Basic Principles of Ultrasound & Basic term
Produced by International Clinical Application Team
PHYSICAL PROPERTIES
Defines Ultrasound (US) as :
sound with frequency greater than 20,000 cycles per second or 20kHz.
Advantages :
US can direct as a beam.
It obeys the laws of reflection and refraction.
It is reflected by objects of small size.
Disadvantages :
It propagates poorly through a gaseous medium.
The amount of US reflected depends on the acoustic mismatch.
Reflection and Propagation :
• Effect of propagation through gaseous zones
- poor propagation, inadequate imaging.
• Effect of propagation through dense zones
- nearly all of the US is reflected. Structures below dense zones are poorly imaged
Examples of dense materials - bone, calcium, metal.
<Definitions>
• Cycle - the combination of one rarefaction and one compression equals one cycle.
• Wavelength
the distance between the onset of peak compression or cycle to the next.
• Velocity - the velocity is the speed at which sound waves travel through a
particular medium. Velocity is equal to the frequency x wavelength.
The velocity of US through human soft tissue is 1540 meters per second.
λ
• Frequency - the number of cycles per unit of time. Frequency and wavelength
are inversely related. The higher the frequency the smaller the wavelength.
• Acoustic Impedance - simply put, acoustic impedance is dependent on the
density of the material in which sound is propagated through. The greater
the impedance the more dense the material.
• Reflection
The portion of a sound that is returned from the boundary of a medium. (echo)
• Refraction
The change of sound direction on passing from one medium to another.
• Acoustic Mismatch
The boundary between two different media where reflection and refraction occurs
• Attenuation
The decrease in amplitude and intensity as a sound wave travels through a medium
<Types of Echoes>
• Specular : echoes originating from relatively large, regularly shaped objects with
smooth surfaces. These echoes are relatively intense and angle dependent.
(i.e. IVS, valves)
• Scattered : echoes originating from relatively small, weakly reflective, irregularly
shaped objects are less angle dependant and less intense.(i.e.. blood cells)
• Frequencies:
Frequencies for adult imaging - 2.0mHz to 3.0mHz.
Frequencies for pediatric imaging - 5.0mHz to 7.5mHz and 10mHz.
Effect of higher frequencies on penetration - the higher the frequency the less
penetration, the lower the frequency the greater the penetration.
• Artifacts:
Acoustic Shadowing - the loss of information below an object because the
greater portion of the sound energy was reflected back by the object. This
occurs in objects like prosthetic valves.
Enhancement - the increase in amplitude from objects that lie behind a weakly
attenuating structure. Enhancement may occur in structures below a cyst.
Reverberation - produced from the multiple reflections from an object as the
sound energy bounces back and forth between the object and the transducers
face or dense structure.
<Resolution>
• Lateral resolution
the ability to resolve objects side by side. Lateral resolution is proportionally affected by
the frequency, the higher the frequency the greater the lateral resolution. Higher
frequency transducers are used in fetal and pediatric echocardiography because the
lateral resolution displays the smaller structures in those patients and there is less need
for depth penetration. Lower frequencies are used for adults where structures are larger
and the need for greater depth penetration is important.
• Axial Resolution
Axial resolution is the ability to resolve objects that lie one above the other. Axial
resolution is inversely proportional to the frequency of the transducer depending on the
size of the patient. The higher the frequency the lower the axial resolution is in large
patients. This state results from the rapid absorption of the ultrasound energy with
lower penetration. Lower frequencies are utilized to increase depth of penetration.
• Depth of Penetration
Higher frequencies are attenuated by tissue more than lower frequencies. This means
that the higher the frequency the greater the resolution but the lower the depth of
penetration. User lower frequencies for adults and higher frequencies for children. The
advent of harmonic imaging allows the use of a lower frequency pulse to be picked up
and sampled at a higher frequency (the second harmonic) where the low frequency
allows greater penetration and high frequency provides better resolution.
2.0mHz 5.0mHz
Axial Decrease Increase
Lateral Decrease Increase
Penetration Increase Decrease
<Basic Components of the Imaging System>
• Transducer
the probe housing the elements, backing material, electrodes, matching layer
and protective face that both sends and receives the sound waves.
• Transmitter - the component that creates the impulses sent to the transducer
to generate sound energy. Also called the pulser.
• Receiver - the component that receives the current generated in the transducer
from the returning sound energy.
• Amplifier - the component that amplifies the returning signals and prepares
them to be displayed on the CRT.
<Instrumentation>
• Depth : Adjusts the depth from which the returning ultrasound signals will be
displayed. Most ultrasound systems have a maximum depth of 24cm.
The deeper the depth the slower the frame rate of the image. A shallow depth
results in a higher frame rate.
• Output gain (power) : Adjusts the amount of power used to send ultrasound into
the body. The higher the output gain the stronger the returning signal and the
better The signal to noise ratio.
• Receiver gain : Adjusts the amount of power used to amplify the returning
ultrasound signal. Higher receiver gains result in higher levels of signal and noise
in the image.
• TGC : Depth gain compensation adjusts the amount of power used to amplify the
returning ultrasound signals at a specific depth. Higher TGC levels result in higher
levels of signal and noise at a given depth. The TGC sliders are used to ensure a
uniform display of the gray scale intensity throughout the display. Strong signals
can be decreased and weak signals can be increased at specific depths
• Focal zone : Modern phased array transducers can be dynamically focused.
As the focal zone control is adjusted up and down a marker can be seen moving
up and down a marker can be seen moving up and down pointing at the
current zone of focus.
• Transmit frequency : Most of today's cardiac ultrasound systems have phased
array transducers which use wide band width technology. These wide band width
Transducers can be set to selectively receive anywhere from 2 to 4 frequencies
within a given range.
• Reject :
Filters out low-level signals so that only the stronger returning signals are
displayed. Reject can be used to decrease the amount of noise in the image if the
noise has a significantly lower intensity than the signals returning from
anatomic structure.
• EE (Edge Enhancement) :
The boundaries between strong and week returning signals are enhanced
by increasing the strong signal and decreasing the weak signal.
• DR (Dynamic range) :
Determines the number of gray shades used to map the gray scale image on
the display. Higher compression results in more shades of gray (a softer looking
image). Lower compression results in a more bi-stable (black and white)
image with fewer shades of gray.
• Post processing :
Sometimes referred to as gamma curves. Post processing
of the gray levels allows the user to determine the gray scale intensity that will
be assigned to the different returning signals.
• Persistence (Frame Average) :
Normally a frame is built by memory and then immediately sent to the display.
Real image data can be enhanced and noise can be averaged out by performing
frame averaging.
• Sweep speed :
The display rate for the M-mode or Doppler data. Common display rates are
12.5, 25, 50 and 100 mm/sec
• Sample size :
The length of the pulsed Doppler sample gate can be adjusted from 1 to 10 mm
on most ultrasound systems. Increasing the sample gate, increases the amount
of time that the system samples (listens for) returning signals.
• Sample position :
The pulsed and continuous wave Doppler cursor position can be adjusted from side
to side across the field of view. The pulsed Doppler sample volume can be adjusted
up and down within the field of view but is limited to the range of the selected PRF.
• PRF :
The pulse repetition frequency is the number of times that a pulse (pulsed wave
Doppler) is repeatedly sent during one second. The PRF determines the scale of the
pulsed Doppler display and is labeled ‘scale’ on most ultrasound systems.
• Filter :
The low velocity component of the pulsed or continuous wave Doppler signal can
be rejected by adjusting the Doppler filters.
• Baseline shift :
Flow towards the transducer is displayed above the spectral Doppler baseline.
Flow away from the transducer is displayed below the spectral Doppler baseline.
The position of the baseline can be adjusted up and down within the spectral Doppler
display.
Pulsed wave Doppler (PW) :
Pulsed wave Doppler uses a single transducer element which sends out short bursts of
ultrasound. Range resolution (the ability to localize and analyze blood flow at specific
depths) can be achieved by sampling the returning bursts of ultrasound at various time
intervals. The major limitation of standard PW Doppler
Is its inability to measure high velocities.
Advantage Is range specific thus allowing for the measurement of various flow
characteristics in small selected regions of interest.
Disadvantage Spectral wrap around (aliasing) at relatively low velocities
Continuous wave Doppler (CW) :
Continuous wave Doppler uses two side by side transducers elements. One element
continuously sends (transmits) while the other element continuously receives. This
method will allow maximum velocities to be measured but has no range resolution CW
Doppler samples all blood flow in its path. A spectrum of Doppler frequencies are
observed. Only the highest velocity in the Doppler path can be clearly measured. It is
possible however to distinguish between flow patterns which have different time and
intensity relationships.
Advantage Can display high velocities without aliasing.
Disadvantage Displays all velocities in its path. The non-imaging probe requires a
High degree of skill to operate.

More Related Content

What's hot

Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
u.surgery
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JH
hari baskar
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
MUBOSScz
 
Physics of Ultrasound Imaging
Physics of Ultrasound ImagingPhysics of Ultrasound Imaging
Physics of Ultrasound Imaging
u.surgery
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolution
Vallabhaneni Bhupal
 
Medical Uses Of Ultrasound
Medical Uses Of UltrasoundMedical Uses Of Ultrasound
Medical Uses Of Ultrasound
u.surgery
 

What's hot (20)

Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer Types
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JH
 
Basics Physics of ultrasound
Basics Physics of ultrasoundBasics Physics of ultrasound
Basics Physics of ultrasound
 
Basics of Ultrasound
Basics of Ultrasound Basics of Ultrasound
Basics of Ultrasound
 
PRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHYPRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHY
 
Ultrasound Basic Knobology & Controls
Ultrasound Basic Knobology & ControlsUltrasound Basic Knobology & Controls
Ultrasound Basic Knobology & Controls
 
Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducers
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
 
Physics of Ultrasound Imaging
Physics of Ultrasound ImagingPhysics of Ultrasound Imaging
Physics of Ultrasound Imaging
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physics
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolution
 
Ultrasound physics and instumentation
Ultrasound physics and instumentationUltrasound physics and instumentation
Ultrasound physics and instumentation
 
Physics of ultrasound
Physics of ultrasoundPhysics of ultrasound
Physics of ultrasound
 
Ultrasound Basics.ppt
Ultrasound Basics.pptUltrasound Basics.ppt
Ultrasound Basics.ppt
 
Ultrasound instrumentation
Ultrasound instrumentationUltrasound instrumentation
Ultrasound instrumentation
 
Basic physics of Ultrasound
Basic physics of UltrasoundBasic physics of Ultrasound
Basic physics of Ultrasound
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matter
 
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
 
Medical Uses Of Ultrasound
Medical Uses Of UltrasoundMedical Uses Of Ultrasound
Medical Uses Of Ultrasound
 

Viewers also liked

Knobology 01 2010 (1)
Knobology 01 2010 (1)Knobology 01 2010 (1)
Knobology 01 2010 (1)
pavan511
 
General-Purpose Ultrasound
General-Purpose UltrasoundGeneral-Purpose Ultrasound
General-Purpose Ultrasound
marianahu123
 
Ultrasound Physics & Knobology
Ultrasound Physics & KnobologyUltrasound Physics & Knobology
Ultrasound Physics & Knobology
u.surgery
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imaging
othman alameen
 
Understanding ultrasound
Understanding ultrasoundUnderstanding ultrasound
Understanding ultrasound
Mohammad Amir
 

Viewers also liked (11)

Computed tomography
Computed tomographyComputed tomography
Computed tomography
 
STANDARD 2D PROBE REPAIR
STANDARD 2D PROBE REPAIRSTANDARD 2D PROBE REPAIR
STANDARD 2D PROBE REPAIR
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 
Knobology 01 2010 (1)
Knobology 01 2010 (1)Knobology 01 2010 (1)
Knobology 01 2010 (1)
 
General-Purpose Ultrasound
General-Purpose UltrasoundGeneral-Purpose Ultrasound
General-Purpose Ultrasound
 
Ultrasound Physics & Knobology
Ultrasound Physics & KnobologyUltrasound Physics & Knobology
Ultrasound Physics & Knobology
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imaging
 
14 Lecture Animation Ppt
14 Lecture Animation Ppt14 Lecture Animation Ppt
14 Lecture Animation Ppt
 
Basics of echo & principles of doppler echocardiography
Basics of echo & principles of doppler echocardiographyBasics of echo & principles of doppler echocardiography
Basics of echo & principles of doppler echocardiography
 
Understanding ultrasound
Understanding ultrasoundUnderstanding ultrasound
Understanding ultrasound
 
Basics of Ultrasound
Basics of UltrasoundBasics of Ultrasound
Basics of Ultrasound
 

Similar to 01 basic principles of ultrasound & basic term

basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
littlealphonsa
 

Similar to 01 basic principles of ultrasound & basic term (20)

Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentation
 
Ultrasound and Doppler physics
Ultrasound and Doppler physicsUltrasound and Doppler physics
Ultrasound and Doppler physics
 
Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentation
 
USphysics.pptx
USphysics.pptxUSphysics.pptx
USphysics.pptx
 
Applications of Ultrasound in Medicine
Applications of Ultrasound in MedicineApplications of Ultrasound in Medicine
Applications of Ultrasound in Medicine
 
Us physics (4)
Us physics (4)Us physics (4)
Us physics (4)
 
Us hand book (1)
Us hand book (1)Us hand book (1)
Us hand book (1)
 
ultra sound.pptx
ultra sound.pptxultra sound.pptx
ultra sound.pptx
 
Us physics (9)
Us physics (9)Us physics (9)
Us physics (9)
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applications
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdf
 
Principles of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxPrinciples of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptx
 
Basics/Physics of Echocardiography
Basics/Physics of EchocardiographyBasics/Physics of Echocardiography
Basics/Physics of Echocardiography
 
BASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptxBASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptx
 
Basic of ultrasound
Basic of ultrasoundBasic of ultrasound
Basic of ultrasound
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
Physics of ultrasound and echocardiography
Physics of ultrasound and echocardiographyPhysics of ultrasound and echocardiography
Physics of ultrasound and echocardiography
 
Ultrasound &amp; properties
Ultrasound &amp;  propertiesUltrasound &amp;  properties
Ultrasound &amp; properties
 
Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...
 

Recently uploaded

Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
ZurliaSoop
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
KarakKing
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 

Recently uploaded (20)

How to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSHow to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POS
 
Jamworks pilot and AI at Jisc (20/03/2024)
Jamworks pilot and AI at Jisc (20/03/2024)Jamworks pilot and AI at Jisc (20/03/2024)
Jamworks pilot and AI at Jisc (20/03/2024)
 
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptxHMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
 
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
SOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning PresentationSOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning Presentation
 
Understanding Accommodations and Modifications
Understanding  Accommodations and ModificationsUnderstanding  Accommodations and Modifications
Understanding Accommodations and Modifications
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
Wellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptxWellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptx
 
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptxHMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Single or Multiple melodic lines structure
Single or Multiple melodic lines structureSingle or Multiple melodic lines structure
Single or Multiple melodic lines structure
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 

01 basic principles of ultrasound & basic term

  • 1. Basic Principles of Ultrasound & Basic term Produced by International Clinical Application Team
  • 2. PHYSICAL PROPERTIES Defines Ultrasound (US) as : sound with frequency greater than 20,000 cycles per second or 20kHz. Advantages : US can direct as a beam. It obeys the laws of reflection and refraction. It is reflected by objects of small size. Disadvantages : It propagates poorly through a gaseous medium. The amount of US reflected depends on the acoustic mismatch. Reflection and Propagation : • Effect of propagation through gaseous zones - poor propagation, inadequate imaging. • Effect of propagation through dense zones - nearly all of the US is reflected. Structures below dense zones are poorly imaged Examples of dense materials - bone, calcium, metal.
  • 3. <Definitions> • Cycle - the combination of one rarefaction and one compression equals one cycle. • Wavelength the distance between the onset of peak compression or cycle to the next. • Velocity - the velocity is the speed at which sound waves travel through a particular medium. Velocity is equal to the frequency x wavelength. The velocity of US through human soft tissue is 1540 meters per second. λ
  • 4. • Frequency - the number of cycles per unit of time. Frequency and wavelength are inversely related. The higher the frequency the smaller the wavelength. • Acoustic Impedance - simply put, acoustic impedance is dependent on the density of the material in which sound is propagated through. The greater the impedance the more dense the material. • Reflection The portion of a sound that is returned from the boundary of a medium. (echo) • Refraction The change of sound direction on passing from one medium to another. • Acoustic Mismatch The boundary between two different media where reflection and refraction occurs • Attenuation The decrease in amplitude and intensity as a sound wave travels through a medium
  • 5. <Types of Echoes> • Specular : echoes originating from relatively large, regularly shaped objects with smooth surfaces. These echoes are relatively intense and angle dependent. (i.e. IVS, valves) • Scattered : echoes originating from relatively small, weakly reflective, irregularly shaped objects are less angle dependant and less intense.(i.e.. blood cells) • Frequencies: Frequencies for adult imaging - 2.0mHz to 3.0mHz. Frequencies for pediatric imaging - 5.0mHz to 7.5mHz and 10mHz. Effect of higher frequencies on penetration - the higher the frequency the less penetration, the lower the frequency the greater the penetration.
  • 6. • Artifacts: Acoustic Shadowing - the loss of information below an object because the greater portion of the sound energy was reflected back by the object. This occurs in objects like prosthetic valves. Enhancement - the increase in amplitude from objects that lie behind a weakly attenuating structure. Enhancement may occur in structures below a cyst. Reverberation - produced from the multiple reflections from an object as the sound energy bounces back and forth between the object and the transducers face or dense structure.
  • 7. <Resolution> • Lateral resolution the ability to resolve objects side by side. Lateral resolution is proportionally affected by the frequency, the higher the frequency the greater the lateral resolution. Higher frequency transducers are used in fetal and pediatric echocardiography because the lateral resolution displays the smaller structures in those patients and there is less need for depth penetration. Lower frequencies are used for adults where structures are larger and the need for greater depth penetration is important. • Axial Resolution Axial resolution is the ability to resolve objects that lie one above the other. Axial resolution is inversely proportional to the frequency of the transducer depending on the size of the patient. The higher the frequency the lower the axial resolution is in large patients. This state results from the rapid absorption of the ultrasound energy with lower penetration. Lower frequencies are utilized to increase depth of penetration. • Depth of Penetration Higher frequencies are attenuated by tissue more than lower frequencies. This means that the higher the frequency the greater the resolution but the lower the depth of penetration. User lower frequencies for adults and higher frequencies for children. The advent of harmonic imaging allows the use of a lower frequency pulse to be picked up and sampled at a higher frequency (the second harmonic) where the low frequency allows greater penetration and high frequency provides better resolution. 2.0mHz 5.0mHz Axial Decrease Increase Lateral Decrease Increase Penetration Increase Decrease
  • 8. <Basic Components of the Imaging System> • Transducer the probe housing the elements, backing material, electrodes, matching layer and protective face that both sends and receives the sound waves. • Transmitter - the component that creates the impulses sent to the transducer to generate sound energy. Also called the pulser. • Receiver - the component that receives the current generated in the transducer from the returning sound energy. • Amplifier - the component that amplifies the returning signals and prepares them to be displayed on the CRT.
  • 9. <Instrumentation> • Depth : Adjusts the depth from which the returning ultrasound signals will be displayed. Most ultrasound systems have a maximum depth of 24cm. The deeper the depth the slower the frame rate of the image. A shallow depth results in a higher frame rate. • Output gain (power) : Adjusts the amount of power used to send ultrasound into the body. The higher the output gain the stronger the returning signal and the better The signal to noise ratio. • Receiver gain : Adjusts the amount of power used to amplify the returning ultrasound signal. Higher receiver gains result in higher levels of signal and noise in the image. • TGC : Depth gain compensation adjusts the amount of power used to amplify the returning ultrasound signals at a specific depth. Higher TGC levels result in higher levels of signal and noise at a given depth. The TGC sliders are used to ensure a uniform display of the gray scale intensity throughout the display. Strong signals can be decreased and weak signals can be increased at specific depths • Focal zone : Modern phased array transducers can be dynamically focused. As the focal zone control is adjusted up and down a marker can be seen moving up and down a marker can be seen moving up and down pointing at the current zone of focus. • Transmit frequency : Most of today's cardiac ultrasound systems have phased array transducers which use wide band width technology. These wide band width Transducers can be set to selectively receive anywhere from 2 to 4 frequencies within a given range.
  • 10. • Reject : Filters out low-level signals so that only the stronger returning signals are displayed. Reject can be used to decrease the amount of noise in the image if the noise has a significantly lower intensity than the signals returning from anatomic structure. • EE (Edge Enhancement) : The boundaries between strong and week returning signals are enhanced by increasing the strong signal and decreasing the weak signal. • DR (Dynamic range) : Determines the number of gray shades used to map the gray scale image on the display. Higher compression results in more shades of gray (a softer looking image). Lower compression results in a more bi-stable (black and white) image with fewer shades of gray. • Post processing : Sometimes referred to as gamma curves. Post processing of the gray levels allows the user to determine the gray scale intensity that will be assigned to the different returning signals. • Persistence (Frame Average) : Normally a frame is built by memory and then immediately sent to the display. Real image data can be enhanced and noise can be averaged out by performing frame averaging.
  • 11. • Sweep speed : The display rate for the M-mode or Doppler data. Common display rates are 12.5, 25, 50 and 100 mm/sec • Sample size : The length of the pulsed Doppler sample gate can be adjusted from 1 to 10 mm on most ultrasound systems. Increasing the sample gate, increases the amount of time that the system samples (listens for) returning signals. • Sample position : The pulsed and continuous wave Doppler cursor position can be adjusted from side to side across the field of view. The pulsed Doppler sample volume can be adjusted up and down within the field of view but is limited to the range of the selected PRF. • PRF : The pulse repetition frequency is the number of times that a pulse (pulsed wave Doppler) is repeatedly sent during one second. The PRF determines the scale of the pulsed Doppler display and is labeled ‘scale’ on most ultrasound systems. • Filter : The low velocity component of the pulsed or continuous wave Doppler signal can be rejected by adjusting the Doppler filters. • Baseline shift : Flow towards the transducer is displayed above the spectral Doppler baseline. Flow away from the transducer is displayed below the spectral Doppler baseline. The position of the baseline can be adjusted up and down within the spectral Doppler display.
  • 12. Pulsed wave Doppler (PW) : Pulsed wave Doppler uses a single transducer element which sends out short bursts of ultrasound. Range resolution (the ability to localize and analyze blood flow at specific depths) can be achieved by sampling the returning bursts of ultrasound at various time intervals. The major limitation of standard PW Doppler Is its inability to measure high velocities. Advantage Is range specific thus allowing for the measurement of various flow characteristics in small selected regions of interest. Disadvantage Spectral wrap around (aliasing) at relatively low velocities Continuous wave Doppler (CW) : Continuous wave Doppler uses two side by side transducers elements. One element continuously sends (transmits) while the other element continuously receives. This method will allow maximum velocities to be measured but has no range resolution CW Doppler samples all blood flow in its path. A spectrum of Doppler frequencies are observed. Only the highest velocity in the Doppler path can be clearly measured. It is possible however to distinguish between flow patterns which have different time and intensity relationships. Advantage Can display high velocities without aliasing. Disadvantage Displays all velocities in its path. The non-imaging probe requires a High degree of skill to operate.