SlideShare a Scribd company logo
1 of 23
QUALITY ASSURANCE OF
BRACHYTHERAPY (Radiation Test)
BY
M.SC. RADIOLOGICAL PHYSICS 3RD SEM.
NIRAB JYOTI SAIKIA
TANI TAKAM DOLEY
Source position accuracy :
Verification of positional accuracy requires that one con firms that the
intended sequence of active sources or dwell positions is delivered to the
correct position in the correct applicator.
METHODS :
 Source positioning accuracy can be determined by directly viewing, by
CCTV, the source moving through a transparent scaled applicator so that
source position can be determined visually with millimeter accuracy .
• Source position accuracy can be checked by taking radiographs of dummy
sources in the applicators with their positions marked on a ready pack film and
combining with autoradiographs of the radioactive sources in the same
applicators. The position of dummy sources and radioactive sources should
correspond within ±1 mm.
• Special test devices have been designed to test source positioning by
autoradiography .The system consists of an acrylic plate with channels drilled
to hold the treatment catheters. Lead sheets of dimensions 20 × 20 × 0.8 mm3
are accurately set into one surface at a regular spacing of 20 mm. The phantom
is placed on a ready pack XV2 film with lead sheets facing the film. The sources
are programmed to stop every 10 mm for intervals suitable for
autoradiography. The autoradiograph thus obtained shows source positions
together with fiducial markers provided by the edges of lead sheets . An
accuracy of ±1 mm in the source positions is acceptable.
Positioning Accuracy and Uniformity (Tolerance: ± 1 mm)
(use autoradiograph / radiograph wherever required)
1. Coincidence between dummy : ------------- mm
and active source positions
2. Accuracy of source position within the : -------------- mm
applicator
3. Reproducibility of dummy source position : -------------- mm
Source strength :
 The source strength for any radionucleide may be specified in terms of
milicuries(mCi).
 The exposure rate at any point is proportional to the product of activity and its
exposure rate constant.
 Measurement of source strength should be measured in term of “Reference air
karma rate” as per ICRU-38.
 Reference air karma rate is the kerma rate in air at a reference distance of 1m,
corrected for air attenuation and scattering. Unit is micro Gy m sq per hour.
 The AAPM recommend air karma strength and Ref. air karma rate numerically
both are the same when reference distance measurement is taken as 1m.
The different methods for source calibration-
A.Open-Air Measurements.
B.Well-Type Ion Chambers.
Open-Air Measurements-
The arrangement consists of a large source to ion chamber distance
relative to source and detector dimensions. The apparatus is set up
as far away as possible potential scattering surfaces. Because the
output from brachytherapy sources is low at large distance, the
chamber volume should be large, For Example, 100 mL or larger, A
signal-to-noise ratio greater than 100:1 should be achievable.
Disadvantage
 Because of difficulty in obtaining “good geometry”
conditions, the open-air-method is a time-consuming
measurement.
 It is not suitable for routine calibration checks required in a
busy department.
Well-Type Ion Chambers :
Routine calibration of brachytherapy sources is usually carried
out with “re-entrant” type ion chamber surrounds the source,
approximately 4 pi Geometry. A source holder is devised to re-
product the source geometry in radiation to the surrounding
chamber walls. As the detection volume is large, it produces
with high ionization current that can be measured with well-
type ion chamber fill with argon gas under high pressure.
Source Strength Verification
1. Quoted source strength (as on ……………) : ----------- mGy. h-1 at 1m
( GBq / Ci )
2. Measured source strength (as on …………) : ---------- mGy. h-1 at 1m
( GBq / Ci )
3 .Deviation between specified : ----------- %
and measured source strength
(Tolerance: ± 3%)
4 .Deviation between measured : ------------- %
mean and individual source strength
(Tolerance: ± 5%)
Source strength verification
• Chamber used: Well type (SourceCheck 4π)
• Electrometer used: PTW Unidos E
• Voltage applied: +400V
• Range: High
 Quoted Source Strength = 22.013 mGym2/hr as on 01/07/2022
 Measured Source Strength = 21.67 mGym2/hr as on 01/07/2022
 Deviation between specified and measured source strength = 1.5%
(Quoted Activity-Measured Activity) *100/Quoted activity
Temporal accuracy :
A treatment system achieves temporal accuracy if each
source sequence or single source dwell position
remains at its intended position for the length of time
specified by the treatment program.
 For manually afterloaded temporary implants, the goal is to develop a
procedure to ensure that the radioactive sources are removed upon
completion of dose delivery.
Contd.
 Remotely afterloaded brachytherapy places treatment duration under
control of an electronic timer. Tests of absolute timer accuracy are required
whenever source calibration is based on an external time standard,
whereas relative tests suffice when the machine timer is used both to
control treatment delivery duration and to integrate charge measurements
during source-strength calibration.
 An accuracy criterion of 62% seems easily achievable, both by manual
afterloading techniques and commercially available remote afterloading
systems.
 In addition, the influence of transit dose on dose delivery accuracy must
be evaluated and corrected for, if necessary. Transit dose is the additional
dose delivered while the source is in motion.
Temporal Accuracy (Tolerance: ± 1 %)
1. Timer error : --------------
2. Timer linearity : ---------------
3. End error : --------------- sec ('or' %)
Timer Accuracy
• Chamber used: Well type (Source Check 4π)
• Electrometer used: PTW Unidos E
• Voltage applied: +400V
• Range: High
• Set time: 60 Sec
 Timer Accuracy = {(Measured - Set)/Set} X 100
 Measured time = 59.41 Sec
 Timer Accuracy = {(Measured - Set)/Set} X 100
% Deviation Timer Accuracy between specified and measured time = -0.99 %
Timer linearity
• Chamber used: Well type (Source Check 4π)
• Electrometer used: PTW Unidos E
• Voltage applied: +400V
• Range: High
• Set time: 60 Sec; Dwell position: 51
 A graph has been plotted with Tset in X-axis and Tmeas in Y-axis.
 From the graph the slope has been found using the most deviated readings on both sides.
 Timer Linearity = (1 - Slope) * 100 = 0.17%
 End Error = 0.2466 Sec
Timer Linearity and End Error
y = 0.9983x + 0.2466
0.00
20.00
40.00
60.00
80.00
100.00
120.00
140.00
160.00
180.00
200.00
0 50 100 150 200
Tmeas
in
seconds
Tset in seconds
Timer Linearity and End Error
Timer error
• Chamber used: Well type (Source Check 4π)
• Electrometer used: PTW Unidos E
• Voltage applied: +400V
• Range: High; Set time: 60 Sec
• Programmed length: 1400 mm
 With interruption at 30Sec = 6.563 nC
 Without interruption = 6.536 nC
 Timer error = (R2-R1)t/(2R1-R2) = 0.24889 Sec
 % Timer error = (timer error in sec * 100)/60
= 0.41%
REFERENCE:
 AAPM TG-40
 AAPM TG-41
 AAPM TG-56
 KHAN’S The Physics of Radiation Therapy
 PROFORMA For ACCEPTANCE/QUALITY ASSURANCE TETS O
REMOTE AFTER LOADING BRACHYTHERAPY UNIT.
Thank You

More Related Content

Similar to brachy Thearapy Quality assurance Radiation test.pptx

Non-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear MedicineNon-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear Medicine
Pawitra Masa-ah
 
Investigation of Anomalous Thrust and Proposal for Future Experimentation
Investigation of Anomalous Thrust and Proposal for Future ExperimentationInvestigation of Anomalous Thrust and Proposal for Future Experimentation
Investigation of Anomalous Thrust and Proposal for Future Experimentation
Brian Kraft
 

Similar to brachy Thearapy Quality assurance Radiation test.pptx (20)

Non-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear MedicineNon-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear Medicine
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
IRJET- Distance Measurement with the Help of Ultrasonic Sensor
IRJET-  	  Distance Measurement with the Help of Ultrasonic SensorIRJET-  	  Distance Measurement with the Help of Ultrasonic Sensor
IRJET- Distance Measurement with the Help of Ultrasonic Sensor
 
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAData Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
 
Investigation of Anomalous Thrust and Proposal for Future Experimentation
Investigation of Anomalous Thrust and Proposal for Future ExperimentationInvestigation of Anomalous Thrust and Proposal for Future Experimentation
Investigation of Anomalous Thrust and Proposal for Future Experimentation
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working ProceduresNon-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461
Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461
Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Validation of IR instrument
Validation of IR instrumentValidation of IR instrument
Validation of IR instrument
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptx
 
Qc
QcQc
Qc
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 
Ayaz_portfolio
Ayaz_portfolioAyaz_portfolio
Ayaz_portfolio
 
Case Studies from 25 Years of Troubleshooting Vibration Problems.pdf
Case Studies from 25 Years of Troubleshooting Vibration Problems.pdfCase Studies from 25 Years of Troubleshooting Vibration Problems.pdf
Case Studies from 25 Years of Troubleshooting Vibration Problems.pdf
 
TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)
 

More from TaushifulHoque

PROJECT modified final of deserdation.pptx
PROJECT modified final of deserdation.pptxPROJECT modified final of deserdation.pptx
PROJECT modified final of deserdation.pptx
TaushifulHoque
 
Multileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptxMultileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptx
TaushifulHoque
 
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
TaushifulHoque
 

More from TaushifulHoque (17)

PROJECT modified final of deserdation.pptx
PROJECT modified final of deserdation.pptxPROJECT modified final of deserdation.pptx
PROJECT modified final of deserdation.pptx
 
Units and measurements in radiation oncologyFINAL.pptx
Units and measurements in radiation oncologyFINAL.pptxUnits and measurements in radiation oncologyFINAL.pptx
Units and measurements in radiation oncologyFINAL.pptx
 
Construction of Radiotherapy centre.pptx
Construction of Radiotherapy centre.pptxConstruction of Radiotherapy centre.pptx
Construction of Radiotherapy centre.pptx
 
New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptxNew Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
 
Presentation film detectors working.pptx
Presentation film detectors working.pptxPresentation film detectors working.pptx
Presentation film detectors working.pptx
 
Basics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptxBasics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptx
 
Basics of transducers phys3360AEP3630.pptx
Basics of transducers phys3360AEP3630.pptxBasics of transducers phys3360AEP3630.pptx
Basics of transducers phys3360AEP3630.pptx
 
RADIOLOGICAL ACCIDENT IN GOIANIA MODULE XIX
RADIOLOGICAL ACCIDENT IN GOIANIA MODULE XIXRADIOLOGICAL ACCIDENT IN GOIANIA MODULE XIX
RADIOLOGICAL ACCIDENT IN GOIANIA MODULE XIX
 
Case history Goiania radiation accident.ppt
Case history Goiania radiation accident.pptCase history Goiania radiation accident.ppt
Case history Goiania radiation accident.ppt
 
Neutron dosimeter personal and area monitoring.pptx
Neutron dosimeter personal and area monitoring.pptxNeutron dosimeter personal and area monitoring.pptx
Neutron dosimeter personal and area monitoring.pptx
 
Multileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptxMultileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptx
 
2014_Blood_Glucose_Measuring_devices.ppt
2014_Blood_Glucose_Measuring_devices.ppt2014_Blood_Glucose_Measuring_devices.ppt
2014_Blood_Glucose_Measuring_devices.ppt
 
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
 
Planning of Nuclear Medicine Facilities.pptx
Planning of Nuclear Medicine Facilities.pptxPlanning of Nuclear Medicine Facilities.pptx
Planning of Nuclear Medicine Facilities.pptx
 
Transport of RAM IAEA
Transport of RAM IAEATransport of RAM IAEA
Transport of RAM IAEA
 
The Radiological accident in Goiânia
The Radiological accident in GoiâniaThe Radiological accident in Goiânia
The Radiological accident in Goiânia
 
CHERNOBYL DISASTER, 1986
CHERNOBYL  DISASTER, 1986CHERNOBYL  DISASTER, 1986
CHERNOBYL DISASTER, 1986
 

Recently uploaded

ONLINE VOTING SYSTEM SE Project for vote
ONLINE VOTING SYSTEM SE Project for voteONLINE VOTING SYSTEM SE Project for vote
ONLINE VOTING SYSTEM SE Project for vote
RaunakRastogi4
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Sérgio Sacani
 
Pteris : features, anatomy, morphology and lifecycle
Pteris : features, anatomy, morphology and lifecyclePteris : features, anatomy, morphology and lifecycle
Pteris : features, anatomy, morphology and lifecycle
Cherry
 
COMPOSTING : types of compost, merits and demerits
COMPOSTING : types of compost, merits and demeritsCOMPOSTING : types of compost, merits and demerits
COMPOSTING : types of compost, merits and demerits
Cherry
 
PODOCARPUS...........................pptx
PODOCARPUS...........................pptxPODOCARPUS...........................pptx
PODOCARPUS...........................pptx
Cherry
 

Recently uploaded (20)

ONLINE VOTING SYSTEM SE Project for vote
ONLINE VOTING SYSTEM SE Project for voteONLINE VOTING SYSTEM SE Project for vote
ONLINE VOTING SYSTEM SE Project for vote
 
Genome Projects : Human, Rice,Wheat,E coli and Arabidopsis.
Genome Projects : Human, Rice,Wheat,E coli and Arabidopsis.Genome Projects : Human, Rice,Wheat,E coli and Arabidopsis.
Genome Projects : Human, Rice,Wheat,E coli and Arabidopsis.
 
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY // USES OF ANTIOBIOTICS TYPES OF ANTIB...
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY  // USES OF ANTIOBIOTICS TYPES OF ANTIB...ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY  // USES OF ANTIOBIOTICS TYPES OF ANTIB...
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY // USES OF ANTIOBIOTICS TYPES OF ANTIB...
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
 
Pteris : features, anatomy, morphology and lifecycle
Pteris : features, anatomy, morphology and lifecyclePteris : features, anatomy, morphology and lifecycle
Pteris : features, anatomy, morphology and lifecycle
 
BHUBANESHWAR ODIA CALL GIRL SEIRVEC ❣️ 72051//37929❣️ CALL GIRL IN ODIA HAND ...
BHUBANESHWAR ODIA CALL GIRL SEIRVEC ❣️ 72051//37929❣️ CALL GIRL IN ODIA HAND ...BHUBANESHWAR ODIA CALL GIRL SEIRVEC ❣️ 72051//37929❣️ CALL GIRL IN ODIA HAND ...
BHUBANESHWAR ODIA CALL GIRL SEIRVEC ❣️ 72051//37929❣️ CALL GIRL IN ODIA HAND ...
 
Selaginella: features, morphology ,anatomy and reproduction.
Selaginella: features, morphology ,anatomy and reproduction.Selaginella: features, morphology ,anatomy and reproduction.
Selaginella: features, morphology ,anatomy and reproduction.
 
Dr. E. Muralinath_ Blood indices_clinical aspects
Dr. E. Muralinath_ Blood indices_clinical  aspectsDr. E. Muralinath_ Blood indices_clinical  aspects
Dr. E. Muralinath_ Blood indices_clinical aspects
 
Role of AI in seed science Predictive modelling and Beyond.pptx
Role of AI in seed science  Predictive modelling and  Beyond.pptxRole of AI in seed science  Predictive modelling and  Beyond.pptx
Role of AI in seed science Predictive modelling and Beyond.pptx
 
GBSN - Microbiology (Unit 3)Defense Mechanism of the body
GBSN - Microbiology (Unit 3)Defense Mechanism of the body GBSN - Microbiology (Unit 3)Defense Mechanism of the body
GBSN - Microbiology (Unit 3)Defense Mechanism of the body
 
Genome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptxGenome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptx
 
Terpineol and it's characterization pptx
Terpineol and it's characterization pptxTerpineol and it's characterization pptx
Terpineol and it's characterization pptx
 
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
 
Efficient spin-up of Earth System Models usingsequence acceleration
Efficient spin-up of Earth System Models usingsequence accelerationEfficient spin-up of Earth System Models usingsequence acceleration
Efficient spin-up of Earth System Models usingsequence acceleration
 
COMPOSTING : types of compost, merits and demerits
COMPOSTING : types of compost, merits and demeritsCOMPOSTING : types of compost, merits and demerits
COMPOSTING : types of compost, merits and demerits
 
PODOCARPUS...........................pptx
PODOCARPUS...........................pptxPODOCARPUS...........................pptx
PODOCARPUS...........................pptx
 
TransientOffsetin14CAftertheCarringtonEventRecordedbyPolarTreeRings
TransientOffsetin14CAftertheCarringtonEventRecordedbyPolarTreeRingsTransientOffsetin14CAftertheCarringtonEventRecordedbyPolarTreeRings
TransientOffsetin14CAftertheCarringtonEventRecordedbyPolarTreeRings
 
Clean In Place(CIP).pptx .
Clean In Place(CIP).pptx                 .Clean In Place(CIP).pptx                 .
Clean In Place(CIP).pptx .
 
Understanding Partial Differential Equations: Types and Solution Methods
Understanding Partial Differential Equations: Types and Solution MethodsUnderstanding Partial Differential Equations: Types and Solution Methods
Understanding Partial Differential Equations: Types and Solution Methods
 
module for grade 9 for distance learning
module for grade 9 for distance learningmodule for grade 9 for distance learning
module for grade 9 for distance learning
 

brachy Thearapy Quality assurance Radiation test.pptx

  • 1. QUALITY ASSURANCE OF BRACHYTHERAPY (Radiation Test) BY M.SC. RADIOLOGICAL PHYSICS 3RD SEM. NIRAB JYOTI SAIKIA TANI TAKAM DOLEY
  • 2. Source position accuracy : Verification of positional accuracy requires that one con firms that the intended sequence of active sources or dwell positions is delivered to the correct position in the correct applicator. METHODS :  Source positioning accuracy can be determined by directly viewing, by CCTV, the source moving through a transparent scaled applicator so that source position can be determined visually with millimeter accuracy .
  • 3.
  • 4. • Source position accuracy can be checked by taking radiographs of dummy sources in the applicators with their positions marked on a ready pack film and combining with autoradiographs of the radioactive sources in the same applicators. The position of dummy sources and radioactive sources should correspond within ±1 mm. • Special test devices have been designed to test source positioning by autoradiography .The system consists of an acrylic plate with channels drilled to hold the treatment catheters. Lead sheets of dimensions 20 × 20 × 0.8 mm3 are accurately set into one surface at a regular spacing of 20 mm. The phantom is placed on a ready pack XV2 film with lead sheets facing the film. The sources are programmed to stop every 10 mm for intervals suitable for autoradiography. The autoradiograph thus obtained shows source positions together with fiducial markers provided by the edges of lead sheets . An accuracy of ±1 mm in the source positions is acceptable.
  • 5. Positioning Accuracy and Uniformity (Tolerance: ± 1 mm) (use autoradiograph / radiograph wherever required) 1. Coincidence between dummy : ------------- mm and active source positions 2. Accuracy of source position within the : -------------- mm applicator 3. Reproducibility of dummy source position : -------------- mm
  • 6. Source strength :  The source strength for any radionucleide may be specified in terms of milicuries(mCi).  The exposure rate at any point is proportional to the product of activity and its exposure rate constant.  Measurement of source strength should be measured in term of “Reference air karma rate” as per ICRU-38.  Reference air karma rate is the kerma rate in air at a reference distance of 1m, corrected for air attenuation and scattering. Unit is micro Gy m sq per hour.  The AAPM recommend air karma strength and Ref. air karma rate numerically both are the same when reference distance measurement is taken as 1m.
  • 7. The different methods for source calibration- A.Open-Air Measurements. B.Well-Type Ion Chambers. Open-Air Measurements- The arrangement consists of a large source to ion chamber distance relative to source and detector dimensions. The apparatus is set up as far away as possible potential scattering surfaces. Because the output from brachytherapy sources is low at large distance, the chamber volume should be large, For Example, 100 mL or larger, A signal-to-noise ratio greater than 100:1 should be achievable.
  • 8.
  • 9. Disadvantage  Because of difficulty in obtaining “good geometry” conditions, the open-air-method is a time-consuming measurement.  It is not suitable for routine calibration checks required in a busy department.
  • 10. Well-Type Ion Chambers : Routine calibration of brachytherapy sources is usually carried out with “re-entrant” type ion chamber surrounds the source, approximately 4 pi Geometry. A source holder is devised to re- product the source geometry in radiation to the surrounding chamber walls. As the detection volume is large, it produces with high ionization current that can be measured with well- type ion chamber fill with argon gas under high pressure.
  • 11.
  • 12. Source Strength Verification 1. Quoted source strength (as on ……………) : ----------- mGy. h-1 at 1m ( GBq / Ci ) 2. Measured source strength (as on …………) : ---------- mGy. h-1 at 1m ( GBq / Ci ) 3 .Deviation between specified : ----------- % and measured source strength (Tolerance: ± 3%) 4 .Deviation between measured : ------------- % mean and individual source strength (Tolerance: ± 5%)
  • 13. Source strength verification • Chamber used: Well type (SourceCheck 4π) • Electrometer used: PTW Unidos E • Voltage applied: +400V • Range: High  Quoted Source Strength = 22.013 mGym2/hr as on 01/07/2022  Measured Source Strength = 21.67 mGym2/hr as on 01/07/2022  Deviation between specified and measured source strength = 1.5% (Quoted Activity-Measured Activity) *100/Quoted activity
  • 14.
  • 15. Temporal accuracy : A treatment system achieves temporal accuracy if each source sequence or single source dwell position remains at its intended position for the length of time specified by the treatment program.  For manually afterloaded temporary implants, the goal is to develop a procedure to ensure that the radioactive sources are removed upon completion of dose delivery.
  • 16. Contd.  Remotely afterloaded brachytherapy places treatment duration under control of an electronic timer. Tests of absolute timer accuracy are required whenever source calibration is based on an external time standard, whereas relative tests suffice when the machine timer is used both to control treatment delivery duration and to integrate charge measurements during source-strength calibration.  An accuracy criterion of 62% seems easily achievable, both by manual afterloading techniques and commercially available remote afterloading systems.  In addition, the influence of transit dose on dose delivery accuracy must be evaluated and corrected for, if necessary. Transit dose is the additional dose delivered while the source is in motion.
  • 17. Temporal Accuracy (Tolerance: ± 1 %) 1. Timer error : -------------- 2. Timer linearity : --------------- 3. End error : --------------- sec ('or' %)
  • 18. Timer Accuracy • Chamber used: Well type (Source Check 4π) • Electrometer used: PTW Unidos E • Voltage applied: +400V • Range: High • Set time: 60 Sec  Timer Accuracy = {(Measured - Set)/Set} X 100  Measured time = 59.41 Sec  Timer Accuracy = {(Measured - Set)/Set} X 100 % Deviation Timer Accuracy between specified and measured time = -0.99 %
  • 19. Timer linearity • Chamber used: Well type (Source Check 4π) • Electrometer used: PTW Unidos E • Voltage applied: +400V • Range: High • Set time: 60 Sec; Dwell position: 51  A graph has been plotted with Tset in X-axis and Tmeas in Y-axis.  From the graph the slope has been found using the most deviated readings on both sides.  Timer Linearity = (1 - Slope) * 100 = 0.17%  End Error = 0.2466 Sec
  • 20. Timer Linearity and End Error y = 0.9983x + 0.2466 0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 180.00 200.00 0 50 100 150 200 Tmeas in seconds Tset in seconds Timer Linearity and End Error
  • 21. Timer error • Chamber used: Well type (Source Check 4π) • Electrometer used: PTW Unidos E • Voltage applied: +400V • Range: High; Set time: 60 Sec • Programmed length: 1400 mm  With interruption at 30Sec = 6.563 nC  Without interruption = 6.536 nC  Timer error = (R2-R1)t/(2R1-R2) = 0.24889 Sec  % Timer error = (timer error in sec * 100)/60 = 0.41%
  • 22. REFERENCE:  AAPM TG-40  AAPM TG-41  AAPM TG-56  KHAN’S The Physics of Radiation Therapy  PROFORMA For ACCEPTANCE/QUALITY ASSURANCE TETS O REMOTE AFTER LOADING BRACHYTHERAPY UNIT.