Standby Valve User's Guide
Version 2011A General Information Page 2
Standby Valve User's Guide
General Information
Isoprime Ltd
Isoprime House
Earl Road
Cheadle Hulme
Cheadle
SK8 6PT
Tel: +44 (0) 161 488 3660
Fax: +44 (0) 161 488 3699
http://www.Isoprime.co.uk/
Assistance
If you encounter difficulties with the operation of your instrument or need
software support, please contact the customer service department.
Customer service will raise a call log to track the progress of your enquiry, which
ensures that the problem will not be forgotten. The call will be passed to the
appropriate technical people for answering. Calls will be handled locally where
possible and escalated to the factory if a local response is not possible. All logged
calls must be dealt with to a successful conclusion before they can be closed.
Contact details are provided on the Isoprime website at www.isoprime.co.uk
Software
This manual is written for users of IonVantage V1.5. If another version of
software is used user operations may differ.
If you feel that more information is required or if you would like to suggest how
the manual could be improved, then please contact the customer service
department at Isoprime Ltd. or your local Isoprime representative.
Standby Valve User's Guide
Page 3 Standby Valve User's Guide Version 2011A
Standby Valve User's Guide
Standby Valve User's Guide ........................................1
General Information .....................................................2
Assistance................................................................................................ 2
Software ................................................................................................... 2
Standby Valve User's Guide ........................................3
Warranty................................................................................................... 4
Safety Information ................................................................................... 4
Safety Symbols........................................................................................ 4
Principle of Operation ............................................................................. 6
Standby Valve Setup ............................................................................... 7
Fitting Standby Valve ........................................................................ 7
Connecting Pump Line...................................................................... 7
Initial Operation ....................................................................................... 8
Leak Checking ........................................................................................10
Procedure.........................................................................................10
Maintenance............................................................................................12
Fused Silica Inlet..............................................................................12
Transfer and Pump Line...................................................................13
Version 2011A Standby Valve User's Guide Page 4
Standby Valve User's Guide
Warranty
The information contained within this document is subject to change without
notice.
Isoprime Ltd. makes no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of merchantability and fitness
for a particular purpose.
Isoprime Ltd. shall not be liable for errors contained herein or for incidental
damages in connection with the furnishing, performance, or use of this material.
Safety Information
The IsoPrime mass spectrometer meets the following IEC (International
Electromechanical Commission) classifications:
Safety Class 1
Transient Over Voltage Category II
Pollution Degree 2
You must comply with all local and national requirements for electrical and
mechanical safety.
The instrument has been designed and tested in accordance with recognized safety
standards. If the instrument is used in a manner not specified by the manufacturer,
the protection provided by the instrument may be impaired.
Whenever the safety protection of the IsoPrime is compromised, disconnect the
instrument from all power sources and secure the unit against unintended
operation.
Suitably qualified personnel only should perform maintenance procedures.
Substituting parts or performing any unauthorized modification to the instrument
may result in a safety hazard.
Disconnect the mains supply before removing covers.
Please contact the customer service department at the above address, or your local
Isoprime representative, should you require any further information.
Safety Symbols
Warnings in this manual or on the instrument must be observed during all phases
of service, repair, installation and operation of the instrument. Failure to comply
with these precautions violates the safety standards of design and the intended use
of the instrument.
Isoprime Ltd. assumes no liability for the customer’s failure to comply with these
requirements.
Standby Valve User's Guide
Page 5 Standby Valve User's Guide Version 2011A
The following safety symbols are used in the manual or on the instrument.
Warnings are given to highlight situations or conditions where failure to observe
the instruction could result in injury or death to persons. Cautions are given to
highlight situations or conditions where failure to observe the instruction could
result in damage to the equipment, associated equipment or process.
Version 2011A Standby Valve User's Guide Page 6
Standby Valve User's Guide
Principle of Operation
The Standby Valve provides a switching mechanism between sample analysis
mode and standby mode or used as an inlet switching device. When in sample
mode, Helium entering the Mass Spectrometer is drawn from the sample
preparation interface and the reference gas injector unit. In standby mode,
however, only Helium from the reference gas injector unit is drawn into the Mass
Spectrometer isolating the entire inlet preparation system. This ensures that only a
continuous stream of dry, clean Helium is supplied into the analyser at all times.
Thus, activities such as water trap back-flushing, furnace tube replacement or any
other intervention on the inlet system can be undertaken without interrupting the
flow of helium into the Mass Spectrometer. This therefore eliminates all possible
disruption to the Mass Spectrometer and ensures immediate stability when
returning to sample mode.
When used as an inlet switching device the Standby Valve Simply acts as a switch
to quickly change between different inlets.
Standby Valve User's Guide
Page 7 Standby Valve User's Guide Version 2011A
Standby Valve Setup
Fitting Standby Valve
The Standby Valve is mounted on top of the source isolation cover. To fit, simply
remove the two screws attaching the source cover to the source flange and place
the Standby Valve in position as shown below. Position the Standby Valve over
the source cover mounting and fix the screws back into position to secure the
Standby Valve.
Using the included ribbon cable, connect the Standby Valve in series with the
cable from J4 of the IsoPrime electronics unit.
Connecting Pump Line
The waste line of the Standby Valve is connected via a tee into the IsoPrime
roughing pump. A second isolation valve is connected to the rear of the
instrument. A short piece of 1/16 0.030" Stainless steel capillary is used to
connect the waste line to the roughing pump as shown below.
Version 2011A Standby Valve User's Guide Page 8
Standby Valve User's Guide
Initial Operation
Once the Standby Valve is fitted, due to the pumping arrangement, careful
operation of the valves and opening of the pump line must be undertaken on first
operation:
Connect the fused silica to the standby and sample sides of the Standby
Valve and connect to desired inlet systems. For example, when
connected to a GC system the standby side will have one length of
75um feeding into the Reference Gas Injector and the sample side will
have both a 100um length from the GC and another piece of 75um again
Standby Valve User's Guide
Page 9 Standby Valve User's Guide Version 2011A
feeding into the Reference Gas Injector allowing for both reference gas
and sample gas to be fed into the Mass Spectrometer.
No changes to the software are required to use the Standby Valve. Once
the valve is connected to the IsoPrime electronics unit via connection to
the ribbon cable from J4, restarting the IonVantage software will
reinitialise the inlet page and the software will auto detect the new
Standby Valve and include it on the inlet page mimic.
Initial control of the Standby Valve should be done using the individual
valve controls on the Custom Inlet page > Inlet Control tab.
Ensure all Standby Valves are closed (S1, S2, G1, G2). Monitor the
instrument high and low vacuum. Slowly open the small Standby Valve
isolation valve on the back of the Mass Spectrometer. As the valve
opens there will be a fast increase in the instrument low vacuum. This is
normal and is just the initial pump out of the Standby Valve pump line.
The instrument high vacuum should not be significantly affected.
If the instrument low vacuum does not return to below 5e-2mBar close
the Standby Valve isolation valve and proceed to check all fittings and
connections are tight on the Standby Valve assembly.
Once the instrument low vacuum has stabilised open S2 and S1. This
will pump out the Standby Valve body.
Again, there will be an increase in instrument low vacuum. Once this
returns to below 5e-2mBar the valve is ready to be opened to the
IsoPrime.
Turn off the source and open the instrument Nupro isolation valve.
Monitor the instrument high and low vacuum. Once this has returned to
an acceptable level, switch the Standby Valve to both sample and
standby side by double clicking on the pink valve as shown below.
Version 2011A Standby Valve User's Guide Page 10
Standby Valve User's Guide
After switching the valve, both sides of the Standby Valve will open to
waste for 10 seconds. The valve will then switch to the alternate side.
Once satisfied that the instrument high vacuum will not move above
9e-6mBar upon switching sides, the source can be switched on and the
valve can be leak checked. Note that the high vacuum should only rise
above 5e-6mBar for a short period of time during valve switching.
Leak Checking
All Standby Valves are leak checked prior to leaving the factory, however some
connections are likely to be remade upon installation and this can cause issues
with regards to leaks.
As a first course of action, it is good practice to always leak check any recently
modified fittings before checking the rest of the assembly.
Procedure
Argon leak checking is highly recommended.
Tune the IsoPrime to Mass 40 and peak centre on Argon.
Setup a 600 second time scan on the IsoPrime tune page.
Double click
to operate
Standby Valve User's Guide
Page 11 Standby Valve User's Guide Version 2011A
Turn the flow of Argon to a very low level and very carefully direct the
Argon over any suspected areas of leaks.
Any rise on the Mass 40 scan window will indicate a leak. Be aware
that there are many connections in a very small area so it is very easy to
identify a leak in the wrong place.
Note: High backgrounds will be observed when the fused silica is
removed/replaced or removed from the flow of Helium. This is expected and not
to be mistaken for a leak. Always leak check the valve when it is setup for
instrument operation.
Version 2011A Standby Valve User's Guide Page 12
Standby Valve User's Guide
Maintenance
Fused Silica Inlet
Warning: Do not remove any fused silica connections to the
Standby Valve until all 4 valves have been closed.
Before removing any connections to the Standby Valve it is important to isolate
all valves from the pumping system. As there is a common backing line for the
IsoPrime and the Standby Valve, the removal of the fused silica will massively
increase the load on the rotary pump and can affect the IsoPrime high vacuum.
Close the 4 Standby Valve connections. To operate valves individually go to
Custom Inlet page > Inlet Control tab:
To close each valve simply double click on each one until they are displayed as
shown above.
At this point it is safe to remove the fused silica connections.
Once the connections have been remade, open valves S2 and G2 for 30seconds.
All valves are
now closed
Standby Valve User's Guide
Page 13 Standby Valve User's Guide Version 2011A
This will pump out the Standby Valve. The Standby Valve is now ready for use
and should be operated from the Custom Inlet page as shown below:
Transfer and Pump Line
Warning: Do not remove any stainless steel tubing connections
from the Standby Valve without closing the Nupro valve and
Speedivalve.
Before removing the transfer or pump line of the Standby Valve is it imperative
that the Nupro isolation valve and the Speedivalve isolation valve are both closed.
Removal of these lines without complete isolation will vent the IsoPrime and
could damage the instrument turbo pump.
Once the valve is isolated remove and replace the tubing and ferrules. Follow the
initial operation section on reintegrating the Standby Valve into the system.

More Related Content

Similar to Standby Valve User Guide

Buddy High Volume Filling and Capping Machines
Buddy High Volume Filling and Capping Machines Buddy High Volume Filling and Capping Machines
Buddy High Volume Filling and Capping Machines RogerQi2
 
Ti 2000 manual de servicio
Ti 2000 manual de servicioTi 2000 manual de servicio
Ti 2000 manual de servicioEduard Vall
 
Liquid Nitrogen Pump Cooling System
Liquid Nitrogen Pump Cooling SystemLiquid Nitrogen Pump Cooling System
Liquid Nitrogen Pump Cooling SystemSima Umrao
 
Cr1.cr1 b instruction manual(bfp-a8054j)
Cr1.cr1 b instruction manual(bfp-a8054j)Cr1.cr1 b instruction manual(bfp-a8054j)
Cr1.cr1 b instruction manual(bfp-a8054j)Abe Paih
 
LK Series Water Treatment Controller
LK Series Water Treatment ControllerLK Series Water Treatment Controller
LK Series Water Treatment ControllerIwaki America Inc.
 
IRJET- Smart Production Line Industry 4.0 - Leak Testing for Fuel Tank
IRJET-  	  Smart Production Line Industry 4.0 - Leak Testing for Fuel TankIRJET-  	  Smart Production Line Industry 4.0 - Leak Testing for Fuel Tank
IRJET- Smart Production Line Industry 4.0 - Leak Testing for Fuel TankIRJET Journal
 
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-Safety
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-SafetyAsco Hazardous Area Solenoid Valves - Sil 3 Functional-Safety
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-SafetyThorne & Derrick UK
 
Bfp a8900x
Bfp a8900xBfp a8900x
Bfp a8900xTan Ngoc
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdffusjekmdjdud3e
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdff8suejjdkddmmd
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdff8usejdkddmmd
 

Similar to Standby Valve User Guide (20)

Buddy High Volume Filling and Capping Machines
Buddy High Volume Filling and Capping Machines Buddy High Volume Filling and Capping Machines
Buddy High Volume Filling and Capping Machines
 
Ti 2000 manual de servicio
Ti 2000 manual de servicioTi 2000 manual de servicio
Ti 2000 manual de servicio
 
Full Automatic Pcb loader Operation manual
Full Automatic Pcb loader Operation manualFull Automatic Pcb loader Operation manual
Full Automatic Pcb loader Operation manual
 
mn03010.pdf
mn03010.pdfmn03010.pdf
mn03010.pdf
 
EHE Series Electronic Metering Pump
EHE Series Electronic Metering PumpEHE Series Electronic Metering Pump
EHE Series Electronic Metering Pump
 
Liquid Nitrogen Pump Cooling System
Liquid Nitrogen Pump Cooling SystemLiquid Nitrogen Pump Cooling System
Liquid Nitrogen Pump Cooling System
 
Md90 user manual
Md90 user manualMd90 user manual
Md90 user manual
 
Mwhe manual
Mwhe manualMwhe manual
Mwhe manual
 
Cr1.cr1 b instruction manual(bfp-a8054j)
Cr1.cr1 b instruction manual(bfp-a8054j)Cr1.cr1 b instruction manual(bfp-a8054j)
Cr1.cr1 b instruction manual(bfp-a8054j)
 
LK Series Water Treatment Controller
LK Series Water Treatment ControllerLK Series Water Treatment Controller
LK Series Water Treatment Controller
 
Vg7 upac manual
Vg7 upac manualVg7 upac manual
Vg7 upac manual
 
IRJET- Smart Production Line Industry 4.0 - Leak Testing for Fuel Tank
IRJET-  	  Smart Production Line Industry 4.0 - Leak Testing for Fuel TankIRJET-  	  Smart Production Line Industry 4.0 - Leak Testing for Fuel Tank
IRJET- Smart Production Line Industry 4.0 - Leak Testing for Fuel Tank
 
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-Safety
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-SafetyAsco Hazardous Area Solenoid Valves - Sil 3 Functional-Safety
Asco Hazardous Area Solenoid Valves - Sil 3 Functional-Safety
 
Pcb automatic vaccum loader user guidance | SunzonTech
Pcb automatic vaccum loader user guidance | SunzonTechPcb automatic vaccum loader user guidance | SunzonTech
Pcb automatic vaccum loader user guidance | SunzonTech
 
Pcb buffer conveyor operation manual
Pcb buffer conveyor operation manualPcb buffer conveyor operation manual
Pcb buffer conveyor operation manual
 
Bfp a8900x
Bfp a8900xBfp a8900x
Bfp a8900x
 
Fr a024 a044
Fr a024 a044Fr a024 a044
Fr a024 a044
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf
 
2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf2009 Nissan Versa Service Repair Manual.pdf
2009 Nissan Versa Service Repair Manual.pdf
 

Recently uploaded

Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.riyadelhic riyadelhic
 
Radhika Call Girls In Jaipur 9358660226 Escorts service
Radhika Call Girls In Jaipur 9358660226 Escorts serviceRadhika Call Girls In Jaipur 9358660226 Escorts service
Radhika Call Girls In Jaipur 9358660226 Escorts servicerahul222jai
 
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...aakahthapa70
 
SURAT CALL GIRL 92628/71154 SURAT CALL G
SURAT CALL GIRL 92628/71154 SURAT CALL GSURAT CALL GIRL 92628/71154 SURAT CALL G
SURAT CALL GIRL 92628/71154 SURAT CALL GNiteshKumar82226
 
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...aakahthapa70
 
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.riyadelhic riyadelhic
 
DIGHA CALL GIRL 92628/1154 DIGHA CALL GI
DIGHA CALL GIRL 92628/1154 DIGHA CALL GIDIGHA CALL GIRL 92628/1154 DIGHA CALL GI
DIGHA CALL GIRL 92628/1154 DIGHA CALL GINiteshKumar82226
 
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Availablenitugupta1209
 
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 60009891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000teencall080
 
Call US Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...
Call US  Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...Call US  Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...
Call US Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...Pooja Nehwal
 
KAKINADA CALL GIRL 92628/71154 KAKINADA C
KAKINADA CALL GIRL 92628/71154 KAKINADA CKAKINADA CALL GIRL 92628/71154 KAKINADA C
KAKINADA CALL GIRL 92628/71154 KAKINADA CNiteshKumar82226
 
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579Best VIP Call Girls Noida Sector 23 Call Me: 8700611579
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579diyaspanoida
 
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...aakahthapa70
 
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...aakahthapa70
 
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579Best VIP Call Girl Noida Sector 48 Call Me: 8700611579
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579diyaspanoida
 
Russian Call Girls in Goa %(9316020077)# Russian Call Girls in Goa By Russi...
Russian Call Girls  in Goa %(9316020077)# Russian Call Girls  in Goa By Russi...Russian Call Girls  in Goa %(9316020077)# Russian Call Girls  in Goa By Russi...
Russian Call Girls in Goa %(9316020077)# Russian Call Girls in Goa By Russi...Goa Call Girls Service Goa escort agency
 
MYSORE CALL GIRLS ESCORT SER 92628/71154
MYSORE CALL GIRLS ESCORT SER 92628/71154MYSORE CALL GIRLS ESCORT SER 92628/71154
MYSORE CALL GIRLS ESCORT SER 92628/71154NiteshKumar82226
 

Recently uploaded (20)

Call Girls In Saket Delhi 9953056974 (Low Price) Escort Service Saket Delhi
Call Girls In Saket Delhi 9953056974 (Low Price) Escort Service Saket DelhiCall Girls In Saket Delhi 9953056974 (Low Price) Escort Service Saket Delhi
Call Girls In Saket Delhi 9953056974 (Low Price) Escort Service Saket Delhi
 
Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 18 Escort Service Noida N.C.R.
 
Radhika Call Girls In Jaipur 9358660226 Escorts service
Radhika Call Girls In Jaipur 9358660226 Escorts serviceRadhika Call Girls In Jaipur 9358660226 Escorts service
Radhika Call Girls In Jaipur 9358660226 Escorts service
 
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...
Call Girls In {{Green Park Delhi}}9667938988 Indian Russian High Profile Esco...
 
SURAT CALL GIRL 92628/71154 SURAT CALL G
SURAT CALL GIRL 92628/71154 SURAT CALL GSURAT CALL GIRL 92628/71154 SURAT CALL G
SURAT CALL GIRL 92628/71154 SURAT CALL G
 
Call Girls In Goa For Fun 9316020077 By Goa Call Girls For Pick Up Night
Call Girls In  Goa  For Fun 9316020077 By  Goa  Call Girls For Pick Up NightCall Girls In  Goa  For Fun 9316020077 By  Goa  Call Girls For Pick Up Night
Call Girls In Goa For Fun 9316020077 By Goa Call Girls For Pick Up Night
 
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...
Call Girls In {Laxmi Nagar Delhi} 9667938988 Indian Russian High Profile Girl...
 
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.
Call Now ☎9870417354|| Call Girls in Noida Sector 12 Escort Service Noida N.C.R.
 
DIGHA CALL GIRL 92628/1154 DIGHA CALL GI
DIGHA CALL GIRL 92628/1154 DIGHA CALL GIDIGHA CALL GIRL 92628/1154 DIGHA CALL GI
DIGHA CALL GIRL 92628/1154 DIGHA CALL GI
 
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available
9811611494,Low Rate Call Girls In Connaught Place Delhi 24hrs Available
 
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 60009891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000
9891550660 Call Girls In Noida Sector 62 Short 1500 Night 6000
 
Call US Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...
Call US  Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...Call US  Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...
Call US Pooja📞 9892124323 ✅Call Girls In Mira Road ( Mumbai ) secure service...
 
KAKINADA CALL GIRL 92628/71154 KAKINADA C
KAKINADA CALL GIRL 92628/71154 KAKINADA CKAKINADA CALL GIRL 92628/71154 KAKINADA C
KAKINADA CALL GIRL 92628/71154 KAKINADA C
 
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579Best VIP Call Girls Noida Sector 23 Call Me: 8700611579
Best VIP Call Girls Noida Sector 23 Call Me: 8700611579
 
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...
Call Girls In {{Laxmi Nagar Delhi}} 9667938988 Indian Russian High Profile Es...
 
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...
Call Girls In {Connaught Place Delhi} 9667938988 IndianRussian High Profile E...
 
9953056974 Call Girls In Ashok Nagar, Escorts (Delhi) NCR.
9953056974 Call Girls In Ashok Nagar, Escorts (Delhi) NCR.9953056974 Call Girls In Ashok Nagar, Escorts (Delhi) NCR.
9953056974 Call Girls In Ashok Nagar, Escorts (Delhi) NCR.
 
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579Best VIP Call Girl Noida Sector 48 Call Me: 8700611579
Best VIP Call Girl Noida Sector 48 Call Me: 8700611579
 
Russian Call Girls in Goa %(9316020077)# Russian Call Girls in Goa By Russi...
Russian Call Girls  in Goa %(9316020077)# Russian Call Girls  in Goa By Russi...Russian Call Girls  in Goa %(9316020077)# Russian Call Girls  in Goa By Russi...
Russian Call Girls in Goa %(9316020077)# Russian Call Girls in Goa By Russi...
 
MYSORE CALL GIRLS ESCORT SER 92628/71154
MYSORE CALL GIRLS ESCORT SER 92628/71154MYSORE CALL GIRLS ESCORT SER 92628/71154
MYSORE CALL GIRLS ESCORT SER 92628/71154
 

Standby Valve User Guide

  • 2. Version 2011A General Information Page 2 Standby Valve User's Guide General Information Isoprime Ltd Isoprime House Earl Road Cheadle Hulme Cheadle SK8 6PT Tel: +44 (0) 161 488 3660 Fax: +44 (0) 161 488 3699 http://www.Isoprime.co.uk/ Assistance If you encounter difficulties with the operation of your instrument or need software support, please contact the customer service department. Customer service will raise a call log to track the progress of your enquiry, which ensures that the problem will not be forgotten. The call will be passed to the appropriate technical people for answering. Calls will be handled locally where possible and escalated to the factory if a local response is not possible. All logged calls must be dealt with to a successful conclusion before they can be closed. Contact details are provided on the Isoprime website at www.isoprime.co.uk Software This manual is written for users of IonVantage V1.5. If another version of software is used user operations may differ. If you feel that more information is required or if you would like to suggest how the manual could be improved, then please contact the customer service department at Isoprime Ltd. or your local Isoprime representative.
  • 3. Standby Valve User's Guide Page 3 Standby Valve User's Guide Version 2011A Standby Valve User's Guide Standby Valve User's Guide ........................................1 General Information .....................................................2 Assistance................................................................................................ 2 Software ................................................................................................... 2 Standby Valve User's Guide ........................................3 Warranty................................................................................................... 4 Safety Information ................................................................................... 4 Safety Symbols........................................................................................ 4 Principle of Operation ............................................................................. 6 Standby Valve Setup ............................................................................... 7 Fitting Standby Valve ........................................................................ 7 Connecting Pump Line...................................................................... 7 Initial Operation ....................................................................................... 8 Leak Checking ........................................................................................10 Procedure.........................................................................................10 Maintenance............................................................................................12 Fused Silica Inlet..............................................................................12 Transfer and Pump Line...................................................................13
  • 4. Version 2011A Standby Valve User's Guide Page 4 Standby Valve User's Guide Warranty The information contained within this document is subject to change without notice. Isoprime Ltd. makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Isoprime Ltd. shall not be liable for errors contained herein or for incidental damages in connection with the furnishing, performance, or use of this material. Safety Information The IsoPrime mass spectrometer meets the following IEC (International Electromechanical Commission) classifications: Safety Class 1 Transient Over Voltage Category II Pollution Degree 2 You must comply with all local and national requirements for electrical and mechanical safety. The instrument has been designed and tested in accordance with recognized safety standards. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the IsoPrime is compromised, disconnect the instrument from all power sources and secure the unit against unintended operation. Suitably qualified personnel only should perform maintenance procedures. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard. Disconnect the mains supply before removing covers. Please contact the customer service department at the above address, or your local Isoprime representative, should you require any further information. Safety Symbols Warnings in this manual or on the instrument must be observed during all phases of service, repair, installation and operation of the instrument. Failure to comply with these precautions violates the safety standards of design and the intended use of the instrument. Isoprime Ltd. assumes no liability for the customer’s failure to comply with these requirements.
  • 5. Standby Valve User's Guide Page 5 Standby Valve User's Guide Version 2011A The following safety symbols are used in the manual or on the instrument. Warnings are given to highlight situations or conditions where failure to observe the instruction could result in injury or death to persons. Cautions are given to highlight situations or conditions where failure to observe the instruction could result in damage to the equipment, associated equipment or process.
  • 6. Version 2011A Standby Valve User's Guide Page 6 Standby Valve User's Guide Principle of Operation The Standby Valve provides a switching mechanism between sample analysis mode and standby mode or used as an inlet switching device. When in sample mode, Helium entering the Mass Spectrometer is drawn from the sample preparation interface and the reference gas injector unit. In standby mode, however, only Helium from the reference gas injector unit is drawn into the Mass Spectrometer isolating the entire inlet preparation system. This ensures that only a continuous stream of dry, clean Helium is supplied into the analyser at all times. Thus, activities such as water trap back-flushing, furnace tube replacement or any other intervention on the inlet system can be undertaken without interrupting the flow of helium into the Mass Spectrometer. This therefore eliminates all possible disruption to the Mass Spectrometer and ensures immediate stability when returning to sample mode. When used as an inlet switching device the Standby Valve Simply acts as a switch to quickly change between different inlets.
  • 7. Standby Valve User's Guide Page 7 Standby Valve User's Guide Version 2011A Standby Valve Setup Fitting Standby Valve The Standby Valve is mounted on top of the source isolation cover. To fit, simply remove the two screws attaching the source cover to the source flange and place the Standby Valve in position as shown below. Position the Standby Valve over the source cover mounting and fix the screws back into position to secure the Standby Valve. Using the included ribbon cable, connect the Standby Valve in series with the cable from J4 of the IsoPrime electronics unit. Connecting Pump Line The waste line of the Standby Valve is connected via a tee into the IsoPrime roughing pump. A second isolation valve is connected to the rear of the instrument. A short piece of 1/16 0.030" Stainless steel capillary is used to connect the waste line to the roughing pump as shown below.
  • 8. Version 2011A Standby Valve User's Guide Page 8 Standby Valve User's Guide Initial Operation Once the Standby Valve is fitted, due to the pumping arrangement, careful operation of the valves and opening of the pump line must be undertaken on first operation: Connect the fused silica to the standby and sample sides of the Standby Valve and connect to desired inlet systems. For example, when connected to a GC system the standby side will have one length of 75um feeding into the Reference Gas Injector and the sample side will have both a 100um length from the GC and another piece of 75um again
  • 9. Standby Valve User's Guide Page 9 Standby Valve User's Guide Version 2011A feeding into the Reference Gas Injector allowing for both reference gas and sample gas to be fed into the Mass Spectrometer. No changes to the software are required to use the Standby Valve. Once the valve is connected to the IsoPrime electronics unit via connection to the ribbon cable from J4, restarting the IonVantage software will reinitialise the inlet page and the software will auto detect the new Standby Valve and include it on the inlet page mimic. Initial control of the Standby Valve should be done using the individual valve controls on the Custom Inlet page > Inlet Control tab. Ensure all Standby Valves are closed (S1, S2, G1, G2). Monitor the instrument high and low vacuum. Slowly open the small Standby Valve isolation valve on the back of the Mass Spectrometer. As the valve opens there will be a fast increase in the instrument low vacuum. This is normal and is just the initial pump out of the Standby Valve pump line. The instrument high vacuum should not be significantly affected. If the instrument low vacuum does not return to below 5e-2mBar close the Standby Valve isolation valve and proceed to check all fittings and connections are tight on the Standby Valve assembly. Once the instrument low vacuum has stabilised open S2 and S1. This will pump out the Standby Valve body. Again, there will be an increase in instrument low vacuum. Once this returns to below 5e-2mBar the valve is ready to be opened to the IsoPrime. Turn off the source and open the instrument Nupro isolation valve. Monitor the instrument high and low vacuum. Once this has returned to an acceptable level, switch the Standby Valve to both sample and standby side by double clicking on the pink valve as shown below.
  • 10. Version 2011A Standby Valve User's Guide Page 10 Standby Valve User's Guide After switching the valve, both sides of the Standby Valve will open to waste for 10 seconds. The valve will then switch to the alternate side. Once satisfied that the instrument high vacuum will not move above 9e-6mBar upon switching sides, the source can be switched on and the valve can be leak checked. Note that the high vacuum should only rise above 5e-6mBar for a short period of time during valve switching. Leak Checking All Standby Valves are leak checked prior to leaving the factory, however some connections are likely to be remade upon installation and this can cause issues with regards to leaks. As a first course of action, it is good practice to always leak check any recently modified fittings before checking the rest of the assembly. Procedure Argon leak checking is highly recommended. Tune the IsoPrime to Mass 40 and peak centre on Argon. Setup a 600 second time scan on the IsoPrime tune page. Double click to operate
  • 11. Standby Valve User's Guide Page 11 Standby Valve User's Guide Version 2011A Turn the flow of Argon to a very low level and very carefully direct the Argon over any suspected areas of leaks. Any rise on the Mass 40 scan window will indicate a leak. Be aware that there are many connections in a very small area so it is very easy to identify a leak in the wrong place. Note: High backgrounds will be observed when the fused silica is removed/replaced or removed from the flow of Helium. This is expected and not to be mistaken for a leak. Always leak check the valve when it is setup for instrument operation.
  • 12. Version 2011A Standby Valve User's Guide Page 12 Standby Valve User's Guide Maintenance Fused Silica Inlet Warning: Do not remove any fused silica connections to the Standby Valve until all 4 valves have been closed. Before removing any connections to the Standby Valve it is important to isolate all valves from the pumping system. As there is a common backing line for the IsoPrime and the Standby Valve, the removal of the fused silica will massively increase the load on the rotary pump and can affect the IsoPrime high vacuum. Close the 4 Standby Valve connections. To operate valves individually go to Custom Inlet page > Inlet Control tab: To close each valve simply double click on each one until they are displayed as shown above. At this point it is safe to remove the fused silica connections. Once the connections have been remade, open valves S2 and G2 for 30seconds. All valves are now closed
  • 13. Standby Valve User's Guide Page 13 Standby Valve User's Guide Version 2011A This will pump out the Standby Valve. The Standby Valve is now ready for use and should be operated from the Custom Inlet page as shown below: Transfer and Pump Line Warning: Do not remove any stainless steel tubing connections from the Standby Valve without closing the Nupro valve and Speedivalve. Before removing the transfer or pump line of the Standby Valve is it imperative that the Nupro isolation valve and the Speedivalve isolation valve are both closed. Removal of these lines without complete isolation will vent the IsoPrime and could damage the instrument turbo pump. Once the valve is isolated remove and replace the tubing and ferrules. Follow the initial operation section on reintegrating the Standby Valve into the system.