L.K. Jambi, J.E. Lees, S.L. Bugby, M.S. Alqahtani, B.S. Bhatia, W.R.
McKnight, N.S. Dawood, A.H. Ng and A.C. Perkins
A High-Resolution Gamma-Optical
Hybrid Camera for Medical
Imaging
12th October 2015
OUTLINES
• Hybrid Compact Gamma Camera (HCGC) design
• Performance characterisation of the HCGC
• Spatial resolution
• Uniformity
• Sensitivity
• Count rate capability
• Conclusion
12th October 2015
NUCLEAR MEDICINE – IMAGING AND TREATMENT
Whole body bone scan
multipurpose single-head LFOV gamma camera
(Nucline™ X-Ring-R (HR), -C)
12th October 2015
HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN
12th October 2015
HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN
12th October 2015
HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN
12th October 2015
HYBRID ALIGNMENT
12th October 2015
Optical and gamma images acquired by the HCGC. A cannula was used to mimic a lymphatic
vessel. The middle cannula was filled with 2MBq of 99mTc mixed with blue dye, the upper and
lower one filled with green dye only. The acquisition time was 1000 frame (~ 120sec).
A) optical image B) gamma image C) fused gamma optical image
A CB
PERFORMANCE CHARACTERISATION OF THE HCGC
• Parameters investigated:
• Spatial resolution
• Uniformity
• Sensitivity
• Count rate capability
• Developed by the Institute of Physics and Engineering in
Medicine (IPEM)
12th October 2015
SPATIAL RESOLUTION
• Intrinsic Spatial Resolution
It is the full width at half
maximum (FWHM) of a line
spread function (LSF) or of a
point spread function (PSF)
without a collimator.
12th October 2015
SPATIAL RESOLUTION
• Intrinsic Spatial Resolution
12th October 2015
Gamma image of 20MBq 99mTc point source, placed at 250mm distance from the 1mm
width slit. The acquisition time was 10000 frames (~ 1200sec).
A) 600µm B) 1500µm
SPATIAL RESOLUTION
• Intrinsic Spatial Resolution
12th October 2015
Edge response function (ERF) for 99mTc source.
BA
A) 600µm B) 1500µm
SPATIAL RESOLUTION
• Intrinsic Spatial Resolution
12th October 2015
Modulus of LSF with fitted Gaussians.
BA
A) 600µm B) 1500µm
SPATIAL RESOLUTION
• Intrinsic Spatial Resolution
12th October 2015
FWHM (mm)
600µm 0.30 ± 0.03
1500µm 0.32 ± 0.05
Better
SPATIAL RESOLUTION
• Extrinsic Spatial Resolution
The FWHM of a LSF or of a PSF with the imaging
collimator in place.
12th October 2015
SPATIAL RESOLUTION
• Extrinsic Spatial Resolution
12th October 2015
A 1.1mm diameter cannula filled with 30MBq 99mTc imaged with both 1.0mm and 0.5mm diameter
pinhole collimators. Layers of Perspex ranging from 5mm to 40mm were placed directly between the
line source and the camera face. The acquisition time was 3000 frames (~ 360sec).
600µm
1500µm
0.5mm Pinhole collimator 1.0mm Pinhole collimator
SPATIAL RESOLUTION
• Extrinsic Spatial Resolution
12th October 2015
Graphs showing the extrinsic spatial resolution vs Perspex thickness. FWHM (black square) and FWTM (red
dots) calculated for 1.1mm diameter cannula filled with 30MBq 99mTc which was used as a line source.
600µm
1500µm
0.5mm Pinhole collimator 1.0mm Pinhole collimator
SPATIAL RESOLUTION
• Extrinsic Spatial Resolution
12th October 2015
CsI scintillator Pinhole
collimator
FWHM (mm)
600µm 0.5mm 1.24 ± 0.08
1.0mm 1.94 ± 0.14
1500µm 0.5mm 1.17 ± 0.15
1.0mm 1.80 ± 0.25
Better
UNIFORMITY
• Intrinsic Spatial Uniformity
Describe the variation in counts per
pixel.
12th October 2015
INTRINSIC SPATIAL UNIFORMITY
12th October 2015
Raw flood images of a 20MBq 99mTc point source, 3mm diameter, placed at 250mm away from the un-
collimated camera to measure the spatial uniformity. The acquisition time was 30000 frames (~ 3600sec).
A) 600µm B) 1500µm
INTRINSIC SPATIAL UNIFORMITY
12th October 2015
Coefficient of variation (COV)
600µm 1.83%
1500µm 1.30%Better
SENSITIVITY
Intrinsic Spatial Sensitivity
• The proportion of
photon flux incident on
the detector that is
recorded within the
photopeak energy
window being used.
12th October 2015
SENSITIVITY
• Intrinsic Spatial Sensitivity
12th October 2015
A 20MBq 99mTc point source placed at 250mm away from the HCGC with increasing layers of Perspex
added in between to show the relationship between the intrinsic sensitivity of the HCGC and the
Perspex thickness. The acquisition time was 5000 frames (~ 600sec).
BA
A) 600µm B) 1500µm
SENSITIVITY
• Intrinsic Spatial Sensitivity
12th October 2015
Intrinsic sensitivity
600µm 16.7%
1500µm 30%Better
COUNT RATE CAPABILITY
• Count rate capability
The ability of the detector to
linearly measure counts.
12th October 2015
COUNT RATE CAPABILITY
12th October 2015
Recorded count rate capability for HCGC by using 10mm diameter point source filled with 30MBq 99mTc
placed 250mm away from the un-collimated camera. Images were taken over the course of 24hrs. The
acquisition time was 2000 frames (~ 240sec).
BA
A) 600µm B) 1500µm
COUNT RATE CAPABILITY
12th October 2015
Count rate capability
600µm 1940
1500µm 2175Better
SUMMARY
12th October 2015
Performance specification 600µm 1500µm
Intrinsic spatial resolution FWHM (mm) 0.30 ± 0.03 0.32 ± 0.05
Extrinsic spatial resolution 0.5mm pinhole collimator FWHM(mm) 1.24 ± 0.08 1.17 ± 0.15
1.0mm pinhole collimator FWHM(mm) 1.94 ± 0.14 1.80 ± 0.25
Intrinsic spatial uniformity Coefficient of variation (%) 1.83 1.30
Intrinsic sensitivity Intrinsic sensitivity (%) 16.7 30
Count rate capability Incident counts/sec 1940 2175
FUTURE WORK
Gadolinium-oxysulfide scintillator Gd2O2S
• Ceramic pixelated scintillator
• 1500µm thick
12th October 2015
A Raw flood image of a 20MBq 99mTc point source,
3mm diameter. The acquisition time was 30000
frames (~ 3600sec).
CONCLUSION
• Preclinical and clinical evaluation of the HCGC
12th October 2015
ACKNOWLEDGEMENTS
12th October 2015
• University of Leicester
Dr. John E. Lees
Sarah Bugby, Mohammed Alqahtani, Bahadar Bhatia, Numan Dawood and William R
McKnight
• University Hospitals Nottingham
Prof. Alan Perkins and A H Ng
• Leicester Royal Infirmary
Chris Hastings and David Monk
FURTHER INFORMATION
12th October 2015
• Lees, J. E., S. L. Bugby, B. S. Bhatia, L. K. Jambi, M. S. Alqahtani, W. R. McKnight, A. H. Ng
and A. C. Perkins. "A Small Field of View Camera for Hybrid Gamma and Optical Imaging."
Journal of Instrumentation 9, no. 12 (2014): C12020-C12020.
• Bhatia, B.S., S.L. Bugby, J.E. Lees, and A.C. Perkins, "A scheme for assessing the
performance characteristics of small field-of-view gamma cameras". Phys Med, 2015. 31(1): p.
98-103.
• Bugby, S.L., J.E. Lees, B.S. Bhatia, and A.C. Perkins, "Characterisation of a high resolution
small field of view portable gamma camera". Phys Med, 2014. 30(3): p. 331-9.
Contact
Layal Jambi
Lj97@le.ac.uk
PRELIMINARY EVALUATION
12th October 2015
Performance
specification
HCGC Typical
LFOV
600µm 1500µm
Field of view Nominal (mm) 40 x 40 210 Ø
Intrinsic spatial resolution FWHM (mm) 0.30 ± 0.03 0.32 ± 0.05 2.7
Extrinsic spatial resolution 0.5mm pinhole collimator FWHM(mm) 1.24 ± 0.08 1.17 ± 0.15
7.8
1.0mm pinhole collimator FWHM(mm) 1.94 ± 0.14 1.80 ± 0.25
Intrinsic spatial uniformity Coefficient of variation 1.83 1.30 x
Intrinsic sensitivity Intrinsic sensitivity (%) 16.7 30 x
Count rate capability Incident counts/sec 1940 2175 25000

More Related Content

Similar to KFMC Presentation - Layal

13.45 o15 n winch
13.45 o15 n winch13.45 o15 n winch
13.45 o15 n winch
NZIP
 
Ct physics – II
Ct physics – IICt physics – II
Ct physics – II
Archana Koshy
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
nivedhanive8
 
ECR2016_C-2206- Layal
ECR2016_C-2206- LayalECR2016_C-2206- Layal
ECR2016_C-2206- Layal
Layal Jambi
 
Task 239766J.pptx
Task 239766J.pptxTask 239766J.pptx
Task 239766J.pptx
TaliyowAlow
 
Layal's 1st year Presentation
Layal's 1st year PresentationLayal's 1st year Presentation
Layal's 1st year Presentation
Layal Jambi
 
11
1111
Comparison of columnar and pixelated scintillators for small field of view hy...
Comparison of columnar and pixelated scintillators for small field of view hy...Comparison of columnar and pixelated scintillators for small field of view hy...
Comparison of columnar and pixelated scintillators for small field of view hy...
Layal Jambi
 
Automatic reading cr39
Automatic reading cr39Automatic reading cr39
Automatic reading cr39
MOAYYAD ALSSABBAGH
 
2016-09-07-IDEAS-presentation
2016-09-07-IDEAS-presentation2016-09-07-IDEAS-presentation
2016-09-07-IDEAS-presentation
Gunnar Maehlum
 
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
IOSR Journals
 
Laser feedback interferometry.pdf
Laser feedback interferometry.pdfLaser feedback interferometry.pdf
Laser feedback interferometry.pdf
bitconcepts
 
Prof. Mrochen "Clinical results with UV-X" (2014)
Prof. Mrochen "Clinical results with UV-X" (2014)Prof. Mrochen "Clinical results with UV-X" (2014)
Prof. Mrochen "Clinical results with UV-X" (2014)
Mediphacos
 
Sparse and Redundant Representations: Theory and Applications
Sparse and Redundant Representations: Theory and ApplicationsSparse and Redundant Representations: Theory and Applications
Sparse and Redundant Representations: Theory and Applications
Distinguished Lecturer Series - Leon The Mathematician
 
Material selection for microcantilever.
Material selection for microcantilever.Material selection for microcantilever.
Material selection for microcantilever.
Shubham Jain
 
Image receptors for radiology seminar
Image receptors   for radiology seminarImage receptors   for radiology seminar
Image receptors for radiology seminar
Puneet Choudhary
 
Computed tomography / dental implant courses
Computed tomography / dental implant coursesComputed tomography / dental implant courses
Computed tomography / dental implant courses
Indian dental academy
 
CBCT in orthodontics
CBCT in orthodontics CBCT in orthodontics
CBCT in orthodontics
DrAkshiSharma
 
Multi spectral imaging
Multi spectral imagingMulti spectral imaging
Multi spectral imaging
Aalaa Khattab
 
image processing
image processingimage processing
image processing
Gayathiri Murugesan
 

Similar to KFMC Presentation - Layal (20)

13.45 o15 n winch
13.45 o15 n winch13.45 o15 n winch
13.45 o15 n winch
 
Ct physics – II
Ct physics – IICt physics – II
Ct physics – II
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
ECR2016_C-2206- Layal
ECR2016_C-2206- LayalECR2016_C-2206- Layal
ECR2016_C-2206- Layal
 
Task 239766J.pptx
Task 239766J.pptxTask 239766J.pptx
Task 239766J.pptx
 
Layal's 1st year Presentation
Layal's 1st year PresentationLayal's 1st year Presentation
Layal's 1st year Presentation
 
11
1111
11
 
Comparison of columnar and pixelated scintillators for small field of view hy...
Comparison of columnar and pixelated scintillators for small field of view hy...Comparison of columnar and pixelated scintillators for small field of view hy...
Comparison of columnar and pixelated scintillators for small field of view hy...
 
Automatic reading cr39
Automatic reading cr39Automatic reading cr39
Automatic reading cr39
 
2016-09-07-IDEAS-presentation
2016-09-07-IDEAS-presentation2016-09-07-IDEAS-presentation
2016-09-07-IDEAS-presentation
 
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...Determination ofRadiological Quality Parameters Using Optical Densitometer an...
Determination ofRadiological Quality Parameters Using Optical Densitometer an...
 
Laser feedback interferometry.pdf
Laser feedback interferometry.pdfLaser feedback interferometry.pdf
Laser feedback interferometry.pdf
 
Prof. Mrochen "Clinical results with UV-X" (2014)
Prof. Mrochen "Clinical results with UV-X" (2014)Prof. Mrochen "Clinical results with UV-X" (2014)
Prof. Mrochen "Clinical results with UV-X" (2014)
 
Sparse and Redundant Representations: Theory and Applications
Sparse and Redundant Representations: Theory and ApplicationsSparse and Redundant Representations: Theory and Applications
Sparse and Redundant Representations: Theory and Applications
 
Material selection for microcantilever.
Material selection for microcantilever.Material selection for microcantilever.
Material selection for microcantilever.
 
Image receptors for radiology seminar
Image receptors   for radiology seminarImage receptors   for radiology seminar
Image receptors for radiology seminar
 
Computed tomography / dental implant courses
Computed tomography / dental implant coursesComputed tomography / dental implant courses
Computed tomography / dental implant courses
 
CBCT in orthodontics
CBCT in orthodontics CBCT in orthodontics
CBCT in orthodontics
 
Multi spectral imaging
Multi spectral imagingMulti spectral imaging
Multi spectral imaging
 
image processing
image processingimage processing
image processing
 

More from Layal Jambi

A hand-held hybrid gamma-near-infrared fluorescence imaging camera
A hand-held hybrid gamma-near-infrared fluorescence imaging cameraA hand-held hybrid gamma-near-infrared fluorescence imaging camera
A hand-held hybrid gamma-near-infrared fluorescence imaging camera
Layal Jambi
 
Festival of PGR
Festival of PGRFestival of PGR
Festival of PGR
Layal Jambi
 
3MT_Layal_final
3MT_Layal_final3MT_Layal_final
3MT_Layal_final
Layal Jambi
 
قسم العلوم الاشعاعية
قسم العلوم الاشعاعيةقسم العلوم الاشعاعية
قسم العلوم الاشعاعيةLayal Jambi
 
EUS3
EUS3EUS3
EUS2_How the Harmful Effects of Radiation can Affect Human Health
EUS2_How the Harmful Effects of Radiation can Affect Human HealthEUS2_How the Harmful Effects of Radiation can Affect Human Health
EUS2_How the Harmful Effects of Radiation can Affect Human Health
Layal Jambi
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetry
Layal Jambi
 
Optimisation of X-Ray CT within SPECTCT Studies
Optimisation of X-Ray CT within SPECTCT StudiesOptimisation of X-Ray CT within SPECTCT Studies
Optimisation of X-Ray CT within SPECTCT Studies
Layal Jambi
 
BIR_Layal_Final
BIR_Layal_FinalBIR_Layal_Final
BIR_Layal_Final
Layal Jambi
 
Liver Ca_Layal
Liver Ca_LayalLiver Ca_Layal
Liver Ca_Layal
Layal Jambi
 

More from Layal Jambi (10)

A hand-held hybrid gamma-near-infrared fluorescence imaging camera
A hand-held hybrid gamma-near-infrared fluorescence imaging cameraA hand-held hybrid gamma-near-infrared fluorescence imaging camera
A hand-held hybrid gamma-near-infrared fluorescence imaging camera
 
Festival of PGR
Festival of PGRFestival of PGR
Festival of PGR
 
3MT_Layal_final
3MT_Layal_final3MT_Layal_final
3MT_Layal_final
 
قسم العلوم الاشعاعية
قسم العلوم الاشعاعيةقسم العلوم الاشعاعية
قسم العلوم الاشعاعية
 
EUS3
EUS3EUS3
EUS3
 
EUS2_How the Harmful Effects of Radiation can Affect Human Health
EUS2_How the Harmful Effects of Radiation can Affect Human HealthEUS2_How the Harmful Effects of Radiation can Affect Human Health
EUS2_How the Harmful Effects of Radiation can Affect Human Health
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetry
 
Optimisation of X-Ray CT within SPECTCT Studies
Optimisation of X-Ray CT within SPECTCT StudiesOptimisation of X-Ray CT within SPECTCT Studies
Optimisation of X-Ray CT within SPECTCT Studies
 
BIR_Layal_Final
BIR_Layal_FinalBIR_Layal_Final
BIR_Layal_Final
 
Liver Ca_Layal
Liver Ca_LayalLiver Ca_Layal
Liver Ca_Layal
 

KFMC Presentation - Layal

  • 1. L.K. Jambi, J.E. Lees, S.L. Bugby, M.S. Alqahtani, B.S. Bhatia, W.R. McKnight, N.S. Dawood, A.H. Ng and A.C. Perkins A High-Resolution Gamma-Optical Hybrid Camera for Medical Imaging 12th October 2015
  • 2. OUTLINES • Hybrid Compact Gamma Camera (HCGC) design • Performance characterisation of the HCGC • Spatial resolution • Uniformity • Sensitivity • Count rate capability • Conclusion 12th October 2015
  • 3. NUCLEAR MEDICINE – IMAGING AND TREATMENT Whole body bone scan multipurpose single-head LFOV gamma camera (Nucline™ X-Ring-R (HR), -C) 12th October 2015
  • 4. HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN 12th October 2015
  • 5. HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN 12th October 2015
  • 6. HYBRID COMPACT GAMMA CAMERA (HCGC) DESIGN 12th October 2015
  • 7. HYBRID ALIGNMENT 12th October 2015 Optical and gamma images acquired by the HCGC. A cannula was used to mimic a lymphatic vessel. The middle cannula was filled with 2MBq of 99mTc mixed with blue dye, the upper and lower one filled with green dye only. The acquisition time was 1000 frame (~ 120sec). A) optical image B) gamma image C) fused gamma optical image A CB
  • 8. PERFORMANCE CHARACTERISATION OF THE HCGC • Parameters investigated: • Spatial resolution • Uniformity • Sensitivity • Count rate capability • Developed by the Institute of Physics and Engineering in Medicine (IPEM) 12th October 2015
  • 9. SPATIAL RESOLUTION • Intrinsic Spatial Resolution It is the full width at half maximum (FWHM) of a line spread function (LSF) or of a point spread function (PSF) without a collimator. 12th October 2015
  • 10. SPATIAL RESOLUTION • Intrinsic Spatial Resolution 12th October 2015 Gamma image of 20MBq 99mTc point source, placed at 250mm distance from the 1mm width slit. The acquisition time was 10000 frames (~ 1200sec). A) 600µm B) 1500µm
  • 11. SPATIAL RESOLUTION • Intrinsic Spatial Resolution 12th October 2015 Edge response function (ERF) for 99mTc source. BA A) 600µm B) 1500µm
  • 12. SPATIAL RESOLUTION • Intrinsic Spatial Resolution 12th October 2015 Modulus of LSF with fitted Gaussians. BA A) 600µm B) 1500µm
  • 13. SPATIAL RESOLUTION • Intrinsic Spatial Resolution 12th October 2015 FWHM (mm) 600µm 0.30 ± 0.03 1500µm 0.32 ± 0.05 Better
  • 14. SPATIAL RESOLUTION • Extrinsic Spatial Resolution The FWHM of a LSF or of a PSF with the imaging collimator in place. 12th October 2015
  • 15. SPATIAL RESOLUTION • Extrinsic Spatial Resolution 12th October 2015 A 1.1mm diameter cannula filled with 30MBq 99mTc imaged with both 1.0mm and 0.5mm diameter pinhole collimators. Layers of Perspex ranging from 5mm to 40mm were placed directly between the line source and the camera face. The acquisition time was 3000 frames (~ 360sec). 600µm 1500µm 0.5mm Pinhole collimator 1.0mm Pinhole collimator
  • 16. SPATIAL RESOLUTION • Extrinsic Spatial Resolution 12th October 2015 Graphs showing the extrinsic spatial resolution vs Perspex thickness. FWHM (black square) and FWTM (red dots) calculated for 1.1mm diameter cannula filled with 30MBq 99mTc which was used as a line source. 600µm 1500µm 0.5mm Pinhole collimator 1.0mm Pinhole collimator
  • 17. SPATIAL RESOLUTION • Extrinsic Spatial Resolution 12th October 2015 CsI scintillator Pinhole collimator FWHM (mm) 600µm 0.5mm 1.24 ± 0.08 1.0mm 1.94 ± 0.14 1500µm 0.5mm 1.17 ± 0.15 1.0mm 1.80 ± 0.25 Better
  • 18. UNIFORMITY • Intrinsic Spatial Uniformity Describe the variation in counts per pixel. 12th October 2015
  • 19. INTRINSIC SPATIAL UNIFORMITY 12th October 2015 Raw flood images of a 20MBq 99mTc point source, 3mm diameter, placed at 250mm away from the un- collimated camera to measure the spatial uniformity. The acquisition time was 30000 frames (~ 3600sec). A) 600µm B) 1500µm
  • 20. INTRINSIC SPATIAL UNIFORMITY 12th October 2015 Coefficient of variation (COV) 600µm 1.83% 1500µm 1.30%Better
  • 21. SENSITIVITY Intrinsic Spatial Sensitivity • The proportion of photon flux incident on the detector that is recorded within the photopeak energy window being used. 12th October 2015
  • 22. SENSITIVITY • Intrinsic Spatial Sensitivity 12th October 2015 A 20MBq 99mTc point source placed at 250mm away from the HCGC with increasing layers of Perspex added in between to show the relationship between the intrinsic sensitivity of the HCGC and the Perspex thickness. The acquisition time was 5000 frames (~ 600sec). BA A) 600µm B) 1500µm
  • 23. SENSITIVITY • Intrinsic Spatial Sensitivity 12th October 2015 Intrinsic sensitivity 600µm 16.7% 1500µm 30%Better
  • 24. COUNT RATE CAPABILITY • Count rate capability The ability of the detector to linearly measure counts. 12th October 2015
  • 25. COUNT RATE CAPABILITY 12th October 2015 Recorded count rate capability for HCGC by using 10mm diameter point source filled with 30MBq 99mTc placed 250mm away from the un-collimated camera. Images were taken over the course of 24hrs. The acquisition time was 2000 frames (~ 240sec). BA A) 600µm B) 1500µm
  • 26. COUNT RATE CAPABILITY 12th October 2015 Count rate capability 600µm 1940 1500µm 2175Better
  • 27. SUMMARY 12th October 2015 Performance specification 600µm 1500µm Intrinsic spatial resolution FWHM (mm) 0.30 ± 0.03 0.32 ± 0.05 Extrinsic spatial resolution 0.5mm pinhole collimator FWHM(mm) 1.24 ± 0.08 1.17 ± 0.15 1.0mm pinhole collimator FWHM(mm) 1.94 ± 0.14 1.80 ± 0.25 Intrinsic spatial uniformity Coefficient of variation (%) 1.83 1.30 Intrinsic sensitivity Intrinsic sensitivity (%) 16.7 30 Count rate capability Incident counts/sec 1940 2175
  • 28. FUTURE WORK Gadolinium-oxysulfide scintillator Gd2O2S • Ceramic pixelated scintillator • 1500µm thick 12th October 2015 A Raw flood image of a 20MBq 99mTc point source, 3mm diameter. The acquisition time was 30000 frames (~ 3600sec).
  • 29. CONCLUSION • Preclinical and clinical evaluation of the HCGC 12th October 2015
  • 30. ACKNOWLEDGEMENTS 12th October 2015 • University of Leicester Dr. John E. Lees Sarah Bugby, Mohammed Alqahtani, Bahadar Bhatia, Numan Dawood and William R McKnight • University Hospitals Nottingham Prof. Alan Perkins and A H Ng • Leicester Royal Infirmary Chris Hastings and David Monk
  • 31. FURTHER INFORMATION 12th October 2015 • Lees, J. E., S. L. Bugby, B. S. Bhatia, L. K. Jambi, M. S. Alqahtani, W. R. McKnight, A. H. Ng and A. C. Perkins. "A Small Field of View Camera for Hybrid Gamma and Optical Imaging." Journal of Instrumentation 9, no. 12 (2014): C12020-C12020. • Bhatia, B.S., S.L. Bugby, J.E. Lees, and A.C. Perkins, "A scheme for assessing the performance characteristics of small field-of-view gamma cameras". Phys Med, 2015. 31(1): p. 98-103. • Bugby, S.L., J.E. Lees, B.S. Bhatia, and A.C. Perkins, "Characterisation of a high resolution small field of view portable gamma camera". Phys Med, 2014. 30(3): p. 331-9. Contact Layal Jambi Lj97@le.ac.uk
  • 32. PRELIMINARY EVALUATION 12th October 2015 Performance specification HCGC Typical LFOV 600µm 1500µm Field of view Nominal (mm) 40 x 40 210 Ø Intrinsic spatial resolution FWHM (mm) 0.30 ± 0.03 0.32 ± 0.05 2.7 Extrinsic spatial resolution 0.5mm pinhole collimator FWHM(mm) 1.24 ± 0.08 1.17 ± 0.15 7.8 1.0mm pinhole collimator FWHM(mm) 1.94 ± 0.14 1.80 ± 0.25 Intrinsic spatial uniformity Coefficient of variation 1.83 1.30 x Intrinsic sensitivity Intrinsic sensitivity (%) 16.7 30 x Count rate capability Incident counts/sec 1940 2175 25000