Santamedical SM-220 Finger Pulse Oximeter

TECHNICAL SPECIFICATIONS: Display-->Type : LED, Parameters : SpO2, PR, Pleth bar., Information : low power ::SPO2-->Measure Range : 35-99%, Resolution : ±1%, Accuracy : 75%-99%:±2, 50%-75%:±3, 0%-50%:unspecified::Pulse Rate-->Range : 30-254 BPM, Resolution : ±1bpm, Accuracy : ±2bpm::Battery--> Type : 2 AAA Alkaline batteries or rechargeable, batteries Power Consumption : less than 30Ma, Voltage : 2.6-3.6V::Environmental-->Operating Temperature : 0'C-50'C, Storage Temperature : -10'C - 60'C, Operating Humidity : 15%RH - 95%RH, Storage Humidity : 10%RH - 95%RH, It can be used by a variety of sports enthusiasts, including mountain climbers, hikers, skiers, bikers, and others interested in measuring oxygen saturation and pulse rate, it is not for medical use. This oximeter is for anyone who is interested in knowing their oxygen saturation and heart rate. Applications would be for use during exercise or daily activities, during air travel for the passenger or pilot, and for extreme sports such as mountain climbing. This oximeter is not a medical device and does not require a prescription from your medical professional.

1 of 2
Download to read offline
OPERATOR’S MANUAL
VER1.0C615
Fingertip Pulse Oximeter
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded
with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it
is consistency of Oxyhemoglobin in blood. It is a very important parameter for the
Respiratory Circulation System. Many respiratory diseases can result in oxygen
saturation being lowered in human blood. Additionally, the following factors can
reduce oxygen saturation: Automatic regulation of organ dysfunction caused by
Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical
examinations. That situation might result in light-headedness, asthenia, and vomiting.
Therefore, it is very important to know the oxygen saturation of a patient so that
doctors can find problems in a timely manner.
The fingertip pulse Oximeter features small size, low power consumption, convenient
operation and portability. It is only necessary for a patient to put one of his fingers into
the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen
saturation. It has been proven in clinical experiments that it also features high
precision and repeatability.
Measurement Principle
Principle of the Oximeter is as follows: A mathematical formula is established making
use of Lambert Beer Law according to Spectrum Absorption Characteristics of
Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared
zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection
Technology is adopted in accordance with Capacity Pulse Scanning and Recording
Technology, so that two beams of different wavelength of lights (660nm glow and
940nm near infrared light) can be focused onto a human nail tip through a clamping
finger-type sensor. A measured signal obtained by a photosensitive element, will be
shown on the Oximeter’s display through process in electronic circuits and
microprocessor shown on the Oximeter’s display through electronic circuits and a
microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Product Operation Scope
Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of
arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home, and
hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care
and etc). The device is not intended for continuously monitoring.
The PULE OXIMETER requires no routine calibration or maintenance other than
replacement of batteries.
Technical Specifications
1. Display: LED
2. SpO2:
Measurement range:70~99%
Accuracy: 70%~99%, ±3%; <70% no definition
3. Pulse Rate:
Measure range:30~235 bpm
Accuracy:30~100bpm ±2bpm; 101~235bpm, ±2%
Pulse Intensity: Bar graph Indicator
4. Power Requirements:
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Low power indication:
Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously
operated as long as 30 hours.
5. Dimension:
Length: 58mm
Width: 32mm
Height: 37mm
Weight: 33g (without batteries)
6. Environment Requirements:
Operation Temperature: 5~40℃
Storage Temperature: -20~55℃
Ambient Humidity: ≤80%, no condensation in operation;
≤93%, no condensation in storage
7.Measurement Performance in Low Perfusion Condition: The pulse wave is
available without failure when the simulation pulse wave amplitude is at 0.6% using
the test equipment (BIO-TEK INDEX Pulse Oximeter Tester).
8.Interference Resistance Capacity against Ambient Light: Device works
normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester.
Precautions for use
1 Do not use the pulse oximeter in an MRI or CT environment
2 Do not use the pulse oximeter in situations where alarms are required. The
device has no alarms.
3 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.
4 The pulse oximeter is intended only as an adjunct in patient assessment. It
must be used in conjunction with other methods of assessing clinical signs and
symptoms.
5 Check the pulse oximeter sensor application site frequently to determine the
positioning of the sensor and circulation and skin sensitivity of the patient.
6 Do not stretch the adhesive tape while applying the pulse oximeter sensor. This
may cause inaccurate readings or skin blisters.
7 Before use, carefully read the manual.
8 The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as
indicated by the symbol.
9 Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and
correct alignment at least every 4 hours.
10 Inaccurate measurements may be caused by autoclaving, ethylene oxide
sterilizing, or immersing the sensors in liquid may cause inaccurate readings.
11 Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin
or methemoglobin).
12 Intravascular dyes such as indocyanine green or methylene blue
13 SpO2 measurements may be adversely affected in the presence of high
ambient light. Shield the sensor area (with a surgical towel, or direct sunlight,
for example) if necessary.
14 Excessive patient movement may cause accurate readings.
15 Venous pulsations may cause accurate readings.
16 Placement of a sensor on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line may cause accurate readings.
17 The patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia.
18 The patient is in cardiac arrest or in shock.
19 Fingernail polish or false fingernails may cause inaccurate SpO2 readings.
Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.
Product Properties
1. Easy to use.
2. Small in volume, light in weight.
3. Lower power consumption.
4. Low voltage warning will be indicated in visual window when battery voltage is low.
5. The product will automatically be powered off when no signal is in the product for
longer than 8 seconds.
Operation Instructions
1. Install two AAA batteries into battery cassette, and then close its cover.
2. Open the clamp as shown in the picture below.
3. Insert one finger into the flexible fingertip of the Oximeter fully.
4. Press the power switch once on the front panel.
5. The patient’s finger should remain still while the Oximeter is working.
6. Read corresponding data from display screen.
When you insert a finger into the Oximeter, your nail surface must be upward.
NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of
Oximeter, and clean the test sensor using alcohol before and after operation. (The
rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the
skin of human being).
Product Accessories
1. One strap
2. Two batteries
3. One user’s manual
Strap Installation
1. Thread thinner end of the strap through the loop.
2. Thread thicker end of the strap through the threaded end before pulling it tightly.
Brief Description of Front Panel
The Pulse Bar graph displays the strength of the pulse rate signal. The height of the
bar graph shows the patient’s pulse strength.
Battery Installation
1. Put the two AAA batteries into battery cassette in correct polarities.
2. Press the button down along the arrow and push the battery cover shown as
below:
Notes:
Battery polarities must be correctly installed to avoid
damage to the device.
Please insert or remove batteries in right order, or
may cause damage to the device bracket.
Please remove batteries if the Oximeter will not be
used for a long time.
Maintenance and Storage
1. Replace batteries in time when low voltage lamp is lighted.
2. Clean surface of the fingertip Oximeter before it is used in diagnosis for patients.
3. Remove batteries inside the battery cassette if the Oximeter will not be operated
for a long time.
4. It is best to preserve the product in a place where ambient temperature –20~55
℃ and humidity is <93%, no condensation.
5. It is recommended that the product should be kept in a dry environment anytime.
A wet ambient might affect its lifetime and even might damage the product.
6. Please follow the law of the local government to deal with used batteries.
Calibrating the Oximeter
1. The functional tester cannot be used to assess the accuracy of the oximeter.
2. The test method used to establish the SpO2 accuracy is clinical testing. The
oximeter used to measure the arterial haemoglobin oxygen saturation levels and
these levels are to be compared to the levels determined from arterial blood
sampling with a CO-oximeter.
3. Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve
to 2, and then users can use this particular calibration curve to measure the
oximeter.
Declaration
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity and no action on
tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Guidance and manufacture’s declaration – electromagnetic
emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Pulse Oximeter should assure
that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Pulse Oximeter uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B The Pulse Oximeter is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Possible Problems and resolutions
Problems Possible reason Solution
SpO2 or PR can
not be shown
normally
1.Finger is not plugged correctly
2.Patient’s Oxyhemoglobin value is too
low to be measured.
1. Retry by plugging the finger.
2. Try some more times, If you can make
sure about no problem existing in the
product. Please go to a hospital for exact
diagnosis.
SpO2 or PR is
shown unstably
1.Finger might not be plugged deeply
enough.
2.Finger is trembling or patient’s body
is in movement status.
1.Retry by plugging the finger.
2.Try not to move.
The Oximeter
can not be
powered on
1.Power of batteries might be
inadequate or not be there at all.
2.Batteries might be installed
incorrectly.
3.The Oximeter might be damaged.
1.Please replace batteries
2.Please reinstall the batteries
3.Please contact with local customer
service centre.
Indication lamps
are suddenly off
1.The product is automatically
powered off when no signal is
detected longer than 8 seconds
2. Lower power
1.Normal
2.Replace the batteries
“Error3”or
“Error4” is
displayed on
screen.
1 Low power
2 Receiving tube being shielded or
damaged together with broken
connector.
3.Mechanical malfunction for
receive-emission tube
4 Amp circuit malfunction.
1 Change new batteries.
2 Please contact with local customer
service center.
3 Please contact with local customer
service center.
4 Please contact with local customer
service center.
Symbol Definitions
Symbol Definition
Type BF applied part
Attention, consult accompanying documents.
SpO2﹪ Oxygen saturation
Pulse rate (BPM)
Low power indication
NOT for continuous monitoring
SN Serial No.
Applicable models
SM-210 SM-220
Note: The illustrations used in this manual may differ slightly from the appearance of the
actual product.
Manufactured for:IRoams Network
Add:713 W. Duarte Rd #G810
Arcadia, CA 91007 USA
ALL RIGHTS RESERVED
Issue date: 2011-2-18

Recommended

Santamedical SM-230 OLED Finger Pulse Oximeter
Santamedical SM-230 OLED Finger Pulse OximeterSantamedical SM-230 OLED Finger Pulse Oximeter
Santamedical SM-230 OLED Finger Pulse OximeterGurin Products, LLC
 
Wearable Bio Sensor PPT
Wearable Bio Sensor PPT Wearable Bio Sensor PPT
Wearable Bio Sensor PPT PRITAM SHARMA
 
Innovative gadgets in anesthesia and medicine
Innovative gadgets in anesthesia and medicine Innovative gadgets in anesthesia and medicine
Innovative gadgets in anesthesia and medicine ikramdr01
 
Iaetsd mobile icu using android
Iaetsd mobile icu using androidIaetsd mobile icu using android
Iaetsd mobile icu using androidIaetsd Iaetsd
 
PM-180 Tens Unit Electronic Pulse Massager, Backlit LCD and Rechargeable Battery
PM-180 Tens Unit Electronic Pulse Massager, Backlit LCD and Rechargeable BatteryPM-180 Tens Unit Electronic Pulse Massager, Backlit LCD and Rechargeable Battery
PM-180 Tens Unit Electronic Pulse Massager, Backlit LCD and Rechargeable BatteryDevendra Tiwari 7838896746
 
Bio instrumentation
Bio instrumentationBio instrumentation
Bio instrumentationMalik Zaeem
 

More Related Content

What's hot

Saline level indicator
Saline level indicatorSaline level indicator
Saline level indicatorAravind Gurram
 
Bio instrumentation
Bio instrumentationBio instrumentation
Bio instrumentationMalik Zaeem
 
Wearable Biosensors Presentation
Wearable Biosensors PresentationWearable Biosensors Presentation
Wearable Biosensors PresentationShubham Rokade
 
Final Project Slides Buchheit Harings Hussein
Final Project Slides Buchheit Harings HusseinFinal Project Slides Buchheit Harings Hussein
Final Project Slides Buchheit Harings HusseinScott Buchheit
 
Sensors and systems by shivam gupta
Sensors and systems   by shivam guptaSensors and systems   by shivam gupta
Sensors and systems by shivam guptaShivam Gupta
 
Newer Gadgets in Anaesthesiology in 2020
Newer Gadgets in Anaesthesiology in 2020Newer Gadgets in Anaesthesiology in 2020
Newer Gadgets in Anaesthesiology in 2020dr tushar chokshi
 
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...IJECEIAES
 
Newer affordable gadgets in anesthesia and medicine
Newer affordable gadgets in anesthesia and medicineNewer affordable gadgets in anesthesia and medicine
Newer affordable gadgets in anesthesia and medicinedr tushar chokshi
 
Wearable Bio-sensors Ppt
Wearable Bio-sensors PptWearable Bio-sensors Ppt
Wearable Bio-sensors PptWinay Choudhary
 
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable batteryPm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable batteryDevendra Tiwari 7838896746
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET Journal
 

What's hot (17)

Saline level indicator
Saline level indicatorSaline level indicator
Saline level indicator
 
use of pulse oximetry
use of pulse oximetryuse of pulse oximetry
use of pulse oximetry
 
Report
ReportReport
Report
 
Bio instrumentation
Bio instrumentationBio instrumentation
Bio instrumentation
 
Wearable Biosensors Presentation
Wearable Biosensors PresentationWearable Biosensors Presentation
Wearable Biosensors Presentation
 
Development of wearable_bio
Development of wearable_bioDevelopment of wearable_bio
Development of wearable_bio
 
Hcl hh20
Hcl hh20Hcl hh20
Hcl hh20
 
Final Project Slides Buchheit Harings Hussein
Final Project Slides Buchheit Harings HusseinFinal Project Slides Buchheit Harings Hussein
Final Project Slides Buchheit Harings Hussein
 
Sensors and systems by shivam gupta
Sensors and systems   by shivam guptaSensors and systems   by shivam gupta
Sensors and systems by shivam gupta
 
Newer Gadgets in Anaesthesiology in 2020
Newer Gadgets in Anaesthesiology in 2020Newer Gadgets in Anaesthesiology in 2020
Newer Gadgets in Anaesthesiology in 2020
 
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
 
Wearable biosensors
Wearable biosensorsWearable biosensors
Wearable biosensors
 
Newer affordable gadgets in anesthesia and medicine
Newer affordable gadgets in anesthesia and medicineNewer affordable gadgets in anesthesia and medicine
Newer affordable gadgets in anesthesia and medicine
 
Wearable Bio-sensors Ppt
Wearable Bio-sensors PptWearable Bio-sensors Ppt
Wearable Bio-sensors Ppt
 
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable batteryPm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level Indicator
 
Bpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitorBpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitor
 

Viewers also liked

Regulation of Respiration
Regulation of RespirationRegulation of Respiration
Regulation of RespirationShahd Al Ali
 
Regulation of respiration
Regulation of respirationRegulation of respiration
Regulation of respirationRASHA
 
Chapter 22: Respiratory System (#4)
Chapter 22: Respiratory System (#4)Chapter 22: Respiratory System (#4)
Chapter 22: Respiratory System (#4)Caroline Tokarski
 
Chemical control of respiration
Chemical control of respirationChemical control of respiration
Chemical control of respirationAnwar Siddiqui
 
Transport of oxygen and carbon dioxide in blood (1)
Transport of oxygen and carbon dioxide in blood (1)Transport of oxygen and carbon dioxide in blood (1)
Transport of oxygen and carbon dioxide in blood (1)Lubna Abu Alrub,DDS
 
Unit 3 respiratory system and gas exchange
Unit 3 respiratory system and gas exchangeUnit 3 respiratory system and gas exchange
Unit 3 respiratory system and gas exchangemaswazi10
 
REGULATION OF RESPIRATION
REGULATION OF RESPIRATIONREGULATION OF RESPIRATION
REGULATION OF RESPIRATIONDr Nilesh Kate
 
Adventitious breath sounds
Adventitious breath soundsAdventitious breath sounds
Adventitious breath soundsjcp1962
 
Pulmonary auscultation
Pulmonary auscultationPulmonary auscultation
Pulmonary auscultationVASS Yukon
 

Viewers also liked (12)

Marieb ch13b
Marieb ch13bMarieb ch13b
Marieb ch13b
 
The Respiratory System
The Respiratory SystemThe Respiratory System
The Respiratory System
 
Regulation of Respiration
Regulation of RespirationRegulation of Respiration
Regulation of Respiration
 
Regulation of respiration
Regulation of respirationRegulation of respiration
Regulation of respiration
 
Chapter 22: Respiratory System (#4)
Chapter 22: Respiratory System (#4)Chapter 22: Respiratory System (#4)
Chapter 22: Respiratory System (#4)
 
Chemical control of respiration
Chemical control of respirationChemical control of respiration
Chemical control of respiration
 
Transport of oxygen and carbon dioxide in blood (1)
Transport of oxygen and carbon dioxide in blood (1)Transport of oxygen and carbon dioxide in blood (1)
Transport of oxygen and carbon dioxide in blood (1)
 
Unit 3 respiratory system and gas exchange
Unit 3 respiratory system and gas exchangeUnit 3 respiratory system and gas exchange
Unit 3 respiratory system and gas exchange
 
REGULATION OF RESPIRATION
REGULATION OF RESPIRATIONREGULATION OF RESPIRATION
REGULATION OF RESPIRATION
 
Adventitious breath sounds
Adventitious breath soundsAdventitious breath sounds
Adventitious breath sounds
 
Pulmonary auscultation
Pulmonary auscultationPulmonary auscultation
Pulmonary auscultation
 
Oxygen hemoglobin dissociation curve
Oxygen hemoglobin dissociation curveOxygen hemoglobin dissociation curve
Oxygen hemoglobin dissociation curve
 

Similar to Santamedical SM-220 Finger Pulse Oximeter

Santamedical SM-240 OLED Finger Pulse Oximeter
Santamedical SM-240 OLED Finger Pulse OximeterSantamedical SM-240 OLED Finger Pulse Oximeter
Santamedical SM-240 OLED Finger Pulse OximeterGurin Products, LLC
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET Journal
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...IRJET Journal
 
IRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET Journal
 
Implementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applicationsImplementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applicationsSree Bhargava Kothapalli
 
FINGUER TIP OXIMETER
FINGUER TIP OXIMETERFINGUER TIP OXIMETER
FINGUER TIP OXIMETERACN
 
Abstract1
Abstract1Abstract1
Abstract1susinu
 
IRJET - A Novel Health Monitoring System for Trance Patients
IRJET -  	  A Novel Health Monitoring System for Trance PatientsIRJET -  	  A Novel Health Monitoring System for Trance Patients
IRJET - A Novel Health Monitoring System for Trance PatientsIRJET Journal
 
Jay 5 bw
Jay 5 bwJay 5 bw
Jay 5 bwJeroxxx
 
Pulse oximeters- Application and Maintenance
Pulse oximeters- Application and MaintenancePulse oximeters- Application and Maintenance
Pulse oximeters- Application and Maintenanceshashi sinha
 
IRJET- The Multiphysiological Parameter Monitoring System
IRJET-  	  The Multiphysiological Parameter Monitoring SystemIRJET-  	  The Multiphysiological Parameter Monitoring System
IRJET- The Multiphysiological Parameter Monitoring SystemIRJET Journal
 
Rigel vital-signs-booklet-uk
Rigel vital-signs-booklet-ukRigel vital-signs-booklet-uk
Rigel vital-signs-booklet-ukCao Xuân Trình
 

Similar to Santamedical SM-220 Finger Pulse Oximeter (20)

Santamedical SM-240 OLED Finger Pulse Oximeter
Santamedical SM-240 OLED Finger Pulse OximeterSantamedical SM-240 OLED Finger Pulse Oximeter
Santamedical SM-240 OLED Finger Pulse Oximeter
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
 
Oximeter
OximeterOximeter
Oximeter
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
 
Md90 user manual
Md90 user manualMd90 user manual
Md90 user manual
 
IRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring System
 
IJET-V3I1P26
IJET-V3I1P26IJET-V3I1P26
IJET-V3I1P26
 
Implementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applicationsImplementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applications
 
66.ppt
66.ppt66.ppt
66.ppt
 
66.pptx
66.pptx66.pptx
66.pptx
 
Ppt0000002
Ppt0000002Ppt0000002
Ppt0000002
 
FINGUER TIP OXIMETER
FINGUER TIP OXIMETERFINGUER TIP OXIMETER
FINGUER TIP OXIMETER
 
Abstract1
Abstract1Abstract1
Abstract1
 
Medical dissection lab pulse oximeter
Medical dissection lab pulse oximeterMedical dissection lab pulse oximeter
Medical dissection lab pulse oximeter
 
IRJET - A Novel Health Monitoring System for Trance Patients
IRJET -  	  A Novel Health Monitoring System for Trance PatientsIRJET -  	  A Novel Health Monitoring System for Trance Patients
IRJET - A Novel Health Monitoring System for Trance Patients
 
Jay 5 bw
Jay 5 bwJay 5 bw
Jay 5 bw
 
Pulse oximeters- Application and Maintenance
Pulse oximeters- Application and MaintenancePulse oximeters- Application and Maintenance
Pulse oximeters- Application and Maintenance
 
IRJET- The Multiphysiological Parameter Monitoring System
IRJET-  	  The Multiphysiological Parameter Monitoring SystemIRJET-  	  The Multiphysiological Parameter Monitoring System
IRJET- The Multiphysiological Parameter Monitoring System
 
Oximeter and CVP.pptx
Oximeter and CVP.pptxOximeter and CVP.pptx
Oximeter and CVP.pptx
 
Rigel vital-signs-booklet-uk
Rigel vital-signs-booklet-ukRigel vital-signs-booklet-uk
Rigel vital-signs-booklet-uk
 

More from Gurin Products, LLC

Scientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse OximeterScientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse OximeterGurin Products, LLC
 
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Gurin Products, LLC
 
Liztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth SpeakerLiztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth SpeakerGurin Products, LLC
 
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Gurin Products, LLC
 
Liztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter CardLiztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter CardGurin Products, LLC
 
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer Gurin Products, LLC
 
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...Gurin Products, LLC
 
Gurin Portable Ultrasonic Massager
Gurin Portable Ultrasonic MassagerGurin Portable Ultrasonic Massager
Gurin Portable Ultrasonic MassagerGurin Products, LLC
 
Bpm 165 w blood pressure monitor
Bpm 165 w blood pressure monitorBpm 165 w blood pressure monitor
Bpm 165 w blood pressure monitorGurin Products, LLC
 

More from Gurin Products, LLC (20)

HDD docking station
HDD docking stationHDD docking station
HDD docking station
 
Scientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse OximeterScientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse Oximeter
 
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
 
Liztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth SpeakerLiztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth Speaker
 
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
 
Liztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter CardLiztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter Card
 
Gurin steam inhaler
Gurin steam inhalerGurin steam inhaler
Gurin steam inhaler
 
Gurin Steam Inhaler
Gurin Steam InhalerGurin Steam Inhaler
Gurin Steam Inhaler
 
Sm 165 pulse oximeter
Sm 165 pulse oximeterSm 165 pulse oximeter
Sm 165 pulse oximeter
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
Ry230 thermometer
Ry230 thermometerRy230 thermometer
Ry230 thermometer
 
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
 
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
 
Gurin Portable Ultrasonic Massager
Gurin Portable Ultrasonic MassagerGurin Portable Ultrasonic Massager
Gurin Portable Ultrasonic Massager
 
Hair straightener
Hair straightenerHair straightener
Hair straightener
 
Ga 2800 d multi adapter card
Ga 2800 d multi adapter cardGa 2800 d multi adapter card
Ga 2800 d multi adapter card
 
Bpm 165 w blood pressure monitor
Bpm 165 w blood pressure monitorBpm 165 w blood pressure monitor
Bpm 165 w blood pressure monitor
 
Bpm 110 blood pressure monitor
Bpm 110 blood pressure monitorBpm 110 blood pressure monitor
Bpm 110 blood pressure monitor
 
Infrared Ear Thermometer Manual
Infrared Ear Thermometer ManualInfrared Ear Thermometer Manual
Infrared Ear Thermometer Manual
 
Gurin Water Flosser manual
Gurin Water Flosser manualGurin Water Flosser manual
Gurin Water Flosser manual
 

Recently uploaded

Introduction to anatomy and physioligy.pptx
Introduction to anatomy and physioligy.pptxIntroduction to anatomy and physioligy.pptx
Introduction to anatomy and physioligy.pptxMarbahunJalaKharbhih
 
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]Brian Peacock
 
(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj
(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj
(DENGUE).pptx Department of Physiotherapy, SHUATS, PrayagrajSurabhi Srivastava
 
Yantropakaranani 2: Heating Devices used in Ayurveda
Yantropakaranani 2: Heating Devices used in AyurvedaYantropakaranani 2: Heating Devices used in Ayurveda
Yantropakaranani 2: Heating Devices used in AyurvedaAparnaNandakumar12
 
Dasha vidha Pareeksha.pptx Dr Akshay Shetty
Dasha vidha Pareeksha.pptx Dr Akshay ShettyDasha vidha Pareeksha.pptx Dr Akshay Shetty
Dasha vidha Pareeksha.pptx Dr Akshay ShettyAkshay Shetty
 
Tuberculosis .pptx Department of Physiotherapy, SHUATS, Prayagraj
Tuberculosis .pptx Department of Physiotherapy, SHUATS, PrayagrajTuberculosis .pptx Department of Physiotherapy, SHUATS, Prayagraj
Tuberculosis .pptx Department of Physiotherapy, SHUATS, PrayagrajSurabhi Srivastava
 
Finger Clubbing by Dr. Samuel2 Alawode.pptx
Finger Clubbing by Dr. Samuel2 Alawode.pptxFinger Clubbing by Dr. Samuel2 Alawode.pptx
Finger Clubbing by Dr. Samuel2 Alawode.pptxAlawode Samuel2
 
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptx
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptxEUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptx
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptxDr. Sumit KUMAR
 
Bioavailability & Bioequivalence Studies- Definitions, Methods of Measureme...
Bioavailability & Bioequivalence  Studies- Definitions, Methods  of Measureme...Bioavailability & Bioequivalence  Studies- Definitions, Methods  of Measureme...
Bioavailability & Bioequivalence Studies- Definitions, Methods of Measureme...Rajshri
 
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICS
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICSPAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICS
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICSDr.Shrikrushna Sobaji
 
Staphylococcus.ppt.........ali.rasool.badr
Staphylococcus.ppt.........ali.rasool.badrStaphylococcus.ppt.........ali.rasool.badr
Staphylococcus.ppt.........ali.rasool.badrssuser06f49d
 
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...PeerVoice
 
Cell :Structure & Functions for Medical and Health allied Students
Cell :Structure & Functions for Medical and Health allied StudentsCell :Structure & Functions for Medical and Health allied Students
Cell :Structure & Functions for Medical and Health allied StudentsRajendra Dev Bhatt
 
Fetal and Neonatal Physiology - Dr Faiza.pdf
Fetal and Neonatal Physiology - Dr Faiza.pdfFetal and Neonatal Physiology - Dr Faiza.pdf
Fetal and Neonatal Physiology - Dr Faiza.pdfMedicoseAcademics
 
Hydrocele and tumors of testis Introduction.pptx
Hydrocele and tumors of testis Introduction.pptxHydrocele and tumors of testis Introduction.pptx
Hydrocele and tumors of testis Introduction.pptxSizan Thapa
 
Ways to Manage Ovarian Cancer Treatment Side Effects
Ways to Manage Ovarian Cancer Treatment Side EffectsWays to Manage Ovarian Cancer Treatment Side Effects
Ways to Manage Ovarian Cancer Treatment Side Effectsbkling
 
Malaria Disease.pptx Department of Physiotherapy, SHUATS, Prayagraj
Malaria Disease.pptx Department of Physiotherapy, SHUATS, PrayagrajMalaria Disease.pptx Department of Physiotherapy, SHUATS, Prayagraj
Malaria Disease.pptx Department of Physiotherapy, SHUATS, PrayagrajSurabhi Srivastava
 
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdf
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdfGuía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdf
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdfJuniorAlexanderCasti
 
Vertigo and Nystagmus - Clinical approach part-2.pptx
Vertigo and Nystagmus - Clinical approach part-2.pptxVertigo and Nystagmus - Clinical approach part-2.pptx
Vertigo and Nystagmus - Clinical approach part-2.pptxYasser Alzainy
 

Recently uploaded (20)

Introduction to anatomy and physioligy.pptx
Introduction to anatomy and physioligy.pptxIntroduction to anatomy and physioligy.pptx
Introduction to anatomy and physioligy.pptx
 
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]
Psychedelic Interventions for Neurological Conditions [NAN 2022 Poster]
 
(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj
(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj
(DENGUE).pptx Department of Physiotherapy, SHUATS, Prayagraj
 
Yantropakaranani 2: Heating Devices used in Ayurveda
Yantropakaranani 2: Heating Devices used in AyurvedaYantropakaranani 2: Heating Devices used in Ayurveda
Yantropakaranani 2: Heating Devices used in Ayurveda
 
Dasha vidha Pareeksha.pptx Dr Akshay Shetty
Dasha vidha Pareeksha.pptx Dr Akshay ShettyDasha vidha Pareeksha.pptx Dr Akshay Shetty
Dasha vidha Pareeksha.pptx Dr Akshay Shetty
 
Tuberculosis .pptx Department of Physiotherapy, SHUATS, Prayagraj
Tuberculosis .pptx Department of Physiotherapy, SHUATS, PrayagrajTuberculosis .pptx Department of Physiotherapy, SHUATS, Prayagraj
Tuberculosis .pptx Department of Physiotherapy, SHUATS, Prayagraj
 
Finger Clubbing by Dr. Samuel2 Alawode.pptx
Finger Clubbing by Dr. Samuel2 Alawode.pptxFinger Clubbing by Dr. Samuel2 Alawode.pptx
Finger Clubbing by Dr. Samuel2 Alawode.pptx
 
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptx
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptxEUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptx
EUTHANASIA: WORLD AND INDIAN PERSPECTIVEpptx
 
Bioavailability & Bioequivalence Studies- Definitions, Methods of Measureme...
Bioavailability & Bioequivalence  Studies- Definitions, Methods  of Measureme...Bioavailability & Bioequivalence  Studies- Definitions, Methods  of Measureme...
Bioavailability & Bioequivalence Studies- Definitions, Methods of Measureme...
 
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICS
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICSPAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICS
PAEDIATRICS RHEMATOLOGY PROTOCOLS SHEET FOR PEDIATRICS ORTHOPAEDICS
 
Staphylococcus.ppt.........ali.rasool.badr
Staphylococcus.ppt.........ali.rasool.badrStaphylococcus.ppt.........ali.rasool.badr
Staphylococcus.ppt.........ali.rasool.badr
 
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...
Disruption at the Neuromuscular Junction: Relevance of Autoantibodies to Ther...
 
Cell :Structure & Functions for Medical and Health allied Students
Cell :Structure & Functions for Medical and Health allied StudentsCell :Structure & Functions for Medical and Health allied Students
Cell :Structure & Functions for Medical and Health allied Students
 
Fetal and Neonatal Physiology - Dr Faiza.pdf
Fetal and Neonatal Physiology - Dr Faiza.pdfFetal and Neonatal Physiology - Dr Faiza.pdf
Fetal and Neonatal Physiology - Dr Faiza.pdf
 
Hydrocele and tumors of testis Introduction.pptx
Hydrocele and tumors of testis Introduction.pptxHydrocele and tumors of testis Introduction.pptx
Hydrocele and tumors of testis Introduction.pptx
 
Ways to Manage Ovarian Cancer Treatment Side Effects
Ways to Manage Ovarian Cancer Treatment Side EffectsWays to Manage Ovarian Cancer Treatment Side Effects
Ways to Manage Ovarian Cancer Treatment Side Effects
 
Malaria Disease.pptx Department of Physiotherapy, SHUATS, Prayagraj
Malaria Disease.pptx Department of Physiotherapy, SHUATS, PrayagrajMalaria Disease.pptx Department of Physiotherapy, SHUATS, Prayagraj
Malaria Disease.pptx Department of Physiotherapy, SHUATS, Prayagraj
 
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdf
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdfGuía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdf
Guía síndrome coronario agudo.pdf Guía síndrome coronario agudo.pdf
 
eye disorders .pptx
eye disorders .pptxeye disorders .pptx
eye disorders .pptx
 
Vertigo and Nystagmus - Clinical approach part-2.pptx
Vertigo and Nystagmus - Clinical approach part-2.pptxVertigo and Nystagmus - Clinical approach part-2.pptx
Vertigo and Nystagmus - Clinical approach part-2.pptx
 

Santamedical SM-220 Finger Pulse Oximeter

  • 1. OPERATOR’S MANUAL VER1.0C615 Fingertip Pulse Oximeter General Description Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems in a timely manner. The fingertip pulse Oximeter features small size, low power consumption, convenient operation and portability. It is only necessary for a patient to put one of his fingers into the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen saturation. It has been proven in clinical experiments that it also features high precision and repeatability. Measurement Principle Principle of the Oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the Oximeter’s display through process in electronic circuits and microprocessor shown on the Oximeter’s display through electronic circuits and a microprocessor. Diagram of Operation Principle 1. Red and Infrared-ray Emission Tube 2. Red and Infrared-ray Receipt Tube Product Operation Scope Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home, and hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care and etc). The device is not intended for continuously monitoring. The PULE OXIMETER requires no routine calibration or maintenance other than replacement of batteries. Technical Specifications 1. Display: LED 2. SpO2: Measurement range:70~99% Accuracy: 70%~99%, ±3%; <70% no definition 3. Pulse Rate: Measure range:30~235 bpm Accuracy:30~100bpm ±2bpm; 101~235bpm, ±2% Pulse Intensity: Bar graph Indicator 4. Power Requirements: Two AAA alkaline Batteries Power consumption: Less than 30mA Low power indication: Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 30 hours. 5. Dimension: Length: 58mm Width: 32mm Height: 37mm Weight: 33g (without batteries) 6. Environment Requirements: Operation Temperature: 5~40℃ Storage Temperature: -20~55℃ Ambient Humidity: ≤80%, no condensation in operation; ≤93%, no condensation in storage 7.Measurement Performance in Low Perfusion Condition: The pulse wave is available without failure when the simulation pulse wave amplitude is at 0.6% using the test equipment (BIO-TEK INDEX Pulse Oximeter Tester). 8.Interference Resistance Capacity against Ambient Light: Device works normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester. Precautions for use 1 Do not use the pulse oximeter in an MRI or CT environment 2 Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. 3 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 4 The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms. 5 Check the pulse oximeter sensor application site frequently to determine the positioning of the sensor and circulation and skin sensitivity of the patient. 6 Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may cause inaccurate readings or skin blisters. 7 Before use, carefully read the manual. 8 The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as indicated by the symbol. 9 Prolonged use or the patient’s condition may require changing the sensor site periodically. Change sensor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours. 10 Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may cause inaccurate readings. 11 Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin or methemoglobin). 12 Intravascular dyes such as indocyanine green or methylene blue 13 SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the sensor area (with a surgical towel, or direct sunlight, for example) if necessary. 14 Excessive patient movement may cause accurate readings. 15 Venous pulsations may cause accurate readings. 16 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line may cause accurate readings. 17 The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia. 18 The patient is in cardiac arrest or in shock. 19 Fingernail polish or false fingernails may cause inaccurate SpO2 readings. Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries. Product Properties 1. Easy to use. 2. Small in volume, light in weight. 3. Lower power consumption. 4. Low voltage warning will be indicated in visual window when battery voltage is low. 5. The product will automatically be powered off when no signal is in the product for longer than 8 seconds. Operation Instructions 1. Install two AAA batteries into battery cassette, and then close its cover. 2. Open the clamp as shown in the picture below. 3. Insert one finger into the flexible fingertip of the Oximeter fully. 4. Press the power switch once on the front panel. 5. The patient’s finger should remain still while the Oximeter is working. 6. Read corresponding data from display screen. When you insert a finger into the Oximeter, your nail surface must be upward. NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter, and clean the test sensor using alcohol before and after operation. (The rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the skin of human being). Product Accessories 1. One strap 2. Two batteries 3. One user’s manual Strap Installation 1. Thread thinner end of the strap through the loop. 2. Thread thicker end of the strap through the threaded end before pulling it tightly. Brief Description of Front Panel
  • 2. The Pulse Bar graph displays the strength of the pulse rate signal. The height of the bar graph shows the patient’s pulse strength. Battery Installation 1. Put the two AAA batteries into battery cassette in correct polarities. 2. Press the button down along the arrow and push the battery cover shown as below: Notes: Battery polarities must be correctly installed to avoid damage to the device. Please insert or remove batteries in right order, or may cause damage to the device bracket. Please remove batteries if the Oximeter will not be used for a long time. Maintenance and Storage 1. Replace batteries in time when low voltage lamp is lighted. 2. Clean surface of the fingertip Oximeter before it is used in diagnosis for patients. 3. Remove batteries inside the battery cassette if the Oximeter will not be operated for a long time. 4. It is best to preserve the product in a place where ambient temperature –20~55 ℃ and humidity is <93%, no condensation. 5. It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and even might damage the product. 6. Please follow the law of the local government to deal with used batteries. Calibrating the Oximeter 1. The functional tester cannot be used to assess the accuracy of the oximeter. 2. The test method used to establish the SpO2 accuracy is clinical testing. The oximeter used to measure the arterial haemoglobin oxygen saturation levels and these levels are to be compared to the levels determined from arterial blood sampling with a CO-oximeter. 3. Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve to 2, and then users can use this particular calibration curve to measure the oximeter. Declaration EMC of this product complies with IEC60601-1-2 standard. The materials which the user can come into contact have no toxicity and no action on tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10. Guidance and manufacture’s declaration – electromagnetic emissions-for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration – electromagnetic emission The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the Pulse Oximeter should assure that it is used in such and environment. Emission test Compliance Electromagnetic environment – guidance RF emissions CISPR 11 Group 1 The Pulse Oximeter uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emission CISPR 11 Class B The Pulse Oximeter is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Possible Problems and resolutions Problems Possible reason Solution SpO2 or PR can not be shown normally 1.Finger is not plugged correctly 2.Patient’s Oxyhemoglobin value is too low to be measured. 1. Retry by plugging the finger. 2. Try some more times, If you can make sure about no problem existing in the product. Please go to a hospital for exact diagnosis. SpO2 or PR is shown unstably 1.Finger might not be plugged deeply enough. 2.Finger is trembling or patient’s body is in movement status. 1.Retry by plugging the finger. 2.Try not to move. The Oximeter can not be powered on 1.Power of batteries might be inadequate or not be there at all. 2.Batteries might be installed incorrectly. 3.The Oximeter might be damaged. 1.Please replace batteries 2.Please reinstall the batteries 3.Please contact with local customer service centre. Indication lamps are suddenly off 1.The product is automatically powered off when no signal is detected longer than 8 seconds 2. Lower power 1.Normal 2.Replace the batteries “Error3”or “Error4” is displayed on screen. 1 Low power 2 Receiving tube being shielded or damaged together with broken connector. 3.Mechanical malfunction for receive-emission tube 4 Amp circuit malfunction. 1 Change new batteries. 2 Please contact with local customer service center. 3 Please contact with local customer service center. 4 Please contact with local customer service center. Symbol Definitions Symbol Definition Type BF applied part Attention, consult accompanying documents. SpO2﹪ Oxygen saturation Pulse rate (BPM) Low power indication NOT for continuous monitoring SN Serial No. Applicable models SM-210 SM-220 Note: The illustrations used in this manual may differ slightly from the appearance of the actual product. Manufactured for:IRoams Network Add:713 W. Duarte Rd #G810 Arcadia, CA 91007 USA ALL RIGHTS RESERVED Issue date: 2011-2-18