Defibrillation is a process that delivers an electric shock to the heart to stop an irregular heartbeat and restore normal rhythm. It is the only effective treatment for cardiac arrest caused by ventricular fibrillation or pulseless ventricular tachycardia. Early defibrillation is critical, as survival rates decrease by 10% each minute without treatment. Different types of defibrillators include automated external defibrillators and implantable cardioverter defibrillators. Key factors that influence defibrillation success include transthoracic impedance, electrode placement and position, and shock waveform and energy level delivered.
Defibrillators are devices that deliver an electrical shock to the heart to terminate life-threatening cardiac arrhythmias like ventricular fibrillation. They were first demonstrated on dogs in 1899 and first used on a human in 1947. There are several types including manual external defibrillators, automated external defibrillators (AEDs) found in public places, implantable cardioverter-defibrillators, and wearable defibrillators. AEDs can analyze the heart rhythm and instruct the user, who needs no training, on whether a shock is needed. Defibrillation aims to stop all cardiac electrical activity so the heart's natural pacemaker can resume normal rhythm, while cardioversion delivers synchronized shocks
Defibrillation is a process that delivers an electric shock to the heart to stop ventricular fibrillation and restore normal rhythm. It involves using a defibrillator to detect and correct dangerous heart rhythms. There are external defibrillators like AEDs that can be used by laypeople, as well as internal defibrillators implanted in the body. The shock delivered must be of sufficient energy to depolarize enough heart muscle to terminate fibrillation. Proper use and troubleshooting of defibrillators is important for reviving someone experiencing cardiac arrest.
The document discusses defibrillation, which uses electric shocks to stop abnormal heart rhythms and allow a normal rhythm to resume. It defines defibrillation and describes the history and mechanisms involved. The types of defibrillators are explained, including automated external defibrillators. Precautions for defibrillation and troubleshooting defibrillators are also reviewed.
Defibrillation is a process that delivers an electric shock to the heart to stop an irregular heartbeat and restore a normal rhythm. It is commonly used to treat life-threatening cardiac arrhythmias like ventricular fibrillation. The document defines defibrillation and describes the history, principle, types of defibrillators including automated external defibrillators, precautions for use, and potential troubleshooting issues.
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, an automated external defibrillator (AED) should be used as soon as it is available, followed by CPR. For unwitnessed SCA, 5 cycles of CPR should be performed before attempting defibrillation. Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher efficacy than monophasic waveforms. A
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease rapidly over time without treatment. For witnessed SCA, survival rates decrease 7-10% per minute without CPR. The three key actions that must occur within the first moments of SCA are activating emergency services, beginning CPR, and using an automated external defibrillator. For out-of-hospital witnessed arrests, rescuers should use the AED as soon as it is available without performing CPR first.
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, the three key steps that must occur within the first few minutes are activating emergency services, beginning CPR, and using an automated external defibrillator (AED). Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher effectiveness than older monophasic devices. Lay rescuer AED programs have increased survival rates for out-of
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, the three key steps that must occur within the first few minutes are activating emergency services, beginning CPR, and using an automated external defibrillator (AED). Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher effectiveness than older monophasic devices. Lay rescuer AED programs have increased survival rates for out-of
Defibrillators are devices that deliver an electrical shock to the heart to terminate life-threatening cardiac arrhythmias like ventricular fibrillation. They were first demonstrated on dogs in 1899 and first used on a human in 1947. There are several types including manual external defibrillators, automated external defibrillators (AEDs) found in public places, implantable cardioverter-defibrillators, and wearable defibrillators. AEDs can analyze the heart rhythm and instruct the user, who needs no training, on whether a shock is needed. Defibrillation aims to stop all cardiac electrical activity so the heart's natural pacemaker can resume normal rhythm, while cardioversion delivers synchronized shocks
Defibrillation is a process that delivers an electric shock to the heart to stop ventricular fibrillation and restore normal rhythm. It involves using a defibrillator to detect and correct dangerous heart rhythms. There are external defibrillators like AEDs that can be used by laypeople, as well as internal defibrillators implanted in the body. The shock delivered must be of sufficient energy to depolarize enough heart muscle to terminate fibrillation. Proper use and troubleshooting of defibrillators is important for reviving someone experiencing cardiac arrest.
The document discusses defibrillation, which uses electric shocks to stop abnormal heart rhythms and allow a normal rhythm to resume. It defines defibrillation and describes the history and mechanisms involved. The types of defibrillators are explained, including automated external defibrillators. Precautions for defibrillation and troubleshooting defibrillators are also reviewed.
Defibrillation is a process that delivers an electric shock to the heart to stop an irregular heartbeat and restore a normal rhythm. It is commonly used to treat life-threatening cardiac arrhythmias like ventricular fibrillation. The document defines defibrillation and describes the history, principle, types of defibrillators including automated external defibrillators, precautions for use, and potential troubleshooting issues.
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, an automated external defibrillator (AED) should be used as soon as it is available, followed by CPR. For unwitnessed SCA, 5 cycles of CPR should be performed before attempting defibrillation. Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher efficacy than monophasic waveforms. A
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease rapidly over time without treatment. For witnessed SCA, survival rates decrease 7-10% per minute without CPR. The three key actions that must occur within the first moments of SCA are activating emergency services, beginning CPR, and using an automated external defibrillator. For out-of-hospital witnessed arrests, rescuers should use the AED as soon as it is available without performing CPR first.
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, the three key steps that must occur within the first few minutes are activating emergency services, beginning CPR, and using an automated external defibrillator (AED). Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher effectiveness than older monophasic devices. Lay rescuer AED programs have increased survival rates for out-of
Defibrillators are electrical devices that deliver a shock to the heart to counteract fibrillation and restore a normal heartbeat. Early defibrillation is critical for survival from sudden cardiac arrest (SCA) as the chances of successful resuscitation decrease by 7-10% per minute without CPR. For witnessed SCA, the three key steps that must occur within the first few minutes are activating emergency services, beginning CPR, and using an automated external defibrillator (AED). Modern defibrillators use biphasic waveforms which are as low as 200J and have equivalent or higher effectiveness than older monophasic devices. Lay rescuer AED programs have increased survival rates for out-of
"Clear!"
7
9.
Shout "Clear!" and press the shock button to deliver the shock.
10. Immediately resume CPR beginning with chest compressions for 2 minutes.
11. Check monitor for rhythm. If still in a shockable rhythm, repeat steps 7-10.
12. If rhythm converts to non-shockable, begin post-resuscitation care.
13. Document all interventions, time, rhythm, response to shocks.
To ensure safety of
caregivers and bystanders
To restore spontaneous
circulation
To determine if additional
shocks are needed
To provide appropriate
care based on patient's
rhythm
Defibrillation uses electrical shocks to restore a normal heart rhythm. It is used for ventricular fibrillation and asystole. Biphasic defibrillators are preferred over monophasic as they cause less damage and have higher success rates. Defibrillators include automated external defibrillators for public use, semi-automated defibrillators for paramedics, and implantable defibrillators. Adhesive patches are now commonly used instead of paddles. Defibrillation procedures involve assessing rhythm, applying pads or paddles, delivering shock, and resuming CPR if needed. Causes of failure include patient condition, prolonged arrest, inadequate CPR, and technical issues.
This document discusses defibrillation and cardioversion. It defines defibrillation as treatment for life-threatening arrhythmias without a pulse using electrical shock, while cardioversion aims to convert arrhythmias to normal rhythm with or without a pulse. Both use electrical energy to allow normal sinus rhythm. Defibrillation is for immediate use in ventricular fibrillation or pulseless ventricular tachycardia, while cardioversion may be used for unstable or failed chemical cardioversion of atrial fibrillation, atrial flutter, ventricular tachycardia with a pulse. The document reviews the history of defibrillation and types of defibrillators, and provides guidance on defibrillation and cardioversion procedures and considerations.
A brief overview of defibrillator,its physical principles, types, its indications & contraindications and maintenance policy.this powerpoint is primarily intended for anaesthesiologists and other health care providers working in critical care centres.
Electrical Therapies in CPR discusses defibrillation and cardioversion as electrical therapies used in emergency situations. Defibrillation involves delivering an electrical shock to the heart to convert ventricular fibrillation or ventricular tachycardia back to normal rhythm, while cardioversion is used to convert abnormal rhythms like atrial fibrillation back to normal sinus rhythm. Biphasic waveforms are now most commonly used because they require less energy than monophasic waveforms. Proper electrode placement on the chest is important for effective defibrillation. Potential complications include skin burns, muscle injuries, or myocardial injuries from high-energy shocks. Immediate CPR and defibrillation within 3 minutes of ventricular fibrillation
Defibrillation is a process that delivers an electric shock to stop an irregular heartbeat and restore normal rhythm. Early defibrillation within the first few minutes of cardiac arrest is crucial for survival. Pioneers like Claude Beck developed the defibrillator and helped revive heart attack victims. Proper defibrillator use requires correct pad placement and force, as well as ensuring safety from fires and electrical hazards. The delivery of the shock must be coordinated to avoid potential dangers and maximize effectiveness.
- The document discusses biomedical instrumentation and focuses on cardiac pacemakers and defibrillators.
- It describes how pacemakers use electrical impulses to regulate abnormal heart rhythms by contracting heart muscles. Pacemakers can be temporary or permanent depending on the cardiac condition.
- Defibrillators deliver electric shocks to the heart in ventricular fibrillation which can be fatal if not corrected. The document discusses different types of defibrillators and how they function to reestablish normal heart rhythm.
Defibrillators are devices that deliver electric shocks to the heart to restore normal heart rhythm and prevent cardiac arrest. They can be external and operated manually, external and automated for public use, or internal and implanted. External defibrillators use adhesive electrode pads or paddles placed on the patient's chest to deliver shocks. Internal defibrillators have electrodes implanted directly on the heart. The electric shock depolarizes heart muscle cells to terminate arrhythmias so the natural pacemaker can restore normal rhythm. Proper placement of defibrillator electrodes is important for effectiveness of treatment.
Advanced cardiac life support, or advanced cardiovascular life support, often referred to by its acronym, "ACLS", refers to a set of clinical algorithms for the urgent treatment of cardiac arrest, stroke, myocardial infarction (also known as a heart attack), and other life-threatening cardiovascular emergencies.
Dr. Dharmendra Joshi provides an overview of defibrillation and cardioversion. Some key points include:
- Defibrillation involves delivering unsynchronized energy during any cardiac cycle phase to terminate arrhythmias like ventricular fibrillation. Cardioversion delivers synchronized energy to large QRS complexes.
- Biphasic waveforms are now preferred over monophasic as they provide effective defibrillation at lower energies, reducing risk of injury.
- Safety is paramount, with operators announcing charges and discharges to avoid contact with patient or equipment. Complications can include arrhythmias, burns, embolism and myocardial necrosis. Troubleshooting focuses on proper equipment connection and settings.
Advanced Cardiovascular Life Support (ACLS).pptxRebilHeiru2
discusses the basic and Advanced Life support according to the AHA guidelines.
ACLS, BLS, defibrillation and Advanced medications at Adama Hospital medical college ICU
Cardioversion and defibrillation both use electrical shocks to treat abnormal heart rhythms, but they differ in their timing, energy level, and use. Cardioversion delivers a lower energy shock synchronized to a specific part of the heartbeat to convert rhythms like atrial fibrillation. Defibrillation provides an unsynchronized higher energy shock to non-organized rhythms like ventricular fibrillation during cardiac arrest. Biphasic shocks are now preferred over monophasic shocks for greater success and safety. Different types of defibrillators include AEDs, professional devices, and implanted defibrillators.
A defibrillator delivers an electric shock to the heart to convert life-threatening abnormal heart rhythms called cardiac arrhythmias back to a normal rhythm. There are two main types - internal defibrillators that are implanted inside the body, and external defibrillators that are used on the outside of the body. A defibrillator works by using electric shocks delivered through electrode pads placed on the chest to depolarize a critical mass of the heart muscle, which terminates the arrhythmia and allows the heart's natural pacemaker to resume control of the heartbeat.
Synchronized cardioversion uses a sensor to deliver a low energy shock that is timed to the peak of the QRS complex. This avoids delivering a shock during cardiac repolarization. Common indications are atrial fibrillation, atrial flutter, and supraventricular tachycardias when medications fail. Unsynchronized defibrillation delivers a high energy shock as soon as the button is pressed, used for pulseless ventricular tachycardia or fibrillation or if synchronization fails in an unstable patient. Biphasic waveforms may be as effective as higher energy monophasic shocks.
Components of Pacemaker and ICDs - understanding the hardwareRaghu Kishore Galla
The document discusses the history and components of cardiac pacing and implantable cardioverter defibrillators (ICDs). It covers the evolution of cardiac pacing from the 1700s to modern devices. It describes the basic components of pacing systems including the pulse generator, leads, electrodes, and batteries. It explains the differences between single chamber, dual chamber, bipolar, and unipolar systems. It provides details on pacemaker functions, concepts of pacing and sensing, and battery chemistries used in pacemakers.
Renal function tests are important for identifying, diagnosing, monitoring renal dysfunction and disease. Key tests include urinalysis to detect abnormalities in urine appearance, composition and sediments. Glomerular filtration rate (GFR) is a key measure of renal function and can be estimated using creatinine clearance from plasma and urine creatinine levels or formulas. Other tests evaluate tubular function like handling of electrolytes, acids, proteins and determine if renal issues are pre-renal or intrinsic to the kidneys. Together these tests provide vital information about renal health and disease.
This document discusses hypoglycemic agents or anti-diabetic drugs used to treat diabetes mellitus. It describes diabetes as a metabolic syndrome caused by insulin deficiency and excess glucagon that can lead to acute manifestations like hyperglycemia and chronic complications affecting small blood vessels and arteries. It outlines the different types of diabetes and treatments including insulin therapies using rapid, short, intermediate and long-acting insulins. It also discusses oral hypoglycemic drugs for type 2 diabetes including sulfonylureas, biguanides, glitazones and alpha-glucosidase inhibitors providing details on their mechanisms, uses, and potential toxicities in treatment.
"Clear!"
7
9.
Shout "Clear!" and press the shock button to deliver the shock.
10. Immediately resume CPR beginning with chest compressions for 2 minutes.
11. Check monitor for rhythm. If still in a shockable rhythm, repeat steps 7-10.
12. If rhythm converts to non-shockable, begin post-resuscitation care.
13. Document all interventions, time, rhythm, response to shocks.
To ensure safety of
caregivers and bystanders
To restore spontaneous
circulation
To determine if additional
shocks are needed
To provide appropriate
care based on patient's
rhythm
Defibrillation uses electrical shocks to restore a normal heart rhythm. It is used for ventricular fibrillation and asystole. Biphasic defibrillators are preferred over monophasic as they cause less damage and have higher success rates. Defibrillators include automated external defibrillators for public use, semi-automated defibrillators for paramedics, and implantable defibrillators. Adhesive patches are now commonly used instead of paddles. Defibrillation procedures involve assessing rhythm, applying pads or paddles, delivering shock, and resuming CPR if needed. Causes of failure include patient condition, prolonged arrest, inadequate CPR, and technical issues.
This document discusses defibrillation and cardioversion. It defines defibrillation as treatment for life-threatening arrhythmias without a pulse using electrical shock, while cardioversion aims to convert arrhythmias to normal rhythm with or without a pulse. Both use electrical energy to allow normal sinus rhythm. Defibrillation is for immediate use in ventricular fibrillation or pulseless ventricular tachycardia, while cardioversion may be used for unstable or failed chemical cardioversion of atrial fibrillation, atrial flutter, ventricular tachycardia with a pulse. The document reviews the history of defibrillation and types of defibrillators, and provides guidance on defibrillation and cardioversion procedures and considerations.
A brief overview of defibrillator,its physical principles, types, its indications & contraindications and maintenance policy.this powerpoint is primarily intended for anaesthesiologists and other health care providers working in critical care centres.
Electrical Therapies in CPR discusses defibrillation and cardioversion as electrical therapies used in emergency situations. Defibrillation involves delivering an electrical shock to the heart to convert ventricular fibrillation or ventricular tachycardia back to normal rhythm, while cardioversion is used to convert abnormal rhythms like atrial fibrillation back to normal sinus rhythm. Biphasic waveforms are now most commonly used because they require less energy than monophasic waveforms. Proper electrode placement on the chest is important for effective defibrillation. Potential complications include skin burns, muscle injuries, or myocardial injuries from high-energy shocks. Immediate CPR and defibrillation within 3 minutes of ventricular fibrillation
Defibrillation is a process that delivers an electric shock to stop an irregular heartbeat and restore normal rhythm. Early defibrillation within the first few minutes of cardiac arrest is crucial for survival. Pioneers like Claude Beck developed the defibrillator and helped revive heart attack victims. Proper defibrillator use requires correct pad placement and force, as well as ensuring safety from fires and electrical hazards. The delivery of the shock must be coordinated to avoid potential dangers and maximize effectiveness.
- The document discusses biomedical instrumentation and focuses on cardiac pacemakers and defibrillators.
- It describes how pacemakers use electrical impulses to regulate abnormal heart rhythms by contracting heart muscles. Pacemakers can be temporary or permanent depending on the cardiac condition.
- Defibrillators deliver electric shocks to the heart in ventricular fibrillation which can be fatal if not corrected. The document discusses different types of defibrillators and how they function to reestablish normal heart rhythm.
Defibrillators are devices that deliver electric shocks to the heart to restore normal heart rhythm and prevent cardiac arrest. They can be external and operated manually, external and automated for public use, or internal and implanted. External defibrillators use adhesive electrode pads or paddles placed on the patient's chest to deliver shocks. Internal defibrillators have electrodes implanted directly on the heart. The electric shock depolarizes heart muscle cells to terminate arrhythmias so the natural pacemaker can restore normal rhythm. Proper placement of defibrillator electrodes is important for effectiveness of treatment.
Advanced cardiac life support, or advanced cardiovascular life support, often referred to by its acronym, "ACLS", refers to a set of clinical algorithms for the urgent treatment of cardiac arrest, stroke, myocardial infarction (also known as a heart attack), and other life-threatening cardiovascular emergencies.
Dr. Dharmendra Joshi provides an overview of defibrillation and cardioversion. Some key points include:
- Defibrillation involves delivering unsynchronized energy during any cardiac cycle phase to terminate arrhythmias like ventricular fibrillation. Cardioversion delivers synchronized energy to large QRS complexes.
- Biphasic waveforms are now preferred over monophasic as they provide effective defibrillation at lower energies, reducing risk of injury.
- Safety is paramount, with operators announcing charges and discharges to avoid contact with patient or equipment. Complications can include arrhythmias, burns, embolism and myocardial necrosis. Troubleshooting focuses on proper equipment connection and settings.
Advanced Cardiovascular Life Support (ACLS).pptxRebilHeiru2
discusses the basic and Advanced Life support according to the AHA guidelines.
ACLS, BLS, defibrillation and Advanced medications at Adama Hospital medical college ICU
Cardioversion and defibrillation both use electrical shocks to treat abnormal heart rhythms, but they differ in their timing, energy level, and use. Cardioversion delivers a lower energy shock synchronized to a specific part of the heartbeat to convert rhythms like atrial fibrillation. Defibrillation provides an unsynchronized higher energy shock to non-organized rhythms like ventricular fibrillation during cardiac arrest. Biphasic shocks are now preferred over monophasic shocks for greater success and safety. Different types of defibrillators include AEDs, professional devices, and implanted defibrillators.
A defibrillator delivers an electric shock to the heart to convert life-threatening abnormal heart rhythms called cardiac arrhythmias back to a normal rhythm. There are two main types - internal defibrillators that are implanted inside the body, and external defibrillators that are used on the outside of the body. A defibrillator works by using electric shocks delivered through electrode pads placed on the chest to depolarize a critical mass of the heart muscle, which terminates the arrhythmia and allows the heart's natural pacemaker to resume control of the heartbeat.
Synchronized cardioversion uses a sensor to deliver a low energy shock that is timed to the peak of the QRS complex. This avoids delivering a shock during cardiac repolarization. Common indications are atrial fibrillation, atrial flutter, and supraventricular tachycardias when medications fail. Unsynchronized defibrillation delivers a high energy shock as soon as the button is pressed, used for pulseless ventricular tachycardia or fibrillation or if synchronization fails in an unstable patient. Biphasic waveforms may be as effective as higher energy monophasic shocks.
Components of Pacemaker and ICDs - understanding the hardwareRaghu Kishore Galla
The document discusses the history and components of cardiac pacing and implantable cardioverter defibrillators (ICDs). It covers the evolution of cardiac pacing from the 1700s to modern devices. It describes the basic components of pacing systems including the pulse generator, leads, electrodes, and batteries. It explains the differences between single chamber, dual chamber, bipolar, and unipolar systems. It provides details on pacemaker functions, concepts of pacing and sensing, and battery chemistries used in pacemakers.
Renal function tests are important for identifying, diagnosing, monitoring renal dysfunction and disease. Key tests include urinalysis to detect abnormalities in urine appearance, composition and sediments. Glomerular filtration rate (GFR) is a key measure of renal function and can be estimated using creatinine clearance from plasma and urine creatinine levels or formulas. Other tests evaluate tubular function like handling of electrolytes, acids, proteins and determine if renal issues are pre-renal or intrinsic to the kidneys. Together these tests provide vital information about renal health and disease.
This document discusses hypoglycemic agents or anti-diabetic drugs used to treat diabetes mellitus. It describes diabetes as a metabolic syndrome caused by insulin deficiency and excess glucagon that can lead to acute manifestations like hyperglycemia and chronic complications affecting small blood vessels and arteries. It outlines the different types of diabetes and treatments including insulin therapies using rapid, short, intermediate and long-acting insulins. It also discusses oral hypoglycemic drugs for type 2 diabetes including sulfonylureas, biguanides, glitazones and alpha-glucosidase inhibitors providing details on their mechanisms, uses, and potential toxicities in treatment.
Echocardiography uses ultrasound to generate images of cardiac anatomy and function. The echocardiogram machine includes a transducer probe that transmits and receives ultrasound waves, and a display monitor. Different echocardiography modes exist, including M-mode for measuring dimensions, 2D/3D for morphology, and various Doppler modes for assessing blood flow velocity and direction. Trans-esophageal echocardiography uses a probe attached to an endoscope to obtain high resolution internal images of the heart. Echocardiography is used to evaluate conditions like thrombi, aneurysms, valves, and septal defects.
DISEASES OF THE THYROID GLAND NOVEMBER 2018.pptxKemi Adaramola
This document discusses diseases of the thyroid gland. It begins by describing the anatomy and physiology of the thyroid, including its location in the neck, histological structure, embryological development, and hormone production process. It then discusses the main thyroid diseases - thyrotoxicosis (Graves' disease, toxic nodular goiter, etc.) and hypothyroidism. For each condition, it outlines the etiology, clinical features, investigations, and treatment approaches. It also describes thyroid storm, a medical emergency representing severe thyrotoxicosis, and myxedema coma, a severe form of hypothyroidism.
This document discusses the management of diabetes mellitus in children beyond typical norms. It begins with an introduction to diabetes and classifications of types of diabetes including type 1, type 2, and maturity onset diabetes of the young (MODY). It then discusses epidemiology and the local experience with childhood diabetes at Ekiti State University Teaching Hospital. The remainder of the document focuses on treatment and management of type 1 diabetes in children, including insulin regimens, therapeutic goals, and sick day rules. It also briefly discusses type 2 diabetes in children and the differences between type 1 and type 2 diabetes.
PULMONARYHYPERTENSION IN HEART FAILURE.pptxKemi Adaramola
This patient presented with bilateral leg swelling, abdominal swelling, facial swelling, and exertional dyspnea over the past 2 years. She has a history of recurrent productive cough as a child treated with antibiotics. On examination she had clubbing, leg swelling up to the knees, enlarged liver, ascites, and signs of right heart failure. Investigations showed features consistent with constrictive pericarditis and severe pulmonary hypertension likely due to tuberculous pericarditis in the past. She was treated unsuccessfully for heart failure and later developed venous thrombosis, remaining in the hospital for over 50 days before leaving against medical advice due to financial constraints.
This document outlines the causes, clinical features, diagnosis, and management of an Addisonian crisis, which is a life-threatening adrenal insufficiency emergency. It may be triggered by stress, infection, trauma, or withdrawal from steroids. Symptoms include nausea, vomiting, abdominal pain, hypotension, and electrolyte imbalances. Treatment involves intravenous hydrocortisone, fluids, glucose, and identifying and treating any precipitating causes. Long term management consists of glucocorticoid and mineralocorticoid replacement therapy and patient education.
This document outlines an introduction to pulmonary hypertension including its epidemiology, etiology, pathogenesis, clinical features, treatment, and future directions. It defines pulmonary hypertension and notes the most common causes are lung diseases like COPD. In Nigeria, common causes include COPD, tuberculosis, connective tissue diseases, and sickle cell disease. The pathogenesis involves remodeling of the pulmonary vasculature from factors like endothelial dysfunction and an imbalance of vasoconstrictors and vasodilators. Over time, this can lead to right heart failure if the right ventricle can no longer compensate for the increased resistance.
This document discusses cardiovascular disease risk factors. It begins by introducing the topic and outlines the sections. The introduction notes that CVD is a leading cause of death worldwide. The epidemiology section describes the prevalence of CVD globally and in certain regions. The pathophysiology section explains the development of atherosclerosis. The traditional risk factors section lists established risks like hypertension, diabetes, and smoking. The document focuses on emerging risk factors, describing biomarkers like lipoprotein(a), apolipoprotein B, and homocysteine that can help identify risk beyond traditional factors. It discusses the evidence supporting these novel factors and their clinical implications.
This document provides an overview of the management of ascites. It discusses the epidemiology, etiology, pathophysiology, evaluation, treatment, and complications of ascites. Ascites is most often caused by portal hypertension from liver cirrhosis. Other causes include malignancy, infection, heart failure, and nephrotic syndrome. Evaluation involves diagnostic paracentesis and ascitic fluid analysis. Treatment depends on the underlying cause but typically involves dietary sodium restriction and diuretic medication. Complications include spontaneous bacterial peritonitis.
Bronchiectasis, lung abscess, and empyema are chronic lung infections that can result from complications of pneumonia. They often occur in people with underlying lung disease or immune disorders. Key features include recurrent chest infections, coughing, sputum production, and life-threatening complications like respiratory failure. Treatment involves identifying and addressing the underlying cause, airway clearance techniques, long-term antibiotics, and surgery in some cases.
This document discusses impetigo, a common bacterial skin infection caused by Staphylococcus aureus or Streptococcus pyogenes. It presents in two forms: non-bullous and bullous. Symptoms include lesions on the face or limbs that may crust over. While usually mild, impetigo can develop into ecthyma if untreated. Treatment involves cleaning the skin, removing crusts, and using topical or oral antibiotics like mupirocin, retapamulin, cloxacillin, or erythromycin. Complications may include regional lymphadenopathy or ecthyma, a deeper skin ulceration.
This document discusses acute leukaemias in adults, including acute myeloid leukemia (AML) and acute lymphoblastic leukemia (ALL). It covers the incidence, risk factors, pathogenesis, clinical features, classification, prognostic factors, treatment for induction and post-remission, and supportive care for AML and ALL. The prognosis for ALL in adults remains poor despite high initial response rates to therapy. Improved supportive care and bone marrow transplantation are utilized while awaiting new, better therapies for these aggressive malignancies.
This document provides an overview of blood components therapy, including their indications and guidelines for use. It discusses the various components that can be derived from whole blood, such as packed red blood cells, platelets, fresh frozen plasma, cryoprecipitate, and granulocytes. Storage conditions, shelf lives, and therapeutic doses are provided. The main reasons for transfusion in Africa are described as childhood malaria, hemoglobinopathies, obstetric bleeding, trauma, and certain surgical procedures. Contraindications and risks of transfusion-transmitted infections are also covered. The document emphasizes considering alternatives to transfusion and whether benefits outweigh risks in each clinical situation.
This document provides an overview of haemoglobinopathies with an emphasis on sickle cell anaemia (SCA). It describes the complications and types of crises seen in SCA, how to identify them, and the modalities for treating various crises in the local environment. The introduction defines qualitative and quantitative haemoglobin abnormalities. It then outlines SCA pathogenesis, epidemiology, complications including vaso-occlusive crisis, management of crises through treatment of pain and infections, and prevention of sickle cell crises.
This document discusses Lassa fever, an endemic viral hemorrhagic fever found in parts of West Africa including Nigeria. It was first identified in 1969 after an outbreak infected and killed healthcare workers. The virus is transmitted from rodents to humans and can also spread between humans. Symptoms include fever, headache and bleeding. Diagnosis involves ELISA, virus isolation or PCR. Treatment is supportive, though the antiviral ribavirin may help if given early. Prevention focuses on rodent control, barrier nursing and safe medical practices. Nosocomial outbreaks remain a risk where infection control is poor. The document provides extensive details on the virus, epidemiology, transmission, clinical features, treatment and control of Lassa fever.
This document provides an overview of pituitary disorders, including:
- The pituitary gland regulates other endocrine glands and is regulated by the hypothalamus. It has anterior and posterior lobes.
- Anterior pituitary disorders include hyposecretion (hypopituitarism), hypersecretion (adenomas like prolactinomas and Cushing's disease), and sella enlargement.
- Causes of hypopituitarism include invasion, infarction, infiltration, injury, infections, immunologic issues, being iatrogenic, or being idiopathic. Symptom onset is gradual and follows a sequence of hormone deficiencies.
- Hyperfunctioning disorders include prolactinomas, acromegaly/
1. Stroke is defined as an acute brain attack caused by a disruption of blood flow, leading to neurological dysfunction lasting more than 24 hours.
2. Stroke is a leading cause of long-term disability worldwide and a major risk factor is hypertension. Diagnosis involves assessing symptoms, risk factors, and imaging of the brain.
3. There are two main types of stroke - ischemia caused by blockage of a blood vessel and hemorrhage caused by bleeding within the brain. Imaging is needed to distinguish between types and guide treatment.
Primary adrenal insufficiency, also known as Addison's disease, is caused by destruction or dysfunction of the adrenal cortex resulting in deficiencies of glucocorticoids and mineralocorticoids. Thomas Addison first described the clinical presentation in 1855. It most commonly presents with hyperpigmentation, dizziness, weakness, and weight loss. Diagnosis involves tests showing a lack of response to ACTH stimulation and electrolyte abnormalities. Treatment is lifelong glucocorticoid and mineralocorticoid hormone replacement to prevent adrenal crises.
This document discusses alopecia (hair loss) and its diagnosis and treatment. It covers the three stages of hair growth, epidemiological data on prevalence in Nigeria, differential diagnosis of hair loss types including diffuse, focal, scarring and non-scarring forms. Causes of hair loss conditions like telogen effluvium are explored. Investigation methods such as hair pull, trichogram and biopsy are outlined. Treatment focuses on treating underlying causes, using steroids, anti-fungals, minoxidil or 5-alpha-reductase inhibitors depending on the condition.
Here is the updated list of Top Best Ayurvedic medicine for Gas and Indigestion and those are Gas-O-Go Syp for Dyspepsia | Lavizyme Syrup for Acidity | Yumzyme Hepatoprotective Capsules etc
These lecture slides, by Dr Sidra Arshad, offer a quick overview of the physiological basis of a normal electrocardiogram.
Learning objectives:
1. Define an electrocardiogram (ECG) and electrocardiography
2. Describe how dipoles generated by the heart produce the waveforms of the ECG
3. Describe the components of a normal electrocardiogram of a typical bipolar lead (limb II)
4. Differentiate between intervals and segments
5. Enlist some common indications for obtaining an ECG
6. Describe the flow of current around the heart during the cardiac cycle
7. Discuss the placement and polarity of the leads of electrocardiograph
8. Describe the normal electrocardiograms recorded from the limb leads and explain the physiological basis of the different records that are obtained
9. Define mean electrical vector (axis) of the heart and give the normal range
10. Define the mean QRS vector
11. Describe the axes of leads (hexagonal reference system)
12. Comprehend the vectorial analysis of the normal ECG
13. Determine the mean electrical axis of the ventricular QRS and appreciate the mean axis deviation
14. Explain the concepts of current of injury, J point, and their significance
Study Resources:
1. Chapter 11, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 9, Human Physiology - From Cells to Systems, Lauralee Sherwood, 9th edition
3. Chapter 29, Ganong’s Review of Medical Physiology, 26th edition
4. Electrocardiogram, StatPearls - https://www.ncbi.nlm.nih.gov/books/NBK549803/
5. ECG in Medical Practice by ABM Abdullah, 4th edition
6. Chapter 3, Cardiology Explained, https://www.ncbi.nlm.nih.gov/books/NBK2214/
7. ECG Basics, http://www.nataliescasebook.com/tag/e-c-g-basics
Integrating Ayurveda into Parkinson’s Management: A Holistic ApproachAyurveda ForAll
Explore the benefits of combining Ayurveda with conventional Parkinson's treatments. Learn how a holistic approach can manage symptoms, enhance well-being, and balance body energies. Discover the steps to safely integrate Ayurvedic practices into your Parkinson’s care plan, including expert guidance on diet, herbal remedies, and lifestyle modifications.
- Video recording of this lecture in English language: https://youtu.be/kqbnxVAZs-0
- Video recording of this lecture in Arabic language: https://youtu.be/SINlygW1Mpc
- Link to download the book free: https://nephrotube.blogspot.com/p/nephrotube-nephrology-books.html
- Link to NephroTube website: www.NephroTube.com
- Link to NephroTube social media accounts: https://nephrotube.blogspot.com/p/join-nephrotube-on-social-media.html
Histololgy of Female Reproductive System.pptxAyeshaZaid1
Dive into an in-depth exploration of the histological structure of female reproductive system with this comprehensive lecture. Presented by Dr. Ayesha Irfan, Assistant Professor of Anatomy, this presentation covers the Gross anatomy and functional histology of the female reproductive organs. Ideal for students, educators, and anyone interested in medical science, this lecture provides clear explanations, detailed diagrams, and valuable insights into female reproductive system. Enhance your knowledge and understanding of this essential aspect of human biology.
Promoting Wellbeing - Applied Social Psychology - Psychology SuperNotesPsychoTech Services
A proprietary approach developed by bringing together the best of learning theories from Psychology, design principles from the world of visualization, and pedagogical methods from over a decade of training experience, that enables you to: Learn better, faster!
Muktapishti is a traditional Ayurvedic preparation made from Shoditha Mukta (Purified Pearl), is believed to help regulate thyroid function and reduce symptoms of hyperthyroidism due to its cooling and balancing properties. Clinical evidence on its efficacy remains limited, necessitating further research to validate its therapeutic benefits.
Does Over-Masturbation Contribute to Chronic Prostatitis.pptxwalterHu5
In some case, your chronic prostatitis may be related to over-masturbation. Generally, natural medicine Diuretic and Anti-inflammatory Pill can help mee get a cure.
NVBDCP.pptx Nation vector borne disease control programSapna Thakur
NVBDCP was launched in 2003-2004 . Vector-Borne Disease: Disease that results from an infection transmitted to humans and other animals by blood-feeding arthropods, such as mosquitoes, ticks, and fleas. Examples of vector-borne diseases include Dengue fever, West Nile Virus, Lyme disease, and malaria.
Recomendações da OMS sobre cuidados maternos e neonatais para uma experiência pós-natal positiva.
Em consonância com os ODS – Objetivos do Desenvolvimento Sustentável e a Estratégia Global para a Saúde das Mulheres, Crianças e Adolescentes, e aplicando uma abordagem baseada nos direitos humanos, os esforços de cuidados pós-natais devem expandir-se para além da cobertura e da simples sobrevivência, de modo a incluir cuidados de qualidade.
Estas diretrizes visam melhorar a qualidade dos cuidados pós-natais essenciais e de rotina prestados às mulheres e aos recém-nascidos, com o objetivo final de melhorar a saúde e o bem-estar materno e neonatal.
Uma “experiência pós-natal positiva” é um resultado importante para todas as mulheres que dão à luz e para os seus recém-nascidos, estabelecendo as bases para a melhoria da saúde e do bem-estar a curto e longo prazo. Uma experiência pós-natal positiva é definida como aquela em que as mulheres, pessoas que gestam, os recém-nascidos, os casais, os pais, os cuidadores e as famílias recebem informação consistente, garantia e apoio de profissionais de saúde motivados; e onde um sistema de saúde flexível e com recursos reconheça as necessidades das mulheres e dos bebês e respeite o seu contexto cultural.
Estas diretrizes consolidadas apresentam algumas recomendações novas e já bem fundamentadas sobre cuidados pós-natais de rotina para mulheres e neonatos que recebem cuidados no pós-parto em unidades de saúde ou na comunidade, independentemente dos recursos disponíveis.
É fornecido um conjunto abrangente de recomendações para cuidados durante o período puerperal, com ênfase nos cuidados essenciais que todas as mulheres e recém-nascidos devem receber, e com a devida atenção à qualidade dos cuidados; isto é, a entrega e a experiência do cuidado recebido. Estas diretrizes atualizam e ampliam as recomendações da OMS de 2014 sobre cuidados pós-natais da mãe e do recém-nascido e complementam as atuais diretrizes da OMS sobre a gestão de complicações pós-natais.
O estabelecimento da amamentação e o manejo das principais intercorrências é contemplada.
Recomendamos muito.
Vamos discutir essas recomendações no nosso curso de pós-graduação em Aleitamento no Instituto Ciclos.
Esta publicação só está disponível em inglês até o momento.
Prof. Marcus Renato de Carvalho
www.agostodourado.com
Rasamanikya is a excellent preparation in the field of Rasashastra, it is used in various Kushtha Roga, Shwasa, Vicharchika, Bhagandara, Vatarakta, and Phiranga Roga. In this article Preparation& Comparative analytical profile for both Formulationon i.e Rasamanikya prepared by Kushmanda swarasa & Churnodhaka Shodita Haratala. The study aims to provide insights into the comparative efficacy and analytical aspects of these formulations for enhanced therapeutic outcomes.
2. Outline
• Introduction
• History of Defibrillators
• Principles and mechanisms of Defibrillators
• Types and classes of Defibrillators
• Automated external defibrillator
• Procedure for Defibrillation
• Safety precautions during Defibrillation
• Conclusion
3. INTRODUCTION
• Defibrillation is a process in which an electronic device sends an
electric shock to the heart to stop an extremely rapid, irregular
heartbeat and restore the normal heart rhythm.
• Electrical defibrillation is the only effective therapy for cardiac arrest
caused by life threatening cardiac dysrhythmias specifically
ventricular fibrillation (VF) or pulseless ventricular tachycardia (VT).
• The guidelines on cardiopulmonary resuscitation of the European
Resuscitation Council and American Heart Association (AHA) strongly
recommend attempting defibrillation with minimal delay in victims of
VF/VT cardiac arrest.
4. INTRODUCTION cont’d
• If defibrillation is delivered promptly, survival rates increases to as
high as 75%.
• The chances of a favorable outcome decline at a rate of about 10%
for each minute cardiac defibrillation is delayed.
• As this event occurs most often in the victim’s private home or in
public spaces away from healthcare facilities, the need for early
defibrillation has led to the development of automatic, portable
defibrillators.
5. HISTORY
• In Switzerland, 1899, Prevost and Batelli discovered that small electric shocks
could induce ventricular fibrillation in experimental animals and that larger
charges would reverse the condition by using AC and DC shocks.
• Wiggers repeated their work in the 1930s, which then prompted Claude Beck, a
surgeon in Cleveland, to attempt defibrillation in humans who developed VF
while undergoing thoracotomy
• His first success came in 1947 when VF developed in a 14 year old boy whose
chest was being closed after surgery for a funnel chest using a homemade AC
defibrillator, developed by Kouwenhoven, with electrodes placed directly on the
heart
6. HISTORY cont’d
• Kouwenhoven was also instrumental in the development of the
external defibrillator, which was first successfully employed by Paul
Zoll in a patient with recurrent VF and pulseless ventricular
tachycardia complicating sinoatrial disease
• Following this breakthrough, direct current defibrillators were
introduced into clinical practice around 1962 when it was
demonstrated that electrical countershock or cardioversion across the
closed chest could abolish other cardiac arrhythmias in addition to
ventricular fibrillation
7. • The first successful defibrillation outside hospital was reported by
Pantridge in 1967
• Later on, Diack et al. described the first clinical experience with an
AED. Subsequently, further studies provided solid evidence on the
potential role of these devices in the early defibrillation and survival.
8. DEFIBRILLATORS
• Defibrillators are devices that restore a normal heartbeat by sending
an electric pulse or shock to the heart.
• They are used to prevent or correct an arrhythmia. They are also used
to restore the hearts beating if the heart suddenly stops.
• They are performed immediately after identifying that the patient is
experiencing a cardiac emergency, has no pulse and is unresponsive.
• A defibrillator delivers a dose of electric current(often called a
counter-shock) to the heart.
9. DEFIBRILLATORS cont’d
• Defibrillation aims to depolarise most of the myocardium
simultaneously ending the dysrhythmia thereby allowing the natural
pacemaker tissue to resume control of the heart in sinus rhythm
• Depolarisation of a critical mass of myocardium is necessary and this
depends on the transmyocardial current flow (measured in Amperes)
rather than the energy of the delivered shock (measured in Joules).
• A heart which is in asystole(flatline) cannot be restarted by a
defibrillator but would be treated by cardiopulmonary resuscitation
10. DEFRIBILLATORS cont’d
• Different types of defibrillators work in different ways. Automated
external defibrillators(AED) which are in many public spaces were
developed to save lives of people experiencing cardiac arrest. Even
untrained bystanders can use these devices in an emergency.
• Other defibrillators can prevent sudden death among people who
have a high risk of a life-threatening arrhythmia. They include
implantable cardioverter defibrillators(ICD) which are surgically
placed inside your body and wearable cardioverter
defibrillators(WCD) which rest on the body.
11. VENTRICULAR FIBRILLATION
• Ventricular fibrillation (VF) is a serious cardiac emergency resulting
from asynchronous contraction of the heart muscles
• Due to ventricular fibrillation, there is an irregular rapid heart rhythm.
• The heart stops functioning as an effective pump and in the absence
of cardiac output, the myocardium becomes more ischaemic and
irreversible anoxic damage occurs within few minutes
12. VENTRICULAR FIBRILLATION cont’d
• Ventricular fibrillation can be converted into a more efficient rhythm by
applying a high energy shock to the heart.
• This sudden surge across the heart causes all muscle fibres to contract
simultaneously. Possibly the fibres may then respond to normal
physiological pace making pulses
• The probability of successful defribillation and subsequent survival to
hospital discharge is inversely related to the time interval between the
onset of VF and delivery of the first countershock.
• The chances of success declines by about 7-10% for each minute delay in
administering the shock
13. • In VF the electrocardiograph shows a bizarre, irregular waveform that
is apparently random in both frequency and amplitude.
• VF is sometimes classified as either coarse or fine, depending on the
amplitude of the complexes
• VF is the commonest initial rhythm leading to cardiac arrest,
particularly in patients with coronary heart disease.
• VF may be preceded by ventricular tachycardia and is seen in up to
80-90% of those patients dying suddenly outside hospital
17. Defibrillator shock waveform
• The effectiveness of a shock in terminating VF depends on the type of
shock waveform discharged by the defibrillator.
• Monophasic waveform. Defibrillators with this type of waveform
deliver current in one polarity.
• They can be further categorized by the rate at which the current pulse
decreases to zero.
• If the monophasic waveform falls to zero gradually, the term damped
sinusoidal is used. If the waveform falls instantaneously, the term
truncated exponential
18. • When using a defibrillator with a monomorphic waveform it is
recommended that the first shock should be at an energy level of
200J.
• Should this be unsuccessful, a second shock at the same energy level
may prove effective because the transthoracic impedance is reduced
by repeated shocks.
• If two shocks at 200 J are unsuccessful, the energy setting should be
increased to 360 J for the third and subsequent attempts
• It gives up to 360 to 400 joules due to which cardiac injury and burns
around the shock pad sites is increased
20. • Biphasic waveform. This type of waveform was developed later. The
delivered current flows in a positive direction for a specified time and
then reverses and flows in a negative direction for the remaining
duration of the electrical discharge
• With biphasic waveforms there is a lower defibrillation threshold
(DFT) that allows reductions of the energy levels administrated and
may cause less myocardial damage.
• It delivers two sequential lower energy shocks of 120-200joules
21. • A biphasic shock of 150J is commonly considered to be at least as
effective as a 200J monophasic shock.
• Defibrillators that deliver biphasic shocks are now in clinical use, and
considerable savings in size and weight result from the reduced
energy levels needed.
• Biphasic shocks have been widely employed in implantable
cardioverter defibrillators (ICDs) because their increased effectiveness
allows more shocks to be given for any particular battery size
22.
23. FACTORS AFFECTING A DEFRIBILLATOR
• Transmyocardial current flow
• Most defibrillators are energy-based, meaning that the device charges a capacitor
to a selected voltage and then delivers a pre-specified amount of energy in
joules.
• The amount of energy which arrives at the myocardium is dependent on the
selected voltage and the transthoracic impedance that is, the resistance to
current flow through the chest wall, lungs, and myocardium (which varies by
patient).
• The optimal shock energy is one that will achieve defibrillation successfully while
causing minimal electrical injury to the myocardium.
• Achieving an appropriate current flow will reduce the number of shocks required
and may limit further myocardial damage.
24. • Determinants of current flow
Energy of delivered shock
Transthoracic impedance
Electrode position
Shock waveform
Body size
Electrode size
25. • Transthoracic impedance
• In adults transthoracic impedance averages about 60Ohms, with 95%
of the population lying in the range of 30-90Ohms.
• Current flow will be highest when transthoracic impedance is at its
lowest.
• To achieve this the operator should press firmly when using handheld
electrode paddles. A conductive electrode gel or defibrillator pads
should be used to reduce the impedance at the electrode and skin
interface. Self-adhesive monitor or defibrillator electrodes do not
require additional pressure.
26. • In patients with considerable chest hair, poor electrode contact and
air trapping will increase the impedance. This can be avoided by
rapidly shaving the chest in the areas where the electrodes are
placed.
• Transthoracic impedance is about 9% lower when the lungs are
empty, so defibrillation is best carried out during the expiratory phase
of ventilation
• It is also important to avoid positioning the electrodes over the
breast tissue of female patients because this causes high impedance
to current flow.
27. • Determinants of transthoracic impedance
Shock energy
Electrode size
Electrical contact
Number of and time since previous shocks
Phase of ventilation
Distance between electrodes
Paddle or electrode pressure
28. Electrode position
• The ideal electrode position allows maximum current to flow through the
myocardium.
• This will occur when the heart lies in the direct path of the current
• The standard position consists of one electrode placed to the right of the
upper sternum below the right clavicle(anterio-lateral) and the other
placed in the mid-axillary line at the level of the fifth left intercostal
space(anterio-posterior)
• An alternative is to place one electrode to the left of the lower sternal
border(anterior left infrascapular) and the other on the posterior chest wall
below the angle of the left scapula(anterior right infrascapular)
• Avoid placing electrodes directly over breast tissue in women
29.
30. • Electrode size or surface area
Low transthoracic impedence is achieved with larger electrodes.
Above an optimum size the transmyocardial current will be reduced
The usual electrode sizes employed are 10-13cm in diameter for
adults and 4.5-8cm for infants and children
• Body size
Infants and children require shocks of lower energy than adults to
achieve defibrillation
Over the usual range of weight encountered in adults, body size does
not greatly influence the energy requirements
31. TYPES OF DEFRIBILLATORS
• Manual defibillators; external manual and internal manual
defibrillators
• Automated defibrillators; external or internal automated defibrillators
Internal defribillator. Electrodes are placed directly to the heart
External defribillator. Electrodes placed directly on the heart
32.
33. • Other types of defibrillators less frequently used in clinical practice:
Impedance-based defibrillators allow selection of the current
applied based upon the transthoracic impedance (TTI). TTI is assessed
initially with a test pulse and subsequently the capacitor charges to
the appropriate voltage.
Current-based defibrillators deliver a fixed dose of current which
results in defibrillation thresholds that are independent of TTI . The
optimal current for ventricular defibrillation appears to be 30 to 40
amperes independently of both TTI and body weight thus achieving
defibrillation with considerably less energy than the conventional
energy-based method
34. Cardioversion
• Cardioversion is one of the possible treatments for arrhythmias that
imply a re-entrant circuit.
• By delivering a synchronized electric shock all excitable tissue of the
circuit is simultaneously depolarised making the tissue refractory and
the circuit no longer able to sustain re-entry.
• As a result, cardioversion terminates arrhythmias resulting from a
single reentrant circuit, such as atrial flutter, atrioventricular nodal
reentrant tachycardia or monomorphic ventricular tachycardia.
35. • Current European Society of Cardiology and AHA guidelines suggest the
following initial energy selection for specific arrhythmias
• For atrial fibrillation, 120 to 200 joules for biphasic devices and 200 joules
for monophasic devices.
• For atrial flutter, 50 to 100 joules for biphasic devices and 100 joules for
monophasic devices.
• For ventricular tachycardia with a pulse, 100 joules for biphasic devices and
200 joules for monophasic devices.
• For ventricular fibrillation or pulseless ventricular tachycardia, at least 150
joules for biphasic devices and 360 joules for monophasic devices.
37. Automated external defibrillator
• The term refers to a portable and lightweight computerized device that
incorporates rhythm analysis and defibrillation systems and uses voice
and/or visual prompts to guide lay rescuers and healthcare providers to
safely defibrillate victims of cardiac arrest due to VF or pulseless VT.
• There are two types of AED: the semi-automatic that indicates the need for
defibrillation but requires that the operator deliver the shock by pushing a
button and the fully automatic AED which is capable of administering a
shock without the need for outside interventions.
• Basically these devices consist of a battery, a capacitor, electrodes and an
electrical circuit designed to analyze the rhythm and send an electric shock
if is needed.
38. Components of an AED
Basically these devices consist of a battery, a capacitor,
electrodes and an electrical circuit designed to analyze
the rhythm and send an electric shock if is needed
39. • Batteries. Essentially they are containers of chemical reactions and
one of the most important parts of the AED system.
• Initially lead batteries and nickel-cadmium were used but lately non-
rechargeable lithium batteries, smaller in size and with longer
duration without maintenance (up to 5 years), are rapidly replacing
them.
• Since extreme temperatures negatively affect the batteries,
defibrillators must be stored in controlled environments
40. • Capacitor. The electrical shock delivered to the patient is generated
by high voltage circuits from energy stored in a capacitor which can
hold up to 7 kV of electricity.
• The energy delivered by this system can be anywhere from 30 to 400
joules.
• Electrodes are the components through which the defibrillator
collects information for rhythm analysis and delivers energy to the
patient's heart.
• Many types of electrodes are available including hand-held paddles,
internal paddles, and self-adhesive disposable electrodes.
41. • Electrical circuit. AEDs are highly sophisticated, microprocessor-based
devices that analyze multiple features of the surface ECG signal
• Controls. The typical controls on an AED include a power button, a
display screen on which trained rescuers can check the heart rhythm
and a discharge button.
• Defibrillators that can be operated manually have also an energy
select control and a charge button including frequency, amplitude,
slope and wave morphology
42. Semi-automatic AEDs Fully automatic AED
Definition Indicates the need for defibrillation
but requires an operator to deliver
the shock by pushing a button
Capable of administering a shock
without the need for outside
interventions
Advantages • Recommended by current
resuscitation guidelines
• Widely used
• Allows healthcare professionals
to override the device and deliver a
shock manually, independently of
prompts.
• Safer, no risk of inappropriate
shocks to the rescuer
• Easier to use and more
appropriate for lay-rescuers
• Better compliance with
resuscitation protocols
Disadvantages • More complex to use for the
untrained responders
• More difficult to synchronize with
CPR maneuvers for lay rescuers
• Longer times until shock delivery
• Risk of electrocution for the
rescuer if inappropriately used
• No possibility to override the
device
• Not recommended by current
43. Semi-automatic AEDs Fully automatic AED
Definition
Indicates the need for defibrillation but requires an operator to
deliver the shock by pushing a button
Capable of administering a shock without the need for outside interventions
Advantages
• Recommended by current resuscitation guidelines
• Widely used
• Allows healthcare professionals to override the device and deliver a
shock manually, independently of prompts.
• Safer, no risk of inappropriate shocks to the rescuer
• Easier to use and more appropriate for lay-rescuers
• Better compliance with resuscitation protocols
Disadvantages
• More complex to use for the untrained responders
• More difficult to synchronize with CPR maneuvers for lay rescuers
• Longer times until shock delivery
• Risk of electrocution for the rescuer if inappropriately used
• No possibility to override the device
• Not recommended by current guidelines except for special situations
44. Sequence of action for AED
• Once cardiac arrest has been confirmed it may be necessary for an
assistant to perform basic life support while the equipment is
prepared and the adhesive electrodes are attached to the patient’s
chest
• Once the AED is ready to use, the following sequence should be used:
45.
46. • Repeat as directed for up to three shocks in any one sequence. Do not
check for a pulse or other signs of a circulation between the three
shocks.
• If no pulse or other sign of a circulation is found, perform CPR for one
minute. This will be timed by the machine, after which it will prompt
the operator to reanalyse the rhythm.
• Alternatively, this procedure may start automatically, depending on
the machine’s individual features or settings. Shocks should be
repeated as indicated by the AED.
47. • If a circulation returns after a shock, check for breathing and
continue to support the patient by rescue breathing if required.
• Check the patient every minute to ensure that signs of a circulation
are still present.
• If the patient shows signs of recovery, place in the recovery position.
48. • Safety factors
• All removable metal objects, such as chains and medallions, should
be removed from the shock pathway—that is, from the front of the
chest.
• Body jewellery that cannot be removed will need to be left in place.
Although this may cause some minor skin burns in the immediate
area, this risk has to be balanced against the delay involved in its
removal
• Clothing should be open or cut to allow access to the patient’s bare
frontal chest area
49. • The patient’s chest should be checked for the presence of self-
medication patches on the front of the chest (these may deflect
energy away from the heart)
• Oxygen that is being used—for example, with a pocket mask—should
be directed away from the patient or turned off during defibrillation
• The environment should be checked for pools of water or metal
surfaces that connect the patient to the operator. It is important to
recognise that volatile atmospheres, such as petrol or aviation fumes,
can ignite with a spark
50. Wearable cardioverter defibrillator(WCD)
• The concept of WCD consists of longterm monitoring, detection of sudden
cardiac death and shock delivery without bystander assistance or an
implanted device to bridge an assessment period or to let optimal medical
therapy deliver its benefits
• The sensing and therapy delivery component consists of 1 anterior and 2
posterior self gelling defibrillation electrodes held together by an elastic
chest garment
• Dry tantalum oxide electrodes provide long term ECG monitoring through 2
non-standard leads (anteroposterior and left-right bipolar signals) whereas
the defibrillation electrodes contain a vibration plate and multiple gel
capsules. The vibration plate is intended to give the patient a tactile
warning of an impending shock once a shockable rhythm detection occurs
51. • The defibrillation gel is released to minimize skin-pad impedance and
prevent skin injury during shock delivery
• When the patient receives tactile, audible and visual alerts, the
therapeutic shock can be aborted by simultaneously pressing 2
buttons
• The WCD is able to deliver shock up to 150j, biphasic with a
programmable response time of 25-180s
• Reported rates of inappropriate therapies range from 0.4-0.5 with fast
supraventricular tachycardia and artifacts as the most common
underlying cause
52. • WCD can be considered in adult patients who present a high
arrhythmic risk for a limited period of time such as reduced LVEF, as a
bridge to heart transplantation or left ventricular assist devices, in the
40days after myocardial infarction or in the 3months after a coronary
artery bypass graft
• WCD can be considered when a transient contraindication to icd is
present such as endocarditis or device related infection
53. Conclusion
• Sudden cardiac arrest, frequently due to VF or pulseless VT, is
traditionally associated with poor survival rates.
• Saving the lives of these patients depends on early cardiac
defibrillation which, with manual defibrillators, is limited only to
qualified rescuers who can interpret ECGs.
• AEDs solve this problem since they are able to analyze rhythm and
inform the rescuers whether a shock is indicated
54. • This approach allows lay rescuers to provide effective early
defibrillation which has been shown to significantly improve survival
and survival with intact neurologic function after out-of-hospital
cardiac arrest.
• One limitation is that AED use requires interruptions in CPR which
was proved to be deleterious especially in patients with non-
shockable rhythms.
• Special efforts are being made in order to improve rhythm analysis
and ‘hands-off’ time during CPR.