SlideShare a Scribd company logo
1 of 23
Qualification of analytical
instruments: Electronic
balance,
UV-Visible
spectrophotometer
Submitted by
Pattan Sohail
M pharm 1st year
• Validation of an analytical procedure is the process
by which it is established, by laboratory studies, that
the performance characteristics of the procedure
meet the requirements for the intended analytical
applications.” (Definition of Validation: USP)
• Qualification is part of validation but the individual
qualification steps alone do not constitute process
validation
• Qualification deals with components or elements
of a process
• Validation deals with the entire manufacturing
process of a product
QUALIFICATION OF BALANCES
“Qualification of Equipment”, which together should
be used when planning, performing and
documenting the qualification process of balances
• The core document contains the introduction and
general forms for Level I and II of qualification, which
are common to all types of instruments.
AIM AND SCOPE OF THE GUIDELINE
This guideline describes the requirements for balances
(electronic - digital) used in chemical and biological
tests in OMCLs. The following types of balances are
considered in this guideline
Type Ordinary name No.of digits after
decimal
position(g)
Accuracy class
1. Ultra micro
balance
7 I
2. Micro balance 6 I
3. Semi-micro
balance
5 I
4. Analytical balance 4 I
5. Precision balance 1 TO 3 II
6. Technical balance 0 TO 1 III
CONSIDERATIONS FOR LEVEL I AND II OF
EQUIPMENT:
QUALIFICATION At Level I of the qualification of
balances (selection of instruments and suppliers), it
is recommended to select a manufacturer of
balances that can satisfy the needs of the laboratory
and works under ISO 9001 certification. At Level II of
the qualification of balances (installation and release
for use), it is recommended to check all the
requirements set during the selection of the
instrument and calibration should be performed
before putting into service by an accredited external
service supplier, or internally by appropriately
qualified personnel, using certified reference weights
according to an approved procedure.
4. REQUIREMENTS FOR BALANCE USE
Location of the Balance
The accuracy and precision of weighing
results are closely associated with the location of the
balance. OMCLs should ensure that the balance can
work under optimal conditions (weighing
room/laboratory, weighing bench, temperature,
light, air, etc,.)
• It should be ensured that the weighing bench is
stable, no matter from what material it is made. The
weighing bench should not deform when work is
carried out on it and it must be vibration-proof or the
transfer of vibrations must not influence the
weighing process.
Temperature
As weighing results are influenced by temperature,
OMCLs should ensure a constant temperature is
maintained in weighing rooms/laboratories. The
deviation should not exceed more than 5 °C per hour.
Atmospheric humidity
The optimum relative humidity (% RH) during a
weighing process is between 40 % to 60 % 80 % in
cases where the accuracy and linearity of
measurements are not affected. Light Balances
should be protected from direct sunlight (heat).
Air
OMCLs should not place balances in the airflow of air
conditioners or devices with ventilators next to
doorways or in areas of high traffic.
Weighing vessel
OMCLs must ensure that:
- the smallest possible weighing vessels are used.
- if the materials that the weighing vessel is made of
have a high degree of electrical insulation they are
not electrostatically charged.
FREQUENCY OF QUALIFICATION
OMCLs must ensure that qualification/calibration of
balances is done on receipt (i.e. immediately after
delivery) or prior to their first use and after any
repair or move.
The frequency of qualification/calibration depends
on the extent of use of the balances and is at the
discretion of individual OMCLs.
Qualification/calibration must be performed in
accordance with a pre-determined protocol in which
acceptance criteria are defined
• k = correlation coefficient
• SD = standard deviation
• RSD = relative standard deviation
Parameter to be checked Frequency Typical tolerance limit
Levelling Every day before weighing
begins
Acceptance limits of the
balance
Internal caliberation Every day before weighing
begins
Automatic acceptance
limits of the balanc
Verification Atleast once a week OMCLs shall define their
own acceptance criteri.
Accuracy Frequency to be defined
by OMCL,typically once a
year
OMCLs shall define their
own acceptance criteri.
Linearity Frequency to be defined
by OMCL,typically once a
year
OMCLs shall define their
own acceptance criteri.
(k= 1±0.0001)
Precission Frequency to be defined
by OMCL,typically once a
year
OMCLs shall define their
own acceptance criteri.
(SD=max 5*d)
Eccentricity Frequency to be defined
by OMCL,typically once a
year
OMCLs shall define their
own acceptance criteri.
(RSD=0.05%)
Verification
Verification of the balance is performed by placing a
suitable weight in the centre of the weighing pan
once and comparing the result with pre-defined
acceptance criteria. The same weight should always
be used in these verifications. The acceptance
criteria shall be defined by each individual OMCL.
Accuracy
The accuracy of the balance is checked by weighing
at least three different certified weights that cover
the usual weighing range of the balance. It is
recommended that the weights have approximately
5%, 50% and 100% of the maximum capacity of the
balance depending on the type of balance. It is
recommended that the weighing is repeated at least
5 times for every weight.
Linearity
The results obtained from a series of accuracy checks can
be used to calculate the correlation coefficient and to
check for linearity. The correlation coefficient is
calculated by comparing the nominal and measured
masses of the weights. The acceptance criteria for the
correlation coefficient shall be defined by each individual
OMCL.
Precision
The precision of the balance should be verified by
weighing at least 5 times a weight that is equivalent to
approximately 50% of the maximum capacity of the
balance. It is recommended to repeat the test with a
weight that is equivalent to approximately 5% of the
maximum capacity of the balance, if the balance is used
at the lower range (Proposed criterion: SD = max 5*d,
where d = (actual) scale interval (e.g. d=0.1 mg).
CALIBRATION OF UV-VISIBLE
SPECTROPHOTOMETER
Tests Performed In UV-Visible
Spectroscopy:
 Baseline flatness test
 UV region wave length accuracy test
 Visible region wave length accuracy
test
 %Transmittance of Cuvette
 Control of absorbance
 Emission lines from deuterium
discharge lamp
 Stray Light measurement
 Resolution power
 Photometric linearity
Make,
Soft ware
& Version
Model
type &
Serial No
Beam
type
(single/d
ouble)
INS/
EQP.ID
No.
SOP No Calibratio
n
Frequenc
y
Every six
months
Instrument/Equipment Calibration Record
INS/ EQP Name: UV-VISIBLE spectrophotometer
Department Name: Central Instrumentation Lab
Equipment Details:
1) Baseline Flatness Test:
Measurement configuration parameters:
Reference : No Cuvette
Sample : No Cuvette
Cuvette material : Quartz
Wave length range : 200-800
Lamp : Deuterium lamp
Detector : PDA detector
WAVE LENGTH MEASURED ABSORBANCE ACCEPTANCE CRITERIA
200 ±0.001
300 ±0.001
400 ±0.001
500 ±0.001
600 ±0.001
700 ±0.001
800 ±0.001
2) Uv Region Wave Length Accuracy Test:
Measurement configuration Parameters
Reference : Air (Empty Cuvette)
Sample : Benzene
Recording range : 220-270nm
Lamp : Deuterium Discharge
Detector : PDA Detector
WAVE LENGTH (nm) OBSERVED PEAKS AT
WAVE LENGTH (nm)
ACCEPTANCE CRITERIA
220-270 nm 235.9±0.5nm
3) Visible Region Wave Length Accuracy Test:
Measurement Parameters:
Reference : Distilled Water
Sample : 0.01% KMno4
Measuring mode : 300 to 600nm
Lamp : Deuterium
Discharge Detector : PDA Detector
WAVE LENGTH (nm) OBSERVED PEAKS AT
WAVE LENGTH(nm)
ACCEPTANCE CRITERIA
300-600nm 310±1nm
525±1nm
4) %Transmittance Of Cuvette:
Measurement configuration Parameters
Reference : Empty Cuvette
Sample : Empty Cuvette
Cuvette material : Quartz
Measuring mode : 240nm
Lamp : Deuterium discharge lamp
Detector : PDADetector
WAVE LENGTH(nm) %TRANSMITTANCE ACCEPTANCE
CRITERIA
240nm Not less then 80%
5) Control of absorbance:
Absorbance:Visible region
 Place dummy Cuvette in sample holder and set %T to
“zero”. Now remove dummy Cuvette, by using fine & coarse
control set a reading exactly 40.0 on the read out. Press
Absorbance push button. If the maximum absorbance
obtained at λ of 485nm is 0.398 ± 0.002, the photometric
calibration of instrument is confirmed to be proper.
 To confirm, repeat above steps, and set 10.00 on read out
 Press Absorbance button. If the λ at 485 nm is 1.000±0.002
then it is confirmed the photometric performance in the
visible region is proper.
Absorbance:U.V region
Place blank0.1N H2SO4 Cuvette and 60ppm K2Cr2O7 as
sample
Set λ exactly to 257 nm, if the value of Absorbance of
sample at the set λ is 0.864±0.005, the instrument is
measuringAbsorbance properly.
6) Emission Lines From Deuterium Discharge Lamp:
Measurement configuration Parameters:
Reference : No Cuvette
Sample : Empty Cuvette
Recording range : 400-500, 600-700
THEORETICAL
VALUE
OBSERVED
VALUE
DIFFERENCE ACCEPTANCE
CRITERIA
486nm 486±0.3nm
656nm 656±0.3nm
7) Stray Light measurement:
 Weigh accurately 1.2g of dried Potassium chloride in
100 ml volumetric flask and makeup to mark with
Double distilled water.
 Measure the absorbance at 200 nm.
Acceptance criteria: Tolerance limit NLT 2.0
8)Resolution Power:
 Prepare 0.02%v/v solution of Toluene and make up
with Hexane.
 Scan the wavelength from 250 to 280nm.
 Maximum absorbance is 269 nm and Minimum
absorbance is 266nm
 Acceptance criteria: Ratio limit NLT 1.5
9) Photometric linearity:
 Weigh accurately 100mg of Potassium chromate in
100ml volumetric flask and dissolve in 0.05N
Potassium hydroxide solution. Make up with the
same solvent.
 From the above solution take 20ml and make up
to 500ml with 0.05N Potassium hydroxide solution.
 Now prepare dilution of 4,8,16,24,32 µg/ml.
 Measure the absorbance at 370nm using blank.
Acceptance criteria: The plot should be linear and
regression coefficient (R2) should NLT 0.999.

More Related Content

What's hot

Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplcANANT NAG
 
Validation of utility system (water system)
Validation of utility system (water system)Validation of utility system (water system)
Validation of utility system (water system)ShameerAbid
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaningKUNDLAJAYALAKSHMI
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solventsMehulJain143
 
Qualification of Analytical Equipments
Qualification of Analytical EquipmentsQualification of Analytical Equipments
Qualification of Analytical EquipmentsDhawal_Raghuvanshi
 
qualification of lcms
qualification of lcmsqualification of lcms
qualification of lcmsAnkush Sule
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumetsChowdaryPavani
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemAkshay Trivedi , Maliba Pharmacy College
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserKUNDLAJAYALAKSHMI
 
Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Dr Raj kumar Kudari
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry personabhishek pandey
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instrumentsSanathoiba Singha
 

What's hot (20)

Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplc
 
Validation of utility system (water system)
Validation of utility system (water system)Validation of utility system (water system)
Validation of utility system (water system)
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
QUALIFICATION OF FT-IR
QUALIFICATION OF FT-IRQUALIFICATION OF FT-IR
QUALIFICATION OF FT-IR
 
CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT
 
Qualification of Gas Chromatography
Qualification of Gas Chromatography Qualification of Gas Chromatography
Qualification of Gas Chromatography
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solvents
 
Qualification of Analytical Equipments
Qualification of Analytical EquipmentsQualification of Analytical Equipments
Qualification of Analytical Equipments
 
Requalification
RequalificationRequalification
Requalification
 
qualification of lcms
qualification of lcmsqualification of lcms
qualification of lcms
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyser
 
Validation qualification
Validation qualificationValidation qualification
Validation qualification
 
Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry person
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Impurity Profile
Impurity ProfileImpurity Profile
Impurity Profile
 

Similar to qualification of analytical instruments..M pharmacy 1st year.validation

Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationMohiniTawade
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instrumentsNavdha Soni
 
Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC LabDr. Amsavel A
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)PRAVADA
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxsanjayudps2016
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
 
Techniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptxTechniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptxAREEBA SHAFIQ
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1Complaint2015
 
calabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfcalabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfsharonmarishkawilfre
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Malarvili14
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjPratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxPratik434909
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...SELINA SRAVANTHI
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniMayuri Soni
 

Similar to qualification of analytical instruments..M pharmacy 1st year.validation (20)

Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instruments
 
Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC Lab
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptx
 
CALIBRATIONS.pptx
CALIBRATIONS.pptxCALIBRATIONS.pptx
CALIBRATIONS.pptx
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Techniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptxTechniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptx
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1
 
calabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfcalabration-principles-chapter1.pdf
calabration-principles-chapter1.pdf
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
 
Calibration
CalibrationCalibration
Calibration
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
QUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLCQUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLC
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri Soni
 

More from SohailPattan

chemistry and classification of amino acids\protiens
chemistry and classification of amino acids\protienschemistry and classification of amino acids\protiens
chemistry and classification of amino acids\protiensSohailPattan
 
Chemistry, classification & properties of carbohydrates
Chemistry, classification & properties of carbohydratesChemistry, classification & properties of carbohydrates
Chemistry, classification & properties of carbohydratesSohailPattan
 
bio luminescence and fluroimmuno assays & Immunological assays
bio luminescence and fluroimmuno assays & Immunological assaysbio luminescence and fluroimmuno assays & Immunological assays
bio luminescence and fluroimmuno assays & Immunological assaysSohailPattan
 
modern pharmaceutical analysis, immunological assays
modern pharmaceutical analysis,  immunological assays modern pharmaceutical analysis,  immunological assays
modern pharmaceutical analysis, immunological assays SohailPattan
 
Electro magnetic resonance & its relation with frequency,wave length and wave...
Electro magnetic resonance & its relation with frequency,wave length and wave...Electro magnetic resonance & its relation with frequency,wave length and wave...
Electro magnetic resonance & its relation with frequency,wave length and wave...SohailPattan
 
Theory of uv spectrophotometer
Theory of uv spectrophotometerTheory of uv spectrophotometer
Theory of uv spectrophotometerSohailPattan
 

More from SohailPattan (6)

chemistry and classification of amino acids\protiens
chemistry and classification of amino acids\protienschemistry and classification of amino acids\protiens
chemistry and classification of amino acids\protiens
 
Chemistry, classification & properties of carbohydrates
Chemistry, classification & properties of carbohydratesChemistry, classification & properties of carbohydrates
Chemistry, classification & properties of carbohydrates
 
bio luminescence and fluroimmuno assays & Immunological assays
bio luminescence and fluroimmuno assays & Immunological assaysbio luminescence and fluroimmuno assays & Immunological assays
bio luminescence and fluroimmuno assays & Immunological assays
 
modern pharmaceutical analysis, immunological assays
modern pharmaceutical analysis,  immunological assays modern pharmaceutical analysis,  immunological assays
modern pharmaceutical analysis, immunological assays
 
Electro magnetic resonance & its relation with frequency,wave length and wave...
Electro magnetic resonance & its relation with frequency,wave length and wave...Electro magnetic resonance & its relation with frequency,wave length and wave...
Electro magnetic resonance & its relation with frequency,wave length and wave...
 
Theory of uv spectrophotometer
Theory of uv spectrophotometerTheory of uv spectrophotometer
Theory of uv spectrophotometer
 

Recently uploaded

100-Concepts-of-AI by Anupama Kate .pptx
100-Concepts-of-AI by Anupama Kate .pptx100-Concepts-of-AI by Anupama Kate .pptx
100-Concepts-of-AI by Anupama Kate .pptxAnupama Kate
 
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Callshivangimorya083
 
From idea to production in a day – Leveraging Azure ML and Streamlit to build...
From idea to production in a day – Leveraging Azure ML and Streamlit to build...From idea to production in a day – Leveraging Azure ML and Streamlit to build...
From idea to production in a day – Leveraging Azure ML and Streamlit to build...Florian Roscheck
 
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝soniya singh
 
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdf
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdfMarket Analysis in the 5 Largest Economic Countries in Southeast Asia.pdf
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdfRachmat Ramadhan H
 
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...Suhani Kapoor
 
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdf
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdfKantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdf
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdfSocial Samosa
 
B2 Creative Industry Response Evaluation.docx
B2 Creative Industry Response Evaluation.docxB2 Creative Industry Response Evaluation.docx
B2 Creative Industry Response Evaluation.docxStephen266013
 
Call Girls In Mahipalpur O9654467111 Escorts Service
Call Girls In Mahipalpur O9654467111  Escorts ServiceCall Girls In Mahipalpur O9654467111  Escorts Service
Call Girls In Mahipalpur O9654467111 Escorts ServiceSapana Sha
 
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service Amravati
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service AmravatiVIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service Amravati
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service AmravatiSuhani Kapoor
 
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...Suhani Kapoor
 
Beautiful Sapna Vip Call Girls Hauz Khas 9711199012 Call /Whatsapps
Beautiful Sapna Vip  Call Girls Hauz Khas 9711199012 Call /WhatsappsBeautiful Sapna Vip  Call Girls Hauz Khas 9711199012 Call /Whatsapps
Beautiful Sapna Vip Call Girls Hauz Khas 9711199012 Call /Whatsappssapnasaifi408
 
RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998YohFuh
 
Customer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxCustomer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxEmmanuel Dauda
 
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...shivangimorya083
 
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130Suhani Kapoor
 

Recently uploaded (20)

E-Commerce Order PredictionShraddha Kamble.pptx
E-Commerce Order PredictionShraddha Kamble.pptxE-Commerce Order PredictionShraddha Kamble.pptx
E-Commerce Order PredictionShraddha Kamble.pptx
 
100-Concepts-of-AI by Anupama Kate .pptx
100-Concepts-of-AI by Anupama Kate .pptx100-Concepts-of-AI by Anupama Kate .pptx
100-Concepts-of-AI by Anupama Kate .pptx
 
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
 
VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
 
From idea to production in a day – Leveraging Azure ML and Streamlit to build...
From idea to production in a day – Leveraging Azure ML and Streamlit to build...From idea to production in a day – Leveraging Azure ML and Streamlit to build...
From idea to production in a day – Leveraging Azure ML and Streamlit to build...
 
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
 
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdf
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdfMarket Analysis in the 5 Largest Economic Countries in Southeast Asia.pdf
Market Analysis in the 5 Largest Economic Countries in Southeast Asia.pdf
 
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...
VIP High Class Call Girls Bikaner Anushka 8250192130 Independent Escort Servi...
 
꧁❤ Aerocity Call Girls Service Aerocity Delhi ❤꧂ 9999965857 ☎️ Hard And Sexy ...
꧁❤ Aerocity Call Girls Service Aerocity Delhi ❤꧂ 9999965857 ☎️ Hard And Sexy ...꧁❤ Aerocity Call Girls Service Aerocity Delhi ❤꧂ 9999965857 ☎️ Hard And Sexy ...
꧁❤ Aerocity Call Girls Service Aerocity Delhi ❤꧂ 9999965857 ☎️ Hard And Sexy ...
 
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdf
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdfKantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdf
Kantar AI Summit- Under Embargo till Wednesday, 24th April 2024, 4 PM, IST.pdf
 
B2 Creative Industry Response Evaluation.docx
B2 Creative Industry Response Evaluation.docxB2 Creative Industry Response Evaluation.docx
B2 Creative Industry Response Evaluation.docx
 
Call Girls In Mahipalpur O9654467111 Escorts Service
Call Girls In Mahipalpur O9654467111  Escorts ServiceCall Girls In Mahipalpur O9654467111  Escorts Service
Call Girls In Mahipalpur O9654467111 Escorts Service
 
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service Amravati
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service AmravatiVIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service Amravati
VIP Call Girls in Amravati Aarohi 8250192130 Independent Escort Service Amravati
 
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...
VIP High Profile Call Girls Amravati Aarushi 8250192130 Independent Escort Se...
 
Beautiful Sapna Vip Call Girls Hauz Khas 9711199012 Call /Whatsapps
Beautiful Sapna Vip  Call Girls Hauz Khas 9711199012 Call /WhatsappsBeautiful Sapna Vip  Call Girls Hauz Khas 9711199012 Call /Whatsapps
Beautiful Sapna Vip Call Girls Hauz Khas 9711199012 Call /Whatsapps
 
RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998
 
Customer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxCustomer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptx
 
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...
Full night 🥵 Call Girls Delhi New Friends Colony {9711199171} Sanya Reddy ✌️o...
 
Deep Generative Learning for All - The Gen AI Hype (Spring 2024)
Deep Generative Learning for All - The Gen AI Hype (Spring 2024)Deep Generative Learning for All - The Gen AI Hype (Spring 2024)
Deep Generative Learning for All - The Gen AI Hype (Spring 2024)
 
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
 

qualification of analytical instruments..M pharmacy 1st year.validation

  • 1. Qualification of analytical instruments: Electronic balance, UV-Visible spectrophotometer Submitted by Pattan Sohail M pharm 1st year
  • 2. • Validation of an analytical procedure is the process by which it is established, by laboratory studies, that the performance characteristics of the procedure meet the requirements for the intended analytical applications.” (Definition of Validation: USP) • Qualification is part of validation but the individual qualification steps alone do not constitute process validation • Qualification deals with components or elements of a process • Validation deals with the entire manufacturing process of a product
  • 3. QUALIFICATION OF BALANCES “Qualification of Equipment”, which together should be used when planning, performing and documenting the qualification process of balances • The core document contains the introduction and general forms for Level I and II of qualification, which are common to all types of instruments.
  • 4. AIM AND SCOPE OF THE GUIDELINE This guideline describes the requirements for balances (electronic - digital) used in chemical and biological tests in OMCLs. The following types of balances are considered in this guideline Type Ordinary name No.of digits after decimal position(g) Accuracy class 1. Ultra micro balance 7 I 2. Micro balance 6 I 3. Semi-micro balance 5 I 4. Analytical balance 4 I 5. Precision balance 1 TO 3 II 6. Technical balance 0 TO 1 III
  • 5. CONSIDERATIONS FOR LEVEL I AND II OF EQUIPMENT: QUALIFICATION At Level I of the qualification of balances (selection of instruments and suppliers), it is recommended to select a manufacturer of balances that can satisfy the needs of the laboratory and works under ISO 9001 certification. At Level II of the qualification of balances (installation and release for use), it is recommended to check all the requirements set during the selection of the instrument and calibration should be performed before putting into service by an accredited external service supplier, or internally by appropriately qualified personnel, using certified reference weights according to an approved procedure.
  • 6. 4. REQUIREMENTS FOR BALANCE USE Location of the Balance The accuracy and precision of weighing results are closely associated with the location of the balance. OMCLs should ensure that the balance can work under optimal conditions (weighing room/laboratory, weighing bench, temperature, light, air, etc,.) • It should be ensured that the weighing bench is stable, no matter from what material it is made. The weighing bench should not deform when work is carried out on it and it must be vibration-proof or the transfer of vibrations must not influence the weighing process.
  • 7. Temperature As weighing results are influenced by temperature, OMCLs should ensure a constant temperature is maintained in weighing rooms/laboratories. The deviation should not exceed more than 5 °C per hour. Atmospheric humidity The optimum relative humidity (% RH) during a weighing process is between 40 % to 60 % 80 % in cases where the accuracy and linearity of measurements are not affected. Light Balances should be protected from direct sunlight (heat). Air OMCLs should not place balances in the airflow of air conditioners or devices with ventilators next to doorways or in areas of high traffic.
  • 8. Weighing vessel OMCLs must ensure that: - the smallest possible weighing vessels are used. - if the materials that the weighing vessel is made of have a high degree of electrical insulation they are not electrostatically charged.
  • 9. FREQUENCY OF QUALIFICATION OMCLs must ensure that qualification/calibration of balances is done on receipt (i.e. immediately after delivery) or prior to their first use and after any repair or move. The frequency of qualification/calibration depends on the extent of use of the balances and is at the discretion of individual OMCLs. Qualification/calibration must be performed in accordance with a pre-determined protocol in which acceptance criteria are defined • k = correlation coefficient • SD = standard deviation • RSD = relative standard deviation
  • 10. Parameter to be checked Frequency Typical tolerance limit Levelling Every day before weighing begins Acceptance limits of the balance Internal caliberation Every day before weighing begins Automatic acceptance limits of the balanc Verification Atleast once a week OMCLs shall define their own acceptance criteri. Accuracy Frequency to be defined by OMCL,typically once a year OMCLs shall define their own acceptance criteri. Linearity Frequency to be defined by OMCL,typically once a year OMCLs shall define their own acceptance criteri. (k= 1±0.0001) Precission Frequency to be defined by OMCL,typically once a year OMCLs shall define their own acceptance criteri. (SD=max 5*d) Eccentricity Frequency to be defined by OMCL,typically once a year OMCLs shall define their own acceptance criteri. (RSD=0.05%)
  • 11. Verification Verification of the balance is performed by placing a suitable weight in the centre of the weighing pan once and comparing the result with pre-defined acceptance criteria. The same weight should always be used in these verifications. The acceptance criteria shall be defined by each individual OMCL. Accuracy The accuracy of the balance is checked by weighing at least three different certified weights that cover the usual weighing range of the balance. It is recommended that the weights have approximately 5%, 50% and 100% of the maximum capacity of the balance depending on the type of balance. It is recommended that the weighing is repeated at least 5 times for every weight.
  • 12. Linearity The results obtained from a series of accuracy checks can be used to calculate the correlation coefficient and to check for linearity. The correlation coefficient is calculated by comparing the nominal and measured masses of the weights. The acceptance criteria for the correlation coefficient shall be defined by each individual OMCL. Precision The precision of the balance should be verified by weighing at least 5 times a weight that is equivalent to approximately 50% of the maximum capacity of the balance. It is recommended to repeat the test with a weight that is equivalent to approximately 5% of the maximum capacity of the balance, if the balance is used at the lower range (Proposed criterion: SD = max 5*d, where d = (actual) scale interval (e.g. d=0.1 mg).
  • 13. CALIBRATION OF UV-VISIBLE SPECTROPHOTOMETER Tests Performed In UV-Visible Spectroscopy:  Baseline flatness test  UV region wave length accuracy test  Visible region wave length accuracy test  %Transmittance of Cuvette  Control of absorbance  Emission lines from deuterium discharge lamp  Stray Light measurement  Resolution power  Photometric linearity
  • 14. Make, Soft ware & Version Model type & Serial No Beam type (single/d ouble) INS/ EQP.ID No. SOP No Calibratio n Frequenc y Every six months Instrument/Equipment Calibration Record INS/ EQP Name: UV-VISIBLE spectrophotometer Department Name: Central Instrumentation Lab Equipment Details:
  • 15. 1) Baseline Flatness Test: Measurement configuration parameters: Reference : No Cuvette Sample : No Cuvette Cuvette material : Quartz Wave length range : 200-800 Lamp : Deuterium lamp Detector : PDA detector
  • 16. WAVE LENGTH MEASURED ABSORBANCE ACCEPTANCE CRITERIA 200 ±0.001 300 ±0.001 400 ±0.001 500 ±0.001 600 ±0.001 700 ±0.001 800 ±0.001
  • 17. 2) Uv Region Wave Length Accuracy Test: Measurement configuration Parameters Reference : Air (Empty Cuvette) Sample : Benzene Recording range : 220-270nm Lamp : Deuterium Discharge Detector : PDA Detector WAVE LENGTH (nm) OBSERVED PEAKS AT WAVE LENGTH (nm) ACCEPTANCE CRITERIA 220-270 nm 235.9±0.5nm
  • 18. 3) Visible Region Wave Length Accuracy Test: Measurement Parameters: Reference : Distilled Water Sample : 0.01% KMno4 Measuring mode : 300 to 600nm Lamp : Deuterium Discharge Detector : PDA Detector WAVE LENGTH (nm) OBSERVED PEAKS AT WAVE LENGTH(nm) ACCEPTANCE CRITERIA 300-600nm 310±1nm 525±1nm
  • 19. 4) %Transmittance Of Cuvette: Measurement configuration Parameters Reference : Empty Cuvette Sample : Empty Cuvette Cuvette material : Quartz Measuring mode : 240nm Lamp : Deuterium discharge lamp Detector : PDADetector WAVE LENGTH(nm) %TRANSMITTANCE ACCEPTANCE CRITERIA 240nm Not less then 80%
  • 20. 5) Control of absorbance: Absorbance:Visible region  Place dummy Cuvette in sample holder and set %T to “zero”. Now remove dummy Cuvette, by using fine & coarse control set a reading exactly 40.0 on the read out. Press Absorbance push button. If the maximum absorbance obtained at λ of 485nm is 0.398 ± 0.002, the photometric calibration of instrument is confirmed to be proper.  To confirm, repeat above steps, and set 10.00 on read out  Press Absorbance button. If the λ at 485 nm is 1.000±0.002 then it is confirmed the photometric performance in the visible region is proper. Absorbance:U.V region Place blank0.1N H2SO4 Cuvette and 60ppm K2Cr2O7 as sample Set λ exactly to 257 nm, if the value of Absorbance of sample at the set λ is 0.864±0.005, the instrument is measuringAbsorbance properly.
  • 21. 6) Emission Lines From Deuterium Discharge Lamp: Measurement configuration Parameters: Reference : No Cuvette Sample : Empty Cuvette Recording range : 400-500, 600-700 THEORETICAL VALUE OBSERVED VALUE DIFFERENCE ACCEPTANCE CRITERIA 486nm 486±0.3nm 656nm 656±0.3nm
  • 22. 7) Stray Light measurement:  Weigh accurately 1.2g of dried Potassium chloride in 100 ml volumetric flask and makeup to mark with Double distilled water.  Measure the absorbance at 200 nm. Acceptance criteria: Tolerance limit NLT 2.0 8)Resolution Power:  Prepare 0.02%v/v solution of Toluene and make up with Hexane.  Scan the wavelength from 250 to 280nm.  Maximum absorbance is 269 nm and Minimum absorbance is 266nm  Acceptance criteria: Ratio limit NLT 1.5
  • 23. 9) Photometric linearity:  Weigh accurately 100mg of Potassium chromate in 100ml volumetric flask and dissolve in 0.05N Potassium hydroxide solution. Make up with the same solvent.  From the above solution take 20ml and make up to 500ml with 0.05N Potassium hydroxide solution.  Now prepare dilution of 4,8,16,24,32 µg/ml.  Measure the absorbance at 370nm using blank. Acceptance criteria: The plot should be linear and regression coefficient (R2) should NLT 0.999.