Trace and Percentage (ATEX approved) Suitable for Hazardous Environments. The EC92DIS portable oxygen analyser will detect levels of oxygen as low as 1ppm, up to high percent levels and can be used on most industrial gases and atmospheres.
This document provides precautions for servicing the 2002 Nissan Maxima. It outlines safety precautions for working with the supplemental restraint system (SRS), Nissan Vehicle Immobilizer System (NVIS), and general vehicle repairs. Proper ventilation, protective equipment, and battery disconnection are recommended. Only specified fluids and tools should be used. Removed parts should be cleaned and replaced with new components as needed.
This document provides instructions for installing and configuring an ultrasonic level meter. It contains the following key points:
1. It describes the meter's wiring, installation parameters, and installation principles, noting the transducer should be parallel to the liquid surface and avoiding installation near inlets/outlets.
2. It explains how to configure the meter's parameters through the keyboard, including calibrating the level (P01), setting the 20mA point (P02), and changing the display mode (P03).
3. It provides notes for installation, such as protecting the meter from sunlight, sealing cables to prevent water, and adding lightning protection for mines.
This document provides instructions for using an ultrasonic level meter. It includes:
1) An overview of the meter's features such as high precision, stability, and applicability to various environments.
2) Technical specifications including measuring range, accuracy, display, and power supply.
3) Guidelines for installing the meter such as wiring, parameter meanings, installation principles, and points of attention.
4) Instructions for configuring the meter including calibration, setting output range, display modes, and testing outputs.
X8/X10/X12
Patient Monitor
Patient Monitor User Manual Product
Specification
- 1 - A Product Specification
NOTE: The performance of the equipment with ☆ mark is determined to be essential
performance.
A.1 Classification Anti-electroshock type Class I equipment and internal powered equipment
Anti-electroshock degree CF
Ingress Protection IPX1
Disinfection/sterilization method Refer to Chapter Care and Cleaning for details.
Working system Continuous operation equipment
Compliant with Standards IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1 :2013; EN 60601-1-2: 2015;
IEC 60601-2-49: 2011
A.2 Physical Specifications
A.2.1 Size and Weight
Product Size Weight
X8 236±2 mm (W)×236±2 mm (H)×147±2 mm (D) < 2.4 kg Standard
configurations,
no battery or
accessories
X10 261±2 mm (W)×246±2 mm (H)×146±2 mm (D) < 2.8 kg
X12 306±2 mm (W)×309±2 mm (H)×151±2 mm (D) < 3.5 kg
A.2.2 Function Configuration
Product Standard Configuration Optional Configuration
X8 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
CO2, Wi-Fi, Recorder
X10 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1, T2), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
IBP, CO2, Wi-Fi,
Recorder
X12 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1, T2), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
IBP, CO2,
The operating manual provides instructions for using the UTD2000/3000 series digital storage oscilloscopes. It discusses general safety procedures, an overview of the instrument's front panel controls, and how to perform basic operations like probe compensation, automatic waveform setup, and inspecting the instrument. The manual also outlines the various chapters that will explain how to set up the vertical, horizontal and trigger systems, as well as auxiliary functions, measurements, storage and more. Application examples are provided to demonstrate how to use the oscilloscope for common tasks.
ShuntCheck is an aid to the detection of flow in implanted cerebrospinal fluid (CSF) shunts. ShuntCheck includes Micro-Pumper, a component which may be used to temporarily increase CSF flow in suspected non-flowing, patent shunts during the ShuntCheck test. ShuntCheck cannot alone diagnose CSF shunt function or malfunction. The clinical diagnosis of CSF shunt function or malfunction, incorporating the flow information from ShuntCheck, should be made only by a qualified neurosurgeon.
Trace and Percentage (ATEX approved) Suitable for Hazardous Environments. The EC92DIS portable oxygen analyser will detect levels of oxygen as low as 1ppm, up to high percent levels and can be used on most industrial gases and atmospheres.
This document provides precautions for servicing the 2002 Nissan Maxima. It outlines safety precautions for working with the supplemental restraint system (SRS), Nissan Vehicle Immobilizer System (NVIS), and general vehicle repairs. Proper ventilation, protective equipment, and battery disconnection are recommended. Only specified fluids and tools should be used. Removed parts should be cleaned and replaced with new components as needed.
This document provides instructions for installing and configuring an ultrasonic level meter. It contains the following key points:
1. It describes the meter's wiring, installation parameters, and installation principles, noting the transducer should be parallel to the liquid surface and avoiding installation near inlets/outlets.
2. It explains how to configure the meter's parameters through the keyboard, including calibrating the level (P01), setting the 20mA point (P02), and changing the display mode (P03).
3. It provides notes for installation, such as protecting the meter from sunlight, sealing cables to prevent water, and adding lightning protection for mines.
This document provides instructions for using an ultrasonic level meter. It includes:
1) An overview of the meter's features such as high precision, stability, and applicability to various environments.
2) Technical specifications including measuring range, accuracy, display, and power supply.
3) Guidelines for installing the meter such as wiring, parameter meanings, installation principles, and points of attention.
4) Instructions for configuring the meter including calibration, setting output range, display modes, and testing outputs.
X8/X10/X12
Patient Monitor
Patient Monitor User Manual Product
Specification
- 1 - A Product Specification
NOTE: The performance of the equipment with ☆ mark is determined to be essential
performance.
A.1 Classification Anti-electroshock type Class I equipment and internal powered equipment
Anti-electroshock degree CF
Ingress Protection IPX1
Disinfection/sterilization method Refer to Chapter Care and Cleaning for details.
Working system Continuous operation equipment
Compliant with Standards IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1 :2013; EN 60601-1-2: 2015;
IEC 60601-2-49: 2011
A.2 Physical Specifications
A.2.1 Size and Weight
Product Size Weight
X8 236±2 mm (W)×236±2 mm (H)×147±2 mm (D) < 2.4 kg Standard
configurations,
no battery or
accessories
X10 261±2 mm (W)×246±2 mm (H)×146±2 mm (D) < 2.8 kg
X12 306±2 mm (W)×309±2 mm (H)×151±2 mm (D) < 3.5 kg
A.2.2 Function Configuration
Product Standard Configuration Optional Configuration
X8 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
CO2, Wi-Fi, Recorder
X10 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1, T2), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
IBP, CO2, Wi-Fi,
Recorder
X12 ECG (3 electrodes), ECG (5 electrodes), RESP,
TEMP (T1, T2), SpO2, NIBP
ECG (6 electrodes),
ECG (10 electrodes),
IBP, CO2,
The operating manual provides instructions for using the UTD2000/3000 series digital storage oscilloscopes. It discusses general safety procedures, an overview of the instrument's front panel controls, and how to perform basic operations like probe compensation, automatic waveform setup, and inspecting the instrument. The manual also outlines the various chapters that will explain how to set up the vertical, horizontal and trigger systems, as well as auxiliary functions, measurements, storage and more. Application examples are provided to demonstrate how to use the oscilloscope for common tasks.
ShuntCheck is an aid to the detection of flow in implanted cerebrospinal fluid (CSF) shunts. ShuntCheck includes Micro-Pumper, a component which may be used to temporarily increase CSF flow in suspected non-flowing, patent shunts during the ShuntCheck test. ShuntCheck cannot alone diagnose CSF shunt function or malfunction. The clinical diagnosis of CSF shunt function or malfunction, incorporating the flow information from ShuntCheck, should be made only by a qualified neurosurgeon.
Catalog design and printing made easy with brand design company in Atlanta. The medical supply catalog specialty formulated design helps medical customers provide a high level of quality care to their patients.
This document provides specifications for the TOSHIBA 6N137 photocoupler, which consists of a high-emitting diode and a one-chip photo IC in an 8-lead DIP package. Key features include LSTTL/TTL compatibility with a 5V supply, ultra-high speed of 10Mbps, guaranteed performance from 0-70°C, high isolation voltage of 2500Vrms minimum, and UL recognition. The document includes pin configurations, electrical characteristics tables, application information, and restrictions on product use.
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COILPT. Siwali Swantika
Pemesanan produk, hubungi PT Siwali Swantika melalui WhatsApp, Jakarta : 0811-1519-949 (chat only) | Surabaya : 0811-1519-948 (chat only). Kunjungi website kami di www.siwali.com, untuk detail informasi spesifikasi dan model alat.
Allied autovent-4000-with-cpap-instruction-manualMario O
This document provides an operation manual for the Autovent 4000CPAP pneumatically powered ventilator and CPAP generator. It describes the key features and specifications of the device, outlines how to connect it to a patient and oxygen source, and explains how to operate it in both ventilation and CPAP modes. The manual also covers maintenance, accessories, and warranty information.
Liquisys M CPM223/253. pH/ORP Measurement. Transmitter for analog and digital glass and ISFET sensors. Email: lam.nguyen@vietan-enviro.com HP: 0945 293292
The S4000C is an intelligent sensor that monitors combustible gases and provides status outputs. It contains electronics in an explosion-proof housing and measures gas concentration through a 4-20mA output signal. The S4000C has a digital display, alarm relays, and optional MODBUS communications. It operates in normal, calibration check, calibration, and setup modes and can self-calibrate using a magnetic switch.
The document provides information on various handheld measurement devices sold by Interface Devices India Pvt. Ltd, including calibrators, multi-meters, clamp meters, insulation testers, earth testers, and other products. It describes the key features and specifications of different product models. The models range from basic analog devices to more advanced digital instruments with additional measurement and data logging capabilities. Interface Devices is an authorized distributor of Yokogawa handheld measurement products in India.
This document provides information on air filter testing equipment from FEC. It describes testing systems that can test various types of air filters according to standards like EN 779 and ASHRAE 52.2. The systems can test parameters like flow rate, differential pressure, dust holding capacity, and efficiency. They include components like test ducts, centrifugal blowers, aerosol generators, and particle counters. FEC offers customizable filter testing solutions at various throughput capacities.
This is a work of the organization subject. Here, we can find the differences of Mercator, Robinson and Azimuthal's Projections. Moreover we can find an example of Treasure Hunt game for little children and finally, an example of a media distortion that can change the vision of students
Soil Management: National Organic Program RegulationsGardening
This document discusses the National Organic Program's soil fertility and crop nutrient management practice standard. It provides:
1) Guidelines for soil management practices that maintain or improve soil physical, chemical, and biological conditions while minimizing erosion as required by the standard.
2) Suggestions for how different soil management practices can meet the objectives of improving soil properties and reducing erosion.
3) Methods for monitoring soil conditions to ensure compliance with the standard's requirements for sustainable organic production systems.
Federal Conservation Resources for Sustainable Farming and Ranching Gardening
This publication provides an overview of major federal conservation programs that provide resources for farmers and ranchers to implement sustainable farming and ranching practices. It describes how the availability of conservation resources has increased dramatically since 2002. The publication aims to help farmers and ranchers navigate the complex application processes to access these resources, which can help preserve agricultural lands, develop sustainable practices, and open new markets. It introduces the story of Anna Jones-Crabtree and Doug Crabtree, organic farmers who have benefited from new NRCS programs like EQIP but have also faced some challenges in working with NRCS.
This document provides a summary of key concepts about sustainable soil management from a publication by the National Sustainable Agriculture Information Service. It discusses the importance of soil organisms in maintaining healthy soil and explains how native ecosystems function without tillage or fertilizers by recycling nutrients through soil food webs and plant litter. Maintaining high levels of organic matter and diverse soil life through appropriate management practices helps soils remain productive over the long term in a sustainable manner.
La psicología es la ciencia que estudia la conducta y los procesos mentales de los individuos. Antes era parte de la filosofía. Trata sobre el hombre como un compuesto de alma y cuerpo, y los diferentes tipos de alma: vital, sensitiva y racional. Los seres humanos son animales racionales que se diferencian de otros animales por su uso de la razón para pensar y resolver problemas. La psicología también estudia conceptos como el instinto, el alma, las potencias del alma como la memoria, el entendimiento y la voluntad.
Clinical practice guidelines for mild traumaticRichard Radecki
This document summarizes new Canadian clinical practice guidelines for managing mild traumatic brain injury (MTBI) and persistent post-concussive symptoms (PPCS). It outlines the development process, which included a systematic review of existing guidelines and research evidence. An expert panel then created 71 recommendations using a modified Delphi consensus process. The recommendations address diagnosis and management of both acute MTBI and long-term PPCS. The full guidelines provide an evidence-based approach to help healthcare professionals properly assess and treat individuals experiencing ongoing symptoms after MTBI.
ATTRA is a national sustainable agriculture information service operated by the National Center for Appropriate Technology through a USDA grant. This document provides information on organic small grain production, which excludes commercial fertilizers and pesticides and relies on crop rotations, cover crops, and biological pest control. It discusses soil fertility management, insect and disease management primarily through integrated pest management, weed control through rotation and cultivation, and organic small grain marketing, which often involves contracting and storage.
Test doubles like stubs and mocks are used to isolate parts of a system under test. Stubs simulate inputs to the system under test, helping to define the "given" part of a test. Mocks are used to verify outputs of the system under test, helping to define the "then" part of a test by setting expectations and verifying interactions. Examples show using stubs to simulate a grades repository input and using mocks to verify that a domotic system closes windows and doors when activating security. It is recommended to test features rather than classes in isolation and to avoid using test doubles for value objects or types not owned.
The document discusses the American Renaissance period in the mid-19th century and the major Transcendentalist writers that helped develop a uniquely American voice and perspective. Key writers mentioned include Henry Wadsworth Longfellow, Harriet Beecher Stowe, and Herman Melville. Transcendentalist literature encouraged individuals to find their own relationship with nature and advocated for positive relationships between people and anti-slavery views. Important works that influenced society include Longfellow's poem "The Slave's Dream", Stowe's novel Uncle Tom's Cabin, and Melville's poem "The Portent" about John Brown.
Catalog design and printing made easy with brand design company in Atlanta. The medical supply catalog specialty formulated design helps medical customers provide a high level of quality care to their patients.
This document provides specifications for the TOSHIBA 6N137 photocoupler, which consists of a high-emitting diode and a one-chip photo IC in an 8-lead DIP package. Key features include LSTTL/TTL compatibility with a 5V supply, ultra-high speed of 10Mbps, guaranteed performance from 0-70°C, high isolation voltage of 2500Vrms minimum, and UL recognition. The document includes pin configurations, electrical characteristics tables, application information, and restrictions on product use.
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COILPT. Siwali Swantika
Pemesanan produk, hubungi PT Siwali Swantika melalui WhatsApp, Jakarta : 0811-1519-949 (chat only) | Surabaya : 0811-1519-948 (chat only). Kunjungi website kami di www.siwali.com, untuk detail informasi spesifikasi dan model alat.
Allied autovent-4000-with-cpap-instruction-manualMario O
This document provides an operation manual for the Autovent 4000CPAP pneumatically powered ventilator and CPAP generator. It describes the key features and specifications of the device, outlines how to connect it to a patient and oxygen source, and explains how to operate it in both ventilation and CPAP modes. The manual also covers maintenance, accessories, and warranty information.
Liquisys M CPM223/253. pH/ORP Measurement. Transmitter for analog and digital glass and ISFET sensors. Email: lam.nguyen@vietan-enviro.com HP: 0945 293292
The S4000C is an intelligent sensor that monitors combustible gases and provides status outputs. It contains electronics in an explosion-proof housing and measures gas concentration through a 4-20mA output signal. The S4000C has a digital display, alarm relays, and optional MODBUS communications. It operates in normal, calibration check, calibration, and setup modes and can self-calibrate using a magnetic switch.
The document provides information on various handheld measurement devices sold by Interface Devices India Pvt. Ltd, including calibrators, multi-meters, clamp meters, insulation testers, earth testers, and other products. It describes the key features and specifications of different product models. The models range from basic analog devices to more advanced digital instruments with additional measurement and data logging capabilities. Interface Devices is an authorized distributor of Yokogawa handheld measurement products in India.
This document provides information on air filter testing equipment from FEC. It describes testing systems that can test various types of air filters according to standards like EN 779 and ASHRAE 52.2. The systems can test parameters like flow rate, differential pressure, dust holding capacity, and efficiency. They include components like test ducts, centrifugal blowers, aerosol generators, and particle counters. FEC offers customizable filter testing solutions at various throughput capacities.
This is a work of the organization subject. Here, we can find the differences of Mercator, Robinson and Azimuthal's Projections. Moreover we can find an example of Treasure Hunt game for little children and finally, an example of a media distortion that can change the vision of students
Soil Management: National Organic Program RegulationsGardening
This document discusses the National Organic Program's soil fertility and crop nutrient management practice standard. It provides:
1) Guidelines for soil management practices that maintain or improve soil physical, chemical, and biological conditions while minimizing erosion as required by the standard.
2) Suggestions for how different soil management practices can meet the objectives of improving soil properties and reducing erosion.
3) Methods for monitoring soil conditions to ensure compliance with the standard's requirements for sustainable organic production systems.
Federal Conservation Resources for Sustainable Farming and Ranching Gardening
This publication provides an overview of major federal conservation programs that provide resources for farmers and ranchers to implement sustainable farming and ranching practices. It describes how the availability of conservation resources has increased dramatically since 2002. The publication aims to help farmers and ranchers navigate the complex application processes to access these resources, which can help preserve agricultural lands, develop sustainable practices, and open new markets. It introduces the story of Anna Jones-Crabtree and Doug Crabtree, organic farmers who have benefited from new NRCS programs like EQIP but have also faced some challenges in working with NRCS.
This document provides a summary of key concepts about sustainable soil management from a publication by the National Sustainable Agriculture Information Service. It discusses the importance of soil organisms in maintaining healthy soil and explains how native ecosystems function without tillage or fertilizers by recycling nutrients through soil food webs and plant litter. Maintaining high levels of organic matter and diverse soil life through appropriate management practices helps soils remain productive over the long term in a sustainable manner.
La psicología es la ciencia que estudia la conducta y los procesos mentales de los individuos. Antes era parte de la filosofía. Trata sobre el hombre como un compuesto de alma y cuerpo, y los diferentes tipos de alma: vital, sensitiva y racional. Los seres humanos son animales racionales que se diferencian de otros animales por su uso de la razón para pensar y resolver problemas. La psicología también estudia conceptos como el instinto, el alma, las potencias del alma como la memoria, el entendimiento y la voluntad.
Clinical practice guidelines for mild traumaticRichard Radecki
This document summarizes new Canadian clinical practice guidelines for managing mild traumatic brain injury (MTBI) and persistent post-concussive symptoms (PPCS). It outlines the development process, which included a systematic review of existing guidelines and research evidence. An expert panel then created 71 recommendations using a modified Delphi consensus process. The recommendations address diagnosis and management of both acute MTBI and long-term PPCS. The full guidelines provide an evidence-based approach to help healthcare professionals properly assess and treat individuals experiencing ongoing symptoms after MTBI.
ATTRA is a national sustainable agriculture information service operated by the National Center for Appropriate Technology through a USDA grant. This document provides information on organic small grain production, which excludes commercial fertilizers and pesticides and relies on crop rotations, cover crops, and biological pest control. It discusses soil fertility management, insect and disease management primarily through integrated pest management, weed control through rotation and cultivation, and organic small grain marketing, which often involves contracting and storage.
Test doubles like stubs and mocks are used to isolate parts of a system under test. Stubs simulate inputs to the system under test, helping to define the "given" part of a test. Mocks are used to verify outputs of the system under test, helping to define the "then" part of a test by setting expectations and verifying interactions. Examples show using stubs to simulate a grades repository input and using mocks to verify that a domotic system closes windows and doors when activating security. It is recommended to test features rather than classes in isolation and to avoid using test doubles for value objects or types not owned.
The document discusses the American Renaissance period in the mid-19th century and the major Transcendentalist writers that helped develop a uniquely American voice and perspective. Key writers mentioned include Henry Wadsworth Longfellow, Harriet Beecher Stowe, and Herman Melville. Transcendentalist literature encouraged individuals to find their own relationship with nature and advocated for positive relationships between people and anti-slavery views. Important works that influenced society include Longfellow's poem "The Slave's Dream", Stowe's novel Uncle Tom's Cabin, and Melville's poem "The Portent" about John Brown.
The steel pan originated in Trinidad from bands that used bamboo instruments called tamboo bamboo bands. As bamboo proved too fragile, metal instruments like oil drums and biscuit tins were used instead to create louder, more durable sounds. By the 1940s, the steel pan had evolved into its current form with different pitched notes and Spree Simon and Ellie Mannette pioneered the ping pong lead pan. In the 1950s, the Trinidad All Steel Percussion Orchestra was formed and helped organize the steel pan movement, led today by Pan Trinbago. The annual Panorama competition began in 1946 and showcases large steel bands with talented arrangers. Over time, the steel pan has spread globally while continuing to
Massive Open Online Courses (MOOCs) provide unlimited free online courses from top universities worldwide. MOOCs can be accessed from any internet-connected device, allowing for flexible self-paced learning. They include videos, lecture notes, assessments, and opportunities for peer-to-peer learning. While MOOCs increase access to education, disadvantages include lack of interaction with instructors, high dropout rates due to lack of structure, and need for reliable internet access.
How to deal with the challenges, problems that happen to you when you get back to your office table every morning!!
Try to manage the things as they usually start occurring when you wake up, one should not take it to the nerves :)
Change arise from the inside of the person, if one can't hold the effects then is not able to confront the assignments
A Brief Overview of Nutrient Cycling in PasturesGardening
This document provides an overview of nutrient cycling in pastures through three key components: soil organisms, pasture plants, and grazing livestock. It describes how nutrients cycle naturally through the soil-plant-animal system when pastures are well-managed. Soil organisms like bacteria, fungi and earthworms break down organic matter and release nutrients in forms available to pasture plants. Plants take up nutrients from the soil and fix nitrogen in the case of legumes. Grazing livestock consume nutrients from forage but most are returned to the pasture through manure and urine, completing the nutrient cycle. Proper management, like rotational grazing and soil testing, can enhance this natural cycling of nutrients in pasture systems.
This document summarizes information about sustainable fire ant management techniques. It discusses the release of fire ant-decapitating phorid flies in several southern states to control fire ant populations. It also discusses experimental releases of a protozoan disease to infect and kill fire ant queens. A number of least-toxic fire ant baits and other control methods such as hot water and traps are also mentioned.
This document provides an overview of bamboo, including its types, uses, cultivation, and markets in the United States. It discusses the two main types of bamboo - clumping and running - and their cold hardiness. Bamboo has multiple uses, including construction material, musical instruments, furniture, and as an ornamental/landscaping plant. The document outlines steps to develop a bamboo business and potential markets. It focuses on bamboo shoots as a food crop and notes their popularity in Asian cuisine, though most are currently imported. Harvesting shoots is recommended when they first emerge from the ground for best nutritional quality.
This document is the Final Environmental Impact Report for the proposed E & B Oil Drilling & Production Project in Hermosa Beach, California. The proposed project includes oil drilling and production operations as well as construction of a new City maintenance yard. Key aspects of the proposed oil project analyzed in the report include site preparation, drilling and testing of up to 10 new wells, construction of processing facilities and pipelines, long-term operations, and eventual decommissioning. The report also evaluates alternatives to the proposed project and their environmental impacts. Potential impacts of the proposed projects on aesthetics, air quality, biology, cultural resources, energy, and other issue areas are identified and mitigation measures are proposed.
The document is a service manual for the BM3 patient monitor that provides:
1) Contact information for product inquiries, service calls, and technical support.
2) Safety precautions and guidelines for using and maintaining the BM3 monitor.
3) Specifications for the monitor's performance, including its display, interfaces, battery life, and the accuracy of its sensors for measuring SpO2, NIBP, ECG, and respiration.
This document introduces the ENE-001 portable gas analyzer. Key details include:
- The ENE-001 can detect PM 2.5, PM 10, combustible gases, carbon monoxide, hydrogen sulfide, and oxygen using advanced sensors and microprocessor control.
- It is compact and easy to use, making it suitable for applications like sewage plants, tunnels, waste sites, and more.
- The document describes the product launch, specifications, technology, charging, calibration procedures, serial communication, and sensors used - including combustible, oxygen, CO, H2S, and PM sensors.
This document is the service manual for the TI-2000 Transport Incubator. It provides instructions on installation, technical specifications, working principles, and preventative maintenance of the incubator. The manual contains warranty information and safety testing procedures to be performed annually on the device. Contact information is provided for technical support and authorized service representatives.
Manual de operación y mantenimiento concentrador de oxigeno jay-5AntonioBarrios54
This document is a user manual for an oxygen concentrator. It provides instructions on safety, operation, maintenance and troubleshooting of the device. The manual contains sections on symbols used, safety notices, product introduction, technical specifications, operating instructions, alarms and safety features, maintenance requirements, troubleshooting tips, and other important information for the proper use of the oxygen concentrator.
This user manual provides instructions for operating an oxygen concentrator. It includes 17 sections that describe safety notices, product introduction, operation conditions, technical parameters, structures and functions, operation instructions, alarms and safety devices, maintenance, troubleshooting, EMC declaration, accessories, transportation and storage conditions, quality warranty, and contact information. The manual contains information to ensure the oxygen concentrator is used safely and efficiently.
Hướng dẫn sử dụng máy đo tốc độ gió, nhiệt độ Extech 407123Tenmars Việt Nam
Hướng dẫn sử dụng máy đo tốc độ gió, nhiệt độ Extech 407123
http://extech.vn/vi/products/do-toc-do-va-luu-luong-gio
http://extech.vn/vi/products/do-toc-do-va-luu-luong-gio/extech-407123---m%C3%A1y-%C4%91o-t%E1%BB%91c-%C4%91%E1%BB%99-gi%C3%B3-nhi%E1%BB%87t-%C4%91%E1%BB%99--407123
Hướng dẫn sử dụng Testo 435
https://testostore.vn/san-pham/dong-ho-do-da-nang-testo-435-3/
https://testostore.vn/danh-muc/do-ap-suat/may-do-ap-suat-vi-sai/
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3Tenmars Việt Nam
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
https://testostore.vn/san-pham/dong-ho-do-da-nang-testo-435-3/
https://testostore.vn/san-pham/may-do-da-nang-testo-435-4/
https://testostore.vn/danh-muc/do-nhiet-do-do-am/do-nhiet-do-be-mat/
The document is an operating and maintenance manual for the E4500 Combustion Analyzer. It provides instructions on safety, startup, configuration, measurements, combustion analysis, sensors, maintenance and troubleshooting. Key features of the E4500 include electrochemical gas sensors to analyze flue gas samples drawn into the device via a sampling probe and pump. The manual describes the components, technical specifications, connection diagrams and step-by-step instructions for operating the E4500 for applications such as emissions monitoring, efficiency calculations, and tightness testing.
Hướng dẫn sử dụng máy đo đa năng Testo 435 đo nhiệt độ độ ẩm, đo áp suất, tốc độ gió...
Máy đo áp suất đa năng Testo 435: https://testostore.vn/san-pham/dong-ho-do-da-nang-testo-435-3/
Máy đo nhiệt độ bề mặt: https://testostore.vn/danh-muc/do-nhiet-do-do-am/do-nhiet-do-be-mat/
Website: testostore.vn
Hotline: 0914222214
Email: info@thietbido.us , info@tenmars.vn
This document provides operation instructions for an oxygen concentrator. It begins with safety notices, then describes the product, operating conditions, name and function of parts, and operation steps. It also covers maintenance including cleaning filters, troubleshooting, transportation/storage conditions, and a 1 year warranty. Key points covered are safety warnings for oxygen use, operating the controls and display, cleaning filters regularly, and basic troubleshooting tips.
The Cloud Point test cell is first flushed through with a fresh sample. A small amount of sample is then isolated in the cell and allowed to settle for a predetermined period to allow any air bubbles or solids to separate. The sample cell is then cooled at a controlled rate by a Peltier cooler with excess heat being removed by a coolant. An optical detector senses when the sample is starting to become cloudy as waxy components start to precipitate. The temperature at which this occurs is taken as the cloud point
We have Baumer make DP Transmitters. DP Transmitters are also known as Differential Pressure Transmitters, Pressure Difference Transmitters, PD Transmitters, DP Transmitters
Hướng dẫn sử dụng testo 435-3
https://testostore.vn/danh-muc/do-ap-suat/may-do-ap-suat-vi-sai/
https://testostore.vn/san-pham/dong-ho-do-da-nang-testo-435-3/
Hướng dẫn sử dụng Testo 435-3
https://testostore.vn/san-pham/dong-ho-do-da-nang-testo-435-3/
https://testostore.vn/danh-muc/do-ap-suat/may-do-ap-suat-vi-sai/
TECHNICAL SPECIFICATIONS: Display-->Type : LED, Parameters : SpO2, PR, Pleth bar., Information : low power ::SPO2-->Measure Range : 35-99%, Resolution : ±1%, Accuracy : 75%-99%:±2, 50%-75%:±3, 0%-50%:unspecified::Pulse Rate-->Range : 30-254 BPM, Resolution : ±1bpm, Accuracy : ±2bpm::Battery--> Type : 2 AAA Alkaline batteries or rechargeable, batteries Power Consumption : less than 30Ma, Voltage : 2.6-3.6V::Environmental-->Operating Temperature : 0'C-50'C, Storage Temperature : -10'C - 60'C, Operating Humidity : 15%RH - 95%RH, Storage Humidity : 10%RH - 95%RH, It can be used by a variety of sports enthusiasts, including mountain climbers, hikers, skiers, bikers, and others interested in measuring oxygen saturation and pulse rate, it is not for medical use. This oximeter is for anyone who is interested in knowing their oxygen saturation and heart rate. Applications would be for use during exercise or daily activities, during air travel for the passenger or pilot, and for extreme sports such as mountain climbing. This oximeter is not a medical device and does not require a prescription from your medical professional.
For oil transporter tank or big tank fuel monitoring, ultrasonic liquid level sensor is needed. Compared to voltage or pressure level sensor, ultrasonic fuel sensor is more accurate and easy to install. Uniguard Technology Limited offers complete fuel monitoring and fuel tracking solutions with GPS Tracker Model UT04.
- The document provides information on using a standby valve with an Isoprime mass spectrometer, including setup instructions, operating procedures, and maintenance tips.
- The standby valve allows switching between a sample analysis mode and standby mode without interrupting helium flow to the mass spectrometer.
- Proper installation and leak checking of connections is important for safe operation of the standby valve system.
RAC-050 Engineer Training V1.0e-202202.pptArnoldoFelix3
The document provides training materials for engineers on the RAC-050 instrument. It covers engineer tasks, instrument installation, daily and preventative maintenance, instrument operation principles, common troubleshooting issues, and how to perform adjustments to instrument positions, temperature, and optics. The training is divided into 5 parts that cover the instrument principle, specifications, installation, maintenance, and troubleshooting.
1. Meditech Equipment Co., Ltd is a medical device company located in Qingdao, China that produces syringe pumps and infusion pumps.
2. The document provides specifications for Meditech's single and double channel syringe pumps, including the MD910C, MD910, and MD910D, which can accurately deliver fluids from 0.1 to 1500 ml/hr and feature alarm systems and up to 2000 memory records.
3. Meditech's MDS50 infusion pump is also described, with a flow rate range of 1 to 1200 ml/hr and safety features such as occlusion detection and audible alarms for issues like air bubbles or low battery.
The DEFI 5T Trainer is a training tool for preparing emergency responders to use the AED defibrillator to treat a victim of sudden cardiac arrest.
The Trainer provides realistic simulation of the defibrillator without actually charging or discharging electrical energy. Voice prompts, simulated energy delivery, pauses for CPR, Training Pads, and the functional system status indicator all accurately mimic the operation of the defibrillator and provide a realistic demonstration.
D-SENSE
The accuracy of its performance parameters related to the safety of patients, must be through the corresponding detection equipment AED production of the performance parameters of the test and regular performance check.
Meditech D-Sense automated external defibrillator analyzer can simulate a variety of normal and abnormal ECG signals for defibrillator testing, test the ability of the defibrillator to analyze ECG signals, and display the defibrillator primary discharge energy and energy via the LED digital tube Discharge voltage values, the selection of ECG signals for different tests, and the display of voltage display or primary discharge energy are achieved by key operation. The overall operation of the instrument is simple and convenient.
Our medical products used in hospitals and clinics, such as EKG machine, Multiparameter Patient Monitor, Defibrillator, Ultrasound scanner, Pulse Oximeter, Fetal Doppler ,Fetal monitor, Blood Pressure, Visual stethoscope ..
All of our products have been approved by ISO13485 and have passed CE Certificates. We stick to delivering our high quality product and best service to our global base of customers.
As one of the leading medical equipment manufacturers, we are involved in R&D, manufacturing and marketing. And as a global enterprise, our products are selling in more than 70 countries worldwide. We produce a wide range of Veterinary medical products used in Veterinary hospitals and clinics
This document provides technical specifications for the Meditech ET3 capnography system, including details on carbon dioxide measurement range and accuracy, sampling rate, safety features, environmental tolerances, and host interface specifications. The ET3 system allows for low-cost end-tidal carbon dioxide monitoring through use of simple connection sample lines. Key specifications include a CO2 measurement range of 0-152 mmHg, sampling rate of 100 Hz, accuracy of ±2 mmHg from 0-40 mmHg and compensation for expired oxygen, anesthetic agents and barometric pressure. The ET3 sensor is splash-proof, shock resistant and compatible with current Respironics airway adapters.
The document lists ultrasound probes including their model, frequency, price for 1-5 pieces, and bulk purchase price with required minimum. It includes a sector probe at 3.5MHz priced at $8300 for 1-5 pieces and $8100 for more than 5 pieces, a convex probe at 3.5MHz priced at $7000 with a minimum of 25 pieces, and a rectal probe at 7.5MHz priced at $5800 for 1-5 pieces and $5500 for more than 5 pieces.
معرفة تأثير تمرينات الايروبك على زيادة نسبة الأوكسجين بالدم لمرضى السكري غير ...Meditech Equipment Co.,Ltd
- معرفة تأثير تمرينات الايروبك على زيادة نسبة الأوكسجين بالدم
لمرضى السكري غير المعتمد على الأنسولين .
إجراءات البحث الميدانية:
وتمت إجراءات البحث باستخدام المنهج التجريبي على عينة عددهم
٤٠ سنة ) من مدينة بغداد. وقد - ٢٠ ) من المصاب ين بمرض السكر بأعمار ( ٣٠ )
استخدم الباحث تمرينات الايروبك وفق متغيرات بايوميكانيكية
استنتاجات البحث: وتوصل الباحث إلى الاستنتاج التالي :
١- ان تمرينات الايروب ك وفق متغيرات بايوميكانيكية ساهمت في زيادة نسبة
الاوكسجين بالدم لمرضى السكري غير المعتمد على الانسولين .
The Summery
The influence of suggested program by using some Aerobic
exercises to development of oxygen ability .
Teacher . ahmed shaker al- aubidy
The aims of the research:
1- To know the development of oxygen ability
The Procedures of the field research:
The procedures of the research have been completed by
using the experimental programme on a sample of (٢0) of the
diabetic patients their ages are about (٣0-٤0) years from
Baquba city. The researcher has carried out the rehabilitae
suggested programme by using the Irobic exercisis, before and
تأثير تمرينات الأيروبك وفق متغيرات بايوميكانيكية
لزيادة نسبة الاوكسجين بالدم للمصابين بمرض السكر
القبلية والبعدية
وتمت الاختبارات
.( Meditech Oxy O باستخدام جهاز الاوكسميتر (نوع
العدد الثالث وا . مجلة الفتح . نیسان لسنة ٢٠١٣ م. احمد شاكر العبیدي
-٢٣٠ -
1- The suggested rehabilitate programme by using the
Irobic exercises has helped to the development of oxygen
ability .
Cuore di bimbo BabySono per ascoltare e registrare il battito cardiaco del fe...Meditech Equipment Co.,Ltd
Come funziona il Doppler?
Il Doppler invia segnali ad alta frequenza (ultrasuoni = frequenze non udibili) e riceve segnali riflessi
(udibili) contenenti informazioni elaborate.
La tecnica è la stessa usata nelle ecografie, ma nel nostro caso, le potenze in gioco sono estremamente
basse dato che utilizziamo solo le informazioni audio e non quelle video.
Queste onde (segnali) riflesse contengono informazioni che ci permettono di sentire diversi rumori
provenienti dall’utero tra cui il battito cardiaco del feto.
Oxygen consumption (VO2) is the amount of oxygen taken up and utilized by the body per minute. The oxygen taken into the body at the level of the lungs is ultimately transported by the cardiovascular system to the systemic tissues and is used for the production of ATP in the mitochondria of our cells. Because most of the energy in the body is produced aerobically, VO2 can be used to determine how much energy a subject is expending.
In vitro canister life of cos absorber prepacks with different fresh gas flow...Meditech Equipment Co.,Ltd
This document presents the results of a study measuring the in vitro canister life of different CO2 absorber prepacks across a range of fresh gas flows. The study found that the fractional canister usage, or amount of each canister used per hour, increased linearly as the fresh gas flow increased for each brand. The canister life was longest for the SpiraLith brand, which is lithium-based and produces no compound A or CO. Of the calcium hydroxide-based brands, Amsorb and LoFloSorb, which do not contain sodium hydroxide, were less efficient but also do not produce compound A or CO.
Defi6 is a portable automatic external defibrillator (AED) that is lightweight, easy to use, and guides the user through the defibrillation process. It analyzes the patient's heart rhythm and will either prompt the user to deliver a shock if needed or begin CPR instructions. The AED is designed to be used in situations where emergency responders may be several minutes away. It uses adhesive pads to deliver the shock and has features like voice prompts, ECG analysis, and event recording for reporting.
This finger pulse oximeter continuously monitors blood oxygen saturation levels (SpO2) and pulse, can operate for 30 hours on AAA batteries, has a low voltage alarm, auto-shuts off after 8 seconds without a signal, and is small, light, and easy to use making it portable for medical professionals.
This multi-parameter veterinary monitor has a 12.1" color display and is lightweight and portable with a built-in rechargeable lithium battery. It can display ECG waveforms from 7 leads simultaneously and store 72 hours of parameter trends and recall 72 alarm events. Additional features include 32 seconds of full-disclosure waveform review, storage of 500 NIBP measurements, adjustable data and waveform colors, arrhythmia and ST segment analysis, and suitability for monitoring various pet types.
This document describes a multi-parameter patient monitor with the following key features:
- 12.1" color display to show ECG waveforms of 7 leads, 72-hour trends, and alarms.
- Built-in rechargeable lithium battery for portable use.
- Monitors ECG, SpO2, NIBP, respiration, temperature and optional modules like IBP and CO2.
- Stores and recalls measurement data, waveforms and alarms for review.
- Compact size of 310 x 140 x 263 mm and lightweight at 3.8 kg.
Meditech SpO2 Module MD100 is a Low-power oxygen monitor module; the SPO2 Module is used in the sampling inspection or measurement of the patient's oxygen saturation and pulse rate. As to the average oxygen saturation, low-power module can send instantaneous blood oxygen saturation. It is very useful, especially in the sleep research. The SPO2 module applies to infants, newborns and adults.
The Meditech SPO2 Module MD200 is a low-power oxygen monitor module that can measure oxygen saturation and pulse rate in infants, children, and adults. It has an operating voltage of 5V, measures oxygen saturation from 50-100% with an accuracy of ±2, and pulse rate from 30-235 beats per minute with an accuracy of greater than ±2 beats per minute or ±2%. The module is compatible with BCI probes and can transmit data in real time.
2. 90E Capnography monitor
Notice
Welcome to use the product of Meditech Equipment Co. Ltd
This manual includes the materials and copyright reserved. It is not allowed to copy,
reduplicate or translate into other languages without our written permission.
Please to read this manual carefully and then to operate by the instruction of this manual.
It is not allowed to open the monitor‟s cover without our permission.
Some changes on the product due to the technologies‟ promotion or due to the special
demands of user which do not influence the monitor‟s work will not be informed further.
Furthermore please pay attention to the difference between the parts or components and
this manual. You may apply to our company for technical documents about the electric circuit
diagram relevant batch and lists of parts or components and so on because the continuous
promotion of the product. You may contact us by means of the following address.
Meditech Equipment Co., Ltd (Meditech Group)
Address: Nanjing Road No.100, Qingdao, Shandong Province, P. R. China
Tel: 86-532-85832673 。85820563
Fax: 86-532-85832155
1
3. 90E Capnography monitor
CONTENTS
Chapter I Preface ............................................................................ 3
Chapter II Technical specifications and characteristics ..................... 4
Chapter III Introduction of Monitor.................................................... 5
Chapter IV Patient connection ......................................................... 7
Chapter V Screen display and Operation ....................................... 10
Chapter VI Charge、Maintenance、sterilization、cleanness ......... 16
Chapter VII Analysis of problems ................................................... 17
Appendix 1 Explications for the appreciated terms in the device .... 18
Appendix 2 How to change compensation of balance gas ............. 19
2
4. 90E Capnography monitor
Chapter I Preface
I. Brief
The purpose of this manual is for quick understanding of characteristics, functions, operation of the
monitor, preventing from mal-operation and mistakes.
This monitor can supervise four physical parameters of patient at the same time: concentration of
EtCO2, respiration rate, pulse and saturation of SPO2 (optional). The monitor you bought may have
two or more functions mentioned above but this manual can be used in common.
II. Warranty and Maintenance
Warranty
This monitor is free of maintenance within one year since the day it has been purchased. But the
accessories such as SpO2 sensor etc. are free only within three months.
The following is out of warranty:
The monitor is damaged due to mal-operation (without following the manual instruction)
The monitor is managed due to incorrect connection with other instrument.
The monitor is damaged due to accident.
The user changes the monitor on his own will without written authority of the company.
The series number is tore off or unreadable.
Maintenance
If the monitor is broken down beyond warranty period, the manufacturer or distributor will
maintain it with charge. The maintenance place depends on actual conditions.
Re-packing
To take out all the sensors and accessories, then to pack them in plastic bag.
It is recommended to use the original packing boxes and packing materials.
It is better to provide description for broken-down phenomenon.
III. Safety Demand
For the purpose of safety, please read the following and abide by the instruction of medical
instrumental products.
Warning: Indicating the possible injury on patient or operator.
·
This monitor is banned to supervise patient for suffocation, and also is banned to apply during
the operation of MRI or CT.
·For the monitor‟s safely grounding, the hospital shall provide three-pole socket.
· use the designated accessories, which complying with standard of manufacturer...
To
· is banned to remove the cover of the monitor.
It
·
This monitor provides Concentration of EtCO2, Respiration rate or Oximeter saturation and
Pulse. These data only do auxiliary help for diagnosis but the real diagnosis shall be made by
doctor through the clinical feature and symptom.
· order to keep the patient‟s skin breathe freely, the SpO2 „s sensor needs to reput on a new
In
place at least every 24 hours.
3
5. 90E Capnography monitor
Chapter II Technical specifications and characteristics
1. The normal working conditions:
Environment temperature: 10℃~40℃
Relative humidity: 30%~75%
Voltage: AC 100V~250V, Frequency: 50Hz/60 Hz
Air pressure: 86kPa~106 kPa
2. Concentration of EtCO2
Range: 0.0-19.7% (V/V), 0-150mmHg,or 0-20kPa
Accuracy: deviation<±
2mmHg when EtCO2 concentration in 0.0~5.0%.
deviation<± (reading) when EtCO2 concentration > 5.0%.
6%
Alarm accuracy: accuracy<±
0.2%
3. Respiration rate
Range: 3 ~120 times/min.
Deviation: when 3-50 times/min±2; when 50-120 times±
5
4. Oximeter Saturation (optional)
Range: 0%~100%
Accuracy: when 50%~79%, ±
3% (absolute value);
when 80%~100%, ±
2%(absolute value)
5. Pulse (optional)
Range: 30 times/min~250times/min
Deviation: when 30times/min~40times/min, ±
2times/min (absolute value);
when 41times/min~250times/min, ±
5%(absolute value)
6. Trend
Parameter: EtCO2 concentration, respiration rate, SpO2 (optional) and Pulse (optional))
Range: 24 hours
7. Power Demand:
Voltage: AC 100V-250V, Frequency 50Hz/60 Hz
Unit Consume: ≤50VA
Fuse: 250V, 2.0A, Ф5×20
8. Battery
Battery pack 12Volt, 1.8Ah
Battery fully charged time: 4h
9. The conditions of transportation and storage:
Transportation Demand: It is not allowed for striking, violent vibration and humidity.
Environment temperature range: -40℃~+55℃
Relative humidity range: ≤93%
Storage demand: In a dry and good ventilating room without corrosive gas and strong magnetic field.
10. Size and weight
Size: 230mm (W) x160mm (H) x 50mm(D)
Weight: 1150g
4
6. 90E Capnography monitor
Chapter III Introduction of Monitor
I. Introduction of front face
Front face as follows:
FIGURE 1
(1) Screen: It can display waves, menu, alarm and all measuring parameters.
(2) ▲/ : Function button:
▲: When menu is activated, to press this button to move cursor.
: On the main display, to press this button to silence in 2 minutes.
(3) ▼: To press this button to move the cursor when menu is activated.
(4)►: To press this button to increase figures.
(5)◄: To press this button to decrease figures.
(6) OK: Confirmation button: To press this button to enter son menu. In the main menu, to press this
button to turn on the pump directly when it is off duty.
(7)
To press this button to enter or quit menu.
(8)
Power button
(9) Indicator AC: AC power indicator, If it is bright, the AC is connected.
(10) Indicator CHARGE: Brightening showing the machine is working, and twinkling shows battery is
charging power.
(11) Indicator LACK: Battery lacking power indicator. If it is twinkling, the monitor is in the condition of
lacking power and please to connect AC in time, otherwise the monitor will alarm and automatic
shutdown will occur.
(12) Indicator ALARM: Alarm indicator. Twinkling shows some parameter is alerting.
(13) CO2: The port of filter of CO2
(14) SpO2: The socket of SpO2 (optional).
II. Introduction of back face
Back face as follows:
5
7. 90E Capnography monitor
FIGURE 2
(1) AC Power socket,
(2) Fuse socket,
(3) Fan. Attention: The fan will not work when the inner surrounding is low temperature in order to
save power consumption.
(4) RS232 port
(5) The exhaust port of gas
(6) The trumpet position
(7) Battery cabin
(8) The connection position with the anesthetic machine or other devices
6
8. 90E Capnography monitor
Chapter IV Patient connection
I .CO2 measurement
The measurement value will be more accurate if the monitor has 20 minutes‟ warming-up. To connect
the CO2 filter with 3-way joint of the loop of anesthetic machine by sampling tube, or to directly connect
patient‟s nose by the hose.
FIGURE 3
1. Installation of water trap:
To take out the water trap from the package, then to hold its surface to put it into the fixing frame until
into the position.
No using the monitor without correct installation lest it should pollute the monitor.
When the water trap is replaced or cleared up, to press the clip so as smoothly to take off or install it,
as the figure3.1 showing:
FIGURE 3.1
2. Connection of sampling tube:
To screw the sampling tube tightly into the gas inlet of water trap.
The measurement values will not be true if leakage occurs.
To use the sampling tube provided by the original manufacturer otherwise the measurement values
will not be correct.
7
9. 90E Capnography monitor
3. To screw the other end of sampling tube into Y tube of loop of anesthetic machine.
II. Respiration rate measurement
The calculation of respiration rate derives from monitoring the wave of CO2.
III. Theory introduction
1 The theory of CO2 probe:
The device working theory is NON-DISPERSIVE INFRA RED GAS ANALIZER. The device has an
AUTO ZERO ADJUSTMENT SYSTEM and GAIN CONTROL
2 Automatic Offset Calibrations
The device was designed to automatically perform calibrations in order to correct for changes in
temperature, altitude and electronic component drift. At this moment, a word with “CAL” appears on
left down corner and respiration wave becomes one straight line.
The air surrounding the device may have elevated concentrations of CO2 present (such as in an
enclosed compartment or room with poor ventilation), we recommend cleaning the CO2 from the
ambient air with a ventilator, otherwise, the device will cause deviation.
3 The Moisture Separation System:
Chamber-type filter is used on the monitor for filter handling large amounts of moisture with a minimal
amount of dead space effect on waveform. It collects the moisture in a bottle and passes the gas to
the monitor. The monitor must be placed flatly to make the filter position is stable, since the filter
depends on gravity to operate properly.
IV. Oximeter density measurement (optional)
1. The finger oximeter sensor is recommended to use. The measurement will begin when your finger is
put into the sensor clip, meanwhile, the PLETH wave will appear on the screen, after several
seconds the saturation and pulse rate appear also. The monitor will utter a sound of “Di” when the
each measurement happens. The tone will vary if the values of SpO2 and Pulse Rate surpass the
alarm scope. The more values of SpO2 and Pulse Rate surpass the sharper the tone will do.
The volume of pulse beep can be modulated and closed down by the item of BEEP VOLUME in the
SOUND SET menu. The pulse beep will disappear under the silent condition.
FIGURE 4
2. The use of SpO2 sensor for neonate (optional)
The SpO2 sensor of neonate has two parts: Y-shape sensor and fixing wrapping and binding part.
When it is being used, Y-shape sensor must be inset into wrapping and binding part. It should be
wrapped and bound on the palm or foot of neonate. Pay attention: the two windows of Y-shape
sensor should be paralleled and face to face, only in this way can the light signal emitted by
light-emitting window, through the body blood, irradiates on the receiving window. If the two windows
are not paralleled and face to face, the monitoring result may be not accurate.
In addition, when SpO2 is not being monitored the probe should be taken off the monitor or the two
windows of sensor should be kept face to face, otherwise the light window will keep shining and
PLETH wave will be disordered and the screen will appear “FAIL SEARCH”
8
10. 90E Capnography monitor
V. Notice
1. Caution:
Showing clearly possible mal-operation to the device
Under the conditions of electromagnetic influence, for example: electrosurgical devices、MRI、CT
etc. will cause negative impact.
The filter should be taken off and replaced when it approaches full of water, otherwise seeping
water will bring irreversible damage for inner detector. Be sure that the collecting pipe is fluent
otherwise it will make inner sampling pump overwork and influence pump life.
2. Attention: indicating other important information.
1). CO2:
The readings will be wrong if the monitor does not warm up.
The sampling pipe provided by the manufacturer, shall be used, otherwise the readings may
be inaccurate.
Under the environment of roughly changed temperature, data measured may not be accurate.
Therefore the screen will show “TEMP IMBALANCE” when quickly changed temperature overruns
a certain range. It is better to use the monitor under the environment stable temperature.
The measured data may be influenced a little bit if the patient is inflicted by kinds of anesthetic
gases. To calibrate some interferential gases if necessary. Please refer to Appendix 2.
Any circumstances of blocking sampling gas, such as bending、 folding、 contaminations‟ blocking
sampling tube 、filter and water trap etc. will lead to inaccurate measurement.
Serious respiratory crock up and so on leading to exhaled CO2 concentration extremely high or
low, e.g. EtCO2 lower than 0.5% or higher than 11%, will generate inaccurate measurement.
No mater what brings loop leak, which will seriously influence accuracy of data measured and
shape of wave.
2). Oximeter:
The monitor‟s measurement to SpO2 may be influenced under outside strong light therefore to
cover the probe (e.g. to use surgical tower etc.) if necessary.
Accuracy of oximeter readings will be influenced if there are some dyestuff in blood.
Please use SpO2 probe attached with the unit or the one specially designated to this unit.
It may be the problem of SoP2 probe if the SpO2 readings or the wave of plethysmogram are
obviously abnormal. Therefore to check if the finger has been put into probe enough deep, if the
direction is right , if the there are some contaminations inside, if the probe is somehow broken or if
the two ends of prolonged cable of SpO2 fall off.
Warning:
It shall not be used if the sensor has some electrically naked and broken interfaces.
The test of high inside pressure of thorax and blood pressure or some other continuous
negative influence to venal circle will lead to vein vibration. The signal of pulse will disappear.
If shock、low blood pressure、serious blood vessel shrinking、serious anemia、over low
temperature、artery block near sensor and incomplete heart shrinking appear, pulse signal will
disappear.
9
11. Chapter V Screen display and Operation
I. Screen main display menu
FIGURE 5
When the power is turned on, the screen will firstly show version and series number for 2 seconds,
then main window shows as FIGURE 5.
The first line of data on the screen shows time (date/month/year, hour, minute, second), closing of
alarm sound (!), silence (
) or not silence (), battery level indicator
The middle area shows waves of plethysmogram and respiration of CO2..
The last line shows in turn the concentration of EtCO2, respiration rate、the concentration of
inhaling CO2(optional), inhaling or exhaling state(when exhaling
turning less).
The right area shows readings of SpO2, histogram and pulse rate (These exist only when SpO2 is
optional)
When the pump is turned off, the central place of the screen appears „PUMP OFF ‟.
If the device is not installed with SpO2, the screen shows as following:
FIGURE 5.1
Alarm indication:
1) If the EtCO2‟s value exceeds the limit of high or low alarm level, the word “EtCO2” will flash
and alert with „Di Di‟ .The same situation will happen too to respiration rate、
SpO2 and pulse.
2) If the battery level is so low that the battery
indicates completely empty, the screen will
flash fast with word of „LOW BATTERY‟, and monitor will utter sound continuously. At this
time the monitor will be closed down automatically.
10
12. Manual
3) When the apnea alarm is turned on and the apnea happens and alerts, the screen center will
appear with “APNEA” (It means to stop breathing) flashing, and if the sound alarm is turned
on , it will alert „DI DI‟.
4) When the SpO2 sensor is disconnected or off from finger, the screen will flash the word
“SENSOR OFF”. If the saturation could not be detected for long time, the screen will flash the
word “FAIL SEARCH”.
5) The volume of continuous or interval sound mentioned above can be modulated and closed
down by the item of ALARM_VOLUME. The sound will disappear under the silent condition.
If the alarm volume is 0, the silence indicator in the main menu will show '!'
6) All the parameters ' alarm for over limits and apnea alarm, will lead to flashing of red alarm
indicator on the panel
II. The main menu
FIGURE 6
When the main menu is open, to press the MENU button to enter the setting menu as FIGURE 6
shows.
In this menu, to press ▲ or ▼button to move the cursor up or down, To press OK button to enter
the next son menu.
This menu includes the following son menu:
The setting menu for CO2: CO2_SETUP
The setting menu for SpO2: SPO2_SETUP
The trend menu: TREND
The time menu: TIME_SETUP
The sound menu: SOUND_SETUP
11
13. Manual
III. CO2 setting menu
FIGURE 7
In this menu, to press ▲ or ▼button to move the cursor up or down, to press ► button or◄
button to change the data highlighted by the cursor.
Regarding some items in this menu, if not to change the data but just to operate them such as
LOAD DEFAULTS or EXIT, then to press OK button.
In this menu, to press MENU button, then to exit this menu and enter the main menu.
This menu includes the following setups:
1).The high alarm limits of EtCO2: EtCO2 ALARM_H: 22-99mmHg, off
2).The low alarm limits of EtCO2: EtCO2 ALARM_L: off, 10-60mmHg
3).The high alarm limits of respiration rate: R_RATE ALARM_H:5-60t/m, off
4).The low alarm limits of respiration rate: R_RATE ALARM_L: off, 4-40t/m
5).The high alarm limits of inCO2(optional): INCO2 ALARM_H: 4-30mmHg, off
6).Pump flux setup: FLOW-SET: 50cc/min-250cc/min
7).The setup of apnea time: APNEA TIME: 15s-44s, off
8).The unit of CO2:CO2 UNIT:%, mmHg or kPA
9).Pump switch: ON or Off
10).Pump auto-closing time: AUTO-OFF-TIME: 10-30min
11)Screen speed of capnograph:SWEEP SPEED: 5.5mm/s, 7.5mm/s or 10mm/s
12)CO2 Wave scale: WAVE SCALE: 54mmHG or 76mmHG
13) EtCO2 average computation time: EtCO2 Averageing: every breath,10sec,20sec,30sec
11).Default reload: LOAD-DEFAULTS
12). Exit: EXIT
Attention:
Pump auto-closing time means that the pump will automatically be closed down when in that
long time no respiration occurs.
The wave range means the maximum value of waveform amplitude display but it does not
mean data on full-scale. Data on full-scale still means 99mmHG.
Default values as follow:
12
14. Manual
EtCO2 alarm high limit: 50mmHg
EtCO2 alarm low limit: 19mmHg
Respiration Rate alarm high limit: 30 times/min
Respiration Rate alarm low limit: 08 times/min
INCO2 alarm high limit: OFF
FLOW_SET: 100 CC/Min
Apnea time:
30S
CO2 unit: %
CO2_PUMP: ON
AUTO_OFF_TIME: 10 Min
SWEEP SPEED: 7.5mm/s
EtCO2 Averaging: 1 Breath
WAVE SCALE: 54mmHG
IV. SpO2 setting menu
FIGURE 8
In this menu, to press ▲ or ▼button to move the cursor up or down, to press ► button or◄
button to change the data highlighted by the cursor.
Regarding some items in this menu, if not to change the data but just to operate them such as
LOAD DEFAULTS or EXIT, then to press OK button.
In this menu, to press MENU button, then to exit this menu and enter the main menu.
This menu includes the following setups:
1). The high alarm limits of SpO2: SPO2 ALARM_H: 50%-100%, off
2). The low alarm limits of SpO2: SPO2 ALARM_L: off, 50%-99%
3). The high alarm limits of pulse rate: P_RATE ALARM_H: 70-250t/m, off
4). The low alarm limits of pulse rate: P_RATE ALARM_L: off, 40-100t/m
5). Wave selection: WAVE: FILL or LINE
6). Renewing of defaults. LOAD DEFAULTS
The wave selection means that: FILL indicates the wave of plethysmogram is solid. LINE
indicates the wave of plethysmogram is linear.
Default values as follow:
Saturation of SPO2 alarm high limit: 100%
Saturation of SPO2 alarm low limit: 92%
Pulse alarm high limit: 130 t/min.
Pulse alarm low limit: 50 t/min.
Wave: Line
13
15. Manual
V. The time setting menu
FIGURE 9
In this menu, to press ▲ or ▼button to move the cursor up or down, to press ► button or◄
button to change the data highlighted by the cursor.
Attention: Any time adjustment will renew the trend, so time adjustment shall be careful.
The procedure is as follow:
1) Changing time;
2) To move the cursor to SAVE & EXIT, then to press the OK button to enter the following
menu FIGURE 9.1;
3) To move the cursor to YES, then to press OK button. Only in this way can the time
adjustment work.
FIGURE 9.1
VI. Sound setting menu
FIGURE 10
In this menu, to press ▲ or ▼button to move the cursor up or down, to press ► button or◄
button to change the data highlighted by the cursor.
This menu includes following setups:
Pulse sound volume: BEEP_VOLUME: 0(OFF)-8
Alarm sound volume: ALARM_VOLUME: 0(OFF)-8
If the alarm volume is 0, the top line in the main menu will show '!'
14
16. Manual
VII. Trend
FIGURE 11
This monitor stores one group data every minute and keeps them even after power off. This
monitor can circularly accumulate trend data for 24 hours. For example, if the monitor is run for
4 hours, then it is powered off, this monitor can store data for its 6 running, but the 7 th running of
this monitor will cover the first stored data.
As FIGURE 11 showing, the time for trend‟s every page is for 2 hours and every point indicates
the result of every minute. The top line of this page indicates the turning on time about its data
of this page (date/month/year hour: minute), for example: TURN ON: 10/03/08 22:15. The
second line indicates the turning off time about its data of this page ( hour、minute) for example,
TURN OFF: 22:15. The time at beginning and ending parts of abscissa in this picture
respectively indicates the beginning and ending time for trend of this page. The ending part of
vertical coordinate indicates the maximum value and unit for its result of parameter.
Parameter of the lowest line indicates parameter type.
Sometime the data is not complete, it shows the monitor is turned off although it has not
completed 2 hours‟ record.
In this menu, to press ▲or ▼button to change the trends of CO2 concentration, respiration
rate, SpO2 and pulse (the latter 2 parameters are selectable).
In this menu, to press ►or ◄ button to change the page of trend. Attention: if the time of the
page exceeds some turning on or off time, the time of turning on or off displayed on the top 2
lines will change.
In this menu, to press MENU button to quit this menu and return to the main menu.
15
17. Manual
Chapter VI Charge、Maintenance、sterilization、cleanness
I. Charge
The battery of this unit adopts reusable lithium battery group. After AC power supply is connected
and unit is turned on, this unit will charge the battery with power at the same time. The battery
charge will end after battery is full.
The replaced battery shall be made for environmental treatment.
II. Maintenance
If the monitor appears abnormal (e.g. system halted), to press the switch the monitor at least more
than 5 seconds, then it can be turned off compulsively.
Water trap:
If the screen shows ‘occlusion' and the pump's noise increases, we need to check if the
water trap and sampling pipe are blocked, otherwise the life of unit will be influenced.
If it is contaminated the water trap should be replaced (If the sampling line and the loop are
already known smooth).
It also should be replaced if its longest usage period is already overrun.
Please do not let alcohol, cleaning reagent or sterilizing reagent into water trap.
To check the storage bottle of water trap if it is empty and clear before it is used and to empty
and clean it after it has been used in order to use next time.
It is advised to use different water trap so as to avoid cross infection
Attention:
The pipe connector or water trap should be not sterilized and used repeatedly if the packing
indication shows that it is disposable.
Attention: For the environment, disposable pipe connector or water trap shall be treated
suitably or recycled.
III. Sterilization、cleanness
Warning:Before to clean the monitor and probe, power shall be turned off and switched off form
AC current.
1. The cleanness and sterilization of SpO2 probe.
Care:
Not to sterilize the probe by the high pressure.
Not to dip the probe into liquid.
Not to use the probe if they have the sign of brokenness or degeneration.
Cleanness and sterilization:
To clean it with the soft cloth moistened with medical alcohol, then to wipe with dry cloth.
2. Cleanness and sterilization of water trap
Attention: To use clear and sterilized water trap.
Cleaning: Only the Fluid Container may be cleaned and/or disinfected
Sterilization:
To use 70% of alcohol or isopropyl alcohol, glutaraldehyde and Chlorhexidine and ethanol
etc. to sterilize.
3. The used water traps and sampling tubes shall be disposed according to national regulations.
16
18. Manual
Chapter VII Analysis of problems
Simple analysis of problems
No.
Phenomena
Causes
Solution
1
The values of CO2 is lower
1.Leaking of water trap or sampling
1. To check and replace water
tube
trap and collector.
2. Occlusion of water trap or
2. To clean the gas loop.
sampling tube, the pump's noise
3. To calibrate by standard gas.
increases
3. Parts overused or excursed.
2
The values of CO2 is zero
1.Heavy leaking inside the gas loop
1. Screen indicating PUMP ERR and
2.The lamp resource of sensor
big noise.
damaged
2. Screen indicating IR-LAMP-BAD
To contact the manufacturer.
3.Sensor broken
3. Screen indicating CO2 SENSOR
ERR
3
Screen indicating CAL-ERR
The last calibration is failed.
To recalibrate by the standard
gas.
4
Screen indicating POWER-ERR
Wrong power supply.
To contact manufacturer.
5
The CO2 wave is not normal.
1. Temperature too high.
1.2. To provide normal
1.
Screen indicating TEMP-HIGH
2. Temperature too low.
environmental temperature.
2.
Screen indicating TEMP-LOW
3.Temperature sharp change
3. Because of environmental
3.
Screen indicating
change or warming up of unit,
TEMP-IMBALANCE
after temperature turning
stable, the data will be
accurate.
1. The clock‟s problem.
Holdup of time
1. To reset time menu and input
2.The chip of clock lack of power
6
the current time.
2.To replace the battery of
clock‟s chip
7
No values of SPO2 or no wave
1.Finger too cold
1.Warm up finger
2.Interference of strong light
2. Avoiding strong light.
3. The measurement test of SpO2
3 .Avoiding such phenomenon.
and blood pressure are done on the
4.Renew sensor of SPO2
same arm.
5.Cleaning
4. Red light in the sensor no flashing.
5. Infrared and collector of sensor is
not clean
8
Flashing „LOW BATTERY‟ and closed
1. Battery lack of power.
1. To connect AC power.
AC power working abnormally.
1. To check AC electric current
down automatically.
9
Still flashing „LOW BATTERY‟ after
the
power
is
supplied
and
AC
and cable.
indicator no light.
2. To check fuse.
Attention: Please to contact the client‟ service center if some problems occurred repeatedly.
17
19. Manual
Appendix 1 Explications for the appreciated terms in the device
MENU
Menu
EtCO2
The co2 concentration of expiration end phase
INCO2
The co2 concentration of inspiration phase
SPO2
Oximeter saturation
RR or R-RATE
Respiration rate
PR or P-RATE
Pulse rate
mmHg
Millimeter hydrargyrum
KPA
Kilopascal
ALARM-H
Alarm high limit
ALARM-L
Alarm low limit
LINE
Linear
FILL
Filled or solid
BEEP_VOLUME
Pulse volume
ALARM_VOLUME
Alarm volume
LOW BATTERY Lack of power
APNEA
Apnea or breath stop
T/Min
Times/ minute
CAL
Offset Calibration
SET
Setup
N2O:
Nitrous oxide
HELIUM
Helium gas
O2 CONCENTRATION
O2 concentration compensation
ANESTHEIC GAS
Anesthetic gas
ZERO GAS
Base point or Zero point
BTPS
Temperature and deep lung air pressure compensation
CALIBRATE
Calibration
ADJUST
Adjustment
CANCEL:
cancellation
OCCLUSION
Block
BLOW-BY
leakage
18
20. Manual
Appendix 2 How to change compensation of balance gas
Attention!Only the trained person can do the following the procedures.
First to enter the engineer menu, the method as follows:
To press ► and
following menu.
two buttons simultaneously when the power is turned on, then to enter the
FIGURE 12
In this menu, to press ▲ or ▼button to move the cursor up or down, to press ► button or◄ button
to change the data highlighted by the cursor.
The some items of this menu can be directly operated, such as LOAD-DEFAULT or EXIT: to press
OK button, then to be executed without through changing data. In this menu, to press MENU button,
then being able to exit this menu and to enter the main menu.
This menu including the following setups
Balance gas choice BALANCE GAS: AIR, N2O, and HELIUM
Compensation of O2 concentration O2 CONCENTRATION: 20%-99%
Compensation of anesthetic gas concentration ANESTHEIC GAS: 0-20%
Zero gas choice
ZERO GAS:AIR,N2
Compensation of temperature and lung air pressure BTPS: ENABLE, DISABLE
Default recovery: LOAD DEFAULTS
CO2 concentration calibration CALIBRATE
Default values as follows:
BALANCE GAS: AIR
O2 CONCENTRATION: 20 %
ANESTHEIC GAS: 0 %
ZERO GAS:AIR
BTPS: DISABLE
19
21. Manual
Appendix 3 How to calibrate precision of the EtCO2 concentration
Attention: Only the trained person can do the following the procedures.
1. Attentive items and preparative work
All the devices have been calibrated before they are out of manufacturer and during every turning
on they will do excursion control and gain adjustment according to working environmental
temperature, pressure and balance gas etc. so generally the user does not need to calibrate
himself or herself. If the unit has been used over one year and clinic readings‟ deviation is big, the
following procedure shall be obeyed.
1.1 Necessary instruments:
CO2 standard gas (Concentration is about 5%, or 5.0-7.0%)
Three- pass connector : inner diameter 1-3 mm, connecting with gas bottle、the water trap
and open air , is used for connection and protection. This is because the device can not bear
1.1 air pressures and it is strictly forbidden to connect the gas bottle directly with the device.
The one end of three-pass connector directly opens to the air so as to release the more gas
and pressure into the air even the standard bottle flows out larger gas.
The two tubes whose length can extend outside room: The standard gas flows into the air
continuously through the three-pass connector and the module pump vents also standard gas,
in the meantime CO2 gas of higher concentration accumulates easily and quickly around the
device. If the two parts of these standard gases can not be vented outside room, the device
will pick up the value with higher concentration as Zero, thus this will influence calibration
accuracy.
1.2 To connect as the FIGURE 13 shows:
FIGURE 13
2. Warm-up: to turn on power and run the unit for 20-30 minutes and adjust the pump flux over
120cc/min. To check if there is a leakage , the method is as following:To squeeze the sampling
tube by hands, the accelerating turning sound of sampling pump increases obviously. If the
sampling pump does not accelerate and its turning sound also does not change under hands‟
squeezing, then there must be a leakage in loop somewhere. At this time, it is not good for
demarcation. It‟s better to find out leakage point and solve it, otherwise, it will leads to short
value measured. After warming-up, to open standard gas, to listen if the sound of pump is as
same as original one. If the pump‟s turning is slow and its turning sound is weak, that means the
20
22. Manual
standard gas export is too large, at this time its export flux needs to turn down until the sound of
sampling pump resumes its original volume.
3. To turn off device, then to enter the engineer menu again (procedure is same as Appendix 2,
FIGURE 12), to move highlight onto CALIBRATE, then to press OK button to enter the next
menu.
FIGURE 14
To move highlight onto STANDARD GAS, to adjust the value as same as the concentration of
CO2 standard gas, if your standard gas concentration is precise enough to behind decimal two
numbers, you can adopt a round method.
To move the highlight onto CAL-BEGIN and to press OK button, in the same time, to open the
standard gas as the demand of FIGURE 13 and attention items, the device begins to calibrate.
Meanwhile, in the middle of FIGURE 14-1 appears „ADJUSTING!‟
FIGURE 14-1
The cross thick bar in this menu will be erased as time passes and the calibration will be ended
when it is completely erased. If the calibration is successful, the menu will show ADJUST OK
and subsequently exit into the main menu. If the calibration is unsuccessful, this menu will show
ADJUST ERR, at this time, the loop needs to be checked if there is a leakage or running out of
standard gas (the pressure indicator
of gas bottle shows 0). This menu still stays here if the
calibration is unsuccessful.
If hoping to exit this menu during the calibration, just to press MENU button or to move the
highlight onto CANCEL, then to press OK button.
Do not forget to close down the switch of the standard gas bottle and depressurized valve lest
the standard gas should be wasted.
21