INVESTIGATING THE USE OF IMPEDANCE
PLETHYSMOGRAPHY FOR DETECTING DECREASED
BLOOD FLOW IN DIABETIC PATIENT LIMBS
Mirza Taimoor Sultan BAIG
Advisor: Assoc. Prof. Mustafa İlker Beyaz
JANUARY 2023
Outline
• Introduction
• Impedance Plethysmography
• Motivation
• Literature Review
• Contribution
• COMSOL Multiphysics 5.6
• Geometrical Model
• Theoretical Calculation
• Tissue properties
• Simulation Result
• Circuit Diagram
• Physical Implementation
• Experimental Result
• Conclusion
Modern Methods of Measuring Blood Flow
Ultrasound transducer
Source: (Wearable Compact Blood Flow Sensor _ MEMS_Microfabricated
Products _ NTTAT Advanced Technology Product Introduction Site, n.d.)
Sphygmomanometer
Source: (Aneroid Vérnyomásmérő És Sztetoszkópkészlet
Egészségügyi Kézi Otthoni Vérnyomásmérő Széria
Mandzsetta És Sztetoszkóp _ Wish, n.d.)
Digital Sphygmomanometer
Source: (Buy Saint Health Arm Automatic Blood Pressure
Monitor BP Sphygmomanometer Pressure Meter Tonometer
for Measuring Arterial Pressure at Affordable Prices — Free
Shipping, Real Reviews Wi, n.d.)
• This system is made up of an app
specifically designed for smartphones
and a sensor gadget.
• Visualizations such as animations of
heartbeats or blood flow graphs are then
generated by the smartphone based on
this data.
• It is a digital blood pressure
monitor.
• It is portable, easy to use, and
rapidly and accurately detects
blood pressure and pulse.
• It is also called blood pressure
cuffs.
• Cuffs are used to stop blood flow
in arteries by wrapping them
around the upper arm and
inflating them.
Modern Methods of Measuring Blood Flow
PPG Wristwatch
Source: Reader, R. (2022). Apple partner is developing a wearable to track blood
sugar. In Fast Company. Retrieved from
https://www.fastcompany.com/90730050/apple-partner-rockley-photonics-
blood-sugar
Pulse Oximeter
Source: (SPO2 Finger Pulse Oximeter Fom a Reputed Online Medical
Supply Store _ HealthConnection, n.d.)
• A Non-Intrusive method for taking blood
pressure using PPG Technique.
• It can be worn on the wrist
• Oximeters are tiny, non-invasive
gadgets that may connect to a finger or
toe.
• It can tell how well the heart is
supplying blood to the hands and feet.
Modern Methods of Measuring Blood Flow
Electrocardiogram ECG
Source: (USA Medical and Surgical Supplies, 2018)
Holter Meter
Source: (Johns Hopkins Medicine, 2019)
• At an early age of minor cardiovascular issues and signs of
initial heart malfunction, an electrocardiogram or ECG is a
simple, fast, and common test used to diagnose and catch
an overview of the problem.
• It is a safe procedure that captures the electrical activity of
the heart.
• A Holter blood pressure device is worn by the
patient for twenty-four hours.
• If the increase in blood pressure readings is to rule
out or confirm the diagnosis of hypertension .
What is impedance plethysmography?
• A method that can measure
plethysmography and
identify the arterial pulse
waveform is called
bioelectrical impedance.
this technique is collectively
known as impedance
plethysmography.
• By infusing a minute
amount of AC into the body
and then detecting the
potential caused by blood
flow, it gauges changes in
blood volume.
Source: (Impedance Plethysmography (IPG) - Medis, n.d.)
Why Impedance
Plethysmography?
Source: Ansari, S. (2013). Motion artifact reduction in impedance plethysmography signal. ProQuest Dissertations and Theses, 116
Motivation
• Monitoring the blood flow device is not affordable
for everyone.
• Expensive devices need to be precise and close to
accuracy for the patient, however, it is directly
proportional to the price they pay.
• Everyone especially diabetic patients can measure
their existence if the blood flow at home.
• When the blood flow stops this device will give a
signal that you should go to a doctor immediately.
• To make a Portable and wearable device that
everyone can afford easily.
Literature Review
• According to Isabel Morales’ research, the foot
impedance plethysmography was carried out with two
distinct kinds of electrodes, using 1mA/10kHz and
finding the impedance of the foot.
• In this study Piuzzi research is an ultimate technique
selected that makes use of a current with a frequency
of 50 kHz that is injected into the thorax of the subject
using a couple of textile electrodes.
• In this study of Ramkumar & Babu, blood flow was
performed using a fuzzy logic toolbox with many
different activities, measured before and after the 25
seconds of breath-holding activity. The
plethysmographic waveform of forearm impedance
can be used to analyze the changes in a heartbeat that
are correlated with the changes in heart stroke
volume.
Source: Morales, I., Gonzalez-Landaeta, R., & Simini, F. . Pressure sensors used as
bioimpedance plantar electrodes: A feasibility study. 2021 IEEE International
Symposium on Medical Measurements and Applications,
Contribution
• In this work we are investigating using multiple electrodes to
see the additional data going to make testing easier.
• Our research is primarily focusing on the amount of blood flow
through the finger of the upper limb for diabetic patients.
COMSOL Multiphysics 5.6
• We are using COMSOL Model
physics to investigate the tests.
• The geometrical model of the
test is on the 2D dimension.
• In the COMSOL we are using
electrical current physics and we
are choosing the frequency
domain.
Geometrical Model
• This geometry model was
constructed to demonstrate the
internal structure of the
forefinger.
• Electrodes numbering 1,2,3 and
4 are attached to the surface of
the forefinger.
• This way it is easier for our
software to detect the best
frequency penetrated across the
layers.
Frequency-Dependent Tissue Properties
• The layers presented in the
geometrical model namely:
skin, fat, muscle, bone, and
artery will produce different
values at different
frequencies.
• A theoretical view of these
values is shown.
• Depicting permittivity and
conductivity.
The dielectric properties of tissues at different frequencies:
(a) Electrical conductivity (S/m)
(a) Electrical permittivity
Source: (Tissue Frequency Chart » IT’IS Foundation, n.d.)
Simulation Results
0
100000
200000
300000
400000
500000
600000
700000
800000
900000
IMPEDANCE
FREQUENCY
ELECTRODE 1,2,3REAL(0%)
z1 Z2 Z3
0
100000
200000
300000
400000
500000
600000
700000
800000
900000
IMPEDANCE
FREQUENCY
ELECTRODE 1,2,3REAL(15%)
Z1 Z2 Z3
• When the heart is diastolic condition. • When the heart is systolic condition.
Artery expands 15% of its original value, because when the heart pumps and
supplies blood to the artery then the artery size will be expanded.
Simulation Results
-6
-5
-4
-3
-2
-1
0
1
2
3
1
3.16227766
10
31.6227766
100
316.227766
1000
3162.27766
10000
31622.7766
100000
316227.766
1000000
3162277.66
10000000
3.16E+07
100000000
3.16E+08
1000000000
3.16E+09
1.00E+10
3.16E+10
1.00E+11
Impedance
Frequency
Electrode 1 (15%-0%)
Real Part
-4
-3
-2
-1
0
1
2
Impedance
Frequency
Electrode 2 (15%-0%)
Real Part
-4
-3
-2
-1
0
1
2
1
3.16227766
10
31.6227766
100
316.227766
1000
3162.27766
10000
31622.7766
100000
316227.766
1000000
3162277.66
10000000
3.16E+07
100000000
3.16E+08
1000000000
3.16E+09
1.00E+10
3.16E+10
1.00E+11
Impedance
Frequency
Electrode 3 (15%-0%)
Real Part
• The simulation result shows us that the output of electrodes 2 and 3 has given
nearly the same values. Therefore, we would choose one out of electrodes 2 and 3.
Simulation Results
• Finally, we took 100 Hz as the optimal frequency to get 1 as our result.
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
Impedance
Frequency
Electrode 1 -Electrode 2 (15%-0%)
Circuit Diagram
Block Diagram of Position of electrodes top-bottom Block Diagram of Position of electrodes Side-bottom
-6
-4
-2
0
2
4
6
0 0.02 0.04 0.06 0.08 0.1 0.12 0.14
Oscilloscope
• We are using 4Vpp with respect to frequency
100Hz /10KHz.
• We try to look at the changes in peak-to-peak
voltages.
Physical Implementation
• Visualization of the electrodes clipped on the forefinger.
Top Bottom Side Bottom
Top-Bottom Results
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
0 0.02 0.04 0.06 0.08 0.1 0.12 0.14
Voltage
Time
Using 100Hz Frequency Using 10KHz Frequency
-0.05
0
0.05
0.1
0.15
0.2
0 5 10 15 20 25 30
voltage
Time
• Resulting graphs of the top bottom with 100Hz and 10KHz.
Side-Bottom Result
Using 100Hz Frequency
-0.15
-0.1
-0.05
0
0.05
0.1
0 0.05 0.1 0.15 0.2 0.25 0.3
Voltage
Time
Using 10KHz Frequency
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
0.6
0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16
Voltage
Time
• Resulting graphs of the side bottom with 100Hz and 10KHz.
Conclusion
• The finger-simplified model was created in model geometry. A framework made up
of five parts was used to build the finger. (i.e., skin, fat, artery, muscle, and bone)
• The study includes COMSOL simulation results compared with experimental results
to obtain the optimum parameters for this technique. The COMSOL simulation uses
4 electrodes in different positions with a current of 10mA giving the optimal
frequency of 100Hz.
• It was shown that multiple electrodes can be useful in obtaining impedance
change.
• This is useful because signals can not be obtained from one pair, or to further
support the results.
Conclusion
• The study was performed under normal resting conditions where an
experimental procedure was set up to measure the impedance
response from the finger.
• Experiment results were not conclusive as more sensitive
impedance analyzers are required.
• This research will be helpful to find out the heartbeat using
impedance plethysmography.
References of Literature Review
• Morales, I., González-Landaeta, R., & Simini, F. (2021). Bioimpedance plethysmography with capacitive
electrodes and sole force sensors: Comparative trial. Journal of Physics: Conference Series, 2008(1).
https://doi.org/10.1088/1742-6596/2008/1/012018
• Piuzzi, E., Pisa, S., Pittella, E., Podestà, L., & Sangiovanni, S. (2020). Wearable belt with built-in textile
electrodes for cardio—respiratory monitoring. Sensors (Switzerland), 20(16), 1–15.
https://doi.org/10.3390/s20164500
• Ramkumar, M., & Babu, C. G. (2017). A Comprehensive Analysis of Forearm Impedance Plethysmography for
the Maximally Non Invasive Monitoring and Measurement of Heart Pumping Function. SSRN Electronic
Journal, 1(2), 178–184. https://doi.org/10.2139/ssrn.2942416
Thank you

More Related Content

What's hot

What's hot (20)

Unit 4 biomedical
Unit 4 biomedicalUnit 4 biomedical
Unit 4 biomedical
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
EMG electromayogram
EMG electromayogramEMG electromayogram
EMG electromayogram
 
Isolation amplifier
Isolation amplifier Isolation amplifier
Isolation amplifier
 
3 defibrilator medical equipment
3 defibrilator medical equipment3 defibrilator medical equipment
3 defibrilator medical equipment
 
Transducers
TransducersTransducers
Transducers
 
Ultrasound & properties
Ultrasound &  propertiesUltrasound &  properties
Ultrasound & properties
 
Cathode Ray Oscilloscope
Cathode Ray OscilloscopeCathode Ray Oscilloscope
Cathode Ray Oscilloscope
 
Ct scan
Ct scanCt scan
Ct scan
 
Basics of CT Scan
Basics of CT ScanBasics of CT Scan
Basics of CT Scan
 
PCG
PCGPCG
PCG
 
ECG Machine
ECG MachineECG Machine
ECG Machine
 
Usg diplay modes
Usg diplay  modesUsg diplay  modes
Usg diplay modes
 
Electrode Interfacing
Electrode InterfacingElectrode Interfacing
Electrode Interfacing
 
Depth of penetration
Depth of penetrationDepth of penetration
Depth of penetration
 
Piezoelectric effect
Piezoelectric effectPiezoelectric effect
Piezoelectric effect
 
Ecg machine
Ecg machineEcg machine
Ecg machine
 
Bio Potential and Bio Electrodes
Bio Potential and Bio ElectrodesBio Potential and Bio Electrodes
Bio Potential and Bio Electrodes
 
Introduction to Oscilloscopes
Introduction to OscilloscopesIntroduction to Oscilloscopes
Introduction to Oscilloscopes
 
Respiration rate measurement
Respiration rate measurement Respiration rate measurement
Respiration rate measurement
 

Similar to INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BLOOD FLOW IN DIABETIC PATIENT LIMBS

Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingjafarfarwa
 
Design of a High Directive Gain Reconfigurable Circular Patch Antenna
Design of a High Directive Gain Reconfigurable Circular Patch AntennaDesign of a High Directive Gain Reconfigurable Circular Patch Antenna
Design of a High Directive Gain Reconfigurable Circular Patch AntennaAssociate Professor in VSB Coimbatore
 
Icasert 2019 pid_230_revised
Icasert 2019 pid_230_revisedIcasert 2019 pid_230_revised
Icasert 2019 pid_230_revisedMd Kafiul Islam
 
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...Meghan Hegarty
 
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...Meghan Hegarty
 
60Click to edit Master title style467.ppt
60Click to edit Master title style467.ppt60Click to edit Master title style467.ppt
60Click to edit Master title style467.pptFahmiOlayah
 
Research Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and ScienceResearch Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and Scienceinventy
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentationlinsstalex
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsIJERA Editor
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Manu Mitra
 
biomedical signal processing and its analysis
biomedical signal processing and its analysisbiomedical signal processing and its analysis
biomedical signal processing and its analysism8171611219
 
Perspectives in blood flow measurement
Perspectives in blood flow measurementPerspectives in blood flow measurement
Perspectives in blood flow measurementSuhas Deshpande
 
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...CSCJournals
 
A simple portable ecg monitor with iot
A simple portable ecg monitor with iotA simple portable ecg monitor with iot
A simple portable ecg monitor with iotArhamSheikh1
 
Telemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsTelemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsSharad Karwa
 

Similar to INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BLOOD FLOW IN DIABETIC PATIENT LIMBS (20)

Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their working
 
I04845861
I04845861I04845861
I04845861
 
Design of a High Directive Gain Reconfigurable Circular Patch Antenna
Design of a High Directive Gain Reconfigurable Circular Patch AntennaDesign of a High Directive Gain Reconfigurable Circular Patch Antenna
Design of a High Directive Gain Reconfigurable Circular Patch Antenna
 
Nanoknife
NanoknifeNanoknife
Nanoknife
 
Icasert 2019 pid_230_revised
Icasert 2019 pid_230_revisedIcasert 2019 pid_230_revised
Icasert 2019 pid_230_revised
 
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...(short presentation) A Wearable Monitoring System for Continuously Assessing ...
(short presentation) A Wearable Monitoring System for Continuously Assessing ...
 
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...A Wearable Monitoring System for Continuously Assessing the Health of the Per...
A Wearable Monitoring System for Continuously Assessing the Health of the Per...
 
60Click to edit Master title style467.ppt
60Click to edit Master title style467.ppt60Click to edit Master title style467.ppt
60Click to edit Master title style467.ppt
 
Research Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and ScienceResearch Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and Science
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart Ailments
 
PPT 1 BIOLOGY EMG ECG.pptx
PPT 1 BIOLOGY EMG ECG.pptxPPT 1 BIOLOGY EMG ECG.pptx
PPT 1 BIOLOGY EMG ECG.pptx
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
 
biomedical signal processing and its analysis
biomedical signal processing and its analysisbiomedical signal processing and its analysis
biomedical signal processing and its analysis
 
Perspectives in blood flow measurement
Perspectives in blood flow measurementPerspectives in blood flow measurement
Perspectives in blood flow measurement
 
Medical dissection lab pulse oximeter
Medical dissection lab pulse oximeterMedical dissection lab pulse oximeter
Medical dissection lab pulse oximeter
 
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...
A Novel Approach for Measuring Electrical Impedance Tomography for Local Tiss...
 
A simple portable ecg monitor with iot
A simple portable ecg monitor with iotA simple portable ecg monitor with iot
A simple portable ecg monitor with iot
 
Telemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsTelemedicine System For Cardiac Patients
Telemedicine System For Cardiac Patients
 
F010513135
F010513135F010513135
F010513135
 

More from Mirza Baig

Hemodynamic monitor by using IPG Technique
Hemodynamic monitor by using IPG TechniqueHemodynamic monitor by using IPG Technique
Hemodynamic monitor by using IPG TechniqueMirza Baig
 
BIOMEDICAL SENSORS.pptx
BIOMEDICAL SENSORS.pptxBIOMEDICAL SENSORS.pptx
BIOMEDICAL SENSORS.pptxMirza Baig
 
Fingerprint Recognition
Fingerprint RecognitionFingerprint Recognition
Fingerprint RecognitionMirza Baig
 
Power Electronics
Power ElectronicsPower Electronics
Power ElectronicsMirza Baig
 
OPTICAL SENSORS AND THEIR APPLICATIONS
OPTICAL SENSORS AND THEIR APPLICATIONS	OPTICAL SENSORS AND THEIR APPLICATIONS
OPTICAL SENSORS AND THEIR APPLICATIONS Mirza Baig
 
Power Electronics
Power Electronics Power Electronics
Power Electronics Mirza Baig
 
Automatic Solar Vertical Car Parking
Automatic Solar Vertical Car ParkingAutomatic Solar Vertical Car Parking
Automatic Solar Vertical Car ParkingMirza Baig
 
state space modeling of electrical system
state space modeling of electrical systemstate space modeling of electrical system
state space modeling of electrical systemMirza Baig
 
optical sensor
 optical sensor optical sensor
optical sensorMirza Baig
 
AUTOMATIC SOLAR VERTICAL CAR PARKING SYSTEM
      AUTOMATIC  SOLAR VERTICAL CAR PARKING SYSTEM      AUTOMATIC  SOLAR VERTICAL CAR PARKING SYSTEM
AUTOMATIC SOLAR VERTICAL CAR PARKING SYSTEMMirza Baig
 
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...Mirza Baig
 
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...Mirza Baig
 
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...Mirza Baig
 
Development of an Adaptive Multi-sensor to Prevent Venous Stasis
Development of an Adaptive Multi-sensor to Prevent Venous StasisDevelopment of an Adaptive Multi-sensor to Prevent Venous Stasis
Development of an Adaptive Multi-sensor to Prevent Venous StasisMirza Baig
 
Automatic digital-analog impedance plethysmography
Automatic digital-analog impedance plethysmographyAutomatic digital-analog impedance plethysmography
Automatic digital-analog impedance plethysmographyMirza Baig
 
Automatic digital-analog impedance plethysmograph
Automatic digital-analog impedance plethysmographAutomatic digital-analog impedance plethysmograph
Automatic digital-analog impedance plethysmographMirza Baig
 
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...Mirza Baig
 
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...Mirza Baig
 

More from Mirza Baig (20)

Hemodynamic monitor by using IPG Technique
Hemodynamic monitor by using IPG TechniqueHemodynamic monitor by using IPG Technique
Hemodynamic monitor by using IPG Technique
 
BIOMEDICAL SENSORS.pptx
BIOMEDICAL SENSORS.pptxBIOMEDICAL SENSORS.pptx
BIOMEDICAL SENSORS.pptx
 
Fingerprint Recognition
Fingerprint RecognitionFingerprint Recognition
Fingerprint Recognition
 
Power Electronics
Power ElectronicsPower Electronics
Power Electronics
 
OPTICAL SENSORS AND THEIR APPLICATIONS
OPTICAL SENSORS AND THEIR APPLICATIONS	OPTICAL SENSORS AND THEIR APPLICATIONS
OPTICAL SENSORS AND THEIR APPLICATIONS
 
wireshark
wiresharkwireshark
wireshark
 
Power Electronics
Power Electronics Power Electronics
Power Electronics
 
GNU Radio
GNU RadioGNU Radio
GNU Radio
 
Automatic Solar Vertical Car Parking
Automatic Solar Vertical Car ParkingAutomatic Solar Vertical Car Parking
Automatic Solar Vertical Car Parking
 
state space modeling of electrical system
state space modeling of electrical systemstate space modeling of electrical system
state space modeling of electrical system
 
optical sensor
 optical sensor optical sensor
optical sensor
 
AUTOMATIC SOLAR VERTICAL CAR PARKING SYSTEM
      AUTOMATIC  SOLAR VERTICAL CAR PARKING SYSTEM      AUTOMATIC  SOLAR VERTICAL CAR PARKING SYSTEM
AUTOMATIC SOLAR VERTICAL CAR PARKING SYSTEM
 
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...
A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximal...
 
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...
Determination of Cardiac Output based on Minimally Invasive Impedance Plethys...
 
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...
BIO-ELECTRICAL IMPEDENCE PLETHYSMOGRAPHYDESIGNING AN EFFICIENT NON-INVASIVE E...
 
Development of an Adaptive Multi-sensor to Prevent Venous Stasis
Development of an Adaptive Multi-sensor to Prevent Venous StasisDevelopment of an Adaptive Multi-sensor to Prevent Venous Stasis
Development of an Adaptive Multi-sensor to Prevent Venous Stasis
 
Automatic digital-analog impedance plethysmography
Automatic digital-analog impedance plethysmographyAutomatic digital-analog impedance plethysmography
Automatic digital-analog impedance plethysmography
 
Automatic digital-analog impedance plethysmograph
Automatic digital-analog impedance plethysmographAutomatic digital-analog impedance plethysmograph
Automatic digital-analog impedance plethysmograph
 
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitor...
 
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...
Studying Peripheral Vascular Pulse Wave Velocity Using Bioimpedance Plethysmo...
 

Recently uploaded

main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024hassan khalil
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and usesDevarapalliHaritha
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2RajaP95
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxPoojaBan
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024Mark Billinghurst
 
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort servicejennyeacort
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx959SahilShah
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxwendy cai
 

Recently uploaded (20)

main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and uses
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptx
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptx
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 

INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BLOOD FLOW IN DIABETIC PATIENT LIMBS

  • 1. INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BLOOD FLOW IN DIABETIC PATIENT LIMBS Mirza Taimoor Sultan BAIG Advisor: Assoc. Prof. Mustafa İlker Beyaz JANUARY 2023
  • 2. Outline • Introduction • Impedance Plethysmography • Motivation • Literature Review • Contribution • COMSOL Multiphysics 5.6 • Geometrical Model • Theoretical Calculation • Tissue properties • Simulation Result • Circuit Diagram • Physical Implementation • Experimental Result • Conclusion
  • 3. Modern Methods of Measuring Blood Flow Ultrasound transducer Source: (Wearable Compact Blood Flow Sensor _ MEMS_Microfabricated Products _ NTTAT Advanced Technology Product Introduction Site, n.d.) Sphygmomanometer Source: (Aneroid Vérnyomásmérő És Sztetoszkópkészlet Egészségügyi Kézi Otthoni Vérnyomásmérő Széria Mandzsetta És Sztetoszkóp _ Wish, n.d.) Digital Sphygmomanometer Source: (Buy Saint Health Arm Automatic Blood Pressure Monitor BP Sphygmomanometer Pressure Meter Tonometer for Measuring Arterial Pressure at Affordable Prices — Free Shipping, Real Reviews Wi, n.d.) • This system is made up of an app specifically designed for smartphones and a sensor gadget. • Visualizations such as animations of heartbeats or blood flow graphs are then generated by the smartphone based on this data. • It is a digital blood pressure monitor. • It is portable, easy to use, and rapidly and accurately detects blood pressure and pulse. • It is also called blood pressure cuffs. • Cuffs are used to stop blood flow in arteries by wrapping them around the upper arm and inflating them.
  • 4. Modern Methods of Measuring Blood Flow PPG Wristwatch Source: Reader, R. (2022). Apple partner is developing a wearable to track blood sugar. In Fast Company. Retrieved from https://www.fastcompany.com/90730050/apple-partner-rockley-photonics- blood-sugar Pulse Oximeter Source: (SPO2 Finger Pulse Oximeter Fom a Reputed Online Medical Supply Store _ HealthConnection, n.d.) • A Non-Intrusive method for taking blood pressure using PPG Technique. • It can be worn on the wrist • Oximeters are tiny, non-invasive gadgets that may connect to a finger or toe. • It can tell how well the heart is supplying blood to the hands and feet.
  • 5. Modern Methods of Measuring Blood Flow Electrocardiogram ECG Source: (USA Medical and Surgical Supplies, 2018) Holter Meter Source: (Johns Hopkins Medicine, 2019) • At an early age of minor cardiovascular issues and signs of initial heart malfunction, an electrocardiogram or ECG is a simple, fast, and common test used to diagnose and catch an overview of the problem. • It is a safe procedure that captures the electrical activity of the heart. • A Holter blood pressure device is worn by the patient for twenty-four hours. • If the increase in blood pressure readings is to rule out or confirm the diagnosis of hypertension .
  • 6. What is impedance plethysmography? • A method that can measure plethysmography and identify the arterial pulse waveform is called bioelectrical impedance. this technique is collectively known as impedance plethysmography. • By infusing a minute amount of AC into the body and then detecting the potential caused by blood flow, it gauges changes in blood volume. Source: (Impedance Plethysmography (IPG) - Medis, n.d.)
  • 8. Source: Ansari, S. (2013). Motion artifact reduction in impedance plethysmography signal. ProQuest Dissertations and Theses, 116
  • 9. Motivation • Monitoring the blood flow device is not affordable for everyone. • Expensive devices need to be precise and close to accuracy for the patient, however, it is directly proportional to the price they pay. • Everyone especially diabetic patients can measure their existence if the blood flow at home. • When the blood flow stops this device will give a signal that you should go to a doctor immediately. • To make a Portable and wearable device that everyone can afford easily.
  • 10. Literature Review • According to Isabel Morales’ research, the foot impedance plethysmography was carried out with two distinct kinds of electrodes, using 1mA/10kHz and finding the impedance of the foot. • In this study Piuzzi research is an ultimate technique selected that makes use of a current with a frequency of 50 kHz that is injected into the thorax of the subject using a couple of textile electrodes. • In this study of Ramkumar & Babu, blood flow was performed using a fuzzy logic toolbox with many different activities, measured before and after the 25 seconds of breath-holding activity. The plethysmographic waveform of forearm impedance can be used to analyze the changes in a heartbeat that are correlated with the changes in heart stroke volume. Source: Morales, I., Gonzalez-Landaeta, R., & Simini, F. . Pressure sensors used as bioimpedance plantar electrodes: A feasibility study. 2021 IEEE International Symposium on Medical Measurements and Applications,
  • 11. Contribution • In this work we are investigating using multiple electrodes to see the additional data going to make testing easier. • Our research is primarily focusing on the amount of blood flow through the finger of the upper limb for diabetic patients.
  • 12. COMSOL Multiphysics 5.6 • We are using COMSOL Model physics to investigate the tests. • The geometrical model of the test is on the 2D dimension. • In the COMSOL we are using electrical current physics and we are choosing the frequency domain.
  • 13. Geometrical Model • This geometry model was constructed to demonstrate the internal structure of the forefinger. • Electrodes numbering 1,2,3 and 4 are attached to the surface of the forefinger. • This way it is easier for our software to detect the best frequency penetrated across the layers.
  • 14. Frequency-Dependent Tissue Properties • The layers presented in the geometrical model namely: skin, fat, muscle, bone, and artery will produce different values at different frequencies. • A theoretical view of these values is shown. • Depicting permittivity and conductivity. The dielectric properties of tissues at different frequencies: (a) Electrical conductivity (S/m) (a) Electrical permittivity Source: (Tissue Frequency Chart » IT’IS Foundation, n.d.)
  • 15. Simulation Results 0 100000 200000 300000 400000 500000 600000 700000 800000 900000 IMPEDANCE FREQUENCY ELECTRODE 1,2,3REAL(0%) z1 Z2 Z3 0 100000 200000 300000 400000 500000 600000 700000 800000 900000 IMPEDANCE FREQUENCY ELECTRODE 1,2,3REAL(15%) Z1 Z2 Z3 • When the heart is diastolic condition. • When the heart is systolic condition. Artery expands 15% of its original value, because when the heart pumps and supplies blood to the artery then the artery size will be expanded.
  • 16. Simulation Results -6 -5 -4 -3 -2 -1 0 1 2 3 1 3.16227766 10 31.6227766 100 316.227766 1000 3162.27766 10000 31622.7766 100000 316227.766 1000000 3162277.66 10000000 3.16E+07 100000000 3.16E+08 1000000000 3.16E+09 1.00E+10 3.16E+10 1.00E+11 Impedance Frequency Electrode 1 (15%-0%) Real Part -4 -3 -2 -1 0 1 2 Impedance Frequency Electrode 2 (15%-0%) Real Part -4 -3 -2 -1 0 1 2 1 3.16227766 10 31.6227766 100 316.227766 1000 3162.27766 10000 31622.7766 100000 316227.766 1000000 3162277.66 10000000 3.16E+07 100000000 3.16E+08 1000000000 3.16E+09 1.00E+10 3.16E+10 1.00E+11 Impedance Frequency Electrode 3 (15%-0%) Real Part • The simulation result shows us that the output of electrodes 2 and 3 has given nearly the same values. Therefore, we would choose one out of electrodes 2 and 3.
  • 17. Simulation Results • Finally, we took 100 Hz as the optimal frequency to get 1 as our result. -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 1.2 Impedance Frequency Electrode 1 -Electrode 2 (15%-0%)
  • 18. Circuit Diagram Block Diagram of Position of electrodes top-bottom Block Diagram of Position of electrodes Side-bottom -6 -4 -2 0 2 4 6 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 Oscilloscope • We are using 4Vpp with respect to frequency 100Hz /10KHz. • We try to look at the changes in peak-to-peak voltages.
  • 19. Physical Implementation • Visualization of the electrodes clipped on the forefinger. Top Bottom Side Bottom
  • 20. Top-Bottom Results -0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0.5 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 Voltage Time Using 100Hz Frequency Using 10KHz Frequency -0.05 0 0.05 0.1 0.15 0.2 0 5 10 15 20 25 30 voltage Time • Resulting graphs of the top bottom with 100Hz and 10KHz.
  • 21. Side-Bottom Result Using 100Hz Frequency -0.15 -0.1 -0.05 0 0.05 0.1 0 0.05 0.1 0.15 0.2 0.25 0.3 Voltage Time Using 10KHz Frequency -0.2 -0.1 0 0.1 0.2 0.3 0.4 0.5 0.6 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 Voltage Time • Resulting graphs of the side bottom with 100Hz and 10KHz.
  • 22. Conclusion • The finger-simplified model was created in model geometry. A framework made up of five parts was used to build the finger. (i.e., skin, fat, artery, muscle, and bone) • The study includes COMSOL simulation results compared with experimental results to obtain the optimum parameters for this technique. The COMSOL simulation uses 4 electrodes in different positions with a current of 10mA giving the optimal frequency of 100Hz. • It was shown that multiple electrodes can be useful in obtaining impedance change. • This is useful because signals can not be obtained from one pair, or to further support the results.
  • 23. Conclusion • The study was performed under normal resting conditions where an experimental procedure was set up to measure the impedance response from the finger. • Experiment results were not conclusive as more sensitive impedance analyzers are required. • This research will be helpful to find out the heartbeat using impedance plethysmography.
  • 24. References of Literature Review • Morales, I., González-Landaeta, R., & Simini, F. (2021). Bioimpedance plethysmography with capacitive electrodes and sole force sensors: Comparative trial. Journal of Physics: Conference Series, 2008(1). https://doi.org/10.1088/1742-6596/2008/1/012018 • Piuzzi, E., Pisa, S., Pittella, E., Podestà, L., & Sangiovanni, S. (2020). Wearable belt with built-in textile electrodes for cardio—respiratory monitoring. Sensors (Switzerland), 20(16), 1–15. https://doi.org/10.3390/s20164500 • Ramkumar, M., & Babu, C. G. (2017). A Comprehensive Analysis of Forearm Impedance Plethysmography for the Maximally Non Invasive Monitoring and Measurement of Heart Pumping Function. SSRN Electronic Journal, 1(2), 178–184. https://doi.org/10.2139/ssrn.2942416