Instrumentation
of
UV-Visible Spectroscopy
Prepared by
Prof.(Dr.) Dinesh Kumar Mehta
MMCP, (MMDU), Mullana
INSTRUMENTATION
Unlike a spectrometer (which is any instrument that can measure the
properties of light over a range of wavelengths), a spectrophotometer
measures only the intensity of light as a function of its wavelength. The
key components of a spectrophotometer are:
• Light source
• Monochromator
• Sample area
• Detector and Recorder
Light Source:
The most suitable sources of light are
1. Deuterium lamp which emits the light in the UV-region (160-375
nm)
2. Tungsten lamp or tungsten-halogen lamp which emits radiation in
the Visible and near infrared regions (350-2500 nm)
3. Xenon arc lamp which emits radiation in the range 190-800 nm
4. Light emitting diodes (LED) which emits radiation in the visible
range 400-800 nm.
The instruments automatically swap lamps when scanning between
the UV and UV-VIS regions.
Monochromator
The main function of the monochromator is to disperse the beam of
light obtained from the primary source, into its component (Figure 3).
The principle components of monochromator are
• An entrance slit
• A collimating lens
• A dispersing device
• A focusing lens
• An exit slit
The radiation emitted from the primary source (polychromatic
radiation) enters the monochromator through the entrance slit. The
beam is collimated and then strikes the dispersing element (Prism or
grit) at a particular angle.
Sample area
One of the two divided beams is passed through the sample solution
and the other beam is passed through the reference solution. Although,
the samples for recording spectra are most commonly liquids, but the
absorbance of gases and even of solids can be measured. Both sample
and reference solutions are placed in a transparent cell, known as
cuvette. The cuvettes are rectangular in shape, and usually have an
internal width of 1 cm.
It is important that the material of cells must be transparent to the
radiation throughout the region under study. The cells are usually made
of glass, plastic as well as silica or quartz
Detector and Recorder
A detector converts a light signal into an electrical signal. After the
beams are passed through the sample under study as well as the
reference cell, the intensities of the respective transmitted beams are
then compared over the whole wavelength range of the instrument.
Generally two photocells are used as detector in UV spectrometer to
record the spectra. One of the photocell receives the beam from
sample cell and second detector receives the beam from the reference.
The intensity of the radiation from the reference cell is stronger than
the beam of sample cell.
This results in the generation of pulsating or alternating currents in the
photocells.
Spectrophotometers consist of either a photomultiplier tube detector
or a photodiode detector.
The commonly used detector in UV-Vis spectroscopy is
Photomultiplier tube. It consists of three components:
• A cathode which emits electrons when struck by photons of
radiation known as photoemissive cathode
• Several dynodes which emit several electrons for each electron
striking them an anode
In its functioning, when a photon of radiation strikes the cathode,
emission of several electrons take place. These emitted electrons are
then accelerated towards the many dynodes. The first dynode is 90V
more positive than the cathode. The electrons strike the first dynode,
causing the emission of several electrons for each incident electron.
These electrons are then accelerated towards the second dynode, to
produce more electrons which are accelerated towards dynode
three and so on. Finally all the electrons are collected at the anode.
Several dynodes are arranged between the anode and the cathode to
produce an amplification effect. Each photon usually produces 106
-107
electrons, resulting in the amplified current that can be measured.
Photomultipliers are very sensitive to both UV and visible radiations
and have fast response times. It is significant to note that Intense light
may damage photomultipliers, hence they are limited to measuring low
power radiation.
Photodiodes are increasingly being used as detectors in modern
spectrophotometers. Photodiode detectors have a wider dynamic range
and are more robust than photomultiplier tube detectors. In a
photodiode, light falling on the semiconductor material allows
electrons to flow through it, thereby depleting the charge in a capacitor
connected across the material. The amount of charge that is required to
recharge the capacitor at regular intervals is proportional to the
intensity of the light. The limits of detection are approximately 170–
1100 nm for silicon-based detectors.
Some modern spectrophotometers contain an array of photodiode
detectors instead of a single detector as mentioned in above figure.
The amount of charge needed to recharge the capacitors is proportional
to the number of photons detected by each diode, which in turn is
proportional to the light intensity. The absorption spectrum is obtained
by measuring the variation in light intensity over the entire wavelength
range.
Sample Preparation
The UV-Vis spectra are usually measured in very dilute solutions.
Usually 1 mg of the compound of molecular weight of 100-200 is
dissolved in 100 ml of suitable solvent and only a portion is used for
recording the spectra. The solvent should not absorb radiation and must
be transparent over the desired range of wavelength. The solvents,
which do not contain conjugated system, are most suitable for
recording the UV-spectra. The solvent should also be inert to the
sample. Some commonly used solvents are water, 95% ethanol,
methanol and hexane.

More Related Content

What's hot

Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopy
Arabinda Changmai
 

What's hot (20)

Fluorimetry/ Fluoroscences
Fluorimetry/ FluoroscencesFluorimetry/ Fluoroscences
Fluorimetry/ Fluoroscences
 
Quenching
QuenchingQuenching
Quenching
 
Quenching In Fluorimetry.
Quenching In Fluorimetry.Quenching In Fluorimetry.
Quenching In Fluorimetry.
 
Derivative spectroscopy
Derivative spectroscopyDerivative spectroscopy
Derivative spectroscopy
 
Spectrophotometric titration by Mr. Pradeep Swarnkar
Spectrophotometric titration by Mr. Pradeep SwarnkarSpectrophotometric titration by Mr. Pradeep Swarnkar
Spectrophotometric titration by Mr. Pradeep Swarnkar
 
Fluorimetery.
Fluorimetery.Fluorimetery.
Fluorimetery.
 
uv -visible spectroscopy
 uv -visible spectroscopy uv -visible spectroscopy
uv -visible spectroscopy
 
GEL CHROMATOGRAPHY.pptx
GEL CHROMATOGRAPHY.pptxGEL CHROMATOGRAPHY.pptx
GEL CHROMATOGRAPHY.pptx
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopy
 
Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopy
 
UV-VISIBLE SPECTROSCOPY
UV-VISIBLE SPECTROSCOPYUV-VISIBLE SPECTROSCOPY
UV-VISIBLE SPECTROSCOPY
 
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
 
Mass fragmentation & rules
Mass fragmentation & rulesMass fragmentation & rules
Mass fragmentation & rules
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
Thin layer Chromatography (TLC)
Thin layer Chromatography (TLC)Thin layer Chromatography (TLC)
Thin layer Chromatography (TLC)
 
Factors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IRFactors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IR
 
Instrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopyInstrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopy
 
Infrared Spectroscopy
Infrared SpectroscopyInfrared Spectroscopy
Infrared Spectroscopy
 
Fluorimetry presentation
Fluorimetry presentationFluorimetry presentation
Fluorimetry presentation
 
Spectrofluorimetry
Spectrofluorimetry Spectrofluorimetry
Spectrofluorimetry
 

Similar to INSTRUMENTATIONS UV-VIS.pdf

uv visible spectrometery ppt
uv visible spectrometery pptuv visible spectrometery ppt
uv visible spectrometery ppt
rajesh kumar
 

Similar to INSTRUMENTATIONS UV-VIS.pdf (20)

Uv vis spectroscopy- part-3
Uv vis spectroscopy- part-3Uv vis spectroscopy- part-3
Uv vis spectroscopy- part-3
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopy
 
Uv
UvUv
Uv
 
Absorbance
AbsorbanceAbsorbance
Absorbance
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
Ultra violet spectrometer
Ultra violet spectrometerUltra violet spectrometer
Ultra violet spectrometer
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopy
 
Instrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometerInstrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometer
 
U.V Spectroscopy
U.V SpectroscopyU.V Spectroscopy
U.V Spectroscopy
 
Spectroscopy.pptx
Spectroscopy.pptxSpectroscopy.pptx
Spectroscopy.pptx
 
uv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopyuv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopy
 
Uv spec
Uv specUv spec
Uv spec
 
Spectropotometer.08.20 - Copy.ppt
Spectropotometer.08.20 - Copy.pptSpectropotometer.08.20 - Copy.ppt
Spectropotometer.08.20 - Copy.ppt
 
performance and specifications of spectrophotometer
performance and specifications of spectrophotometerperformance and specifications of spectrophotometer
performance and specifications of spectrophotometer
 
U.V Spectroscopy.
U.V Spectroscopy.U.V Spectroscopy.
U.V Spectroscopy.
 
uv visible spectrometery ppt
uv visible spectrometery pptuv visible spectrometery ppt
uv visible spectrometery ppt
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Uv spectroscopy
Uv spectroscopyUv spectroscopy
Uv spectroscopy
 
Spectrophotometer introducing
Spectrophotometer  introducingSpectrophotometer  introducing
Spectrophotometer introducing
 

More from Dr. Dinesh Mehta

More from Dr. Dinesh Mehta (20)

MASS SPECTROMETRY=.pdf
MASS SPECTROMETRY=.pdfMASS SPECTROMETRY=.pdf
MASS SPECTROMETRY=.pdf
 
MS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdfMS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdf
 
MASS SPECTROMETRY INTRODUCTION.pdf
MASS SPECTROMETRY INTRODUCTION.pdfMASS SPECTROMETRY INTRODUCTION.pdf
MASS SPECTROMETRY INTRODUCTION.pdf
 
MS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdfMS Fragmentation Process and Application of MS.pdf
MS Fragmentation Process and Application of MS.pdf
 
UV-VIS.pdf
UV-VIS.pdfUV-VIS.pdf
UV-VIS.pdf
 
Quality System and Audit.pptx
Quality System and Audit.pptxQuality System and Audit.pptx
Quality System and Audit.pptx
 
QMS.pptx
QMS.pptxQMS.pptx
QMS.pptx
 
AUDIT Ph Manufacturing.pptx
AUDIT Ph Manufacturing.pptxAUDIT Ph Manufacturing.pptx
AUDIT Ph Manufacturing.pptx
 
Auditing Manufacturing Process and Product and Process Information.pdf
Auditing Manufacturing Process and Product and Process Information.pdfAuditing Manufacturing Process and Product and Process Information.pdf
Auditing Manufacturing Process and Product and Process Information.pdf
 
Building of Raw Material.pdf
Building of Raw Material.pdfBuilding of Raw Material.pdf
Building of Raw Material.pdf
 
CHAPTER-3 Product and Process Information.pdf
CHAPTER-3 Product and Process Information.pdfCHAPTER-3 Product and Process Information.pdf
CHAPTER-3 Product and Process Information.pdf
 
CHAPTER-1 DEFICIENCIES.pdf
CHAPTER-1 DEFICIENCIES.pdfCHAPTER-1 DEFICIENCIES.pdf
CHAPTER-1 DEFICIENCIES.pdf
 
CHAPTER-1 Information Gathering and Administration.pdf
CHAPTER-1 Information Gathering and Administration.pdfCHAPTER-1 Information Gathering and Administration.pdf
CHAPTER-1 Information Gathering and Administration.pdf
 
CHAPTER-1 Management Audit and Planning procedure.pdf
CHAPTER-1 Management Audit and Planning procedure.pdfCHAPTER-1 Management Audit and Planning procedure.pdf
CHAPTER-1 Management Audit and Planning procedure.pdf
 
HPTLC MS=
HPTLC MS=HPTLC MS=
HPTLC MS=
 
GC -MS/MS and LC MS/MS
GC -MS/MS and LC MS/MSGC -MS/MS and LC MS/MS
GC -MS/MS and LC MS/MS
 
Benchmarking
BenchmarkingBenchmarking
Benchmarking
 
Application of hyphenated techniques
Application of hyphenated techniques Application of hyphenated techniques
Application of hyphenated techniques
 
Application of hyphenated techniques(GC-MS)
Application of hyphenated techniques(GC-MS)Application of hyphenated techniques(GC-MS)
Application of hyphenated techniques(GC-MS)
 
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)Hyphenated techniques(GC-MS/MS, LC-MS/MS,  HPTLC-MS)
Hyphenated techniques(GC-MS/MS, LC-MS/MS, HPTLC-MS)
 

Recently uploaded

Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
chaddageeta79
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
jualobat34
 
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
chaddageeta79
 
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
chaddageeta79
 
Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
MedicoseAcademics
 

Recently uploaded (20)

Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
Call Girl In Mysore 💯Niamh 📲🔝7427069034🔝Call Girls No💰Advance Cash On Deliver...
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
 
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedPart I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
 
Shazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdfShazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdf
 
Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...
 
The Clean Living Project Episode 23 - Journaling
The Clean Living Project Episode 23 - JournalingThe Clean Living Project Episode 23 - Journaling
The Clean Living Project Episode 23 - Journaling
 
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
👉 Gulbarga Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl S...
 
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
👉 Saharanpur Call Girls Service Just Call 🍑👄7427069034 🍑👄 Top Class Call Girl...
 
Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024
 
Drug development life cycle indepth overview.pptx
Drug development life cycle indepth overview.pptxDrug development life cycle indepth overview.pptx
Drug development life cycle indepth overview.pptx
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
 
Face and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptxFace and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptx
 
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptxCreeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
Premium Call Girls Jammu 🧿 7427069034 🧿 High Class Call Girl Service Available
Premium Call Girls Jammu 🧿 7427069034 🧿 High Class Call Girl Service AvailablePremium Call Girls Jammu 🧿 7427069034 🧿 High Class Call Girl Service Available
Premium Call Girls Jammu 🧿 7427069034 🧿 High Class Call Girl Service Available
 
Call Now ☎ 9549551166 || Call Girls in Dehradun Escort Service Dehradun
Call Now ☎ 9549551166  || Call Girls in Dehradun Escort Service DehradunCall Now ☎ 9549551166  || Call Girls in Dehradun Escort Service Dehradun
Call Now ☎ 9549551166 || Call Girls in Dehradun Escort Service Dehradun
 
Dr. A Sumathi - LINEARITY CONCEPT OF SIGNIFICANCE.pdf
Dr. A Sumathi - LINEARITY CONCEPT OF SIGNIFICANCE.pdfDr. A Sumathi - LINEARITY CONCEPT OF SIGNIFICANCE.pdf
Dr. A Sumathi - LINEARITY CONCEPT OF SIGNIFICANCE.pdf
 
Premium Call Girls Kochi 🧿 7427069034 🧿 High Class Call Girl Service Available
Premium Call Girls Kochi 🧿 7427069034 🧿 High Class Call Girl Service AvailablePremium Call Girls Kochi 🧿 7427069034 🧿 High Class Call Girl Service Available
Premium Call Girls Kochi 🧿 7427069034 🧿 High Class Call Girl Service Available
 
Porur Escorts (Chennai) 9632533318 Women seeking Men Real Service
Porur Escorts (Chennai) 9632533318 Women seeking Men Real ServicePorur Escorts (Chennai) 9632533318 Women seeking Men Real Service
Porur Escorts (Chennai) 9632533318 Women seeking Men Real Service
 

INSTRUMENTATIONS UV-VIS.pdf

  • 1. Instrumentation of UV-Visible Spectroscopy Prepared by Prof.(Dr.) Dinesh Kumar Mehta MMCP, (MMDU), Mullana
  • 2. INSTRUMENTATION Unlike a spectrometer (which is any instrument that can measure the properties of light over a range of wavelengths), a spectrophotometer measures only the intensity of light as a function of its wavelength. The key components of a spectrophotometer are: • Light source • Monochromator • Sample area • Detector and Recorder Light Source: The most suitable sources of light are 1. Deuterium lamp which emits the light in the UV-region (160-375 nm) 2. Tungsten lamp or tungsten-halogen lamp which emits radiation in the Visible and near infrared regions (350-2500 nm) 3. Xenon arc lamp which emits radiation in the range 190-800 nm 4. Light emitting diodes (LED) which emits radiation in the visible range 400-800 nm.
  • 3. The instruments automatically swap lamps when scanning between the UV and UV-VIS regions. Monochromator The main function of the monochromator is to disperse the beam of light obtained from the primary source, into its component (Figure 3). The principle components of monochromator are • An entrance slit • A collimating lens • A dispersing device • A focusing lens • An exit slit The radiation emitted from the primary source (polychromatic radiation) enters the monochromator through the entrance slit. The beam is collimated and then strikes the dispersing element (Prism or grit) at a particular angle.
  • 4. Sample area One of the two divided beams is passed through the sample solution and the other beam is passed through the reference solution. Although, the samples for recording spectra are most commonly liquids, but the absorbance of gases and even of solids can be measured. Both sample and reference solutions are placed in a transparent cell, known as cuvette. The cuvettes are rectangular in shape, and usually have an internal width of 1 cm. It is important that the material of cells must be transparent to the radiation throughout the region under study. The cells are usually made of glass, plastic as well as silica or quartz Detector and Recorder A detector converts a light signal into an electrical signal. After the beams are passed through the sample under study as well as the reference cell, the intensities of the respective transmitted beams are
  • 5. then compared over the whole wavelength range of the instrument. Generally two photocells are used as detector in UV spectrometer to record the spectra. One of the photocell receives the beam from sample cell and second detector receives the beam from the reference. The intensity of the radiation from the reference cell is stronger than the beam of sample cell. This results in the generation of pulsating or alternating currents in the photocells. Spectrophotometers consist of either a photomultiplier tube detector or a photodiode detector. The commonly used detector in UV-Vis spectroscopy is Photomultiplier tube. It consists of three components: • A cathode which emits electrons when struck by photons of radiation known as photoemissive cathode • Several dynodes which emit several electrons for each electron striking them an anode In its functioning, when a photon of radiation strikes the cathode, emission of several electrons take place. These emitted electrons are then accelerated towards the many dynodes. The first dynode is 90V
  • 6. more positive than the cathode. The electrons strike the first dynode, causing the emission of several electrons for each incident electron. These electrons are then accelerated towards the second dynode, to produce more electrons which are accelerated towards dynode three and so on. Finally all the electrons are collected at the anode. Several dynodes are arranged between the anode and the cathode to produce an amplification effect. Each photon usually produces 106 -107 electrons, resulting in the amplified current that can be measured. Photomultipliers are very sensitive to both UV and visible radiations and have fast response times. It is significant to note that Intense light
  • 7. may damage photomultipliers, hence they are limited to measuring low power radiation. Photodiodes are increasingly being used as detectors in modern spectrophotometers. Photodiode detectors have a wider dynamic range and are more robust than photomultiplier tube detectors. In a photodiode, light falling on the semiconductor material allows electrons to flow through it, thereby depleting the charge in a capacitor connected across the material. The amount of charge that is required to recharge the capacitor at regular intervals is proportional to the intensity of the light. The limits of detection are approximately 170– 1100 nm for silicon-based detectors.
  • 8. Some modern spectrophotometers contain an array of photodiode detectors instead of a single detector as mentioned in above figure. The amount of charge needed to recharge the capacitors is proportional to the number of photons detected by each diode, which in turn is proportional to the light intensity. The absorption spectrum is obtained by measuring the variation in light intensity over the entire wavelength range. Sample Preparation The UV-Vis spectra are usually measured in very dilute solutions. Usually 1 mg of the compound of molecular weight of 100-200 is dissolved in 100 ml of suitable solvent and only a portion is used for recording the spectra. The solvent should not absorb radiation and must be transparent over the desired range of wavelength. The solvents, which do not contain conjugated system, are most suitable for recording the UV-spectra. The solvent should also be inert to the sample. Some commonly used solvents are water, 95% ethanol, methanol and hexane.