SlideShare a Scribd company logo
1 of 5
Download to read offline
IOSR Journal of Electronics and Communication Engineering (IOSR-JECE)
e-ISSN: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 5, Ver. I (Sep - Oct .2015), PP 31-35
www.iosrjournals.org
DOI: 10.9790/2834-10513135 www.iosrjournals.org 31 | Page
Blood Flow Analysis Using Ultrasound Doppler Technique
K.Venketesh
1
(Center for Bio-Medical Engineering, A.U. College of Engineering,Visakhapatnam, India)
Abstract : An in vitro system was used for measurement of frequency and amplitude of ultrasonic signals
scattered from a moving column of blood analog under physiologic flow conditions. Flow rate and probe
orientations (angle, vertical and lateral displacement) were varied in dependently. With the probe centered
above the blood vessel model the frequency shift (F) of the scattered signal were fit to the equation:
F=Cx +C 𝜽 𝜽+Cv +C0 (10°< 𝜽<40°), where 𝜽 is the angle (in degrees) to the normal to the flow, V is
average velocity of flow (measured indirectly) and Cx, C 𝜽, Cv and C0 are constants. Using stepwise regression
the constant Cv was found to be insignificant; including the other constants produced a standard error of 31 Hz,
about 5-l0% of observed values of F. It was not possible to obtain the diameter of physiologic sized blood
vessels using the CW technique because of the low resolution of the system (about 0.5cm). I t was found
practical to locate the proper probe position and estimate angle using only variation in observed amplitude to
guide positioning. This opens the possibility of accurate transcutaneous blood velocity measurements in animals
and humans
Keywords: Blood velocity, ultrasound, flow rate.
I. Introduction
Doppler ultrasound signals have become vital in measuring the disturbances in the instantaneous blood
flow velocity profile. In ultrasound flow device, a series of pulses is transmitted to detect movement of blood.
Echoes from stationary tissue are the same from pulse to pulse. Echoes from moving scatterers exhibit slight
differences in the time for the signal to be returned to the receiver. These differences can be measured as a direct
time difference or, more usually, in terms of a phase shift from which the „Doppler frequency‟ is obtained.
Blood Flow waveform shape and spectrum has proved to be an useful technique in the investigation of many
vascular beds. It has the advantage that derived indices are independent of the beam/flow angle. Changes in flow
waveform shape have been used to investigate both proximal disease (e.g. in the adult peripheral arterial
circulation) and distal changes (in the fetal circulation and uterine arteries). We develop method to analyze the
frequency (f) and flow rate (Q) for various ultrasound frequencies and also get vital results in echo amplitude Vs
probe angle.
Several methods have developed for analyzing the blood flow velocity profiles produced by ultrasound
signals. Some of the methods are Accuracy of continuous wave Doppler measurements of fluid velocity [1], in
this paper describe for measuring the average velocity of flow indirectly and also found proper probe position
and estimate the angle. Another technique is Influence of Beam Profile and Degree of lnsonation on the CW
Doppler Ultrasound Spectrum[2], using this method ultrasound beam profile and the degree to which a vessel is
uniformly insonated influences both the spectral density and the estimation of the mean velocity. A computer
model that enables these effects to be conveniently studied is described for any axisymmetric flow velocity
profile and any probe field response that is symmetric. Specifically, the model provides estimates of the error in
the mean velocity compared to the true mean, the spectral density graph, and the total received power. Another
method, Theoretical Analysis of the Ultrasonic Doppler Flow meter for Measurements of High Flow Velocities
[3], this is based on the calculation of the transit time difference between the ultrasonic waves that are reflected
from a moving particle at its various positions. Beam divergence is taken into account, and each path of the
ultrasonic wave propagation is approximated by two rectilinear components. In our technique frequency against
flow rate is obtained using multiple regression. Subsequent results also obtained with probe angle from 5 to 40
degrees.
Materials and Methods
In the experiments to be described, an -vitro f low model system was used. This system supported
continuous flow in vessels of variable known diameter at directly measured physiologic flow rates. The
commercial Doppler instrument was connected to a frequency and amplitude detector, which produced voltages
proportional to the amp1itude and frequency of the Doppler shifted signal.
Several computer programs were written to model the performance of the ultrasonic flow detection
system, and to correct for the frequency and amplitude sensitivities of the frequency measuring equipment. One
program was used to correct the frequency meter reading for low amplitude input signals. A second computer
Blood Flow Analysis Using Ultrasound Doppler Technique
DOI: 10.9790/2834-10513135 www.iosrjournals.org 32 | Page
program was used to produce a digital signal having the same characteristics as the signal produced by the
ultrasound detector: that is, a band-limited signal with a flat frequency spectrum, but with randomly-varying
phase angles. This model enabled comparison of the operation of the flow system with a mathematical analog.
There are basically four kinds of experiments described in this report:
1. Calibration of the velocity detector for varying probe angle, blood vessel diameter, and probe-to-blood
vessel distance. This calibration was performed using an indirect approach. The flow was measured and the
velocity computed assuming non-turbulent flow. (Measurement of the viscosity of the blood analog and
calculation o f Reynolds numbers showed that this assumption was reasonable).
2. Determination of the spatial response of the probe. This measurement was made using a microscope stage
to position the ultrasonic transducer accurately and reproducibly with respect to the blood vessel analog.
3. Estimation of vessel diameter using CW ultrasound. Again, the microscope stage was used. In addition, a
position transducer was constructed so that plots of probe position versus returned signal amp1itude and/or
frequency could be displayed on a storage oscilloscope.
4. Comparison studies of blood analog (evaporated milk and water) to blood and to digital simulation. Blood
studies were performed to validate the use of evaporated milk as a blood analog, since it was believed that
evaporated milk was much more convenient to use in experimentation
II. Results
Probe calibration. Because the amp1itude of the Doppler shifted signal had significant variation with
probe-to-vessel distance, it was necessary to determine the effect of variation of the signal on measured
frequencies.
By varying the gain of the detector circuit (Figure I) while flow and probe position were held constant,
experimental data (values o f observed frequency "Fobs" and observed "A" were fit to an equation of the form:
Fobs = Flim –ae-bA (1)
„Flim‟ represents the frequency that presumably would be measured by a frequency meter of infinite sensitivity.
The “a” and "b" are empirically determined constants. Experiments were performed using a constant flow rate
while the amp1itde of the input signal to the amp1itude velocity detector was varied. For each flow rate the
arbitrary constants "a" and "b" were determined using the transformed equation:
ln (Flim – Fobs) = ln (a) – bA (2)
The determination was carried out by a linear regression using the amp1itude "A" as the independent
variable and the logarithm of the frequency difference as the dependent variable. The value o f „Flim‟ had to be
selected (by trial). It was chosen to maximize the correlation coefficient between the amplitude and the log of
the frequency difference. A complete table showing the extrapolated relationships between these variables was
produced. When the observed amplitude was greater than 1 volt, the observed frequency was atleast 90% of the
limiting frequency. Accordingly, there was no effort made routinely to calculate explicit corrections on the
frequency values.
As expected, the relationship of velocity of flow to measured frequency shift was linear. Initially, for
constant angles of the probe, a least squares fit was obtained in the equation:
In this equation "F" is the observed frequency shift, “Q” is the volume flow rate, and "m" and "b" are
constants. One unexpected result in the experiments performed was that the constant “b” was never zero, nor
was "b" attributable to experimental error. The value "b" did not appear to depend on tube diameter. Analysis of
the situation revealed that the constant took a value of approximately half of the low frequency cut-off of the
Doppler shift detector(“b” was typically about 200 Hz). Values for the constants "m" and "b" are shown in
Table1.
Table 1. Values for Flow Constants
Angle of Probe to Normal (𝜽) degrees m
Hz/cm3
/min
b
Hz
5 1.412 54.71
10 0.941 109.43
20 1.618 218
30 2.719 244.4
40 3.766 437
50 4.711 547.05
Blood Flow Analysis Using Ultrasound Doppler Technique
DOI: 10.9790/2834-10513135 www.iosrjournals.org 33 | Page
Note that they vary with the angle of the probe. Points are plotted in Figure 2.
Note that increasing the angle to the normal of the probe increases the slope of regression line. This is
an expected result from Doppler theory. The more nearly parallel the probe is to the flow, the greater should be
the shifted frequency of the returned signals.
In subsequent experiments the effect of angle in determining the frequency shift of the Doppler signal
was included quantitatively using data from experiments with angles from 3˚ to 40˚. The relationship between
average flow velocity, angle of probe, and frequency shift was modeled as:
F=Cx +C 𝜽 𝜽+Cv +C0 (4)
Using mu1tiple regression it was found that the constant “Cv” was not significant but that other
constants were. Further, restricting “θ” to the range between 10˚ and 40°, values f o r the constants were
determined to be those listed in Table. Note that the standard error f o r the estimate of frequency shift is only 31
Hz. Thus, for all frequencies observed above 300 Hz, the expected error of estimation using this equation is less
than 10%.
Table 2. Constants of equation (4)
 
Hz31estimateoferrorStandard
0.996PtCoefficiennCorrelatio
125HzC
Hz/degree6.891C
ree.cmHz.sec/deg1.034C
o
e
x





Spatial response of the probe. Measurement of the beam dispersion indicated that the region of
maximum sensitivity of the probe lay in its center line approximately 0.6cm from the probe tip. Unfortunately, it
was quickly discovered that the region of maximum sensitivity encompassed an area which was about 0.75cm
along the axis of the probe and about 0.6cm in diameter, forming a cy1inder around the extended probe axis.
Sensitivity tapered off gradually from this region of maximum sensitivity.
Fig. 1. flowrate Vs frequency.
Blood Flow Analysis Using Ultrasound Doppler Technique
DOI: 10.9790/2834-10513135 www.iosrjournals.org 34 | Page
0 5 10 15 20 25 30
0
0.5
1
1.5
2
2.5
3
3.5
4
Amplitude Vs probe Angle in Degrees
Angle in Degrees
AmplitudeinVolts
Fig. 2. Amplitude VS Angle in degrees
0 5 10 15 20 25 30
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Frequency Vs probe angle
Angle in Degrees
DopplerFrequency,Hz
Fig. 3. Doppler frequency Vs angle in degrees
Brackets represent observed ranges of the amp1itude signal. The relationship of signal amplitude to
probe angle is illustrated in Figure 3. Note that the signal diminishes near the angle zero as expected from
Doppler theory. If the system were ideal there should be no signal at a zero angle because the fluid is moving
perpendicularly to the ultrasonic signal.
Blood Flow Analysis Using Ultrasound Doppler Technique
DOI: 10.9790/2834-10513135 www.iosrjournals.org 35 | Page
Diameter estimation. Since the velocity calibration (Equation 4) contains no constants related to vessel
diameter, it might be anticipated that it is independent of diameter. Indeed, for diameters from 1/8th
to 1/4th
of an
inch this appeared to be the case. Encouraged by these findings, we attempted to measure vessel diameter so that
by combining the estimate of velocity with an estimated diameter we could then obtain a figure for blood flow
without having to place a probe around a dissected vessel. Initially it appeared that the spatial sensitivity was so
diffuse that it would not be possible to determine the size of blood vessel in which velocity was being estimated
by simple probe manipulation. This was experimentally confirmed. Neither the velocity signal nor the amplitude
signal was affected in any way except by a scale factor when the diameter of the blood vessel analog was
changed. Comparison studies. Experiments using blood as the test fluid and digital simulation produced
agreement with results predicted by Equations 1 and 3.
III. Discussion
We have shown that it is possible to estimate the velocity of the flow o f blood, or blood analog, with
reasonable accuracy using CW ultrasound. Thus far, it has not been possible to estimate the diameter of the
vessel using the same technique. However, i f the angle of .the probe to the blood vessel is known, then the
possibility of transcutaneous estimation of velocity alone exists. In order to perform such an estimation one
would have to construct a probe holder which would enable accurate measurement of the angle o f the probe to
the surface of the body or to some other constant reference line. The probe could be manipulated through
different angles until the angle of minimum signal was determined. This angle would then be used as the value θ
= 0. From this direction relative angle measurements could be made and velocity determined from Equation 4.
The ultimate clinical usefulness of these measurements will depend upon the establishment of standards for
blood velocity just as there are now standards for electrocardiography, spirometry, etc. A recent survey of
established investigators in ultrasound revealed a high priority on verification of the validity of diagnosis
obtained with ultrasonic diagnostic equipment in clinical use. It might be useful to undertake a long-range
research project aimed at establishing such standards.
IV. Conclusion
In this experiment results are shown that velocity of the flow of blood using ultrasound Doppler
technique. It was also observed alternate detection of frequency and amplitude for various probe angles.
Finally concluded that values for flow constants were derived from linear regression results observed frequency
90% of limiting frequency. So detected frequency accuracy can be increased by other techniques instead of
linear regression.
References
[1] Baker, D.W., "Pulsed Ultrasonic Doppler Blood Flow Sensing" lEEE Transactions on Sonics and Ultrasonics, SU 17: I70 (July
1970).
[2] Bristman, R., Grossman, B.L., and Correll, J.W., "Accuracy of Transcutaneous Doppler Ultrasonics in Evaluating Extra Cranial
Vascular Disease". Journal of Neurosurqery 32: 529 (Hay 1970).
[3] Sigel, B., e t al, "Evaluation of Doppler Ultrasound Examinations", Archives of Surqery 100: 535 (b y 1970).
[4] Kindt, G.W., Youmans, J.R. and Conway, L.W., "The Use of Ultrasound to Determine Cerebral Arterial Reserve", Journal of
Neurosurqery.
[5] Schade and Duff, "In Vitro Model for Ultrasonic How Measurement," Proc. 24th ACEMB (1971). p.355.
[6] Edman, P.D... and Rushmer, R.F., “Priorities in Basic Research on Diagnostic and Therapeutic Ultrasonic Radiation,” IEEE
Transactions on Biomedical Enqineerinq, BME 19: 382 (Sept.1972).

More Related Content

What's hot

Advances in ultrasound
Advances in ultrasoundAdvances in ultrasound
Advances in ultrasoundSangeeta Jha
 
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...Satellite-and Ground-based Red Tide Detection Method and System by Means of P...
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...Waqas Tariq
 
Airborne Lidar Bathymetry
Airborne Lidar BathymetryAirborne Lidar Bathymetry
Airborne Lidar BathymetryAmar .
 
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...idescitation
 

What's hot (12)

Doppler
DopplerDoppler
Doppler
 
Advances in ultrasound
Advances in ultrasoundAdvances in ultrasound
Advances in ultrasound
 
WorldComp2016_Jenny
WorldComp2016_JennyWorldComp2016_Jenny
WorldComp2016_Jenny
 
Jw3617091714
Jw3617091714Jw3617091714
Jw3617091714
 
Writing Sample2
Writing Sample2Writing Sample2
Writing Sample2
 
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...Satellite-and Ground-based Red Tide Detection Method and System by Means of P...
Satellite-and Ground-based Red Tide Detection Method and System by Means of P...
 
atta ppt
atta pptatta ppt
atta ppt
 
U34129131
U34129131U34129131
U34129131
 
Doppler us (2)
Doppler us (2)Doppler us (2)
Doppler us (2)
 
Airborne Lidar Bathymetry
Airborne Lidar BathymetryAirborne Lidar Bathymetry
Airborne Lidar Bathymetry
 
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...
 
1 ijsrms 02516 (1)
1 ijsrms 02516 (1)1 ijsrms 02516 (1)
1 ijsrms 02516 (1)
 

Viewers also liked

Madcom analyzes the need for broadband in eastern pa
Madcom analyzes the need for broadband in eastern paMadcom analyzes the need for broadband in eastern pa
Madcom analyzes the need for broadband in eastern paRich Frank
 
Teachers' Training Gallery 2015
Teachers' Training Gallery 2015Teachers' Training Gallery 2015
Teachers' Training Gallery 2015Amir Qureshi
 
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...IOSR Journals
 
“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna
“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna
“Designed and Simulation Modified H Shaped” Microstrip Patch AntennaIOSR Journals
 
Performance Improvement of IEEE 802.22 WRAN Physical Layer
Performance Improvement of IEEE 802.22 WRAN Physical LayerPerformance Improvement of IEEE 802.22 WRAN Physical Layer
Performance Improvement of IEEE 802.22 WRAN Physical LayerIOSR Journals
 
Bossung, Kenneth W-NYC-18
Bossung, Kenneth W-NYC-18Bossung, Kenneth W-NYC-18
Bossung, Kenneth W-NYC-18Ken Bossung
 

Viewers also liked (20)

I013117984
I013117984I013117984
I013117984
 
D010432125
D010432125D010432125
D010432125
 
K012127784
K012127784K012127784
K012127784
 
Madcom analyzes the need for broadband in eastern pa
Madcom analyzes the need for broadband in eastern paMadcom analyzes the need for broadband in eastern pa
Madcom analyzes the need for broadband in eastern pa
 
Teachers' Training Gallery 2015
Teachers' Training Gallery 2015Teachers' Training Gallery 2015
Teachers' Training Gallery 2015
 
L010527175
L010527175L010527175
L010527175
 
C010242128
C010242128C010242128
C010242128
 
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...
Design, Electrostatic and Eigen Frequency Analysis of Fixed– Fixed Beam MEMS ...
 
“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna
“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna
“Designed and Simulation Modified H Shaped” Microstrip Patch Antenna
 
Performance Improvement of IEEE 802.22 WRAN Physical Layer
Performance Improvement of IEEE 802.22 WRAN Physical LayerPerformance Improvement of IEEE 802.22 WRAN Physical Layer
Performance Improvement of IEEE 802.22 WRAN Physical Layer
 
G012516065
G012516065G012516065
G012516065
 
A017630109
A017630109A017630109
A017630109
 
F1101043356
F1101043356F1101043356
F1101043356
 
Bossung, Kenneth W-NYC-18
Bossung, Kenneth W-NYC-18Bossung, Kenneth W-NYC-18
Bossung, Kenneth W-NYC-18
 
G1303014145
G1303014145G1303014145
G1303014145
 
K012265965
K012265965K012265965
K012265965
 
A010510109
A010510109A010510109
A010510109
 
C012621520
C012621520C012621520
C012621520
 
H010244147
H010244147H010244147
H010244147
 
K1303046164
K1303046164K1303046164
K1303046164
 

Similar to Blood flow analysis using ultrasound Doppler

Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmetersAJAL A J
 
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...Sheikh R Manihar Ahmed
 
Yoav Levy PHD Thesis - innovative techniques for US imaging
Yoav Levy PHD Thesis - innovative techniques for US imagingYoav Levy PHD Thesis - innovative techniques for US imaging
Yoav Levy PHD Thesis - innovative techniques for US imagingYoav Levy
 
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...Meghan Hegarty
 
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...Meghan Hegarty
 
Luigi Giubbolini | Time/Space-Probing Interferometer for Plasma Diagnostics
Luigi Giubbolini  | Time/Space-Probing Interferometer for Plasma DiagnosticsLuigi Giubbolini  | Time/Space-Probing Interferometer for Plasma Diagnostics
Luigi Giubbolini | Time/Space-Probing Interferometer for Plasma DiagnosticsLuigi Giubbolini
 
Research Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and ScienceResearch Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and Scienceinventy
 
Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingjafarfarwa
 
New Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformNew Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformCSCJournals
 
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...Mirza Baig
 
Research Poster
Research PosterResearch Poster
Research PosterYan Teng
 
Doppler Flow Velocity: Applications in Cardiovascular Research
Doppler Flow Velocity: Applications in Cardiovascular ResearchDoppler Flow Velocity: Applications in Cardiovascular Research
Doppler Flow Velocity: Applications in Cardiovascular ResearchScintica Instrumentation
 

Similar to Blood flow analysis using ultrasound Doppler (20)

Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmeters
 
Presentation2
Presentation2Presentation2
Presentation2
 
Doppler phys (2)
Doppler phys (2)Doppler phys (2)
Doppler phys (2)
 
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...
A Smart Flow Measurement System Adaptive to Different Variation Using Ultraso...
 
Yoav Levy PHD Thesis - innovative techniques for US imaging
Yoav Levy PHD Thesis - innovative techniques for US imagingYoav Levy PHD Thesis - innovative techniques for US imaging
Yoav Levy PHD Thesis - innovative techniques for US imaging
 
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
 
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
 
Journal club
Journal clubJournal club
Journal club
 
Hypertension XCEL PWV validation
Hypertension XCEL PWV validationHypertension XCEL PWV validation
Hypertension XCEL PWV validation
 
Luigi Giubbolini | Time/Space-Probing Interferometer for Plasma Diagnostics
Luigi Giubbolini  | Time/Space-Probing Interferometer for Plasma DiagnosticsLuigi Giubbolini  | Time/Space-Probing Interferometer for Plasma Diagnostics
Luigi Giubbolini | Time/Space-Probing Interferometer for Plasma Diagnostics
 
Research Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and ScienceResearch Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and Science
 
Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their working
 
06509465
0650946506509465
06509465
 
New Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformNew Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet Transform
 
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
 
Icp monitoring seminar
Icp monitoring seminarIcp monitoring seminar
Icp monitoring seminar
 
Research Poster
Research PosterResearch Poster
Research Poster
 
Doppler Flow Velocity: Applications in Cardiovascular Research
Doppler Flow Velocity: Applications in Cardiovascular ResearchDoppler Flow Velocity: Applications in Cardiovascular Research
Doppler Flow Velocity: Applications in Cardiovascular Research
 

More from IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Recently uploaded

Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxMaking_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxnull - The Open Security Community
 
"Federated learning: out of reach no matter how close",Oleksandr Lapshyn
"Federated learning: out of reach no matter how close",Oleksandr Lapshyn"Federated learning: out of reach no matter how close",Oleksandr Lapshyn
"Federated learning: out of reach no matter how close",Oleksandr LapshynFwdays
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupFlorian Wilhelm
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024BookNet Canada
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024The Digital Insurer
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDGMarianaLemus7
 
Snow Chain-Integrated Tire for a Safe Drive on Winter Roads
Snow Chain-Integrated Tire for a Safe Drive on Winter RoadsSnow Chain-Integrated Tire for a Safe Drive on Winter Roads
Snow Chain-Integrated Tire for a Safe Drive on Winter RoadsHyundai Motor Group
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubKalema Edgar
 
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024BookNet Canada
 
Science&tech:THE INFORMATION AGE STS.pdf
Science&tech:THE INFORMATION AGE STS.pdfScience&tech:THE INFORMATION AGE STS.pdf
Science&tech:THE INFORMATION AGE STS.pdfjimielynbastida
 
Artificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraArtificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraDeakin University
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):comworks
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024BookNet Canada
 
Unlocking the Potential of the Cloud for IBM Power Systems
Unlocking the Potential of the Cloud for IBM Power SystemsUnlocking the Potential of the Cloud for IBM Power Systems
Unlocking the Potential of the Cloud for IBM Power SystemsPrecisely
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitecturePixlogix Infotech
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...Fwdays
 
Build your next Gen AI Breakthrough - April 2024
Build your next Gen AI Breakthrough - April 2024Build your next Gen AI Breakthrough - April 2024
Build your next Gen AI Breakthrough - April 2024Neo4j
 
Scanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsScanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsRizwan Syed
 

Recently uploaded (20)

Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxMaking_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
 
"Federated learning: out of reach no matter how close",Oleksandr Lapshyn
"Federated learning: out of reach no matter how close",Oleksandr Lapshyn"Federated learning: out of reach no matter how close",Oleksandr Lapshyn
"Federated learning: out of reach no matter how close",Oleksandr Lapshyn
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project Setup
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC BiblioShare - Tech Forum 2024
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDG
 
Snow Chain-Integrated Tire for a Safe Drive on Winter Roads
Snow Chain-Integrated Tire for a Safe Drive on Winter RoadsSnow Chain-Integrated Tire for a Safe Drive on Winter Roads
Snow Chain-Integrated Tire for a Safe Drive on Winter Roads
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding Club
 
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
#StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
 
Science&tech:THE INFORMATION AGE STS.pdf
Science&tech:THE INFORMATION AGE STS.pdfScience&tech:THE INFORMATION AGE STS.pdf
Science&tech:THE INFORMATION AGE STS.pdf
 
Artificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraArtificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning era
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
 
Unlocking the Potential of the Cloud for IBM Power Systems
Unlocking the Potential of the Cloud for IBM Power SystemsUnlocking the Potential of the Cloud for IBM Power Systems
Unlocking the Potential of the Cloud for IBM Power Systems
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC Architecture
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
 
Build your next Gen AI Breakthrough - April 2024
Build your next Gen AI Breakthrough - April 2024Build your next Gen AI Breakthrough - April 2024
Build your next Gen AI Breakthrough - April 2024
 
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptxE-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
 
Scanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsScanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL Certs
 

Blood flow analysis using ultrasound Doppler

  • 1. IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) e-ISSN: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 5, Ver. I (Sep - Oct .2015), PP 31-35 www.iosrjournals.org DOI: 10.9790/2834-10513135 www.iosrjournals.org 31 | Page Blood Flow Analysis Using Ultrasound Doppler Technique K.Venketesh 1 (Center for Bio-Medical Engineering, A.U. College of Engineering,Visakhapatnam, India) Abstract : An in vitro system was used for measurement of frequency and amplitude of ultrasonic signals scattered from a moving column of blood analog under physiologic flow conditions. Flow rate and probe orientations (angle, vertical and lateral displacement) were varied in dependently. With the probe centered above the blood vessel model the frequency shift (F) of the scattered signal were fit to the equation: F=Cx +C 𝜽 𝜽+Cv +C0 (10°< 𝜽<40°), where 𝜽 is the angle (in degrees) to the normal to the flow, V is average velocity of flow (measured indirectly) and Cx, C 𝜽, Cv and C0 are constants. Using stepwise regression the constant Cv was found to be insignificant; including the other constants produced a standard error of 31 Hz, about 5-l0% of observed values of F. It was not possible to obtain the diameter of physiologic sized blood vessels using the CW technique because of the low resolution of the system (about 0.5cm). I t was found practical to locate the proper probe position and estimate angle using only variation in observed amplitude to guide positioning. This opens the possibility of accurate transcutaneous blood velocity measurements in animals and humans Keywords: Blood velocity, ultrasound, flow rate. I. Introduction Doppler ultrasound signals have become vital in measuring the disturbances in the instantaneous blood flow velocity profile. In ultrasound flow device, a series of pulses is transmitted to detect movement of blood. Echoes from stationary tissue are the same from pulse to pulse. Echoes from moving scatterers exhibit slight differences in the time for the signal to be returned to the receiver. These differences can be measured as a direct time difference or, more usually, in terms of a phase shift from which the „Doppler frequency‟ is obtained. Blood Flow waveform shape and spectrum has proved to be an useful technique in the investigation of many vascular beds. It has the advantage that derived indices are independent of the beam/flow angle. Changes in flow waveform shape have been used to investigate both proximal disease (e.g. in the adult peripheral arterial circulation) and distal changes (in the fetal circulation and uterine arteries). We develop method to analyze the frequency (f) and flow rate (Q) for various ultrasound frequencies and also get vital results in echo amplitude Vs probe angle. Several methods have developed for analyzing the blood flow velocity profiles produced by ultrasound signals. Some of the methods are Accuracy of continuous wave Doppler measurements of fluid velocity [1], in this paper describe for measuring the average velocity of flow indirectly and also found proper probe position and estimate the angle. Another technique is Influence of Beam Profile and Degree of lnsonation on the CW Doppler Ultrasound Spectrum[2], using this method ultrasound beam profile and the degree to which a vessel is uniformly insonated influences both the spectral density and the estimation of the mean velocity. A computer model that enables these effects to be conveniently studied is described for any axisymmetric flow velocity profile and any probe field response that is symmetric. Specifically, the model provides estimates of the error in the mean velocity compared to the true mean, the spectral density graph, and the total received power. Another method, Theoretical Analysis of the Ultrasonic Doppler Flow meter for Measurements of High Flow Velocities [3], this is based on the calculation of the transit time difference between the ultrasonic waves that are reflected from a moving particle at its various positions. Beam divergence is taken into account, and each path of the ultrasonic wave propagation is approximated by two rectilinear components. In our technique frequency against flow rate is obtained using multiple regression. Subsequent results also obtained with probe angle from 5 to 40 degrees. Materials and Methods In the experiments to be described, an -vitro f low model system was used. This system supported continuous flow in vessels of variable known diameter at directly measured physiologic flow rates. The commercial Doppler instrument was connected to a frequency and amplitude detector, which produced voltages proportional to the amp1itude and frequency of the Doppler shifted signal. Several computer programs were written to model the performance of the ultrasonic flow detection system, and to correct for the frequency and amplitude sensitivities of the frequency measuring equipment. One program was used to correct the frequency meter reading for low amplitude input signals. A second computer
  • 2. Blood Flow Analysis Using Ultrasound Doppler Technique DOI: 10.9790/2834-10513135 www.iosrjournals.org 32 | Page program was used to produce a digital signal having the same characteristics as the signal produced by the ultrasound detector: that is, a band-limited signal with a flat frequency spectrum, but with randomly-varying phase angles. This model enabled comparison of the operation of the flow system with a mathematical analog. There are basically four kinds of experiments described in this report: 1. Calibration of the velocity detector for varying probe angle, blood vessel diameter, and probe-to-blood vessel distance. This calibration was performed using an indirect approach. The flow was measured and the velocity computed assuming non-turbulent flow. (Measurement of the viscosity of the blood analog and calculation o f Reynolds numbers showed that this assumption was reasonable). 2. Determination of the spatial response of the probe. This measurement was made using a microscope stage to position the ultrasonic transducer accurately and reproducibly with respect to the blood vessel analog. 3. Estimation of vessel diameter using CW ultrasound. Again, the microscope stage was used. In addition, a position transducer was constructed so that plots of probe position versus returned signal amp1itude and/or frequency could be displayed on a storage oscilloscope. 4. Comparison studies of blood analog (evaporated milk and water) to blood and to digital simulation. Blood studies were performed to validate the use of evaporated milk as a blood analog, since it was believed that evaporated milk was much more convenient to use in experimentation II. Results Probe calibration. Because the amp1itude of the Doppler shifted signal had significant variation with probe-to-vessel distance, it was necessary to determine the effect of variation of the signal on measured frequencies. By varying the gain of the detector circuit (Figure I) while flow and probe position were held constant, experimental data (values o f observed frequency "Fobs" and observed "A" were fit to an equation of the form: Fobs = Flim –ae-bA (1) „Flim‟ represents the frequency that presumably would be measured by a frequency meter of infinite sensitivity. The “a” and "b" are empirically determined constants. Experiments were performed using a constant flow rate while the amp1itde of the input signal to the amp1itude velocity detector was varied. For each flow rate the arbitrary constants "a" and "b" were determined using the transformed equation: ln (Flim – Fobs) = ln (a) – bA (2) The determination was carried out by a linear regression using the amp1itude "A" as the independent variable and the logarithm of the frequency difference as the dependent variable. The value o f „Flim‟ had to be selected (by trial). It was chosen to maximize the correlation coefficient between the amplitude and the log of the frequency difference. A complete table showing the extrapolated relationships between these variables was produced. When the observed amplitude was greater than 1 volt, the observed frequency was atleast 90% of the limiting frequency. Accordingly, there was no effort made routinely to calculate explicit corrections on the frequency values. As expected, the relationship of velocity of flow to measured frequency shift was linear. Initially, for constant angles of the probe, a least squares fit was obtained in the equation: In this equation "F" is the observed frequency shift, “Q” is the volume flow rate, and "m" and "b" are constants. One unexpected result in the experiments performed was that the constant “b” was never zero, nor was "b" attributable to experimental error. The value "b" did not appear to depend on tube diameter. Analysis of the situation revealed that the constant took a value of approximately half of the low frequency cut-off of the Doppler shift detector(“b” was typically about 200 Hz). Values for the constants "m" and "b" are shown in Table1. Table 1. Values for Flow Constants Angle of Probe to Normal (𝜽) degrees m Hz/cm3 /min b Hz 5 1.412 54.71 10 0.941 109.43 20 1.618 218 30 2.719 244.4 40 3.766 437 50 4.711 547.05
  • 3. Blood Flow Analysis Using Ultrasound Doppler Technique DOI: 10.9790/2834-10513135 www.iosrjournals.org 33 | Page Note that they vary with the angle of the probe. Points are plotted in Figure 2. Note that increasing the angle to the normal of the probe increases the slope of regression line. This is an expected result from Doppler theory. The more nearly parallel the probe is to the flow, the greater should be the shifted frequency of the returned signals. In subsequent experiments the effect of angle in determining the frequency shift of the Doppler signal was included quantitatively using data from experiments with angles from 3˚ to 40˚. The relationship between average flow velocity, angle of probe, and frequency shift was modeled as: F=Cx +C 𝜽 𝜽+Cv +C0 (4) Using mu1tiple regression it was found that the constant “Cv” was not significant but that other constants were. Further, restricting “θ” to the range between 10˚ and 40°, values f o r the constants were determined to be those listed in Table. Note that the standard error f o r the estimate of frequency shift is only 31 Hz. Thus, for all frequencies observed above 300 Hz, the expected error of estimation using this equation is less than 10%. Table 2. Constants of equation (4)   Hz31estimateoferrorStandard 0.996PtCoefficiennCorrelatio 125HzC Hz/degree6.891C ree.cmHz.sec/deg1.034C o e x      Spatial response of the probe. Measurement of the beam dispersion indicated that the region of maximum sensitivity of the probe lay in its center line approximately 0.6cm from the probe tip. Unfortunately, it was quickly discovered that the region of maximum sensitivity encompassed an area which was about 0.75cm along the axis of the probe and about 0.6cm in diameter, forming a cy1inder around the extended probe axis. Sensitivity tapered off gradually from this region of maximum sensitivity. Fig. 1. flowrate Vs frequency.
  • 4. Blood Flow Analysis Using Ultrasound Doppler Technique DOI: 10.9790/2834-10513135 www.iosrjournals.org 34 | Page 0 5 10 15 20 25 30 0 0.5 1 1.5 2 2.5 3 3.5 4 Amplitude Vs probe Angle in Degrees Angle in Degrees AmplitudeinVolts Fig. 2. Amplitude VS Angle in degrees 0 5 10 15 20 25 30 0 200 400 600 800 1000 1200 1400 1600 1800 2000 Frequency Vs probe angle Angle in Degrees DopplerFrequency,Hz Fig. 3. Doppler frequency Vs angle in degrees Brackets represent observed ranges of the amp1itude signal. The relationship of signal amplitude to probe angle is illustrated in Figure 3. Note that the signal diminishes near the angle zero as expected from Doppler theory. If the system were ideal there should be no signal at a zero angle because the fluid is moving perpendicularly to the ultrasonic signal.
  • 5. Blood Flow Analysis Using Ultrasound Doppler Technique DOI: 10.9790/2834-10513135 www.iosrjournals.org 35 | Page Diameter estimation. Since the velocity calibration (Equation 4) contains no constants related to vessel diameter, it might be anticipated that it is independent of diameter. Indeed, for diameters from 1/8th to 1/4th of an inch this appeared to be the case. Encouraged by these findings, we attempted to measure vessel diameter so that by combining the estimate of velocity with an estimated diameter we could then obtain a figure for blood flow without having to place a probe around a dissected vessel. Initially it appeared that the spatial sensitivity was so diffuse that it would not be possible to determine the size of blood vessel in which velocity was being estimated by simple probe manipulation. This was experimentally confirmed. Neither the velocity signal nor the amplitude signal was affected in any way except by a scale factor when the diameter of the blood vessel analog was changed. Comparison studies. Experiments using blood as the test fluid and digital simulation produced agreement with results predicted by Equations 1 and 3. III. Discussion We have shown that it is possible to estimate the velocity of the flow o f blood, or blood analog, with reasonable accuracy using CW ultrasound. Thus far, it has not been possible to estimate the diameter of the vessel using the same technique. However, i f the angle of .the probe to the blood vessel is known, then the possibility of transcutaneous estimation of velocity alone exists. In order to perform such an estimation one would have to construct a probe holder which would enable accurate measurement of the angle o f the probe to the surface of the body or to some other constant reference line. The probe could be manipulated through different angles until the angle of minimum signal was determined. This angle would then be used as the value θ = 0. From this direction relative angle measurements could be made and velocity determined from Equation 4. The ultimate clinical usefulness of these measurements will depend upon the establishment of standards for blood velocity just as there are now standards for electrocardiography, spirometry, etc. A recent survey of established investigators in ultrasound revealed a high priority on verification of the validity of diagnosis obtained with ultrasonic diagnostic equipment in clinical use. It might be useful to undertake a long-range research project aimed at establishing such standards. IV. Conclusion In this experiment results are shown that velocity of the flow of blood using ultrasound Doppler technique. It was also observed alternate detection of frequency and amplitude for various probe angles. Finally concluded that values for flow constants were derived from linear regression results observed frequency 90% of limiting frequency. So detected frequency accuracy can be increased by other techniques instead of linear regression. References [1] Baker, D.W., "Pulsed Ultrasonic Doppler Blood Flow Sensing" lEEE Transactions on Sonics and Ultrasonics, SU 17: I70 (July 1970). [2] Bristman, R., Grossman, B.L., and Correll, J.W., "Accuracy of Transcutaneous Doppler Ultrasonics in Evaluating Extra Cranial Vascular Disease". Journal of Neurosurqery 32: 529 (Hay 1970). [3] Sigel, B., e t al, "Evaluation of Doppler Ultrasound Examinations", Archives of Surqery 100: 535 (b y 1970). [4] Kindt, G.W., Youmans, J.R. and Conway, L.W., "The Use of Ultrasound to Determine Cerebral Arterial Reserve", Journal of Neurosurqery. [5] Schade and Duff, "In Vitro Model for Ultrasonic How Measurement," Proc. 24th ACEMB (1971). p.355. [6] Edman, P.D... and Rushmer, R.F., “Priorities in Basic Research on Diagnostic and Therapeutic Ultrasonic Radiation,” IEEE Transactions on Biomedical Enqineerinq, BME 19: 382 (Sept.1972).