SlideShare a Scribd company logo
1 of 24
Download to read offline
PST351-POLYMER CHARACTERISATION
MIGRATION RATES OF SOLUTE
PST351-POLYMER CHARACTERISATION
BACKGROU
ND
• Chromatogram is useful for both qualitative and
quantitative analysis.
• The positions of peaks on the time axis - identify
the components of the sample;
• the areas under the peaks-quantitative measure
of the amount of each component.
PST351-POLYMER CHARACTERISATION
BACKGROU
ND
Graph showing detector response as a function of
elution time.
PST351-POLYMER CHARACTERISATION
• The effectiveness of a chromatographic column in
separating two solutes depends in part on the
relative rates at which the two species are eluted.
• These rates are determined by the magnitude of the
equilibrium constants for the reactions by which the
solutes distribute themselves between the mobile
and stationary phases.
BACKGROUND
PST351-POLYMER CHARACTERISATION
DISTRIBUTION
CONSTANT
The distribution equilibria involved in chromatography involve the
transfer of an analyte between the mobile and stationary phases.
The equilibrium constant (Kc ) for this reaction is called the
distribution constant, which is defined as;
Where;
cs is the molar concentration of the solute in the stationary phase
cm is its molar concentration in the mobile phase
K is constant over a wide range of solute concentrations.
PST351-POLYMER CHARACTERISATION
The Dead Time (tM), Stationary Time
(tS), & Retention Time (tR)
▪ The time between sample injection and the appearance of the
non-retained species peak is called as the dead or void time, (tM).
▪ It is the time it takes for an unretained species to
pass through chromatography column.
▪ All components spend the dead time in the mobile
phase
▪ Separation are based on the different times, ts that
components spend in stationary phase
▪ The dead time provides a measure of the average rate of
migration of a mobile phase and is an important parameter to
identifying analyte peaks.
PST351-POLYMER CHARACTERISATION
The time it takes after sample injection for the analyte peak to
reach the detector is called the retention time , (tR).
The analyte has been retained because it spends a time in the
stationary phase, (ts)
The retention time is then calculated by;
The Dead Time (tM), Stationary Time
(tS), & Retention Time (tR)
PST351-POLYMER CHARACTERISATION
The Dead Time (tM), Stationary Time
(tS), & Retention Time (tR)
PST351-POLYMER CHARACTERISATION
Retention Factor (k’)
The retention factor (capacity factor), k’ is an important parameter that is
widely used to describe the migration rates of solutes on columns
For a solute A, the retention factor k`A is defined as;
where;
k’A is the distribution constant for the species A.
tR and tM are readily obtained from a chromatogram
PST351-POLYMER CHARACTERISATION
Retention Factor (k’)
When the retention factor for a solute is much less than this range (1-
5), seperation occurs so rapidly that accurate determination of the
retention times is difficult.
When the retention factor is larger (20 to 30) , seperation times become
inordinately long.
PST351-POLYMER CHARACTERISATION
Selectivity Factor, 
The selectivity factor,  of a column for the two species A and B is defined as;
where;
KB is the distribution constant for species B
KA is the distribution constant for species A
A relationship between the selectivity factor and retention factors can be described
as:
Where;
k`B is the retention factor of species A.
k`A is the retention factor of species B.
PST351-POLYMER CHARACTERISATION
The theoretical Plate Model of Chromatography
• The plate model supposes that the chromatographic column is
contains a large number of separate layers, called theoretical
plates.
• Separate equilibrations of the sample between the stationary and
mobile phase occur in these "plates".
• The analyte moves down the column by transfer of equilibrated
mobile phase from one plate to the next.
PST351-POLYMER CHARACTERISATION
Column Efficiency
To obtain optimal separations, sharp, symmetrical
chromatographic peaks must be obtained. This means that band
broadening must be limited. It is also beneficial to measure the
efficiency of the column.
Two related terms are widely used as quantitative measures of
chromatographic column efficiency
Plate Height, H Number of Plate, N
The two are related by the equation;
where L is the length (usually in centimeters) of the column packing.
The efficiency of chromatographic columns can be increases by
• Increase the number of plates, N
• Reduce the plate height
N= L/H
PST351-POLYMER CHARACTERISATION
Column
Efficiency • No of plate, N = the
more plates the
better
• Plate height, H = the
smaller the better
PST351-POLYMER CHARACTERISATION
The number of theoretical plates that a real column possesses can
be found by examining a chromatographic peak after elution;
Column Efficiency
As can be seen from this equation, columns behave as if they
have different numbers of plates for different solutes in a mixture.
where w1/2 is the peak width
at half-height.
PST351-POLYMER CHARACTERISATION
Column
Resolution (Rs)
The resolution Rs of a column provides a quantitative measure of
its ability to separate two analytes.
The resolution of two species, A and B, is defined as
Baseline resolution is achieved when R = 1.5
PST351-POLYMER CHARACTERISATION
It is useful to relate the resolution to the number of plates in the
column, the selectivity factor and the retention factors of the two
solutes;
Column
Resolution (Rs)
From this equation, it is known that to obtain high resolution, the
three terms must be maximised.
• An increase in N, the number of theoretical plates,
• Reduce the column height by
- reducing the size of the stationary phase particles.
PST351-POLYMER CHARACTERISATION
Column
Resolution
(Rs)
Apart from improving column resolution to improve
separation, it is often found that by controlling the capacity
factor, k', separations can be greatly improved.
This can be achieved by:
• Changing the temperature (in Gas Chromatography) or
• Changing the composition of the mobile phase (in Liquid
Chromatography).
PST351-POLYMER CHARACTERISATION
The selectivity factor,  can also be manipulated to improve separations.
In these cases, k' is optimised first, and then a is increased by one of the
following procedures:
1. Changing mobile phase composition
2. Changing column temperature
3. Changing composition of stationary phase
4. Using special chemical effects (such as incorporating a
species which complexes with one of the solutes into the
stationary phase)
Column
Resolution (Rs)
PST351-POLYMER CHARACTERISATION
Column
Temperature
• The retention time of a compound depends not only on the
type of stationary phase but also the column temperature.
• In general you can analyze your sample under isothermal
conditions in which the temp of the column remains constant
throughout the analysis.
• As the temperature increases, compounds move through the
column faster and consequently, the retention time decreases.
• Thus, we can potentially reduce the analysis time by
increasing column temperature; however, we need to allow
sufficient time for the compounds to interact with the
stationary to provide the required separation of sample
components.
PST351-POLYMER CHARACTERISATION
Column
Temperature
In the simplest GC analysis, the column is maintained at a constant
temperature and is termed “isothermal” analysis.
The chromatogram in this figure shows the separation of a mixture of
normal alkanes (straight-chained hydrocarbons) ranging from C6 to C21
using isothermal conditions at 150 C.
The low molecular weight alkanes <C10 elute very quickly and are not
resolved.
The higher molecular weight components >C16 do not elute from the
column at this temperature and C14 and C15 are broad, asymmetrical
peaks.
PST351-POLYMER CHARACTERISATION
Column
Temperature
The chromatogram shows the same alkane mixture analyzed
using “temperature programmed” conditions.
In this case, the column temperature is increased from 50 to 250
C at 8 C/ min during the analysis.
All of the alkane components are completely separated into
narrow, symmetrical peaks.
PST351-POLYMER CHARACTERISATION
Column
Temperature Comparing that separation of the same
mixture with the lower chromatogram
which is temperature programmed from
50 degrees to 250 degrees at 8 degrees
per minute , it can be seen that;
- The separation efficiency is much
greater here (graph b).
- Higher molecular weigh components
longer chain hydrocarbons that being
eluted
In a conclusion, it can be said that;
- The longer the compound is in the
column the broader the peaks become.
- These peaks become quite broad and
present some integration difficulties as
they become broader (graph a).
1.2 Migration Rates of Solute pdf version.pdf

More Related Content

Similar to 1.2 Migration Rates of Solute pdf version.pdf

Gas Chromatography by Dr. A. Amsavel
Gas Chromatography  by Dr. A. AmsavelGas Chromatography  by Dr. A. Amsavel
Gas Chromatography by Dr. A. AmsavelDr. Amsavel A
 
Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysispatil17
 
Lecture 3 analytical chemistry 4-second stage
Lecture 3 analytical chemistry 4-second stageLecture 3 analytical chemistry 4-second stage
Lecture 3 analytical chemistry 4-second stageAshraf Saad
 
Qualitative errors in hplc (1)
Qualitative errors in hplc (1)Qualitative errors in hplc (1)
Qualitative errors in hplc (1)Gaurav Kr
 
Chromatography - General and Theory.ppt
Chromatography - General and Theory.pptChromatography - General and Theory.ppt
Chromatography - General and Theory.pptNagoya University
 
Lec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.pptLec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.pptShahinaPanah
 
Hplc parameters, factors affecting resolution
Hplc parameters, factors affecting resolution  Hplc parameters, factors affecting resolution
Hplc parameters, factors affecting resolution DHINESHKUMAR V
 
Reverse phsase chromatography 2
Reverse phsase chromatography 2Reverse phsase chromatography 2
Reverse phsase chromatography 2Amol Sagulale
 
Gas chromshhd .ppt.pptx
Gas chromshhd .ppt.pptxGas chromshhd .ppt.pptx
Gas chromshhd .ppt.pptxssuser3be34e
 
Gas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfGas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfSherif Taha
 
GC_Introduce_and_starter_for_beginer.ppt
GC_Introduce_and_starter_for_beginer.pptGC_Introduce_and_starter_for_beginer.ppt
GC_Introduce_and_starter_for_beginer.pptAnonymousw7gZFE8
 
Theory of chromatographic separations
Theory of chromatographic separationsTheory of chromatographic separations
Theory of chromatographic separationsprashik shimpi
 

Similar to 1.2 Migration Rates of Solute pdf version.pdf (20)

Chromatography
Chromatography Chromatography
Chromatography
 
Gas Chromatography by Dr. A. Amsavel
Gas Chromatography  by Dr. A. AmsavelGas Chromatography  by Dr. A. Amsavel
Gas Chromatography by Dr. A. Amsavel
 
Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysis
 
Lecture 3 analytical chemistry 4-second stage
Lecture 3 analytical chemistry 4-second stageLecture 3 analytical chemistry 4-second stage
Lecture 3 analytical chemistry 4-second stage
 
Qualitative errors in hplc (1)
Qualitative errors in hplc (1)Qualitative errors in hplc (1)
Qualitative errors in hplc (1)
 
Chromatography - General and Theory.ppt
Chromatography - General and Theory.pptChromatography - General and Theory.ppt
Chromatography - General and Theory.ppt
 
Lec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.pptLec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.ppt
 
Hplc parameters, factors affecting resolution
Hplc parameters, factors affecting resolution  Hplc parameters, factors affecting resolution
Hplc parameters, factors affecting resolution
 
HPLC.6487473.powerpoint.pptx
HPLC.6487473.powerpoint.pptxHPLC.6487473.powerpoint.pptx
HPLC.6487473.powerpoint.pptx
 
Reverse phsase chromatography 2
Reverse phsase chromatography 2Reverse phsase chromatography 2
Reverse phsase chromatography 2
 
Gas chromshhd .ppt.pptx
Gas chromshhd .ppt.pptxGas chromshhd .ppt.pptx
Gas chromshhd .ppt.pptx
 
GC.ppt.ppt
GC.ppt.pptGC.ppt.ppt
GC.ppt.ppt
 
GC.ppt.ppt
GC.ppt.pptGC.ppt.ppt
GC.ppt.ppt
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Gas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfGas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdf
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
GC_Introduce_and_starter_for_beginer.ppt
GC_Introduce_and_starter_for_beginer.pptGC_Introduce_and_starter_for_beginer.ppt
GC_Introduce_and_starter_for_beginer.ppt
 
Theory of chromatographic separations
Theory of chromatographic separationsTheory of chromatographic separations
Theory of chromatographic separations
 
Fritz & scott
Fritz & scottFritz & scott
Fritz & scott
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 

Recently uploaded

Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.aasikanpl
 
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.aasikanpl
 
Temporomandibular joint Muscles of Mastication
Temporomandibular joint Muscles of MasticationTemporomandibular joint Muscles of Mastication
Temporomandibular joint Muscles of Masticationvidulajaib
 
Artificial Intelligence In Microbiology by Dr. Prince C P
Artificial Intelligence In Microbiology by Dr. Prince C PArtificial Intelligence In Microbiology by Dr. Prince C P
Artificial Intelligence In Microbiology by Dr. Prince C PPRINCE C P
 
Heredity: Inheritance and Variation of Traits
Heredity: Inheritance and Variation of TraitsHeredity: Inheritance and Variation of Traits
Heredity: Inheritance and Variation of TraitsCharlene Llagas
 
Gas_Laws_powerpoint_notes.ppt for grade 10
Gas_Laws_powerpoint_notes.ppt for grade 10Gas_Laws_powerpoint_notes.ppt for grade 10
Gas_Laws_powerpoint_notes.ppt for grade 10ROLANARIBATO3
 
Solution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsSolution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsHajira Mahmood
 
‏‏VIRUS - 123455555555555555555555555555555555555555
‏‏VIRUS -  123455555555555555555555555555555555555555‏‏VIRUS -  123455555555555555555555555555555555555555
‏‏VIRUS - 123455555555555555555555555555555555555555kikilily0909
 
insect anatomy and insect body wall and their physiology
insect anatomy and insect body wall and their  physiologyinsect anatomy and insect body wall and their  physiology
insect anatomy and insect body wall and their physiologyDrAnita Sharma
 
Twin's paradox experiment is a meassurement of the extra dimensions.pptx
Twin's paradox experiment is a meassurement of the extra dimensions.pptxTwin's paradox experiment is a meassurement of the extra dimensions.pptx
Twin's paradox experiment is a meassurement of the extra dimensions.pptxEran Akiva Sinbar
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxpriyankatabhane
 
Bentham & Hooker's Classification. along with the merits and demerits of the ...
Bentham & Hooker's Classification. along with the merits and demerits of the ...Bentham & Hooker's Classification. along with the merits and demerits of the ...
Bentham & Hooker's Classification. along with the merits and demerits of the ...Nistarini College, Purulia (W.B) India
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfSELF-EXPLANATORY
 
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.aasikanpl
 
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxTHE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxNandakishor Bhaurao Deshmukh
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxFarihaAbdulRasheed
 
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tanta
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tantaDashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tanta
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tantaPraksha3
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.PraveenaKalaiselvan1
 
Transposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptTransposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptArshadWarsi13
 

Recently uploaded (20)

Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Munirka Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
 
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Aiims Metro Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
 
Engler and Prantl system of classification in plant taxonomy
Engler and Prantl system of classification in plant taxonomyEngler and Prantl system of classification in plant taxonomy
Engler and Prantl system of classification in plant taxonomy
 
Temporomandibular joint Muscles of Mastication
Temporomandibular joint Muscles of MasticationTemporomandibular joint Muscles of Mastication
Temporomandibular joint Muscles of Mastication
 
Artificial Intelligence In Microbiology by Dr. Prince C P
Artificial Intelligence In Microbiology by Dr. Prince C PArtificial Intelligence In Microbiology by Dr. Prince C P
Artificial Intelligence In Microbiology by Dr. Prince C P
 
Heredity: Inheritance and Variation of Traits
Heredity: Inheritance and Variation of TraitsHeredity: Inheritance and Variation of Traits
Heredity: Inheritance and Variation of Traits
 
Gas_Laws_powerpoint_notes.ppt for grade 10
Gas_Laws_powerpoint_notes.ppt for grade 10Gas_Laws_powerpoint_notes.ppt for grade 10
Gas_Laws_powerpoint_notes.ppt for grade 10
 
Solution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsSolution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutions
 
‏‏VIRUS - 123455555555555555555555555555555555555555
‏‏VIRUS -  123455555555555555555555555555555555555555‏‏VIRUS -  123455555555555555555555555555555555555555
‏‏VIRUS - 123455555555555555555555555555555555555555
 
insect anatomy and insect body wall and their physiology
insect anatomy and insect body wall and their  physiologyinsect anatomy and insect body wall and their  physiology
insect anatomy and insect body wall and their physiology
 
Twin's paradox experiment is a meassurement of the extra dimensions.pptx
Twin's paradox experiment is a meassurement of the extra dimensions.pptxTwin's paradox experiment is a meassurement of the extra dimensions.pptx
Twin's paradox experiment is a meassurement of the extra dimensions.pptx
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptx
 
Bentham & Hooker's Classification. along with the merits and demerits of the ...
Bentham & Hooker's Classification. along with the merits and demerits of the ...Bentham & Hooker's Classification. along with the merits and demerits of the ...
Bentham & Hooker's Classification. along with the merits and demerits of the ...
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
 
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
Call Girls in Hauz Khas Delhi 💯Call Us 🔝9953322196🔝 💯Escort.
 
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptxTHE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
THE ROLE OF PHARMACOGNOSY IN TRADITIONAL AND MODERN SYSTEM OF MEDICINE.pptx
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
 
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tanta
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tantaDashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tanta
Dashanga agada a formulation of Agada tantra dealt in 3 Rd year bams agada tanta
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
 
Transposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptTransposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.ppt
 

1.2 Migration Rates of Solute pdf version.pdf

  • 2. PST351-POLYMER CHARACTERISATION BACKGROU ND • Chromatogram is useful for both qualitative and quantitative analysis. • The positions of peaks on the time axis - identify the components of the sample; • the areas under the peaks-quantitative measure of the amount of each component.
  • 3. PST351-POLYMER CHARACTERISATION BACKGROU ND Graph showing detector response as a function of elution time.
  • 4. PST351-POLYMER CHARACTERISATION • The effectiveness of a chromatographic column in separating two solutes depends in part on the relative rates at which the two species are eluted. • These rates are determined by the magnitude of the equilibrium constants for the reactions by which the solutes distribute themselves between the mobile and stationary phases. BACKGROUND
  • 5. PST351-POLYMER CHARACTERISATION DISTRIBUTION CONSTANT The distribution equilibria involved in chromatography involve the transfer of an analyte between the mobile and stationary phases. The equilibrium constant (Kc ) for this reaction is called the distribution constant, which is defined as; Where; cs is the molar concentration of the solute in the stationary phase cm is its molar concentration in the mobile phase K is constant over a wide range of solute concentrations.
  • 6. PST351-POLYMER CHARACTERISATION The Dead Time (tM), Stationary Time (tS), & Retention Time (tR) ▪ The time between sample injection and the appearance of the non-retained species peak is called as the dead or void time, (tM). ▪ It is the time it takes for an unretained species to pass through chromatography column. ▪ All components spend the dead time in the mobile phase ▪ Separation are based on the different times, ts that components spend in stationary phase ▪ The dead time provides a measure of the average rate of migration of a mobile phase and is an important parameter to identifying analyte peaks.
  • 7. PST351-POLYMER CHARACTERISATION The time it takes after sample injection for the analyte peak to reach the detector is called the retention time , (tR). The analyte has been retained because it spends a time in the stationary phase, (ts) The retention time is then calculated by; The Dead Time (tM), Stationary Time (tS), & Retention Time (tR)
  • 8. PST351-POLYMER CHARACTERISATION The Dead Time (tM), Stationary Time (tS), & Retention Time (tR)
  • 9. PST351-POLYMER CHARACTERISATION Retention Factor (k’) The retention factor (capacity factor), k’ is an important parameter that is widely used to describe the migration rates of solutes on columns For a solute A, the retention factor k`A is defined as; where; k’A is the distribution constant for the species A. tR and tM are readily obtained from a chromatogram
  • 10. PST351-POLYMER CHARACTERISATION Retention Factor (k’) When the retention factor for a solute is much less than this range (1- 5), seperation occurs so rapidly that accurate determination of the retention times is difficult. When the retention factor is larger (20 to 30) , seperation times become inordinately long.
  • 11. PST351-POLYMER CHARACTERISATION Selectivity Factor,  The selectivity factor,  of a column for the two species A and B is defined as; where; KB is the distribution constant for species B KA is the distribution constant for species A A relationship between the selectivity factor and retention factors can be described as: Where; k`B is the retention factor of species A. k`A is the retention factor of species B.
  • 12. PST351-POLYMER CHARACTERISATION The theoretical Plate Model of Chromatography • The plate model supposes that the chromatographic column is contains a large number of separate layers, called theoretical plates. • Separate equilibrations of the sample between the stationary and mobile phase occur in these "plates". • The analyte moves down the column by transfer of equilibrated mobile phase from one plate to the next.
  • 13. PST351-POLYMER CHARACTERISATION Column Efficiency To obtain optimal separations, sharp, symmetrical chromatographic peaks must be obtained. This means that band broadening must be limited. It is also beneficial to measure the efficiency of the column. Two related terms are widely used as quantitative measures of chromatographic column efficiency Plate Height, H Number of Plate, N The two are related by the equation; where L is the length (usually in centimeters) of the column packing. The efficiency of chromatographic columns can be increases by • Increase the number of plates, N • Reduce the plate height N= L/H
  • 14. PST351-POLYMER CHARACTERISATION Column Efficiency • No of plate, N = the more plates the better • Plate height, H = the smaller the better
  • 15. PST351-POLYMER CHARACTERISATION The number of theoretical plates that a real column possesses can be found by examining a chromatographic peak after elution; Column Efficiency As can be seen from this equation, columns behave as if they have different numbers of plates for different solutes in a mixture. where w1/2 is the peak width at half-height.
  • 16. PST351-POLYMER CHARACTERISATION Column Resolution (Rs) The resolution Rs of a column provides a quantitative measure of its ability to separate two analytes. The resolution of two species, A and B, is defined as Baseline resolution is achieved when R = 1.5
  • 17. PST351-POLYMER CHARACTERISATION It is useful to relate the resolution to the number of plates in the column, the selectivity factor and the retention factors of the two solutes; Column Resolution (Rs) From this equation, it is known that to obtain high resolution, the three terms must be maximised. • An increase in N, the number of theoretical plates, • Reduce the column height by - reducing the size of the stationary phase particles.
  • 18. PST351-POLYMER CHARACTERISATION Column Resolution (Rs) Apart from improving column resolution to improve separation, it is often found that by controlling the capacity factor, k', separations can be greatly improved. This can be achieved by: • Changing the temperature (in Gas Chromatography) or • Changing the composition of the mobile phase (in Liquid Chromatography).
  • 19. PST351-POLYMER CHARACTERISATION The selectivity factor,  can also be manipulated to improve separations. In these cases, k' is optimised first, and then a is increased by one of the following procedures: 1. Changing mobile phase composition 2. Changing column temperature 3. Changing composition of stationary phase 4. Using special chemical effects (such as incorporating a species which complexes with one of the solutes into the stationary phase) Column Resolution (Rs)
  • 20. PST351-POLYMER CHARACTERISATION Column Temperature • The retention time of a compound depends not only on the type of stationary phase but also the column temperature. • In general you can analyze your sample under isothermal conditions in which the temp of the column remains constant throughout the analysis. • As the temperature increases, compounds move through the column faster and consequently, the retention time decreases. • Thus, we can potentially reduce the analysis time by increasing column temperature; however, we need to allow sufficient time for the compounds to interact with the stationary to provide the required separation of sample components.
  • 21. PST351-POLYMER CHARACTERISATION Column Temperature In the simplest GC analysis, the column is maintained at a constant temperature and is termed “isothermal” analysis. The chromatogram in this figure shows the separation of a mixture of normal alkanes (straight-chained hydrocarbons) ranging from C6 to C21 using isothermal conditions at 150 C. The low molecular weight alkanes <C10 elute very quickly and are not resolved. The higher molecular weight components >C16 do not elute from the column at this temperature and C14 and C15 are broad, asymmetrical peaks.
  • 22. PST351-POLYMER CHARACTERISATION Column Temperature The chromatogram shows the same alkane mixture analyzed using “temperature programmed” conditions. In this case, the column temperature is increased from 50 to 250 C at 8 C/ min during the analysis. All of the alkane components are completely separated into narrow, symmetrical peaks.
  • 23. PST351-POLYMER CHARACTERISATION Column Temperature Comparing that separation of the same mixture with the lower chromatogram which is temperature programmed from 50 degrees to 250 degrees at 8 degrees per minute , it can be seen that; - The separation efficiency is much greater here (graph b). - Higher molecular weigh components longer chain hydrocarbons that being eluted In a conclusion, it can be said that; - The longer the compound is in the column the broader the peaks become. - These peaks become quite broad and present some integration difficulties as they become broader (graph a).