Environnement SA Graphite 52M

Heated FID Volatile Organic
Compounds Analyser

Datasheet
Equipped with one burner placed in a heated block, the Graphite
52M allows continuous and accurate Total VOC monitoring.
The FID is temperature controlled up to 191°C to eliminate
condensation. Extremely compact and with a very short response
time, this heated FID provides continuous monitoring of total VOC’s,
in compliance with the Waste Incineration Directive and EN12619
standard as written in most WID permits.
Features:

The Graphite 52M is fully supported by MCERTS
certified CDAS Elite Data Aquisition Software and
a comprehensive service and support package

• Detector placed in a heated oven, up to 191°C allowing 	
	
measurement of high concentrations of heavy hydrocarbons
• Fast response time
• Real time graphic display with interactive menu driven software
• Integrated zero air catalyst
• Dual stream monitoring by independent detectors available
• User selectable ranging
• Very low limits of detection
• Hydrogen/Helium fuel supply thus reducing oxygen interference
• Fully compatible with the
Environnement S.A. MIR-FT & MIR-9000 CEM systems
• MCERTS Certified

Specification
Applications:
• Waste Incinerators
• Thermal Oxidisers
• Exhaust Gas Analysis
• Standard Reference Method for VOC Monitoring
• Laboratories and Research Centres
• Process Control
• Combustion Monitoring
MCERTS Certification Range:
TOC:	 0-15mg/m3

026

Certificate Number 996/96

Specification:
• Ranges: 0-10/100/1000/10000 ppm or user programmable 	 	
range up to 50 000 ppm
• Accuracy: < 1% of the displayed value between 15% and 	 	
100% of the F.S.
• Lower detectable limit: 0.05 ppm of the 10 ppm range
• Response time: < 1 sec. (T90)
• Zero drift: <1%/24h F.S. <2%/7 days F.S.
• Span drift: <1% / 7 days F.S.
• Linearity: <1% for concentration between 10 % and 90% of 	 	
the scale
• Temperature of the heated block: up to 191°C
• Sample flow rate: 0.7 to 2 l/min
• Dimensions: 483 x 470 x 177mm (WxDxH)
• Weight: 22kg approximately
• Power supply: 230 VAC, 50Hz or 115 VAC, 60 Hz
• Power consumption: 500VA during start up
• Operating temperature: +5°C to +40°C

t: +44(0)151 666 8300 f: +44(0)151 666 8329 e: sales@a1-cbiss.com www.a1-cbiss.com

Data sheet g52 m print version

  • 1.
    Environnement SA Graphite52M Heated FID Volatile Organic Compounds Analyser Datasheet Equipped with one burner placed in a heated block, the Graphite 52M allows continuous and accurate Total VOC monitoring. The FID is temperature controlled up to 191°C to eliminate condensation. Extremely compact and with a very short response time, this heated FID provides continuous monitoring of total VOC’s, in compliance with the Waste Incineration Directive and EN12619 standard as written in most WID permits. Features: The Graphite 52M is fully supported by MCERTS certified CDAS Elite Data Aquisition Software and a comprehensive service and support package • Detector placed in a heated oven, up to 191°C allowing measurement of high concentrations of heavy hydrocarbons • Fast response time • Real time graphic display with interactive menu driven software • Integrated zero air catalyst • Dual stream monitoring by independent detectors available • User selectable ranging • Very low limits of detection • Hydrogen/Helium fuel supply thus reducing oxygen interference • Fully compatible with the Environnement S.A. MIR-FT & MIR-9000 CEM systems • MCERTS Certified Specification Applications: • Waste Incinerators • Thermal Oxidisers • Exhaust Gas Analysis • Standard Reference Method for VOC Monitoring • Laboratories and Research Centres • Process Control • Combustion Monitoring MCERTS Certification Range: TOC: 0-15mg/m3 026 Certificate Number 996/96 Specification: • Ranges: 0-10/100/1000/10000 ppm or user programmable range up to 50 000 ppm • Accuracy: < 1% of the displayed value between 15% and 100% of the F.S. • Lower detectable limit: 0.05 ppm of the 10 ppm range • Response time: < 1 sec. (T90) • Zero drift: <1%/24h F.S. <2%/7 days F.S. • Span drift: <1% / 7 days F.S. • Linearity: <1% for concentration between 10 % and 90% of the scale • Temperature of the heated block: up to 191°C • Sample flow rate: 0.7 to 2 l/min • Dimensions: 483 x 470 x 177mm (WxDxH) • Weight: 22kg approximately • Power supply: 230 VAC, 50Hz or 115 VAC, 60 Hz • Power consumption: 500VA during start up • Operating temperature: +5°C to +40°C t: +44(0)151 666 8300 f: +44(0)151 666 8329 e: sales@a1-cbiss.com www.a1-cbiss.com