FAULTS IN X-RAY TUBE
AND ITS CARE
PRESENTED BY
Saru Gosain
BSc.MIT 2nd year
(2017)
COMMON CAUSES OF
TUBE FAILURE
• Extremely high voltage
• Excessive heat generation
• Poor cooling system
• Rough-handling/ Careless
• Aging
COMMON FAULTS, THEIR
CAUSES & REMEDY
A. FAULTS IN X-RAY TUBE
HOUSING
1. Cracking of x-ray tube
housing
 Escape of oil from the housing
 Insulation level &Cooling rate
are reduced
Causes
• Expansion of air & oil present
inside it and increased in its
internal pressure.
Remedy
• Excluding air
• Use of metal bellows to allow oil
expansion as well as to operate
as “microswitch”
Leakage oil testing devices
Cont.…
2. Radiation leakage
 Increases patient’s as well as
personnel radiation dose.
 Should not exceed 100 mR/hr
at 1m
Causes
• Cracking of the tube housing
• Improper shielding
Remedy
• Proper shielding should be
done .
• Replace the tube with new
one as soon as possible.
Cont.…
B. FAULTS IN GLASS/METAL
ENVELOPE
1. Loss of vacuum
 Production of secondary electrons
 Result in the variation in the number,
reduced speed of electrons impinging
on the target, energy of x-ray
produced
 Production couldn’t be controlled
independently.
Causes
• Crazing of the glass enclosure due to
difference in coefficients of expansion
of glass and metal
• Inactivity
Cont.…
Remedy
 Use of special alloys having the
same coefficients of linear
expansion as Pyrex glass are used.
C. FAULTS IN FILAMENT
1. Vaporization of filament
a) Breakage of filament
 No exposure takes place
Causes
• Excessive heating of filament for
the longer period of time
Cont.…
Remedy
• Use of automatic filament-boosting
circuit
• Use of 1-2 % Thorium with tungsten
filament
b) Tube arcing
 Produces a bronzed
-colored “sunburn” color
 Filters the x-ray beam
 “Crazing” of the tube
Causes
• Evaporation of the filament
Remedy
• Use of metal enclosure instead of
glass enclosure
Cont.…
D. FAULTS IN ANODE
1. Kinking of the target material
 Results in distortion of the anode angle, narrowing of the
beam, greater geometric unsharpness.
Causes
• Difference in thermal expansion of metal on the surface
& metal immediately beneath the surface.
Remedy
• 10% Rhenium is added to 90% of tungsten target
• Manufacture anode disc with slits in the target surface to
allow the material in the focal track to expand without
producing the mechanical tension.
Cont….
2. Roughening and pitting of anode surface
 Scattering of the x-rays
 Absorption of x-rays by the
target itself.
 Results in diminish x-ray output
Cause
• Excessive thermal stress
Remedy
• Use of lowest necessary power
• Largest possible focal spot
• Longer exposures at reduced power
For
FIGURE 6-14 Comparison of smooth, shiny appearances of rotating anodes when new (A) versus
their appearance after failure
(B–D). Examples of anode separation and surface melting shown were caused by slow rotation
caused by bearing damage (B),
repeated overload (C), and exceeding of maximum heat storage capacity (D). (Courtesy Philips
Medical Systems.)
A B
C D
Cont….
3. Imbalance ball bearings
 Jamming
 Increased bearing noise
 Anode stops rotating
Causes
• Expansion due to excessive
heating of the rotor bearings
• Increased rotational friction
Remedy
• Poor conductor of heat(Mo) is
used as anode stem
• Metallic Lubricants such as Silver
& liquid Gallium are used.
Cont…
4. Broken stator windings
 No power supply to make the
anode rotate.
Causes
• Heavy radiography exposures
• Excessive heat in the target
material, rotor and its bearings,
• Difference in thermal expansion of
the various components used in the
anode.
Remedy
• Provide safety circuit to terminate
exposure.
OTHER FAULTS
A. DEFICIENCIES IN MANUFACTURING
1. Unlike tubes
2. Weed out by test:- Check high voltage stability, cathode
emission, focal spot size, thermal loading and for
rotating anode tubes additional tests such as noise,
vibration, coast time etc. are tested. Not meeting these
specifications are rejected.
3. Hold period:- If tubes are held for 2-4 weeks vacuum
may leak
4. Improper materials:- O2 free Cu bar can cause vacuum
leaks
Cont….
B. APPLICATION MISMATCH:- e.g., poor diagnosis and
radiation burns while using standard diagnostic tubes to
produce mammograms
C. IMPROPER DRIVE BY THE POWER SUPPLY:-
1. Filament life is reduced 2-10 times by DC filament than
AC filament.
2. Magnetic materials ( e.g., iron, nickel, cobalt used for
glass-to-metal seal) can be subjected to magnetic
hysteresis, eddy currents etc.)
3. Filament limit:- maximum output current of the filament
is limited
CARE OF THE X-RAY
TUBE
1. WARM UP THE ANODE FOLLOWING MANUFACTURER’S
RECOMMENDATIONS
 Performed if the system not been energized for ≥ 2 hrs
 To prevent tube damage from thermal shock
WARM UP PROCEDURES
Make 1st 3-exposures waiting approximately 30s between
each exposures @
75kVp 200mA 1s
Wait 60s, then take 1 exposure @
85kVp 200mA ½ s
Wait 60s, then take the last exposure @
100kVp 200mA ¼ s
Cont….
2. MINIMIZE FILAMENT BOOST(“PREP”) TIME
 Current applied for too long will shorten filament life and
will lead to unstable operation.
Cont…
3. USE LOWER TUBE CURRENT (mA)
 High tube current will cause evaporation of the tungsten
 Whenever possible, use a lower mA station and a longer
exposure time to arrive at the desired mAs.
Cont….
4. FOLLOW RATING CHARTS AND ANODE AND
HOUSING HEATING/COOLING CHARTS
 Operation beyond published ratings will result in
premature focal track wear and tear.
TUBE RATING CHART
o The chart which provides the information about the
details of the electrical loads which may be applied to the
tube without damaging it & the safe duration of such
loads is called tube rating chart.
Cont….
o In considering the tube rating three characteristics are
encountered:
 Ability of x-ray tube to withstand a single exposure
 Ability of x-ray tube to function despite rapid exposure
 Ability of x-ray tube to withstand multiple exposure during
several hours of heavy use.
o There are 3 types
A. RADIOGRAPHIC RATING CHARTS
B. ANODE HEATING/COOLING CHART
C. HOUSING HEATING/COOLING CHART
RADIOGRAPHIC
RATING CHART
• The safe limit within
which x-ray tubes can be
operated for a single
exposure can be easily
determined by it.
• It is the most imp
because it conveys
which radiographic
techniques are safe or
unsafe for tube
operation.
• Microprocessors are
used to allow only safe
exposures.
Rating chart of x-ray room
no.1
Rating chart of x-ray room
no.1
ANODE
HEATING/COOLING
CHART
• Determines the
maximum heat
capacity of the
anode.
• Determines the
length of time the
tube must be
allowed to cool
before additional
exposures can be
made.
HOUSING
HEATING/COOLING
CHART
• Determines the
maximum heat
capacity(1-1.5
million HU) of the
tube housing.
• Complete cooling
after maximum
heat capacity
may take 1-2 hrs.
Cont.…
• Thermal rating can be determined by calculating
Energy deposited in the target(E)
=kVp × mA × exposure time in sec
• Unit is heat units(HU)
• SI unit is Joule(J)
1 HU= 0.71 J
1 J= 1.4 HU
• Now the Energy deposited in the target(E)
=kVp × mA × s J
CARE OF THE X-RAY
TUBE
5. LIMIT OPERATION TO 80% OF MAXIMUM SINGLE
EXPOSURE RATINGS
 It will minimize the reduction in radiation output
associated with a roughened focal track.
6. DO NOT EXCEED ANODE THERMAL CAPACITY OR
DISSIPATION RATE OF THE TARGET.
7. DO NOT MAKE HIGH mA EXPOSURES ON A COLD
TARGET.
8. AVOID LONG INTERVALS BETWEEN SPOT-FILMS.
o Most common “holdover period”(during which the rotor is
kept at high speed before the rotor brake cycle is
allowed to occur) is approximately 25 s between spot
films
Cont.….
9. LIMIT ROTOR START/STOP OPERATIONS
o Minimum of 30-40 s should be between starts
10.DONOT ROTATE THE TUBE HOUSING RAPIDLY
o Tension or pressure will produce to the x-ray tube.
o Gyroscopic effect( resistance to turns in any direction)
may crack or otherwise damage the rotor.
Cont.…
11.SOUND COOLING SYSTEM SHOULD BE PROVIDED
o Enormous heat(99%) is generated in the anode during
the exposure.
o Heat can be dissipated one of the 3 ways:-
RADIATION
CONDUCTION
CONVECTION.
o During exposures , heat is primarily dissipated by
radiation whereas between exposures it is primarily
dissipated by conduction.
 General life of x-ray tubes is ten of thousands of
exposures.
 And CT tubes are now guaranteed for 50,000 exposures
ANY
QUESTIONS???
Reference
• Christensen’s PHYSICS OF DIAGNOSTIC
RADIOLOGY(4TH EDITION)
• Stewart Carlyle Bushong’s RADIOLOGIC SCIENCE
FOR TECHNOLOGISTS (10TH EDITION)
• En.m.Wikipedia.org
• www.studentradiographer.com
• www.Radiopaedia.org
• www.spellmahv.com
• www.scribd.com
• www.slideshare.com
• www.sprawls.com
Faults in x ray tube and its care

More Related Content

What's hot

Computed radiography
Computed radiographyComputed radiography
Computed radiographyIllyasmk
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
Computed radiography
Computed radiographyComputed radiography
Computed radiographyAnjan Dangal
 
responsibility of radiographer
responsibility of radiographerresponsibility of radiographer
responsibility of radiographerBISHAL KHANAL
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyTarun Goyal
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiographyPriyanka Parimala
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
Beam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarBeam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarSushilPattar
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tubeairwave12
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxThejaTej6
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screensTarun Goyal
 
Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectJoaquinLuisRinehart1
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageSHASHI BHUSHAN
 
Beam restriction
Beam restrictionBeam restriction
Beam restrictionaslam bs
 

What's hot (20)

Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Modern x-ray tube
Modern x-ray tubeModern x-ray tube
Modern x-ray tube
 
responsibility of radiographer
responsibility of radiographerresponsibility of radiographer
responsibility of radiographer
 
digital radiography
digital radiographydigital radiography
digital radiography
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and Fluoroscopy
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiography
 
Beam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarBeam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattar
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tube
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screens
 
Filters
FiltersFilters
Filters
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effect
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic image
 
Beam restriction
Beam restrictionBeam restriction
Beam restriction
 
Grids
GridsGrids
Grids
 

Similar to Faults in x ray tube and its care

Similar to Faults in x ray tube and its care (20)

Faults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarFaults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magar
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Cttubeanddetectors
Cttubeanddetectors Cttubeanddetectors
Cttubeanddetectors
 
7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).ppt7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).ppt
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
X-Ray Tube.pptx
X-Ray Tube.pptxX-Ray Tube.pptx
X-Ray Tube.pptx
 
X ray tube
X ray tube X ray tube
X ray tube
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Conventional X-RAY Tube
 Conventional X-RAY Tube Conventional X-RAY Tube
Conventional X-RAY Tube
 
Introduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxIntroduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptx
 
Digital testing of high voltage circuit breaker
Digital testing of high voltage circuit breakerDigital testing of high voltage circuit breaker
Digital testing of high voltage circuit breaker
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray Tube
 
X ray tube
X ray tubeX ray tube
X ray tube
 
3 cp corrosion control
3 cp corrosion control3 cp corrosion control
3 cp corrosion control
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptx
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptx
 
medical equipment lec 2 x-rays
medical equipment lec 2 x-raysmedical equipment lec 2 x-rays
medical equipment lec 2 x-rays
 

More from SaruGosain

Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skullSaruGosain
 
Jugular venography
Jugular venographyJugular venography
Jugular venographySaruGosain
 
Frontal venography
 Frontal venography Frontal venography
Frontal venographySaruGosain
 
Abdomen & Pelvis II
Abdomen & Pelvis IIAbdomen & Pelvis II
Abdomen & Pelvis IISaruGosain
 
Abdomen & pelvis part I
Abdomen & pelvis part IAbdomen & pelvis part I
Abdomen & pelvis part ISaruGosain
 
Oral cholecystography
Oral cholecystographyOral cholecystography
Oral cholecystographySaruGosain
 

More from SaruGosain (7)

Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skull
 
Jugular venography
Jugular venographyJugular venography
Jugular venography
 
Frontal venography
 Frontal venography Frontal venography
Frontal venography
 
Abdomen & Pelvis II
Abdomen & Pelvis IIAbdomen & Pelvis II
Abdomen & Pelvis II
 
Abdomen & pelvis part I
Abdomen & pelvis part IAbdomen & pelvis part I
Abdomen & pelvis part I
 
Oral cholecystography
Oral cholecystographyOral cholecystography
Oral cholecystography
 
X ray filters
X ray filtersX ray filters
X ray filters
 

Recently uploaded

Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipurparulsinha
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 

Recently uploaded (20)

Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 

Faults in x ray tube and its care

  • 1. { FAULTS IN X-RAY TUBE AND ITS CARE PRESENTED BY Saru Gosain BSc.MIT 2nd year (2017)
  • 2. COMMON CAUSES OF TUBE FAILURE • Extremely high voltage • Excessive heat generation • Poor cooling system • Rough-handling/ Careless • Aging
  • 3. COMMON FAULTS, THEIR CAUSES & REMEDY A. FAULTS IN X-RAY TUBE HOUSING 1. Cracking of x-ray tube housing  Escape of oil from the housing  Insulation level &Cooling rate are reduced Causes • Expansion of air & oil present inside it and increased in its internal pressure. Remedy • Excluding air • Use of metal bellows to allow oil expansion as well as to operate as “microswitch”
  • 5. Cont.… 2. Radiation leakage  Increases patient’s as well as personnel radiation dose.  Should not exceed 100 mR/hr at 1m Causes • Cracking of the tube housing • Improper shielding Remedy • Proper shielding should be done . • Replace the tube with new one as soon as possible.
  • 6. Cont.… B. FAULTS IN GLASS/METAL ENVELOPE 1. Loss of vacuum  Production of secondary electrons  Result in the variation in the number, reduced speed of electrons impinging on the target, energy of x-ray produced  Production couldn’t be controlled independently. Causes • Crazing of the glass enclosure due to difference in coefficients of expansion of glass and metal • Inactivity
  • 7. Cont.… Remedy  Use of special alloys having the same coefficients of linear expansion as Pyrex glass are used. C. FAULTS IN FILAMENT 1. Vaporization of filament a) Breakage of filament  No exposure takes place Causes • Excessive heating of filament for the longer period of time
  • 8. Cont.… Remedy • Use of automatic filament-boosting circuit • Use of 1-2 % Thorium with tungsten filament b) Tube arcing  Produces a bronzed -colored “sunburn” color  Filters the x-ray beam  “Crazing” of the tube Causes • Evaporation of the filament Remedy • Use of metal enclosure instead of glass enclosure
  • 9. Cont.… D. FAULTS IN ANODE 1. Kinking of the target material  Results in distortion of the anode angle, narrowing of the beam, greater geometric unsharpness. Causes • Difference in thermal expansion of metal on the surface & metal immediately beneath the surface. Remedy • 10% Rhenium is added to 90% of tungsten target • Manufacture anode disc with slits in the target surface to allow the material in the focal track to expand without producing the mechanical tension.
  • 10. Cont…. 2. Roughening and pitting of anode surface  Scattering of the x-rays  Absorption of x-rays by the target itself.  Results in diminish x-ray output Cause • Excessive thermal stress Remedy • Use of lowest necessary power • Largest possible focal spot • Longer exposures at reduced power For
  • 11. FIGURE 6-14 Comparison of smooth, shiny appearances of rotating anodes when new (A) versus their appearance after failure (B–D). Examples of anode separation and surface melting shown were caused by slow rotation caused by bearing damage (B), repeated overload (C), and exceeding of maximum heat storage capacity (D). (Courtesy Philips Medical Systems.) A B C D
  • 12. Cont…. 3. Imbalance ball bearings  Jamming  Increased bearing noise  Anode stops rotating Causes • Expansion due to excessive heating of the rotor bearings • Increased rotational friction Remedy • Poor conductor of heat(Mo) is used as anode stem • Metallic Lubricants such as Silver & liquid Gallium are used.
  • 13. Cont… 4. Broken stator windings  No power supply to make the anode rotate. Causes • Heavy radiography exposures • Excessive heat in the target material, rotor and its bearings, • Difference in thermal expansion of the various components used in the anode. Remedy • Provide safety circuit to terminate exposure.
  • 14. OTHER FAULTS A. DEFICIENCIES IN MANUFACTURING 1. Unlike tubes 2. Weed out by test:- Check high voltage stability, cathode emission, focal spot size, thermal loading and for rotating anode tubes additional tests such as noise, vibration, coast time etc. are tested. Not meeting these specifications are rejected. 3. Hold period:- If tubes are held for 2-4 weeks vacuum may leak 4. Improper materials:- O2 free Cu bar can cause vacuum leaks
  • 15. Cont…. B. APPLICATION MISMATCH:- e.g., poor diagnosis and radiation burns while using standard diagnostic tubes to produce mammograms C. IMPROPER DRIVE BY THE POWER SUPPLY:- 1. Filament life is reduced 2-10 times by DC filament than AC filament. 2. Magnetic materials ( e.g., iron, nickel, cobalt used for glass-to-metal seal) can be subjected to magnetic hysteresis, eddy currents etc.) 3. Filament limit:- maximum output current of the filament is limited
  • 16. CARE OF THE X-RAY TUBE 1. WARM UP THE ANODE FOLLOWING MANUFACTURER’S RECOMMENDATIONS  Performed if the system not been energized for ≥ 2 hrs  To prevent tube damage from thermal shock WARM UP PROCEDURES Make 1st 3-exposures waiting approximately 30s between each exposures @ 75kVp 200mA 1s Wait 60s, then take 1 exposure @ 85kVp 200mA ½ s Wait 60s, then take the last exposure @ 100kVp 200mA ¼ s
  • 17. Cont…. 2. MINIMIZE FILAMENT BOOST(“PREP”) TIME  Current applied for too long will shorten filament life and will lead to unstable operation.
  • 18. Cont… 3. USE LOWER TUBE CURRENT (mA)  High tube current will cause evaporation of the tungsten  Whenever possible, use a lower mA station and a longer exposure time to arrive at the desired mAs.
  • 19. Cont…. 4. FOLLOW RATING CHARTS AND ANODE AND HOUSING HEATING/COOLING CHARTS  Operation beyond published ratings will result in premature focal track wear and tear. TUBE RATING CHART o The chart which provides the information about the details of the electrical loads which may be applied to the tube without damaging it & the safe duration of such loads is called tube rating chart.
  • 20. Cont…. o In considering the tube rating three characteristics are encountered:  Ability of x-ray tube to withstand a single exposure  Ability of x-ray tube to function despite rapid exposure  Ability of x-ray tube to withstand multiple exposure during several hours of heavy use. o There are 3 types A. RADIOGRAPHIC RATING CHARTS B. ANODE HEATING/COOLING CHART C. HOUSING HEATING/COOLING CHART
  • 21. RADIOGRAPHIC RATING CHART • The safe limit within which x-ray tubes can be operated for a single exposure can be easily determined by it. • It is the most imp because it conveys which radiographic techniques are safe or unsafe for tube operation. • Microprocessors are used to allow only safe exposures.
  • 22. Rating chart of x-ray room no.1
  • 23. Rating chart of x-ray room no.1
  • 24. ANODE HEATING/COOLING CHART • Determines the maximum heat capacity of the anode. • Determines the length of time the tube must be allowed to cool before additional exposures can be made.
  • 25. HOUSING HEATING/COOLING CHART • Determines the maximum heat capacity(1-1.5 million HU) of the tube housing. • Complete cooling after maximum heat capacity may take 1-2 hrs.
  • 26. Cont.… • Thermal rating can be determined by calculating Energy deposited in the target(E) =kVp × mA × exposure time in sec • Unit is heat units(HU) • SI unit is Joule(J) 1 HU= 0.71 J 1 J= 1.4 HU • Now the Energy deposited in the target(E) =kVp × mA × s J
  • 27. CARE OF THE X-RAY TUBE 5. LIMIT OPERATION TO 80% OF MAXIMUM SINGLE EXPOSURE RATINGS  It will minimize the reduction in radiation output associated with a roughened focal track. 6. DO NOT EXCEED ANODE THERMAL CAPACITY OR DISSIPATION RATE OF THE TARGET. 7. DO NOT MAKE HIGH mA EXPOSURES ON A COLD TARGET. 8. AVOID LONG INTERVALS BETWEEN SPOT-FILMS. o Most common “holdover period”(during which the rotor is kept at high speed before the rotor brake cycle is allowed to occur) is approximately 25 s between spot films
  • 28. Cont.…. 9. LIMIT ROTOR START/STOP OPERATIONS o Minimum of 30-40 s should be between starts 10.DONOT ROTATE THE TUBE HOUSING RAPIDLY o Tension or pressure will produce to the x-ray tube. o Gyroscopic effect( resistance to turns in any direction) may crack or otherwise damage the rotor.
  • 29. Cont.… 11.SOUND COOLING SYSTEM SHOULD BE PROVIDED o Enormous heat(99%) is generated in the anode during the exposure. o Heat can be dissipated one of the 3 ways:- RADIATION CONDUCTION CONVECTION. o During exposures , heat is primarily dissipated by radiation whereas between exposures it is primarily dissipated by conduction.
  • 30.
  • 31.  General life of x-ray tubes is ten of thousands of exposures.  And CT tubes are now guaranteed for 50,000 exposures
  • 33. Reference • Christensen’s PHYSICS OF DIAGNOSTIC RADIOLOGY(4TH EDITION) • Stewart Carlyle Bushong’s RADIOLOGIC SCIENCE FOR TECHNOLOGISTS (10TH EDITION) • En.m.Wikipedia.org • www.studentradiographer.com • www.Radiopaedia.org • www.spellmahv.com • www.scribd.com • www.slideshare.com • www.sprawls.com