecular weight of NO2, 46 lb/lb-mole
C -- Conversion factor, 1.58 x loo7
-- (loo6 ppm)(60 min./hr) (379 cu ft/lb-

mole) 

D -- Dry standard stack gas flow rate per
minute measured, dscfm (follow Method 2.1)
DETERKINZATIO~QOF SITRoGEH OXIDE EXISSIOWS
PBOM STATIONARY~S O W S
Section 5 of 5
5 - Alternative Laboratory Procedures .
5.1 Apparatus 

-
5.1.1 Preparation of Sample Collection Flask 

See Section 3.4- 

Same as Section 2-1 b. 

5.1-2 Sample Recovery
a. Class A 50 ml Volumetric Flask 

One for each sample and blank-

b, Polyethylene Wash Bottle
c, Manometer 

Capable of reading to Imm within the 0 

to 900mm Hg range. 

d, NBS Traceable Thermometer
e. Sample Shaker 

5,1.3 Analysis
. a, Volumetric Pipets
Class A, one 10 m l for each sample and
blank, 100 m l for working standard
preparation, 

10 ml, 15 ml, 20 ml, 25 ml, 50 m l , and
graduated 5 m l Class A pipets, or
microburet for calibration standard 

dilutions, 

b, volumetric Flasks 

Class A, one 50 ml flask for each
sample and blank, two 50 ml for each
calibration standard, and tvo 100 m l
for stock and vorking standard 

preparation. 

c. Ion Chromatograph 

Capable of detecting and resolving 0.1 

ppm nitrate in the presence of dilute 

(<O.OlN) sulfuric acid. 

d. Recorder or Integrator 

e. Analytical Balance
Capable of veighing to 0.1 mg. 

5.2 Reagents 

Unless otherwise indicated, it is intended that 

all reagents conform to the specification 

established by the Committee on Analytical 

Reagents of the American Chemical Society, where 

such specifications are available, otherwise use 

the best available grade.
5.2.1 Sampling
Absorbing solution. Same as Section
Sample Recovery
See Section 3.2.2.
Analysis
a. Potassium N i t r a t e (KN03)
Dried a t 105 t o 1 1 0 ~ ~f o r a minimum of
2 hours, -and desiccated t o room
temperature just before use.
b. Standard KN03 Solution
0.5495 g dried RN03 dissolved i n 1
l i t e r of w a t e r .
c- Working Standard KN03 Solution
D i l u t e 100 m l of KN03 standard
solution t o 1000 m l with water.
d. Water
See section 3.2.2.
5.3 Preparation of Sample Collection Flask
See Section 3.4.
5.4 Sample Collection Flask Leak Check
See Section 3.5.
5.5 Sample Recovery
Shake samples and allow to stand overnight as in
Section 3.6. Measure and record the temperature
and absolute gas pressure in each flask. Recover .
the contents of each flask into a 50 m l
volumetric flask with at least three rinsings,
then dilute to volume and mix well.
5. 6 Sample Analysis
Dilute a 10 m l aliquot of each sample to 50 ml.
Set the instrument response as close as possible
to maximum on the highest standard, then run the
calibration curve, the reagent blank, and each 

sample once. Dilute samples that are too high 

into range. Samples that are too low require an 

additional standard dilution and more sensitive 

instrument settings. Run a control for each set 

of samples or each ten samples, whichever is 

fewer, Rerun the calibration curve at the end of 

the analysis. 

5.7 Calibration 

A five-point calibration curve in O.OlN H2S04 is 

made from the working standard as follows: 

2 m l KN03 working standard solution diluted to 50
ml. Take 2 ml of the working solution and 5 ml
of 0-lN H2S04, and dilute to 50 ml with water.
5 ml KN03 working standard solution diluted to 50
ml. Take 5 ml of the working solution and 5 ml
of 0,lN H2S04, and dilute to 50 ml with water.
10 ml KN03 working standard solution diluted to
50 ml- Take 10 m l of the working solution and 5
ml of O.lN H2S04, and dilute to 50 ml with water.
'15 ml,RN03 working standard solution diluted to q.
50 ml. Take 15 ml of the working solution and 5 

ml of O.lN H2S04, and dilute to 50 ml with water. 

25 ml KN03 working standard solution diluted to 

50 ml. Take 25 m l of the working solution and 5
ml of 0.1N H2S04, and dilute to 50 ml with water. 

This corresponds to a sample range of 50 pg NO2 

to 625 pg NO2 per sample or approximately 12 to
150 ppm. 

The calibration curve must bracket all sample 

1*, 

concentrations. Dilute a sample that is too high 

to fall within the range of the calibration 

curve. For samples that are too low, make 

another standard solution using the 

working standard solution (25 pg), delete the
highest standard (625 pg), and maximize the
instrument response on the next highest standard 

(375 pg) . Rerun the analysis.
Calculate linear regression of total pg NO2 and
peak height using all standard points. Remake 

and rerun any standard point that deviates by 7 

percent or more from expected. A change of T
instrument response over the course of the run by
10 percent or more invalidates the analysis. A 

difference of the measured control concentration 

from the theoretical concentration by 10 percent 

or more invalidates the analysis. 

5.8 Calculations .
Calculate the NO2 in pg for each sample using the 

linear regression, expressed as: 

where:
m = Slope
b = Intercept of the regression
h = Sample peak height
Then calculate NO2 as ppm for each sample, using 

the following equation: 

pg NO2 X 0.515 X 760 X T
NO2 =
Net Pressure x Net Volume x 288.6
where:
Pg NO2 = Micrograms of NO2 of the sample
being calculated 

T = , Temperature of the flask, OK
Net Pressure = Measured pressure of the flask,
minus the water pressure, at 

temperature TI mm Rg 
 ,
Net Volume = Volume of the flask, liters
Report NOx as NO2 (ppm) using three significant
figures.
Round the third digit to the nearest 5. 

5 - 9 ~ualityControl
Run suitable blank and control samples with each 

set of samples or each ten samples, whichever is 

fewer. Blank consists of reagent reserved from 

sample preparation.
Suitable control consists of an EPA.QA vial, 

prepared as directed, and analyzed with the 

samples, Measured value must be within 10 

percent of the theoretical value. A control that 

falls outside of the range requires instrument 

recalibration and reanalysis of the sample(s). 

Change in instrument response of more than 10 

percent from the beginning to the end of the run 

also requires instrument recalibration and 

reanalysis of the sample(s),
Figure 7.1-1
N4, Sampling Train
Figure 7.1-2
Gas Sampling Data Sheet

More Related Content

What's hot

Water test methods
Water test methodsWater test methods
Water test methodsSoumen Basu
 
Automated Nitrogen Drying
Automated Nitrogen DryingAutomated Nitrogen Drying
Automated Nitrogen DryingNathan Schaefer
 
Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)Bakhtiar Mahmood
 
Steam and water analysis system 1
Steam and water analysis system 1Steam and water analysis system 1
Steam and water analysis system 1Sanjana Khound
 
Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)khashayar Ataee
 
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNET
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNETItem 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNET
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNETExternalEvents
 
Bitumen extraction saskatchewan
Bitumen extraction saskatchewanBitumen extraction saskatchewan
Bitumen extraction saskatchewanKamran Abrar
 
Autoclave Steam Sterilizer - Pharmaceutical
Autoclave Steam Sterilizer - PharmaceuticalAutoclave Steam Sterilizer - Pharmaceutical
Autoclave Steam Sterilizer - Pharmaceuticalclientscomp
 
calibration
calibrationcalibration
calibrationLeila Sh
 
Reservoir Rock Properties Laboratory Manual Exp (#3)
 Reservoir Rock Properties Laboratory Manual Exp (#3) Reservoir Rock Properties Laboratory Manual Exp (#3)
Reservoir Rock Properties Laboratory Manual Exp (#3)Muhammed Fuad Al-Barznji
 
253651195 astm-c939
253651195 astm-c939253651195 astm-c939
253651195 astm-c939Socolap Pham
 

What's hot (19)

Water test methods
Water test methodsWater test methods
Water test methods
 
INCAT (NTE)
INCAT (NTE)INCAT (NTE)
INCAT (NTE)
 
Core cleaning
Core cleaningCore cleaning
Core cleaning
 
Liquid permeability
Liquid permeabilityLiquid permeability
Liquid permeability
 
Automated Nitrogen Drying
Automated Nitrogen DryingAutomated Nitrogen Drying
Automated Nitrogen Drying
 
Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)
 
Steam and water analysis system 1
Steam and water analysis system 1Steam and water analysis system 1
Steam and water analysis system 1
 
Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)
 
Lab Report
Lab ReportLab Report
Lab Report
 
DEGASSED CATION CONDUCTIVITY
DEGASSED CATION CONDUCTIVITYDEGASSED CATION CONDUCTIVITY
DEGASSED CATION CONDUCTIVITY
 
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNET
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNETItem 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNET
Item 7 :Comparación de los *SOPs entre LATSOLAN y **SEALNET
 
RAE Systems Gas Detection And Tube Sampling Handbook
RAE Systems Gas Detection And Tube Sampling HandbookRAE Systems Gas Detection And Tube Sampling Handbook
RAE Systems Gas Detection And Tube Sampling Handbook
 
Capillary suction-timers
Capillary suction-timersCapillary suction-timers
Capillary suction-timers
 
Bitumen extraction saskatchewan
Bitumen extraction saskatchewanBitumen extraction saskatchewan
Bitumen extraction saskatchewan
 
Autoclave Steam Sterilizer - Pharmaceutical
Autoclave Steam Sterilizer - PharmaceuticalAutoclave Steam Sterilizer - Pharmaceutical
Autoclave Steam Sterilizer - Pharmaceutical
 
calibration
calibrationcalibration
calibration
 
Reservoir Rock Properties Laboratory Manual Exp (#3)
 Reservoir Rock Properties Laboratory Manual Exp (#3) Reservoir Rock Properties Laboratory Manual Exp (#3)
Reservoir Rock Properties Laboratory Manual Exp (#3)
 
253651195 astm-c939
253651195 astm-c939253651195 astm-c939
253651195 astm-c939
 
IWC 16-49
IWC 16-49IWC 16-49
IWC 16-49
 

Viewers also liked (8)

Welcome to-ewrtaxiservice
Welcome to-ewrtaxiserviceWelcome to-ewrtaxiservice
Welcome to-ewrtaxiservice
 
Versalia Avante And Woodwinds 3 and 4 bhk Luxury Living Floors Sector-67, Gu...
Versalia Avante And Woodwinds 3 and 4 bhk  Luxury Living Floors Sector-67, Gu...Versalia Avante And Woodwinds 3 and 4 bhk  Luxury Living Floors Sector-67, Gu...
Versalia Avante And Woodwinds 3 and 4 bhk Luxury Living Floors Sector-67, Gu...
 
New year lakhtionova_na
New year lakhtionova_naNew year lakhtionova_na
New year lakhtionova_na
 
MTC IATSE
MTC IATSEMTC IATSE
MTC IATSE
 
new_year_BroevMG
new_year_BroevMGnew_year_BroevMG
new_year_BroevMG
 
New year lakhtionova_na
New year lakhtionova_naNew year lakhtionova_na
New year lakhtionova_na
 
New year lakhtionova_na
New year lakhtionova_naNew year lakhtionova_na
New year lakhtionova_na
 
New year broyev_mg
New year broyev_mgNew year broyev_mg
New year broyev_mg
 

Similar to Epa 7.1

Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
D 548 – 97 ;rdu0 oc05nw _
D 548 – 97  ;rdu0 oc05nw _D 548 – 97  ;rdu0 oc05nw _
D 548 – 97 ;rdu0 oc05nw _hamfrost
 
Flash EA 1112 , Organic Elemental Analyzer
Flash EA 1112 , Organic Elemental AnalyzerFlash EA 1112 , Organic Elemental Analyzer
Flash EA 1112 , Organic Elemental AnalyzerGamal Abdel Hamid
 
Biol390-Lab3IntroductiontotheBioreactor Mary.docx
Biol390-Lab3IntroductiontotheBioreactor Mary.docxBiol390-Lab3IntroductiontotheBioreactor Mary.docx
Biol390-Lab3IntroductiontotheBioreactor Mary.docxmoirarandell
 
Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide
 Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide
Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation GuideClean Water Systems
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...hydrologywebsite1
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...hydrologyproject001
 
Determination of Inert Gas in Anhydrous Ammonia
Determination of Inert Gas in  Anhydrous AmmoniaDetermination of Inert Gas in  Anhydrous Ammonia
Determination of Inert Gas in Anhydrous AmmoniaGerard B. Hawkins
 
Pentair #320 chlorinator manual
Pentair #320 chlorinator manualPentair #320 chlorinator manual
Pentair #320 chlorinator manualDavid Hackney
 
introduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesintroduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesLgIrvin
 
Anastasia machine reparing
Anastasia machine reparingAnastasia machine reparing
Anastasia machine reparingRABAH ABUABKER
 
API682 Mechanical Seal Plan and Requirements of Mechanical Seals
API682 Mechanical Seal Plan and Requirements of Mechanical SealsAPI682 Mechanical Seal Plan and Requirements of Mechanical Seals
API682 Mechanical Seal Plan and Requirements of Mechanical SealsElango Palaniappan
 
API 682 Seal piping plans.pdf
API 682 Seal piping plans.pdfAPI 682 Seal piping plans.pdf
API 682 Seal piping plans.pdfTeddy Setiady
 
Human orosomucoid-2-elisa-kit
Human orosomucoid-2-elisa-kitHuman orosomucoid-2-elisa-kit
Human orosomucoid-2-elisa-kitDavid tong
 

Similar to Epa 7.1 (20)

Method 23 USEPA PCDD-PCDF.pdf
Method 23 USEPA PCDD-PCDF.pdfMethod 23 USEPA PCDD-PCDF.pdf
Method 23 USEPA PCDD-PCDF.pdf
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
D 548 – 97 ;rdu0 oc05nw _
D 548 – 97  ;rdu0 oc05nw _D 548 – 97  ;rdu0 oc05nw _
D 548 – 97 ;rdu0 oc05nw _
 
Flash EA 1112 , Organic Elemental Analyzer
Flash EA 1112 , Organic Elemental AnalyzerFlash EA 1112 , Organic Elemental Analyzer
Flash EA 1112 , Organic Elemental Analyzer
 
Detail product SGS test 2015 - VIB factory.PDF
Detail product SGS test  2015 - VIB factory.PDFDetail product SGS test  2015 - VIB factory.PDF
Detail product SGS test 2015 - VIB factory.PDF
 
Biol390-Lab3IntroductiontotheBioreactor Mary.docx
Biol390-Lab3IntroductiontotheBioreactor Mary.docxBiol390-Lab3IntroductiontotheBioreactor Mary.docx
Biol390-Lab3IntroductiontotheBioreactor Mary.docx
 
Miscellaneous test
Miscellaneous testMiscellaneous test
Miscellaneous test
 
Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide
 Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide
Dry Pellet Chlorinator In-Line Chlorinator Model 400 Installation Guide
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 
Determination of Inert Gas in Anhydrous Ammonia
Determination of Inert Gas in  Anhydrous AmmoniaDetermination of Inert Gas in  Anhydrous Ammonia
Determination of Inert Gas in Anhydrous Ammonia
 
Pentair #320 chlorinator manual
Pentair #320 chlorinator manualPentair #320 chlorinator manual
Pentair #320 chlorinator manual
 
Bs 1881 112 1983
Bs 1881 112 1983Bs 1881 112 1983
Bs 1881 112 1983
 
introduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesintroduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gases
 
Anastasia machine reparing
Anastasia machine reparingAnastasia machine reparing
Anastasia machine reparing
 
API682 Mechanical Seal Plan and Requirements of Mechanical Seals
API682 Mechanical Seal Plan and Requirements of Mechanical SealsAPI682 Mechanical Seal Plan and Requirements of Mechanical Seals
API682 Mechanical Seal Plan and Requirements of Mechanical Seals
 
API 682 Seal piping plans.pdf
API 682 Seal piping plans.pdfAPI 682 Seal piping plans.pdf
API 682 Seal piping plans.pdf
 
Assignment 1 SOP (1)
Assignment 1 SOP (1)Assignment 1 SOP (1)
Assignment 1 SOP (1)
 
Human orosomucoid-2-elisa-kit
Human orosomucoid-2-elisa-kitHuman orosomucoid-2-elisa-kit
Human orosomucoid-2-elisa-kit
 

Recently uploaded

High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxwendy cai
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024hassan khalil
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Christo Ananth
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 

Recently uploaded (20)

High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptx
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCRCall Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 

Epa 7.1

  • 1. SOUTHCOASTAIR QUALITYMANAGEMENTDISI'RICI' DETERMINATIONOFNITROGEN OXIDEEMISSIONS FROM STATIONARY SOURCES TECHNICALSUPPORTSERVICES APPLIED SCIENCEAND TECHNOLOGY MARCH 1989
  • 2.
  • 3. D E T E ~ T I O ~OF HITROGEN OXIDE EKISSIOblS FROM STATZObfaRY SOURCES TABLE OF CONTENTS Section 1, Overview 1.1 Principle 1.2 Applicability 2 , F i e l d P r o c e d u r e s 2.1 Sampling Apparatus 2.2 Sampling Reagents 2.3 Preparation of Sampling Equipment 2.4 Sampling Procedure 3, L a b o r a t o r y Procedures 3.1 Apparatus 3.2 Reagents 3.3 Pretest Preparation 3.4 Preparation of Sampling Collection Flask 3.5 Sample Collection Flask Leak Check 3.6 Sample Recovery 3.7 Analysis
  • 4.
  • 5. 3.8 Calculation and Reporting 3.9 Calibration 4. Engineering Calculations and Reporting 5. Alternative Laboratory Procedures 5.1 Apparatus 5.2 Reagents 5.3 Preparation of Sample Collection Flask 5.4 Sample Collection Flask Leak Check 5.5 Recovery 5.6 Sample Analysis 5.7 Calibrations 5.8 Calculations 5.9 Quality Control
  • 6. Section 1 of 5 1. overview' 1-1 Principle A grab sample is collected in an evacuated flask containing a dilute sulfuric acid-hydrogen peroxide absorbing solution, and tpe nitrogen oxides except nitrous oxide are measured colorimetrically using the phenoldisulfonic acid (PDS) procedure or ion chromatograph procedures (see Section 5 ) . 1.2 Applicability This method measures nitrogen oxides (NOx) emitted from stationary sources, Its range is 2 to 400 mg NOx (as NOZ) per dry standard cubic meter, without diluting the sample.
  • 7. DETERMIHATIOX OF OXIDE EMISSIOHS FROM STATIOHBRP SOURCES Section 2 of 5 2. Field Procedures 2.1 Sampling Apparatus Sample collection requires the equipment l i s t e d below and shown i n Figure 7.1-1. T a- Probe ~ o r o s i l i c a t eglass tubing, sufficiently heated t o prevent water condensation and equipped with an in-stack or out-stack f i l t e r t o r&ove particulate matter. (A plug of glass wool'is satisfactory for t h i s purpose.) Stainless steel or Teflon tubing also may be used for the probe. Heating is not necessary i f the probe remains dry during the purging -period, or less than 2 percent of the NOx present is NO2 .
  • 8. b. Collection Flask ' Two-liter borosilicate, round bottom flask, with short neck and 24/40 standard taper opening, protected against implosion or breakage with a T-bore stopcock connected to a 24/40 standard taper joint. Alternatively, this flask may have the short neck replaced with 48 mm (1 1/2 in.) long 12 mm (1/2 in.) glass tubing sealed with rubber tubing and screw clamp. Flasks must be calibrated to the nearest 10 m l . No difference in NO, results has been observed between the use of rubbertubing and the use of ground joints. Ground glass joints tend to have more leakage problems than do rubber tubing and screw clamps. When tubing is used, the surface area exposed to the sample should be minimized. See Section 3.9 for calibration procedure. c. Vacuum Line Tubing capable of withstanding a vacuum of 75 mm (3 in.) Hg absolute pressure, with "Tn
  • 9. connection and T-bore stopcock o r rubber tubing and s c r e w clamp. d. Vacuum Gauge The collection flask must be equipped with a vacuum gauge capable of measuring t o 760 mm (30 in,) Hg with 2.5 mm (0.1 in,) Hg division, e. One-way Squeeze Bulb f, Stopcock and Ground Joint Grease A high vacuum, high temperature chlorofluorcarbon grease is required when ground glass joints are used, Halocarbon 25- SS is effective,
  • 10. 2 - 2 Sampling Reagents a. Sulfuric Acid-Hydrogen Peroxide Solution See Section 3.2.1 f o r preparation procedure. The solution should be prepared daily and not exposed t o extreme heat o r direct sunlight- 2.3 Preparation of Sampling Equipment This may be done in the laboratory prior t o the test. U s e two-liter round bottam flasks, Add 25 m l of absorbing solution t o each flask- Evacuate the flask t o the vapor pressure of the solution, tightly close the s c r e w c l a p and insert a solid glass plug (optional) into the open end of the tubing u n t i l ready f o r sampling, 2.4 Sampling Procedure Assemble the sampling t r a i n a s shown i n Figure 7-1-1 and place the probe a t t h e sampling point- (Select the sampling point a s i n Method 3-1, Section 2.2.) Make sure a l l f i t t i n g s are tight and leak-free, and i f necessary a l l glass joints have been properly greased with a high-
  • 11. temperature, chlorfluorocarbon-based stopcock 444, grease. Turn the flask valve and the pump to the "evacuatem1positions. Evacuate the flask to a pressure approaching the vapor pressure of water at the existing temperature. Turn the pump valve to its "ventn position and turn off the pump. Check for leakage by observing the manometer for any pressure fluctuation. A variation more than 10 mm (0.4 in.) Hg over a period of one minute is - unacceptable. Do not use the flask until the leakage problem is'wrrected. If flaskk are evacuated prior to setup, the evacuation steps are unnecessary. Turn the flask valve counterclockwise to its "purgen position. Purge the probe and the vacui tube using the squeeze bulb. If condensation occurs in the probe and the flask valve area, heat the probe and purge until the condensation disappears. Turn the pump valve to its nventn position. Turn the flask valve clockwise to its "evacuaten position and record the difference in the mercury levels in the manometer. The absolute internal pressure in the flask (Pi) is equal to -=%
  • 12. the barometric pressure, minus the manometer reading. Immediately turn the flask valve to the wsamplen position and permit the gas to enter the flask until pressure in the flask and sample line (i.e. duct stack) are equal. (Vacuum gauge should read near zero.) This will usually require about 15 seconds; a longer period indicates a mplugn in the probe, which must be corrected before sampl&g is continued. After collecting the sample, turn valve to its npurgewposition and disconnect the flask sampling train. Shake the flask for at least 5 minutes. Alternatively, open or.closethe screw clamps in the appropriate manner to achieve the same results, i.e. (1) purge, (2) vacuum check, (3) sample, (4) seal flask, (5) shake, When an alternate setup is used, perform all the above operations by opening and closing the appropriate screw clamps, Record data on a form similar to Figure 7.1-2.
  • 13. DETERKINATIOH OF 10ITBOOE# OXIDE EKI88IOEI8 FROM STATIOXARY SOURCES Section 3 of 4 3. Laboratory Procedures 3..1 Apparatus . . 3.1 .1 Sampling Apparatus a. Probe, Collection Flask, vacuum' Line, Vacuum Gauge, Squeeze Bulb See Section 2.1. b. Pipet Class A, 25 ml 3.1.2 Sample Recovery a. Graduated Cylinder, 50 ml with 1 m l Divisions
  • 14. b. Polyethylene Wash Bottle c. Glass Stirring Rod d. Test Paper for Indicating pH, Range 7 to 14 e. Mercury Manometer, or Calibrated Electronic Manometer, with Fitting for Sample Bulb Attachment f. Thermometer Ambient range, OC, traceable to NBS thermometer. 3.1.3 Analysis For the analysis, including calibration, the following equipment is needed: a. volumetric Pipets Class A, or a microburet to measure-2, 4, 6, 8, 10, 12, 16 and 20 ml of working calibration solution; one loo ml, class A pipet.
  • 15. b. Porcelain Evaporating Dishes 250 m l capacity with lip for pouring, one for each sample and each standard. The Coors NO-45006 (shallow-form, 196 ml) is satisfactory. c. 'Steam Bath Low-temperature oven or thermostatically controlled hot plates -kept below 70°c ( 1 5 8 ~ ~ )are acceptable alternatives. d. Dropping Pipet or Dropper Three required. e. Glass Rod One for each sample, blank, and standard. f. Graduated Cylinder 100 ml with 1 m l divisions.
  • 16. g. volumetric Flasks Class A, 1000 ml (four), h, Spectrophotometer To measure absorbance at 405 nm, using a 10 mm cell, i, Graduated Pipet, 10 ml, 0-1 ml divisions j, Test Paper for Indicating pH, Range 7 to 14 pH k. Analytical Balance To measure to within 0.1 mg. , . 1, Centrifuge To remove solid contained in sample, m. Graduated Centrifuge Tube, 50 ml, 1 ml Divisions
  • 17. 3-2 Reagents Unless otherwise indicated, it is intended that all reagent conform to the specification established by the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available; otherwise, use the best available grade. 3.2.1 Sampling To prepare the absorbing solution, cautiously add 2.8 ml concentrated H2S04 -Y to 1 liter of water. Mix well and add 2 ml of 30 percent H202. The absorbing solution should be prepared daily. Do not expose to extreme heat or direct sunlight. 3.2.2 Sample Recovery a. Sodium Hydroxide (NaOH), 1N Dissolve 40 g NaOH in water and dilute to 1 liter.
  • 18. b. Water Deionized, distilled to conform to ASTM specification D1193-77, Type 3. At the option of the chemist, the KMn04 test for oxidizable organic matter may be omitted when high concentrations of organic matters are not expected to be present. Reference to water throughout this method implies deionized, distilled water, 3.2-3 Analysis a. Fuming Sulfuric Acid (H2S04) 15 to 18 percent, by weight, free SO3. HANDLE WITH CAUTION. Used to make phenoldisulfonic acid (PDSA) reagent: See Section 3.2.3.h. b. Phenol White solid. Used to make phenoldisulfonic acid (PDSA) reagent: See Section 3.2.3.h.
  • 19. c. S u l f u r i c Acid ' (H2S04) Concentrated, 95 percent minimum assay. Handle with care. d. Potassium N i t r a t e (KN03) D r i e d a t 105 t o 1 1 0 ~ ~(220 t o 2 3 0 ~ ~ ) f o r a minimum of 2 hours j u s t p r i o r t o preparation of standard solution. e. Standard KN03 Solution Dissolve exactly 0.5495 g of d r i e d KN03 i n w a t e r and d i l u t e t o 1 liter w i t h water i n a 1000 m l volumetric flask.' One m i l l i l i t e r of t h i s standard solution contains the .equivalent o f 0.250 m g of NO2. .. f . Working Standard KN03 S o l u t i o n ' Dilute 100 m l of the standard solution t o 1000 m l with water. One m i l l i l i t e r of the working standard solution is -'% equivalent t o 0.025 mg of NO2. 7.1-14
  • 20. g. Water See Section 2.2.2. h. Phenoldisulf onic Acid Solution Dissolve 25 g of pure white phenol i n 150 m l concentrated H2S04, and heat a t 1 0 0 ~ ~( 2 1 2 ~ ~ )f o r 2 hours. Store i n a dark stoppered bottle, Protect from ambient moisture a t a l l t i m e s . Phenoldisulfonic acid purchased from a commercial source should be rejected i f blank absorbance is greater than 0,030, and i f linearity is not obtained during calibration (see Section 3,8) .. 3.3 Pretest Preparaticn A l l equipment; including t h e balance, steam bath, centrifuge, spectrophotometer, cleaned glassware, reagents, and safety equipment should be checked f o r readiness before conducting .the following procedure,
  • 21. 3.4 Preparation of Sample Collection Flask Use two-liter round bottom flasks for sampling. Add 25 m l of absorbing solution to each flask. Reserve a portion of this reagent for a blank and for control preparation. Evacuate the flask to the vapor pressure of the solution, tightly close the screw clamp, and insert a solid glass plug (optional) into the open end of the tubing. Pack the flasks in a sturdy box for transport to the field, Prepare fresh flasks each day. NOTE: No difference in NOx results have been observed with use of rubber tubing as compared to T glass joints, In addition, rubber tubing has been virtually leak-free as compared to glass joints. 3.5 Sample Collection Flask Leak Check Check for sample flask leakage during the flask evacuation step (see Section 3.4). The contents of the flask will boil when evacuated if no leak is present. More leak checks may be performed in the field with a manometer or vacuum gauge, though these checks may be avoided by sampling in triplicate or quadruplicate. The analytical
  • 22. results will indicate which flasks may have leaked and should be rejected. Check the flask screw clamps (if used) as soon as flasks are returned to the laboratory. Record the presence of slightly loose clamps for consideration in comparing flask analytical results. Very loose clamps are clear cause for rejection. 3.6 Sample Recovery Shake the sealed collection flasks for 15 minutes on a mechanical shaker, and rotate frequently to provide a thorough scrubbing action. Allow the flasks to stand overnight (or at least 16 hours) to ensure complete oxidation and absorption of the oxides of nitrogen by the solution. When absorption is complete, the oxides of nitrogen (except nitrous oxide) will be converted to HN03. Record the temperature and absolute gas pressure in each flask. Rince the contents into a 200 ml evaporating dish, using three 15 ml portions of water. Prepare a blank using the same amount of absorbing solution and wash water as used for each sampling flask.
  • 23. 3.7 Analysis Add 1 N sodium hydroxide to.the solution in each evaporating dish, including the blank, until it is barely alkaline (<pH 9) to litmus paper. Evaporate the solution in the evaporating dish to . dryness on a steam bath and allow to cool. Add 2 m l phenoldisulfonic acid solution to the dried residue and titrate thoroughly with a glass rod. Make sure the solution contacts all the residue. Add 1 ml water and four drops of concentrated H2S04. Heat the solution on a steam bath for 3 minutes with occasional stirring. Allow the solution to cool, add 20 ml'water, mix well by stirring, and add 10 ml of concentrated NHqOH, dropwise with constant stirring. Transfer the solution to a 50 ml graduated centrifuge tube, washing the evaporating dish three times with 5 m l portions of water. Dilute to the mark with water, stir thoroughly, and centrifuge. Transfer the clear solution to a spectrophotometer cell. Measure absorbance of each solution against the reagent blank at a wavelength of 405 nm. .If a greater dilution of the sample is required, dilute the blank to the same volume. Thinner cells or dilution blocks may be used. Process
  • 24. and dilute each sample determination in a manner similar to that used in preparing the corresponding concentration range of the standard curves. In each case, subtract the bldnlr reading from the sample reading. The dilution used for obtaining the calibration curves in the ranges given may have to be varied, depending on the make of spectrophotometer used. Determine the weight of NOZ, in milligrams, from the light absorption reading on the calibration curve corresponding to dilution used. See Section 3.9 for calibrations. 3.8 Calculations and Reporting Net ODxCalipration Factom515x760xT NO2 = Net Pressure x Net Volume x 288.6 where: N02 = NOx as NO2, ppm Net OD = Sample light absorption - blank light absorption Calibration = pg N02/OD, from calibration Factor c u m e
  • 25. T = Temperature in the flask, OK Net = Measured pressure in the flask, Pressure minus the water vapor pressure at temperature T, mm Hg Net Volume = Volume of flask - 25 m l (volume of absorbing solution in the flask), liters Report NOx as NO2 (ppm), using three significant figures. Discard decimal digits. Round the third digit to the nearest 5. 3.9 Calibration 3.9.1 PDSA For a 0-10 ppm calibration curve, use pipets or a microburet to transfer 0.0 (blank), 2.0, 4.0, 6.0, 8.0, 10.0, 12.0, 16.0, and 20.0 ml of potassium nitrate working standard solution into separate 200 ml evaporating dishes, and add 25 ml of absorbing solution to each evaporating dish. Follow as directed in Section 3.7.
  • 26. Read the light absorption of each solution and subtract the blank reading. If absorption is too high, dilute appropriately with distilled water. Construct the curve for the concentration range by plotting the net light absorption of the solution as ordinate against milligrams of NO2 as abscissa, or use a calculator programmed for linear regression analysis: where: m = slope b = intercept
  • 27. 3.9.2 Flask Calibration The temperature of the water is measured. Volume is calculated according to the formula. where: v = Volume of the bulb, m l W1 = Weight of flask and water, g W2 = Weight of empty fiask, g Dt = Density of water, g/ml at temperature, t
  • 28. 4 DETERHINATION OF ~ T R O G ~OXIDE ZXISSIOHS PBOH STAT1Ob;QBY 80uBcE8 Section 4 of 5 Engineering Calculations and Reporting When mass.emission rate of NO, is required, follow this calculation. lb of NOx per hour = A-B.C-D -- (ppm NO^) (46)(1.58 x (dscfm) A -- NOX concentration in ppm B -- Molecular weight of NO2, 46 lb/lb-mole C -- Conversion factor, 1.58 x loo7 -- (loo6 ppm)(60 min./hr) (379 cu ft/lb- mole) D -- Dry standard stack gas flow rate per minute measured, dscfm (follow Method 2.1)
  • 29. DETERKINZATIO~QOF SITRoGEH OXIDE EXISSIOWS PBOM STATIONARY~S O W S Section 5 of 5 5 - Alternative Laboratory Procedures . 5.1 Apparatus - 5.1.1 Preparation of Sample Collection Flask See Section 3.4- Same as Section 2-1 b. 5.1-2 Sample Recovery a. Class A 50 ml Volumetric Flask One for each sample and blank- b, Polyethylene Wash Bottle
  • 30. c, Manometer Capable of reading to Imm within the 0 to 900mm Hg range. d, NBS Traceable Thermometer e. Sample Shaker 5,1.3 Analysis . a, Volumetric Pipets Class A, one 10 m l for each sample and blank, 100 m l for working standard preparation, 10 ml, 15 ml, 20 ml, 25 ml, 50 m l , and graduated 5 m l Class A pipets, or microburet for calibration standard dilutions, b, volumetric Flasks Class A, one 50 ml flask for each sample and blank, two 50 ml for each
  • 31. calibration standard, and tvo 100 m l for stock and vorking standard preparation. c. Ion Chromatograph Capable of detecting and resolving 0.1 ppm nitrate in the presence of dilute (<O.OlN) sulfuric acid. d. Recorder or Integrator e. Analytical Balance Capable of veighing to 0.1 mg. 5.2 Reagents Unless otherwise indicated, it is intended that all reagents conform to the specification established by the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available, otherwise use the best available grade.
  • 32. 5.2.1 Sampling Absorbing solution. Same as Section Sample Recovery See Section 3.2.2. Analysis a. Potassium N i t r a t e (KN03) Dried a t 105 t o 1 1 0 ~ ~f o r a minimum of 2 hours, -and desiccated t o room temperature just before use. b. Standard KN03 Solution 0.5495 g dried RN03 dissolved i n 1 l i t e r of w a t e r . c- Working Standard KN03 Solution D i l u t e 100 m l of KN03 standard solution t o 1000 m l with water.
  • 33. d. Water See section 3.2.2. 5.3 Preparation of Sample Collection Flask See Section 3.4. 5.4 Sample Collection Flask Leak Check See Section 3.5. 5.5 Sample Recovery Shake samples and allow to stand overnight as in Section 3.6. Measure and record the temperature and absolute gas pressure in each flask. Recover . the contents of each flask into a 50 m l volumetric flask with at least three rinsings, then dilute to volume and mix well. 5. 6 Sample Analysis Dilute a 10 m l aliquot of each sample to 50 ml. Set the instrument response as close as possible to maximum on the highest standard, then run the
  • 34. calibration curve, the reagent blank, and each sample once. Dilute samples that are too high into range. Samples that are too low require an additional standard dilution and more sensitive instrument settings. Run a control for each set of samples or each ten samples, whichever is fewer, Rerun the calibration curve at the end of the analysis. 5.7 Calibration A five-point calibration curve in O.OlN H2S04 is made from the working standard as follows: 2 m l KN03 working standard solution diluted to 50 ml. Take 2 ml of the working solution and 5 ml of 0-lN H2S04, and dilute to 50 ml with water. 5 ml KN03 working standard solution diluted to 50 ml. Take 5 ml of the working solution and 5 ml of 0,lN H2S04, and dilute to 50 ml with water. 10 ml KN03 working standard solution diluted to 50 ml- Take 10 m l of the working solution and 5 ml of O.lN H2S04, and dilute to 50 ml with water.
  • 35. '15 ml,RN03 working standard solution diluted to q. 50 ml. Take 15 ml of the working solution and 5 ml of O.lN H2S04, and dilute to 50 ml with water. 25 ml KN03 working standard solution diluted to 50 ml. Take 25 m l of the working solution and 5 ml of 0.1N H2S04, and dilute to 50 ml with water. This corresponds to a sample range of 50 pg NO2 to 625 pg NO2 per sample or approximately 12 to 150 ppm. The calibration curve must bracket all sample 1*, concentrations. Dilute a sample that is too high to fall within the range of the calibration curve. For samples that are too low, make another standard solution using the working standard solution (25 pg), delete the highest standard (625 pg), and maximize the instrument response on the next highest standard (375 pg) . Rerun the analysis. Calculate linear regression of total pg NO2 and peak height using all standard points. Remake and rerun any standard point that deviates by 7 percent or more from expected. A change of T instrument response over the course of the run by
  • 36. 10 percent or more invalidates the analysis. A difference of the measured control concentration from the theoretical concentration by 10 percent or more invalidates the analysis. 5.8 Calculations . Calculate the NO2 in pg for each sample using the linear regression, expressed as: where: m = Slope b = Intercept of the regression h = Sample peak height Then calculate NO2 as ppm for each sample, using the following equation: pg NO2 X 0.515 X 760 X T NO2 = Net Pressure x Net Volume x 288.6
  • 37. where: Pg NO2 = Micrograms of NO2 of the sample being calculated T = , Temperature of the flask, OK Net Pressure = Measured pressure of the flask, minus the water pressure, at temperature TI mm Rg , Net Volume = Volume of the flask, liters Report NOx as NO2 (ppm) using three significant figures. Round the third digit to the nearest 5. 5 - 9 ~ualityControl Run suitable blank and control samples with each set of samples or each ten samples, whichever is fewer. Blank consists of reagent reserved from sample preparation.
  • 38. Suitable control consists of an EPA.QA vial, prepared as directed, and analyzed with the samples, Measured value must be within 10 percent of the theoretical value. A control that falls outside of the range requires instrument recalibration and reanalysis of the sample(s). Change in instrument response of more than 10 percent from the beginning to the end of the run also requires instrument recalibration and reanalysis of the sample(s),