1. OCT-Angiography
By Optovue
A Quick Review
Science and Technology
Abilities
Advances
M. Taherolghalb
Technical and Application Department
Fajr Shahed Co.
2. Financial Disclosure
I have the following financial interests or
relationships to disclose:
Mehdi Taherolghalb has a significant financial
interest in Optovue a company that has a
commercial interest in the market of ophthalmic
devices.
3. About Company
Incorporated in December 2003
Located in Fermont, California, USA
Optovue Founded to develop the next generation ophthalmic
diagnosis instrument based on the Fourier-Domain OCT
technology
5. RTVue XR (Avanti)
General Specs
OCT Camera : 70,000 A-SCAN/SECOND
Optical Resolution: (in tissue)
Depth:5.0μm
Beam Spot Size: 22μm
Image Sampling Rate:
Depth: 3.0μm Digital Resolution
Scan Range:
Depth: ~3mm
Transverse: 2mm to 12mm
Scan Beam Wavelength:840±10nm
Exposure Power at pupil: 750μW maximum
Patient Interface:
Working Distance: 22mm
Motorized Focus Range: -15D to +20D
11. RTVue XR (Avanti)
Innovations
Epithelial Mapping
Keratoconus Management
• Epithelial thickness may be an important parameter
in the early detection of keratoconus diagnosis and
progression2
• ETM™ provides an efficient way to quantify epithelial
thickness and to track change between visits
Refractive Surgery Planning
• An abnormal epithelium may interfere with
Scheimpflug imaging, which could bias refractive
surgery outcomes3
• ETM provides new information on the distribution of
epithelial thickness that may improve visual
outcomes3
15. RTVue XR (Avanti)
Innovations
GCC: Gangelion Cell Complex Analysis
FLV: Focal Loss Volumes
GLV:Global Loss Volume
FLV and GLV are most reliable parameters for very early
glaucoma detection
16. RTVue XR (Avanti)
New Approach to Micro Vascular Detection
Imaging Individual Layers of the Retina in a Fast and
Non-Invasive Way
17. RTVue XR (Avanti)
What is the technology/science
behind AngioVue
Two key IP protected technologies allow for
OCTA to be used in the daily clinical practice
1) SSADA (Split Spectrum Amplitude
Decorrelation Algorithm) from OHSU (David
Huang, PhD)
2) MCT (Motion Correction Technology) from
MIT (James Fujimoto, PhD)
20. RTVue XR (Avanti)
Role of SSADA
2 repeat OCT frames
Can generate 11
decorrelation sets by
splitting the OCT spectrum
Allowing to build a 304 x 304
Angio-OCT volume in only 3 secs
21. RTVue XR (Avanti)
Role of SSADA
Without split-spectrum, 12 repeat OCT frames
would be required at each B-scan location
and a 400 kHz OCT system would be needed
to generate a 300 x 300 Angio-OCT volume in
3 seconds
Current Swept-Source OCT technology at
100KHz is then not fast enough to achieve
high-quality Angio-OCT imaging without
SSADA.
22. RTVue XR (Avanti)
Role of MCT
MCT corrects the almost inevitable saccade artifact, thereby significantly
improving the image quality and optimizing clinical interpretation.
Fast X Fast Y Merged