Image Quality Impact of SmartGrid Processing in
Bedside Chest Imaging
Authors: Lori L. Barski (MS), Dong Yang (Ph.D.), Mary Couwenhoven (MS)
Introduction
Scattered radiation is known to degrade image quality in
diagnostic X-ray imaging. Traditional methods of reducing
scatter are collimation, utilizing an air-gap, and/or utilizing an
anti-scatter grid [1]. Best practices in radiography prescribe
collimation on every exam [2] and, while this is helpful, it is not
completely sufficient due to scatter that occurs within the
patient. Likewise, air-gap techniques are helpful [1], but are
not practical at the bedside. Anti-scatter grids are by far the
most popular means of reducing scatter in portable imaging,
but they present challenges to radiographers (e.g., positioning
and alignment) and increases dose to the patient [1,3].
A new image processing tool, SmartGrid, has been developed
that compensates for the effects of X-ray scatter in an image,
and produces results comparable to those of a physical anti-
scatter grid. Figure 1 shows a pair of portable chest images
captured with a grid (top left), without a grid (top right), and
the results with SmartGrid (bottom left).
The scatter distribution image (bottom right) demonstrates
regions of the image that are most impacted by scatter (lighter
tone). SmartGrid processing estimates the scatter distribution
and removes it, resulting in an image with improved contrast.
Many physical factors affect scatter: energy spectrum of the
beam, thickness and material composition of the object, and
collimation, to name a few. SmartGrid accommodates these
variables automatically and results in image quality that
approximates anti-scatter grid visual performance.
The SmartGrid Algorithm
The SmartGrid algorithm (Figure 2) is an enhancement
algorithm that improves image contrast by suppressing scatter
in the image. The fundamental concept is one of developing a
scatter distribution image, which is a representation of the
scatter contained in the image, and then subtracting it from
the original input image.
Figure 1. Portable chest exam captured with a grid (top left)
and without a grid (top right, grid exposure 1.25X non-grid),
and then processed with SmartGrid (bottom left) with the
corresponding scatter distribution image (bottom right).
The scatter distribution image is developed using information
from the image in both linear exposure space as well as
attenuation space, which is a log transformation of the linear
data. Segmentation is done to focus the development of the
scatter field on relevant anatomical data and is used to
compute the mean linear exposure of the input image.
Parameters used to estimate scatter are determined. These
include the scatter-to-primary ratio (SPR) and a curvature
parameter used to control perturbation of the scatter
distribution as the scatter field is developed.
White Paper | SmartGrid
2
Figure 2. Flow diagram of the SmartGrid algorithm.
The scatter distribution image is computed based upon the
assumption that every object exposed by X-rays has a basic
scatter distribution that is characterized by a certain level of
energy and scatter intensity variation across the whole object
field of view. Adaptive updating of the scatter intensity across
the entire object field of view is performed in a repetitive
fashion based upon the SPR parameter for a prescribed
number of iterations.
SmartGrid Image Quality
In order to assess the image quality impact of SmartGrid
processing, a clinical study evaluating paired grid and non-grid
portable chest images was conducted at the University of
Rochester Medical Center with IRB oversight. Patients and
healthy volunteers, under informed consent and who met the
inclusion criteria (age greater than 18, body mass index greater
than 25 and a chest circumference greater than 90 cm, non-
pregnant females, and no history of high-radiation exposure),
had portable chest exams taken with and without anti-scatter
grids. Grid/non-grid usage was alternated for in-patient
subjects over a time span of days, whereas healthy volunteers
were imaged within a time span of minutes. Techniques were
selected and documented by the radiographer at the time of
the exam, following department guidelines for the site's
standard of care.
Two sets of pair-wise comparisons were evaluated (side-by-
side) for each grid/non-grid image pair: SmartGrid (SG) vs. Grid
(n=44) and SmartGrid vs. standard default processing (EVP Plus
processing available in V5.7, n=44). Pairs were presented on a
PACS in randomized order to five board-certified thoracic
radiologists (3 - 57 yrs. experience) and anatomical regions
(retro-cardiac, carina/airways, sub-diaphragm, chest wall, and
lung parenchyma) were rated using a +/- 3 point preference
scale. For each pair, the reader picked the image with the best
region (e.g., retro-cardiac) and rated it as +/- 0, 1, 2, or 3,
where the sign indicated the left or right monitor (i.e., left-
preferred was indicated with negative numbers) and the
magnitude corresponded to the degree of clinical impact
(Table 1). Readers were free to pan, zoom, and adjust window
width and window level as desired. Upon completion of the
reader study, response data was decoded such that positive
values indicated preference in favor of SmartGrid.
Additionally, each image of the pair was rated for diagnostic
quality using a 4-point RadLex scale [4] (Table 2), allowing
decimal ratings (e.g., 3.3) between the integer anchors.
Score Definition
Left Preferred
-3 Image displayed on left is strongly preferred
Clinical Impact
-2 Image displayed on left is moderately preferred
-1 Image displayed on left is slightly preferred
No Clinical Impact0 No preference between the images
Right Preferred
1 Image displayed on right is slightly preferred
2 Image displayed on right is moderately preferred
Clinical Impact
3 Image displayed on right is strongly preferred
Table 1. Preference rating scale.
White Paper | SmartGrid
3
Score Term Definition
1
Non-
diagnostic
Unacceptable for diagnostic purposes. Little or no clinically-usable diagnostic information (e.g., gross
underexposure or extensive motion artifact) and should be repeated.
2 Limited
Acceptable, with some technical defect (motion artifact, body habitus/poor X-ray penetration, or patient
positioning, but still adequate for diagnostic purposes).
3 Diagnostic Image quality that would be expected routinely when imaging cooperative patients.
4 Exemplary
Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be
emulated.
Table 2. Diagnostic quality RadLex rating scale.
Clinical Study Results
On average, no significant overall preference difference was
detected between the SmartGrid and anti-scatter grid pairs
(mean difference = 0.0, 95% CI [-0.12, 0.12], p =0.941) with
an average grid dose factor of 2.1x (i.e., SmartGrid exposures
were, on average, acquired with approximately half of the grid
dose). Figure 3 is an example of SmartGrid processing (left)
compared to the grid image (right), which was acquired at
approx 3.5x the SmartGrid dose. The average overall
preference difference was 1.0. Table 3 summarizes the pair-
wise preference differences for the anatomical regions.
Figure 3. SmartGrid image (left): 95 kVp, 2.5 mAs, 61 in. SID, compared to grid image (right): 110 kVp, 4 mAs, 58 in. SID.
Average overall preference difference is 1.0.
Grid vs. SG Preference Mean Difference Confidence Interval p-value Conclusion
Over All 0.0 -0.12, 0.12 0.941 No difference
Retrocardiac -0.2 -0.28, -0.05 0.007* In favor of grid
Carina/Airways -0.2 -0.29, -0.06 0.003* In favor of grid
Subdiaphragm -0.1 -0.18, 0.04 0.186 No difference
Chest Wall 0.1 0.04, 0.23 0.004* In favor of SG
Parenchyma 0.1 -0.29, 0.17 0.170 No difference
Table 3. SmartGrid vs. Grid Preference rating results (* indicates statistical significance at an alpha risk of 5%).
White Paper | SmartGrid
4
The difference in diagnostic quality ratings of SmartGrid vs.
grids was significant in favor of SmartGrid processing (mean
difference = 0.03, 95% CI [0.00, 0.05], p = 0.044).
Not surprising, is the result that the SmartGrid processing has
significantly higher preference overall compared to images
without SmartGrid enhancement (mean difference = 0.8, 95%
CI [0.69, 0.84], p < 10-3
) as well as for all anatomical regions
(all p-values < 10-3
).
Likewise, diagnostic quality differences were significantly in
favor of SmartGrid (p < 10-3
).
Figure 4 illustrates a typical chest wall appearance as seen in a
SmartGrid (left) image and the corresponding grid image
(right). Scatter in this region has the potential to reduce the
visualization of the pleura and the boney structures of the
outer rib cage.
Figure 4. Chest wall example. SmartGrid (Left): 90 kVp, 2.6 mAs, 60 in. SID; Grid (right): 110 kVp, 4 mAs, 68 in. SID.
Figure 5 demonstrates the retrocardiac region of a SmartGrid
image (left) and grid (right) pair. The apparent lack of contrast
in the grid image is likely caused by grid misalignment,
illustrating exactly why a software solution to scatter
compensation is of great benefit in the ICU.
Figure 6 is another portable chest comparison with the
SmartGrid processing on the right, default processing of the
same image without SmartGrid enabled, and the grid exposure
shown on the right. Note that the non-grid image was
acquired at approximately half of the grid exposure.
Figure 5. Retrocardic example. SmartGrid (left): 95 kVp, 3.2 mAs, 72 in. SID; Grid (right) 112 kVp, 7.2 mAs, 72 in. SID.
White Paper | SmartGrid
carestream.com
ยฉCarestream Health, Inc., 2017. CARESTREAM is a trademark of
Carestream Health. CAT 9K3979 10/17
Figure 6. SmartGrid (left, 95 kVp, 2.2 mAs, 65 in. SID) and processing without SmartGrid (middle) compared to grid (right, 105
kVp, 3.2 mAs, 65 in. SID).
Conclusion
SmartGrid processing provides image quality comparable to images acquired with an anti-scatter grid at a reduced patient dose in
bedside chest imaging. The benefits of grid-like image quality without the use of an anti-scatter grid can lead to improved work
flow and ease of imaging for radiographers, producing a win-win for a busy hospital.
References
1
Christensen, Edward E., et al. Christensen's Physics of Diagnostic Radiology. Lea & Febiger, 1990.
2
T.L. Herrmann, T. L. Fauber, et. al. "Best Practices in Digital Radiography."(2012) (https://www.asrt.org/docs/default-
source/publications/whitepapers/asrt12_bstpracdigradwhp_final.pdf).
3
J. C. Wandtke, โ€œBedside Chest Radiography,โ€ Radiology, 190:1-10 (1994).
4
Radiological Society of North America (2010) RadLex: a lexicon for uniform indexing and retrieval of radiology information resources. Available
via http://www.rsna.org/radlex/. Accessed 22 Oct 2010.

More Related Content

What's hot

An Efficient Thresholding Neural Network Technique for High Noise Densities E...
An Efficient Thresholding Neural Network Technique for High Noise Densities E...An Efficient Thresholding Neural Network Technique for High Noise Densities E...
An Efficient Thresholding Neural Network Technique for High Noise Densities E...CSCJournals
ย 
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...IRJET Journal
ย 
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT sipij
ย 
Restoration of Old Documents that Suffer from Degradation
Restoration of Old Documents that Suffer from DegradationRestoration of Old Documents that Suffer from Degradation
Restoration of Old Documents that Suffer from DegradationIRJET Journal
ย 
P180203105108
P180203105108P180203105108
P180203105108IOSR Journals
ย 
Ijarcet vol-2-issue-7-2369-2373
Ijarcet vol-2-issue-7-2369-2373Ijarcet vol-2-issue-7-2369-2373
Ijarcet vol-2-issue-7-2369-2373Editor IJARCET
ย 
FPGA Implementation of Glaucoma Detection using Neural Networks
FPGA Implementation of Glaucoma Detection using Neural NetworksFPGA Implementation of Glaucoma Detection using Neural Networks
FPGA Implementation of Glaucoma Detection using Neural NetworksIRJET Journal
ย 
Visual Quality for both Images and Display of Systems by Visual Enhancement u...
Visual Quality for both Images and Display of Systems by Visual Enhancement u...Visual Quality for both Images and Display of Systems by Visual Enhancement u...
Visual Quality for both Images and Display of Systems by Visual Enhancement u...IJMER
ย 
Brain Tumor Area Calculation in CT-scan image using Morphological Operations
Brain Tumor Area Calculation in CT-scan image using Morphological OperationsBrain Tumor Area Calculation in CT-scan image using Morphological Operations
Brain Tumor Area Calculation in CT-scan image using Morphological Operationsiosrjce
ย 
Deep learning for image super resolution
Deep learning for image super resolutionDeep learning for image super resolution
Deep learning for image super resolutionPrudhvi Raj
ย 
An adaptive threshold segmentation for detection of nuclei in cervical cells ...
An adaptive threshold segmentation for detection of nuclei in cervical cells ...An adaptive threshold segmentation for detection of nuclei in cervical cells ...
An adaptive threshold segmentation for detection of nuclei in cervical cells ...csandit
ย 
Survey on Image Integration of Misaligned Images
Survey on Image Integration of Misaligned ImagesSurvey on Image Integration of Misaligned Images
Survey on Image Integration of Misaligned ImagesIRJET Journal
ย 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisCSCJournals
ย 
Multimodal Medical Image Fusion Based On SVD
Multimodal Medical Image Fusion Based On SVDMultimodal Medical Image Fusion Based On SVD
Multimodal Medical Image Fusion Based On SVDIOSR Journals
ย 
IRJET- A Review on Various Restoration Techniques in Digital Image Processing
IRJET- A Review on Various Restoration Techniques in Digital Image ProcessingIRJET- A Review on Various Restoration Techniques in Digital Image Processing
IRJET- A Review on Various Restoration Techniques in Digital Image ProcessingIRJET Journal
ย 
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...iosrjce
ย 
Tomosurgery Presentation
Tomosurgery PresentationTomosurgery Presentation
Tomosurgery Presentationngowdar
ย 

What's hot (18)

An Efficient Thresholding Neural Network Technique for High Noise Densities E...
An Efficient Thresholding Neural Network Technique for High Noise Densities E...An Efficient Thresholding Neural Network Technique for High Noise Densities E...
An Efficient Thresholding Neural Network Technique for High Noise Densities E...
ย 
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...
Deblurring Image and Removing Noise from Medical Images for Cancerous Disease...
ย 
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT
A NOVEL ALGORITHM FOR IMAGE DENOISING USING DT-CWT
ย 
Restoration of Old Documents that Suffer from Degradation
Restoration of Old Documents that Suffer from DegradationRestoration of Old Documents that Suffer from Degradation
Restoration of Old Documents that Suffer from Degradation
ย 
P180203105108
P180203105108P180203105108
P180203105108
ย 
Ijarcet vol-2-issue-7-2369-2373
Ijarcet vol-2-issue-7-2369-2373Ijarcet vol-2-issue-7-2369-2373
Ijarcet vol-2-issue-7-2369-2373
ย 
FPGA Implementation of Glaucoma Detection using Neural Networks
FPGA Implementation of Glaucoma Detection using Neural NetworksFPGA Implementation of Glaucoma Detection using Neural Networks
FPGA Implementation of Glaucoma Detection using Neural Networks
ย 
Visual Quality for both Images and Display of Systems by Visual Enhancement u...
Visual Quality for both Images and Display of Systems by Visual Enhancement u...Visual Quality for both Images and Display of Systems by Visual Enhancement u...
Visual Quality for both Images and Display of Systems by Visual Enhancement u...
ย 
Brain Tumor Area Calculation in CT-scan image using Morphological Operations
Brain Tumor Area Calculation in CT-scan image using Morphological OperationsBrain Tumor Area Calculation in CT-scan image using Morphological Operations
Brain Tumor Area Calculation in CT-scan image using Morphological Operations
ย 
Deep learning for image super resolution
Deep learning for image super resolutionDeep learning for image super resolution
Deep learning for image super resolution
ย 
An adaptive threshold segmentation for detection of nuclei in cervical cells ...
An adaptive threshold segmentation for detection of nuclei in cervical cells ...An adaptive threshold segmentation for detection of nuclei in cervical cells ...
An adaptive threshold segmentation for detection of nuclei in cervical cells ...
ย 
Survey on Image Integration of Misaligned Images
Survey on Image Integration of Misaligned ImagesSurvey on Image Integration of Misaligned Images
Survey on Image Integration of Misaligned Images
ย 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
ย 
Multimodal Medical Image Fusion Based On SVD
Multimodal Medical Image Fusion Based On SVDMultimodal Medical Image Fusion Based On SVD
Multimodal Medical Image Fusion Based On SVD
ย 
IRJET- A Review on Various Restoration Techniques in Digital Image Processing
IRJET- A Review on Various Restoration Techniques in Digital Image ProcessingIRJET- A Review on Various Restoration Techniques in Digital Image Processing
IRJET- A Review on Various Restoration Techniques in Digital Image Processing
ย 
Sersc 3.org (1)
Sersc 3.org (1)Sersc 3.org (1)
Sersc 3.org (1)
ย 
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...
A Novel Adaptive Denoising Method for Removal of Impulse Noise in Images usin...
ย 
Tomosurgery Presentation
Tomosurgery PresentationTomosurgery Presentation
Tomosurgery Presentation
ย 

Similar to Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Imaging

Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...
Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...
Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...IRJET Journal
ย 
Classification of mammograms based on features extraction techniques using su...
Classification of mammograms based on features extraction techniques using su...Classification of mammograms based on features extraction techniques using su...
Classification of mammograms based on features extraction techniques using su...CSITiaesprime
ย 
76201966
7620196676201966
76201966IJRAT
ย 
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...IJTET Journal
ย 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Hauke Sann
ย 
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...IJCSIS Research Publications
ย 
IRJET- Detection of Breast Asymmetry by Active Contour Segmentation Techn...
IRJET-  	  Detection of Breast Asymmetry by Active Contour Segmentation Techn...IRJET-  	  Detection of Breast Asymmetry by Active Contour Segmentation Techn...
IRJET- Detection of Breast Asymmetry by Active Contour Segmentation Techn...IRJET Journal
ย 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System Carestream
ย 
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...csandit
ย 
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and Thresholding
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and ThresholdingIRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and Thresholding
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and ThresholdingIRJET Journal
ย 
Comparative performance analysis of segmentation techniques
Comparative performance analysis of segmentation techniquesComparative performance analysis of segmentation techniques
Comparative performance analysis of segmentation techniquesIAEME Publication
ย 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)IJERD Editor
ย 
MRI_processing
MRI_processingMRI_processing
MRI_processingLitu Rout
ย 
Az4102375381
Az4102375381Az4102375381
Az4102375381IJERA Editor
ย 
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...IRJET Journal
ย 
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...IRJET Journal
ย 
TL-DTC_Michael Ravoo.pdf
TL-DTC_Michael Ravoo.pdfTL-DTC_Michael Ravoo.pdf
TL-DTC_Michael Ravoo.pdfFlorinSpainhour
ย 
Trends and Techniques of Medical Image Analysis and Brain Tumor Detection
Trends and Techniques of Medical Image Analysis and Brain Tumor DetectionTrends and Techniques of Medical Image Analysis and Brain Tumor Detection
Trends and Techniques of Medical Image Analysis and Brain Tumor DetectionIRJET Journal
ย 
Brain Tumor Detection From MRI Image Using Deep Learning
Brain Tumor Detection From MRI Image Using Deep LearningBrain Tumor Detection From MRI Image Using Deep Learning
Brain Tumor Detection From MRI Image Using Deep LearningIRJET Journal
ย 
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...IRJET Journal
ย 

Similar to Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Imaging (20)

Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...
Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...
Breast Cancer Detection and Classification using Ultrasound and Ultrasound El...
ย 
Classification of mammograms based on features extraction techniques using su...
Classification of mammograms based on features extraction techniques using su...Classification of mammograms based on features extraction techniques using su...
Classification of mammograms based on features extraction techniques using su...
ย 
76201966
7620196676201966
76201966
ย 
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
Supervised Blood Vessel Segmentation in Retinal Images Using Gray level and M...
ย 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
ย 
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...
A Hybrid Approach for Classification of MRI Brain Tumors Using Genetic Algori...
ย 
IRJET- Detection of Breast Asymmetry by Active Contour Segmentation Techn...
IRJET-  	  Detection of Breast Asymmetry by Active Contour Segmentation Techn...IRJET-  	  Detection of Breast Asymmetry by Active Contour Segmentation Techn...
IRJET- Detection of Breast Asymmetry by Active Contour Segmentation Techn...
ย 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System
ย 
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...
Hybrid Technique Based on N-GRAM and Neural Networks for Classification of Ma...
ย 
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and Thresholding
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and ThresholdingIRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and Thresholding
IRJET- Brain Tumor Detection using Hybrid Model of DCT DWT and Thresholding
ย 
Comparative performance analysis of segmentation techniques
Comparative performance analysis of segmentation techniquesComparative performance analysis of segmentation techniques
Comparative performance analysis of segmentation techniques
ย 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)
ย 
MRI_processing
MRI_processingMRI_processing
MRI_processing
ย 
Az4102375381
Az4102375381Az4102375381
Az4102375381
ย 
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...
Augmentation of Multimodal 3D Magnetic Resonance Imaging using Generative Adv...
ย 
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...
Precise Foreground Extraction, Bias Correction and Segmentation of Intensity ...
ย 
TL-DTC_Michael Ravoo.pdf
TL-DTC_Michael Ravoo.pdfTL-DTC_Michael Ravoo.pdf
TL-DTC_Michael Ravoo.pdf
ย 
Trends and Techniques of Medical Image Analysis and Brain Tumor Detection
Trends and Techniques of Medical Image Analysis and Brain Tumor DetectionTrends and Techniques of Medical Image Analysis and Brain Tumor Detection
Trends and Techniques of Medical Image Analysis and Brain Tumor Detection
ย 
Brain Tumor Detection From MRI Image Using Deep Learning
Brain Tumor Detection From MRI Image Using Deep LearningBrain Tumor Detection From MRI Image Using Deep Learning
Brain Tumor Detection From MRI Image Using Deep Learning
ย 
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...
IRJET- Analysis of Alzheimerโ€™s Disease on MRI Image using Transform and Featu...
ย 

More from Carestream

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Carestream
ย 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationCarestream
ย 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareCarestream
ย 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...Carestream
ย 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsCarestream
ย 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-rayCarestream
ย 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentCarestream
ย 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityCarestream
ย 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Carestream
ย 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesCarestream
ย 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seramCarestream
ย 
Cloud Security
Cloud Security Cloud Security
Cloud Security Carestream
ย 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Carestream
ย 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareCarestream
ย 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundCarestream
ย 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Carestream
ย 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Carestream
ย 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Carestream
ย 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16Carestream
ย 
Casa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCasa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCarestream
ย 

More from Carestream (20)

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
ย 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of Operation
ย 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization Software
ย 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
ย 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
ย 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-ray
ย 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray Equipment
ย 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image Quality
ย 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions
ย 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
ย 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seram
ย 
Cloud Security
Cloud Security Cloud Security
Cloud Security
ย 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
ย 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
ย 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
ย 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
ย 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
ย 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
ย 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
ย 
Casa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCasa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital Italy
ย 

Recently uploaded

VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Memriyagarg453
ย 
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetTirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
Call Girls Service In Goa ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls By Russian Call Girl...russian goa call girl and escorts service
ย 
VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171Call Girls Service Gurgaon
ย 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthanindiancallgirl4rent
ย 
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetMangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012adityaroy0215
ย 
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...Ahmedabad Call Girls
ย 
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…gragmanisha42
ย 
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetJalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…gragmanisha42
ย 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...indiancallgirl4rent
ย 
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetChandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meetpriyashah722354
ย 
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetHubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...Sheetaleventcompany
ย 
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR Call G...
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR   Call G...โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR Call G...Gfnyt.com
ย 
9316020077๐Ÿ“žGoa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
9316020077๐Ÿ“žGoa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa9316020077๐Ÿ“žGoa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa
9316020077๐Ÿ“žGoa Call Girls Numbers, Call Girls Whatsapp Numbers Goarussian goa call girl and escorts service
ย 
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetOzhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetNanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetCall Girls Service
ย 
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near MeRussian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Memriyagarg453
ย 

Recently uploaded (20)

VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
ย 
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetTirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Tirupati Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Call Girls Service In Goa ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa ๐Ÿ’‹ 9316020077๐Ÿ’‹ Goa Call Girls By Russian Call Girl...
ย 
VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171
ย 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
ย 
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetMangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Mangalore Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
ย 
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) ๐Ÿ’“ 9920725232 ๐Ÿ’“High Profile Call Girls Navi Mumbai You Can Get The S...
ย 
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…
Call Girl Gorakhpur * 8250192130 Service starts from just โ‚น9999 โœ…
ย 
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetJalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Jalna Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…
Russian Call Girls Kota * 8250192130 Service starts from just โ‚น9999 โœ…
ย 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
ย 
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetChandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls ๐Ÿ‘™ 7001035870 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetHubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Hubli Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...
Call Girl In Zirakpur โค๏ธโ™€๏ธ@ 9988299661 Zirakpur Call Girls Near Me โค๏ธโ™€๏ธ@ Sexy...
ย 
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR Call G...
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR   Call G...โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
โค๏ธโ™€๏ธ@ Jaipur Call Girls โค๏ธโ™€๏ธ@ Meghna Jaipur Call Girls Number CRTHNR Call G...
ย 
9316020077๐Ÿ“žGoa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
9316020077๐Ÿ“žGoa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa9316020077๐Ÿ“žGoa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa
9316020077๐Ÿ“žGoa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
ย 
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetOzhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real MeetNanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
Nanded Call Girls ๐Ÿ‘™ 6297143586 ๐Ÿ‘™ Genuine WhatsApp Number for Real Meet
ย 
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near MeRussian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
ย 

Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Imaging

  • 1. White Paper | SmartGrid Image Quality Impact of SmartGrid Processing in Bedside Chest Imaging Authors: Lori L. Barski (MS), Dong Yang (Ph.D.), Mary Couwenhoven (MS) Introduction Scattered radiation is known to degrade image quality in diagnostic X-ray imaging. Traditional methods of reducing scatter are collimation, utilizing an air-gap, and/or utilizing an anti-scatter grid [1]. Best practices in radiography prescribe collimation on every exam [2] and, while this is helpful, it is not completely sufficient due to scatter that occurs within the patient. Likewise, air-gap techniques are helpful [1], but are not practical at the bedside. Anti-scatter grids are by far the most popular means of reducing scatter in portable imaging, but they present challenges to radiographers (e.g., positioning and alignment) and increases dose to the patient [1,3]. A new image processing tool, SmartGrid, has been developed that compensates for the effects of X-ray scatter in an image, and produces results comparable to those of a physical anti- scatter grid. Figure 1 shows a pair of portable chest images captured with a grid (top left), without a grid (top right), and the results with SmartGrid (bottom left). The scatter distribution image (bottom right) demonstrates regions of the image that are most impacted by scatter (lighter tone). SmartGrid processing estimates the scatter distribution and removes it, resulting in an image with improved contrast. Many physical factors affect scatter: energy spectrum of the beam, thickness and material composition of the object, and collimation, to name a few. SmartGrid accommodates these variables automatically and results in image quality that approximates anti-scatter grid visual performance. The SmartGrid Algorithm The SmartGrid algorithm (Figure 2) is an enhancement algorithm that improves image contrast by suppressing scatter in the image. The fundamental concept is one of developing a scatter distribution image, which is a representation of the scatter contained in the image, and then subtracting it from the original input image. Figure 1. Portable chest exam captured with a grid (top left) and without a grid (top right, grid exposure 1.25X non-grid), and then processed with SmartGrid (bottom left) with the corresponding scatter distribution image (bottom right). The scatter distribution image is developed using information from the image in both linear exposure space as well as attenuation space, which is a log transformation of the linear data. Segmentation is done to focus the development of the scatter field on relevant anatomical data and is used to compute the mean linear exposure of the input image. Parameters used to estimate scatter are determined. These include the scatter-to-primary ratio (SPR) and a curvature parameter used to control perturbation of the scatter distribution as the scatter field is developed.
  • 2. White Paper | SmartGrid 2 Figure 2. Flow diagram of the SmartGrid algorithm. The scatter distribution image is computed based upon the assumption that every object exposed by X-rays has a basic scatter distribution that is characterized by a certain level of energy and scatter intensity variation across the whole object field of view. Adaptive updating of the scatter intensity across the entire object field of view is performed in a repetitive fashion based upon the SPR parameter for a prescribed number of iterations. SmartGrid Image Quality In order to assess the image quality impact of SmartGrid processing, a clinical study evaluating paired grid and non-grid portable chest images was conducted at the University of Rochester Medical Center with IRB oversight. Patients and healthy volunteers, under informed consent and who met the inclusion criteria (age greater than 18, body mass index greater than 25 and a chest circumference greater than 90 cm, non- pregnant females, and no history of high-radiation exposure), had portable chest exams taken with and without anti-scatter grids. Grid/non-grid usage was alternated for in-patient subjects over a time span of days, whereas healthy volunteers were imaged within a time span of minutes. Techniques were selected and documented by the radiographer at the time of the exam, following department guidelines for the site's standard of care. Two sets of pair-wise comparisons were evaluated (side-by- side) for each grid/non-grid image pair: SmartGrid (SG) vs. Grid (n=44) and SmartGrid vs. standard default processing (EVP Plus processing available in V5.7, n=44). Pairs were presented on a PACS in randomized order to five board-certified thoracic radiologists (3 - 57 yrs. experience) and anatomical regions (retro-cardiac, carina/airways, sub-diaphragm, chest wall, and lung parenchyma) were rated using a +/- 3 point preference scale. For each pair, the reader picked the image with the best region (e.g., retro-cardiac) and rated it as +/- 0, 1, 2, or 3, where the sign indicated the left or right monitor (i.e., left- preferred was indicated with negative numbers) and the magnitude corresponded to the degree of clinical impact (Table 1). Readers were free to pan, zoom, and adjust window width and window level as desired. Upon completion of the reader study, response data was decoded such that positive values indicated preference in favor of SmartGrid. Additionally, each image of the pair was rated for diagnostic quality using a 4-point RadLex scale [4] (Table 2), allowing decimal ratings (e.g., 3.3) between the integer anchors. Score Definition Left Preferred -3 Image displayed on left is strongly preferred Clinical Impact -2 Image displayed on left is moderately preferred -1 Image displayed on left is slightly preferred No Clinical Impact0 No preference between the images Right Preferred 1 Image displayed on right is slightly preferred 2 Image displayed on right is moderately preferred Clinical Impact 3 Image displayed on right is strongly preferred Table 1. Preference rating scale.
  • 3. White Paper | SmartGrid 3 Score Term Definition 1 Non- diagnostic Unacceptable for diagnostic purposes. Little or no clinically-usable diagnostic information (e.g., gross underexposure or extensive motion artifact) and should be repeated. 2 Limited Acceptable, with some technical defect (motion artifact, body habitus/poor X-ray penetration, or patient positioning, but still adequate for diagnostic purposes). 3 Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4 Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated. Table 2. Diagnostic quality RadLex rating scale. Clinical Study Results On average, no significant overall preference difference was detected between the SmartGrid and anti-scatter grid pairs (mean difference = 0.0, 95% CI [-0.12, 0.12], p =0.941) with an average grid dose factor of 2.1x (i.e., SmartGrid exposures were, on average, acquired with approximately half of the grid dose). Figure 3 is an example of SmartGrid processing (left) compared to the grid image (right), which was acquired at approx 3.5x the SmartGrid dose. The average overall preference difference was 1.0. Table 3 summarizes the pair- wise preference differences for the anatomical regions. Figure 3. SmartGrid image (left): 95 kVp, 2.5 mAs, 61 in. SID, compared to grid image (right): 110 kVp, 4 mAs, 58 in. SID. Average overall preference difference is 1.0. Grid vs. SG Preference Mean Difference Confidence Interval p-value Conclusion Over All 0.0 -0.12, 0.12 0.941 No difference Retrocardiac -0.2 -0.28, -0.05 0.007* In favor of grid Carina/Airways -0.2 -0.29, -0.06 0.003* In favor of grid Subdiaphragm -0.1 -0.18, 0.04 0.186 No difference Chest Wall 0.1 0.04, 0.23 0.004* In favor of SG Parenchyma 0.1 -0.29, 0.17 0.170 No difference Table 3. SmartGrid vs. Grid Preference rating results (* indicates statistical significance at an alpha risk of 5%).
  • 4. White Paper | SmartGrid 4 The difference in diagnostic quality ratings of SmartGrid vs. grids was significant in favor of SmartGrid processing (mean difference = 0.03, 95% CI [0.00, 0.05], p = 0.044). Not surprising, is the result that the SmartGrid processing has significantly higher preference overall compared to images without SmartGrid enhancement (mean difference = 0.8, 95% CI [0.69, 0.84], p < 10-3 ) as well as for all anatomical regions (all p-values < 10-3 ). Likewise, diagnostic quality differences were significantly in favor of SmartGrid (p < 10-3 ). Figure 4 illustrates a typical chest wall appearance as seen in a SmartGrid (left) image and the corresponding grid image (right). Scatter in this region has the potential to reduce the visualization of the pleura and the boney structures of the outer rib cage. Figure 4. Chest wall example. SmartGrid (Left): 90 kVp, 2.6 mAs, 60 in. SID; Grid (right): 110 kVp, 4 mAs, 68 in. SID. Figure 5 demonstrates the retrocardiac region of a SmartGrid image (left) and grid (right) pair. The apparent lack of contrast in the grid image is likely caused by grid misalignment, illustrating exactly why a software solution to scatter compensation is of great benefit in the ICU. Figure 6 is another portable chest comparison with the SmartGrid processing on the right, default processing of the same image without SmartGrid enabled, and the grid exposure shown on the right. Note that the non-grid image was acquired at approximately half of the grid exposure. Figure 5. Retrocardic example. SmartGrid (left): 95 kVp, 3.2 mAs, 72 in. SID; Grid (right) 112 kVp, 7.2 mAs, 72 in. SID.
  • 5. White Paper | SmartGrid carestream.com ยฉCarestream Health, Inc., 2017. CARESTREAM is a trademark of Carestream Health. CAT 9K3979 10/17 Figure 6. SmartGrid (left, 95 kVp, 2.2 mAs, 65 in. SID) and processing without SmartGrid (middle) compared to grid (right, 105 kVp, 3.2 mAs, 65 in. SID). Conclusion SmartGrid processing provides image quality comparable to images acquired with an anti-scatter grid at a reduced patient dose in bedside chest imaging. The benefits of grid-like image quality without the use of an anti-scatter grid can lead to improved work flow and ease of imaging for radiographers, producing a win-win for a busy hospital. References 1 Christensen, Edward E., et al. Christensen's Physics of Diagnostic Radiology. Lea & Febiger, 1990. 2 T.L. Herrmann, T. L. Fauber, et. al. "Best Practices in Digital Radiography."(2012) (https://www.asrt.org/docs/default- source/publications/whitepapers/asrt12_bstpracdigradwhp_final.pdf). 3 J. C. Wandtke, โ€œBedside Chest Radiography,โ€ Radiology, 190:1-10 (1994). 4 Radiological Society of North America (2010) RadLex: a lexicon for uniform indexing and retrieval of radiology information resources. Available via http://www.rsna.org/radlex/. Accessed 22 Oct 2010.