Determination of optical density
(absorbance ) of bromophenol blue dye using
spectrophotometer
Prepared by
Sana Abbas
lecturer
DPT & Nursing
Spectophotometry
Spectrophotometry
A method in which the absorption or
transmission properties of a material is
quantitatively measured as a function of
wavelength.
The basic principle behind this method is that :
“Each compound absorbs or transmits light
over a certain range of wavelength”
Spectrophotometer
An instrument which measures light absorption as
a function of wavelength in the UV as well as
visible regions. It also follows essentially the laws
of light absorption viz the beer-lambert’s law.
Unlike colorimeters in spectrophotometers the
compound can be measured at precise
wavelength. The light absorbed by the sample is
directly proportional to the concentration of
sample in the solution. As concentration increases
, absorption increases exponentially.
Basic concepts and Principle
Principle behind this method is that : “Each
compound absorbs or transmits light over a
certain range of wavelength. When light passes
through a solution, a certain fraction is being
absorbed.
This fraction is detected, measured and used to
relate the light absorbed or transmitted to the
concentration of the substance.
This enables both qualitative and quantitative
analyses of substances.
The spectrophotometric technique measures
the light intensity due to : – Diffracting the
light beam into a spectrum of wavelengths –
Direct it to an object – Receiving the light
reflected or returned from the object –
Detecting the intensities with a charge-
coupled device – Displaying the results as a
graph on the detector and then the display
device.
The light absorption is directly related to the
concentration of the compound in the sample.
• As Concentration increases, light
Absorption increases linearly and light
Transmission decreases, exponentially.
Instrumentation
A spectrophotometer consists of
A light source
A prism that separate the lights into different
wavelengths
A slit through which a narrow beam of a
desire wavelength passes
A sample holder
A detector and a recording device.
Transmittance and Absorbance
When a sample is illuminated, it absorbs
some of the light and transmits the rest.
• The transmitted light (Is ) is of lower
intensity than the incident light (Io ), and the
transmitted light is defined as:
T = Is / Io
To ensure accuracy (by eliminating effects of reflection
by surface of the cell, absorption by the cell wall and
by solvent) an identical reference cell without the
compound of interest is also used.
• Thus, the amount of light absorbed (A) as the
incident light passes through the sample is equivalent
to: A = - log Is / IR = - log T
• In practice, the Reference cell is inserted, and the
instrument adjusted to an arbitrary scale corresponding
to 100% transmittance, after which the percentage
transmittance reading is made on the sample.
Types of spectrometry
Spectrophotometer can be classified into two
different types :
• Single Beam Spectrometer
• Double Beam Spectrometer
Single Beam Spectrometer
To measure the intensity of the incident light
the sample must be removed so that the
reference can be placed each time.
• This type of spectrometer is usually less
expensive and less complicated.
Double Beam Spectrometer
In this type, before it reaches the sample, the
light source is split into two separate beams. •
From these one passes through the sample
and second one is used for reference. • This
gives an advantage because the reference
reading and sample reading can take place at
the same time.
Based on wavelength of light used it can be classified into:
(A)VisibleSpectrometer:-
• Uses visible range (400 – 700nm) of electromagnetic
radiation spectrum. • Visible spectrophotometers vary in
accuracy. • Plastic and glass cuvettes can be used for visible
light spectroscopy.
(B)UV Spectrometer:-
• Uses light over the UV range (180 - 400 nm). • UV
spectroscopy is used for fluids, and even solids. • Cuvettes,
only made of quartz, are used for placing the samples.
(C) IR Spectrophotometer:-
• Uses light over infra red range (700 -15000) of
electromagnetic radiation spectra.
(D) Flourescent Spectrometer
These are instruments used to scan the fluorescence
spectrum emitted by liquid fluorescent labels, which used
in scientific research, chemical
industry,medicine,biochemistry,environmental
protection,clinical testing,food testing,teaching
experiments and other fields. (E)Atomic absorption
Spectrometer:-
• This method is mainly applied to detect trace components
in the sample analysis.
• It is powerful tool for material analysis and elemental
analysis of trace metals(semimetals).
Device and Mechanism
The spectrophotometer, in general, consists of two
devices.
They are the following :
1.Spectrometer : A device that produce, typically
disperse and measure the light.
2.Photometer : Indicates the photoelectric detector
that measures the light.
• The spectrometer consists of the following
parts :
(i) Light source :It produces a desired range of
wavelength of light.
(ii) Collimator : It transmits a straight beam of
light.
(iii)Monochromator : It split the light into its
component wavelength.
(iv)Wavelength selector : transmits only the
desired wavelength.
Cuvettes:
The optically transparent cells (cuvette) are made up
of glass, plastic, silica or quartz, glass and plastic
absorb UV light below 310 nm. Photocell and photo-
multiplier tubes: It a photocell is a photoelectric
device which converts light energy into electrical
energy, which is then amplified, detected and
recorded.’’ • The photometer detects the light
absorbed by the sample as the light from the slit is
passed through the solution and then it sends signal
to the galvanometer or digital display.
Beer Lambert Law
• It states that the absorbance of light by a material
in a solution is directly proportional to its
concentration in that solution.
A = lc Where,
ϵ
A -absorbance
-molar absorptivity
ϵ
l -length of solution
c -concentration
Standardization Graph.
Standard Solution of known concentration) of the
compound of interest are made, treated, and their
absorbance's (ABS) and concentration values are
used to create a Standardization Graph.
Absorption Spectra
A spectrum of electromagnetic radiation
transmitted through a substance, showing dark
lines or bands due to absorption at specific
wavelengths.
Applications
Concentration measurement
Detection of impurities
Chemical kinetics
Detection of functional group
Molecular weight determination
The spectrophotometer can measure the intensity
of light absorbed indirectly by the solutes in
solution because each solution has its own
characteristic absorption chattels. The
spectrophotometer can be used to measure the
concentration of compounds in a certain solution.
Differentiating compounds is another uses of
spectrophotometer .It works by analyzing the
pattern of wavelengths absorbed by the sample.
The absorbance can be calculated using:
Beer-Lambert Law, Absorbance, A= λbc
where,
λ is the molar absorbtivity coefficient in L
mol-1
cm-1
b is the path length (in cm) of the cuvette in
which the sample is contained
c is the concentration of compound in
solution, in mol L­
­
-1
If two compounds are present in a solution, the
total absorbance of the solution is the sum of the
two separate contributions according to the
formula Atotal = K1C1+K2C2 where,
C1 and C2 are the concentrations of
bromophenol blue and methyl orange
respectively in the mixture
K1 and K2 are the molar absorbtivity coefficient
of each compound of the respective wavelength.
Material
Bromophenol blue solution
Micropipette
Tip droplet
Test tubes .
Method
Determination of Amax of bromophenol blue
A cuvette with distilled water was placed into the
spectrophotometer. The wavelength input is set to 470 nm. Auto
zero button was pressed to set the absorbance into zero.
The blank is removed. Absorbance of bromophenol blue is read at
different wavelengths.
After a certain wavelength is tested against bromophenol blue, the
spectrophotometer is set blank using distilled water.
A new wavelength is set to obtain a new absorbance value.
A graph of absorption spectrum was plotted and the wavelength with
maximum absorbance reading was determined from the graph.
Absorbance of bromophenol blue at different
wavelength
Table 1
Wavelength (nm) Absorbance
470 0.093
500 0.143
530 0.282
560 0.535
590 0.981
620 0.211
650 0.013
680 0.002
Graph 1
Absorbance of bromophenol blue at different wavelength
Results
From the graph 1, the highest peak is found
to be at the wavelength 590nm. Therefore,
the Amax of bromophenol blue is in
wavelength of 590nm.
THANK YOU

bromophenol protocol of nursing (1).pptx

  • 1.
    Determination of opticaldensity (absorbance ) of bromophenol blue dye using spectrophotometer Prepared by Sana Abbas lecturer DPT & Nursing
  • 2.
  • 3.
    Spectrophotometry A method inwhich the absorption or transmission properties of a material is quantitatively measured as a function of wavelength. The basic principle behind this method is that : “Each compound absorbs or transmits light over a certain range of wavelength”
  • 4.
    Spectrophotometer An instrument whichmeasures light absorption as a function of wavelength in the UV as well as visible regions. It also follows essentially the laws of light absorption viz the beer-lambert’s law. Unlike colorimeters in spectrophotometers the compound can be measured at precise wavelength. The light absorbed by the sample is directly proportional to the concentration of sample in the solution. As concentration increases , absorption increases exponentially.
  • 5.
    Basic concepts andPrinciple Principle behind this method is that : “Each compound absorbs or transmits light over a certain range of wavelength. When light passes through a solution, a certain fraction is being absorbed. This fraction is detected, measured and used to relate the light absorbed or transmitted to the concentration of the substance. This enables both qualitative and quantitative analyses of substances.
  • 6.
    The spectrophotometric techniquemeasures the light intensity due to : – Diffracting the light beam into a spectrum of wavelengths – Direct it to an object – Receiving the light reflected or returned from the object – Detecting the intensities with a charge- coupled device – Displaying the results as a graph on the detector and then the display device.
  • 7.
    The light absorptionis directly related to the concentration of the compound in the sample. • As Concentration increases, light Absorption increases linearly and light Transmission decreases, exponentially.
  • 8.
    Instrumentation A spectrophotometer consistsof A light source A prism that separate the lights into different wavelengths A slit through which a narrow beam of a desire wavelength passes A sample holder A detector and a recording device.
  • 9.
    Transmittance and Absorbance Whena sample is illuminated, it absorbs some of the light and transmits the rest. • The transmitted light (Is ) is of lower intensity than the incident light (Io ), and the transmitted light is defined as: T = Is / Io
  • 10.
    To ensure accuracy(by eliminating effects of reflection by surface of the cell, absorption by the cell wall and by solvent) an identical reference cell without the compound of interest is also used. • Thus, the amount of light absorbed (A) as the incident light passes through the sample is equivalent to: A = - log Is / IR = - log T • In practice, the Reference cell is inserted, and the instrument adjusted to an arbitrary scale corresponding to 100% transmittance, after which the percentage transmittance reading is made on the sample.
  • 11.
    Types of spectrometry Spectrophotometercan be classified into two different types : • Single Beam Spectrometer • Double Beam Spectrometer
  • 12.
    Single Beam Spectrometer Tomeasure the intensity of the incident light the sample must be removed so that the reference can be placed each time. • This type of spectrometer is usually less expensive and less complicated.
  • 13.
    Double Beam Spectrometer Inthis type, before it reaches the sample, the light source is split into two separate beams. • From these one passes through the sample and second one is used for reference. • This gives an advantage because the reference reading and sample reading can take place at the same time.
  • 14.
    Based on wavelengthof light used it can be classified into: (A)VisibleSpectrometer:- • Uses visible range (400 – 700nm) of electromagnetic radiation spectrum. • Visible spectrophotometers vary in accuracy. • Plastic and glass cuvettes can be used for visible light spectroscopy. (B)UV Spectrometer:- • Uses light over the UV range (180 - 400 nm). • UV spectroscopy is used for fluids, and even solids. • Cuvettes, only made of quartz, are used for placing the samples. (C) IR Spectrophotometer:- • Uses light over infra red range (700 -15000) of electromagnetic radiation spectra.
  • 15.
    (D) Flourescent Spectrometer Theseare instruments used to scan the fluorescence spectrum emitted by liquid fluorescent labels, which used in scientific research, chemical industry,medicine,biochemistry,environmental protection,clinical testing,food testing,teaching experiments and other fields. (E)Atomic absorption Spectrometer:- • This method is mainly applied to detect trace components in the sample analysis. • It is powerful tool for material analysis and elemental analysis of trace metals(semimetals).
  • 16.
    Device and Mechanism Thespectrophotometer, in general, consists of two devices. They are the following : 1.Spectrometer : A device that produce, typically disperse and measure the light. 2.Photometer : Indicates the photoelectric detector that measures the light.
  • 17.
    • The spectrometerconsists of the following parts : (i) Light source :It produces a desired range of wavelength of light. (ii) Collimator : It transmits a straight beam of light. (iii)Monochromator : It split the light into its component wavelength. (iv)Wavelength selector : transmits only the desired wavelength.
  • 18.
    Cuvettes: The optically transparentcells (cuvette) are made up of glass, plastic, silica or quartz, glass and plastic absorb UV light below 310 nm. Photocell and photo- multiplier tubes: It a photocell is a photoelectric device which converts light energy into electrical energy, which is then amplified, detected and recorded.’’ • The photometer detects the light absorbed by the sample as the light from the slit is passed through the solution and then it sends signal to the galvanometer or digital display.
  • 19.
    Beer Lambert Law •It states that the absorbance of light by a material in a solution is directly proportional to its concentration in that solution. A = lc Where, ϵ A -absorbance -molar absorptivity ϵ l -length of solution c -concentration
  • 20.
    Standardization Graph. Standard Solutionof known concentration) of the compound of interest are made, treated, and their absorbance's (ABS) and concentration values are used to create a Standardization Graph. Absorption Spectra A spectrum of electromagnetic radiation transmitted through a substance, showing dark lines or bands due to absorption at specific wavelengths.
  • 21.
    Applications Concentration measurement Detection ofimpurities Chemical kinetics Detection of functional group Molecular weight determination
  • 22.
    The spectrophotometer canmeasure the intensity of light absorbed indirectly by the solutes in solution because each solution has its own characteristic absorption chattels. The spectrophotometer can be used to measure the concentration of compounds in a certain solution. Differentiating compounds is another uses of spectrophotometer .It works by analyzing the pattern of wavelengths absorbed by the sample.
  • 23.
    The absorbance canbe calculated using: Beer-Lambert Law, Absorbance, A= λbc where, λ is the molar absorbtivity coefficient in L mol-1 cm-1 b is the path length (in cm) of the cuvette in which the sample is contained c is the concentration of compound in solution, in mol L­ ­ -1
  • 24.
    If two compoundsare present in a solution, the total absorbance of the solution is the sum of the two separate contributions according to the formula Atotal = K1C1+K2C2 where, C1 and C2 are the concentrations of bromophenol blue and methyl orange respectively in the mixture K1 and K2 are the molar absorbtivity coefficient of each compound of the respective wavelength.
  • 25.
  • 26.
    Method Determination of Amaxof bromophenol blue A cuvette with distilled water was placed into the spectrophotometer. The wavelength input is set to 470 nm. Auto zero button was pressed to set the absorbance into zero. The blank is removed. Absorbance of bromophenol blue is read at different wavelengths. After a certain wavelength is tested against bromophenol blue, the spectrophotometer is set blank using distilled water. A new wavelength is set to obtain a new absorbance value. A graph of absorption spectrum was plotted and the wavelength with maximum absorbance reading was determined from the graph.
  • 27.
    Absorbance of bromophenolblue at different wavelength Table 1 Wavelength (nm) Absorbance 470 0.093 500 0.143 530 0.282 560 0.535 590 0.981 620 0.211 650 0.013 680 0.002
  • 28.
    Graph 1 Absorbance ofbromophenol blue at different wavelength
  • 29.
    Results From the graph1, the highest peak is found to be at the wavelength 590nm. Therefore, the Amax of bromophenol blue is in wavelength of 590nm.
  • 30.