Intra Aortic Balloon Pump
(IABP)
Yan Wing Wa
Department of Intensive Care
Pamela Youde Nethersole Eastern Hospital
Spring 2009
2
What is an IABP?
• The Intra-Aortic Balloon Counterpulsation
system is a volume displacement device.
• It should be positioned so that the tip is
approximately 1 to 2 cm below the origin
of the left subclavian artery and above the
renal arteries.
• On chest x-ray the tip should be visible in
the 2nd or 3rd intercostal space.
3
Purpose of IABP
• Hemodynamic support in medical
conditions
• Prophylactic hemodynamic support
• Cardiac support for hemodynamically
challenged patients with mechanical
defects PRIOR to correction
4
Surgical Indications
• Post surgical myocardial dysfunction
• Support for weaning from cardiopulmonary
bypass (CPB)
• Cardiac support following correction of
anatomical defects
• Maintenance of graft patency post CABG
• Pulsatile flow during CPB
5
• Introducer needle
• Guide wire
• Vessel dilators
• Sheath
• IABP (34 or 40cc)
• Gas tubing
• 60-mL syringe
• Three-way stopcock
• Arterial pressure tubing (not in kit)
IABP Kit Contents
6
• The IAB Should be selected according to the
following chart (chart located on every box).
Note: The 50cc balloons are no longer made by Datascope
IAB Sizing Chart
7
• Percutaneous
–Sheath less
• Surgical insertion
–Femoral cut down
–Trans-thoracic
Insertion Techniques
8
• The end of the balloon should be just
distal to the takeoff of the left subclavian
artery
• Position should be confirmed by
fluoroscopy or chest x-ray
Positioning
9
Insertion Position
• On chest x-ray the
tip should be
visible in the 2nd
or 3rd intercostal
space
10
Pressure Waveforms
DEFLATED INFLATED DEFLATED INFLATED
Red wave = normal arterial pressure
trace
Blue wave = arterial pressure trace on
IABP
11
Desired Outcome
• Appropriately timed blood volume
displacement (30 – 50 mL) in the aorta by
the rapid shuttling of helium gas in and out
of the balloon chamber, resulting in
changes in inflation and deflation
hemodynamics
12
Inflation
• The goal of inflation is to increase or
augment perfusion
13
Inflation of IABP Results In:
• Increased coronary perfusion pressure
• Increased systemic perfusion pressure
• Increased O2 supply to both the
coronary and peripheral tissue
• Increased baroreceptor response
• Decreased sympathetic stimulation
causing decreased HR, decreased SVR,
and increased LV function
14
Deflation
• The goal of deflation is to reduce LV
workload
• Deflation creates a "potential space"
in the aorta, reducing aortic volume
and pressure
15
Deflation of the IABP Results In:
• Afterload reduction and therefore a
reduction in MVO2
• Reduction in peak systolic pressure,
therefore a reduction in LV work
• Increased cardiac output
• Improved ejection fraction and forward
flow
16
How do you Know?
• Afterload reduction?
–Change in diastolic pressures
–Unassisted and assisted
• Decreased workload?
–Change in systolic pressures
–Unassisted and assisted
• Improved cardiac output?
–Increase in pulse pressures
17
Review of Arterial Pressure
Landmarks
AVO = Aortic valve
opens, beginning of
systole
PSP = Peak systolic
pressure, 65-75% of
stroke volume has
been delivered
DN = Dicrotic notch,
signifies aortic valve
closure and the
beginning of diastole
AEDP = Aortic end
diastolic pressure
18
IABP Waveform
1. Zero baseline (on console)
2. Balloon pressure baseline
3. Rapid inflation
4. Peak inflation artifact
5. Balloon pressure plateau
(IAB fully inflated)
6. Rapid deflation
7. Balloon deflation artifact
8. Return to baseline (IAB fully
deflated)
9. Duration of balloon cycle
19
Inflate just prior to the dicrotic
notch
• To accomplish the
goals of inflation, the
balloon must be
inflated at the onset
of diastole
• The result of properly
timed inflation is a
pressure rise during
diastole
20
Review of Arterial Pressure
Landmarks in 1:2 Assist
• Patient AEDP
• PSP (unassisted
systole)
• PDP/DA (diastolic
augmentation)
• Balloon AEDP
• APSP (assisted
systole)
21
Inflation Hemodynamics
• Coronary artery blood flow and pressure
are increased
–Increased renal and cerebral blood flow
• Increased diastolic pressure increases
perfusion to distal organs and tissues
22
Inflation Hemodynamics
• Coronary collateral circulation is potentially
increased from the increase in coronary
perfusion pressure (CPP)
• Systemic perfusion pressure is increased
23
Coronary Perfusion Pressure
• The pressure at which blood perfuses
through the coronary circulation, mainly in
diastole
• Blood flow to the left ventricle occurs
principally during diastole and is largely
determined by the driving pressure
(diastolic pressure minus left ventricular
diastolic pressure), the diastolic interval
and the state of coronary vasodilatation
24
Deflation Hemodynamics
• The pressure that the LV must generate
(afterload) is less throughout the systolic
phase
–Decreases myocardial O2 demands
• Isovolumetric contraction (IVC) phase of
systole is shortened
–Decreases myocardial O2 demands
25
Deflation Hemodynamics
• Reduced afterload allows the LV to empty
more effectively so stroke volume (SV) is
increased
• Enhanced forward cardiac output
decreases the amount of blood shunted
from left to right in cases of intraventricular
septal defects and incompetent mitral
valve
26
Triggering
• It is necessary to establish a reliable
trigger signal before balloon pumping can
begin
• The computer in the IAB console needs a
stimulus to cycle the pneumatic system,
which inflates and deflates the balloon
• The trigger signal tells the computer that
another cardiac cycle has begun
27
Triggering
• In most cases it is preferable to use the R
wave of the ECG as the trigger signal
• There are other trigger options for
instances when the R wave cannot be
used or is not appropriate
28
Trigger Loss
• The console MUST see a trigger to initiate
an inflate/deflate cycle
• If no trigger is seen when the clinician
attempts to start pumping, no pumping will
occur and an alarm will be sounded
• If the trigger is lost after pumping starts, no
further pumping will occur until a trigger is
re-established
29
Trigger Loss
• If the current trigger is lost the clinician can
choose an alternate, available trigger to
resume pumping
• For example, if the ECG lead becomes
disconnected the arterial pressure (AP)
trigger may be selected until the ECG is
re-established
30
ECG Trigger
• Since triggering on the R wave of the ECG
is preferred, it is very important to give the
IABP a good quality ECG signal and lead
31
Poor ECG Choices
Note: changing QRS
morphology may cause
wandering timing
Note: tall T waves may
cause double triggering
or may alter previously
set timing points
Note: wandering baseline
may cause skipped
trigger
Note: artifact may cause
inappropriate triggering
32
ECG Gain
• In addition to selecting a lead with a QRS
morphology that provides consistent,
appropriate triggering, it is important to
ensure the QRS complex has adequate
amplitude
33
Triggering on the Arterial
Pressure Waveform
• Arterial pressure provides another signal
to the IABP to determine where the
cardiac cycle begins and ends
• It is used when the ECG has too much
interference from patient movement or
poor lead connection
34
CPR
• AP trigger is the appropriate choice during
CPR
• Once chest compressions are started an
arterial pressure waveform will be
generated
• Triggering on the AP will produce pumping
in synchrony to chest compression and
has been shown to assist with coronary
and carotid perfusion
35
TRIGGER MODES: ECG
• The computer analyzes the SLOPE and
HEIGHT of a positively or negatively deflected
QRS complex.
• Rejects pacer spikes on the basis of sharp rising
edge.
• Automatically initiates ARRHYTHMIA TIMING
when at least 8 out of 16 beats are detected as
irregular.
• In most instances this is the trigger of
choice.
36
TRIGGER MODES: ECG
• In AUTOPILOT mode, if no QRS or AP trigger is
found, it will trigger on the V SPIKE of an AV
paced rhythm.
• In OPERATOR mode, the ECG trigger key can
be used to toggle between triggering on the R
WAVE and the V SPIKE of a V or AV paced
rhythm.
37
TRIGGER MODES: ARTERIAL
PRESSURE (AP)
• The computer uses the SYSTOLIC UPSTROKE of an
arterial pressure waveform as the trigger signal.
• A 14mmHg minimum pulse pressure is required to initially
recognize the trigger.
• Once pumping begins the subsequent pulse pressure
requirement is reduced to 7mmHg.
• Every 64th beat is unassisted and reassessed.
• This mode is available as an option for clinical situations
where an ECG is unavailable or distorted.
• Not recommended for irregular rhythms or irregular
pulse pressures.
38
Gas Alarms/Balloon Pressure
Waveform
• During a cycle of inflation/deflation, helium
is rapidly moved in and out of the balloon.
• The environment within the balloon and
the surrounding forces that affect balloon
behavior all contribute to a predictable
pattern of gas flow and pressure.
• The gas pressure characteristics are
converted into a waveform that is reflective
of the behavior of the gas.
39
• This transduced waveform tells much
about the interaction of the balloon within
the patient's aorta.
• Understanding the balloon pressure
waveform is important for troubleshooting
of the console as most alarms are based
on the gas surveillance system.
40
IABP Waveform
• Zero baseline (on console)
• Balloon pressure baseline
• Rapid inflation
• Peak inflation artifact
• Balloon pressure plateau
(IAB fully inflated)
• Rapid deflation
• Balloon deflation artifact
• Return to baseline (IAB fully
deflated)
• Duration of balloon cycle
41
Normal Waveform Variations
Tachycardia Bradycardia
Hypertension Hypotension
42
Abnormal Waveform Variation: Wide
Inflation and/or Deflation Artifact
• Check for kinks, as they may trap gas in
the IAB.
• If water is present in the gas tubing,
remove the condensation.
–Pushing the helium through the water
during inflation and deflation slows down
gas transition.
–If gas transition is prolonged too much, it
can create alarms.
43
Abnormal Waveform Variation: Helium Loss
/ Gas Loss / Gas Leakage Alarms
• Observe for blood in the gas tubing. If
even a slight amount is present, it may
indicate a balloon rupture.
–Do not resume pumping.
–Notify physician immediately and
prepare for IAB removal.
• Check connections where gas tubing
connects to IAB and to pump.
–Secure if loose.
44
Abnormal Waveform Variation: High
Pressure / Kinked Line Alarm
45
• Reposition patient. Keep affected leg straight.
• Use rolled towel under hip to hyperextend hip.
• Apply slight traction to the catheter if kinking
at the insertion site or in the artery is
suspected.
• The distal portion of the IAB may be in the
sheath if a long introducer sheath was used.
Pull sheath back until IAB bladder has exited
the sheath.
• Introducer sheath may be kinked which in turn
is kinking the balloon. Suspect this particularly
if placement of the sheath was difficult. Pull
sheath back or rotate sheath a partial turn.
46
• Check placement of the balloon; it may be
too high or too low.
• IAB may be partially wrapped if alarm
occurs shortly after insertion. Take steps to
facilitate unwrapping (consult IAB
manufacturer).
• The balloon may be too large for the
patient. Reduce the helium volume the
balloon is inflated with. (It is recommended
to not reduce the volume below 2/3 of
maximum. For example, do not decrease
volume in a 40cc IAB below 27cc)
47
• Check for intermittent obstruction of gas
lumen.
• Check for overfill of system.
• This condition may occur during ascent
in air transport since gas expands as you
go up in altitude. Reset the alarm and
restart pumping. The volume will be
adjusted automatically for current
barometric pressure.
Abnormal Waveform Variation: High
Baseline / Fill Pressure
48
Complications of IABP
• The following patients are at the greatest
risk of developing complications
associated with IABP:
• Females, diabetics, smokers, obese
patients
• Patients with PVD, HTN, high SVR, shock
49
Complications of IABP
• Aortic wall dissection, rupture or local
vascular injury
–Care as indicated
• Emboli: thrombus, plaque or air
–Care as indicated
–Treatment of an air embolism is as
follows: 100% oxygen, intubate,
Trendelenberg in left lateral decubitus
position, fluid resuscitation, etc.
50
Complications of IABP
• IABP Rupture:
• COFFEE GROUNDS seen in the drive line
is a precursor to a rupture (make sure they
are on the inside and not the outside of the
tubing)
–NOTIFY PHYSICIAN!!!!!
–IF THERE IS A FLAGRANT RUPURE
OF THE IABP CLAMP THE GAS
LINE!!!!!
51
Complications of IABP
• Infection
–Check catheter insertion site often
–STRICT ASEPTIC TECHNIQUE
–Restrict movement while IABP in place
52
Complications of IABP
• Obstruction
• Malposition
–Too high – obstruction of left subclavian
and carotids
• CHECK LEFT RADIAL ARTERY PULSE
–Too low – obstruction of renal and
mesenteric arteries
• MONITOR URINE OUTPUT
53
Complications of IABP
• Compromised circulation due to catheter
• Ischemia
–Routine nursing care and monitoring
• Compartment syndrome
–Rare complication seen in the LE,
usually related to infection
–Monitor calf circumference
54
Complications of IABP
• Hematologic
–ALL PATIENTS typed & crossmatched!!!
• Bleeding
–REMOVE THE DRESSING!!!
–PUT ON STERILE GLOVES!!!
–HOLD PRESSURE!!!
• Thrombocytopenia
–Routine monitoring
55
• Timing of weaning
–Patient should be stable for 24-48 hours
• Decreasing inotropic support
• Decreasing pump ratio
–From 1:1 to 1:2 or 1:3
• Decrease augmentation
• Monitor patient closely
–If patient becomes unstable, weaning should
be immediately discontinued
Weaning should only be attempted on order of a physician.
56
• Discontinue heparin six hours prior
• Check platelets and coagulation factors
• Deflate the balloon
• Apply manual pressure above and below IABP
insertion site
• Remove and alternate pressure to expel any
clots
• Apply constant pressure to the insertion site for
a minimum of 30 minutes
• Check distal pulses frequently
IAB should only be pulled by a physician or the cardiac P.A.
57
Thank you for your attention

More Related Content

Similar to Intra Aortic Balloon Pump (IABP) 2009.ppt

Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherizationvasanth7pv
 
Iabp presentation
Iabp presentationIabp presentation
Iabp presentationIndia CTVS
 
Cardiovascular monitoring Part II
Cardiovascular monitoring Part IICardiovascular monitoring Part II
Cardiovascular monitoring Part IISiddhanta Choudhury
 
Anaesthesia for cardiopulmonary bypass surgery
Anaesthesia for cardiopulmonary bypass surgeryAnaesthesia for cardiopulmonary bypass surgery
Anaesthesia for cardiopulmonary bypass surgeryZIKRULLAH MALLICK
 
CARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSCARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSSreevidya V S
 
Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Suheil Dhanse
 
Extra aortic balloon counterpulsation
Extra aortic balloon counterpulsationExtra aortic balloon counterpulsation
Extra aortic balloon counterpulsationVijay Anand
 
Advanced treatment in hf ppt
Advanced treatment in hf pptAdvanced treatment in hf ppt
Advanced treatment in hf pptShalini Garg
 
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman editIabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman editIsman Firdaus
 
Cardiovascular physiology REVISION NOTES
Cardiovascular physiology REVISION NOTES Cardiovascular physiology REVISION NOTES
Cardiovascular physiology REVISION NOTES TONY SCARIA
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicineanjika
 
CARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSCARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSManu Jacob
 
Cardiopulmonary Bypass overview for beginners
Cardiopulmonary Bypass overview for beginnersCardiopulmonary Bypass overview for beginners
Cardiopulmonary Bypass overview for beginnersNICS, Bangalore
 

Similar to Intra Aortic Balloon Pump (IABP) 2009.ppt (20)

IABP - DEV BUCHE
IABP - DEV BUCHEIABP - DEV BUCHE
IABP - DEV BUCHE
 
Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherization
 
Iabp presentation
Iabp presentationIabp presentation
Iabp presentation
 
Cardiovascular monitoring Part II
Cardiovascular monitoring Part IICardiovascular monitoring Part II
Cardiovascular monitoring Part II
 
INTRA AORTIC BALLON PUMP (IABP)
INTRA AORTIC BALLON PUMP (IABP)INTRA AORTIC BALLON PUMP (IABP)
INTRA AORTIC BALLON PUMP (IABP)
 
CVP.pptx
CVP.pptxCVP.pptx
CVP.pptx
 
Anaesthesia for cardiopulmonary bypass surgery
Anaesthesia for cardiopulmonary bypass surgeryAnaesthesia for cardiopulmonary bypass surgery
Anaesthesia for cardiopulmonary bypass surgery
 
IABP chetan
IABP chetanIABP chetan
IABP chetan
 
CARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSCARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASS
 
Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)
 
Extra aortic balloon counterpulsation
Extra aortic balloon counterpulsationExtra aortic balloon counterpulsation
Extra aortic balloon counterpulsation
 
Advanced treatment in hf ppt
Advanced treatment in hf pptAdvanced treatment in hf ppt
Advanced treatment in hf ppt
 
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman editIabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
 
Cardiovascular physiology REVISION NOTES
Cardiovascular physiology REVISION NOTES Cardiovascular physiology REVISION NOTES
Cardiovascular physiology REVISION NOTES
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicine
 
CARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSCARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASS
 
Cardiopulmonary Bypass overview for beginners
Cardiopulmonary Bypass overview for beginnersCardiopulmonary Bypass overview for beginners
Cardiopulmonary Bypass overview for beginners
 
Open heart surgery uday
Open heart surgery udayOpen heart surgery uday
Open heart surgery uday
 
IABP
IABPIABP
IABP
 
IABP
IABPIABP
IABP
 

Recently uploaded

Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...jaredbarbolino94
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsanshu789521
 
How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17Celine George
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxAvyJaneVismanos
 
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxHistory Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxsocialsciencegdgrohi
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceSamikshaHamane
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
MARGINALIZATION (Different learners in Marginalized Group
MARGINALIZATION (Different learners in Marginalized GroupMARGINALIZATION (Different learners in Marginalized Group
MARGINALIZATION (Different learners in Marginalized GroupJonathanParaisoCruz
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatYousafMalik24
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...M56BOOKSTORE PRODUCT/SERVICE
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTiammrhaywood
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentInMediaRes1
 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfMahmoud M. Sallam
 
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Celine George
 

Recently uploaded (20)

Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha elections
 
How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptx
 
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxHistory Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in Pharmacovigilance
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
MARGINALIZATION (Different learners in Marginalized Group
MARGINALIZATION (Different learners in Marginalized GroupMARGINALIZATION (Different learners in Marginalized Group
MARGINALIZATION (Different learners in Marginalized Group
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice great
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...
KSHARA STURA .pptx---KSHARA KARMA THERAPY (CAUSTIC THERAPY)————IMP.OF KSHARA ...
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media Component
 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdf
 
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
 
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
 

Intra Aortic Balloon Pump (IABP) 2009.ppt

  • 1. 1 Intra Aortic Balloon Pump (IABP) Yan Wing Wa Department of Intensive Care Pamela Youde Nethersole Eastern Hospital Spring 2009
  • 2. 2 What is an IABP? • The Intra-Aortic Balloon Counterpulsation system is a volume displacement device. • It should be positioned so that the tip is approximately 1 to 2 cm below the origin of the left subclavian artery and above the renal arteries. • On chest x-ray the tip should be visible in the 2nd or 3rd intercostal space.
  • 3. 3 Purpose of IABP • Hemodynamic support in medical conditions • Prophylactic hemodynamic support • Cardiac support for hemodynamically challenged patients with mechanical defects PRIOR to correction
  • 4. 4 Surgical Indications • Post surgical myocardial dysfunction • Support for weaning from cardiopulmonary bypass (CPB) • Cardiac support following correction of anatomical defects • Maintenance of graft patency post CABG • Pulsatile flow during CPB
  • 5. 5 • Introducer needle • Guide wire • Vessel dilators • Sheath • IABP (34 or 40cc) • Gas tubing • 60-mL syringe • Three-way stopcock • Arterial pressure tubing (not in kit) IABP Kit Contents
  • 6. 6 • The IAB Should be selected according to the following chart (chart located on every box). Note: The 50cc balloons are no longer made by Datascope IAB Sizing Chart
  • 7. 7 • Percutaneous –Sheath less • Surgical insertion –Femoral cut down –Trans-thoracic Insertion Techniques
  • 8. 8 • The end of the balloon should be just distal to the takeoff of the left subclavian artery • Position should be confirmed by fluoroscopy or chest x-ray Positioning
  • 9. 9 Insertion Position • On chest x-ray the tip should be visible in the 2nd or 3rd intercostal space
  • 10. 10 Pressure Waveforms DEFLATED INFLATED DEFLATED INFLATED Red wave = normal arterial pressure trace Blue wave = arterial pressure trace on IABP
  • 11. 11 Desired Outcome • Appropriately timed blood volume displacement (30 – 50 mL) in the aorta by the rapid shuttling of helium gas in and out of the balloon chamber, resulting in changes in inflation and deflation hemodynamics
  • 12. 12 Inflation • The goal of inflation is to increase or augment perfusion
  • 13. 13 Inflation of IABP Results In: • Increased coronary perfusion pressure • Increased systemic perfusion pressure • Increased O2 supply to both the coronary and peripheral tissue • Increased baroreceptor response • Decreased sympathetic stimulation causing decreased HR, decreased SVR, and increased LV function
  • 14. 14 Deflation • The goal of deflation is to reduce LV workload • Deflation creates a "potential space" in the aorta, reducing aortic volume and pressure
  • 15. 15 Deflation of the IABP Results In: • Afterload reduction and therefore a reduction in MVO2 • Reduction in peak systolic pressure, therefore a reduction in LV work • Increased cardiac output • Improved ejection fraction and forward flow
  • 16. 16 How do you Know? • Afterload reduction? –Change in diastolic pressures –Unassisted and assisted • Decreased workload? –Change in systolic pressures –Unassisted and assisted • Improved cardiac output? –Increase in pulse pressures
  • 17. 17 Review of Arterial Pressure Landmarks AVO = Aortic valve opens, beginning of systole PSP = Peak systolic pressure, 65-75% of stroke volume has been delivered DN = Dicrotic notch, signifies aortic valve closure and the beginning of diastole AEDP = Aortic end diastolic pressure
  • 18. 18 IABP Waveform 1. Zero baseline (on console) 2. Balloon pressure baseline 3. Rapid inflation 4. Peak inflation artifact 5. Balloon pressure plateau (IAB fully inflated) 6. Rapid deflation 7. Balloon deflation artifact 8. Return to baseline (IAB fully deflated) 9. Duration of balloon cycle
  • 19. 19 Inflate just prior to the dicrotic notch • To accomplish the goals of inflation, the balloon must be inflated at the onset of diastole • The result of properly timed inflation is a pressure rise during diastole
  • 20. 20 Review of Arterial Pressure Landmarks in 1:2 Assist • Patient AEDP • PSP (unassisted systole) • PDP/DA (diastolic augmentation) • Balloon AEDP • APSP (assisted systole)
  • 21. 21 Inflation Hemodynamics • Coronary artery blood flow and pressure are increased –Increased renal and cerebral blood flow • Increased diastolic pressure increases perfusion to distal organs and tissues
  • 22. 22 Inflation Hemodynamics • Coronary collateral circulation is potentially increased from the increase in coronary perfusion pressure (CPP) • Systemic perfusion pressure is increased
  • 23. 23 Coronary Perfusion Pressure • The pressure at which blood perfuses through the coronary circulation, mainly in diastole • Blood flow to the left ventricle occurs principally during diastole and is largely determined by the driving pressure (diastolic pressure minus left ventricular diastolic pressure), the diastolic interval and the state of coronary vasodilatation
  • 24. 24 Deflation Hemodynamics • The pressure that the LV must generate (afterload) is less throughout the systolic phase –Decreases myocardial O2 demands • Isovolumetric contraction (IVC) phase of systole is shortened –Decreases myocardial O2 demands
  • 25. 25 Deflation Hemodynamics • Reduced afterload allows the LV to empty more effectively so stroke volume (SV) is increased • Enhanced forward cardiac output decreases the amount of blood shunted from left to right in cases of intraventricular septal defects and incompetent mitral valve
  • 26. 26 Triggering • It is necessary to establish a reliable trigger signal before balloon pumping can begin • The computer in the IAB console needs a stimulus to cycle the pneumatic system, which inflates and deflates the balloon • The trigger signal tells the computer that another cardiac cycle has begun
  • 27. 27 Triggering • In most cases it is preferable to use the R wave of the ECG as the trigger signal • There are other trigger options for instances when the R wave cannot be used or is not appropriate
  • 28. 28 Trigger Loss • The console MUST see a trigger to initiate an inflate/deflate cycle • If no trigger is seen when the clinician attempts to start pumping, no pumping will occur and an alarm will be sounded • If the trigger is lost after pumping starts, no further pumping will occur until a trigger is re-established
  • 29. 29 Trigger Loss • If the current trigger is lost the clinician can choose an alternate, available trigger to resume pumping • For example, if the ECG lead becomes disconnected the arterial pressure (AP) trigger may be selected until the ECG is re-established
  • 30. 30 ECG Trigger • Since triggering on the R wave of the ECG is preferred, it is very important to give the IABP a good quality ECG signal and lead
  • 31. 31 Poor ECG Choices Note: changing QRS morphology may cause wandering timing Note: tall T waves may cause double triggering or may alter previously set timing points Note: wandering baseline may cause skipped trigger Note: artifact may cause inappropriate triggering
  • 32. 32 ECG Gain • In addition to selecting a lead with a QRS morphology that provides consistent, appropriate triggering, it is important to ensure the QRS complex has adequate amplitude
  • 33. 33 Triggering on the Arterial Pressure Waveform • Arterial pressure provides another signal to the IABP to determine where the cardiac cycle begins and ends • It is used when the ECG has too much interference from patient movement or poor lead connection
  • 34. 34 CPR • AP trigger is the appropriate choice during CPR • Once chest compressions are started an arterial pressure waveform will be generated • Triggering on the AP will produce pumping in synchrony to chest compression and has been shown to assist with coronary and carotid perfusion
  • 35. 35 TRIGGER MODES: ECG • The computer analyzes the SLOPE and HEIGHT of a positively or negatively deflected QRS complex. • Rejects pacer spikes on the basis of sharp rising edge. • Automatically initiates ARRHYTHMIA TIMING when at least 8 out of 16 beats are detected as irregular. • In most instances this is the trigger of choice.
  • 36. 36 TRIGGER MODES: ECG • In AUTOPILOT mode, if no QRS or AP trigger is found, it will trigger on the V SPIKE of an AV paced rhythm. • In OPERATOR mode, the ECG trigger key can be used to toggle between triggering on the R WAVE and the V SPIKE of a V or AV paced rhythm.
  • 37. 37 TRIGGER MODES: ARTERIAL PRESSURE (AP) • The computer uses the SYSTOLIC UPSTROKE of an arterial pressure waveform as the trigger signal. • A 14mmHg minimum pulse pressure is required to initially recognize the trigger. • Once pumping begins the subsequent pulse pressure requirement is reduced to 7mmHg. • Every 64th beat is unassisted and reassessed. • This mode is available as an option for clinical situations where an ECG is unavailable or distorted. • Not recommended for irregular rhythms or irregular pulse pressures.
  • 38. 38 Gas Alarms/Balloon Pressure Waveform • During a cycle of inflation/deflation, helium is rapidly moved in and out of the balloon. • The environment within the balloon and the surrounding forces that affect balloon behavior all contribute to a predictable pattern of gas flow and pressure. • The gas pressure characteristics are converted into a waveform that is reflective of the behavior of the gas.
  • 39. 39 • This transduced waveform tells much about the interaction of the balloon within the patient's aorta. • Understanding the balloon pressure waveform is important for troubleshooting of the console as most alarms are based on the gas surveillance system.
  • 40. 40 IABP Waveform • Zero baseline (on console) • Balloon pressure baseline • Rapid inflation • Peak inflation artifact • Balloon pressure plateau (IAB fully inflated) • Rapid deflation • Balloon deflation artifact • Return to baseline (IAB fully deflated) • Duration of balloon cycle
  • 41. 41 Normal Waveform Variations Tachycardia Bradycardia Hypertension Hypotension
  • 42. 42 Abnormal Waveform Variation: Wide Inflation and/or Deflation Artifact • Check for kinks, as they may trap gas in the IAB. • If water is present in the gas tubing, remove the condensation. –Pushing the helium through the water during inflation and deflation slows down gas transition. –If gas transition is prolonged too much, it can create alarms.
  • 43. 43 Abnormal Waveform Variation: Helium Loss / Gas Loss / Gas Leakage Alarms • Observe for blood in the gas tubing. If even a slight amount is present, it may indicate a balloon rupture. –Do not resume pumping. –Notify physician immediately and prepare for IAB removal. • Check connections where gas tubing connects to IAB and to pump. –Secure if loose.
  • 44. 44 Abnormal Waveform Variation: High Pressure / Kinked Line Alarm
  • 45. 45 • Reposition patient. Keep affected leg straight. • Use rolled towel under hip to hyperextend hip. • Apply slight traction to the catheter if kinking at the insertion site or in the artery is suspected. • The distal portion of the IAB may be in the sheath if a long introducer sheath was used. Pull sheath back until IAB bladder has exited the sheath. • Introducer sheath may be kinked which in turn is kinking the balloon. Suspect this particularly if placement of the sheath was difficult. Pull sheath back or rotate sheath a partial turn.
  • 46. 46 • Check placement of the balloon; it may be too high or too low. • IAB may be partially wrapped if alarm occurs shortly after insertion. Take steps to facilitate unwrapping (consult IAB manufacturer). • The balloon may be too large for the patient. Reduce the helium volume the balloon is inflated with. (It is recommended to not reduce the volume below 2/3 of maximum. For example, do not decrease volume in a 40cc IAB below 27cc)
  • 47. 47 • Check for intermittent obstruction of gas lumen. • Check for overfill of system. • This condition may occur during ascent in air transport since gas expands as you go up in altitude. Reset the alarm and restart pumping. The volume will be adjusted automatically for current barometric pressure. Abnormal Waveform Variation: High Baseline / Fill Pressure
  • 48. 48 Complications of IABP • The following patients are at the greatest risk of developing complications associated with IABP: • Females, diabetics, smokers, obese patients • Patients with PVD, HTN, high SVR, shock
  • 49. 49 Complications of IABP • Aortic wall dissection, rupture or local vascular injury –Care as indicated • Emboli: thrombus, plaque or air –Care as indicated –Treatment of an air embolism is as follows: 100% oxygen, intubate, Trendelenberg in left lateral decubitus position, fluid resuscitation, etc.
  • 50. 50 Complications of IABP • IABP Rupture: • COFFEE GROUNDS seen in the drive line is a precursor to a rupture (make sure they are on the inside and not the outside of the tubing) –NOTIFY PHYSICIAN!!!!! –IF THERE IS A FLAGRANT RUPURE OF THE IABP CLAMP THE GAS LINE!!!!!
  • 51. 51 Complications of IABP • Infection –Check catheter insertion site often –STRICT ASEPTIC TECHNIQUE –Restrict movement while IABP in place
  • 52. 52 Complications of IABP • Obstruction • Malposition –Too high – obstruction of left subclavian and carotids • CHECK LEFT RADIAL ARTERY PULSE –Too low – obstruction of renal and mesenteric arteries • MONITOR URINE OUTPUT
  • 53. 53 Complications of IABP • Compromised circulation due to catheter • Ischemia –Routine nursing care and monitoring • Compartment syndrome –Rare complication seen in the LE, usually related to infection –Monitor calf circumference
  • 54. 54 Complications of IABP • Hematologic –ALL PATIENTS typed & crossmatched!!! • Bleeding –REMOVE THE DRESSING!!! –PUT ON STERILE GLOVES!!! –HOLD PRESSURE!!! • Thrombocytopenia –Routine monitoring
  • 55. 55 • Timing of weaning –Patient should be stable for 24-48 hours • Decreasing inotropic support • Decreasing pump ratio –From 1:1 to 1:2 or 1:3 • Decrease augmentation • Monitor patient closely –If patient becomes unstable, weaning should be immediately discontinued Weaning should only be attempted on order of a physician.
  • 56. 56 • Discontinue heparin six hours prior • Check platelets and coagulation factors • Deflate the balloon • Apply manual pressure above and below IABP insertion site • Remove and alternate pressure to expel any clots • Apply constant pressure to the insertion site for a minimum of 30 minutes • Check distal pulses frequently IAB should only be pulled by a physician or the cardiac P.A.
  • 57. 57 Thank you for your attention