SlideShare a Scribd company logo
1 of 33
LECTURE TWO
CHM 2203
BY
Godfrey Muhwezi
INSTRUMENTAL CHEMICAL ANALYSIS
Tuesday, February 27, 2024 1
Outline of lecture Two
Figures of merit
Calibrations
Standardizations
Blank correction
2
Tuesday, February 27, 2024
Figures of merit
• A figure of merit is a numerical expression
representing the efficiency of a given instrument or
procedure. Figures of merit are a way for analytical
chemists to characterize a method. The six figures
of merit we use are;
Precision-A measure of the reproducibility(Relative
SD, Absolute SD)
Accuracy-How close a measured value is to the
true value
 Sensitivity-A measure of the methods ability to
distinguish between two samples reported as
change in signal per unit change in the amount of
analyte
Tuesday, February 27, 2024 3
Figures of merit cont’d
Linear dynamic range-Concentration range-(range
over which measurements can be made)
 Detection limit -smallest amount of analyte that
can be determined with confidence or the smallest
amount of analyte that has a signal larger than the
signal from the reagent blank
Selectivity-A measure of a methods freedom from
interferences
Tuesday, February 27, 2024 4
Calibrations
• Calibration is defined as the act of making sure that a
scientific process or instrument will produce results
which are accurate. In more complex terms, calibration
is the act which determines the functional relationship
between measured values and analytical quantities.
Tuesday, February 27, 2024 5
Calibration cont’d
•The goal of calibration is to minimize any
measurement uncertainty by ensuring the
accuracy of test equipment. Calibration
quantifies and controls errors or uncertainties
within measurement processes to an acceptable
level.
Tuesday, February 27, 2024 6
When do you calibrate instruments?
After an event e.g. if there is a knock,
bump anything that can impact accuracy
When measurements don’t seem right
When instructed by manufacturer
Tuesday, February 27, 2024 7
Methods of calibration
•Calibration curve method
•Standard addition method
•Internal standard method
Tuesday, February 27, 2024 8
Calibration curve method
•In the calibration method, also known as
working curve method, a set of
standards must be prepared. They will each
contain a known amount of the analyte being
measured. These standards are then measured
using the instrument in question, and
a calibration curve will be plotted. This curve
will show the relationship between the
response of the instrument and the
concentration of the analyte. An example of a
calibration curve can be found below.
Tuesday, February 27, 2024 9
Calibration curve
Tuesday, February 27, 2024 10
Calibration cont’d
• When using the working curve method, it is important that each standard is
prepared individually and not all from the same stock solution. Any errors
in the stock solution will carry through the entire calibration process, and
thus the instrument will not be calibrated correctly.
• Also, the calibration curve should be checked for any outliers. An outlier
is a measurement which is significantly different from the other
measurements. Put simply, these results will shift the regression line (line of
best fit) and give inaccurate results, and should, therefore, be removed.
Tuesday, February 27, 2024 11
Steps to follow during Calibration
•The calibration standards should cover the range
of interest – this is so, during your actual
experiment, you are sure to get the most accurate
results from your curve
•A ‘blank’ should be included in your calibration –
this is a standard which contains no analyte
•Don’t automatically set your regression line
intercept to zero! Only do this if you have enough
evidence to show that the intercept point is not
actually statistically different from zero.
Tuesday, February 27, 2024 12
Example: Pb in Blood by GFAAS
9(Students plot)
[Pb] Signal
(ppb) (mAbs)
0.50 3.76
1.50 9.16
2.50 15.03
3.50 20.42
4.50 25.33
5.50 31.87
Tuesday, February 27, 2024 13
0
5
10
15
20
25
30
35
0 1 2 3 4 5 6
Pb Concentration (ppb)
mAbs
y = 5.56x + 0.93
Tuesday, February 27, 2024 14
.EXAMPLE
A sample containing an unknown amount of
Pb gives a signal of 27.5 mAbs. Calculate
the Pb concentration
Tuesday, February 27, 2024 15
Standard addition calibration
The standard-addition method of calibration helps to remove bias that may arise
from a number of factors, including the temperature and composition of the actual
matrix.
For the standard-addition method of calibration, two requirements must be met:
1.The calibration curve has to be linear
2.The regression line must pass through zero
• In this method, the signal intensity of the sample solution is measured. Then, the
analyte is added to this solution at known concentrations – the signal intensity is
measured after each addition of the analyte. This, therefore, gives a calibration
curve which is linear and shows signal intensity vs. added concentration.
The concentration of the analyte is determined from the point where
the regression line crosses the axis at zero.
Tuesday, February 27, 2024 16
Standard addition calibration
Tuesday, February 27, 2024 17
Standard addition cont’d
•In this method, the matrix itself remains
completely unchanged – it is for this
reason that this method is useful in cases
where the matrix is either very
complicated or hard to reproduce.
Tuesday, February 27, 2024 18
Standard Addition Method
1. Most convenient when a small number of
samples are to be analyzed.
2. Useful when the analyte is present in a
complicated matrix and no ideal blank is
available.
Tuesday, February 27, 2024 19
Standard Addition Procedure
1. Add one or more increments of a standard
solution to sample aliquots of the same size. Each
mixture is then diluted to the same volume.
2. Prepare a plot of Analytical Signal versus:
a) volume of standard solution added, or
b) concentration of analyte added.
Tuesday, February 27, 2024 20
Standard Addition Procedure
3. The x-intercept of the standard addition plot
corresponds to the amount of analyte that must
have been present in the sample (after accounting
for dilution).
4. The standard addition method assumes:
a) the curve is linear over the concentration range
b) the y-intercept of a calibration curve would be 0
Tuesday, February 27, 2024 21
Example: Fe in Drinking Water
Sample
Volume
(mL)
Standard
Volume
(mL) Signal (V)
10 0 0.215
10 5 0.424
10 10 0.685
10 15 0.826
10 20 0.967
The concentration of
the Fe standard
solution is 11.1 ppm
All solutions are
diluted to a final
volume of 50 mL
Tuesday, February 27, 2024 22
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
-10 -5 0 5 10 15 20 25
Volume of standard added (mL)
Signal
(V)
-6.08 mL
Tuesday, February 27, 2024 23
[Fe] = ?
x-intercept = -6.08 mL
Therefore, 10 mL of sample diluted to 50 mL would give a
signal equivalent to 6.08 mL of standard diluted to 50 mL.
Vsam x [Fe]sam = Vstd x [Fe]std
10.0 mL x [Fe] = 6.08 mL x 11.1 ppm
[Fe] = 6.75 ppm
Tuesday, February 27, 2024 24
Internal Standard Method
1. Most convenient when variations in
analytical sample size, position, or matrix
limit the precision of a technique.
2. May correct for certain types of noise.
Tuesday, February 27, 2024 25
Internal Standard Procedure
1. Prepare a set of standard solutions for
analyte (A) as with the calibration curve
method, but add a constant amount of a
second species (B) to each solution.
2. Prepare a plot of SA/SB versus [A].
Tuesday, February 27, 2024 26
Notes
1. The resulting measurement will be
independent of sample size and position.
2. Species A & B must not produce signals that
interfere with each other. Usually they are
separated by wavelength or time.
Tuesday, February 27, 2024 27
Example: Pb by ICP Emission
Each Pb solution contains 100
ppm Cu.
[Pb]
(ppm) Pb Cu Pb/Cu
20 112 1347 0.083
40 243 1527 0.159
60 326 1383 0.236
80 355 1135 0.313
100 558 1440 0.388
Signal
Tuesday, February 27, 2024 28
Standardization
• Standardization is the process of determining the
exact concentration of a solution. It is the process
of experimentally determining the relationship
between the signal and the amount of analyte
Tuesday, February 27, 2024 29
Blank correction
•In blank determination all steps of the analysis are
performed in the absence of a sample. A blank solution
contains the solvent and all the reagents in an analysis
but none of the sample. The results from the blank are
then applied as a correction to the sample
measurements. Blank determinations reveal errors due
to interfering contaminants from the reagents and
vessels employed in analysis.
Tuesday, February 27, 2024 30
Questions that follow are based on this Excel
sheet
Tuesday, February 27, 2024 31
Calculations
1. Using the data on the Precision and Accuracy tab, calculate the
average, standard deviation, percent relative standard deviation,
and 95% confidence interval.
a. What is the precision of the data?
b. The known value for the data is 0.1500. Is there evidence of
bias? Support your answer.
2. A calibration curve is presented in the Calibration Curve tab, where
the response of a method is plotted versus the analyte
concentration.
a. Estimate the linear region of the data. One way to approach this
problem is to selectively remove data points from the plot until
you maximize the R2 value.
b. What is the sensitivity in the linear region?
Tuesday, February 27, 2024 32
Questions continued
3. Data are presented in the Limit of Detection tab for 10
replicate measurements of a 4.00 mM sample ([A] =
4.00 mM) and 10 replicate measurements of a blank
([A] = 0 mM). Perform the following calculations in the
spreadsheet. Use the calibration curve from part (2)
to determine the slope.
a.Calculate the signal detection limit.
b.Calculate the limit of detection.
c. Calculate the signal quantitation limit.
d.Calculate the limit of quantitation.
Tuesday, February 27, 2024 33

More Related Content

Similar to Instrumental CAtomic absorption Spectrophotometryhemicl analysis_Lecture 2.pptx

Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
 Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
hydrologywebsite1
 
Quality Control and Batch Processing.pptx
Quality Control and Batch Processing.pptxQuality Control and Batch Processing.pptx
Quality Control and Batch Processing.pptx
KrisnaBagtasos1
 
Points to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological ProductsPoints to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological Products
Weijun Li
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
hydrologywebsite1
 

Similar to Instrumental CAtomic absorption Spectrophotometryhemicl analysis_Lecture 2.pptx (20)

Quantitation techniques used in chromatography
Quantitation techniques used in chromatographyQuantitation techniques used in chromatography
Quantitation techniques used in chromatography
 
How to establish QC reference ranges - Randox QC Educational Guide
How to establish QC reference ranges - Randox QC Educational GuideHow to establish QC reference ranges - Randox QC Educational Guide
How to establish QC reference ranges - Randox QC Educational Guide
 
Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
 Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
Download-manuals-water quality-technicalpapers-inter-labfindingsaqci-indround
 
Item 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical MethodsItem 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical Methods
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
Method validation
Method validationMethod validation
Method validation
 
Quality Control and Batch Processing.pptx
Quality Control and Batch Processing.pptxQuality Control and Batch Processing.pptx
Quality Control and Batch Processing.pptx
 
How to Measure Uncertainty
How to Measure UncertaintyHow to Measure Uncertainty
How to Measure Uncertainty
 
Introduzione ai differenti approcci alla stima dell'incertezza di misura Nari...
Introduzione ai differenti approcci alla stima dell'incertezza di misura Nari...Introduzione ai differenti approcci alla stima dell'incertezza di misura Nari...
Introduzione ai differenti approcci alla stima dell'incertezza di misura Nari...
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
Practical work in groups
Practical work in groupsPractical work in groups
Practical work in groups
 
Points to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological ProductsPoints to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological Products
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_IntroductionAnalytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_Introduction
 
Quality of Analytical Procedures
Quality of Analytical ProceduresQuality of Analytical Procedures
Quality of Analytical Procedures
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 
Attribute MSA
Attribute MSAAttribute MSA
Attribute MSA
 
Attribute MSA
Attribute MSA Attribute MSA
Attribute MSA
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality Control
 

More from abusunabakali

GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptxGROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
abusunabakali
 
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
abusunabakali
 
Lecture- 01- 2024 E amnnmb mmmchem..pptx
Lecture- 01- 2024 E amnnmb mmmchem..pptxLecture- 01- 2024 E amnnmb mmmchem..pptx
Lecture- 01- 2024 E amnnmb mmmchem..pptx
abusunabakali
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptxaaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
abusunabakali
 
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
abusunabakali
 
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
abusunabakali
 
3219atomicemissionspectroscopyslides-210225084128.pdf
3219atomicemissionspectroscopyslides-210225084128.pdf3219atomicemissionspectroscopyslides-210225084128.pdf
3219atomicemissionspectroscopyslides-210225084128.pdf
abusunabakali
 

More from abusunabakali (8)

GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptxGROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
GROUP 3 BTMbnvvvvvbcccccccccbxfhffdddhhhhhhhggggggggggggggggg.pptx
 
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
Curriculum and syllabusrrrrdrfgsgdfgggggggggggggggggggggggddddddddddddddddddd...
 
Lecture- 01- 2024 E amnnmb mmmchem..pptx
Lecture- 01- 2024 E amnnmb mmmchem..pptxLecture- 01- 2024 E amnnmb mmmchem..pptx
Lecture- 01- 2024 E amnnmb mmmchem..pptx
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptxaaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
aaaaaaaaaaaaaaaaaaaaaaaaaaaadsrrrreeeeeeettttttttttttr.pptx
 
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
Lecture- 01- 2024 E and G chemOne of the most important chemical water qualit...
 
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
UV-VIS molecular spectroscopy.202004201521035685ranvijay_engg_UV_and_VISIBLE_...
 
3219atomicemissionspectroscopyslides-210225084128.pdf
3219atomicemissionspectroscopyslides-210225084128.pdf3219atomicemissionspectroscopyslides-210225084128.pdf
3219atomicemissionspectroscopyslides-210225084128.pdf
 
1_Aromaticity year-2..pdf
1_Aromaticity year-2..pdf1_Aromaticity year-2..pdf
1_Aromaticity year-2..pdf
 

Recently uploaded

The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 

Recently uploaded (20)

Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docx
 
Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 
psychiatric nursing HISTORY COLLECTION .docx
psychiatric  nursing HISTORY  COLLECTION  .docxpsychiatric  nursing HISTORY  COLLECTION  .docx
psychiatric nursing HISTORY COLLECTION .docx
 
How to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSHow to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POS
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 

Instrumental CAtomic absorption Spectrophotometryhemicl analysis_Lecture 2.pptx

  • 1. LECTURE TWO CHM 2203 BY Godfrey Muhwezi INSTRUMENTAL CHEMICAL ANALYSIS Tuesday, February 27, 2024 1
  • 2. Outline of lecture Two Figures of merit Calibrations Standardizations Blank correction 2 Tuesday, February 27, 2024
  • 3. Figures of merit • A figure of merit is a numerical expression representing the efficiency of a given instrument or procedure. Figures of merit are a way for analytical chemists to characterize a method. The six figures of merit we use are; Precision-A measure of the reproducibility(Relative SD, Absolute SD) Accuracy-How close a measured value is to the true value  Sensitivity-A measure of the methods ability to distinguish between two samples reported as change in signal per unit change in the amount of analyte Tuesday, February 27, 2024 3
  • 4. Figures of merit cont’d Linear dynamic range-Concentration range-(range over which measurements can be made)  Detection limit -smallest amount of analyte that can be determined with confidence or the smallest amount of analyte that has a signal larger than the signal from the reagent blank Selectivity-A measure of a methods freedom from interferences Tuesday, February 27, 2024 4
  • 5. Calibrations • Calibration is defined as the act of making sure that a scientific process or instrument will produce results which are accurate. In more complex terms, calibration is the act which determines the functional relationship between measured values and analytical quantities. Tuesday, February 27, 2024 5
  • 6. Calibration cont’d •The goal of calibration is to minimize any measurement uncertainty by ensuring the accuracy of test equipment. Calibration quantifies and controls errors or uncertainties within measurement processes to an acceptable level. Tuesday, February 27, 2024 6
  • 7. When do you calibrate instruments? After an event e.g. if there is a knock, bump anything that can impact accuracy When measurements don’t seem right When instructed by manufacturer Tuesday, February 27, 2024 7
  • 8. Methods of calibration •Calibration curve method •Standard addition method •Internal standard method Tuesday, February 27, 2024 8
  • 9. Calibration curve method •In the calibration method, also known as working curve method, a set of standards must be prepared. They will each contain a known amount of the analyte being measured. These standards are then measured using the instrument in question, and a calibration curve will be plotted. This curve will show the relationship between the response of the instrument and the concentration of the analyte. An example of a calibration curve can be found below. Tuesday, February 27, 2024 9
  • 11. Calibration cont’d • When using the working curve method, it is important that each standard is prepared individually and not all from the same stock solution. Any errors in the stock solution will carry through the entire calibration process, and thus the instrument will not be calibrated correctly. • Also, the calibration curve should be checked for any outliers. An outlier is a measurement which is significantly different from the other measurements. Put simply, these results will shift the regression line (line of best fit) and give inaccurate results, and should, therefore, be removed. Tuesday, February 27, 2024 11
  • 12. Steps to follow during Calibration •The calibration standards should cover the range of interest – this is so, during your actual experiment, you are sure to get the most accurate results from your curve •A ‘blank’ should be included in your calibration – this is a standard which contains no analyte •Don’t automatically set your regression line intercept to zero! Only do this if you have enough evidence to show that the intercept point is not actually statistically different from zero. Tuesday, February 27, 2024 12
  • 13. Example: Pb in Blood by GFAAS 9(Students plot) [Pb] Signal (ppb) (mAbs) 0.50 3.76 1.50 9.16 2.50 15.03 3.50 20.42 4.50 25.33 5.50 31.87 Tuesday, February 27, 2024 13
  • 14. 0 5 10 15 20 25 30 35 0 1 2 3 4 5 6 Pb Concentration (ppb) mAbs y = 5.56x + 0.93 Tuesday, February 27, 2024 14
  • 15. .EXAMPLE A sample containing an unknown amount of Pb gives a signal of 27.5 mAbs. Calculate the Pb concentration Tuesday, February 27, 2024 15
  • 16. Standard addition calibration The standard-addition method of calibration helps to remove bias that may arise from a number of factors, including the temperature and composition of the actual matrix. For the standard-addition method of calibration, two requirements must be met: 1.The calibration curve has to be linear 2.The regression line must pass through zero • In this method, the signal intensity of the sample solution is measured. Then, the analyte is added to this solution at known concentrations – the signal intensity is measured after each addition of the analyte. This, therefore, gives a calibration curve which is linear and shows signal intensity vs. added concentration. The concentration of the analyte is determined from the point where the regression line crosses the axis at zero. Tuesday, February 27, 2024 16
  • 18. Standard addition cont’d •In this method, the matrix itself remains completely unchanged – it is for this reason that this method is useful in cases where the matrix is either very complicated or hard to reproduce. Tuesday, February 27, 2024 18
  • 19. Standard Addition Method 1. Most convenient when a small number of samples are to be analyzed. 2. Useful when the analyte is present in a complicated matrix and no ideal blank is available. Tuesday, February 27, 2024 19
  • 20. Standard Addition Procedure 1. Add one or more increments of a standard solution to sample aliquots of the same size. Each mixture is then diluted to the same volume. 2. Prepare a plot of Analytical Signal versus: a) volume of standard solution added, or b) concentration of analyte added. Tuesday, February 27, 2024 20
  • 21. Standard Addition Procedure 3. The x-intercept of the standard addition plot corresponds to the amount of analyte that must have been present in the sample (after accounting for dilution). 4. The standard addition method assumes: a) the curve is linear over the concentration range b) the y-intercept of a calibration curve would be 0 Tuesday, February 27, 2024 21
  • 22. Example: Fe in Drinking Water Sample Volume (mL) Standard Volume (mL) Signal (V) 10 0 0.215 10 5 0.424 10 10 0.685 10 15 0.826 10 20 0.967 The concentration of the Fe standard solution is 11.1 ppm All solutions are diluted to a final volume of 50 mL Tuesday, February 27, 2024 22
  • 23. -0.2 0 0.2 0.4 0.6 0.8 1 1.2 -10 -5 0 5 10 15 20 25 Volume of standard added (mL) Signal (V) -6.08 mL Tuesday, February 27, 2024 23
  • 24. [Fe] = ? x-intercept = -6.08 mL Therefore, 10 mL of sample diluted to 50 mL would give a signal equivalent to 6.08 mL of standard diluted to 50 mL. Vsam x [Fe]sam = Vstd x [Fe]std 10.0 mL x [Fe] = 6.08 mL x 11.1 ppm [Fe] = 6.75 ppm Tuesday, February 27, 2024 24
  • 25. Internal Standard Method 1. Most convenient when variations in analytical sample size, position, or matrix limit the precision of a technique. 2. May correct for certain types of noise. Tuesday, February 27, 2024 25
  • 26. Internal Standard Procedure 1. Prepare a set of standard solutions for analyte (A) as with the calibration curve method, but add a constant amount of a second species (B) to each solution. 2. Prepare a plot of SA/SB versus [A]. Tuesday, February 27, 2024 26
  • 27. Notes 1. The resulting measurement will be independent of sample size and position. 2. Species A & B must not produce signals that interfere with each other. Usually they are separated by wavelength or time. Tuesday, February 27, 2024 27
  • 28. Example: Pb by ICP Emission Each Pb solution contains 100 ppm Cu. [Pb] (ppm) Pb Cu Pb/Cu 20 112 1347 0.083 40 243 1527 0.159 60 326 1383 0.236 80 355 1135 0.313 100 558 1440 0.388 Signal Tuesday, February 27, 2024 28
  • 29. Standardization • Standardization is the process of determining the exact concentration of a solution. It is the process of experimentally determining the relationship between the signal and the amount of analyte Tuesday, February 27, 2024 29
  • 30. Blank correction •In blank determination all steps of the analysis are performed in the absence of a sample. A blank solution contains the solvent and all the reagents in an analysis but none of the sample. The results from the blank are then applied as a correction to the sample measurements. Blank determinations reveal errors due to interfering contaminants from the reagents and vessels employed in analysis. Tuesday, February 27, 2024 30
  • 31. Questions that follow are based on this Excel sheet Tuesday, February 27, 2024 31
  • 32. Calculations 1. Using the data on the Precision and Accuracy tab, calculate the average, standard deviation, percent relative standard deviation, and 95% confidence interval. a. What is the precision of the data? b. The known value for the data is 0.1500. Is there evidence of bias? Support your answer. 2. A calibration curve is presented in the Calibration Curve tab, where the response of a method is plotted versus the analyte concentration. a. Estimate the linear region of the data. One way to approach this problem is to selectively remove data points from the plot until you maximize the R2 value. b. What is the sensitivity in the linear region? Tuesday, February 27, 2024 32
  • 33. Questions continued 3. Data are presented in the Limit of Detection tab for 10 replicate measurements of a 4.00 mM sample ([A] = 4.00 mM) and 10 replicate measurements of a blank ([A] = 0 mM). Perform the following calculations in the spreadsheet. Use the calibration curve from part (2) to determine the slope. a.Calculate the signal detection limit. b.Calculate the limit of detection. c. Calculate the signal quantitation limit. d.Calculate the limit of quantitation. Tuesday, February 27, 2024 33

Editor's Notes

  1. Answer is m=5.56mAbs/ppb. B= 0.93( Y= Mc +b
  2. Calculate ( Answer, C = 4.78 ppb) ( Answer, C = 4.78 ppb) (3 significant figures)