2. HIPPA Compliance
Our image is HIPPA
compliant
Nowhere on our image does
it violate any patient
confidentiality
3. Marker & Patient ID
The correct anatomical side
marker is visible in the
image.
The side marker is placed
correctly with the right
marker displaying on the
viewers left.
There are no markers
superimposed on any
pertinent anatomy.
4. Marker & Patient ID
There were no additional
side markers used but an
arrow marker could have
been used to indicate sight
of pain/injury etc.
The image is displayed
correctly based on marker
placement.
5. Radiation Hygiene
Acceptable beam restriction
requires at least 3 or more sides
of collimation. There is no
evidence of beam restriction
located on our image.
Shielding must be provided if
the gonads are within 5 cm of
the primary beam. There is no
evidence of primary shielding
on the side closest to the
gonads.
It is always a good habit to
place a lead shield on any
patient of child bearing age.
6. Completeness of Position/Projection
Routine positions/projections
for a hand include:
PA
PA Oblique 45 degree rotation
Left fan lateral
This is a PA hand which
complies with the routine
projections/positions.
All anatomical parts are
correctly visualized on the
image.
7. Artifact Identification
There does not appear
that there are any of the
following in our image:
Physical artifacts
Body parts superimposed
Hospital paraphernalia
Patient clothing/belongings
Indwelling artifacts/foreign
bodies
Our patient does have a
fracture of the 5th
metacarpal.
8. Artifact Identification
There does not appear
that excess fog is visible
and/or degrading overall
image quality.
There also does not
appear that there are
any CR/DR artifacts
visible in the image.
9. Image Sharpness
There does not appear that
any of the following are in
the image:
Gross voluntary motion
Excessive quantum
mottle/image noise
Evidence of double
exposure
Grid lines/cut-off, or
artifacts
When imaging a hand, it is
unnecessary to use a grid.
10. Image Sharpness
Size distortion does not
appear greater than
expected because when
imaging a hand, the hand lies
directly on the IR.
The CR should be entering at
the third MCP joint. Because
the CR is more than 1 cm off-
centered to the part, shape
distortion is being caused but
only to a minimal.
11. Accurate Part Positioning
The part is adequately aligned to
the image media.
The part is not accurately centered.
The part would have to be moved
about an inch to the left.
The CR is not centered within 1 cm
of the anatomical part.
The CR is aligned to the imaging
media.
The CR does conform to an
accepted IR exposure field
recognition template.
12. Accurate Part Positioning
According to Merrill’s Atlas or Radiographic Positioning and
Procedures for a PA hand state:
Rest the patient’s forearm on the table and place the hand with
the palmar surface down on the IR.
Center the IR to the MCP joints, and adjust the long axis of the
IR with the long axis of the hand and forearm.
Spread fingers slightly.
Shield gonads.
Ask patient to relax hand
to avoid motion.
13. Evaluation Criteria
Evidence of proper collimation
No rotation of hand
Open MCP and IP joints,
indicating hand is placed flat on
the IR
Slightly separate digits with no
soft tissue overlap
All anatomy distal to the radius
and ulna
Soft tissue and bony
trabeculation
14. Judicious ExposureTechnique
Based on criteria, the
anatomical part is correctly
positioned.
The most radiolucent
structure visible in the image
is the soft tissue of the hand.
The most radiopaque
structure that is visible on
the image is the bony cortex
of the hand.
15. Judicious ExposureTechnique
I think the image displays
adequate contrast by
displaying short scale
contrast.
The brightness of our image
is adequate as well.
Even though we do not know
our EI value, I would say our
EI value was most likely
within normal range.
16. Accept or Reject?
Based on the evaluation
criteria, I would ACCEPT this
image.
However, for next time:
I would make sure I
increased my beam
restriction especially on the
side closest to the gonads.
Also appropriately center
the CR to correct
anatomical part.
17. Sources
Frank, Eugene D., Bruce W. Long, Barbara J. Smith,
Vinita Merrill, and Philip W. Ballinger. Merrill's Atlas
of Radiographic Positioning & Procedures. 12th ed.
Vol. 1. St. Louis, MO: Mosby/Elsevier, 2007. Print.