SlideShare a Scribd company logo
1 of 75
CHAPTER 15 Pressure Standards
We took then a long glass tube, which, by a
dexterous hand and the help of a lamp, was in
such a manner crooked at the bottom, that the
part turned up was almost parallel to the rest
of the tube.
Robert Boyle (1660)
No definition of pressure is really useful to the engineer
until it is translated into measurable characteristics.
Pressure standards are the basis of all pressure
measurements.
Those generally available are deadweight piston gauges,
manometers, barometers and McLeod gauges.
Each is discussed briefly as to its principle of operation, its
range of usefulness, and the more important corrections that
must be applied for its proper interpretation.
15.1 DEADWEIGHT PISTON GAUGE
Use of the deadweight free-piston gauge (Figure 15.1) for
the precise determination of steady pressures was reported as
early as 1893 by Amagat[1].
The gauge serves to define pressures in the range from 0.01
to upward of 10,000 psig, in steps as small as 0.01% of range
within a calibration uncertainty of from 0.01 to 0.05% of the
reading.
15.1.1 Principle
The gauge consists of an accurately machined piston
(sometimes honed micro-inch tolerances) that is inserted
into a close-fitting cylinder, both of known cross-sectional
areas.
In use [2]-[4] a number of masses of known weight are
first loaded on one end of the free piston. Fluid pressure is
then applied to the other end of the piston until enough
force is developed to lift the piston-weight combination.
When the piston is floating freely within the cylinder
(between limit stops), the piston gauge is in equilibrium
with the unknown system pressure and hence defines this
pressure in terms of equation (14.1) as
E
DW
E
F
P
A
=
where FE, the equivalent force of the piston-weight
combination, depends on such factors as local gravity and air
buoyancy, whereas AE, the equivalent area of the piston-
cylinder combination, depends on such factors as piston-
cylinder clearance, pressure level, and temperature. The
subscript DW indicates deadweight.
There will be fluid leakage out of the system through the
piston-cylinder clearance. Such a fluid film provides the
necessary lubrication between these two surfaces. The piston
(or, less frequently, the cylinder) is also rotated or oscillated to
reduce the friction further.
Because of fluid leakage, system pressure must be
continuously trimmed upward to keep the piston-weight
combination floating.
This is often achieved in a gas gauge by decreasing system
volume by a Boyle’s law apparatus (Figure 15.2). As long as
the piston is freely balanced, the system pressure is defined by
equation (15.1).
(a) High-pressure hydraulic gauge
FIGURE 15.1 Various deadweight piston gauges.
(b) Low-pressure gas gauge.
.
FIGURE 15.2 Pressure-volume regulator to compensate for gas
leakage in a deadweight gauge. As gas leaks, the mass and
hence the pressure decrease. As the system volume is
decreased, the pressure is reestablished according to pV=MRT.
Corrections
The two most important corrections to be applied to the
deadweight piston gauge indication pl to obtain the system
pressure of equation (15.1) concern air buoyancy and local
gravity [5].
According to Archimedes’ principle, the air displaced by
the weights and the piston exerts a buoyant force that causes
the gauge indicate too high a pressure. The correction term
for this effect is
air
tb
weights
w
C
w
 
= − ÷ ÷
 
Weights are normally given in terms of the standard gravity
value of 32.1740ft/s2
.
Whenever the gravity value differs because of latitude or
altitude variations, a gravity correction term must be applied.
It is given according to [2] and [10] as
where φ is the latitude in degrees, and h is the altitude above
sea level in feet.
( )
tan
3 8 5
1
2.637 10 cos2 9.6 10 5 10
local
g
s dard
g
C
g
hφ− − −
 
= − − ÷
 
= − × + × + ×
The corrected deadweight piston gauge pressure is given in
terms of equations (15.2) and (15.3) as
The effective area of the deadweight piston gauge is
normally taken as the mean of the cylinder and piston
areas, but temperature affects this dimension.
The effective area increases between 13 and 18 ppm/℉
for commonly used materials, and a suitable correction for
this effect may also be applied [6].
( )1DW l tb gP P C C= + +
Example 1
In Philadelphia, at a latitude of 40°N and altitude of 50 ft
above sea level, the indicated piston gauge pressure was 1000
psig. The specific weights of the ambient air and the piston
weights were 0.076 and 494 lbf/ft3
, respectively.
The dimensions of the piston and cylinder were determined
at the temperature of use (75 ) so that no temperature℉
correction was required for the effective piston gauge area.
The corrected pressure according to equations (15.2)-(15.4)
was therefore
3
8 5
0.076
1 2.637 10 cos80
1000 494
9.6 10 50 5 10
1000(0.999333) 999.333
DWP
psig
−
− −
 
− − × °− ÷=
 ÷ ÷× × − × 
= =
Monographs are often used to simplify the correction procedure
FIGURE 15.3 Monographs for temperature/air-buoyancy
correction Ctb and gravity correction Cg for deadweight gauge
measurement.
A variation on the conventional deadweight piston gauges of
Figure 15.1 is given in Figure 15.4.
Here a force balance system with a binary-coded decimal set
of deadweights is used in conjunction with two free pistons
moving in two cylinder domes.
The highly sensitive equal-arm force balance indicates when
the weights plus the reference pressure times the piston area
on one arm are precisely balanced by the system pressure
times the piston area on the other arm.
The pistons in this system are continuously rotated by
electric motors, which are integral parts of the beam balance,
thus eliminating mechanical linkages.
The corrections of equation (15.4) apply equally well to the
force balance piston gauge apparatus.
FIGURE 15.4 Equal-arm force balance piston gauge. (Source:
After [7])
15.2 MANOMETER
The manometer (Figure 15.5) was used as early as 1662 by
Boyle[8] for the precise determination of steady fluid
pressures.
Because it is founded on a basic principle of hydraulics, and
because of its inherent simplicity, the U-tube manometer
serves as a pressure standard in the range from 0.1 in of water
to 100 psig, within a calibration uncertainty of from 0.02 to
0.2% of the reading.
15.2.1 Principle
The manometer consists of a transparent tube (usually of
glass) bent or otherwise constructed in the form of an
elongated U and partially filled with a suitable liquid.
Mercury and water are the most commonly preferred
manometric fluids because detailed information is available on
their specific weights.
To measure the pressure of a fluid that is less dense than and
immiscible with the manometer fluid, it is applied to the top of
one of the tubes of the manometer while a reference fluid
pressure is applied to the other tube.
In the steady state, the difference between the unknown
pressure and the reference pressure is balanced by the weight
per unit area of the equivalent displaced manometer liquid
column according to a form of equation (14.2)
where wM, the corrected specific weight of the manometer
fluid, depends on such factors as temperature and local
gravity, and
man m EP W h∆ = ∆ (15.5)
ΔhE, the equivalent manometer fluid height, depends on such
factors as scale variations with temperature, relative specific
weights and heights of the fluids involved, and capillary
effects and is defined in greater detail by equation(15.11).
As long as the manometer fluid is in equilibrium (i.e.,
exhibiting a constant manometer fluid displacement Δhl
), the
applied pressure difference is defined by equation (15.5).
15.2.2 Corrections
The variations of specific weights of mercury and water
with temperature in the range of manometer usage are well
described by the relations
( ) 3
, 4
0.491154
/
1 1.01( 32)10
s t mercury
w lbf in
t −
=
+ −
(15.6)
where the subscript s,t signifies evaluation at the standard
gravity value and at the Fahrenheit temperature of the
manometric fluid.
These equations are the basis of the tabulations given in
Table 15.1. The specific weight called for in equation (15.5),
however, must be based on the local value of gravity.
Hence it is clear that the gravity correction term of equation
(15.3) or Figure 15.3, introduced for the deadweights of the
piston gauge, also applies to the specific weight of any fluid
involved in manometer usage according to the relation
( )
2 3 2 6 3
3
,
62.2523 0.978476 10 0.145 10 0.217 10
/
1728
s t water
t t t
w lbf in
− − −
+ × − × + ×
=
(15.7)
where wc is the corrected specific weight of any fluid of
standard specific weight wst.
Temperature gradients along the manometer can cause local
variations in the specific weight of the manometer fluid, and
these are to be avoided in any reliable pressure measurement
because of the uncertainties they necessarily introduce.
Evaporation of the manometer fluid will cause a shift in the
manometer zero, but this is easily accounted for and is not
deemed a problem.
However distillation may cause an unknown change in the
specific weight of the mixture.
( ), 1c s t gw w C= + (15.8)
The most important correction [9], [10] to be applied to the
manometer indication Δhl is that associated with the relative
specific weights and heights of the fluids involved.
According to the notation of Figure 15.5, the hydraulic
correction factor is, in general,
where all specific weights are to be corrected in accordance
with equation (15.8).
The effect of temperature on the scale calibration is not
considered significant in manometry, since these scales are
usually calibrated and used at near-room temperatures.
1 1B B A A B
h
M l M l
w h w h h
C
w h w h
         + 
= + − +   ÷  ÷ ÷  ÷
∆ ∆         
(15.9)
As for capillary effects, it is well known that the shape of the
interface between two fluids at rest depends on their relative
gravity, and on cohesion and adhesion forces between the
fluids and the containing walls.
In water-air-glass combinations, the crescent shape of the
liquid surface (called meniscus) is concave upward, and water
is said to wet the glass.
In this situation, adhesive forces dominate, and water in a
tube will be elevated by capillary action.
Conversely, for mercury-air-glass combinations, cohesive
forces dominate, the mercury meniscus is concave
downward, and the mercury level in a tube will be depressed
by capillary action (Figure 15.6).
From elementary physics, the capillary correction factor for
manometers is
where θM is the angle of contact between the manometer fluid
and the glass, σA-M and σA-M are the surface tension
coefficients of the manometer fluid M with respect to the
fluids A and B above it, and rA and rB are the radii of the tubes
containing fluids A and B.
2cos M A M B M
c
M A B
C
w r r
θ σ σ− −
 
= − ÷
 
Typical values for these capillary corrections are taken into
account, the equivalent manometer fluid height is given in
terms of equations (15.9) and (15.10) as
E h l Ch C h C∆ = ∆ ± (15.11)
15.2.3 Sign Convention
For mercury manometers, Cc
is positive when the larger
capillary effect occurs in the tube showing the larger height of
manometer fluid.
For water manometers under the same conditions, Cc
is
negative. When the same fluid is applied to both legs of the
manometer, the capillary effect is often neglected.
This can be done because the tube bores are approximately
equal in standard U-tube manometers, and hence capillarity in
one tube just counterbalances that in the other.
The capillary effect can be extremely important, however,
and must always be considered in manometer-type
instruments.
To minimize the effect of a variable meniscus, which can be
caused by the presence of dirt, the method of approaching
equilibrium, the tube bore, and so on, the tubes are always
tapped before reading, and the measured liquid height is
always based on readings taken at the center of the meniscus
in each leg of the manometer.
To reduce the capillary effect itself, the use of large-bore
tubes (over 3/8-in diameter) is most effective.
Example 2
In Denver, Colorado, at a latitude of 39°40’N, an altitude of
5380 ft above sea level, and a temperature of 76 , the℉
indicated manometer fluid height was 50 in mercury in
uniform 1/8-in bore tubing.
The reference fluid was air at atmospheric pressure, and the
higher pressure fluid was water at an elevation of 10 ft above
the water-mercury interface.
According to equations(15.3), (15.6), (15.8), the corrected
specific weight of mercury was
The corrected specific weight of water was similarly
( )
( )
( )
4
3 ' 8 5
3
0.491154
1 1.01 76 32 10
1 2.637 10 cos79 20 9.6 10 5380 5 10
0.488981 0.998945 0.488465 /
mercuryw
lbf in
−
− − −
 
= × 
+ − 
− × ° − × × − ×
= =
2 3 2 6 3
2
62.2523 0.978476 10 76 0.145 10 76 0.217 10 76
0.998945
1728
0.036026 0.998945 0.035998 /
waterw
lbf in
− − −
+ × × − × × + × ×
= ×
= × =
and, following the notation of Figure 15.5, wwater=wA.
Because of the extremely small specific weight of air wB
compared to that of the manometer fluid wM, the air column
effect in equation (15.9) is neglected. Hence according to
equations(15.9)-(15.11), the equivalent manometer fluid
height was
Now that the plus sign is used in the capillary correction
since the larger effect is on the air side, which also shows the
larger height of manometer fluid.
The corrected manometer pressure difference, according to
equation (15.5), was
( )0.488465 41.1846 20.1172manp∆ = =
( )
3 3
2 0.76600.35998 10 12 50
1 1 50
0.488465 30 0.488405
2.17 10 2.68 10
1 16 1 16
41.1846
h
in
ε
− −
−  × −    
∆ = − + +  ÷  ÷ 
     
 × ×
× − 
 
=
FIGURE 15.5 Hydraulic correction factor Ch
for generalized
manometer (capillarity neglected).
Given p1
=p2
(i.e., same level in same fluid at rest has same
pressure), then
pA
+ wA
(hA
+ hB
+Δhl
)= pB
+ wA
hB
+ wM
Δhl
(all specific weights are those corrected for temperature and
gravity), and
( ) ( ) ( ) ( ) ( ){ }1 1A B M B M B l A M A B l lP P w w w h h w w h h h h− = + ∆ − + ∆ + ∆  
Thus, in general,
If wA= wB, Ch=1-( wA/ wM) (hA/Δhl+1).
If, in addition, hA=0, Ch=1- wB/ wM.
FIGURE 15.6 Capillary effects in water and mercury.
( ) ( ) ( ) ( ){ }1 1H B M B l A M A B lC w w h h w w h h h= + ∆ − + ∆ +  
Pressure difference inside and outside tubes is zero so that
variations in liquid heights because of capillarity must be
accounted for in pressure measurements. The single-tube
correction factor is Cc
=2σcosθ/wM
r.
15.3 MICROMANOMETERS
Extending the capabilities of conventional U-tube
manometers are various types of micromanometers that serve
as pressure standards in the range from 0.0002 to 20 in of
water at pressure levels from 0 absolute to 100 psig.
Three of these micromanometer types are discussed next.
These have been chosen on the basis of simplicity of
operation.
A very complete and authoritative survey of
micromanometers is given in [11] 。
15.3.1 Prandtl type
In the Prandtl-type micromanometer (Figure 15.7), capillary
and meniscus errors are minimized by returning the meniscus
to a reference null position (within a transparent portion of the
manometer tube) before measuring the applied pressure
difference.
A reservoir, which forms one side of the manometer, is
moved vertically with respect to an inclined portion of the
tube, which forms the other side of the manometer, to achieve
the null position.
This position is reached when the meniscus falls within two
closely scribed marks on the near horizontal portion of the
micromanometer tube.
Either the reservoir or the inclined tube is moved by means
of a precision lead screw arrangement, and the
micromanometer liquid displacement Δh, corresponding to the
applied pressure difference, is determined by noting the
rotation of the lead screw on a calibrated dial.
The Prandtl-type miromanometer is generally accepted as a
pressure standard within a calibration uncertainty of 0.001 in
of water.
15.3.1 Micrometer Type
Another method of minimizing capillary and meniscus
effects in manometry is to measure liquid displacements with
micrometer heads fitted with adjustable, sharp index points
located at or near the centers of large-bore transparent tubes
that are joined at their bases to form a U.
In some commercial micromanometers [12], contact with
the surface of the manometric liquid may be sensed visually
by dimpling the surface with the index point, or by electrical
contact.
Micrometer type micromanometers also serve as pressure
standards within a calibration uncertainty of 0.001 in of water
(Figure 15.8).
15.3.3 Air Micromanometer
An extremely sensitive high-response micromanometer uses
air as its working fluid, and thus avoids all capillary and
meniscus effects usually encountered in liquid manometry.
Such an instrument has been described by Kemp [13]
(Figure 15.9). In this device the reference pressure is
mechanically amplified by centrifugal action in a rotating disk.
This disk speed is adjusted until the amplified reference
pressure just balances the unknown pressure. This null
position is recognized by observing the lack of movement of
minute oil droplets sprayed into a glass indicator tube located
between the unknown and amplified pressure lines.
At balance, the air micromanometer yields the applied
pressure difference through the relation
where ρ is the reference air density, n is the rotational speed of
the disk, and K is a constant that depends on disk radius and
annular clearance between disk and housing.
Measurements of pressure differences as small as 0.0002 in
of water can be made with this type of micromanometer,
within an uncertainty of 1%. n
2
microP K nρ∆ = (15.12)
15.3.4 Reference Pressure
A word on the reference pressure employed in manometry
is pertinent at this point in the discussion. If atmospheric
pressure is used as a reference, the manometer yields gauge
pressures.
Such pressures vary with time, altitude, latitude, and
temperature, because of the variability of air pressure (Figure
15.10).
If, however, a vacuum is used as reference, the manometer
yields absolute pressures directly, and it could serve as a
barometer (which is considered in the next section).
In any case, the obvious but important relation between
gauge and absolute pressures is
where by ambient pressure we mean pressure surrounding the
gauge.
Most often, ambient pressure is simply the atmospheric
pressure.
FIGURE 15.7 Two variations of Prandtl-type manometer.
absolute gaoge ambientP P P= + (15.13)
After application of pressure difference, either the reservoir
or the inclined tube is moved by a precision lead screw to
achieve the null position of the meniscus.
FIGURE 15.8 Micrometer-type manometer.
FIGURE 15.9 Air-type centrifugal micromanometer.
(Source: After Kemp [13])
FIGURE 15.10 Relations among terms used in pressure measurements:
—locus of constant positive gauge pressure;
——gauge pressure datum =ambient;
…locus of constant negative gauge pressures:
—.—locus of constant absolute pressure.
15.4 BAROMETER
As already indicated, the barometer was used as early as
1643 by Torricelli for the precise determination of steady
atmospheric pressures and continues today to define such
pressures within a calibration uncertainty of from 0.001 to
0.03% of the reading.
15.4.1 Principle
The cistern barometer consists of a vacuum-referred
mercury column immersed in a large-diameter ambient-vented
mercury column that serves as a reservoir (the cistern).
The most common cistern barometer in general use is the
Fortin type [after Nicolas Fortin (1750-1831)] in which the
height of the mercury surface in the cistern can be adjusted
(Figure 15.11).
The cistern in this case is essentially a leather bag supported
in a bakelite housing.
The cistern level adjustment provides a fixed zero reference
for the plated-brass mercury height scale that is adjustably
attached, but fixed at the factory during calibration, to a metal
tube.
The metal tube, in turn, is rigidly fastened to the solid parts
of the cistern assembly and, except for reading slits, surrounds
the glass tube containing the barometer mercury.
A short tube, which is movable up and down within the
first tube, carries a vernier scale and a ring used for sighting
on the mercury meniscus in the glass tube.
In use, the datum-adjusting screw is turned until the
mercury in the cistern just makes contact with the ivory
index, at which point the mercury surface is aligned with
zero on the instrument scale.
Next, the indicated height of the mercury column in the
glass tube is determined.
The lower edge of the sighting ring is lined up with the
top of the meniscus in the tube.
A scale reading and vernier reading are taken and
combined to yield the indicated mercury column height hd at
the barometer temperature t.
The atmospheric pressure exerted on the mercury in the
cistern is just balanced by the weight per unit area of the
vacuum-referred mercury column in the barometer tube
according to a form of equation (14.2), that is,
where wHg, the referred specific weight of mercury, depends on
such factors as temperature and local gravity, whereas hl0, the
referred height of mercury, depends on such factors as thermal
expansions of the scale and of mercury.
0baro Hg tP w h= (15.14)
15.4.2 Corrections
When reading mercury height in a Fortin barometer with
the scale zero adjusted to agree with the level of mercury in
the cistern, the correct height of mercury at temperature t,
called ht
, will be greater than the indicated height of mercury
at temperature t, called htI
, whenever t>ts
,where ts
is the
temperature at which the scale was calibrated.
This difference can be expressed in terms of the scale
expansion in going from ts
to t as
where S is the linear coefficient of thermal expansion of the
scale per degree [10].
( )1t tl sh h S t t= + −   (15.15)
If, as is usual, the height of mercury is desired at some
reference temperature t0, the correct height of mercury at t will
be greater than the referred height of mercury at t0, called hl0,
whenever t>ts .
This difference can be expressed in terms of the mercury
expansion in going from t0 to t as
where m is the cubical coefficient of thermal expansion of
mercury per degree.
A temperature correction factor can be defined in terms of
the indicated reading at t and the referred height at t0 as
( )0 01t th h m t t= + −   (15.16)
0t t tlC h h= − (15.17)
In terms of equations (15.15) and (15.16), this correction is
Replacement of hl of equation (15.18) by equation (15.15)
results in
When standard values of S=10.2×10-6
/℉ ,
m=101×10-6
/℉ , ts
=62 , and℉ t0
=32 are substituted in℉
equation (15.19), the result is
( ) ( )0
1 1
1 1
t t
s
C h
m t t S t t
 
= − ÷ ÷+ − + − 
(15.18)
( ) ( )
( )
0
01
s
t tl
S t t m t t
C h
m t t
 − − −
=  ÷ ÷+ − 
(15.19)
This temperature correction is zero at a barometer
temperature of 28.63 for all values of℉ htl
.
For usual values of htl
and t, the algebraically additive
temperature correction factor of equation (15.20) is presented
in Table 15.3.
The temperature correction factor of equation (15.20) can
be approximated with very little loss of accuracy as
This equation is useful for both hand and machine
calculations.
The uncertainty introduced in hr0
is always less than
( )
( )
5
1
9.08 28.63 10
1 1.01 32 10
t tl
t
C h
t
−
−
 − ×
= − ÷ ÷+ − × 
(15.20)
( ) 5
9 28.6 10tapprox tlC t h−
= − − × (15.21)
The specific weight called for in equation (15.14) must be
based on the local value of gravity and on the reference
temperature t0.
Thus once again the gravity correction term of equation
(15.3) or figure 15.3 must be applied. This time it is according
to the relation
where, from Table 15.1, ws.t0 is specifically 0.491154 lbf/in3
.
Atmospheric pressure is now obtained in straightforward
manner by combining equations (15.14), (15.17), and
(15.22).
(The gravity correction is sometimes applied instead to the
referred height of mercury, and in this role it is often
fallaciously looked on as a gravity correction to height.)
( )( ), 0 1Hg s t gw w C= + (15.22)
Several other factors that could contribute to the uncertainty
in htl
are detailed in [10] and are discussed briefly here.
These factors introduce no additional correction terms.
1 . Lighting.
Proper illumination is essential to define the location of
crown of the meniscus.
Precision meniscus sighting under optimum viewing
conditions can approach ±0.001 in.
Contact between index and mercury surface in the cistern,
judged to be made when a small dimple in the mercury first
disappears during adjustment, can be detected with proper
lighting to much better than ±0.001 in.
2 . Alignment.
Vertical alignment of the barometer tube is required for an
accurate pressure determination.
The Fortin barometer, designed to hang from a hook, does
not of itself hang vertically. This must be accomplished by
a separately supported ring encircling the cistern.
Adjustment screws control the horizontal position.
3 . Capillary effects.
Depression of the mercury column in commercial
barometers is accounted for in the initial calibration setting
at the factory, since such effects could not be applied
conveniently during use.
The quality of the barometer is largely determined by the
bore of the glass tube.
Barometers with a bore of 1/4 in are suitable for readings of
0.01 in, whereas barometers with a bore of 1/2 in are suitable
for readings of 0.002 in.
Finally, whenever the barometer is read at an elevation other
than that of the test site, an latitude correction factor must be
applied to the local absolute barometric pressure of equation
(15.14).
This is necessary because of the variation in atmospheric
pressure with elevation as expressed by the relation
site site
baro baro
air
dp
dz
w
= −∫ ∫ (15.23)
An altitude correction factor similar to Ct of equation
(15.17), which can be added directly to the local barometric
pressure, can be defied as
Using equation (15.23) and applying the realistic
isothermal assumption between barometer and test sites, and
the perfect gas relation, this is, p/ρ=RT=constant, and the
usual w=(g/gs)ρ in the lbm-lbf system, there results from
equation (15.23)
where
z site baroC P P= − (15.24)
x
site baroP P e=
( )
( )
baro site
sair
z z g
x
RT g
−
=
Thus the altitude correction factor of equation (15.24)
becomes
where z is the altitude in feet, R is the gas constant of
Table 2.1 in and T is the absolute
temperature in °R.
Example 3.
A Fortin barometer indicates 29.52 in Hg at 75.2 at Troy,℉
NY, at a latitude of 42°41’N, and at an altitude of 945 ft above
sea level. The test site is 100 ft below the barometer location.
Equations (15.20), (15.21), and Table 15.3 all agree that Ct
=-
0.124 in Hg. Equations (15.3) and (15.22) indicate that
wHg
=0.490980 lbf/in2
. Hence according to equations (15.14)
and (15.17), the corrected barometric pressure was
( )1x
t baroC P e= −
/ft lbf R lbm− ° −
( )0.490980 29.52 0.124 14.433baroP Psia= − =
The altitude correction factor to account for the test site
elevation is obtained from equation (15.25), with R=53.35
from Table 2.1, and with the factor
g/gs
=0.999646 from equation (15.3), as
The corrected site pressure is, via equation (15.24)
/ft lbf R lbm− ° −
100
14.433 exp 0.999646 1 0.05065
53.35 534.9
zC Ps
  
= × − = ÷ ×  
0.051 14.433 14.484site z baroP C P Psia= + = + =
FIGURE 15.11 Forting-type barometer.
(Source: ASME PTC 19.2[3])
15.5 McLEOD GAUGE
A special mercury-in-glass manometer was described by
McLeod [14] in 1874 for the precise determination of very low
absolute pressures of permanent gases.
Based on an elementary principle of thermodynamics
(Boyle’s law), the McLeod gauge serves as the pressure
standard in the range from 1mm Hg above absolute zero to
about 0.01μm (where 1μm=10-3
mmHg), with a calibration
uncertainty of from 0.5% above 1μm to about 3% at 0.1μm.
15.1.1 Principle
The McLeod gauge consists of glass tubing arranged so
that a sample of gas at the unknown pressure can be
trapped and then isothermally compressed by a rising
mercury column [15].
This amplifies the unknown pressure and allows
measurement by conventional manometric means.
The apparatus is illustrated in Figure 15.12.
All of the mercury is initially contained in the volume
below the cutoff level.
The McLeod gauge is first exposed to the unknown gas
pressure p1.
FIGURE 15.12 McLeod gauge.
Before gas compression takes place the mercury is contained in the
reservoir.
The cross-hatched area indicates the location of the mercury after the
trapped gas is compressed.
The mercury is then raised in tube A beyond the cutoff,
trapping a gas sample of initial volume V1=V+ahc , where a is
the area of the measuring capillary.
The mercury is continuously forced upward until it reaches
the zero level in the reference capillary B.
At this time, the mercury in the measuring capillary C
reached a level h where the gas sample is at its final volume
V2=ah, and at the final amplified manometric pressure
p2= p1+h.
The relevant equations at these pressures are
and
1 1 2 2PV PV= (15.26)
2
1
1
ah
P
V ah
=
− (15.27)
If ah<V1, as is usually the case, then
It is clear from equations (15.26)-(15.28) that the larger the
volume ratio V1
/V2
, the greater will be the magnification of the
pressure p1
and of the manometer reading h.
Hence it is desirable that measuring tube C have a small
bore.
Unfortunately for tube bores under 1mm, the compression
gain is offset by the increased reading uncertainty caused by
capillary effects [see equation (15.10)].
2
1
1
ah
P
V
= (15.28)
In fact, the reference tube B is introduced just to provide a
meaningful zero for the measuring tube.
If the zero is fixed, equation (15.28) indicates that
manometer indication h varies nonlinearly with initial pressure
p1.
A McLeod gauge with such an expanded scale at the lower
pressure will naturally exhibit a higher sensitivity in this
region.
The McLeod pressure scale, once established, serves equally
well for all the permanent gases.
15.5.1 Corrections
There are no corrections to be applied to the McLeod gauge
reading, but certain precautions should be taken [16], [17].
Moisture traps must be provided to avoid taking any
condensable vapors into the gauge because such condensable
vapors occupy a larger volume when in the vapor phase at the
initial low pressures than they occupy when in the liquid phase
at the high pressures of reading.
Thus the presence of condensable vapors always causes
pressure readings to be too low. Capillary effects,
although partly counterbalanced by using a reference
capillary, can still introduce significant uncertainties, since the
angle of contact between mercury and glass can vary
30°depending on how the mercury approaches its final
position.
Finally, since the McLeod gauge does not give continuous
readings, steady-state conditions must prevail for the
The mercury piston of the McLeod gauge can be motivated a
number of ways.
A mechanical plunger can force the mercury up tube A. A
partial vacuum over the mercury reservoir can hold the
mercury below the cutoff until the gauge is charged, and then
the mercury can be allowed to rise by bleeding dry gas into the
reservoir.
There are also several types of swivel gauges [18] in which
the mercury reservoir is located above the gauge zero during
charging.
A 90°rotation of the gauge causes mercury to rise in tube A
by the action of gravity alone.
In a variation of the McLeod principle, the gas sample is
compressed between two mercury columns, thus avoiding the
need for a reference capillary and a sealed off measuring
capillary.
A McLeod gauge with an automatic zeroing reference
capillary has also been described [9].
A summary of the characteristics of the various pressure
standards is given in Table 15.4.
15.5.3 Pressure Scales and Units
A word on pressure scales seems necessary here, since there
is indeed a confusing array of scales and units to choose from
in expressing pressures. Some of the more common include:
Conversion factors between these units are given in Table
15.5.
Two useful approximations to help sense orders of
magnitude are
Other units that have also been used to express pressures are
torr (1 torr ≈1 mmHg), pascal, deciboyle, and stress-press, to
mention but a few.
In general, equations (14.1) and (14.2) serve to relate all
these various pressure units if proper attention is given to
dimensional analysis.
For further detail on the various systems of units, the
literature should be consulted [20]-[22].
1 1000 ~ 0.04
1 0.001 ~ 0.00004
mm m inHg
m mm inHg
µ
µ
=
=
Pressure Calibrations and Standard

More Related Content

What's hot

Fluid Mechanics Pp
Fluid Mechanics PpFluid Mechanics Pp
Fluid Mechanics Pp
XU
 
Fluid mechanics prolem hour 3
Fluid mechanics prolem hour 3Fluid mechanics prolem hour 3
Fluid mechanics prolem hour 3
drguven
 
Lecture notes djj20073 t1 (1)
Lecture notes djj20073 t1 (1)Lecture notes djj20073 t1 (1)
Lecture notes djj20073 t1 (1)
AlpNjmi
 

What's hot (20)

Fluid Mechanics - Fluid Pressure and its measurement
Fluid Mechanics - Fluid Pressure and its measurementFluid Mechanics - Fluid Pressure and its measurement
Fluid Mechanics - Fluid Pressure and its measurement
 
fluid-mechanics
fluid-mechanicsfluid-mechanics
fluid-mechanics
 
fluid statics
fluid staticsfluid statics
fluid statics
 
Fluid Mechanics Pp
Fluid Mechanics PpFluid Mechanics Pp
Fluid Mechanics Pp
 
Fluid Mechanics: fluid statics: pascals law, hydrostatic law
Fluid Mechanics: fluid statics: pascals law, hydrostatic lawFluid Mechanics: fluid statics: pascals law, hydrostatic law
Fluid Mechanics: fluid statics: pascals law, hydrostatic law
 
Measurement of liq pressure
Measurement of liq pressureMeasurement of liq pressure
Measurement of liq pressure
 
Fluid mechanics prolem hour 3
Fluid mechanics prolem hour 3Fluid mechanics prolem hour 3
Fluid mechanics prolem hour 3
 
Manometer
ManometerManometer
Manometer
 
Unit 5 Fluid Mechanics
Unit 5 Fluid MechanicsUnit 5 Fluid Mechanics
Unit 5 Fluid Mechanics
 
Lecture notes djj20073 t1 (1)
Lecture notes djj20073 t1 (1)Lecture notes djj20073 t1 (1)
Lecture notes djj20073 t1 (1)
 
Static fluids
Static fluidsStatic fluids
Static fluids
 
Pressure measurement - Manometers
Pressure measurement - ManometersPressure measurement - Manometers
Pressure measurement - Manometers
 
Flow of fluids
Flow of fluidsFlow of fluids
Flow of fluids
 
Unit 3 Fluid Static
Unit 3 Fluid StaticUnit 3 Fluid Static
Unit 3 Fluid Static
 
Differential Manometers History
Differential Manometers History Differential Manometers History
Differential Manometers History
 
Hydraulic Principles & Applications
Hydraulic Principles & ApplicationsHydraulic Principles & Applications
Hydraulic Principles & Applications
 
Fluids static
Fluids staticFluids static
Fluids static
 
fluid mechanics- pressure measurement
fluid mechanics- pressure measurementfluid mechanics- pressure measurement
fluid mechanics- pressure measurement
 
Dymanics of open channel flow
Dymanics of open channel flowDymanics of open channel flow
Dymanics of open channel flow
 
Unit 1.2 mano class
Unit 1.2 mano classUnit 1.2 mano class
Unit 1.2 mano class
 

Similar to Pressure Calibrations and Standard

Flow measurement basics
Flow measurement basicsFlow measurement basics
Flow measurement basics
Salman1011
 
oil-and-gas-water-coning-in vertical-well.pdf
oil-and-gas-water-coning-in vertical-well.pdfoil-and-gas-water-coning-in vertical-well.pdf
oil-and-gas-water-coning-in vertical-well.pdf
HosnaRahman1
 
Flow around a bent duct theory
Flow around a bent duct theoryFlow around a bent duct theory
Flow around a bent duct theory
Gaurav Vaibhav
 

Similar to Pressure Calibrations and Standard (20)

7
77
7
 
flow_and_pressure.pdf
flow_and_pressure.pdfflow_and_pressure.pdf
flow_and_pressure.pdf
 
Flow visualization
Flow visualizationFlow visualization
Flow visualization
 
Pressure and Manometers
 Pressure and Manometers Pressure and Manometers
Pressure and Manometers
 
Using Computational Fluid Dynamics as a tool for improved prediction of press...
Using Computational Fluid Dynamics as a tool for improved prediction of press...Using Computational Fluid Dynamics as a tool for improved prediction of press...
Using Computational Fluid Dynamics as a tool for improved prediction of press...
 
Industrial Air Controls
Industrial Air ControlsIndustrial Air Controls
Industrial Air Controls
 
PRESSURE.pptx
PRESSURE.pptxPRESSURE.pptx
PRESSURE.pptx
 
Pressure Measurement
Pressure MeasurementPressure Measurement
Pressure Measurement
 
E1072850
E1072850E1072850
E1072850
 
International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and Development
 
Flow measurement basics
Flow measurement basicsFlow measurement basics
Flow measurement basics
 
oil-and-gas-water-coning-in vertical-well.pdf
oil-and-gas-water-coning-in vertical-well.pdfoil-and-gas-water-coning-in vertical-well.pdf
oil-and-gas-water-coning-in vertical-well.pdf
 
Pressure and Manometers
 Pressure and Manometers  Pressure and Manometers
Pressure and Manometers
 
Flow around a bent duct theory
Flow around a bent duct theoryFlow around a bent duct theory
Flow around a bent duct theory
 
Boiler Drum level measurement in Thermal Power Stations
Boiler Drum level measurement in Thermal Power StationsBoiler Drum level measurement in Thermal Power Stations
Boiler Drum level measurement in Thermal Power Stations
 
Power2010_27048_Final
Power2010_27048_FinalPower2010_27048_Final
Power2010_27048_Final
 
Axial Flow and Radial Flow.pdf
Axial Flow and Radial Flow.pdfAxial Flow and Radial Flow.pdf
Axial Flow and Radial Flow.pdf
 
HYDROSTATICS.pptx
HYDROSTATICS.pptxHYDROSTATICS.pptx
HYDROSTATICS.pptx
 
pressure sensors
pressure sensorspressure sensors
pressure sensors
 
5_Pressure_Measurement.pdf
5_Pressure_Measurement.pdf5_Pressure_Measurement.pdf
5_Pressure_Measurement.pdf
 

Recently uploaded

Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Christo Ananth
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
dollysharma2066
 
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
dharasingh5698
 
notes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.pptnotes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.ppt
MsecMca
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
dharasingh5698
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 

Recently uploaded (20)

Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
 
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdfONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Palanpur 7001035870 Whatsapp Number, 24/07 Booking
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
Unleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapUnleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leap
 
notes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.pptnotes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.ppt
 
NFPA 5000 2024 standard .
NFPA 5000 2024 standard                                  .NFPA 5000 2024 standard                                  .
NFPA 5000 2024 standard .
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
 
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdf
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torque
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 

Pressure Calibrations and Standard

  • 2. We took then a long glass tube, which, by a dexterous hand and the help of a lamp, was in such a manner crooked at the bottom, that the part turned up was almost parallel to the rest of the tube. Robert Boyle (1660)
  • 3. No definition of pressure is really useful to the engineer until it is translated into measurable characteristics. Pressure standards are the basis of all pressure measurements. Those generally available are deadweight piston gauges, manometers, barometers and McLeod gauges. Each is discussed briefly as to its principle of operation, its range of usefulness, and the more important corrections that must be applied for its proper interpretation.
  • 4. 15.1 DEADWEIGHT PISTON GAUGE Use of the deadweight free-piston gauge (Figure 15.1) for the precise determination of steady pressures was reported as early as 1893 by Amagat[1]. The gauge serves to define pressures in the range from 0.01 to upward of 10,000 psig, in steps as small as 0.01% of range within a calibration uncertainty of from 0.01 to 0.05% of the reading.
  • 5. 15.1.1 Principle The gauge consists of an accurately machined piston (sometimes honed micro-inch tolerances) that is inserted into a close-fitting cylinder, both of known cross-sectional areas. In use [2]-[4] a number of masses of known weight are first loaded on one end of the free piston. Fluid pressure is then applied to the other end of the piston until enough force is developed to lift the piston-weight combination. When the piston is floating freely within the cylinder (between limit stops), the piston gauge is in equilibrium with the unknown system pressure and hence defines this pressure in terms of equation (14.1) as E DW E F P A =
  • 6. where FE, the equivalent force of the piston-weight combination, depends on such factors as local gravity and air buoyancy, whereas AE, the equivalent area of the piston- cylinder combination, depends on such factors as piston- cylinder clearance, pressure level, and temperature. The subscript DW indicates deadweight. There will be fluid leakage out of the system through the piston-cylinder clearance. Such a fluid film provides the necessary lubrication between these two surfaces. The piston (or, less frequently, the cylinder) is also rotated or oscillated to reduce the friction further. Because of fluid leakage, system pressure must be continuously trimmed upward to keep the piston-weight combination floating.
  • 7. This is often achieved in a gas gauge by decreasing system volume by a Boyle’s law apparatus (Figure 15.2). As long as the piston is freely balanced, the system pressure is defined by equation (15.1). (a) High-pressure hydraulic gauge
  • 8. FIGURE 15.1 Various deadweight piston gauges. (b) Low-pressure gas gauge.
  • 9. . FIGURE 15.2 Pressure-volume regulator to compensate for gas leakage in a deadweight gauge. As gas leaks, the mass and hence the pressure decrease. As the system volume is decreased, the pressure is reestablished according to pV=MRT.
  • 10. Corrections The two most important corrections to be applied to the deadweight piston gauge indication pl to obtain the system pressure of equation (15.1) concern air buoyancy and local gravity [5]. According to Archimedes’ principle, the air displaced by the weights and the piston exerts a buoyant force that causes the gauge indicate too high a pressure. The correction term for this effect is air tb weights w C w   = − ÷ ÷  
  • 11. Weights are normally given in terms of the standard gravity value of 32.1740ft/s2 . Whenever the gravity value differs because of latitude or altitude variations, a gravity correction term must be applied. It is given according to [2] and [10] as where φ is the latitude in degrees, and h is the altitude above sea level in feet. ( ) tan 3 8 5 1 2.637 10 cos2 9.6 10 5 10 local g s dard g C g hφ− − −   = − − ÷   = − × + × + ×
  • 12. The corrected deadweight piston gauge pressure is given in terms of equations (15.2) and (15.3) as The effective area of the deadweight piston gauge is normally taken as the mean of the cylinder and piston areas, but temperature affects this dimension. The effective area increases between 13 and 18 ppm/℉ for commonly used materials, and a suitable correction for this effect may also be applied [6]. ( )1DW l tb gP P C C= + +
  • 13. Example 1 In Philadelphia, at a latitude of 40°N and altitude of 50 ft above sea level, the indicated piston gauge pressure was 1000 psig. The specific weights of the ambient air and the piston weights were 0.076 and 494 lbf/ft3 , respectively. The dimensions of the piston and cylinder were determined at the temperature of use (75 ) so that no temperature℉ correction was required for the effective piston gauge area. The corrected pressure according to equations (15.2)-(15.4) was therefore 3 8 5 0.076 1 2.637 10 cos80 1000 494 9.6 10 50 5 10 1000(0.999333) 999.333 DWP psig − − −   − − × °− ÷=  ÷ ÷× × − ×  = =
  • 14. Monographs are often used to simplify the correction procedure FIGURE 15.3 Monographs for temperature/air-buoyancy correction Ctb and gravity correction Cg for deadweight gauge measurement.
  • 15. A variation on the conventional deadweight piston gauges of Figure 15.1 is given in Figure 15.4. Here a force balance system with a binary-coded decimal set of deadweights is used in conjunction with two free pistons moving in two cylinder domes. The highly sensitive equal-arm force balance indicates when the weights plus the reference pressure times the piston area on one arm are precisely balanced by the system pressure times the piston area on the other arm. The pistons in this system are continuously rotated by electric motors, which are integral parts of the beam balance, thus eliminating mechanical linkages.
  • 16. The corrections of equation (15.4) apply equally well to the force balance piston gauge apparatus. FIGURE 15.4 Equal-arm force balance piston gauge. (Source: After [7])
  • 17. 15.2 MANOMETER The manometer (Figure 15.5) was used as early as 1662 by Boyle[8] for the precise determination of steady fluid pressures. Because it is founded on a basic principle of hydraulics, and because of its inherent simplicity, the U-tube manometer serves as a pressure standard in the range from 0.1 in of water to 100 psig, within a calibration uncertainty of from 0.02 to 0.2% of the reading.
  • 18. 15.2.1 Principle The manometer consists of a transparent tube (usually of glass) bent or otherwise constructed in the form of an elongated U and partially filled with a suitable liquid. Mercury and water are the most commonly preferred manometric fluids because detailed information is available on their specific weights.
  • 19. To measure the pressure of a fluid that is less dense than and immiscible with the manometer fluid, it is applied to the top of one of the tubes of the manometer while a reference fluid pressure is applied to the other tube. In the steady state, the difference between the unknown pressure and the reference pressure is balanced by the weight per unit area of the equivalent displaced manometer liquid column according to a form of equation (14.2) where wM, the corrected specific weight of the manometer fluid, depends on such factors as temperature and local gravity, and man m EP W h∆ = ∆ (15.5)
  • 20. ΔhE, the equivalent manometer fluid height, depends on such factors as scale variations with temperature, relative specific weights and heights of the fluids involved, and capillary effects and is defined in greater detail by equation(15.11). As long as the manometer fluid is in equilibrium (i.e., exhibiting a constant manometer fluid displacement Δhl ), the applied pressure difference is defined by equation (15.5). 15.2.2 Corrections The variations of specific weights of mercury and water with temperature in the range of manometer usage are well described by the relations ( ) 3 , 4 0.491154 / 1 1.01( 32)10 s t mercury w lbf in t − = + − (15.6)
  • 21. where the subscript s,t signifies evaluation at the standard gravity value and at the Fahrenheit temperature of the manometric fluid. These equations are the basis of the tabulations given in Table 15.1. The specific weight called for in equation (15.5), however, must be based on the local value of gravity. Hence it is clear that the gravity correction term of equation (15.3) or Figure 15.3, introduced for the deadweights of the piston gauge, also applies to the specific weight of any fluid involved in manometer usage according to the relation ( ) 2 3 2 6 3 3 , 62.2523 0.978476 10 0.145 10 0.217 10 / 1728 s t water t t t w lbf in − − − + × − × + × = (15.7)
  • 22. where wc is the corrected specific weight of any fluid of standard specific weight wst. Temperature gradients along the manometer can cause local variations in the specific weight of the manometer fluid, and these are to be avoided in any reliable pressure measurement because of the uncertainties they necessarily introduce. Evaporation of the manometer fluid will cause a shift in the manometer zero, but this is easily accounted for and is not deemed a problem. However distillation may cause an unknown change in the specific weight of the mixture. ( ), 1c s t gw w C= + (15.8)
  • 23.
  • 24. The most important correction [9], [10] to be applied to the manometer indication Δhl is that associated with the relative specific weights and heights of the fluids involved. According to the notation of Figure 15.5, the hydraulic correction factor is, in general, where all specific weights are to be corrected in accordance with equation (15.8). The effect of temperature on the scale calibration is not considered significant in manometry, since these scales are usually calibrated and used at near-room temperatures. 1 1B B A A B h M l M l w h w h h C w h w h          +  = + − +   ÷  ÷ ÷  ÷ ∆ ∆          (15.9)
  • 25. As for capillary effects, it is well known that the shape of the interface between two fluids at rest depends on their relative gravity, and on cohesion and adhesion forces between the fluids and the containing walls. In water-air-glass combinations, the crescent shape of the liquid surface (called meniscus) is concave upward, and water is said to wet the glass. In this situation, adhesive forces dominate, and water in a tube will be elevated by capillary action.
  • 26. Conversely, for mercury-air-glass combinations, cohesive forces dominate, the mercury meniscus is concave downward, and the mercury level in a tube will be depressed by capillary action (Figure 15.6). From elementary physics, the capillary correction factor for manometers is where θM is the angle of contact between the manometer fluid and the glass, σA-M and σA-M are the surface tension coefficients of the manometer fluid M with respect to the fluids A and B above it, and rA and rB are the radii of the tubes containing fluids A and B. 2cos M A M B M c M A B C w r r θ σ σ− −   = − ÷  
  • 27. Typical values for these capillary corrections are taken into account, the equivalent manometer fluid height is given in terms of equations (15.9) and (15.10) as E h l Ch C h C∆ = ∆ ± (15.11)
  • 28. 15.2.3 Sign Convention For mercury manometers, Cc is positive when the larger capillary effect occurs in the tube showing the larger height of manometer fluid. For water manometers under the same conditions, Cc is negative. When the same fluid is applied to both legs of the manometer, the capillary effect is often neglected. This can be done because the tube bores are approximately equal in standard U-tube manometers, and hence capillarity in one tube just counterbalances that in the other. The capillary effect can be extremely important, however, and must always be considered in manometer-type instruments.
  • 29. To minimize the effect of a variable meniscus, which can be caused by the presence of dirt, the method of approaching equilibrium, the tube bore, and so on, the tubes are always tapped before reading, and the measured liquid height is always based on readings taken at the center of the meniscus in each leg of the manometer. To reduce the capillary effect itself, the use of large-bore tubes (over 3/8-in diameter) is most effective.
  • 30. Example 2 In Denver, Colorado, at a latitude of 39°40’N, an altitude of 5380 ft above sea level, and a temperature of 76 , the℉ indicated manometer fluid height was 50 in mercury in uniform 1/8-in bore tubing. The reference fluid was air at atmospheric pressure, and the higher pressure fluid was water at an elevation of 10 ft above the water-mercury interface. According to equations(15.3), (15.6), (15.8), the corrected specific weight of mercury was
  • 31. The corrected specific weight of water was similarly ( ) ( ) ( ) 4 3 ' 8 5 3 0.491154 1 1.01 76 32 10 1 2.637 10 cos79 20 9.6 10 5380 5 10 0.488981 0.998945 0.488465 / mercuryw lbf in − − − −   = ×  + −  − × ° − × × − × = = 2 3 2 6 3 2 62.2523 0.978476 10 76 0.145 10 76 0.217 10 76 0.998945 1728 0.036026 0.998945 0.035998 / waterw lbf in − − − + × × − × × + × × = × = × = and, following the notation of Figure 15.5, wwater=wA.
  • 32. Because of the extremely small specific weight of air wB compared to that of the manometer fluid wM, the air column effect in equation (15.9) is neglected. Hence according to equations(15.9)-(15.11), the equivalent manometer fluid height was
  • 33. Now that the plus sign is used in the capillary correction since the larger effect is on the air side, which also shows the larger height of manometer fluid. The corrected manometer pressure difference, according to equation (15.5), was ( )0.488465 41.1846 20.1172manp∆ = = ( ) 3 3 2 0.76600.35998 10 12 50 1 1 50 0.488465 30 0.488405 2.17 10 2.68 10 1 16 1 16 41.1846 h in ε − − −  × −     ∆ = − + +  ÷  ÷         × × × −    =
  • 34. FIGURE 15.5 Hydraulic correction factor Ch for generalized manometer (capillarity neglected). Given p1 =p2 (i.e., same level in same fluid at rest has same pressure), then pA + wA (hA + hB +Δhl )= pB + wA hB + wM Δhl (all specific weights are those corrected for temperature and gravity), and ( ) ( ) ( ) ( ) ( ){ }1 1A B M B M B l A M A B l lP P w w w h h w w h h h h− = + ∆ − + ∆ + ∆  
  • 35. Thus, in general, If wA= wB, Ch=1-( wA/ wM) (hA/Δhl+1). If, in addition, hA=0, Ch=1- wB/ wM. FIGURE 15.6 Capillary effects in water and mercury. ( ) ( ) ( ) ( ){ }1 1H B M B l A M A B lC w w h h w w h h h= + ∆ − + ∆ +  
  • 36. Pressure difference inside and outside tubes is zero so that variations in liquid heights because of capillarity must be accounted for in pressure measurements. The single-tube correction factor is Cc =2σcosθ/wM r. 15.3 MICROMANOMETERS Extending the capabilities of conventional U-tube manometers are various types of micromanometers that serve as pressure standards in the range from 0.0002 to 20 in of water at pressure levels from 0 absolute to 100 psig. Three of these micromanometer types are discussed next. These have been chosen on the basis of simplicity of operation.
  • 37. A very complete and authoritative survey of micromanometers is given in [11] 。 15.3.1 Prandtl type In the Prandtl-type micromanometer (Figure 15.7), capillary and meniscus errors are minimized by returning the meniscus to a reference null position (within a transparent portion of the manometer tube) before measuring the applied pressure difference. A reservoir, which forms one side of the manometer, is moved vertically with respect to an inclined portion of the tube, which forms the other side of the manometer, to achieve the null position.
  • 38. This position is reached when the meniscus falls within two closely scribed marks on the near horizontal portion of the micromanometer tube. Either the reservoir or the inclined tube is moved by means of a precision lead screw arrangement, and the micromanometer liquid displacement Δh, corresponding to the applied pressure difference, is determined by noting the rotation of the lead screw on a calibrated dial. The Prandtl-type miromanometer is generally accepted as a pressure standard within a calibration uncertainty of 0.001 in of water.
  • 39. 15.3.1 Micrometer Type Another method of minimizing capillary and meniscus effects in manometry is to measure liquid displacements with micrometer heads fitted with adjustable, sharp index points located at or near the centers of large-bore transparent tubes that are joined at their bases to form a U. In some commercial micromanometers [12], contact with the surface of the manometric liquid may be sensed visually by dimpling the surface with the index point, or by electrical contact. Micrometer type micromanometers also serve as pressure standards within a calibration uncertainty of 0.001 in of water (Figure 15.8).
  • 40. 15.3.3 Air Micromanometer An extremely sensitive high-response micromanometer uses air as its working fluid, and thus avoids all capillary and meniscus effects usually encountered in liquid manometry. Such an instrument has been described by Kemp [13] (Figure 15.9). In this device the reference pressure is mechanically amplified by centrifugal action in a rotating disk. This disk speed is adjusted until the amplified reference pressure just balances the unknown pressure. This null position is recognized by observing the lack of movement of minute oil droplets sprayed into a glass indicator tube located between the unknown and amplified pressure lines.
  • 41. At balance, the air micromanometer yields the applied pressure difference through the relation where ρ is the reference air density, n is the rotational speed of the disk, and K is a constant that depends on disk radius and annular clearance between disk and housing. Measurements of pressure differences as small as 0.0002 in of water can be made with this type of micromanometer, within an uncertainty of 1%. n 2 microP K nρ∆ = (15.12)
  • 42. 15.3.4 Reference Pressure A word on the reference pressure employed in manometry is pertinent at this point in the discussion. If atmospheric pressure is used as a reference, the manometer yields gauge pressures. Such pressures vary with time, altitude, latitude, and temperature, because of the variability of air pressure (Figure 15.10). If, however, a vacuum is used as reference, the manometer yields absolute pressures directly, and it could serve as a barometer (which is considered in the next section). In any case, the obvious but important relation between gauge and absolute pressures is
  • 43. where by ambient pressure we mean pressure surrounding the gauge. Most often, ambient pressure is simply the atmospheric pressure. FIGURE 15.7 Two variations of Prandtl-type manometer. absolute gaoge ambientP P P= + (15.13)
  • 44. After application of pressure difference, either the reservoir or the inclined tube is moved by a precision lead screw to achieve the null position of the meniscus. FIGURE 15.8 Micrometer-type manometer.
  • 45. FIGURE 15.9 Air-type centrifugal micromanometer. (Source: After Kemp [13])
  • 46. FIGURE 15.10 Relations among terms used in pressure measurements: —locus of constant positive gauge pressure; ——gauge pressure datum =ambient; …locus of constant negative gauge pressures: —.—locus of constant absolute pressure.
  • 47. 15.4 BAROMETER As already indicated, the barometer was used as early as 1643 by Torricelli for the precise determination of steady atmospheric pressures and continues today to define such pressures within a calibration uncertainty of from 0.001 to 0.03% of the reading. 15.4.1 Principle The cistern barometer consists of a vacuum-referred mercury column immersed in a large-diameter ambient-vented mercury column that serves as a reservoir (the cistern). The most common cistern barometer in general use is the Fortin type [after Nicolas Fortin (1750-1831)] in which the height of the mercury surface in the cistern can be adjusted (Figure 15.11).
  • 48. The cistern in this case is essentially a leather bag supported in a bakelite housing. The cistern level adjustment provides a fixed zero reference for the plated-brass mercury height scale that is adjustably attached, but fixed at the factory during calibration, to a metal tube. The metal tube, in turn, is rigidly fastened to the solid parts of the cistern assembly and, except for reading slits, surrounds the glass tube containing the barometer mercury. A short tube, which is movable up and down within the first tube, carries a vernier scale and a ring used for sighting on the mercury meniscus in the glass tube.
  • 49. In use, the datum-adjusting screw is turned until the mercury in the cistern just makes contact with the ivory index, at which point the mercury surface is aligned with zero on the instrument scale. Next, the indicated height of the mercury column in the glass tube is determined. The lower edge of the sighting ring is lined up with the top of the meniscus in the tube. A scale reading and vernier reading are taken and combined to yield the indicated mercury column height hd at the barometer temperature t.
  • 50. The atmospheric pressure exerted on the mercury in the cistern is just balanced by the weight per unit area of the vacuum-referred mercury column in the barometer tube according to a form of equation (14.2), that is, where wHg, the referred specific weight of mercury, depends on such factors as temperature and local gravity, whereas hl0, the referred height of mercury, depends on such factors as thermal expansions of the scale and of mercury. 0baro Hg tP w h= (15.14)
  • 51. 15.4.2 Corrections When reading mercury height in a Fortin barometer with the scale zero adjusted to agree with the level of mercury in the cistern, the correct height of mercury at temperature t, called ht , will be greater than the indicated height of mercury at temperature t, called htI , whenever t>ts ,where ts is the temperature at which the scale was calibrated. This difference can be expressed in terms of the scale expansion in going from ts to t as where S is the linear coefficient of thermal expansion of the scale per degree [10]. ( )1t tl sh h S t t= + −   (15.15)
  • 52. If, as is usual, the height of mercury is desired at some reference temperature t0, the correct height of mercury at t will be greater than the referred height of mercury at t0, called hl0, whenever t>ts . This difference can be expressed in terms of the mercury expansion in going from t0 to t as where m is the cubical coefficient of thermal expansion of mercury per degree. A temperature correction factor can be defined in terms of the indicated reading at t and the referred height at t0 as ( )0 01t th h m t t= + −   (15.16) 0t t tlC h h= − (15.17)
  • 53. In terms of equations (15.15) and (15.16), this correction is Replacement of hl of equation (15.18) by equation (15.15) results in When standard values of S=10.2×10-6 /℉ , m=101×10-6 /℉ , ts =62 , and℉ t0 =32 are substituted in℉ equation (15.19), the result is ( ) ( )0 1 1 1 1 t t s C h m t t S t t   = − ÷ ÷+ − + −  (15.18) ( ) ( ) ( ) 0 01 s t tl S t t m t t C h m t t  − − − =  ÷ ÷+ −  (15.19)
  • 54. This temperature correction is zero at a barometer temperature of 28.63 for all values of℉ htl . For usual values of htl and t, the algebraically additive temperature correction factor of equation (15.20) is presented in Table 15.3. The temperature correction factor of equation (15.20) can be approximated with very little loss of accuracy as This equation is useful for both hand and machine calculations. The uncertainty introduced in hr0 is always less than ( ) ( ) 5 1 9.08 28.63 10 1 1.01 32 10 t tl t C h t − −  − × = − ÷ ÷+ − ×  (15.20) ( ) 5 9 28.6 10tapprox tlC t h− = − − × (15.21)
  • 55. The specific weight called for in equation (15.14) must be based on the local value of gravity and on the reference temperature t0. Thus once again the gravity correction term of equation (15.3) or figure 15.3 must be applied. This time it is according to the relation where, from Table 15.1, ws.t0 is specifically 0.491154 lbf/in3 . Atmospheric pressure is now obtained in straightforward manner by combining equations (15.14), (15.17), and (15.22). (The gravity correction is sometimes applied instead to the referred height of mercury, and in this role it is often fallaciously looked on as a gravity correction to height.) ( )( ), 0 1Hg s t gw w C= + (15.22)
  • 56. Several other factors that could contribute to the uncertainty in htl are detailed in [10] and are discussed briefly here. These factors introduce no additional correction terms. 1 . Lighting. Proper illumination is essential to define the location of crown of the meniscus. Precision meniscus sighting under optimum viewing conditions can approach ±0.001 in. Contact between index and mercury surface in the cistern, judged to be made when a small dimple in the mercury first disappears during adjustment, can be detected with proper lighting to much better than ±0.001 in.
  • 57. 2 . Alignment. Vertical alignment of the barometer tube is required for an accurate pressure determination. The Fortin barometer, designed to hang from a hook, does not of itself hang vertically. This must be accomplished by a separately supported ring encircling the cistern. Adjustment screws control the horizontal position. 3 . Capillary effects. Depression of the mercury column in commercial barometers is accounted for in the initial calibration setting at the factory, since such effects could not be applied conveniently during use.
  • 58. The quality of the barometer is largely determined by the bore of the glass tube. Barometers with a bore of 1/4 in are suitable for readings of 0.01 in, whereas barometers with a bore of 1/2 in are suitable for readings of 0.002 in. Finally, whenever the barometer is read at an elevation other than that of the test site, an latitude correction factor must be applied to the local absolute barometric pressure of equation (15.14). This is necessary because of the variation in atmospheric pressure with elevation as expressed by the relation site site baro baro air dp dz w = −∫ ∫ (15.23)
  • 59. An altitude correction factor similar to Ct of equation (15.17), which can be added directly to the local barometric pressure, can be defied as Using equation (15.23) and applying the realistic isothermal assumption between barometer and test sites, and the perfect gas relation, this is, p/ρ=RT=constant, and the usual w=(g/gs)ρ in the lbm-lbf system, there results from equation (15.23) where z site baroC P P= − (15.24) x site baroP P e= ( ) ( ) baro site sair z z g x RT g − =
  • 60. Thus the altitude correction factor of equation (15.24) becomes where z is the altitude in feet, R is the gas constant of Table 2.1 in and T is the absolute temperature in °R. Example 3. A Fortin barometer indicates 29.52 in Hg at 75.2 at Troy,℉ NY, at a latitude of 42°41’N, and at an altitude of 945 ft above sea level. The test site is 100 ft below the barometer location. Equations (15.20), (15.21), and Table 15.3 all agree that Ct =- 0.124 in Hg. Equations (15.3) and (15.22) indicate that wHg =0.490980 lbf/in2 . Hence according to equations (15.14) and (15.17), the corrected barometric pressure was ( )1x t baroC P e= − /ft lbf R lbm− ° − ( )0.490980 29.52 0.124 14.433baroP Psia= − =
  • 61. The altitude correction factor to account for the test site elevation is obtained from equation (15.25), with R=53.35 from Table 2.1, and with the factor g/gs =0.999646 from equation (15.3), as The corrected site pressure is, via equation (15.24) /ft lbf R lbm− ° − 100 14.433 exp 0.999646 1 0.05065 53.35 534.9 zC Ps    = × − = ÷ ×   0.051 14.433 14.484site z baroP C P Psia= + = + =
  • 62. FIGURE 15.11 Forting-type barometer. (Source: ASME PTC 19.2[3])
  • 63.
  • 64. 15.5 McLEOD GAUGE A special mercury-in-glass manometer was described by McLeod [14] in 1874 for the precise determination of very low absolute pressures of permanent gases. Based on an elementary principle of thermodynamics (Boyle’s law), the McLeod gauge serves as the pressure standard in the range from 1mm Hg above absolute zero to about 0.01μm (where 1μm=10-3 mmHg), with a calibration uncertainty of from 0.5% above 1μm to about 3% at 0.1μm.
  • 65. 15.1.1 Principle The McLeod gauge consists of glass tubing arranged so that a sample of gas at the unknown pressure can be trapped and then isothermally compressed by a rising mercury column [15]. This amplifies the unknown pressure and allows measurement by conventional manometric means. The apparatus is illustrated in Figure 15.12. All of the mercury is initially contained in the volume below the cutoff level. The McLeod gauge is first exposed to the unknown gas pressure p1.
  • 66. FIGURE 15.12 McLeod gauge. Before gas compression takes place the mercury is contained in the reservoir. The cross-hatched area indicates the location of the mercury after the trapped gas is compressed.
  • 67. The mercury is then raised in tube A beyond the cutoff, trapping a gas sample of initial volume V1=V+ahc , where a is the area of the measuring capillary. The mercury is continuously forced upward until it reaches the zero level in the reference capillary B. At this time, the mercury in the measuring capillary C reached a level h where the gas sample is at its final volume V2=ah, and at the final amplified manometric pressure p2= p1+h. The relevant equations at these pressures are and 1 1 2 2PV PV= (15.26) 2 1 1 ah P V ah = − (15.27)
  • 68. If ah<V1, as is usually the case, then It is clear from equations (15.26)-(15.28) that the larger the volume ratio V1 /V2 , the greater will be the magnification of the pressure p1 and of the manometer reading h. Hence it is desirable that measuring tube C have a small bore. Unfortunately for tube bores under 1mm, the compression gain is offset by the increased reading uncertainty caused by capillary effects [see equation (15.10)]. 2 1 1 ah P V = (15.28)
  • 69. In fact, the reference tube B is introduced just to provide a meaningful zero for the measuring tube. If the zero is fixed, equation (15.28) indicates that manometer indication h varies nonlinearly with initial pressure p1. A McLeod gauge with such an expanded scale at the lower pressure will naturally exhibit a higher sensitivity in this region. The McLeod pressure scale, once established, serves equally well for all the permanent gases. 15.5.1 Corrections There are no corrections to be applied to the McLeod gauge reading, but certain precautions should be taken [16], [17].
  • 70. Moisture traps must be provided to avoid taking any condensable vapors into the gauge because such condensable vapors occupy a larger volume when in the vapor phase at the initial low pressures than they occupy when in the liquid phase at the high pressures of reading. Thus the presence of condensable vapors always causes pressure readings to be too low. Capillary effects, although partly counterbalanced by using a reference capillary, can still introduce significant uncertainties, since the angle of contact between mercury and glass can vary 30°depending on how the mercury approaches its final position. Finally, since the McLeod gauge does not give continuous readings, steady-state conditions must prevail for the
  • 71. The mercury piston of the McLeod gauge can be motivated a number of ways. A mechanical plunger can force the mercury up tube A. A partial vacuum over the mercury reservoir can hold the mercury below the cutoff until the gauge is charged, and then the mercury can be allowed to rise by bleeding dry gas into the reservoir. There are also several types of swivel gauges [18] in which the mercury reservoir is located above the gauge zero during charging. A 90°rotation of the gauge causes mercury to rise in tube A by the action of gravity alone.
  • 72. In a variation of the McLeod principle, the gas sample is compressed between two mercury columns, thus avoiding the need for a reference capillary and a sealed off measuring capillary. A McLeod gauge with an automatic zeroing reference capillary has also been described [9]. A summary of the characteristics of the various pressure standards is given in Table 15.4.
  • 73. 15.5.3 Pressure Scales and Units A word on pressure scales seems necessary here, since there is indeed a confusing array of scales and units to choose from in expressing pressures. Some of the more common include:
  • 74. Conversion factors between these units are given in Table 15.5. Two useful approximations to help sense orders of magnitude are Other units that have also been used to express pressures are torr (1 torr ≈1 mmHg), pascal, deciboyle, and stress-press, to mention but a few. In general, equations (14.1) and (14.2) serve to relate all these various pressure units if proper attention is given to dimensional analysis. For further detail on the various systems of units, the literature should be consulted [20]-[22]. 1 1000 ~ 0.04 1 0.001 ~ 0.00004 mm m inHg m mm inHg µ µ = =