Heart–Lung Machine
Extracorporeal Circulation
ECC or perfusion is defined as a procedure
carried out with the aid of HLM
In the early stages, cardiac surgery was
mainly concerned with the treatment of
congenital heart defects. Today, the main
focus is on acquired heart defects.
Access path to the operation area is kept
as small as possible to reduce the patient’s
surgical trauma this surgical technique
usually has a direct influence on the
configuration of the HLM and on the way
the perfusion is carried out
OPCAB (off-pump coronary artery bypass)
refer to cardiac surgeries carried out
without the support of an HLM
The HLM can never be looked at in
isolation additional components, mainly in
the form of sterile disposables such as an
oxygenator, heat exchanger, or reservoir,
take on other important functions during
ECC
Introduction
The Heart-lung machine is a device which is used to
provide blood circulation and oxygenation while
the heart is stopped during the open heart surgery.
Heart-lung machine consists of three main functional
units, the pump, the oxygenator and heat exchanger.
To function, the heart-lung machine must be connected to
the patient in a way that allows blood to be removed,
processed, and returned to the body.
History
The researchers searched mainly for a means to provide
oxygenated blood to organ preparations. They filtered the
blood through various screens and membranes and even
pumped it through the lungs of dogs or monkeys, but their
problem was not to be solved, though this may be
considered the beginning of research into a heart-lung
device.
The first successful open heart procedure on a human
utilizing the heart lung machine was performed by John
Gibbon in 1953. He repaired an atrial septal defect in an
18-year-old woman.
Purpose of Heart-lung machine:-
The heart-lung machine provides the benefit of a
motionless heart during open heart surgery.
The heart-lung machine is valuable during this time since
the patient is unable to maintain blood flow to the lungs or
the body.
Surgical procedures in which
Heart-lung machine is used:-
Coronary artery bypasses surgery.
Cardiac valve repair and/or replacement (aortic
valve, mitral valve, tricuspid valve, pulmonic valve)
Repair of large septum defects (atrial septum defect,
ventricular septum defect, atrioventricular septum
defect)
Transplantation (heart transplantation, lung
transplantation, heart–lung transplantation)
Implantation of heart.
Block diagram:-
Components of Heart-lung
Machine:-
Cardioplegia
Cannula
Blood reservoir
Heparin Pump
Roller pumps/Centrifugal pump
Oxygenators
Heat Exchanger
Structure and Function of the Heart–Lung Machine
The HLM is the basis of ECC and cover
two important organ functions :-
1. Pump function of the heart
2. Gas exchange function of the lungs
It must ensure :-
 A sufficient perfusion volume that
corresponds to the normal cardiac
output of the patient under
anesthetic
 An adequate perfusion pressure (50–
90 mmHg)
 It must also ensure sufficient
oxygenation, elimination of CO2, and
control of the blood temperature
The following components make up the
basic equipment of the ECC that is used
during modern cardiac surgery:
1. Blood pumps
2. Oxygenator
3. Tubing system with various tubing
diameters
4. Blood filters with various functions
5. Cardiotomy reservoir
6. Cannulae and intracardiac suction
tubes
Blood Pumps and Their Function
Blood pumps can generally be
categorized into two groups, depending
on their function:
1. Roller pumps
2. Centrifugal pumps
 Essentially, both pump types should
meet certain criteria when used with
the HLM:
1. Dosed delivery of liquids, with
precise display of the actual flow rate
and delivered volume
2. External control mechanism (e.g.
monitoring functions)
3. Sufficient pressure or vacuum
generation
5. Minimal blood damage
6. Adjustable occlusion settings
(roller pumps only)
7. Pulsatile flow generation
8. High efficiency
9. High reliability and safety
10 . Option for emergency
operation (e.g., manual
operation)
Blood Pumps and Their Function
Roller Pumps:-
De Bakey blood pumps are based on the
displacement principle and deliver blood
through a tubing segment from the pump
housing using rotating rollers
The roller pump consists of :-
1. Rotating pump arm with two attached
cylindrical rollers
2. Pump housing into which a semicircular
silicone tubing segment is inserted and
then secured using special tubing
inserts
 The rotating rollers alternately compress
the tubing segment and deliver the liquid
contained in the tubing in accordance
with the rotational speed and direction
 The exact adjustment of the two rollers
determines the delivery
 rollers move outward symmetrically and
block the inserted tube evenly to reduces
erythrocyte damage caused by shear
stress or direct crushing.
Blood Pumps and Their Function
Centrifugal Pumps:-
Such pumps use centrifugal
forces to transport the blood instead
of tubing compression.
Because of the technical principle of
operation, the centrifugal pump has
a limited application.
Centrifugal pumps have the
advantage of :-
delivering limited amounts of air
causing less blood damage in the
long run.
An additional flow meter is required
to determine the delivered flow is a
disadvantage.
Centrifugal Pump from www.britannica.com
Oxygenator and Gas Exchange Function
Oxygenator:-
The artificial lung, also called the
oxygenator, takes on the lung function
during ECC and is therefore responsible for
the exchange of vital gases.
The membrane oxygenators:-
Are now used as standard.
They contain a semipermeable membrane
in the shape of a microporous hollow fiber.
This liquid-impermeable membrane
separates the gas side from the
bloodstream.
Due to the partial pressure gradients, O2
and CO2 diffuse through the microporous
membrane
Today, most membranes are made of
polypropylene or polyethylene and permit
operating times of 6–8 h
 The oxygenator must be able to
oxygenate about 5 litres per minute
of venous blood from 65% oxygen
saturation to above 95% oxygen
saturation before the blood enters
the body systems.
oxygenator
The oxygenator
The oxygenator consists of :-
1. A cylindrical glass vessel of 15 cm
internal diameter, 38 cm long
2. 80–100 stainless steel discs of 0.6 mm
thickness and 14 cm diameter are
mounted axially with 3 mm spacers
between discs.
The shaft on which the discs are mounted
is hollow and is supported by three ball
bearings at the ends.
The oxygen is fed through the axis of the
shaft and it enters the oxygenator through
the distributing apertures on the
circumference of the shaft.
The oxygen feed system permits use of disc
diameters close to the diameter of the
oxygenator cylinder so that for a given
blood priming volume maximum utilization
of the available surface area is made.
The optimum blood level in the oxygenator
that provides the maximum film area for a
given volume is 0.7 R where R is the radius
of the disc
Tubing Systems for Extracorporeal Circulation
The tubing system is used to connect the
individual components of the
extracorporeal system and forms a closed
circuit with the vascular system of the
patient.
Depending on the location, either PVC or
silicone tubing is used
Tubes are available in different diameters
and wall thicknesses.
Tubing sizes range from a diameter of
1/8" and a wall thickness of 1/16" to
3/16" and 1/4 " for pediatric applications
(with correspondingly low flow rates).
Tubing for adult applications is usually
3/8 " or 1/2 " with a wall thickness of
3/32", facilitating flow rates of more than
10 l/min
Completely preassembled tubing
system with infant oxygenator
(Sorin Group S.p.A., Milan)
Blood Filters
Blood filters are integrated into the
ECC system mainly to avoid micro-
embolisms caused by autologous
effect , foreign particles, and
microbubbles
Depth Filters
made of Dacron wool or
polyurethane foam, are inserted into
cardiotomy reservoirs and are used
mainly for particle filtration.
The pore size can vary from 80 -
100μm (coarse separation of
particles) to 20–40μm (micropore
range).
Mesh Filters:-
Made of a mesh of woven polyester
strands.
Used as arterial blood filters and work
on the principle of sieves.
The pore size of mesh filters ranges
from 20 to 40μm.
Unlike depth filters, mesh filters have
little adhesive force.
Their air retention properties are
excellent as air bubbles can only pass
the filter medium when a certain
pressure difference (bubble point
pressure) is reached.
Cardiotomy Reservoir
The cardiotomy reservoir functions:-
1. Collects and filters the blood
aspirated from the operation area
and feeds it back into the ECC as
required.
2. Provides volume storage
3. The transparent housing facilitates
continuous level control
 A detailed scale makes it easy to
quantify the level and detect
corresponding changes in volume
 A minimum residual volume always
remains in the reservoir during the
entire ECC to prevent air delivery
into the extracorporeal system
Cannulae
Cannulae are the interface between the
ECC system and the vascular system of the
patient.
They are inserted into the relevant vessels
by a surgeon, where they are secured and
deaerated before their sterile connection
to the tubing system of the HLM.
Two types of cannulae:--
1. Arterial Cannulation
2. Venous Cannulation
Arterial Cannulation:-
 Arterial cannulae are used to return the
oxygenated blood to the systemic
circulation of the patient
 The aorta ascendens is the most
frequently used site.
 The type and size of the selected cannula
depends on both the required blood flow
and the anatomic conditions
Venous Cannulation:-
Venous cannulae drain the patient’s blood
from the venous vascular system to the
HLM.
The type and size of the selected cannula is
determined by conditions that are similar
to those on the arterial side.
Vent Catheters:-
Inserted to protect the temporarily
arrested heart and especially the left
ventricle from overextension due to blood
flowing back from the bronchial
circulation.
Using this vent suction tube, excessive
blood is drained usually via the auricle of
the left atrium or into the cardiotomy
reservoir of the HLM
Components of the Heart–Lung Machine
HLM Basic Components:-
1. Mobile console for mounting
multiple pumps
2. Adjustable mast system for
mounting Blood pumps
3. Control and monitoring devices
4. Display and control panel
5. Electronic or mechanical gas
blender
6. Anesthetic gas vaporizer
7. Electronic documentation system
Additional HLM Components:-
1. water mattress
2. Heater–Cooler Device
During ECC, the heat exchanger of the
oxygenator indirectly regulates the
patient’s body temperature by heating or
cooling the blood flow. Patients can also
be placed on a water mattress that
increases or decreases body temperature
Components of the Heart–Lung Machine
Mobile console:-
for mounting :-
1. Multiple pumps
2. The power supply
3. Emergency power supply
4. Electronics parts
Adjustable mast system:-
for mounting the holders for:-
1. The oxygenator
2. Filters
3. Cardiotomy reservoirs
4. External pumps, and additional
devices
Figure : shows a modern HLM.
Blood pumps:-
Large roller pump:
pump with a longer tubing path in the
pump housing that produces higher flow
rates, required for the arterial blood flow,
suction, or vent pump
Small roller pump:
pump with a shorter tubing path in the
pump housing for lower flow rates, e.g.,
for the perfusion of infants and children
or for the administration of cardioplegic
solutions, usually a double pump
Centrifugal pumps:
pumps for the arterial blood flow.
Components of the Heart–Lung Machine
Components of the Heart–Lung Machine
Control and monitoring devices:-
Include :-
1. Pressure monitor:- including
sensors, for measuring different
pressures in the extracorporeal
system and for controlling the flow
rates.
2. Temperature monitor:- including
sensors, for measuring and
displaying different system
temperatures and, if required,
patient temperatures.
3. Level monitor :- including sensors,
for measuring and controlling the
volume level in the cardiotomy
reservoir.
4. Bubble monitor :- in the form of
ultrasonic sensors that regulate the
flow rate of the affected pump
when air is detected in the system.
5. Pulsatile flow control :-for creating
and controlling a pulsatile flow
profile.
6. Cardioplegia control:- including pressure
and bubble sensor, for controlling
cardioplegia delivery.
7. Timer for measuring important times
and intervals during perfusion such as total
perfusion time, aortic clamping time, and
reperfusion time
Display and control panel:-
providing access to all display and
control elements described above as
well as to additional system
information and alarm management
Electronic or mechanical gas
blender:-
1. For controlling and displaying the
ventilation gases (air, O2, and CO2)
that are delivered to the gas side of
the oxygenator; vacuum controller.
2. For the precise control and display
of a permanent vacuum on the
venous reservoir
Components of the Heart–Lung Machine
Figure : shows a modern HLM.
Components of the Heart–Lung Machine
Anesthetic gas vaporizer:-
Allowing and displaying the
administration of a precisely dosed
amount of anesthetic gas to the
oxygenator.
Electronic documentation system
That displays and stores all relevant data
during perfusion.
All data are stored centrally and
displayed instantly as a perfusion report
which allows for data evaluation at a later
stage for statistical and scientific
purposes.
perfusion report
Additional HLM Components:-
Heater–Cooler Device (Heat exchanger ):-
The heater–cooler device delivers water with
a temperature ranging from 2 to 42°C.
The Heat exchanger contain:-
1. Two chamber pumps ensure that the
delivered water does not cause a
buildup of overpressure in the circuits.
2. Two circuits supply the oxygenator and
the heating/cooling mattress
3. Third circuit is connected to the heat
exchanger in the cardioplegia system.
4. Two independent tanks with different
temperatures, which allows instant
switching from cold to warm or vice
versa
 Modern heater–cooler devices feature
several circuits with different
temperatures
Heater–cooler devices have always presented
a hygiene challenge because
1. The system parts that transport water
inevitably become contaminated by
germs.
 Regular water changes and the use of
appropriate chemical cleaning agents are
the two most effective cleaning and
maintenance strategies.
Components of the Heart–Lung Machine
References
Springer Handbook of Medical Technology
Rüdiger Kramme, Klaus-Peter Hoffmann,
Robert S. Pozos
Springer-Verlag Berlin Heidelberg 2011

More Related Content

What's hot

CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
CARDIO PULMONARYBYPASS BLOOD PUMPS  DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...CARDIO PULMONARYBYPASS BLOOD PUMPS  DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...DR NIKUNJ SHEKHADA
 
Heart lung machine also referred to as extracorporeal circulation...
Heart lung machine also referred to as extracorporeal circulation...Heart lung machine also referred to as extracorporeal circulation...
Heart lung machine also referred to as extracorporeal circulation...Sharmin Susiwala
 
Cell Saver -1.pptx
Cell Saver -1.pptxCell Saver -1.pptx
Cell Saver -1.pptxssuserb836a1
 
Pulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionPulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionNahas N
 
Conduct of cardio pulmonary bypass
Conduct of cardio pulmonary bypassConduct of cardio pulmonary bypass
Conduct of cardio pulmonary bypassManu Jacob
 
Microemboli gaseous and particulate
Microemboli gaseous and particulateMicroemboli gaseous and particulate
Microemboli gaseous and particulateManu Jacob
 
cardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetycardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetyIda Simanjuntak
 
CARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSCARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSSreevidya V S
 
Pediatric cardiopulmonary bypass
Pediatric cardiopulmonary bypassPediatric cardiopulmonary bypass
Pediatric cardiopulmonary bypasskp gourav
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTRanjith Thampi
 
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASS
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASSEMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASS
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASSGLORY MINI MOL. A
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in ORYamini Dhengle
 
Cardiopulmonary bypass effect to others organs
Cardiopulmonary bypass effect to others organsCardiopulmonary bypass effect to others organs
Cardiopulmonary bypass effect to others organsIda Simanjuntak
 
Principles Of Intra Aortic Balloon Pump Counterpulsation
Principles Of Intra Aortic Balloon Pump CounterpulsationPrinciples Of Intra Aortic Balloon Pump Counterpulsation
Principles Of Intra Aortic Balloon Pump Counterpulsationhospital
 
Cardiopulmonary Bypass and Valvular Surgery
Cardiopulmonary Bypass and Valvular SurgeryCardiopulmonary Bypass and Valvular Surgery
Cardiopulmonary Bypass and Valvular SurgeryMuhammad Eimaduddin
 

What's hot (20)

CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
CARDIO PULMONARYBYPASS BLOOD PUMPS  DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...CARDIO PULMONARYBYPASS BLOOD PUMPS  DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
 
Heart lung machine also referred to as extracorporeal circulation...
Heart lung machine also referred to as extracorporeal circulation...Heart lung machine also referred to as extracorporeal circulation...
Heart lung machine also referred to as extracorporeal circulation...
 
Cell Saver -1.pptx
Cell Saver -1.pptxCell Saver -1.pptx
Cell Saver -1.pptx
 
Pulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionPulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusion
 
Conduct of cardio pulmonary bypass
Conduct of cardio pulmonary bypassConduct of cardio pulmonary bypass
Conduct of cardio pulmonary bypass
 
Microemboli gaseous and particulate
Microemboli gaseous and particulateMicroemboli gaseous and particulate
Microemboli gaseous and particulate
 
Suction
SuctionSuction
Suction
 
CPB(CARDIO PULMONARY BYPASS)
CPB(CARDIO PULMONARY BYPASS)CPB(CARDIO PULMONARY BYPASS)
CPB(CARDIO PULMONARY BYPASS)
 
cardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetycardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safety
 
CARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASSCARDIO-PULMONARY BYPASS
CARDIO-PULMONARY BYPASS
 
Pediatric cardiopulmonary bypass
Pediatric cardiopulmonary bypassPediatric cardiopulmonary bypass
Pediatric cardiopulmonary bypass
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRT
 
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASS
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASSEMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASS
EMBOLISM AND FILTERS USED IN CARDIOPULMONARY BYPASS
 
Cardiopulmonarybypass
CardiopulmonarybypassCardiopulmonarybypass
Cardiopulmonarybypass
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in OR
 
Cardiopulmonary bypass effect to others organs
Cardiopulmonary bypass effect to others organsCardiopulmonary bypass effect to others organs
Cardiopulmonary bypass effect to others organs
 
PROSTHETIC VALVES
PROSTHETIC VALVESPROSTHETIC VALVES
PROSTHETIC VALVES
 
Principles Of Intra Aortic Balloon Pump Counterpulsation
Principles Of Intra Aortic Balloon Pump CounterpulsationPrinciples Of Intra Aortic Balloon Pump Counterpulsation
Principles Of Intra Aortic Balloon Pump Counterpulsation
 
Cardiopulmonary Bypass and Valvular Surgery
Cardiopulmonary Bypass and Valvular SurgeryCardiopulmonary Bypass and Valvular Surgery
Cardiopulmonary Bypass and Valvular Surgery
 
IABP
IABPIABP
IABP
 

Similar to Heart lung machine.pdf

heartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfheartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfvegadof539
 
heartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfheartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfvegadof539
 
Heart lung by pass unit
Heart lung by pass unitHeart lung by pass unit
Heart lung by pass unitIzzad Samir
 
Heart Lung Machine.pptx
Heart Lung Machine.pptxHeart Lung Machine.pptx
Heart Lung Machine.pptxPSNAVIJAY
 
Energy transmission system for artificial heart
Energy transmission system for artificial heartEnergy transmission system for artificial heart
Energy transmission system for artificial heartIshwar Bhoge
 
HEART-LUNG MACHINE (heart lung machine).pptx
HEART-LUNG MACHINE (heart lung machine).pptxHEART-LUNG MACHINE (heart lung machine).pptx
HEART-LUNG MACHINE (heart lung machine).pptxBorn13
 
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptx
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptxenergytransmissionsystemforartificialheart-180707175101 (1) (1).pptx
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptxMarkaVikaskumar
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationprapulla chandra
 
Physiology of Cardiopulmonary Bypass
Physiology of Cardiopulmonary BypassPhysiology of Cardiopulmonary Bypass
Physiology of Cardiopulmonary Bypasskshama_db
 
Blood pump perfusion
Blood pump perfusionBlood pump perfusion
Blood pump perfusionAkshayKatoch4
 
ppt39sak23-171123100040.pdf
ppt39sak23-171123100040.pdfppt39sak23-171123100040.pdf
ppt39sak23-171123100040.pdfvegadof539
 
Pumps, oxygenators and priming solution
Pumps, oxygenators and priming solutionPumps, oxygenators and priming solution
Pumps, oxygenators and priming solutionIndia CTVS
 
heart-lung-machine-pump.ppt
heart-lung-machine-pump.pptheart-lung-machine-pump.ppt
heart-lung-machine-pump.pptssuser1ecccc
 
Cardiac surgeries
Cardiac surgeriesCardiac surgeries
Cardiac surgeriesAmruta Pai
 
CARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSCARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSManu Jacob
 
Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherizationvasanth7pv
 

Similar to Heart lung machine.pdf (20)

heartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfheartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdf
 
heartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdfheartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdf
 
Heart–lung machine
Heart–lung machineHeart–lung machine
Heart–lung machine
 
Heart Lung Mechine
Heart Lung MechineHeart Lung Mechine
Heart Lung Mechine
 
10735108.ppt
10735108.ppt10735108.ppt
10735108.ppt
 
Heart lung by pass unit
Heart lung by pass unitHeart lung by pass unit
Heart lung by pass unit
 
Heart Lung Machine.pptx
Heart Lung Machine.pptxHeart Lung Machine.pptx
Heart Lung Machine.pptx
 
Energy transmission system for artificial heart
Energy transmission system for artificial heartEnergy transmission system for artificial heart
Energy transmission system for artificial heart
 
HEART-LUNG MACHINE (heart lung machine).pptx
HEART-LUNG MACHINE (heart lung machine).pptxHEART-LUNG MACHINE (heart lung machine).pptx
HEART-LUNG MACHINE (heart lung machine).pptx
 
Sinz cpb
Sinz cpbSinz cpb
Sinz cpb
 
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptx
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptxenergytransmissionsystemforartificialheart-180707175101 (1) (1).pptx
energytransmissionsystemforartificialheart-180707175101 (1) (1).pptx
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenation
 
Physiology of Cardiopulmonary Bypass
Physiology of Cardiopulmonary BypassPhysiology of Cardiopulmonary Bypass
Physiology of Cardiopulmonary Bypass
 
Blood pump perfusion
Blood pump perfusionBlood pump perfusion
Blood pump perfusion
 
ppt39sak23-171123100040.pdf
ppt39sak23-171123100040.pdfppt39sak23-171123100040.pdf
ppt39sak23-171123100040.pdf
 
Pumps, oxygenators and priming solution
Pumps, oxygenators and priming solutionPumps, oxygenators and priming solution
Pumps, oxygenators and priming solution
 
heart-lung-machine-pump.ppt
heart-lung-machine-pump.pptheart-lung-machine-pump.ppt
heart-lung-machine-pump.ppt
 
Cardiac surgeries
Cardiac surgeriesCardiac surgeries
Cardiac surgeries
 
CARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASSCARDIOPULMONARY BYPASS
CARDIOPULMONARY BYPASS
 
Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherization
 

More from ssuser1ecccc

Intro class. U1 M1.pptx
Intro class. U1 M1.pptxIntro class. U1 M1.pptx
Intro class. U1 M1.pptxssuser1ecccc
 
TASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxTASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxssuser1ecccc
 
4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptxssuser1ecccc
 
5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptxssuser1ecccc
 
unit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxunit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxssuser1ecccc
 
unit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxunit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxssuser1ecccc
 
MWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxMWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxssuser1ecccc
 
3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptxssuser1ecccc
 
Biomedical Waste Management.pptx
Biomedical Waste Management.pptxBiomedical Waste Management.pptx
Biomedical Waste Management.pptxssuser1ecccc
 
unit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxunit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxssuser1ecccc
 
2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptxssuser1ecccc
 
projectmanagement.ppt
projectmanagement.pptprojectmanagement.ppt
projectmanagement.pptssuser1ecccc
 

More from ssuser1ecccc (18)

Intro class. U1 M1.pptx
Intro class. U1 M1.pptxIntro class. U1 M1.pptx
Intro class. U1 M1.pptx
 
NGP BME Final.ppt
NGP BME Final.pptNGP BME Final.ppt
NGP BME Final.ppt
 
ES.pptx
ES.pptxES.pptx
ES.pptx
 
TASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxTASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptx
 
4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx
 
5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx
 
unit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxunit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptx
 
unit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxunit-iv-facility-safety.pptx
unit-iv-facility-safety.pptx
 
MWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxMWM Unit 3 (1).pptx
MWM Unit 3 (1).pptx
 
day9.ppt
day9.pptday9.ppt
day9.ppt
 
100014228.ppt
100014228.ppt100014228.ppt
100014228.ppt
 
3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx
 
Biomedical Waste Management.pptx
Biomedical Waste Management.pptxBiomedical Waste Management.pptx
Biomedical Waste Management.pptx
 
UNIT 1 MWM.pptx
UNIT 1 MWM.pptxUNIT 1 MWM.pptx
UNIT 1 MWM.pptx
 
unit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxunit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptx
 
Oxygenartors.pdf
Oxygenartors.pdfOxygenartors.pdf
Oxygenartors.pdf
 
2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx
 
projectmanagement.ppt
projectmanagement.pptprojectmanagement.ppt
projectmanagement.ppt
 

Recently uploaded

SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSCAESB
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxwendy cai
 

Recently uploaded (20)

9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentation
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptx
 

Heart lung machine.pdf

  • 2. Extracorporeal Circulation ECC or perfusion is defined as a procedure carried out with the aid of HLM In the early stages, cardiac surgery was mainly concerned with the treatment of congenital heart defects. Today, the main focus is on acquired heart defects. Access path to the operation area is kept as small as possible to reduce the patient’s surgical trauma this surgical technique usually has a direct influence on the configuration of the HLM and on the way the perfusion is carried out OPCAB (off-pump coronary artery bypass) refer to cardiac surgeries carried out without the support of an HLM The HLM can never be looked at in isolation additional components, mainly in the form of sterile disposables such as an oxygenator, heat exchanger, or reservoir, take on other important functions during ECC
  • 3. Introduction The Heart-lung machine is a device which is used to provide blood circulation and oxygenation while the heart is stopped during the open heart surgery. Heart-lung machine consists of three main functional units, the pump, the oxygenator and heat exchanger. To function, the heart-lung machine must be connected to the patient in a way that allows blood to be removed, processed, and returned to the body.
  • 4. History The researchers searched mainly for a means to provide oxygenated blood to organ preparations. They filtered the blood through various screens and membranes and even pumped it through the lungs of dogs or monkeys, but their problem was not to be solved, though this may be considered the beginning of research into a heart-lung device. The first successful open heart procedure on a human utilizing the heart lung machine was performed by John Gibbon in 1953. He repaired an atrial septal defect in an 18-year-old woman.
  • 5. Purpose of Heart-lung machine:- The heart-lung machine provides the benefit of a motionless heart during open heart surgery. The heart-lung machine is valuable during this time since the patient is unable to maintain blood flow to the lungs or the body.
  • 6. Surgical procedures in which Heart-lung machine is used:- Coronary artery bypasses surgery. Cardiac valve repair and/or replacement (aortic valve, mitral valve, tricuspid valve, pulmonic valve) Repair of large septum defects (atrial septum defect, ventricular septum defect, atrioventricular septum defect) Transplantation (heart transplantation, lung transplantation, heart–lung transplantation) Implantation of heart.
  • 8. Components of Heart-lung Machine:- Cardioplegia Cannula Blood reservoir Heparin Pump Roller pumps/Centrifugal pump Oxygenators Heat Exchanger
  • 9. Structure and Function of the Heart–Lung Machine The HLM is the basis of ECC and cover two important organ functions :- 1. Pump function of the heart 2. Gas exchange function of the lungs It must ensure :-  A sufficient perfusion volume that corresponds to the normal cardiac output of the patient under anesthetic  An adequate perfusion pressure (50– 90 mmHg)  It must also ensure sufficient oxygenation, elimination of CO2, and control of the blood temperature The following components make up the basic equipment of the ECC that is used during modern cardiac surgery: 1. Blood pumps 2. Oxygenator 3. Tubing system with various tubing diameters 4. Blood filters with various functions 5. Cardiotomy reservoir 6. Cannulae and intracardiac suction tubes
  • 10. Blood Pumps and Their Function Blood pumps can generally be categorized into two groups, depending on their function: 1. Roller pumps 2. Centrifugal pumps  Essentially, both pump types should meet certain criteria when used with the HLM: 1. Dosed delivery of liquids, with precise display of the actual flow rate and delivered volume 2. External control mechanism (e.g. monitoring functions) 3. Sufficient pressure or vacuum generation 5. Minimal blood damage 6. Adjustable occlusion settings (roller pumps only) 7. Pulsatile flow generation 8. High efficiency 9. High reliability and safety 10 . Option for emergency operation (e.g., manual operation)
  • 11. Blood Pumps and Their Function Roller Pumps:- De Bakey blood pumps are based on the displacement principle and deliver blood through a tubing segment from the pump housing using rotating rollers The roller pump consists of :- 1. Rotating pump arm with two attached cylindrical rollers 2. Pump housing into which a semicircular silicone tubing segment is inserted and then secured using special tubing inserts  The rotating rollers alternately compress the tubing segment and deliver the liquid contained in the tubing in accordance with the rotational speed and direction  The exact adjustment of the two rollers determines the delivery  rollers move outward symmetrically and block the inserted tube evenly to reduces erythrocyte damage caused by shear stress or direct crushing.
  • 12. Blood Pumps and Their Function Centrifugal Pumps:- Such pumps use centrifugal forces to transport the blood instead of tubing compression. Because of the technical principle of operation, the centrifugal pump has a limited application. Centrifugal pumps have the advantage of :- delivering limited amounts of air causing less blood damage in the long run. An additional flow meter is required to determine the delivered flow is a disadvantage. Centrifugal Pump from www.britannica.com
  • 13. Oxygenator and Gas Exchange Function Oxygenator:- The artificial lung, also called the oxygenator, takes on the lung function during ECC and is therefore responsible for the exchange of vital gases. The membrane oxygenators:- Are now used as standard. They contain a semipermeable membrane in the shape of a microporous hollow fiber. This liquid-impermeable membrane separates the gas side from the bloodstream. Due to the partial pressure gradients, O2 and CO2 diffuse through the microporous membrane Today, most membranes are made of polypropylene or polyethylene and permit operating times of 6–8 h  The oxygenator must be able to oxygenate about 5 litres per minute of venous blood from 65% oxygen saturation to above 95% oxygen saturation before the blood enters the body systems. oxygenator
  • 14. The oxygenator The oxygenator consists of :- 1. A cylindrical glass vessel of 15 cm internal diameter, 38 cm long 2. 80–100 stainless steel discs of 0.6 mm thickness and 14 cm diameter are mounted axially with 3 mm spacers between discs. The shaft on which the discs are mounted is hollow and is supported by three ball bearings at the ends. The oxygen is fed through the axis of the shaft and it enters the oxygenator through the distributing apertures on the circumference of the shaft. The oxygen feed system permits use of disc diameters close to the diameter of the oxygenator cylinder so that for a given blood priming volume maximum utilization of the available surface area is made. The optimum blood level in the oxygenator that provides the maximum film area for a given volume is 0.7 R where R is the radius of the disc
  • 15. Tubing Systems for Extracorporeal Circulation The tubing system is used to connect the individual components of the extracorporeal system and forms a closed circuit with the vascular system of the patient. Depending on the location, either PVC or silicone tubing is used Tubes are available in different diameters and wall thicknesses. Tubing sizes range from a diameter of 1/8" and a wall thickness of 1/16" to 3/16" and 1/4 " for pediatric applications (with correspondingly low flow rates). Tubing for adult applications is usually 3/8 " or 1/2 " with a wall thickness of 3/32", facilitating flow rates of more than 10 l/min Completely preassembled tubing system with infant oxygenator (Sorin Group S.p.A., Milan)
  • 16. Blood Filters Blood filters are integrated into the ECC system mainly to avoid micro- embolisms caused by autologous effect , foreign particles, and microbubbles Depth Filters made of Dacron wool or polyurethane foam, are inserted into cardiotomy reservoirs and are used mainly for particle filtration. The pore size can vary from 80 - 100μm (coarse separation of particles) to 20–40μm (micropore range). Mesh Filters:- Made of a mesh of woven polyester strands. Used as arterial blood filters and work on the principle of sieves. The pore size of mesh filters ranges from 20 to 40μm. Unlike depth filters, mesh filters have little adhesive force. Their air retention properties are excellent as air bubbles can only pass the filter medium when a certain pressure difference (bubble point pressure) is reached.
  • 17. Cardiotomy Reservoir The cardiotomy reservoir functions:- 1. Collects and filters the blood aspirated from the operation area and feeds it back into the ECC as required. 2. Provides volume storage 3. The transparent housing facilitates continuous level control  A detailed scale makes it easy to quantify the level and detect corresponding changes in volume  A minimum residual volume always remains in the reservoir during the entire ECC to prevent air delivery into the extracorporeal system
  • 18. Cannulae Cannulae are the interface between the ECC system and the vascular system of the patient. They are inserted into the relevant vessels by a surgeon, where they are secured and deaerated before their sterile connection to the tubing system of the HLM. Two types of cannulae:-- 1. Arterial Cannulation 2. Venous Cannulation Arterial Cannulation:-  Arterial cannulae are used to return the oxygenated blood to the systemic circulation of the patient  The aorta ascendens is the most frequently used site.  The type and size of the selected cannula depends on both the required blood flow and the anatomic conditions Venous Cannulation:- Venous cannulae drain the patient’s blood from the venous vascular system to the HLM. The type and size of the selected cannula is determined by conditions that are similar to those on the arterial side. Vent Catheters:- Inserted to protect the temporarily arrested heart and especially the left ventricle from overextension due to blood flowing back from the bronchial circulation. Using this vent suction tube, excessive blood is drained usually via the auricle of the left atrium or into the cardiotomy reservoir of the HLM
  • 19. Components of the Heart–Lung Machine HLM Basic Components:- 1. Mobile console for mounting multiple pumps 2. Adjustable mast system for mounting Blood pumps 3. Control and monitoring devices 4. Display and control panel 5. Electronic or mechanical gas blender 6. Anesthetic gas vaporizer 7. Electronic documentation system Additional HLM Components:- 1. water mattress 2. Heater–Cooler Device During ECC, the heat exchanger of the oxygenator indirectly regulates the patient’s body temperature by heating or cooling the blood flow. Patients can also be placed on a water mattress that increases or decreases body temperature
  • 20. Components of the Heart–Lung Machine Mobile console:- for mounting :- 1. Multiple pumps 2. The power supply 3. Emergency power supply 4. Electronics parts Adjustable mast system:- for mounting the holders for:- 1. The oxygenator 2. Filters 3. Cardiotomy reservoirs 4. External pumps, and additional devices Figure : shows a modern HLM.
  • 21. Blood pumps:- Large roller pump: pump with a longer tubing path in the pump housing that produces higher flow rates, required for the arterial blood flow, suction, or vent pump Small roller pump: pump with a shorter tubing path in the pump housing for lower flow rates, e.g., for the perfusion of infants and children or for the administration of cardioplegic solutions, usually a double pump Centrifugal pumps: pumps for the arterial blood flow. Components of the Heart–Lung Machine
  • 22. Components of the Heart–Lung Machine Control and monitoring devices:- Include :- 1. Pressure monitor:- including sensors, for measuring different pressures in the extracorporeal system and for controlling the flow rates. 2. Temperature monitor:- including sensors, for measuring and displaying different system temperatures and, if required, patient temperatures. 3. Level monitor :- including sensors, for measuring and controlling the volume level in the cardiotomy reservoir. 4. Bubble monitor :- in the form of ultrasonic sensors that regulate the flow rate of the affected pump when air is detected in the system. 5. Pulsatile flow control :-for creating and controlling a pulsatile flow profile. 6. Cardioplegia control:- including pressure and bubble sensor, for controlling cardioplegia delivery. 7. Timer for measuring important times and intervals during perfusion such as total perfusion time, aortic clamping time, and reperfusion time
  • 23. Display and control panel:- providing access to all display and control elements described above as well as to additional system information and alarm management Electronic or mechanical gas blender:- 1. For controlling and displaying the ventilation gases (air, O2, and CO2) that are delivered to the gas side of the oxygenator; vacuum controller. 2. For the precise control and display of a permanent vacuum on the venous reservoir Components of the Heart–Lung Machine Figure : shows a modern HLM.
  • 24. Components of the Heart–Lung Machine Anesthetic gas vaporizer:- Allowing and displaying the administration of a precisely dosed amount of anesthetic gas to the oxygenator. Electronic documentation system That displays and stores all relevant data during perfusion. All data are stored centrally and displayed instantly as a perfusion report which allows for data evaluation at a later stage for statistical and scientific purposes. perfusion report
  • 25. Additional HLM Components:- Heater–Cooler Device (Heat exchanger ):- The heater–cooler device delivers water with a temperature ranging from 2 to 42°C. The Heat exchanger contain:- 1. Two chamber pumps ensure that the delivered water does not cause a buildup of overpressure in the circuits. 2. Two circuits supply the oxygenator and the heating/cooling mattress 3. Third circuit is connected to the heat exchanger in the cardioplegia system. 4. Two independent tanks with different temperatures, which allows instant switching from cold to warm or vice versa  Modern heater–cooler devices feature several circuits with different temperatures Heater–cooler devices have always presented a hygiene challenge because 1. The system parts that transport water inevitably become contaminated by germs.  Regular water changes and the use of appropriate chemical cleaning agents are the two most effective cleaning and maintenance strategies. Components of the Heart–Lung Machine
  • 26. References Springer Handbook of Medical Technology Rüdiger Kramme, Klaus-Peter Hoffmann, Robert S. Pozos Springer-Verlag Berlin Heidelberg 2011