How to assess cardiac
contractility using langendorff
apparatus connected to a
powerlab.
Dina Hamdy Merzeban
Physiology Lecturer
Fayoum University
dhm00@Fayoum.edu.eg
Aim of the work
• How to isolate heart.
• How to calibrate powerlab from millivolt recording to mmHg.
• How to assess contractility using langendorff apparatus
connected to a powerlab.
Introduction
• As isolated/ hanging heart models lack fresh blood
circulation, hormonal and autonomic nervous
responses and otherwise very complex in-vivo factors
are decoupled which helps to perform a variety of
basic analyses of fundamental cardiac properties
unmasking the potential direct action of studied
compounds
Types of isolated heart models:
• Langendorff: (retrograde perfusion).
• Working Heart: (antegrade perfusion).
APPLICATIONS OF LANGENDORFF SET-UP USING
PRESSURE CATHETER MEASUREMENTS
• Assessment of contractility (inotropy).
• Assessment of HR (chronotropy).
• Coronary vasculature dilations.
• Arrhythmogenic, anti-arrhythmic, anti-fibrillatory effects .
• Gradual determination of hypoxic damage.
• Calcium antagonism.
Fig 1: langendorff apparatus.
Langendorff apparatus:
• Principle : is to maintain the heart perfused at constant
temperature.
• Langendorff apparatus consists of two tube systems, the first
system contains perfusate (Kreb-Henseleit KH buffer) that states in
a reservoir and ends in the aotic cannula passing through a pump
to maintain constant flow to the isolated heart (non recirculating
constant flow Langendorff apparatus)
• A second tube system that jacket the first system with water
maitained at 37 ºC using a heater unit that the fluid recirculates to
it.
• After aortic cannulation of the heart, it is placed in water jacketed
heart chamber maintained at 37 ºC
Types of flow in langendorff
CONSTANT PRESSURE LANGENDORFF
• Pressure is held constant and changes in coronary resistance are
detected as changes in blood flow.
• Constant pressure is achieved by elevating the aortic bubble trap
chamber as 1 mmHg = 13.6 mm of water column therefore
0.816 m would be needed to create 60 mmHg.
CONSTANT FLOW LANGENDORFF:
• Flow is held constant at fixed flow rate via peristaltic pump and
changes in coronary resistance are detected as changes in
pressure.
Excision of the heart and its
cannulation (rat):
• Male Wistar rats weighing 350-450 g were used. They were
anaesthetized using ketamine hydrochloride (25 mg/kg,
intraperitonial) and heparinized by intraperitoneal injection of
Heparin(1000IU). A left thoracotomy was performed and
hearts were rapidly exposed, excised by cutting across the arch
of the aorta (leaving enough space for mounting on the
cannula) along with all other vascular structures in the area. The
heart is removed into ice-cold heparinized phosphate buffered
saline (4°C) to arrest beating. The ascending aorta is then
mounted on the aortic cannula (Louch et.al., 2011).
•
• Mounted heart should be perfused by flow of up to 15
ml/min/g of heart (Bell et.al.,2011).
Introduction of fluid-filled balloon into
the LV :
Caliberation of powerlab:
Experimental preparation
1. Frog Pithing Procedure
2. Dissecting out the Frog Sciatic Nerve:
Frog’s sciatic nerve in nerve bath
Powerlab 4/30
Proper Connections to the Powerlab
Experimental procedure
3. Setup and calibration of equipment:
•the nerve bath above, stimulating
electrode in black wire,
•two stimulating electrodes in white
color,
• red is anode +,
•black is cathode -,
•green is ground
LabChart 7 welcome center
The test connection data shown in the Scope
view.
Using the Scope view to show
increasing CAP amplitude.
Waveform cursor on the
maximum CAP amplitude
CAP amplitude versus stimulus intensity (data obtained from
the present experiment).
Proximal CAP amplitude
Stimulus
amplitude
Proximal CAP
amplitude
Stimulus amplitude
Δ(mV) mV Δ(mV) mV
2.1109 20 42.0859 400
4.2641 40 63.125 500
6.3141 60 103.875 1000
8.3594 80 253.8125 2000
10.6297 100 382.9375 3000
12.8141 120 510.5313 4000
14.5984 140 767.1563 6000
16.8875 160 774.7 7000
19.025 180 882.5 8000
21.0953 200 1003 9000
23.1406 220 1051.7 9500
25.2953 240 1117.2 10000
27.4719 260 1113.8 10500
29.5641 280 1122.2 11000
31.8297 300 1129.2 11500
33.4781 320 1130.4 12000
35.7344 340 1131.4 12500
38.0953 360 1131 13000
40.2172 380 1131 13500
Threshold stimulus voltage: 20 mV
Maximum CAP amplitude: 11500 mV
CAP amplitude versus stimulus intensity (as
obtained from the threshold test)
0
2000
4000
6000
8000
10000
12000
14000
16000
020040060080010001200
stimulusamplitude
CAP amplitude
CAP amplitude versus stimulus intensity
Proximal CAP amplitude Δ(mV)
Stimulus amplitude mV
Determination of nerve conduction
velocity
• Using a ruler, measure the distance in millimeters between
the black negative leads of each of the two recording
electrodes.
• Read the value for time differential (Δt) from the Cursor
display in the LabChart application window in milliseconds.
Measure the distance in millimeters between the
black negative leads of each of the two recording
electrodes.
Zoom window in overlay mode showing analysis
procedure for calculating conduction velocity.
Waveform Cursor information is displayed at the
top of the window.
(overlay in zoom view not activated!!!)
Calculation of conduction velocity
The conduction velocity of nerve is normally reported in meters per second. It is
more easily recorded as millimeters per millisecond, which yields the same result.
Distance between recording
electrodes:
10 Mm
Time interval between CAP1
and CAP2:
0.02 Ms
Conduction velocity 500 m/s
Methods for Calculation of
conduction velocity
• Conduction velocity is systematically related to fiber
diameter
• Velocity (m/s) = Diameter (µmeters) x 2.5
• Two methods for Calculation of conduction velocity
• Difference method
• Absolute method
The difference method:
• The stimulus duration is set to 0.2 ms, and two pairs
of recording leads are connected
• A maximal CAP is elicited, and the latencies (in
ms) to the peaks of the responses are measured.
• This latency signifies the time it took for average
fibers in the nerve to transmit their APs from the
stimulating electrodes to the first recording electrode
Zoom window in overlay mode showing analysis
procedure for calculating conduction velocity.
Waveform Cursor information is displayed at the
top of the window.
(overlay in zoom view not activated!!!)
• Now subtract the latency determined proximally from the latency
determined distally,
• a value for the time it took for the CAP to travel from proximal
electrode R1 to distal electrode R7.
• Conduction velocity (difference method)
• = (d2-d1)/(latencydistal - latencyproximal)
• =(20mm-10mm)/ (0.08msec- 0.05 msec)= 10/0.03= 333m/sec
The absolute method:
• The conduction velocity can also be calculated by the
absolute method, which means that the velocity is
calculated using a single latency and distance
measurement.
Calculation using proximal electrode
only
Measure from the cathode (the stimulating
electrode nearest the recording electrodes) to the
first recording electrode = 0.5 cm
the stimulating electrode
nearest the recording
electrodes
The first recording
electrode
• Measure from the cathode (the stimulating electrode nearest
the recording electrodes) to the first recording electrode R1.
• Latency = latency at electrode R1 (ms)
d = distance (mm) from stimulating electrode to recording
electrode R1
• Velocity = d / latency (m/s) or (mm/ms)
• One can also calculate the velocity for distal electrode R7: the same
formula applies except that the distance d and the latency are of
course longer.
Bi-Directionality of Nerve
Conduction
Characteristics of the CAP:
• The peak amplitude of the CAP .
• The latency of the onset of the CAP.
• The latency of the peak of the CAP
• The duration of the CAP
• The threshold stimulus voltage
• The maximal stimulus voltage
Stimulator Settings
Stimulator Settings• Stimulus Duration: 0.2 ms
• Stimulus Voltage: 0.05 V
• Delay: 1 ms
• Rate: 1 Hz
• Mode: Repeat
The peak amplitude of the CAP:
is the voltage value of the peak of the CAP
response.
The peak amplitude of the CAP is 63 mV (as revealed in the
present experiment).
the latency of the onset of CAP is 0.00002 sec (=0.02 msec) (as revealed in
the present experiment).
The latency of the onset of the CAP:
is the time from the onset of the stimulus
artifact to the onset of the CAP.
The latency of the peak of the CAP:
is the time from the onset of the stimulus
artifact to the peak of the CAP.
The latency of the peak of the CAP is 0.00004 sec (=0.04msec) (as
revealed from the present experiment).
The duration of the CAP:
is the time from the beginning of the positive
phase to the end of the negative phase of the
CAP.
the duration of the CAP is 0.00029 sec (=0.29 msec) (as revealed from
the present experiment).
The threshold stimulus voltage:
is determined by raising and lowering the stimulus voltage
a little to find the voltage at which the CAP is just
discernible.
The threshold stimulus voltage is 20 mv and the CAP peak is 2.09 mV
(as revealed from the present experiment table 1).
The maximal stimulus voltage:
is the point at which a further increase in
stimulus voltage produces no further increase in
the CAP amplitude.
The maximal stimulus voltage is 11.5 V and the CAP peak is 1149
mV (as revealed from the present experiment table 1)
The significance of measuring the latency
to the beginning of the CAP, versus
measuring it to the peak of the CAP:
• The latency of the beginning of the CAP reflects
how long it takes for the fastest fibers to conduct
action potentials from the stimulus source to the
recording electrodes.
• When the latency is measured to the peak of the
CAP, we obtain the latency of an average fiber in
the nerve.
The significance of the threshold
voltage
• the threshold voltage, measured from the stimulator, is the
voltage needed to generate at least one AP from a fiber in
the sciatic nerve bundle.
• (Remember when we speak of CAP thresholds, we are
dealing with a whole nerve and not an individual fibre.)
• As the resolution of the screen is poor, it is hard to
measure an accurate threshold. The true thresholds are in
reality much lower.
Assessment of cardiac contractility using langendorff apparatus connected to a powerlab

More Related Content

What's hot

screening methods for Antiepileptic activity
screening methods for Antiepileptic activityscreening methods for Antiepileptic activity
screening methods for Antiepileptic activity
Sravanthi Shetty
 
Screening of antihypertensive agents
Screening of antihypertensive agentsScreening of antihypertensive agents
Screening of antihypertensive agents
Kanthlal SK
 
Screening of anti anxiety drugs
Screening of anti anxiety drugsScreening of anti anxiety drugs
Screening of anti anxiety drugs
Bindu Pulugurtha
 

What's hot (20)

Screening of antiepileptic drugs
Screening of antiepileptic drugsScreening of antiepileptic drugs
Screening of antiepileptic drugs
 
Screening Methods of Antihypertensive Agents
Screening Methods of Antihypertensive AgentsScreening Methods of Antihypertensive Agents
Screening Methods of Antihypertensive Agents
 
pre clinical Screening for anti asthmatic drugs
pre clinical Screening  for anti asthmatic drugspre clinical Screening  for anti asthmatic drugs
pre clinical Screening for anti asthmatic drugs
 
OECD Guidelines
OECD GuidelinesOECD Guidelines
OECD Guidelines
 
screening methods for Antiepileptic activity
screening methods for Antiepileptic activityscreening methods for Antiepileptic activity
screening methods for Antiepileptic activity
 
Pharmacological screening of anti arrhythmic drug
Pharmacological screening of anti arrhythmic drugPharmacological screening of anti arrhythmic drug
Pharmacological screening of anti arrhythmic drug
 
Multiple sclerosis screening methods
Multiple sclerosis screening methodsMultiple sclerosis screening methods
Multiple sclerosis screening methods
 
Screening of antidepressant
Screening of antidepressantScreening of antidepressant
Screening of antidepressant
 
Screening models for inflammatory drugs
Screening models for inflammatory drugsScreening models for inflammatory drugs
Screening models for inflammatory drugs
 
Analgesic screening methods
Analgesic screening methodsAnalgesic screening methods
Analgesic screening methods
 
SCREENING OF DRUGS USED IN ANTIARRYTHMIA
SCREENING OF DRUGS USED IN ANTIARRYTHMIASCREENING OF DRUGS USED IN ANTIARRYTHMIA
SCREENING OF DRUGS USED IN ANTIARRYTHMIA
 
Screening of antihypertensive agents
Screening of antihypertensive agentsScreening of antihypertensive agents
Screening of antihypertensive agents
 
Screening of hepatoprotective drugs
Screening of hepatoprotective drugsScreening of hepatoprotective drugs
Screening of hepatoprotective drugs
 
Expt 3 Bracketing bioassay
Expt 3 Bracketing bioassayExpt 3 Bracketing bioassay
Expt 3 Bracketing bioassay
 
Screening models for immunomodulator
Screening models for immunomodulatorScreening models for immunomodulator
Screening models for immunomodulator
 
Screening Models of Antidepressants Drugs
Screening Models of Antidepressants DrugsScreening Models of Antidepressants Drugs
Screening Models of Antidepressants Drugs
 
Screening of anti anxiety drugs
Screening of anti anxiety drugsScreening of anti anxiety drugs
Screening of anti anxiety drugs
 
Dose selection
Dose selectionDose selection
Dose selection
 
Cns stimulants and depressants screening models
Cns stimulants and depressants screening modelsCns stimulants and depressants screening models
Cns stimulants and depressants screening models
 
Screening antianginal (1)
Screening antianginal (1)Screening antianginal (1)
Screening antianginal (1)
 

Similar to Assessment of cardiac contractility using langendorff apparatus connected to a powerlab

Digital blood pressure meter
Digital blood pressure meterDigital blood pressure meter
Digital blood pressure meter
Culun Habis
 
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-CouplingGate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
Nusrat Mary
 
A new precision peak detector full wave rectifier
A new precision peak detector full wave rectifierA new precision peak detector full wave rectifier
A new precision peak detector full wave rectifier
Vishal kakade
 

Similar to Assessment of cardiac contractility using langendorff apparatus connected to a powerlab (20)

Electronic instrumentation in NMT
Electronic instrumentation in NMTElectronic instrumentation in NMT
Electronic instrumentation in NMT
 
SENSING WITH CHAOS
SENSING WITH CHAOSSENSING WITH CHAOS
SENSING WITH CHAOS
 
Digital blood pressure meter
Digital blood pressure meterDigital blood pressure meter
Digital blood pressure meter
 
blood flow measurement
blood flow measurementblood flow measurement
blood flow measurement
 
Basic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacingBasic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacing
 
ECG
ECGECG
ECG
 
STSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptxSTSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptx
 
VISHAV snp.pptx
VISHAV snp.pptxVISHAV snp.pptx
VISHAV snp.pptx
 
Digial instrumentation fnal
Digial instrumentation fnalDigial instrumentation fnal
Digial instrumentation fnal
 
Unit 4 Pulse Modulation.pdf
Unit 4 Pulse Modulation.pdfUnit 4 Pulse Modulation.pdf
Unit 4 Pulse Modulation.pdf
 
enhancement of ecg signal using wavelet transfform
enhancement of ecg signal using wavelet transfformenhancement of ecg signal using wavelet transfform
enhancement of ecg signal using wavelet transfform
 
biomedical signal processing and its analysis
biomedical signal processing and its analysisbiomedical signal processing and its analysis
biomedical signal processing and its analysis
 
IAS Report Final
IAS Report FinalIAS Report Final
IAS Report Final
 
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-CouplingGate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
Gate Pulse Triggering of Single Phase Thyristor Circuit through Opto-Coupling
 
Amplifiers, filters and digital recording systems
Amplifiers, filters and digital recording systemsAmplifiers, filters and digital recording systems
Amplifiers, filters and digital recording systems
 
MR Spectroscopy
MR SpectroscopyMR Spectroscopy
MR Spectroscopy
 
Design of Two CMOS Differential Amplifiers
Design of Two CMOS Differential AmplifiersDesign of Two CMOS Differential Amplifiers
Design of Two CMOS Differential Amplifiers
 
A new precision peak detector full wave rectifier
A new precision peak detector full wave rectifierA new precision peak detector full wave rectifier
A new precision peak detector full wave rectifier
 
Saecg
SaecgSaecg
Saecg
 
Blood Flow Meter
Blood Flow MeterBlood Flow Meter
Blood Flow Meter
 

More from dina merzeban

Bodyfluids , measurement of compartmental volumes, and types of cell junctions
Bodyfluids , measurement of compartmental volumes, and types of cell junctionsBodyfluids , measurement of compartmental volumes, and types of cell junctions
Bodyfluids , measurement of compartmental volumes, and types of cell junctions
dina merzeban
 

More from dina merzeban (20)

Endocrine Physiology ca homeostasis.pptx
Endocrine Physiology  ca homeostasis.pptxEndocrine Physiology  ca homeostasis.pptx
Endocrine Physiology ca homeostasis.pptx
 
Endocrine Physiology pituitary.
Endocrine Physiology  pituitary.Endocrine Physiology  pituitary.
Endocrine Physiology pituitary.
 
Endocrine Physiology introduction.pptx
Endocrine Physiology  introduction.pptxEndocrine Physiology  introduction.pptx
Endocrine Physiology introduction.pptx
 
Endocrine Physiology ADRENAL CORTEX.pptx
Endocrine Physiology  ADRENAL CORTEX.pptxEndocrine Physiology  ADRENAL CORTEX.pptx
Endocrine Physiology ADRENAL CORTEX.pptx
 
Endocrine Physiology thyroid.
Endocrine Physiology  thyroid.Endocrine Physiology  thyroid.
Endocrine Physiology thyroid.
 
physiology of Sensory nervous system, updated 2021
physiology of Sensory nervous system,  updated 2021physiology of Sensory nervous system,  updated 2021
physiology of Sensory nervous system, updated 2021
 
physiology of trigeminal nerve, UMNL& LMNL facial paralysis, taste
physiology of  trigeminal nerve, UMNL& LMNL facial paralysis, tastephysiology of  trigeminal nerve, UMNL& LMNL facial paralysis, taste
physiology of trigeminal nerve, UMNL& LMNL facial paralysis, taste
 
Muscle physiology
Muscle physiologyMuscle physiology
Muscle physiology
 
Nerve physiology
Nerve physiologyNerve physiology
Nerve physiology
 
Bodyfluids , measurement of compartmental volumes, and types of cell junctions
Bodyfluids , measurement of compartmental volumes, and types of cell junctionsBodyfluids , measurement of compartmental volumes, and types of cell junctions
Bodyfluids , measurement of compartmental volumes, and types of cell junctions
 
Cerebellum
Cerebellum Cerebellum
Cerebellum
 
Motor cortex areas and physiology of stroke
Motor cortex areas and physiology of strokeMotor cortex areas and physiology of stroke
Motor cortex areas and physiology of stroke
 
Basalganglia
Basalganglia   Basalganglia
Basalganglia
 
Renal physiology lecture 2 dina
Renal physiology lecture 2 dinaRenal physiology lecture 2 dina
Renal physiology lecture 2 dina
 
Lecture 2 leucocytes
Lecture 2 leucocytesLecture 2 leucocytes
Lecture 2 leucocytes
 
Lecture 1 platelet and hemostasis
Lecture 1 platelet and hemostasisLecture 1 platelet and hemostasis
Lecture 1 platelet and hemostasis
 
Nervous system
Nervous system Nervous system
Nervous system
 
Renal physiology lecture 1 dina
Renal physiology lecture 1 dinaRenal physiology lecture 1 dina
Renal physiology lecture 1 dina
 
Peptic ulcer
Peptic ulcer Peptic ulcer
Peptic ulcer
 
Digestive system
Digestive systemDigestive system
Digestive system
 

Recently uploaded

Recently uploaded (20)

Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
 
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadO898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 

Assessment of cardiac contractility using langendorff apparatus connected to a powerlab

  • 1. How to assess cardiac contractility using langendorff apparatus connected to a powerlab. Dina Hamdy Merzeban Physiology Lecturer Fayoum University dhm00@Fayoum.edu.eg
  • 2. Aim of the work • How to isolate heart. • How to calibrate powerlab from millivolt recording to mmHg. • How to assess contractility using langendorff apparatus connected to a powerlab.
  • 4. • As isolated/ hanging heart models lack fresh blood circulation, hormonal and autonomic nervous responses and otherwise very complex in-vivo factors are decoupled which helps to perform a variety of basic analyses of fundamental cardiac properties unmasking the potential direct action of studied compounds
  • 5. Types of isolated heart models: • Langendorff: (retrograde perfusion). • Working Heart: (antegrade perfusion).
  • 6. APPLICATIONS OF LANGENDORFF SET-UP USING PRESSURE CATHETER MEASUREMENTS • Assessment of contractility (inotropy). • Assessment of HR (chronotropy). • Coronary vasculature dilations. • Arrhythmogenic, anti-arrhythmic, anti-fibrillatory effects . • Gradual determination of hypoxic damage. • Calcium antagonism.
  • 7. Fig 1: langendorff apparatus.
  • 8. Langendorff apparatus: • Principle : is to maintain the heart perfused at constant temperature. • Langendorff apparatus consists of two tube systems, the first system contains perfusate (Kreb-Henseleit KH buffer) that states in a reservoir and ends in the aotic cannula passing through a pump to maintain constant flow to the isolated heart (non recirculating constant flow Langendorff apparatus) • A second tube system that jacket the first system with water maitained at 37 ºC using a heater unit that the fluid recirculates to it. • After aortic cannulation of the heart, it is placed in water jacketed heart chamber maintained at 37 ºC
  • 9. Types of flow in langendorff CONSTANT PRESSURE LANGENDORFF • Pressure is held constant and changes in coronary resistance are detected as changes in blood flow. • Constant pressure is achieved by elevating the aortic bubble trap chamber as 1 mmHg = 13.6 mm of water column therefore 0.816 m would be needed to create 60 mmHg. CONSTANT FLOW LANGENDORFF: • Flow is held constant at fixed flow rate via peristaltic pump and changes in coronary resistance are detected as changes in pressure.
  • 10. Excision of the heart and its cannulation (rat): • Male Wistar rats weighing 350-450 g were used. They were anaesthetized using ketamine hydrochloride (25 mg/kg, intraperitonial) and heparinized by intraperitoneal injection of Heparin(1000IU). A left thoracotomy was performed and hearts were rapidly exposed, excised by cutting across the arch of the aorta (leaving enough space for mounting on the cannula) along with all other vascular structures in the area. The heart is removed into ice-cold heparinized phosphate buffered saline (4°C) to arrest beating. The ascending aorta is then mounted on the aortic cannula (Louch et.al., 2011). •
  • 11. • Mounted heart should be perfused by flow of up to 15 ml/min/g of heart (Bell et.al.,2011).
  • 12. Introduction of fluid-filled balloon into the LV :
  • 14. Experimental preparation 1. Frog Pithing Procedure 2. Dissecting out the Frog Sciatic Nerve:
  • 15. Frog’s sciatic nerve in nerve bath
  • 17. Proper Connections to the Powerlab
  • 18. Experimental procedure 3. Setup and calibration of equipment:
  • 19. •the nerve bath above, stimulating electrode in black wire, •two stimulating electrodes in white color, • red is anode +, •black is cathode -, •green is ground
  • 21. The test connection data shown in the Scope view.
  • 22. Using the Scope view to show increasing CAP amplitude.
  • 23. Waveform cursor on the maximum CAP amplitude
  • 24. CAP amplitude versus stimulus intensity (data obtained from the present experiment). Proximal CAP amplitude Stimulus amplitude Proximal CAP amplitude Stimulus amplitude Δ(mV) mV Δ(mV) mV 2.1109 20 42.0859 400 4.2641 40 63.125 500 6.3141 60 103.875 1000 8.3594 80 253.8125 2000 10.6297 100 382.9375 3000 12.8141 120 510.5313 4000 14.5984 140 767.1563 6000 16.8875 160 774.7 7000 19.025 180 882.5 8000 21.0953 200 1003 9000 23.1406 220 1051.7 9500 25.2953 240 1117.2 10000 27.4719 260 1113.8 10500 29.5641 280 1122.2 11000 31.8297 300 1129.2 11500 33.4781 320 1130.4 12000 35.7344 340 1131.4 12500 38.0953 360 1131 13000 40.2172 380 1131 13500 Threshold stimulus voltage: 20 mV Maximum CAP amplitude: 11500 mV
  • 25. CAP amplitude versus stimulus intensity (as obtained from the threshold test) 0 2000 4000 6000 8000 10000 12000 14000 16000 020040060080010001200 stimulusamplitude CAP amplitude CAP amplitude versus stimulus intensity Proximal CAP amplitude Δ(mV) Stimulus amplitude mV
  • 26. Determination of nerve conduction velocity • Using a ruler, measure the distance in millimeters between the black negative leads of each of the two recording electrodes. • Read the value for time differential (Δt) from the Cursor display in the LabChart application window in milliseconds.
  • 27. Measure the distance in millimeters between the black negative leads of each of the two recording electrodes.
  • 28. Zoom window in overlay mode showing analysis procedure for calculating conduction velocity. Waveform Cursor information is displayed at the top of the window. (overlay in zoom view not activated!!!)
  • 29.
  • 30. Calculation of conduction velocity The conduction velocity of nerve is normally reported in meters per second. It is more easily recorded as millimeters per millisecond, which yields the same result. Distance between recording electrodes: 10 Mm Time interval between CAP1 and CAP2: 0.02 Ms Conduction velocity 500 m/s
  • 31. Methods for Calculation of conduction velocity • Conduction velocity is systematically related to fiber diameter • Velocity (m/s) = Diameter (µmeters) x 2.5 • Two methods for Calculation of conduction velocity • Difference method • Absolute method
  • 32. The difference method: • The stimulus duration is set to 0.2 ms, and two pairs of recording leads are connected
  • 33. • A maximal CAP is elicited, and the latencies (in ms) to the peaks of the responses are measured. • This latency signifies the time it took for average fibers in the nerve to transmit their APs from the stimulating electrodes to the first recording electrode
  • 34. Zoom window in overlay mode showing analysis procedure for calculating conduction velocity. Waveform Cursor information is displayed at the top of the window. (overlay in zoom view not activated!!!)
  • 35. • Now subtract the latency determined proximally from the latency determined distally, • a value for the time it took for the CAP to travel from proximal electrode R1 to distal electrode R7. • Conduction velocity (difference method) • = (d2-d1)/(latencydistal - latencyproximal) • =(20mm-10mm)/ (0.08msec- 0.05 msec)= 10/0.03= 333m/sec
  • 36. The absolute method: • The conduction velocity can also be calculated by the absolute method, which means that the velocity is calculated using a single latency and distance measurement.
  • 37. Calculation using proximal electrode only
  • 38. Measure from the cathode (the stimulating electrode nearest the recording electrodes) to the first recording electrode = 0.5 cm the stimulating electrode nearest the recording electrodes The first recording electrode
  • 39. • Measure from the cathode (the stimulating electrode nearest the recording electrodes) to the first recording electrode R1. • Latency = latency at electrode R1 (ms) d = distance (mm) from stimulating electrode to recording electrode R1 • Velocity = d / latency (m/s) or (mm/ms) • One can also calculate the velocity for distal electrode R7: the same formula applies except that the distance d and the latency are of course longer.
  • 41. Characteristics of the CAP: • The peak amplitude of the CAP . • The latency of the onset of the CAP. • The latency of the peak of the CAP • The duration of the CAP • The threshold stimulus voltage • The maximal stimulus voltage
  • 43. Stimulator Settings• Stimulus Duration: 0.2 ms • Stimulus Voltage: 0.05 V • Delay: 1 ms • Rate: 1 Hz • Mode: Repeat
  • 44. The peak amplitude of the CAP: is the voltage value of the peak of the CAP response. The peak amplitude of the CAP is 63 mV (as revealed in the present experiment).
  • 45. the latency of the onset of CAP is 0.00002 sec (=0.02 msec) (as revealed in the present experiment). The latency of the onset of the CAP: is the time from the onset of the stimulus artifact to the onset of the CAP.
  • 46. The latency of the peak of the CAP: is the time from the onset of the stimulus artifact to the peak of the CAP. The latency of the peak of the CAP is 0.00004 sec (=0.04msec) (as revealed from the present experiment).
  • 47. The duration of the CAP: is the time from the beginning of the positive phase to the end of the negative phase of the CAP. the duration of the CAP is 0.00029 sec (=0.29 msec) (as revealed from the present experiment).
  • 48. The threshold stimulus voltage: is determined by raising and lowering the stimulus voltage a little to find the voltage at which the CAP is just discernible. The threshold stimulus voltage is 20 mv and the CAP peak is 2.09 mV (as revealed from the present experiment table 1).
  • 49. The maximal stimulus voltage: is the point at which a further increase in stimulus voltage produces no further increase in the CAP amplitude. The maximal stimulus voltage is 11.5 V and the CAP peak is 1149 mV (as revealed from the present experiment table 1)
  • 50. The significance of measuring the latency to the beginning of the CAP, versus measuring it to the peak of the CAP: • The latency of the beginning of the CAP reflects how long it takes for the fastest fibers to conduct action potentials from the stimulus source to the recording electrodes. • When the latency is measured to the peak of the CAP, we obtain the latency of an average fiber in the nerve.
  • 51. The significance of the threshold voltage • the threshold voltage, measured from the stimulator, is the voltage needed to generate at least one AP from a fiber in the sciatic nerve bundle. • (Remember when we speak of CAP thresholds, we are dealing with a whole nerve and not an individual fibre.) • As the resolution of the screen is poor, it is hard to measure an accurate threshold. The true thresholds are in reality much lower.