・High durability PPS (polyphenylene sulfide) body
・Adopts a newly developed membrane against hydrofluoric acid
・The replaceable tip (glass <--> liquid junction) improves maintenance cost.
・Lowered running costs
Tip/glass combination electrodes
Hydrofluoric
acid proof
High-alkali
proof
Extensive line up of industrial pH electrodes enabling 50% reduction maintainance load.
Almost
(compared with
other HORIBA products)
double 2
1
Maintenance
cost and load
(compared with
other HORIBA products)
Longer life
pH electrodes/ORP electrodes
for industrial use
Tough enough to withstand brine electrolysis
and highly alkaline waste fluid.
Tough enough to withstand semiconductor processing
and waste fluid from hydrofluoric acid and other strong acids.
Standardized
Glass tip
Liquid junction tip
Built in temperature sensor liquid grounded
190
166.5±5
18
87±3
14.5
127
12
19
23
21.5
2
Notes on electrode selection
If you are using one of the samples below, please consult our sales department.
● A strong oxidizing solution such as aqua regia, chromic acid, hypochlorous acid, and perchloric acid.
● When containing corrosive gas such as ammonia, chlorine, and hydrogen sulfide.
pH electrode
Model
Liquid
junction
Multi
zirconia
ceramic
–10 to 100°C 0 to 0.6MPa 1kΩ
Useable
temperature range
Usable
pressure range
Temperature
compensation Traits
Rugged construction and easy-to-clean dome shape.
Double ceramic liquid junction helps prevent soiling.
Suitable for high-temperature (100°C) and
high-pressure (0.6 MPa) processes.
Good acid resistance.
Can be combined with jet, brush and ultrasonic
cleaning units.
Dome type pH electrode
6108-50B (Integrated combination)
Fixed
sleeve
–10 to 80°C
0 to 0.03MPa
(内部液加圧にて)
1kΩ
New response membrane provides even greater
toughness.
Uses easy-to-attach fixed sleeve liquid junction.
Sleeve type pH electrode
6109-50B (Integrated combination)
Porous
ceramic
–10 to 60°C 0 to 0.03MPa 1kΩ
It is an pH sensor using lead-free glass that
is environmentally friendly.
Ion trap is adopted as the reference electrode,
making clogging by silver ion less likely to occur.
pH electrode
6110-50B (Integrated combination)
Multi
alimina
ceramic
–10 to 60°C 0 to 0.2MPa 1kΩ
Uses newly developed response membrane for
alkali-resistant applications
Highly chemical resistant PSU (polysufone) body.
Provides service for one month in 1000 ppm
hydrofluoric acid at 25°C with pH 3-4 (general guide)
Double ceramic liquid junction.
Multi
zirconia
ceramic
–10 to 60°C 0 to 0.2MPa 1kΩ
Uses newly developed response membrane for
alkali-resistant applications.
Highly chemical resistant PSU (polysufone) body.
Provides service for three months in 0.1N caustic soda
at 60°C (general guide)
Double ceramic liquid junction.
Reduces alkali errors.
Porous
ceramic
&
Porous
Polyethylene
–10 to 100°C 0 to 0.7MPa
Pt
1000 RTD
1000Ω
(at 0°C)
Measurement of low conductivity samples is possible
from 10 μS/cm.
Internal solution of reference electrode is
water-insoluble polymer gel that is strong against
heat, pressure, acid and alkali.
Both cable and connector type are available.
Alimina
ceramic
–10 to 60°C 0 to 0.03MPa 1kΩ
Uses newly developed response membrane for
alkali-resistant applications
Highly chemical resistant PSU (polysufone) body.
Provides service for one month in 1000 ppm
hydrofluoric acid at 25°C with pH 3-4 (general guide)
Changeable glass tip and liquid junction tip improve
ease of maintenance and reduces running costs.
Zirconia
ceramic
–10 to 60°C 0 to 0.03MPa 1kΩ
Uses newly developed response membrane for
alkali-resistant applications.
Highly chemical resistant PSU (polysufone) body.
Provides service for three months in 0.1N caustic soda
at 60°C (general guide)
Changeable glass tip and liquid junction tip improves
ease of maintenance and reduces running costs.
Zirconia
ceramic
–10 to 100°C 0 to 0.03MPa 1kΩ
It is made of PPS which has heat resistance and
chemical resistance comparable to fluororesin, and
there is no worry of breakage etc. In addition, pH tip
and liquid junction tip are easy to replace cartridge
type, excellent in maintainability.
pH electrode
6174-50B (Tip replaceable)
Porous
plastic
0 to 50°C
Atmospheric
release
–
It is a pH sensor dedicated to pH meter "HP-8A" for
pure water. PPS is used for the body, and high
durability is realized. Also, since the pH tip is a
connector type, it can be replaced with a single touch.
Cable
length
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
5m
(standard)
0.3m
(standard)
Hydrofluoric acid-resistant pH electrode
6151-50B (Integrated combination)
High alkali-resistant pH electrode
6152-50B (Integrated combination)
Gel-filled electrode
6155 series (Cable type/Connector type)
High alkali-resistant pH electrode
6171-50B (Tip replaceable)
High alkali-resistant pH electrode
6172-50B (Tip replaceable)
EL sensor for pure water
EL-6121-60B
For HP-8A
Dome-shaped pH electrode
Fixed sleeve electrode
pH electrode
Hydrofluoric-acid proof pH electrode
High alkali proof pH electrode
Gel-filled electrode
Hydrofluoric-acid proof pH electrode
High alkali proof pH electrode
Replaceable tip electrode
EL sensor for pure water
6108-50B (Integrated combination)
6109-50B (Integrated combination)
6110-50B (Integrated combination)
6151-50B (Integrated combination)
6152-50B (Integrated combination)
6155 series (Cable type/Connector type)
6171-50B (Combination with replaceable tip)
6172-50B (Combination with replaceable tip)
6174-50B (Combination with replaceable tip)
EL-6121-60B
ORP platinum electrode
ORP platinum + Gold plate electrode
Platinum plated sensor
Gel-filled electrode
6805-50B
6815-50B
6870-60B (Tip replaceable)
6855 series (Cable type/Connector type)
◉Recommended combination −Unable to connect
pH electrode/ORP electrode
Electrode
combination table
−
−
−
−
−
−
−
−
−
−
−
−
HP-8 (for pure water)
◉
◉
◉
◉
◉
Panel mount type 48/96 series
◉
◉
◉
◉
◉
◉
◉
◉
◉
◉
◉
−
−
−
◉
Field type H-1 series
◉
◉
◉
◉
◉
◉
◉
◉
◉
◉
190
127
23
21.5
OPR electrode
Movable sleeve
Makes cleaning of liquid junction easy.
ToupH Electrode
Increases strength and sensitivity significantly.
Silver Ion Trap
Controls outflow of silver ion significantly.
Lead-free glass
Lead-free glass membrane and body.
Model
Liquid
junction
Fixed
sleeve
0 to 80°C 0 to 0.03MPa –
Useable
temperature range
Usable
pressure range
Temperature
compensation Traits
It is an ORP sensor using lead-free glass that is
environmentally friendly. Can be combined with jet,
brush and ultrasonic cleaning units. Ion trap is
adopted as the reference electrode, making clogging
by silver ion less likely to occur.
Platinum plated sensor
6805-50B
Fixed
sleeve
0 to 80°C 0 to 0.03MPa –
Added gold plating to 6805 platinum electrodes. With
this addition, it may be effective for samples that
could not be measured stably previously.
Gold plated sensor
6815-50B
Porous
ceramic
&
Porous
Polyethylene
–10 to 100°C 0 to 0.7MPa –
Internal solution of reference electrode is
water-insoluble polymer gel that is strong against
heat, pressure, acid and alkali.
Both cable and connector type are available.
Gel-filled electrode
6855 series (Cable type/Connector type)
Cable
length
5m
(standard)
5m
(standard)
5m
(standard)
Ceramic –10 to 105°C 0 to 0.03MPa 6.8kΩ
The ORP electrode, reference electrode and
temperature compensation electrode are integrated. It
is made of PPS which has heat resistance and
chemical resistance comparable to fluororesin, and
there is no worry of breakage etc. In addition, ORP tip
and liquid junction chip are easy to replace cartridge
type, excellent in maintainability.
Platinum plated sensor
6870-60B (Tip replaceable)
6m
(standard)
Combination
table
of
pH
meter
and
electrode
Combination
table
of
ORP
meter
and
electrode
For HP-8A
1994
・
1997
D-20, F-20 (Laboratory use)
1988
・
1989
D-10, F-10 (Laboratory use)
1951
The first domestic glass electrode
pH meter “H type” birth
2003
D-50, F-50 (Laboratory use)
2003
Panel mount type 48
・
96 series (Industrial use)
2010
H-1 series
(Industrial use)
2018
6155/6855 series
(Industrial use)
1982
K-8 (Industrial use)
1992
K-10 (Industrial use)
1970
K-100 (Industrial use)
1957
K-5 (Industrial use)
2011
DS-70, F-70 (Laboratory use)
HORIBA began with a pH meter
and is now the worldwide standard.
01
18
Printed in Japan 1910SK00
Please read the operation manual before using this product to assure safe and proper handling of the product.
OThe specifications, appearance or other aspects of products in this catalog are subject to change without notice.
OPlease contact us with enquiries concerning further details on the products in this catalog.
OThe color of the actual products may differ from the color pictured in this catalog due to printing limitations.
OIt is strictly forbidden to copy the content of this catalog in part or in full.
OThe screen displays shown on products in this catalog have been inserted into the photographs through compositing.
OAll brand names, product names and service names in this catalog are trademarks or registered trademarks of their respective companies.
Jl. Jalur Sutera Blok 20A, No.16-17, Kel. Kunciran, Kec. Pinang
Tangerang-15144, Indonesia
Phone: 62 (21) 3044-8525 Fax: 62 (21) 3044-8521
Unit 6, 10 Floor, CMC Tower, Duy Tan Street, Dich Vong
Hau Ward, Cau Giay District, Hanoi, Vietnam
Phone: 84 (24) 3795-8552 Fax: 84 (24) 3795-8553
3 Changi Business Park Vista #01-01 Akzonobel House,
Singapore 486051
Phone: 65 (6) 745-8300 Fax: 65 (6) 745-8155
East Office
850 / 7 Soi Lat Krabang 30 / 5, Lat Krabang Road, Lat Krabang,
Bangkok 10520, Thailand
Phone: 66 (0) 2734 4434 Fax: 66 (0) 2734 4438
Unit D, 1F, Building A, Synnex International Park, 1068 West
Tianshan Road, Shanghai, 200335, China
Phone: 86 (21) 6289-6060 Fax: 86 (21) 6289-5553
Beijing Office
12F, Metropolis Tower, No.2, Haidian Dong 3 Street, Beijing,
100080, China
Phone: 86 (10) 8567-9966 Fax: 86 (10) 8567-9066
HORIBA (China) Trading Co., Ltd.
HORIBA Instruments (Singapore) Pte Ltd.
PT HORIBA Indonesia
HORIBA (Thailand) Limited
China
Singapore
HORIBA Vietnam Co., Ltd. Vietnam
Indonesia
Thailand
9755 Research Drive, Irvine, CA 92618, U.S.A.
Phone: 1 (949) 250-4811 Fax: 1 (949) 250-0924
Houston Office
5390 Bay Oaks Drive, Pasadena, TX 77505
Phone: 1 (281) 482-4334 Fax: 1 (281) 674-6058
246, Okhla Industrial Estate, Phase 3 New Delhi-110020, India
Phone: 91 (11) 4646-5000 Fax: 91 (11) 4646-5020
Technical Center
D-255, Chakan MIDC Phase-II, Bhamboli Village, Pune-410501, India
Phone: 91 (21) 3567-6000
Bangalore Office
No.55, 12th Main, Behind BDA Complex, 6th sector, HSR Layout,
Bangalore South, Bangalore-560102, India
Phone: 91 (80) 4127-3637
HORIBA India Private Limited
HORIBA Instruments Incorporated
India
USA
Kaplanstrasse 5 A-3430 Tulln, Austria
Phone: 43 (2272) 65225 Fax: 43 (2272) 65230
Prague Office
Prumyslova 1306 / 7, CZ-10200, Praha 10, Czech Republic
Phone: 420 (2) 460-392-65
Hans-Mess-Str.6 D-61440 Oberursel Germany
Phone: 49 (6172) 1396-0 Fax: 49 (6172) 1373-85
Leichlingen Office
Julius-kronenberg Str.9 D-42799 Leichlingen Germany
Phone: 49 (2175) 8978-0 Fax: 49 (2175) 8978-50
Kyoto Close Moulton Park, Northampton NN3 6FL, UK
Phone: 44 (1604) 542-500 Fax: 44 (1604) 542-699
HORIBA UK Limited
HORIBA Europe GmbH
HORIBA Czech
UK
Germany
Czech
Austria
HORIBA (Austria) GmbH
Head Office
2 Miyanohigashi-cho, Kisshoin Minami-ku, Kyoto, Japan
Phone: 81 (75) 313-8123 Fax: 81 (75) 321-5725
http://www.horiba.com e-mail: info@horiba.co.jp
Manufactured by
25, 94-Gil, Iljik-Ro, Manan-Gu, Anyang-Si, Gyeonggi-Do,
13901, Korea
Phone: 82 (31) 296-7911 Fax: 82 (31) 296-7913
HORIBA KOREA Ltd. Korea
Avenue de la Vauve - Passage Jobin Yvon CS 45002 - 91120
Palaiseau - France
Phone: 33 (1) 69-74-72-00 Fax: 33 (1) 69-31-32-20
HORIBA Europe Research Center France
Romania
HORIBA (Austria) GmbH
Romania Branch
B-dul.Republicii, nr. 164, Etaj Parter, Birourile nr. 3 si 4
Pitestijudetul Arges110177 Romania
Phone: 40 (348) 807117 Fax: 40 (348) 807118
Rua Presbitero Plinio Alves de Souza, 645, Parte A, Loteamento
Multivias, Jardin Ermida II - Jundiai Sao Paulo - CEP
13.212-181 Brazil
Phone: 55 (11) 2923-5400 Fax: 55 (11) 2923-5490
HORIBA Instruments Brasil, Ltda. Brasil
The HORIBA Group adopts IMS (Integrated Management System) which integrates Quality Management System ISO9001, Environmental Management System ISO14001,
and Occupational Health and Safety Management System OHSAS18001.
We have now integrated Business Continuity Management System ISO22301 in order to provide our products and services in a stable manner, even in emergencies.
Bulletin:HAE-T0247B

More Related Content

Similar to HORIBA_pH_6151_more pH/ORP electrodes for industrial use.pdf

Nanotech Scope of Technology
Nanotech Scope of TechnologyNanotech Scope of Technology
Nanotech Scope of Technology
Jakes Jacobs
 
Fina lindustrial tour report
Fina lindustrial tour reportFina lindustrial tour report
Fina lindustrial tour report
manish katara
 

Similar to HORIBA_pH_6151_more pH/ORP electrodes for industrial use.pdf (20)

Catalogue OBLl water division 2017
Catalogue OBLl  water division 2017Catalogue OBLl  water division 2017
Catalogue OBLl water division 2017
 
Ferroli electric water heater all compressed.pdf
Ferroli electric water heater all compressed.pdfFerroli electric water heater all compressed.pdf
Ferroli electric water heater all compressed.pdf
 
GE LED Tubes - Product presentation
GE LED Tubes - Product presentationGE LED Tubes - Product presentation
GE LED Tubes - Product presentation
 
Changchun CCP PBT Overview Presentation
Changchun CCP PBT Overview PresentationChangchun CCP PBT Overview Presentation
Changchun CCP PBT Overview Presentation
 
Nanotech Scope of Technology
Nanotech Scope of TechnologyNanotech Scope of Technology
Nanotech Scope of Technology
 
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdfATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
 
Trustin Pipe Brochure designed by Balikpapandesign.com
Trustin Pipe Brochure designed by Balikpapandesign.comTrustin Pipe Brochure designed by Balikpapandesign.com
Trustin Pipe Brochure designed by Balikpapandesign.com
 
Box DC Link in industrial and automotive applications
Box DC Link in industrial and automotive applicationsBox DC Link in industrial and automotive applications
Box DC Link in industrial and automotive applications
 
Domestic purification
Domestic purificationDomestic purification
Domestic purification
 
Domestic purification pdf document aqua middle east fzc
Domestic purification pdf document aqua middle east fzcDomestic purification pdf document aqua middle east fzc
Domestic purification pdf document aqua middle east fzc
 
Iron_Removal_Filter
Iron_Removal_FilterIron_Removal_Filter
Iron_Removal_Filter
 
Webinar 85 85-box_dc_link
Webinar 85 85-box_dc_linkWebinar 85 85-box_dc_link
Webinar 85 85-box_dc_link
 
Electronics - Selector guide
Electronics - Selector guideElectronics - Selector guide
Electronics - Selector guide
 
Fina lindustrial tour report
Fina lindustrial tour reportFina lindustrial tour report
Fina lindustrial tour report
 
HORIBA_pH_6155_6855 Industrial pH/ORP electrode with Gel-filled.pdf
HORIBA_pH_6155_6855 Industrial pH/ORP electrode with Gel-filled.pdfHORIBA_pH_6155_6855 Industrial pH/ORP electrode with Gel-filled.pdf
HORIBA_pH_6155_6855 Industrial pH/ORP electrode with Gel-filled.pdf
 
1 Potentiometry
1  Potentiometry1  Potentiometry
1 Potentiometry
 
CORTECS Columns Infographic
CORTECS Columns InfographicCORTECS Columns Infographic
CORTECS Columns Infographic
 
Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)Steam &amp; water analysis system (swas)
Steam &amp; water analysis system (swas)
 
Summer Trg
Summer TrgSummer Trg
Summer Trg
 
Samsung Office Lighting LED Solutions
Samsung Office Lighting LED SolutionsSamsung Office Lighting LED Solutions
Samsung Office Lighting LED Solutions
 

More from ENVIMART

More from ENVIMART (20)

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdf
 

Recently uploaded

LaPlace Transforms 2 with use of Matlab.pptx
LaPlace Transforms 2 with use of Matlab.pptxLaPlace Transforms 2 with use of Matlab.pptx
LaPlace Transforms 2 with use of Matlab.pptx
joshuaclack73
 

Recently uploaded (20)

NO1 Pakistan online istikhara for love marriage vashikaran specialist love pr...
NO1 Pakistan online istikhara for love marriage vashikaran specialist love pr...NO1 Pakistan online istikhara for love marriage vashikaran specialist love pr...
NO1 Pakistan online istikhara for love marriage vashikaran specialist love pr...
 
LaPlace Transforms 2 with use of Matlab.pptx
LaPlace Transforms 2 with use of Matlab.pptxLaPlace Transforms 2 with use of Matlab.pptx
LaPlace Transforms 2 with use of Matlab.pptx
 
Carbon Stock Assessment in Banten Province and Demak, Central Java, Indonesia
Carbon Stock Assessment in Banten Province and Demak, Central Java, IndonesiaCarbon Stock Assessment in Banten Province and Demak, Central Java, Indonesia
Carbon Stock Assessment in Banten Province and Demak, Central Java, Indonesia
 
Production, dispersal, sedimentation and taphonomy of spores/pollen
Production, dispersal, sedimentation and taphonomy of spores/pollenProduction, dispersal, sedimentation and taphonomy of spores/pollen
Production, dispersal, sedimentation and taphonomy of spores/pollen
 
EFFLUENT TREATMENT PLANT For PHARMA INDUSTRY
EFFLUENT TREATMENT PLANT For PHARMA INDUSTRYEFFLUENT TREATMENT PLANT For PHARMA INDUSTRY
EFFLUENT TREATMENT PLANT For PHARMA INDUSTRY
 
slidesgo-maximizing-sustainability-the-case-for-plastic-reuse
slidesgo-maximizing-sustainability-the-case-for-plastic-reuseslidesgo-maximizing-sustainability-the-case-for-plastic-reuse
slidesgo-maximizing-sustainability-the-case-for-plastic-reuse
 
National Tree Planting Day May 2024 Nairobi
National Tree Planting Day May 2024 NairobiNational Tree Planting Day May 2024 Nairobi
National Tree Planting Day May 2024 Nairobi
 
Elemental Analysis of Plants using ICP-OES(2023)
Elemental Analysis of Plants using ICP-OES(2023)Elemental Analysis of Plants using ICP-OES(2023)
Elemental Analysis of Plants using ICP-OES(2023)
 
A Review on Integrated River Basin Management and Development Master Plan of ...
A Review on Integrated River Basin Management and Development Master Plan of ...A Review on Integrated River Basin Management and Development Master Plan of ...
A Review on Integrated River Basin Management and Development Master Plan of ...
 
Coastal and mangrove vulnerability assessment In the Northern Coast of Java, ...
Coastal and mangrove vulnerability assessment In the Northern Coast of Java, ...Coastal and mangrove vulnerability assessment In the Northern Coast of Java, ...
Coastal and mangrove vulnerability assessment In the Northern Coast of Java, ...
 
Impacts of agriculture on the environment.
Impacts of agriculture on the environment.Impacts of agriculture on the environment.
Impacts of agriculture on the environment.
 
Lab Investigation.pptxjjjjjjjjjjjjjjjjjj
Lab Investigation.pptxjjjjjjjjjjjjjjjjjjLab Investigation.pptxjjjjjjjjjjjjjjjjjj
Lab Investigation.pptxjjjjjjjjjjjjjjjjjj
 
BIG DEVELOPMENT SHAPING' LESOTHO FUTURE.
BIG DEVELOPMENT SHAPING' LESOTHO FUTURE.BIG DEVELOPMENT SHAPING' LESOTHO FUTURE.
BIG DEVELOPMENT SHAPING' LESOTHO FUTURE.
 
NO1 Pakistan Black magic/kala jadu,manpasand shadi in lahore,karachi rawalpin...
NO1 Pakistan Black magic/kala jadu,manpasand shadi in lahore,karachi rawalpin...NO1 Pakistan Black magic/kala jadu,manpasand shadi in lahore,karachi rawalpin...
NO1 Pakistan Black magic/kala jadu,manpasand shadi in lahore,karachi rawalpin...
 
Up to 40% of food crops are lost to plant pests and diseases annually.
Up to 40% of food crops are lost to plant pests and diseases annually.Up to 40% of food crops are lost to plant pests and diseases annually.
Up to 40% of food crops are lost to plant pests and diseases annually.
 
Rising temperatures also mean that more plant pests are appearing earlier and...
Rising temperatures also mean that more plant pests are appearing earlier and...Rising temperatures also mean that more plant pests are appearing earlier and...
Rising temperatures also mean that more plant pests are appearing earlier and...
 
TEST BANK For Geosystems An Introduction to Physical Geography, 5th Canadian ...
TEST BANK For Geosystems An Introduction to Physical Geography, 5th Canadian ...TEST BANK For Geosystems An Introduction to Physical Geography, 5th Canadian ...
TEST BANK For Geosystems An Introduction to Physical Geography, 5th Canadian ...
 
NO1 Best Amil Baba In Pakistan Authentic Amil In pakistan Best Amil In Pakist...
NO1 Best Amil Baba In Pakistan Authentic Amil In pakistan Best Amil In Pakist...NO1 Best Amil Baba In Pakistan Authentic Amil In pakistan Best Amil In Pakist...
NO1 Best Amil Baba In Pakistan Authentic Amil In pakistan Best Amil In Pakist...
 
Global warming, Types, Causes and Effects.
Global warming, Types, Causes and Effects.Global warming, Types, Causes and Effects.
Global warming, Types, Causes and Effects.
 
Cooperative Mangrove Project: Introduction, Scope, and Perspectives
Cooperative Mangrove Project: Introduction, Scope, and PerspectivesCooperative Mangrove Project: Introduction, Scope, and Perspectives
Cooperative Mangrove Project: Introduction, Scope, and Perspectives
 

HORIBA_pH_6151_more pH/ORP electrodes for industrial use.pdf

  • 1. ・High durability PPS (polyphenylene sulfide) body ・Adopts a newly developed membrane against hydrofluoric acid ・The replaceable tip (glass <--> liquid junction) improves maintenance cost. ・Lowered running costs Tip/glass combination electrodes Hydrofluoric acid proof High-alkali proof Extensive line up of industrial pH electrodes enabling 50% reduction maintainance load. Almost (compared with other HORIBA products) double 2 1 Maintenance cost and load (compared with other HORIBA products) Longer life pH electrodes/ORP electrodes for industrial use Tough enough to withstand brine electrolysis and highly alkaline waste fluid. Tough enough to withstand semiconductor processing and waste fluid from hydrofluoric acid and other strong acids. Standardized Glass tip Liquid junction tip Built in temperature sensor liquid grounded
  • 2. 190 166.5±5 18 87±3 14.5 127 12 19 23 21.5 2 Notes on electrode selection If you are using one of the samples below, please consult our sales department. ● A strong oxidizing solution such as aqua regia, chromic acid, hypochlorous acid, and perchloric acid. ● When containing corrosive gas such as ammonia, chlorine, and hydrogen sulfide. pH electrode Model Liquid junction Multi zirconia ceramic –10 to 100°C 0 to 0.6MPa 1kΩ Useable temperature range Usable pressure range Temperature compensation Traits Rugged construction and easy-to-clean dome shape. Double ceramic liquid junction helps prevent soiling. Suitable for high-temperature (100°C) and high-pressure (0.6 MPa) processes. Good acid resistance. Can be combined with jet, brush and ultrasonic cleaning units. Dome type pH electrode 6108-50B (Integrated combination) Fixed sleeve –10 to 80°C 0 to 0.03MPa (内部液加圧にて) 1kΩ New response membrane provides even greater toughness. Uses easy-to-attach fixed sleeve liquid junction. Sleeve type pH electrode 6109-50B (Integrated combination) Porous ceramic –10 to 60°C 0 to 0.03MPa 1kΩ It is an pH sensor using lead-free glass that is environmentally friendly. Ion trap is adopted as the reference electrode, making clogging by silver ion less likely to occur. pH electrode 6110-50B (Integrated combination) Multi alimina ceramic –10 to 60°C 0 to 0.2MPa 1kΩ Uses newly developed response membrane for alkali-resistant applications Highly chemical resistant PSU (polysufone) body. Provides service for one month in 1000 ppm hydrofluoric acid at 25°C with pH 3-4 (general guide) Double ceramic liquid junction. Multi zirconia ceramic –10 to 60°C 0 to 0.2MPa 1kΩ Uses newly developed response membrane for alkali-resistant applications. Highly chemical resistant PSU (polysufone) body. Provides service for three months in 0.1N caustic soda at 60°C (general guide) Double ceramic liquid junction. Reduces alkali errors. Porous ceramic & Porous Polyethylene –10 to 100°C 0 to 0.7MPa Pt 1000 RTD 1000Ω (at 0°C) Measurement of low conductivity samples is possible from 10 μS/cm. Internal solution of reference electrode is water-insoluble polymer gel that is strong against heat, pressure, acid and alkali. Both cable and connector type are available. Alimina ceramic –10 to 60°C 0 to 0.03MPa 1kΩ Uses newly developed response membrane for alkali-resistant applications Highly chemical resistant PSU (polysufone) body. Provides service for one month in 1000 ppm hydrofluoric acid at 25°C with pH 3-4 (general guide) Changeable glass tip and liquid junction tip improve ease of maintenance and reduces running costs. Zirconia ceramic –10 to 60°C 0 to 0.03MPa 1kΩ Uses newly developed response membrane for alkali-resistant applications. Highly chemical resistant PSU (polysufone) body. Provides service for three months in 0.1N caustic soda at 60°C (general guide) Changeable glass tip and liquid junction tip improves ease of maintenance and reduces running costs. Zirconia ceramic –10 to 100°C 0 to 0.03MPa 1kΩ It is made of PPS which has heat resistance and chemical resistance comparable to fluororesin, and there is no worry of breakage etc. In addition, pH tip and liquid junction tip are easy to replace cartridge type, excellent in maintainability. pH electrode 6174-50B (Tip replaceable) Porous plastic 0 to 50°C Atmospheric release – It is a pH sensor dedicated to pH meter "HP-8A" for pure water. PPS is used for the body, and high durability is realized. Also, since the pH tip is a connector type, it can be replaced with a single touch. Cable length 5m (standard) 5m (standard) 5m (standard) 5m (standard) 5m (standard) 5m (standard) 5m (standard) 5m (standard) 5m (standard) 0.3m (standard) Hydrofluoric acid-resistant pH electrode 6151-50B (Integrated combination) High alkali-resistant pH electrode 6152-50B (Integrated combination) Gel-filled electrode 6155 series (Cable type/Connector type) High alkali-resistant pH electrode 6171-50B (Tip replaceable) High alkali-resistant pH electrode 6172-50B (Tip replaceable) EL sensor for pure water EL-6121-60B For HP-8A
  • 3. Dome-shaped pH electrode Fixed sleeve electrode pH electrode Hydrofluoric-acid proof pH electrode High alkali proof pH electrode Gel-filled electrode Hydrofluoric-acid proof pH electrode High alkali proof pH electrode Replaceable tip electrode EL sensor for pure water 6108-50B (Integrated combination) 6109-50B (Integrated combination) 6110-50B (Integrated combination) 6151-50B (Integrated combination) 6152-50B (Integrated combination) 6155 series (Cable type/Connector type) 6171-50B (Combination with replaceable tip) 6172-50B (Combination with replaceable tip) 6174-50B (Combination with replaceable tip) EL-6121-60B ORP platinum electrode ORP platinum + Gold plate electrode Platinum plated sensor Gel-filled electrode 6805-50B 6815-50B 6870-60B (Tip replaceable) 6855 series (Cable type/Connector type) ◉Recommended combination −Unable to connect pH electrode/ORP electrode Electrode combination table − − − − − − − − − − − − HP-8 (for pure water) ◉ ◉ ◉ ◉ ◉ Panel mount type 48/96 series ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ − − − ◉ Field type H-1 series ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ ◉ 190 127 23 21.5 OPR electrode Movable sleeve Makes cleaning of liquid junction easy. ToupH Electrode Increases strength and sensitivity significantly. Silver Ion Trap Controls outflow of silver ion significantly. Lead-free glass Lead-free glass membrane and body. Model Liquid junction Fixed sleeve 0 to 80°C 0 to 0.03MPa – Useable temperature range Usable pressure range Temperature compensation Traits It is an ORP sensor using lead-free glass that is environmentally friendly. Can be combined with jet, brush and ultrasonic cleaning units. Ion trap is adopted as the reference electrode, making clogging by silver ion less likely to occur. Platinum plated sensor 6805-50B Fixed sleeve 0 to 80°C 0 to 0.03MPa – Added gold plating to 6805 platinum electrodes. With this addition, it may be effective for samples that could not be measured stably previously. Gold plated sensor 6815-50B Porous ceramic & Porous Polyethylene –10 to 100°C 0 to 0.7MPa – Internal solution of reference electrode is water-insoluble polymer gel that is strong against heat, pressure, acid and alkali. Both cable and connector type are available. Gel-filled electrode 6855 series (Cable type/Connector type) Cable length 5m (standard) 5m (standard) 5m (standard) Ceramic –10 to 105°C 0 to 0.03MPa 6.8kΩ The ORP electrode, reference electrode and temperature compensation electrode are integrated. It is made of PPS which has heat resistance and chemical resistance comparable to fluororesin, and there is no worry of breakage etc. In addition, ORP tip and liquid junction chip are easy to replace cartridge type, excellent in maintainability. Platinum plated sensor 6870-60B (Tip replaceable) 6m (standard) Combination table of pH meter and electrode Combination table of ORP meter and electrode For HP-8A
  • 4. 1994 ・ 1997 D-20, F-20 (Laboratory use) 1988 ・ 1989 D-10, F-10 (Laboratory use) 1951 The first domestic glass electrode pH meter “H type” birth 2003 D-50, F-50 (Laboratory use) 2003 Panel mount type 48 ・ 96 series (Industrial use) 2010 H-1 series (Industrial use) 2018 6155/6855 series (Industrial use) 1982 K-8 (Industrial use) 1992 K-10 (Industrial use) 1970 K-100 (Industrial use) 1957 K-5 (Industrial use) 2011 DS-70, F-70 (Laboratory use) HORIBA began with a pH meter and is now the worldwide standard. 01 18 Printed in Japan 1910SK00 Please read the operation manual before using this product to assure safe and proper handling of the product. OThe specifications, appearance or other aspects of products in this catalog are subject to change without notice. OPlease contact us with enquiries concerning further details on the products in this catalog. OThe color of the actual products may differ from the color pictured in this catalog due to printing limitations. OIt is strictly forbidden to copy the content of this catalog in part or in full. OThe screen displays shown on products in this catalog have been inserted into the photographs through compositing. OAll brand names, product names and service names in this catalog are trademarks or registered trademarks of their respective companies. Jl. Jalur Sutera Blok 20A, No.16-17, Kel. Kunciran, Kec. Pinang Tangerang-15144, Indonesia Phone: 62 (21) 3044-8525 Fax: 62 (21) 3044-8521 Unit 6, 10 Floor, CMC Tower, Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi, Vietnam Phone: 84 (24) 3795-8552 Fax: 84 (24) 3795-8553 3 Changi Business Park Vista #01-01 Akzonobel House, Singapore 486051 Phone: 65 (6) 745-8300 Fax: 65 (6) 745-8155 East Office 850 / 7 Soi Lat Krabang 30 / 5, Lat Krabang Road, Lat Krabang, Bangkok 10520, Thailand Phone: 66 (0) 2734 4434 Fax: 66 (0) 2734 4438 Unit D, 1F, Building A, Synnex International Park, 1068 West Tianshan Road, Shanghai, 200335, China Phone: 86 (21) 6289-6060 Fax: 86 (21) 6289-5553 Beijing Office 12F, Metropolis Tower, No.2, Haidian Dong 3 Street, Beijing, 100080, China Phone: 86 (10) 8567-9966 Fax: 86 (10) 8567-9066 HORIBA (China) Trading Co., Ltd. HORIBA Instruments (Singapore) Pte Ltd. PT HORIBA Indonesia HORIBA (Thailand) Limited China Singapore HORIBA Vietnam Co., Ltd. Vietnam Indonesia Thailand 9755 Research Drive, Irvine, CA 92618, U.S.A. Phone: 1 (949) 250-4811 Fax: 1 (949) 250-0924 Houston Office 5390 Bay Oaks Drive, Pasadena, TX 77505 Phone: 1 (281) 482-4334 Fax: 1 (281) 674-6058 246, Okhla Industrial Estate, Phase 3 New Delhi-110020, India Phone: 91 (11) 4646-5000 Fax: 91 (11) 4646-5020 Technical Center D-255, Chakan MIDC Phase-II, Bhamboli Village, Pune-410501, India Phone: 91 (21) 3567-6000 Bangalore Office No.55, 12th Main, Behind BDA Complex, 6th sector, HSR Layout, Bangalore South, Bangalore-560102, India Phone: 91 (80) 4127-3637 HORIBA India Private Limited HORIBA Instruments Incorporated India USA Kaplanstrasse 5 A-3430 Tulln, Austria Phone: 43 (2272) 65225 Fax: 43 (2272) 65230 Prague Office Prumyslova 1306 / 7, CZ-10200, Praha 10, Czech Republic Phone: 420 (2) 460-392-65 Hans-Mess-Str.6 D-61440 Oberursel Germany Phone: 49 (6172) 1396-0 Fax: 49 (6172) 1373-85 Leichlingen Office Julius-kronenberg Str.9 D-42799 Leichlingen Germany Phone: 49 (2175) 8978-0 Fax: 49 (2175) 8978-50 Kyoto Close Moulton Park, Northampton NN3 6FL, UK Phone: 44 (1604) 542-500 Fax: 44 (1604) 542-699 HORIBA UK Limited HORIBA Europe GmbH HORIBA Czech UK Germany Czech Austria HORIBA (Austria) GmbH Head Office 2 Miyanohigashi-cho, Kisshoin Minami-ku, Kyoto, Japan Phone: 81 (75) 313-8123 Fax: 81 (75) 321-5725 http://www.horiba.com e-mail: info@horiba.co.jp Manufactured by 25, 94-Gil, Iljik-Ro, Manan-Gu, Anyang-Si, Gyeonggi-Do, 13901, Korea Phone: 82 (31) 296-7911 Fax: 82 (31) 296-7913 HORIBA KOREA Ltd. Korea Avenue de la Vauve - Passage Jobin Yvon CS 45002 - 91120 Palaiseau - France Phone: 33 (1) 69-74-72-00 Fax: 33 (1) 69-31-32-20 HORIBA Europe Research Center France Romania HORIBA (Austria) GmbH Romania Branch B-dul.Republicii, nr. 164, Etaj Parter, Birourile nr. 3 si 4 Pitestijudetul Arges110177 Romania Phone: 40 (348) 807117 Fax: 40 (348) 807118 Rua Presbitero Plinio Alves de Souza, 645, Parte A, Loteamento Multivias, Jardin Ermida II - Jundiai Sao Paulo - CEP 13.212-181 Brazil Phone: 55 (11) 2923-5400 Fax: 55 (11) 2923-5490 HORIBA Instruments Brasil, Ltda. Brasil The HORIBA Group adopts IMS (Integrated Management System) which integrates Quality Management System ISO9001, Environmental Management System ISO14001, and Occupational Health and Safety Management System OHSAS18001. We have now integrated Business Continuity Management System ISO22301 in order to provide our products and services in a stable manner, even in emergencies. Bulletin:HAE-T0247B