SlideShare a Scribd company logo
1 of 54
FOLLOW UP AND MANAGEMENT OF
PACEMAKER PROGRAMMING AND
TROUBLESHOOTING
SOUMYA KANTI DUTTA.
 Pacemaker components
 Battery(Lithium iodide or Vanadium silver)
 Pacing impedance
 Pulse generator
1. Output circuit
2. Sensing circuit
3. Timing circuit
4. Rate adaptive sensor
5. Modes and mode switching
 Pacing impedance :
 Pacing impedance refers to the opposition to
current flow.
 Sources that contribute to pacing impedance:
1. Pacing lead conductor coil
2. Electrode-tissue interface
 Normal lead impedance vary from 250-
1200ohms.
 Single impedance value may be of little use
with out previous values for comparison.
Pulse generator output circuit
 Capture threshold or Pacing threshold .
 Minimum amount of energy required to constantly
cause depolarization
 Voltage(mA) and pulse duration(mSec)
 V=IR
2.0 v 1.5 v 1 v
1. Pulse generator output circuit
1. Pulse generator output
circuit
 High Pacemaker Output can
cause
◦ Reduce longevity
◦ Diaphragmatic stimulation
◦ Muscle Sti. in Unipolar
pacemakers
◦ Patient may “feel” heart
beat
Electrical Testing Of Pacemaker
Electrical Testing Of Pacemaker
2. Pulse generator sensing circuit :
◦ Ability of the device to detect intrinsic beat of
the heart
◦ Measured - peak to peak magnitude (mv) &
slew rate(mv/ms)
Pulse generator timing circuit :
a. Lower rate limit (LRL)
b. Hysteresis rate
c. Refractory and blanking periods
d. Ventricular safety pacing interval .
e. Upper rate response .
Electrical Testing Of Pacemaker
Pulse generator timing circuit :
 Lower rate interval - lowest rate that the pacemaker
will pace .
 Escape interval:sensed beat to next paced beat.
 A paced or non-refractory sensed event restarts the rate
timer at the programmed rate.
Condition LRL (beats/mt)
Infrequent pauses 40-50
Chronic persistent bradycardia 60-70
Relative bradycardia
detrimental (long QT)
70-80
Detrimental fast heart rates
(angina)
50-60
VVI 60-70
Pulse generator timing circuit :
Pulse generator timing circuit :
 Hysteresis :
 Hysteresis allows the rate to drop below the programmed
pacing LRL.
 Advantages of hysteresis :
1. Encourages native rhythm – maintain AV sync in VVI ,
prolong battery life
2. Prevent retrograde conduction – avoids pacemaker
syndrome
 AV delay (AVI) – pacemaker equivalent of PR interval.
 Sensed vs paced AVI – paced AVI is programmed at 125-
200ms , sensed AV interval is programmed at 20-50ms
shorter than paced.
 Dynamic/adaptive AV delay :shorter AVI with increase in
heart rate,allow pacemaker to respond to exercise and
have better haemodynamics.
Pulse generator timing circuit :
sAVI – 150ms
pAVI – 200ms
 AV delay (AVI)
 Longer AVI :
Good AV conduction – maintains AV synchrony ,
long battery life
Achieved by following methods :Programming
longer AVI , managed ventricular pacing , AV delay
hysteresis .
 Shorter AVI:
HOCM – RV apical pacing decreases HOCM gradient
CRT – usually 80-120ms , for 100%ventricular pacing
and optimize CO
Physiological response to faster heart rates can be
answered
Pulse generator timing circuit :
 Refractory and
blanking
periods :
 Refractory
period –
sensing present
but no action
 Blanking period
- sensing
absent and
hence no action
Pulse generator timing circuit :
Blanking
period
Time Importance
Atrial blanking
period
50-100ms Non programmable ,
Avoid atrial sensing of its own
paced beat
Post ventricular
atrial blanking
period
220ms Avoid sensing of ventricular
beat
Long PVAB decreases
detection of AF,AFL
Ventricular
blanking period
50-100ms Non programmable,
Avoid ventricular sensing of
its paced beat
Post atrial
ventricular
blanking period
28ms if the PAvB period is too long,
R on T - ventricular
tachyarrhythmia.
Refractory period Importance
Ventricular refractory period
(VRP)
Prevent sensing of T wave
.
Atrial refractory period
(ARP)
AVI (120-200ms) .
Post ventricular atrial
refractory period
Avoid sensing retrograde P
waves (PMT) , far field R
waves .
Pulse generator timing circuit :
Pulse generator timing circuit :
Ventricular safety pacing/ventricular
triggered period/cross talk sensing
window :
Atrial pacing in DDD
Trigger ventricular sensing
PAVB -pAVI
False inhibition of
ventricular pacing circuit
Asystole
 Rate responsive pacing refer to ability of pacemaker
to increase its lower rate in response to physiological
stimulus
 Upper rate limit:important to prevent tracking of
rapid atrial activity like in AF
 HRR should start with in 10s of exercise , peak at 90
– 120s and should return to baseline with in 60 –
120s after exercise.
 If intrinsic atrial rate exceeds URR then wenckebach
or 2:1 AVB
 Choosing URR : young patients (150b/mt) , old
angina (<110b/mt).
 Various sensors (activity , minute ventilation , QT)
 Rate smoothing: allows variation of R-R interval <6%
Pulse generator rate responsive pacing:
Pulse generator rate responsive pacing:
Wenckebach
2:1 AVB
PACING MODES
 BEPG(British pacing and electrophysiology
group)
 NASPE(North American society of pacing
and electrophysiology)
 DD I/T R
 III-D(I+R) :dual tracking of atrial activity while
inhibited by ventricular activity
 5th position: antitachycardia function(pacing/shock)
5. Pulse generator modes switching:
Electrical Testing Of Pacemaker
DDD / VDD
Atial tachyarrythmias
Sensed atrial events
Trigger fast ventricular rates
Palpitations. Dyspnoea. And Fatigue.
DDIR /
VVIR
Pulse generator modes switching:
Programming mode switching Mode switching
occurs when the sensed atrial rate exceeds a programmed
atrial tachycardia detection rate. By definition, this value
must be faster than the URL (maximum tracking rate).
Atrial tachycardia detection is typically programmed to
175-l88bpm or thereabouts.AMS base rate is higher than
LRI.
Electrical Testing Of Pacemaker
Pacemaker follow up guidelines:
Single chambered
pacing
Dual chambered
pacing
1st month q 2 week 1st month q 2 weeks
2nd -48th month q 12
week
2nd – 30th month q
12weeks
49th – 72nd month q 8
week
31st – 48th month q 8
weeks
73rd month and later q 4
weeks
49th month and later q
4week
Pacemaker complications
Pocket
complications
Pocket hematoma
Infection
Erosion
Wound pain
Allergic reactions
Pacemaker
complications
Lead dislodgement
Pneumothorax /air
embolism
Cardiac perforation
Extracardiac stimulation
Venous thrombosis
Coronary sinus
dissection
Twidller syndrome
Pacemaker malfunction
Pocket hematoma :
 The risk of haematoma is increased in patients
taking antithrombotic or anticoagulant drugs
(Goldstein et al., 1998).
 Most small hematomas can be managed
conservatively with cold compress and withdrawal
of antiplatelet or antithrombotic agents.
 Occasionally, large hematomas that compromise
the suture line or skin integrity may have to be
surgically evacuated.
 Needle aspiration increases risk of infection and
should not be done.
Pocket hematoma :
 In patients requiring oral anticoagulants (warfarin), to
take INR of about 2.0 at the time of implantation is
safe (Belott & Reynolds, 2000).
 Unfractionated heparin or low-molecular-weight
heparin are always discontinued prior to device implant
and ideally avoided for a minimum of 24 hours post
implantation.
 Administration of anticoagulants can be resumed
within 48-72 h after implantation if there is no
evidence of substantial hematoma formation.
Device-related infections :
 The reported incidence of pacemaker-related
infection ranges from 0.5% to 6% in early series.
 The use of prophylactic antibiotics has decreased
the rate of acute infections following pacemaker
implantations to <1 to 2 percent in most series.
 The mortality of persistent infection when infected
leads are not removed can be as high as 66%.
 DM, malignancy, operator inexperience, advanced
age, corticosteroid use, anticoagulation, recent
device manipulation, CRF, and bacteremia from a
distant focus of infection.
Device related infection :
2007;49;1851-1859 J. Am. Co
Twiddler syndrome:
Obese women with loose, fatty subcutaneous tissue
Small size of the implanted generator with a large pocket
Twisting of pulse generator in long axis
Lead dislodgement and lead fracture
Failure to capture
 The prevelance of this
syndrome is 0.07% (Gungor
et al., 2009)
 Rotated along the transverse
axis it is referred by us as
the reel syndrome.
 Pocket should be revised.
 Avoid by
◦ Limit the pocket size,
◦ Suture the device to the
fascia
◦ The patients not to
manipulate their device
pocket
Twiddler syndrome:
 Failure to capture
 Failure to output
 Sensing abnormalities(under and over
sensing)
 Specific mode complications
1. Pacemaker related tachycardia
2. Pacemaker syndrome
PACEMAKER MALFUNCTION
Failure to capture:
 Pacing artifact present but no evoked potential .
 Causes
1. Lead dislodgement or perforation
2. Lead maturation(inflammation/fibrosis)(exit
block)
3. Battery depletion
4. Circuit failure(coil fracture , insulation defect)
5. Inappropriate programming(low output /pulse
width)
6. Pseudo malfunction(pacing during refractory
period)
7. Functional non capture(oversensing).
8. Metabolic , drugs , cardiomyopathies
Medication effect on Capture
Drugs that increase
capture threshold
Drugs that decrease
capture threshold
Amiodarone
Flecainide
Propafenone
Sotalol
Procainamide
lidocaine
Atropine
Epinephrine
Isoproterenol
corticosteroids
Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular
pacing artifacts but one failed to result in ventricular depolarization—
that is, failure to capture
Failure to capture:
Failure to capture:
Pacing threshold
Normal Increased
Battery depletion
Functional non
capture
Impedance
Normal
Dislodgement
Exit block
Decreased
Insulation
failure/
break
Increased
Lead
fracture
Loose screw
Failure to output:
 Absence of pacing stimuli and hence no capture .
 Causes
1. Pseudo malfunction - hysteresis , PMT
termination , sleep rate
2. Over sensing - EMI ; T P R over sensing ;
Myopotential/diaphragmatic ; Cross talk .
3. Open circuit - lead fracture , loose screw ,
incompatible lead .
4. Battery depletion
5. Recording artifact.
 Failure to output:
VVIR pacemaker This patient had a pacemaker programmed to a
unipolar sensing configuration. The sensing of myopotentials led to
symptomatic pauses, and reprogramming the pacemaker to a
bipolar sensing configuration prevented subsequent myopotential
over sensing.
Application of magnet
Failure to output:
Eliminates pauses Pauses persistent
Impedance
Normal
Decreased
Insulation
failure/break
Increased
Lead fracture
Loose screw
Battery
depletionOver sensing
Pseudo malfunction
Battery depletion :
 Elective replacement indicators (ERI)
1. Low voltage(2.1-2.4V)
2. Low pacing rate on magnet application
3. Elevated battery impedance
4. Increased pulse width duration
5. Restricted programmability
6. Change to simpler pacing mode
 End of life (EOL)
1. Low voltage(≤2.1vol)
Pacemaker undersensing :
 Pulse artifact or intrinsic complex present but no
sensing(sensed beat doesn’t reset escape interval or
cycle).
 Causes are
1. Defect in signal production – scar /fibrosis
following MI , ectopic , cardioversion,defibrillation ,
metabolic.
2. Defect in signal transmission – lead fracture
/dislodgement , insulation failure , partial open
circuit.
3. Defect in pacemaker – battery depletion , sensing
circuit abnormalities , committed DVI.
Pacemaker over sensing :
 Inappropriate inhibition as
failure to pace
 Causes
1. EMI
2. T , P , R over sensing .
3. Far field R wave
sensing(atrial).
4. Cross talk
5. Myopotential
(unipolar)
6. Make break signals
Cross talk :
High atrial output
High ventricular sensitivity
Low VBP
Ventricular sensing of paced
atrial impulse
Ventricular Asystole
Pacemaker Syndrome
“An assortment of symptoms
related to the adverse
hemodynamic impact from the
loss of AV synchrony.”
Pacemaker Syndrome
Symptoms include:
 Dizziness
 Presyncope
 Chest tightness
 Shortness of breath
 Neck pulsations(cannon A wave)
 Apprehension/malaise
 Fatigue
Pacemaker Syndrome May Be Caused By:
 Loss of capture, sensing
 A-V intervals of long duration
 Onset of 2:1 block
 Single chamber system(VVI)
 Absence of rate increase with exercise
 ECG:Retrogradely conducted inverted P over T
 CORRECTIVE ACTION:
◦ Reduce pacing rate.
◦ Hysteresis on to allow more sinus rhythm
◦ Shift to dual chamber pacing.
Pacemaker-Mediated Tachycardia
(PMT)
 A rapid paced rhythm that can occur
with atrial tracking pacemakers
PMT is the Result of:
 Also known as pacemaker re-entry
tachycardia or endless-loop tachycardia
(like other re-entry dysrhythmias,
pacemaker is now a part of the re-entry
circuit,Mostly anterograde limb of
reentrant tachycardia)
 Retrograde conduction
 PVARP
 Tracking fast atrial rates (physiologic or
non-physiologic)
 only in patients with dual chamber devices
 The ventricular impulse is retrogradely
conducted via the AV node to the atria, now the
atrial lead senses the atrial impulse and
generates a ventriculair pulse (completed loop)
 This pacemaker re-entry tachycardia will not
exceed the programmed upper rate, however
can be significant enough to cause symptoms
 magnet therapy.
 Pacemakers atrial sensing tresholds need to be
adjusted
 Mode switch can cause the pacemaker to
discontinue atrial tracking when the atrial upper
rate is exceeded
Rate Drop Response
 Delivers pacing at high rate when
episodic drop in rate occurs
 Pacing therapy indicated for patients
with neurocardiogenic syncope
Rate Drop Response
Venous thrombosis :
 Venous thrombosis occurs in 30% to 50% of patients
and only 1-3% of patients become symptomatic.
 Manifestations vary from usually asymptomatic,
acute symptomatic thrombosis, and even SVCS .
 Early or late after pacemaker implantation.
 Predictors of severe stenosis are multiple pacemaker
leads , previous pacing , double coils , hormone
therapy .
 Asymptomatic (no treatment) , symptomatic
(anticoagulants – endovascular stents – surgical
correction ).
Extracardiac stimulation
 The diaphragm or pectoral or intercostal muscles
 Diaphragmatic stimulation - direct stimulation of the
diaphragm (left) or stimulation of the phrenic nerve
(right).
 Early postimplantation period , dislodgment of the
pacing lead.
 MC in patients with LV coronary vein branch lead
placement for CRT
 Output pacing importance (testing and treatment)
 Pectoral stimulation - incorrect orientation of the
pacemaker or a current leak from a lead insulation
failure or exposed connector.
PERIOPERATIVE MANAGEMENT
 Require perioperative determination of pacemaker
dependency.
 An external defibrillator should be close by.
 Intraoperative management:
Application of magnet.
Perioperative reprogramming to
assynchronous mode.
Rate adaptive sensors should be disabled.
Oversensing of electrocautery is more common with
unipolar leads.
 When external cardioversion is required ,defibrillation pads
should be placed at least 8 inches from the pulse
generator.
THANK YOU

More Related Content

What's hot

How to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerHow to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerYasmeen Kamal
 
Evaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionEvaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionPRAVEEN GUPTA
 
Speckle Tracking Echocardiography
Speckle Tracking EchocardiographySpeckle Tracking Echocardiography
Speckle Tracking EchocardiographyMADHURJAIN78
 
Aortic stenosis Echo
Aortic stenosis Echo Aortic stenosis Echo
Aortic stenosis Echo madhusiva03
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part iisalah_atta
 
Electrophysiology study basics
Electrophysiology study basicsElectrophysiology study basics
Electrophysiology study basicsSatyam Rajvanshi
 
Collection of cath tracings by navin
Collection of cath tracings by navinCollection of cath tracings by navin
Collection of cath tracings by navinNavin Agrawal
 
Cardiac catheteriztion, Oximetery study in a patient with VSD
Cardiac catheteriztion, Oximetery study in a patient with VSDCardiac catheteriztion, Oximetery study in a patient with VSD
Cardiac catheteriztion, Oximetery study in a patient with VSDPRAVEEN GUPTA
 

What's hot (20)

How to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerHow to follow up a patient with a pacemaker
How to follow up a patient with a pacemaker
 
ICD troubleshooting
ICD troubleshootingICD troubleshooting
ICD troubleshooting
 
Evaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionEvaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunction
 
Role of CRT and CRTD in CHF
Role of CRT and CRTD in CHFRole of CRT and CRTD in CHF
Role of CRT and CRTD in CHF
 
Speckle Tracking Echocardiography
Speckle Tracking EchocardiographySpeckle Tracking Echocardiography
Speckle Tracking Echocardiography
 
Aortic stenosis Echo
Aortic stenosis Echo Aortic stenosis Echo
Aortic stenosis Echo
 
ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
 
Cardiac dyssynchrony ppt by dr awadhesh
Cardiac dyssynchrony ppt   by dr awadheshCardiac dyssynchrony ppt   by dr awadhesh
Cardiac dyssynchrony ppt by dr awadhesh
 
FRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVEFRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVE
 
CRT Case-Based Troubleshooting
CRT Case-Based TroubleshootingCRT Case-Based Troubleshooting
CRT Case-Based Troubleshooting
 
ECG tracing in pacemaker patients
ECG tracing in pacemaker patientsECG tracing in pacemaker patients
ECG tracing in pacemaker patients
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
 
Electrophysiology study basics
Electrophysiology study basicsElectrophysiology study basics
Electrophysiology study basics
 
Stent thrombosis
Stent thrombosisStent thrombosis
Stent thrombosis
 
Collection of cath tracings by navin
Collection of cath tracings by navinCollection of cath tracings by navin
Collection of cath tracings by navin
 
Trouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRTTrouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRT
 
Strain and strain rate
Strain  and strain rateStrain  and strain rate
Strain and strain rate
 
Trans septal puncture
Trans septal punctureTrans septal puncture
Trans septal puncture
 
Crt
CrtCrt
Crt
 
Cardiac catheteriztion, Oximetery study in a patient with VSD
Cardiac catheteriztion, Oximetery study in a patient with VSDCardiac catheteriztion, Oximetery study in a patient with VSD
Cardiac catheteriztion, Oximetery study in a patient with VSD
 

Viewers also liked

Understanding pacemakers
Understanding pacemakersUnderstanding pacemakers
Understanding pacemakersdibufolio
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemakersalman habeeb
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iiisalah_atta
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)Hardik Patel
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1salah_atta
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1tulasi27
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slidesmaltiaziz24
 

Viewers also liked (15)

Pacemaker
PacemakerPacemaker
Pacemaker
 
Understanding pacemakers
Understanding pacemakersUnderstanding pacemakers
Understanding pacemakers
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker Rhythms
Pacemaker RhythmsPacemaker Rhythms
Pacemaker Rhythms
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Pacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen KamalPacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen Kamal
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Pacemaker basics
Pacemaker basicsPacemaker basics
Pacemaker basics
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
pacemaker
pacemakerpacemaker
pacemaker
 

Similar to Pacemaker Programming and Troubleshooting Guide

Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabgEhab Khairy
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsGowri Shankar
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaDrSravaniVishnubhatl
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemakerVishal Cb
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAmr Moustafa Kamel
 
Pacer ppt
Pacer pptPacer ppt
Pacer pptEM OMSB
 
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區Taiwan Heart Rhythm Society
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptSheliDuya2
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiologyjmlafroscia
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxTulikaSingh60
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsAndrewCrofton
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerHASSAN RASHID
 
Defibrillators
DefibrillatorsDefibrillators
DefibrillatorsSapan Jena
 
Pacemaker and anaesthetic implications
Pacemaker and anaesthetic implicationsPacemaker and anaesthetic implications
Pacemaker and anaesthetic implicationsRahul Chauhan
 

Similar to Pacemaker Programming and Troubleshooting Guide (20)

Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabg
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani Vishnubhatla
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemaker
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devices
 
Pacer ppt
Pacer pptPacer ppt
Pacer ppt
 
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.ppt
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptx
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillators
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemaker
 
TACHY ADITYA-2.pptx
TACHY ADITYA-2.pptxTACHY ADITYA-2.pptx
TACHY ADITYA-2.pptx
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker and anaesthetic implications
Pacemaker and anaesthetic implicationsPacemaker and anaesthetic implications
Pacemaker and anaesthetic implications
 
pace1.ppt
pace1.pptpace1.ppt
pace1.ppt
 
Jp's pacemaker
Jp's pacemakerJp's pacemaker
Jp's pacemaker
 
Lvad
LvadLvad
Lvad
 
Oversensing In Is
Oversensing In  IsOversensing In  Is
Oversensing In Is
 

Recently uploaded

Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxVishalSingh1417
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...fonyou31
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfagholdier
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhikauryashika82
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfchloefrazer622
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfAdmir Softic
 

Recently uploaded (20)

Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdf
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 

Pacemaker Programming and Troubleshooting Guide

  • 1. FOLLOW UP AND MANAGEMENT OF PACEMAKER PROGRAMMING AND TROUBLESHOOTING SOUMYA KANTI DUTTA.
  • 2.  Pacemaker components  Battery(Lithium iodide or Vanadium silver)  Pacing impedance  Pulse generator 1. Output circuit 2. Sensing circuit 3. Timing circuit 4. Rate adaptive sensor 5. Modes and mode switching
  • 3.  Pacing impedance :  Pacing impedance refers to the opposition to current flow.  Sources that contribute to pacing impedance: 1. Pacing lead conductor coil 2. Electrode-tissue interface  Normal lead impedance vary from 250- 1200ohms.  Single impedance value may be of little use with out previous values for comparison.
  • 4. Pulse generator output circuit  Capture threshold or Pacing threshold .  Minimum amount of energy required to constantly cause depolarization  Voltage(mA) and pulse duration(mSec)  V=IR
  • 5. 2.0 v 1.5 v 1 v 1. Pulse generator output circuit
  • 6. 1. Pulse generator output circuit  High Pacemaker Output can cause ◦ Reduce longevity ◦ Diaphragmatic stimulation ◦ Muscle Sti. in Unipolar pacemakers ◦ Patient may “feel” heart beat Electrical Testing Of Pacemaker
  • 7. Electrical Testing Of Pacemaker 2. Pulse generator sensing circuit : ◦ Ability of the device to detect intrinsic beat of the heart ◦ Measured - peak to peak magnitude (mv) & slew rate(mv/ms)
  • 8. Pulse generator timing circuit : a. Lower rate limit (LRL) b. Hysteresis rate c. Refractory and blanking periods d. Ventricular safety pacing interval . e. Upper rate response . Electrical Testing Of Pacemaker
  • 9. Pulse generator timing circuit :  Lower rate interval - lowest rate that the pacemaker will pace .  Escape interval:sensed beat to next paced beat.  A paced or non-refractory sensed event restarts the rate timer at the programmed rate.
  • 10. Condition LRL (beats/mt) Infrequent pauses 40-50 Chronic persistent bradycardia 60-70 Relative bradycardia detrimental (long QT) 70-80 Detrimental fast heart rates (angina) 50-60 VVI 60-70 Pulse generator timing circuit :
  • 11. Pulse generator timing circuit :  Hysteresis :  Hysteresis allows the rate to drop below the programmed pacing LRL.  Advantages of hysteresis : 1. Encourages native rhythm – maintain AV sync in VVI , prolong battery life 2. Prevent retrograde conduction – avoids pacemaker syndrome
  • 12.  AV delay (AVI) – pacemaker equivalent of PR interval.  Sensed vs paced AVI – paced AVI is programmed at 125- 200ms , sensed AV interval is programmed at 20-50ms shorter than paced.  Dynamic/adaptive AV delay :shorter AVI with increase in heart rate,allow pacemaker to respond to exercise and have better haemodynamics. Pulse generator timing circuit : sAVI – 150ms pAVI – 200ms
  • 13.  AV delay (AVI)  Longer AVI : Good AV conduction – maintains AV synchrony , long battery life Achieved by following methods :Programming longer AVI , managed ventricular pacing , AV delay hysteresis .  Shorter AVI: HOCM – RV apical pacing decreases HOCM gradient CRT – usually 80-120ms , for 100%ventricular pacing and optimize CO Physiological response to faster heart rates can be answered Pulse generator timing circuit :
  • 14.  Refractory and blanking periods :  Refractory period – sensing present but no action  Blanking period - sensing absent and hence no action Pulse generator timing circuit :
  • 15. Blanking period Time Importance Atrial blanking period 50-100ms Non programmable , Avoid atrial sensing of its own paced beat Post ventricular atrial blanking period 220ms Avoid sensing of ventricular beat Long PVAB decreases detection of AF,AFL Ventricular blanking period 50-100ms Non programmable, Avoid ventricular sensing of its paced beat Post atrial ventricular blanking period 28ms if the PAvB period is too long, R on T - ventricular tachyarrhythmia.
  • 16. Refractory period Importance Ventricular refractory period (VRP) Prevent sensing of T wave . Atrial refractory period (ARP) AVI (120-200ms) . Post ventricular atrial refractory period Avoid sensing retrograde P waves (PMT) , far field R waves . Pulse generator timing circuit :
  • 17. Pulse generator timing circuit : Ventricular safety pacing/ventricular triggered period/cross talk sensing window : Atrial pacing in DDD Trigger ventricular sensing PAVB -pAVI False inhibition of ventricular pacing circuit Asystole
  • 18.  Rate responsive pacing refer to ability of pacemaker to increase its lower rate in response to physiological stimulus  Upper rate limit:important to prevent tracking of rapid atrial activity like in AF  HRR should start with in 10s of exercise , peak at 90 – 120s and should return to baseline with in 60 – 120s after exercise.  If intrinsic atrial rate exceeds URR then wenckebach or 2:1 AVB  Choosing URR : young patients (150b/mt) , old angina (<110b/mt).  Various sensors (activity , minute ventilation , QT)  Rate smoothing: allows variation of R-R interval <6% Pulse generator rate responsive pacing:
  • 19. Pulse generator rate responsive pacing: Wenckebach 2:1 AVB
  • 20. PACING MODES  BEPG(British pacing and electrophysiology group)  NASPE(North American society of pacing and electrophysiology)  DD I/T R  III-D(I+R) :dual tracking of atrial activity while inhibited by ventricular activity  5th position: antitachycardia function(pacing/shock)
  • 21. 5. Pulse generator modes switching: Electrical Testing Of Pacemaker DDD / VDD Atial tachyarrythmias Sensed atrial events Trigger fast ventricular rates Palpitations. Dyspnoea. And Fatigue. DDIR / VVIR
  • 22. Pulse generator modes switching: Programming mode switching Mode switching occurs when the sensed atrial rate exceeds a programmed atrial tachycardia detection rate. By definition, this value must be faster than the URL (maximum tracking rate). Atrial tachycardia detection is typically programmed to 175-l88bpm or thereabouts.AMS base rate is higher than LRI.
  • 23. Electrical Testing Of Pacemaker Pacemaker follow up guidelines: Single chambered pacing Dual chambered pacing 1st month q 2 week 1st month q 2 weeks 2nd -48th month q 12 week 2nd – 30th month q 12weeks 49th – 72nd month q 8 week 31st – 48th month q 8 weeks 73rd month and later q 4 weeks 49th month and later q 4week
  • 24.
  • 25. Pacemaker complications Pocket complications Pocket hematoma Infection Erosion Wound pain Allergic reactions Pacemaker complications Lead dislodgement Pneumothorax /air embolism Cardiac perforation Extracardiac stimulation Venous thrombosis Coronary sinus dissection Twidller syndrome Pacemaker malfunction
  • 26. Pocket hematoma :  The risk of haematoma is increased in patients taking antithrombotic or anticoagulant drugs (Goldstein et al., 1998).  Most small hematomas can be managed conservatively with cold compress and withdrawal of antiplatelet or antithrombotic agents.  Occasionally, large hematomas that compromise the suture line or skin integrity may have to be surgically evacuated.  Needle aspiration increases risk of infection and should not be done.
  • 27. Pocket hematoma :  In patients requiring oral anticoagulants (warfarin), to take INR of about 2.0 at the time of implantation is safe (Belott & Reynolds, 2000).  Unfractionated heparin or low-molecular-weight heparin are always discontinued prior to device implant and ideally avoided for a minimum of 24 hours post implantation.  Administration of anticoagulants can be resumed within 48-72 h after implantation if there is no evidence of substantial hematoma formation.
  • 28. Device-related infections :  The reported incidence of pacemaker-related infection ranges from 0.5% to 6% in early series.  The use of prophylactic antibiotics has decreased the rate of acute infections following pacemaker implantations to <1 to 2 percent in most series.  The mortality of persistent infection when infected leads are not removed can be as high as 66%.  DM, malignancy, operator inexperience, advanced age, corticosteroid use, anticoagulation, recent device manipulation, CRF, and bacteremia from a distant focus of infection.
  • 29. Device related infection : 2007;49;1851-1859 J. Am. Co
  • 30. Twiddler syndrome: Obese women with loose, fatty subcutaneous tissue Small size of the implanted generator with a large pocket Twisting of pulse generator in long axis Lead dislodgement and lead fracture Failure to capture
  • 31.  The prevelance of this syndrome is 0.07% (Gungor et al., 2009)  Rotated along the transverse axis it is referred by us as the reel syndrome.  Pocket should be revised.  Avoid by ◦ Limit the pocket size, ◦ Suture the device to the fascia ◦ The patients not to manipulate their device pocket Twiddler syndrome:
  • 32.  Failure to capture  Failure to output  Sensing abnormalities(under and over sensing)  Specific mode complications 1. Pacemaker related tachycardia 2. Pacemaker syndrome PACEMAKER MALFUNCTION
  • 33. Failure to capture:  Pacing artifact present but no evoked potential .  Causes 1. Lead dislodgement or perforation 2. Lead maturation(inflammation/fibrosis)(exit block) 3. Battery depletion 4. Circuit failure(coil fracture , insulation defect) 5. Inappropriate programming(low output /pulse width) 6. Pseudo malfunction(pacing during refractory period) 7. Functional non capture(oversensing). 8. Metabolic , drugs , cardiomyopathies
  • 34. Medication effect on Capture Drugs that increase capture threshold Drugs that decrease capture threshold Amiodarone Flecainide Propafenone Sotalol Procainamide lidocaine Atropine Epinephrine Isoproterenol corticosteroids
  • 35. Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular pacing artifacts but one failed to result in ventricular depolarization— that is, failure to capture Failure to capture:
  • 36. Failure to capture: Pacing threshold Normal Increased Battery depletion Functional non capture Impedance Normal Dislodgement Exit block Decreased Insulation failure/ break Increased Lead fracture Loose screw
  • 37. Failure to output:  Absence of pacing stimuli and hence no capture .  Causes 1. Pseudo malfunction - hysteresis , PMT termination , sleep rate 2. Over sensing - EMI ; T P R over sensing ; Myopotential/diaphragmatic ; Cross talk . 3. Open circuit - lead fracture , loose screw , incompatible lead . 4. Battery depletion 5. Recording artifact.
  • 38.  Failure to output: VVIR pacemaker This patient had a pacemaker programmed to a unipolar sensing configuration. The sensing of myopotentials led to symptomatic pauses, and reprogramming the pacemaker to a bipolar sensing configuration prevented subsequent myopotential over sensing.
  • 39. Application of magnet Failure to output: Eliminates pauses Pauses persistent Impedance Normal Decreased Insulation failure/break Increased Lead fracture Loose screw Battery depletionOver sensing Pseudo malfunction
  • 40. Battery depletion :  Elective replacement indicators (ERI) 1. Low voltage(2.1-2.4V) 2. Low pacing rate on magnet application 3. Elevated battery impedance 4. Increased pulse width duration 5. Restricted programmability 6. Change to simpler pacing mode  End of life (EOL) 1. Low voltage(≤2.1vol)
  • 41. Pacemaker undersensing :  Pulse artifact or intrinsic complex present but no sensing(sensed beat doesn’t reset escape interval or cycle).  Causes are 1. Defect in signal production – scar /fibrosis following MI , ectopic , cardioversion,defibrillation , metabolic. 2. Defect in signal transmission – lead fracture /dislodgement , insulation failure , partial open circuit. 3. Defect in pacemaker – battery depletion , sensing circuit abnormalities , committed DVI.
  • 42. Pacemaker over sensing :  Inappropriate inhibition as failure to pace  Causes 1. EMI 2. T , P , R over sensing . 3. Far field R wave sensing(atrial). 4. Cross talk 5. Myopotential (unipolar) 6. Make break signals Cross talk : High atrial output High ventricular sensitivity Low VBP Ventricular sensing of paced atrial impulse Ventricular Asystole
  • 43. Pacemaker Syndrome “An assortment of symptoms related to the adverse hemodynamic impact from the loss of AV synchrony.”
  • 44. Pacemaker Syndrome Symptoms include:  Dizziness  Presyncope  Chest tightness  Shortness of breath  Neck pulsations(cannon A wave)  Apprehension/malaise  Fatigue
  • 45. Pacemaker Syndrome May Be Caused By:  Loss of capture, sensing  A-V intervals of long duration  Onset of 2:1 block  Single chamber system(VVI)  Absence of rate increase with exercise  ECG:Retrogradely conducted inverted P over T  CORRECTIVE ACTION: ◦ Reduce pacing rate. ◦ Hysteresis on to allow more sinus rhythm ◦ Shift to dual chamber pacing.
  • 46. Pacemaker-Mediated Tachycardia (PMT)  A rapid paced rhythm that can occur with atrial tracking pacemakers
  • 47. PMT is the Result of:  Also known as pacemaker re-entry tachycardia or endless-loop tachycardia (like other re-entry dysrhythmias, pacemaker is now a part of the re-entry circuit,Mostly anterograde limb of reentrant tachycardia)  Retrograde conduction  PVARP  Tracking fast atrial rates (physiologic or non-physiologic)
  • 48.  only in patients with dual chamber devices  The ventricular impulse is retrogradely conducted via the AV node to the atria, now the atrial lead senses the atrial impulse and generates a ventriculair pulse (completed loop)  This pacemaker re-entry tachycardia will not exceed the programmed upper rate, however can be significant enough to cause symptoms  magnet therapy.  Pacemakers atrial sensing tresholds need to be adjusted  Mode switch can cause the pacemaker to discontinue atrial tracking when the atrial upper rate is exceeded
  • 49. Rate Drop Response  Delivers pacing at high rate when episodic drop in rate occurs  Pacing therapy indicated for patients with neurocardiogenic syncope Rate Drop Response
  • 50. Venous thrombosis :  Venous thrombosis occurs in 30% to 50% of patients and only 1-3% of patients become symptomatic.  Manifestations vary from usually asymptomatic, acute symptomatic thrombosis, and even SVCS .  Early or late after pacemaker implantation.  Predictors of severe stenosis are multiple pacemaker leads , previous pacing , double coils , hormone therapy .  Asymptomatic (no treatment) , symptomatic (anticoagulants – endovascular stents – surgical correction ).
  • 51. Extracardiac stimulation  The diaphragm or pectoral or intercostal muscles  Diaphragmatic stimulation - direct stimulation of the diaphragm (left) or stimulation of the phrenic nerve (right).  Early postimplantation period , dislodgment of the pacing lead.  MC in patients with LV coronary vein branch lead placement for CRT  Output pacing importance (testing and treatment)  Pectoral stimulation - incorrect orientation of the pacemaker or a current leak from a lead insulation failure or exposed connector.
  • 52.
  • 53. PERIOPERATIVE MANAGEMENT  Require perioperative determination of pacemaker dependency.  An external defibrillator should be close by.  Intraoperative management: Application of magnet. Perioperative reprogramming to assynchronous mode. Rate adaptive sensors should be disabled. Oversensing of electrocautery is more common with unipolar leads.  When external cardioversion is required ,defibrillation pads should be placed at least 8 inches from the pulse generator.