Biopesticide (2).pptx .This slides helps to know the different types of biop...
Energy resolution of several scintillating crystals using different readout systems.
1. Energy resolution of several scintillating crystals
using differents readout systems
Measurements performed during Jul-Sept 2007 at the GSI
Mart´ın Gasc´on
Particle Physics Departament
University of Santiago de Compostela
5th. October, 2007
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 1
2. Contents
Tests performed
1 Energy resolution calibration using 1 cm3 CsI(Tl)
2 Energy resolution of CsI(Na) and BGO using APDs
3 Comparison between APDs and PIN Diodes
4 Energy resolution of LaBr3 using PMTs
5 Energy resolution of LaCl3 using PMTs
6 Non-Uniformities measurements
7 Best values obtained for energy resolution
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 2
3. 1. Energy resolution calibration
using 1 cm3
CsI(Tl)
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 3
4. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Experimental setup
The detector is electrically isolated by a grounded Faraday box
The detector and Preamplifier are placed in a metalic box with a humidity control system
The system is temperature monitorized
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 4
5. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Bias Voltage Optimization
Bias Voltage Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.
300 937.3 ± 0.3 4.87 ± 0.08 937.1 ± 0.4 4.90 ± 0.08
320 1254 ± 0.5 5.12 ± 0.09 1258 ± 0.5 5.16 ± 0.10
340 1698 ± 0.7 5.37 ± 0.10 1705 ± 0.6 4.83 ± 0.09
360 2357 ± 0.9 5.23 ± 0.09 2366 ± 0.9 4.79 ± 0.09
380 3255 ± 1.3 5.14 ± 0.09 3266 ± 1.2 4.75 ± 0.08
400 4517 ± 1.8 4.97 ± 0.08 4527 ± 1.6 4.57 ± 0.08
420 6211 ± 2.3 4.62 ± 0.08 6222 ± 2.2 4.54 ± 0.08
430 7275 ± 2.7 4.74 ± 0.09 7285 ± 2.7 4.71 ± 0.08
Table: Photopeak position vs Bias Voltage
Spectra for each Bias Voltage are taken twice,
in order to check stability and to reduce
uncertainties
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 5
6. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Bias Voltage Optimization
Not saturated between 300 V and 420 V
There is no minimum in the energy resolution
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 6
7. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Bias Voltage Optimization
Bias Voltage Amp. gain Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.
360 135 6307 ± 2.3 4.51 ± 0.08 6304 ± 2.3 4.81 ± 0.09
370 113 6330 ± 2.3 4.71 ± 0.08 6332 ± 2.3 4.59 ± 0.08
380 96 6330 ± 2.2 4.61 ± 0.08 6326 ± 2.4 4.62 ± 0.08
390 81 6335 ± 2.4 4.85 ± 0.09 6340 ± 2.3 4.74 ± 0.08
400 69 6296 ± 2.4 4.92 ± 0.09 6302 ± 2.3 4.74 ± 0.09
410 59 6333 ± 2.3 4.71 ± 0.09 6333 ± 2.3 4.83 ± 0.09
Table: Photopeak position vs Bias Voltage
Varying Bias Voltage and Amplifier gain to keep
constant the photopeak channel
There is minimum of energy resolution around
380 V
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 7
8. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Amplifier Gain
Linear along the whole dynamic range
the higher the Amplifier gain, the better the energy resolution
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 8
9. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Acquisition Time
There is a statistical deficit below 40 seconds
60 seconds give us a good energy resolution
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 9
10. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Shaping Time
3 µs and 6 µs are a good trade-off
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 10
11. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Final energy resolution
Temp. o
C Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.
24.3 ± 0.1 6156 ± 2.3 4.96 ± 0.09 6154 ± 2.4 4.87 ± 0.09
24.3 ± 0.1 6159 ± 2.4 5.08 ± 0.09 6156 ± 2.4 5.04 ± 0.09
24.3 ± 0.1 6159 ± 2.3 4.75 ± 0.09 6152 ± 2.3 4.65 ± 0.09
24.3 ± 0.1 6157 ± 2.5 4.97 ± 0.09 6155 ± 2.4 4.93 ± 0.09
24.3 ± 0.1 6151 ± 2.3 4.81 ± 0.09 6155 ± 2.4 4.86 ± 0.09
Table: Ten spectra acquired for st = 3µs, t = 50s, HV = 380V and G = 105
Temp. o
C Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.
24.3 ± 0.1 6220 ± 2.2 4.59 ± 0.08 6221 ± 2.3 4.68 ± 0.08
24.3 ± 0.1 6219 ± 2.3 4.74 ± 0.09 6219 ± 2.2 4.68 ± 0.08
24.3 ± 0.1 6219 ± 2.2 4.60 ± 0.08 6222 ± 2.2 4.59 ± 0.08
24.3 ± 0.1 6220 ± 2.3 4.65 ± 0.08 6221 ± 2.2 4.63 ± 0.08
24.3 ± 0.1 6216 ± 2.2 4.55 ± 0.08 6219 ± 2.3 4.73 ± 0.09
Table: Ten spectra acquired for st = 6µs, t = 50s, HV = 380V and G = 96
3 µs: Mean value: 4.89 ± 0.09 - Best value: 4.65 ± 0.09
6 µs: Mean value: 4.64 ± 0.09 - Best value: 4.55 ± 0.08
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 11
12. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Energy resolution vs Energy
1170 keV is part of Comptom edge of 1332 keV of 60Co
The energy resolution is function of the square root of the incident energy photon
We used a 60Co, 137Cs and 152Eu sources
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 12
13. Energy resolution calibration using 1 cubic cm CsI(Tl) + APD
Energy resolution
The reported energy resolution obtained is 4.55 % is similar to the obtained in
”Optimization of energy resolution obtained with CsI(Tl) crystals for the R3B
Calorimeter” M. Gasc´on et al., Submited to IEEE Transactions on Nuclear
Science.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 13
14. Albert Einstein
A theory is
something nobody
believes, except the
person who made it.
An experiment is
something everybody
believes, except the
person who made it.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 14
15. 2. Energy resolution of
CsI(Na) and BGO using APDs
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 15
16. Energy resolution of CsI(Na) and BGO using APDs
Experimental setup
Frustum shape crystals (170 mm length)
All are encapsulated into aluminium shielding
9x9 and 15x15 windows (1mm thick) (Face A and B)
Delivered by Scionix Holland.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 16
17. Energy resolution of CsI(Na) and BGO using APDs
Experimental setup
The detector is electrically isolated by a grounded Faraday box
The detector and Preamplifier are placed in a metalic box with a humidity control system
The system is temperature monitorized
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 17
18. Energy resolution of CsI(Na) and BGO using APDs
CsI(Na) - Shaping Time - Face A
between 1 and 3 µs the energy resolution is approximately constant
3 µs requires the lowest amplifier gain
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 18
19. Energy resolution of CsI(Na) and BGO using APDs
CsI(Na) - Acquisition Time - Face A
80 seconds is a good trade-off
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 19
20. Energy resolution of CsI(Na) and BGO using APDs
CsI(Na) - Amplifier Gain - Face A
Linear along the whole amplifier dynamic range
500 is the situable one
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 20
21. Energy resolution of CsI(Na) and BGO using APDs
CsI(Na) - Amplifier Gain - Face A and B
Face B provides more light collection
The higher the light collection, the better the energy resolution
300 is selected for Face B
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 21
22. Energy resolution of CsI(Na) and BGO using APDs
CsI(Na) - Acquisition Time - Face A and B
The higher the light collection, the better the energy resolution
80 seconds is a good trade-off
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 22
23. Energy resolution of CsI(Na) and BGO using APDs
BGO - Shaping Time - Face B
between 1 and 2 µs the energy resolution is approximately constant
1 µs requires lower amplifier gain
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 23
24. Energy resolution of CsI(Na) and BGO using APDs
BGO - Amplifier Gain - Face B
Linear along the whole dynamic range
500 is selected, above this value noise is amplified.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 24
25. Energy resolution of CsI(Na) and BGO using APDs
BGO - Acquisition Time - Face B
60 seconds is a good trade-off
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 25
26. African Proverb
Not to know
is bad. Not to
wish to know is
worse.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 26
27. 3. Comparison between
APDs and PIN Diodes
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 27
28. Comparison between APDs and PIN Diodes
Experimental setup
Filter + Preamp made by GSI
No chance of varying Bias Voltage
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 28
29. Comparison between APDs and PIN Diodes
Amplifier Gain
Bad energy resolution
PINs Don’t need High Voltage
Faster stabilization
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 29
30. Comparison between APDs and PIN Diodes
Acquisition Time
similar behavior to Hamamatsu
70 seconds is appropriated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 30
31. Comparison between APDs and PIN Diodes
Shaping Time
1 µs and 6 µs are minima
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 31
32. Comparison between APDs and PIN Diodes
Experimental setup
Canberra 2001A Preamplifier
PIN Diode S3590-08 from Hamamatsu
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 32
33. Comparison between APDs and PIN Diodes
Bias Voltage
PINs require much lower Bias Voltage
Faster stabilization (stable between 40 V and 80 V)
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 33
34. Comparison between APDs and PIN Diodes
Bias Voltage
60 V is selected
Energy resolution is 2 % worse for PIN diode
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 34
35. Comparison between APDs and PIN Diodes
Amplifier Gain
Linear along the whole dynamic range
750 is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 35
36. Comparison between APDs and PIN Diodes
Acquisition Time
Similar behavior to Hamamatsu
60 seconds is adecuated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 36
37. Comparison between APDs and PIN Diodes
Shaping Time
3 µs and 6 µs are good trade off for both
Pin Diode are too noisy, so that they can not resolve energies below 400 keV
Pin Diode has worse energy resolution, but similar trend for energies above 500
keV
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 37
38. 4. Energy resolution of LaBr3
using PMTs
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 38
39. Energy resolution of LaBr3 using PMTs
R329 - Experimental setup
Brillance 380 (Saint Gobain)
R329 Hamamatsu PMT
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 39
40. Energy resolution of LaBr3 using PMTs
R329 - Bias Voltage
Last point is out of the curve
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 40
41. Energy resolution of LaBr3 using PMTs
R329 - Bias Voltage
Varying Bias Voltage and Amplifier gain to keep constant the photopeak channel
1050 V is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 41
42. Energy resolution of LaBr3 using PMTs
R329 - Amplifier Gain
Not linear for gain higher than 40
36 is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 42
43. Energy resolution of LaBr3 using PMTs
R329 - Acquisition Time
120 seconds is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 43
44. Energy resolution of LaBr3 using PMTs
R329 - Shaping Time
1 µs and 2 µs are good trade off
after 2 µs saturates
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 44
45. Energy resolution of LaBr3 using PMTs
XP1918 - Experimental setup
Brillance 380 (Saint Gobain)
XP1918 Photonis
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 45
46. Energy resolution of LaBr3 using PMTs
XP1918 - Bias Voltage
After 1100 V is saturated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 46
47. Energy resolution of LaBr3 using PMTs
XP1918 - Amplifier Gain
Not linear from 40
36 is adecuated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 47
48. Energy resolution of LaBr3 using PMTs
XP1918 - Acquisition Time
80 seconds is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 48
49. Energy resolution of LaBr3 using PMTs
XP1918 - Shaping Time
1 or 2 µs are selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 49
50. Energy resolution of LaBr3 using PMTs
XP1918 - Energy resolution
4.11 % is the lowest energy resolution without saturation.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 50
51. Energy resolution of LaBr3 using PMTs
XP1918 - Energy resolution
The presented by Saint Gobain resolutions could be saturated
The relation between Photopeak channel/Compton is higher
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 51
52. 5. Energy resolution of LaCl3
using PMTs
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 52
53. Energy resolution of LaCl3 using PMTs
Experimental setup
R329 PMT from Hamamatsu
Brillance 350 - 10 cm LaCl3 (1 and 2 windows)
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 53
54. Energy resolution of LaCl3 using PMTs
R329 - Bias Voltage
After 1200 V is saturated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 54
55. Energy resolution of LaCl3 using PMTs
R329 - Bias Voltage
Varying the gain and Bias Voltage
1050 V is adecuated
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 55
56. Energy resolution of LaCl3 using PMTs
R329 - Amplifier Gain
linear
40 is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 56
57. Energy resolution of LaCl3 using PMTs
R329 - Acquisition Time
120 seconds is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 57
58. Energy resolution of LaCl3 using PMTs
R329 - Shaping Time
1 or 2 µs are selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 58
59. Energy resolution of LaCl3 using PMTs
R329 - Amplifier Gain
Covering the window with aluminized mylar
linear
60 is selected
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 59
60. Energy resolution of LaCl3 using PMTs
R329 -Energy resolution
5.78 % is the best energy resolution achieved.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 60
63. Non-Uniformities measurements
Non-Uniformity definitions
Non Uniformity in the Ligth Output
is one of the factors worsening the energy resolution
at least, 2 methods are used to define it: G and δ
Definition of G
G =
LOmax − LOmin
LOmed
(1)
Definition of δ
LO
LOmed
= 1 + δ(
x − xmed
xmed
) (2)
Notes
δ-value is independent of the number of measuring points, but G does not
infinite number of measuring points means G = 2 · δ
Example:
For a 10 cm lenght crystal, if we measure 9 points every 1 cm, we get:
G = 1.6 · δ
(1) D.M. Beylin, et al., Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.
(2) G. Ren, X et al.,Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 63
64. Non-Uniformities measurements
Non-Uniformity definitions
Non Uniformity in the Ligth Output
is one of the factors worsening the energy resolution
at least, 2 methods are used to define it: G and δ
Definition of G
G =
LOmax − LOmin
LOmed
(1)
Definition of δ
LO
LOmed
= 1 + δ(
x − xmed
xmed
) (2)
Notes
δ-value is independent of the number of measuring points, but G does not
infinite number of measuring points means G = 2 · δ
Example:
For a 10 cm lenght crystal, if we measure 9 points every 1 cm, we get:
G = 1.6 · δ
(1) D.M. Beylin, et al., Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.
(2) G. Ren, X et al.,Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 63
65. Non-Uniformities measurements
10 cm CsI(Tl) + APD + PMT
Spectra were taken using Cs-137 and Co-60 sources
Fitting the photopeak channel from the 1332 MeV gamma of a Co-60 source
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 64
66. Non-Uniformities measurements
10 cm CsI(Tl) + APD + PMT
G = 16.7 %; δ = 9.28% for 10 cm CsI(Tl) + HAMA
G = 9.6 %; δ = 5.33% for 10 cm CsI(Tl) + XP1918
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 65
67. Non-Uniformities measurements
10 cm CsI(Tl) + APD + PMT
Increasing the PMT gain and plotting as a function of distance to APD
Using a 1332 keV of Cobaltum 60 source (5 min/point)
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 66
68. Non-Uniformities measurements
10 cm CsI(Tl) + APD + PMT
Doing the sum of the photopeak channel for both
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 67
69. Non-Uniformities measurements
CsI(Na) - Non Uniformity in light collection - Face B
Collimated source of 60Co by 5 cm of Lead
Face A produced few light to perform this test
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 68
70. 7. Best values obtained
for energy resolution
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 69
71. Best values obtained for energy resolution
Energy resolution
Crystal lenght Readout Premplifiera
Shaping Bias Volt. Energy Resolutionb
CsI(Tl) 1 cm APDc
Canb. 2001 6 µs e-m 4.50%
CsI(Tl) 1 cm Pin GSI-Preamp 6 µs - 18.7%
CsI(Tl) 1 cm Pin Canb. 2001 6 µs e-m 6.45%
CsI(Tl) 10 cm APD Canb. 2001 6 µs e-m 5.36%
LaBr3 3 cm R329 - 2 µs Wenzel 4.54%
LaBr3 3 cm XP1918 - 2 µs Wenzel 4.11%
CsI(Na) 17 cm APD Canb. 2001 3 µs e-m 17.1%
BGO 17 cm APD Canb. 2001 1 µs e-m 23.7%
LaCl3 10 cm R329 - 1 µs Wenzel 5.70%
LaCl3 (2w) 10 cm R329 - 1 µs Wenzel 8.35%
a
The Amplifier used was Ortec 572
b
Energy Resolution for the 662 keV photon of a Cs-137 source
c
APD is S8664-1010 from Hamamatsu
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 70
72. Conclusions
The calibration showed that the energy resolution value of 4.5 % is repeated with different electronic
chains
The enery resolution of CsI(Na) and BGO are bad because its bad quality
The used window for the frustum like crystals produced an energy resolution worsening
There is difference of aproximately 4 % in the energy resolution between Face A and B in frustum like
crystals
The Filter and Preamplifier made by GSI for Pin Diodes don’t offer an energy resolution comparable to
Canberra Preamplifier
Pin Diodes don’t need High Voltage, are stable between 40 and 80 V, have an energy resolution of 6.5
% for 662 keV photons
Pin Diodes has worse energy resolution for low shaping times and for incident energies below 400 keV
than APDs.
LaBr3 produces so many light in a so short time that saturates common PMT working at tipical voltage
values
Working at linear gain, the best energy resolution obtained for LaBr3 was 4.11 % for XP1918 PMT from
Photonis
The best energy resolution for LaCl3 was 5.78 % for the R329 PMT from Hamamatsu
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 71
73. Bibliography
Technical Proposal for the Design, Construction, Commissioning and Operation of R3B.
http://www-land.gsi.de/r3b/
H. Alvarez−Pol et al. A first proposal for the geometry of the Total Absorption Calorimeter
design at R3B. Internal Note: R3B CAL 01/05. http://www.usc.es/genp/
C.W.E. van Eijk, P. Dorenbos, E.V.D. van Loef, K. Kramer, H. Gudel ”Energy resolution of
some new inorganic-scintillator in gamma-ray detectors” Radiation Measurements, vol. 33, pp
521-525, 2001.
Hamamatsu Photonics: Photomultiplier Tubes and Related Products
D.M. Beylin, A.I. Korchagin, A.S. Kuzmin, L.M. Kurdadze, S.B. Oreshkin, S.E. Petrov, B.A.
Shwartz, ”Study of the radiation hardness of CsI(Tl) scintillation crystals,”
Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.
G. Ren, X. Chen, S. Lu, Z.Li, X. Xue, D. Shen, ”Non-uniformity of light output in large-sized
CsI(Tl) crystals grown by non-vacuum Bridgman method”,
Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 72
74. Albert Einstein
Put your hand on a hot stove
for a minute,
and it seems like an hour.
Sit with a pretty girl
for an hour,
and it seems like a minute.
That’s relativity.
Mart´ın Gasc´on Energy Resolution of several scintillating crystals 5th. October, 2007 73