Dose Measurement of Photon
Beam
Dr. Saira Siddique
MSc. RMO fellow
4/5/2022 Photon Dosimetry 1
Presentation Layout
• Introduction
• Phantom
• Depth Dose Distribution
• PDD
• Definition
• It’s Dependence
• TAR & BSF
• TMR & TPR
4/5/2022 Photon Dosimetry 2
Introduction
It is seldom possible to measure dose distribution directly in patients
treated with radiation.
Data on dose distribution are almost entirely derived from measurements
in phantoms—tissue-equivalent materials, usually large enough in
volume to provide fullscatter conditions for the given beam.
These basic data are used in a dose calculation system devised to
predict dose distribution in an actual patient. Various dosimetric quantities
and methodologies have been devised to facilitate dose calculation in
clinical situations.
4/5/2022 Photon Dosimetry 3
The methods based on quantities such as percent depth
doses (PDDs), tissue–air ratios (TARs), and scatter–air
ratios (SARs) have been used traditionally for dose
calculation involving low energy beams (up to 60Co),
which were usually calibrated in terms of exposure rate in
air or dose rate in free space.
Current methods of dose calculation use tissue–phantom
ratios (TPRs) or tissue–maximum ratios (TMRs), which
are better suited for higher energy beams and involve
measurements in phantom rather than in air.
4/5/2022 Photon Dosimetry 4
Low energy – Exposure Rate/ Dose rate in FREE AIR  PDD, TAR, SAR
Higher energy – in phantom  TPR, TMR
TMT Various Phantom materials
4/5/2022 Photon Dosimetry 5
Basic dose distribution data are usually
measured in a WATER PHANTOM b/c:
• it closely approximates the radiation absorption and scattering properties of
muscle and other soft tissues.
• water is universally available with reproducible radiation properties.
Ideally, for a given material to be tissue or water equivalent, it must have:
the same effective atomic number,
number of electrons per gram,
and mass density.
However, in MV photon beam (clinical range)  the most predominant mode
is Compton effect so the necessary condition for water equivalence for such
beams is to have the same electron density (number of electrons per cubic
centimeter) as that of water.
4/5/2022 Photon Dosimetry 6
For clinical dosimetry 
anthropomorphic phantoms are
frequently used.
One such commercially available
system, known as Alderson Radiation
Therapy Phantom (formerly the
Alderson Rando Phantom),
it incorporates materials to simulate
various body tissues—muscle, bone,
lung, and air cavities.
It is shaped into a human torso and is
sectioned transversely into slices for
inserting film or other dosimeters
4/5/2022 Photon Dosimetry 7
DEPTH DOSE DISTRIBUTION
As the beam is incident on a patient (or a phantom), the absorbed dose in the
patient varies with depth.
This variation depends on many conditions: beam energy, depth, field size,
distance from source, and beam collimation system.
Thus, the calculation of dose in a patient involves considerations in regard to these
parameters and others as they affect depth dose distribution.
An essential step in the dose calculation system is to establish depth dose variation
along the central axis of the beam.
A number of quantities, are defined for this purpose, major among these being
PDDs, TARs, TPRs, and TMRs.
These quantities are usually derived from measurements made in water phantoms
using small ionization chambers. Although other dosimetry systems such as
thermoluminescent dosimeters (TLDs), diodes, and film are occasionally used, ion
chambers are preferred because of their better precision and smaller energy
dependence.
4/5/2022 Photon Dosimetry 8
PERCENTAGE DEPTH DOSE
One way of characterizing the central axis dose distribution is to
normalize dose at depth with respect to dose at a reference
depth.
The quantity percentage depth dose is defined as
“the quotient, expressed as a percentage, of the absorbed
dose at any depth d to the absorbed dose at a reference
depth d0 , along the central axis of the beam.”
• For orthovoltage (up to 400 kV) & lower-energy x-rays, the
reference depth is usually the surface (d0 = 0).
• For higher energies, the reference depth is usually taken at the
position of the peak absorbed dose (d0 = dmax)
4/5/2022 Photon Dosimetry 9
PDD
4/5/2022 Photon Dosimetry 10
PDD depends on:
• BEAM QUALITY AND DEPTH
• FIELD SIZE AND SHAPE
• SSD
4/5/2022 Photon Dosimetry 11
PDD _ Dependence on Beam QUALITY & DEPTH
PDD (beyond the depth of maximum dose) decreases with
depth and increases with beam energy.
Higher-energy beams have greater penetrating power and
thus deliver a higher PDD at a given depth
the beam quality affects the PDD by virtue of the average
attenuation coefficient . As it decreases, the beam becomes
more penetrating, resulting in a higher PDD at any given
depth beyond the buildup region.
4/5/2022 Photon Dosimetry 12
4/5/2022 Photon Dosimetry 13
Initial Dose Buildup
PDD decreases with depth beyond the depth of maximum dose. However, there
is an initial buildup of dose that becomes more and more pronounced as the
energy is increased. In the case of the orthovoltage or lower-energy x-rays, the
dose builds up to a maximum on or very close to the surface. But for higher-
energy beams, the point of maximum dose lies deeper into the tissue or
phantom.
“The region between the surface and the point of maximum dose is called
the dose buildup region.”
dose buildup effect (higher-energy)  the skin-sparing effect.
For MV beams (cobalt-60 & higher energies), the surface dose is much smaller
than the Dmax . This offers a distinct advantage over the lower-energy beams
for which the Dmax occurs at or very close to the skin surface.
Thus, in MV beams, higher doses can be delivered to deep-seated tumors
without exceeding the tolerance of the skin. This, of course, is possible because
of both the higher PDD at the tumor and the lower surface dose at the skin.
4/5/2022 Photon Dosimetry 14
4/5/2022 Photon Dosimetry 15
PDD _ EFFECT OF FIELD SIZE AND SHAPE
Field size may be specified either
geometrically
dosimetrically.
The geometric field size: “the projection, on a plane
perpendicular to the beam axis, of the distal end of the
collimator as seen from the front center of the source”
The dosimetric, or the physical, field size is the distance
intercepted by a given isodose curve (usually 50% isodose)
on a plane perpendicular to the beam axis at a stated
distance from the source.
4/5/2022 Photon Dosimetry 16
PDD _ EFFECT OF FIELD SIZE AND SHAPE
For a sufficiently small field, the depth dose at a point is primarily the result
of the primary radiation (the photons that have traversed the overlying
medium without interacting).
The contribution of the scattered photons to the depth dose in this case is
negligibly small and may be assumed to be 0.
But as the field size is increased, the contribution of the scattered
photons to the absorbed dose increases because the volume that can
scatter radiation gets larger with field size. Also, this increase in scattered
dose will be greater at greater depths.
The increase in PDD caused by increase in field size depends on beam
quality. Because the scattering probability or scatter cross section decreases
with increase in energy and the higher-energy photons are scattered more
predominantly in the forward direction, the field size dependence of PDD is
less pronounced for the higher-energy than for the lower-energy beams.
4/5/2022 Photon Dosimetry 17
PDD _ DEPENDENCE ON SSD
Photon fluence emitted by a point source of radiation varies inversely as a
square of the distance from the source. The source is considered as a point
source (at large SSDs – 80cm). Thus, the exposure rate or “dose rate in free
space” from such a source varies inversely as the square of the distance
(assuming it as a Primary beam – no scatter).
PDD increases with SSD as a result of the inverse square law.
Although the actual dose rate at a point decreases with increase in distance
from the source, the PDD, which is a relative dose with respect to a
reference point, increases with SSD.
4/5/2022 Photon Dosimetry 18
Relative dose rate from a point
source is plotted as a function of
distance from the source, following
the inverse square law.
The plot shows that:
the drop in dose rate
between two points is much
greater at smaller distances
from the source than at large
distances.
This means that the PDD, which
represents depth dose relative to a
reference point, decreases more
rapidly near the source than far away
from the source.
4/5/2022 Photon Dosimetry 19
TISSUE–AIR RATIO
TAR may be defined as:
“the ratio of the dose at a given point in the phantom (Dd) to the
dose in free space at the same point (Dfs).”
For a given quality beam, TAR depends on
 depth d
 field size at that depth.
Because the PDD depends on the SSD, the SSD correction to the
PDD will have to be applied to correct for the varying SSD—a
cumbersome procedure to apply routinely in clinical practice. TAR—
has been defined to remove the SSD dependence.
concept of TAR has been refined to facilitate calculations not only for
rotation therapy but also for stationary isocentric techniques as well
as irregular fields. (specifically for rotation therapy calculation in past)
4/5/2022 Photon Dosimetry 20
4/5/2022 Photon Dosimetry 21
TAR
EFFECT OF DISTANCE:
The most important properties attributed to TAR is that it is
independent of the distance from the source. TAR for the primary
beam is only a function of depth, not of SSD.
VARIATION WITH ENERGY, DEPTH, AND FIELD SIZE
TAR varies with energy, depth, and field size very much like the PDD.
For the megavoltage beams, the TAR builds up to a maximum at the
depth of maximum dose (dm ) and then decreases with depth.
4/5/2022 Photon Dosimetry 22
BSF
The term backscatter factor or peak scatter factor (PSF) is simply
the TAR at the reference depth of maximum dose on central axis
of the beam. It may be defined as:
“the ratio of the dose on central axis at the reference depth of
maximum dose to the dose at the same point in free space.”
The BSF, like the TAR, is independent of distance from the source
and depends only on the beam quality and field size.
4/5/2022 Photon Dosimetry 23
TMR & TPR
4/5/2022 Photon Dosimetry 24
TMR & TPR
TPR (Tissue Phantom Ratio) is defined as the ratio of the dose
rate at a given depth in phantom to the dose rate at the
same source-point distance, but at a reference depth.
If dmax is adopted as a fixed reference depth, the quantity TPR
gives rise to the TMR.
Thus, TMR is a special case of TPR
TMR (Tissue Maximum Ratio) is defined as the ratio of the
dose rate at a given point in phantom to the dose rate at the
same source-point distance and at the reference depth of
maximum dose.
4/5/2022 Photon Dosimetry 25
THANK YOU!
4/5/2022 Photon Dosimetry 26

More Related Content

What's hot

Cavity theory-Radiation physics
Cavity theory-Radiation physicsCavity theory-Radiation physics
Cavity theory-Radiation physicsKathiravan E M
 
proton therapy
proton therapyproton therapy
proton therapydinadawi
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetryfondas vakalis
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetrySabari Kumar
 
Brachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .RBrachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .Rraazvarma
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsDrAyush Garg
 
Particle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyParticle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyAswathi c p
 
ICRU 89 summary & beyond converted
ICRU 89 summary & beyond convertedICRU 89 summary & beyond converted
ICRU 89 summary & beyond convertedDr. Abhishek Basu
 
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in RadiotherapyDosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in RadiotherapyAYMAN G. STOHY
 
Linear quadratic model ppt
Linear quadratic model pptLinear quadratic model ppt
Linear quadratic model pptNilesh Kucha
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62Dhiman Das
 

What's hot (20)

Cavity theory-Radiation physics
Cavity theory-Radiation physicsCavity theory-Radiation physics
Cavity theory-Radiation physics
 
Icru 58.
Icru 58.Icru 58.
Icru 58.
 
Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
proton therapy
proton therapyproton therapy
proton therapy
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetry
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetry
 
Brachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .RBrachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .R
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 Radiophysics
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Dose volume histogram
Dose volume histogramDose volume histogram
Dose volume histogram
 
Mlc
MlcMlc
Mlc
 
Particle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyParticle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapy
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
ICRU 89 summary & beyond converted
ICRU 89 summary & beyond convertedICRU 89 summary & beyond converted
ICRU 89 summary & beyond converted
 
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in RadiotherapyDosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
 
Linear quadratic model ppt
Linear quadratic model pptLinear quadratic model ppt
Linear quadratic model ppt
 
Co60 vs Ir192
Co60 vs Ir192Co60 vs Ir192
Co60 vs Ir192
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Electron beam radiotherapy
Electron beam radiotherapyElectron beam radiotherapy
Electron beam radiotherapy
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62
 

Similar to Photon dosimetry 31 01-2022

Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxDr. Dheeraj Kumar
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYPaul George
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapyjyotimannath
 
Gowtham's 4th phsysics photons
Gowtham's 4th phsysics   photonsGowtham's 4th phsysics   photons
Gowtham's 4th phsysics photonsGowtham Manimaran
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdfNASERALHAQ
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapyajjitchandran
 
Teletherapy dosage data ii
Teletherapy dosage data  iiTeletherapy dosage data  ii
Teletherapy dosage data iiSneha George
 
4. Isodose Distribution In Radiation Oncology .pptx
4. Isodose Distribution In Radiation Oncology .pptx4. Isodose Distribution In Radiation Oncology .pptx
4. Isodose Distribution In Radiation Oncology .pptxAbhishekMewara2
 
Electron beam radiotherapy
Electron beam radiotherapyElectron beam radiotherapy
Electron beam radiotherapyankitapandey63
 
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiationsehlawi
 
Laser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thLaser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thGazy Khatmi
 
Treatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set UpTreatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set Upfondas vakalis
 

Similar to Photon dosimetry 31 01-2022 (20)

Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Icru 71 electrons
Icru 71 electrons Icru 71 electrons
Icru 71 electrons
 
RADIOTHERAPY CALCULATION
RADIOTHERAPY CALCULATIONRADIOTHERAPY CALCULATION
RADIOTHERAPY CALCULATION
 
Isodose lines
Isodose linesIsodose lines
Isodose lines
 
Gowtham's 4th phsysics photons
Gowtham's 4th phsysics   photonsGowtham's 4th phsysics   photons
Gowtham's 4th phsysics photons
 
RON BEAM THEELECT
RON BEAM THEELECTRON BEAM THEELECT
RON BEAM THEELECT
 
Lecture (5)
Lecture (5)Lecture (5)
Lecture (5)
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Gowtham's 5th electrons
Gowtham's 5th   electronsGowtham's 5th   electrons
Gowtham's 5th electrons
 
Teletherapy dosage data ii
Teletherapy dosage data  iiTeletherapy dosage data  ii
Teletherapy dosage data ii
 
4. Isodose Distribution In Radiation Oncology .pptx
4. Isodose Distribution In Radiation Oncology .pptx4. Isodose Distribution In Radiation Oncology .pptx
4. Isodose Distribution In Radiation Oncology .pptx
 
Electron beam radiotherapy
Electron beam radiotherapyElectron beam radiotherapy
Electron beam radiotherapy
 
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
 
Tseb
TsebTseb
Tseb
 
Laser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thLaser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 th
 
Treatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set UpTreatment Planning Ii Patient Data, Corrections, And Set Up
Treatment Planning Ii Patient Data, Corrections, And Set Up
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 

Recently uploaded

Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCR
Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCRCall Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCR
Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCRlizamodels9
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationColumbia Weather Systems
 
Analytical Profile of Coleus Forskohlii | Forskolin .pptx
Analytical Profile of Coleus Forskohlii | Forskolin .pptxAnalytical Profile of Coleus Forskohlii | Forskolin .pptx
Analytical Profile of Coleus Forskohlii | Forskolin .pptxSwapnil Therkar
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxBerniceCayabyab1
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxFarihaAbdulRasheed
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPirithiRaju
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPirithiRaju
 
Base editing, prime editing, Cas13 & RNA editing and organelle base editing
Base editing, prime editing, Cas13 & RNA editing and organelle base editingBase editing, prime editing, Cas13 & RNA editing and organelle base editing
Base editing, prime editing, Cas13 & RNA editing and organelle base editingNetHelix
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024AyushiRastogi48
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.PraveenaKalaiselvan1
 
Pests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPirithiRaju
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxpriyankatabhane
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentationtahreemzahra82
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringPrajakta Shinde
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologycaarthichand2003
 
OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024innovationoecd
 
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.aasikanpl
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensorsonawaneprad
 
FREE NURSING BUNDLE FOR NURSES.PDF by na
FREE NURSING BUNDLE FOR NURSES.PDF by naFREE NURSING BUNDLE FOR NURSES.PDF by na
FREE NURSING BUNDLE FOR NURSES.PDF by naJASISJULIANOELYNV
 
Is RISC-V ready for HPC workload? Maybe?
Is RISC-V ready for HPC workload? Maybe?Is RISC-V ready for HPC workload? Maybe?
Is RISC-V ready for HPC workload? Maybe?Patrick Diehl
 

Recently uploaded (20)

Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCR
Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCRCall Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCR
Call Girls In Nihal Vihar Delhi ❤️8860477959 Looking Escorts In 24/7 Delhi NCR
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather Station
 
Analytical Profile of Coleus Forskohlii | Forskolin .pptx
Analytical Profile of Coleus Forskohlii | Forskolin .pptxAnalytical Profile of Coleus Forskohlii | Forskolin .pptx
Analytical Profile of Coleus Forskohlii | Forskolin .pptx
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
 
Base editing, prime editing, Cas13 & RNA editing and organelle base editing
Base editing, prime editing, Cas13 & RNA editing and organelle base editingBase editing, prime editing, Cas13 & RNA editing and organelle base editing
Base editing, prime editing, Cas13 & RNA editing and organelle base editing
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
 
Pests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdf
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptx
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentation
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical Engineering
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technology
 
OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024
 
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensor
 
FREE NURSING BUNDLE FOR NURSES.PDF by na
FREE NURSING BUNDLE FOR NURSES.PDF by naFREE NURSING BUNDLE FOR NURSES.PDF by na
FREE NURSING BUNDLE FOR NURSES.PDF by na
 
Is RISC-V ready for HPC workload? Maybe?
Is RISC-V ready for HPC workload? Maybe?Is RISC-V ready for HPC workload? Maybe?
Is RISC-V ready for HPC workload? Maybe?
 

Photon dosimetry 31 01-2022

  • 1. Dose Measurement of Photon Beam Dr. Saira Siddique MSc. RMO fellow 4/5/2022 Photon Dosimetry 1
  • 2. Presentation Layout • Introduction • Phantom • Depth Dose Distribution • PDD • Definition • It’s Dependence • TAR & BSF • TMR & TPR 4/5/2022 Photon Dosimetry 2
  • 3. Introduction It is seldom possible to measure dose distribution directly in patients treated with radiation. Data on dose distribution are almost entirely derived from measurements in phantoms—tissue-equivalent materials, usually large enough in volume to provide fullscatter conditions for the given beam. These basic data are used in a dose calculation system devised to predict dose distribution in an actual patient. Various dosimetric quantities and methodologies have been devised to facilitate dose calculation in clinical situations. 4/5/2022 Photon Dosimetry 3
  • 4. The methods based on quantities such as percent depth doses (PDDs), tissue–air ratios (TARs), and scatter–air ratios (SARs) have been used traditionally for dose calculation involving low energy beams (up to 60Co), which were usually calibrated in terms of exposure rate in air or dose rate in free space. Current methods of dose calculation use tissue–phantom ratios (TPRs) or tissue–maximum ratios (TMRs), which are better suited for higher energy beams and involve measurements in phantom rather than in air. 4/5/2022 Photon Dosimetry 4 Low energy – Exposure Rate/ Dose rate in FREE AIR  PDD, TAR, SAR Higher energy – in phantom  TPR, TMR
  • 5. TMT Various Phantom materials 4/5/2022 Photon Dosimetry 5
  • 6. Basic dose distribution data are usually measured in a WATER PHANTOM b/c: • it closely approximates the radiation absorption and scattering properties of muscle and other soft tissues. • water is universally available with reproducible radiation properties. Ideally, for a given material to be tissue or water equivalent, it must have: the same effective atomic number, number of electrons per gram, and mass density. However, in MV photon beam (clinical range)  the most predominant mode is Compton effect so the necessary condition for water equivalence for such beams is to have the same electron density (number of electrons per cubic centimeter) as that of water. 4/5/2022 Photon Dosimetry 6
  • 7. For clinical dosimetry  anthropomorphic phantoms are frequently used. One such commercially available system, known as Alderson Radiation Therapy Phantom (formerly the Alderson Rando Phantom), it incorporates materials to simulate various body tissues—muscle, bone, lung, and air cavities. It is shaped into a human torso and is sectioned transversely into slices for inserting film or other dosimeters 4/5/2022 Photon Dosimetry 7
  • 8. DEPTH DOSE DISTRIBUTION As the beam is incident on a patient (or a phantom), the absorbed dose in the patient varies with depth. This variation depends on many conditions: beam energy, depth, field size, distance from source, and beam collimation system. Thus, the calculation of dose in a patient involves considerations in regard to these parameters and others as they affect depth dose distribution. An essential step in the dose calculation system is to establish depth dose variation along the central axis of the beam. A number of quantities, are defined for this purpose, major among these being PDDs, TARs, TPRs, and TMRs. These quantities are usually derived from measurements made in water phantoms using small ionization chambers. Although other dosimetry systems such as thermoluminescent dosimeters (TLDs), diodes, and film are occasionally used, ion chambers are preferred because of their better precision and smaller energy dependence. 4/5/2022 Photon Dosimetry 8
  • 9. PERCENTAGE DEPTH DOSE One way of characterizing the central axis dose distribution is to normalize dose at depth with respect to dose at a reference depth. The quantity percentage depth dose is defined as “the quotient, expressed as a percentage, of the absorbed dose at any depth d to the absorbed dose at a reference depth d0 , along the central axis of the beam.” • For orthovoltage (up to 400 kV) & lower-energy x-rays, the reference depth is usually the surface (d0 = 0). • For higher energies, the reference depth is usually taken at the position of the peak absorbed dose (d0 = dmax) 4/5/2022 Photon Dosimetry 9
  • 11. PDD depends on: • BEAM QUALITY AND DEPTH • FIELD SIZE AND SHAPE • SSD 4/5/2022 Photon Dosimetry 11
  • 12. PDD _ Dependence on Beam QUALITY & DEPTH PDD (beyond the depth of maximum dose) decreases with depth and increases with beam energy. Higher-energy beams have greater penetrating power and thus deliver a higher PDD at a given depth the beam quality affects the PDD by virtue of the average attenuation coefficient . As it decreases, the beam becomes more penetrating, resulting in a higher PDD at any given depth beyond the buildup region. 4/5/2022 Photon Dosimetry 12
  • 14. Initial Dose Buildup PDD decreases with depth beyond the depth of maximum dose. However, there is an initial buildup of dose that becomes more and more pronounced as the energy is increased. In the case of the orthovoltage or lower-energy x-rays, the dose builds up to a maximum on or very close to the surface. But for higher- energy beams, the point of maximum dose lies deeper into the tissue or phantom. “The region between the surface and the point of maximum dose is called the dose buildup region.” dose buildup effect (higher-energy)  the skin-sparing effect. For MV beams (cobalt-60 & higher energies), the surface dose is much smaller than the Dmax . This offers a distinct advantage over the lower-energy beams for which the Dmax occurs at or very close to the skin surface. Thus, in MV beams, higher doses can be delivered to deep-seated tumors without exceeding the tolerance of the skin. This, of course, is possible because of both the higher PDD at the tumor and the lower surface dose at the skin. 4/5/2022 Photon Dosimetry 14
  • 16. PDD _ EFFECT OF FIELD SIZE AND SHAPE Field size may be specified either geometrically dosimetrically. The geometric field size: “the projection, on a plane perpendicular to the beam axis, of the distal end of the collimator as seen from the front center of the source” The dosimetric, or the physical, field size is the distance intercepted by a given isodose curve (usually 50% isodose) on a plane perpendicular to the beam axis at a stated distance from the source. 4/5/2022 Photon Dosimetry 16
  • 17. PDD _ EFFECT OF FIELD SIZE AND SHAPE For a sufficiently small field, the depth dose at a point is primarily the result of the primary radiation (the photons that have traversed the overlying medium without interacting). The contribution of the scattered photons to the depth dose in this case is negligibly small and may be assumed to be 0. But as the field size is increased, the contribution of the scattered photons to the absorbed dose increases because the volume that can scatter radiation gets larger with field size. Also, this increase in scattered dose will be greater at greater depths. The increase in PDD caused by increase in field size depends on beam quality. Because the scattering probability or scatter cross section decreases with increase in energy and the higher-energy photons are scattered more predominantly in the forward direction, the field size dependence of PDD is less pronounced for the higher-energy than for the lower-energy beams. 4/5/2022 Photon Dosimetry 17
  • 18. PDD _ DEPENDENCE ON SSD Photon fluence emitted by a point source of radiation varies inversely as a square of the distance from the source. The source is considered as a point source (at large SSDs – 80cm). Thus, the exposure rate or “dose rate in free space” from such a source varies inversely as the square of the distance (assuming it as a Primary beam – no scatter). PDD increases with SSD as a result of the inverse square law. Although the actual dose rate at a point decreases with increase in distance from the source, the PDD, which is a relative dose with respect to a reference point, increases with SSD. 4/5/2022 Photon Dosimetry 18
  • 19. Relative dose rate from a point source is plotted as a function of distance from the source, following the inverse square law. The plot shows that: the drop in dose rate between two points is much greater at smaller distances from the source than at large distances. This means that the PDD, which represents depth dose relative to a reference point, decreases more rapidly near the source than far away from the source. 4/5/2022 Photon Dosimetry 19
  • 20. TISSUE–AIR RATIO TAR may be defined as: “the ratio of the dose at a given point in the phantom (Dd) to the dose in free space at the same point (Dfs).” For a given quality beam, TAR depends on  depth d  field size at that depth. Because the PDD depends on the SSD, the SSD correction to the PDD will have to be applied to correct for the varying SSD—a cumbersome procedure to apply routinely in clinical practice. TAR— has been defined to remove the SSD dependence. concept of TAR has been refined to facilitate calculations not only for rotation therapy but also for stationary isocentric techniques as well as irregular fields. (specifically for rotation therapy calculation in past) 4/5/2022 Photon Dosimetry 20
  • 22. TAR EFFECT OF DISTANCE: The most important properties attributed to TAR is that it is independent of the distance from the source. TAR for the primary beam is only a function of depth, not of SSD. VARIATION WITH ENERGY, DEPTH, AND FIELD SIZE TAR varies with energy, depth, and field size very much like the PDD. For the megavoltage beams, the TAR builds up to a maximum at the depth of maximum dose (dm ) and then decreases with depth. 4/5/2022 Photon Dosimetry 22
  • 23. BSF The term backscatter factor or peak scatter factor (PSF) is simply the TAR at the reference depth of maximum dose on central axis of the beam. It may be defined as: “the ratio of the dose on central axis at the reference depth of maximum dose to the dose at the same point in free space.” The BSF, like the TAR, is independent of distance from the source and depends only on the beam quality and field size. 4/5/2022 Photon Dosimetry 23
  • 24. TMR & TPR 4/5/2022 Photon Dosimetry 24
  • 25. TMR & TPR TPR (Tissue Phantom Ratio) is defined as the ratio of the dose rate at a given depth in phantom to the dose rate at the same source-point distance, but at a reference depth. If dmax is adopted as a fixed reference depth, the quantity TPR gives rise to the TMR. Thus, TMR is a special case of TPR TMR (Tissue Maximum Ratio) is defined as the ratio of the dose rate at a given point in phantom to the dose rate at the same source-point distance and at the reference depth of maximum dose. 4/5/2022 Photon Dosimetry 25