IOSR Journal of Applied Physics (IOSR-JAP)
e-ISSN: 2278-4861.Volume 7, Issue 1 Ver. III (Jan.-Feb. 2015), PP 27-43
www.iosrjournals.org
DOI: 10.9790/4861-07132743 www.iosrjournals.org 27 | Page
Efficacy of Use of A-Si EPID as Imaging Device in IMRT QA
Dina Abdelaziz, Wafaa Khalifa and Nader El Sherbini
Abstract: When IMRT treatment is considered to be a routine work, one should seek for an accurate, time
saving, fast and simple QA procedure. Electronic portal imaging device (EPID), especially amorphous silicon
type, is a promising tool for IMRT dosimetry. In this work a calibration of an amorphous silicon flat panel-type
imager (Elekta iViewGT, release 3.2) is done, which was incorrect (figure 3.1.) due to the EPID
renormalization to the target value (Elekta instructions). The calibration of the amorphous silicon flat panel-
type imager (Elekta iViewGT) was repeated after considering the Pixel Scaling Factor (PSF), which gave the
correct response (figure 3.2.). By using Omni Pro software the EPID images are converted into doses and a
correction for the dose reading from these images is done to obtain the same results as that measured at the
isocenter in 20 cm PMMA slab phantom. A correction for the EPID profiles for different field sizes is done to be
like those measured in the full scatter water phantom. A pre-treatment verification for a clinical step & shoot
IMRT plan is done for a prostate cancer case generated by Xio (version 4.2.2) treatment planning system (TPS).
The pre-treatment verification is done by EPID and a comparison in dose distribution against TPS and Films
are performed. This study show that the combination between iViewGT and Omni Pro software with the
corrections provided in the present study gave an accurate method to verify the dose of IMRT fields in two
dimensions inside a homogenous slab phantom. The results indicated also that the EPID (Elekta iViewGT) is an
accurate and can replace film for field-by field pretreatment verification of IMRT inside a phantom.
I. Introduction
The challenge of external beam radiotherapy for cancer treatment is to irradiate the tumor with a
desired dose, while the surrounding healthy tissue suffers as little as possible radiation damage.
Due to increase of complexity in modern radiotherapy techniques the needs for checking accuracy of
the delivered dose to patient has been also increased, either pretreatment or in vivo. This reflects the importance
of quality assurance (QA) as a step in radiation therapy treatment.
Quality assurance procedures of a patient treatment can be divided into two main parts:
Quality control of the technical aspects of the linear accelerator (Linac) itself includes verification of:
Dose output, the transferred parameter between treatment planning system (TPS) to linac (collimator settings,
the direction of wedges and multileaf collimator (MLC) shape) and Isodose check.
Verification of the actual patient treatment, including verification of both patient setup and delivery of
correct dose.
Ionization chamber is the predominant tool used to check the dose accuracy at certain points, while the
dose distribution of certain plan could be measured using different tools, e. g. Films, 2D diode array. The
increasing of using IMRT to deliver conformal radiation treatment has prompted the search for a faster and more
cost effective quality assurance (QA) system. The standard technique relies on the use of film for two-
dimensional dose distribution verification. Although film is considered the standard and most widely used for
this purpose, the procedures involved are relatively lengthy, labor intensive and costly for multiple field IMRT
verification. (Lim S. 2008).
QA of intensity modulated radiation therapy (IMRT) is a sophisticated and time consuming task,
where one should check not only the delivered dose and dose distribution but also need to check the field
segments shape, number and machine capability to deliver IMRT plan. When IMRT treatment is considered to
be a routine work, one should seek for an accurate, time saving, fast and simple QA procedure. Electronic portal
imaging device (EPID), especially amorphous silicon type, is a promising tool for IMRT dosimetry. A
commercial a-Si EPID has been studied by Talamonti C., et al, (2006) to investigate its potential in the field of
pretreatment verifications of step and shoot, IMRT for 6 MV photon beams.
An a-Si EPID has been investigated by Budgell G., et al, (2005) to determine its usefulness and
efficiency for performing linear accelerator quality control checks specific to step and shoot IMRT. Several
dosimetric parameters were measured using the EPID: dose linearity and segment to segment reproducibility of
low dose segments, and delivery accuracy of fractions of monitor units. Xu C., et al, (2002) have developed an
automated image based method for MLC positioning verification using images acquired from an EPID. Yeo I.
(2007) developed a method based on dose-response function, which verifies the beamlet intensity in IMRT from
dose image in EPID and reconstructs dose in a patient. Greer P., et al, (2007) determined the difference in
response between an a-Si EPID to the open and MLC transmitted beam components of IMRT beams. Ansbacher
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 28 | Page
W. (2006) developed a method for rapid evaluation of IMRT plans, using portal images for reconstruction of the
dose delivered to a virtual 3D phantom.
II. Materials And Methods
Elekta Precise (Crawley, UK) linear accelerator, which support IMRT deliver, was used in this study
(figure 2.1a). This linear accelerator (LINAC) produces photon energies of 6 and 15MV, also it produces
electrons with energies of 6, 8, 10, 12 and 15 Mev. Only 6 MV photon beam is used in this study. The LINAC is
equipped with MLC. The MLC consists of 40 leaf pairs with 1 cm leaf width at the isocenter.The type of EPID
used throughout this study is an amorphous silicon flat panel-type imager (Elekta iViewGT, release 3.2).The
detector panel is a PerkinElmer Amorphous Silicon (a-Si) detector, and provides a resolution of 1024 × 1024
16-bit pixel images, with a detector panel size of 41 × 41 cm2
(approximately 26 × 26 cm2
at isocenter).Display
pixel factor information;When images are acquired by iViewGTTM
, the pixel values are re-normalized or
averaged before the image data is saved to the database. At the time of acquisition, the pixel scaling factor is
saved with the images in the database. This feature allows us to determine the original accumulated pixel value
by dividing the pixel value by the pixel scaling factor.Polymethyl methacrylate (PMMA), also known as
acrylic. Trade names are Lucite, Plexiglas or Perspex. It consists of 1 cm for each sheet of density 1.19 g/cm3
and area 30 х 30 cm2
.GAFCHROMIC® EBT2 Dosimetry Film (8x10 in size) was used. The film spatial
resolution >5,000 dpi (dots per inch) with dose range 1cGy – 10Gy (in red color channel).Omni ProTM
IMRT is
software used for complete dosimetric verification and QA of IMRT treatment cycle. This software function is
to import and compare calculated doses fromTPS planned data with measured dose distributions from films that
were exposed in the IMRT phantom. It includes 1D profile, 2D isodose profiles as well as fully automated
comparisons using analysis tools such as Gamma index method.A computerized welhÖfer WP 700 water
phantom version 3.5 was used in the present study for acquiring the beam data.XiO treatment planning system
(TPS) (from CMS Inc., USA) version 4.2.2 implemented superposition algorithm is used in this study.
Calibration of the EPID Several steps are necessary to reconstruct the dose in the phantom or patient from the
pixel values of the EPID. All measurements of this study were performed at gantry and collimator angle 0°.
A- Calibration of EPID response to different doses:EPID images of reference field size (typically 10x10
cm2
) were recorded for different MU (2, 5, 10, 15, 25, 50, 100, 150 and 200 MU).The acquired images by
iViewGTTM
are re-normalized (automatically by the iViewGTTM
) before saving process; which means that
all images will appear with the same optical density (OD) for all different doses images. Sofor each image
the pixel scaling factor (PSF) is recorded to determine the original accumulated pixel value by dividing the
pixel value by this factor
Originalaccumulatedpixelvalue=
Pixelvalue
PixelScalingFactor (PSF )
(1)
To convert these images into doses, all images were exported to Omni Pro system.The Omni Pro
software can read the Optical Density OD of the image.Each EPID image imported to Omni Pro will be
corrected for its own PSF.even after dividing each image by its corresponding PSF it will still incorrect because
of several reasons:The scattered radiation from EPID;The scatter within the phantom;The scatter from the
phantom to the EPID;The attenuation of the beam by the phantom; andThe distance from the radiation source to
the EPID plane and to the dose reconstruction plane. (Wendling M., et al, 2006).Record the Omni Pro reading
value for each EPID dose, and get the Dose correction factor (DF), which is the Omni Pro reading value for a
specific dose divided by the Actual irradiated dose at isocenter
DF =
OmniProreadingvalueforaspecificdose
Actualirradiateddose
(2)
B- Calibration of the EPID profile
In order to determine the necessary parameters for correcting the dose profile reconstruction, i.e.,
relating pixel values in the EPID images with absolute dose values in the phantom, the EPID images at different
square field sizes are recorded with PMMA slab phantom of thickness 20 cm(an average thickness to encompass
the phantom (or patients) thickness at 6 MV photon beam). For all phantom measurements an isocentric setup
[SAD (source –axis distance) setup] was used. Dose profiles for the same field sizeswere obtained using
semiflex 0.147 ion chamber in the full-scatter water phantom [Source-Skin Distance (SSD) = 88.6cm] were
measured with ion chamber at 11.4cm depth.Where 11.4 cm water is equivalent to 10 cm of Perspex. These dose
profiles will be the reference for EPID dose profile correction. The EPID images were taken and corrected by
PSF & DF. After the conversion of EPID pixel values into dose values according to the dose response relation,
the resulting image is called dose image DEPID
. The corrected portal pixel dose according to its position
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 29 | Page
regarding to central axis (CAX) is PDEPID
, which is the image after profile correction, is obtained by the
predicted equation based on the previous work by Wendling, et al, 2006 and Renner, et al, 2005.
PDij
EPID
= Dij
EPID
× e−ar
3
Where r is the distance of a pixel ij from the central axis, a is the exponential fitting constant.This constant is
fitted properly for all field sizes.
Pre-treatment verification: EPID versus filmA clinical step & shoot IMRT plan is used for a prostate cancer
treatment generated by Xio (version 4.2.2) TPS. This plan consists of seven fields, and each field had between
8-14 segments.With EPID, one image was acquired for each segment using single exposure option in the image
acquisition mode for IMRT (Pre- segment image acquisition and storage also allows verification of MLC leaf
position), where the dose construction was also done separately for each segment. All of these images were sent
to the Omni Pro and corrected according to the previous corrections.The film irradiated simultaneously with the
corresponding EPID image acquisition for each field using PMMA phantom. A film is placed at 10 cm depth in
the PMMA phantom on the Linac table and 10 cm of PMMA slabs are placed under the film to achieve the
backscatter. This simultaneous irradiationto avoid any fluctuation in doses between film and EPID irradiation
like dose, dose rate fluctuation and MLC reposition. These images are compared with the simultaneously
irradiated films and a γ evaluation was done in two dimensions.
III. Results &Discussion
3.1 EPID Response
Wendling, et al, (2006) stated that the resulting response of the a-Si EPID has been shown to be linear with
dose, although a small ghosting effect remains, which is mainly a function of the number of exposed frames and
no correction for ghosting effect was made in their study. So, in the present study the original accumulated pixel
values of the raw images data which are obtained from iViewGTTM
software, by using equation (2-1) as
described in section (2) were used.
Figure (3.1.) central axis original accumulated pixel values versus monitor units for a 10x10 cm2
field.
3.2 Pixel Scaling Factor
Table (3-1) shows the signal reading in Omni Pro software of the doses calculated at the isocenter at
10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm.
Table (3-1)the signal reading in Omni Pro software of the doses calculated at the isocenter at 10cm depth in a
20cm thick PMMA slab phantom at SSD = 90cm.
Dose mGy at the isocenter of the PMMA phantom Signal reading 100%
20 479
50 513
100 520
150 521.5
200 521.5
250 521.7
500 521.6
1000 521.6
2000 521.6
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 30 | Page
These signals readings were obtained without any correction where one can found that there is
incorrect response,so all of these images numerical values will be divided by the PSF values listed in table (3-2).
These PSF are obtained from EPID row image data.
Table (3-2) The PSF for different doses at the isocenter at 10cm depth in a 20cm thick PMMA slab
phantom at SSD = 90cm for reference field size 10x10 cm2
Dose (mGy) PSF
20 4.61207
50 1.98999
100 0.93866
150 0.67392
200 0.51005
250 0.41549
500 0.20685
1000 0.10175
2000 0.05091
The resulting signal reading (100%) that irradiated to different doses 2,5,10, 50,100 and 200 cGy at the
isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm for reference field size 10x10
cm2
after dividing all images values by the corresponding PSF using the equation (2-1), that are obtained by
Omni Pro show different response for the different doses, as indicated in table (3-3).
Table (3-3)the signal reading in Omni Pro software after dividing each image numerical values by its
own PSF using equation (2-1).
Dose mGy at the isocenter of the PMMA phantom Signal reading 100%
20 103.4
50 257.5
100 554
150 773
200 1021
250 1254
500 2519
1000 5120
2000 10230
3.3 Dose Correction Factor
Although the Omni Pro gave a linear relationship between the dose and the signal reading for the EPID
images after dividing by PSF, the Omni Pro signal reading is not the same as the incident dose values. These
results agree with those obtained by Wendling,et al, (2006).
Wendling,et al, (2006) attributes these incorrect results to that the scatter within the EPID, the scatter within the
phantom, the scatter from the phantom to the EPID and the attenuation of the beam by the phantom are not
considered. Also the distance from the radiation source to the EPID plane and to the dose reconstruction plane is
not considered.
From our results the discrepancies may also be attributed to the introduction of a proposed
sensitometric curve which is the relation that consolidates Omni Pro to read the EPID images.
Figure (3.2.) The relation between Omni Pro reading and doses.
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 31 | Page
It is clear from figure (3.2.) that the relation between Omni Pro reading and the delivered doses is a
linear relationship, and its slope = 51.26. By using equation (2-2) then the Omni Pro signal reading is the dose at
the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm.
Table (3-4) Dose correction factor (DF) factor for different doses that obtained using equation (2-2).
Dose mGy at the isocenter of the Perspex phantom DF
20 51.70
50 51.50
100 55.00
150 51.53
200 51.05
250 50.16
500 50.38
1000 51.20
2000 51.15
Average value of DF 51.51
It is obvious from table (3-4) that the average DF factor is nearly the same as the slope of linear
relationship of figure (3.2.), this slope = 51.26.
So the slope value of the linear relationship between the dose and the signal reading for the EPID
images is used as DF value for different doses.
In this case the signal reading in Omni Pro is the same as that irradiated at the isocenter at 10cm depth
in a 20cm thick PMMA slab phantom at SSD = 90cm.
Figure (3.3.) shows the two dimensional dose distributions for the EPID images after PSF and DF correction.
(a) 2CGy (b) 5CGy
(c) 10CGy (d) 50CGy
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 32 | Page
(e)100CGy (f) 200CGy
Figure (3.3.) Two dimensional dose distributions for the EPID images for 10x10 cm2
of 6MV photon beam
after PSF & DF correction
3.4 Dose correction for different field size
Table (3-5) shows the PSF for different field sizes with the same delivered doses (500 mGy).
Although all the field sizes were irradiated to the same dose at the isocenter at 10cm depth in a 20cm thick
PMMA slab phantom and at SSD = 90cm, there was difference in the PSF values, which is due to the different
contribution of scatter by changing the field size.
Table (3-5): PSF for different field sizes using 500 MGY.
Field size PSF
3×3 0.1688
5×5 0.20515
10×10 0.20683
15×15 0.20022
These differences in PSF values for the different field sizes receiving the same dose at the isocenter at
10cm depth in a 20cm thick PMMA slab phantom and at SSD = 90cm may be due to the dose calculation for
each field size is in full scatter conditions, that takes into consideration the collimator scatter factor and the
phantom scatter factor (calculated under full scatter situation), on the other hand EPID is not a full scatter
phantom (Varian Portal Vision instructions).
Also the EPID is known to have a lateral scatter (Wendling,et al, 2005) which is not considered in the
full scatter phantom calculation.
By trial it was found that the PSF for the reference field 10x10 cm2
gave approximately the same dose
value for different field sizes, as tabulated in table (3-6).
Table (3-6): OmniPro maximum signal reading for different field size images obtained by EPID after
dividing by the PSF of the reference field (10×10 cm2
).
Field size Signal reading 100%
3×3 2563
5×5 2562.9
10×10 2562.9
15×15 2562.9
Then after dividing the signal reading for each image of the different field sizes by the DF (51.26) the correct
dose at the isocenter, at 10cm depth in a 20cm thick PMMA slab phantom and SSD = 90cm is obtained with
maximum difference ± 0.2 % as shown by the red arrow in figure (3.4.).
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 33 | Page
(a) 3×3 cm2
(b) 5×5 cm2
Figure (3.4.)
Generally after DF correction either for different doses of the reference field size 10×10cm2
or for the
same dose at different field size the average dose values agreed with the ionization chamber measurements at
the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm with ± 0.2 % (which agree
with work done by Wendling,et al, 2006 and Renner,et al, 2005).
3.2 Profile Correction:
Profiles of square fields from EPID images were compared with those measured using an ionization
chamber that was located in a full-scatter water phantom. Figure (3.5.) shows the actual dose profiles of square
fields of different sizes obtained by EPID against the water phantom and TPS. The measurements are performed
in water phantom at 11.4 cm water (which is equivalent to 10 cm of PMMA), and source skin distance (SSD) =
88.6 cm.
(a) Field size 3×3 cm2
(b) Field size 5×5 cm2
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 34 | Page
(c) Field size 10×10 cm2
Figure (3.5.) dose profile as measured by water phantom and from TPS against the reconstructed dose
profile from the EPID images without profile correction (a) 3×3 cm2 field size. (b) 5×5 cm2field size. (c)
10×10 cm2 field size. All profiles are normalized to 100% of the measured dose.
Figure (3.5.) shows a difference between water phantom measurements and the reconstructed profiles from the
EPID images for the studied field sizes.
Table (3-7) show the maximum percent of error between profiles measured and t
he EPID at different field sizes.
Table (3-7) (% error) between EPID and the water phantom dose profiles.
Field size(cm2
) Error (%)
3×3 3%
5×5 3.5%
10×10 3%
It is clear from figure (3.5.) and table (3-7); that the profile behaviour along the central axis region by
the three methods is good, and became worse in the penumbra region and the tails. This is because within the
EPID mainly lateral X-ray scatters and optical photon scatter occurs, resulting in more flattened at the shoulders
more than in water phantom &TPS. These findings are in agreement with the results of Wendling, et al, 2006.
The fit parameter [a] in equation (2-3) correct for these behaviour where it is obtained from the water phantom
measurements.
Figure (3.6.) shows the dose profiles of square fields for different sizes obtained by EPID after correction using
equation (2-3) and water phantom and TPS.
(a) Field size 3×3 cm2
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 35 | Page
(b) Field size 5×5 cm2
(c) Field size 10×10 cm2
Figure (3.6.) dose profiles for different square field sizes as measured by water phantom at the isocenter at 11.4
cm depth which is equivalent to 10 cm Perspex, TPS and profiles reconstructed by the EPID using the Omni Pro
software at 6MV photon beam.
The fitting constant (a) in equation (2-3) was got by iteration to minimize the variation between the
measured dose by the water phantom and the reconstructed dose from the EPID, and it was found to be 0.03 for
all field sizes studied.
Figure (3.6.) shows that the agreement has improved and become in agreement with the other methods
(water phantom and TPS).
Among all field sizes the error does not exceed 0.2% between the EPID profiles and the profiles
measured by the water phantom and calculated by the TPS except for field size 10×10 cm2
where the agreement
is down to the 30 % dose on the central axis. These findings were found to be in qualitative agreement with
Wendling M., et al, 2006 who stated that the agreement was very good with the central axis, but remain worse
in the penumbra and in the tail of the profiles due to blurring of the measurements by the ionization chamber)
but in this study the most importance is for the small field sizes in order that the small segments in the IMRT
verification and the entire profiles for small field sizes agree with the measured ones.
Figure (3.7.) shows the difference between the EPID after dose and profile correction against the TPS
for different field size with the gamma evaluation (with criteria 3mm and 3%) using the Omni Pro software
and after implementing the dose and profile correction factor.
The plane dose profile comparison between the corrected EPID images and TPS is taken arbitrary on
the periphery of the field, to be sure that it correct through out the field.
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 36 | Page
(a) (b)
(c)
Figure (3.7) Workspace – Omni Pro IMRT QA program reveal the dose profile of the EPID against that of
TPS (upper) and the γ index (upper) figure (a) for 3×3 field size, (b) for 5×5 field size and (c) for 10×10 field
size. Be sure that there is no red area (γ index > 1) within the field.
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 37 | Page
3.3 Pre-treatment IMRT Verification
a) Segment by segment:
Now, instead of using ordinary open fields, an IMRT plan distribution is displayed by EPID & films in
comparison with the TPS.
The plan is done using superposition algorithm. For an IMRT case, 7 fields were used. The number of
segments in each field is illustrated in the plan report, (table 3-8).
Plan Report:
Table (3-8): number of segments for a prostate IMRT plan using Xio TPS version (4.2.2).
An IMRT plan has 7 fields
Beam number Number of segments
1 14
2 9
3 9
4 12
5 13
6 8
7 8
The correction factors (PSF, DF & profile correction equation) will be applied to each segment for every field.
EPID in comparison with Films versus TPS
Figures from (3.8.) to (3.10.) show some of the seventh field segments as an examples with their dose
distribution comparisons in terms of dose profile and theγ evaluation for EPID,as compared with film versus
TPS. The γ evaluation issued to judge about the comparison on the basis of the standard deviations in the γ
index between EPID & films using the Omni Pro software after all correction factors.
Since the dose for each segment was within 5-10 MUs which are under the response of Gafchromic films, the
doses were normalized to 50 cGy within the isocenter.
Figure (3.8, a) Workspace – Omni Pro IMRT QA program for segment 2 reveal the dose profile of the EPID
against that of TPS (upper) and the γ index (down).
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 38 | Page
Figure (3.8, b) Workspace – Omni Pro IMRT QA program for segment 2 reveal the dose profile of the film
against that of TPS (upper) and the γ index (down).
Figure (3.9, a) Workspace – Omni Pro IMRT QA program for segment 3 reveal the dose profile of the EPID
against that of TPS (upper) and the γ index (down).
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 39 | Page
Figure (3.9,b) Workspace – Omni Pro IMRT QA program for segment 3 reveal the dose profile of the film
against that of TPS (upper) and the γ index (down).
Figure (3.10, a) Workspace – Omni Pro IMRT QA program for segment 7 reveal the dose profile of the EPID
against that of TPS (upper) and the γ index (down).
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 40 | Page
Figure (3.10,b) Workspace – Omni Pro IMRT QA program for segment 7 reveal the dose profile of the film
against that of TPS (upper) and the γ index (down).
Table (3-11) show the gamma evaluation for segments as comparison between EPID & films of the
present study.
Table (3-11): the standard deviation (SD) for both EPID and films in comparison to TPS using Omni Pro
software for the segments of field number 7
Segment number EPID SD % Film SD %
1 0.24 0.31
2 0.14 0.23
3 0.15 0.29
4 0.19 0.30
5 0.23 0.28
6 0.36 0.38
7 0.23 0.24
8 0.27 0.29
Average 0.22 0.29
As seen from table (3-11); the values of the SD and the average among the 8 segments for EPID are
less than that for Gafchromic films. This may be due to that the scanning of the film is not reproducible as the
EPID, and even by using the protocol in the method of film scanning (taken from Estro booklet 9 (2008)) there
is still a deviation, where scanning is done for the film more than 5 times which is an evidence on the instability
and turning the scanner off between scanning this may be contributed to the total error because the condition
will not be the same for all films, especially that the results will be affected if the orientation of the film is
inversed. Also Alber M., et al, (2008) stated that if there is any defect in the film manufacturing it will strongly
affect the results.
If the values of standard deviation for EPID and films in table (3-11) are analyzed separately, it will be
clear that as the shaping of the segment increase with the MLC the SD relatively increases. Renner,et al, 2005
stated that the superposition algorithm is generally known to be more accurate at tissue interfaces but it does not
model either multileaf collimator leaf-end leakage or inter-leaf leakage. So increasing the number of segments
in a plan will demonstrate any inaccuracy of the planning system.
On the other hand the difference between Films and EPID can also be compared by the percentage of
pixels passing the gamma test in the region of interest for each segment of the field as demonstrated by the
histogram in figure (3.19.).
Khan F., et al, (2008) also used the percentage of pixels passing the digital gamma test to compare between
EPID and the Matrixx (in this study films used instead of MatriXX, IMRT MatriXX is a two dimensional ion
chamber array (IBA Dosimetry, Schwarzenburck,Germany))
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 41 | Page
Figure (3.11.) shows the percent of pixel passing the gamma test for EPID against films.
From figure (3.11.) it is clear that for all segments the percent of pixels passing the gamma test for
EPID is better than that for films. Also the minimum number of pixels passing the gamma test for EPID and
film were 99% and 98% respectively for segment 8. For segments 2 and 3, all pixels pass the gamma test for
both EPID and films.
Figures (3.8.) to (3.10.) show the comparison between EPID and films for samples of the segments in the
seventh field. These results prove that the EPID can replace films in IMRT QA.
b) Segment combination: Finally, it important to show the difference between reconstructed dose distribution
from EPID after correction and the calculated dose by TPS in the fields as a total with the same prescribed doses
for the planed prostate case plan. The two-dimensional midplan (isocentric-plan) dose distributions for all
segments of each field were then summed to obtain the total midplan dose of that specific field. Figure (3.12, a)
show Workspace – Omni Pro IMRT QA program for the a field from IMRT plan for the prostate case with its
prescribed doses that calculated by the TPS in the PMMA slab phantom( at 10 cm depth and SSD=90) at gantry
zero. The difference in dose distribution between the calculated dose by TPS and the reconstructed dose
distribution by EPID is also presented in the figures. Figure (3.12, b) show the histograms of the γ index,
indicating the maximum, minimum signal, the average signal and the standard deviation for each field using
EPID.
Figure (3.12, a) Workspace – Omni Pro IMRT QA program for field 7, reveal the dose distribution of the
EPID(down-left),the TPS dose distribution (upper-left),the dose profile of the EPID against that of TPS (upper-
right) and the γ index (down-right).
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 42 | Page
Figure (3.12, b) Gamma index histogram for field 7.
Table (3-12) shows the difference in dose distribution between the measured films and the
reconstructed ones from EPID in comparison with the TPS in terms of SD.
Table (3-12) gamma evaluation comparison for the fields as a total between EPID and films.
Field number EPID SD % Film SD % Number of segments
1 0.30 0.33 14
2 0.25 0.26 9
3 0.28 0.27 9
4 0.16 0.31 12
5 0.20 0.32 13
6 0.18 0.30 8
7 0.15 0.28 8
Average 0.21 0.29
From table (3-12) if the data is analyzed separately either for EPID or for films it will be clear that
increasing the number of segments in the plan, increase the value of the SD for both EPID and films that
indicate that the deviation is due to the TPS calculation accuracy for either multileaf collimator leaf-end leakage
or inter-leaf leakage.
For films and EPID the percentage of pixels passing the gamma test in the region of interest for each field as are
shown in figure (3.13.).
Figure (3.13.) shows the percent of pixel passing the gamma test for EPID against films for each field.
Efficacy of use of a-Si EPID as imaging device in IMRT QA
DOI: 10.9790/4861-07132743 www.iosrjournals.org 43 | Page
From figure (3.13.) it is clear that all the fields reconstructed from EPID are above 99% (for the criteria 3% dose
or 3mm distance to agreement).
Khan,et al, (2008) used 9 beam step-and-shoot IMRT plan for head phantom irradiated with 6MV X-ray and
the gamma evaluation histograms calculated for all9 beams indicates that 95% of all points pass the preset
criteria 3% dose or 3mm distance to agreement for each field.
IV. Conclusion
From this study one may conclude the following:
 The time required to perform dose reconstruction with the EPID (iViewGT, release 3.2) is probably the
same as the established method using film. Yet, EPID as QA device has advantages over film QA in that it
is linear, more reproducible, and eliminates film problems, with scanners.
 The total annual cost of imaging with film is higher than thatusing EPID concerning IMRT QA.
 EPID (iViewGT, release 3.2) can replace the ionization chamber for measuring dose for the studied field
sizes at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm after correction
for PSF and DF.
 From the study the combination between iViewGTand Omni Pro software with the used corrections can be
applied and provides an accurate method to verify the dose of IMRT fields in two dimensions inside a
homogenous slab phantom.
 The EPID is an accurate and fast alternative to film for field-by field pretreatment verification of IMRT
inside a phantom.
Generally, Michael G. Herman – AAPM Task Group 58 – Electronic Portal Imaging – 2001 shows that
the annual cost of imaging with EPID as a function of increasing use of portal imaging is cheaper than that with
films. When IMRT QA by films taken into consideration especially that the cost of QA films for IMRT is more
expensive than that for portal imaging, sure this cost will be more effective.

More Related Content

What's hot

Performance of a Novel SUV Calculation Scheme for PET Study
Performance of a Novel SUV Calculation Scheme for PET StudyPerformance of a Novel SUV Calculation Scheme for PET Study
Performance of a Novel SUV Calculation Scheme for PET StudyPawitra Masa-ah
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographyCarestream
 
Automatic target detection in
Automatic target detection inAutomatic target detection in
Automatic target detection inijistjournal
 
Recognition of optical images based on the
Recognition of optical images based on theRecognition of optical images based on the
Recognition of optical images based on theijcsa
 
IRJET- Analysis of Plant Diseases using Image Processing Method
IRJET- Analysis of Plant Diseases using Image Processing MethodIRJET- Analysis of Plant Diseases using Image Processing Method
IRJET- Analysis of Plant Diseases using Image Processing MethodIRJET Journal
 
Multimodality medical image fusion using improved contourlet transformation
Multimodality medical image fusion using improved contourlet transformationMultimodality medical image fusion using improved contourlet transformation
Multimodality medical image fusion using improved contourlet transformationIAEME Publication
 
IRJET - Lung Cancer Detection using GLCM and Convolutional Neural Network
IRJET -  	  Lung Cancer Detection using GLCM and Convolutional Neural NetworkIRJET -  	  Lung Cancer Detection using GLCM and Convolutional Neural Network
IRJET - Lung Cancer Detection using GLCM and Convolutional Neural NetworkIRJET Journal
 
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...TELKOMNIKA JOURNAL
 
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...IJEEE
 
Basic Medical Imaging Processing and Analysis
Basic Medical Imaging Processing and AnalysisBasic Medical Imaging Processing and Analysis
Basic Medical Imaging Processing and AnalysisKyla De Chavez
 
my poster presentation in the jcms2011 conference
my poster presentation in the jcms2011 conferencemy poster presentation in the jcms2011 conference
my poster presentation in the jcms2011 conferencePawitra Masa-ah
 
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...IDES Editor
 
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...IDES Editor
 
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...Journal For Research
 

What's hot (20)

P180203105108
P180203105108P180203105108
P180203105108
 
Performance of a Novel SUV Calculation Scheme for PET Study
Performance of a Novel SUV Calculation Scheme for PET StudyPerformance of a Novel SUV Calculation Scheme for PET Study
Performance of a Novel SUV Calculation Scheme for PET Study
 
[IJCT-V3I2P37] Authors: Amritpal Singh, Prithvipal Singh
[IJCT-V3I2P37] Authors: Amritpal Singh, Prithvipal Singh[IJCT-V3I2P37] Authors: Amritpal Singh, Prithvipal Singh
[IJCT-V3I2P37] Authors: Amritpal Singh, Prithvipal Singh
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
 
Automatic target detection in
Automatic target detection inAutomatic target detection in
Automatic target detection in
 
Recognition of optical images based on the
Recognition of optical images based on theRecognition of optical images based on the
Recognition of optical images based on the
 
IRJET- Analysis of Plant Diseases using Image Processing Method
IRJET- Analysis of Plant Diseases using Image Processing MethodIRJET- Analysis of Plant Diseases using Image Processing Method
IRJET- Analysis of Plant Diseases using Image Processing Method
 
Viva201393(1).pptxbaru
Viva201393(1).pptxbaruViva201393(1).pptxbaru
Viva201393(1).pptxbaru
 
34 107-1-pb
34 107-1-pb34 107-1-pb
34 107-1-pb
 
Multimodality medical image fusion using improved contourlet transformation
Multimodality medical image fusion using improved contourlet transformationMultimodality medical image fusion using improved contourlet transformation
Multimodality medical image fusion using improved contourlet transformation
 
IRJET - Lung Cancer Detection using GLCM and Convolutional Neural Network
IRJET -  	  Lung Cancer Detection using GLCM and Convolutional Neural NetworkIRJET -  	  Lung Cancer Detection using GLCM and Convolutional Neural Network
IRJET - Lung Cancer Detection using GLCM and Convolutional Neural Network
 
Session 7 interventional radiology
Session 7 interventional radiologySession 7 interventional radiology
Session 7 interventional radiology
 
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...
An Image Enhancement Approach to Achieve High Speed using Adaptive Modified B...
 
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...
A Flexible Scheme for Transmission Line Fault Identification Using Image Proc...
 
Basic Medical Imaging Processing and Analysis
Basic Medical Imaging Processing and AnalysisBasic Medical Imaging Processing and Analysis
Basic Medical Imaging Processing and Analysis
 
my poster presentation in the jcms2011 conference
my poster presentation in the jcms2011 conferencemy poster presentation in the jcms2011 conference
my poster presentation in the jcms2011 conference
 
Vol2no2 17
Vol2no2 17Vol2no2 17
Vol2no2 17
 
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...
Image De-noising and Enhancement for Salt and Pepper Noise using Genetic Algo...
 
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...
Authentication of Degraded Fingerprints Using Robust Enhancement and Matching...
 
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...
ROI BASED MEDICAL IMAGE COMPRESSION WITH AN ADVANCED APPROACH SPIHT CODING AL...
 

Viewers also liked

Auxin induced germination and plantlet regeneration via Rhizome section cultu...
Auxin induced germination and plantlet regeneration via Rhizome section cultu...Auxin induced germination and plantlet regeneration via Rhizome section cultu...
Auxin induced germination and plantlet regeneration via Rhizome section cultu...IOSR Journals
 
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...IOSR Journals
 
A Study on the Relationship between Nutrition Status and Physical Fitness of ...
A Study on the Relationship between Nutrition Status and Physical Fitness of ...A Study on the Relationship between Nutrition Status and Physical Fitness of ...
A Study on the Relationship between Nutrition Status and Physical Fitness of ...IOSR Journals
 
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...IOSR Journals
 
The Performance Analysis of a Fettling Shop Using Simulation
The Performance Analysis of a Fettling Shop Using SimulationThe Performance Analysis of a Fettling Shop Using Simulation
The Performance Analysis of a Fettling Shop Using SimulationIOSR Journals
 
Network Lifetime Analysis of Routing Protocols of Short Network in Qualnet
Network Lifetime Analysis of Routing Protocols of Short Network in QualnetNetwork Lifetime Analysis of Routing Protocols of Short Network in Qualnet
Network Lifetime Analysis of Routing Protocols of Short Network in QualnetIOSR Journals
 
Comparative Study of End Moments Regarding Application of Rotation Contributi...
Comparative Study of End Moments Regarding Application of Rotation Contributi...Comparative Study of End Moments Regarding Application of Rotation Contributi...
Comparative Study of End Moments Regarding Application of Rotation Contributi...IOSR Journals
 
Congestion Control in Manets Using Hybrid Routing Protocol
Congestion Control in Manets Using Hybrid Routing ProtocolCongestion Control in Manets Using Hybrid Routing Protocol
Congestion Control in Manets Using Hybrid Routing ProtocolIOSR Journals
 
Pollution Due To Noise from Selected Places
Pollution Due To Noise from Selected PlacesPollution Due To Noise from Selected Places
Pollution Due To Noise from Selected PlacesIOSR Journals
 
LC-MS/MS method for the quantification of carbinoxamine in human plasma
LC-MS/MS method for the quantification of carbinoxamine in human plasmaLC-MS/MS method for the quantification of carbinoxamine in human plasma
LC-MS/MS method for the quantification of carbinoxamine in human plasmaIOSR Journals
 
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...IOSR Journals
 
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...IOSR Journals
 
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAl
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAlPrediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAl
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAlIOSR Journals
 

Viewers also liked (20)

A010140106
A010140106A010140106
A010140106
 
G010124245
G010124245G010124245
G010124245
 
Auxin induced germination and plantlet regeneration via Rhizome section cultu...
Auxin induced germination and plantlet regeneration via Rhizome section cultu...Auxin induced germination and plantlet regeneration via Rhizome section cultu...
Auxin induced germination and plantlet regeneration via Rhizome section cultu...
 
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...
Modeling and Analysis for Cutting Temperature in Turning of Aluminium 6063 Us...
 
A Study on the Relationship between Nutrition Status and Physical Fitness of ...
A Study on the Relationship between Nutrition Status and Physical Fitness of ...A Study on the Relationship between Nutrition Status and Physical Fitness of ...
A Study on the Relationship between Nutrition Status and Physical Fitness of ...
 
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...
Quality Assurancein Technical Vocational Education (TVE)for Sustainable Natio...
 
The Performance Analysis of a Fettling Shop Using Simulation
The Performance Analysis of a Fettling Shop Using SimulationThe Performance Analysis of a Fettling Shop Using Simulation
The Performance Analysis of a Fettling Shop Using Simulation
 
Network Lifetime Analysis of Routing Protocols of Short Network in Qualnet
Network Lifetime Analysis of Routing Protocols of Short Network in QualnetNetwork Lifetime Analysis of Routing Protocols of Short Network in Qualnet
Network Lifetime Analysis of Routing Protocols of Short Network in Qualnet
 
F012234450
F012234450F012234450
F012234450
 
Comparative Study of End Moments Regarding Application of Rotation Contributi...
Comparative Study of End Moments Regarding Application of Rotation Contributi...Comparative Study of End Moments Regarding Application of Rotation Contributi...
Comparative Study of End Moments Regarding Application of Rotation Contributi...
 
Congestion Control in Manets Using Hybrid Routing Protocol
Congestion Control in Manets Using Hybrid Routing ProtocolCongestion Control in Manets Using Hybrid Routing Protocol
Congestion Control in Manets Using Hybrid Routing Protocol
 
J010346268
J010346268J010346268
J010346268
 
A010120103
A010120103A010120103
A010120103
 
Pollution Due To Noise from Selected Places
Pollution Due To Noise from Selected PlacesPollution Due To Noise from Selected Places
Pollution Due To Noise from Selected Places
 
LC-MS/MS method for the quantification of carbinoxamine in human plasma
LC-MS/MS method for the quantification of carbinoxamine in human plasmaLC-MS/MS method for the quantification of carbinoxamine in human plasma
LC-MS/MS method for the quantification of carbinoxamine in human plasma
 
H010245763
H010245763H010245763
H010245763
 
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...
Monitoring Of Macronutrients Uptake by Soil and Potato Plants – A Comparative...
 
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...
Impact of Frequency Offset on Interference between Zigbee and Wifi for Smart ...
 
R01226111114
R01226111114R01226111114
R01226111114
 
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAl
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAlPrediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAl
Prediction of electronic and magnetic properties of Full Heusler Alloy – Ir2CrAl
 

Similar to Efficacy of Use of A-Si EPID as Imaging Device in IMRT QA

PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...
PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...
PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...AM Publications
 
IRJET- Fabric Defect Detection using Discrete Wavelet Transform
IRJET- Fabric Defect Detection using Discrete Wavelet TransformIRJET- Fabric Defect Detection using Discrete Wavelet Transform
IRJET- Fabric Defect Detection using Discrete Wavelet TransformIRJET Journal
 
IRJET- Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...
IRJET-  	  Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...IRJET-  	  Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...
IRJET- Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...IRJET Journal
 
IRJET-Retina Image Decomposition using Variational Mode Decomposition
IRJET-Retina Image Decomposition using Variational Mode DecompositionIRJET-Retina Image Decomposition using Variational Mode Decomposition
IRJET-Retina Image Decomposition using Variational Mode DecompositionIRJET Journal
 
Lung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesLung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesIRJET Journal
 
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...IRJET Journal
 
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...IJTET Journal
 
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...gerogepatton
 
Image Denoising of various images Using Wavelet Transform and Thresholding Te...
Image Denoising of various images Using Wavelet Transform and Thresholding Te...Image Denoising of various images Using Wavelet Transform and Thresholding Te...
Image Denoising of various images Using Wavelet Transform and Thresholding Te...IRJET Journal
 
Speckle noise reduction using hybrid tmav based fuzzy filter
Speckle noise reduction using hybrid tmav based fuzzy filterSpeckle noise reduction using hybrid tmav based fuzzy filter
Speckle noise reduction using hybrid tmav based fuzzy filtereSAT Publishing House
 
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...
IRJET - Change Detection in Satellite Images using Convolutional Neural N...IRJET Journal
 
Ijarcet vol-2-issue-3-891-896
Ijarcet vol-2-issue-3-891-896Ijarcet vol-2-issue-3-891-896
Ijarcet vol-2-issue-3-891-896Editor IJARCET
 
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...IRJET Journal
 
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...IRJET Journal
 
A Survey on Ultrasound Beamforming Strategies
A Survey on Ultrasound Beamforming StrategiesA Survey on Ultrasound Beamforming Strategies
A Survey on Ultrasound Beamforming StrategiesIRJET Journal
 
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATION
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATIONMULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATION
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATIONprj_publication
 
Wavelet based Image Coding Schemes: A Recent Survey
Wavelet based Image Coding Schemes: A Recent Survey  Wavelet based Image Coding Schemes: A Recent Survey
Wavelet based Image Coding Schemes: A Recent Survey ijsc
 

Similar to Efficacy of Use of A-Si EPID as Imaging Device in IMRT QA (20)

PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...
PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...
PROFILE DOSE ANALYSIS OF 6MV LINEAR ACCELERATOR WITH CCD ELECTRONIC PORTAL IM...
 
C04631115
C04631115C04631115
C04631115
 
IRJET- Fabric Defect Detection using Discrete Wavelet Transform
IRJET- Fabric Defect Detection using Discrete Wavelet TransformIRJET- Fabric Defect Detection using Discrete Wavelet Transform
IRJET- Fabric Defect Detection using Discrete Wavelet Transform
 
IRJET- Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...
IRJET-  	  Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...IRJET-  	  Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...
IRJET- Jeevn-Net: Brain Tumor Segmentation using Cascaded U-Net & Overall...
 
IRJET-Retina Image Decomposition using Variational Mode Decomposition
IRJET-Retina Image Decomposition using Variational Mode DecompositionIRJET-Retina Image Decomposition using Variational Mode Decomposition
IRJET-Retina Image Decomposition using Variational Mode Decomposition
 
Lung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesLung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing Techniques
 
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...
IRJET - Automated 3-D Segmentation of Lung with Lung Cancer in CT Data using ...
 
F04433538
F04433538F04433538
F04433538
 
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
 
Epid
EpidEpid
Epid
 
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...
A DEEP LEARNING APPROACH FOR DEFECT DETECTION AND SEGMENTATION IN X-RAY COMPU...
 
Image Denoising of various images Using Wavelet Transform and Thresholding Te...
Image Denoising of various images Using Wavelet Transform and Thresholding Te...Image Denoising of various images Using Wavelet Transform and Thresholding Te...
Image Denoising of various images Using Wavelet Transform and Thresholding Te...
 
Speckle noise reduction using hybrid tmav based fuzzy filter
Speckle noise reduction using hybrid tmav based fuzzy filterSpeckle noise reduction using hybrid tmav based fuzzy filter
Speckle noise reduction using hybrid tmav based fuzzy filter
 
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
 
Ijarcet vol-2-issue-3-891-896
Ijarcet vol-2-issue-3-891-896Ijarcet vol-2-issue-3-891-896
Ijarcet vol-2-issue-3-891-896
 
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
 
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...
Retinal Vessel Segmentation using Infinite Perimeter Active Contour with Hybr...
 
A Survey on Ultrasound Beamforming Strategies
A Survey on Ultrasound Beamforming StrategiesA Survey on Ultrasound Beamforming Strategies
A Survey on Ultrasound Beamforming Strategies
 
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATION
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATIONMULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATION
MULTI WAVELET BASED IMAGE COMPRESSION FOR TELE MEDICAL APPLICATION
 
Wavelet based Image Coding Schemes: A Recent Survey
Wavelet based Image Coding Schemes: A Recent Survey  Wavelet based Image Coding Schemes: A Recent Survey
Wavelet based Image Coding Schemes: A Recent Survey
 

More from IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Recently uploaded

GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)Areesha Ahmad
 
Forensic Biology & Its biological significance.pdf
Forensic Biology & Its biological significance.pdfForensic Biology & Its biological significance.pdf
Forensic Biology & Its biological significance.pdfrohankumarsinghrore1
 
Botany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfBotany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfSumit Kumar yadav
 
Isotopic evidence of long-lived volcanism on Io
Isotopic evidence of long-lived volcanism on IoIsotopic evidence of long-lived volcanism on Io
Isotopic evidence of long-lived volcanism on IoSérgio Sacani
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticssakshisoni2385
 
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...Sérgio Sacani
 
Spermiogenesis or Spermateleosis or metamorphosis of spermatid
Spermiogenesis or Spermateleosis or metamorphosis of spermatidSpermiogenesis or Spermateleosis or metamorphosis of spermatid
Spermiogenesis or Spermateleosis or metamorphosis of spermatidSarthak Sekhar Mondal
 
Pests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPirithiRaju
 
Botany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsBotany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsSumit Kumar yadav
 
Unlocking the Potential: Deep dive into ocean of Ceramic Magnets.pptx
Unlocking  the Potential: Deep dive into ocean of Ceramic Magnets.pptxUnlocking  the Potential: Deep dive into ocean of Ceramic Magnets.pptx
Unlocking the Potential: Deep dive into ocean of Ceramic Magnets.pptxanandsmhk
 
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.Nitya salvi
 
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisRaman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisDiwakar Mishra
 
Biopesticide (2).pptx .This slides helps to know the different types of biop...
Biopesticide (2).pptx  .This slides helps to know the different types of biop...Biopesticide (2).pptx  .This slides helps to know the different types of biop...
Biopesticide (2).pptx .This slides helps to know the different types of biop...RohitNehra6
 
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencyHire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencySheetal Arora
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)Areesha Ahmad
 
Biological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfBiological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfmuntazimhurra
 
VIRUSES structure and classification ppt by Dr.Prince C P
VIRUSES structure and classification ppt by Dr.Prince C PVIRUSES structure and classification ppt by Dr.Prince C P
VIRUSES structure and classification ppt by Dr.Prince C PPRINCE C P
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsSérgio Sacani
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000Sapana Sha
 
Zoology 4th semester series (krishna).pdf
Zoology 4th semester series (krishna).pdfZoology 4th semester series (krishna).pdf
Zoology 4th semester series (krishna).pdfSumit Kumar yadav
 

Recently uploaded (20)

GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Forensic Biology & Its biological significance.pdf
Forensic Biology & Its biological significance.pdfForensic Biology & Its biological significance.pdf
Forensic Biology & Its biological significance.pdf
 
Botany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfBotany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdf
 
Isotopic evidence of long-lived volcanism on Io
Isotopic evidence of long-lived volcanism on IoIsotopic evidence of long-lived volcanism on Io
Isotopic evidence of long-lived volcanism on Io
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...
PossibleEoarcheanRecordsoftheGeomagneticFieldPreservedintheIsuaSupracrustalBe...
 
Spermiogenesis or Spermateleosis or metamorphosis of spermatid
Spermiogenesis or Spermateleosis or metamorphosis of spermatidSpermiogenesis or Spermateleosis or metamorphosis of spermatid
Spermiogenesis or Spermateleosis or metamorphosis of spermatid
 
Pests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdf
 
Botany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsBotany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questions
 
Unlocking the Potential: Deep dive into ocean of Ceramic Magnets.pptx
Unlocking  the Potential: Deep dive into ocean of Ceramic Magnets.pptxUnlocking  the Potential: Deep dive into ocean of Ceramic Magnets.pptx
Unlocking the Potential: Deep dive into ocean of Ceramic Magnets.pptx
 
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
 
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisRaman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
 
Biopesticide (2).pptx .This slides helps to know the different types of biop...
Biopesticide (2).pptx  .This slides helps to know the different types of biop...Biopesticide (2).pptx  .This slides helps to know the different types of biop...
Biopesticide (2).pptx .This slides helps to know the different types of biop...
 
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencyHire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)
 
Biological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfBiological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdf
 
VIRUSES structure and classification ppt by Dr.Prince C P
VIRUSES structure and classification ppt by Dr.Prince C PVIRUSES structure and classification ppt by Dr.Prince C P
VIRUSES structure and classification ppt by Dr.Prince C P
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
 
Zoology 4th semester series (krishna).pdf
Zoology 4th semester series (krishna).pdfZoology 4th semester series (krishna).pdf
Zoology 4th semester series (krishna).pdf
 

Efficacy of Use of A-Si EPID as Imaging Device in IMRT QA

  • 1. IOSR Journal of Applied Physics (IOSR-JAP) e-ISSN: 2278-4861.Volume 7, Issue 1 Ver. III (Jan.-Feb. 2015), PP 27-43 www.iosrjournals.org DOI: 10.9790/4861-07132743 www.iosrjournals.org 27 | Page Efficacy of Use of A-Si EPID as Imaging Device in IMRT QA Dina Abdelaziz, Wafaa Khalifa and Nader El Sherbini Abstract: When IMRT treatment is considered to be a routine work, one should seek for an accurate, time saving, fast and simple QA procedure. Electronic portal imaging device (EPID), especially amorphous silicon type, is a promising tool for IMRT dosimetry. In this work a calibration of an amorphous silicon flat panel-type imager (Elekta iViewGT, release 3.2) is done, which was incorrect (figure 3.1.) due to the EPID renormalization to the target value (Elekta instructions). The calibration of the amorphous silicon flat panel- type imager (Elekta iViewGT) was repeated after considering the Pixel Scaling Factor (PSF), which gave the correct response (figure 3.2.). By using Omni Pro software the EPID images are converted into doses and a correction for the dose reading from these images is done to obtain the same results as that measured at the isocenter in 20 cm PMMA slab phantom. A correction for the EPID profiles for different field sizes is done to be like those measured in the full scatter water phantom. A pre-treatment verification for a clinical step & shoot IMRT plan is done for a prostate cancer case generated by Xio (version 4.2.2) treatment planning system (TPS). The pre-treatment verification is done by EPID and a comparison in dose distribution against TPS and Films are performed. This study show that the combination between iViewGT and Omni Pro software with the corrections provided in the present study gave an accurate method to verify the dose of IMRT fields in two dimensions inside a homogenous slab phantom. The results indicated also that the EPID (Elekta iViewGT) is an accurate and can replace film for field-by field pretreatment verification of IMRT inside a phantom. I. Introduction The challenge of external beam radiotherapy for cancer treatment is to irradiate the tumor with a desired dose, while the surrounding healthy tissue suffers as little as possible radiation damage. Due to increase of complexity in modern radiotherapy techniques the needs for checking accuracy of the delivered dose to patient has been also increased, either pretreatment or in vivo. This reflects the importance of quality assurance (QA) as a step in radiation therapy treatment. Quality assurance procedures of a patient treatment can be divided into two main parts: Quality control of the technical aspects of the linear accelerator (Linac) itself includes verification of: Dose output, the transferred parameter between treatment planning system (TPS) to linac (collimator settings, the direction of wedges and multileaf collimator (MLC) shape) and Isodose check. Verification of the actual patient treatment, including verification of both patient setup and delivery of correct dose. Ionization chamber is the predominant tool used to check the dose accuracy at certain points, while the dose distribution of certain plan could be measured using different tools, e. g. Films, 2D diode array. The increasing of using IMRT to deliver conformal radiation treatment has prompted the search for a faster and more cost effective quality assurance (QA) system. The standard technique relies on the use of film for two- dimensional dose distribution verification. Although film is considered the standard and most widely used for this purpose, the procedures involved are relatively lengthy, labor intensive and costly for multiple field IMRT verification. (Lim S. 2008). QA of intensity modulated radiation therapy (IMRT) is a sophisticated and time consuming task, where one should check not only the delivered dose and dose distribution but also need to check the field segments shape, number and machine capability to deliver IMRT plan. When IMRT treatment is considered to be a routine work, one should seek for an accurate, time saving, fast and simple QA procedure. Electronic portal imaging device (EPID), especially amorphous silicon type, is a promising tool for IMRT dosimetry. A commercial a-Si EPID has been studied by Talamonti C., et al, (2006) to investigate its potential in the field of pretreatment verifications of step and shoot, IMRT for 6 MV photon beams. An a-Si EPID has been investigated by Budgell G., et al, (2005) to determine its usefulness and efficiency for performing linear accelerator quality control checks specific to step and shoot IMRT. Several dosimetric parameters were measured using the EPID: dose linearity and segment to segment reproducibility of low dose segments, and delivery accuracy of fractions of monitor units. Xu C., et al, (2002) have developed an automated image based method for MLC positioning verification using images acquired from an EPID. Yeo I. (2007) developed a method based on dose-response function, which verifies the beamlet intensity in IMRT from dose image in EPID and reconstructs dose in a patient. Greer P., et al, (2007) determined the difference in response between an a-Si EPID to the open and MLC transmitted beam components of IMRT beams. Ansbacher
  • 2. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 28 | Page W. (2006) developed a method for rapid evaluation of IMRT plans, using portal images for reconstruction of the dose delivered to a virtual 3D phantom. II. Materials And Methods Elekta Precise (Crawley, UK) linear accelerator, which support IMRT deliver, was used in this study (figure 2.1a). This linear accelerator (LINAC) produces photon energies of 6 and 15MV, also it produces electrons with energies of 6, 8, 10, 12 and 15 Mev. Only 6 MV photon beam is used in this study. The LINAC is equipped with MLC. The MLC consists of 40 leaf pairs with 1 cm leaf width at the isocenter.The type of EPID used throughout this study is an amorphous silicon flat panel-type imager (Elekta iViewGT, release 3.2).The detector panel is a PerkinElmer Amorphous Silicon (a-Si) detector, and provides a resolution of 1024 × 1024 16-bit pixel images, with a detector panel size of 41 × 41 cm2 (approximately 26 × 26 cm2 at isocenter).Display pixel factor information;When images are acquired by iViewGTTM , the pixel values are re-normalized or averaged before the image data is saved to the database. At the time of acquisition, the pixel scaling factor is saved with the images in the database. This feature allows us to determine the original accumulated pixel value by dividing the pixel value by the pixel scaling factor.Polymethyl methacrylate (PMMA), also known as acrylic. Trade names are Lucite, Plexiglas or Perspex. It consists of 1 cm for each sheet of density 1.19 g/cm3 and area 30 х 30 cm2 .GAFCHROMIC® EBT2 Dosimetry Film (8x10 in size) was used. The film spatial resolution >5,000 dpi (dots per inch) with dose range 1cGy – 10Gy (in red color channel).Omni ProTM IMRT is software used for complete dosimetric verification and QA of IMRT treatment cycle. This software function is to import and compare calculated doses fromTPS planned data with measured dose distributions from films that were exposed in the IMRT phantom. It includes 1D profile, 2D isodose profiles as well as fully automated comparisons using analysis tools such as Gamma index method.A computerized welhÖfer WP 700 water phantom version 3.5 was used in the present study for acquiring the beam data.XiO treatment planning system (TPS) (from CMS Inc., USA) version 4.2.2 implemented superposition algorithm is used in this study. Calibration of the EPID Several steps are necessary to reconstruct the dose in the phantom or patient from the pixel values of the EPID. All measurements of this study were performed at gantry and collimator angle 0°. A- Calibration of EPID response to different doses:EPID images of reference field size (typically 10x10 cm2 ) were recorded for different MU (2, 5, 10, 15, 25, 50, 100, 150 and 200 MU).The acquired images by iViewGTTM are re-normalized (automatically by the iViewGTTM ) before saving process; which means that all images will appear with the same optical density (OD) for all different doses images. Sofor each image the pixel scaling factor (PSF) is recorded to determine the original accumulated pixel value by dividing the pixel value by this factor Originalaccumulatedpixelvalue= Pixelvalue PixelScalingFactor (PSF ) (1) To convert these images into doses, all images were exported to Omni Pro system.The Omni Pro software can read the Optical Density OD of the image.Each EPID image imported to Omni Pro will be corrected for its own PSF.even after dividing each image by its corresponding PSF it will still incorrect because of several reasons:The scattered radiation from EPID;The scatter within the phantom;The scatter from the phantom to the EPID;The attenuation of the beam by the phantom; andThe distance from the radiation source to the EPID plane and to the dose reconstruction plane. (Wendling M., et al, 2006).Record the Omni Pro reading value for each EPID dose, and get the Dose correction factor (DF), which is the Omni Pro reading value for a specific dose divided by the Actual irradiated dose at isocenter DF = OmniProreadingvalueforaspecificdose Actualirradiateddose (2) B- Calibration of the EPID profile In order to determine the necessary parameters for correcting the dose profile reconstruction, i.e., relating pixel values in the EPID images with absolute dose values in the phantom, the EPID images at different square field sizes are recorded with PMMA slab phantom of thickness 20 cm(an average thickness to encompass the phantom (or patients) thickness at 6 MV photon beam). For all phantom measurements an isocentric setup [SAD (source –axis distance) setup] was used. Dose profiles for the same field sizeswere obtained using semiflex 0.147 ion chamber in the full-scatter water phantom [Source-Skin Distance (SSD) = 88.6cm] were measured with ion chamber at 11.4cm depth.Where 11.4 cm water is equivalent to 10 cm of Perspex. These dose profiles will be the reference for EPID dose profile correction. The EPID images were taken and corrected by PSF & DF. After the conversion of EPID pixel values into dose values according to the dose response relation, the resulting image is called dose image DEPID . The corrected portal pixel dose according to its position
  • 3. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 29 | Page regarding to central axis (CAX) is PDEPID , which is the image after profile correction, is obtained by the predicted equation based on the previous work by Wendling, et al, 2006 and Renner, et al, 2005. PDij EPID = Dij EPID × e−ar 3 Where r is the distance of a pixel ij from the central axis, a is the exponential fitting constant.This constant is fitted properly for all field sizes. Pre-treatment verification: EPID versus filmA clinical step & shoot IMRT plan is used for a prostate cancer treatment generated by Xio (version 4.2.2) TPS. This plan consists of seven fields, and each field had between 8-14 segments.With EPID, one image was acquired for each segment using single exposure option in the image acquisition mode for IMRT (Pre- segment image acquisition and storage also allows verification of MLC leaf position), where the dose construction was also done separately for each segment. All of these images were sent to the Omni Pro and corrected according to the previous corrections.The film irradiated simultaneously with the corresponding EPID image acquisition for each field using PMMA phantom. A film is placed at 10 cm depth in the PMMA phantom on the Linac table and 10 cm of PMMA slabs are placed under the film to achieve the backscatter. This simultaneous irradiationto avoid any fluctuation in doses between film and EPID irradiation like dose, dose rate fluctuation and MLC reposition. These images are compared with the simultaneously irradiated films and a γ evaluation was done in two dimensions. III. Results &Discussion 3.1 EPID Response Wendling, et al, (2006) stated that the resulting response of the a-Si EPID has been shown to be linear with dose, although a small ghosting effect remains, which is mainly a function of the number of exposed frames and no correction for ghosting effect was made in their study. So, in the present study the original accumulated pixel values of the raw images data which are obtained from iViewGTTM software, by using equation (2-1) as described in section (2) were used. Figure (3.1.) central axis original accumulated pixel values versus monitor units for a 10x10 cm2 field. 3.2 Pixel Scaling Factor Table (3-1) shows the signal reading in Omni Pro software of the doses calculated at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm. Table (3-1)the signal reading in Omni Pro software of the doses calculated at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm. Dose mGy at the isocenter of the PMMA phantom Signal reading 100% 20 479 50 513 100 520 150 521.5 200 521.5 250 521.7 500 521.6 1000 521.6 2000 521.6
  • 4. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 30 | Page These signals readings were obtained without any correction where one can found that there is incorrect response,so all of these images numerical values will be divided by the PSF values listed in table (3-2). These PSF are obtained from EPID row image data. Table (3-2) The PSF for different doses at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm for reference field size 10x10 cm2 Dose (mGy) PSF 20 4.61207 50 1.98999 100 0.93866 150 0.67392 200 0.51005 250 0.41549 500 0.20685 1000 0.10175 2000 0.05091 The resulting signal reading (100%) that irradiated to different doses 2,5,10, 50,100 and 200 cGy at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm for reference field size 10x10 cm2 after dividing all images values by the corresponding PSF using the equation (2-1), that are obtained by Omni Pro show different response for the different doses, as indicated in table (3-3). Table (3-3)the signal reading in Omni Pro software after dividing each image numerical values by its own PSF using equation (2-1). Dose mGy at the isocenter of the PMMA phantom Signal reading 100% 20 103.4 50 257.5 100 554 150 773 200 1021 250 1254 500 2519 1000 5120 2000 10230 3.3 Dose Correction Factor Although the Omni Pro gave a linear relationship between the dose and the signal reading for the EPID images after dividing by PSF, the Omni Pro signal reading is not the same as the incident dose values. These results agree with those obtained by Wendling,et al, (2006). Wendling,et al, (2006) attributes these incorrect results to that the scatter within the EPID, the scatter within the phantom, the scatter from the phantom to the EPID and the attenuation of the beam by the phantom are not considered. Also the distance from the radiation source to the EPID plane and to the dose reconstruction plane is not considered. From our results the discrepancies may also be attributed to the introduction of a proposed sensitometric curve which is the relation that consolidates Omni Pro to read the EPID images. Figure (3.2.) The relation between Omni Pro reading and doses.
  • 5. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 31 | Page It is clear from figure (3.2.) that the relation between Omni Pro reading and the delivered doses is a linear relationship, and its slope = 51.26. By using equation (2-2) then the Omni Pro signal reading is the dose at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm. Table (3-4) Dose correction factor (DF) factor for different doses that obtained using equation (2-2). Dose mGy at the isocenter of the Perspex phantom DF 20 51.70 50 51.50 100 55.00 150 51.53 200 51.05 250 50.16 500 50.38 1000 51.20 2000 51.15 Average value of DF 51.51 It is obvious from table (3-4) that the average DF factor is nearly the same as the slope of linear relationship of figure (3.2.), this slope = 51.26. So the slope value of the linear relationship between the dose and the signal reading for the EPID images is used as DF value for different doses. In this case the signal reading in Omni Pro is the same as that irradiated at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm. Figure (3.3.) shows the two dimensional dose distributions for the EPID images after PSF and DF correction. (a) 2CGy (b) 5CGy (c) 10CGy (d) 50CGy
  • 6. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 32 | Page (e)100CGy (f) 200CGy Figure (3.3.) Two dimensional dose distributions for the EPID images for 10x10 cm2 of 6MV photon beam after PSF & DF correction 3.4 Dose correction for different field size Table (3-5) shows the PSF for different field sizes with the same delivered doses (500 mGy). Although all the field sizes were irradiated to the same dose at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom and at SSD = 90cm, there was difference in the PSF values, which is due to the different contribution of scatter by changing the field size. Table (3-5): PSF for different field sizes using 500 MGY. Field size PSF 3×3 0.1688 5×5 0.20515 10×10 0.20683 15×15 0.20022 These differences in PSF values for the different field sizes receiving the same dose at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom and at SSD = 90cm may be due to the dose calculation for each field size is in full scatter conditions, that takes into consideration the collimator scatter factor and the phantom scatter factor (calculated under full scatter situation), on the other hand EPID is not a full scatter phantom (Varian Portal Vision instructions). Also the EPID is known to have a lateral scatter (Wendling,et al, 2005) which is not considered in the full scatter phantom calculation. By trial it was found that the PSF for the reference field 10x10 cm2 gave approximately the same dose value for different field sizes, as tabulated in table (3-6). Table (3-6): OmniPro maximum signal reading for different field size images obtained by EPID after dividing by the PSF of the reference field (10×10 cm2 ). Field size Signal reading 100% 3×3 2563 5×5 2562.9 10×10 2562.9 15×15 2562.9 Then after dividing the signal reading for each image of the different field sizes by the DF (51.26) the correct dose at the isocenter, at 10cm depth in a 20cm thick PMMA slab phantom and SSD = 90cm is obtained with maximum difference ± 0.2 % as shown by the red arrow in figure (3.4.).
  • 7. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 33 | Page (a) 3×3 cm2 (b) 5×5 cm2 Figure (3.4.) Generally after DF correction either for different doses of the reference field size 10×10cm2 or for the same dose at different field size the average dose values agreed with the ionization chamber measurements at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm with ± 0.2 % (which agree with work done by Wendling,et al, 2006 and Renner,et al, 2005). 3.2 Profile Correction: Profiles of square fields from EPID images were compared with those measured using an ionization chamber that was located in a full-scatter water phantom. Figure (3.5.) shows the actual dose profiles of square fields of different sizes obtained by EPID against the water phantom and TPS. The measurements are performed in water phantom at 11.4 cm water (which is equivalent to 10 cm of PMMA), and source skin distance (SSD) = 88.6 cm. (a) Field size 3×3 cm2 (b) Field size 5×5 cm2
  • 8. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 34 | Page (c) Field size 10×10 cm2 Figure (3.5.) dose profile as measured by water phantom and from TPS against the reconstructed dose profile from the EPID images without profile correction (a) 3×3 cm2 field size. (b) 5×5 cm2field size. (c) 10×10 cm2 field size. All profiles are normalized to 100% of the measured dose. Figure (3.5.) shows a difference between water phantom measurements and the reconstructed profiles from the EPID images for the studied field sizes. Table (3-7) show the maximum percent of error between profiles measured and t he EPID at different field sizes. Table (3-7) (% error) between EPID and the water phantom dose profiles. Field size(cm2 ) Error (%) 3×3 3% 5×5 3.5% 10×10 3% It is clear from figure (3.5.) and table (3-7); that the profile behaviour along the central axis region by the three methods is good, and became worse in the penumbra region and the tails. This is because within the EPID mainly lateral X-ray scatters and optical photon scatter occurs, resulting in more flattened at the shoulders more than in water phantom &TPS. These findings are in agreement with the results of Wendling, et al, 2006. The fit parameter [a] in equation (2-3) correct for these behaviour where it is obtained from the water phantom measurements. Figure (3.6.) shows the dose profiles of square fields for different sizes obtained by EPID after correction using equation (2-3) and water phantom and TPS. (a) Field size 3×3 cm2
  • 9. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 35 | Page (b) Field size 5×5 cm2 (c) Field size 10×10 cm2 Figure (3.6.) dose profiles for different square field sizes as measured by water phantom at the isocenter at 11.4 cm depth which is equivalent to 10 cm Perspex, TPS and profiles reconstructed by the EPID using the Omni Pro software at 6MV photon beam. The fitting constant (a) in equation (2-3) was got by iteration to minimize the variation between the measured dose by the water phantom and the reconstructed dose from the EPID, and it was found to be 0.03 for all field sizes studied. Figure (3.6.) shows that the agreement has improved and become in agreement with the other methods (water phantom and TPS). Among all field sizes the error does not exceed 0.2% between the EPID profiles and the profiles measured by the water phantom and calculated by the TPS except for field size 10×10 cm2 where the agreement is down to the 30 % dose on the central axis. These findings were found to be in qualitative agreement with Wendling M., et al, 2006 who stated that the agreement was very good with the central axis, but remain worse in the penumbra and in the tail of the profiles due to blurring of the measurements by the ionization chamber) but in this study the most importance is for the small field sizes in order that the small segments in the IMRT verification and the entire profiles for small field sizes agree with the measured ones. Figure (3.7.) shows the difference between the EPID after dose and profile correction against the TPS for different field size with the gamma evaluation (with criteria 3mm and 3%) using the Omni Pro software and after implementing the dose and profile correction factor. The plane dose profile comparison between the corrected EPID images and TPS is taken arbitrary on the periphery of the field, to be sure that it correct through out the field.
  • 10. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 36 | Page (a) (b) (c) Figure (3.7) Workspace – Omni Pro IMRT QA program reveal the dose profile of the EPID against that of TPS (upper) and the γ index (upper) figure (a) for 3×3 field size, (b) for 5×5 field size and (c) for 10×10 field size. Be sure that there is no red area (γ index > 1) within the field.
  • 11. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 37 | Page 3.3 Pre-treatment IMRT Verification a) Segment by segment: Now, instead of using ordinary open fields, an IMRT plan distribution is displayed by EPID & films in comparison with the TPS. The plan is done using superposition algorithm. For an IMRT case, 7 fields were used. The number of segments in each field is illustrated in the plan report, (table 3-8). Plan Report: Table (3-8): number of segments for a prostate IMRT plan using Xio TPS version (4.2.2). An IMRT plan has 7 fields Beam number Number of segments 1 14 2 9 3 9 4 12 5 13 6 8 7 8 The correction factors (PSF, DF & profile correction equation) will be applied to each segment for every field. EPID in comparison with Films versus TPS Figures from (3.8.) to (3.10.) show some of the seventh field segments as an examples with their dose distribution comparisons in terms of dose profile and theγ evaluation for EPID,as compared with film versus TPS. The γ evaluation issued to judge about the comparison on the basis of the standard deviations in the γ index between EPID & films using the Omni Pro software after all correction factors. Since the dose for each segment was within 5-10 MUs which are under the response of Gafchromic films, the doses were normalized to 50 cGy within the isocenter. Figure (3.8, a) Workspace – Omni Pro IMRT QA program for segment 2 reveal the dose profile of the EPID against that of TPS (upper) and the γ index (down).
  • 12. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 38 | Page Figure (3.8, b) Workspace – Omni Pro IMRT QA program for segment 2 reveal the dose profile of the film against that of TPS (upper) and the γ index (down). Figure (3.9, a) Workspace – Omni Pro IMRT QA program for segment 3 reveal the dose profile of the EPID against that of TPS (upper) and the γ index (down).
  • 13. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 39 | Page Figure (3.9,b) Workspace – Omni Pro IMRT QA program for segment 3 reveal the dose profile of the film against that of TPS (upper) and the γ index (down). Figure (3.10, a) Workspace – Omni Pro IMRT QA program for segment 7 reveal the dose profile of the EPID against that of TPS (upper) and the γ index (down).
  • 14. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 40 | Page Figure (3.10,b) Workspace – Omni Pro IMRT QA program for segment 7 reveal the dose profile of the film against that of TPS (upper) and the γ index (down). Table (3-11) show the gamma evaluation for segments as comparison between EPID & films of the present study. Table (3-11): the standard deviation (SD) for both EPID and films in comparison to TPS using Omni Pro software for the segments of field number 7 Segment number EPID SD % Film SD % 1 0.24 0.31 2 0.14 0.23 3 0.15 0.29 4 0.19 0.30 5 0.23 0.28 6 0.36 0.38 7 0.23 0.24 8 0.27 0.29 Average 0.22 0.29 As seen from table (3-11); the values of the SD and the average among the 8 segments for EPID are less than that for Gafchromic films. This may be due to that the scanning of the film is not reproducible as the EPID, and even by using the protocol in the method of film scanning (taken from Estro booklet 9 (2008)) there is still a deviation, where scanning is done for the film more than 5 times which is an evidence on the instability and turning the scanner off between scanning this may be contributed to the total error because the condition will not be the same for all films, especially that the results will be affected if the orientation of the film is inversed. Also Alber M., et al, (2008) stated that if there is any defect in the film manufacturing it will strongly affect the results. If the values of standard deviation for EPID and films in table (3-11) are analyzed separately, it will be clear that as the shaping of the segment increase with the MLC the SD relatively increases. Renner,et al, 2005 stated that the superposition algorithm is generally known to be more accurate at tissue interfaces but it does not model either multileaf collimator leaf-end leakage or inter-leaf leakage. So increasing the number of segments in a plan will demonstrate any inaccuracy of the planning system. On the other hand the difference between Films and EPID can also be compared by the percentage of pixels passing the gamma test in the region of interest for each segment of the field as demonstrated by the histogram in figure (3.19.). Khan F., et al, (2008) also used the percentage of pixels passing the digital gamma test to compare between EPID and the Matrixx (in this study films used instead of MatriXX, IMRT MatriXX is a two dimensional ion chamber array (IBA Dosimetry, Schwarzenburck,Germany))
  • 15. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 41 | Page Figure (3.11.) shows the percent of pixel passing the gamma test for EPID against films. From figure (3.11.) it is clear that for all segments the percent of pixels passing the gamma test for EPID is better than that for films. Also the minimum number of pixels passing the gamma test for EPID and film were 99% and 98% respectively for segment 8. For segments 2 and 3, all pixels pass the gamma test for both EPID and films. Figures (3.8.) to (3.10.) show the comparison between EPID and films for samples of the segments in the seventh field. These results prove that the EPID can replace films in IMRT QA. b) Segment combination: Finally, it important to show the difference between reconstructed dose distribution from EPID after correction and the calculated dose by TPS in the fields as a total with the same prescribed doses for the planed prostate case plan. The two-dimensional midplan (isocentric-plan) dose distributions for all segments of each field were then summed to obtain the total midplan dose of that specific field. Figure (3.12, a) show Workspace – Omni Pro IMRT QA program for the a field from IMRT plan for the prostate case with its prescribed doses that calculated by the TPS in the PMMA slab phantom( at 10 cm depth and SSD=90) at gantry zero. The difference in dose distribution between the calculated dose by TPS and the reconstructed dose distribution by EPID is also presented in the figures. Figure (3.12, b) show the histograms of the γ index, indicating the maximum, minimum signal, the average signal and the standard deviation for each field using EPID. Figure (3.12, a) Workspace – Omni Pro IMRT QA program for field 7, reveal the dose distribution of the EPID(down-left),the TPS dose distribution (upper-left),the dose profile of the EPID against that of TPS (upper- right) and the γ index (down-right).
  • 16. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 42 | Page Figure (3.12, b) Gamma index histogram for field 7. Table (3-12) shows the difference in dose distribution between the measured films and the reconstructed ones from EPID in comparison with the TPS in terms of SD. Table (3-12) gamma evaluation comparison for the fields as a total between EPID and films. Field number EPID SD % Film SD % Number of segments 1 0.30 0.33 14 2 0.25 0.26 9 3 0.28 0.27 9 4 0.16 0.31 12 5 0.20 0.32 13 6 0.18 0.30 8 7 0.15 0.28 8 Average 0.21 0.29 From table (3-12) if the data is analyzed separately either for EPID or for films it will be clear that increasing the number of segments in the plan, increase the value of the SD for both EPID and films that indicate that the deviation is due to the TPS calculation accuracy for either multileaf collimator leaf-end leakage or inter-leaf leakage. For films and EPID the percentage of pixels passing the gamma test in the region of interest for each field as are shown in figure (3.13.). Figure (3.13.) shows the percent of pixel passing the gamma test for EPID against films for each field.
  • 17. Efficacy of use of a-Si EPID as imaging device in IMRT QA DOI: 10.9790/4861-07132743 www.iosrjournals.org 43 | Page From figure (3.13.) it is clear that all the fields reconstructed from EPID are above 99% (for the criteria 3% dose or 3mm distance to agreement). Khan,et al, (2008) used 9 beam step-and-shoot IMRT plan for head phantom irradiated with 6MV X-ray and the gamma evaluation histograms calculated for all9 beams indicates that 95% of all points pass the preset criteria 3% dose or 3mm distance to agreement for each field. IV. Conclusion From this study one may conclude the following:  The time required to perform dose reconstruction with the EPID (iViewGT, release 3.2) is probably the same as the established method using film. Yet, EPID as QA device has advantages over film QA in that it is linear, more reproducible, and eliminates film problems, with scanners.  The total annual cost of imaging with film is higher than thatusing EPID concerning IMRT QA.  EPID (iViewGT, release 3.2) can replace the ionization chamber for measuring dose for the studied field sizes at the isocenter at 10cm depth in a 20cm thick PMMA slab phantom at SSD = 90cm after correction for PSF and DF.  From the study the combination between iViewGTand Omni Pro software with the used corrections can be applied and provides an accurate method to verify the dose of IMRT fields in two dimensions inside a homogenous slab phantom.  The EPID is an accurate and fast alternative to film for field-by field pretreatment verification of IMRT inside a phantom. Generally, Michael G. Herman – AAPM Task Group 58 – Electronic Portal Imaging – 2001 shows that the annual cost of imaging with EPID as a function of increasing use of portal imaging is cheaper than that with films. When IMRT QA by films taken into consideration especially that the cost of QA films for IMRT is more expensive than that for portal imaging, sure this cost will be more effective.