PART I
Dose Distribution Measurement
TOPIC 3
EXTERNAL PHOTON BEAMS THERAPY
128/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
PART 1
3.0 Introduction Of Treatment Planning
3.1 Dose Distribution Measurement
3.1.1 Beam Profile
3.1.2 Isodose curve
3.1.2.1 Properties of Isodose Charts
3.1.2.2 Measurement of Isodose Curves
3.1.2.3 Parameters Affecting Isodose Curves
3.1.2.4.1 Beam Quality
3.1.2.4.2 SSD
3.1.2.4.3 Collimation and Flattening Filter
3.1.2.4.4 Field Size
3.1.2.45. Wedge Filter
2
3.2 Central Axis Depth Doses In Water
3.2.1 Central Axis Depth Doses In Water SSD Technique
3.2.1.1 Percentage Depth Dose (PDD)
3.2.1.1.1 Effect of Beam Quality & Depth
3.2.1.1.2 Effect of Field Size & Shape
3.2.1.1.3 Effect of SSD
3.2.2 Central Axis Depth Doses In Water SAD Technique
3.2.2.1 Tissue Air Ratio (TAR)
3.2.2.1.1 Effect of Distance
3.2.2.1.2 Effect of energy, depth, and field size
3.2.2.1.3 TAR and SAD
3.2.2.1.4 BSF
3.2.2.2 Tissue Phantom Ratio (TPR)
3.2.2.3 Tissue Maximum Ratio (TMR)
3.2.2.4 Scatter Air Ratio (SAR)
PART 2
28/1/2018 3
PART 3
3.3 Treatment Technique
3.3.1 Single Field Technique
3.3.2 Parallel Opposed Field
3.3.3 Multiple Field
3.3.4 Rotation Therapy
3.3.5 Wedge Field
428/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.0 Introduction of treatment planning
528/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
TREATMENT PLANNING
Positioning
Immobilization
Localization
Field Selection
Dose distribution
Calculations
Verification
Execution
3.0 Introduction of Treatment Planning
• Water phantom:
• closely approximates the radiation
absorption and scattering properties of
muscle and other soft tissues;
• universally available with reproducible
radiation properties
water phantom
1 Phantom
• Solid dry phantoms:
• same effective atomic number, number of electrons per gram and mass density
• E.g: Lucite (perspex), polystyrene
• Solid Water Phantom:
• epoxy resin--based solid substitute for water
• E.g: acrylic
1 Phantom
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 9
• Alderson Rando Phantom:
• incorporates materials to simulate various
body tissues---muscle, bone, lung, and air
cavities
Alderson Rando phantom
1 Phantom
2 Treatment Planning
1128/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 12
2 Treatment Planning
13
1
• Treatment planning can be described as the iterative process
whereby the treatment strategy of the radiation oncologist is
quantified as a set of treatment instructions including a
description of the expected dose distribution in the patient
2
• Planning is based on predictions of dosage delivered to the
individual patient by the proposed arrangement of radiation
beams
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
2 Treatment Planning
14
3
• External photon beam radiotherapy is usually carried out
with more than one radiation beam in order to achieve a
uniform dose distribution inside the target volume and an as low
as possible a dose in healthy tissues surrounding the target
4
• ICRU Report No. 50 recommends a target dose uniformity within
+7% and –5% of the dose delivered to a well defined prescription
point within the target
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 15
The main parameters in external beam dose delivery with
photon beams are:
Depth of treatment
Fields size
Source-skin distance (SSD) in SSD
setups.
Number of beams used in dose
delivery to the patient.
Photon beam energy.
Source-axis distance (SAD) in SAD
setup
Treatment time for orthovoltage and
teletherapy machines.
Number of monitor units (MUs) for
linacs.
2 Treatment Planning
3.1 Measurement of Dose Distribution
3.1.1 Beam Profile
3.1.2 Isodose curve
3.1.3 PDD
1628/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
Dose distribution &
scatter analysis
Off axis ratios &
beam profiles
Central axis depth dose in
water
Isodose distribution
Beam flatness
Isodose
correctionSource to
surface
distance
Source to
axis
distance
Beam
symmetry
Percentage
depth dose
Tissue air
ratio
Relationship
Tissue
phantom ratio
Tissue maximum ratio
Scatter air ratio
Scatter maximum ratio
3.1.1 Beam Profile
1828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.1.3 Penumbra
3.1.1.1Beam Flatness (F)
3.1.1.2 Beam symmetry (S)
3.1.1 Beam Profile
A beam profile is
measured at multiple
points on a plane
perpendicular to the
central beam axis.
Measurement is usually
performed in a water
phantom using a
cylindrical ionisation
chamber
Beam profile may
also be determined
by film dosimetry or
other dosimeters,
particularly TLD or
silicon diodes which
have a small
detection area.
A beam profile can
be one dimensional
(along one axis) or
two dimensional
(measuring in the x
and y axes).
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 20
3.1.1 Beam Profile3.1.1 Beam Profile
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 21
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 22
3.1.1 Beam Profile
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 23
3.1.1 Beam Profile
Beam Profiles
(dose variation across the field at a specified depth)
3.1.1 Beam Profile
• usually specified at 10 cm
• within 3% over 80% of the field
flatness
• usually specified at 10 cm
• within 2% over 80% of the fieldsymmetry
• dose falls off rapidly at the beam edge,
between a dose of 20-80% of the central
beam axis
penumbra region
• dose is minimal (under 20% of the central
beam dose)umbra region
• Is assessed by finding the maximum Dmax and
Dmin dose point values on the beam profile
within the central 80% of
F=100 x Dmax-Dmin
Dmax+Dmin
the beam width
•
Standard linac specification generally require that F be less than 3% when
measured in a water phantom at a depth of 10 cm and an SSD of 100cm for
the largest field size available
3.1.1.1Beam Flatness (F)
• Determined at Dmax which represents the most
sensitive depth
A typical symmetry specification is that any
two dose points on a beam profile, equidistant
from the central axis point, are within 2 % of
each other
•
3.1.1.2 Beam symmetry (S)
• Alternately area under the dmax beam profiles
on each side (left and right) of central axis
extending to the 50% dose level (normalized
to 100% at the central axis point)
determined.
S = 100 x arealeft-arearight
arealeft + arearight
are
•
3.1.1.2 Beam symmetry (S)
• Penumbra is the unsharp edge of the radiation beam created mainly by the
finite source size
• Penumbra could be classified as
– Geometric Penumbra which depends on
• Source Size
• Source to Diaphragm distance
• Source to skin distance
– Radiological Penumbra which is
• Geometric penumbra + Scatter
Part VIII.3.5 Determination of Dose to a Patient-I Slide 28
3.1.1.3 Penumbra
The Geometric Penumbra
Slide 29
SAD Field size at surface
3.1.1.3 Penumbra
Physical Penumbra / Radiological
penumbra
• Radiation beam penumbra
– Lateral scatter increases the radiation beam penumbra
– Lower energy, more side scatter and hence larger
penumbra
– Density of the scattering medium affects the penumbra
• Issues : larger the penumbra, greater the dose to normal tissues
Part VIII.3.5 Determination of Dose to a Patient-I Slide 30
3.1.1.3 Penumbra
Scattered radiation
Part VIII.3.5 Determination of Dose to a Patient-I Slide 31
Scatter
3.1.1.3 Penumbra
3.1.2 Isodose curve
3.1.2.1 Properties of Isodose Charts
3.1.2.2 Measurement of Isodose Curves
3.1.2.3 Parameters Affecting Isodose Curves
3.1.2.4.1 Beam Quality
3.1.2.4.2 Source Size, SSD, and SDD
3.1.2.4.3 Collimation and Flattening Filter
3.1.2.4.4 Field Size
3.1.2.4.5 Wedge Filter
3328/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
Beams of ionising
radiation have
characteristic
process of energy
deposition, in order
to represent
volumetric & planar
variations in
absorbed dose are
depicted by isodose
curves
3.2 Isodose curve3.2 Isodose curve
 Isodose curves
– Lines joining the points of equal percentage depth
dose(PDD)
– Curves are usually drawn at regular intervals of
absorbed dose & expressed as a percentage of dose
• Isodose charts
– Family of isodose curves
– PDD values are normalised at Dmax or reference
depth
3.2 Isodose curve
3.2 Isodose curve
36
1
• Isodose curves are lines that join points of equal dose
(drawn at interval absorbed dose & Expressed as a
percentage of the dose at a reference point).
2
• They offer a planar representation of the dose distribution and
easily show the behaviour of one beam or a combination of
beams with different shielding, wedges, bolus, etc.
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
Isodose Distribution
Isodose is calculated by using 2 normalization:
– For SSD set ups, all isodose values are normalized
to 100 at point P on the central beam axis
– For SAD set-ups, the isodose values are
normalized to 100 at the isocentre
•
• The isodose charts for an SSD set up are
of PDD values
plots
• For SAD set up, plots are either TAR or TMR
values
3.2 Isodose curve
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 38
3.2 Isodose curve
Part VIII.3.7 Operational Considerations – Planning of physical
treatment
Slide 39
Isodose curves and chart
2 3 4 5 6 7 8 9 10 10 10 9 8 7 6 5 4 3 2
2 3 4 5 6 7 8 9 9.5 9.5 9.5 9 8 7 6 5 4 3 2
2 3 4 5 6 7 8 9 9 9 9 9 8 7 6 5 4 3 2
2 3 4 5 6 7 8 8.5 8.5 8.5 8.5 8.5 8 7 6 5 4 3 2
2 3 4 5 6 7 8 8 8 8 8 8 8 7 6 5 4 3 2
2 3 4 5 6 7 7.5 7.5 7.5 7.5 7.5 7.5 7.5 7 6 5 4 3 2
2 3 4 5 6 7 7 7 7 7 7 7 7 7 6 5 4 3 2
2 3 4 5 6 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6 5 4 3 2
2 3 4 5 6 6 6 6 6 6 6 6 6 6 6 5 4 3 2
2 3 4 5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5 4 3 2
2 3 4 5 5 5 5 5 5 5 5 5 5 5 5 5 4 3 2
2 3 4 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4 3 2
2 3 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 3 2
2 3 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3 2
2 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 2
2 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
Isodose curve or line; line joining points of equal dose value
a representative diagram
A typical isodose chart
90
80
70
60
50
40
30
20
3.2 Isodose curve
40
3
• Isodose curves can be measured in water directly or can be
calculated from PDD and beam profile data
4
• A set of isodose curves is valid for a given treatment machine,
beam energy, SSD and field size
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.2 Isodose curve
41
5
• While isodose curves can be made to display the actual dose in
grays, it is more common to present them normalized to 100%
at a fixed point
6
• Two such common point normalizations are as follows:
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 42
3.2 Isodose curve
3.1.2.1 Properties of Isodose Charts
4428/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.1 Properties of Isodose Charts
2.Dose greatest on the CA at any depth and dec
rease toward the edges of the
beam with the exception
of LINAC beams with horns at shallow depths
1.Horns created by flattening filter which is designed to
overcompensate near the
urface in order to obtain flat isodose curves at depth
(10 cm)
3. The penumbra region:
The dose rate decreases rapidly as a function of lateral distance
from the beam axis.
The width of geometric penumbra depends on source size(d), distance from
the source (SD), and source-to-collimator diaphragm distance (SDD).
The falloff near the edge of the beam is caused by geometric
penumbra and reduced side scatter
This falloff can be described using physical (total) penumbra:
Physical penumbra is defined as lateral distance between two specified isodose curves at a specified dept
h (e.g., lateral distance between 90% and 20% isodose lines at the depth of Dmax).
3.1.2.1 Properties of Isodose Charts
4. Outside the geometric limits of the beam
and the penumbra
• The dose variation is the result of side
scatter from the field and both leakage
and scatter from the collimator system.
 The dose in this region is generally
low results from radiation transmitted
through the collimator and head
shielding
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 48
3.1.2.1 Properties of Isodose Charts
3.1.2.2 Measurement of Isodose Curves
4928/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.2 Measurement of Isodose Curves
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 51
3.1.2.2 Measurement of Isodose Curves
1. Measurement in water phantom
Ionization chamberWater phantom
IC position in water phantom
2. Measurement in solid water phantom
2. Measurement in solid water phantom
3.1.2.3 Parameters Affecting Isodose Curves
3.1.2.3.1 Beam Quality
3.1.2.3.2 Source Size and SSD
3.1.2.3.3 Collimation and Flattening Filter
3.1.2.3.4 Field Size
3.1.2.3.5 Wedge Filter
5628/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.2.4 Parameters Affecting Isodose Curves
Beam
quality
Source
Size and
SSD
Collimation
and
flattening
filter
Field size
Parameters
affecting
the single-
beam
isodose
3.1.2.3.1 Beam Quality
5828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.3.1 Beam Quality
Depth
The depth of a
given isodose
curve increases
with beam
quality
Beam energy
Beam energy
also influences
isodose curve
shape near the
field border
Lateral scatter
Greater lateral scatter
associated with lower-energy
beams causes the isodose curves
outside the field to bulge out
In other words, the absorbed
dose in the medium outside the
primary beam is greater for low
energy beams than for those of
higher energy
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 60
3.1.2.3.1 Beam Quality3.1.2.3.1 Beam Quality
4 MV 6MV 15MV
3.1.2.3.1 Beam Quality
One disadvantage of the
orthovoltage beams is increased
scattered dose to tissue outside the
treatment region.
For megavoltage beams, the scatter
outside the field is minimized as a
result of forward scattering and
becomes more a function of
collimation than energy.
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 63
3.1.2.3.1 Beam Quality
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 64
3.1.2.3.1 Beam Quality
3.1.2.3.2 SSD
6528/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.3.2 SSD
SSD affect the isodose curves by the geometric
penumbra.
The SSD affects the PDD and the depth of the
isodose curves.
The dose variation across the field border is a
complex function of geometric penumbra, lateral
scatter, and collimation.
SSD
3.1.2.3.2 SSD
3.1.2.4.3 Collimation and Flattening Filter
6828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.3.3 Collimation and Flattening Filter
Collimation
If no filter used, the result is conical
field distribution
Makes field uniform across whole
width, with the same intensity
Flattening filter thickest in the middle
and tapers to sides
Filter causes change in beam quality
Different photon spectrum at edge than
in middle, harder in middle
1. Flattening filter
2.Collimator blocks
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 70
3.1.2.3.3 Collimation and Flattening Filter
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 71
3.1.2.3.3 Collimation and Flattening Filter
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 72
3.1.2.3.4 Field Size
• Field size is the width and length of the radiation beam at SSD or SAD
• Field size at any depth is usually defined by the 50% width of the profile at
that depth
Part VIII.3.5 Determination of Dose to a Patient-I Slide 73
3.1.2.3.4 Field Size
Field size (dosimetrical):
lateral distance between the 50%
isodose lines at a reference depth
Field size (geometrical) :
field defining light is made to coincide
with 50% isodose lines of the radiation
beam projected on a plane
perpendicular to the beam axis and at
standard SSD or SAD
3.1.2.3.4 Field Size
3.1.2.3.4 Field Size
Field size
< 6 cm
TPS should be
mandatory for
small field size.
Bell shape
Relative
large
penumbra
region
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 76
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 77
3.1.2.3.5 Wedge Filter
7828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
3.1.2.5 Wedge Filter
Wedge Filters
Figure A:normalized to Dmax,
B:normalized to Dmax without wedge
3.1.2.3.5 Wedge Filter
the degree of tilt being
dependent on the slope
of the wedge filter
Wedge shaped: cause
the isodose curve
tilted from their
normal position
The most
commonly used
beam-
modifying
device
Mounted on a (transparent
plastic) tray
Arranged at a distance of at
least 15 cm from the skin
surface, avoid destroying
the skin-sparing effect of
the megavoltage beam
3.1.2.3.5 Wedge Filter
10 cm
wedge angle
3.1.2.3.5 Wedge Filter
Wedge angle
the angle through which an isodose curve is titled at the
central ray of a beam at a specified depth (10 cm)
The angle between the isodose curve and the normal to the
central axis
The angle of isodose tilt decreases with increasing depth in
the phantom due to presence of scattered radiation
28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 83
3.1.2.3.5 Wedge Filter
3.1.2.3.5 Wedge Filter
WEDGE FACTOR
The ratio of doses
with and without the
wedge, at a point in
phantom along the
central axis of the
beam
Measured at a
suitable depth beyond
dmax
(5 to 10 cm) either in
air or in phantom
Sometimes
incorporated into the
isodose curves
The presence of a wedge filter decreases the output of the machine.
Hinge angle,φ
It is the angle between central axes of two
beams passing through the wedge
Relationship b/w φ & θ
Wedge angle,θ= 90 – φ/2
3.1.2.3.5 Wedge Filter
Dose Distribution Measurement (part 1)

More Related Content

What's hot

Dosimetric calculations
Dosimetric calculationsDosimetric calculations
Dosimetric calculations
CSULB
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
fondas vakalis
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
Santam Chakraborty
 

What's hot (20)

Helical Tomotherapy
Helical TomotherapyHelical Tomotherapy
Helical Tomotherapy
 
Epid
EpidEpid
Epid
 
Immobilization device in radiotherapy
Immobilization device in radiotherapyImmobilization device in radiotherapy
Immobilization device in radiotherapy
 
TSET
TSETTSET
TSET
 
Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Mlc
MlcMlc
Mlc
 
Tomotherapy
TomotherapyTomotherapy
Tomotherapy
 
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
 
Dosimetric calculations
Dosimetric calculationsDosimetric calculations
Dosimetric calculations
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 
Role of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapyRole of immobilisation and devices in radiotherapy
Role of immobilisation and devices in radiotherapy
 
Gap correction
Gap correctionGap correction
Gap correction
 
Beam modification
Beam modification Beam modification
Beam modification
 
Photon dosimetry 31 01-2022
Photon dosimetry 31 01-2022Photon dosimetry 31 01-2022
Photon dosimetry 31 01-2022
 
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in RadiotherapyDosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
 
ICRU 83
ICRU 83ICRU 83
ICRU 83
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetry
 
Total body irradiation
Total body irradiationTotal body irradiation
Total body irradiation
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 

Similar to Dose Distribution Measurement (part 1)

Design study of stressed mirror polishing (smp) fixture
Design study of stressed mirror polishing (smp) fixtureDesign study of stressed mirror polishing (smp) fixture
Design study of stressed mirror polishing (smp) fixture
eSAT Journals
 
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
Simona Liukaityte
 
Verification of a treatment planning system using an in-house designedtrunk p...
Verification of a treatment planning system using an in-house designedtrunk p...Verification of a treatment planning system using an in-house designedtrunk p...
Verification of a treatment planning system using an in-house designedtrunk p...
IOSR Journals
 

Similar to Dose Distribution Measurement (part 1) (20)

RADIOTHERAPY CALCULATION
RADIOTHERAPY CALCULATIONRADIOTHERAPY CALCULATION
RADIOTHERAPY CALCULATION
 
ICRU 71
ICRU 71ICRU 71
ICRU 71
 
Hrt II
Hrt IIHrt II
Hrt II
 
Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...
PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...
PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...
 
Manuale fantoccio mri 76 903
Manuale fantoccio mri 76 903Manuale fantoccio mri 76 903
Manuale fantoccio mri 76 903
 
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
 
Design study of stressed mirror polishing (smp) fixture
Design study of stressed mirror polishing (smp) fixtureDesign study of stressed mirror polishing (smp) fixture
Design study of stressed mirror polishing (smp) fixture
 
Corneal topography wavefront analysis
Corneal topography wavefront analysisCorneal topography wavefront analysis
Corneal topography wavefront analysis
 
MRI Image Compression
MRI Image CompressionMRI Image Compression
MRI Image Compression
 
Cone beam computed tomography (2)
Cone beam computed tomography (2)Cone beam computed tomography (2)
Cone beam computed tomography (2)
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
MEASUREMENT OF SCATTERING LOSSES OF LASER COMPONENTS BY TOTAL INTEGRATED SCAT...
 
Feature extraction from asthma patient’s voice using mel frequency cepstral c...
Feature extraction from asthma patient’s voice using mel frequency cepstral c...Feature extraction from asthma patient’s voice using mel frequency cepstral c...
Feature extraction from asthma patient’s voice using mel frequency cepstral c...
 
Image guided surgery
Image guided surgeryImage guided surgery
Image guided surgery
 
BASICS RADIOBIOLOGY FOR RADIOTHERAPY
BASICS RADIOBIOLOGY FOR RADIOTHERAPYBASICS RADIOBIOLOGY FOR RADIOTHERAPY
BASICS RADIOBIOLOGY FOR RADIOTHERAPY
 
Verification of a treatment planning system using an in-house designedtrunk p...
Verification of a treatment planning system using an in-house designedtrunk p...Verification of a treatment planning system using an in-house designedtrunk p...
Verification of a treatment planning system using an in-house designedtrunk p...
 
Resolving filter grating etc
Resolving filter grating etcResolving filter grating etc
Resolving filter grating etc
 
Paper id 312201526
Paper id 312201526Paper id 312201526
Paper id 312201526
 
Technique 1 inroduction
Technique 1 inroductionTechnique 1 inroduction
Technique 1 inroduction
 

More from Nik Noor Ashikin Nik Ab Razak (12)

BASIC RADIOBIOLOGY FOR RADIOTHERAPY
BASIC RADIOBIOLOGY FOR RADIOTHERAPYBASIC RADIOBIOLOGY FOR RADIOTHERAPY
BASIC RADIOBIOLOGY FOR RADIOTHERAPY
 
Introduction to Radiation Therapy
Introduction to Radiation TherapyIntroduction to Radiation Therapy
Introduction to Radiation Therapy
 
MRI
MRIMRI
MRI
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
X-RAY TUBE (ANODE)
X-RAY TUBE (ANODE)X-RAY TUBE (ANODE)
X-RAY TUBE (ANODE)
 
BASICS RADIOBIOLOGY FOR RADIOTHERAPY
BASICS RADIOBIOLOGY FOR RADIOTHERAPYBASICS RADIOBIOLOGY FOR RADIOTHERAPY
BASICS RADIOBIOLOGY FOR RADIOTHERAPY
 
introduction of Radiotherapy
introduction of Radiotherapyintroduction of Radiotherapy
introduction of Radiotherapy
 
LINEAR ACCELERATOR
LINEAR ACCELERATORLINEAR ACCELERATOR
LINEAR ACCELERATOR
 
Radiotherapy Equipment
Radiotherapy EquipmentRadiotherapy Equipment
Radiotherapy Equipment
 
Topic 2: Structure of matter
Topic 2: Structure of matterTopic 2: Structure of matter
Topic 2: Structure of matter
 
TOPIC 1: HISTORY OF RADIATION (PART 2)
TOPIC 1: HISTORY OF RADIATION (PART 2)TOPIC 1: HISTORY OF RADIATION (PART 2)
TOPIC 1: HISTORY OF RADIATION (PART 2)
 
TOPIC 1: HISTORY OF RADIATION
TOPIC 1: HISTORY OF RADIATIONTOPIC 1: HISTORY OF RADIATION
TOPIC 1: HISTORY OF RADIATION
 

Recently uploaded

The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
seri bangash
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
Silpa
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
Silpa
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 

Recently uploaded (20)

Chemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdfChemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdf
 
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
 
The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
 
Zoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdfZoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdf
 
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIACURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
 
An introduction on sequence tagged site mapping
An introduction on sequence tagged site mappingAn introduction on sequence tagged site mapping
An introduction on sequence tagged site mapping
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
 
GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
 
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort ServiceCall Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
 
Grade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its FunctionsGrade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its Functions
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)
 
Velocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.pptVelocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.ppt
 
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICEPATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
 
Factory Acceptance Test( FAT).pptx .
Factory Acceptance Test( FAT).pptx       .Factory Acceptance Test( FAT).pptx       .
Factory Acceptance Test( FAT).pptx .
 
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS ESCORT SERVICE In Bhiwan...
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS  ESCORT SERVICE In Bhiwan...Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS  ESCORT SERVICE In Bhiwan...
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS ESCORT SERVICE In Bhiwan...
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate ProfessorThyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
 
300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx
 

Dose Distribution Measurement (part 1)

  • 1. PART I Dose Distribution Measurement TOPIC 3 EXTERNAL PHOTON BEAMS THERAPY 128/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 2. PART 1 3.0 Introduction Of Treatment Planning 3.1 Dose Distribution Measurement 3.1.1 Beam Profile 3.1.2 Isodose curve 3.1.2.1 Properties of Isodose Charts 3.1.2.2 Measurement of Isodose Curves 3.1.2.3 Parameters Affecting Isodose Curves 3.1.2.4.1 Beam Quality 3.1.2.4.2 SSD 3.1.2.4.3 Collimation and Flattening Filter 3.1.2.4.4 Field Size 3.1.2.45. Wedge Filter 2
  • 3. 3.2 Central Axis Depth Doses In Water 3.2.1 Central Axis Depth Doses In Water SSD Technique 3.2.1.1 Percentage Depth Dose (PDD) 3.2.1.1.1 Effect of Beam Quality & Depth 3.2.1.1.2 Effect of Field Size & Shape 3.2.1.1.3 Effect of SSD 3.2.2 Central Axis Depth Doses In Water SAD Technique 3.2.2.1 Tissue Air Ratio (TAR) 3.2.2.1.1 Effect of Distance 3.2.2.1.2 Effect of energy, depth, and field size 3.2.2.1.3 TAR and SAD 3.2.2.1.4 BSF 3.2.2.2 Tissue Phantom Ratio (TPR) 3.2.2.3 Tissue Maximum Ratio (TMR) 3.2.2.4 Scatter Air Ratio (SAR) PART 2 28/1/2018 3
  • 4. PART 3 3.3 Treatment Technique 3.3.1 Single Field Technique 3.3.2 Parallel Opposed Field 3.3.3 Multiple Field 3.3.4 Rotation Therapy 3.3.5 Wedge Field 428/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 5. 3.0 Introduction of treatment planning 528/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 6. TREATMENT PLANNING Positioning Immobilization Localization Field Selection Dose distribution Calculations Verification Execution 3.0 Introduction of Treatment Planning
  • 7. • Water phantom: • closely approximates the radiation absorption and scattering properties of muscle and other soft tissues; • universally available with reproducible radiation properties water phantom 1 Phantom
  • 8. • Solid dry phantoms: • same effective atomic number, number of electrons per gram and mass density • E.g: Lucite (perspex), polystyrene • Solid Water Phantom: • epoxy resin--based solid substitute for water • E.g: acrylic 1 Phantom
  • 9. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 9
  • 10. • Alderson Rando Phantom: • incorporates materials to simulate various body tissues---muscle, bone, lung, and air cavities Alderson Rando phantom 1 Phantom
  • 11. 2 Treatment Planning 1128/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 12. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 12
  • 13. 2 Treatment Planning 13 1 • Treatment planning can be described as the iterative process whereby the treatment strategy of the radiation oncologist is quantified as a set of treatment instructions including a description of the expected dose distribution in the patient 2 • Planning is based on predictions of dosage delivered to the individual patient by the proposed arrangement of radiation beams 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 14. 2 Treatment Planning 14 3 • External photon beam radiotherapy is usually carried out with more than one radiation beam in order to achieve a uniform dose distribution inside the target volume and an as low as possible a dose in healthy tissues surrounding the target 4 • ICRU Report No. 50 recommends a target dose uniformity within +7% and –5% of the dose delivered to a well defined prescription point within the target 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 15. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 15 The main parameters in external beam dose delivery with photon beams are: Depth of treatment Fields size Source-skin distance (SSD) in SSD setups. Number of beams used in dose delivery to the patient. Photon beam energy. Source-axis distance (SAD) in SAD setup Treatment time for orthovoltage and teletherapy machines. Number of monitor units (MUs) for linacs. 2 Treatment Planning
  • 16. 3.1 Measurement of Dose Distribution 3.1.1 Beam Profile 3.1.2 Isodose curve 3.1.3 PDD 1628/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 17. Dose distribution & scatter analysis Off axis ratios & beam profiles Central axis depth dose in water Isodose distribution Beam flatness Isodose correctionSource to surface distance Source to axis distance Beam symmetry Percentage depth dose Tissue air ratio Relationship Tissue phantom ratio Tissue maximum ratio Scatter air ratio Scatter maximum ratio
  • 18. 3.1.1 Beam Profile 1828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 3.1.1.3 Penumbra 3.1.1.1Beam Flatness (F) 3.1.1.2 Beam symmetry (S)
  • 19. 3.1.1 Beam Profile A beam profile is measured at multiple points on a plane perpendicular to the central beam axis. Measurement is usually performed in a water phantom using a cylindrical ionisation chamber Beam profile may also be determined by film dosimetry or other dosimeters, particularly TLD or silicon diodes which have a small detection area. A beam profile can be one dimensional (along one axis) or two dimensional (measuring in the x and y axes).
  • 20. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 20 3.1.1 Beam Profile3.1.1 Beam Profile
  • 21. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 21
  • 22. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 22 3.1.1 Beam Profile
  • 23. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 23 3.1.1 Beam Profile
  • 24. Beam Profiles (dose variation across the field at a specified depth) 3.1.1 Beam Profile • usually specified at 10 cm • within 3% over 80% of the field flatness • usually specified at 10 cm • within 2% over 80% of the fieldsymmetry • dose falls off rapidly at the beam edge, between a dose of 20-80% of the central beam axis penumbra region • dose is minimal (under 20% of the central beam dose)umbra region
  • 25. • Is assessed by finding the maximum Dmax and Dmin dose point values on the beam profile within the central 80% of F=100 x Dmax-Dmin Dmax+Dmin the beam width • Standard linac specification generally require that F be less than 3% when measured in a water phantom at a depth of 10 cm and an SSD of 100cm for the largest field size available 3.1.1.1Beam Flatness (F)
  • 26. • Determined at Dmax which represents the most sensitive depth A typical symmetry specification is that any two dose points on a beam profile, equidistant from the central axis point, are within 2 % of each other • 3.1.1.2 Beam symmetry (S)
  • 27. • Alternately area under the dmax beam profiles on each side (left and right) of central axis extending to the 50% dose level (normalized to 100% at the central axis point) determined. S = 100 x arealeft-arearight arealeft + arearight are • 3.1.1.2 Beam symmetry (S)
  • 28. • Penumbra is the unsharp edge of the radiation beam created mainly by the finite source size • Penumbra could be classified as – Geometric Penumbra which depends on • Source Size • Source to Diaphragm distance • Source to skin distance – Radiological Penumbra which is • Geometric penumbra + Scatter Part VIII.3.5 Determination of Dose to a Patient-I Slide 28 3.1.1.3 Penumbra
  • 29. The Geometric Penumbra Slide 29 SAD Field size at surface 3.1.1.3 Penumbra
  • 30. Physical Penumbra / Radiological penumbra • Radiation beam penumbra – Lateral scatter increases the radiation beam penumbra – Lower energy, more side scatter and hence larger penumbra – Density of the scattering medium affects the penumbra • Issues : larger the penumbra, greater the dose to normal tissues Part VIII.3.5 Determination of Dose to a Patient-I Slide 30 3.1.1.3 Penumbra
  • 31. Scattered radiation Part VIII.3.5 Determination of Dose to a Patient-I Slide 31 Scatter 3.1.1.3 Penumbra
  • 32.
  • 33. 3.1.2 Isodose curve 3.1.2.1 Properties of Isodose Charts 3.1.2.2 Measurement of Isodose Curves 3.1.2.3 Parameters Affecting Isodose Curves 3.1.2.4.1 Beam Quality 3.1.2.4.2 Source Size, SSD, and SDD 3.1.2.4.3 Collimation and Flattening Filter 3.1.2.4.4 Field Size 3.1.2.4.5 Wedge Filter 3328/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 34. Beams of ionising radiation have characteristic process of energy deposition, in order to represent volumetric & planar variations in absorbed dose are depicted by isodose curves 3.2 Isodose curve3.2 Isodose curve
  • 35.  Isodose curves – Lines joining the points of equal percentage depth dose(PDD) – Curves are usually drawn at regular intervals of absorbed dose & expressed as a percentage of dose • Isodose charts – Family of isodose curves – PDD values are normalised at Dmax or reference depth 3.2 Isodose curve
  • 36. 3.2 Isodose curve 36 1 • Isodose curves are lines that join points of equal dose (drawn at interval absorbed dose & Expressed as a percentage of the dose at a reference point). 2 • They offer a planar representation of the dose distribution and easily show the behaviour of one beam or a combination of beams with different shielding, wedges, bolus, etc. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 37. Isodose Distribution Isodose is calculated by using 2 normalization: – For SSD set ups, all isodose values are normalized to 100 at point P on the central beam axis – For SAD set-ups, the isodose values are normalized to 100 at the isocentre • • The isodose charts for an SSD set up are of PDD values plots • For SAD set up, plots are either TAR or TMR values 3.2 Isodose curve
  • 38. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 38 3.2 Isodose curve
  • 39. Part VIII.3.7 Operational Considerations – Planning of physical treatment Slide 39 Isodose curves and chart 2 3 4 5 6 7 8 9 10 10 10 9 8 7 6 5 4 3 2 2 3 4 5 6 7 8 9 9.5 9.5 9.5 9 8 7 6 5 4 3 2 2 3 4 5 6 7 8 9 9 9 9 9 8 7 6 5 4 3 2 2 3 4 5 6 7 8 8.5 8.5 8.5 8.5 8.5 8 7 6 5 4 3 2 2 3 4 5 6 7 8 8 8 8 8 8 8 7 6 5 4 3 2 2 3 4 5 6 7 7.5 7.5 7.5 7.5 7.5 7.5 7.5 7 6 5 4 3 2 2 3 4 5 6 7 7 7 7 7 7 7 7 7 6 5 4 3 2 2 3 4 5 6 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6.5 6 5 4 3 2 2 3 4 5 6 6 6 6 6 6 6 6 6 6 6 5 4 3 2 2 3 4 5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5 4 3 2 2 3 4 5 5 5 5 5 5 5 5 5 5 5 5 5 4 3 2 2 3 4 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4.5 4 3 2 2 3 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 3 2 2 3 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3 2 2 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 2 2 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2.5 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 Isodose curve or line; line joining points of equal dose value a representative diagram A typical isodose chart 90 80 70 60 50 40 30 20
  • 40. 3.2 Isodose curve 40 3 • Isodose curves can be measured in water directly or can be calculated from PDD and beam profile data 4 • A set of isodose curves is valid for a given treatment machine, beam energy, SSD and field size 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 41. 3.2 Isodose curve 41 5 • While isodose curves can be made to display the actual dose in grays, it is more common to present them normalized to 100% at a fixed point 6 • Two such common point normalizations are as follows: 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 42. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 42 3.2 Isodose curve
  • 43.
  • 44. 3.1.2.1 Properties of Isodose Charts 4428/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 45. 3.1.2.1 Properties of Isodose Charts 2.Dose greatest on the CA at any depth and dec rease toward the edges of the beam with the exception of LINAC beams with horns at shallow depths 1.Horns created by flattening filter which is designed to overcompensate near the urface in order to obtain flat isodose curves at depth (10 cm)
  • 46. 3. The penumbra region: The dose rate decreases rapidly as a function of lateral distance from the beam axis. The width of geometric penumbra depends on source size(d), distance from the source (SD), and source-to-collimator diaphragm distance (SDD). The falloff near the edge of the beam is caused by geometric penumbra and reduced side scatter This falloff can be described using physical (total) penumbra: Physical penumbra is defined as lateral distance between two specified isodose curves at a specified dept h (e.g., lateral distance between 90% and 20% isodose lines at the depth of Dmax).
  • 47. 3.1.2.1 Properties of Isodose Charts 4. Outside the geometric limits of the beam and the penumbra • The dose variation is the result of side scatter from the field and both leakage and scatter from the collimator system.  The dose in this region is generally low results from radiation transmitted through the collimator and head shielding
  • 48. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 48 3.1.2.1 Properties of Isodose Charts
  • 49. 3.1.2.2 Measurement of Isodose Curves 4928/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 50. 3.1.2.2 Measurement of Isodose Curves
  • 51. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 51 3.1.2.2 Measurement of Isodose Curves
  • 52. 1. Measurement in water phantom Ionization chamberWater phantom
  • 53. IC position in water phantom
  • 54. 2. Measurement in solid water phantom
  • 55. 2. Measurement in solid water phantom
  • 56. 3.1.2.3 Parameters Affecting Isodose Curves 3.1.2.3.1 Beam Quality 3.1.2.3.2 Source Size and SSD 3.1.2.3.3 Collimation and Flattening Filter 3.1.2.3.4 Field Size 3.1.2.3.5 Wedge Filter 5628/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 57. 3.2.4 Parameters Affecting Isodose Curves Beam quality Source Size and SSD Collimation and flattening filter Field size Parameters affecting the single- beam isodose
  • 58. 3.1.2.3.1 Beam Quality 5828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 59. 3.1.2.3.1 Beam Quality Depth The depth of a given isodose curve increases with beam quality Beam energy Beam energy also influences isodose curve shape near the field border Lateral scatter Greater lateral scatter associated with lower-energy beams causes the isodose curves outside the field to bulge out In other words, the absorbed dose in the medium outside the primary beam is greater for low energy beams than for those of higher energy
  • 60. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 60 3.1.2.3.1 Beam Quality3.1.2.3.1 Beam Quality
  • 61. 4 MV 6MV 15MV
  • 62. 3.1.2.3.1 Beam Quality One disadvantage of the orthovoltage beams is increased scattered dose to tissue outside the treatment region. For megavoltage beams, the scatter outside the field is minimized as a result of forward scattering and becomes more a function of collimation than energy.
  • 63. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 63 3.1.2.3.1 Beam Quality
  • 64. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 64 3.1.2.3.1 Beam Quality
  • 65. 3.1.2.3.2 SSD 6528/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 66. 3.1.2.3.2 SSD SSD affect the isodose curves by the geometric penumbra. The SSD affects the PDD and the depth of the isodose curves. The dose variation across the field border is a complex function of geometric penumbra, lateral scatter, and collimation. SSD
  • 68. 3.1.2.4.3 Collimation and Flattening Filter 6828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 69. 3.1.2.3.3 Collimation and Flattening Filter Collimation If no filter used, the result is conical field distribution Makes field uniform across whole width, with the same intensity Flattening filter thickest in the middle and tapers to sides Filter causes change in beam quality Different photon spectrum at edge than in middle, harder in middle 1. Flattening filter 2.Collimator blocks
  • 70. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 70 3.1.2.3.3 Collimation and Flattening Filter
  • 71. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 71 3.1.2.3.3 Collimation and Flattening Filter
  • 72. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 72 3.1.2.3.4 Field Size
  • 73. • Field size is the width and length of the radiation beam at SSD or SAD • Field size at any depth is usually defined by the 50% width of the profile at that depth Part VIII.3.5 Determination of Dose to a Patient-I Slide 73 3.1.2.3.4 Field Size
  • 74. Field size (dosimetrical): lateral distance between the 50% isodose lines at a reference depth Field size (geometrical) : field defining light is made to coincide with 50% isodose lines of the radiation beam projected on a plane perpendicular to the beam axis and at standard SSD or SAD 3.1.2.3.4 Field Size
  • 75. 3.1.2.3.4 Field Size Field size < 6 cm TPS should be mandatory for small field size. Bell shape Relative large penumbra region
  • 76. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 76
  • 77. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 77
  • 78. 3.1.2.3.5 Wedge Filter 7828/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak
  • 80. Wedge Filters Figure A:normalized to Dmax, B:normalized to Dmax without wedge 3.1.2.3.5 Wedge Filter the degree of tilt being dependent on the slope of the wedge filter Wedge shaped: cause the isodose curve tilted from their normal position The most commonly used beam- modifying device
  • 81. Mounted on a (transparent plastic) tray Arranged at a distance of at least 15 cm from the skin surface, avoid destroying the skin-sparing effect of the megavoltage beam 3.1.2.3.5 Wedge Filter
  • 82. 10 cm wedge angle 3.1.2.3.5 Wedge Filter Wedge angle the angle through which an isodose curve is titled at the central ray of a beam at a specified depth (10 cm) The angle between the isodose curve and the normal to the central axis The angle of isodose tilt decreases with increasing depth in the phantom due to presence of scattered radiation
  • 83. 28/1/2018 Dr. Nik Noor Ashikin Bt Nik Ab Razak 83 3.1.2.3.5 Wedge Filter
  • 84. 3.1.2.3.5 Wedge Filter WEDGE FACTOR The ratio of doses with and without the wedge, at a point in phantom along the central axis of the beam Measured at a suitable depth beyond dmax (5 to 10 cm) either in air or in phantom Sometimes incorporated into the isodose curves The presence of a wedge filter decreases the output of the machine.
  • 85. Hinge angle,φ It is the angle between central axes of two beams passing through the wedge Relationship b/w φ & θ Wedge angle,θ= 90 – φ/2 3.1.2.3.5 Wedge Filter