4. Engineer Training
Part 1 Instrument Principle
Part 2 Specification of RAC-050
Part 3 Instrument Installing
Part 4 Maintenance
Part 5 Common Troubleshooting
7. After Open the Light-Shield Cover
Normally, when the instrument is working, the cover should be closed in
advance, or else, the test can not go further. For Engineer, you can place the
Magnetic on the cover sensor to disable it. Or you can disconnect the sensor
cable and connect the Pin 1 and Pin3.
8. Instrument Requirement
1. Instrument has Inbuilt Main board inside. It doesn’t
need to have PC working with.
2. Online UPS is Necessary.
3.RAC-Cleanser and cuvettes should be original from
Rayto, or else, it will have the problem that the Probe
cannot wash totally and bring Carry-over contamination,
and the cuvette is not mixed properly, or even the optical
system cannot work properly.
9. RAC-Cleanser
RAC-Cleanser (30ml)
Expired Date: 12 Month
Expired Date after open: 30 Day
Storage Temp: 2-35 Celcius
Consumption for Long Rinse and
Test preparation is 200ul
Consumption in testing is Reagent
volume+20ul
10. Part 1 Instrument Principle
Principle of light-path
Electronic circuit principle
Liquid Path
16. Clotting/Coagulation Method
Scatter Light intersity refects due to the Turbidity Strength. And it’s Linear with
the Testing Material Concentration.
Wavelength: 660nm
No Filter inside, Only
Red Interference Plastic
17. PT Actual Reaction Curve
X axis: Time(s)*10, 200 means 20s
Y axis: AD, Max Value is 65535
Coagulation Time (S)
18. PT ratio= PT/Normal
INR= (PT ratio)^(ISI)
DFbg need to calibration with sample
PT% need to do calibration with blood pool
Note: For the Software V2.4e, the DFbg Unit is same with FIB.
19. Immunologic Method
Light resource is LED and there is 575nm filter in the
Optical System. It will produce yellow light
go through cuvette.
The Turdimitry is more intense, the less light
will transit through the Cuvette.
LED
Photo-
Detector
20. D-Dimer Reaction Curve
X axis: Time(s)*10, 200 means 20s
Y axis: AD, Max Value is 65535
Reaction Rate
21. Chromogenic Method
Light resource is LED and there is 405nm filter in the
Optical System.
It’s Purple color in the Reaction cuvette. The
Turdimitry is more intense, the less light
will transit through the Cuvette.
LED
Photo-
Detector
30. Item Testing Workflow
• Temperature
Stabilized
Add Sample
• Incubate
Add R1
• Mix and Test
Or Incubate
Add R2
Note: If the Sample is need to predilution, the probe will add the
Diluent firstly and then add the Sample, After that, Dispense into
the cuvette and then add R1 and R2.
If sample dilution is not possible to complete in one time, it will
use one Cuvette to diluent the sample. After that, the probe will
add diluent and then add the diluted sample and dispense to the
new cuvette and further add R1 and R2
Test Cycle Time is
30 second
31. Structure of Hardware
Main Board with
CF card Main Control Board
Motor Driving Board
Temperature
Control Board
Touch
Screen
Driving
Board
Display Driver
Board
Printer Driving
Board
Touch
Screen
LCD Screen
Printer
Head
Input &output System of the instrument
4 Way Pre-Amplify
Board
3 Way Pre-Amplify
Board
Five Pcs
Boards for
Light
Resource
Optical System
Big Driving board
For Catcher
System
Sampling Driving
board
Liquid Detection
Board
Temp-Control System
Mechanic
System
Small Driving
board For
Catcher System
32. Main board
1. Battery is used to save the BIOS settings.
2. Memory Bank can use the Erase to clean the Pins.
3. CF card is installed on the same side of Memory Bank.
44. Part 2
Instrument Specification
1. 3 Testing Method
2. 3. Reagent Mixing around Position A1
3. LIS Data Transfer
4. 8 inch Screen with Touch Screen
5. Reaction Volume: 150-300ul, Vortex Mixing
6. Laser Printer can installed on the instrument, Built-in
Thermal printer
7. Optional Function: Sample Barcode Reader, Reagent
Card Reader (RFID).
45. Part 3 Instrument Installation
1. Installation Environment
2. Installation Process
46. Installation Environment
Well-ventilated; Free dust as possible around it;
Prevent from direct exposure to sunlight;
Prevent it placing around heat source or air one;
Free from corrosive or flammable gas;
UPS is necessary;
Do not be close to brush generator or frequently-switching-on electric
equipment;
Do not be close to electric equipment sending electromagnetic wave;
Ambient temperature: 15-30℃, operation temperature fluctuation: less than
+/-2℃/H;
Ambient humidity: 35%-80%RH, non-condensed;
This needs to install air conditioner if indoor temperature fails to meet
requirement.
47. Power supply Requirement:
Power supply: 220VAC, 50/60Hz, 1000VA, three-core
power supply wire, good grounding;
Needs independently have grounding wire;
Neutral-grounding voltage: less than 5V;
AC power supply must be stable, prohibit sharing in
power supply with high-power electric appliance, stable
power supply should be set at most;
48. Unpack
1. Wear glove
2. Remove the lock from
the box bottom.
52. Switch On and Software Login
1. For User, The password is 888888; For Engineer, the password is h-201
2. Click the Company Logo, and you can see the versions of the software and Main
control board.
53. Reset all the positions and Then Long Rinse
Note: Long Rinse is necessary for new installation. It make the liquid go
into the instrument.
56. Daily Maintenance
1. Empty the Waste Bottles
2. After daily test, need to turn off the
instrument and put the reagent back to the
fridge.
57. Weekly Maintenance
1. Clean the probe surface with alcohol
2. Clean Reagent area
3. Clean the instrument outside surface
with soft cloth, includes the sample Rack
4. Clean the water bottle
58. Half Year Maintenance
1. Grease the Probe arm and Catch Arm
2. Clean the water bottle filter
3. Clean the dust on the Fans
59. Yearly Maintenance
1. Check whether the fan have abnormal
sound. If yes, please replace it.
2. Check the optical path. Normally it’s ok
to check every 2 years. LEDs are work
stable.
3. Check the tube is dirty or not. If yes,
please replace.
60. Part 5 Adjustment
1. Probe Position Adjustment
2. Catcher Position Adjustment
3. Temperature adjustment
4. Optical adjustment
61. Engineering Adjust Interface
Note: The Maintenance Button is enabled when Login Engineer.
The Maintenance interface is mainly used to check the Positions, Check
Movement is correct or not, Parts status, etc.
62. Probe Adjustment
Debug Area
Position Check
Area
For Dispend and Sample/Reagent Pos., the probe needs to be in the center,
for Vertical position, the probe should touch bottom.
65. Air Pos.
Please Reset probe and catcher arm firstly, then you can choose Air
position for Catcher and click Move XY, then click “Move Z”. In the
next, please choose Air for sampling probe and click Move XY, Click
Move Z and then check whether the probe almost touch the cuvette
bottom.
66.
67. Optical Path Adjustment
Original Value:
Low Sensitivity default value: Offset 139, Gain 128.
Middle Sensitivity default value: Offset 139, Gain 128.
High Sensitivity default value: Offset 128, Gain 128.
68. Tools for Optical check
1. Turbidity 100 Better to store in Dark Environment.
2. Turbidity 400 Better to store in Dark Environment
3. Black Cuvette
69. 400 Turbidity Tool:
CH1-CH5, Low Sensitivit’s AD reference range: 20000±2000.
Ch6-CH7, Middle Sensitivit’s AD reference range: 30000±3000.
Ch6-CH7, High Sensitivit’s AD reference range: 55000±10000.
CH1-CH5, Middle Sensitivit’s AD reference range: 22000±2000.
CH1-CH5, High Sensitivit’s AD reference range: 55000±5000.
Cuvette with Distilled water
Ch6-CH7, Low Sensitivit’s AD reference range: 42000±2000.
Use 400 Turbidity Tool and Change Gain value:
100 Turbidity Tool:
Use 100 Turbidity Tool and Change Gain value:
Black Cuvette:
Change Offset value:
AD value: around 400-750
70. Pressure adjustment
Displayed Pressure: Get into “Temperature” menu and click Open Sensor button. Click
Open liquid pump button. Pay attention to the Liquid Pressure value (Unit is mmHg). We need the
difference between software and air pressure meter less than 10 mmHg. If the value is big, please
click Close liquid pump button and adjust the Gain value. Then test again till it is in the reference
range.
Actual Pressure Adjustment:
Click Open Liquid Pump button and connect air pressure meter to WASH connector
which color is blue. Adjust VR501 to make the value of air pressure meter to
280±20mmHg. After adjust PR5, please disconnect the tube of air pressure meter. After
the air releases, please connect air pressure meter and adjust again. Click Close Liquid
Pump after finishing adjustment.
71. Temperature adjustment
Click “open sensor” and see the displayed temperature. Normally the probe temperature
is around 43 celcius. Reagent Temp. is less than 16 celcius.
73. Blue Screen
How to check whether the system is working ok:
Click “ctrl”+“Alt”+“Del” to ensure whether
the task manager can appear
Possible Reason:
1. System is lost
2. Operation software is corrupted
Note: When the system is working ok, you can see the mouse curse on the
screen after connecting USB mouse.
After turning on the instrument, if you can see the Counting down 3, 2,1, please
click R button several times to recover the system.
74. Black Screen
Check the voltage on the LCD interface board
Note: You can disconnect the cable of other
boards, i.e, front board, High voltage board to
locate the problem
75. Mechanic Parameter Lost
Solution: Copy back the original parameter
file in The software folder.
1. The Mechanic Parameter is stored in
Sysparam.dat file. When installing the new
instrument, it’s required to backup the
Sysparam.dat for future use.
Note: The USB disk need to kill Virus in
advance.
76. Common Errors
(001) error alarm for optical coupling reset
for sampling arm X-axis
1
(1) Failure for optical coupling or cluster
(2) Incorrect or in place for optical coupling piece
(3) Cluster for drive board
(4) Aged or broken flat wire
(002) error alarm for optical coupling reset
for sampling arm Y-axis
1
(003)error alarm for optical coupling reset
for sampling arm Z-axis
1
(004) error place for the reagent area for
sampling needle liquid level sensor
2
(1) liquid level test board
(2) Sampling needle
(3) Cluster
(005 or 006) sampling needle fails to test
error existing in liquid level! The place is
at **
5
(1) Lack of reagent or sample
(2) Failure for liquid level test board, replace such
board.
(3) Aged cluster for sampling needle
(4) Broke flat line at X-axis or Y-axis for sampling arm,
replace small or big flat wire
(007) error for impact in sampling needle!
The place is at **
3
(1) J1 port for liquid level test board has two pins
conducted
(2) Failure for micro-switch, replace it
(3) Aged cluster for micro-switch
(4) Broken flat wire at X-axis or Y-axis for
sampling arm
(5) Unreasonable debugging for mechanical arm
place data
(6) Sampling arm fails to move specified position
(7) Painting lubrication oil onto XYZ-axis for
sampling arm
77. Common Errors
(010)Error for download place data
for sampling arm
1 Error for serial port
(017) error alarm for optical coupling
reset for catcher arm X-axis
1
(1) Failure for optical coupling or cluster
(2) incorrect or in place for optical coupling piece
(3) aged or broken flat wire
(4) Damaged motor drive board
(018) Error alarm for optical coupling
reset for catcher arm Y-axis
1
(1) Failure for optical coupling or cluster
(2) Incorrect or in place for optical coupling piece
(3) aged or broken flat wire
(4) Damaged motor drive board
(019) Error alarm for optical coupling
reset for catcher arm Z-axis
1
(1) Failure for optical coupling or cluster
(2) Incorrect or in place for optical coupling piece
(3) Aged or broken flat wire
(4) Damaged motor drive board
(020) Error for catcher arm cup
failing to detect cuvette! The place is
at **
4
(1) No cuvette exists at make-up place
(2) Some smear exists at reflecting sensor, which affects detection
or damage
(3) Aged or damaged flat wire at X-axis or Y-axis of catcher arm
(4) catcher arm peels out
(5) Failure for drive board
(6) Not vertical reassembly for grab
78. Common Errors
(024) Error for download place data
for catcher arm
1
(028)Lack of reaction cup 4
(102) Error for communication with
sampling arm
1
(103) Error for communication with
catcher arm
1
(104) Error for communication with
temperature control
1
(105) Abnormal temperature at
reagent area, temperature is **℃
6
(1) Waiting for running machine for less than 20 minutes
(2) Failure for Pletier, replace Pletier
(3) Air fan stops running, resulting in failure for heat radiation.
Replace air fan for cold storage
(4) Failure for temperature sensor
(5) Failure for temperature-control board
79. How to Claim Problem to Rayto?
1. Fill the Problem or spare parts Claim
Form with the problem details, including
SN, Photos, Video, software folder, what is
done (if necessary).
2. Send the claim form to
Support@rayto.com which is available to
all Service Engineers in Rayto.
80. Collect The problem Record
1. Each Engineer need to fill in the
problem record after complete the
troubleshooting. For the Maintenance Trip,
there is no need to mentioned.
2. Please send back to
support@rayto.com every 3 Month.