3. What is an ECG?
The electrocardiogram (ECG) is a representation
of the electrical events of the cardiac cycle.
Each event has a distinctive waveform, the
study of which can lead to greater insight into a
patient’s cardiac pathophysiology.
3
6. ECG
ECG is a surface measurement of the
electrical potential generated by
electrical activity in cardiac tissue.
6
7. ECG
The human heart can be considered as a
large muscle whose beating is simply
muscular contraction. Therefore
contractions of the heart cause a potential
to be developed. The measurement of the
potential produced by cardiac muscle is
called electrocardiology.
7
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
8. What types of pathology can we identify and
study from ECGs?
• Arrhythmias
• Myocardial ischemia and infarction
• Pericarditis
• Chamber hypertrophy
• Electrolyte disturbances (i.e. hyperkalemia,
hypokalemia)
• Drug toxicity (i.e. digoxin and drugs which
prolong the QT interval)
8
11. Different Segments of ECG
P wave: the sequential activation (depolarization) of the right and left atria
QRS complex: right and left ventricular depolarization (normally the ventricles
are activated simultaneously)
ST-T wave: ventricular repolarization
U wave: origin for this wave is not clear - but probably represents
"afterdepolarizations" in the ventricles
PR interval: time interval from onset of atrial depolarization (P wave) to onset
of ventricular depolarization (QRS complex)
QRS duration: duration of ventricular muscle depolarization
QT interval: duration of ventricular depolarization and repolarization
RR interval: duration of ventricular cardiac cycle (an indicator of ventricular
rate)
PP interval: duration of atrial cycle (an indicator or atrial rate
11
13. ECG Leads
Leads are electrode setups which measure the
difference in electrical potential between either:
13
1. Two different points on the body (bipolar leads)
2. One point on the body and a virtual reference point
with zero electrical potential, located in the center of
the heart (unipolar leads)
16. Each lead gives a slightly different
representation of electrical activity of heart
16
17. ECG Leads
The standard ECG has 12 leads:
17
3 Standard Limb Leads
3 Augmented Limb Leads
6 Precordial Leads
The axis of a particular lead represents the viewpoint from
which it looks at the heart.
20. 12-Lead ECG measurement
• Most widely used ECG measurement setup in clinical environment
• Signal is measured non-invasively with 9 electrodes + 1 reference electrode (right
leg)
Einthoven leads: I, II & III Goldberger augmented leads: VR, VL & VF Precordial leads: V1-V6
20
23. Block Diagram of a Basic ECG System
23
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
Isolated Power
Source
Lead Selector
24. ECG Pre-Amplifier
ECG pre-amplifier is a bioelectric differential amplifier.
It includes;
• High impedance input of bioelectric amplifier
• Lead selector switch
• 1mV calibration pulse source
• Means of protecting amplifier from high voltage
discharge such as a defibrillator discharge used on a
patient
• Pre-Amplifier contains an instrumentation amplifier as
well as an isolation amplifier for patient safety
24
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
25. Isolation Amplifier
• Needed for safety! Isolates the patient from high voltages
and currents to prevent electric shock by introducing a
specific barrier between passage of current from the power
line to the patient.
• Can be done by using light (photo emitter and photo
detector) or a transformer (set of inductors that are used in
a step up / step down configuration)
25
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
28. Common Mode Voltage (CMV)
• If 2 inputs are hooked together into a differential
amplifier driven by a common source with respect to
ground, then the common mode voltage should be
the same and the ideal output should be zero.
However practically you will see a voltage.
• CMV is composed of 2 parts:
– DC electrode offset potential
– 50Hz AC induced interference caused by magnetic and
electric fields from power lines and transformers
• Capacitively coupled into circuit
28
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
29. Analysis to reduce noise in ECG
• Right leg driver circuit is used in a feedback
configuration to reduce 50 Hz noise and to drive the
noise on the patient to a lower level.
29
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
30. “6” ways to reduce Noise in ECG
• Common Mode Rejection (differential Amplifier)
• Right Leg Drive (feedback loop to decrease noise)
• Shielding of wires
• Isolation amplifier
• Notch filter to reduce 50 Hz noise
• Bandpass filter to reduce noise below and above ECG
bandwidth (0.05-100 Hz)
30
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
31. High Pass Active Filters
Attenuates frequencies lower than the cutoff frequency.
cutoff frequency is 1/(2Πτ) =1/ 2ΠRiCi
-
+ Voutput
Vinput
Ri
A
Rf
Ci
Rf
Ri Voutput
0
IRf
Ii
Vinput
Ci
31
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
32. Low Pass Active Filters = Integrator
Attenuates frequencies higher
than the cutoff frequency
Voutput
Voutput
IRf
Ii
Cf
-
+
Vinpu
t
Ri
A
Rf
Rf
Ri
0
Vinpu
t
C
f
ICf
32
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
33. Defibrillator Protection Circuit
• Defibrillator = a high voltage electrical heart
stimulator used to resuscitate heart attack victims
• When the physician uses defibrillator, the high
voltages and currents discharged onto patient can
cause damage to medical equipment(specificly pre-
amplifier circuitry), BUT physician still needs to view
ECG of the patient while defibrillating.
• How do you protect your medical equipment from
excessive voltages and currents?
33
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
34. Protection Devices in ECGs: Neon Glow Lamps
• Neon Glow Lamps are pair of electrodes mounted in a glass envelope
filled with low pressured neon gas or a mix of other inert gases.
• Normally impedance across the electrodes is very high (so lamps are not
conductive) but when the potential across the electrodes reaches to the
ionization point of the gas, impedance suddenly drops making the
lamps conductive (they light up) creating a short circuit to ground,
transferring excess energy safely to the ground without damaging the
amplifier.
34
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
35. Protection Devices in ECGs: Zener Diodes
• Zener Diode: is a diode which allows current to flow in the forward
direction in the same manner as an ideal diode, but also permits it to
flow in the reverse direction when the voltage is above a certain
value known as the breakdown voltage.
• In this configuration when a larger potential than the breakdown
voltage is applied to the system (i.e. Defibrillator discharge) the
zener diode allow the current to flow in reverse direction and shunts
it to the ground thus preventing any damage to the amplifier.
35
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
36. Protection Devices in ECGs: Current-Limiting Diodes
• Current Limiting Diodes are electronic devices that limit current to a
maximum specified value for the device.
• These diodes consists of a n-channel JFET-transistor
• They act as a resistor as long as the current level remains below the
limiting value. If current tends to rise above the limit (as in the case of a
Defibrillator discharge), it will be clamped and limited preventing any
excess current reaching the amplifier thus protecting it.
36
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
37. Electro-Surgery Unit (ESU) Filtering
• During surgery simultaneous ECG recording is also
required.
• ESU can introduce high frequency signals with
frequencies ranging from 100KHz to 100 MHz and
with magnitudes up to few kVolts into the ECG
signal. This interferance can heavily distort the ECG
signal.
• ESU introduces:
– DC offsets
– and obscures the ECG signal
37
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
38. • Even though the ECG frequency range is between 0.05-100 Hz
and all the higher frequency signals are attenuated by using
low-pass filters at filtering stage (slide 32), ESU signal still
manages to disort ECG.
• Because;
⮚ Internal junction points of the ECG amplifier can rectify high
frequency signals like ESU signal and the parasitic capacitance
between the leg connections of the amplifier further filters
this signal to create a dc ofset potential. (Whenever the ESU
device is triggered during the surgery ECG baseline oscilates
up and down.)
BMI LECTURE NOTE 8 – SCE SUPAUL 2022 38
39. ⮚And the ESU signal can leak through amplifier and
low pass filter layers obscuring the ECG signal.
• ECG needs to be of diagnostic quality so the ESU
noise should be eliminated
• Common technique is to use a pi-type 3-layered
RC filter
BMI LECTURE NOTE 8 – SCE SUPAUL 2022 39
40. RC Filters
• Low Pass Filters will pass frequencies lower than cutoff frequency of FH
=1/2πRC
40
Vs
Frequency
• High Pass Filters will pass frequencies greater than cutoff frequency of FL
=1/2πRC
FH
Vs
FL
BMI LECTURE NOTE 8 – SCE SUPAUL 2022
41. 3-Layered RC ESU Filter
BMI LECTURE NOTE 8 – SCE SUPAUL 2022 41
Defibrillator
Protection
Circuit
ECG
Differential
Amplifier
RA
LA
RC Filter
LC filter design can also be used
42. Some Sample ECG Circuits
BMI LECTURE NOTE 8 – SCE SUPAUL 2022 42
High-precision analog front end of a portable ECG application. 3 patient
electrodes (only 3 bipolar leads)
43. • ECG system with 5-patient electrodes (7-leads)
BMI LECTURE NOTE 8 – SCE SUPAUL 2022 43