Gas chromatography coupled to mass spectrometry is a
versatile tool to separate, quantify and identify unknown
(volatile) organic compounds and permanent gases. By
combining sensitivity and a high resolving power, complex
mixtures can be analyzed. The information obtained can be
used for detection of impurities, contamination control and
improvement of, for example, semiconductor manufacturing
processes
FUNDAMENTAL OF GC
Gas Chromatography (GC) uses a carrier gas to transport sample
components through either packed columns or hollow capillary columns
containing the stationary phase. In most cases, GC columns have smaller
internal diameter and are longer than HPLC columns.
In Gas Chromatography, the mobile phase is a gas and the stationary
phase is either:
• A solid, commonly termed “Gas solid chromatography (GSC)”.
• An immobilised polymeric liquid, referred as “Gas Liquid
Chromatography (GLC)”.
6
• Carrier – (H2, He, N2 – helium is most usual with MS detection)
• Make-up gas– (H2, He, N2 - if using an FID detector in parallel to the
MS detector)
• Detector gas – (H2 & Air – if using an FID detector in parallel to the
MS detector)
1. Gas Supply: Gas is fed from cylinders through regulators and
tubing to the instrument. It is usual to filter gases to ensure high gas
purity and the gas supply pressure. Required gases might include:
2. Interface: After separation in the GC system, analyte species have
to be transported to the mass spectrometer to be ionised, mass filtered
and detected. The interface in modern instruments is heated to
prevent analyte condensation and in some instruments will contain
a device to remove carrier gas molecules to allow analyte pre-
concentration.
5. Column: Gas Chromatography uses a gaseous mobile phase to
transport sample components through columns either packed with coated
silica particles or hollow capillary columns containing the stationary
phase coated onto the inner wall. Capillary GC columns
are usually several meters long (10 – 120m is typical) with an internal
diameter of 0.10 – 0.50mm, whilst packed GC columns tend be 1 – 5
meters in length with either 2 or 4mm internal diameter.
6. Injector: Here the sample is volatilised and the resulting gas
entrained into the carrier
3. Pneumatic controls: The gas supply is regulated to the correct pressure
(or flow) and then fed to the required part of the instrument. Modern GC
instruments have Electronic Pneumatic pressure controllers –older
instruments may have manual pressure control via regulators.
4. Oven: Gas chromatography have ovens that are temperature
programmable, the temperature of the gas chromatographic ovens
typically range from 5oC to 400oC but can go as low as -25oC with
cryogenic cooling.
Weaknesses of GC:
• Requires the analyte to have significant vapour
pressure between about 30atm and 300oC.
• Presents a lack of definitive proof of the nature of
detected compounds. The identification process is
based on retention time matching, which may be
inaccurate or at worst misleading.
INSTRUMENT FUNDAMENTALS –MS
The mass spectrometer is an instrument designed to
separate gas phase ions according to their mass to charge
ratio (m/z) value.
mass to charge ratio (m/z):
This represents the mass of a given particle (Da) to the
number (z) of electrostatic charges (e) that the particle
carries. The term m/z is measured in Da/e.
1. Control Electronics: The MS parameters can be selected and
controlled from this panel. Modern instruments will also allow to
control MS parameters from a computer by using specially
designed software.
2. Detector: The ion beam that emerges from the mass analyzer, have
to be detected and transformed into a usable signal.
The detector is an important element of the mass spectrometer that
generates a signal from incident ions by either generating secondary
electrons, which are further amplified, or by inducing a current
(generated by moving charges).
3. Vacuum System: Mass analysers require high levels of
vacuum in order to operate in a predictable and efficient way.
The vacuum systems of most modern GC-MS systems consist of
a differentially pumped system, usually with a foreline pump
establishing a ‘rough’ vacuum and a high vacuum pump or
pumps situated on the analyser body to establish the high levels of
vacuum required for effective mass to charge ratio measurement.
4. Mass Analyser: There are several very popular types of mass
analyser associated with routine GC-MS analysis and all differ
in the fundamental way in which they separate species on a
mass-to-charge basis.
5. Ion Source: In the ion source, the products are ionized prior to
analysis in the mass spectrometer.
The Coupling of GC/MS
GC-MS has the advantages of higher chromatographic resolution and
higher peak capacity, a single mobile phase (helium), fewer issues with
solubility and separations can be adjusted by electronic controls such
as temperature and flow programming.
Today with the advent of capillary columns using relatively low carrier
gas flow rates, the need for vapour concentrator devices has been
eliminated. The mass spectrometer pumping system can easily deal
with gas flows for direct injection into the mass spectrometer.
The Bieman Concentrator
The Bieman concentrator device is used with packed columns
or with wide bore capillary columns at higher flow rates. In this
device the carrier gas is removed in preference to the analyte
and hence analyte enrichment occurs.
Direct Introduction
Direct introduction is typically used with capillary GC columns
and most modern instruments can easily cope with flow rates up
to 2 mL/min
Ionisation
Mass spectrometry involves the separation of charged species which are
produced by a variety of ionisation methods in GC-MS. These include:
• Electron impact Ionisation (EI) - where analyte molecules are directly
ionised through collision with a bombarding electron stream resulting in the
removal of an electron to form a radical cation species.
• Chemical Ionisation (CI) - where analyte molecules are charged through
reaction processes with a charged reagent gas plasma producing either anion
or cation species depending upon the analyte and analyser polarity.
Electron Impact (EI)
In the Electron Impact (EI) process, electrons are emitted from a
heated filament (usually made of tungsten or rhenium) and are
accelerated across the source by using an appropriate potential
(5-100V) to achieve the required electron energy (sufficient to
ionize the molecule).
M (g) + e- M+ + 2e-
Chemical Ionisation (CI)
Chemical ionisation involves the ionisation of a reagent gas,
such as methane at relatively high pressure (~1 mbar) in a
simple electron impact source. Once produced, the reagent gas
ions collide with the analyte molecules producing ions through
gas phase reaction processes such as proton transfer.
nNH3 + e- NH4+ + 2e-
NH4
+ + M NH3 + [M+ H]+
NH4
+ + M [M+ NH4]+
GC-MS Process
To achieve the identification of different compounds,
three steps can be distinguished in a GC-MS system:
• Sample injection
• Separation.
• Detection
INJECTION
Depending on the sample (gas, liquid or solid), compounds in a
mixture need to be volatilized or extracted from the matrix
INJECTION TECHNIQUE USED FOR
Split High concentration samples (liquid/gas)
Only small amounts of sample required
Split-less Trace analysis (liquid/gas)
Thermal Desorption Trace analysis (gases) and solid samples
Desorption of volatile compounds in a solid
matrix
Headspace Volatile compounds in difficult matrices
For collection of vapors above dirty samples
Solid Phase Micro Extraction
(SPME) Volatile compounds in difficult matrices
with pre- concentration
SEPARATION
After injection of a mixture, separation is achieved in the
capillary column. This column is coated with a fluid or a
solid support, the stationary phase. An inert gas, also
called the mobile phase, is flowing through the column.
Depending on the phase equilibrium between the
stationary and mobile phase, compounds travel with
different velocities through the column. The mixture
becomes separated, and as a result, individual
compounds reach the detector with a different retention
time. By choosing a column, which separates on boiling
point, polarity, size or stereochemistry, a wide range of
compounds can be separated.
DETECTION
Many different detectors can be used for detection of the separated
compounds. The mass spectrometer combines a high sensitivity with the
unique property of being able to determine the molecular composition.
DETECTOR USED F OR
Flame Ionization Detector (FID) Compounds containing Carbon
Thermal Conductivity Detector (TCD) Universal detection for gases without C
Nitrogen Phosphorous Detector (NPD) Selective detection of N and P containing
organic compounds
Electron Capture Detector (ECD) Selective detection of halogen containing
compounds
Atomic Emission Detector (AED) Selective detection of elemental composition
APPLICATIONS
The use of GC-MS in many application areas within analytical
science continues to grow almost exponentially. Listed below are
some pointers as to the applicability of both GC as a separative
technique and MS as a means of detecting analyte species
GC analysis of urine sample know to contain cocaine
MS spectrum of cocaine
GC Separations
• Produce fast analyses because of the highly efficient nature of
the separations achieved
• By using a combination of oven temperature and stationary
phase very difficult separations may also be undertaken
• Excellent for quantitative analysis with a range of sensitive and
linear detectors to choose from (including the MS detector)
• A practical upper temperature limit for conventional GC columns
is around 350-380oC.
In GC analysis, analyte boiling points rarely exceed 400oC and
the upper molecular weight is usually around 500 Da
MS Detection
• Allows specific compound identification. Structural
elucidation via spectral interpretation combined with
elemental composition from accurate mass analysers is
possible.
• Allows high sensitivity detection - femto-gram amounts
have been detected by certain mass analyzer types.
• Is highly selective, certain analyzer and experiment
combinations can lead to highly selective and sensitive
analysis of a wide range of analyte types.
Cleanroom air
The presence of airborne molecular contamination (AMC) in cleanroom air
can have a negative effect on the health. To assess the level of AMC in a
cleanroom several samples are taken using adsorption tubes and sent to the
laboratory for analysis. At the laboratory, the adsorption tubes are desorbed
at elevated temperature and the organic contamination present is analyzed
using GC-MS. Typically, over 100 different compounds are detected,
including compounds like cleanroom solvents (acetone, ethyl acetate etc.)
and large hydrocarbons. Quantification of the obtained data yields the
exact composition of the gas/cleanroom air.
Chromatogram of
cleanroom air.
Some of the characteristic
peaks are indicated:
1= ethanol
2= acetone
3= ethylacetate
4= cyclohexanone
5= dodecane
REFERENCES
• Chromacademy – e- learning for analytical chemistry
community : Fundamentals of GC-MS
• Material analysis by GC-Ms : www.innovatechlabs.com
› analytical services
• Google Search engine: Wikipedia
Gc ms (Revised)

More Related Content

What's hot

Chemical ionization
Chemical ionizationChemical ionization
Chemical ionizationAarif Khan
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry Sailee Gurav
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Sajjad Ullah
 
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]Shikha Popali
 
Fragmentation rule MASS
Fragmentation rule MASSFragmentation rule MASS
Fragmentation rule MASSMahendra G S
 
Comparison of sfc, gc and hplc
Comparison of sfc, gc and hplcComparison of sfc, gc and hplc
Comparison of sfc, gc and hplcVrushali Tambe
 
Hyphenated techniques
Hyphenated techniquesHyphenated techniques
Hyphenated techniquesBiji Saro
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.Dr. Ravi Sankar
 
Gas chromatography-mass spectrometry (GC-MS)-an introduction
Gas chromatography-mass spectrometry (GC-MS)-an introductionGas chromatography-mass spectrometry (GC-MS)-an introduction
Gas chromatography-mass spectrometry (GC-MS)-an introductionRaj Kumar
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometryMISHUSINGH1
 
Gc ms applications
Gc ms applicationsGc ms applications
Gc ms applications9829686702
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYRaju Sanghvi
 
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)Dr. Dinesh Mehta
 
Detectors used in gas chromatography by darshan b j
Detectors used in gas chromatography by darshan b jDetectors used in gas chromatography by darshan b j
Detectors used in gas chromatography by darshan b jDarshan B J
 

What's hot (20)

Chemical ionization
Chemical ionizationChemical ionization
Chemical ionization
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry
 
Gc
GcGc
Gc
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
 
Fragmentation rule MASS
Fragmentation rule MASSFragmentation rule MASS
Fragmentation rule MASS
 
Comparison of sfc, gc and hplc
Comparison of sfc, gc and hplcComparison of sfc, gc and hplc
Comparison of sfc, gc and hplc
 
Hyphenated techniques
Hyphenated techniquesHyphenated techniques
Hyphenated techniques
 
Chromatography(gc ms & lc ms)
Chromatography(gc ms & lc ms)Chromatography(gc ms & lc ms)
Chromatography(gc ms & lc ms)
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
 
Gas chromatography-mass spectrometry (GC-MS)-an introduction
Gas chromatography-mass spectrometry (GC-MS)-an introductionGas chromatography-mass spectrometry (GC-MS)-an introduction
Gas chromatography-mass spectrometry (GC-MS)-an introduction
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Gc ms applications
Gc ms applicationsGc ms applications
Gc ms applications
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRY
 
GC-MS
GC-MSGC-MS
GC-MS
 
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
 
Detectors used in gas chromatography by darshan b j
Detectors used in gas chromatography by darshan b jDetectors used in gas chromatography by darshan b j
Detectors used in gas chromatography by darshan b j
 
Double resonance
Double resonanceDouble resonance
Double resonance
 
Detectors in Gas Chromatography
Detectors in Gas ChromatographyDetectors in Gas Chromatography
Detectors in Gas Chromatography
 
Lc ms
Lc msLc ms
Lc ms
 

Similar to Gc ms (Revised)

Gas chromatography mass spectrometry (GC-MS)
Gas chromatography mass spectrometry (GC-MS)Gas chromatography mass spectrometry (GC-MS)
Gas chromatography mass spectrometry (GC-MS)D.R. Chandravanshi
 
gaschromatography-massspectrometrygc-ms-191115202105.pdf
gaschromatography-massspectrometrygc-ms-191115202105.pdfgaschromatography-massspectrometrygc-ms-191115202105.pdf
gaschromatography-massspectrometrygc-ms-191115202105.pdfssuser98cc4c
 
Modern chromatographic technique.pptx
Modern chromatographic technique.pptxModern chromatographic technique.pptx
Modern chromatographic technique.pptxSilpa87
 
GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .Chinnu Suresh
 
Gas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal DakhaleGas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal DakhaleVishalDakhale
 
Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry Swapnil Baraskar
 
Gas Chromatograph & Analyzers
Gas Chromatograph & AnalyzersGas Chromatograph & Analyzers
Gas Chromatograph & AnalyzersENG AL AMERY
 
Gas Chromatography-Mass Spectrometry
Gas Chromatography-Mass SpectrometryGas Chromatography-Mass Spectrometry
Gas Chromatography-Mass SpectrometryAyesha Ghafoor
 
Gas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptGas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptArunsinghlohhra
 
Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)Saira Fatima
 

Similar to Gc ms (Revised) (20)

Gc Ms
Gc MsGc Ms
Gc Ms
 
Gas chromatography mass spectrometry (GC-MS)
Gas chromatography mass spectrometry (GC-MS)Gas chromatography mass spectrometry (GC-MS)
Gas chromatography mass spectrometry (GC-MS)
 
gaschromatography-massspectrometrygc-ms-191115202105.pdf
gaschromatography-massspectrometrygc-ms-191115202105.pdfgaschromatography-massspectrometrygc-ms-191115202105.pdf
gaschromatography-massspectrometrygc-ms-191115202105.pdf
 
Modern chromatographic technique.pptx
Modern chromatographic technique.pptxModern chromatographic technique.pptx
Modern chromatographic technique.pptx
 
GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .
 
Gas chromatography
Gas chromatography Gas chromatography
Gas chromatography
 
GC-MS .pptx
GC-MS .pptxGC-MS .pptx
GC-MS .pptx
 
Mahesh g ame
Mahesh g ameMahesh g ame
Mahesh g ame
 
Gas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal DakhaleGas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal Dakhale
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gc ms
Gc msGc ms
Gc ms
 
Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry
 
Gas Chromatography and Mass Spectrometry-GCMS
Gas Chromatography and Mass Spectrometry-GCMSGas Chromatography and Mass Spectrometry-GCMS
Gas Chromatography and Mass Spectrometry-GCMS
 
Gas Chromatograph & Analyzers
Gas Chromatograph & AnalyzersGas Chromatograph & Analyzers
Gas Chromatograph & Analyzers
 
GAS CHROMATOGRAPHY
GAS CHROMATOGRAPHYGAS CHROMATOGRAPHY
GAS CHROMATOGRAPHY
 
Gas Chromatography-Mass Spectrometry
Gas Chromatography-Mass SpectrometryGas Chromatography-Mass Spectrometry
Gas Chromatography-Mass Spectrometry
 
Gas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptGas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry ppt
 
GC MASS SPECTROMETERY
GC MASS SPECTROMETERYGC MASS SPECTROMETERY
GC MASS SPECTROMETERY
 
Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 

More from Anamika Banerjee

Introduction to Photoelectrochemical (PEC) Water Splitting
Introduction to Photoelectrochemical (PEC) Water SplittingIntroduction to Photoelectrochemical (PEC) Water Splitting
Introduction to Photoelectrochemical (PEC) Water SplittingAnamika Banerjee
 
Mossbauer- Nuclear quadrupole effect (basics)
Mossbauer- Nuclear quadrupole effect (basics)Mossbauer- Nuclear quadrupole effect (basics)
Mossbauer- Nuclear quadrupole effect (basics)Anamika Banerjee
 
Greenhouse gases and their effect
Greenhouse gases and their effectGreenhouse gases and their effect
Greenhouse gases and their effectAnamika Banerjee
 
Introduction to PAHs in mosquito coil emissions
Introduction to PAHs in mosquito coil emissionsIntroduction to PAHs in mosquito coil emissions
Introduction to PAHs in mosquito coil emissionsAnamika Banerjee
 
Brief introduction to perturbation theory
Brief introduction to perturbation theoryBrief introduction to perturbation theory
Brief introduction to perturbation theoryAnamika Banerjee
 
Biosynthesis of fatty acids
Biosynthesis of fatty acidsBiosynthesis of fatty acids
Biosynthesis of fatty acidsAnamika Banerjee
 

More from Anamika Banerjee (11)

Introduction to Photoelectrochemical (PEC) Water Splitting
Introduction to Photoelectrochemical (PEC) Water SplittingIntroduction to Photoelectrochemical (PEC) Water Splitting
Introduction to Photoelectrochemical (PEC) Water Splitting
 
Mossbauer- Nuclear quadrupole effect (basics)
Mossbauer- Nuclear quadrupole effect (basics)Mossbauer- Nuclear quadrupole effect (basics)
Mossbauer- Nuclear quadrupole effect (basics)
 
Drug dna interaction
Drug dna interactionDrug dna interaction
Drug dna interaction
 
Greenhouse gases and their effect
Greenhouse gases and their effectGreenhouse gases and their effect
Greenhouse gases and their effect
 
SEM,TEM & AFM
SEM,TEM & AFMSEM,TEM & AFM
SEM,TEM & AFM
 
Introduction to PAHs in mosquito coil emissions
Introduction to PAHs in mosquito coil emissionsIntroduction to PAHs in mosquito coil emissions
Introduction to PAHs in mosquito coil emissions
 
Brief introduction to perturbation theory
Brief introduction to perturbation theoryBrief introduction to perturbation theory
Brief introduction to perturbation theory
 
Biosynthesis of fatty acids
Biosynthesis of fatty acidsBiosynthesis of fatty acids
Biosynthesis of fatty acids
 
Literature survey
Literature surveyLiterature survey
Literature survey
 
Synthetic fuel cells
Synthetic   fuel cellsSynthetic   fuel cells
Synthetic fuel cells
 
Gain media & q switching
Gain media & q  switchingGain media & q  switching
Gain media & q switching
 

Recently uploaded

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
 
What is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPWhat is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPCeline George
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Celine George
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxthorishapillay1
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTiammrhaywood
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceSamikshaHamane
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxChelloAnnAsuncion2
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersSabitha Banu
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.arsicmarija21
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfMr Bounab Samir
 
ENGLISH6-Q4-W3.pptxqurter our high choom
ENGLISH6-Q4-W3.pptxqurter our high choomENGLISH6-Q4-W3.pptxqurter our high choom
ENGLISH6-Q4-W3.pptxqurter our high choomnelietumpap1
 
Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Celine George
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfUjwalaBharambe
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfSpandanaRallapalli
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentInMediaRes1
 

Recently uploaded (20)

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
 
9953330565 Low Rate Call Girls In Rohini Delhi NCR
9953330565 Low Rate Call Girls In Rohini  Delhi NCR9953330565 Low Rate Call Girls In Rohini  Delhi NCR
9953330565 Low Rate Call Girls In Rohini Delhi NCR
 
What is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPWhat is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERP
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptx
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in Pharmacovigilance
 
Rapple "Scholarly Communications and the Sustainable Development Goals"
Rapple "Scholarly Communications and the Sustainable Development Goals"Rapple "Scholarly Communications and the Sustainable Development Goals"
Rapple "Scholarly Communications and the Sustainable Development Goals"
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginners
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
 
ENGLISH6-Q4-W3.pptxqurter our high choom
ENGLISH6-Q4-W3.pptxqurter our high choomENGLISH6-Q4-W3.pptxqurter our high choom
ENGLISH6-Q4-W3.pptxqurter our high choom
 
Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17
 
OS-operating systems- ch04 (Threads) ...
OS-operating systems- ch04 (Threads) ...OS-operating systems- ch04 (Threads) ...
OS-operating systems- ch04 (Threads) ...
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdf
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media Component
 
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
 

Gc ms (Revised)

  • 1. Presented by – ANAMIKA BANERJEE (RESEARCH SCHOLAR)
  • 2. Gas chromatography coupled to mass spectrometry is a versatile tool to separate, quantify and identify unknown (volatile) organic compounds and permanent gases. By combining sensitivity and a high resolving power, complex mixtures can be analyzed. The information obtained can be used for detection of impurities, contamination control and improvement of, for example, semiconductor manufacturing processes
  • 3. FUNDAMENTAL OF GC Gas Chromatography (GC) uses a carrier gas to transport sample components through either packed columns or hollow capillary columns containing the stationary phase. In most cases, GC columns have smaller internal diameter and are longer than HPLC columns. In Gas Chromatography, the mobile phase is a gas and the stationary phase is either: • A solid, commonly termed “Gas solid chromatography (GSC)”. • An immobilised polymeric liquid, referred as “Gas Liquid Chromatography (GLC)”. 6
  • 4. • Carrier – (H2, He, N2 – helium is most usual with MS detection) • Make-up gas– (H2, He, N2 - if using an FID detector in parallel to the MS detector) • Detector gas – (H2 & Air – if using an FID detector in parallel to the MS detector) 1. Gas Supply: Gas is fed from cylinders through regulators and tubing to the instrument. It is usual to filter gases to ensure high gas purity and the gas supply pressure. Required gases might include: 2. Interface: After separation in the GC system, analyte species have to be transported to the mass spectrometer to be ionised, mass filtered and detected. The interface in modern instruments is heated to prevent analyte condensation and in some instruments will contain a device to remove carrier gas molecules to allow analyte pre- concentration.
  • 5. 5. Column: Gas Chromatography uses a gaseous mobile phase to transport sample components through columns either packed with coated silica particles or hollow capillary columns containing the stationary phase coated onto the inner wall. Capillary GC columns are usually several meters long (10 – 120m is typical) with an internal diameter of 0.10 – 0.50mm, whilst packed GC columns tend be 1 – 5 meters in length with either 2 or 4mm internal diameter. 6. Injector: Here the sample is volatilised and the resulting gas entrained into the carrier 3. Pneumatic controls: The gas supply is regulated to the correct pressure (or flow) and then fed to the required part of the instrument. Modern GC instruments have Electronic Pneumatic pressure controllers –older instruments may have manual pressure control via regulators. 4. Oven: Gas chromatography have ovens that are temperature programmable, the temperature of the gas chromatographic ovens typically range from 5oC to 400oC but can go as low as -25oC with cryogenic cooling.
  • 6. Weaknesses of GC: • Requires the analyte to have significant vapour pressure between about 30atm and 300oC. • Presents a lack of definitive proof of the nature of detected compounds. The identification process is based on retention time matching, which may be inaccurate or at worst misleading.
  • 7. INSTRUMENT FUNDAMENTALS –MS The mass spectrometer is an instrument designed to separate gas phase ions according to their mass to charge ratio (m/z) value. mass to charge ratio (m/z): This represents the mass of a given particle (Da) to the number (z) of electrostatic charges (e) that the particle carries. The term m/z is measured in Da/e.
  • 8. 1. Control Electronics: The MS parameters can be selected and controlled from this panel. Modern instruments will also allow to control MS parameters from a computer by using specially designed software. 2. Detector: The ion beam that emerges from the mass analyzer, have to be detected and transformed into a usable signal. The detector is an important element of the mass spectrometer that generates a signal from incident ions by either generating secondary electrons, which are further amplified, or by inducing a current (generated by moving charges).
  • 9. 3. Vacuum System: Mass analysers require high levels of vacuum in order to operate in a predictable and efficient way. The vacuum systems of most modern GC-MS systems consist of a differentially pumped system, usually with a foreline pump establishing a ‘rough’ vacuum and a high vacuum pump or pumps situated on the analyser body to establish the high levels of vacuum required for effective mass to charge ratio measurement. 4. Mass Analyser: There are several very popular types of mass analyser associated with routine GC-MS analysis and all differ in the fundamental way in which they separate species on a mass-to-charge basis. 5. Ion Source: In the ion source, the products are ionized prior to analysis in the mass spectrometer.
  • 10. The Coupling of GC/MS GC-MS has the advantages of higher chromatographic resolution and higher peak capacity, a single mobile phase (helium), fewer issues with solubility and separations can be adjusted by electronic controls such as temperature and flow programming. Today with the advent of capillary columns using relatively low carrier gas flow rates, the need for vapour concentrator devices has been eliminated. The mass spectrometer pumping system can easily deal with gas flows for direct injection into the mass spectrometer.
  • 11. The Bieman Concentrator The Bieman concentrator device is used with packed columns or with wide bore capillary columns at higher flow rates. In this device the carrier gas is removed in preference to the analyte and hence analyte enrichment occurs.
  • 12. Direct Introduction Direct introduction is typically used with capillary GC columns and most modern instruments can easily cope with flow rates up to 2 mL/min
  • 13. Ionisation Mass spectrometry involves the separation of charged species which are produced by a variety of ionisation methods in GC-MS. These include: • Electron impact Ionisation (EI) - where analyte molecules are directly ionised through collision with a bombarding electron stream resulting in the removal of an electron to form a radical cation species. • Chemical Ionisation (CI) - where analyte molecules are charged through reaction processes with a charged reagent gas plasma producing either anion or cation species depending upon the analyte and analyser polarity.
  • 14. Electron Impact (EI) In the Electron Impact (EI) process, electrons are emitted from a heated filament (usually made of tungsten or rhenium) and are accelerated across the source by using an appropriate potential (5-100V) to achieve the required electron energy (sufficient to ionize the molecule). M (g) + e- M+ + 2e-
  • 15. Chemical Ionisation (CI) Chemical ionisation involves the ionisation of a reagent gas, such as methane at relatively high pressure (~1 mbar) in a simple electron impact source. Once produced, the reagent gas ions collide with the analyte molecules producing ions through gas phase reaction processes such as proton transfer. nNH3 + e- NH4+ + 2e- NH4 + + M NH3 + [M+ H]+ NH4 + + M [M+ NH4]+
  • 16. GC-MS Process To achieve the identification of different compounds, three steps can be distinguished in a GC-MS system: • Sample injection • Separation. • Detection
  • 17. INJECTION Depending on the sample (gas, liquid or solid), compounds in a mixture need to be volatilized or extracted from the matrix INJECTION TECHNIQUE USED FOR Split High concentration samples (liquid/gas) Only small amounts of sample required Split-less Trace analysis (liquid/gas) Thermal Desorption Trace analysis (gases) and solid samples Desorption of volatile compounds in a solid matrix Headspace Volatile compounds in difficult matrices For collection of vapors above dirty samples Solid Phase Micro Extraction (SPME) Volatile compounds in difficult matrices with pre- concentration
  • 18. SEPARATION After injection of a mixture, separation is achieved in the capillary column. This column is coated with a fluid or a solid support, the stationary phase. An inert gas, also called the mobile phase, is flowing through the column. Depending on the phase equilibrium between the stationary and mobile phase, compounds travel with different velocities through the column. The mixture becomes separated, and as a result, individual compounds reach the detector with a different retention time. By choosing a column, which separates on boiling point, polarity, size or stereochemistry, a wide range of compounds can be separated.
  • 19. DETECTION Many different detectors can be used for detection of the separated compounds. The mass spectrometer combines a high sensitivity with the unique property of being able to determine the molecular composition. DETECTOR USED F OR Flame Ionization Detector (FID) Compounds containing Carbon Thermal Conductivity Detector (TCD) Universal detection for gases without C Nitrogen Phosphorous Detector (NPD) Selective detection of N and P containing organic compounds Electron Capture Detector (ECD) Selective detection of halogen containing compounds Atomic Emission Detector (AED) Selective detection of elemental composition
  • 20. APPLICATIONS The use of GC-MS in many application areas within analytical science continues to grow almost exponentially. Listed below are some pointers as to the applicability of both GC as a separative technique and MS as a means of detecting analyte species GC analysis of urine sample know to contain cocaine
  • 21. MS spectrum of cocaine
  • 22. GC Separations • Produce fast analyses because of the highly efficient nature of the separations achieved • By using a combination of oven temperature and stationary phase very difficult separations may also be undertaken • Excellent for quantitative analysis with a range of sensitive and linear detectors to choose from (including the MS detector) • A practical upper temperature limit for conventional GC columns is around 350-380oC. In GC analysis, analyte boiling points rarely exceed 400oC and the upper molecular weight is usually around 500 Da
  • 23. MS Detection • Allows specific compound identification. Structural elucidation via spectral interpretation combined with elemental composition from accurate mass analysers is possible. • Allows high sensitivity detection - femto-gram amounts have been detected by certain mass analyzer types. • Is highly selective, certain analyzer and experiment combinations can lead to highly selective and sensitive analysis of a wide range of analyte types.
  • 24. Cleanroom air The presence of airborne molecular contamination (AMC) in cleanroom air can have a negative effect on the health. To assess the level of AMC in a cleanroom several samples are taken using adsorption tubes and sent to the laboratory for analysis. At the laboratory, the adsorption tubes are desorbed at elevated temperature and the organic contamination present is analyzed using GC-MS. Typically, over 100 different compounds are detected, including compounds like cleanroom solvents (acetone, ethyl acetate etc.) and large hydrocarbons. Quantification of the obtained data yields the exact composition of the gas/cleanroom air. Chromatogram of cleanroom air. Some of the characteristic peaks are indicated: 1= ethanol 2= acetone 3= ethylacetate 4= cyclohexanone 5= dodecane
  • 25. REFERENCES • Chromacademy – e- learning for analytical chemistry community : Fundamentals of GC-MS • Material analysis by GC-Ms : www.innovatechlabs.com › analytical services • Google Search engine: Wikipedia