CONTENTS
• Introduction
• Basic principles of elastography
• Methods and Types of elastography
• Applications of elastography
• Pitfalls and limitations
• Recent Advances and future prospects
• Conclusion
• References and further reading
INTRODUCTION
Ever since Hippocrates, physicians have used palpation to
detect differences in tissue stiffness as an aid to diagnosis.
Elastography is a novel imaging modality which depends on
the same differences in mechanical properties between
healthy and abnormal tissues but uses imaging to detect
these differences at depths not reachable by manual
palpation.
Elastography techniques have been developed for both
ultrasound and MR imaging.
–Sono-Elasticity Lerner 1990
–Elastography Ophir 1991
–Shear Waves Elastography Sarvazyan1997
–Vibro-acoustography Fatemi 1998
–Transient Elastography Sandrin 1999
–Acoustic Radiation Force Imaging(ARFI) Nightingale 2001
–Supersonic Shear Imaging Elastography Bercoff 2004
– MR Elastography Muthupillai1995
TIMELINE REVIEW
BASIC PRINCIPLES OF ELASTOGRAPHY
Elastogarphy is a tissue elasticity imaging, hence it is important
to understand the general principles which govern the
behaviour of an elastic material when acted upon by an external
force.
When external forces are applied to objects made up of elastic
materials, they undergo a change in shape and size.
STRAIN is a relative change in shape or size of an object due to
externally applied forces.
STRESS is the internal force ( per unit area) associated with
strain
HOOK’S LAW states that strain is directly proportional to stress
for a linearly isotropic elastic material
e= (h-ho)/ho
Strain = Change in length/original length
e= / E
e = Strain (dimensionless)
 = Stress (Pascal) = Force / Area
E = Young’s Modulus (Pascal)
Elastic Moduli
YOUNG’S MODULUS ( E ) : It is the ratio of stress along an axis to
the strain along that axis in the range of stress in which hooke’s law
holds.
BULK MODULUS (K or B) : It is the ratio of infinitesimal pressure
increase to the resultant relative volume decrease.
SHEAR MODULUS ( G or S or m ) : It is the ratio of shear stress to
shear strain. It concerns with deformation of solids when two opposing
surfaces experience forces in opposing directions.
YOUNG’S MODULUS (E) BULK MODULUS ( K OR B)
SHEAR MODULUS ( G or m)
Conventional ultrasound
• Propagation of ultrasound depends on bulk modulus
• Bulk modulus is homogeneous in different tissues
• No information provided regarding tissue stiffness
Elastography
• Imaging of tissue stiffness
• Estimation of strain in tissues under stress
• Imaging of shear waves whose propagation is governed
by tissue stiffness or Young’s modulus(E)
• Young’s modulus exhibits important differences in
tissues which is ideal for characterization of tissues with
excellent contrast.
• Young’s modulus is a quantitative estimation of
clinician’s palpation and has diagnostic value
ILLUSTRATIVE EXAMPLE
dA dB
tA= 2dA/ c tB= 2dB/ c
c = Speed of sound
Displacement ( d) = dB – dA = c ( t B-t A)/2
d = c/2 ( = t B – t A )
METHODS AND TYPES OF ELASTOGRAPHY
Two established methods of elastography
• ULTRASOUND ELASTOGRAPHY
• MAGNETIC RESONANCE ELASTOGRAPHY
ULTRASOUND ELASTOGRAPHY
METHODS
• QUASI-STATIC
• DYNAMIC
QUASI-STATIC METHODS
In the case of quasi-static elastography, a constant stress is applied to
the tissue. The displacement and the generated strain are then
estimated using two-dimensional correlation of ultrasound images. The
Young’s modulus is then given via the Hooke’s law , which links stress
and strain in a purely elastic medium.
In practice, since the applied stress is unknown, only the strain is
displayed, this strain map is sometimes called the elastogram.
This was the first elastography technique, developed by the Ophir Group
at the beginning of the 1990s. Today, the technique is being tested in
breast lesion characterization with interesting results.
The main limitations of this technique are
• the control of the stress applied, which remains operator
dependent,
• the absence of a specific quantification.
• the use of a stress applied by the operator limits the technique to
superficial organs, mainly the breast or the thyroid.
This technique is simple to implement and is widely spread in the
world of radiology, and has been mainly validated for breast lesions
classification
DYNAMIC METHODS
In dynamic methods, a time-varying force is applied to the tissue, it can
either be a short transient mechanical force or an oscillatory force with a
fixed frequency.
A time-varying mechanical perturbation will propagate as mechanical
waves which in a solid body can be compressional waves or shear
waves.
The compressional waves propagate very quickly in the human body
(∼1500 m/s), and at high frequencies, these waves, also known as
ultrasounds, can be used to image the body.
Shear waves, which are only generated at low frequencies (10 Hz to
2000 Hz) due to absorption at higher frequencies, propagate more
slowly, and their speed (∼1—50 m/s) is directly related to the medium
shear modulus.
In biological tissues, which are almost incompressible, the Young’s
modulus can be approximated as three times the shear modulus. The
shear wave propagation speed can thus be used to map the Young’s
modulus quantitatively.
Dynamic elastography techniques, which rely on shear waves
propagation, can produce quantitative and higher resolution Young’s
modulus map compared to quasi-static methods.
The use of shear waves, however, requires a more complex system,
able to generate the shear wave (mechanical vibrator or ultrasound
radiation pressure) and to image the small displacements induced by the
shear wave (ultrafast or stroboscopic ultrasound).
VIBRO-ACOUSTOGRAPHY
Vibro-acoustography is a dynamic elastography method based on
the ultrasound radiation pressure and developed by the American
Group of James Greenleaf
The radiation pressure is a volumic force created by transfer of
momentum within the medium This momentum is related to the
absorption of the ultrasound wave.
Vibro-accoustography uses two confocal ultrasound beams with
slightly different frequencies w and w0 + w. It results in beats at the
frequency w , which give rise to a modulated force at frequency w,
only at the focal spot. Therefore, everything looks like if
the target vibrates at the frequency w.
It is then sufficient to listen to the sound produced by this excitation
to deduce the mechanical properties of the target, particularly its
stiffness.
For this, one can place a hydrophone that records the response of
the target at the frequency w . In order to create an image, the
entire area is swept by moving the focal spot and listening to the
response for each point of the image.
However the sweeping of the entire image causes an important
deposit of energy within the medium and a long acquisition time,
which makes its implementation in real-time and in vivo difficult.
In particular, the measured parameter depends on the stiffness
of the zone, but also on the amplitude of the generated force and
the geometry of the vibrating object. Therefore images
correspond to a mixture of several physical parameters,
including elasticity.
Greenleaf’s team is now trying to combine this method with
mammography systems for the diagnosis of breast cancer
and with conventional ultrasound machines
TRANSIENT ELASTOGRAPHY METHODS
Acoustic Radiation Force Impulse Imaging (ARFI) or ‘‘Acoustic Radiation
Force Imaging’’, is a method developed by the American team of Kathy
Nightingale.
This technique uses the acoustic radiation force but, unlike vibro -
acoustography, ARFI only uses one focalized ultrasound beam.
The radiation force slightly displaces the tissue at the focal spot according
to Hooke’s law. Then the transducer switches into imaging mode and
detect displacements of the focal spot by tracking of the ultrasound signal
(called ‘‘speckle’’). The ultrasound speckle-tracking , already used in static
elastography, allows to correlate the ultrasound signal window by window
in order to detect the displacement of a tissue with sensitivity of less than a
micrometer. By window, I mean a piece of an ultrasound signal.
Today, Nightingale’s team is interested in the propagation of shear
waves generated by the radiation force and has recently proposed a
new ARFI model called ‘‘ARFI-SWS’’.
Based on this concept, it allows to quantitatively measure the Young’s
modulus in a small region of interest. This variation is currently being
evaluated for the liver staging and is available on commercial
ultrasound systems, such as the Siemens Acuson S2000.
1D transient elastography: the 1D shear elasticity probe
The 1D transient elastography probe was first developed at the Institut
Langevin in 1995 by Catheline et al.
It consists of generating a transient impulse (little shock) on the medium
and recording the shear wave that propagates within the medium by using
an ultrasound transducer.
First, the front face of the transducer acting as a piston gives a slight
mechanical impulse on the surface of medium, which generates a
spherical compression wave as well as a spherical shear wave.
The displacement generated, which is a function of depth and of time, is
thus estimated by correlations of retro-diffused echoes (via ultrasound
speckle) recorded at a framerate higher than one thousand time per
second with a mono-dimensional ultrasound transducer (5 MHz).
This device was the first to use the principle of ultrafast imaging in one
dimension to visualize in a transient manner the propagation of shear
waves.
Finally, by measuring the phase for each depth, we extract the phase
speed of the shear wave at the central frequency, leading to an
estimation of Young’s modulus by considering the medium to be
homogeneous and nonviscous.
Since 2001, the company Echosens has been commercializing this
technique, 1D transient elastography, under the name FibroScan.
2D transient elastography
In 1997, at the ‘‘Institut Langevin’’, the 1D transient elastography
technique was extended to 2D, allowing the creation of elasticity maps of
biological tissues.
A programmable ultrasound electronic device used for time reversal
experiments (process of refocalizing acoustic waves) was modified to be
able to perform ultrafast imaging based on ultrasound plane wave
emission.
It allowed storing raw data acquired with a frame rate of more than 5000
images per second. A vibrator was fixed to the ultrasound imaging array,
which is then used as an impactor to generate a quasi-plane shear wave.
Once the movie of the propagating shear wave is reconstructed, the wave
equation is inverted to recover a map of Young’s modulus.
The first in vivo tests were carried out in 2003 with volunteers at Institut
Curie, the results were encouraging, but the device was bulky, heavy and
difficult to use in practice.
The study of shear wave by using ultrafast imaging
The Supersonic Shear Imaging technique is the outcome of researches at
the Institute Langevin in transient elastography.
The idea of combining radiation pressure or acoustic radiation force and the
study of shear waves generated comes from Armen Sarvazyan, which can
be considered as a precursor of elastography techniques based on
ultrasonic radiation pressure via its technique: Shear Wave Elasticity
Imaging.
In 2004, two fundamental ideas are developed to overcome the limitations
of 2D elastography technique:
• acoustic radiation force
• ultrasound ultrafast imaging.
These two concepts described below are at the heart of the Supersonic
Shear Imaging technique
• a Mach cone: ultrasound beams are successively focused at different
depths. The different spherical waves generated for each focal beam
interfere like a Mach cone in which the source propagates more quickly
than the generated shear wave and creates a quasi-plane wave front in
the imaging plane (cylindrical in three dimensions).
The use of constructive interfaces makes it possible to increase the
amplitude of the wave and thus the signal to noise ratio of the
displacement field.
The generated quasi-plane shear wave in the imaging plane allows also
simplifying the propagation hypotheses, which is of great interesting to
solve inverse problems.
Finally, only one Mach cone allows illuminating almost all of the medium
with one plane shear wave; shear wave being generated, the propagation
equation of the waves is inverted to rebuild a map of Young’s modulus;
• ultrafast imaging: complete acquisition all at once. Ultrafast imaging
allows scanning the entire imaging plane with very good temporal
resolution in one single acquisition, typically with a frame rate of 5000
images per second, and up to 30,000 images per second in the case
of tissues such as the peripheral arteries or the eye.
Therefore there is no need to repeat the acquisition several times by
stroboscopy to acquire the entire displacement field. This allows, not
only imaging in real-time, which makes the examination easier, but
also averaging the very rapidly acquired images to improve image
quality.
Therefore, the Supersonic Shear Imaging technique uses the technology of
the 2D transient elastography, but substitutes vibrator by the acoustic
radiation pressure.
The whole excitation-imaging method is then integrated onto one single
component: the ultrasound imaging transducer array. Amplified by the Mach
cone, the generated shear wave has amplitude of tens of microns. This latter
is detectable with a good signal to noise ratio by ultrasound speckletracking
algorithm and ultrafast imaging.
Thanks to ultrafast imaging, the acquisition of the shear wave propagation can
be carried out all at once in less than 30 milliseconds. The technique is
therefore slightly sensitive to patient movements (as an example breathing)
and can be displayed in real-time, like a conventional ultrasound image.
The Young’s modulus maps are then reconstructed by estimating the speed
of the shear wave between two points of the image, using a time of flight
algorithm.
The first in vivo experiences with this technique were conducted at the Institut
Curie where, in approximately 50 patients, the use of the technique to
differentiate benign tissue from malignant tissue was demonstrated
This technique was implemented on an ultrasound diagnostic imaging
device called, the Aixplorer (Supersonic Imagine, Aix-en-Provence,
France) and its diagnosis performance as well as its reproducibility were
demonstrated in several organs, and more particularly in the breast with
the addition of the elasticity parameters to the classical BIRADs criteria
Shear wave
excitation
Measurement
of shear wave
propagation
Shear modulus
reconstruction
Shear
wave
Force
Probe Probe
Tx Echo Rx
Displacement
of
reflectors
Tissue displacement map
Shear wave
speed
calculation
Shear modulus
Shear modulus
Shear wave speed
Density
Tx = transmission
Rx = receiver
Multiple M-mode or B-mode images: tissue displacement map
MAGNETIC RESONANCE ELASTOGRAPHY
Magnetic Resonance Elastography (MRE) is a rapidly developing technology
for quantitatively assessing the mechanical properties of tissue.
MRE obtains information about the stiffness of tissue by assessing the
propagation of mechanical waves through the tissue with a special magnetic
resonance imaging (MRI) technique. The technique essentially involves three
steps:
1. generating shear waves in the tissue,
2. acquiring MR images depicting the propagation of the induced shear waves
and
3. processing the images of the shear waves to generate quantitative maps of
tissue stiffness, called elastograms.
Generating Mechanical Waves in Tissue
MRE typically uses vibrations of a single frequency (within the audio frequency
range) generated by external driver devices. The electrical signal for these devices
is created by a signal generator triggered by and synchronized to the MR pulse
sequence and is amplified by an audio amplifier before being fed into the
mechanical driver
Over the years, several driving mechanisms have been developed, each with their
own advantages and limitations. Three of the most commonly used driver
systems are:
• an electromechanical driver that works via the Lorentz force and
utilizes the magnetic field of the main MRI magnet.
• A piezoelectric stack driver system where the motion created is based
on the piezoelectric property of certain materials. Focused-ultrasound-based
(FUSbased) radiation force has also been investigated as a means to create
mechanical motion for various elasticity imaging strategies including MRE,
where shear waves are created directly within tissue with externally placed
ultrasound transducers.
• pressure-activated driver another widely used method of creating the
required vibrations for MRE utilizes the motion of the voice coils used in
acoustic speaker systems. The required vibrations are again produced by the
Lorentz force, but the static magnetic field is from a devoted permanent magnet
present in the acoustic speaker .
Imaging the Propagating Waves
Measuring the tissue motion produced by a driver with MRE is based on an MR
imaging technique called phase-contrast MRI.
Muthupillai et al. developed the technique of dynamic phase-contrast MRE
where propagating shear waves in tissue are encoded into the phase of the MR
images with the help of motion encoding gradient (MEG) pair.
After continuous harmonic motion is induced in the tissue, a MEG oscillating at
the same frequency as the motion is applied and conventional MR imaging is
performed. The phase contribution to the MR image ϕ due to the motion and the
applied magnetic field gradient at a given position vector r⃗ and phase offset θ
between the motion and the MEG can be written as.
where γ is the gyromagnetic ratio of the tissue protons , N is the number of
gradient pairs used to sensitize the motion, T is the period of the motion
encoding gradient, G⃗ is its amplitude, ξ⃗0 is the peak amplitude of motion,
and k⃗ is the wave number. This equation states that the phase of
harmonically vibrating tissue is directly proportional to its displacement.
Figure shows a typical MRE pulse sequence using a gradient-recalled echo with
the conventional radiofrequency (RF) pulse waveform, slice-selection gradient,
phase-encoding gradient and frequency-encoding gradient.
The motion-encoding gradient (shown here only in the frequency-encoding
direction) is placed after the RF excitation of the sample and before the
measurement of the induced signal.
Motion occurring in any direction can be encoded into the phase of the MR
image by manipulating the axes on which the MEGs are placed. In this example,
only the motion occurring in the frequency-encoding direction will be sensitized
to and encoded into the image phase.
The motion-encoding capability of this technique is very sensitive and can detect
motion on the order of 100’s of nanometers.
An MR image thus obtained containing information about the propagating wave
in its phase is called a wave image. Typically two such wave images are collected
with opposite polarity of the MEG and a phase-difference image is calculated to
remove non-motion-related phase information.
Since the motion-encoding gradients necessary for MRE are inserted into
conventional MR pulse sequences, MRE can be implemented with many MR
imaging sequences, each with its own advantages and limitations. MRE pulse
sequences based on spin echo (SE), gradient-recalled echo (GRE), balanced
steady-state free precision (bSSFP) and echo planar imaging (EPI) techniques
exist.
Mechanical Parameter Estimation
From the wave images indicating the propagation of shear waves in the tissue,
mathematical inversion algorithms based on equations of motion, with
simplifying assumptions like isotropy, homogeneity, and incompressibility, allow
for the calculation of mechanical properties like the shear modulus to be used for
clinical interpretations.
The images of the mechanical properties of tissue calculated in MRE are often
referred to as elastograms. Depending on the technique used to derive the
elastograms from the original MR images, elastograms can theoretically (e.g.,
absent any noise) have half the resolution of the native MR images (which can
range from 50 μm to 10 mm depending on the application), however they more
typically have one-third to one-fifth of the MRI resolution.
Shear wave interference can cause artifacts in the stiffness calculations using
several of the above techniques, and a preprocessing technique called
directional filtering has been developed to reduce these artifacts
MRE of an inclusion phantom
(a) MR magnitude image of an inclusion phantom with soft and stiff inclusions
seen as the hyperintense and hypointense regions, respectively.
(b) A single wave image from the MRE acquisition performed at 100 Hz. The
difference in the wavelengths in the different regions is evident.
(c) An elastogram obtained from these data showing the stiff and soft regions.
APPLICATIONS OF ELASTOGRAPHY
ULTRASOUND ELASTOGRAPHY
Breast US Elastography
Compression elastography may have a role for further evaluating
abnormal findings on conventional breast US images and differentiating
benign from malignant breast lesions. Thus, the eventual goal of
incorporating US compression elastography into routine practice would
be to reduce the biopsy rate of benign lesions.
The two most important elastographic characteristics in evaluating breast
lesions are size and stiffness criteria
Normal Breast Tissue
Biomechanical testing has shown normal fibroglandular breast tissue to be
markedly stiffer than normal fatty breast tissue by as much as two orders of
magnitude . Therefore, at elastography, fatty tissue will appear bright with
respect to the adjacent glandular tissue, and normal fibrous parenchyma
appears darker
(a) B-mode US image of the breast shows that a lobule of normal fatty tissue (f) is
hypoechoic with respect to the surrounding glandular tissue (g).
(b) US elastogram shows that the fatty tissue (f) surrounded by dense breast
tissue appears bright because it is appreciably “softer” than the surrounding
glandular tissue (g).
Simple cyst. (a) B-mode US image of the breast shows a classic
cyst (arrowheads). (b) US elastogram demonstrates a bull’s-eye
configuration within the lesion (arrowheads), which contains
bright compressible material (thin arrow), and a bright spot
behind the cyst (thick arrow).
Complex cyst mimicking a solid lesion at US. (a) B-mode US image
demonstrates a round hypoechoic breast lesion without associated
through transmission. (b) US elastogram shows that the lesion that
appeared “solid” on the B-mode image is a cyst. .
Fibrocystic change. (a) B-mode US image demonstrates an irregular
heterogeneous lesion in a patient who presented with a palpable mass
in the right breast. (b) On the US elastogram, the lesion appears soft
and smaller than on the B-mode image. The findings from a subsequent
biopsy showed fibrocystic change and periductal inflammation. Dotted
lines indicate measured lesion diameters: 1 = 7.16 mm; 2 = 5.55 mm; 3
= 5.02 mm; and 4 = 4.29 mm. Distance ratio was calculated as 5.02 
7.16 = 0.70.
Fibroadenoma
Fibroadenoma. (a) B-mode US image of the left breast shows a
lobulated hypoechoic lesion that is taller than wide, with posterior
acoustic shadowing.
(b) US elastogram, however, shows the lesion to be smaller than on
the B-mode image. The findings from biopsy revealed a fibroadenoma.
Fibroadenoma with elasticity score of 1 in 51-year-old woman. US images were
obtained in transverse plane. Left: On conventional B-mode image, lesion was
classified as BI-RADS category 2.
Right: On elasticity image, the entire hypoechoic lesion was evenly shaded green,
as was the surrounding breast tissue.
Fibroadenoma with elasticity score of 2 in 39-year-old woman. US images were
obtained in transverse plane.
Left: On conventional B-mode image, lesion was classified as BI-RADS category 3.
Right: On elasticity image, hypoechoic lesion shows mosaic pattern of green and
blue.
Invasive Ductal Carcinoma
Invasive ductal carcinoma. The breast lesion measures smaller on the B-mode
US image (a) than on the US elastogram (b) in the transverse plane. The
lesion is too large to measure accurately in the anteroposterior plane; dotted
lines indicate measurements of the lesion diameter: 21.9  34.2 mm in a and
24.9  29.6 mm in b. In addition, the lesion is stiffer than the surrounding
tissues.
invasive ductal carcinoma typically is appreciably darker than normal
tissues or benign lesions and is substantially larger on the elastogram
than on B-mode US images
Nonscirrhous type invasive ductal carcinoma with elasticity score of 4 in 29-year
old woman. US images were obtained in transverse plane.
Left: On conventional B-mode image, lesion was classified as BI-RADS category
5.
Middle: On elasticity image, the entire hypoechoic lesion was blue.
Right: Pathologic section of lesion is shown. (Hematoxylin-eosin stain; original
magnification,1.)
Scirrhous type invasive ductal carcinoma with elasticity score of 5 in 55-year-old
woman. US images were obtained in sagittal plane. Left: On conventional B-mode
Image, lesion was classified as BIRADS category 5. Middle: On elasticity image,
both the entire hypoechoic lesion and its surrounding area were blue. Right:
Pathologic section of lesion is shown. (Hematoxylin-eosin stain; original
magnification X 1.),
Lobular carcinoma in situ with elasticity score of 3 in 46-year-old woman. US
images were obtained in transverse plane. Left: On conventional B-mode image,
lesion was classified as BI-RADS category 3. Right: On elasticity image, the
central part of the hypoechoic lesion was blue, and the peripheral part of the lesion
was green.
Lymph node with metastatic involvement. (a) B-mode US image demonstrates a
lymph node that appears benign on the basis of its reniform shape and
echogenicity. Dotted line indicates a lesion diameter of 20.4 mm. (b)
Corresponding US elastogram, however, shows a stiff area that turned out to be
a metastatic focus (arrows). Dotted line indicates a lesion diameter of 24.0 mm.
Hematoma. (a) B-mode US image of the breast demonstrates a superficial
heterogeneous mass. (b) US elastogram shows that the lesion is stiff but
measures smaller than on the B-mode image. Dotted lines indicate
measurements of the lesion diameter: 11.0  5.3 mm in a and 10.2  4.2 mm in b.
The distance ratios were calculated as 10.2  11.0 = 0.93 (for the long-axis
dimension) and as 4.2  5.3 = 0.79 (for the short-axis dimension). At physical
examination, skin discoloration was noted in the corresponding area, a finding
consistent with a hematoma.
Prostate US Elastography
Prostate Cancer
Endorectal US continues to be the most commonly used imaging modality for
assessing the prostate; however, it is relatively inaccurate for cancer
detection. Endorectal US has a role in guiding prostate biopsies. Magnetic
resonance imaging and computed tomography are mainly implemented for
staging.
At B-mode US, most prostate cancers are hypoechoic, but many are
isoechoic, and a few are hyperechoic. As a result, the overall sensitivity of
endorectal US for prostate cancer detection is about 50%. The added use of
Doppler imaging increases the detection rate by only 5%. In contrast,
endorectal real-time elastography enables the diagnosis of prostate cancer
with a reported accuracy of 76%. However, the major role of elastography is
to improve the results of image-directed biopsy and therapy, compared with
the use of endorectal US alone.
Prostate adenocarcinoma with Gleason score of 6/10. (a) Sagittal endorectal US
Image with the prostate gland outlined shows no discrete lesion (arrow). (b) Sagittal
US elastogram shows strong vibration in the normal prostate tissue and also an area
of no vibration (arrow) anteriorly from the midportion of the gland to the base. (c)
Photomicrograph (hematoxylineosin stain) of a transverse histologic section that
included the corresponding lesion (arrow) at the base shows good correlation, with
the deficit noted anteriorly on the elastogram through the selected plane of section
(dashed line in a and b) in the midportion of the gland. The smaller lesion outlined in
blue in c is outside the plane of the elastogram.
Prostate adenocarcinoma with Gleason score of 9/10, involving two of two core
samples and approximately 60% of the tissue. (a) Right parasagittal conventional
US image shows that the lesion (arrow) is mildly hypoechoic. (b) Corresponding
right parasagittal US elastogram better depicts the lesion (arrow) as a dark area of
low strain. (c) Photomicrograph (hematoxylin-eosin stain) of a transverse histologic
section (dashed line in a and b indicates the plane of section) shows that there is
good size correlation between the histologic extent of the lesion (arrow) and the
lesion extent delineated on the elastogram.
Benign Prostatic Hyperplasia
The appearance of benign prostatic hyperplasia at endorectal US is variable but
usually consists of a heterogeneous hypoechoic area or areas in the transitional
zone .
In general, foci of benign prostatic hyperplasia have elastic moduli (stiffness)
that are an order of magnitude greater than those of normal prostate tissues but
are less than those of prostate carcinomas. As a result,on elastograms, benign
prostatic hyperplasia will appear darker than normal prostate tissue.
However, the difference between benign prostatic hyperplasia and prostate
carcinoma can be difficult to discern because benign prostatic hyperplasia also
appears darker than the background tissues. Consequently, benign prostatic
hyperplasia can represent a false-positive finding for cancer .
Benign prostatic hyperplasia. (a) Transverse B-mode US image demonstrates a
mildly prominent isoechoic area (arrow) in the central zone of the prostate gland.
(b) Corresponding transverse elastogram shows that
the area of benign prostatic hyperplasia (arrow) is more conspicuous. The lesion
demonstrates greater stiffness than the surrounding normal prostate tissue.
(c) Photomicrograph (hematoxylin-eosin stain) shows that there is good
correlation between the extent of the lesion (arrow) on the histologic section
and the lesion extent on the elastogram.
Elastography of Thyroid Lesions
A thyroid lesion may have different levels of stiffness within it, depending
on the cellularity and composition of the nodule. Information from these
elastograms helps assess the relative stiffness of the lesion compared
with its surrounding tissues and within itself. However, this information
cannot be used to compare the stiffness of thyroid lesions from different
patients because strain changes with applied compression.
When pulsation of the carotid artery is used as the compression source,
strain near the carotid artery can indicate the amount of compression
applied by carotid arterial pulsation.
Because the thyroid gland is located between the trachea and the carotid
artery, lateral expansion of the carotid artery during systole compresses
the thyroid gland against the trachea. As a result, the thyroid gland
expands in the anteroposterior direction, that is, in the direction of the US
probe (beam axis).
Because US is sensitive to detecting motion along the beam axis,
deformation of the thyroid gland in this direction can be readily detected.
Because the thyroid stiffness index is a ratio between the carotid strain
and the strain in the thyroid nodule, the varying strain from change in the
blood pressure did not affect the ratio.
Example of selecting region of interest (ROI) to compute thyroid stiffness index
within a nodule.
(a) Transverse US image showing a nodule (arrows) within thyroid gland (Tg),
the trachea (Tr), and the carotid artery (C).
(b) Combined US image and elastogram of this nodule; average strain of
stiffest region inside a thyroid nodule (ROI #1) is used as denominator of
thyroid stiffness index.
(c) Combined US image and elastogram of this nodule with a different scale to
better visualize the areas of highest strain near the carotid artery. Average
strain of the highest-strain regions near the carotid artery (ROI #2, ROI #3) is
used as numerator of thyroid stiffness index. Range of the color map is
adjusted to better visualize ROIs.
(a) Left: Transverse US image in a 62-year-old man shows a predominantly solid
nodule (arrows) occupying entire thyroid gland. Right: Elastogram shows the
predominantly homogeneous appearance within nodule, with small areas of
decreased stiffness (blue indicates stiffer part; red indicates softer part). This was
diagnosed as a follicular lesion at FNA.
(b) Left: Transverse US image in a 60-year-old woman shows a predominantly
solid nodule with small cystic areas (arrows). Right: Elastogram shows the
heterogeneous appearance within nodule. This was diagnosed as a nodular goiter
at FNA.
(c) Left: Transverse US image in a 36-year-old woman shows a hypoechoic
nodule (arrows) with irregular borders and tiny punctate calcifications in it. Right:
Elastogram at same level shows stiff areas within lesion (blue region). This was
diagnosed as papillary carcinoma at FNA. Ccarotid artery; Trtrachea.
Ultrasound elastography for musculoskeletal applications
Disease in the musculoskeletal system results in alterations to its
biomechanical properties. Although EUS techniques have been
extensively employed for in vitro research of muscle and tendon
biomechanics since the early 1990s, the recent introduction of EUS
into commercially available ultrasound systems has driven research
activity towards potential clinical applications of this novel method in
the musculoskeletal system
(a) (b)
(c) (d)
(a, b) Longitudinal and (c, d) transverse free-hand strain elastograms of the
middle third of asymptomatic Achilles tendons (T) showing the two distinct
ultrasound elastography patterns of normal tendons.
Type 1 (a, c) tendons appear homogeneously stiff, (green/blue) with no distinct
soft (red) areas.
Type 2 (b, d) tendons appear considerably inhomogeneous with soft (red) areas
(longitudinal bands or spots), which did not correspond to any changes in B-
mode ultrasound.
The retro- Achilles fat appears as a mosaic of green, red and blue. Note the red
areas at the lateral and medial sides of the tendon in the transverse plane (c, d),
corresponding to an artefact, secondary to difficulty in stabilising the hand-held
transducer.
(a) (b)
(c)
(a) Longitudinal colour Doppler ultrasound image and (b) free-hand strain
elastogram of a 23-year-old recreational runner with insertional Achilles
tendinopathy. There is hypoechogenicity and neovascularity at the tendinopathic
area [asterisk in (a)], which appears softer (red) compared with the stiff
(blue/green) normal-appearing remaining tendon (T). The retro-
Achilles fat appears as a mosaic of various levels of stiffness. (c) The
elastogram of a normal (asymptomatic) Achilles tendon calcaneal insertion is
presented for comparison. The retrocalcaneal bursa (asterisk) appears
considerably softer (red) compared with the stiffer tendon (green/yellow). Note
the presence of soft (red) foci beneath the calcaneal bone (C), which correspond
to artefacts. LT, left.
(a) (b)
(c) (d)
Longitudinal shear wave elastograms of a normal (a) Achilles and (c) patella
tendon, as well as (b, d) a case of distal patella tendinopathy in a 23-year-old
football player.
The elasticity qualitative and quantitative scale is presented at the upper right
corner of the images. Measurements (mean, minimum, maximum and
standard deviation) within the circular region of interest (ROI) are presented in
kilopascals ranging from 0 (dark blue) to 300 (dark red).
(a, c) The normal Achilles and patella tendons (T) appear as homogeneous
stiff (red) structures, as opposed to fat, which is homogeneously soft (blue).
(a) The mean stiffness of a representative area at the mid-portion of the
Achilles free tendon is 300 kPa. (d) In the case of distal patella
tendinopathy, the tendinopathic area appears hypoechoic with
neovascularity (asterisk).
(b) In the corresponding elastogram, the abnormal area appears softer (blue;
mean elasticity 40.94 kPa) compared with the stiffer normal tendon (red;
mean elasticity 261.16 kPa). The small amount of fluid in the deep infrapatella
bursa appears softer than the tendinopathic area (blue, mean elasticity 34.38
kPa).
(a) (b)
Axial free-hand strain elastograms of (a) normal relaxed calf muscles and (b) vastus
lateralis muscle.
The medial and lateral heads of gastrocnemius (M and L, respectively), as well as
the vastus lateralis (VL), appear as a mosaic of intermediate or increased stiffness
(green or blue colour, respectively), with scattered softer (red) areas near muscle
boundaries. The subcutaneous fat appears soft (red/yellow). The homogeneous stiff
appearance of soleus (S) in (a) is probably because of inadequate B-mode and
ultrasound elastography data due to increased depth.
ultrasound elastography for diffuse liver disease
Accurate assessment of the degree of liver fibrosis is important for
estimating prognosis and deciding on an appropriate course of
treatment for cases of chronic liver disease (CLD) with various
etiologies. Because of the inherent limitations of liver biopsy, there is a
great need for noninvasive and reliable tests that accurately estimate
the degree of liver fibrosis. Ultrasound (US) elastography is considered
a non-invasive, convenient, and precise technique to grade the degree
of liver fibrosis by measuring liver stiffness.
Transient elastography of a normal and a cirrhotic patient.
A
B
The displacement M-mode image located in the center of the monitor shows
axial displacement as a function of depth (y-axis) and time
(x-axis). In the healthy patient
(A), the shear wave is relatively slow and the liver stiffness is low. However,
in the patient with cirrhosis (B), the shear wave can propagate more rapidly
though a hard tissue, and the time-depth gradient is very steep. The
controlled attenuation parameter is displayed, which is known to correlate
with the fatty liver severity
Acoustic radiation force impulse imaging of a normal and a cirrhotic
patient.
The cylindrical region of interest (ROI) in the middle of the
ultrasonogram was used as the sample volume for the measurement of
stiffness. Instead of Young modulus, the propagating velocity of the
shear wave is displayed. Although the grayscale ultrasonograms are
similar to each other, the propagating velocity is different: 1.2 m/sec in
the healthy patient (A) and 1.6 m/sec in the patient with cirrhosis (B)
Supersonic shear-wave imaging of a normal and a cirrhotic patient.
A large trapezoidal color box display of the distribution of the elastic properties of
the liver. The severity of stiffness is depicted with the colored look-up table. Dark
blue color represents normal liver tissue (A), and bright blue-green color
represents increased liver stiffness such as liver cirrhosis (B). The round region
of interest (ROI) in the color box is the Q-Box and the mean Young modulus and
standard deviation in the ROI have been calculated
A B
D
Acute exacerbation of hepatitis B virus-induced hepatitis.
Grayscale ultrasonography (A) shows mild parenchymal coarseness of
the liver; the liver stiffness is 10.3 kPa according to the shear wave
elastography (B). Six months later, the patients recovered from liver
function deterioration. The grayscale ultrasonographic feature (C) is
similar to the previous study, but the liver stiffness decreases to 6 kPa
(D).
Elastography in the Differential Diagnosis of Enlarged Cervical Lymph Nodes
Drawings show typical diagrammatic appearance of five patterns of lymph nodes. Elastographic
patterns were determined on distribution and percentage of lymph node area having high elasticity
(hard): pattern 1, absent or small hard area; pattern 2, hard area < 45% of lymph node; pattern 3,
hard area ≥ 45%; pattern 4, peripheral hard and central soft areas; pattern 5, hard area occupying
entire lymph node. Increasing tissue hardness appears in ascending order as red, yellow, green,
and blue.
Transverse sonogram of level 1 lymph node in 30-year-old man with left submandibular
lymphadenopathy. Elastography image on left shows pattern 1, absent or small hard area. B-mode
sonographic image on right shows score of 5, reactive. Final diagnosis from clinical and serologic
findings was reactive lymph node.
Longitudinal sonogram of level 2 lymph node in 33-year-old woman with left posterior
auricular lymphadenopathy. Elastography image on left shows pattern 2, hard area <
45% of node. B-mode sonographic image on right shows score 5, reactive. Final
diagnosis from clinical and serologic findings was reactive lymph node.
Transverse sonogram of level 2 lymph node in 79-year-old man with laryngeal
carcinoma. Elastography image on left shows pattern 3, hard area ≥ 45%. B-mode
sonographic image on right shows score 7, metastatic. Final diagnosis by
histopathology was metastatic lymph node.
Longitudinal sonogram of level 5 lymph node in 52-year-old man with nasopharyngeal
carcinoma. Elastography image on left shows pattern 4, peripheral hard and central soft areas.
B-mode sonographic image on right shows score 7, metastatic. Final diagnosis by CT was
metastatic lymph node.
MR ELASTOGRAPHY
Following are the some of the current applications of MRE which are being
investigated :
Liver MR Elastography—Magnetic resonance elastography has been
widely investigated for the diagnosis of hepatic diseases and is currently
used in clinical practice for fibrosis and cirrhosis assessment where the
stiffness of the diseased liver is significantly higher than normal liver tissue
stiffness
Clinical hepatic MRE is performed at a frequency of 60 Hz using pneumatic-
based pressure-activated drivers. Wave data are acquired with four phase
offsets and a modified direct-inversion algorithm with multiscale
capabilities is used for the estimation of the liver stiffness.
MRE-derived stiffness of healthy liver tissue at 60 Hz compared to the stiffness of
liver tissue at various stages of fibrosis. The stiffness increases gradually with the
progression of the fibrosis. A cutoff of 2.93 kPa well differentiates the healthy and
fibrotic livers.
Hepatic MRE
Results are shown from clinical hepatic MRE exams of a patient with a normal
liver (top row) and a patient with a cirrhotic liver. (a,d)
Conventional abdominal MR magnitude images of the two patients, showing no
significant difference between the two livers. (b,e)
Wave images from the MRE acquisition at 60 Hz showing shear waves with a
shorter wavelength in the first patient, and a substantially longer wavelength in
the second patient. (c,f)
The corresponding elastograms indicating that the two livers were normal (1.7
kPa) and cirrhotic (18.83 kPa), respectively.
MRE is also being investigated as a means to characterize hepatic tumors and it
has been found that malignant liver tumors have significantly greater mean shear
stiffness than benign tumors and normal liver tissue and a cutoff value of 5 kPa
can differentiate malignant tumors from benign tumors and normal liver
parenchyma.
Breast MR Elastography—Another application of MRE that is being
investigated with great interest is for the assessment of breast cancer .
Breast tumors are known to be typically stiffer than benign lesions and normal
breast tissue . Manual palpation is a recommended part of routine screening for
breast cancer and helps in the detection of these hard masses.
Contrast-enhanced MR imaging (CE-MRI) has proven to have a very high
sensitivity for the detection of tumor nodules, but the specificity of the technique
can be a problem leading to numerous false positives and unnecessary biopsies.
MRE is being investigated as a complementary technique to CE-MRI to provide
additional information about these suspicious regions and the combined
technique has shown promise to increase diagnostic specificity
Breast MRE
(a) An axial MR magnitude image of the right breast of a patient volunteer is
shown. A large adenocarcinoma is shown as the outlined, mildly
hyperintense region on the lateral side of the breast.
(b) A single wave image from MRE performed at 100 Hz is shown along with
the corresponding elastogram (c).
(d) An overlay image of the elastogram and the magnitude image shows good
correlation between the tumor and the stiff region detected by MRE.
Skeletal Muscle MR Elastography—MRE has also been extensively
investigated for studying the stiffness of skeletal muscle since it is well known
that the stiffness changes significantly depending upon the contractile state of the
muscle . Skeletal muscle MRE can be used for studying the physiological
response of diseased and damaged muscles. For instance, it has been found that
there is a difference in the stiffness of muscles with and without neuromuscular
disease
Skeletal muscle MRE
(a) A sagittal MR image of the calf soleus muscle with the location of the
driver indicated by the arrow is shown. 100-Hz MRE wave images of the
muscle are shown while exerting 0 (b), 5 (c) and 10 N/m (d) of force.
The increase in the wavelength (and thus stiffness) with
the increase in muscle force is easily visible and is indicated by the double
sided arrows.
Brain MR Elastography—Assessment of the mechanical properties of
brain tissue with MRE is another area of significant research and clinical
interest due to the diagnostic potential of brain tissue stiffness information
as it may be related to diseases like Alzheimer’s disease, hydrocephalus, brain
cancer and multiple sclerosis. While it would be difficult to use ultrasound
based approaches to noninvasively assess brain mechanical properties, MRE is
well suited for this application.
Brain MRE
(a) Shown is an axial MR magnitude image of the brain showing white matter, gray
matter, and cerebrospinal fluid. (b) A single wave image from MRE performed
at 60 Hz. (c) The corresponding elastogram is shown and a good correlation
between the magnitude image and the stiffness estimate is evident.
MR Elastography for Visualizing Anatomy—The primary applications of
MRE that have been developed have focused on the study of the mechanical
properties of different tissues. However, the capability of MRE to measure the
displacement of tissues has also been investigated to address other clinical questions.
Such applications include a shearline imaging technique designed to investigate the
functionality of slip interfaces and a vibration imaging technique used to localize the
functional compartments of the forearm flexor muscles
As an example of these applications, the vibration imaging technique is introduced
here, the goal of which is to locate the finger-specific functional compartments of
the extrinsic forearm flexor muscles that partly control the flexion of the
interphalangeal joints of the index, middle, ring and little fingers.
Even though these flexor muscles (flexor digitorum superficialis, FDS and flexor
digitorum profundus, FDP) do not possess anatomically distinct compartments,
they can exert almost independent action on specific fingers through selective
activation of functional compartments specific for each finger.
The knowledge of the location and boundaries of these segments is useful for
localizing techniques like MR spectroscopy and biopsy, but is difficult to obtain
currently.
With the proposed vibration imaging technique based on MRE principles, this can
be achieved by vibrating a finger of interest independently of the others and
measuring the amount of motion within the forearm musculature that is transferred
through the structurally connected tendons.
Functional compartments of the flexor muscles
(a) MR magnitude image of the right forearm of a healthy volunteer.
(b) An example wave image obtained using MRE motion encoding of tissue
vibrations in the forearm induced by selectively vibrating the ring finger at
90 Hz.
(c) An amplitude map obtained from wave images indicating localized regions
of higher displacement corresponding to the compartments of the flexor and
extensor muscles (arrows).
(d) A phase map obtained from the wave images indicating that the
compartments of the flexor and extensor muscles of the activated finger are
moving out of phase with each other.
In vivo difficulties:
•Uncontrolled Motion
–Respiratory
–Cardiac
–Patient
•Requirement
–Fast Acquisition
RECENT ADVANCES AND FUTURE PROSPECTS
• Intracardiac Myocardial elastography
• Acute Pancreatitis risk assessment on admission
• Pancreatic Masses and chronic pancreatitis
• Renal elastography
• Obstretics & Gynaecology e.g. Cervical stiffness, Prediction of PTL,
Gynaecological cancer etc.
• Vascular imaging, Non-Invasive Vascular Elastography
• Optical Coherence Elastography
• Organ transplantation
.
.
.
.
.
.and many more……
CONCLUSION
Elastography is an important research field It has also been evaluated
in the clinical field for more than 10 years. It has provided a new
feature to assess tissue stiffness and has shown that tissue elasticity
is of great value for diagnosis.
Indeed the variation of the Young’s modulus in biological tissues offers
a contrast that is potentially more interesting than conventional
ultrasound.
If the elastography techniques developed in research laboratories
start making their commercial appearance, they are not all
quantitative or operator independent, the word ‘‘elastography’’ can
hide very different physical phenomena. It is therefore important to
know how to decrypt the underlying physics of each elastography
method in order to understand their advantages and physical
limitations.
Shear wave based techniques have strong advantages over quasi-
static techniques, as they are more reproducible, quantitative, rely on
automatic shear wave generation and provide good elasticity contrast.
The availability of true quantitative numerical data allows adjustment
of the dynamic range to optimize the visualization of structures. These
advantages should certainly lead to new applications of shear wave
elasticity imaging, not only for diagnosis but also for follow-up.
The real-time capability of some of these SW techniques also allows
the development of 3D elastography imaging that should facilitate the
clinical use for detection, therapy planning and monitoring in the
routine clinical practice.
Finally, the integration of elastography techniques in conventional
ultrasound systems open the door for routine application during
ultrasound examination and will allow information fusion with other
imaging techniques to strengthen their diagnostic performance.
REFERENCES AND FURTHER READING
• Ultrasound elastography: Principles and techniques BY J.-
L. Gennisson∗, T. Deffieux, M. Fink, M. Tanter : Diagnostic
and Interventional Imaging (2013) 94, 487—495
• Tissue Elasticity Imaging BYJean Martial Mari Department
of Medical Physics and Bioengineering,University College
London
• Ultrasound elastography for musculoskeletal applications
BY E E DRAKONAK, G M ALLEN and D J WILSON : The
British Journal of Radiology, 85 (2012), 1435–1445
• MAGNETIC RESONANCE ELASTOGRAPHY: A REVIEW BY
Yogesh K Mariappan, Kevin J Glaser, and Richard L Ehman
Clin Anat. 2010 July ; 23(5): 497–511
• Principles and clinical application of ultrasound
elastography for diffuse liver disease BY Woo Kyoung
Jeong, Hyo K. Lim, Hyoung-Ki Lee, Jae Moon Jo, Yongsoo
Kim : Ultrasonography 2014;33:149-160
• US Elastography ofBreast and Prostate Lesions BY Daniel T.
Ginat, Stamatia V. Destounis, Richard G.Barr, Benjamin
Castaneda, John G. Strang,Deborah J. Rubens : RadioGraphics
2009; 29:2007–2016
• Differential Diagnosis of Thyroid Nodules with US Elastography
Using Carotid Artery Pulsation BY Manjiri Dighe, Unmin Bae,
Michael L. Richardson, Theodore J. Dubinsky Satoshi
Minoshima, Yongmin Kim : Radiology: Volume 248: Number 2—
August 2008
• Accuracy of Sonographic Elastography in the Differential
Diagnosis of Enlarged Cervical Lymph Nodes: Comparison with
Conventional B-Mode Sonography BY Farzana Alam, Kumiko
Naito, Jun Horiguchi, Hiroshi Fukuda, Toshihiro Tachikake,
Katsuhide Ito : AJR 2008; 191:604–610
• RUMAC’s Diagnostic Ultrasound 4th Edition.
HOPE WE SEE ELASTOGRAPHY BEING DONE IN OUR
DEPARTMENT SOON!!!!!

More Related Content

What's hot

Role of Imaging in Male Infertility
Role of Imaging in Male InfertilityRole of Imaging in Male Infertility
Role of Imaging in Male InfertilityNiranjan Chavan
 
Contrast enhanced ultrasound
Contrast enhanced ultrasoundContrast enhanced ultrasound
Contrast enhanced ultrasoundNeeraj Patange
 
Diffusion weighted imaging
Diffusion  weighted imagingDiffusion  weighted imaging
Diffusion weighted imagingSharath Raj
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging MAMTA PANDA
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkDr pradeep Kumar
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance DiffusionRahman Ud Din
 
Advances in ultrasound
Advances in ultrasoundAdvances in ultrasound
Advances in ultrasoundSangeeta Jha
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and techniqueArchana Koshy
 
Ultrasound elastography of breast and prostate lesions
Ultrasound elastography of breast and prostate lesionsUltrasound elastography of breast and prostate lesions
Ultrasound elastography of breast and prostate lesionsRajesh Venunath
 
MRI-Magnetic Resonance Elastography (MRE) -Avinesh Shrestha
MRI-Magnetic Resonance Elastography (MRE) -Avinesh ShresthaMRI-Magnetic Resonance Elastography (MRE) -Avinesh Shrestha
MRI-Magnetic Resonance Elastography (MRE) -Avinesh ShresthaAvinesh Shrestha
 
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEDose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEhaijaypee_dan
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
 

What's hot (20)

MR Elastography
MR ElastographyMR Elastography
MR Elastography
 
Ultrasound Elastography
Ultrasound ElastographyUltrasound Elastography
Ultrasound Elastography
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
Role of Imaging in Male Infertility
Role of Imaging in Male InfertilityRole of Imaging in Male Infertility
Role of Imaging in Male Infertility
 
Contrast enhanced ultrasound
Contrast enhanced ultrasoundContrast enhanced ultrasound
Contrast enhanced ultrasound
 
Diffusion weighted imaging
Diffusion  weighted imagingDiffusion  weighted imaging
Diffusion weighted imaging
 
Post Processing of CT Thorax
Post Processing of CT ThoraxPost Processing of CT Thorax
Post Processing of CT Thorax
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
 
Mri diffusion
Mri diffusionMri diffusion
Mri diffusion
 
CT artifact
CT artifact CT artifact
CT artifact
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance Diffusion
 
Advances in ultrasound
Advances in ultrasoundAdvances in ultrasound
Advances in ultrasound
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and technique
 
Ultrasound elastography of breast and prostate lesions
Ultrasound elastography of breast and prostate lesionsUltrasound elastography of breast and prostate lesions
Ultrasound elastography of breast and prostate lesions
 
MRI-Magnetic Resonance Elastography (MRE) -Avinesh Shrestha
MRI-Magnetic Resonance Elastography (MRE) -Avinesh ShresthaMRI-Magnetic Resonance Elastography (MRE) -Avinesh Shrestha
MRI-Magnetic Resonance Elastography (MRE) -Avinesh Shrestha
 
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEDose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 

Viewers also liked

Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...
Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...
Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...Respibron Mili
 
Остеопороз - профилактика и лечение
Остеопороз - профилактика и лечениеОстеопороз - профилактика и лечение
Остеопороз - профилактика и лечениеИрина Головач
 
Космос
КосмосКосмос
Космосlukyluk05
 
Chronic obstructive pulmonary disease
Chronic obstructive pulmonary diseaseChronic obstructive pulmonary disease
Chronic obstructive pulmonary diseaseБачка Рүбби
 
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артрит
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артритEULAR критерии артралгии, клинически подозреваемой как ревматоидный артрит
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артритИрина Головач
 
Клинический случай поздней диагностики болезни Такаясу
Клинический случай поздней диагностики болезни ТакаясуКлинический случай поздней диагностики болезни Такаясу
Клинический случай поздней диагностики болезни ТакаясуИрина Головач
 
Антифосфоліпідний синдром
Антифосфоліпідний синдромАнтифосфоліпідний синдром
Антифосфоліпідний синдромИрина Головач
 
Коморбидность при остеоартрите - выбор терапии
Коморбидность при остеоартрите - выбор терапииКоморбидность при остеоартрите - выбор терапии
Коморбидность при остеоартрите - выбор терапииИрина Головач
 
Болезнь XXI века - аллергия
Болезнь XXI века - аллергияБолезнь XXI века - аллергия
Болезнь XXI века - аллергияallergomed
 
Поздняя диагностика неспецифического аортоартериита
Поздняя диагностика неспецифического аортоартериитаПоздняя диагностика неспецифического аортоартериита
Поздняя диагностика неспецифического аортоартериитаИрина Головач
 
Вопросы иммуногенности и безопасности биологической терапии
Вопросы иммуногенности и безопасности биологической терапииВопросы иммуногенности и безопасности биологической терапии
Вопросы иммуногенности и безопасности биологической терапииИрина Головач
 
Современные подходы к лечению подагры
Современные подходы к лечению подагрыСовременные подходы к лечению подагры
Современные подходы к лечению подагрыИрина Головач
 
Ключевая роль метотрексата в лечение ревматоидного артрита
Ключевая роль метотрексата в лечение ревматоидного артритаКлючевая роль метотрексата в лечение ревматоидного артрита
Ключевая роль метотрексата в лечение ревматоидного артритаИрина Головач
 
Ревматическая полимиалгия - стандарты диагностики и лечения
Ревматическая полимиалгия - стандарты диагностики и леченияРевматическая полимиалгия - стандарты диагностики и лечения
Ревматическая полимиалгия - стандарты диагностики и леченияИрина Головач
 
Кардиоренальный синдром
Кардиоренальный синдромКардиоренальный синдром
Кардиоренальный синдромPopovSI
 
Увеиты при спондилоартитах
Увеиты при спондилоартитахУвеиты при спондилоартитах
Увеиты при спондилоартитахИрина Головач
 
Болезнь модифицирующая терапия остеоартрита
Болезнь модифицирующая терапия остеоартритаБолезнь модифицирующая терапия остеоартрита
Болезнь модифицирующая терапия остеоартритаИрина Головач
 
На приеме пациент с болью в спине
На приеме пациент с болью в спинеНа приеме пациент с болью в спине
На приеме пациент с болью в спинеИрина Головач
 
Liver stiffness measurement (fibroscan®)
Liver stiffness measurement (fibroscan®)Liver stiffness measurement (fibroscan®)
Liver stiffness measurement (fibroscan®)Samir Haffar
 

Viewers also liked (20)

Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...
Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...
Аспекты применения бактериальных лизатов при ХОЗЛ и БА в практике аллерголога...
 
Остеопороз - профилактика и лечение
Остеопороз - профилактика и лечениеОстеопороз - профилактика и лечение
Остеопороз - профилактика и лечение
 
Космос
КосмосКосмос
Космос
 
Chronic obstructive pulmonary disease
Chronic obstructive pulmonary diseaseChronic obstructive pulmonary disease
Chronic obstructive pulmonary disease
 
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артрит
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артритEULAR критерии артралгии, клинически подозреваемой как ревматоидный артрит
EULAR критерии артралгии, клинически подозреваемой как ревматоидный артрит
 
Клинический случай поздней диагностики болезни Такаясу
Клинический случай поздней диагностики болезни ТакаясуКлинический случай поздней диагностики болезни Такаясу
Клинический случай поздней диагностики болезни Такаясу
 
Антифосфоліпідний синдром
Антифосфоліпідний синдромАнтифосфоліпідний синдром
Антифосфоліпідний синдром
 
Коморбидность при остеоартрите - выбор терапии
Коморбидность при остеоартрите - выбор терапииКоморбидность при остеоартрите - выбор терапии
Коморбидность при остеоартрите - выбор терапии
 
Болезнь XXI века - аллергия
Болезнь XXI века - аллергияБолезнь XXI века - аллергия
Болезнь XXI века - аллергия
 
Поздняя диагностика неспецифического аортоартериита
Поздняя диагностика неспецифического аортоартериитаПоздняя диагностика неспецифического аортоартериита
Поздняя диагностика неспецифического аортоартериита
 
Вопросы иммуногенности и безопасности биологической терапии
Вопросы иммуногенности и безопасности биологической терапииВопросы иммуногенности и безопасности биологической терапии
Вопросы иммуногенности и безопасности биологической терапии
 
Современные подходы к лечению подагры
Современные подходы к лечению подагрыСовременные подходы к лечению подагры
Современные подходы к лечению подагры
 
подагра как эклектика
подагра как эклектикаподагра как эклектика
подагра как эклектика
 
Ключевая роль метотрексата в лечение ревматоидного артрита
Ключевая роль метотрексата в лечение ревматоидного артритаКлючевая роль метотрексата в лечение ревматоидного артрита
Ключевая роль метотрексата в лечение ревматоидного артрита
 
Ревматическая полимиалгия - стандарты диагностики и лечения
Ревматическая полимиалгия - стандарты диагностики и леченияРевматическая полимиалгия - стандарты диагностики и лечения
Ревматическая полимиалгия - стандарты диагностики и лечения
 
Кардиоренальный синдром
Кардиоренальный синдромКардиоренальный синдром
Кардиоренальный синдром
 
Увеиты при спондилоартитах
Увеиты при спондилоартитахУвеиты при спондилоартитах
Увеиты при спондилоартитах
 
Болезнь модифицирующая терапия остеоартрита
Болезнь модифицирующая терапия остеоартритаБолезнь модифицирующая терапия остеоартрита
Болезнь модифицирующая терапия остеоартрита
 
На приеме пациент с болью в спине
На приеме пациент с болью в спинеНа приеме пациент с болью в спине
На приеме пациент с болью в спине
 
Liver stiffness measurement (fibroscan®)
Liver stiffness measurement (fibroscan®)Liver stiffness measurement (fibroscan®)
Liver stiffness measurement (fibroscan®)
 

Similar to Elastography

Medical imaging modalities(2)
Medical imaging modalities(2)Medical imaging modalities(2)
Medical imaging modalities(2)UmarKhan68
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imagingothman alameen
 
Rehabilitative Ultrasound Imaging: A musculoskeletal Perspective
Rehabilitative Ultrasound Imaging: A musculoskeletal PerspectiveRehabilitative Ultrasound Imaging: A musculoskeletal Perspective
Rehabilitative Ultrasound Imaging: A musculoskeletal PerspectiveZinat Ashnagar
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdflittlealphonsa
 
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniUltrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfMohamedAli690824
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonographyRiyad Banayot
 
Sonography
SonographySonography
SonographyFLI
 
Elastosonografia
ElastosonografiaElastosonografia
ElastosonografiaASMaD
 
Lecture - 2 MBBS (x-ray modalities)
Lecture - 2 MBBS (x-ray modalities)Lecture - 2 MBBS (x-ray modalities)
Lecture - 2 MBBS (x-ray modalities)Dr.Bijay Yadav
 
J0362069073
J0362069073J0362069073
J0362069073theijes
 

Similar to Elastography (20)

Liver Fibrosis - SWE and MRE
Liver Fibrosis - SWE and MRE Liver Fibrosis - SWE and MRE
Liver Fibrosis - SWE and MRE
 
Physics
PhysicsPhysics
Physics
 
Medical imaging modalities(2)
Medical imaging modalities(2)Medical imaging modalities(2)
Medical imaging modalities(2)
 
ultrasonography.pptx
ultrasonography.pptxultrasonography.pptx
ultrasonography.pptx
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imaging
 
ultrasound JP.pptx
ultrasound JP.pptxultrasound JP.pptx
ultrasound JP.pptx
 
Rehabilitative Ultrasound Imaging: A musculoskeletal Perspective
Rehabilitative Ultrasound Imaging: A musculoskeletal PerspectiveRehabilitative Ultrasound Imaging: A musculoskeletal Perspective
Rehabilitative Ultrasound Imaging: A musculoskeletal Perspective
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniUltrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonography
 
Sonography
SonographySonography
Sonography
 
Elastography
ElastographyElastography
Elastography
 
Elastosonografia
ElastosonografiaElastosonografia
Elastosonografia
 
ULTRA.pptx
ULTRA.pptxULTRA.pptx
ULTRA.pptx
 
Lecture - 2 MBBS (x-ray modalities)
Lecture - 2 MBBS (x-ray modalities)Lecture - 2 MBBS (x-ray modalities)
Lecture - 2 MBBS (x-ray modalities)
 
Ultrasound1
Ultrasound1Ultrasound1
Ultrasound1
 
Imaging.ppt
Imaging.pptImaging.ppt
Imaging.ppt
 
Radiosurgery Revised
Radiosurgery RevisedRadiosurgery Revised
Radiosurgery Revised
 
J0362069073
J0362069073J0362069073
J0362069073
 

More from Sameer Peer

Contrast reactions
Contrast reactionsContrast reactions
Contrast reactionsSameer Peer
 
Approach to a vomiting Infant
Approach to a vomiting InfantApproach to a vomiting Infant
Approach to a vomiting InfantSameer Peer
 
Plain X-ray SKULL
Plain X-ray SKULLPlain X-ray SKULL
Plain X-ray SKULLSameer Peer
 
Current trends in imaging of Epilepsy
Current trends in imaging of EpilepsyCurrent trends in imaging of Epilepsy
Current trends in imaging of EpilepsySameer Peer
 
IMAGING OF TEMPORAL BONE
IMAGING OF TEMPORAL BONEIMAGING OF TEMPORAL BONE
IMAGING OF TEMPORAL BONESameer Peer
 
Imaging of Bone Tumors
Imaging of Bone TumorsImaging of Bone Tumors
Imaging of Bone TumorsSameer Peer
 

More from Sameer Peer (6)

Contrast reactions
Contrast reactionsContrast reactions
Contrast reactions
 
Approach to a vomiting Infant
Approach to a vomiting InfantApproach to a vomiting Infant
Approach to a vomiting Infant
 
Plain X-ray SKULL
Plain X-ray SKULLPlain X-ray SKULL
Plain X-ray SKULL
 
Current trends in imaging of Epilepsy
Current trends in imaging of EpilepsyCurrent trends in imaging of Epilepsy
Current trends in imaging of Epilepsy
 
IMAGING OF TEMPORAL BONE
IMAGING OF TEMPORAL BONEIMAGING OF TEMPORAL BONE
IMAGING OF TEMPORAL BONE
 
Imaging of Bone Tumors
Imaging of Bone TumorsImaging of Bone Tumors
Imaging of Bone Tumors
 

Recently uploaded

Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 

Recently uploaded (20)

Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 

Elastography

  • 1. SHER-I-KASHMIR INSTITUTE OF MEDICAL SCIENCES SOURA DEPARTMENT OF RADIODIAGNOSIS & IMAGING
  • 2. PRESENTATION BY Dr SAMEER PEER 1ST YEAR PG STUDENT
  • 3.
  • 4. CONTENTS • Introduction • Basic principles of elastography • Methods and Types of elastography • Applications of elastography • Pitfalls and limitations • Recent Advances and future prospects • Conclusion • References and further reading
  • 5. INTRODUCTION Ever since Hippocrates, physicians have used palpation to detect differences in tissue stiffness as an aid to diagnosis. Elastography is a novel imaging modality which depends on the same differences in mechanical properties between healthy and abnormal tissues but uses imaging to detect these differences at depths not reachable by manual palpation. Elastography techniques have been developed for both ultrasound and MR imaging.
  • 6. –Sono-Elasticity Lerner 1990 –Elastography Ophir 1991 –Shear Waves Elastography Sarvazyan1997 –Vibro-acoustography Fatemi 1998 –Transient Elastography Sandrin 1999 –Acoustic Radiation Force Imaging(ARFI) Nightingale 2001 –Supersonic Shear Imaging Elastography Bercoff 2004 – MR Elastography Muthupillai1995 TIMELINE REVIEW
  • 7. BASIC PRINCIPLES OF ELASTOGRAPHY Elastogarphy is a tissue elasticity imaging, hence it is important to understand the general principles which govern the behaviour of an elastic material when acted upon by an external force. When external forces are applied to objects made up of elastic materials, they undergo a change in shape and size. STRAIN is a relative change in shape or size of an object due to externally applied forces. STRESS is the internal force ( per unit area) associated with strain HOOK’S LAW states that strain is directly proportional to stress for a linearly isotropic elastic material
  • 8. e= (h-ho)/ho Strain = Change in length/original length e= / E e = Strain (dimensionless)  = Stress (Pascal) = Force / Area E = Young’s Modulus (Pascal)
  • 9. Elastic Moduli YOUNG’S MODULUS ( E ) : It is the ratio of stress along an axis to the strain along that axis in the range of stress in which hooke’s law holds. BULK MODULUS (K or B) : It is the ratio of infinitesimal pressure increase to the resultant relative volume decrease. SHEAR MODULUS ( G or S or m ) : It is the ratio of shear stress to shear strain. It concerns with deformation of solids when two opposing surfaces experience forces in opposing directions.
  • 10. YOUNG’S MODULUS (E) BULK MODULUS ( K OR B) SHEAR MODULUS ( G or m)
  • 11. Conventional ultrasound • Propagation of ultrasound depends on bulk modulus • Bulk modulus is homogeneous in different tissues • No information provided regarding tissue stiffness Elastography • Imaging of tissue stiffness • Estimation of strain in tissues under stress • Imaging of shear waves whose propagation is governed by tissue stiffness or Young’s modulus(E) • Young’s modulus exhibits important differences in tissues which is ideal for characterization of tissues with excellent contrast. • Young’s modulus is a quantitative estimation of clinician’s palpation and has diagnostic value
  • 13.
  • 14.
  • 15.
  • 16. dA dB tA= 2dA/ c tB= 2dB/ c c = Speed of sound
  • 17. Displacement ( d) = dB – dA = c ( t B-t A)/2 d = c/2 ( = t B – t A )
  • 18. METHODS AND TYPES OF ELASTOGRAPHY Two established methods of elastography • ULTRASOUND ELASTOGRAPHY • MAGNETIC RESONANCE ELASTOGRAPHY
  • 19.
  • 21. QUASI-STATIC METHODS In the case of quasi-static elastography, a constant stress is applied to the tissue. The displacement and the generated strain are then estimated using two-dimensional correlation of ultrasound images. The Young’s modulus is then given via the Hooke’s law , which links stress and strain in a purely elastic medium. In practice, since the applied stress is unknown, only the strain is displayed, this strain map is sometimes called the elastogram. This was the first elastography technique, developed by the Ophir Group at the beginning of the 1990s. Today, the technique is being tested in breast lesion characterization with interesting results.
  • 22.
  • 23.
  • 24. The main limitations of this technique are • the control of the stress applied, which remains operator dependent, • the absence of a specific quantification. • the use of a stress applied by the operator limits the technique to superficial organs, mainly the breast or the thyroid. This technique is simple to implement and is widely spread in the world of radiology, and has been mainly validated for breast lesions classification
  • 25. DYNAMIC METHODS In dynamic methods, a time-varying force is applied to the tissue, it can either be a short transient mechanical force or an oscillatory force with a fixed frequency. A time-varying mechanical perturbation will propagate as mechanical waves which in a solid body can be compressional waves or shear waves. The compressional waves propagate very quickly in the human body (∼1500 m/s), and at high frequencies, these waves, also known as ultrasounds, can be used to image the body. Shear waves, which are only generated at low frequencies (10 Hz to 2000 Hz) due to absorption at higher frequencies, propagate more slowly, and their speed (∼1—50 m/s) is directly related to the medium shear modulus.
  • 26.
  • 27. In biological tissues, which are almost incompressible, the Young’s modulus can be approximated as three times the shear modulus. The shear wave propagation speed can thus be used to map the Young’s modulus quantitatively. Dynamic elastography techniques, which rely on shear waves propagation, can produce quantitative and higher resolution Young’s modulus map compared to quasi-static methods. The use of shear waves, however, requires a more complex system, able to generate the shear wave (mechanical vibrator or ultrasound radiation pressure) and to image the small displacements induced by the shear wave (ultrafast or stroboscopic ultrasound).
  • 28. VIBRO-ACOUSTOGRAPHY Vibro-acoustography is a dynamic elastography method based on the ultrasound radiation pressure and developed by the American Group of James Greenleaf The radiation pressure is a volumic force created by transfer of momentum within the medium This momentum is related to the absorption of the ultrasound wave. Vibro-accoustography uses two confocal ultrasound beams with slightly different frequencies w and w0 + w. It results in beats at the frequency w , which give rise to a modulated force at frequency w, only at the focal spot. Therefore, everything looks like if the target vibrates at the frequency w. It is then sufficient to listen to the sound produced by this excitation to deduce the mechanical properties of the target, particularly its stiffness.
  • 29. For this, one can place a hydrophone that records the response of the target at the frequency w . In order to create an image, the entire area is swept by moving the focal spot and listening to the response for each point of the image.
  • 30. However the sweeping of the entire image causes an important deposit of energy within the medium and a long acquisition time, which makes its implementation in real-time and in vivo difficult. In particular, the measured parameter depends on the stiffness of the zone, but also on the amplitude of the generated force and the geometry of the vibrating object. Therefore images correspond to a mixture of several physical parameters, including elasticity. Greenleaf’s team is now trying to combine this method with mammography systems for the diagnosis of breast cancer and with conventional ultrasound machines
  • 31. TRANSIENT ELASTOGRAPHY METHODS Acoustic Radiation Force Impulse Imaging (ARFI) or ‘‘Acoustic Radiation Force Imaging’’, is a method developed by the American team of Kathy Nightingale. This technique uses the acoustic radiation force but, unlike vibro - acoustography, ARFI only uses one focalized ultrasound beam. The radiation force slightly displaces the tissue at the focal spot according to Hooke’s law. Then the transducer switches into imaging mode and detect displacements of the focal spot by tracking of the ultrasound signal (called ‘‘speckle’’). The ultrasound speckle-tracking , already used in static elastography, allows to correlate the ultrasound signal window by window in order to detect the displacement of a tissue with sensitivity of less than a micrometer. By window, I mean a piece of an ultrasound signal.
  • 32.
  • 33. Today, Nightingale’s team is interested in the propagation of shear waves generated by the radiation force and has recently proposed a new ARFI model called ‘‘ARFI-SWS’’. Based on this concept, it allows to quantitatively measure the Young’s modulus in a small region of interest. This variation is currently being evaluated for the liver staging and is available on commercial ultrasound systems, such as the Siemens Acuson S2000.
  • 34. 1D transient elastography: the 1D shear elasticity probe The 1D transient elastography probe was first developed at the Institut Langevin in 1995 by Catheline et al. It consists of generating a transient impulse (little shock) on the medium and recording the shear wave that propagates within the medium by using an ultrasound transducer. First, the front face of the transducer acting as a piston gives a slight mechanical impulse on the surface of medium, which generates a spherical compression wave as well as a spherical shear wave. The displacement generated, which is a function of depth and of time, is thus estimated by correlations of retro-diffused echoes (via ultrasound speckle) recorded at a framerate higher than one thousand time per second with a mono-dimensional ultrasound transducer (5 MHz). This device was the first to use the principle of ultrafast imaging in one dimension to visualize in a transient manner the propagation of shear waves.
  • 35. Finally, by measuring the phase for each depth, we extract the phase speed of the shear wave at the central frequency, leading to an estimation of Young’s modulus by considering the medium to be homogeneous and nonviscous. Since 2001, the company Echosens has been commercializing this technique, 1D transient elastography, under the name FibroScan.
  • 36. 2D transient elastography In 1997, at the ‘‘Institut Langevin’’, the 1D transient elastography technique was extended to 2D, allowing the creation of elasticity maps of biological tissues. A programmable ultrasound electronic device used for time reversal experiments (process of refocalizing acoustic waves) was modified to be able to perform ultrafast imaging based on ultrasound plane wave emission. It allowed storing raw data acquired with a frame rate of more than 5000 images per second. A vibrator was fixed to the ultrasound imaging array, which is then used as an impactor to generate a quasi-plane shear wave. Once the movie of the propagating shear wave is reconstructed, the wave equation is inverted to recover a map of Young’s modulus. The first in vivo tests were carried out in 2003 with volunteers at Institut Curie, the results were encouraging, but the device was bulky, heavy and difficult to use in practice.
  • 37.
  • 38.
  • 39. The study of shear wave by using ultrafast imaging The Supersonic Shear Imaging technique is the outcome of researches at the Institute Langevin in transient elastography. The idea of combining radiation pressure or acoustic radiation force and the study of shear waves generated comes from Armen Sarvazyan, which can be considered as a precursor of elastography techniques based on ultrasonic radiation pressure via its technique: Shear Wave Elasticity Imaging. In 2004, two fundamental ideas are developed to overcome the limitations of 2D elastography technique: • acoustic radiation force • ultrasound ultrafast imaging. These two concepts described below are at the heart of the Supersonic Shear Imaging technique
  • 40. • a Mach cone: ultrasound beams are successively focused at different depths. The different spherical waves generated for each focal beam interfere like a Mach cone in which the source propagates more quickly than the generated shear wave and creates a quasi-plane wave front in the imaging plane (cylindrical in three dimensions). The use of constructive interfaces makes it possible to increase the amplitude of the wave and thus the signal to noise ratio of the displacement field. The generated quasi-plane shear wave in the imaging plane allows also simplifying the propagation hypotheses, which is of great interesting to solve inverse problems. Finally, only one Mach cone allows illuminating almost all of the medium with one plane shear wave; shear wave being generated, the propagation equation of the waves is inverted to rebuild a map of Young’s modulus;
  • 41. • ultrafast imaging: complete acquisition all at once. Ultrafast imaging allows scanning the entire imaging plane with very good temporal resolution in one single acquisition, typically with a frame rate of 5000 images per second, and up to 30,000 images per second in the case of tissues such as the peripheral arteries or the eye. Therefore there is no need to repeat the acquisition several times by stroboscopy to acquire the entire displacement field. This allows, not only imaging in real-time, which makes the examination easier, but also averaging the very rapidly acquired images to improve image quality.
  • 42.
  • 43. Therefore, the Supersonic Shear Imaging technique uses the technology of the 2D transient elastography, but substitutes vibrator by the acoustic radiation pressure. The whole excitation-imaging method is then integrated onto one single component: the ultrasound imaging transducer array. Amplified by the Mach cone, the generated shear wave has amplitude of tens of microns. This latter is detectable with a good signal to noise ratio by ultrasound speckletracking algorithm and ultrafast imaging. Thanks to ultrafast imaging, the acquisition of the shear wave propagation can be carried out all at once in less than 30 milliseconds. The technique is therefore slightly sensitive to patient movements (as an example breathing) and can be displayed in real-time, like a conventional ultrasound image. The Young’s modulus maps are then reconstructed by estimating the speed of the shear wave between two points of the image, using a time of flight algorithm. The first in vivo experiences with this technique were conducted at the Institut Curie where, in approximately 50 patients, the use of the technique to differentiate benign tissue from malignant tissue was demonstrated
  • 44.
  • 45. This technique was implemented on an ultrasound diagnostic imaging device called, the Aixplorer (Supersonic Imagine, Aix-en-Provence, France) and its diagnosis performance as well as its reproducibility were demonstrated in several organs, and more particularly in the breast with the addition of the elasticity parameters to the classical BIRADs criteria
  • 46. Shear wave excitation Measurement of shear wave propagation Shear modulus reconstruction Shear wave Force Probe Probe Tx Echo Rx Displacement of reflectors Tissue displacement map Shear wave speed calculation Shear modulus Shear modulus Shear wave speed Density Tx = transmission Rx = receiver Multiple M-mode or B-mode images: tissue displacement map
  • 47. MAGNETIC RESONANCE ELASTOGRAPHY Magnetic Resonance Elastography (MRE) is a rapidly developing technology for quantitatively assessing the mechanical properties of tissue. MRE obtains information about the stiffness of tissue by assessing the propagation of mechanical waves through the tissue with a special magnetic resonance imaging (MRI) technique. The technique essentially involves three steps: 1. generating shear waves in the tissue, 2. acquiring MR images depicting the propagation of the induced shear waves and 3. processing the images of the shear waves to generate quantitative maps of tissue stiffness, called elastograms.
  • 48. Generating Mechanical Waves in Tissue MRE typically uses vibrations of a single frequency (within the audio frequency range) generated by external driver devices. The electrical signal for these devices is created by a signal generator triggered by and synchronized to the MR pulse sequence and is amplified by an audio amplifier before being fed into the mechanical driver Over the years, several driving mechanisms have been developed, each with their own advantages and limitations. Three of the most commonly used driver systems are: • an electromechanical driver that works via the Lorentz force and utilizes the magnetic field of the main MRI magnet. • A piezoelectric stack driver system where the motion created is based on the piezoelectric property of certain materials. Focused-ultrasound-based (FUSbased) radiation force has also been investigated as a means to create mechanical motion for various elasticity imaging strategies including MRE, where shear waves are created directly within tissue with externally placed ultrasound transducers.
  • 49. • pressure-activated driver another widely used method of creating the required vibrations for MRE utilizes the motion of the voice coils used in acoustic speaker systems. The required vibrations are again produced by the Lorentz force, but the static magnetic field is from a devoted permanent magnet present in the acoustic speaker .
  • 50. Imaging the Propagating Waves Measuring the tissue motion produced by a driver with MRE is based on an MR imaging technique called phase-contrast MRI. Muthupillai et al. developed the technique of dynamic phase-contrast MRE where propagating shear waves in tissue are encoded into the phase of the MR images with the help of motion encoding gradient (MEG) pair. After continuous harmonic motion is induced in the tissue, a MEG oscillating at the same frequency as the motion is applied and conventional MR imaging is performed. The phase contribution to the MR image ϕ due to the motion and the applied magnetic field gradient at a given position vector r⃗ and phase offset θ between the motion and the MEG can be written as.
  • 51. where γ is the gyromagnetic ratio of the tissue protons , N is the number of gradient pairs used to sensitize the motion, T is the period of the motion encoding gradient, G⃗ is its amplitude, ξ⃗0 is the peak amplitude of motion, and k⃗ is the wave number. This equation states that the phase of harmonically vibrating tissue is directly proportional to its displacement.
  • 52. Figure shows a typical MRE pulse sequence using a gradient-recalled echo with the conventional radiofrequency (RF) pulse waveform, slice-selection gradient, phase-encoding gradient and frequency-encoding gradient. The motion-encoding gradient (shown here only in the frequency-encoding direction) is placed after the RF excitation of the sample and before the measurement of the induced signal. Motion occurring in any direction can be encoded into the phase of the MR image by manipulating the axes on which the MEGs are placed. In this example, only the motion occurring in the frequency-encoding direction will be sensitized to and encoded into the image phase. The motion-encoding capability of this technique is very sensitive and can detect motion on the order of 100’s of nanometers. An MR image thus obtained containing information about the propagating wave in its phase is called a wave image. Typically two such wave images are collected with opposite polarity of the MEG and a phase-difference image is calculated to remove non-motion-related phase information.
  • 53. Since the motion-encoding gradients necessary for MRE are inserted into conventional MR pulse sequences, MRE can be implemented with many MR imaging sequences, each with its own advantages and limitations. MRE pulse sequences based on spin echo (SE), gradient-recalled echo (GRE), balanced steady-state free precision (bSSFP) and echo planar imaging (EPI) techniques exist.
  • 54. Mechanical Parameter Estimation From the wave images indicating the propagation of shear waves in the tissue, mathematical inversion algorithms based on equations of motion, with simplifying assumptions like isotropy, homogeneity, and incompressibility, allow for the calculation of mechanical properties like the shear modulus to be used for clinical interpretations. The images of the mechanical properties of tissue calculated in MRE are often referred to as elastograms. Depending on the technique used to derive the elastograms from the original MR images, elastograms can theoretically (e.g., absent any noise) have half the resolution of the native MR images (which can range from 50 μm to 10 mm depending on the application), however they more typically have one-third to one-fifth of the MRI resolution. Shear wave interference can cause artifacts in the stiffness calculations using several of the above techniques, and a preprocessing technique called directional filtering has been developed to reduce these artifacts
  • 55. MRE of an inclusion phantom (a) MR magnitude image of an inclusion phantom with soft and stiff inclusions seen as the hyperintense and hypointense regions, respectively. (b) A single wave image from the MRE acquisition performed at 100 Hz. The difference in the wavelengths in the different regions is evident. (c) An elastogram obtained from these data showing the stiff and soft regions.
  • 57. ULTRASOUND ELASTOGRAPHY Breast US Elastography Compression elastography may have a role for further evaluating abnormal findings on conventional breast US images and differentiating benign from malignant breast lesions. Thus, the eventual goal of incorporating US compression elastography into routine practice would be to reduce the biopsy rate of benign lesions. The two most important elastographic characteristics in evaluating breast lesions are size and stiffness criteria
  • 58.
  • 59. Normal Breast Tissue Biomechanical testing has shown normal fibroglandular breast tissue to be markedly stiffer than normal fatty breast tissue by as much as two orders of magnitude . Therefore, at elastography, fatty tissue will appear bright with respect to the adjacent glandular tissue, and normal fibrous parenchyma appears darker
  • 60. (a) B-mode US image of the breast shows that a lobule of normal fatty tissue (f) is hypoechoic with respect to the surrounding glandular tissue (g). (b) US elastogram shows that the fatty tissue (f) surrounded by dense breast tissue appears bright because it is appreciably “softer” than the surrounding glandular tissue (g).
  • 61. Simple cyst. (a) B-mode US image of the breast shows a classic cyst (arrowheads). (b) US elastogram demonstrates a bull’s-eye configuration within the lesion (arrowheads), which contains bright compressible material (thin arrow), and a bright spot behind the cyst (thick arrow).
  • 62. Complex cyst mimicking a solid lesion at US. (a) B-mode US image demonstrates a round hypoechoic breast lesion without associated through transmission. (b) US elastogram shows that the lesion that appeared “solid” on the B-mode image is a cyst. .
  • 63. Fibrocystic change. (a) B-mode US image demonstrates an irregular heterogeneous lesion in a patient who presented with a palpable mass in the right breast. (b) On the US elastogram, the lesion appears soft and smaller than on the B-mode image. The findings from a subsequent biopsy showed fibrocystic change and periductal inflammation. Dotted lines indicate measured lesion diameters: 1 = 7.16 mm; 2 = 5.55 mm; 3 = 5.02 mm; and 4 = 4.29 mm. Distance ratio was calculated as 5.02  7.16 = 0.70.
  • 64. Fibroadenoma Fibroadenoma. (a) B-mode US image of the left breast shows a lobulated hypoechoic lesion that is taller than wide, with posterior acoustic shadowing. (b) US elastogram, however, shows the lesion to be smaller than on the B-mode image. The findings from biopsy revealed a fibroadenoma.
  • 65. Fibroadenoma with elasticity score of 1 in 51-year-old woman. US images were obtained in transverse plane. Left: On conventional B-mode image, lesion was classified as BI-RADS category 2. Right: On elasticity image, the entire hypoechoic lesion was evenly shaded green, as was the surrounding breast tissue.
  • 66. Fibroadenoma with elasticity score of 2 in 39-year-old woman. US images were obtained in transverse plane. Left: On conventional B-mode image, lesion was classified as BI-RADS category 3. Right: On elasticity image, hypoechoic lesion shows mosaic pattern of green and blue.
  • 67. Invasive Ductal Carcinoma Invasive ductal carcinoma. The breast lesion measures smaller on the B-mode US image (a) than on the US elastogram (b) in the transverse plane. The lesion is too large to measure accurately in the anteroposterior plane; dotted lines indicate measurements of the lesion diameter: 21.9  34.2 mm in a and 24.9  29.6 mm in b. In addition, the lesion is stiffer than the surrounding tissues.
  • 68. invasive ductal carcinoma typically is appreciably darker than normal tissues or benign lesions and is substantially larger on the elastogram than on B-mode US images Nonscirrhous type invasive ductal carcinoma with elasticity score of 4 in 29-year old woman. US images were obtained in transverse plane. Left: On conventional B-mode image, lesion was classified as BI-RADS category 5. Middle: On elasticity image, the entire hypoechoic lesion was blue. Right: Pathologic section of lesion is shown. (Hematoxylin-eosin stain; original magnification,1.)
  • 69. Scirrhous type invasive ductal carcinoma with elasticity score of 5 in 55-year-old woman. US images were obtained in sagittal plane. Left: On conventional B-mode Image, lesion was classified as BIRADS category 5. Middle: On elasticity image, both the entire hypoechoic lesion and its surrounding area were blue. Right: Pathologic section of lesion is shown. (Hematoxylin-eosin stain; original magnification X 1.),
  • 70. Lobular carcinoma in situ with elasticity score of 3 in 46-year-old woman. US images were obtained in transverse plane. Left: On conventional B-mode image, lesion was classified as BI-RADS category 3. Right: On elasticity image, the central part of the hypoechoic lesion was blue, and the peripheral part of the lesion was green.
  • 71. Lymph node with metastatic involvement. (a) B-mode US image demonstrates a lymph node that appears benign on the basis of its reniform shape and echogenicity. Dotted line indicates a lesion diameter of 20.4 mm. (b) Corresponding US elastogram, however, shows a stiff area that turned out to be a metastatic focus (arrows). Dotted line indicates a lesion diameter of 24.0 mm.
  • 72. Hematoma. (a) B-mode US image of the breast demonstrates a superficial heterogeneous mass. (b) US elastogram shows that the lesion is stiff but measures smaller than on the B-mode image. Dotted lines indicate measurements of the lesion diameter: 11.0  5.3 mm in a and 10.2  4.2 mm in b. The distance ratios were calculated as 10.2  11.0 = 0.93 (for the long-axis dimension) and as 4.2  5.3 = 0.79 (for the short-axis dimension). At physical examination, skin discoloration was noted in the corresponding area, a finding consistent with a hematoma.
  • 73. Prostate US Elastography Prostate Cancer Endorectal US continues to be the most commonly used imaging modality for assessing the prostate; however, it is relatively inaccurate for cancer detection. Endorectal US has a role in guiding prostate biopsies. Magnetic resonance imaging and computed tomography are mainly implemented for staging. At B-mode US, most prostate cancers are hypoechoic, but many are isoechoic, and a few are hyperechoic. As a result, the overall sensitivity of endorectal US for prostate cancer detection is about 50%. The added use of Doppler imaging increases the detection rate by only 5%. In contrast, endorectal real-time elastography enables the diagnosis of prostate cancer with a reported accuracy of 76%. However, the major role of elastography is to improve the results of image-directed biopsy and therapy, compared with the use of endorectal US alone.
  • 74. Prostate adenocarcinoma with Gleason score of 6/10. (a) Sagittal endorectal US Image with the prostate gland outlined shows no discrete lesion (arrow). (b) Sagittal US elastogram shows strong vibration in the normal prostate tissue and also an area of no vibration (arrow) anteriorly from the midportion of the gland to the base. (c) Photomicrograph (hematoxylineosin stain) of a transverse histologic section that included the corresponding lesion (arrow) at the base shows good correlation, with the deficit noted anteriorly on the elastogram through the selected plane of section (dashed line in a and b) in the midportion of the gland. The smaller lesion outlined in blue in c is outside the plane of the elastogram.
  • 75. Prostate adenocarcinoma with Gleason score of 9/10, involving two of two core samples and approximately 60% of the tissue. (a) Right parasagittal conventional US image shows that the lesion (arrow) is mildly hypoechoic. (b) Corresponding right parasagittal US elastogram better depicts the lesion (arrow) as a dark area of low strain. (c) Photomicrograph (hematoxylin-eosin stain) of a transverse histologic section (dashed line in a and b indicates the plane of section) shows that there is good size correlation between the histologic extent of the lesion (arrow) and the lesion extent delineated on the elastogram.
  • 76. Benign Prostatic Hyperplasia The appearance of benign prostatic hyperplasia at endorectal US is variable but usually consists of a heterogeneous hypoechoic area or areas in the transitional zone . In general, foci of benign prostatic hyperplasia have elastic moduli (stiffness) that are an order of magnitude greater than those of normal prostate tissues but are less than those of prostate carcinomas. As a result,on elastograms, benign prostatic hyperplasia will appear darker than normal prostate tissue. However, the difference between benign prostatic hyperplasia and prostate carcinoma can be difficult to discern because benign prostatic hyperplasia also appears darker than the background tissues. Consequently, benign prostatic hyperplasia can represent a false-positive finding for cancer .
  • 77. Benign prostatic hyperplasia. (a) Transverse B-mode US image demonstrates a mildly prominent isoechoic area (arrow) in the central zone of the prostate gland. (b) Corresponding transverse elastogram shows that the area of benign prostatic hyperplasia (arrow) is more conspicuous. The lesion demonstrates greater stiffness than the surrounding normal prostate tissue. (c) Photomicrograph (hematoxylin-eosin stain) shows that there is good correlation between the extent of the lesion (arrow) on the histologic section and the lesion extent on the elastogram.
  • 78. Elastography of Thyroid Lesions A thyroid lesion may have different levels of stiffness within it, depending on the cellularity and composition of the nodule. Information from these elastograms helps assess the relative stiffness of the lesion compared with its surrounding tissues and within itself. However, this information cannot be used to compare the stiffness of thyroid lesions from different patients because strain changes with applied compression. When pulsation of the carotid artery is used as the compression source, strain near the carotid artery can indicate the amount of compression applied by carotid arterial pulsation. Because the thyroid gland is located between the trachea and the carotid artery, lateral expansion of the carotid artery during systole compresses the thyroid gland against the trachea. As a result, the thyroid gland expands in the anteroposterior direction, that is, in the direction of the US probe (beam axis).
  • 79. Because US is sensitive to detecting motion along the beam axis, deformation of the thyroid gland in this direction can be readily detected. Because the thyroid stiffness index is a ratio between the carotid strain and the strain in the thyroid nodule, the varying strain from change in the blood pressure did not affect the ratio.
  • 80. Example of selecting region of interest (ROI) to compute thyroid stiffness index within a nodule.
  • 81. (a) Transverse US image showing a nodule (arrows) within thyroid gland (Tg), the trachea (Tr), and the carotid artery (C). (b) Combined US image and elastogram of this nodule; average strain of stiffest region inside a thyroid nodule (ROI #1) is used as denominator of thyroid stiffness index. (c) Combined US image and elastogram of this nodule with a different scale to better visualize the areas of highest strain near the carotid artery. Average strain of the highest-strain regions near the carotid artery (ROI #2, ROI #3) is used as numerator of thyroid stiffness index. Range of the color map is adjusted to better visualize ROIs.
  • 82.
  • 83. (a) Left: Transverse US image in a 62-year-old man shows a predominantly solid nodule (arrows) occupying entire thyroid gland. Right: Elastogram shows the predominantly homogeneous appearance within nodule, with small areas of decreased stiffness (blue indicates stiffer part; red indicates softer part). This was diagnosed as a follicular lesion at FNA. (b) Left: Transverse US image in a 60-year-old woman shows a predominantly solid nodule with small cystic areas (arrows). Right: Elastogram shows the heterogeneous appearance within nodule. This was diagnosed as a nodular goiter at FNA. (c) Left: Transverse US image in a 36-year-old woman shows a hypoechoic nodule (arrows) with irregular borders and tiny punctate calcifications in it. Right: Elastogram at same level shows stiff areas within lesion (blue region). This was diagnosed as papillary carcinoma at FNA. Ccarotid artery; Trtrachea.
  • 84. Ultrasound elastography for musculoskeletal applications Disease in the musculoskeletal system results in alterations to its biomechanical properties. Although EUS techniques have been extensively employed for in vitro research of muscle and tendon biomechanics since the early 1990s, the recent introduction of EUS into commercially available ultrasound systems has driven research activity towards potential clinical applications of this novel method in the musculoskeletal system
  • 86. (a, b) Longitudinal and (c, d) transverse free-hand strain elastograms of the middle third of asymptomatic Achilles tendons (T) showing the two distinct ultrasound elastography patterns of normal tendons. Type 1 (a, c) tendons appear homogeneously stiff, (green/blue) with no distinct soft (red) areas. Type 2 (b, d) tendons appear considerably inhomogeneous with soft (red) areas (longitudinal bands or spots), which did not correspond to any changes in B- mode ultrasound. The retro- Achilles fat appears as a mosaic of green, red and blue. Note the red areas at the lateral and medial sides of the tendon in the transverse plane (c, d), corresponding to an artefact, secondary to difficulty in stabilising the hand-held transducer.
  • 88. (a) Longitudinal colour Doppler ultrasound image and (b) free-hand strain elastogram of a 23-year-old recreational runner with insertional Achilles tendinopathy. There is hypoechogenicity and neovascularity at the tendinopathic area [asterisk in (a)], which appears softer (red) compared with the stiff (blue/green) normal-appearing remaining tendon (T). The retro- Achilles fat appears as a mosaic of various levels of stiffness. (c) The elastogram of a normal (asymptomatic) Achilles tendon calcaneal insertion is presented for comparison. The retrocalcaneal bursa (asterisk) appears considerably softer (red) compared with the stiffer tendon (green/yellow). Note the presence of soft (red) foci beneath the calcaneal bone (C), which correspond to artefacts. LT, left.
  • 90. Longitudinal shear wave elastograms of a normal (a) Achilles and (c) patella tendon, as well as (b, d) a case of distal patella tendinopathy in a 23-year-old football player. The elasticity qualitative and quantitative scale is presented at the upper right corner of the images. Measurements (mean, minimum, maximum and standard deviation) within the circular region of interest (ROI) are presented in kilopascals ranging from 0 (dark blue) to 300 (dark red). (a, c) The normal Achilles and patella tendons (T) appear as homogeneous stiff (red) structures, as opposed to fat, which is homogeneously soft (blue). (a) The mean stiffness of a representative area at the mid-portion of the Achilles free tendon is 300 kPa. (d) In the case of distal patella tendinopathy, the tendinopathic area appears hypoechoic with neovascularity (asterisk). (b) In the corresponding elastogram, the abnormal area appears softer (blue; mean elasticity 40.94 kPa) compared with the stiffer normal tendon (red; mean elasticity 261.16 kPa). The small amount of fluid in the deep infrapatella bursa appears softer than the tendinopathic area (blue, mean elasticity 34.38 kPa).
  • 91. (a) (b) Axial free-hand strain elastograms of (a) normal relaxed calf muscles and (b) vastus lateralis muscle. The medial and lateral heads of gastrocnemius (M and L, respectively), as well as the vastus lateralis (VL), appear as a mosaic of intermediate or increased stiffness (green or blue colour, respectively), with scattered softer (red) areas near muscle boundaries. The subcutaneous fat appears soft (red/yellow). The homogeneous stiff appearance of soleus (S) in (a) is probably because of inadequate B-mode and ultrasound elastography data due to increased depth.
  • 92. ultrasound elastography for diffuse liver disease Accurate assessment of the degree of liver fibrosis is important for estimating prognosis and deciding on an appropriate course of treatment for cases of chronic liver disease (CLD) with various etiologies. Because of the inherent limitations of liver biopsy, there is a great need for noninvasive and reliable tests that accurately estimate the degree of liver fibrosis. Ultrasound (US) elastography is considered a non-invasive, convenient, and precise technique to grade the degree of liver fibrosis by measuring liver stiffness.
  • 93. Transient elastography of a normal and a cirrhotic patient. A
  • 94. B
  • 95. The displacement M-mode image located in the center of the monitor shows axial displacement as a function of depth (y-axis) and time (x-axis). In the healthy patient (A), the shear wave is relatively slow and the liver stiffness is low. However, in the patient with cirrhosis (B), the shear wave can propagate more rapidly though a hard tissue, and the time-depth gradient is very steep. The controlled attenuation parameter is displayed, which is known to correlate with the fatty liver severity
  • 96. Acoustic radiation force impulse imaging of a normal and a cirrhotic patient. The cylindrical region of interest (ROI) in the middle of the ultrasonogram was used as the sample volume for the measurement of stiffness. Instead of Young modulus, the propagating velocity of the shear wave is displayed. Although the grayscale ultrasonograms are similar to each other, the propagating velocity is different: 1.2 m/sec in the healthy patient (A) and 1.6 m/sec in the patient with cirrhosis (B)
  • 97. Supersonic shear-wave imaging of a normal and a cirrhotic patient. A large trapezoidal color box display of the distribution of the elastic properties of the liver. The severity of stiffness is depicted with the colored look-up table. Dark blue color represents normal liver tissue (A), and bright blue-green color represents increased liver stiffness such as liver cirrhosis (B). The round region of interest (ROI) in the color box is the Q-Box and the mean Young modulus and standard deviation in the ROI have been calculated
  • 98. A B D
  • 99. Acute exacerbation of hepatitis B virus-induced hepatitis. Grayscale ultrasonography (A) shows mild parenchymal coarseness of the liver; the liver stiffness is 10.3 kPa according to the shear wave elastography (B). Six months later, the patients recovered from liver function deterioration. The grayscale ultrasonographic feature (C) is similar to the previous study, but the liver stiffness decreases to 6 kPa (D).
  • 100.
  • 101. Elastography in the Differential Diagnosis of Enlarged Cervical Lymph Nodes Drawings show typical diagrammatic appearance of five patterns of lymph nodes. Elastographic patterns were determined on distribution and percentage of lymph node area having high elasticity (hard): pattern 1, absent or small hard area; pattern 2, hard area < 45% of lymph node; pattern 3, hard area ≥ 45%; pattern 4, peripheral hard and central soft areas; pattern 5, hard area occupying entire lymph node. Increasing tissue hardness appears in ascending order as red, yellow, green, and blue.
  • 102. Transverse sonogram of level 1 lymph node in 30-year-old man with left submandibular lymphadenopathy. Elastography image on left shows pattern 1, absent or small hard area. B-mode sonographic image on right shows score of 5, reactive. Final diagnosis from clinical and serologic findings was reactive lymph node.
  • 103. Longitudinal sonogram of level 2 lymph node in 33-year-old woman with left posterior auricular lymphadenopathy. Elastography image on left shows pattern 2, hard area < 45% of node. B-mode sonographic image on right shows score 5, reactive. Final diagnosis from clinical and serologic findings was reactive lymph node.
  • 104. Transverse sonogram of level 2 lymph node in 79-year-old man with laryngeal carcinoma. Elastography image on left shows pattern 3, hard area ≥ 45%. B-mode sonographic image on right shows score 7, metastatic. Final diagnosis by histopathology was metastatic lymph node.
  • 105. Longitudinal sonogram of level 5 lymph node in 52-year-old man with nasopharyngeal carcinoma. Elastography image on left shows pattern 4, peripheral hard and central soft areas. B-mode sonographic image on right shows score 7, metastatic. Final diagnosis by CT was metastatic lymph node.
  • 106. MR ELASTOGRAPHY Following are the some of the current applications of MRE which are being investigated : Liver MR Elastography—Magnetic resonance elastography has been widely investigated for the diagnosis of hepatic diseases and is currently used in clinical practice for fibrosis and cirrhosis assessment where the stiffness of the diseased liver is significantly higher than normal liver tissue stiffness Clinical hepatic MRE is performed at a frequency of 60 Hz using pneumatic- based pressure-activated drivers. Wave data are acquired with four phase offsets and a modified direct-inversion algorithm with multiscale capabilities is used for the estimation of the liver stiffness.
  • 107. MRE-derived stiffness of healthy liver tissue at 60 Hz compared to the stiffness of liver tissue at various stages of fibrosis. The stiffness increases gradually with the progression of the fibrosis. A cutoff of 2.93 kPa well differentiates the healthy and fibrotic livers.
  • 108.
  • 109. Hepatic MRE Results are shown from clinical hepatic MRE exams of a patient with a normal liver (top row) and a patient with a cirrhotic liver. (a,d) Conventional abdominal MR magnitude images of the two patients, showing no significant difference between the two livers. (b,e) Wave images from the MRE acquisition at 60 Hz showing shear waves with a shorter wavelength in the first patient, and a substantially longer wavelength in the second patient. (c,f) The corresponding elastograms indicating that the two livers were normal (1.7 kPa) and cirrhotic (18.83 kPa), respectively.
  • 110. MRE is also being investigated as a means to characterize hepatic tumors and it has been found that malignant liver tumors have significantly greater mean shear stiffness than benign tumors and normal liver tissue and a cutoff value of 5 kPa can differentiate malignant tumors from benign tumors and normal liver parenchyma. Breast MR Elastography—Another application of MRE that is being investigated with great interest is for the assessment of breast cancer . Breast tumors are known to be typically stiffer than benign lesions and normal breast tissue . Manual palpation is a recommended part of routine screening for breast cancer and helps in the detection of these hard masses. Contrast-enhanced MR imaging (CE-MRI) has proven to have a very high sensitivity for the detection of tumor nodules, but the specificity of the technique can be a problem leading to numerous false positives and unnecessary biopsies. MRE is being investigated as a complementary technique to CE-MRI to provide additional information about these suspicious regions and the combined technique has shown promise to increase diagnostic specificity
  • 111.
  • 112. Breast MRE (a) An axial MR magnitude image of the right breast of a patient volunteer is shown. A large adenocarcinoma is shown as the outlined, mildly hyperintense region on the lateral side of the breast. (b) A single wave image from MRE performed at 100 Hz is shown along with the corresponding elastogram (c). (d) An overlay image of the elastogram and the magnitude image shows good correlation between the tumor and the stiff region detected by MRE. Skeletal Muscle MR Elastography—MRE has also been extensively investigated for studying the stiffness of skeletal muscle since it is well known that the stiffness changes significantly depending upon the contractile state of the muscle . Skeletal muscle MRE can be used for studying the physiological response of diseased and damaged muscles. For instance, it has been found that there is a difference in the stiffness of muscles with and without neuromuscular disease
  • 113.
  • 114. Skeletal muscle MRE (a) A sagittal MR image of the calf soleus muscle with the location of the driver indicated by the arrow is shown. 100-Hz MRE wave images of the muscle are shown while exerting 0 (b), 5 (c) and 10 N/m (d) of force. The increase in the wavelength (and thus stiffness) with the increase in muscle force is easily visible and is indicated by the double sided arrows. Brain MR Elastography—Assessment of the mechanical properties of brain tissue with MRE is another area of significant research and clinical interest due to the diagnostic potential of brain tissue stiffness information as it may be related to diseases like Alzheimer’s disease, hydrocephalus, brain cancer and multiple sclerosis. While it would be difficult to use ultrasound based approaches to noninvasively assess brain mechanical properties, MRE is well suited for this application.
  • 115.
  • 116. Brain MRE (a) Shown is an axial MR magnitude image of the brain showing white matter, gray matter, and cerebrospinal fluid. (b) A single wave image from MRE performed at 60 Hz. (c) The corresponding elastogram is shown and a good correlation between the magnitude image and the stiffness estimate is evident. MR Elastography for Visualizing Anatomy—The primary applications of MRE that have been developed have focused on the study of the mechanical properties of different tissues. However, the capability of MRE to measure the displacement of tissues has also been investigated to address other clinical questions. Such applications include a shearline imaging technique designed to investigate the functionality of slip interfaces and a vibration imaging technique used to localize the functional compartments of the forearm flexor muscles
  • 117. As an example of these applications, the vibration imaging technique is introduced here, the goal of which is to locate the finger-specific functional compartments of the extrinsic forearm flexor muscles that partly control the flexion of the interphalangeal joints of the index, middle, ring and little fingers. Even though these flexor muscles (flexor digitorum superficialis, FDS and flexor digitorum profundus, FDP) do not possess anatomically distinct compartments, they can exert almost independent action on specific fingers through selective activation of functional compartments specific for each finger. The knowledge of the location and boundaries of these segments is useful for localizing techniques like MR spectroscopy and biopsy, but is difficult to obtain currently. With the proposed vibration imaging technique based on MRE principles, this can be achieved by vibrating a finger of interest independently of the others and measuring the amount of motion within the forearm musculature that is transferred through the structurally connected tendons.
  • 118.
  • 119. Functional compartments of the flexor muscles (a) MR magnitude image of the right forearm of a healthy volunteer. (b) An example wave image obtained using MRE motion encoding of tissue vibrations in the forearm induced by selectively vibrating the ring finger at 90 Hz. (c) An amplitude map obtained from wave images indicating localized regions of higher displacement corresponding to the compartments of the flexor and extensor muscles (arrows). (d) A phase map obtained from the wave images indicating that the compartments of the flexor and extensor muscles of the activated finger are moving out of phase with each other.
  • 120.
  • 121. In vivo difficulties: •Uncontrolled Motion –Respiratory –Cardiac –Patient •Requirement –Fast Acquisition
  • 122. RECENT ADVANCES AND FUTURE PROSPECTS • Intracardiac Myocardial elastography • Acute Pancreatitis risk assessment on admission • Pancreatic Masses and chronic pancreatitis • Renal elastography • Obstretics & Gynaecology e.g. Cervical stiffness, Prediction of PTL, Gynaecological cancer etc. • Vascular imaging, Non-Invasive Vascular Elastography • Optical Coherence Elastography • Organ transplantation . . . . . .and many more……
  • 123. CONCLUSION Elastography is an important research field It has also been evaluated in the clinical field for more than 10 years. It has provided a new feature to assess tissue stiffness and has shown that tissue elasticity is of great value for diagnosis. Indeed the variation of the Young’s modulus in biological tissues offers a contrast that is potentially more interesting than conventional ultrasound. If the elastography techniques developed in research laboratories start making their commercial appearance, they are not all quantitative or operator independent, the word ‘‘elastography’’ can hide very different physical phenomena. It is therefore important to know how to decrypt the underlying physics of each elastography method in order to understand their advantages and physical limitations. Shear wave based techniques have strong advantages over quasi- static techniques, as they are more reproducible, quantitative, rely on automatic shear wave generation and provide good elasticity contrast.
  • 124. The availability of true quantitative numerical data allows adjustment of the dynamic range to optimize the visualization of structures. These advantages should certainly lead to new applications of shear wave elasticity imaging, not only for diagnosis but also for follow-up. The real-time capability of some of these SW techniques also allows the development of 3D elastography imaging that should facilitate the clinical use for detection, therapy planning and monitoring in the routine clinical practice. Finally, the integration of elastography techniques in conventional ultrasound systems open the door for routine application during ultrasound examination and will allow information fusion with other imaging techniques to strengthen their diagnostic performance.
  • 125. REFERENCES AND FURTHER READING • Ultrasound elastography: Principles and techniques BY J.- L. Gennisson∗, T. Deffieux, M. Fink, M. Tanter : Diagnostic and Interventional Imaging (2013) 94, 487—495 • Tissue Elasticity Imaging BYJean Martial Mari Department of Medical Physics and Bioengineering,University College London • Ultrasound elastography for musculoskeletal applications BY E E DRAKONAK, G M ALLEN and D J WILSON : The British Journal of Radiology, 85 (2012), 1435–1445 • MAGNETIC RESONANCE ELASTOGRAPHY: A REVIEW BY Yogesh K Mariappan, Kevin J Glaser, and Richard L Ehman Clin Anat. 2010 July ; 23(5): 497–511 • Principles and clinical application of ultrasound elastography for diffuse liver disease BY Woo Kyoung Jeong, Hyo K. Lim, Hyoung-Ki Lee, Jae Moon Jo, Yongsoo Kim : Ultrasonography 2014;33:149-160
  • 126. • US Elastography ofBreast and Prostate Lesions BY Daniel T. Ginat, Stamatia V. Destounis, Richard G.Barr, Benjamin Castaneda, John G. Strang,Deborah J. Rubens : RadioGraphics 2009; 29:2007–2016 • Differential Diagnosis of Thyroid Nodules with US Elastography Using Carotid Artery Pulsation BY Manjiri Dighe, Unmin Bae, Michael L. Richardson, Theodore J. Dubinsky Satoshi Minoshima, Yongmin Kim : Radiology: Volume 248: Number 2— August 2008 • Accuracy of Sonographic Elastography in the Differential Diagnosis of Enlarged Cervical Lymph Nodes: Comparison with Conventional B-Mode Sonography BY Farzana Alam, Kumiko Naito, Jun Horiguchi, Hiroshi Fukuda, Toshihiro Tachikake, Katsuhide Ito : AJR 2008; 191:604–610 • RUMAC’s Diagnostic Ultrasound 4th Edition.