This document provides an overview of cone-beam computed tomography (CBCT) imaging. It discusses the principles of CBCT imaging, including how CBCT uses a rotating x-ray source and detector to obtain multiple 2D images that are reconstructed into a 3D volume. It describes the components of image production, clinical considerations for CBCT scans, common artifacts, and applications of CBCT imaging such as implant planning and assessment of pathology. In conclusion, CBCT is presented as an effective diagnostic imaging technique that produces high resolution 3D images of the maxillofacial region at lower radiation doses and costs compared to medical CT.
2. CONTENTS
INTRODUCTION
PRINCIPLES OF CBCT IMAGING
COMPONENTS OF IMAGE PRODUCTION
CLINICAL CONSIDERATIONS
IMAGE ARTIFACTS
ADVANTAGES AND DISADVANTAGES
APPLICATIONS
CONCLUSION
3. INTRODUCTION
CBCT IMAGING
Most significant technology advancement in
maxillofacial imaging
In this, the imaging shifts from 2D to a
volumetric approach.
4. PRINCIPLES
CBCT imaging is performed
using a rotating platform
carrying an x-ray source
and detector
5. A divergent cone shaped or pyramidal source of radiation
is directed through region of interest (ROI)
X-ray source and detector rotate around a rotation
center, fixed within center of the ROI
6. • During rotation, multiple sequential planer
projection images are obtained while the x-ray
source and detector move through an arc of 180
to 360 degree
• Single projection image form raw primary data,
which is individually known as - basis, frame or
raw image
• Usually several hundred 2-D basic images are
formed from which the image volume is
calculated.
• Complete series of images is called PROJECTION
DATA
9. X-RAY GENERATION
Patient Stabilization
Sitting, Standing, Supine
With all system, immobilization of the patient’s head
is more important than position because any
movement degrades the final image
Immobilization of head by -
Chin cup OR Bite fork
11. X-ray generator
X-ray generation
continuous or pulsed
When pulsed- exposure time is up to 50% less than
scanning time (this technique reduces patients radiation
dose)
ALARA (As Low As Reasonable Achievable) principle of dose
optimization states that CBCT exposure factor should be
adjusted on the basis of patient’s size.
12. Scan Volume / field of view(FOV)
It is the amount of area to be exposed in a single
scan
DEPENDS ON -
* Detector size and shape
* Beam projection geometry
* Ability to collimate the beam
Shape – cylindrical or spherical
13. It is desirable to limit the field size to the
smallest volume that images the ROI.
This procedure reduces unnecessary
exposure to the patient and produces the
best image by minimum scattered radiation,
which degrade image quality.
16. Scan factor
Number of images forming the “projection data”
throughout the scan is determined by-
1. Detector frame rate (no. of image acquired per sec.)
2. Completeness of the trajectory arc (180 to 360)
3. Rotation speed of source and detector
17. IMAGE DETECTOR
Larger and bulkier Lighter in weight
Circular basis image area Rectangular
Spherical volume Clindrical volume
Cesium iodide scintillator
CBCT units
Image intensifier
tube/charge-coupled
device(II/CCD)
Flat panel
detector(FPDs)
18. Voxel- Volume element
Individual volume element is VOXEL
Voxels form the volumetric data set
CBCT units provide voxel resolution
that are isotropic - equal in all 3 dimension
Determinant of voxel is- Pixel size of detector
Detector with small pixel
Capture few x-ray photon per voxel
19. 3. RECONSTRUCTION
basis projection frames
volumetric data
a single CBCT rotation take less than 20 sec
produce 100 to 600 individual projection frames
Each with more than 1 million pixel with 12 to
16 bits of data assigned to each pixel
These data processed to create volumetric data set(voxel) by a
sequence of software algorithms
a process known as RECONSTRUCTION
20. 2 STAGES OF RECONSTRUCTION PROCESS
1) Preprocessing stage-
performed at acquisition computer
Inherent pixel imperfections should be corrected
Exposure normalization
2) Reconstruction stage-
Corrected images are converted into a special
representation called a sinogram
Sinogram is a composite image developed from multiple
projection images
The final image is constructed from the sinogram with
a filtered back-projection algorithm.
21. CLINICAL CONSIDERATION
1.Patient selection criteria
It provides a radiation dose to the patient higher than
radiation dose of other dental radiograph
When periapical or panoramic view cannot provide the
necessary information
Used as adjunctive diagnostic tool
22. 2. Patient preparation
Appropriate personal radiation barrier protection
Leaded apron - for pregnant patients and children
Lead thyroid collar- to reduce thyroid exposure
Before scan, remove all the
Metallic object
Eyeglass
Jewelry
Metallic partial denture
23. Patient motion can be minimized by
Head stabilization
Chin cups to posterior or lateral head support
Patient should be directed to remain still as
possible before exposure, to breathe slowly
through nose, and to close the eyes.
24. 3. Imaging protocol
Develop to produce image of optimal quality with the
least amount of radiation exposure to the patient
PARAMETERS
Exposure settings
Spatial resolution
Scan time and number of projections
25. 1. Exposure setting
Quality and quantity of x-ray beam depend on
i. Tube voltage(kVp)
ii. Tube current( mA)
CBCT unit manufacturers approach setting exposure in 2
ways-
1. Selection of fixed exposure setting
2. Allow operator manual adjustment of kVp or mA
26. 2. Spatial resolution
Ability of an image to reveal fine detail
Determined by
i. Pixel size
ii. Beam projection geometry
iii. Patient scatter
iv. Focal spot size
v. Number of basis images
vi. Reconstruction algorithm
27. 3. Scan time and number of projection
adjusting the detector frame rate
increase the number of basis image projections
reconstructed image with fewer artifacts and better image
quality
28. 4. Archiving, export, and distribution
Process of CBCT imaging produces 2 data products
1. Volumetric image data from the scan
2. Image report generated by the operator
Both set of data must be archived and distributed
However the export of image data is in Digital Imaging
And Communications in Medicine standard version 3
(DICOM v3) file format.
29. IMAGE ARTIFACTS
An artifact is any distortion or error in the image
Image
artifacts
Inherent
Procedure
related
Introduced
Patient
motion
artifact
30. 1. INHERENT ARTIFACTS
Can arise from limitations in the physical processes
Beam projection geometry, reduced trajectory rotational
arcs, and image reconstruction methods produce 3 type
of artifacts
Scatter
Partial volume averaging
Cone beam effect
31. Scatter-
Result from x-ray photons that are diffracted from
their original path after interaction with matter
Partial volume averaging-
It occur when the selected voxel size of the scan is
larger than the size of the object being imaged
32.
33. Cone beam effect-
Is a potential source of artifacts, especially in the
peripheral portion of scan volume
Can result in
i. Image distortion
ii. Greater peripheral noise
Clinically, the effect can be reduced by positioning of ROI
in the horizontal plane of x-ray beam.
34. 2. Procedure related artifacts
Under sampling of the object can occur when too few basis
projections are provided for image reconstruction or when
rotational trajectory arc are incomplete
Reduced data sample leads to:-
1. misregistration
2. noisier image
which appear as fine striations in the image
35.
36.
37. 3. Introduced artifacts
An x-ray beam pass through an object ,lower energy photons
are absorbed in preference to higher energy photons, this
phenomenon is known as beam hardening
Can result in 2 type of artifacts
1. Distortion of metallic structure as a result of differential
absorption, known as cupping artifact
2. Streaks and dark bands, which when present b/w 2 dense
objects ,create extinction or missing value artifacts
38.
39. 4. Patient motion artifacts
Can cause misregistration of data which appear
as double contours in the reconstructed image
Problem can be minimized by restraining the
head and using a short scan time as possible
40.
41. ADVANTAGES OF CBCT
Less cost
Less space required rapid, quick scanning time
Radiation dose reduction
Image accuracy
Reduced image artifacts
Unlimited number of views
Imaging can be obtained at any angle
Superior representation of bony structure
Powerful diagnostic 3D planning tool
42. DISADVANTAGES OF CBCT
Image noise –
Because radiation from the source transmitted
through tissue in the body, the receptor receives
non uniform information from radiation scattered in
many directions-termed as noise
Poor soft tissue contrast-
Scattered radiation contributes to increased noise of
the image which reduces the contrast of the cone
beam system
44. IMPLANT SITE ASSESSMENT
Provides cross section view of
i. alveolar bone height, width, and angulations
ii. accurate distance from vital structure such as inferior
alveolar canal in mandible and maxillary sinus
45. ORTHODONTICS
1. Used in identification of root resorption
2. Display of position of impacted or supernumerary teeth
3. Relation to adjacent structure
4. Cephalometric analysis
47. MAXILLOFACIAL PATHOSIS
Useful in assessment of trauma
Visualizing the extent and degree of involvement of
benign odontogenic or non odontogenic as well as
osteomyelitis
48. MANDIBULAR THIRD MOLAR
POSITION
To check the relationship of the third molar with the
inferior alveolar canal during extraction
To prevent the nerve damage
49. CONCLUSION
CBCT imaging is an effective volumetric diagnostic
imaging technology that produces accurate, high
resolution images of diagnostic quality in formats
enabling volumetric visualisationof the osseous
structures of the maxillofacial region at lower doses
and costs.