Atomic Spectroscopy based on
     Flame Atomization;
 Theory, Instrumentation and
          Application
    Nurul Auni binti Zainal Abidin
     Faculty of Applied Science
       UiTM Negeri Sembilan
ATOMIC SPECTROSCOPY

• Three techniques (methods) included in atomic
  spectroscopy:

  1) atomic absorption spectroscopy
  2) atomic emisssion spectroscopy
  3) atomic fluorescene spectroscopy
Atomic spectroscopy
• In order to perform atomic spectroscopy, atoms of
  the analyte must first be formed, usually in the form
  of an atomic vapor.

• Atomization – the process by which a sample is
  converted to an atomic vapor.

• Atomizer – a device used to converted a sample to
  an atomic vapor.
Types of atomizer

1) flame atomizer
2) plasma atomizer
3) electrothermal atomizer
Atomization process
• Solution of the analyte is evaporated rapidly at an
  elevated temperature to yield a finely divided solid.

• Further heating will break down into gaseous atoms.
Atomic Absorption Spectroscopy
• Uses absorption of light to measure the
  concentration of gas-phase atoms.

• Since samples are usually liquids or solids, the
  analyte atoms must be vapourised in a flame
  (or graphite furnace).




                                                     6
Atomic Absorption Spectroscopy
• The analyte concentration is determined
  from the amount of absorption.




                                            7
Atomic Absorption Spectroscopy
• The analyte concentration is determined from the
  amount of absorption.




                                                     8
Atomization process
• Solution of the analyte is evaporated rapidly at an
  elevated temperature to yield a finely divided solid.

• Further heating will break down into gaseous atoms.
Sample Atomization
Flame Atomization
• Nebulization – Conversion of the
  liquid sample to a fine spray.

• Desolvation - Solid atoms are
  mixed with the gaseous fuel.

• Volatilization – Solid atoms are
  converted to a vapor in the flame.

• There are three types of particles
  that exist in the flame:
  1) Atoms
  2) Ions
  3) Molecules
1. Types of Flames
               Fuel / Oxidant        Temperature
H–C C-H        acetylene / air      2100 – 2400 C (most common)
               acetylene / N2O      2600 – 2800 C
               acetylene / O2       3050 – 3150 C

• Selection of flame type depends on the volatilization of interest.

2. Flame Structure
• Interzonal region is the hottest part of the flame
   and best for atomic absorption.

• Oxidation of the atoms occurs in the secondary
  combustion zone where the atoms will form
  molecular oxides and are dispersed into the
  surroundings.
FUELS/OXIDANTS
• Low T flames : easily reduced elements
  (Cu, Pb, Zn, Cd)
• High T flames: difficult to reduce elements
  (e.g. alkaline earths).
• Fuels: natural gas, propane, butane, H2,
  and acetylene;
• Oxidants- Air and O2 (low temperature
  flames). N2O (high temperature flames).
• Flame characteristics:
• Sample enters flame, is vaporized, reduced
  and eventually oxidized.
• Exact region of flame in which each of
  these occurs depends upon:
• flow rate,
• drop size, and
• oxidizability of sample.
• Optimum position for flame for many
  metals.
Temperature Profile




Temperature profile in degrees Celsius for a natural gas-air flame
3. Burners

• Two types of burners in flame spectroscopy

  i. Turbulent flow (total consumption burner)
  ii. Laminar flow (premix burner)
Turbulent Flow Burner

• Nebulizer & burner are combined into a single unit.

• Sample is drawn up the capillary & nebulized

• Sample flow rate: 1 to 3 mL/min
Turbulent Flow Burner

• Advantage:
   – introduce relatively large & representative sample
     into the flame.

• Disadvantages:
   – A relatively short path length through flame
   – Problems with clogging of the tip
   – Burners noisy from electronic and auditory stand
     point
Laminar Flow Burner
• Sample is nebulized by the flow of
  oxidant past a capillary tip.

• Resulting aerosol then mixed with
  fuel & flows past a series of baffles
  that remove all but the finest
  droplets.

• Majority of the sample collects in
  the bottom of mixing chamber,
  drained to a waste container.

• Aerosol, oxidant & fuel are burned
  in a slotted burner that provided a
  flame, 5 – 10 cm in length.
Laminar Flow Burner

• Advantages:
  i. Provide quiet flame
  ii. Longer path length, enhance sensitivity &
      reproducibility

• Disadvantages:
  i. Lower rate of sample introduction
  ii. Possibility of selective evaporation of mixed
       solvents in the mixing chamber, create analytical
       uncertainties.
  iii. Mixing chamber contains a potentially explosive
       mixture that can flash back if the flow rates are
       too low.
Atomic Absorption Spectroscopy
(based on flame)

• Currently the most widely used of all the atomic
  methods.

• Simplicity, effectiveness, relatively low cost.

• Flame atomization best for reproducibility
  (precision)(<1%)
Atomic Absorption Spectroscopy
• AAS is commonly used for metal analysis

• A solution of a metal compound is sprayed
  into a flame and vaporises

• The metal atoms absorb light of a specific
  frequency, and the amount of light absorbed
  is a direct measure of the number of atoms of
  the metal in the solution

                                                  21
Atomic Absorption Spectroscopy
(based on flame)

• The spectra result form the atomized sample absorbing
  photons of radiation of the appropriate energy
  (wavelength).

• Energy of radiation absorbed by a vaporized atom is
  identical with that needed to bring about excitation to a
  higher electronic state.

• Transitions: excitation of an electron from ground state
  to a higher energy level                            E
                                                      1

                                                      E
INSTRUMENTATION
Spectrophotometers

• Contain same basic components as an instrument
  designed for absorption analysis of solution.

• Source        sample container (flame or hot surface)
            selector      detector              readout
  system
AA Spectrophotometer Design
1. Single Beam Instrument




    •The modulated power source can be replaced by a
    chopper
AA Spectrophotometer Design
2. Double-Beam Instrument
Double-Beam Instrument

• Radiation from hollow cathode lamp is split into 2 beams
   – One passes through the flame
   – The other around the flame

• A half-silvered mirror returns both beams to a single path
  then pass through the monochromator then detector.

• Note: monochromator placed between sample and
  detector
   – Eliminates most of the radiation originating from the
     flame.
A.     Line Sources

 LINE SOURCE – is required because width of the
 absorption line is very narrow

          Examples of line source used in AAS:




     Hollow cathode           Electrodeless discharge
      lamp (HCL)                    lamp (EDL)



                                                 done
1. Hollow Cathode Lamp (HCL)




• Light source in AA instrument

• Light from this lamp exactly light required for the
  analysis, even no monochromator is used.

• Atoms of metal tested are present within the lamp.
How does the Hollow Cathode Lamp works?

• When lamp on, atoms are supplied with energy

• Causes atoms elevate to excited states

• Upon returning to ground state, exactly the same
  wavelength that are useful in the analysis are emitted

• Since the analyzed metal with exactly the same energy
  levels that undergoes excitation

• Hollow cathode lamp MUST contain the element to be
  determined
Mechanism of excitation process in Hollow
  Cathode Lamp

• Lamp is a sealed glass envelope filled with argon (or
  Neon) gas

• When lamp is ON, Ne atoms are ionized with electrons
  drawn to anode (+ charge electrode)

• The Ne ions, Ne+ bombard the surface of the cathode (-
  charge electrode)

• Metal atoms, M, in the cathode are elevated to the
  excited state and are rejected from the surface as a
  result of this bombardment
Mechanism of excitation process in Hollow
  Cathode Lamp

• When atoms return to ground state, characteristic line
  spectrum of that atom is emitted

• This light is directed at the flame, where unexcited
  atoms of the same element absorb the radiation and
  are themselves raised to the excited state

• Absorbance is measured, and related to the
  concentration
Illustration of mechanism of excitation
             process in HCL
1. Electrodeless Discharge Lamp (EDL)
• Constructed of a metal or salt of interest sealed in a quartz tube
  filled with a noble gas (Ne or Ar) at low pressure (1 – 5 torr).




• The noble gas is ionized and accelerated by a strong radio-
  frequency (RF) or microwave field and in turn excite the metal or
  salt of interest.

• Provide radiant intensities usually one to two orders of magnitude
  greater than HCL.
Source Modulation


• Function:
     Employed in AAS to distinguish between
     the component of radiation arising from the
     source and the component of arising from
     the flame background

• Device: light chopper
Source Modulation
               Light chopper

• A chopper or a modulated power supply is used to
  modulate the source radiation that passes through the
  atomizer (flame).

• The chopper rotates resulting in an alternating atomic
  absorption and atomic emission signal.

• The signal from emission of radiation will be continuous
  and can be substracted from the modulated AA signal.
Light chopper

• Eliminate the effects of radiation from the flame
• Light is “chopped” with a rotating half-mirror so that
  detector sees alternating light intensities
• One moment, only light emitted by flame is read
  since the light from the source is cut off
• Next moment, light from both the flame emission &
  transmission of the source‟s light is measured since
  the source‟s light is allowed to pass
• The elements of the detector is such that the
  emission signal is substracted from the total signal &
  this difference is what we measured.
Questions

1. In flame atomic absorption spectroscopy, briefly
   describe the „atomization‟ process which the analyte
   undergoes.

2. Why is source modulation employed in atomic
   absorption spectroscopy? Name the device used for this
   purpose.

3. Name the common line source used in AAS.
INTERFERENCES


  Spectral
interference                    Chemical
                              interference


                Ionization
               interference
Spectral interference

• Arise when particulate matter from atomization
  scatters the incident radiation from the source.

• Or when absorption or emission of an
  interfering species either overlaps or lies so
  close to the analyte absorption that resolution
  by the monochromator becomes impossible.
Spectral interference (simple words)

• Is one in which the spectral line of the
  elements being determined overlaps with a
  spectral line (or band) from another element
  present in the sample.

• The effect of the element will also be measured
  & thus the results will be incorrect.
Spectral interference

The most common method of solving this problem

 i) Tune the monochromator to a different
 spectral line for the element of interest so that there is
 no overlap.

 ii) secondary lines (can be found in the literature)
Chemical interferences

• 2 common types of chemical interferences (that
  reduce the concentration of free gaseous atoms)

  (i) ionization
  (ii) formation of molecular species
Chemical interference – Ionization

• Sample (liquid/solid) must be vaporized & atomized in a
  high temperature source such as flame.
• This high temperature environment lead to ionization of
  the analyte atoms.
• Analyte ionization can be supressed by adding a source
  of electrons, which shifts the equilibrium of the analyte
  from the ionic to the atomic form.
• Cesium and potassium are common ionization
  supressors that are added to analyte solutions.


     Analyte          Analyte+ +     e-
Chemical interference – Ionization

• Cesium and potassium are common ionization
  suppressors that are added to analyte solutions

• These atoms (Cs & K) are easily ionized &
  produce a high concentration of free electrons in
  the flame.
Chemical interference – Refractory
  Formation

• Some elements can form refractory compounds that are
  not atomized in flames.

• E.g. in the presence of phosphates, which interferes
  with calcium measurements due to formation of
  refractory calcium phosphate.

• Formation of refractory compounds can be prevented or
  reduced by adding a releasing agent.
Chemical interference – Refractory
 Formation

• For calcium measurement:
  – Adding lanthanum to the sample (and
    standard) solutions binds the phosphates as
    LaPO4.

  – LaPO4 has a very high formation constant, Kf
    and effectively ties up the phosphate
    interferent.
QUANTITATIVE ANALYSIS BY ATOMIC
  ABSORPTION SPECTROSCOPY

   – Calibration Curve
   – Standard Addition Method

 Absorption behavior follows Beer‟s Law &
  concentration of unknown are determined the same
  way.

 All atomic species have an absorptivity, a

 Pathlength, b is width of the flame

 A = abc
Calibration Curve Method

• A general method for determining the concentration of a
  substance in an unknown sample by comparing the
  unknown to a set of std sample of known concentration.

• Plot is linear over a significant concentration range (the
  dynamic range).

• Analysis should never be based on the measurement of
  a single standard with assumption that Beer‟s law is
  being followed.
Standard Addition Method

• Extensively used in AAS
• Compensate for variation caused by physical & chemical
  interferences in the analyte solution
• The most common way involves adding one or more
  increments of a standard solution to sample aliquots of
  the same size.
• This process is often called spiking the sample.
• Each solution is then diluted to a fixed volume before
  measurement.
Example: Standard Addition Method
The chromium in an aqueous sample was determined by
pipetting 10.0 mL of the unknown into each of five 50.0 mL
volumetric flasks. Various volumes of a standard containing
12.2 ppm Cr were added to the flasks, following which the
solutions were diluted to the volume.
        Volume of        Volume of       Absorbance
      unknown (mL)     standard (mL)

           10.0             0.0             0.201
           10.0             10.0            0.292
           10.0             20.0            0.378
           10.0             30.0            0.467
           10.0             40.0            0.554
1.   Plot the data.
2.   Calculate the concentration of Cr in the sample.
Standard Addition
- if Beer‟s law is obeyed,

   A=    bVstdCstd +   bVxCx
         Vt             Vt
      = kVstdCstd    +    kVxCx

 k is a constant equal to   b
                            Vt
• Standard Addition
  - Plot a graph: A vs Vstd (should yield a straight line)

      A = mVstd + b             slope;        m = kCstd
                                intercept;   b = kVxCx



  Concentration of unknown:

              Cx = bCstd
                   mVx
• Detection Limits (ng/mL)

   Element    AAS Flame    Element    AAS Flame

      Al         30          Cu             2

      As         100         Fe             5

      Ca          1          Hg          500

      Cd          1          Pb             10

      Cr          3          Zn             2




Nanogram/mililiter = 10-3 g/mL = 10-3 ppm
Accuracy


– Relative error associated with a flame
  absorption analysis is the order of 1% to 2%

– With special precautions, this figure can be
  lowered to a few tenths of one percent.
Standard addition method

- used to overcome matrix effect
- involves adding one or more increments
of a standard solution to sample aliquots
of the same size.
- Each solution is diluted to a fixed volume
before measuring its absorbance.
Standard Addition Plot

Absorbance
How to produce standard addition
                curve?




1. Add same quantity of unknown sample to a series of flasks.

2. Add varying amounts of standard (made in solvent) to each
   flasks, e.g. 0, 5, 10, 15 mL).

3. Fill each flask to line, mix and measure.
Standard Addition
                Methods

Single-point standard   Multiple standard
   addition method      addition method
Standard addition
- if Beer‟s Law is obeyed,

A = bVstdCstd   +    bVxCx
      Vt             Vt

  = kVstdCstd   +   kVxCx

k is a constant equal to         b
                            Vt
Standard Addition
- Plot a graph: A vs Vstd

   A = mVstd + b

where the slope m and intercept b are:

m = kCstd   ;    b = kVxCx
• Cx can be obtained from the ratio of these
  two quantities: m and b

     b = kVxCx
     m kCstd

     Cx = bCstd
          mVx
Example:
• 10 ml aliquots of raw-water sample were pipetted into
  50.0           ml            volumetric          flasks.
  Then, 0.00, 5.00, 10.00, 15.00 and 20.00 ml
  respectively of a standard solution containing 10 ppm
  of Fe3+ were added to the flasks, followed by an excess
  of aqueous potassium thiocyanate in order to produce
  the red iron-thiocyanate complex. All the resultant
  solutions were diluted to volume and the absorbance of
  each solution was measured at the same.
The results obtained:
        Vol. of std added   Absorbance
               (ml)            (A)
               0              0.215

              5.00            0.424

             10.00            0.625

             15.00            0.836

             20.00            1.040



Calculate the concentration of Fe3+ (in ppm) in
the raw-water sample
Absorbance vs Vol. of std added
                                   1.2


                                    1


                                   0.8
 Absorbance




                                             b = 0.24
                                   0.6

                                                                           Slope, m = 0.0382
              (Vstd)0 = -6.31 ml   0.4


                                   0.2


                                    0
    -10                 -5               0         5                  10    15            20   25
                                                        Vol. of std




Note: From the graph, extrapolated value represents
the volume of reagent corresponding to zero
instrument response.
• The unknown concentration of the analyte
  in the solution is then calculated:

    Csample = -(Vstd)0Cstd
                Vsample


     Cx = bCstd
          mVx
SELF-EXERCISE
 The chromium in an aqueous sample was determined by pipetting
 10.0 ml of the unknown into each of 50.0 mL volumetric flasks.
 Various volumes of a standard containing 12.2 ppm Cr were added
 to the flasks, following which the solutions were diluted to the mark.

            Volume of          Volume of        Absorbance
          unknown (mL)       standard (mL)
               10.0               0.0              0.201
               10.0              10.0              0.292
               10.0              20.0              0.378
               10.0              30.0              0.467
               10.0              40.0              0.554


 i) Plot a suitable graph to determine the concentration
 of Cr in the aqueous sample.

More Related Content

What's hot

Atomic emission spectroscopy In Pharmacy Education ppt
Atomic emission spectroscopy In Pharmacy Education ppt Atomic emission spectroscopy In Pharmacy Education ppt
Atomic emission spectroscopy In Pharmacy Education ppt Dr Duggirala Mahendra
 
Light sources for atomic absorption spectroscopy (aas)
Light sources for atomic absorption spectroscopy (aas)Light sources for atomic absorption spectroscopy (aas)
Light sources for atomic absorption spectroscopy (aas)Dr. Mallikarjunaswamy C
 
inductively coupled mass spectrometer
inductively coupled mass spectrometerinductively coupled mass spectrometer
inductively coupled mass spectrometerMANU MOHAN
 
Atomic Fluorescence Spectroscopy
Atomic Fluorescence SpectroscopyAtomic Fluorescence Spectroscopy
Atomic Fluorescence SpectroscopyFairuz Ajlaa Azmin
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYMusa Khan
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyAnuradha Verma
 
Atomic spectroscopy (AAS & AES)
Atomic spectroscopy  (AAS & AES) Atomic spectroscopy  (AAS & AES)
Atomic spectroscopy (AAS & AES) Rohan Jagdale
 
Raman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarRaman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarNITISH KUMAR
 
Voltammetry and Polarography
Voltammetry and PolarographyVoltammetry and Polarography
Voltammetry and PolarographyMelakuMetto
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy Zainab&Sons
 
Analytical class atomic absorption spectroscopy, P K MANI
Analytical class   atomic absorption spectroscopy, P K MANIAnalytical class   atomic absorption spectroscopy, P K MANI
Analytical class atomic absorption spectroscopy, P K MANIP.K. Mani
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyWasla Anum
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopymanishasoral
 
Atomic emision spectroscopy
Atomic emision spectroscopy Atomic emision spectroscopy
Atomic emision spectroscopy Nabeel Ahmad
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Sajjad Ullah
 

What's hot (20)

Atomic emission spectroscopy In Pharmacy Education ppt
Atomic emission spectroscopy In Pharmacy Education ppt Atomic emission spectroscopy In Pharmacy Education ppt
Atomic emission spectroscopy In Pharmacy Education ppt
 
Light sources for atomic absorption spectroscopy (aas)
Light sources for atomic absorption spectroscopy (aas)Light sources for atomic absorption spectroscopy (aas)
Light sources for atomic absorption spectroscopy (aas)
 
inductively coupled mass spectrometer
inductively coupled mass spectrometerinductively coupled mass spectrometer
inductively coupled mass spectrometer
 
Atomic Fluorescence Spectroscopy
Atomic Fluorescence SpectroscopyAtomic Fluorescence Spectroscopy
Atomic Fluorescence Spectroscopy
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Atomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptxAtomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptx
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopy
 
Atomic spectroscopy (AAS & AES)
Atomic spectroscopy  (AAS & AES) Atomic spectroscopy  (AAS & AES)
Atomic spectroscopy (AAS & AES)
 
Raman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarRaman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumar
 
Voltammetry and Polarography
Voltammetry and PolarographyVoltammetry and Polarography
Voltammetry and Polarography
 
Uv seminar ppt
Uv seminar pptUv seminar ppt
Uv seminar ppt
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Analytical class atomic absorption spectroscopy, P K MANI
Analytical class   atomic absorption spectroscopy, P K MANIAnalytical class   atomic absorption spectroscopy, P K MANI
Analytical class atomic absorption spectroscopy, P K MANI
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic emision spectroscopy
Atomic emision spectroscopy Atomic emision spectroscopy
Atomic emision spectroscopy
 
UV Spectroscopy
UV Spectroscopy UV Spectroscopy
UV Spectroscopy
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
ICP-AES.pptx
ICP-AES.pptxICP-AES.pptx
ICP-AES.pptx
 

Viewers also liked

Atomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsAtomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsVasiliy V. Rosen (Ph.D.)
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopysharmasuriti
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometryBasil "Lexi" Bruno
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry Sailee Gurav
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYnadeem akhter
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYParimi Anuradha
 
5 memberd heterocyclic compound pyrrol
5 memberd heterocyclic compound pyrrol5 memberd heterocyclic compound pyrrol
5 memberd heterocyclic compound pyrrolPOURNIMA BHALEKAR
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYSUBHASISH DAS
 
Basic instrumentation of ir and vibration modes
Basic instrumentation of ir and vibration modesBasic instrumentation of ir and vibration modes
Basic instrumentation of ir and vibration modesamnatahir1991
 
Ir principle and factors affecting-lakshmi priya
Ir  principle and factors affecting-lakshmi priyaIr  principle and factors affecting-lakshmi priya
Ir principle and factors affecting-lakshmi priyasuhasini
 
Attenuated total reflectance spectroscopy
Attenuated total reflectance spectroscopyAttenuated total reflectance spectroscopy
Attenuated total reflectance spectroscopychaitanya kolli
 

Viewers also liked (20)

Atomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsAtomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and Instruments
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
CHM260 - UV-VIS
CHM260 - UV-VISCHM260 - UV-VIS
CHM260 - UV-VIS
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Isal air proses
Isal air prosesIsal air proses
Isal air proses
 
Flame A E S
Flame   A E SFlame   A E S
Flame A E S
 
Q canalytic aas2
Q canalytic aas2Q canalytic aas2
Q canalytic aas2
 
Atomic absorption
Atomic absorption Atomic absorption
Atomic absorption
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
 
APA seminar
APA seminarAPA seminar
APA seminar
 
Diffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary deviceDiffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary device
 
Standard methods
Standard methodsStandard methods
Standard methods
 
5 memberd heterocyclic compound pyrrol
5 memberd heterocyclic compound pyrrol5 memberd heterocyclic compound pyrrol
5 memberd heterocyclic compound pyrrol
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Basic instrumentation of ir and vibration modes
Basic instrumentation of ir and vibration modesBasic instrumentation of ir and vibration modes
Basic instrumentation of ir and vibration modes
 
Ir detectors
Ir detectorsIr detectors
Ir detectors
 
Ir principle and factors affecting-lakshmi priya
Ir  principle and factors affecting-lakshmi priyaIr  principle and factors affecting-lakshmi priya
Ir principle and factors affecting-lakshmi priya
 
Attenuated total reflectance spectroscopy
Attenuated total reflectance spectroscopyAttenuated total reflectance spectroscopy
Attenuated total reflectance spectroscopy
 

Similar to CHM260 - AAS (i)

Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAhnaf maznun
 
Flame photometry
Flame photometryFlame photometry
Flame photometryZainab&Sons
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometryGovardhan Joshi
 
Mohd wahid (aes)
Mohd wahid (aes)Mohd wahid (aes)
Mohd wahid (aes)Mohd Wahid
 
Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry Hari Sharan Makaju
 
atomic absorption spectroscopy _Amany_elshamy.pdf
atomic absorption spectroscopy _Amany_elshamy.pdfatomic absorption spectroscopy _Amany_elshamy.pdf
atomic absorption spectroscopy _Amany_elshamy.pdfAmany Elshamy
 
instrumente3Newin Microsoft PowerPoint Presentation (2).pptx
instrumente3Newin Microsoft PowerPoint Presentation (2).pptxinstrumente3Newin Microsoft PowerPoint Presentation (2).pptx
instrumente3Newin Microsoft PowerPoint Presentation (2).pptxMohammedYousef71
 
Flame Spectroscopy Analysis B.Pharma and M/Pharma
Flame Spectroscopy Analysis B.Pharma and M/PharmaFlame Spectroscopy Analysis B.Pharma and M/Pharma
Flame Spectroscopy Analysis B.Pharma and M/PharmaAmanKapoorMPharma
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lecZainab&Sons
 
flame photometry.pptx
flame photometry.pptxflame photometry.pptx
flame photometry.pptxYubrajBhatta1
 
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Muhammad Asif Shaheeen
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyVignesh .V
 

Similar to CHM260 - AAS (i) (20)

Aas
AasAas
Aas
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
 
Flame Photometry
Flame PhotometryFlame Photometry
Flame Photometry
 
Flame photometry
Flame photometryFlame photometry
Flame photometry
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Mohd wahid (aes)
Mohd wahid (aes)Mohd wahid (aes)
Mohd wahid (aes)
 
AAS.ppt
AAS.pptAAS.ppt
AAS.ppt
 
Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry
 
atomic absorption spectroscopy _Amany_elshamy.pdf
atomic absorption spectroscopy _Amany_elshamy.pdfatomic absorption spectroscopy _Amany_elshamy.pdf
atomic absorption spectroscopy _Amany_elshamy.pdf
 
Chetan
ChetanChetan
Chetan
 
instrumente3Newin Microsoft PowerPoint Presentation (2).pptx
instrumente3Newin Microsoft PowerPoint Presentation (2).pptxinstrumente3Newin Microsoft PowerPoint Presentation (2).pptx
instrumente3Newin Microsoft PowerPoint Presentation (2).pptx
 
Flame photometer
Flame photometerFlame photometer
Flame photometer
 
Flame Spectroscopy Analysis B.Pharma and M/Pharma
Flame Spectroscopy Analysis B.Pharma and M/PharmaFlame Spectroscopy Analysis B.Pharma and M/Pharma
Flame Spectroscopy Analysis B.Pharma and M/Pharma
 
Flame photometry Dr. Abhinav Golla MBBS.MD Pathology Assistant Professor.
Flame photometry Dr. Abhinav Golla MBBS.MD Pathology Assistant Professor.Flame photometry Dr. Abhinav Golla MBBS.MD Pathology Assistant Professor.
Flame photometry Dr. Abhinav Golla MBBS.MD Pathology Assistant Professor.
 
Flame photometry
Flame photometryFlame photometry
Flame photometry
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lec
 
flame photometry.pptx
flame photometry.pptxflame photometry.pptx
flame photometry.pptx
 
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic spectroscopy
Atomic spectroscopy Atomic spectroscopy
Atomic spectroscopy
 

More from Alia Najiha

Guideline report format
Guideline report formatGuideline report format
Guideline report formatAlia Najiha
 
Guideline for etr presentation
Guideline for etr presentationGuideline for etr presentation
Guideline for etr presentationAlia Najiha
 
ENT300 Presentation
ENT300 Presentation ENT300 Presentation
ENT300 Presentation Alia Najiha
 
ENT300 Business Proposal Jeruk madu tip top
ENT300 Business Proposal Jeruk madu tip topENT300 Business Proposal Jeruk madu tip top
ENT300 Business Proposal Jeruk madu tip topAlia Najiha
 
ENT300 Business Proposal
ENT300 Business ProposalENT300 Business Proposal
ENT300 Business ProposalAlia Najiha
 
Chapter 2 – normal flora
Chapter 2 – normal floraChapter 2 – normal flora
Chapter 2 – normal floraAlia Najiha
 
basic principles and protocol in plant tissue culture
basic principles and protocol in plant tissue culturebasic principles and protocol in plant tissue culture
basic principles and protocol in plant tissue cultureAlia Najiha
 
plant disease control
plant disease controlplant disease control
plant disease controlAlia Najiha
 
plant disease development
plant disease developmentplant disease development
plant disease developmentAlia Najiha
 
causes of plant disease
causes of plant diseasecauses of plant disease
causes of plant diseaseAlia Najiha
 
introduction to plant pathology
introduction to plant pathologyintroduction to plant pathology
introduction to plant pathologyAlia Najiha
 
organic matter decomposition
organic matter decompositionorganic matter decomposition
organic matter decompositionAlia Najiha
 
Gene Expresssion
Gene ExpresssionGene Expresssion
Gene ExpresssionAlia Najiha
 
Recombinant DNA Technology
Recombinant DNA TechnologyRecombinant DNA Technology
Recombinant DNA TechnologyAlia Najiha
 

More from Alia Najiha (20)

Guideline report format
Guideline report formatGuideline report format
Guideline report format
 
Guideline for etr presentation
Guideline for etr presentationGuideline for etr presentation
Guideline for etr presentation
 
ENT300 Presentation
ENT300 Presentation ENT300 Presentation
ENT300 Presentation
 
ENT300 Business Proposal Jeruk madu tip top
ENT300 Business Proposal Jeruk madu tip topENT300 Business Proposal Jeruk madu tip top
ENT300 Business Proposal Jeruk madu tip top
 
ENT300 Business Proposal
ENT300 Business ProposalENT300 Business Proposal
ENT300 Business Proposal
 
Chapter 2 – normal flora
Chapter 2 – normal floraChapter 2 – normal flora
Chapter 2 – normal flora
 
Biofertilizer
BiofertilizerBiofertilizer
Biofertilizer
 
basic principles and protocol in plant tissue culture
basic principles and protocol in plant tissue culturebasic principles and protocol in plant tissue culture
basic principles and protocol in plant tissue culture
 
Mycorrhizae
MycorrhizaeMycorrhizae
Mycorrhizae
 
nutrients cycle
nutrients cyclenutrients cycle
nutrients cycle
 
plant disease control
plant disease controlplant disease control
plant disease control
 
C4 mic319
C4 mic319C4 mic319
C4 mic319
 
plant disease development
plant disease developmentplant disease development
plant disease development
 
causes of plant disease
causes of plant diseasecauses of plant disease
causes of plant disease
 
introduction to plant pathology
introduction to plant pathologyintroduction to plant pathology
introduction to plant pathology
 
organic matter decomposition
organic matter decompositionorganic matter decomposition
organic matter decomposition
 
PCR
PCRPCR
PCR
 
DNA Cloning
DNA CloningDNA Cloning
DNA Cloning
 
Gene Expresssion
Gene ExpresssionGene Expresssion
Gene Expresssion
 
Recombinant DNA Technology
Recombinant DNA TechnologyRecombinant DNA Technology
Recombinant DNA Technology
 

Recently uploaded

Exploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone ProcessorsExploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone Processorsdebabhi2
 
Histor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slideHistor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slidevu2urc
 
Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024The Digital Insurer
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptxHampshireHUG
 
presentation ICT roal in 21st century education
presentation ICT roal in 21st century educationpresentation ICT roal in 21st century education
presentation ICT roal in 21st century educationjfdjdjcjdnsjd
 
Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Enterprise Knowledge
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024Rafal Los
 
Boost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdfBoost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdfsudhanshuwaghmare1
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Miguel Araújo
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking MenDelhi Call girls
 
How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonetsnaman860154
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdfhans926745
 
08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking Men08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking MenDelhi Call girls
 
A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)Gabriella Davis
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityPrincipled Technologies
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherRemote DBA Services
 
GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdflior mazor
 
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonAnna Loughnan Colquhoun
 
Scaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationScaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationRadu Cotescu
 

Recently uploaded (20)

Exploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone ProcessorsExploring the Future Potential of AI-Enabled Smartphone Processors
Exploring the Future Potential of AI-Enabled Smartphone Processors
 
Histor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slideHistor y of HAM Radio presentation slide
Histor y of HAM Radio presentation slide
 
Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024Partners Life - Insurer Innovation Award 2024
Partners Life - Insurer Innovation Award 2024
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
 
presentation ICT roal in 21st century education
presentation ICT roal in 21st century educationpresentation ICT roal in 21st century education
presentation ICT roal in 21st century education
 
Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...Driving Behavioral Change for Information Management through Data-Driven Gree...
Driving Behavioral Change for Information Management through Data-Driven Gree...
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024
 
Boost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdfBoost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdf
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men
 
How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonets
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf
 
08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking Men08448380779 Call Girls In Greater Kailash - I Women Seeking Men
08448380779 Call Girls In Greater Kailash - I Women Seeking Men
 
A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivity
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdf
 
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
Scaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationScaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organization
 

CHM260 - AAS (i)

  • 1. Atomic Spectroscopy based on Flame Atomization; Theory, Instrumentation and Application Nurul Auni binti Zainal Abidin Faculty of Applied Science UiTM Negeri Sembilan
  • 2. ATOMIC SPECTROSCOPY • Three techniques (methods) included in atomic spectroscopy: 1) atomic absorption spectroscopy 2) atomic emisssion spectroscopy 3) atomic fluorescene spectroscopy
  • 3. Atomic spectroscopy • In order to perform atomic spectroscopy, atoms of the analyte must first be formed, usually in the form of an atomic vapor. • Atomization – the process by which a sample is converted to an atomic vapor. • Atomizer – a device used to converted a sample to an atomic vapor.
  • 4. Types of atomizer 1) flame atomizer 2) plasma atomizer 3) electrothermal atomizer
  • 5. Atomization process • Solution of the analyte is evaporated rapidly at an elevated temperature to yield a finely divided solid. • Further heating will break down into gaseous atoms.
  • 6. Atomic Absorption Spectroscopy • Uses absorption of light to measure the concentration of gas-phase atoms. • Since samples are usually liquids or solids, the analyte atoms must be vapourised in a flame (or graphite furnace). 6
  • 7. Atomic Absorption Spectroscopy • The analyte concentration is determined from the amount of absorption. 7
  • 8. Atomic Absorption Spectroscopy • The analyte concentration is determined from the amount of absorption. 8
  • 9. Atomization process • Solution of the analyte is evaporated rapidly at an elevated temperature to yield a finely divided solid. • Further heating will break down into gaseous atoms.
  • 11. Flame Atomization • Nebulization – Conversion of the liquid sample to a fine spray. • Desolvation - Solid atoms are mixed with the gaseous fuel. • Volatilization – Solid atoms are converted to a vapor in the flame. • There are three types of particles that exist in the flame: 1) Atoms 2) Ions 3) Molecules
  • 12. 1. Types of Flames Fuel / Oxidant Temperature H–C C-H acetylene / air 2100 – 2400 C (most common) acetylene / N2O 2600 – 2800 C acetylene / O2 3050 – 3150 C • Selection of flame type depends on the volatilization of interest. 2. Flame Structure • Interzonal region is the hottest part of the flame and best for atomic absorption. • Oxidation of the atoms occurs in the secondary combustion zone where the atoms will form molecular oxides and are dispersed into the surroundings.
  • 13. FUELS/OXIDANTS • Low T flames : easily reduced elements (Cu, Pb, Zn, Cd) • High T flames: difficult to reduce elements (e.g. alkaline earths). • Fuels: natural gas, propane, butane, H2, and acetylene; • Oxidants- Air and O2 (low temperature flames). N2O (high temperature flames). • Flame characteristics: • Sample enters flame, is vaporized, reduced and eventually oxidized. • Exact region of flame in which each of these occurs depends upon: • flow rate, • drop size, and • oxidizability of sample. • Optimum position for flame for many metals.
  • 14. Temperature Profile Temperature profile in degrees Celsius for a natural gas-air flame
  • 15. 3. Burners • Two types of burners in flame spectroscopy i. Turbulent flow (total consumption burner) ii. Laminar flow (premix burner)
  • 16. Turbulent Flow Burner • Nebulizer & burner are combined into a single unit. • Sample is drawn up the capillary & nebulized • Sample flow rate: 1 to 3 mL/min
  • 17. Turbulent Flow Burner • Advantage: – introduce relatively large & representative sample into the flame. • Disadvantages: – A relatively short path length through flame – Problems with clogging of the tip – Burners noisy from electronic and auditory stand point
  • 18. Laminar Flow Burner • Sample is nebulized by the flow of oxidant past a capillary tip. • Resulting aerosol then mixed with fuel & flows past a series of baffles that remove all but the finest droplets. • Majority of the sample collects in the bottom of mixing chamber, drained to a waste container. • Aerosol, oxidant & fuel are burned in a slotted burner that provided a flame, 5 – 10 cm in length.
  • 19. Laminar Flow Burner • Advantages: i. Provide quiet flame ii. Longer path length, enhance sensitivity & reproducibility • Disadvantages: i. Lower rate of sample introduction ii. Possibility of selective evaporation of mixed solvents in the mixing chamber, create analytical uncertainties. iii. Mixing chamber contains a potentially explosive mixture that can flash back if the flow rates are too low.
  • 20. Atomic Absorption Spectroscopy (based on flame) • Currently the most widely used of all the atomic methods. • Simplicity, effectiveness, relatively low cost. • Flame atomization best for reproducibility (precision)(<1%)
  • 21. Atomic Absorption Spectroscopy • AAS is commonly used for metal analysis • A solution of a metal compound is sprayed into a flame and vaporises • The metal atoms absorb light of a specific frequency, and the amount of light absorbed is a direct measure of the number of atoms of the metal in the solution 21
  • 22. Atomic Absorption Spectroscopy (based on flame) • The spectra result form the atomized sample absorbing photons of radiation of the appropriate energy (wavelength). • Energy of radiation absorbed by a vaporized atom is identical with that needed to bring about excitation to a higher electronic state. • Transitions: excitation of an electron from ground state to a higher energy level E 1 E
  • 24. Spectrophotometers • Contain same basic components as an instrument designed for absorption analysis of solution. • Source sample container (flame or hot surface) selector detector readout system
  • 25. AA Spectrophotometer Design 1. Single Beam Instrument •The modulated power source can be replaced by a chopper
  • 26. AA Spectrophotometer Design 2. Double-Beam Instrument
  • 27. Double-Beam Instrument • Radiation from hollow cathode lamp is split into 2 beams – One passes through the flame – The other around the flame • A half-silvered mirror returns both beams to a single path then pass through the monochromator then detector. • Note: monochromator placed between sample and detector – Eliminates most of the radiation originating from the flame.
  • 28. A. Line Sources LINE SOURCE – is required because width of the absorption line is very narrow Examples of line source used in AAS: Hollow cathode Electrodeless discharge lamp (HCL) lamp (EDL) done
  • 29. 1. Hollow Cathode Lamp (HCL) • Light source in AA instrument • Light from this lamp exactly light required for the analysis, even no monochromator is used. • Atoms of metal tested are present within the lamp.
  • 30. How does the Hollow Cathode Lamp works? • When lamp on, atoms are supplied with energy • Causes atoms elevate to excited states • Upon returning to ground state, exactly the same wavelength that are useful in the analysis are emitted • Since the analyzed metal with exactly the same energy levels that undergoes excitation • Hollow cathode lamp MUST contain the element to be determined
  • 31. Mechanism of excitation process in Hollow Cathode Lamp • Lamp is a sealed glass envelope filled with argon (or Neon) gas • When lamp is ON, Ne atoms are ionized with electrons drawn to anode (+ charge electrode) • The Ne ions, Ne+ bombard the surface of the cathode (- charge electrode) • Metal atoms, M, in the cathode are elevated to the excited state and are rejected from the surface as a result of this bombardment
  • 32. Mechanism of excitation process in Hollow Cathode Lamp • When atoms return to ground state, characteristic line spectrum of that atom is emitted • This light is directed at the flame, where unexcited atoms of the same element absorb the radiation and are themselves raised to the excited state • Absorbance is measured, and related to the concentration
  • 33. Illustration of mechanism of excitation process in HCL
  • 34. 1. Electrodeless Discharge Lamp (EDL) • Constructed of a metal or salt of interest sealed in a quartz tube filled with a noble gas (Ne or Ar) at low pressure (1 – 5 torr). • The noble gas is ionized and accelerated by a strong radio- frequency (RF) or microwave field and in turn excite the metal or salt of interest. • Provide radiant intensities usually one to two orders of magnitude greater than HCL.
  • 35. Source Modulation • Function: Employed in AAS to distinguish between the component of radiation arising from the source and the component of arising from the flame background • Device: light chopper
  • 36. Source Modulation Light chopper • A chopper or a modulated power supply is used to modulate the source radiation that passes through the atomizer (flame). • The chopper rotates resulting in an alternating atomic absorption and atomic emission signal. • The signal from emission of radiation will be continuous and can be substracted from the modulated AA signal.
  • 37. Light chopper • Eliminate the effects of radiation from the flame • Light is “chopped” with a rotating half-mirror so that detector sees alternating light intensities • One moment, only light emitted by flame is read since the light from the source is cut off • Next moment, light from both the flame emission & transmission of the source‟s light is measured since the source‟s light is allowed to pass • The elements of the detector is such that the emission signal is substracted from the total signal & this difference is what we measured.
  • 38. Questions 1. In flame atomic absorption spectroscopy, briefly describe the „atomization‟ process which the analyte undergoes. 2. Why is source modulation employed in atomic absorption spectroscopy? Name the device used for this purpose. 3. Name the common line source used in AAS.
  • 39. INTERFERENCES Spectral interference Chemical interference Ionization interference
  • 40. Spectral interference • Arise when particulate matter from atomization scatters the incident radiation from the source. • Or when absorption or emission of an interfering species either overlaps or lies so close to the analyte absorption that resolution by the monochromator becomes impossible.
  • 41. Spectral interference (simple words) • Is one in which the spectral line of the elements being determined overlaps with a spectral line (or band) from another element present in the sample. • The effect of the element will also be measured & thus the results will be incorrect.
  • 42. Spectral interference The most common method of solving this problem i) Tune the monochromator to a different spectral line for the element of interest so that there is no overlap. ii) secondary lines (can be found in the literature)
  • 43. Chemical interferences • 2 common types of chemical interferences (that reduce the concentration of free gaseous atoms) (i) ionization (ii) formation of molecular species
  • 44. Chemical interference – Ionization • Sample (liquid/solid) must be vaporized & atomized in a high temperature source such as flame. • This high temperature environment lead to ionization of the analyte atoms. • Analyte ionization can be supressed by adding a source of electrons, which shifts the equilibrium of the analyte from the ionic to the atomic form. • Cesium and potassium are common ionization supressors that are added to analyte solutions. Analyte  Analyte+ + e-
  • 45. Chemical interference – Ionization • Cesium and potassium are common ionization suppressors that are added to analyte solutions • These atoms (Cs & K) are easily ionized & produce a high concentration of free electrons in the flame.
  • 46. Chemical interference – Refractory Formation • Some elements can form refractory compounds that are not atomized in flames. • E.g. in the presence of phosphates, which interferes with calcium measurements due to formation of refractory calcium phosphate. • Formation of refractory compounds can be prevented or reduced by adding a releasing agent.
  • 47. Chemical interference – Refractory Formation • For calcium measurement: – Adding lanthanum to the sample (and standard) solutions binds the phosphates as LaPO4. – LaPO4 has a very high formation constant, Kf and effectively ties up the phosphate interferent.
  • 48. QUANTITATIVE ANALYSIS BY ATOMIC ABSORPTION SPECTROSCOPY – Calibration Curve – Standard Addition Method  Absorption behavior follows Beer‟s Law & concentration of unknown are determined the same way.  All atomic species have an absorptivity, a  Pathlength, b is width of the flame  A = abc
  • 49. Calibration Curve Method • A general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of std sample of known concentration. • Plot is linear over a significant concentration range (the dynamic range). • Analysis should never be based on the measurement of a single standard with assumption that Beer‟s law is being followed.
  • 50. Standard Addition Method • Extensively used in AAS • Compensate for variation caused by physical & chemical interferences in the analyte solution • The most common way involves adding one or more increments of a standard solution to sample aliquots of the same size. • This process is often called spiking the sample. • Each solution is then diluted to a fixed volume before measurement.
  • 51. Example: Standard Addition Method The chromium in an aqueous sample was determined by pipetting 10.0 mL of the unknown into each of five 50.0 mL volumetric flasks. Various volumes of a standard containing 12.2 ppm Cr were added to the flasks, following which the solutions were diluted to the volume. Volume of Volume of Absorbance unknown (mL) standard (mL) 10.0 0.0 0.201 10.0 10.0 0.292 10.0 20.0 0.378 10.0 30.0 0.467 10.0 40.0 0.554 1. Plot the data. 2. Calculate the concentration of Cr in the sample.
  • 52. Standard Addition - if Beer‟s law is obeyed, A= bVstdCstd + bVxCx Vt Vt = kVstdCstd + kVxCx k is a constant equal to b Vt
  • 53. • Standard Addition - Plot a graph: A vs Vstd (should yield a straight line) A = mVstd + b slope; m = kCstd intercept; b = kVxCx Concentration of unknown: Cx = bCstd mVx
  • 54. • Detection Limits (ng/mL) Element AAS Flame Element AAS Flame Al 30 Cu 2 As 100 Fe 5 Ca 1 Hg 500 Cd 1 Pb 10 Cr 3 Zn 2 Nanogram/mililiter = 10-3 g/mL = 10-3 ppm
  • 55. Accuracy – Relative error associated with a flame absorption analysis is the order of 1% to 2% – With special precautions, this figure can be lowered to a few tenths of one percent.
  • 56. Standard addition method - used to overcome matrix effect - involves adding one or more increments of a standard solution to sample aliquots of the same size. - Each solution is diluted to a fixed volume before measuring its absorbance.
  • 58. How to produce standard addition curve? 1. Add same quantity of unknown sample to a series of flasks. 2. Add varying amounts of standard (made in solvent) to each flasks, e.g. 0, 5, 10, 15 mL). 3. Fill each flask to line, mix and measure.
  • 59. Standard Addition Methods Single-point standard Multiple standard addition method addition method
  • 60. Standard addition - if Beer‟s Law is obeyed, A = bVstdCstd + bVxCx Vt Vt = kVstdCstd + kVxCx k is a constant equal to b Vt
  • 61. Standard Addition - Plot a graph: A vs Vstd A = mVstd + b where the slope m and intercept b are: m = kCstd ; b = kVxCx
  • 62. • Cx can be obtained from the ratio of these two quantities: m and b b = kVxCx m kCstd Cx = bCstd mVx
  • 63. Example: • 10 ml aliquots of raw-water sample were pipetted into 50.0 ml volumetric flasks. Then, 0.00, 5.00, 10.00, 15.00 and 20.00 ml respectively of a standard solution containing 10 ppm of Fe3+ were added to the flasks, followed by an excess of aqueous potassium thiocyanate in order to produce the red iron-thiocyanate complex. All the resultant solutions were diluted to volume and the absorbance of each solution was measured at the same.
  • 64. The results obtained: Vol. of std added Absorbance (ml) (A) 0 0.215 5.00 0.424 10.00 0.625 15.00 0.836 20.00 1.040 Calculate the concentration of Fe3+ (in ppm) in the raw-water sample
  • 65. Absorbance vs Vol. of std added 1.2 1 0.8 Absorbance b = 0.24 0.6 Slope, m = 0.0382 (Vstd)0 = -6.31 ml 0.4 0.2 0 -10 -5 0 5 10 15 20 25 Vol. of std Note: From the graph, extrapolated value represents the volume of reagent corresponding to zero instrument response.
  • 66. • The unknown concentration of the analyte in the solution is then calculated: Csample = -(Vstd)0Cstd Vsample Cx = bCstd mVx
  • 67. SELF-EXERCISE The chromium in an aqueous sample was determined by pipetting 10.0 ml of the unknown into each of 50.0 mL volumetric flasks. Various volumes of a standard containing 12.2 ppm Cr were added to the flasks, following which the solutions were diluted to the mark. Volume of Volume of Absorbance unknown (mL) standard (mL) 10.0 0.0 0.201 10.0 10.0 0.292 10.0 20.0 0.378 10.0 30.0 0.467 10.0 40.0 0.554 i) Plot a suitable graph to determine the concentration of Cr in the aqueous sample.