SlideShare a Scribd company logo
1 of 120
Basic Technical Concepts inBasic Technical Concepts in
Cardiac PacingCardiac Pacing
Dr D Sunil Reddy
Consultant Cardiologist
KIMS Hospital
TopicTopic
 Electrical Stimulation of Cardiac Tissue
Myocardial StimulationMyocardial Stimulation
 An artificial electrical stimulus
excites cardiac tissue by the
creation of an electrical field at the
interface of the stimulating
electrode and the myocardium
 The electric field should be strong
enough and should last long
enough to initiate action potentials
in the cells at the electrode-tissue
interface
Myocardial StimulationMyocardial Stimulation
 The AP’s at the site of stimulation
result in AP’s in the neighbouring
areas of the myocardium
resulting in a wave of AP’s
(depolarization) propagating
away from the site of stimulation
Time (Milliseconds)
100 200 300 400 500
Phase 2
Phase 1
Phase 3
Phase 4
TransmembranePotential
(Millivolts)
-50
0
50
-100
Phase0
Threshold
 An artificial electrical stimulus excites
cardiac tissue by the creation of an
electrical field at the interface of the
stimulating electrode and the myocardium
 The electric field should be strong enough
and should last long enough to initiate
action potentials in the cells at the
electrode-tissue interface
Electric Field – Current DensityElectric Field – Current Density
TheoryTheory
 Electric Field Theory
– A Minimum Voltage/cmis required to trigger a
self-propagating wave of depolarization
 Current Density Theory
– A Minimum Current/cm2
is required to trigger a
self-propagating wave of depolarization
 Stimulation threshold is a function of Voltage/cm or
Current/ cm2
that is induced in the myocardium
beneath the stimulating electrode
 The two theories are related by Ohm’s Law
The Implantable Pacemaker SystemThe Implantable Pacemaker System
 The Implantable Pulse Generator (IPG) : metal can
(titanium) containing electronics/battery & an electrode or
lead connector header
 Lead : Electrical connection between the pacemaker & the
heart
The Pacemaker CircuitThe Pacemaker Circuit
 Stimulation of cardiac tissue using electric pulses
– IPG
– Lead – insulated current conductor (s) &
electrodes to transmit pulses to heart tissue
and measure or sense electrical activity in the
heart
Heart
Lead
IPG
Completion of circuit
through
Lead or Body tissue
A Unipolar Pacing System Contains a Lead withA Unipolar Pacing System Contains a Lead with
Only One Electrode Within the Heart; In ThisOnly One Electrode Within the Heart; In This
System, the Impulse:System, the Impulse:
 Flows through the tip
electrode (cathode)
 Stimulates the heart
 Returns through
body fluid and tissue
to the IPG (anode)
Cathode
Anode
-
Anode
 Flows through the
tip electrode located
at the end of the
lead wire
 Stimulates the heart
 Returns to the ring
electrode above the
lead tip
A Bipolar Pacing System Contains a Lead with TwoA Bipolar Pacing System Contains a Lead with Two
Electrodes Within the Heart. In This System, the Impulse:Electrodes Within the Heart. In This System, the Impulse:
Cathode
Tip electrode coil
Indifferent electrode
coil
Lead Components
Connector
Lead body
Electrode(s)
Insulation
Conductors
Fixation
Threshold
Sensing
Connector
Standards
Electrodes -- Fixation Mechanism
 Passive Fixation Mechanism – Endocardial
– Tined
– Finned
– Canted/curved
Electrodes – Fixation Mechanism
 Active Fixation Mechanism – Endocardial
– Fixed screw
– Extendible/retractable
Electrodes -- Fixation Mechanism
 Fixation Mechanism – Myocardial/Epicardial
– Stab-in
– Screw-in
– Suture-on
Single-Chamber SystemSingle-Chamber System
 The pacing lead is
implanted in the
atrium or ventricle,
depending on the
chamber to be paced
and sensed
 One lead implanted
in the atrium and
one in the ventricle
Dual-Chamber Systems Have Two Leads:Dual-Chamber Systems Have Two Leads:
A pacing system can be thought of a standard electrical
circuit:
The pacemaker supplies the voltage.
Current (electrons) flow down the
conductor to the lead tip or cathode (-)
Where the lead tip touches the
myocardium, electrical resistance is
produced.
The current then flows through the
body tissues to the anode (+) and back
to the battery.
All of these things are required for
current to flow.
Pacing Circuit ParametersPacing Circuit Parameters
 Voltage
 Current
 Impedance (resistance)
 Energy
VoltageVoltage
Pushes electrons through an electric circuit resulting in electric
current through the circuit
VoltageVoltage
 Voltage is the force or “push” that causes
electrons to move through a circuit
 In a pacing system, voltage is:
– Measured in volts
– Represented by the letter “V”
– Provided by the pacemaker battery
– Often referred to as amplitude
CurrentCurrent
Current is the flow of charge or electrons through a circuit
CurrentCurrent
 The flow of electrons in a completed circuit
 In a pacing system, current is:
– Measured in mA (milliamps)
– Represented by the letter “I”
– Determined by the amount of electrons that
move through a circuit
Resistance or ImpedanceResistance or Impedance
Opposition offered by a circuit to the flow of current
ImpedanceImpedance
 The opposition to current flow in a circuit
 In a pacing system, impedance is:
– Measured in ohms
– Represented by the letter “R” (Ω for
numerical values)
– The measurement of the sum of all
resistance to the flow of current
Ohm’s LawOhm’s Law
Voltage = Current X Resistance
Current = Voltage ÷Resistance
Resistance = Voltage ÷ Current
The Pacemaker StimulusThe Pacemaker Stimulus
Time
5 Volts 5 Volts
0.5 ms
1 sec
Pacing Stimulus Voltage or Amplitude – 5 Volts
Pulse Width – 0.0005 seconds or 0.5 milliseconds
Pacing Rate – One stimulus per second or 60 stimuli (beats)
per minute
Voltage
The Pacing PulseThe Pacing Pulse
tt
Pacing Pulse
Pulse Duration (Width)
OutputVoltage
V = Pulse Amplitude in Volts (V) (say 2.5 V)
t = Pulse Duration or Width in milliseconds (ms)
(say 0.5 ms)
R = Impedance of Pacing Circuit (ohms)
(say 500 ohms)
I = V/R = Current through pacing circuit (mA)
= 2.5 V/ 500 ohms = 0.005 A = 5 mA
E = Energy supplied by Pulse to the Pacing Circuit
and Cardiac Tissue = V . I . t = I2
Rt = V2
t/R
= 2.5 V . 5 mA . 0.5 ms = 6.25 micro Joules
V
t
Stimulation ThresholdStimulation Threshold
 Pacing Voltage Threshold – The minimum
pacing voltage at any given pulse width
required to consistently achieve myocardial
depolarization outside the heart’s refractory
period
Loss of Capture
The Strength-Duration RelationThe Strength-Duration Relation
 The intensity of an electrical stimulus (Energy,
Voltage, Current, Charge) required to capture
(non-refractory) cardiac tissue is dependent on
the duration for which the electrical stimulus is
applied (i.e. pulse width)
The Voltage-Strength DurationThe Voltage-Strength Duration
CurveCurve
 Stimulus Voltage & Pulse Width
have an exponential relationship
 At short pulse widths (<0.25 ms)the
curve rises sharply (i.e. small
reductions in pulse width result in
large increases in the voltage
threshold)
 At long pulse widths (>1.0 ms)the
curve is flat (i.e. the voltage
threshold does not reduce with
increasing pulse width) Duration
Pulse Width (ms)
.50
1.0
1.5
2.0
.25StimulationThreshold(Volts)
0.25 1.0 1.5
Capture
Rheobase & ChronaxieRheobase & Chronaxie
Rheobase Voltage = Voltage
Threshold at infinite Pulse widths
(e.g. 2 ms) – also known as
fundamental threshold
Chronaxie Point = Pulse Width
threshold at twice the rheobase
voltage (i.e. 1 V & 0.3 ms)
Energy of Pacing Pulse
At rheobase = 0.5V. 2ms. 1mA = 1
MicroJoule
At Chronaxie = 1V. 0.3 ms. 2mA
= 0.6 MicroJoules
At PW=0.25 ms, = 1.5V. 0.25ms.
3mA = 1.125 MicroJoules
The chronaxie point approximates the point of minimum threshold energy on the Strength-Duration Curve
Duration
Pulse Width (ms)
StimulationThreshold(Volts)
0.2 0.6 1.0
Energy Strength Duration
Curve
Capture
0.4
0.5
1.0
2.0
0.8 1.2 1.4
Rheobase Voltage
Chronaxie Point
Goals of Pacemaker OutputGoals of Pacemaker Output
ProgrammingProgramming
 Consistent capture & Patient Safety is ensured
 Battery drain minimized, Pacemaker longevity
maximized
Programming Chronic PacemakerProgramming Chronic Pacemaker
OutputOutput
Duration
Pulse Width (ms)
StimulationThreshold(Volts)
0.2 0.6 1.0
Energy Strength Duration Curve
• Capture
Adequate Safety Margin while minimizing
Pulse Energy
Operate around the chronaxie point
Programming long pulse widths (& low
Voltages) increases Pulse energy but
hardly increases safety margin
Programming high voltages (with short
Pulse widths) increases Pulse energy but
hardly increases safety margin
For a 2 times safety margin :
Output = 2 x chronaxie voltage at
chronaxie pulse width (2V, 0.35 ms)
Output = 3 x pulse duration threshold
at twice chronaxie voltage (2V, 0.51 ms)
0.4
0.5
1.0
2.0
0.8 1.2 1.4
Rheobase
Chronaxie
Determination of StimulationDetermination of Stimulation
Threshold during Implant or Follow-upThreshold during Implant or Follow-up
 Pace the heart at
– Rate higher than sinus or intrinsic rate
– Stimulus amplitude and pulse width that ensure
capture (usually 5 V and 0.5 ms)
 Gradually reduce stimulus amplitude while maintaining
pulse width constant till capture is lost
 The Stimulation threshold is specified by the minimum
stimulus amplitude at which capture consistently occurs
at a given stimulus pulse width
Programming Pacemaker OutputProgramming Pacemaker Output
 Acute – Immediately post-implant
 Chronic – Approx. 6 to 8 weeks post implant
 Goal : To ensure consistent capture despite potential
changes in the SD curve while minimizing the energy
delivered by the pulse
Evolution of Pacing ThresholdEvolution of Pacing ThresholdVoltage
Threshold(V)
Observation Time (weeks)
Acute Phase
Chronic Phase
Safety Margin
6
5
4
3
2
1
0 4 8 12 16
Programmable ParametersProgrammable Parameters
 Pacemaker Pulse Output or Amplitude
– 0.5 V, 1.0 V, 1.5 V…., 7.5 V
 Pacemaker Pulse Width
– 0.03, 0.06, …., 0.25, 0.5,…, 1.0 milliseconds
Pacing Impedance - ConsiderationsPacing Impedance - Considerations
 Maximize Pacemaker Longevity
– Reduce current drain
– Maintain relatively high impedance
 Ensure consistent capture
– Appropriate voltage & current are available
at the electrode-tissue interface required to
stimulate tissue
I pacing
Components of Pacing ImpedanceComponents of Pacing Impedance
Rcoil
Zpolarisation -electrode tissue
interface
Rtissue
Electrode
Tissue
Interface
Pulse
Generator
V pulse
I pacing
RRcoil -coil - Lead Conductor ResistanceLead Conductor Resistance
– Lead Conductor
Impedance
• Reduces the voltage &
energy available at the
tissue for pacing
• Generates waste heat
• Designed to be low
conductor
resistance
Conductor ResistanceConductor Resistance
Total Coil Voltage Voltage
impedance resistance at tip loss
(Ω)(Ω) (V) %
600 50 2.3 8.3
750 200 1.8 26.7
1200 650 1.1 54.2
1200 50 2.4 4.2
Output Voltage = 2.5 V
Impact of conductor coil resistance upon
available tip electrode voltage
Conductor Resistance
V mV
2.5 2500
Total imp 600
Current I =V/R 4.166666667 mAConductor
imp 50
Voltage drop V = IR 208.3333333 mV
0.208333333 V
What is the Polarization Effect – PolarizationWhat is the Polarization Effect – Polarization
Impedance ?Impedance ?
As the pacing pulse begins, electrons from the pacemaker battery flow to the lead
tip and positively charged ions from the tissue are attracted to the lead tip.
Initially, the movement of these negatively
charged ions results in the flow of current
from the electrode into the myocardium.
As the pacing pulse continues, positively charged ions surround the electrode tip. This
produces a positively charged layer on the electrode called polarization. Polarization can
impede current flow from the electrode into the tissue.
Polarization Impedance &Polarization Impedance &
Electrode Surface AreaElectrode Surface Area
 As electrode surface area goes up
– Polarization impedance decreases
 As electrode surface area goes down
– Polarization impedance increases
Polarization ImpedancePolarization Impedance
Pacing Pulse
Polarization
Factors affecting PolarizationFactors affecting Polarization
ImpedanceImpedance
 Pulse Width
 Surface area of the electrode tip
– Larger the surface area – Lower the
Polarization
Electrode – Tissue InterfaceElectrode – Tissue Interface
ImpedanceImpedance
 Resistance to current flow from body tissue at the
electrode-tissue interface
Electrode – Tissue InterfaceElectrode – Tissue Interface
ImpedanceImpedance
 Smaller the surface area of the tip electrode, the higher
the current density (or electric field strength) at the
electrode-tissue interface
 The higher the field strength, the greater the pacing
efficiency & lower the threshold
Electrode – Tissue InterfaceElectrode – Tissue Interface
ImpedanceImpedance
 Smaller the geometric size of the tip electrode, the
higher the resistance between electrode and tissue
PacingImpedance(Ohms)
0
500
1000
1500
0 1 2 3 4 5.5 6
Geometric Tip Electrode Surface Area (mm2
)
Size = Impedance
The Ideal Stimulating ElectrodeThe Ideal Stimulating Electrode
 Needs large surface area to reduce Polarization
Impedance
 Needs small size to maximize Electric Field
Strength and Stimulation Efficacy
 Needs small size to maximize Electrode-tissue
resistance and minimize pacemaker current
drain
Electrodes -- Surface StructureElectrodes -- Surface Structure
 Porous Electrode Surface
CapSure®
8.0 mm2
Porous Electrode
CapSure®
SP Novus
5.8 mm2
Platinized
Porous Electrode
CapSure®
Z Novus
1.2 mm2
Platinized
Porous Electrode
15KV x2500 12.0V MDT
Analogy:
Lead with low impedance
and poor efficiency
Lead Impedance I
Lead Impedance II
(Conductor-) Resistance
Analogy:
Lead with high impedance in the
conductor and poor efficiency
Lead Impedance III
Analogy:
Lead with big tip-tissue impedance
low conductor impedance
and increased efficiency
The Chronic Virtual ElectrodeThe Chronic Virtual Electrode
IMPLANT CHRONIC
(8 weeks or longer)
Excitable
Cardiac
Tissue
Non-Excitable
Fibrotic
Tissue
Excitable
Cardiac
Tissue
Evolution of Pacing ThresholdEvolution of Pacing ThresholdVoltage
Threshold(V)
Observation Time (weeks)
Acute Phase
Chronic Phase
Safety Margin
6
5
4
3
2
1
0 4 8 12 16
Electrodes -- Steroid ElutionElectrodes -- Steroid Elution
Tines for
Stable Fixation
Silicone Rubber Plug
Containing Steroid
Porous,
Platinized Tip
for Steroid
Elution
 Type - Steroid in matrix
Electrodes -- Steroid ElutionElectrodes -- Steroid Elution
 Effect of Steroid on Stimulation Thresholds
Pulse Width = 0.5 msec
0
3 6
Implant Time (Weeks)
Textured Metal Electrode
Smooth Metal Electrode
1
2
3
4
5
Steroid-Eluting Electrode
0 1 2 4 5 7 8 9 10 11 12
Volts
Factors that affect StimulationFactors that affect Stimulation
ThresholdThreshold
 Eating, Sleeping, Exercise, Medications, Changes in Cardiac condition –
30 to 50 % during the day
 Drugs e.g. Steroids reduce stimulation threshold by reducing
inflammation
 Sympathomimetic drugs decrease threshold
 Amiodarone, Class I A (quinidine, procainamide), Class I B (Mexilitene)
 Hypokalemia increases threshold (diuretics)
 Hypocalcemia increases threshold
 Hypoxia & Hypercapnia – Increase threshold
 Electrolyte imbalance & pH e.g. acidosis and alkalosis increase threshold
Typical Pacing CircuitTypical Pacing Circuit
ImpedancesImpedances
 300 to 1200 ohms
 May be measured at implant time with a
Pacing System Analyzer (PSA) at implant time
 May be measured through telemetry with a
pacemaker programmer
Pacing Impedance Values Will Change DuePacing Impedance Values Will Change Due
to:to:
 Insulation breaks
 Wire fractures
An Insulation Break Around the Lead WireAn Insulation Break Around the Lead Wire
Can Cause Impedance Values to FallCan Cause Impedance Values to Fall
 Insulation breaks expose the
wire to body fluids which
have a low resistance and
cause impedance values to fall
 Current drains through the
insulation break into the body
which depletes the battery
 An insulation break can cause
impedance values to fall
below 300 Ω
Insulation break
Decreased resistance
A Wire Fracture Within the Insulating SheathA Wire Fracture Within the Insulating Sheath
May Cause Impedance Values to RiseMay Cause Impedance Values to Rise
 Impedance values
across a break in the
wire will increase
 Current flow may be
too low to be effective
 Impedance values
may exceed 3,000 Ω
Lead wire fracture
Increased resistance
Unipolar &Bipolar StimulationUnipolar &Bipolar Stimulation
+
-
+ -
-
+
Anode is IPG
Case
Cathode is lead tip
electrode
-
+
Anode is lead
proximal
electrode
Cathode is lead tip
electrode
Unipolar & Bipolar StimulationUnipolar & Bipolar Stimulation
Unipolar Stimulation– Large Stimulus Artifact
Bipolar Stimulation – Small Stimulus Artifact
There is a greater chance of pacemaker pocket
muscle stimulation with Unipolar Stimulation
The Pacemaker BatteryThe Pacemaker Battery
 Pacemakers typically use Lithium-Iodide batteries
 Most LiI batteries have a Beginning of Life (BOL) value
of 2.8 volts
 Recommended Replacement Time (ERI/RRT) value of
2.6 V
 End of Service (EOL/EOS) value of 2.5 V
 As battery depletes its internal resistance goes up – (BOL
= 100 ohms , EOL > 5000 ohms )
Recommended Replacement Time
RRT (ERI)
2.8V-
2.6V-
2.5V-
BOL
RRT 2.6V
EOS 2.5V
Time
Lithium Iodine Battery depletion
BOL – Beginning of Life
RRT – Recommended Replacement Time
EOS - End of Service
Battery LifeBattery Life
 Battery Life
– Battery Life = Battery Capacity/Current Drain
– 2.0 Ah/25microamps = 80,000 hours = 9.3 years
 Battery properties
– Reliability – no premature failure
– High volumetric energy density – Small battery
volume with high storage capacity
– Low self-discharge rate
– High hermiticity – no gas generation during
operation
Pacemaker LongevityPacemaker Longevity
 Energy = (V.I.t) = (V2
/R) . t Joules
 Reducing V by a factor of 2 reduces E by a
factor of 4
 Increasing Z by a factor by 2 reduces E by a
factor of 2
Pacemaker LongevityPacemaker Longevity
 High Outputs and pulse widths are the primary
cause for reduced pacemaker longevity
• Output Voltages higher than the Battery Voltage (2.8 V)
require the use of voltage doublers that use high battery
energy
• Long Pulse Widths Reduce pacing efficiency due to
increased polarization impedance
Sensing Intracardiac ElectricalSensing Intracardiac Electrical
ActivityActivity
A Pacemaker Must Be Able to Sense
and Respond to Cardiac Rhythms
 Accurate sensing enables the pacemaker to determine
whether or not the heart has created a beat on its own
 The pacemaker is usually programmed to respond with
a pacing impulse only when the heart fails to produce
an intrinsic beat
Intracardiac Electrical SignalsIntracardiac Electrical Signals
 Electrical currents that arise in the myocardium during
depolarization and repolarization
 A myocardial electrode - records voltage difference
wrt reference electrode when the myocardium under
the electrode undergoes depolarization or
repolarization
 The electrical activity measured by such an electrode
(which is in direct contact with cardiac tissue) – local
tissue electrical activity- Intracardiac Electrogram or
EGM
Sensing the EGMSensing the EGM
Depolarization WaveDepolarization Wave
Processed byProcessed by
DeviceDevice
Processed byProcessed by
DeviceDevice
The EGM Signal
The signal from a depolarization wave passing between two electrodes
Intracardiac ElectrogramIntracardiac Electrogram
R wave of the EGM indicating
depolarisation of cardiac
ventricular tissue at lead tip –
Roughly corresponds to R
wave of the ECG that
represents depolarisation of the
ventricles
T wave of the EGM indicating
repolarisation of cardiac
ventricular tissue at lead tip –
Roughly corresponds to T wave
of the ECG that represents
repolarisation of the ventricles
Voltage Deflections of the SensedVoltage Deflections of the Sensed
EGM in a PacemakerEGM in a Pacemaker
Pacemaker
Stimulus
Paced R wave
Post-pace T wave
Intrinsic R wave
T wave corresponding to
intrinsic R wave
Undersensing . . .
 Pacemaker does not “see” the intrinsic beat, and therefore
does not respond appropriately
Intrinsic beat
not sensed
Scheduled pace
delivered
VVI / 60
Oversensing
 An electrical signal other than the intended P or R wave
is detected
Marker channel
shows intrinsic
activity...
...though no
activity is
present
VVI / 60
 Measured by:
– Amplitude
• Peak-to-peak measurement (height)
of deflection
• Measured in Millivolts (mV)
– Slew Rate
• Speed of deflection change over time
• Measured in volts per second (V/s)
EGM amplitude
EGM Amplitude & Slew RateEGM Amplitude & Slew Rate
Amplitude of the EGM R wave = dV
Slew Rate of the EGM R wave= dV/dt
EGM amplitude & slew rate – ideal
values
Atrium Ventricle
Slew Rate 0.5 volts/sec 0.75 volts/sec
Amplitude >1.5 mV >5.0 mV
Which of the these complexes shows the
highest slew rate?
a)1
b)2
c)3
d)4
Frequency (Hz)
Accurate Sensing Requires That
Extraneous Signals Be Filtered Out
Effect of Filtering on EGMEffect of Filtering on EGM
FILTER
•Filters out frequencies below 5 Hz (T waves) and above 50 Hz
(myopotentials)
•Enhances frequencies between 10 Hz and 50 Hz (R waves) with maximum
enhancement at around 30 Hz frequency
• Low Slew Rate implies Low Frequency implies Less enhancement due to
filtering
Level
Detector
Level Detector or SensitivityLevel Detector or Sensitivity
SettingSetting
This is a value specified to the pacemaker in millivolts through programming.
All ventricular EGM deflections AFTER FILTERING that exceed the sensitivity
setting will be identified by the pacemaker as intrinsic R waves
The typical sensitivity setting that is programmed for Ventricular sensing is 2.5 mV
Sensitivity
The Greater the Number, the Less Sensitive the Device to Intracardiac Events
Sensitivity
Amplitude(mV)
Time
5.0
2.5
1.25
Sensitivity SettingSensitivity Setting
Sensitivity settings less than 2.5 mv – High sensitivity – can lead
to oversensing
Sensitivity settings greater than 2.5 mV – Low sensitivity – can
lead to undersensing
Amplitude(mV)
Amplitude(mV)
Time Time
5.0
2.5
1.25
5.0
2.5
1.25
Factors That May Affect Sensing Are:
 Lead polarity (unipolar vs. bipolar)
 Lead integrity
– Insulation break
– Wire fracture
 EMI – Electromagnetic Interference
Unipolar Sensing
 Produces a large
potential difference due
to:
– A cathode and
anode that are
farther apart than
in a bipolar
system
_
Bipolar Sensing
 Produces a smaller potential
difference due to the short
interelectrode distance
– Electrical signals from
outside the heart such
as myopotentials are
less likely to be sensed
An Insulation Break May Cause Both
Undersensing or Oversensing
 Undersensing occurs when inner and outer conductor coils are in
continuous contact
– Signals from intrinsic beats are reduced at the sense
amplifier and amplitude no longer meets the programmed
sensing value
–
 Oversensing occurs when inner and outer conductor coils make
intermittent contact
– Signals are incorrectly interpreted as P or R waves
Wire Fracture Can Cause Both Undersensing and
Oversensing
 Undersensing occurs when the cardiac signal is unable to get back to the
pacemaker – intrinsic signals cannot cross the wire fracture
 Oversensing occurs when the severed ends of the wire intermittently make
contact, which creates potentials interpreted by the pacemaker as P or R waves
Fracture in one filament leads to an
increase in resistance
Electromagnetic Interference
 Interference is caused by electromagnetic energy with a source that is
outside the body
 Electromagnetic fields that may affect pacemakers are radio-frequency
waves
– 50-60 Hz are most frequently associated with
pacemaker interference
 Few sources of EMI are found in the home or office but several exist in
hospitals
Oversensing May Occur When EMI Signals Are
Incorrectly Interpreted as P Waves or R Waves
 Pacing rates will vary as a result of EMI:
– Rates will accelerate if sensed as P waves in
dual-chamber systems (P waves are “tracked”)
– Rates will be low or inhibited if sensed in single-
chamber systems, or on ventricular lead in
dual-chamber systems
Electrocautery is the Most Common Hospital
Source of Pacemaker EMI
 Outcomes
– Oversensing–inhibition
– Undersensing (noise
reversion)
– Power on Reset
– Permanent loss of
pacemaker output
(if battery
voltage is low)
 Precautions
– Reprogram mode to
VOO/DOO, or place a
magnet over device
– Strategically place the
grounding plate
– Limit electrocautery
bursts to 1-second burst
every 10 seconds
– Use bipolar
electrocautery forceps
Transthoracic Defibrillation
 Outcome
– Inappropriate reprogramming
of the pulse
generator (POR)
– Damage to
pacemaker circuitry
 Precautions
– Position defibrillation paddles
apex-posterior (AP) and as far
from the pacemaker and leads
as possible
Magnetic Resonance Imaging (MRI) is Generally
Contraindicated in Patients with Pacemakers
 Outcomes
– Extremely high
pacing rate
– Reversion to
asynchronous pacing
 Precautions
– Program pacemaker
output low enough to
create persistent
non-capture, ODO
or OVO mode
Radiation Energy May Cause Permanent
Damage
 Certain kinds of radiation energy may cause damage to the semi-
conductor circuitry
– Ionizing radiation used for breast or
lung cancer therapy
 Damage can be permanent and requires
replacement of the pacemaker
Therapeutic Radiation May Cause Severe
Damage
 Outcomes:
– Pacemaker circuit damage
– Loss of output
– “Runaway”
 Precautions:
– Keep cumulative radiation
absorbed by the pacemaker
to less than 500 rads;
shielding may be required
– Check pacemaker after
radiation sessions for changes
in pacemaker function (can
be done transtelephonically)
Refractory & Blanking PeriodsRefractory & Blanking Periods
Voltage Deflections of the SensedVoltage Deflections of the Sensed
EGM in a PacemakerEGM in a Pacemaker
Pacemaker
Stimulus
Paced R wave
Post-pace T wave
Intrinsic R wave
T wave corresponding to
intrinsic R wave
2.5 mV
Refractory PeriodRefractory Period
Refractory Period Refractory Period Refractory Period
NO SENSING NO SENSING NO SENSING
Refractory PeriodRefractory Period
 A programmable period immediately following a
pacemaker stimulus or a sensed intrinsic R wave
during which the pacemaker does not react to sensed
events
 To prevent repeated sensing of the same intrinsic R
wave
 To prevent misidentification of T waves as intrinsic R
waves
 To prevent misidentification of effects of pacemaker
stimulus/evoked R wave
Usually programmed to 325 ms
To Prevent Oversensing
Afterpotential due to PolarizationAfterpotential due to Polarization
Afterpotential
Blanking PeriodBlanking Period
 The first portion of every refractory period
 Pacemaker is “blind” to any activity and no events can
be sensed
 Designed to prevent oversensing of pacing stimulus &
after-potential
Blanking Period
Refractory Period
ConductorTip Electrode Insulation Connector Pin
Pacing Lead ComponentsPacing Lead Components
 Conductor
 Connector Pin
 Insulation
 Electrode
ConnectorConnector
 Purpose
– Connects lead to IPG, and provides a conduit to:
• Deliver current from IPG to lead
• Return sensed cardiac signals to IPG
Connector
Connector -- IS-1 StandardConnector -- IS-1 Standard
 IS-1 Standard Connectors
 Sizes Prior to IS-1 Standard
– 3.2 mm low-profile connectors
– 5/6 mm connectors
Insulation -- TypeInsulation -- Type
 Insulation Types
– Silicone
– Polyurethane
– Fluoropolymers (PTFE, ETFE)
Electrodes -- Fixation MechanismElectrodes -- Fixation Mechanism
 Passive Fixation Mechanism – Endocardial
– Tined
– Finned
– Canted/curved
Electrodes – Fixation MechanismElectrodes – Fixation Mechanism
 Active Fixation Mechanism – Endocardial
– Fixed screw
– Extendible/retractable
Electrodes -- Fixation/VisualizationElectrodes -- Fixation/Visualization
Fluoroscopic Visual Quality of Active Fixation Leads
SureFixCapSureFix®
Extended Retracted Fixed Screw
space
Electrodes -- Fixation MechanismElectrodes -- Fixation Mechanism
 Fixation Mechanism – Myocardial/Epicardial
– Stab-in
– Screw-in
– Suture-on
Battery CapacityBattery Capacity
 A battery is a reservoir of electrical charge measured in
Coulombs
 Current is the amount of charge delivered per unit time
– 1 Ampere = 1 Coulomb per second
– 1 Coulomb = 1 Ampere x 1 second
 Qc = Battery Capacity is specified as the quantity of
charge it can deliver in AmpereHours (0.5 to 3
Amperehours)
Battery LifeBattery Life
 Battery Life
– Battery Life = Battery Capacity/Current Drain
– 2.0 Ah/25microamps = 80,000 hours = 9.3 years
 Battery properties
– Reliability – no premature failure
– High volumetric energy density – Small battery
volume with high storage capacity
– Low self-discharge rate
– High hermiticity – no gas generation during
operation
On the figure, the zone of non capture is
indicated by which number?
a)1
b)2
c)3
d)4
 Which of the following output settings best
represents the Chronaxie point on the strength-
duration curve when the Rheobase is 0.5V @
1.5ms?
a) 0.5V @ 1.5ms
b) 1.0V @ 0.5ms
c) 1.5V @ 0.1 ms
d) 2.0V @ 0.05ms
Sensing
 Sensing is the ability of the pacemaker to “see” when a natural (intrinsic)
depolarization is occurring
– Pacemakers sense cardiac depolarization by measuring
changes in electrical potential of myocardial cells between the
anode and cathode
 Intrinsic deflection on an EGM
occurs when a depolarization wave
passes directly under the
electrodes
 Two characteristics of the EGM
are:
– Signal amplitude
– Slew rate

More Related Content

What's hot

Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part iisalah_atta
 
Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)SolidaSakhan
 
ECG, step by step approach (Updated)
ECG, step by step approach (Updated)ECG, step by step approach (Updated)
ECG, step by step approach (Updated)Kerolus Shehata
 
Ventricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographyVentricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographySruthi Meenaxshi
 
Basics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerBasics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerPallab Nath
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1salah_atta
 
Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)salah_atta
 
LEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONLEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONPraveen Nagula
 
Components of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareComponents of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareRaghu Kishore Galla
 
Right heart catheters
Right heart cathetersRight heart catheters
Right heart cathetersRohitWalse2
 
Percutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPercutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPraveen Nagula
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming anddrskd6
 
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANBMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANDr Virbhan Balai
 

What's hot (20)

Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
 
Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)
 
ECG, step by step approach (Updated)
ECG, step by step approach (Updated)ECG, step by step approach (Updated)
ECG, step by step approach (Updated)
 
Cath hemodynamics vir
Cath hemodynamics virCath hemodynamics vir
Cath hemodynamics vir
 
Asd device closure
Asd device closureAsd device closure
Asd device closure
 
ECG
ECGECG
ECG
 
Ventricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographyVentricular Septal defects Echocardiography
Ventricular Septal defects Echocardiography
 
Electrophysiologic Study
Electrophysiologic StudyElectrophysiologic Study
Electrophysiologic Study
 
Temporary pacemaker toufiqur rahman
Temporary pacemaker toufiqur rahmanTemporary pacemaker toufiqur rahman
Temporary pacemaker toufiqur rahman
 
Basics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerBasics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & Pacemaker
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)
 
LEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONLEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATION
 
Components of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareComponents of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardware
 
Right heart catheters
Right heart cathetersRight heart catheters
Right heart catheters
 
fundamentals of pacemaker
fundamentals of pacemaker  fundamentals of pacemaker
fundamentals of pacemaker
 
Electrophysiology AVNRT
Electrophysiology AVNRTElectrophysiology AVNRT
Electrophysiology AVNRT
 
Percutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPercutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve Interventions
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
 
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANBMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
 

Viewers also liked

Pacemaker timing &amp; advanced dual chamber concepts
Pacemaker timing &amp; advanced dual chamber conceptsPacemaker timing &amp; advanced dual chamber concepts
Pacemaker timing &amp; advanced dual chamber conceptsSunil Reddy D
 
MMUH reg teaching: pacemakers falls blackouts
MMUH reg teaching: pacemakers falls blackoutsMMUH reg teaching: pacemakers falls blackouts
MMUH reg teaching: pacemakers falls blackoutsAndy Neill
 
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)MedicineAndFamily
 
Primary Prevention Of Sudden Cardiac Death - Role Of Devices
Primary Prevention Of Sudden Cardiac Death - Role Of DevicesPrimary Prevention Of Sudden Cardiac Death - Role Of Devices
Primary Prevention Of Sudden Cardiac Death - Role Of DevicesArindam Pande
 
Common presentations of pacemaker malfunction
Common presentations of pacemaker malfunction Common presentations of pacemaker malfunction
Common presentations of pacemaker malfunction Rashid Abuelhassan
 
History of ICDs (Internal Cardiac Defibrillators)
History of ICDs (Internal Cardiac Defibrillators)History of ICDs (Internal Cardiac Defibrillators)
History of ICDs (Internal Cardiac Defibrillators)Jose Osorio
 
Pacer ppt
Pacer pptPacer ppt
Pacer pptEM OMSB
 
Guidelines cardiac-pacing-slides
Guidelines cardiac-pacing-slidesGuidelines cardiac-pacing-slides
Guidelines cardiac-pacing-slidesBasem Enany
 
Implantable Cardioverter Defibrillator - the evidence behind the guidelines
Implantable Cardioverter Defibrillator - the evidence behind the guidelinesImplantable Cardioverter Defibrillator - the evidence behind the guidelines
Implantable Cardioverter Defibrillator - the evidence behind the guidelinesJunhao Koh
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemakersalman habeeb
 
Electrical testing of pacemaker
Electrical testing of pacemakerElectrical testing of pacemaker
Electrical testing of pacemakerRamachandra Barik
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N GSherry Knowles
 

Viewers also liked (20)

Pacemaker timing &amp; advanced dual chamber concepts
Pacemaker timing &amp; advanced dual chamber conceptsPacemaker timing &amp; advanced dual chamber concepts
Pacemaker timing &amp; advanced dual chamber concepts
 
Temporary pacing
Temporary pacingTemporary pacing
Temporary pacing
 
pacemaker
pacemakerpacemaker
pacemaker
 
MMUH reg teaching: pacemakers falls blackouts
MMUH reg teaching: pacemakers falls blackoutsMMUH reg teaching: pacemakers falls blackouts
MMUH reg teaching: pacemakers falls blackouts
 
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)
Implantable Cardioverter Defibrillators in Heart Failure (2005-11-02)
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Primary Prevention Of Sudden Cardiac Death - Role Of Devices
Primary Prevention Of Sudden Cardiac Death - Role Of DevicesPrimary Prevention Of Sudden Cardiac Death - Role Of Devices
Primary Prevention Of Sudden Cardiac Death - Role Of Devices
 
Pacemaker Rhythms
Pacemaker RhythmsPacemaker Rhythms
Pacemaker Rhythms
 
Common presentations of pacemaker malfunction
Common presentations of pacemaker malfunction Common presentations of pacemaker malfunction
Common presentations of pacemaker malfunction
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
History of ICDs (Internal Cardiac Defibrillators)
History of ICDs (Internal Cardiac Defibrillators)History of ICDs (Internal Cardiac Defibrillators)
History of ICDs (Internal Cardiac Defibrillators)
 
Pacer ppt
Pacer pptPacer ppt
Pacer ppt
 
Guidelines cardiac-pacing-slides
Guidelines cardiac-pacing-slidesGuidelines cardiac-pacing-slides
Guidelines cardiac-pacing-slides
 
Implantable Cardioverter Defibrillator - the evidence behind the guidelines
Implantable Cardioverter Defibrillator - the evidence behind the guidelinesImplantable Cardioverter Defibrillator - the evidence behind the guidelines
Implantable Cardioverter Defibrillator - the evidence behind the guidelines
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
Electrical testing of pacemaker
Electrical testing of pacemakerElectrical testing of pacemaker
Electrical testing of pacemaker
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemakers
PacemakersPacemakers
Pacemakers
 

Similar to Basic technical concepts in cardiac pacing

Biopotentials
BiopotentialsBiopotentials
Biopotentialsstooty s
 
NEUROMUSCULAR JUNCTION.ppt
NEUROMUSCULAR JUNCTION.pptNEUROMUSCULAR JUNCTION.ppt
NEUROMUSCULAR JUNCTION.pptSYED MASOOD
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptopijics
 
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區Taiwan Heart Rhythm Society
 
Electrical Agents
Electrical AgentsElectrical Agents
Electrical AgentsJLS10
 
Assessment of cardiac contractility using langendorff apparatus connected to ...
Assessment of cardiac contractility using langendorff apparatus connected to ...Assessment of cardiac contractility using langendorff apparatus connected to ...
Assessment of cardiac contractility using langendorff apparatus connected to ...dina merzeban
 
Electrocardiogram
ElectrocardiogramElectrocardiogram
ElectrocardiogramTarek Selim
 
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerPacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerMathavan N
 
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)DikshaGondkar
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)nishanthi p
 
Electronic instrumentation in NMT
Electronic instrumentation in NMTElectronic instrumentation in NMT
Electronic instrumentation in NMTSUMAN GOWNDER
 
Vskills certified it support professional reading material
Vskills certified it support professional reading materialVskills certified it support professional reading material
Vskills certified it support professional reading materialVskills
 

Similar to Basic technical concepts in cardiac pacing (20)

PACEMAKER BASIC AND TIMING CYCLE .pptx
PACEMAKER BASIC AND TIMING CYCLE .pptxPACEMAKER BASIC AND TIMING CYCLE .pptx
PACEMAKER BASIC AND TIMING CYCLE .pptx
 
Presentation 21 4-11electrotherapy
Presentation 21 4-11electrotherapyPresentation 21 4-11electrotherapy
Presentation 21 4-11electrotherapy
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
NEUROMUSCULAR JUNCTION.ppt
NEUROMUSCULAR JUNCTION.pptNEUROMUSCULAR JUNCTION.ppt
NEUROMUSCULAR JUNCTION.ppt
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
 
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130907北區
 
Electrical Agents
Electrical AgentsElectrical Agents
Electrical Agents
 
Basics of Electrocardiography(ECG)
Basics of Electrocardiography(ECG)Basics of Electrocardiography(ECG)
Basics of Electrocardiography(ECG)
 
pacemaker
pacemakerpacemaker
pacemaker
 
Basic ECG notes
Basic ECG notesBasic ECG notes
Basic ECG notes
 
Assessment of cardiac contractility using langendorff apparatus connected to ...
Assessment of cardiac contractility using langendorff apparatus connected to ...Assessment of cardiac contractility using langendorff apparatus connected to ...
Assessment of cardiac contractility using langendorff apparatus connected to ...
 
Electrocardiogram
ElectrocardiogramElectrocardiogram
Electrocardiogram
 
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerPacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
 
Eg24842846
Eg24842846Eg24842846
Eg24842846
 
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)
Current Electricity- Dr. Diksha Gondkar, MPT (Neuro)
 
ECG (1).pptx
ECG (1).pptxECG (1).pptx
ECG (1).pptx
 
ECG (1).pptx
ECG (1).pptxECG (1).pptx
ECG (1).pptx
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
 
Electronic instrumentation in NMT
Electronic instrumentation in NMTElectronic instrumentation in NMT
Electronic instrumentation in NMT
 
Vskills certified it support professional reading material
Vskills certified it support professional reading materialVskills certified it support professional reading material
Vskills certified it support professional reading material
 

More from Sunil Reddy D

Sudden cardiac death
Sudden cardiac deathSudden cardiac death
Sudden cardiac deathSunil Reddy D
 
Risk stratification of scd
Risk stratification of scdRisk stratification of scd
Risk stratification of scdSunil Reddy D
 
Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmiasNon invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmiasSunil Reddy D
 
Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmias Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmias Sunil Reddy D
 
Non invasive evaluation of arrhyhtmias
Non invasive evaluation of arrhyhtmiasNon invasive evaluation of arrhyhtmias
Non invasive evaluation of arrhyhtmiasSunil Reddy D
 
Newer antiarrhythmics
Newer antiarrhythmicsNewer antiarrhythmics
Newer antiarrhythmicsSunil Reddy D
 

More from Sunil Reddy D (10)

Sudden cardiac death
Sudden cardiac deathSudden cardiac death
Sudden cardiac death
 
Risk stratification of scd
Risk stratification of scdRisk stratification of scd
Risk stratification of scd
 
Proarrhythmia
ProarrhythmiaProarrhythmia
Proarrhythmia
 
Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmiasNon invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmias
 
Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmias Non invasive evaluation of arrhythmias
Non invasive evaluation of arrhythmias
 
Non invasive evaluation of arrhyhtmias
Non invasive evaluation of arrhyhtmiasNon invasive evaluation of arrhyhtmias
Non invasive evaluation of arrhyhtmias
 
Newer antiarrhythmics
Newer antiarrhythmicsNewer antiarrhythmics
Newer antiarrhythmics
 
Antiarrhythmics
AntiarrhythmicsAntiarrhythmics
Antiarrhythmics
 
Antiarrhythmics
AntiarrhythmicsAntiarrhythmics
Antiarrhythmics
 
Antiarrhythmics
AntiarrhythmicsAntiarrhythmics
Antiarrhythmics
 

Recently uploaded

College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Suratnarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 

Recently uploaded (20)

College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 

Basic technical concepts in cardiac pacing

  • 1. Basic Technical Concepts inBasic Technical Concepts in Cardiac PacingCardiac Pacing Dr D Sunil Reddy Consultant Cardiologist KIMS Hospital
  • 3. Myocardial StimulationMyocardial Stimulation  An artificial electrical stimulus excites cardiac tissue by the creation of an electrical field at the interface of the stimulating electrode and the myocardium  The electric field should be strong enough and should last long enough to initiate action potentials in the cells at the electrode-tissue interface
  • 4. Myocardial StimulationMyocardial Stimulation  The AP’s at the site of stimulation result in AP’s in the neighbouring areas of the myocardium resulting in a wave of AP’s (depolarization) propagating away from the site of stimulation Time (Milliseconds) 100 200 300 400 500 Phase 2 Phase 1 Phase 3 Phase 4 TransmembranePotential (Millivolts) -50 0 50 -100 Phase0 Threshold  An artificial electrical stimulus excites cardiac tissue by the creation of an electrical field at the interface of the stimulating electrode and the myocardium  The electric field should be strong enough and should last long enough to initiate action potentials in the cells at the electrode-tissue interface
  • 5. Electric Field – Current DensityElectric Field – Current Density TheoryTheory  Electric Field Theory – A Minimum Voltage/cmis required to trigger a self-propagating wave of depolarization  Current Density Theory – A Minimum Current/cm2 is required to trigger a self-propagating wave of depolarization  Stimulation threshold is a function of Voltage/cm or Current/ cm2 that is induced in the myocardium beneath the stimulating electrode  The two theories are related by Ohm’s Law
  • 6. The Implantable Pacemaker SystemThe Implantable Pacemaker System  The Implantable Pulse Generator (IPG) : metal can (titanium) containing electronics/battery & an electrode or lead connector header  Lead : Electrical connection between the pacemaker & the heart
  • 7. The Pacemaker CircuitThe Pacemaker Circuit  Stimulation of cardiac tissue using electric pulses – IPG – Lead – insulated current conductor (s) & electrodes to transmit pulses to heart tissue and measure or sense electrical activity in the heart Heart Lead IPG Completion of circuit through Lead or Body tissue
  • 8. A Unipolar Pacing System Contains a Lead withA Unipolar Pacing System Contains a Lead with Only One Electrode Within the Heart; In ThisOnly One Electrode Within the Heart; In This System, the Impulse:System, the Impulse:  Flows through the tip electrode (cathode)  Stimulates the heart  Returns through body fluid and tissue to the IPG (anode) Cathode Anode -
  • 9. Anode  Flows through the tip electrode located at the end of the lead wire  Stimulates the heart  Returns to the ring electrode above the lead tip A Bipolar Pacing System Contains a Lead with TwoA Bipolar Pacing System Contains a Lead with Two Electrodes Within the Heart. In This System, the Impulse:Electrodes Within the Heart. In This System, the Impulse: Cathode Tip electrode coil Indifferent electrode coil
  • 11. Electrodes -- Fixation Mechanism  Passive Fixation Mechanism – Endocardial – Tined – Finned – Canted/curved
  • 12. Electrodes – Fixation Mechanism  Active Fixation Mechanism – Endocardial – Fixed screw – Extendible/retractable
  • 13. Electrodes -- Fixation Mechanism  Fixation Mechanism – Myocardial/Epicardial – Stab-in – Screw-in – Suture-on
  • 14. Single-Chamber SystemSingle-Chamber System  The pacing lead is implanted in the atrium or ventricle, depending on the chamber to be paced and sensed
  • 15.  One lead implanted in the atrium and one in the ventricle Dual-Chamber Systems Have Two Leads:Dual-Chamber Systems Have Two Leads:
  • 16. A pacing system can be thought of a standard electrical circuit: The pacemaker supplies the voltage. Current (electrons) flow down the conductor to the lead tip or cathode (-) Where the lead tip touches the myocardium, electrical resistance is produced. The current then flows through the body tissues to the anode (+) and back to the battery. All of these things are required for current to flow.
  • 17. Pacing Circuit ParametersPacing Circuit Parameters  Voltage  Current  Impedance (resistance)  Energy
  • 18. VoltageVoltage Pushes electrons through an electric circuit resulting in electric current through the circuit
  • 19. VoltageVoltage  Voltage is the force or “push” that causes electrons to move through a circuit  In a pacing system, voltage is: – Measured in volts – Represented by the letter “V” – Provided by the pacemaker battery – Often referred to as amplitude
  • 20. CurrentCurrent Current is the flow of charge or electrons through a circuit
  • 21. CurrentCurrent  The flow of electrons in a completed circuit  In a pacing system, current is: – Measured in mA (milliamps) – Represented by the letter “I” – Determined by the amount of electrons that move through a circuit
  • 22. Resistance or ImpedanceResistance or Impedance Opposition offered by a circuit to the flow of current
  • 23. ImpedanceImpedance  The opposition to current flow in a circuit  In a pacing system, impedance is: – Measured in ohms – Represented by the letter “R” (Ω for numerical values) – The measurement of the sum of all resistance to the flow of current
  • 24. Ohm’s LawOhm’s Law Voltage = Current X Resistance Current = Voltage ÷Resistance Resistance = Voltage ÷ Current
  • 25. The Pacemaker StimulusThe Pacemaker Stimulus Time 5 Volts 5 Volts 0.5 ms 1 sec Pacing Stimulus Voltage or Amplitude – 5 Volts Pulse Width – 0.0005 seconds or 0.5 milliseconds Pacing Rate – One stimulus per second or 60 stimuli (beats) per minute Voltage
  • 26. The Pacing PulseThe Pacing Pulse tt Pacing Pulse Pulse Duration (Width) OutputVoltage V = Pulse Amplitude in Volts (V) (say 2.5 V) t = Pulse Duration or Width in milliseconds (ms) (say 0.5 ms) R = Impedance of Pacing Circuit (ohms) (say 500 ohms) I = V/R = Current through pacing circuit (mA) = 2.5 V/ 500 ohms = 0.005 A = 5 mA E = Energy supplied by Pulse to the Pacing Circuit and Cardiac Tissue = V . I . t = I2 Rt = V2 t/R = 2.5 V . 5 mA . 0.5 ms = 6.25 micro Joules V t
  • 27. Stimulation ThresholdStimulation Threshold  Pacing Voltage Threshold – The minimum pacing voltage at any given pulse width required to consistently achieve myocardial depolarization outside the heart’s refractory period
  • 29. The Strength-Duration RelationThe Strength-Duration Relation  The intensity of an electrical stimulus (Energy, Voltage, Current, Charge) required to capture (non-refractory) cardiac tissue is dependent on the duration for which the electrical stimulus is applied (i.e. pulse width)
  • 30. The Voltage-Strength DurationThe Voltage-Strength Duration CurveCurve  Stimulus Voltage & Pulse Width have an exponential relationship  At short pulse widths (<0.25 ms)the curve rises sharply (i.e. small reductions in pulse width result in large increases in the voltage threshold)  At long pulse widths (>1.0 ms)the curve is flat (i.e. the voltage threshold does not reduce with increasing pulse width) Duration Pulse Width (ms) .50 1.0 1.5 2.0 .25StimulationThreshold(Volts) 0.25 1.0 1.5 Capture
  • 31. Rheobase & ChronaxieRheobase & Chronaxie Rheobase Voltage = Voltage Threshold at infinite Pulse widths (e.g. 2 ms) – also known as fundamental threshold Chronaxie Point = Pulse Width threshold at twice the rheobase voltage (i.e. 1 V & 0.3 ms) Energy of Pacing Pulse At rheobase = 0.5V. 2ms. 1mA = 1 MicroJoule At Chronaxie = 1V. 0.3 ms. 2mA = 0.6 MicroJoules At PW=0.25 ms, = 1.5V. 0.25ms. 3mA = 1.125 MicroJoules The chronaxie point approximates the point of minimum threshold energy on the Strength-Duration Curve Duration Pulse Width (ms) StimulationThreshold(Volts) 0.2 0.6 1.0 Energy Strength Duration Curve Capture 0.4 0.5 1.0 2.0 0.8 1.2 1.4 Rheobase Voltage Chronaxie Point
  • 32. Goals of Pacemaker OutputGoals of Pacemaker Output ProgrammingProgramming  Consistent capture & Patient Safety is ensured  Battery drain minimized, Pacemaker longevity maximized
  • 33. Programming Chronic PacemakerProgramming Chronic Pacemaker OutputOutput Duration Pulse Width (ms) StimulationThreshold(Volts) 0.2 0.6 1.0 Energy Strength Duration Curve • Capture Adequate Safety Margin while minimizing Pulse Energy Operate around the chronaxie point Programming long pulse widths (& low Voltages) increases Pulse energy but hardly increases safety margin Programming high voltages (with short Pulse widths) increases Pulse energy but hardly increases safety margin For a 2 times safety margin : Output = 2 x chronaxie voltage at chronaxie pulse width (2V, 0.35 ms) Output = 3 x pulse duration threshold at twice chronaxie voltage (2V, 0.51 ms) 0.4 0.5 1.0 2.0 0.8 1.2 1.4 Rheobase Chronaxie
  • 34. Determination of StimulationDetermination of Stimulation Threshold during Implant or Follow-upThreshold during Implant or Follow-up  Pace the heart at – Rate higher than sinus or intrinsic rate – Stimulus amplitude and pulse width that ensure capture (usually 5 V and 0.5 ms)  Gradually reduce stimulus amplitude while maintaining pulse width constant till capture is lost  The Stimulation threshold is specified by the minimum stimulus amplitude at which capture consistently occurs at a given stimulus pulse width
  • 35. Programming Pacemaker OutputProgramming Pacemaker Output  Acute – Immediately post-implant  Chronic – Approx. 6 to 8 weeks post implant  Goal : To ensure consistent capture despite potential changes in the SD curve while minimizing the energy delivered by the pulse
  • 36. Evolution of Pacing ThresholdEvolution of Pacing ThresholdVoltage Threshold(V) Observation Time (weeks) Acute Phase Chronic Phase Safety Margin 6 5 4 3 2 1 0 4 8 12 16
  • 37. Programmable ParametersProgrammable Parameters  Pacemaker Pulse Output or Amplitude – 0.5 V, 1.0 V, 1.5 V…., 7.5 V  Pacemaker Pulse Width – 0.03, 0.06, …., 0.25, 0.5,…, 1.0 milliseconds
  • 38. Pacing Impedance - ConsiderationsPacing Impedance - Considerations  Maximize Pacemaker Longevity – Reduce current drain – Maintain relatively high impedance  Ensure consistent capture – Appropriate voltage & current are available at the electrode-tissue interface required to stimulate tissue
  • 39. I pacing Components of Pacing ImpedanceComponents of Pacing Impedance Rcoil Zpolarisation -electrode tissue interface Rtissue Electrode Tissue Interface Pulse Generator V pulse I pacing
  • 40. RRcoil -coil - Lead Conductor ResistanceLead Conductor Resistance – Lead Conductor Impedance • Reduces the voltage & energy available at the tissue for pacing • Generates waste heat • Designed to be low conductor resistance
  • 41. Conductor ResistanceConductor Resistance Total Coil Voltage Voltage impedance resistance at tip loss (Ω)(Ω) (V) % 600 50 2.3 8.3 750 200 1.8 26.7 1200 650 1.1 54.2 1200 50 2.4 4.2 Output Voltage = 2.5 V Impact of conductor coil resistance upon available tip electrode voltage
  • 42. Conductor Resistance V mV 2.5 2500 Total imp 600 Current I =V/R 4.166666667 mAConductor imp 50 Voltage drop V = IR 208.3333333 mV 0.208333333 V
  • 43. What is the Polarization Effect – PolarizationWhat is the Polarization Effect – Polarization Impedance ?Impedance ? As the pacing pulse begins, electrons from the pacemaker battery flow to the lead tip and positively charged ions from the tissue are attracted to the lead tip. Initially, the movement of these negatively charged ions results in the flow of current from the electrode into the myocardium. As the pacing pulse continues, positively charged ions surround the electrode tip. This produces a positively charged layer on the electrode called polarization. Polarization can impede current flow from the electrode into the tissue.
  • 44. Polarization Impedance &Polarization Impedance & Electrode Surface AreaElectrode Surface Area  As electrode surface area goes up – Polarization impedance decreases  As electrode surface area goes down – Polarization impedance increases
  • 46. Factors affecting PolarizationFactors affecting Polarization ImpedanceImpedance  Pulse Width  Surface area of the electrode tip – Larger the surface area – Lower the Polarization
  • 47. Electrode – Tissue InterfaceElectrode – Tissue Interface ImpedanceImpedance  Resistance to current flow from body tissue at the electrode-tissue interface
  • 48. Electrode – Tissue InterfaceElectrode – Tissue Interface ImpedanceImpedance  Smaller the surface area of the tip electrode, the higher the current density (or electric field strength) at the electrode-tissue interface  The higher the field strength, the greater the pacing efficiency & lower the threshold
  • 49. Electrode – Tissue InterfaceElectrode – Tissue Interface ImpedanceImpedance  Smaller the geometric size of the tip electrode, the higher the resistance between electrode and tissue PacingImpedance(Ohms) 0 500 1000 1500 0 1 2 3 4 5.5 6 Geometric Tip Electrode Surface Area (mm2 ) Size = Impedance
  • 50. The Ideal Stimulating ElectrodeThe Ideal Stimulating Electrode  Needs large surface area to reduce Polarization Impedance  Needs small size to maximize Electric Field Strength and Stimulation Efficacy  Needs small size to maximize Electrode-tissue resistance and minimize pacemaker current drain
  • 51. Electrodes -- Surface StructureElectrodes -- Surface Structure  Porous Electrode Surface CapSure® 8.0 mm2 Porous Electrode CapSure® SP Novus 5.8 mm2 Platinized Porous Electrode CapSure® Z Novus 1.2 mm2 Platinized Porous Electrode 15KV x2500 12.0V MDT
  • 52. Analogy: Lead with low impedance and poor efficiency Lead Impedance I
  • 53. Lead Impedance II (Conductor-) Resistance Analogy: Lead with high impedance in the conductor and poor efficiency
  • 54. Lead Impedance III Analogy: Lead with big tip-tissue impedance low conductor impedance and increased efficiency
  • 55. The Chronic Virtual ElectrodeThe Chronic Virtual Electrode IMPLANT CHRONIC (8 weeks or longer) Excitable Cardiac Tissue Non-Excitable Fibrotic Tissue Excitable Cardiac Tissue
  • 56. Evolution of Pacing ThresholdEvolution of Pacing ThresholdVoltage Threshold(V) Observation Time (weeks) Acute Phase Chronic Phase Safety Margin 6 5 4 3 2 1 0 4 8 12 16
  • 57. Electrodes -- Steroid ElutionElectrodes -- Steroid Elution Tines for Stable Fixation Silicone Rubber Plug Containing Steroid Porous, Platinized Tip for Steroid Elution  Type - Steroid in matrix
  • 58. Electrodes -- Steroid ElutionElectrodes -- Steroid Elution  Effect of Steroid on Stimulation Thresholds Pulse Width = 0.5 msec 0 3 6 Implant Time (Weeks) Textured Metal Electrode Smooth Metal Electrode 1 2 3 4 5 Steroid-Eluting Electrode 0 1 2 4 5 7 8 9 10 11 12 Volts
  • 59. Factors that affect StimulationFactors that affect Stimulation ThresholdThreshold  Eating, Sleeping, Exercise, Medications, Changes in Cardiac condition – 30 to 50 % during the day  Drugs e.g. Steroids reduce stimulation threshold by reducing inflammation  Sympathomimetic drugs decrease threshold  Amiodarone, Class I A (quinidine, procainamide), Class I B (Mexilitene)  Hypokalemia increases threshold (diuretics)  Hypocalcemia increases threshold  Hypoxia & Hypercapnia – Increase threshold  Electrolyte imbalance & pH e.g. acidosis and alkalosis increase threshold
  • 60. Typical Pacing CircuitTypical Pacing Circuit ImpedancesImpedances  300 to 1200 ohms  May be measured at implant time with a Pacing System Analyzer (PSA) at implant time  May be measured through telemetry with a pacemaker programmer
  • 61. Pacing Impedance Values Will Change DuePacing Impedance Values Will Change Due to:to:  Insulation breaks  Wire fractures
  • 62. An Insulation Break Around the Lead WireAn Insulation Break Around the Lead Wire Can Cause Impedance Values to FallCan Cause Impedance Values to Fall  Insulation breaks expose the wire to body fluids which have a low resistance and cause impedance values to fall  Current drains through the insulation break into the body which depletes the battery  An insulation break can cause impedance values to fall below 300 Ω Insulation break Decreased resistance
  • 63. A Wire Fracture Within the Insulating SheathA Wire Fracture Within the Insulating Sheath May Cause Impedance Values to RiseMay Cause Impedance Values to Rise  Impedance values across a break in the wire will increase  Current flow may be too low to be effective  Impedance values may exceed 3,000 Ω Lead wire fracture Increased resistance
  • 64. Unipolar &Bipolar StimulationUnipolar &Bipolar Stimulation + - + - - + Anode is IPG Case Cathode is lead tip electrode - + Anode is lead proximal electrode Cathode is lead tip electrode
  • 65. Unipolar & Bipolar StimulationUnipolar & Bipolar Stimulation Unipolar Stimulation– Large Stimulus Artifact Bipolar Stimulation – Small Stimulus Artifact There is a greater chance of pacemaker pocket muscle stimulation with Unipolar Stimulation
  • 66. The Pacemaker BatteryThe Pacemaker Battery  Pacemakers typically use Lithium-Iodide batteries  Most LiI batteries have a Beginning of Life (BOL) value of 2.8 volts  Recommended Replacement Time (ERI/RRT) value of 2.6 V  End of Service (EOL/EOS) value of 2.5 V  As battery depletes its internal resistance goes up – (BOL = 100 ohms , EOL > 5000 ohms )
  • 67. Recommended Replacement Time RRT (ERI) 2.8V- 2.6V- 2.5V- BOL RRT 2.6V EOS 2.5V Time Lithium Iodine Battery depletion BOL – Beginning of Life RRT – Recommended Replacement Time EOS - End of Service
  • 68. Battery LifeBattery Life  Battery Life – Battery Life = Battery Capacity/Current Drain – 2.0 Ah/25microamps = 80,000 hours = 9.3 years  Battery properties – Reliability – no premature failure – High volumetric energy density – Small battery volume with high storage capacity – Low self-discharge rate – High hermiticity – no gas generation during operation
  • 69. Pacemaker LongevityPacemaker Longevity  Energy = (V.I.t) = (V2 /R) . t Joules  Reducing V by a factor of 2 reduces E by a factor of 4  Increasing Z by a factor by 2 reduces E by a factor of 2
  • 70. Pacemaker LongevityPacemaker Longevity  High Outputs and pulse widths are the primary cause for reduced pacemaker longevity • Output Voltages higher than the Battery Voltage (2.8 V) require the use of voltage doublers that use high battery energy • Long Pulse Widths Reduce pacing efficiency due to increased polarization impedance
  • 71. Sensing Intracardiac ElectricalSensing Intracardiac Electrical ActivityActivity
  • 72. A Pacemaker Must Be Able to Sense and Respond to Cardiac Rhythms  Accurate sensing enables the pacemaker to determine whether or not the heart has created a beat on its own  The pacemaker is usually programmed to respond with a pacing impulse only when the heart fails to produce an intrinsic beat
  • 73. Intracardiac Electrical SignalsIntracardiac Electrical Signals  Electrical currents that arise in the myocardium during depolarization and repolarization  A myocardial electrode - records voltage difference wrt reference electrode when the myocardium under the electrode undergoes depolarization or repolarization  The electrical activity measured by such an electrode (which is in direct contact with cardiac tissue) – local tissue electrical activity- Intracardiac Electrogram or EGM
  • 75. Depolarization WaveDepolarization Wave Processed byProcessed by DeviceDevice Processed byProcessed by DeviceDevice The EGM Signal The signal from a depolarization wave passing between two electrodes
  • 76. Intracardiac ElectrogramIntracardiac Electrogram R wave of the EGM indicating depolarisation of cardiac ventricular tissue at lead tip – Roughly corresponds to R wave of the ECG that represents depolarisation of the ventricles T wave of the EGM indicating repolarisation of cardiac ventricular tissue at lead tip – Roughly corresponds to T wave of the ECG that represents repolarisation of the ventricles
  • 77. Voltage Deflections of the SensedVoltage Deflections of the Sensed EGM in a PacemakerEGM in a Pacemaker Pacemaker Stimulus Paced R wave Post-pace T wave Intrinsic R wave T wave corresponding to intrinsic R wave
  • 78. Undersensing . . .  Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately Intrinsic beat not sensed Scheduled pace delivered VVI / 60
  • 79. Oversensing  An electrical signal other than the intended P or R wave is detected Marker channel shows intrinsic activity... ...though no activity is present VVI / 60
  • 80.  Measured by: – Amplitude • Peak-to-peak measurement (height) of deflection • Measured in Millivolts (mV) – Slew Rate • Speed of deflection change over time • Measured in volts per second (V/s) EGM amplitude
  • 81. EGM Amplitude & Slew RateEGM Amplitude & Slew Rate Amplitude of the EGM R wave = dV Slew Rate of the EGM R wave= dV/dt
  • 82. EGM amplitude & slew rate – ideal values Atrium Ventricle Slew Rate 0.5 volts/sec 0.75 volts/sec Amplitude >1.5 mV >5.0 mV
  • 83. Which of the these complexes shows the highest slew rate? a)1 b)2 c)3 d)4
  • 84. Frequency (Hz) Accurate Sensing Requires That Extraneous Signals Be Filtered Out
  • 85. Effect of Filtering on EGMEffect of Filtering on EGM FILTER •Filters out frequencies below 5 Hz (T waves) and above 50 Hz (myopotentials) •Enhances frequencies between 10 Hz and 50 Hz (R waves) with maximum enhancement at around 30 Hz frequency • Low Slew Rate implies Low Frequency implies Less enhancement due to filtering Level Detector
  • 86. Level Detector or SensitivityLevel Detector or Sensitivity SettingSetting This is a value specified to the pacemaker in millivolts through programming. All ventricular EGM deflections AFTER FILTERING that exceed the sensitivity setting will be identified by the pacemaker as intrinsic R waves The typical sensitivity setting that is programmed for Ventricular sensing is 2.5 mV
  • 87. Sensitivity The Greater the Number, the Less Sensitive the Device to Intracardiac Events
  • 89. Sensitivity SettingSensitivity Setting Sensitivity settings less than 2.5 mv – High sensitivity – can lead to oversensing Sensitivity settings greater than 2.5 mV – Low sensitivity – can lead to undersensing Amplitude(mV) Amplitude(mV) Time Time 5.0 2.5 1.25 5.0 2.5 1.25
  • 90. Factors That May Affect Sensing Are:  Lead polarity (unipolar vs. bipolar)  Lead integrity – Insulation break – Wire fracture  EMI – Electromagnetic Interference
  • 91. Unipolar Sensing  Produces a large potential difference due to: – A cathode and anode that are farther apart than in a bipolar system _
  • 92. Bipolar Sensing  Produces a smaller potential difference due to the short interelectrode distance – Electrical signals from outside the heart such as myopotentials are less likely to be sensed
  • 93. An Insulation Break May Cause Both Undersensing or Oversensing  Undersensing occurs when inner and outer conductor coils are in continuous contact – Signals from intrinsic beats are reduced at the sense amplifier and amplitude no longer meets the programmed sensing value –  Oversensing occurs when inner and outer conductor coils make intermittent contact – Signals are incorrectly interpreted as P or R waves
  • 94. Wire Fracture Can Cause Both Undersensing and Oversensing  Undersensing occurs when the cardiac signal is unable to get back to the pacemaker – intrinsic signals cannot cross the wire fracture  Oversensing occurs when the severed ends of the wire intermittently make contact, which creates potentials interpreted by the pacemaker as P or R waves Fracture in one filament leads to an increase in resistance
  • 95. Electromagnetic Interference  Interference is caused by electromagnetic energy with a source that is outside the body  Electromagnetic fields that may affect pacemakers are radio-frequency waves – 50-60 Hz are most frequently associated with pacemaker interference  Few sources of EMI are found in the home or office but several exist in hospitals
  • 96. Oversensing May Occur When EMI Signals Are Incorrectly Interpreted as P Waves or R Waves  Pacing rates will vary as a result of EMI: – Rates will accelerate if sensed as P waves in dual-chamber systems (P waves are “tracked”) – Rates will be low or inhibited if sensed in single- chamber systems, or on ventricular lead in dual-chamber systems
  • 97. Electrocautery is the Most Common Hospital Source of Pacemaker EMI  Outcomes – Oversensing–inhibition – Undersensing (noise reversion) – Power on Reset – Permanent loss of pacemaker output (if battery voltage is low)  Precautions – Reprogram mode to VOO/DOO, or place a magnet over device – Strategically place the grounding plate – Limit electrocautery bursts to 1-second burst every 10 seconds – Use bipolar electrocautery forceps
  • 98. Transthoracic Defibrillation  Outcome – Inappropriate reprogramming of the pulse generator (POR) – Damage to pacemaker circuitry  Precautions – Position defibrillation paddles apex-posterior (AP) and as far from the pacemaker and leads as possible
  • 99. Magnetic Resonance Imaging (MRI) is Generally Contraindicated in Patients with Pacemakers  Outcomes – Extremely high pacing rate – Reversion to asynchronous pacing  Precautions – Program pacemaker output low enough to create persistent non-capture, ODO or OVO mode
  • 100. Radiation Energy May Cause Permanent Damage  Certain kinds of radiation energy may cause damage to the semi- conductor circuitry – Ionizing radiation used for breast or lung cancer therapy  Damage can be permanent and requires replacement of the pacemaker
  • 101. Therapeutic Radiation May Cause Severe Damage  Outcomes: – Pacemaker circuit damage – Loss of output – “Runaway”  Precautions: – Keep cumulative radiation absorbed by the pacemaker to less than 500 rads; shielding may be required – Check pacemaker after radiation sessions for changes in pacemaker function (can be done transtelephonically)
  • 102. Refractory & Blanking PeriodsRefractory & Blanking Periods
  • 103. Voltage Deflections of the SensedVoltage Deflections of the Sensed EGM in a PacemakerEGM in a Pacemaker Pacemaker Stimulus Paced R wave Post-pace T wave Intrinsic R wave T wave corresponding to intrinsic R wave 2.5 mV
  • 104. Refractory PeriodRefractory Period Refractory Period Refractory Period Refractory Period NO SENSING NO SENSING NO SENSING
  • 105. Refractory PeriodRefractory Period  A programmable period immediately following a pacemaker stimulus or a sensed intrinsic R wave during which the pacemaker does not react to sensed events  To prevent repeated sensing of the same intrinsic R wave  To prevent misidentification of T waves as intrinsic R waves  To prevent misidentification of effects of pacemaker stimulus/evoked R wave Usually programmed to 325 ms To Prevent Oversensing
  • 106. Afterpotential due to PolarizationAfterpotential due to Polarization Afterpotential
  • 107. Blanking PeriodBlanking Period  The first portion of every refractory period  Pacemaker is “blind” to any activity and no events can be sensed  Designed to prevent oversensing of pacing stimulus & after-potential Blanking Period Refractory Period
  • 108. ConductorTip Electrode Insulation Connector Pin Pacing Lead ComponentsPacing Lead Components  Conductor  Connector Pin  Insulation  Electrode
  • 109. ConnectorConnector  Purpose – Connects lead to IPG, and provides a conduit to: • Deliver current from IPG to lead • Return sensed cardiac signals to IPG Connector
  • 110. Connector -- IS-1 StandardConnector -- IS-1 Standard  IS-1 Standard Connectors  Sizes Prior to IS-1 Standard – 3.2 mm low-profile connectors – 5/6 mm connectors
  • 111. Insulation -- TypeInsulation -- Type  Insulation Types – Silicone – Polyurethane – Fluoropolymers (PTFE, ETFE)
  • 112. Electrodes -- Fixation MechanismElectrodes -- Fixation Mechanism  Passive Fixation Mechanism – Endocardial – Tined – Finned – Canted/curved
  • 113. Electrodes – Fixation MechanismElectrodes – Fixation Mechanism  Active Fixation Mechanism – Endocardial – Fixed screw – Extendible/retractable
  • 114. Electrodes -- Fixation/VisualizationElectrodes -- Fixation/Visualization Fluoroscopic Visual Quality of Active Fixation Leads SureFixCapSureFix® Extended Retracted Fixed Screw space
  • 115. Electrodes -- Fixation MechanismElectrodes -- Fixation Mechanism  Fixation Mechanism – Myocardial/Epicardial – Stab-in – Screw-in – Suture-on
  • 116. Battery CapacityBattery Capacity  A battery is a reservoir of electrical charge measured in Coulombs  Current is the amount of charge delivered per unit time – 1 Ampere = 1 Coulomb per second – 1 Coulomb = 1 Ampere x 1 second  Qc = Battery Capacity is specified as the quantity of charge it can deliver in AmpereHours (0.5 to 3 Amperehours)
  • 117. Battery LifeBattery Life  Battery Life – Battery Life = Battery Capacity/Current Drain – 2.0 Ah/25microamps = 80,000 hours = 9.3 years  Battery properties – Reliability – no premature failure – High volumetric energy density – Small battery volume with high storage capacity – Low self-discharge rate – High hermiticity – no gas generation during operation
  • 118. On the figure, the zone of non capture is indicated by which number? a)1 b)2 c)3 d)4
  • 119.  Which of the following output settings best represents the Chronaxie point on the strength- duration curve when the Rheobase is 0.5V @ 1.5ms? a) 0.5V @ 1.5ms b) 1.0V @ 0.5ms c) 1.5V @ 0.1 ms d) 2.0V @ 0.05ms
  • 120. Sensing  Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring – Pacemakers sense cardiac depolarization by measuring changes in electrical potential of myocardial cells between the anode and cathode  Intrinsic deflection on an EGM occurs when a depolarization wave passes directly under the electrodes  Two characteristics of the EGM are: – Signal amplitude – Slew rate

Editor's Notes

  1. Permanent pacing leads have five major components: 1. The conductor(s) 2. The connector pin 3. The insulation 4. The electrode(s) 5. The Fixation mechanism Each of these components has critical design considerations We already discussed the conductors and insulation. Now we will address also connector, electrode and fixation mechanism.
  2. Passive fixation in relation to endocardial leads means that no part of the lead itself is actually embedded in the endocardium. Rather, the lead tip is trapped within the trabeculae and/or is held in position by its pre-formed shape (e.g., J-lead in atrium). Passive fixation leads commonly use tines or fins to &amp;quot;catch&amp;quot; trabeculi in the heart, or the lead is canted or curved to help place and hold the lead tip in a certain position. Passive anchoring devices, such as flanges or wedge tips, cages and balloons were early attempts to solve the dislodgment problems of early pacing leads. However, only after tined leads were introduced by Medtronic in 1976 did a true solution to dislodgment emerge with regard to passive fixation. Today, there is approximately a 1-2% dislodgment rate 1, 2, 3, 4, 5, 6 with passive fixation leads: 1 - 2% in the atria,1, 2, 6 1 - 2% in the ventricle.1, 3, 4, 5 References: 1 Gammage MD, Marshall HJ, Harris JI. Five-year experience with polyurethane, passive fixation, steroid-eluting leads. PACE. 1998;(Pt II):842. Abstract. 2 Hua W, Mond HG, Strathmore N. Chronic steroid-eluting lead performance: a comparison of atrial and ventricular pacing. Pacing Clinical Electrophysiology. 1995;20(1 Pt 1):17-24. 3 Kazama S, Nishiyama K, Machii M, Tanaka K, Amano T, Nomura T, Ohuchi M, Kasahara S, Nie M, Ishihara A. Long-term follow-up of ventricular endocardial pacing leads. Jpn Heart J. 1993;34(2):193-200. 4 Mayer DA, Tsapogas MJ. Pacemakers: dual or single chamber implantation. Vasc Surg. 1992;26(5):400-7. 5 Miller GB, Leman RB, Kratz JM, Gillette PC. Comparison of lead dislodgment and pocket infection rates after pacemaker implantation in the operating room versus the catheterization laboratory. Am Heart J. 1998;115(5):1048-51. 6 Mond HG, Hua W, Wang CC. Atrial pacing leads: the clinical contribution of steroid elution. Pacing Clinical Electrophysiology. 1995;18(9 Pt 1):1601-8. 7 Glikson M, von Feldt LK, Suman VJ, Hayes DL. Short- and long-term results with an active fixation, bipolar, polyurethane-insulated atrial pacing lead. Pacing Clinical Electrophysiology. 1996;19(10):1469-73. 8 Stirbys P. Implantation of double screw-in leads. Pacing Clinical Electrophysiology. 1988;11(10):1482-4.
  3. Active fixation means that part of the lead actually embeds in the heart tissue for fixation via ascrew-in helix electrode. Extendible/retractable and fixed screw mechanisms are the most common active fixation methods. Fixed Screw leads provide excellent stability. The lead body must be turned in a counter-clockwise rotation during insertion of the lead. Sense mapping can be done prior to fixing the screw to the myocardium. The lead body is then turned in a clockwise rotation to fix the screw to the myocardium. Extendible/Retractable screw-in leads provide excellent stability. The screw is retracted to prevent damage to the veins and cardiac structures during lead advancement. The ability to retract the screw makes entanglement in cardiac structures less likely. The screw is retracted during transvenous introduction and during sense mapping and extended for lead fixation. Care must be taken not to over-extend or over-retract the screw with this type of screw mechanism. Today, there is approximately a 1 - 4% dislodgment rate 6, 7, 8 with active fixation leads: 1 - 4% in the atria 6, 7, 8 &amp;lt;1% in the ventricle.8 References: 6 Mond HG, Hua W, Wang CC. Atrial pacing leads: the clinical contribution of steroid elution. Pacing Clinical Electrophysiology. 1995;18(9 Pt 1):1601-8. 7 Glikson M, von Feldt LK, Suman VJ, Hayes DL. Short- and long-term results with an active fixation, bipolar, polyurethane-insulated atrial pacing lead. Pacing Clinical Electrophysiology. 1996;19(10):1469-73. 8 Stirbys P. Implantation of double screw-in leads. Pacing Clinical Electrophysiology. 1988;11(10):1482-4.
  4. Myocardial/Epicardial leads require an open-chest, tunneling, or minimally invasively sub-xyphoid to attach the lead tip to the outside of the heart. They are used in special situations, such as: - pediatric applications, - congenital malformations, - when a mechanical heart valve is present, - where multiple abandoned endocardial leads are present, - or concurrent with open-chest surgery. Stab-in and screw-in leads are considered myocardial leads because they are attached to the outside of the heart and the “hook” electrode and the screw electrode actually pass into the myocardium when fixated. Suture-on leads are considered epicardial leads because the electrode is attached to the outside of the heart and does not penetrate the myocardium.
  5. Speakers Notes What to say: In a basic electrical circuit the voltage drops around the circuit at each resistance. Assuming the heart/electrode interface is the only significant impedance, all voltage should be “spent” there What to Do: What to Ask: Optional Information: Optional Resources: References:
  6. The terms “amplitude” and “voltage” are often used interchangeably, undoubtedly to express “Voltage Amplitude” which refers to the voltage output.
  7. Current: Measured in amperes (I) 1 Ampere =1000 milliamps Movement of electricity or free electrons through a circuit One ampere is a unit of electrical current produced by 1 volt acting through a resistance of 1 ohm
  8. Impedance is the sum of all resistance to the flow of current. The resistive factors to a pacing system include: Lead conductor resistance The resistance to current flow from the electrode to the myocardium Polarization impedance, which is the accumulation of charges of opposite polarity in the myocardium at the electrode-tissue interface. Resistance is a term used to refer to simple electric circuits without capacitors and with constant voltage and current. Impedance is a term used to describe more complex circuits with capacitors and with varying voltage and current. Therefore, the use of the term impedance is more appropriate than resistance when discussing pacing circuits.
  9. Possible Causes Increased threshold QRS complex not visible Faulty cable connection Dislodged/fractured lead Tissue is refractory Battery depletion Corrective Measures Increased output (mA) Change EKG lead or gain Check cable connections Reposition/replace lead Switch polarity Change battery or pulse generator Turn patient
  10. Pacing impedance refers to the opposition to current flow. Three sources contribute to pacing impedance: 1. pacing lead conductor coil 2. electrode-tissue interface 3. tissue Impedance at the electrode-tissue interface derives from two factors: electrode resistance and electrode capacitance. The latter is due to polarization. Increasing the value of any of these three components will decrease lead current and enhance battery longevity. But, the safety and efficiency of increasing each component must also be considered. The effect of increasing impedance of each component will be examined.
  11. The table shows that increasing the conductor coil resistance reduces the amount of the output voltage that is available at the tip electrode for myocardial stimulation. When the conductor coil resistance is low, more of the pacemaker’s output voltage (and current) is made available at the cathode for myocardial stimulation. Not only does the voltage at the tip electrode change as a result of increased coil impedance, but the safety factor also changes.
  12. Animated Slides. To advance the animation, choose one: Press the “N” key on your keyboard Push the “Page Down” key Press the down arrow or Left click your mouse. A pacing pulse may be thought of as negative ions moving down the lead. These negative ions attract positive ions to the lead tip. Initially, the electrical current (negative ions) move freely into the myocardium. As the pacing pulse continues for the programmed pulse duration (pulse width), positively charged ions surround the electrode tip, forming a layer called polarization. This layer may impede the movement of negative ions into the myocardium, and affect the ability of the pacemaker to sense the evoked response.
  13. A smaller geometric size can increase impedance but, as we have seen, it can also increase polarization. By altering the surface structure to be more porous you increase the geometric surface area of the electrode and therefore lower the polarization. 3The platinized porus electrodes have a larger functional surface area. The collective spaces within the electrode pores make up the larger functional surface area, which controls polarization for this geometrically smaller electrode. In summary, smaller geometric size, larger functional surface area.
  14. We can compare the pacemaker with this system formed by a reservoir with water (the pacemaker’s battery), a pipe (the lead) and a waterwheel which is rotated by the water pressure (the heart muscle captured by the pacing pulse). Notice on the picture the waterwheel turns but a significant amount of water is wasted, (a lot of current depleted from the battery that is not used).
  15. To reduce the water flow (the current stream from the battery), we can narrow the pipeline (increasing the resistance of the entire lead), but this will significantly decrease the water pressure at the waterwheel (the voltage applied directly on the heart muscle).
  16. Finally, the best option is to narrow the conductor only at the end of the pipeline (increasing the impedance to the tip electrode only, and maintaining a small impedance on the lead conductor) The result in our analogy is to have good water pressure at the waterwheel but also reduce the water flow from the reservoir. Therefore a high impedance for the tip electrode, and a low impedance for the lead conductor ensures a low the current drain from the pacemaker’s battery.
  17. Steroid benefits begin immediately upon implantation. Fewer and less active inflammatory cells congregate at the electrode-tissue site. As a result of the decreased inflammatory response, there is less fibrotic development around the electrode. The intrinsic conductivity of the myocardial cells that surround the electrode appears to be maintained.
  18. A pacing lead is a foreign body that can produce an inflammatory response. Steroid-eluting electrodes provide a continuous elution of minute doses of dexamethasone sodium phosphate from a silicone rubber binder. Steroid elution limits the inflammatory process at the electrode-tissue interface. Steroid in matrix: This approach provides long-term steroid elution by providing steroid in a silicone rubber plug. 3 Medtronic uses this method. Steroid coated: Another approach to steroid elution is to coat the electrode. This approach provides an acute benefit, but is not yet proven for chronic performance. 3 Medtronic uses this method in conjunction with the steroid plug.
  19. This graph compares the stimulation thresholds of contemporary pacing leads. Older electrodes exhibited higher threshold peaking than that of steroid leads shown on the slide. The different types of electrodes exhibit a wide range of threshold peaking. Steroid-eluting electrodes continue to show lower chronic stimulation thresholds and no significant peaking. Threshold changes are shown here over a 12-week period post-implant, where a comparison is made between: • smooth metal electrode • textured metal electrode • steroid-eluting electrode Traditionally, implant stimulation thresholds are relatively low. Non-steroid-eluting electrodes exhibit a peaking phase from week 1 to approximately week 6, due to the maturation process at the electrode-tissue interface. Steroid-eluting electrodes exhibit virtually no peaking. The chronic phase of stimulation threshold occurs 8-12 weeks post-implant which is characterized by a plateau. This plateau is higher than the acute phase, due to fibrotic encapsulation of the electrode. Steroid-eluting lead chronic thresholds remain close to implant values.
  20. Insulation around the lead wire prevents current loss from the lead wire. Electrical current seeks the path of least resistance. An insulation break that exposes wire to body fluids which have low resistance causes: Lead impedance to fall Current to drain into the body Battery depletion Impedance values below 300  Insulation breaks are often marked by a trend of falling impedance values. An impedance reading that changes suddenly or one that is &amp;gt;30% is considered significant and should be watched closely.
  21. Insulation may remain intact but the wire may break within the insulating sheath. Impedance may exceed 3,000  Current flow may be too low to be effective. If a complete fracture of the wire occurs: No current will flow Impedance number will be “infinite” When suspecting a wire break, look for a trend in an increase in impedance values rather than a single lead impedance value.
  22. Speakers Notes What to say: When the heart functions normally, there is no need for the pacemaker to deliver artificial pacing impulses. A pacemaker must be able to sense and respond to normal and abnormal cardiac rhythms What to Do: What to Ask: Optional Information: Optional Resources: References:
  23. Speakers Notes What to say: As the heart muscle depolarizes, a potential difference (measured in millivolts) is created between the tip electrode (cathode or negative electrode) and the ring in a bipolar system, or the coil in an integrated bipolar system (anode or positive electrode). It is this net signal that goes to the device for processing. What to Do: What to Ask: Optional Information: Optional Resources: References:
  24. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  25. Speakers Notes What to say: Oversensing will exhibit pauses in single chamber systems. In dual chamber systems, atrial oversensing may cause fast ventricular pacing without P waves preceding the paced ventricular events. What to Do: What to Ask: Optional Information: Optional Resources: References:
  26. Speakers Notes What to say: The intracardiac EGM is characterized in clinical practice in terms of its amplitude (measured in millivolts), and slew rate (measured in volts per second). We will expand on these in the next few slides. What to Do: What to Ask: Optional Information: Optional Resources: References:
  27. Speakers Notes What to say: Typically, slew rate measurements at implant should exceed .5 volts per second for P waves; .75 volts per second for R wave measurements. What to Do: What to Ask: Optional Information: Optional Resources: References:
  28. Speakers Notes What to say: Different signal types occur in different frequency bands. Pacemaker sense amplifiers will attempt to filter out signal that are “non-physiologic”. P waves, R waves and PVC occur in the 5 to 60Hz range. What to Do: What to Ask: Optional Information: Optional Resources: References:
  29. Speakers Notes What to say: Pacemakers have programmable sensitivity settings that can be thought of like a fence: with a lower fence more of the signal is seen; with a higher fence less of the signal is seen. What to Do: What to Ask: Optional Information: Optional Resources: References:
  30. Speakers Notes What to say: If the system is sensing myopotentials, then raise the fence or increase the number of the sensitivity setting. The pacemaker will &amp;quot;see less&amp;quot; of the incoming signal. If the pacing system is not “seeing” intrinsic cardiac events, set the fence lower or decrease the number of the sensitivity setting. The pacemaker will then &amp;quot;see” more of the incoming signal. What to Do: What to Ask: Optional Information: Optional Resources: References:
  31. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  32. Speakers Notes What to say: Unipolar sensing produces a large potential difference due to a cathode and anode that are farther apart than a bipolar system. Because both electrodes may contribute to the electrical signal that is sensed, the unipolar electrode configuration may detect electrical signals that occur near the pacemaker pocket as well as those inside the heart. What to Do: What to Ask: Optional Information: Optional Resources: References:
  33. Speakers Notes What to say: Produces a small potential difference: Electrodes are close to one another Intracardiac signal arrives at each electrode at almost the same time Less likely to sense: Myopotentials: depolarization of muscles near the heart or the anode (the wave produced after the action potential wave passes along a nerve). Afterpotentials Far-field intracardiac signals Noise EMI What to Do: What to Ask: Optional Information: Optional Resources: References:
  34. Speakers Notes What to say: When two metals make intermittent contact, a small voltage is produced. This can be inappropriately sensed by the pacemaker. An inner insulation break may allow the two conductors to touch. What to Do: What to Ask: Optional Information: Optional Resources: References:
  35. Speakers Notes What to say: Lead fractures that are intermittent are also referred to as “make and break” fractures, due to the artifacts in the electrogram that are produced as the conductor wires make and break contact. What to Do: What to Ask: Optional Information: Optional Resources: References:
  36. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  37. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  38. What to say: Electrosurgery used within six inches of an implanted pacemaker/lead system has the potential to cause permanent loss of pacemaker output. Earlier designed pacemakers are more susceptible to loss of output as the battery voltage decreases. Precautions: Monitor the patient’s pulse during application of the cautery. Program the pacemaker to VOO/DOO if the patient is pacemaker dependent, or secure a magnet over the device. Place the grounding plate as close to the operative site as possible—usually under buttocks or thighs—and as far from pacemaker as possible (a minimum of 15 cm from pacemaker). Limit electrocautery to 1-second bursts every 10 seconds. Use bipolar electrocautery forceps where practical. Speakers Notes What to Do: What to Ask: Optional Information: Optional Resources: References:
  39. Speakers Notes What to say: Defibrillation concerns for IPGs are similar to those mentioned for use with electrocautery. If possible, position the electrodes so that currents are not passing through the pacing system. Place the defibrillator electrodes at least thirteen centimeters or five inches from the IPG. Use the least amount of energy to satisfactorily revert the patient. Medtronic IPGs are designed to withstand 400 watt-seconds of defibrillation energy. Always check the operation of the IPG following defibrillator discharges. Damage may be to various components of the circuitry. What to Do: What to Ask: Optional Information: Optional Resources: References:
  40. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  41. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  42. Speakers Notes What to say: Diagnostic x-ray exposure poses no risk. Therapeutic radiation may cause severe damage. Patients receiving therapeutic radiation in treatment of malignant thoracic disease are particularly at risk. Appropriate shielding of the pulse generator is essential. If adequate shielding is not possible, repositioning of the IPG may have to be carried out. Patients should be checked after each session. What to Do: What to Ask: Optional Information: Optional Resources: References:
  43. This slide illustrates the essential components of a pacing lead. The following topics will be discussed for each component: · Purpose · Design factors · Performance factors
  44. This is the connector of the lead. It connects the pacing lead to the IPG. An accurate mechanical fit between the connector pin and the pin cavity of the pulse generator is crucial for the safe transmission of current from the pulse generator without current leakage. Pass around samples of IS-1 connector pins and labels.
  45. IS-1 is a formal International Standard for pacing connectors developed jointly by the International Organization for Standardization and the International Electrotechnical Commission in 1992. Both bipolar and unipolar IS-1 lead connectors are 3.2 mm in diameter and have sealing rings and a short connector pin. Sealing rings provide proper insulation of the pacing lead electrical connection. The leads are labeled IS-1 UNI or IS-1 BI. You will also notice that unipolar connectors have a blue connector ring as a visual indication that the lead is unipolar. Prior to IS-1, other lead sizes include 3.2 mm low-profile and 5/6 mm. These connector sizes are only seen today during IPG replacements. Adapters may be required.
  46. Silicone and polyurethane are the most commonly used insulation materials for pacing leads. Silicone is a soft, flexible material and has been used for over 30+ years for pacing leads. It is used for the outer and inner insulations in today’s pacing leads. Currently, silicone is used on half in the world’s pacing leads. Polyurethane is a firmer, stiffer material than silicone and has been used for over 20 years for pacing leads. It is used for the outer and inner insulations in today’s pacing leads. Currently, polyurethane is used in half of the world’s pacing leads. Fluoropolymers (PTFE - poly tetra fluoro ethylene, ETFE - Ethyl tetra fluoro ethylene) is a soft, stiff material and is used primarily as a coating to protect conductor wires from corrosion. In addition, fluoropolymers are used to make the conductors more slippery so as not to cut through the primary silicone or polyurethane insulation . Fluoropolymers are used in less than 5% of today’s pacing leads.
  47. Passive fixation in relation to endocardial leads means that no part of the lead itself is actually embedded in the endocardium. Rather, the lead tip is trapped within the trabeculae and/or is held in position by its pre-formed shape (e.g., J-lead in atrium). Passive fixation leads commonly use tines or fins to &amp;quot;catch&amp;quot; trabeculi in the heart, or the lead is canted or curved to help place and hold the lead tip in a certain position. Passive anchoring devices, such as flanges or wedge tips, cages and balloons were early attempts to solve the dislodgment problems of early pacing leads. However, only after tined leads were introduced by Medtronic in 1976 did a true solution to dislodgment emerge with regard to passive fixation. Today, there is approximately a 1-2% dislodgment rate 1, 2, 3, 4, 5, 6 with passive fixation leads: 1 - 2% in the atria,1, 2, 6 1 - 2% in the ventricle.1, 3, 4, 5 References: 1 Gammage MD, Marshall HJ, Harris JI. Five-year experience with polyurethane, passive fixation, steroid-eluting leads. PACE. 1998;(Pt II):842. Abstract. 2 Hua W, Mond HG, Strathmore N. Chronic steroid-eluting lead performance: a comparison of atrial and ventricular pacing. Pacing Clinical Electrophysiology. 1995;20(1 Pt 1):17-24. 3 Kazama S, Nishiyama K, Machii M, Tanaka K, Amano T, Nomura T, Ohuchi M, Kasahara S, Nie M, Ishihara A. Long-term follow-up of ventricular endocardial pacing leads. Jpn Heart J. 1993;34(2):193-200. 4 Mayer DA, Tsapogas MJ. Pacemakers: dual or single chamber implantation. Vasc Surg. 1992;26(5):400-7. 5 Miller GB, Leman RB, Kratz JM, Gillette PC. Comparison of lead dislodgment and pocket infection rates after pacemaker implantation in the operating room versus the catheterization laboratory. Am Heart J. 1998;115(5):1048-51. 6 Mond HG, Hua W, Wang CC. Atrial pacing leads: the clinical contribution of steroid elution. Pacing Clinical Electrophysiology. 1995;18(9 Pt 1):1601-8. 7 Glikson M, von Feldt LK, Suman VJ, Hayes DL. Short- and long-term results with an active fixation, bipolar, polyurethane-insulated atrial pacing lead. Pacing Clinical Electrophysiology. 1996;19(10):1469-73. 8 Stirbys P. Implantation of double screw-in leads. Pacing Clinical Electrophysiology. 1988;11(10):1482-4.
  48. Active fixation means that part of the lead actually embeds in the heart tissue for fixation via ascrew-in helix electrode. Extendible/retractable and fixed screw mechanisms are the most common active fixation methods. Fixed Screw leads provide excellent stability. The lead body must be turned in a counter-clockwise rotation during insertion of the lead. Sense mapping can be done prior to fixing the screw to the myocardium. The lead body is then turned in a clockwise rotation to fix the screw to the myocardium. Extendible/Retractable screw-in leads provide excellent stability. The screw is retracted to prevent damage to the veins and cardiac structures during lead advancement. The ability to retract the screw makes entanglement in cardiac structures less likely. The screw is retracted during transvenous introduction and during sense mapping and extended for lead fixation. Care must be taken not to over-extend or over-retract the screw with this type of screw mechanism. Today, there is approximately a 1 - 4% dislodgment rate 6, 7, 8 with active fixation leads: 1 - 4% in the atria 6, 7, 8 &amp;lt;1% in the ventricle.8 References: 6 Mond HG, Hua W, Wang CC. Atrial pacing leads: the clinical contribution of steroid elution. Pacing Clinical Electrophysiology. 1995;18(9 Pt 1):1601-8. 7 Glikson M, von Feldt LK, Suman VJ, Hayes DL. Short- and long-term results with an active fixation, bipolar, polyurethane-insulated atrial pacing lead. Pacing Clinical Electrophysiology. 1996;19(10):1469-73. 8 Stirbys P. Implantation of double screw-in leads. Pacing Clinical Electrophysiology. 1988;11(10):1482-4.
  49. Fluorovisibility of pacing leads: • Visual quality can vary widely among different fluoroscopy units and techniques. • Plastics and less dense materials, such as titanium, are not visible in fluoroscopy. • Denser metals, such as platinum, are more easily viewed. 3In this image a SureFix and CapSureFix active fixation lead is show under normal fluoroscopic conditions. You will notice a space is present when the helix is retracted as pointed out in the slide above.
  50. Myocardial/Epicardial leads require an open-chest, tunneling, or minimally invasively sub-xyphoid to attach the lead tip to the outside of the heart. They are used in special situations, such as: - pediatric applications, - congenital malformations, - when a mechanical heart valve is present, - where multiple abandoned endocardial leads are present, - or concurrent with open-chest surgery. Stab-in and screw-in leads are considered myocardial leads because they are attached to the outside of the heart and the “hook” electrode and the screw electrode actually pass into the myocardium when fixated. Suture-on leads are considered epicardial leads because the electrode is attached to the outside of the heart and does not penetrate the myocardium.
  51. This slide illustrates the effects of lead impedance upon longevity. Increasing pacing impedance enhances longevity. 3500 ohms approximates the CapSure lead pacing impedance. 600 ohms approximates the CapSure SP Novus lead pacing impedance. The 1,000 to 1,200 ohms approximates the CapSure Z Novus lead pacing impedance.
  52. Speakers Notes What to say: What to Do: What to Ask: Optional Information: Optional Resources: References:
  53. Due to the excellent electrode-tissue biocompatibility: •there is no significant threshold peaking and no substantial increase in chronic thresholds, virtually eliminating exit block. •the amplitude of sensed intracardiac signals remains relatively constantpost-implant, especially chronic P-wave sensing.