SlideShare a Scribd company logo
Cat. No. 2285 R3
EasyLyte
Service Manual
EDIC@
Na/K
Na/K/Cl
Na/K/Li
Analyzers, Serial Numbers 5000 and above
EasySampler
Copyright ©1995 Medica
Information contained herein is believed by Medica to be accurate; in any event, no responsibility, whether express or implied, is assumed hereby
by Medica for or in connection with the use thereof, or for infringement of any third party rights which might arise therefrom, or from any represen-
tation or omissions contained therein. Information is subject to change and/or updating without notice.
Flow Chart Yes
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
Yes No
CALIBRATE NOW?
ANALYZE BLOOD?
DAILY CLEANER?
STANDBY MODE?
✽SECOND MENU?✽
✽ SEE RESULTS? ✽
✽ PACK USAGE? ✽
✽ QUAL CONTROL? ✽
✽ OPER FUNCTS? ✽
✽ DIAGNOSTICS? ✽
✽ DATE/RANGES? ✽
Whole Blood, Serum, or Plasma Analysis
End of Day Cleaning
Fluid Conservation, Maintenance
Stored Results Displayed, Printed, Deleted
Check % Pack Usage, Install New Pack,
Reset % Countdown
Run Controls, See Last Control,
See/Print Statistics, Delete Results
"On Demand" Calibration, Wash, Solutions Purge,
User Options, Probe Wiping, I.D.#, Syringe Sample,
Set Correlation,Vet Mode, Select Language,
AutoSampler, Printer Off
Check: Calibration Values
Electrode Values
Fluid Flow
Sample Detector
Pump Calibration
Change Day/Time, AutoCal,
Set Na/K/Cl/Li Ranges
Yes No
✽ ANALYZE URINE? ✽
Urine Analysis
Service Manual Contents
Product Description
Product Overview..........................................................................................................................1
System Overview ..........................................................................................................................1
Table of Symbols ..........................................................................................................................2
Front Assembly Components ......................................................................................................3
Front Assembly Components – Analyzer Series below 6100 ....................................................7
Electrode Housing Components ................................................................................................11
Rear Assembly Components ....................................................................................................12
Replacement List ........................................................................................................................13
Troubleshooting
Troubleshooting Overview ..........................................................................................................15
Troubleshooting Guide................................................................................................................16
AC Power Verification ................................................................................................................17
DC Power Verification ................................................................................................................20
Display Verification......................................................................................................................21
Probe Motor Verification..............................................................................................................22
Pump Motor Verification..............................................................................................................23
Printer Verification ......................................................................................................................24
Sample Detector Verification ......................................................................................................25
Signal Input Verification ..............................................................................................................26
Serial Communication Verification ..............................................................................................28
QC Tests......................................................................................................................................29
Repair
Repair Overview........................................................................................................................33
Tools ..........................................................................................................................................34
Rear Housing Removal ............................................................................................................35
Rear Housing Installation..........................................................................................................37
Power Supply Removal ............................................................................................................39
Power Supply Installation..........................................................................................................41
Printer Removal ........................................................................................................................43
Printer Installation ....................................................................................................................47
CPU Board Removal ................................................................................................................50
CPU Board Installation..............................................................................................................54
Display Removal ......................................................................................................................57
Display Installation ..................................................................................................................59
Pump Motor Removal ..............................................................................................................62
Pump Motor Installation ............................................................................................................63
Probe Motor Assembly Removal ..............................................................................................64
Probe Motor Assembly Installation............................................................................................67
Guide Rod Removal..................................................................................................................73
Guide Rod Installation ..............................................................................................................75
Electrode Jack Removal ..........................................................................................................76
Electrode Jack Installation ........................................................................................................77
EasySampler
Product Description ....................................................................................................................79
Assembly and Component Overview..........................................................................................80
Replacement List ........................................................................................................................81
Troubleshooting Guide................................................................................................................82
Alignment Verification..................................................................................................................83
Tray Placement Verification ........................................................................................................84
Carousel Test ..............................................................................................................................85
Tray Cycle Verification ................................................................................................................86
Sensor Verification ......................................................................................................................87
QC Tests......................................................................................................................................88
Repair Overview..........................................................................................................................90
Recommended Tools ..................................................................................................................90
Housing Removal........................................................................................................................91
Housing Installation ....................................................................................................................93
Spindle Column Removal............................................................................................................95
Spindle Column Installation ........................................................................................................97
Sensor Removal..........................................................................................................................99
Sensor Installation ....................................................................................................................100
Motor Removal..........................................................................................................................101
Motor Installation ......................................................................................................................103
PC Board Removal ..................................................................................................................105
PC Board Installation ................................................................................................................107
Specifications
Product Specifications ..............................................................................................................109
EasySampler Specifications ....................................................................................................110
Appendix
Display Messages......................................................................................................................111
QC TEST? Flowcharts ..............................................................................................................118
Schematics
CPU Board
CPU Board – Analyzer Series below 6100
Power Supply – Analyzer Series below 6100
Transformer Assembly
EasySampler PCB
Product Description 1
Product Description 1
Product Overview
The EasyLyte analyzer is a direct measurement ion-selective electrode (ISE) system which uses
Sodium (Na), Chloride (Cl), Potassium (K), and Lithium (Li) sensitive components to measure
electrolytes in either whole blood, serum, plasma or urine.
The analyzer software allows the operator to perform sample analysis, analyzer diagnostics and
maintenance by simply pressing the "YES" or "NO" buttons in response to prompts displayed on
the LCD screen. The EasyLyte analyzer performs the analysis both quickly and cost effectively.
Analysis takes only one minute and utilizes approximately one hundred microliters of whole
blood, serum or plasma; or four hundred microliters of diluted urine.
The analyzer can re-calibrate automatically every four hours and can perform emergency (STAT)
analysis on demand. Results are always displayed on the LCD screen and can also be docu-
mented on paper by an integrated printer.
The analyzer contains a patented reagent system including a modular reagent pack. This pack
provides a minimum of two months supply of solution and also serves as a waste receptacle for
solution and samples after analysis.
System Overview
The EasyLyte analyzer is engineered to make local service feasible. Modular construction is
used throughout the design. The basic operating system consists of a motor which drives a peri-
staltic pump. The peristaltic pump pulls the desired fluid through the system, disposing waste into
a designated compartment inside the solutions pack. A second motor drives a lead screw which
in turn drives a mechanical link-arm assembly. This link-arm assembly moves the sample probe
vertically allowing the probe inside the solutions valve to aspirate the appropriate fluid for pump-
ing: sample; standards; or air. The sample detector in the fluid line positions the fluids.
Please note that mechanical and electrical assemblies can be replaced. Follow the instructions in
the TROUBLE SHOOTING SECTION to determine if assembly replacement is necessary. Refer
to the Replacement List in this section which lists replaceable assemblies, components, and
appropriate catalog numbers.
If repair to any electrical assembly is necessary, consult Schematics in the APPENDIX
SECTION for further detail. Service personnel are encouraged to contact Medica directly with
any problems or concerns relating to electrical assemblies.
2 Product Description
NOTES Helpful information or references to other related
sections of the manual.
KEY INFORMATION Could take the EasyLyte analyzer out of operation.
WARNING Possible biological hazard to the operator.
EASYLYTE ANALYZER’S Information for EasyLyte analyzers with serial num-
bers BELOW 6100 below 6100.
ELECTRICAL HAZARD Possible electrical hazard to the operator.
Table of Symbols
6100
Product Description 3
Front Assembly Components
This section describes the main assemblies and components of the EasyLyte front assembly.
Refer to Figures P1, P2, and P3 for the location of each component.
Front Housing front case of the analyzer
Power Supply Assembly powers the analyzer
Entry Module connection for the power cable
Fuse Bay holds 2 fuses
Power Supply Cable connects the power supply to the CPU assembly
Voltage Selector Switch to select 110V or 220V power source
CPU Assembly contains the CPU board with digital components and analog sensor inputs
Firmware (EPROM) contains the software
Interface Ports for EasySampler and RS-232 for serial devices
Display Assembly shows EasyLyte prompts/results on LCD screen
LCD Screen available in 1- line or 2- line display
Display Cable connects the display to the CPU assembly
Sample Detector Cable connects the sample detector to the display board
Pump Assembly peristaltic pump to move fluids through the analyzer
Pump Motor powers pump
Pump Head Assembly contains pump body, rollers
Pump Motor Cable connects the pump to the CPU assembly
Probe Motor Assembly moves the probe arm
Probe Motor with Lead Screw powers lead screw
Mounting Bracket holds probe lead screw motor assembly in place
Probe Position Sensor positions the probe position scale
Sensor Cable connects the sensor to the CPU assembly
Probe Position Scale positions the probe arm
Probe Motor Cable connects the motor to the CPU assembly
Printer Assembly prints a hard copy of analysis, diagnostics or maintenance results
Printer Cable connects the printer to the CPU assembly
Probe Arm holds and moves the probe to aspirate samples and calibrants
Guide Rod allows probe arm to move vertically
Electrode Jack(s) connects electrodes to the CPU assembly
4 Product Description
Figure P1 Front Assembly Back View
Printer Assembly
CPU
Assembly
Power Supply Assembly
AUTOSAMPLER
P3 P1 P2 P4 P13
P6
Power Supply
Cable
Printer Cable
Pump Motor
Cable
Sensor
Cable
Probe
Motor Cable
Display Cable
Front Housing
Display
Assembly
Sample
Detector Cable
Probe Position
Sensor
Probe Position
Scale
Electrode Jack(s)
Firmware
(EPROM)
Interface Ports
Voltage Selector
Switch
Entry Module
Fuse Bay
Product Description 5
Figure P2 Front Assembly - Right Side View
Front Housing
Probe
Motor
Assembly
Mounting
Bracket
Probe
Position Scale
Probe
Motor
Guide Rod
Lead Screw
Display Cable
Printer Assembly
Probe Arm
6 Product Description
Figure P3 Front Assembly - Left Side View
Front Housing
Pump Head
Assembly
LCD Screen
Pump
Assembly
Pump Motor
Display Cable
Printer Cable
Product Description 7
Front Assembly Components – Analyzers Series Below 6100
This section describes the main assemblies and components of the EasyLyte front assembly,
analyzer series below 6100. Refer to Figures P4, P5, P6 for the location of each component.
Front Housing front case of the analyzer
Power Supply Assembly powers the analyzer
Entry Module connection for the power cable
Fuse Bay holds 2 fuses
Power Supply Cable connects the power supply to the CPU assembly
Voltage Selector Switch to select 110V or 220V power supply
CPU Assembly contains the CPU board with digital components and analog sensor inputs
Firmware (EPROM) contains the software
Interface Ports for EasySampler and RS-232 for serial devices
Display Assembly shows EasyLyte prompts/results on the LCD screen
LCD Screen available in 1-line or 2-line display
Display Cable connects the display to the CPU assembly
Sample Detector Cable connects the sample detector to the display board
Pump Assembly peristaltic pump to move fluids through the analyzer
Pump Motor powers pump
Pump Head Assembly contains pump body, rollers
Pump Motor Cable connects the pump to the CPU assembly
Probe Motor Assembly moves the probe arm
Probe Motor with Lead Screw powers lead screw
Mounting Bracket holds probe lead screw motor assembly in place
Probe Position Sensor positions the probe position scale
Sensor Cable connects the sensor to the CPU assembly
Probe Position Scale positions the probe arm
Probe Motor Cable connects the motor to the CPU assembly
Printer Assembly prints a hard copy of analysis, diagnostics or maintenance results
Printer Cable connects the printer to the CPU assembly
Probe Arm holds and moves the probe to aspirate samples and calibrants
Guide Rod allows the probe arm to move vertically
Electrode Jack(s) connects electrode to the CPU assembly
6100
8 Product Description
Figure P4 Front Assembly Back View
– Analyzers with Serial Numbers below 6100
6100
Printer Assembly
CPU
Assembly
Power Supply Assembly
AUTOSAMPLER
P3 P1 P2 P4 P13
P6
Power Supply
Cable
Printer Cable
Pump Motor
Cable
Sensor
Cable
Probe Motor
Cable
Front Housing
Display
Assembly
Sample
Detector Cable
Probe Position
Sensor
Probe Position
Scale
Electrode Jack(s) Display Cable
Entry Module
Fuse Bay
Firmware
(EPROM)
Voltage Selector
Switch
Interface Ports
Product Description 9
Figure P5 Front Assembly Right Side View
- Analyzers with Serial Numbers below 6100
Front Housing
Probe Motor
Assembly
Mounting
Bracket
Probe
Position Scale
Probe
Motor
Guide Rod
Lead Screw
Printer
Assembly
Voltage
Selector
Switch
6100
10 Product Description
Figure P6 Front Assembly Left Side View
– Analyzers with Serial Numbers below 6100
Front Housing
Pump Head
Assembly
LCD Screen
Pump
Assembly
Pump Motor
Display
Cable
Printer
Cable
6100
Product Description 11
Figure P7 - Electrode Housing Assembly
Electrode Housing Assembly Components
This section describes the components of the EasyLyte electrode assembly. Refer to Figure P7
for the location of each component.
Electrodes sodium, potassium, chloride, lithium, reference
Sample Detector detects presence of air/liquid, sets sample position inside electrodes
Internal Filling Solution 2 molar KCI solution, acts as 'salt bridge’ with reference electrode
Fill Plug prevents the internal filling solution from evaporating
Membrane Assembly connects reference electrode with other electrodes
Fill Plug
Sample Detector
Chloride Electrode
(Green Dot) or
Lithium Electrode
(Yellow Dot)
Potassium Electrode
(Red Dot)
Sodium Electrode
(Blue Dot)
Internal Filling
Solution
Reference Electrode
Membrane Assembly
12 Product Description
On EasyLyte Analyzers with serial numbers below 6100 the voltage selector is located
inside the front housing on the power supply.
Figure P8 - Rear Assembly
Rear Assembly Components
This section describes the components of the EasyLyte Rear Assembly. Refer to Figure P8 for
the location of each component.
Rear Housing rear case
Label identifies serial number
Autosampler
Port
Serial
Port
Power Cord
Entry Module
Label with
Serial Number
Voltage
Selector
Switch
Rear Housing
6100
Product Description 13
Replacement List
Consumable and assemblies items can be purchased directly from Medica.
CONSUMABLE ITEMS
Description Catalog No.
Na Electrode 2102
K Electrode 2101
Cl Electrode 2113
Li Electrode 2106
Reference Electrode 2103
Sample Detector 2257
Sample Probe 2107
Solutions Valve 2108
Fill Plug 0195
Membrane Assembly 2258
Tubing Kit 2104
EasyLyte Operator's Manual 2277
ASSEMBLIES
Description Part No.
Electrode Housing Na/K/Cl 000885
Electrode Housing Na/K/Li 000985
Electrode Housing Na/K 000177
Power Supply 001676
with CPU 001937
or CPU (Japanese or Chinese software) 001957
Power Supply 001287
with CPU 001081
Display, 1 Line 001284
Display, 2 Line 001901
Display, Graphic (Japanese or Chinese characters) 001956
Pump 001271
Probe Motor Assembly 001392
Probe Position Sensor 001306
Motor Driver 001148
– CPU 001937, positions U35, U36
– CPU 001081, positions U22, U23
Printer 001296
Electrode Jack 000173
The EasyLyte CPU Assembly has ten (10) variations:
EasyLyte Lithium, 1-line display 001937-003
EasyLyte Chloride, 1-line display 001937-004
EasyLyte Lithium, 2-line display 001937-005
EasyLyte Chloride, 2-line display 001937-006
EasyLyte Na/K, 1-line display 001937-007
EasyLyte Na/K, 2-line display 001937-008
EasyLyte Na/K, Chinese 001957-001
EasyLyte Na/K/Cl or Li, Chinese 001957-002
EasyLyte Na/K, Japanese 001957-003
EasyLyte Na/K/Cl or Li, Japanese 001957-004
The CPU Part Numbers include the EasyLyte CPU Board with software installed, programmed
to operate specifically in EasyLytes described as above. When ordering this item, please list the
specific part number (-003, -004, -005, etc.) that you require.
CPU series 001957 functions correctly only when used with the Graphic Display (displays
Japanese or Chinese characters), part number 001956.
Please note that EasyLyte Analyzers with serial Numbers below 6100 are assembled
using Power Supply, 001287 and CPU, 001081. These two items work together as a
set. Power Supply, 001676 and CPU, 001937 also work together as a set. These parts
cannot be mixed on the EasyLyte.
14 Product Description
6100
6100
Troubleshooting 15
Troubleshooting Overview
This section describes troubleshooting the mechanical and electrical functions of the EasyLyte
analyzer. This section is based on the principle of replacing suspected faulty assemblies with
new assemblies. Medica recommends keeping an inventory of functioning test assemblies
for troubleshooting the EasyLyte. Refer to the Troubleshooting Guide in this section to deter-
mine the cause and corrective action that must be taken to properly service the EasyLyte
Analyzer.
Since EasyLyte Analyzers with serial numbers below 6100 are assembled using Power
Supply, part no. 05001065 and CPU, part no. 05001063 the troubleshooting instructions
differ in certain areas. Please refer to the serial number, located on the back panel of the
analyzer, to identify which instructions to follow.
For information on troubleshooting the EasyLyte Analyzer flow system and consumable items,
consult the EasyLyte Operator’s Manual. See the Replacement List in the PRODUCT
DESCRIPTION SECTION of this manual for a complete listing of all consumable items and cata-
log numbers. For information on troubleshooting the EasyLyte EasySampler consult
Troubleshooting in the EASYSAMPLER SECTION.
6100
16 Troubleshooting
Troubleshooting Guide
Refer to the chart below to help determine the problem and corrective action needed.
Problem Description Problem Source Corrective Action Steps
Blank Display Power Malfunction or • AC Power Verification
Defective Display • DC Power Verification
• Display Verification
No response Power Malfunction or • Display Assembly Verification
to YES/NO buttons Defective Display • DC Power Verification
Blank Display, Faulty EPROM Installation • Inspect EPROM Intallation
Beeping or Power Malfunction • DC Power Verification
“MTS” Errors Probe Motor • Probe Motor Verification
Malfunction
“MTP” Errors or Pump Motor • Perform PUMP CAL test
No Pump Movement Malfunction in DIAGNOSTICS? Menu
• Pump Motor Verification
Printer Does Not Print Printer Malfunction • Printer Verification
Sample Detector Test Failure Sample Detector/Circuit • Sample Detector
Result 255 or Ø Malfunction Verification
Bad Electrode Voltages Electrodes or • Check Electrode Installation
Signal Input Malfunction and Performance
• Signal Input Verification
Serial Transmission Failure Serial Cable or • Serial Communication
Transmission Protocol Verification
Troubleshooting 17
Hot
Earth Ground
Neutral
P3 P1 P2 P4 P13
Entry Module
Fuse Bay
Figure T1
AC Power Verification
WARNING: Careless handling of AC Power, especially 220V, can be hazardous or fatal.
Only qualified personnel should work with AC Power.
1. Check that the voltage selector switch is set at the proper voltage (110V or 220V). This volt-
age must match the power outlet voltage.
2. Unplug the power cord from the entry module and plug into the outlet.
3. Check the voltage at hot and neutral at the end of the power cord. Refer to Figure T1.
For EasyLyte analyzers set to 110V, the voltage at hot should be 90V–125V.
For EasyLyte analyzers set to 220V, the voltage at hot should be 210V–240V.
If the voltage is incorrect, replace the power cord and return to normal operation. Otherwise,
proceed to STEP 4.
4. Using a small flat-tip screwdriver pry open the fuse bay. Refer to Figure T2 for the location of
fuse bay. Remove and inspect both fuses as follows:
For 110V EasyLyte analyzers, the fuses must be rated 0.8A, 250 Volts
For 220V EasyLyte analyzers, the fuses must be rated 0.4A, 250 Volts
Visually inspect the wire conductor inside the fuses. If the wire is broken or missing,
replace the fuse.
Figure T2
18 Troubleshooting
Figure T3
5. Remove the rear housing. Refer to Rear Housing Removal in the REPAIR SECTION.
6. Check for short circuits between the test points listed below:
TP12 and TP2
TP12 and TP1
TP12 and TP3
TP12 and TP10
TP12 and TP11
If a short exists, replace the CPU board. Otherwise, return to normal operation.
7. Disconnect the Power Supply cable (6-pin connector) from the CPU board, see Figure T3.
Connect the EasyLyte entry module to a wall outlet with a known good power cord.
8. Connect the reference side of the voltmeter to Pin 1 of the power supply cable (6-pin
connector).
9. Check the AC voltages at Pins 2 and 5. see Figure T3 for the location of the pins. The voltage
for each should be 7.6 ± 2V.
10. Check the voltaage across Pins 3 and 4. The voltage should be 26± 2V.
11. Measure the resistance between Pin 6 and power supply chassis. The resistance should be
zero.
If any of the voltages are incorrect, replace the power supply module and return to normal
operation.
AUTOSAMPLER
P4 P13
Power Supply
Cable
Pin 6 –> 1
Troubleshooting 19
If any of the voltages are incorrect, replace the power supply and return to normal operation.
12. Check for short circuits between the test points below:
TP11 and TP5
TP11 and TP8
TP9 and TP6
TP9 and TP7
If a short exists, replace the CPU Board and return to normal
13. Disconnect the power supply cable (13-pin connector) from the CPU board, see Figure T4.
14. Connect the reference side of the voltmeter to Pin 5 of the powr supply cable. Avoid
shorting the adjacent pins.
15. Check the DC voltages at the pin placements listed below. See Figure T4 for the pin
locations.
Test Point Value Tolerance
Pins 1,2, and 3 +5.0V ±0.15V
Pin 4 through 8 0.0V ±0.15V
Pins 9 and 10 35.0V ±3.00V
Pin 11 –5.0V ±0.15V
Pin 12 +5.0V ±0.15V
Pin 13 0.0V ±0.15V
6100
6100
6100
P4 P13
Power Supply
Cable
Pin 13 –> 1
Figure T4
20 Troubleshooting
DC Power Verification
1. Remove the rear housing. See Rear Housing Removal in the REPAIR SECTION for details.
2. Connect the power cord first to an outlet, then to the entry module.
3. Connect the reference side of the voltage meter to TP12 on the CPU board.
4. Verify the voltage at each test point listed below.
Test Point Value Tolerance
TP1 5.00 V +/- 0.15 V
TP2 34.0 V +/- 2.0 V
TP3 5.00 V +/- 0.15 V
TP10 2.50 V +/- 0.10 V
TP11 -5.00 V +/- 0.15 V
5. For EasyLyte serial numbers below 6100, use TP11 as the reference point. Verify the
voltage at each test point listed below.
Test Point Value Tolerance
TP5 5.00 V +/- 0.15 V
TP8 34.0 V +/- 2.0 V
TP9 0.00 V +/- 0.15 V
TP6 5.00 V +/- 0.10 V
TP7 -5.00 V +/- 0.15 V
6. If any of the voltage values in STEP 4 OR STEP 5 are not within specification, AND
the AC power verification is successful, replace the CPU board.
6100
Troubleshooting 21
Display Verification
1. Disconnect the display cable from the CPU board. See Display Assembly Removal in the
REPAIR SECTION for details.
2. Connect a known good display assembly to the CPU board and connect the main power cord.
Verify that the test display is operative. Based on the test results, take the appropriate
corrective action.
• Display Blank and YES/NO buttons DO NOT operate:
Remove the test display cable. Perform AC and DC power verification. If the AC
and DC power tests are successful, replace the CPU board with a new CPU
board, and return to normal operation. See CPU Assembly in the REPAIR SEC-
TION for details.
• Display On and YES/NO buttons function:
Remove the original display assembly and replace with a new display assembly.
See Display Assembly in the REPAIR SECTION for details.
22 Troubleshooting
Probe Motor Verification
1. Disconnect the power cord, wait 5 seconds, then reinstall the power. The probe arm should
move down and back up. If there are no error messages, resume normal operation. If an
“MTS” error occurs, proceed to STEP 2.
2. Remove the sample probe from the probe arm. For details refer to EasyLyte Operator’s
Manual. Remove and reinstall the power cord. The probe arm should move down and back
up. If there are no error messages examine and if necessary replace the sample probe, solu-
tions valve and probe wiper. If an “MTS” error occurs, proceed to STEP 3.
3. If the probe arm does not move at all perform the following operations.
A. Visually inspect that the actuator and retaining ring are correctly installed on the probe
motor lead screw. Refer to Figure R64 in the REPAIR SECTION. If the installation is
incorrect, refer to CPU Removal and Probe Motor/Lead Screw Installation in the
REPAIR SECTION for details.
B. Check that the probe motor wires are completely seated inside the 4-pin connector as
follows.
– Remove the probe motor connector from the bottom left of the CPU Board
(position P3).
– Compress the strain-relief cap against the probe motor connector to firmly seat
the wires.
– Reinstall the probe motor connector into the CPU Board.
C. Repeat STEP 1. If no errors occur, resume normal operation.
4. If the probe arm moves and stops anywhere except the top position, or if the probe arm
vibrates against the top of the EasyLyte housing, replace the probe position sensor. To
replace the sensor, first remove the CPU Board and the probe position scale. Refer to Probe
Motor/Lead Screw in the REPAIR SECTION for details.
5. If the Probe does not move at all, check for a short circuit on the CPU Board between TP12
and TP2. If there is a short circuit, replace the motor driver at position U36.See Assemblies in
the PRODUCT DESCRIPTION SECTION.
For EasyLyte serial numbers below 6100, check for a short circuit on the CPU Board
between TP8 and TP11. If there is a short circuit, replace the motor driver at position
U23. See Assemblies in the PRODUCT DESCRIPTION SECTION.
6. If no short circuit is found, replace the probe motor assembly. Refer to the Probe Motor
Assembly in the REPAIR SECTION for details.
6100
Troubleshooting 23
Pump Motor Verification
1. Enter the QC TEST? function and proceed to the PUMP TEST? Refer to QC Tests in this
section for further instructions.
2. At the PUMP TEST? prompt, press YES.
3. The pump should rotate ten revolutions.
4. If the pump vibrates or does not run smoothly, ensure that the pump motor wires are properly
seated inside the 6 pin connector as detailed below.
A. Remove the pump motor connector from the bottom left of the CPU board
See Figure P1 in the PRODUCT DESCRIPTION SECTION.
B. Compress the strain-relief cap against the pump motor connector to firmly seat the wires.
C. Reinstall the pump motor connector into the CPU board.
5. Repeat the Pump Test. If the pump malfunctions, replace the pump motor with a test motor.
See Pump Motor in the REPAIR SECTION for futher instructions.
6. Repeat the Pump Test with the test motor. Based on the test results, take the appropriate cor-
rective action.
• Motor Operative: Replace the test motor with a new motor. See Pump Motor in the
REPAIR SECTION.
• Motor Malfunction: Check for a short on the CPU board between TP12 and TP2. If there is
a short circuit, replace the pump motor driver at U35. See Assemblies
in the PRODUCT DESCRIPTION SECTION.
7. For EasyLytes with serial numbers below 6100, check for a short circuit between TP8
and TP11.
• Short Circuit: Replace the pump motor driver at position U22.
• No Short Circuit: Replace the pump motor assembly. See Pump Motor in the
REPAIR SECTION for details.
6100
24 Troubleshooting
Printer Verification
1. Press the printer paper advance button. If the printer head moves proceed to STEP 2. If the
printer head does not move, proceed to STEP 3.
2. Verify that the printer is on. Turn the printer on by pressing NO to the PRINTER OFF? ques-
tion in the USER OPTIONS section.
3. If the printer is still not working, remove the power cord from the analyzer and wait for ten sec-
onds, then install the power cord back into the entry module. Press the printer paper advance
button. The printer head should move across the paper and the paper should advance for-
ward.
4. If the printer is still not working, remove the printer cover and check for any paper obstruction
in the printer head area. Remove any obstructions if possible, without damaging the printer
assembly or paper tear bar. If an obstruction cannot be removed, replace the Printer
Assembly. Refer to Printer Removal and Installation in the REPAIR SECTION for details.
5. If there are no obstructions, but the printer does not work, perform the following steps.
A. Remove the rear housing as detailed in the REPAIR SECTION .
B. Remove the printer cable at the bottom right of the CPU board.
C. Compress the strain-relief cap against the probe motor connector to firmly seat the printer
cable.
D. Reinstall the printer cable into the CPU Board.
6. If the printer still does not print, connect a test printer assembly to the CPU Board. If the test
printer works, install a new printer assembly. If the test printer does not work, replace the CPU
board. Refer to CPU Installation in the REPAIR SECTION for details.
Troubleshooting 25
Sample Detector Verification
1. Remove the electrode housing from the analyzer. See the EasyLyte Operators Manual for fur-
ther details.
2. Remove the sample detector from the front housing.
3. Install the sample detector into the sample detector receptacle.
Do not install the electrode housing or pump tubing at this time.
4. Proceed to the DIAGNOSTICS? menu and perform the SAMPLE DETECTOR TEST?
• With the sample detector installed, the value on the display screen should be less than 200.
• With the test in progress, remove the sample detector. The display value should rapidly
change to 255.
If the sample detector test meets these specifications, resume normal operation.
A value of 255 with the sample detector installed indicates a bad connection. Proceed to
STEP 5 to check the sample detector circuit.
5. Reinstall the sample detector. Look at the left side of the sample detector. A red light must
shine from the LED hole. If there is no red light, install a known good sample detector. If there
is still no red light, replace the Display Assembly. Refer to the REPAIR SECTION for informa-
tion.
6. If the TEST sample detector red light is on, check the four connecting pins on the ORIGINAL
sample detector. The pins must face straight out forming a rectangle. If the pins are bent
against the black plastic body, straighten the pins using a small flat head screwdriver.
7. Reinstall the original sample detector and repeat STEP 4.
• If the red light comes on: Reassemble the EasyLyte analyzer and perform a SOLUTIONS
PURGE and a SAMPLE DET. TEST. Resume normal operation.
• If the red light remains OFF: Install a new sample detector.
Signal Input Verification
1. Remove and inspect electrodes and electrode housing for liquid leakage and salt build-up. All
electrodes and the electrode housing surfaces MUST be completely dry. See EasyLyte
Operator’s Manual for further information on consumable items.
2. Inspect the electrode pin receptacles on the EasyLyte analyzer front housing. Remove any liq-
uid or salt from these areas.
3. Enter * DIAGNOSTICS? * and answer YES to ELECTRODE VALUES? The display most
likely will show “249 249 249”.
The following steps require the use of the Mini-Test High Impedance Voltage Source,
part no. 001374. If this tool is not available, the remaining tests cannot be
performed.
4. Connect the Voltage Source to the EasyLyte analyzer as follows. Refer to Figure T5.
A. Connect the three red wires from the Voltage Source to the K+, Na+, and CL-/Li+
electrode receptacles
B. Connect the black wire from the Voltage Source to the reference electrode receptacle.
C. Connect the remaining black wire into the back connector on the Cl-/Li+ red wire.
This step will short out all three electrode channels.
Figure T5
5. Read the EasyLyte display. The electrode value specification is 0 mV to 12 mV.
For analyzers with serial numbers below 6100, the electrode specification is
-12 mV to +12 mV.
26 Troubleshooting
6100
Troubleshooting 27
6. Disconnect the black wire from the back connector on the Cl-/Li+ red wire.
7. Turn on the Voltage Source.
8. Set the Voltage Source to the group 1 settings listed in the table below. Three new individual
values will appear on the screen. Wait a few seconds for the values to stabilize then verify that
the values are within specification. Press YES to print the values.
9. Set the Voltage Source to the group 2 settings, listed in the table below, and repeat STEP 8.
10. Calculate the DIFFERENCE between the group 1 values and group 2 values. Verify that the
values are within specification.
Settings Group 1 Group 2
Polarity setting acid (positive) acid (positive)
Impedance setting HI HI
Offset setting 8-6pH 10-4 pH
Voltage Values
Individual values 59 mV± 10mV 177 mV ± 10 mV
Difference between
group 1 and 2 118 mV ± 3 mV
11. If any of the voltage values are out of specification, examine the electrode jacks for corrosion
and verify the connection between the electrode jack and CPU board. Refer to
Electrode Jack Removal in the REPAIR SECTION. If the electrode jacks are not corroded
and are properly connected, replace the CPU board with a new CPU board. See CPU
Assembly in the REPAIR SECTION for details.
12. Turn off the Voltage Source.
If the technician has any questions about the reliability of the mini-tester, install a new 9
volt battery and repeat the above procedure.
28 Troubleshooting
Serial
Cable
Settings
RS-232
Protocols
Serial Communication Verification
1. Connect the EasyLyte analyzer to a device with a known working serial port. Use a known
good EasyLyte serial cable, part no. 05001334. Make cross connections as required (e.g. 2 to
3; 4 to 5).
2. Refer to the following serial interface specifications.
Serial Interface Specifications
Connector 25-pin, D-Sub male,
Pin type RS-232
Pin assignments
Pin 1 Chassis, Not Connected
Pin 2 TxD
Pin 3 RxD
Pin 4 RTS
Pin 5 CTS
Pin 7 Signal Ground
Transmission Rate 1200 baud
Bit Configuration 8 data bits, 1 stop
Bit parity None
Control Codes None
All lines end with Carriage return line feed (CRLF)
All transmission is ASCII string
3. When correctly linked the EasyLyte will transmit to the external device all of the data that is
printed by the EasyLyte printer, in the same format.
Troubleshooting 29
QC Tests
1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis-
play, as shown in the chart below.
LCD Screen Displays Press
CALIBRATE NOW? NO
DAILY CLEANER? NO
STANDBY MODE? NO
SECOND MENU? YES
* SEE RESULTS? * NO
* PACK USAGE? * NO
* QUAL CONTROLS? * NO
* OPER FUNCTS? * NO
* DIAGNOSTICS? * NO
* DATE/RANGES? * YES
* CHANGE DAY/TIME? * NO
CAL FREQUENCY? NO
Na/K/Cl or Na/K/Li RANGES? YES
BLOOD Na LIMITS? YES
“Numeric value” —>“Numeric value” wait
2. Two numbers appear on the display screen separated by an arrow. A cursor will be posi-
tioned underneath the first digit of the first number.
The second number is not applicable in this procedure.
3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the
digit is correct press YES to advance to the next digit.
4. Continue until the first number is changed to “246.0”.
5. Press YES when finished. QC TEST? will appear on the screen. If an erroneous value is
input the screen displays CORRECT?. Press NO to return to STEP 2 and correct the
number.
6. Press YES to QC TEST?
30 Troubleshooting
7. Consult the table below which lists all of the QC tests and their functions. To select a test
press YES. To bypass a test press NO.
QC Test QC Tests Function
QC TEST? Access to the QC Test functions.
TURN ON MV‘S? Prints the registered Na, K, Cl or Li electrode
calibration and analysis voltages.
Calibration: Stds. A & B voltages
Blood analysis: Sample and Std. A voltages
Urine analysis: Sample and Std. B voltages
To end printing, return to TURN ON MV’s and press YES.
At PLEASE CONFIRM!, press NO.
START CLOCK? Permits changing date and time if analyzer setting is
incorrect. Press YES to PLEASE CONFIRM! to
change date and time. Please note that this
automatically turns on the electrode voltages. To turn off
the electrode voltages, press NO to all messages until
START CLOCK? appears; press YES. At PLEASE
CONFIRM!, press NO.
CHECK CAL CONSTS? Displays the factory set electrode constants. The
acceptable ranges are:
Left column number: 0.92 to 1.08
Right column number: -800 to +800
If values are out of specification contact Medica.
RAM/PROM TEST? Displays current software version and date installed.
RAM CHECK OUT? Clears the EasyLyte memory. This will erase all
stored EasyLyte calibration and sample data. After
testing this function exits the operator out of the QC
TEST? menu and resets the analyzer.
SIPPER TEST? Moves the sample probe all the way down, then up to
test mechanical performance. Abnormal movement is
indicated by the display messages:
“MTS”, “Na/K/CL” or “Na/K/Li.”
PUMP TEST? Rotates the pump 10 revolutions to test mechanical
performance. Abnormal movement is indicated by the
display message “MTP”.
TEST CAROUSEL? EasySampler QC Test Functions.
CALIB CAROUSEL? Refer to the EASYSAMPLER SECTION
FIND HOME? for details.
CYCLE TRAY?
SOL’N PURGE? Purges any air in the fluid system.
Calibration must be performed after a solution purge.
CALIBRATE NOW? Performs analyzer calibration.
6100
Troubleshooting 31
QC Test Function QC Test Function
DIAGNOSTICS? Performs diagnostic testing of the EasyLyte Analyzer
(See EasyLyte Operator’s Manual).
LAST CAL VALUES? Checks last calibration values. Acceptable ranges are
Na: 50-64, K: 50-64, Cl: 40-64, Li: 50-64.
ELECTR’D VALS? Direct electrode millivolt readings. Typical values are
Na: 35-125, K: 35-125 Cl: 35-125, Li: 30-60.
FLUID FLOW TEST? Confirms that fluid is aspirating through the analyzer.
SAMPLE DET. TEST? Tests the sample detector, which detects the differ-
ence between air and liquid in the fluid system.
PUMP CAL TEST? Checks the pump calibration.
Expected values are between 500-800.
TEST VALVE? Verifies probe/valve alignment and pump efficiency.
Calculates the number of pump motor steps to move
each solution from the valve through the probe to the
sample detector. Typical values are Flush: 800-1200,
Std B 800-1200, Std A 800-1200, Air 500-800. The
three reagent values should be 100-200% of the Air
value. In most cases, abnormal results indicate a
faulty solutions valve.
NOTE: The operator must perform a successful SOL’
N PURGE before this test.
RUN STATISTICS? Analyzes standards A & B (in the EasyLyte Solutions
Pack) as samples.
RUN CONTROLS? Analyzes 8 normal or 8 abnormal control samples.
Calculates the mean, standard deviation and coeffi-
cient of variation for control sets. NOTE: The operator
must perform a successful CALIBRATION before run-
ning this test.
TEST ALIGNMENT? Verifies the number of probe motor steps to move the
probe through each solutions valve position. Hex val-
ues are Std. A: 3F, Flush: 3F, Std. B: 3F, Air: 3F. This
test is useful only when assembling the EasyLyte
analyzer.
BURN IN? Manufacturer’s procedure to test electronic compo-
nents of the CPU assembly. Not applicable for servic-
ing the analyzer.
CONTINUOUS? Part of Burn In sequence.
QC TEST? Access to the QC Test functions. Press NO to return
to the main EasyLyte menu.
Repair 33
Repair Overview
Always consult the TROUBLESHOOTING SECTION first to determine the problem before
attempting to repair the analyzer.
Always remove the solutions pack and the electrode housing from the analyzer before
beginning any repair work. Refer to the EasyLyte Operator’s Manual for detailed instruc-
tions.
Follow the assembly and removal instructions in the order outlined in this section unless other-
wise noted. Always unplug the power cord from the analyzer before beginning any repair work.
Note that only complete mechanical and electronic assemblies can be replaced. Refer to the
Replacement List in the PRODUCT DESCRIPTION SECTION which lists replaceable assem-
blies as well as consumable items. Consult the list to determine the appropriate replacement
item and part number.
If repair to any electrical assembly is necessary, consult the Schematics in the APPENDIX SEC-
TION for further detail. Service personnel are encouraged to contact Medica directly with any
problems or concerns relating to electrical assemblies.
34 Repair
Recommended Tools
The following is a list of the tools required to perform repairs on the EasyLyte analyzer. Please
refer to this list to determine the appropriate tool and size needed.
Phillips Screw Driver, #1
Phillips Screw Driver, #2
Allen Wrench, 1/16 inch
Scissors
Hex Nut Driver, 11/32 inch, hollow stemmed
Retaining Ring Pliers
Resistor 10 OHM, 5 watts
Volt Meter
Optional Tools
The following is a list of tools recommended to help perform repairs on the EasyLyte analyzer.
Please refer to this list to determine the tool and part number when ordering.
Alignment Valve 000104
Mini-Test High Impedance Voltage Source 001374
Repair 35
Screws
Screws
Entry Module
Power Cord
Rear Housing Removal
1. Disconnect the power cord from the entry module, see Figure R1.
2. Detach the four screws from the rear housing using a #2 Phillips screw driver, as shown in
Figure R2.
Figure R1
Figure R2
36 Repair
Figure R3
3. Separate the front and back housing. See Figure R3.
Rear Housing Front Housing
Repair 37
Figure R4
Rear Housing Installation
1. Align and attach the front and back housing. See Figure R4. Verify that all cable
connectors are attached properly. Refer to Figures P1 and P6 in the PRODUCT
DESCRIPTION SECTION for detail. Verify that all wires are contained within the
front housing before proceeding.
Rear Housing Front Housing
38 Repair
Figure R6
Figure R5
Entry Module
Power Cord
Screws
Screws
2. Attach the four screws to the rear housing using a #2 Phillips screw driver, as shown in
Figure R5.
3. Connect the power cord to the entry module, see Figure R6.
Repair 39
Power Supply
Cable
Connector to
CPU Board
Power Supply
Cable
Connector to
CPU Board
Screws
Figure R9
Figure R8
Figure R7
Power Supply Removal
1. Disconnect the power supply cable from the CPU board, see Figure R7. Refer to
Figure R8 for EasyLyte analyzers with serial numbers below 6100.
2. Detach the four screws from the bottom of the front housing, using a #2 Phillips screw driver.
See Figure R9.
6100
40 Repair
Power Supply
Assembly
Power Supply
Assembly
Figure R11
Figure R10
3. Lift and remove the power supply assembly from under the CPU board, as shown in
Figure R10. Refer to Figure R11 for EasyLyte analyzers with serial numbers below 6100.
6100
Repair 41
Power Supply
Assembly
Figure R12
Power Supply
Assembly
Screws
Figure R14
Figure R13
Power Supply Installation
1. Slide the power supply assembly under the CPU board, as shown in Figure R12. Fit into the
alignment holes in the bottom of the front housing. Refer to Figure R13 for EasyLyte
analyzers with serial numbers below 6100.
2. Attach the four screws through the bottom of the front housing using a #2 Phillips screw
driver. See Figure R14.
6100
42 Repair
Power Supply
Cable
Connector to
CPU Board
Figure R16
Power Supply
Cable
Connector to
CPU Board
Figure R15
3. Connect the power supply cable to the CPU board, as shown in Figure R15. Refer to
Figure R16 for EasyLyte analyzers with serial numbers below 6100.
6100
Repair 43
Printer Bay
Paper Roll
Printer Lid
Printer
Figure R17
Figure R18
Printer Removal
1. Remove the printer lid from the printer, see Figure R17.
2. Remove the paper roll from the printer bay, as shown in Figure R18.
44 Repair
Figure R20
Printer Cable
Connector to
CPU Board
Printer Cable
Connector to
CPU Board
Figure R19
3. Disconnect the printer cable from the CPU board, see Figure R19. Refer to Figure R20 for
EasyLyte analyzers with serial numbers below 6100.
6100
Repair 45
Screws
Retaining
Clamps
Printer Cable
Figure R22
Figure R21
4. Detach the printer cable from the retaining clamps located along the right side of the CPU
board, as shown in Figure R21.
5. Detach the two screws using a # 2 Phillips screw driver. See Figure R22.
46 Repair
CPU Board
Printer Cable
Printer Body
Printer Body
Figure R24
Figure R23
6. Remove the printer body as shown in Figure R23.
7. Pull the printer cable up from behind the CPU board, while holding the printer body as shown
in Figure R24.
Repair 47
Printer Body
Screws
CPU Board
Printer Cable
Printer Body
Figure R27
Figure R26
Figure R25
Printer Installation
1. Slide the printer cable behind the CPU board, as shown in Figure R25.
2. Install the printer body into place. See Figure R26.
3. Attach the two screws using a #2 Phillips screw driver. See Figure R27.
48 Repair
Figure R30
Figure R28
Printer Cable
Connector to
CPU Board
Printer Cable
Connector to
CPU Board
Figure R29
Paper Feed Button
Printer Bay
Paper Roll
4. Connect the printer cable to the CPU board, see Figure R28. Refer to Figure R29 for
EasyLyte analyzers with serial numbers below 6100.
5. Using scissors, cut the beginning of the paper roll into a “V” shape. See Figure R30. Insert
the paper into the feeder slot inside the printer bay, and press the paper feed button.
6. Place the printer roll into the printer bay, as shown in Figure R31.
Figure 31
6100
Repair 49
Printer Cable
Retaining Clamps
Printer
Printer Lid
Figure R33
Figure R32
7. Place the printer lid onto the printer, see Figure R32.
8. Attach the printer cable to the retaining clamps located along the right side of the CPU
board, as shown in Figure R33.
50 Repair
Pump Motor Cable
Sensor Cable
Probe Motor Cable
Power Supply
Cable
Printer Cable
Sensor Cable
Pump Motor Cable
Probe Motor Cable
CPU Board
Connector to
CPU Board
Display Cable
Power Supply
Cable
Printer Cable
Figure R36
Figure R35
Figure R34
CPU Board Removal
1. Disconnect the probe motor, sensor, pump motor, power supply and printer cables from the
CPU board. See Figure R34. Refer to Figure R35 for EasyLyte analyzers with serial num-
bers below 6100.
2. Remove the display cable from the top of the CPU board as shown in Figure R36.
6100
Repair 51
CPU Board
Display Cable Tabs
Display Cable
Figure R37
Electrode Jacks
Connector Pin
to CPU Board
Figure R38
3. For EasyLyte Analyzers with serial numbers below 6100, push the display cable tabs to the
side to unlock. Remove the cable from the connector. See Figure R37.
4. Disconnect the electrode jacks from behind the CPU board. See Figure R38 for the location
of the electrode jacks. To disconnect, gently pull electrode jack straight away from the con-
nector pin of the CPU Board.
The connector pins are extremely delicate. To avoid pin breakage when removing the
electrode jacks, gently pull the electrode jacks straight away from the pins. If a pin breaks
the entire CPU board may require replacement.
6100
52 Repair
Screw
Figure R40
5. Disengage the CPU board from the four plastic mounting posts, located at each corner of the
CPU board. To disengage, press the mounting post tab in and pull the CPU board over the
tab, as shown in Figure R39.
6. Disengage the center plastic mounting post from the CPU board, located above the serial
and EasySampler ports.
7. For EasyLyte analyzers with serial numbers below 6100, detach the screw from the center of
the board using a # 2 Phillips screw driver. See Figure R40.
6100
Pull Board
Over Tab
Plastic
Mounting
Post
Press
In Tab
CPU
Board
Plastic
Mounting Post
Tab
Figure R39
Repair 53
Mounting
Posts
Mounting
Posts
Figure R41
7. Slide the CPU board off the mounting posts. See Figure R41.
54 Repair
Mounting Posts
Figure R42
CPU Board Installation
1. Align the CPU board with the plastic mounting posts. Snap the right side mounting posts of
the CPU board into place. See Figure R42.
Do not snap the left or center mounting posts into place at this time.
Repair 55
Figure R45
Screw
Figure R44
R42
P11
P11
P8
P8
P9 P9
P10
P10
(Reference) Connect 1st
(Cl-/Li+) Connect 4th
(Na+) Connect 3rd
(K+) Connect 2nd
Figure R43
P11
AGND
P9
P8
P8
P10
P9
P10
(Reference)
Connect 1st
(Cl-/Li+) Connect 4th
(Na+) Connect 3rd
(K+) Connect 2nd
2. Connect the electrode jacks to the back of the CPU board. To connect, gently push the elec-
trode jack straight into the connector pin. See Figure R43 to determine the proper connector
pin position on the CPU board and the recommended installation order.
See Figure R44 for EasyLyte Analyzer with serial numbers below 6100. On some earlier
EasyLyte analyzers below 6100, the reference electrode jack is connected to R42 in place of P11.
Do not bend the connector pin when installing the electrode jack. If a pin breaks the
entire CPU board may require replacement.
3. Snap the left and center mounting posts of the CPU board into place. For EasyLyte analyz-
ers with serial numbers below 6100, attach the center screw to the CPU board using a
#2 Phillips screw driver. See Figure R45.
6100
6100
56 Repair
Figure R46
Probe Motor Cable
(4 Pin Connector)
Sensor Cable
(5 Pin Connector)
Pump Motor Cable
(6 Pin Connector)
Pump Motor Cable
(6 Pin Connector)
Probe Motor Cable
(4 Pin Connector)
Sensor Cable
(5 Pin Connector)
Display Cable
Display Cable Tabs
Power Supply
Cable
Printer Cable
Power Supply
Cable
Printer Cable
Figure R48
Figure R47
4. Attach the display cable to the CPU board. Push the display cable tabs to the front to secure
cable in place. See Figure R46.
5. Connect the probe motor, sensor, pump motor, power supply and printer cables to the CPU
board, see Figure R47. Refer to Figure R48 for EasyLyte analyzers with serial numbers below
6100.
6100
Repair 57
Sample Detector
Board
Screws
Figure R49
Display Cable
Screws
Figure R50
Display Removal
1. Detach the two short self tapping screws from the sample detector board using a #2 Phillips
screw driver. See Figure R49.
2. Move and hold the display cable to expose the three display assembly screws. Detach the
three short screws from the display assembly using a #2 Phillips screw driver. See
Figure R50.
58 Repair
Display
Adapter
Display
Assembly LCD Screen
Figure R51
3. Carefully remove the display assembly and the LCD screen, holding both items together.
See Figure R51.
Notice the orientation of the LCD screen which is extremely important when installing the
display assembly. The one-line display has soldered leads located on the top of the LCD
screen; the two-line display has soldered leads located on the bottom of the LCD screen.
Refer to Figure R52 in the Display Installation for further details.
4. Remove the display adapter from the front housing window. Refer to Figure R51 for the loca-
tion of the display adapter.
For two-line displays there is an orientation notch which is important when installing the
display adapter. The orientation notch is located on the top inside edge of the adapter.
Repair 59
LCD Screen
Display Installation
1. Place the display adapter into the front housing window. The smooth edge should be facing
the front housing of the analyzer.
For two-line displays there is an additional orientation notch. To install correctly, arrange
the adapter so that the notch is located on the top inside edge of the adapter.
2. Place the foam pad on the display board above the YES/NO buttons.
3. Place the LCD screen on top of the foam pad. The orientation of the LCD screen is extreme-
ly important, see Figure R52.
The one-line display has soldered leads located on the top of the LCD screen, the two-
line display has soldered leads located on the bottom of the LCD screen.
Soldered Leads
Buttons Foam Pad
Figure R52 (One-line Display)
60 Repair
Figure R54
Screws
Display
Assembly
4. Position the front housing so that the front of the analyzer is facing down. Align the display
assembly and insert into place, as shown in Figure R53, using the YES/NO buttons as a
guide.
5. Hold the display in place and attach the three screws using a #2 Phillips screw driver. See
Figure R54.
Figure R53
Repair 61
Sample
Detector
Board
Sample
Detector
Cable
Screws
Figure R55
6. Align the sample detector board so the sample detector cable is positioned on the right side.
Attach the two screws lightly, using a #2 Phillips screw driver, see Figure R55.
7. Install the sample detector in the front of the analyzer to verify the sample detector board
alignment. See Figure R56. When properly aligned fully tighten the sample detector screws.
Sample Detector
Figure R56
62 Repair
Screws
Screws
Pump Motor
Assembly
Pump Motor Removal
1. Remove the pump tubing from the front of the analyzer.
2. Using a wire cutter, cut the tie wrap that secures the pump motor cable wires, as shown in
Figure R57.
3. Detach the four screws from the front of the analyzer, using a #1 Phillips screw driver. See
Figure R58. Some earlier EasyLyte analyzers may use socket screws that require a 1/16
inch Allen wrench for removal.
4. Remove the pump motor assembly, as shown in Figure R59.
Pump Motor
Assembly
Tie Wrap
Cable
Figure R57
Figure R59
Figure R58
Repair 63
Screws
Screws
Pump Motor
Wires
Pump Motor
Assembly
Figure R60
Figure R61
Pump Motor Installation
1. Install the pump motor assembly through the front housing, as shown in Figure R60. The
pump motor wires are positioned below the pump when properly aligned.
2. Attach the four screws into the front of the analyzer, using a #2 Phillips screw driver. See
Figure R61. Earlier EasyLyte analyzers may use socket screws that require a 1/16 inch Allen
wrench for installation.
3. Using a tie wrap cable, secure the loose wires under the pump motor to the lead screw/probe
motor wires and the sensor wires.
4. Install the pump tubing.
64 Repair
Probe Position
Scale
Screw
Probe Position
Scale
Figure R63
Figure R62
Probe Motor Assembly Removal
1. Detach the screw at the top of the probe position scale, using a #2 Phillips screw driver, as
shown in Figure R62. Remove the washer. Some EasyLyte analyzers use a white spacer
that should be removed also.
2. Remove the probe position scale, as shown in Figure R63.
Repair 65
Top of
Lead Screw
Retaining Ring
Figure R64
Actuator
Tie Wrap Cable
Retaining
Clamp
Probe Motor
Cable Wires
Figure R65
Figure R66
3. Remove the C-shaped retaining ring from the actuator, using the retaining ring pliers. DO
NOT stretch out the retaining ring with the pliers. Disengage the actuator from the top of the
lead screw assembly. See Figure R64.
Save the retaining ring.
4. Cut the tie wrap using a wire cutter, see Figure R65. Remove the probe motor cable wires
from the retaining clamp. See Figure R66.
66 Repair
Probe Motor/
Lead Screw
Assembly
Probe Motor
Bracket
Figure R67
Figure R68
5. Detach the three screws from the probe motor bracket, using a #2 Phillips screw driver, as
shown in Figure R67. Remove the washers.
The other two screws are located on the bottom edge of the bracket.
6. Remove the Probe Motor Assembly, as shown in Figure R68.
Repair 67
Figure R69
Screw
Probe Motor
Assembly
Probe Motor
Bracket
Figure R70
Probe Motor Assembly Installation
1. Place the probe motor assembly into place. See Figure R69.
2. Attach the three washers and screws to the probe motor assembly bracket, using a #2
Phillips screw driver. See Figure R70.
The other two screws are located on the bottom edge of the bracket.
68 Repair
Lead Screw
Assembly
Retaining Ring
Actuator
Tie Wrap
Cable
Location
Retaining
Clamp
Probe Motor
Cable
Wires
Figure R72
Figure R71
3. Insert the actuator through the top of the lead screw assembly as shown in Figure R71.
Install the retaining ring onto the actuator using the retaining ring pliers. DO NOT stretch out
the retaining ring with the pliers.
4. Place the probe motor cable wires into the retaining clamp. See Figure R72. Secure the
wires to the sensor cable wires with a tie wrap cable. The optimum location for the tie wrap
cable is slightly to the left of the retaining clamp.
5. Install the alignment valve tool, part number 000104, in place of the solutions valve.
Repair 69
6. Install the sample probe as follows. Refer to Figure R73 for details.
A. With the probe hole facing out, align the probe collar with the white line on the top of the
probe arm. When aligned, snap the probe into the probe arm.
B. Manually move the probe arm until the sample probe hole is aligned with the top line
marked on the alignment valve.
When the sample probe is aligned, DO NOT move the probe arm until the Probe Motor
installation is complete.
A
B
Probe Collar
aligned with
Slot
Probe Hole facing
out and aligned with
top line
Sample
Probe
Figure R73
Probe Arm
Alignment
Valve
70 Repair
Bottom of
Scale
Probe Position
Alignment Slot
Probe Position
Scale
Figure R74
7. Hold the sample probe against the alignment valve tool to keep in place. Install the probe
position scale through the top of the housing, see Figure R74. Verify the probe position
alignment slot is located on the bottom right side of the scale.
Repair 71
Note:
Alignment of
Scale and
Sensor
Figure R75
Probe Position
Screw – Keep
loose until
scale in align-
ment
8. Place the scale into position. The scale should be seated inside the probe position sensor
and inside the mounting bracket slot. See Figure R75.
9. Attach and slightly tighten the washer and screw at the top of the position scale. Do not fully
tighten the screw. Some EasyLyte analyzers use a white spacer that should be attached to
the back surface of the scale.
10. Align the probe position scale slot within the center of the probe position sensor. Gently
move the scale to the left and visually confirm the alignment. Refer to Figure R75 for cor-
rect alignment.
Confirm that the sample probe hole is aligned with the top line marked on the alignment
valve, on the front of the analyzer, before continuing.
Probe Position
Sensor
Probe Scale
Alignment Slot
Probe Position
Scale
Mounting Bracket
CPU Board
72 Repair
Screw
Figure R76
11. When alignment is complete, fully tighten the probe position screw into place using a #2
Phillips screw driver, as shown in Figure R76.
Repair 73
Figure R78
Guide Rod
Screw
Solutions
Valve
Sample
Probe
Probe Arm
Figure R77
Guide Rod Removal
1. Remove the sample probe from the probe arm and the solutions valve. See Figure R77.
Remove the solutions valve.
2. Detach the screw from the bottom of the guide rod, using a #2 Phillips screw driver, as
shown in Figure R78. Remove the washer.
74 Repair
Probe Arm
Guide Rod
Figure R79
3. Remove the guide rod and probe arm assembly, down and away from the housing.
See Figure R79.
Repair 75
Screw
Guide Rod
Probe Arm
Alignment Hole
Guide Rod
Figure R81
Guide Rod Installation
1. Install the guide rod as shown in Figure R80. To install, first insert the guide rod into the
alignment hole located in the top wall of the front assembly.
2. Attach the washer and screw to the bottom of the guide rod, using a #2 Phillips screw
driver. See Figure R81.
Figure R80
76 Repair
Electrode Jack
Analyzer Front
Housing
Electrode Jack
Wires
Nut Driver
Figure R84
Figure R83
Figure R82
Electrode Jack Removal
1. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R82.
2. Remove the hex nuts using the nut driver. See Figure R83.
3. Pull the electrode jacks from the front housing of the analyzer, as shown in Figure R84.
Repair 77
Electrode Jack
Wires
Nut Driver
Electrode Jack
Holes
Electrode Jack
Analyzer Front
Housing
Figure R86
Figure R85
Figure R87
Electrode Jack Installation
1. Push the electrode jacks through the holes in the front of the analyzer, see Figure R85.
2. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R86.
3. Attach the hex nut using the nut driver, and tighten, as shown in Figure R87.
EasySampler 79
Product Description
The EasySampler when used with the EasyLyte analyzer provides automated batch sampling of
serum, plasma or diluted urine for electrolyte analysis. The EasySampler can also perform emer-
gency (STAT) analysis of blood samples on demand.
The EasySampler contains a 25 position sample tray which includes a STAT and two quality con-
trol positions. The tray holds either 500 μl or 2 ml sample cups.
The EasyLyte analyzer software allows the operator to perform EasySampler diagnostics by sim-
ply pressing the “YES” or “NO” buttons in response to prompts displayed on the analyzer LCD
screen. Results are always displayed on the LCD screen and can also be documented on paper
by the EasyLyte analyzer printer.
80 EasySampler
Figure A1 - EasySampler Assemblies and Components
Assembly and Component Overview
The EasySampler is made up of the following mechanical assemblies and associated compo-
nents. Refer to Figure A1 to determine the location of each assembly or component in relation to
the EasySampler.
EasySampler Housing case of the EasySampler
IR Windows covers the IR beam transmitter and receiver
Chassis Assembly frame which supports the mechanical assemblies
PC Board Assembly contains the PC board and IR transmitter and receiver
Motor Assembly powers the spindle column assembly
Drive Belt connects the motor assembly to the spindle column assembly
Spindle Column Assembly rotates the sample tray
Sensor Assembly detects the sample tray position
Sample Tray holds the samples for analysis
Sample Tray
EasySampler
Housing
Spindle Column
Assembly
Drive Belt
Chassis
Assembly
IR Transmitter
Motor
Assembly
Sensor
Assembly
PC Board
Assembly
IR Windows
IR Receiver
Replacement List
The following assemblies and components can be purchased directly from Medica.
Description Part No.
EasySampler Housing 001074
IR Windows 001192
Chassis Assembly 001370
PC Board Assembly 001233
Motor Assembly 001328
Spindle Column Assembly 001680
Sensor Assembly 001306
Drive Belt 001197
EasySampler 81
EasySampler 81
82 EasySampler
EasySampler Troubleshooting Guide
Refer to the chart below to determine the problem and corrective action needed.
Problem Description Problem Source Corrective Action Steps
LCD screen displays SAFETY BEAM • Verify there is no IR obstruction
“OBSTRUCTION” and OBSTRUCTION • Sensor Verification
analyzer is beeping
LCD screen displays “MTS” INCORRECT PROBE or • Alignment Verification
“TURN TRAY HOME” or TRAY MOVEMENT • Tray Placement Verification
“MTC” • Carousel Test
• Tray Cycle Verification
• Sensor Verification
EasySampler 83
Alignment Verification
1. Remove and reinstall the power cord. The sample tray will rotate and then stop.
2. ALIGN PROBE/CUP? appears on the LCD screen. Press YES to verify the probe
alignment. The sample probe will descend and stop above the “Daily Cleaner” cup position.
3. ALIGN CUP!/CUP ALIGNED? appears on the LCD screen; do not press YES or NO at this
time.
4. Verify that the tip of the sample probe is centered over the “Daily Cleaner” cup position.
Based on the inspection results, take the appropriate corrective action.
• Probe Aligned: If sample probe is aligned in the proper position, press YES to the
ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on
the LCD screen; press YES. CALIBRATE NOW? appears on the LCD
screen; press YES. Return to normal operation.
• Probe Misaligned: If the probe is misaligned, proceed to STEP 5.
5. Grasp the EasySampler and analyzer at the base and carefully slide forward until the
EasySampler extends over the edge of the work surface.
6. Turn the knob underneath the EasySampler base clockwise to loosen. DO NOT remove the
knob.
7. Locate the hub on top of the EasySampler tray. Move the hub and the knob together, until
the sample probe aligns with the “Daily Cleaner” cup position.
8. When aligned, hold the hub in place and turn the knob counterclockwise to tighten.
9. Grasp the EasySampler and analyzer at the base and carefully slide back to the original work
area.
10. Press YES to the ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on the
LCD screen; press YES. CALIBRATE NOW? appears on the LCD screen; press YES.
84 EasySampler
Tray Placement Verification
1. Advance to the DAILY CLEANER? display; press YES. TRAY IN PLACE? appears on the
LCD screen; press YES.
2. The sample tray should rotate and stop with the probe at cup position “12”.
3. CLEANER IN TRAY? appears on the LCD screen; press NO.
4. The sample tray should rotate and then stop with the probe at “Daily Cleaner” position.
5. Based on the test results, take the appropriate corrective action.
• Correct Position: If the tray stops at the correct position, return to normal operation.
• Incorrect Position: If the tray stops at an incorrect position, perform the Carousel Test.
EasySampler 85
Carousel Test
1. Enter the QC TEST? menu and proceed to TEST CAROUSEL?. Refer to QC Test in this
section for further instructions.
2. At the TEST CAROUSEL? display, press YES.
3. CALIB CAROUSEL? appears on the LCD screen; press YES. The sample tray should spin
one revolution; then stop.
4. FIND HOME? appears on the LCD screen; press YES. The sample tray should spin and
stop with the probe at the “Daily Cleaner” position.
5. Based on the CALIB CAROUSEL? test results, take the appropriate corrective action.
• Error Message: If “MTC” or “CAROUSEL PROBLEM” appears on the LCD screen
perform Sensor Verification.
• No Error Message: If no error message appears on the LCD screen, perform
Tray Cycle Verification.
86 EasySampler
Tray Cycle Verification
1. Enter the QC TEST? menu and proceed to the TEST CAROUSEL? test. Refer to QC Test
in this section for further instructions.
2. At the TEST CAROUSEL? display, press YES. At the CALIB CAROUSEL? display, press
NO. At the FIND HOME? display, press YES. The tray should spin and stop with the sample
probe at the “Daily Cleaner” position.
3. CYCLE TRAY? appears on the LCD screen; press YES.
4. The sample tray should rotate from the “Daily Cleaner” position to each cup position and
back for cup positions “NORMAL through 10”. Then the sample tray should reverse direction
and rotate from the “Daily Cleaner” position to each cup position and back for cup positions
“11 through STAT”.
5. The sample tray will perform four complete rotation cycles as described above.
6. The analyzer will print the test results during the fourth cycle, as illustrated in the
example below.
Cup Number of Steps Number of Steps
Position to Cup Position to Daily Cleaner
1 0097 0100
2 0196 0196
3 0292 0292
4 0388 0388
7. Verify that the number of steps between adjacent cup positions is 0096 +/-0006 for both
columns of test data.
8. Based on the tray cycle test results, take the appropriate corrective action.
• Error Message: If “MTC” or “CAROUSEL PROBLEM” appears on the LCD screen during
testing, perform Sensor Verification.
• In Specification: If tray cycle readings are in specification, exit the QC TEST? menu and
return to normal operation.
EasySampler 87
Sensor Verification
1. Remove the rear housing of the EasyLyte analyzer as detailed in Rear Housing Assembly
Removal in the REPAIR SECTION.
2. Connect a voltage meter to the EasyLyte analyzer. Attach one reference lead to TP11
(DGND) and the other to U13 PIN #2 on the analyzer CPU board.
3. Manually turn the sample tray so the probe is at the “Daily Cleaner” position.
4. Verify that the measured voltage is greater than 4.50 volts.
5. Manually turn the sample tray off the “Daily Cleaner” position.
6. Verify that the measured voltage is less than 0.50 volts.
7. Based on the voltage test results, take the appropriate corrective action.
• In Specification: If voltage readings are in specification, adjust the drive belt tension and
repeat Tray Cycle Verification. If “MTC” or “CAROUSEL PROBLEM”
errors continue, replace the spindle column assembly. Refer to Spindle
Column Removal and Installation in this section for details.
• Out of Specification: If voltage readings are out of specification, replace the sensor assem-
bly. Refer to Sensor Removal and Installation in this section for
details.
88 EasySampler
QC Tests
1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis-
play, as shown on the chart below.
LCD Screen Displays Press
CALIBRATE NOW? NO
DAILY CLEANER? NO
STANDBY MODE? NO
SECOND MENU? YES
* SEE RESULTS? * NO
* PACK USAGE? * NO
* QUAL CONTROLS? * NO
* OPER FUNCTS? * NO
* DIAGNOSTICS? * NO
* DATE/RANGES? * YES
* CHANGE DAY/TIME? * NO
CAL FREQUENCY? NO
Na/K/Cl or Na/K/Li RANGES? YES
BLOOD Na LIMITS? YES
“Numeric value” —> “Numeric value” wait
2. Two numbers appear on the screen separated by an arrow. A cursor will be positioned
underneath the first digit of the first number.
The second number is not applicable in this procedure.
3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the
digit is correct press YES to advance to the next digit.
4. Continue until the first number is changed to 246.0.
5. Press YES when finished. QC TEST? will appear on the screen. If a value other than “246”
is input the screen displays CORRECT?. Press NO to return to STEP 2 and correct any
error.
6. Press YES to QC TEST?
EasySampler 89
7. Consult the table below which lists all the QC tests and their functions. To select a test press
YES. To bypass a test press NO.
QC Test QC Test Function
QC TEST? Access to the QC Test Functions.
TURN ON MV’S?
START CLOCK? Analyzer QC Test Functions.
Refer to the main
CHECK CAL CONSTS? TROUBLESHOOTING SECTION
RAM/PROM TEST? for information on QC testing
SIPPER TEST? the EasyLyte analyzer.
PUMP TEST?
TEST CAROUSEL? Verifies EasySampler movement functions.
CALIB CAROUSEL? Moves sample tray one revolution.
FIND HOME? Returns sample tray to “Daily Cleaner” position.
CYCLE TRAY? Performs four complete tray cycles to each cup position.
SOL’N PURGE?
CALIBRATE NOW?
DIAGNOSTICS?
LAST CAL VALUES?
ELECTR’D VALS? Analyzer QC Test Functions.
FLUID FLOW TEST?
SAMPLE DET. TEST? Refer to the main
PUMP CAL TEST? TROUBLESHOOTING SECTION
TEST VALVE? for information on QC testing
RUN STATISTICS? the EasyLyte analyzer.
RUN CONTROLS?
TEST ALIGNMENT?
BURN IN?
CONTINUOUS?
QC TEST? Access to the QC Test Functions. Press NO to return to
the main EasyLyte menu.
6100
90 EasySampler
Repair Overview
Always consult Troubleshooting in this section first to determine the problem before attempting to
service the EasySampler.
Please note that complete mechanical assemblies can be replaced. Refer to the Replacement
List in this section which lists replaceable assemblies and appropriate catalog numbers.
Recommended Tools
The following is a list of the tools required to perform repairs on the EasySampler. Please refer
to this list to determine the appropriate tool and size needed.
Phillips Screw Driver, #1
Phillips Screw Driver, #2
Flat Head Screw Driver, 7/32 inch
Allen Wrench, 1/16 inch
Open End Wrench, 1/4 inch
Volt Meter
EasySampler 91
Housing Removal
1. Turn the EasySampler upside down and gently place on a flat surface.
2. Remove the two screws from the cable retaining plate using a #2 Phillips screw driver, as
shown in Figure A2.
Figure A2
Screws
Cable Retaining
Plate
92 EasySampler
Chasis Assembly
Housing
Screws
Metal Base
Figure A4
Figure A3
3. Remove the EasySampler cable from the groove in the bottom of the housing.
4. Remove the three screws from the metal base using a #2 Phillips screw driver, as shown in
Figure A3.
5. Turn the EasySampler to the upright position.
6. Lift the housing, as shown in Figure A4, to remove from the chassis assembly.
EasySampler 93
Chasis
Assembly
Figure A5
Screws
Metal Base
Housing
Figure A6
Housing Installation
1. Align and place the housing over the chassis assembly, as shown in Figure A5.
Be careful not to damage the IR receiver or transmitter on the PCB.
2. Turn the EasySampler upside down and gently place on a flat surface.
3. Attach the three screws to the metal base using a #2 Phillips screw driver, see Figure A6.
94 EasySampler
4. Thread the EasySampler cable into the groove in the bottom of the housing. See Figure A7.
5. Attach the two screws to the cable retaining plate using a #2 Phillips screw driver, as shown
in Figure A7.
Screws
EasySampler
Cable
Cable Retaining
Plate
Figure A7
EasySampler 95
PC Board
Chassis
Assembly
Adjustment
Knob
Motor Cable
Sensor Cable
Figure A9
Figure A8
Spindle Column Removal
1. Disconnect the sensor cable and the motor cables from the PC Board. See Figure A8.
2. Turn the adjustment knob counterclockwise to loosen. Remove the adjustment knob from
the bottom of the chassis assembly, as shown in Figure A9.
Some EasySamplers use a spacing disk that should be removed and retained.
96 EasySampler
Spindle Column
Assembly
Figure A10
3. Remove the spindle column assembly, as shown in Figure A10.
EasySampler 97
Adjustment Knob
Spindle Column
Assembly
Figure A12
Figure A11
Spindle Column Installation
1. Place the spindle column assembly down into place, as shown in Figure A11.
2. Attach the adjustment knob, see Figure A12, and turn clockwise to tighten.
Some EasySamplers use a spacing disk that should be attached before the knob is
installed.
98 EasySampler
Motor Connector (J1)
Sensor Connector
(J3)
Sensor
Cable
Motor
Cable
Figure A13
3. Attach the sensor cable and the motor cable to positions J1 and J3 on the PC board, see
Figure A13.
EasySampler 99
Motor
Bracket
Sensor
and Mounting
Block
Screws
Figure A15
Figure A14
Sensor Removal
1. Loosen the two slotted screws from the motor bracket, using a #2 Phillips screw driver, as
shown in Figure A14.
2. Remove the sensor and mounting block, as shown in Figure A15.
3. Remove the two slotted screws from the mounting block, using a #2 Phillips screw driver.
4. Remove the sensor from the mounting block.
100 EasySampler
Screws
Motor
Bracket
Sensor
and Mounting
Block
Figure A17
Figure A16
Sensor Installation
1. Place the sensor into the mounting block.
2. Attach the two slotted screws to the mounting block, using a #2 Phillips screw driver.
3. Place the sensor and mounting block into the motor bracket. See Figure A16.
4. Attach the two slotted screws to the motor bracket, using a #2 Phillips screw driver, as shown
in Figure A17.
EasySampler 101
Motor
Shaft
Damper
Damper
Socket Screw
Figure A19
Figure A18
Motor Removal
1. Loosen the socket screw from the brass damper using a 1/16 inch Allen wrench. See Figure
A18.
2. Remove the damper from the top of the motor shaft, as shown in Figure A19.
102 EasySampler
Motor Shelf
Motor
Slotted
Screws
Motor
Figure A21
Figure A20
3. Remove the four slotted screws from the top of the motor using a 7/32” flat head screw driv-
er. See Figure A20.
4. Remove the motor from the motor shelf, See Figure A21.
EasySampler 103
Belt
Motor
Slotted
Screws
Motor Gear
Damper
Motor
Shaft
Figure A23
Figure A22
Motor Installation
1. Place the motor into the motor shelf.
2. Pull the belt around the gear of the motor, see Figure A22 for location of gear and belt.
3. Attach the four slotted screws to the top of the motor using a 7/32” flat head screw driver, as
shown in Figure A22.
4. Attach the damper to the top of the motor shaft, as shown in Figure A23.
104 EasySampler
Socket Screw
Damper
Figure A24
5. Tighten the socket screw to the damper using a 1/16 inch Allen wrench. See Figure A24.
EasySampler 105
Heat Sink
Plate
Hex Nut
Power Cable
Figure A26
Figure A25
PC Board Removal
1. Disconnect the power cable from the PC board. See Figure A25.
2. Remove the hex nut from the heat-sink plate using a 1/4 inch open end wrench.
See Figure A26.
106 EasySampler
PC
Board
Screws
Screws
Base
Housing
PC
Board
Figure A28
Figure A27
3. Remove the four screws from the PC board using a #2 Phillips screw driver.
See Figure A27.
The lower left screw may be hidden by the EMI suppressor box. If so, peel back the tape
to reveal the screw.
4. Lift the PC Board, as shown in Figure A28, to remove from the base housing.
EasySampler 107
Screws
Screws
Figure A30
Figure A29
PC Board Installation
1. Place the PC Board, as shown in Figure A29, onto the base housing.
2. Install the four screws to the CPU board using a #2 Phillips screw driver,
see Figure A30.
108 EasySampler
PC Board
Power
Connector
(J2)
Heat Sink
Plate
Hex Nut
Power
Cable
Figure A32
Figure A31
3. Install the hex nut to the heat-sink plate using a 1/4 inch open end wrench,
see Figure A31.
4. Connect the power cable to position J2 on the PC board, see Figure A32.
Analyzer Specifications
CLIA Classification: Moderate Complexity
Sample: Whole Blood, Serum, Plasma or Urine.
Direct (undiluted) measurement Ion Selective Electrode
Sample Size: 100 μl Whole Blood, Serum, Plasma or 400 μL diluted (1:10) Urine
Detection Blood Na+ : 20-200 mmol/L Urine Na+ : 40-1000 mmol/L
Range: K+ : 0.2-40 mmol/L K+ : 1.0-5000 mmol/L
Cl- : 25-200 mmol/L Cl
- : 25-500 mmol/L
Li+ : 0.2-5.0 mmol/L Li+ :(N/A)
Resolution: Na+: 0.1 mmol/L K+: 0.01 mmol/L Li+: 0.01 mmol/L Cl-:0.1mmol/L
Performance: Na+ K+ Li+ Cl-
BLOOD, SERUM, PLASMA 80-200 mmol/L 1-10 mmol/L 0.2-2.0 mmol/L 80-200 mmol/L
Reproducibility (typical)
within run (n=20) C.V. <1% C.V. <2% SD≤.03 mmol/L C.V. <2%
between run (10 day) C.V. <2% C.V. <2.5% SD≤.05 mmol/L C.V. <2.5%
Accuracy (typical) Flame Photometer Amperometric
correlation coeff 0.99 1.00 0.99 0.97
slope 0.99 1.07 1.09 0.95
URINE
Reproducibility (typical)
within run (n=20) C.V. <5% C.V. <5% Not C.V. <5%
between run (10 day) C.V. <5% C.V. <5% Applicable C.V. <5%
Accuracy (typical)
Correlation coeff 1.00 1.00 Not 0.98
slope 1.05 1.00 Applicable 1.07
Analysis Time 60 sec (Blood) / 90 sec (Urine)
Data Storage: 125 Patient results
QC - up to 20 results Normal and Abnormal
Calibration: Automatic or On-Demand
Output: 32 character, 2- line alphanumeric display (or)
16 character, 1- line alphanumeric display
40 column thermal printer
RS-232 Serial port (optional cable)
EasySampler port
continued on next page
Specifications 109
Specifications 109
110 Specifications
Analyzer Specifications (continued)
Ambient Conditions: 15-32oC (60-90oF), <85% humidity
Power: 100/115~VAC, 50-60 Hz, 0.8A/250 V fuse or 220~VAC, 50-60 Hz 0.4A/250 V fuse
Refer to the label on the back of the analyzer for preset power and fuse
settings.
Size & Weight: 9.5"W x 16.5"H x 8.0"D, 13 lbs.
24 cm x 42 cm x 20 cm, 5.8 kg.
EasySampler Specifications
Size: Width: 12.2 inches (31.0 cm)
Depth:14.0 inches (35.6 cm)
Height: 8.8 inches (22.4 cm)
Weight: 7.0 lbs (3.2 kg)
Sample: Serum, Plasma, Urine (diluted)
Throughput: 1 sample per minute
Sample Cup Capacity: 500μl or 2ml
Sample Tray Positions: 25 positions
21 sample
1 normal QC
1 abnormal QC
1 stat
1 Daily Cleaner position
Power Source: EasyLyte analyzer
Appendix 111
Display Messages
“!!”
Na/K/Cl result(s) out of analyzer measurement range.
*****
Result not reported because electrode is not calibrated.
ABNORMAL CONTROL?
Allows operator to select analysis of abnormal control.
ABNORMAL STATS?
Allows operator to display or print statistics for stored abnormal control results.
AIR IN CLEANER
Cleaning solution not detected correctly.
AIR IN CONTROL
Control sample not detected correctly.
AIR IN SAMPLE
Sample not detected correctly.
AIR IN STD A
Standard A not detected correctly.
AIR IN STD B
Standard B not detected correctly.
AIR IN FLUSH
Flush not detected correctly.
ALL ABNORMAL?
Allows operator to delete all stored abnormal control results.
ALL NORMAL?
Allows operator to delete all stored normal control results.
ALL RESULTS?
Allows operator to display all stored results.
ANALYZE BLOOD?
Sample analysis mode for Blood, Serum or Plasma. The message will only appear
when the analyzer is calibrated.
*ANALYZE URINE?*
Urine analysis mode for Na/K/Cl results only.
ANALYZING...
Analyzer is measuring sample.
ASPIRATING...
Sample is being drawn into probe.
AUTOCAL?
Allows operator to activate and to set time for initial daily calibration.
BLOOD Cl LIMITS?
Allows operator to set normal Cl range.
112 Appendix
BLOOD K LIMITS?
Allows operator to set normal K range.
BLOOD Li LIMITS?
Allows operator to set normal Li range.
BLOOD NA LIMITS?
Allows operator to set normal Na range.
CAL VALUES?
Allows operator to display last calibration values.
CALIBRATE NOW?
Allows operator to calibrate analyzer.
CALIBRATING...
Analyzer is calibrating.
CALIBR’TING STD A
Electrodes being calibrated in relation to Standard A.
CALIBR’TNG STD B
Electrodes being calibrated in relation to Standard B.
CHANGE DAY/TIME?
Allows operator to change date and time.
CHANGE PACK!
Occurs at 99% pack usage. Pack must be replaced!
CLEANING...
Analyzer is performing daily cleaning.
Cl HIGH...
Chloride result obtained is higher than the preset normal range (flashing display).
Cl LOW...
Chloride result obtained is lower than the preset normal range (flashing display).
CONTINUE?
Analyzer requests confirmation to proceed.
C.V. VALUES?
Allows operator to display % C.V.’s of Quality Control results.
DAILY CLEANER?
Allows operator to perform daily cleaning.
DAILY SUMMARY?
Allows operator to print summary of daily results.
*DATE/RANGES?*
Allows operator to set date, time, and ranges.
DELETE CONTROLS?
Allows operator to delete stored control results from memory.
DELETE RESULTS?
Allows operator to delete patient sample results from memory.
Appendix 113
*DIAGNOSTICS?*
Allows operator to check key analyzer functions.
DILUTED 1:10?
Analyzer requests confirmation that urine sample has been diluted.
DRIFT Cl
Chloride electrode is drifting outside of preset electrode operating range.
DRIFT K
Potassium electrode is drifting outside of preset electrode operating range.
DRIFT Li
Lithium electrode is drifting outside of preset electrode operating range.
DRIFT NA
Sodium electrode is drifting outside of preset electrode operating range.
DRIFT, RECAL..
Analyzer is automatically recalibrating because of electrode drift.
ELECTR’D VALUES?
Allows operator to display real time electrode millivolt readings.
ENTIRE SUMMARY?
Allows operator to display summary of all results in storage.
***ERROR***
Denotes an error has occurred. Press YES or NO to determine error.
FLOW TEST FAIL
Unsuccessful Fluid Flow Test.
FLUID FLOW OK
Successful Fluid Flow Test.
FLUID FLOW TEST?
Allows operator to run red Test Dye to check fluid flow.
FLUID PATH!
Analyzer is not detecting fluid and/or air segments when expected.
FLUSH?
Allows operator to perform a Flush cycle on demand.
FLUSHING...
Analyzer is flushing electrodes.
FLUSH, RECAL ...
If calibration fails, analyzer does a Flush and recalibration automatically.
I.D. #- - - -
Allows operator to assign an identification number to patient result.
K HIGH...
Potassium result obtained is higher than preset normal range (flashing display).
K LOW...
Potassium result obtained is lower than preset normal range (flashing display).
LAST CONTROL?
Allows operator to display last control result.
114 Appendix
LAST RESULT?
Allows operator to display last analysis result.
Li BELOW 0.2mM/L
Lithium sample concentration is below EasyLyte measurement range.
Li HIGH..
Lithium result obtained is higher than the preset normal range (flashing display).
Li LOW...
Lithium result obtained is lower than the preset normal range (flashing display).
MEAN VALUES?
Allows operator to display mean values of quality control results.
MTP
Mechanical problem with pump.
MTS
Mechanical problem with probe.
MUST USE CLEANR!
Cleaning Solution must be used before operating the analyzer.
MV RANGE Cl
Measured Chloride electrode voltage is outside operating limits.
MV RANGE K
Measured Potassium electrode voltage is outside operating limits.
MV RANGE Li
Measured Lithium electrode voltage is outside operating limits.
MV RANGE Na
Measured Sodium electrode voltage is outside operating limits.
NA HIGH...
Sodium result obtained is higher than the preset normal range (flashing display).
NA LOW...
Sodium result obtained is lower than the preset normal range (flashing display).
**NA/K*** or ***Na/K/Cl*** or ***Na/K/Li***
Appears when power is connected or restored after interruption.
Na/K/Cl or Na/K/Li RANGE?
Allows operator to set normal ranges.
NEW PACK INSTLD?
Asks operator to confirm that a new pack has just been installed.
NEXT SAMPLE?
Allows operator to proceed to next analysis.
NOISE
Erratic signal coming from electrodes.
NOISE, RECAL..
Analyzer automatically recalibrates after detecting noise.
NORMAL CONTROL?
Allows operator to select analysis of normal control.
Appendix 115
NORMAL STATS?
Allows operator to display or print statistics for stored normal control results.
NOT ENOUGH DATA!
In order to calculate statistics, the Analyzer requires 5 stored control results of both normal
and abnormal levels.
*OPER FUNCTS?*
Allows operator to select “On Demand” flush, purge or calibration functions.
*PACK USAGE?*
Allows operator to display % pack expended.
PATIENT ID#?
Allows operator to assign identification number to patient’s result.
% OF PACK USED
Percentage of solutions pack that has been expended.
PRINTER OFF?
Allows operator to turn printer function on or off.
PRINTING...
Printer is printing requested data.
PRINT RESULTS?
Allows operator to print all results stored in memory.
PRINT STAT’TICS?
Allows operator to print normal or abnormal control summaries.
PRINT SUMMARY?
Prints daily summary of results.
PRINT WKLY SUM?
Prints summary of last five days results.
PROBE IN BLOOD?
Requests operator to confirm that probe is in blood.
PROBE IN CLEANR?
Requests operator to confirm that probe is in daily cleaning solution.
PROBE IN CONTROL?
Requests operator to confirm that probe is in control sample.
PROBE IN DYE?
Requests operator to confirm that probe is in Test Dye solution.
PROBE IN URINE?
Requests operator to confirm that probe is in urine.
PROBE IN FLUSH?
Requests operator to confirm that probe is in Flush Solution.
PROBE WIPING?
Allows operator time to manually wipe the probe after sample is removed.
PUMP CAL...
Pump is being calibrated.
PUMPCAL FAILURE!
Failure of Pumpcal Test.
116 Appendix
PUMPCAL OK
Successful completion of Pumpcal Test.
PUMPCAL TEST?
Allows operator to test pump calibration.
PURGING STD A
Standard A fluid line is being purged.
PURGING STD B
Standard B fluid line is being purged.
PURGING FLUSH
Flush Solution fluid line is being purged.
*QUAL CONTROL?
Allows operator to choose quality control functions.
REMOVE CONTROL
Requests operator to remove control sample from probe.
REMOVE SAMPLE
Requests operator to remove sample from probe.
RESET TO 0%?
Sets pack usage percentage counter to 0%.
RUN CONTROL?
Quality control analysis mode.
SAMPLE CAL...
Analyzer is calibrating the Sample Detector.
SAMPLE DETECT OK
Successful completion of Sample Detector Test.
SAMPLE DETECTOR!
Sample Detector readings out of range.
SAMPLE DET. TEST?
Allows operator to test sample detector.
S.D. VALUES?
Allows operator to display Standard Deviation values of Quality control results.
*SECOND MENU?*
Allows operator access to secondary menu.
SEE MANUAL!
See Operator’s Manual for assistance or explanation.
*SEE RESULTS?*
Allows operator to display or print stored results.
SEE STATISTICS?
Allows operator to display Mean, S.D., C.V. values of Quality Control results.
SET CORRELATION?
Allows operator to select dilution equivalent correlations.
SOL’N PACK LOW!
Appears when solutions pack is 88% expended. Continue operating analyzer.
Appendix 117
SOL’N PURGE?
Allows operator to purge air from fluid lines.
***STANDBY***
Appears when Analyzer is in Standby mode.
STANDBY IN 1 MIN
Analyzer will go into standby in 1 minute (Flashing Display).
STANDBY MODE?
Allows operator to put analyzer in standby mode to conserve fluids.
STATISTICS?
Allows operator to display statistics for all control results.
STORE RESULT?
Allows operator to store results in memory.
SYRINGE SAMPLE?
Allows operator to adjust probe extension during analysis to accommodate a syringe.
TESTING...
Analyzer in process of completing a Fluid Flow Test.
TRY PURGE AGAIN?
Allows operator to perform additional purge after initial failure.
URINE Cl LIMITS?
Allows operator to set normal Cl range for urine.
URINE K LIMITS?
Allows operator to set normal K range for urine.
URINE NA LIMITS?
Allows operator to set normal Na range for urine.
USE EXT. FLUSH!
Flush solution in pack is empty. Operator should use external flush.
USER OPTIONS
Allows operator to access the following functions: Probe wiping, Patient I.D. #,
Syringe Sample, Set Correlation, EasySampler and Printer Off.
WEEKLY SUMMARY?
Allows operator to print last five days of stored results.
WIPE PROBE
Allows operator time to wipe off sample from exterior of probe.
400 mL PACK?
Allows operator to select for correct solutions pack size.
800 mL PACK?
Allows operator to select for correct solutions pack size.
118 QC TEST? Flowchart (Service Diagnostics)
PUMPCAL TEST?
DATE/RANGES?
BLOOD Na LIMITS?
246.0
QC TEST?
TURN ON MV’S?
CHECK CAL
CONSTS?
RAM/PROM TEST?
SIPPER TEST?
PUMP TEST?
TEST CAROUSEL?
SOLUTION PURGE?
CALIBRATE NOW?
RUN CONTROLS?
DIAGNOSTICS?
TEST VALVE?
TEST ALIGNMENT?
BURN IN?
QC TEST?
CAL VALUES?
ELECTRODES VALUES?
FLUID FLOW TEST?
SAMPLE DET. TEST?
EASYSAMPLER
5 NORMAL AND 5 ABNORMAL CONTROLS
AT CUP #1 TO #10 IN SAMPLE TRAY
NORMAL CONTROL?
ABNORMAL CONTROL?
8 NORMAL AND
STATS
8 ABNORMAL AND
STATS
CALIB CAROUSEL?
FIND HOME?
CYCLE TRAY?
CHECK SUM
RELEASE DATE
RAM CHECKOUT? CLEAR MEMORY
QC TEST?
Yes/No
CPU Board – Analyzer Series below 6100
CPU Board
Power Supply – Analyzer Series under 6100
Transformer Assembly
EasySampler PCB
001462-001
R3
EDIC@
Medica Corporation 5 Oak Park Drive Bedford, MA 01730
For Service Or Technical Assistance Call: 1-800-777-5983 (In Continental US)
1-617-275-4892 (International)

More Related Content

Similar to Service Manual Easylite electrolite device

Aespire 6X Participant Guide (1).pdf
Aespire 6X Participant Guide (1).pdfAespire 6X Participant Guide (1).pdf
Aespire 6X Participant Guide (1).pdfJUANRAMONMONCADABACA
 
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdf
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdfJohn Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdf
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdfExcavator
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualudfjjjskekmdm
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualfusjefjjskekmem
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualjfjskekdhnejd
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualfujjsefksekmem
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualfjsjekkmdmme
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualfusjefjjskekmem
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualudfjjjskekmdm
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualfujjsefksekmem
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualjfjskekdhnejd
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualfjsjekkmdmme
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualfjsjekkmdmme
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualjfjskekdhnejd
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualfujjsefksekmem
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualudfjjjskekmdm
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualfusjefjjskekmem
 
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...ufuusjjfjkskekmem
 
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...ufjjdjjkskemme
 

Similar to Service Manual Easylite electrolite device (20)

Aespire 6X Participant Guide (1).pdf
Aespire 6X Participant Guide (1).pdfAespire 6X Participant Guide (1).pdf
Aespire 6X Participant Guide (1).pdf
 
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdf
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdfJohn Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdf
John Deere Tractor 6215, 6415, 6615, 6715 Diagnostic & Test Service Manual.pdf
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manual
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manual
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manual
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manual
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manual
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manual
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manual
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manual
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manual
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manual
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manual
 
John deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manualJohn deere 4700 compact utility tractor service repair manual
John deere 4700 compact utility tractor service repair manual
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manual
 
John deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manualJohn deere 4600 compact utility tractor service repair manual
John deere 4600 compact utility tractor service repair manual
 
John deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manualJohn deere 4500 compact utility tractor service repair manual
John deere 4500 compact utility tractor service repair manual
 
Cr 10x user manual
Cr 10x user manualCr 10x user manual
Cr 10x user manual
 
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd750 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
 
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
Kawasaki fd771 d (dfi) 4 stroke liquid-cooled v-twin gasoline engine service ...
 

Recently uploaded

Gauri Gawande(9) Constipation Final.pptx
Gauri Gawande(9) Constipation Final.pptxGauri Gawande(9) Constipation Final.pptx
Gauri Gawande(9) Constipation Final.pptxgauripg8
 
Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Catherine Liao
 
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1DR SETH JOTHAM
 
Aptopadesha Pramana / Pariksha: The Verbal Testimony
Aptopadesha Pramana / Pariksha: The Verbal TestimonyAptopadesha Pramana / Pariksha: The Verbal Testimony
Aptopadesha Pramana / Pariksha: The Verbal TestimonyDr KHALID B.M
 
Arterial health throughout cancer treatment and exercise rehabilitation in wo...
Arterial health throughout cancer treatment and exercise rehabilitation in wo...Arterial health throughout cancer treatment and exercise rehabilitation in wo...
Arterial health throughout cancer treatment and exercise rehabilitation in wo...Catherine Liao
 
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Catherine Liao
 
Muscle Energy Technique (MET) with variant and techniques.
Muscle Energy Technique (MET) with variant and techniques.Muscle Energy Technique (MET) with variant and techniques.
Muscle Energy Technique (MET) with variant and techniques.Anjali Parmar
 
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Catherine Liao
 
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)Dr. Aryan (Anish Dhakal)
 
End Feel -joint end feel - Normal and Abnormal end feel
End Feel -joint end feel - Normal and Abnormal end feelEnd Feel -joint end feel - Normal and Abnormal end feel
End Feel -joint end feel - Normal and Abnormal end feeldranji1
 
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...Catherine Liao
 
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.Gawad
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.GawadHemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.Gawad
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.GawadNephroTube - Dr.Gawad
 
Why invest into infodemic management in health emergencies
Why invest into infodemic management in health emergenciesWhy invest into infodemic management in health emergencies
Why invest into infodemic management in health emergenciesTina Purnat
 
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAntiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAkashGanganePatil1
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Sujoy Dasgupta
 
Cervical screening – taking care of your health flipchart (Vietnamese)
Cervical screening – taking care of your health flipchart (Vietnamese)Cervical screening – taking care of your health flipchart (Vietnamese)
Cervical screening – taking care of your health flipchart (Vietnamese)Cancer Institute NSW
 
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...Non-Invasive assessment of arterial stiffness in advanced heart failure patie...
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...Catherine Liao
 
linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...KavyasriPuttamreddy
 
Effects of vaping e-cigarettes on arterial health
Effects of vaping e-cigarettes on arterial healthEffects of vaping e-cigarettes on arterial health
Effects of vaping e-cigarettes on arterial healthCatherine Liao
 
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t..."Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...Catherine Liao
 

Recently uploaded (20)

Gauri Gawande(9) Constipation Final.pptx
Gauri Gawande(9) Constipation Final.pptxGauri Gawande(9) Constipation Final.pptx
Gauri Gawande(9) Constipation Final.pptx
 
Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...Scientificity and feasibility study of non-invasive central arterial pressure...
Scientificity and feasibility study of non-invasive central arterial pressure...
 
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
THORACOTOMY . SURGICAL PERSPECTIVES VOL 1
 
Aptopadesha Pramana / Pariksha: The Verbal Testimony
Aptopadesha Pramana / Pariksha: The Verbal TestimonyAptopadesha Pramana / Pariksha: The Verbal Testimony
Aptopadesha Pramana / Pariksha: The Verbal Testimony
 
Arterial health throughout cancer treatment and exercise rehabilitation in wo...
Arterial health throughout cancer treatment and exercise rehabilitation in wo...Arterial health throughout cancer treatment and exercise rehabilitation in wo...
Arterial health throughout cancer treatment and exercise rehabilitation in wo...
 
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
Impact of cancers therapies on the loss in cardiac function, myocardial fffic...
 
Muscle Energy Technique (MET) with variant and techniques.
Muscle Energy Technique (MET) with variant and techniques.Muscle Energy Technique (MET) with variant and techniques.
Muscle Energy Technique (MET) with variant and techniques.
 
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...Relationship between vascular system disfunction, neurofluid flow and Alzheim...
Relationship between vascular system disfunction, neurofluid flow and Alzheim...
 
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)
Book Trailer: PGMEE in a Nutshell (CEE MD/MS PG Entrance Examination)
 
End Feel -joint end feel - Normal and Abnormal end feel
End Feel -joint end feel - Normal and Abnormal end feelEnd Feel -joint end feel - Normal and Abnormal end feel
End Feel -joint end feel - Normal and Abnormal end feel
 
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
 
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.Gawad
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.GawadHemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.Gawad
Hemodialysis: Chapter 2, Extracorporeal Blood Circuit - Dr.Gawad
 
Why invest into infodemic management in health emergencies
Why invest into infodemic management in health emergenciesWhy invest into infodemic management in health emergencies
Why invest into infodemic management in health emergencies
 
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPTAntiplatelets in IHD, Dose Duration, DAPT vs SAPT
Antiplatelets in IHD, Dose Duration, DAPT vs SAPT
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
 
Cervical screening – taking care of your health flipchart (Vietnamese)
Cervical screening – taking care of your health flipchart (Vietnamese)Cervical screening – taking care of your health flipchart (Vietnamese)
Cervical screening – taking care of your health flipchart (Vietnamese)
 
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...Non-Invasive assessment of arterial stiffness in advanced heart failure patie...
Non-Invasive assessment of arterial stiffness in advanced heart failure patie...
 
linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...linearity concept of significance, standard deviation, chi square test, stude...
linearity concept of significance, standard deviation, chi square test, stude...
 
Effects of vaping e-cigarettes on arterial health
Effects of vaping e-cigarettes on arterial healthEffects of vaping e-cigarettes on arterial health
Effects of vaping e-cigarettes on arterial health
 
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t..."Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...
"Central Hypertension"‚ in China: Towards the nation-wide use of SphygmoCor t...
 

Service Manual Easylite electrolite device

  • 1. Cat. No. 2285 R3 EasyLyte Service Manual EDIC@ Na/K Na/K/Cl Na/K/Li Analyzers, Serial Numbers 5000 and above EasySampler
  • 2. Copyright ©1995 Medica Information contained herein is believed by Medica to be accurate; in any event, no responsibility, whether express or implied, is assumed hereby by Medica for or in connection with the use thereof, or for infringement of any third party rights which might arise therefrom, or from any represen- tation or omissions contained therein. Information is subject to change and/or updating without notice. Flow Chart Yes Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No CALIBRATE NOW? ANALYZE BLOOD? DAILY CLEANER? STANDBY MODE? ✽SECOND MENU?✽ ✽ SEE RESULTS? ✽ ✽ PACK USAGE? ✽ ✽ QUAL CONTROL? ✽ ✽ OPER FUNCTS? ✽ ✽ DIAGNOSTICS? ✽ ✽ DATE/RANGES? ✽ Whole Blood, Serum, or Plasma Analysis End of Day Cleaning Fluid Conservation, Maintenance Stored Results Displayed, Printed, Deleted Check % Pack Usage, Install New Pack, Reset % Countdown Run Controls, See Last Control, See/Print Statistics, Delete Results "On Demand" Calibration, Wash, Solutions Purge, User Options, Probe Wiping, I.D.#, Syringe Sample, Set Correlation,Vet Mode, Select Language, AutoSampler, Printer Off Check: Calibration Values Electrode Values Fluid Flow Sample Detector Pump Calibration Change Day/Time, AutoCal, Set Na/K/Cl/Li Ranges Yes No ✽ ANALYZE URINE? ✽ Urine Analysis
  • 3. Service Manual Contents Product Description Product Overview..........................................................................................................................1 System Overview ..........................................................................................................................1 Table of Symbols ..........................................................................................................................2 Front Assembly Components ......................................................................................................3 Front Assembly Components – Analyzer Series below 6100 ....................................................7 Electrode Housing Components ................................................................................................11 Rear Assembly Components ....................................................................................................12 Replacement List ........................................................................................................................13 Troubleshooting Troubleshooting Overview ..........................................................................................................15 Troubleshooting Guide................................................................................................................16 AC Power Verification ................................................................................................................17 DC Power Verification ................................................................................................................20 Display Verification......................................................................................................................21 Probe Motor Verification..............................................................................................................22 Pump Motor Verification..............................................................................................................23 Printer Verification ......................................................................................................................24 Sample Detector Verification ......................................................................................................25 Signal Input Verification ..............................................................................................................26 Serial Communication Verification ..............................................................................................28 QC Tests......................................................................................................................................29 Repair Repair Overview........................................................................................................................33 Tools ..........................................................................................................................................34 Rear Housing Removal ............................................................................................................35 Rear Housing Installation..........................................................................................................37 Power Supply Removal ............................................................................................................39 Power Supply Installation..........................................................................................................41 Printer Removal ........................................................................................................................43 Printer Installation ....................................................................................................................47 CPU Board Removal ................................................................................................................50 CPU Board Installation..............................................................................................................54 Display Removal ......................................................................................................................57 Display Installation ..................................................................................................................59 Pump Motor Removal ..............................................................................................................62 Pump Motor Installation ............................................................................................................63 Probe Motor Assembly Removal ..............................................................................................64 Probe Motor Assembly Installation............................................................................................67 Guide Rod Removal..................................................................................................................73 Guide Rod Installation ..............................................................................................................75 Electrode Jack Removal ..........................................................................................................76 Electrode Jack Installation ........................................................................................................77
  • 4. EasySampler Product Description ....................................................................................................................79 Assembly and Component Overview..........................................................................................80 Replacement List ........................................................................................................................81 Troubleshooting Guide................................................................................................................82 Alignment Verification..................................................................................................................83 Tray Placement Verification ........................................................................................................84 Carousel Test ..............................................................................................................................85 Tray Cycle Verification ................................................................................................................86 Sensor Verification ......................................................................................................................87 QC Tests......................................................................................................................................88 Repair Overview..........................................................................................................................90 Recommended Tools ..................................................................................................................90 Housing Removal........................................................................................................................91 Housing Installation ....................................................................................................................93 Spindle Column Removal............................................................................................................95 Spindle Column Installation ........................................................................................................97 Sensor Removal..........................................................................................................................99 Sensor Installation ....................................................................................................................100 Motor Removal..........................................................................................................................101 Motor Installation ......................................................................................................................103 PC Board Removal ..................................................................................................................105 PC Board Installation ................................................................................................................107 Specifications Product Specifications ..............................................................................................................109 EasySampler Specifications ....................................................................................................110 Appendix Display Messages......................................................................................................................111 QC TEST? Flowcharts ..............................................................................................................118 Schematics CPU Board CPU Board – Analyzer Series below 6100 Power Supply – Analyzer Series below 6100 Transformer Assembly EasySampler PCB
  • 5. Product Description 1 Product Description 1 Product Overview The EasyLyte analyzer is a direct measurement ion-selective electrode (ISE) system which uses Sodium (Na), Chloride (Cl), Potassium (K), and Lithium (Li) sensitive components to measure electrolytes in either whole blood, serum, plasma or urine. The analyzer software allows the operator to perform sample analysis, analyzer diagnostics and maintenance by simply pressing the "YES" or "NO" buttons in response to prompts displayed on the LCD screen. The EasyLyte analyzer performs the analysis both quickly and cost effectively. Analysis takes only one minute and utilizes approximately one hundred microliters of whole blood, serum or plasma; or four hundred microliters of diluted urine. The analyzer can re-calibrate automatically every four hours and can perform emergency (STAT) analysis on demand. Results are always displayed on the LCD screen and can also be docu- mented on paper by an integrated printer. The analyzer contains a patented reagent system including a modular reagent pack. This pack provides a minimum of two months supply of solution and also serves as a waste receptacle for solution and samples after analysis. System Overview The EasyLyte analyzer is engineered to make local service feasible. Modular construction is used throughout the design. The basic operating system consists of a motor which drives a peri- staltic pump. The peristaltic pump pulls the desired fluid through the system, disposing waste into a designated compartment inside the solutions pack. A second motor drives a lead screw which in turn drives a mechanical link-arm assembly. This link-arm assembly moves the sample probe vertically allowing the probe inside the solutions valve to aspirate the appropriate fluid for pump- ing: sample; standards; or air. The sample detector in the fluid line positions the fluids. Please note that mechanical and electrical assemblies can be replaced. Follow the instructions in the TROUBLE SHOOTING SECTION to determine if assembly replacement is necessary. Refer to the Replacement List in this section which lists replaceable assemblies, components, and appropriate catalog numbers. If repair to any electrical assembly is necessary, consult Schematics in the APPENDIX SECTION for further detail. Service personnel are encouraged to contact Medica directly with any problems or concerns relating to electrical assemblies.
  • 6. 2 Product Description NOTES Helpful information or references to other related sections of the manual. KEY INFORMATION Could take the EasyLyte analyzer out of operation. WARNING Possible biological hazard to the operator. EASYLYTE ANALYZER’S Information for EasyLyte analyzers with serial num- bers BELOW 6100 below 6100. ELECTRICAL HAZARD Possible electrical hazard to the operator. Table of Symbols 6100
  • 7. Product Description 3 Front Assembly Components This section describes the main assemblies and components of the EasyLyte front assembly. Refer to Figures P1, P2, and P3 for the location of each component. Front Housing front case of the analyzer Power Supply Assembly powers the analyzer Entry Module connection for the power cable Fuse Bay holds 2 fuses Power Supply Cable connects the power supply to the CPU assembly Voltage Selector Switch to select 110V or 220V power source CPU Assembly contains the CPU board with digital components and analog sensor inputs Firmware (EPROM) contains the software Interface Ports for EasySampler and RS-232 for serial devices Display Assembly shows EasyLyte prompts/results on LCD screen LCD Screen available in 1- line or 2- line display Display Cable connects the display to the CPU assembly Sample Detector Cable connects the sample detector to the display board Pump Assembly peristaltic pump to move fluids through the analyzer Pump Motor powers pump Pump Head Assembly contains pump body, rollers Pump Motor Cable connects the pump to the CPU assembly Probe Motor Assembly moves the probe arm Probe Motor with Lead Screw powers lead screw Mounting Bracket holds probe lead screw motor assembly in place Probe Position Sensor positions the probe position scale Sensor Cable connects the sensor to the CPU assembly Probe Position Scale positions the probe arm Probe Motor Cable connects the motor to the CPU assembly Printer Assembly prints a hard copy of analysis, diagnostics or maintenance results Printer Cable connects the printer to the CPU assembly Probe Arm holds and moves the probe to aspirate samples and calibrants Guide Rod allows probe arm to move vertically Electrode Jack(s) connects electrodes to the CPU assembly
  • 8. 4 Product Description Figure P1 Front Assembly Back View Printer Assembly CPU Assembly Power Supply Assembly AUTOSAMPLER P3 P1 P2 P4 P13 P6 Power Supply Cable Printer Cable Pump Motor Cable Sensor Cable Probe Motor Cable Display Cable Front Housing Display Assembly Sample Detector Cable Probe Position Sensor Probe Position Scale Electrode Jack(s) Firmware (EPROM) Interface Ports Voltage Selector Switch Entry Module Fuse Bay
  • 9. Product Description 5 Figure P2 Front Assembly - Right Side View Front Housing Probe Motor Assembly Mounting Bracket Probe Position Scale Probe Motor Guide Rod Lead Screw Display Cable Printer Assembly Probe Arm
  • 10. 6 Product Description Figure P3 Front Assembly - Left Side View Front Housing Pump Head Assembly LCD Screen Pump Assembly Pump Motor Display Cable Printer Cable
  • 11. Product Description 7 Front Assembly Components – Analyzers Series Below 6100 This section describes the main assemblies and components of the EasyLyte front assembly, analyzer series below 6100. Refer to Figures P4, P5, P6 for the location of each component. Front Housing front case of the analyzer Power Supply Assembly powers the analyzer Entry Module connection for the power cable Fuse Bay holds 2 fuses Power Supply Cable connects the power supply to the CPU assembly Voltage Selector Switch to select 110V or 220V power supply CPU Assembly contains the CPU board with digital components and analog sensor inputs Firmware (EPROM) contains the software Interface Ports for EasySampler and RS-232 for serial devices Display Assembly shows EasyLyte prompts/results on the LCD screen LCD Screen available in 1-line or 2-line display Display Cable connects the display to the CPU assembly Sample Detector Cable connects the sample detector to the display board Pump Assembly peristaltic pump to move fluids through the analyzer Pump Motor powers pump Pump Head Assembly contains pump body, rollers Pump Motor Cable connects the pump to the CPU assembly Probe Motor Assembly moves the probe arm Probe Motor with Lead Screw powers lead screw Mounting Bracket holds probe lead screw motor assembly in place Probe Position Sensor positions the probe position scale Sensor Cable connects the sensor to the CPU assembly Probe Position Scale positions the probe arm Probe Motor Cable connects the motor to the CPU assembly Printer Assembly prints a hard copy of analysis, diagnostics or maintenance results Printer Cable connects the printer to the CPU assembly Probe Arm holds and moves the probe to aspirate samples and calibrants Guide Rod allows the probe arm to move vertically Electrode Jack(s) connects electrode to the CPU assembly 6100
  • 12. 8 Product Description Figure P4 Front Assembly Back View – Analyzers with Serial Numbers below 6100 6100 Printer Assembly CPU Assembly Power Supply Assembly AUTOSAMPLER P3 P1 P2 P4 P13 P6 Power Supply Cable Printer Cable Pump Motor Cable Sensor Cable Probe Motor Cable Front Housing Display Assembly Sample Detector Cable Probe Position Sensor Probe Position Scale Electrode Jack(s) Display Cable Entry Module Fuse Bay Firmware (EPROM) Voltage Selector Switch Interface Ports
  • 13. Product Description 9 Figure P5 Front Assembly Right Side View - Analyzers with Serial Numbers below 6100 Front Housing Probe Motor Assembly Mounting Bracket Probe Position Scale Probe Motor Guide Rod Lead Screw Printer Assembly Voltage Selector Switch 6100
  • 14. 10 Product Description Figure P6 Front Assembly Left Side View – Analyzers with Serial Numbers below 6100 Front Housing Pump Head Assembly LCD Screen Pump Assembly Pump Motor Display Cable Printer Cable 6100
  • 15. Product Description 11 Figure P7 - Electrode Housing Assembly Electrode Housing Assembly Components This section describes the components of the EasyLyte electrode assembly. Refer to Figure P7 for the location of each component. Electrodes sodium, potassium, chloride, lithium, reference Sample Detector detects presence of air/liquid, sets sample position inside electrodes Internal Filling Solution 2 molar KCI solution, acts as 'salt bridge’ with reference electrode Fill Plug prevents the internal filling solution from evaporating Membrane Assembly connects reference electrode with other electrodes Fill Plug Sample Detector Chloride Electrode (Green Dot) or Lithium Electrode (Yellow Dot) Potassium Electrode (Red Dot) Sodium Electrode (Blue Dot) Internal Filling Solution Reference Electrode Membrane Assembly
  • 16. 12 Product Description On EasyLyte Analyzers with serial numbers below 6100 the voltage selector is located inside the front housing on the power supply. Figure P8 - Rear Assembly Rear Assembly Components This section describes the components of the EasyLyte Rear Assembly. Refer to Figure P8 for the location of each component. Rear Housing rear case Label identifies serial number Autosampler Port Serial Port Power Cord Entry Module Label with Serial Number Voltage Selector Switch Rear Housing 6100
  • 17. Product Description 13 Replacement List Consumable and assemblies items can be purchased directly from Medica. CONSUMABLE ITEMS Description Catalog No. Na Electrode 2102 K Electrode 2101 Cl Electrode 2113 Li Electrode 2106 Reference Electrode 2103 Sample Detector 2257 Sample Probe 2107 Solutions Valve 2108 Fill Plug 0195 Membrane Assembly 2258 Tubing Kit 2104 EasyLyte Operator's Manual 2277
  • 18. ASSEMBLIES Description Part No. Electrode Housing Na/K/Cl 000885 Electrode Housing Na/K/Li 000985 Electrode Housing Na/K 000177 Power Supply 001676 with CPU 001937 or CPU (Japanese or Chinese software) 001957 Power Supply 001287 with CPU 001081 Display, 1 Line 001284 Display, 2 Line 001901 Display, Graphic (Japanese or Chinese characters) 001956 Pump 001271 Probe Motor Assembly 001392 Probe Position Sensor 001306 Motor Driver 001148 – CPU 001937, positions U35, U36 – CPU 001081, positions U22, U23 Printer 001296 Electrode Jack 000173 The EasyLyte CPU Assembly has ten (10) variations: EasyLyte Lithium, 1-line display 001937-003 EasyLyte Chloride, 1-line display 001937-004 EasyLyte Lithium, 2-line display 001937-005 EasyLyte Chloride, 2-line display 001937-006 EasyLyte Na/K, 1-line display 001937-007 EasyLyte Na/K, 2-line display 001937-008 EasyLyte Na/K, Chinese 001957-001 EasyLyte Na/K/Cl or Li, Chinese 001957-002 EasyLyte Na/K, Japanese 001957-003 EasyLyte Na/K/Cl or Li, Japanese 001957-004 The CPU Part Numbers include the EasyLyte CPU Board with software installed, programmed to operate specifically in EasyLytes described as above. When ordering this item, please list the specific part number (-003, -004, -005, etc.) that you require. CPU series 001957 functions correctly only when used with the Graphic Display (displays Japanese or Chinese characters), part number 001956. Please note that EasyLyte Analyzers with serial Numbers below 6100 are assembled using Power Supply, 001287 and CPU, 001081. These two items work together as a set. Power Supply, 001676 and CPU, 001937 also work together as a set. These parts cannot be mixed on the EasyLyte. 14 Product Description 6100 6100
  • 19. Troubleshooting 15 Troubleshooting Overview This section describes troubleshooting the mechanical and electrical functions of the EasyLyte analyzer. This section is based on the principle of replacing suspected faulty assemblies with new assemblies. Medica recommends keeping an inventory of functioning test assemblies for troubleshooting the EasyLyte. Refer to the Troubleshooting Guide in this section to deter- mine the cause and corrective action that must be taken to properly service the EasyLyte Analyzer. Since EasyLyte Analyzers with serial numbers below 6100 are assembled using Power Supply, part no. 05001065 and CPU, part no. 05001063 the troubleshooting instructions differ in certain areas. Please refer to the serial number, located on the back panel of the analyzer, to identify which instructions to follow. For information on troubleshooting the EasyLyte Analyzer flow system and consumable items, consult the EasyLyte Operator’s Manual. See the Replacement List in the PRODUCT DESCRIPTION SECTION of this manual for a complete listing of all consumable items and cata- log numbers. For information on troubleshooting the EasyLyte EasySampler consult Troubleshooting in the EASYSAMPLER SECTION. 6100
  • 20. 16 Troubleshooting Troubleshooting Guide Refer to the chart below to help determine the problem and corrective action needed. Problem Description Problem Source Corrective Action Steps Blank Display Power Malfunction or • AC Power Verification Defective Display • DC Power Verification • Display Verification No response Power Malfunction or • Display Assembly Verification to YES/NO buttons Defective Display • DC Power Verification Blank Display, Faulty EPROM Installation • Inspect EPROM Intallation Beeping or Power Malfunction • DC Power Verification “MTS” Errors Probe Motor • Probe Motor Verification Malfunction “MTP” Errors or Pump Motor • Perform PUMP CAL test No Pump Movement Malfunction in DIAGNOSTICS? Menu • Pump Motor Verification Printer Does Not Print Printer Malfunction • Printer Verification Sample Detector Test Failure Sample Detector/Circuit • Sample Detector Result 255 or Ø Malfunction Verification Bad Electrode Voltages Electrodes or • Check Electrode Installation Signal Input Malfunction and Performance • Signal Input Verification Serial Transmission Failure Serial Cable or • Serial Communication Transmission Protocol Verification
  • 21. Troubleshooting 17 Hot Earth Ground Neutral P3 P1 P2 P4 P13 Entry Module Fuse Bay Figure T1 AC Power Verification WARNING: Careless handling of AC Power, especially 220V, can be hazardous or fatal. Only qualified personnel should work with AC Power. 1. Check that the voltage selector switch is set at the proper voltage (110V or 220V). This volt- age must match the power outlet voltage. 2. Unplug the power cord from the entry module and plug into the outlet. 3. Check the voltage at hot and neutral at the end of the power cord. Refer to Figure T1. For EasyLyte analyzers set to 110V, the voltage at hot should be 90V–125V. For EasyLyte analyzers set to 220V, the voltage at hot should be 210V–240V. If the voltage is incorrect, replace the power cord and return to normal operation. Otherwise, proceed to STEP 4. 4. Using a small flat-tip screwdriver pry open the fuse bay. Refer to Figure T2 for the location of fuse bay. Remove and inspect both fuses as follows: For 110V EasyLyte analyzers, the fuses must be rated 0.8A, 250 Volts For 220V EasyLyte analyzers, the fuses must be rated 0.4A, 250 Volts Visually inspect the wire conductor inside the fuses. If the wire is broken or missing, replace the fuse. Figure T2
  • 22. 18 Troubleshooting Figure T3 5. Remove the rear housing. Refer to Rear Housing Removal in the REPAIR SECTION. 6. Check for short circuits between the test points listed below: TP12 and TP2 TP12 and TP1 TP12 and TP3 TP12 and TP10 TP12 and TP11 If a short exists, replace the CPU board. Otherwise, return to normal operation. 7. Disconnect the Power Supply cable (6-pin connector) from the CPU board, see Figure T3. Connect the EasyLyte entry module to a wall outlet with a known good power cord. 8. Connect the reference side of the voltmeter to Pin 1 of the power supply cable (6-pin connector). 9. Check the AC voltages at Pins 2 and 5. see Figure T3 for the location of the pins. The voltage for each should be 7.6 ± 2V. 10. Check the voltaage across Pins 3 and 4. The voltage should be 26± 2V. 11. Measure the resistance between Pin 6 and power supply chassis. The resistance should be zero. If any of the voltages are incorrect, replace the power supply module and return to normal operation. AUTOSAMPLER P4 P13 Power Supply Cable Pin 6 –> 1
  • 23. Troubleshooting 19 If any of the voltages are incorrect, replace the power supply and return to normal operation. 12. Check for short circuits between the test points below: TP11 and TP5 TP11 and TP8 TP9 and TP6 TP9 and TP7 If a short exists, replace the CPU Board and return to normal 13. Disconnect the power supply cable (13-pin connector) from the CPU board, see Figure T4. 14. Connect the reference side of the voltmeter to Pin 5 of the powr supply cable. Avoid shorting the adjacent pins. 15. Check the DC voltages at the pin placements listed below. See Figure T4 for the pin locations. Test Point Value Tolerance Pins 1,2, and 3 +5.0V ±0.15V Pin 4 through 8 0.0V ±0.15V Pins 9 and 10 35.0V ±3.00V Pin 11 –5.0V ±0.15V Pin 12 +5.0V ±0.15V Pin 13 0.0V ±0.15V 6100 6100 6100 P4 P13 Power Supply Cable Pin 13 –> 1 Figure T4
  • 24. 20 Troubleshooting DC Power Verification 1. Remove the rear housing. See Rear Housing Removal in the REPAIR SECTION for details. 2. Connect the power cord first to an outlet, then to the entry module. 3. Connect the reference side of the voltage meter to TP12 on the CPU board. 4. Verify the voltage at each test point listed below. Test Point Value Tolerance TP1 5.00 V +/- 0.15 V TP2 34.0 V +/- 2.0 V TP3 5.00 V +/- 0.15 V TP10 2.50 V +/- 0.10 V TP11 -5.00 V +/- 0.15 V 5. For EasyLyte serial numbers below 6100, use TP11 as the reference point. Verify the voltage at each test point listed below. Test Point Value Tolerance TP5 5.00 V +/- 0.15 V TP8 34.0 V +/- 2.0 V TP9 0.00 V +/- 0.15 V TP6 5.00 V +/- 0.10 V TP7 -5.00 V +/- 0.15 V 6. If any of the voltage values in STEP 4 OR STEP 5 are not within specification, AND the AC power verification is successful, replace the CPU board. 6100
  • 25. Troubleshooting 21 Display Verification 1. Disconnect the display cable from the CPU board. See Display Assembly Removal in the REPAIR SECTION for details. 2. Connect a known good display assembly to the CPU board and connect the main power cord. Verify that the test display is operative. Based on the test results, take the appropriate corrective action. • Display Blank and YES/NO buttons DO NOT operate: Remove the test display cable. Perform AC and DC power verification. If the AC and DC power tests are successful, replace the CPU board with a new CPU board, and return to normal operation. See CPU Assembly in the REPAIR SEC- TION for details. • Display On and YES/NO buttons function: Remove the original display assembly and replace with a new display assembly. See Display Assembly in the REPAIR SECTION for details.
  • 26. 22 Troubleshooting Probe Motor Verification 1. Disconnect the power cord, wait 5 seconds, then reinstall the power. The probe arm should move down and back up. If there are no error messages, resume normal operation. If an “MTS” error occurs, proceed to STEP 2. 2. Remove the sample probe from the probe arm. For details refer to EasyLyte Operator’s Manual. Remove and reinstall the power cord. The probe arm should move down and back up. If there are no error messages examine and if necessary replace the sample probe, solu- tions valve and probe wiper. If an “MTS” error occurs, proceed to STEP 3. 3. If the probe arm does not move at all perform the following operations. A. Visually inspect that the actuator and retaining ring are correctly installed on the probe motor lead screw. Refer to Figure R64 in the REPAIR SECTION. If the installation is incorrect, refer to CPU Removal and Probe Motor/Lead Screw Installation in the REPAIR SECTION for details. B. Check that the probe motor wires are completely seated inside the 4-pin connector as follows. – Remove the probe motor connector from the bottom left of the CPU Board (position P3). – Compress the strain-relief cap against the probe motor connector to firmly seat the wires. – Reinstall the probe motor connector into the CPU Board. C. Repeat STEP 1. If no errors occur, resume normal operation. 4. If the probe arm moves and stops anywhere except the top position, or if the probe arm vibrates against the top of the EasyLyte housing, replace the probe position sensor. To replace the sensor, first remove the CPU Board and the probe position scale. Refer to Probe Motor/Lead Screw in the REPAIR SECTION for details. 5. If the Probe does not move at all, check for a short circuit on the CPU Board between TP12 and TP2. If there is a short circuit, replace the motor driver at position U36.See Assemblies in the PRODUCT DESCRIPTION SECTION. For EasyLyte serial numbers below 6100, check for a short circuit on the CPU Board between TP8 and TP11. If there is a short circuit, replace the motor driver at position U23. See Assemblies in the PRODUCT DESCRIPTION SECTION. 6. If no short circuit is found, replace the probe motor assembly. Refer to the Probe Motor Assembly in the REPAIR SECTION for details. 6100
  • 27. Troubleshooting 23 Pump Motor Verification 1. Enter the QC TEST? function and proceed to the PUMP TEST? Refer to QC Tests in this section for further instructions. 2. At the PUMP TEST? prompt, press YES. 3. The pump should rotate ten revolutions. 4. If the pump vibrates or does not run smoothly, ensure that the pump motor wires are properly seated inside the 6 pin connector as detailed below. A. Remove the pump motor connector from the bottom left of the CPU board See Figure P1 in the PRODUCT DESCRIPTION SECTION. B. Compress the strain-relief cap against the pump motor connector to firmly seat the wires. C. Reinstall the pump motor connector into the CPU board. 5. Repeat the Pump Test. If the pump malfunctions, replace the pump motor with a test motor. See Pump Motor in the REPAIR SECTION for futher instructions. 6. Repeat the Pump Test with the test motor. Based on the test results, take the appropriate cor- rective action. • Motor Operative: Replace the test motor with a new motor. See Pump Motor in the REPAIR SECTION. • Motor Malfunction: Check for a short on the CPU board between TP12 and TP2. If there is a short circuit, replace the pump motor driver at U35. See Assemblies in the PRODUCT DESCRIPTION SECTION. 7. For EasyLytes with serial numbers below 6100, check for a short circuit between TP8 and TP11. • Short Circuit: Replace the pump motor driver at position U22. • No Short Circuit: Replace the pump motor assembly. See Pump Motor in the REPAIR SECTION for details. 6100
  • 28. 24 Troubleshooting Printer Verification 1. Press the printer paper advance button. If the printer head moves proceed to STEP 2. If the printer head does not move, proceed to STEP 3. 2. Verify that the printer is on. Turn the printer on by pressing NO to the PRINTER OFF? ques- tion in the USER OPTIONS section. 3. If the printer is still not working, remove the power cord from the analyzer and wait for ten sec- onds, then install the power cord back into the entry module. Press the printer paper advance button. The printer head should move across the paper and the paper should advance for- ward. 4. If the printer is still not working, remove the printer cover and check for any paper obstruction in the printer head area. Remove any obstructions if possible, without damaging the printer assembly or paper tear bar. If an obstruction cannot be removed, replace the Printer Assembly. Refer to Printer Removal and Installation in the REPAIR SECTION for details. 5. If there are no obstructions, but the printer does not work, perform the following steps. A. Remove the rear housing as detailed in the REPAIR SECTION . B. Remove the printer cable at the bottom right of the CPU board. C. Compress the strain-relief cap against the probe motor connector to firmly seat the printer cable. D. Reinstall the printer cable into the CPU Board. 6. If the printer still does not print, connect a test printer assembly to the CPU Board. If the test printer works, install a new printer assembly. If the test printer does not work, replace the CPU board. Refer to CPU Installation in the REPAIR SECTION for details.
  • 29. Troubleshooting 25 Sample Detector Verification 1. Remove the electrode housing from the analyzer. See the EasyLyte Operators Manual for fur- ther details. 2. Remove the sample detector from the front housing. 3. Install the sample detector into the sample detector receptacle. Do not install the electrode housing or pump tubing at this time. 4. Proceed to the DIAGNOSTICS? menu and perform the SAMPLE DETECTOR TEST? • With the sample detector installed, the value on the display screen should be less than 200. • With the test in progress, remove the sample detector. The display value should rapidly change to 255. If the sample detector test meets these specifications, resume normal operation. A value of 255 with the sample detector installed indicates a bad connection. Proceed to STEP 5 to check the sample detector circuit. 5. Reinstall the sample detector. Look at the left side of the sample detector. A red light must shine from the LED hole. If there is no red light, install a known good sample detector. If there is still no red light, replace the Display Assembly. Refer to the REPAIR SECTION for informa- tion. 6. If the TEST sample detector red light is on, check the four connecting pins on the ORIGINAL sample detector. The pins must face straight out forming a rectangle. If the pins are bent against the black plastic body, straighten the pins using a small flat head screwdriver. 7. Reinstall the original sample detector and repeat STEP 4. • If the red light comes on: Reassemble the EasyLyte analyzer and perform a SOLUTIONS PURGE and a SAMPLE DET. TEST. Resume normal operation. • If the red light remains OFF: Install a new sample detector.
  • 30. Signal Input Verification 1. Remove and inspect electrodes and electrode housing for liquid leakage and salt build-up. All electrodes and the electrode housing surfaces MUST be completely dry. See EasyLyte Operator’s Manual for further information on consumable items. 2. Inspect the electrode pin receptacles on the EasyLyte analyzer front housing. Remove any liq- uid or salt from these areas. 3. Enter * DIAGNOSTICS? * and answer YES to ELECTRODE VALUES? The display most likely will show “249 249 249”. The following steps require the use of the Mini-Test High Impedance Voltage Source, part no. 001374. If this tool is not available, the remaining tests cannot be performed. 4. Connect the Voltage Source to the EasyLyte analyzer as follows. Refer to Figure T5. A. Connect the three red wires from the Voltage Source to the K+, Na+, and CL-/Li+ electrode receptacles B. Connect the black wire from the Voltage Source to the reference electrode receptacle. C. Connect the remaining black wire into the back connector on the Cl-/Li+ red wire. This step will short out all three electrode channels. Figure T5 5. Read the EasyLyte display. The electrode value specification is 0 mV to 12 mV. For analyzers with serial numbers below 6100, the electrode specification is -12 mV to +12 mV. 26 Troubleshooting 6100
  • 31. Troubleshooting 27 6. Disconnect the black wire from the back connector on the Cl-/Li+ red wire. 7. Turn on the Voltage Source. 8. Set the Voltage Source to the group 1 settings listed in the table below. Three new individual values will appear on the screen. Wait a few seconds for the values to stabilize then verify that the values are within specification. Press YES to print the values. 9. Set the Voltage Source to the group 2 settings, listed in the table below, and repeat STEP 8. 10. Calculate the DIFFERENCE between the group 1 values and group 2 values. Verify that the values are within specification. Settings Group 1 Group 2 Polarity setting acid (positive) acid (positive) Impedance setting HI HI Offset setting 8-6pH 10-4 pH Voltage Values Individual values 59 mV± 10mV 177 mV ± 10 mV Difference between group 1 and 2 118 mV ± 3 mV 11. If any of the voltage values are out of specification, examine the electrode jacks for corrosion and verify the connection between the electrode jack and CPU board. Refer to Electrode Jack Removal in the REPAIR SECTION. If the electrode jacks are not corroded and are properly connected, replace the CPU board with a new CPU board. See CPU Assembly in the REPAIR SECTION for details. 12. Turn off the Voltage Source. If the technician has any questions about the reliability of the mini-tester, install a new 9 volt battery and repeat the above procedure.
  • 32. 28 Troubleshooting Serial Cable Settings RS-232 Protocols Serial Communication Verification 1. Connect the EasyLyte analyzer to a device with a known working serial port. Use a known good EasyLyte serial cable, part no. 05001334. Make cross connections as required (e.g. 2 to 3; 4 to 5). 2. Refer to the following serial interface specifications. Serial Interface Specifications Connector 25-pin, D-Sub male, Pin type RS-232 Pin assignments Pin 1 Chassis, Not Connected Pin 2 TxD Pin 3 RxD Pin 4 RTS Pin 5 CTS Pin 7 Signal Ground Transmission Rate 1200 baud Bit Configuration 8 data bits, 1 stop Bit parity None Control Codes None All lines end with Carriage return line feed (CRLF) All transmission is ASCII string 3. When correctly linked the EasyLyte will transmit to the external device all of the data that is printed by the EasyLyte printer, in the same format.
  • 33. Troubleshooting 29 QC Tests 1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis- play, as shown in the chart below. LCD Screen Displays Press CALIBRATE NOW? NO DAILY CLEANER? NO STANDBY MODE? NO SECOND MENU? YES * SEE RESULTS? * NO * PACK USAGE? * NO * QUAL CONTROLS? * NO * OPER FUNCTS? * NO * DIAGNOSTICS? * NO * DATE/RANGES? * YES * CHANGE DAY/TIME? * NO CAL FREQUENCY? NO Na/K/Cl or Na/K/Li RANGES? YES BLOOD Na LIMITS? YES “Numeric value” —>“Numeric value” wait 2. Two numbers appear on the display screen separated by an arrow. A cursor will be posi- tioned underneath the first digit of the first number. The second number is not applicable in this procedure. 3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the digit is correct press YES to advance to the next digit. 4. Continue until the first number is changed to “246.0”. 5. Press YES when finished. QC TEST? will appear on the screen. If an erroneous value is input the screen displays CORRECT?. Press NO to return to STEP 2 and correct the number. 6. Press YES to QC TEST?
  • 34. 30 Troubleshooting 7. Consult the table below which lists all of the QC tests and their functions. To select a test press YES. To bypass a test press NO. QC Test QC Tests Function QC TEST? Access to the QC Test functions. TURN ON MV‘S? Prints the registered Na, K, Cl or Li electrode calibration and analysis voltages. Calibration: Stds. A & B voltages Blood analysis: Sample and Std. A voltages Urine analysis: Sample and Std. B voltages To end printing, return to TURN ON MV’s and press YES. At PLEASE CONFIRM!, press NO. START CLOCK? Permits changing date and time if analyzer setting is incorrect. Press YES to PLEASE CONFIRM! to change date and time. Please note that this automatically turns on the electrode voltages. To turn off the electrode voltages, press NO to all messages until START CLOCK? appears; press YES. At PLEASE CONFIRM!, press NO. CHECK CAL CONSTS? Displays the factory set electrode constants. The acceptable ranges are: Left column number: 0.92 to 1.08 Right column number: -800 to +800 If values are out of specification contact Medica. RAM/PROM TEST? Displays current software version and date installed. RAM CHECK OUT? Clears the EasyLyte memory. This will erase all stored EasyLyte calibration and sample data. After testing this function exits the operator out of the QC TEST? menu and resets the analyzer. SIPPER TEST? Moves the sample probe all the way down, then up to test mechanical performance. Abnormal movement is indicated by the display messages: “MTS”, “Na/K/CL” or “Na/K/Li.” PUMP TEST? Rotates the pump 10 revolutions to test mechanical performance. Abnormal movement is indicated by the display message “MTP”. TEST CAROUSEL? EasySampler QC Test Functions. CALIB CAROUSEL? Refer to the EASYSAMPLER SECTION FIND HOME? for details. CYCLE TRAY? SOL’N PURGE? Purges any air in the fluid system. Calibration must be performed after a solution purge. CALIBRATE NOW? Performs analyzer calibration. 6100
  • 35. Troubleshooting 31 QC Test Function QC Test Function DIAGNOSTICS? Performs diagnostic testing of the EasyLyte Analyzer (See EasyLyte Operator’s Manual). LAST CAL VALUES? Checks last calibration values. Acceptable ranges are Na: 50-64, K: 50-64, Cl: 40-64, Li: 50-64. ELECTR’D VALS? Direct electrode millivolt readings. Typical values are Na: 35-125, K: 35-125 Cl: 35-125, Li: 30-60. FLUID FLOW TEST? Confirms that fluid is aspirating through the analyzer. SAMPLE DET. TEST? Tests the sample detector, which detects the differ- ence between air and liquid in the fluid system. PUMP CAL TEST? Checks the pump calibration. Expected values are between 500-800. TEST VALVE? Verifies probe/valve alignment and pump efficiency. Calculates the number of pump motor steps to move each solution from the valve through the probe to the sample detector. Typical values are Flush: 800-1200, Std B 800-1200, Std A 800-1200, Air 500-800. The three reagent values should be 100-200% of the Air value. In most cases, abnormal results indicate a faulty solutions valve. NOTE: The operator must perform a successful SOL’ N PURGE before this test. RUN STATISTICS? Analyzes standards A & B (in the EasyLyte Solutions Pack) as samples. RUN CONTROLS? Analyzes 8 normal or 8 abnormal control samples. Calculates the mean, standard deviation and coeffi- cient of variation for control sets. NOTE: The operator must perform a successful CALIBRATION before run- ning this test. TEST ALIGNMENT? Verifies the number of probe motor steps to move the probe through each solutions valve position. Hex val- ues are Std. A: 3F, Flush: 3F, Std. B: 3F, Air: 3F. This test is useful only when assembling the EasyLyte analyzer. BURN IN? Manufacturer’s procedure to test electronic compo- nents of the CPU assembly. Not applicable for servic- ing the analyzer. CONTINUOUS? Part of Burn In sequence. QC TEST? Access to the QC Test functions. Press NO to return to the main EasyLyte menu.
  • 36. Repair 33 Repair Overview Always consult the TROUBLESHOOTING SECTION first to determine the problem before attempting to repair the analyzer. Always remove the solutions pack and the electrode housing from the analyzer before beginning any repair work. Refer to the EasyLyte Operator’s Manual for detailed instruc- tions. Follow the assembly and removal instructions in the order outlined in this section unless other- wise noted. Always unplug the power cord from the analyzer before beginning any repair work. Note that only complete mechanical and electronic assemblies can be replaced. Refer to the Replacement List in the PRODUCT DESCRIPTION SECTION which lists replaceable assem- blies as well as consumable items. Consult the list to determine the appropriate replacement item and part number. If repair to any electrical assembly is necessary, consult the Schematics in the APPENDIX SEC- TION for further detail. Service personnel are encouraged to contact Medica directly with any problems or concerns relating to electrical assemblies.
  • 37. 34 Repair Recommended Tools The following is a list of the tools required to perform repairs on the EasyLyte analyzer. Please refer to this list to determine the appropriate tool and size needed. Phillips Screw Driver, #1 Phillips Screw Driver, #2 Allen Wrench, 1/16 inch Scissors Hex Nut Driver, 11/32 inch, hollow stemmed Retaining Ring Pliers Resistor 10 OHM, 5 watts Volt Meter Optional Tools The following is a list of tools recommended to help perform repairs on the EasyLyte analyzer. Please refer to this list to determine the tool and part number when ordering. Alignment Valve 000104 Mini-Test High Impedance Voltage Source 001374
  • 38. Repair 35 Screws Screws Entry Module Power Cord Rear Housing Removal 1. Disconnect the power cord from the entry module, see Figure R1. 2. Detach the four screws from the rear housing using a #2 Phillips screw driver, as shown in Figure R2. Figure R1 Figure R2
  • 39. 36 Repair Figure R3 3. Separate the front and back housing. See Figure R3. Rear Housing Front Housing
  • 40. Repair 37 Figure R4 Rear Housing Installation 1. Align and attach the front and back housing. See Figure R4. Verify that all cable connectors are attached properly. Refer to Figures P1 and P6 in the PRODUCT DESCRIPTION SECTION for detail. Verify that all wires are contained within the front housing before proceeding. Rear Housing Front Housing
  • 41. 38 Repair Figure R6 Figure R5 Entry Module Power Cord Screws Screws 2. Attach the four screws to the rear housing using a #2 Phillips screw driver, as shown in Figure R5. 3. Connect the power cord to the entry module, see Figure R6.
  • 42. Repair 39 Power Supply Cable Connector to CPU Board Power Supply Cable Connector to CPU Board Screws Figure R9 Figure R8 Figure R7 Power Supply Removal 1. Disconnect the power supply cable from the CPU board, see Figure R7. Refer to Figure R8 for EasyLyte analyzers with serial numbers below 6100. 2. Detach the four screws from the bottom of the front housing, using a #2 Phillips screw driver. See Figure R9. 6100
  • 43. 40 Repair Power Supply Assembly Power Supply Assembly Figure R11 Figure R10 3. Lift and remove the power supply assembly from under the CPU board, as shown in Figure R10. Refer to Figure R11 for EasyLyte analyzers with serial numbers below 6100. 6100
  • 44. Repair 41 Power Supply Assembly Figure R12 Power Supply Assembly Screws Figure R14 Figure R13 Power Supply Installation 1. Slide the power supply assembly under the CPU board, as shown in Figure R12. Fit into the alignment holes in the bottom of the front housing. Refer to Figure R13 for EasyLyte analyzers with serial numbers below 6100. 2. Attach the four screws through the bottom of the front housing using a #2 Phillips screw driver. See Figure R14. 6100
  • 45. 42 Repair Power Supply Cable Connector to CPU Board Figure R16 Power Supply Cable Connector to CPU Board Figure R15 3. Connect the power supply cable to the CPU board, as shown in Figure R15. Refer to Figure R16 for EasyLyte analyzers with serial numbers below 6100. 6100
  • 46. Repair 43 Printer Bay Paper Roll Printer Lid Printer Figure R17 Figure R18 Printer Removal 1. Remove the printer lid from the printer, see Figure R17. 2. Remove the paper roll from the printer bay, as shown in Figure R18.
  • 47. 44 Repair Figure R20 Printer Cable Connector to CPU Board Printer Cable Connector to CPU Board Figure R19 3. Disconnect the printer cable from the CPU board, see Figure R19. Refer to Figure R20 for EasyLyte analyzers with serial numbers below 6100. 6100
  • 48. Repair 45 Screws Retaining Clamps Printer Cable Figure R22 Figure R21 4. Detach the printer cable from the retaining clamps located along the right side of the CPU board, as shown in Figure R21. 5. Detach the two screws using a # 2 Phillips screw driver. See Figure R22.
  • 49. 46 Repair CPU Board Printer Cable Printer Body Printer Body Figure R24 Figure R23 6. Remove the printer body as shown in Figure R23. 7. Pull the printer cable up from behind the CPU board, while holding the printer body as shown in Figure R24.
  • 50. Repair 47 Printer Body Screws CPU Board Printer Cable Printer Body Figure R27 Figure R26 Figure R25 Printer Installation 1. Slide the printer cable behind the CPU board, as shown in Figure R25. 2. Install the printer body into place. See Figure R26. 3. Attach the two screws using a #2 Phillips screw driver. See Figure R27.
  • 51. 48 Repair Figure R30 Figure R28 Printer Cable Connector to CPU Board Printer Cable Connector to CPU Board Figure R29 Paper Feed Button Printer Bay Paper Roll 4. Connect the printer cable to the CPU board, see Figure R28. Refer to Figure R29 for EasyLyte analyzers with serial numbers below 6100. 5. Using scissors, cut the beginning of the paper roll into a “V” shape. See Figure R30. Insert the paper into the feeder slot inside the printer bay, and press the paper feed button. 6. Place the printer roll into the printer bay, as shown in Figure R31. Figure 31 6100
  • 52. Repair 49 Printer Cable Retaining Clamps Printer Printer Lid Figure R33 Figure R32 7. Place the printer lid onto the printer, see Figure R32. 8. Attach the printer cable to the retaining clamps located along the right side of the CPU board, as shown in Figure R33.
  • 53. 50 Repair Pump Motor Cable Sensor Cable Probe Motor Cable Power Supply Cable Printer Cable Sensor Cable Pump Motor Cable Probe Motor Cable CPU Board Connector to CPU Board Display Cable Power Supply Cable Printer Cable Figure R36 Figure R35 Figure R34 CPU Board Removal 1. Disconnect the probe motor, sensor, pump motor, power supply and printer cables from the CPU board. See Figure R34. Refer to Figure R35 for EasyLyte analyzers with serial num- bers below 6100. 2. Remove the display cable from the top of the CPU board as shown in Figure R36. 6100
  • 54. Repair 51 CPU Board Display Cable Tabs Display Cable Figure R37 Electrode Jacks Connector Pin to CPU Board Figure R38 3. For EasyLyte Analyzers with serial numbers below 6100, push the display cable tabs to the side to unlock. Remove the cable from the connector. See Figure R37. 4. Disconnect the electrode jacks from behind the CPU board. See Figure R38 for the location of the electrode jacks. To disconnect, gently pull electrode jack straight away from the con- nector pin of the CPU Board. The connector pins are extremely delicate. To avoid pin breakage when removing the electrode jacks, gently pull the electrode jacks straight away from the pins. If a pin breaks the entire CPU board may require replacement. 6100
  • 55. 52 Repair Screw Figure R40 5. Disengage the CPU board from the four plastic mounting posts, located at each corner of the CPU board. To disengage, press the mounting post tab in and pull the CPU board over the tab, as shown in Figure R39. 6. Disengage the center plastic mounting post from the CPU board, located above the serial and EasySampler ports. 7. For EasyLyte analyzers with serial numbers below 6100, detach the screw from the center of the board using a # 2 Phillips screw driver. See Figure R40. 6100 Pull Board Over Tab Plastic Mounting Post Press In Tab CPU Board Plastic Mounting Post Tab Figure R39
  • 56. Repair 53 Mounting Posts Mounting Posts Figure R41 7. Slide the CPU board off the mounting posts. See Figure R41.
  • 57. 54 Repair Mounting Posts Figure R42 CPU Board Installation 1. Align the CPU board with the plastic mounting posts. Snap the right side mounting posts of the CPU board into place. See Figure R42. Do not snap the left or center mounting posts into place at this time.
  • 58. Repair 55 Figure R45 Screw Figure R44 R42 P11 P11 P8 P8 P9 P9 P10 P10 (Reference) Connect 1st (Cl-/Li+) Connect 4th (Na+) Connect 3rd (K+) Connect 2nd Figure R43 P11 AGND P9 P8 P8 P10 P9 P10 (Reference) Connect 1st (Cl-/Li+) Connect 4th (Na+) Connect 3rd (K+) Connect 2nd 2. Connect the electrode jacks to the back of the CPU board. To connect, gently push the elec- trode jack straight into the connector pin. See Figure R43 to determine the proper connector pin position on the CPU board and the recommended installation order. See Figure R44 for EasyLyte Analyzer with serial numbers below 6100. On some earlier EasyLyte analyzers below 6100, the reference electrode jack is connected to R42 in place of P11. Do not bend the connector pin when installing the electrode jack. If a pin breaks the entire CPU board may require replacement. 3. Snap the left and center mounting posts of the CPU board into place. For EasyLyte analyz- ers with serial numbers below 6100, attach the center screw to the CPU board using a #2 Phillips screw driver. See Figure R45. 6100 6100
  • 59. 56 Repair Figure R46 Probe Motor Cable (4 Pin Connector) Sensor Cable (5 Pin Connector) Pump Motor Cable (6 Pin Connector) Pump Motor Cable (6 Pin Connector) Probe Motor Cable (4 Pin Connector) Sensor Cable (5 Pin Connector) Display Cable Display Cable Tabs Power Supply Cable Printer Cable Power Supply Cable Printer Cable Figure R48 Figure R47 4. Attach the display cable to the CPU board. Push the display cable tabs to the front to secure cable in place. See Figure R46. 5. Connect the probe motor, sensor, pump motor, power supply and printer cables to the CPU board, see Figure R47. Refer to Figure R48 for EasyLyte analyzers with serial numbers below 6100. 6100
  • 60. Repair 57 Sample Detector Board Screws Figure R49 Display Cable Screws Figure R50 Display Removal 1. Detach the two short self tapping screws from the sample detector board using a #2 Phillips screw driver. See Figure R49. 2. Move and hold the display cable to expose the three display assembly screws. Detach the three short screws from the display assembly using a #2 Phillips screw driver. See Figure R50.
  • 61. 58 Repair Display Adapter Display Assembly LCD Screen Figure R51 3. Carefully remove the display assembly and the LCD screen, holding both items together. See Figure R51. Notice the orientation of the LCD screen which is extremely important when installing the display assembly. The one-line display has soldered leads located on the top of the LCD screen; the two-line display has soldered leads located on the bottom of the LCD screen. Refer to Figure R52 in the Display Installation for further details. 4. Remove the display adapter from the front housing window. Refer to Figure R51 for the loca- tion of the display adapter. For two-line displays there is an orientation notch which is important when installing the display adapter. The orientation notch is located on the top inside edge of the adapter.
  • 62. Repair 59 LCD Screen Display Installation 1. Place the display adapter into the front housing window. The smooth edge should be facing the front housing of the analyzer. For two-line displays there is an additional orientation notch. To install correctly, arrange the adapter so that the notch is located on the top inside edge of the adapter. 2. Place the foam pad on the display board above the YES/NO buttons. 3. Place the LCD screen on top of the foam pad. The orientation of the LCD screen is extreme- ly important, see Figure R52. The one-line display has soldered leads located on the top of the LCD screen, the two- line display has soldered leads located on the bottom of the LCD screen. Soldered Leads Buttons Foam Pad Figure R52 (One-line Display)
  • 63. 60 Repair Figure R54 Screws Display Assembly 4. Position the front housing so that the front of the analyzer is facing down. Align the display assembly and insert into place, as shown in Figure R53, using the YES/NO buttons as a guide. 5. Hold the display in place and attach the three screws using a #2 Phillips screw driver. See Figure R54. Figure R53
  • 64. Repair 61 Sample Detector Board Sample Detector Cable Screws Figure R55 6. Align the sample detector board so the sample detector cable is positioned on the right side. Attach the two screws lightly, using a #2 Phillips screw driver, see Figure R55. 7. Install the sample detector in the front of the analyzer to verify the sample detector board alignment. See Figure R56. When properly aligned fully tighten the sample detector screws. Sample Detector Figure R56
  • 65. 62 Repair Screws Screws Pump Motor Assembly Pump Motor Removal 1. Remove the pump tubing from the front of the analyzer. 2. Using a wire cutter, cut the tie wrap that secures the pump motor cable wires, as shown in Figure R57. 3. Detach the four screws from the front of the analyzer, using a #1 Phillips screw driver. See Figure R58. Some earlier EasyLyte analyzers may use socket screws that require a 1/16 inch Allen wrench for removal. 4. Remove the pump motor assembly, as shown in Figure R59. Pump Motor Assembly Tie Wrap Cable Figure R57 Figure R59 Figure R58
  • 66. Repair 63 Screws Screws Pump Motor Wires Pump Motor Assembly Figure R60 Figure R61 Pump Motor Installation 1. Install the pump motor assembly through the front housing, as shown in Figure R60. The pump motor wires are positioned below the pump when properly aligned. 2. Attach the four screws into the front of the analyzer, using a #2 Phillips screw driver. See Figure R61. Earlier EasyLyte analyzers may use socket screws that require a 1/16 inch Allen wrench for installation. 3. Using a tie wrap cable, secure the loose wires under the pump motor to the lead screw/probe motor wires and the sensor wires. 4. Install the pump tubing.
  • 67. 64 Repair Probe Position Scale Screw Probe Position Scale Figure R63 Figure R62 Probe Motor Assembly Removal 1. Detach the screw at the top of the probe position scale, using a #2 Phillips screw driver, as shown in Figure R62. Remove the washer. Some EasyLyte analyzers use a white spacer that should be removed also. 2. Remove the probe position scale, as shown in Figure R63.
  • 68. Repair 65 Top of Lead Screw Retaining Ring Figure R64 Actuator Tie Wrap Cable Retaining Clamp Probe Motor Cable Wires Figure R65 Figure R66 3. Remove the C-shaped retaining ring from the actuator, using the retaining ring pliers. DO NOT stretch out the retaining ring with the pliers. Disengage the actuator from the top of the lead screw assembly. See Figure R64. Save the retaining ring. 4. Cut the tie wrap using a wire cutter, see Figure R65. Remove the probe motor cable wires from the retaining clamp. See Figure R66.
  • 69. 66 Repair Probe Motor/ Lead Screw Assembly Probe Motor Bracket Figure R67 Figure R68 5. Detach the three screws from the probe motor bracket, using a #2 Phillips screw driver, as shown in Figure R67. Remove the washers. The other two screws are located on the bottom edge of the bracket. 6. Remove the Probe Motor Assembly, as shown in Figure R68.
  • 70. Repair 67 Figure R69 Screw Probe Motor Assembly Probe Motor Bracket Figure R70 Probe Motor Assembly Installation 1. Place the probe motor assembly into place. See Figure R69. 2. Attach the three washers and screws to the probe motor assembly bracket, using a #2 Phillips screw driver. See Figure R70. The other two screws are located on the bottom edge of the bracket.
  • 71. 68 Repair Lead Screw Assembly Retaining Ring Actuator Tie Wrap Cable Location Retaining Clamp Probe Motor Cable Wires Figure R72 Figure R71 3. Insert the actuator through the top of the lead screw assembly as shown in Figure R71. Install the retaining ring onto the actuator using the retaining ring pliers. DO NOT stretch out the retaining ring with the pliers. 4. Place the probe motor cable wires into the retaining clamp. See Figure R72. Secure the wires to the sensor cable wires with a tie wrap cable. The optimum location for the tie wrap cable is slightly to the left of the retaining clamp. 5. Install the alignment valve tool, part number 000104, in place of the solutions valve.
  • 72. Repair 69 6. Install the sample probe as follows. Refer to Figure R73 for details. A. With the probe hole facing out, align the probe collar with the white line on the top of the probe arm. When aligned, snap the probe into the probe arm. B. Manually move the probe arm until the sample probe hole is aligned with the top line marked on the alignment valve. When the sample probe is aligned, DO NOT move the probe arm until the Probe Motor installation is complete. A B Probe Collar aligned with Slot Probe Hole facing out and aligned with top line Sample Probe Figure R73 Probe Arm Alignment Valve
  • 73. 70 Repair Bottom of Scale Probe Position Alignment Slot Probe Position Scale Figure R74 7. Hold the sample probe against the alignment valve tool to keep in place. Install the probe position scale through the top of the housing, see Figure R74. Verify the probe position alignment slot is located on the bottom right side of the scale.
  • 74. Repair 71 Note: Alignment of Scale and Sensor Figure R75 Probe Position Screw – Keep loose until scale in align- ment 8. Place the scale into position. The scale should be seated inside the probe position sensor and inside the mounting bracket slot. See Figure R75. 9. Attach and slightly tighten the washer and screw at the top of the position scale. Do not fully tighten the screw. Some EasyLyte analyzers use a white spacer that should be attached to the back surface of the scale. 10. Align the probe position scale slot within the center of the probe position sensor. Gently move the scale to the left and visually confirm the alignment. Refer to Figure R75 for cor- rect alignment. Confirm that the sample probe hole is aligned with the top line marked on the alignment valve, on the front of the analyzer, before continuing. Probe Position Sensor Probe Scale Alignment Slot Probe Position Scale Mounting Bracket CPU Board
  • 75. 72 Repair Screw Figure R76 11. When alignment is complete, fully tighten the probe position screw into place using a #2 Phillips screw driver, as shown in Figure R76.
  • 76. Repair 73 Figure R78 Guide Rod Screw Solutions Valve Sample Probe Probe Arm Figure R77 Guide Rod Removal 1. Remove the sample probe from the probe arm and the solutions valve. See Figure R77. Remove the solutions valve. 2. Detach the screw from the bottom of the guide rod, using a #2 Phillips screw driver, as shown in Figure R78. Remove the washer.
  • 77. 74 Repair Probe Arm Guide Rod Figure R79 3. Remove the guide rod and probe arm assembly, down and away from the housing. See Figure R79.
  • 78. Repair 75 Screw Guide Rod Probe Arm Alignment Hole Guide Rod Figure R81 Guide Rod Installation 1. Install the guide rod as shown in Figure R80. To install, first insert the guide rod into the alignment hole located in the top wall of the front assembly. 2. Attach the washer and screw to the bottom of the guide rod, using a #2 Phillips screw driver. See Figure R81. Figure R80
  • 79. 76 Repair Electrode Jack Analyzer Front Housing Electrode Jack Wires Nut Driver Figure R84 Figure R83 Figure R82 Electrode Jack Removal 1. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R82. 2. Remove the hex nuts using the nut driver. See Figure R83. 3. Pull the electrode jacks from the front housing of the analyzer, as shown in Figure R84.
  • 80. Repair 77 Electrode Jack Wires Nut Driver Electrode Jack Holes Electrode Jack Analyzer Front Housing Figure R86 Figure R85 Figure R87 Electrode Jack Installation 1. Push the electrode jacks through the holes in the front of the analyzer, see Figure R85. 2. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R86. 3. Attach the hex nut using the nut driver, and tighten, as shown in Figure R87.
  • 81. EasySampler 79 Product Description The EasySampler when used with the EasyLyte analyzer provides automated batch sampling of serum, plasma or diluted urine for electrolyte analysis. The EasySampler can also perform emer- gency (STAT) analysis of blood samples on demand. The EasySampler contains a 25 position sample tray which includes a STAT and two quality con- trol positions. The tray holds either 500 μl or 2 ml sample cups. The EasyLyte analyzer software allows the operator to perform EasySampler diagnostics by sim- ply pressing the “YES” or “NO” buttons in response to prompts displayed on the analyzer LCD screen. Results are always displayed on the LCD screen and can also be documented on paper by the EasyLyte analyzer printer.
  • 82. 80 EasySampler Figure A1 - EasySampler Assemblies and Components Assembly and Component Overview The EasySampler is made up of the following mechanical assemblies and associated compo- nents. Refer to Figure A1 to determine the location of each assembly or component in relation to the EasySampler. EasySampler Housing case of the EasySampler IR Windows covers the IR beam transmitter and receiver Chassis Assembly frame which supports the mechanical assemblies PC Board Assembly contains the PC board and IR transmitter and receiver Motor Assembly powers the spindle column assembly Drive Belt connects the motor assembly to the spindle column assembly Spindle Column Assembly rotates the sample tray Sensor Assembly detects the sample tray position Sample Tray holds the samples for analysis Sample Tray EasySampler Housing Spindle Column Assembly Drive Belt Chassis Assembly IR Transmitter Motor Assembly Sensor Assembly PC Board Assembly IR Windows IR Receiver
  • 83. Replacement List The following assemblies and components can be purchased directly from Medica. Description Part No. EasySampler Housing 001074 IR Windows 001192 Chassis Assembly 001370 PC Board Assembly 001233 Motor Assembly 001328 Spindle Column Assembly 001680 Sensor Assembly 001306 Drive Belt 001197 EasySampler 81 EasySampler 81
  • 84. 82 EasySampler EasySampler Troubleshooting Guide Refer to the chart below to determine the problem and corrective action needed. Problem Description Problem Source Corrective Action Steps LCD screen displays SAFETY BEAM • Verify there is no IR obstruction “OBSTRUCTION” and OBSTRUCTION • Sensor Verification analyzer is beeping LCD screen displays “MTS” INCORRECT PROBE or • Alignment Verification “TURN TRAY HOME” or TRAY MOVEMENT • Tray Placement Verification “MTC” • Carousel Test • Tray Cycle Verification • Sensor Verification
  • 85. EasySampler 83 Alignment Verification 1. Remove and reinstall the power cord. The sample tray will rotate and then stop. 2. ALIGN PROBE/CUP? appears on the LCD screen. Press YES to verify the probe alignment. The sample probe will descend and stop above the “Daily Cleaner” cup position. 3. ALIGN CUP!/CUP ALIGNED? appears on the LCD screen; do not press YES or NO at this time. 4. Verify that the tip of the sample probe is centered over the “Daily Cleaner” cup position. Based on the inspection results, take the appropriate corrective action. • Probe Aligned: If sample probe is aligned in the proper position, press YES to the ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on the LCD screen; press YES. CALIBRATE NOW? appears on the LCD screen; press YES. Return to normal operation. • Probe Misaligned: If the probe is misaligned, proceed to STEP 5. 5. Grasp the EasySampler and analyzer at the base and carefully slide forward until the EasySampler extends over the edge of the work surface. 6. Turn the knob underneath the EasySampler base clockwise to loosen. DO NOT remove the knob. 7. Locate the hub on top of the EasySampler tray. Move the hub and the knob together, until the sample probe aligns with the “Daily Cleaner” cup position. 8. When aligned, hold the hub in place and turn the knob counterclockwise to tighten. 9. Grasp the EasySampler and analyzer at the base and carefully slide back to the original work area. 10. Press YES to the ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on the LCD screen; press YES. CALIBRATE NOW? appears on the LCD screen; press YES.
  • 86. 84 EasySampler Tray Placement Verification 1. Advance to the DAILY CLEANER? display; press YES. TRAY IN PLACE? appears on the LCD screen; press YES. 2. The sample tray should rotate and stop with the probe at cup position “12”. 3. CLEANER IN TRAY? appears on the LCD screen; press NO. 4. The sample tray should rotate and then stop with the probe at “Daily Cleaner” position. 5. Based on the test results, take the appropriate corrective action. • Correct Position: If the tray stops at the correct position, return to normal operation. • Incorrect Position: If the tray stops at an incorrect position, perform the Carousel Test.
  • 87. EasySampler 85 Carousel Test 1. Enter the QC TEST? menu and proceed to TEST CAROUSEL?. Refer to QC Test in this section for further instructions. 2. At the TEST CAROUSEL? display, press YES. 3. CALIB CAROUSEL? appears on the LCD screen; press YES. The sample tray should spin one revolution; then stop. 4. FIND HOME? appears on the LCD screen; press YES. The sample tray should spin and stop with the probe at the “Daily Cleaner” position. 5. Based on the CALIB CAROUSEL? test results, take the appropriate corrective action. • Error Message: If “MTC” or “CAROUSEL PROBLEM” appears on the LCD screen perform Sensor Verification. • No Error Message: If no error message appears on the LCD screen, perform Tray Cycle Verification.
  • 88. 86 EasySampler Tray Cycle Verification 1. Enter the QC TEST? menu and proceed to the TEST CAROUSEL? test. Refer to QC Test in this section for further instructions. 2. At the TEST CAROUSEL? display, press YES. At the CALIB CAROUSEL? display, press NO. At the FIND HOME? display, press YES. The tray should spin and stop with the sample probe at the “Daily Cleaner” position. 3. CYCLE TRAY? appears on the LCD screen; press YES. 4. The sample tray should rotate from the “Daily Cleaner” position to each cup position and back for cup positions “NORMAL through 10”. Then the sample tray should reverse direction and rotate from the “Daily Cleaner” position to each cup position and back for cup positions “11 through STAT”. 5. The sample tray will perform four complete rotation cycles as described above. 6. The analyzer will print the test results during the fourth cycle, as illustrated in the example below. Cup Number of Steps Number of Steps Position to Cup Position to Daily Cleaner 1 0097 0100 2 0196 0196 3 0292 0292 4 0388 0388 7. Verify that the number of steps between adjacent cup positions is 0096 +/-0006 for both columns of test data. 8. Based on the tray cycle test results, take the appropriate corrective action. • Error Message: If “MTC” or “CAROUSEL PROBLEM” appears on the LCD screen during testing, perform Sensor Verification. • In Specification: If tray cycle readings are in specification, exit the QC TEST? menu and return to normal operation.
  • 89. EasySampler 87 Sensor Verification 1. Remove the rear housing of the EasyLyte analyzer as detailed in Rear Housing Assembly Removal in the REPAIR SECTION. 2. Connect a voltage meter to the EasyLyte analyzer. Attach one reference lead to TP11 (DGND) and the other to U13 PIN #2 on the analyzer CPU board. 3. Manually turn the sample tray so the probe is at the “Daily Cleaner” position. 4. Verify that the measured voltage is greater than 4.50 volts. 5. Manually turn the sample tray off the “Daily Cleaner” position. 6. Verify that the measured voltage is less than 0.50 volts. 7. Based on the voltage test results, take the appropriate corrective action. • In Specification: If voltage readings are in specification, adjust the drive belt tension and repeat Tray Cycle Verification. If “MTC” or “CAROUSEL PROBLEM” errors continue, replace the spindle column assembly. Refer to Spindle Column Removal and Installation in this section for details. • Out of Specification: If voltage readings are out of specification, replace the sensor assem- bly. Refer to Sensor Removal and Installation in this section for details.
  • 90. 88 EasySampler QC Tests 1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis- play, as shown on the chart below. LCD Screen Displays Press CALIBRATE NOW? NO DAILY CLEANER? NO STANDBY MODE? NO SECOND MENU? YES * SEE RESULTS? * NO * PACK USAGE? * NO * QUAL CONTROLS? * NO * OPER FUNCTS? * NO * DIAGNOSTICS? * NO * DATE/RANGES? * YES * CHANGE DAY/TIME? * NO CAL FREQUENCY? NO Na/K/Cl or Na/K/Li RANGES? YES BLOOD Na LIMITS? YES “Numeric value” —> “Numeric value” wait 2. Two numbers appear on the screen separated by an arrow. A cursor will be positioned underneath the first digit of the first number. The second number is not applicable in this procedure. 3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the digit is correct press YES to advance to the next digit. 4. Continue until the first number is changed to 246.0. 5. Press YES when finished. QC TEST? will appear on the screen. If a value other than “246” is input the screen displays CORRECT?. Press NO to return to STEP 2 and correct any error. 6. Press YES to QC TEST?
  • 91. EasySampler 89 7. Consult the table below which lists all the QC tests and their functions. To select a test press YES. To bypass a test press NO. QC Test QC Test Function QC TEST? Access to the QC Test Functions. TURN ON MV’S? START CLOCK? Analyzer QC Test Functions. Refer to the main CHECK CAL CONSTS? TROUBLESHOOTING SECTION RAM/PROM TEST? for information on QC testing SIPPER TEST? the EasyLyte analyzer. PUMP TEST? TEST CAROUSEL? Verifies EasySampler movement functions. CALIB CAROUSEL? Moves sample tray one revolution. FIND HOME? Returns sample tray to “Daily Cleaner” position. CYCLE TRAY? Performs four complete tray cycles to each cup position. SOL’N PURGE? CALIBRATE NOW? DIAGNOSTICS? LAST CAL VALUES? ELECTR’D VALS? Analyzer QC Test Functions. FLUID FLOW TEST? SAMPLE DET. TEST? Refer to the main PUMP CAL TEST? TROUBLESHOOTING SECTION TEST VALVE? for information on QC testing RUN STATISTICS? the EasyLyte analyzer. RUN CONTROLS? TEST ALIGNMENT? BURN IN? CONTINUOUS? QC TEST? Access to the QC Test Functions. Press NO to return to the main EasyLyte menu. 6100
  • 92. 90 EasySampler Repair Overview Always consult Troubleshooting in this section first to determine the problem before attempting to service the EasySampler. Please note that complete mechanical assemblies can be replaced. Refer to the Replacement List in this section which lists replaceable assemblies and appropriate catalog numbers. Recommended Tools The following is a list of the tools required to perform repairs on the EasySampler. Please refer to this list to determine the appropriate tool and size needed. Phillips Screw Driver, #1 Phillips Screw Driver, #2 Flat Head Screw Driver, 7/32 inch Allen Wrench, 1/16 inch Open End Wrench, 1/4 inch Volt Meter
  • 93. EasySampler 91 Housing Removal 1. Turn the EasySampler upside down and gently place on a flat surface. 2. Remove the two screws from the cable retaining plate using a #2 Phillips screw driver, as shown in Figure A2. Figure A2 Screws Cable Retaining Plate
  • 94. 92 EasySampler Chasis Assembly Housing Screws Metal Base Figure A4 Figure A3 3. Remove the EasySampler cable from the groove in the bottom of the housing. 4. Remove the three screws from the metal base using a #2 Phillips screw driver, as shown in Figure A3. 5. Turn the EasySampler to the upright position. 6. Lift the housing, as shown in Figure A4, to remove from the chassis assembly.
  • 95. EasySampler 93 Chasis Assembly Figure A5 Screws Metal Base Housing Figure A6 Housing Installation 1. Align and place the housing over the chassis assembly, as shown in Figure A5. Be careful not to damage the IR receiver or transmitter on the PCB. 2. Turn the EasySampler upside down and gently place on a flat surface. 3. Attach the three screws to the metal base using a #2 Phillips screw driver, see Figure A6.
  • 96. 94 EasySampler 4. Thread the EasySampler cable into the groove in the bottom of the housing. See Figure A7. 5. Attach the two screws to the cable retaining plate using a #2 Phillips screw driver, as shown in Figure A7. Screws EasySampler Cable Cable Retaining Plate Figure A7
  • 97. EasySampler 95 PC Board Chassis Assembly Adjustment Knob Motor Cable Sensor Cable Figure A9 Figure A8 Spindle Column Removal 1. Disconnect the sensor cable and the motor cables from the PC Board. See Figure A8. 2. Turn the adjustment knob counterclockwise to loosen. Remove the adjustment knob from the bottom of the chassis assembly, as shown in Figure A9. Some EasySamplers use a spacing disk that should be removed and retained.
  • 98. 96 EasySampler Spindle Column Assembly Figure A10 3. Remove the spindle column assembly, as shown in Figure A10.
  • 99. EasySampler 97 Adjustment Knob Spindle Column Assembly Figure A12 Figure A11 Spindle Column Installation 1. Place the spindle column assembly down into place, as shown in Figure A11. 2. Attach the adjustment knob, see Figure A12, and turn clockwise to tighten. Some EasySamplers use a spacing disk that should be attached before the knob is installed.
  • 100. 98 EasySampler Motor Connector (J1) Sensor Connector (J3) Sensor Cable Motor Cable Figure A13 3. Attach the sensor cable and the motor cable to positions J1 and J3 on the PC board, see Figure A13.
  • 101. EasySampler 99 Motor Bracket Sensor and Mounting Block Screws Figure A15 Figure A14 Sensor Removal 1. Loosen the two slotted screws from the motor bracket, using a #2 Phillips screw driver, as shown in Figure A14. 2. Remove the sensor and mounting block, as shown in Figure A15. 3. Remove the two slotted screws from the mounting block, using a #2 Phillips screw driver. 4. Remove the sensor from the mounting block.
  • 102. 100 EasySampler Screws Motor Bracket Sensor and Mounting Block Figure A17 Figure A16 Sensor Installation 1. Place the sensor into the mounting block. 2. Attach the two slotted screws to the mounting block, using a #2 Phillips screw driver. 3. Place the sensor and mounting block into the motor bracket. See Figure A16. 4. Attach the two slotted screws to the motor bracket, using a #2 Phillips screw driver, as shown in Figure A17.
  • 103. EasySampler 101 Motor Shaft Damper Damper Socket Screw Figure A19 Figure A18 Motor Removal 1. Loosen the socket screw from the brass damper using a 1/16 inch Allen wrench. See Figure A18. 2. Remove the damper from the top of the motor shaft, as shown in Figure A19.
  • 104. 102 EasySampler Motor Shelf Motor Slotted Screws Motor Figure A21 Figure A20 3. Remove the four slotted screws from the top of the motor using a 7/32” flat head screw driv- er. See Figure A20. 4. Remove the motor from the motor shelf, See Figure A21.
  • 105. EasySampler 103 Belt Motor Slotted Screws Motor Gear Damper Motor Shaft Figure A23 Figure A22 Motor Installation 1. Place the motor into the motor shelf. 2. Pull the belt around the gear of the motor, see Figure A22 for location of gear and belt. 3. Attach the four slotted screws to the top of the motor using a 7/32” flat head screw driver, as shown in Figure A22. 4. Attach the damper to the top of the motor shaft, as shown in Figure A23.
  • 106. 104 EasySampler Socket Screw Damper Figure A24 5. Tighten the socket screw to the damper using a 1/16 inch Allen wrench. See Figure A24.
  • 107. EasySampler 105 Heat Sink Plate Hex Nut Power Cable Figure A26 Figure A25 PC Board Removal 1. Disconnect the power cable from the PC board. See Figure A25. 2. Remove the hex nut from the heat-sink plate using a 1/4 inch open end wrench. See Figure A26.
  • 108. 106 EasySampler PC Board Screws Screws Base Housing PC Board Figure A28 Figure A27 3. Remove the four screws from the PC board using a #2 Phillips screw driver. See Figure A27. The lower left screw may be hidden by the EMI suppressor box. If so, peel back the tape to reveal the screw. 4. Lift the PC Board, as shown in Figure A28, to remove from the base housing.
  • 109. EasySampler 107 Screws Screws Figure A30 Figure A29 PC Board Installation 1. Place the PC Board, as shown in Figure A29, onto the base housing. 2. Install the four screws to the CPU board using a #2 Phillips screw driver, see Figure A30.
  • 110. 108 EasySampler PC Board Power Connector (J2) Heat Sink Plate Hex Nut Power Cable Figure A32 Figure A31 3. Install the hex nut to the heat-sink plate using a 1/4 inch open end wrench, see Figure A31. 4. Connect the power cable to position J2 on the PC board, see Figure A32.
  • 111. Analyzer Specifications CLIA Classification: Moderate Complexity Sample: Whole Blood, Serum, Plasma or Urine. Direct (undiluted) measurement Ion Selective Electrode Sample Size: 100 μl Whole Blood, Serum, Plasma or 400 μL diluted (1:10) Urine Detection Blood Na+ : 20-200 mmol/L Urine Na+ : 40-1000 mmol/L Range: K+ : 0.2-40 mmol/L K+ : 1.0-5000 mmol/L Cl- : 25-200 mmol/L Cl - : 25-500 mmol/L Li+ : 0.2-5.0 mmol/L Li+ :(N/A) Resolution: Na+: 0.1 mmol/L K+: 0.01 mmol/L Li+: 0.01 mmol/L Cl-:0.1mmol/L Performance: Na+ K+ Li+ Cl- BLOOD, SERUM, PLASMA 80-200 mmol/L 1-10 mmol/L 0.2-2.0 mmol/L 80-200 mmol/L Reproducibility (typical) within run (n=20) C.V. <1% C.V. <2% SD≤.03 mmol/L C.V. <2% between run (10 day) C.V. <2% C.V. <2.5% SD≤.05 mmol/L C.V. <2.5% Accuracy (typical) Flame Photometer Amperometric correlation coeff 0.99 1.00 0.99 0.97 slope 0.99 1.07 1.09 0.95 URINE Reproducibility (typical) within run (n=20) C.V. <5% C.V. <5% Not C.V. <5% between run (10 day) C.V. <5% C.V. <5% Applicable C.V. <5% Accuracy (typical) Correlation coeff 1.00 1.00 Not 0.98 slope 1.05 1.00 Applicable 1.07 Analysis Time 60 sec (Blood) / 90 sec (Urine) Data Storage: 125 Patient results QC - up to 20 results Normal and Abnormal Calibration: Automatic or On-Demand Output: 32 character, 2- line alphanumeric display (or) 16 character, 1- line alphanumeric display 40 column thermal printer RS-232 Serial port (optional cable) EasySampler port continued on next page Specifications 109 Specifications 109
  • 112. 110 Specifications Analyzer Specifications (continued) Ambient Conditions: 15-32oC (60-90oF), <85% humidity Power: 100/115~VAC, 50-60 Hz, 0.8A/250 V fuse or 220~VAC, 50-60 Hz 0.4A/250 V fuse Refer to the label on the back of the analyzer for preset power and fuse settings. Size & Weight: 9.5"W x 16.5"H x 8.0"D, 13 lbs. 24 cm x 42 cm x 20 cm, 5.8 kg. EasySampler Specifications Size: Width: 12.2 inches (31.0 cm) Depth:14.0 inches (35.6 cm) Height: 8.8 inches (22.4 cm) Weight: 7.0 lbs (3.2 kg) Sample: Serum, Plasma, Urine (diluted) Throughput: 1 sample per minute Sample Cup Capacity: 500μl or 2ml Sample Tray Positions: 25 positions 21 sample 1 normal QC 1 abnormal QC 1 stat 1 Daily Cleaner position Power Source: EasyLyte analyzer
  • 113. Appendix 111 Display Messages “!!” Na/K/Cl result(s) out of analyzer measurement range. ***** Result not reported because electrode is not calibrated. ABNORMAL CONTROL? Allows operator to select analysis of abnormal control. ABNORMAL STATS? Allows operator to display or print statistics for stored abnormal control results. AIR IN CLEANER Cleaning solution not detected correctly. AIR IN CONTROL Control sample not detected correctly. AIR IN SAMPLE Sample not detected correctly. AIR IN STD A Standard A not detected correctly. AIR IN STD B Standard B not detected correctly. AIR IN FLUSH Flush not detected correctly. ALL ABNORMAL? Allows operator to delete all stored abnormal control results. ALL NORMAL? Allows operator to delete all stored normal control results. ALL RESULTS? Allows operator to display all stored results. ANALYZE BLOOD? Sample analysis mode for Blood, Serum or Plasma. The message will only appear when the analyzer is calibrated. *ANALYZE URINE?* Urine analysis mode for Na/K/Cl results only. ANALYZING... Analyzer is measuring sample. ASPIRATING... Sample is being drawn into probe. AUTOCAL? Allows operator to activate and to set time for initial daily calibration. BLOOD Cl LIMITS? Allows operator to set normal Cl range.
  • 114. 112 Appendix BLOOD K LIMITS? Allows operator to set normal K range. BLOOD Li LIMITS? Allows operator to set normal Li range. BLOOD NA LIMITS? Allows operator to set normal Na range. CAL VALUES? Allows operator to display last calibration values. CALIBRATE NOW? Allows operator to calibrate analyzer. CALIBRATING... Analyzer is calibrating. CALIBR’TING STD A Electrodes being calibrated in relation to Standard A. CALIBR’TNG STD B Electrodes being calibrated in relation to Standard B. CHANGE DAY/TIME? Allows operator to change date and time. CHANGE PACK! Occurs at 99% pack usage. Pack must be replaced! CLEANING... Analyzer is performing daily cleaning. Cl HIGH... Chloride result obtained is higher than the preset normal range (flashing display). Cl LOW... Chloride result obtained is lower than the preset normal range (flashing display). CONTINUE? Analyzer requests confirmation to proceed. C.V. VALUES? Allows operator to display % C.V.’s of Quality Control results. DAILY CLEANER? Allows operator to perform daily cleaning. DAILY SUMMARY? Allows operator to print summary of daily results. *DATE/RANGES?* Allows operator to set date, time, and ranges. DELETE CONTROLS? Allows operator to delete stored control results from memory. DELETE RESULTS? Allows operator to delete patient sample results from memory.
  • 115. Appendix 113 *DIAGNOSTICS?* Allows operator to check key analyzer functions. DILUTED 1:10? Analyzer requests confirmation that urine sample has been diluted. DRIFT Cl Chloride electrode is drifting outside of preset electrode operating range. DRIFT K Potassium electrode is drifting outside of preset electrode operating range. DRIFT Li Lithium electrode is drifting outside of preset electrode operating range. DRIFT NA Sodium electrode is drifting outside of preset electrode operating range. DRIFT, RECAL.. Analyzer is automatically recalibrating because of electrode drift. ELECTR’D VALUES? Allows operator to display real time electrode millivolt readings. ENTIRE SUMMARY? Allows operator to display summary of all results in storage. ***ERROR*** Denotes an error has occurred. Press YES or NO to determine error. FLOW TEST FAIL Unsuccessful Fluid Flow Test. FLUID FLOW OK Successful Fluid Flow Test. FLUID FLOW TEST? Allows operator to run red Test Dye to check fluid flow. FLUID PATH! Analyzer is not detecting fluid and/or air segments when expected. FLUSH? Allows operator to perform a Flush cycle on demand. FLUSHING... Analyzer is flushing electrodes. FLUSH, RECAL ... If calibration fails, analyzer does a Flush and recalibration automatically. I.D. #- - - - Allows operator to assign an identification number to patient result. K HIGH... Potassium result obtained is higher than preset normal range (flashing display). K LOW... Potassium result obtained is lower than preset normal range (flashing display). LAST CONTROL? Allows operator to display last control result.
  • 116. 114 Appendix LAST RESULT? Allows operator to display last analysis result. Li BELOW 0.2mM/L Lithium sample concentration is below EasyLyte measurement range. Li HIGH.. Lithium result obtained is higher than the preset normal range (flashing display). Li LOW... Lithium result obtained is lower than the preset normal range (flashing display). MEAN VALUES? Allows operator to display mean values of quality control results. MTP Mechanical problem with pump. MTS Mechanical problem with probe. MUST USE CLEANR! Cleaning Solution must be used before operating the analyzer. MV RANGE Cl Measured Chloride electrode voltage is outside operating limits. MV RANGE K Measured Potassium electrode voltage is outside operating limits. MV RANGE Li Measured Lithium electrode voltage is outside operating limits. MV RANGE Na Measured Sodium electrode voltage is outside operating limits. NA HIGH... Sodium result obtained is higher than the preset normal range (flashing display). NA LOW... Sodium result obtained is lower than the preset normal range (flashing display). **NA/K*** or ***Na/K/Cl*** or ***Na/K/Li*** Appears when power is connected or restored after interruption. Na/K/Cl or Na/K/Li RANGE? Allows operator to set normal ranges. NEW PACK INSTLD? Asks operator to confirm that a new pack has just been installed. NEXT SAMPLE? Allows operator to proceed to next analysis. NOISE Erratic signal coming from electrodes. NOISE, RECAL.. Analyzer automatically recalibrates after detecting noise. NORMAL CONTROL? Allows operator to select analysis of normal control.
  • 117. Appendix 115 NORMAL STATS? Allows operator to display or print statistics for stored normal control results. NOT ENOUGH DATA! In order to calculate statistics, the Analyzer requires 5 stored control results of both normal and abnormal levels. *OPER FUNCTS?* Allows operator to select “On Demand” flush, purge or calibration functions. *PACK USAGE?* Allows operator to display % pack expended. PATIENT ID#? Allows operator to assign identification number to patient’s result. % OF PACK USED Percentage of solutions pack that has been expended. PRINTER OFF? Allows operator to turn printer function on or off. PRINTING... Printer is printing requested data. PRINT RESULTS? Allows operator to print all results stored in memory. PRINT STAT’TICS? Allows operator to print normal or abnormal control summaries. PRINT SUMMARY? Prints daily summary of results. PRINT WKLY SUM? Prints summary of last five days results. PROBE IN BLOOD? Requests operator to confirm that probe is in blood. PROBE IN CLEANR? Requests operator to confirm that probe is in daily cleaning solution. PROBE IN CONTROL? Requests operator to confirm that probe is in control sample. PROBE IN DYE? Requests operator to confirm that probe is in Test Dye solution. PROBE IN URINE? Requests operator to confirm that probe is in urine. PROBE IN FLUSH? Requests operator to confirm that probe is in Flush Solution. PROBE WIPING? Allows operator time to manually wipe the probe after sample is removed. PUMP CAL... Pump is being calibrated. PUMPCAL FAILURE! Failure of Pumpcal Test.
  • 118. 116 Appendix PUMPCAL OK Successful completion of Pumpcal Test. PUMPCAL TEST? Allows operator to test pump calibration. PURGING STD A Standard A fluid line is being purged. PURGING STD B Standard B fluid line is being purged. PURGING FLUSH Flush Solution fluid line is being purged. *QUAL CONTROL? Allows operator to choose quality control functions. REMOVE CONTROL Requests operator to remove control sample from probe. REMOVE SAMPLE Requests operator to remove sample from probe. RESET TO 0%? Sets pack usage percentage counter to 0%. RUN CONTROL? Quality control analysis mode. SAMPLE CAL... Analyzer is calibrating the Sample Detector. SAMPLE DETECT OK Successful completion of Sample Detector Test. SAMPLE DETECTOR! Sample Detector readings out of range. SAMPLE DET. TEST? Allows operator to test sample detector. S.D. VALUES? Allows operator to display Standard Deviation values of Quality control results. *SECOND MENU?* Allows operator access to secondary menu. SEE MANUAL! See Operator’s Manual for assistance or explanation. *SEE RESULTS?* Allows operator to display or print stored results. SEE STATISTICS? Allows operator to display Mean, S.D., C.V. values of Quality Control results. SET CORRELATION? Allows operator to select dilution equivalent correlations. SOL’N PACK LOW! Appears when solutions pack is 88% expended. Continue operating analyzer.
  • 119. Appendix 117 SOL’N PURGE? Allows operator to purge air from fluid lines. ***STANDBY*** Appears when Analyzer is in Standby mode. STANDBY IN 1 MIN Analyzer will go into standby in 1 minute (Flashing Display). STANDBY MODE? Allows operator to put analyzer in standby mode to conserve fluids. STATISTICS? Allows operator to display statistics for all control results. STORE RESULT? Allows operator to store results in memory. SYRINGE SAMPLE? Allows operator to adjust probe extension during analysis to accommodate a syringe. TESTING... Analyzer in process of completing a Fluid Flow Test. TRY PURGE AGAIN? Allows operator to perform additional purge after initial failure. URINE Cl LIMITS? Allows operator to set normal Cl range for urine. URINE K LIMITS? Allows operator to set normal K range for urine. URINE NA LIMITS? Allows operator to set normal Na range for urine. USE EXT. FLUSH! Flush solution in pack is empty. Operator should use external flush. USER OPTIONS Allows operator to access the following functions: Probe wiping, Patient I.D. #, Syringe Sample, Set Correlation, EasySampler and Printer Off. WEEKLY SUMMARY? Allows operator to print last five days of stored results. WIPE PROBE Allows operator time to wipe off sample from exterior of probe. 400 mL PACK? Allows operator to select for correct solutions pack size. 800 mL PACK? Allows operator to select for correct solutions pack size.
  • 120. 118 QC TEST? Flowchart (Service Diagnostics) PUMPCAL TEST? DATE/RANGES? BLOOD Na LIMITS? 246.0 QC TEST? TURN ON MV’S? CHECK CAL CONSTS? RAM/PROM TEST? SIPPER TEST? PUMP TEST? TEST CAROUSEL? SOLUTION PURGE? CALIBRATE NOW? RUN CONTROLS? DIAGNOSTICS? TEST VALVE? TEST ALIGNMENT? BURN IN? QC TEST? CAL VALUES? ELECTRODES VALUES? FLUID FLOW TEST? SAMPLE DET. TEST? EASYSAMPLER 5 NORMAL AND 5 ABNORMAL CONTROLS AT CUP #1 TO #10 IN SAMPLE TRAY NORMAL CONTROL? ABNORMAL CONTROL? 8 NORMAL AND STATS 8 ABNORMAL AND STATS CALIB CAROUSEL? FIND HOME? CYCLE TRAY? CHECK SUM RELEASE DATE RAM CHECKOUT? CLEAR MEMORY QC TEST? Yes/No
  • 121. CPU Board – Analyzer Series below 6100
  • 122.
  • 124.
  • 125. Power Supply – Analyzer Series under 6100
  • 128.
  • 129.
  • 130. 001462-001 R3 EDIC@ Medica Corporation 5 Oak Park Drive Bedford, MA 01730 For Service Or Technical Assistance Call: 1-800-777-5983 (In Continental US) 1-617-275-4892 (International)