This document discusses potentiometric analysis for water quality assessment. It begins with an introduction and review of basic redox reaction principles. It then summarizes the key concepts of potentiometric analysis, including reference electrodes that maintain a constant potential, indicator electrodes whose potential depends on analyte concentration, and measuring cell potential with instruments like pH meters and ion meters. The document provides examples of different types of reference electrodes like calomel and indicator electrodes including metal, redox, and membrane electrodes. It concludes with a brief overview of probes that can be used for potentiometric analysis.
There is no redox chemistry at the membrane; potential is
determined by the relative concentrations of the analyte on each side of
the membrane. By convention, the indicator electrode is considered to
be the cathode in a potentiometric device
For my senior CEU pharmacy students in QC 2 with Instrumentation.
The different types and examples of indicator electrodes used in potentiometric titration method of drug analysis.
There is no redox chemistry at the membrane; potential is
determined by the relative concentrations of the analyte on each side of
the membrane. By convention, the indicator electrode is considered to
be the cathode in a potentiometric device
For my senior CEU pharmacy students in QC 2 with Instrumentation.
The different types and examples of indicator electrodes used in potentiometric titration method of drug analysis.
GraphRAG is All You need? LLM & Knowledge GraphGuy Korland
Guy Korland, CEO and Co-founder of FalkorDB, will review two articles on the integration of language models with knowledge graphs.
1. Unifying Large Language Models and Knowledge Graphs: A Roadmap.
https://arxiv.org/abs/2306.08302
2. Microsoft Research's GraphRAG paper and a review paper on various uses of knowledge graphs:
https://www.microsoft.com/en-us/research/blog/graphrag-unlocking-llm-discovery-on-narrative-private-data/
Kubernetes & AI - Beauty and the Beast !?! @KCD Istanbul 2024Tobias Schneck
As AI technology is pushing into IT I was wondering myself, as an “infrastructure container kubernetes guy”, how get this fancy AI technology get managed from an infrastructure operational view? Is it possible to apply our lovely cloud native principals as well? What benefit’s both technologies could bring to each other?
Let me take this questions and provide you a short journey through existing deployment models and use cases for AI software. On practical examples, we discuss what cloud/on-premise strategy we may need for applying it to our own infrastructure to get it to work from an enterprise perspective. I want to give an overview about infrastructure requirements and technologies, what could be beneficial or limiting your AI use cases in an enterprise environment. An interactive Demo will give you some insides, what approaches I got already working for real.
"Impact of front-end architecture on development cost", Viktor TurskyiFwdays
I have heard many times that architecture is not important for the front-end. Also, many times I have seen how developers implement features on the front-end just following the standard rules for a framework and think that this is enough to successfully launch the project, and then the project fails. How to prevent this and what approach to choose? I have launched dozens of complex projects and during the talk we will analyze which approaches have worked for me and which have not.
LF Energy Webinar: Electrical Grid Modelling and Simulation Through PowSyBl -...DanBrown980551
Do you want to learn how to model and simulate an electrical network from scratch in under an hour?
Then welcome to this PowSyBl workshop, hosted by Rte, the French Transmission System Operator (TSO)!
During the webinar, you will discover the PowSyBl ecosystem as well as handle and study an electrical network through an interactive Python notebook.
PowSyBl is an open source project hosted by LF Energy, which offers a comprehensive set of features for electrical grid modelling and simulation. Among other advanced features, PowSyBl provides:
- A fully editable and extendable library for grid component modelling;
- Visualization tools to display your network;
- Grid simulation tools, such as power flows, security analyses (with or without remedial actions) and sensitivity analyses;
The framework is mostly written in Java, with a Python binding so that Python developers can access PowSyBl functionalities as well.
What you will learn during the webinar:
- For beginners: discover PowSyBl's functionalities through a quick general presentation and the notebook, without needing any expert coding skills;
- For advanced developers: master the skills to efficiently apply PowSyBl functionalities to your real-world scenarios.
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdf91mobiles
91mobiles recently conducted a Smart TV Buyer Insights Survey in which we asked over 3,000 respondents about the TV they own, aspects they look at on a new TV, and their TV buying preferences.
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...UiPathCommunity
💥 Speed, accuracy, and scaling – discover the superpowers of GenAI in action with UiPath Document Understanding and Communications Mining™:
See how to accelerate model training and optimize model performance with active learning
Learn about the latest enhancements to out-of-the-box document processing – with little to no training required
Get an exclusive demo of the new family of UiPath LLMs – GenAI models specialized for processing different types of documents and messages
This is a hands-on session specifically designed for automation developers and AI enthusiasts seeking to enhance their knowledge in leveraging the latest intelligent document processing capabilities offered by UiPath.
Speakers:
👨🏫 Andras Palfi, Senior Product Manager, UiPath
👩🏫 Lenka Dulovicova, Product Program Manager, UiPath
Let's dive deeper into the world of ODC! Ricardo Alves (OutSystems) will join us to tell all about the new Data Fabric. After that, Sezen de Bruijn (OutSystems) will get into the details on how to best design a sturdy architecture within ODC.
Connector Corner: Automate dynamic content and events by pushing a buttonDianaGray10
Here is something new! In our next Connector Corner webinar, we will demonstrate how you can use a single workflow to:
Create a campaign using Mailchimp with merge tags/fields
Send an interactive Slack channel message (using buttons)
Have the message received by managers and peers along with a test email for review
But there’s more:
In a second workflow supporting the same use case, you’ll see:
Your campaign sent to target colleagues for approval
If the “Approve” button is clicked, a Jira/Zendesk ticket is created for the marketing design team
But—if the “Reject” button is pushed, colleagues will be alerted via Slack message
Join us to learn more about this new, human-in-the-loop capability, brought to you by Integration Service connectors.
And...
Speakers:
Akshay Agnihotri, Product Manager
Charlie Greenberg, Host
Search and Society: Reimagining Information Access for Radical FuturesBhaskar Mitra
The field of Information retrieval (IR) is currently undergoing a transformative shift, at least partly due to the emerging applications of generative AI to information access. In this talk, we will deliberate on the sociotechnical implications of generative AI for information access. We will argue that there is both a critical necessity and an exciting opportunity for the IR community to re-center our research agendas on societal needs while dismantling the artificial separation between the work on fairness, accountability, transparency, and ethics in IR and the rest of IR research. Instead of adopting a reactionary strategy of trying to mitigate potential social harms from emerging technologies, the community should aim to proactively set the research agenda for the kinds of systems we should build inspired by diverse explicitly stated sociotechnical imaginaries. The sociotechnical imaginaries that underpin the design and development of information access technologies needs to be explicitly articulated, and we need to develop theories of change in context of these diverse perspectives. Our guiding future imaginaries must be informed by other academic fields, such as democratic theory and critical theory, and should be co-developed with social science scholars, legal scholars, civil rights and social justice activists, and artists, among others.
1. World Bank & Government of The Netherlands funded
Training module # WQ - 32
Potentiometric Analysis
New Delhi, January 2000
CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas,
New Delhi – 11 00 16 India
Tel: 68 61 681 / 84 Fax: (+ 91 11) 68 61 685
E-Mail: dhvdelft@del2.vsnl.net.in
DHV Consultants BV & DELFT HYDRAULICS
with
HALCROW, TAHAL, CES, ORG & JPS
2. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 1
Table of contents
Page
1 Module context 2
2 Module profile 3
3 Session plan 4
4 Overhead/flipchart master 5
5 Evaluation sheets 25
6 Handout 27
7 Additional handout 32
8 Main text 34
3. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 2
1. Module context
This module discusses principles of potentiometric analysis and different types of indicating
electrodes used in such analysis. Modules in which prior training is required to complete this
module successfully and other, available related modules are listed in the table below.
While designing a training course, the relationship between this module and the others,
would be maintained by keeping them close together in the syllabus and place them in a
logical sequence. The actual selection of the topics and the depth of training would, of
course, depend on the training needs of the participants, i.e. their knowledge level and skills
performance upon the start of the course.
No. Module title Code Objectives
1. Basic chemistry concepts WQ - 02 • Convert units from one to another
• Discuss the basic concepts of
quantitative chemistry
• Report analytical results with the
correct number of significant digits.
2. Understanding the hydrogen
ion concentration (pH)
WQ - 06 • Discuss about the concept of pH
• Calculate pH
3. The chemistry of dissolved
oxygen measurement
WQ - 11 • Appreciate significance of DO
measurement.
• Understand the chemistry of DO
measurement by Winkler method.
4. Understanding chemical
oxygen demand test
WQ - 18 • Appreciate significance of COD
measurement
• Understand the chemistry of COD
measurement
5. Basic aquatic chemistry
concepts
WQ - 24 • Understand equilibrium chemistry and
ionisation constants.
• Understand basis of pH and buffers
• Calculate different types of alkalinity.
6. Advanced aquatic chemistry:
solubility equilibria
WQ - 29 • Explain the principles of chemical
equilibrium
• Define solubility product and explain
how this relates to water quality
assessment
• Define the octanol-water partition
coefficient and explain how this
relates to water quality assessment.
7. Advanced aquatic chemistry
concepts: redox equilibria
WQ - 30 • Understand the principles of redox
reactions
• Understand the significance of redox
potential measurement
4. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 3
2. Module profile
Title : Potentiometric Analysis
Target group : HIS function(s): Q2, Q3, Q5, Q6
Duration : One session of 45 min
Objectives : After the training the participants will be able to understand:
• the principles of potentiometric analysis
• construction of reference and indicating electrodes.
Key concepts : • reference electrodes
• indicating electrodes
Training methods : Lecture, exercises
Training tools
required
: Board, flipchart, OHS
Handouts : As provided in this module
Further reading
and references
: • Chemistry for environmental engineers, C. N. Sawyer, P. L.
McCarty & G. F. Parkin, Mc Graw Hill, Inc., 1994
• Process chemistry for water and wastewater treatment, L. D.
Benefield, J. P. Judkins & B. L. Weand, Prentice-Hall, Inc.,
1982
5. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 4
3. Session plan
No Activities Time Tools
1 Preparations
2 Introduction:
• Review fundamentals of redox reactions, requirement of
potentiometric analysis
10 min OHS
3 Reference Electrodes
• As trainees about electrodes used in pH measurement
• Explain different reference electrodes
10 min OHS
4 Indicator electrodes
• Different types of indicator electrodes
20 min OHS
5 Wrap up 5 min
6. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 5
4. Overhead/flipchart master
OHS format guidelines
Type of text Style Setting
Headings: OHS-Title Arial 30-36, with bottom border line (not:
underline)
Text: OHS-lev1
OHS-lev2
Arial 24-26, maximum two levels
Case: Sentence case. Avoid full text in UPPERCASE.
Italics: Use occasionally and in a consistent way
Listings: OHS-lev1
OHS-lev1-Numbered
Big bullets.
Numbers for definite series of steps. Avoid
roman numbers and letters.
Colours: None, as these get lost in photocopying and
some colours do not reproduce at all.
Formulas/Equat
ions
OHS-Equation Use of a table will ease horizontal alignment
over more lines (columns)
Use equation editor for advanced formatting
only
7. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 6
Potentiometric analysis
• Review of principles of redox reactions
• Reference electrodes
• Indicating electrodes
8. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 7
Review (1)
• Electrode
• Half-cell
• Electrode potential
• Electrochemical cell
9. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 8
Review (2)
10. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 9
Review (3)
• Nernst Equation
aA + bB … + ne ↔ cC + dD
E E
RT
nF
C D
A B
c d
a b
= −
L
N
MM
O
Q
PP
0
ln
11. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 10
Potentiometric analysis
• Reference electrodes
- constant electrode potential
• Indicating electrodes
- electrode potential depends on analyte concentration
• Measurement of potential
- pH meter
- ion meter
12. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 11
Reference Electrodes (1)
13. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 12
Reference electrodes (2)
• By definition potential of hydrogen electrode is zero and acts
as anode
• Inconvenient to maintain
• Hazardours
14. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 13
Reference Electrodes (3)
Figure 3: Schematic diagram of the Calomel reference
electrode
Hg.Hg2Cl2
for filling with
KCl
15. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 14
Reference electrodes (4)
• Calomel electrode reaction
Hg2Cl2(s) + 2e ↔ 2Hg(l) + 2Cl-
• Standard potential at 250
C
KCl EH
0.1N -0.3358
1.0N -0.3834
Saturated -0.2444
16. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 15
Indicator Electrodes
• Electrode potential depends on analyte concentration
• Types of indicating electrodes
- Metal electrodes, type I and type II
- Redox electrodes
- Membrane electrodes, crystalline & liquid
17. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 16
E E
Cu
= − +
0
2
0 059
2
1.
log
Metal Electrode: type I
• For copper electrode dipped in Cu2+
solution
Cu2+
+ 2e ↔ Cu(s)
•
• E vs. log [Cu2+
] has slope = + 0.0295
18. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 17
Metal Electrode: type II
• For anion which forms sparingly soluble precipitate with
electrode metal.
AgI(s) ↔ Ag+
+ I-
Ag+
+ e ↔ Ag(s)
------------------------
AgI(s) +e ↔ Ag(s) + I-
• E = E0
AgI - 0.059 log [I-
]
• E vs. log [I-
] has slope = -0.059
19. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 18
Redox electrode
• Non-reactive metal electrode, e.g. platinum
• Electrode potential depends on relative conc. of reduced and
oxidised specie in equilibrium
Fe3+
+ e ↔ Fe2+
E
Fe
Fe
= −
+
+
0 77
0 059
1
2
3
.
.
log
20. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 19
Membrane Electrodes (1)
• Membrane sensitive to ion of interest
• Ion conc. within electrode is constant (std.)
• Potential depends on conc. in analyte sample
21. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 20
Membrane Electrode (2)
Figure 4: Glass electrode system for measuring pH
22. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 21
Membrane Electrode (3)
Eind = K+ 0.059 pH
- [Ha
+
] and [He
+
] = internal electrode and external analyte conc.
(activity), respectively.
- K = constant
E k
H
H
ind
a
e
= +
+
+
0 059. log
23. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 22
Membrane Electrode (4)
Figure 5: Liquid membrane electrode for Ca2+
24. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 23
Probes (1)
• Respond to molecular specie
• Complete cell
• Polarographic type, need external current
• Galvanic cell type
25. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 24
Probes (2)
Figure 6: Schematic diagram of oxygen probe
26. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 25
5. Evaluation sheets
27. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 26
28. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 27
6. Handout
29. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 28
Potentiometric analysis
• Review of principles of redox reactions
• Reference electrodes
• Indicating electrodes
Review (1)
• Electrode
• Half-cell
• Electrode potential
• Electrochemical cell
Review (3)
• Nernst Equation
aA + bB … + ne ↔ cC + dD
Potentiometric analysis
• Reference electrodes
- constant electrode potential
• Indicating electrodes
- electrode potential depends on analyte concentration
• Measurement of potential
- pH meter
- ion meter
Reference electrodes (2)
• By definition potential of hydrogen electrode is zero and acts as anode
• Inconvenient to maintain
• Hazardours
Reference electrodes (4)
• Calomel electrode reaction
Hg2Cl2(s) + 2e ↔ 2Hg(l) + 2Cl-
• Standard potential at 250
C
KCl EH
0.1N -0.3358
1.0N -0.3834
Saturated -0.2444
E E
RT
nF
C D
A B
c d
a b
= −
L
N
MM
O
Q
PP
0
ln
30. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 29
E E
Cu
= − +
0
2
0 059
2
1.
log
Indicator Electrodes
• Electrode potential depends on analyte concentration
• Types of indicating electrodes
- Metal electrodes, type I and type II
- Redox electrodes
- Membrane electrodes, crystalline & liquid
Metal Electrode: type I
• For copper electrode dipped in Cu2+
solution
Cu2+
+ 2e ↔ Cu(s)
•
• E vs. log [Cu2+
] has slope = + 0.0295
Metal Electrode: type II
• For anion which forms sparingly soluble precipitate with electrode metal.
AgI(s) ↔ Ag+
+ I-
Ag+
+ e ↔ Ag(s)
------------------------
AgI(s) +e ↔ Ag(s) + I-
• E = E0
AgI - 0.059 log [I-
]
• E vs. log [I-
] has slope = -0.059
Redox electrode
• Non-reactive metal electrode, e.g. platinum
• Electrode potential depends on relative conc. of reduced and oxidised specie in
equilibrium
Fe3+
+ e ↔ Fe2+
Membrane Electrodes (1)
• Membrane sensitive to ion of interest
• Ion conc. within electrode is constant (std.)
• Potential depends on conc. in analyte sample
E
Fe
Fe
= −
+
+
0 77
0 059
1
2
3
.
.
log
31. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 30
Membrane Electrode (3)
Eind = K+ 0.059 pH
- [Ha
+
] and [He
+
] = internal electrode and external analyte conc. (activity),
respectively.
- K = constant
Probes (1)
• Respond to molecular specie
• Complete cell
• Polarographic type, need external current
• Galvanic cell type
E k
H
H
ind
a
e
= +
+
+
0 059. log
32. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 31
Add copy of Main text in chapter 8, for all participants.
33. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 32
7. Additional handout
These handouts are distributed during delivery and contain test questions, answers to
questions, special worksheets, optional information, and other matters you would not like to
be seen in the regular handouts.
It is a good practice to pre-punch these additional handouts, so the participants can easily
insert them in the main handout folder.
34. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 33
35. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 34
8. Main text
Contents
1. Introduction 1
2. Basic principles of redox reactions 1
3. Potentiometric analysis: 3
3.1 Reference Electrodes 3
3.2 Indicator Electrodes 5
3.3 Measurement of Cell potential 8
36. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 1
Potentiometric Analysis
1. Introduction
Potentiometric and Conductance analyses instruments both rely on electrochemistry: the
relationship between electric and chemical conditions in a solution.
The pH meter is probably the most widely used electrical method of analysis. The pH
measurement is a Potentiometric analysis. In this method, a set of electrodes is immersed
in a solution and direct measurement of difference in electrode potential (voltage) between
the electrodes is carried out. From this, the concentration of an active ion may be derived
(e.g. H+
).
Conductance analysis is employed for measurement of Electric Conductivity (EC). In this
method, the ability of a solution to carry an electrical current is measured using two platinum
electrodes in a reference cell.
Potentiometric analysis and Conductance analysis are two different types of analytical
methods, both based on electric-chemical principles. pH and EC are two of the most
common basic water quality parameters which are routinely measured. Theory and
laboratory determination of EC was discussed in modules # 08 and 09. This module
discusses the principles of potentiometric analysis.
2. Basic principles of redox reactions
Redox reactions and basic electrical-chemical principles were discussed in Module # 30 -
"Advanced Aquatic Chemistry Concepts: Redox Equilibria". These are briefly reviewed here.
Electrode: A simple electrode is a strip of metal immersed in a solution of its own ions. A
chemical change will occur at an electrode, as certain ions in solution migrate towards the
electrode. The reaction occurring at the electrode is a redox-reaction, where electron
transfer is involved. Thus, a potential difference is established between the metal and the
solution.
Half-cell: a single electrode dipping into a solution is also known as a half-cell. Each half-cell
has a specific redox reaction occurring there.
Electrode potential: The electric potential of an electrode is the potential due to electron
transfer in a redox reaction occurring at the electrode. The potential is measured against a
reference half-cell or electrode.
Electrochemical cell (also known as ‘galvanic cell’): The combination of 2 electrodes
(half-cells), creates an electrochemical cell. If a voltmeter is connected across the
electrodes, it can measure the electromotive force being generated by the electrochemical
cell (Ecell).
Cell Potential: The electrochemical cell potential, Ecell, is equal to the difference of the 2
electrode potentials and can be measured in volts, see Figure 1.
37. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 2
Ecell = Ec – Ea
Where:
• Ecell = electrochemical cell potential (volts)
• Ec = the electrode potentials of the cathode
• Ea = the electrode potentials of the anode
Nernst Equation: For the generalised half-reaction
aA + bB… + ne ↔ cC + dD…
the potential is related to the ionic activity of the solutes (approximated by their
concentration) by the expression:
Where:
• E = Electrode Potential, volts
• Eo
= a constant called standard electrode potential; it is equal to E when
the activities of all reactants and products are one.
• R = the Gas Constant, 8.316 J oK-1 mol-1
• T = Absolute temperature, 0
K
• n = number of moles of electrons that appear in the half reaction for the
electrode process
• F = The faraday = 96487 coulombs
• [A], [B], [C], [D] = Active ions in the solution
• a, b, c, d = mole concentrations of the respective ions
Substituting numerical values, simplifying and converting to base-10 logarithm provides:
E E
RT
nF
C D
A B
o
c d
a b
= −
ln
[ ] [ ]
[ ] [ ]
E E
n
C D
A B
o
c d
a b
= −
0 059.
log
[ ] [ ]
[ ] [ ]
38. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 3
The Nernst equation indicates that the electrode potential varies with the log of ionic
activities (concentration). In an electrochemical cell where the potential of one electrode is
known (reference electrode), the measured cell potential can be related to the electrode
potential at the indicator electrode. The indicator electrode potential, is in turn related to the
ionic concentration of the species of interest.
If the cell potential is measured for standard solutions of the ion of interest, using its indicator
electrode, and a calibration is made, the cell can be then used to measure the ion in
unknown solutions.
3. Potentiometric analysis:
In order to determine the potential in an electrochemical cell, a potential measuring device
and two electrodes are necessary. These two electrodes are generally referred as reference
electrode and indicator electrode. The potential of the reference electrode is independent of
concentration of analyte or any other ion present in the sample. Any instrument used in
potentiometric methods includes:
• reference electrode,
• indicator electrode, and
• a potential measuring device(voltmeter).
3.1 Reference Electrodes
There are 3 common reference electrodes used for potentiometric analysis
• Hydrogen electrode
• Calomel electrode
• Silver – silver chloride electrode
Hydrogen Electrode
Figure 2 shows a standard hydrogen electrode. This reference electrode is used as the
standard to which the potentials of all other electrodes are related. The hydrogen ion activity
in a standard hydrogen-electrode is maintained at unity and the hydrogen gas partial
pressure at one atmosphere. The electrode reaction is:
H2(g) + 2e ↔ 2H+
39. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 4
By convention the potential of this half-cell is assigned a value of exactly zero at all
temperatures. The flowing stream of gas needed for the operation of a hydrogen electrode
is inconvenient as well as hazardous. Therefore, ordinarily such an electrode is not used.
Calomel Electrode
The calomel electrode is most widely used reference electrode due to its ease of preparation
and constancy of potential. The calomel electrode is the reference electrode used in the
electronic pH measurement. The calomel electrode is also used in oxidation-reduction
measurements, and in most other electrochemical analyses for which a stable easy-to-use
reference electrode is desired. A simplified diagram of the calomel electrode is shown in
Figure 3.
It has a calomel half-cell in which mercury and calomel (mercury(I) chloride) are covered
with potassium chloride solution of definite concentration (0.1 M,1 M or saturated). The
inner tube containing Hg.Hg2Cl2 is connected to the outer tube containing KCl through a
small opening. Contact with the analyte solution is made through a porous fibre wick or a
fritted disc at the bottom of the outer tube. These electrodes are referred to as the
decimolar, molar and the saturated calomel electrode depending on the concentration of KCl
solution. The potential of each with respect to the hydrogen electrode is given below:
Standard potentials of calomel reference electrodes at 25o
Concentration of KCl EH in volts
0.1 N -0.3358
1.0 N - 0.2824
Saturated -0.2444
Figure 3. Schematic diagram of the Calomel reference electrode
The electrode reaction is: Hg2Cl2 (s) + 2e-
↔ 2Hg (liq) + 2Cl-
Silver-Silver Chloride Electrode
This electrode is another reference electrode consisting of a silver wire or a silver plated
platinum electrode dipped in saturated solution of KCl in a tube. The tube is connected to
analyte through a porous plug at the bottom. In practice, it is no longer commonly used and
therefore is not discussed here.
Hg.Hg2Cl2
for filling
with KCl
40. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 5
3.2 Indicator Electrodes
An indicator electrode of a cell is one whose potential is dependent upon the activity
(concentration) of a particular ionic specie whose concentration is to be determined. Various
types of electrodes used for determination of different ions are discribed below.
Metal electrodes: A metal electrode consists of a metal in contact with its ions. For
example, the half-reaction for the Cu electrode in Figure 1 is,
Cu2+
+ 2e ↔ Cu(s)
and the Nernst expression is written as:
It is seen that the electrode potential is directly related to logarithm of inverse of
concentration of Cu2+
ion in solution in contact with the electrode. A plot of E vs. log [Cu2+
]
would have a slope of +0.0295.
This type of electrode might be used for analysis of metals such as Ag, Cu, Hg, Pb and Cd.
Metal electrodes for other elements cannot be used as they do not behave ideally and do not
give reproducible results.
Another kind of metal electrode may be used for determination of an anion that forms a
sparingly soluble precipitate on reaction with the electrode metal. For example, I-
can be
determined using an Ag electrode. Here two equilibria are involved:
AgI(s) ↔ Ag+
+ I-
and Ag+
+e ↔ Ag
Combining the two equations one gets:
AgI(s) + e ↔ Ag(s) + I-
The Nernst expression for this process is:
E = E0
AgI - 0.0591 log [I-
]
The above equation shows that the potential of a silver electrode immersed in a tube
containing a saturated solution of AgI is directly proportional to the logarithm of iodide
concentration. A plot of E vs. log [1-
] would have a slope of -0.0591.
The electrode is constructed by providing the tube containing the electrode and AgI solution
with an agar plug saturated with AgI and supported on a porous disc at the bottom of the
tube. This prevents the loss of AgI solution from the electrode tube and at the same time
provides contact with the analyte solution.
Redox electrode:
A non-reactive metal electrode, such as platinum, may be used as indicator electrode for
redox systems. Of itself, the electrode is inert; its potential depends solely on the potential of
the chemical system with which it is in contact. As in the case of other indicator electrodes,
it is used in conjunction with a reference electrode.
E E
Cu
= − +
0
2
0 059
2
1.
log
41. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 6
Glass membrane electrodes:
Determination of pH involves measurement of the potential developed across a thin glass
membrane that separates two solutions with different hydrogen ion concentrations. Indicator
glass membrane electrodes for other univalent ions, such as Na+
, K+
, NH+
4, etc., are also
available.
The glass membrane electrode and also other types of membrane electrodes, which are
discussed later, are fundamentally different from metal electrodes both in design and
principle. Since pH is measured routinely potentiometrically in all laboratories, glass
membrane electrode for pH measurement is described in detail.
Figure 4 shows an electrode system, comprising a reference saturated calomel electrode
and a glass indicator electrode, for measurement of pH. The indicator electrode consists of
a thin, pH-sensititve glass membrane sealed onto one end of a heavy-walled glass or plastic
tube. A small amount of dilute HCl that is saturated with silver chloride is contained in the
membrane bulb. A silver wire in this solution forms an internal silver-silver chloride
reference electrode. An internal calomel reference electrode may also be used within the
glass electrode in place of the silver-silver chloride electrode.
Figure 4: Glass electrode system for measuring pH
The thin glass membrane responds to pH. The process involves ion exchange reactions on
the two sides of the membrane in which H+
ions in the solutions are exchanged with
monovalent cations in the glass lattice. The electrode potential established on the pH
electrode is a function of H+
ion activities of the analyte, H a
+
, and the solution within the
electrode, He
+
, on the two sides of the glass membrane:
since [H e
+
] is constant
Eind = K - 0.059 pH
The glass membrane of the pH electrode is of 0.03 to 0.1mm thickness. It is of silicate glass
within which there are singly charged cations, such as sodium and lithium, that are mobile in
the lattice and are responsible for electrical conduction. The surface of the membrane is
hygroscopic and must be hydrated before it will function as a pH electrode. It loses pH
E k
H
H
ind
a
e
= +
+
+
0 059. log
42. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 7
sensitivity after dehydration by storage. The effect is reversible, however, and the response
is restored by soaking in water.
Liquid membrane electrode:
Liquid membrane electrodes have been developed for the direct potentiometric
measurement of both polyvalent cations and polyvalent anions.
Figure 5 shows a schematic diagram of a liquid membrane electrode for calcium. It consists
of a conducting porous membrane impregnated with a liquid ion exchanger that selectively
bonds calcium ions, an internal solution containing a fixed concentration of CaCl2 and a
silver-silver chloride electrode. The similarity to the glass membrane electrode is obvious.
The exchange of Ca2+
ions at the two surfaces of the membrane differs as a consequence of
the difference in Ca++
activity of the internal electrode solution and the external analyte. This
results in the development of the electrode potential as in the case of the pH electrode.
Figure 5: Liquid membrane electrode for Ca2+
Crystalline membrane electrodes:
Many solid crystalline membranes, other than glass, have been developed for specific
anions and cations, such as CN-
, F-
, Pb2+
, Cd2+
, etc. These also function on the same
principle as the glass membrane and liquid membrane electrodes.
Gas-sensing probe:
The electrodes belonging to this class are not strictly ion electrodes because they respond to
dissolved molecular species.
Further, they represent a complete electrochemical cell rather than an electrode. They are,
therefore, more properly called probes.
Figure 6 shows a schematic diagram of an oxygen probe based on principle of polarography.
An inert metal such as gold or platinum serves as cathode and silver is used for the anode.
These are electrically connected through an electrolyte solution. The cell is separated from
the sample by means of a replaceable gas permeable membrane. Any oxygen that passes
43. Hydrology Project Training Module File: “ 32 Potentiometric Analysis of Water Quality.doc” Version 06/11/02 Page 8
through the membrane is reduced at the cathode by externally applying a potential of about
0.5 to 0.8 volt across the anode and cathode. This causes a current to flow. The magnitude
of the current is proportional to the amount of oxygen in the sample.
Figure 6: Schematic diagram for oxygen probe
In a galvanic cell type probe, a basic electrolyte such as KOH and a lead cathod is used in
place of the inert platinum cathode. This produces enough potential to reduce the oxygen
and the magnitude of the current is proportional to the dissolved oxygen concentration in the
analyte solution.
3.3 Measurement of Cell potential
An instrument for potentiometric measurements must not draw any current from the cell as it
will change its potential. Historically, potential measurements were performed with a
potentiometer, a nul-point instrument in which the unknown potential was just balanced by a
standard reference potential. At null, no electricity is withdrawn from the cell whose potential
is being measured.
The potentiometer has now been supplanted by electronic voltmeters, commonly called pH
meters when used for measurement of pH or ion meters when used for measurement of
other ions. These meters directly read the concentration or p-concentration of the chemical
specie being measured after suitable calibration.