Components of cardiac pacing and ICD-
Understand the hardware
Dr Raghu Kishore Galla
History: 1700’s
Cardiac electrostimulation began in the
mid-18th century with the use of currents
from the Leyden jar or Voltaic Pile
(Allessandra Volta, 1799) to stimulate
cardiac nerves and muscles in animals and
to attempt resuscitation of intact dead
animals.
History: 1800’s
Dr.DeSanctis used “Re-animation Chair.” as
described by Richmond Reece in his The Medical
Guide (1820). It had 3 pertinent features: a
bellows to give forced ventilation, a metallic tube
to be inserted into the esophagus and a voltaic
pile attached at one pole to the esophageal tube
and at the other to an electrode. The electrode
was to be successively touched to “the regions of
the heart, the diaphragm and the stomach….
History: 1900’s
In 1899, Prevost and Battelli
demonstrated that electrical currents could
cause ventricular fibrillation that often
could be reversed by another powerful
stimulus of either alternating or direct
current.
Robinvitch in a series of reports from
1907 through 1909 confirmed this work and
designed the first portable electrical
resuscitative apparatus for ambulances.
1905 – Einthoven
Published first human
AV block using string
galvanometer
History: 1930’s
In 1932, Albert Hyman developed
a machine for controlled repetitive
electrostimulation of heart and
named his device the “artificial
cardiac pacemaker”
• 1950 - John Hopps - Canadian
electrical engineer paired with Dr.
Wilfred Bigelow to create a
pacemaker that could be used for long
periods and treat many of the new
dysarhythmias that had been
discovered.
• The device that sat on a tabletop was
large and heavy.
• 1951 - Dr. Paul Zoll a cardiologist from
Boston created pacemaker using modern
transistors but still was large, heavy and
relied on AC power.
• Despite these shortfalls the Zoll
pacemaker remained a standard for
pacing patients for several years.
History: 1950’s
• The first clinical implantation into a
human of a fully implantable pacemaker
was in 1958 at the Karolinska Institute in
Solna, Sweden, using a pacemaker
designed by Rune Elmqvist and surgeon
Åke Senning.
• It connected to electrodes attached to the
myocardium of the heart by thoracotomy.
The device failed after three hours. A
second device was then implanted which
lasted for two days.
• The world's first implantable pacemaker
patient, Arne Larsson, went on to receive
5 different lead systems and 22 pulse
generators of 11 different models of
pacemakers during his lifetime.
• He died in 2001 with cancer, at the age of
86, outliving the inventor as well as the
surgeon
• 1981 – Rate responsive pacing by QT interval,
respiration, and movement
• 1994 – Cardiac resynchronization pacing
• 1998 – Automatic capture detection
• 2016 - Leadless pacemaker
• Now
– Approximately 3 million with pacemakers
– Approximately 1 million with ICD device
Introduction
• Modern day pacemakers serve the primary function of either maintaining a
minimum heart rate to avoid symptomatic or potentially life-threatening brady or
tachyarrhythmias or offering resynchronization between the left and right
ventricles in the setting of heart failure.
• Multiple advances in device design, programming, sensor technology, and
materials science have afforded the ability to offer an ever-widening range of
devices with a variety of specialized features.
Pacemaker Functions
1. Stimulate cardiac depolarization
2. Sense intrinsic cardiac function
3. Respond to increased metabolic demand by providing rate
responsive pacing
4. Provide diagnostic information stored by the pacemaker
Basic Concepts of Pacing
• Cardiac myocytes may be “activated” by the delivery of an electrical pacing
stimulus.
• This stimulus creates an electrical field that allows for the generation of a self-
propagating wave front of action potentials that may then advance from the
stimulation site.
• The minimum amplitude and duration required to generate the self-propagating
wave front that results in cardiac activation is referred to as the threshold.
• In order to provide for consistent myocardial stimulation, there needs to be a
constant source of energy for pulse delivery, which is provided by the pulse
generator, a conductor that will deliver the stimulus from the source, which is
the lead itself, an electrode at the end of the conductor that delivers the pulse,
and underlying myocardium that is excitable.
• Failure at any of these points (e.g. due to fracture in the lead or loss of contact
between the electrode and the underlying myocardium) will result in failure of
myocardial stimulation.
• Currently available pacing systems comprise a pulse generator and one or more
pacemaker leads with lead tips positioned in the cardiac chamber of interest.
Basic Concepts of Pacing
Basic Concepts of Pacing
• Sensing of local electrograms (EGMs) is essential to the proper function of
permanent pacemakers.
• Whether sensing takes place in a unipolar or bipolar system and despite the
differences between the two, the principles of sensing remain the same.
• Two electrodes, a cathode and an anode, are required to complete the electrical
circuit between the body and the pacemaker.
• In a bipolar system, both anode and cathode are located in the heart, whereas in a
unipolar system only the cathode is located in the heart and the pacemaker
generator can serve as the anode.
Basic Concepts of Pacing
• As a wavefront of depolarization travels toward an endocardial electrode in
contact with resting myocardium, the electrode becomes positively charged
relative to the depolarized region. This is manifest in the intracardiac EGM as a
positive deflection.
• As the wave front of depolarization passes under the recording electrode, the
outside of the cell suddenly becomes negatively charged relative to resting
myocardium, and a brisk negative deflection is seen in the intracardiac EGM.
• The positive and negative deflections that precede and follow the intrinsic
deflection represent activation in neighboring regions of myocardium relative to
the recording electrode.
Basic Concepts of Pacing
A typical bipolar ventricular electrogram (EGM)
in a normal individual. The sharp downward
deflection in the EGM represents the intrinsic
deflection and indicates the moment of
activation under the recording electrode. The
slope of the intrinsic deflection (dV/dt) is
expressed in volts per second and is referred to
as the slew rate. For an EGM to be sensed by a
sensing amplifier, the amplitude and slew rate
must exceed the sensing thresholds.
Bipolar Versus Unipolar Stimulation
• Modern lead design has largely eliminated any significant clinical difference
between bipolar and unipolar stimulation, but some inherent differences remain.
• Historically, the unipolar threshold was lower than the bipolar threshold in the
same system due to a markedly larger anode and resulting reduced impedance in
the unipolar system.
• With the contemporary lead tip design, this difference is attenuated as impedance
is mainly determined by the lead tip properties.
Bipolar Versus Unipolar Stimulation
• The clinically important implication of unipolar pacing is that the pacemaker
generator may cause pectoral muscle stimulation (muscle twitching) if implanted
adjacent to the pectoral muscle.
• Because the distance between the cathode and anode is so much greater during
unipolar pacing, the pacing stimulus on the surface ECG is also much larger than
during bipolar pacing.
• The increased size of the “pacer spike” may be visually helpful to determine proper
pacemaker function when interpreting cardiac telemetry recordings or trans-
telephonic ECG tracings during remote pacemaker monitoring.
Bipolar Versus Unipolar Stimulation
• The larger amplitude signal from unipolar pacing is far more likely to interfere with
appropriate sensing of the cardiac rhythm by an ICD in the rare instance that a
patient has a separate pacemaker and ICD implanted.
• Specifically, a sensed large-amplitude unipolar pacing stimulus will result in
subsequent under-detection of low-amplitude intracardiac signals, such as those
that occur during ventricular fibrillation.
• It is for this reason that the use of bipolar pacing is critical when a separate
pacemaker is present in an ICD patient.
• Bipolar pacing leads are also necessary for some rate-adaptive sensors, such as the
minute-ventilation sensor, as well as for some types of automatic threshold
algorithms, where the ring electrode is used to sense myocardial capture.
Bipolar Versus Unipolar Stimulation
• Due to the much smaller field of view, bipolar electrodes tend to be minimally
influenced by electrical signals that originate outside the heart, while unipolar
leads may detect electrical signals that originate near the pulse generator pocket.
• Thus, unipolar sensing is more susceptible to interference from electrical signals
originating in skeletal muscle (myopotentials), which can result in inappropriate
inhibition of pacing (if sensed in the ventricular channel) or triggering of pacing
output (if sensed in the atrial channel).
• Thus, bipolar sensing is less prone to oversensing myopotentials, far-field cardiac
signals, or electromagnetic interference from environmental sources.
• Shielding of the device and signal processing in the pacemaker circuitry, such as
rectification and filtering of lower or higher frequency components, also help to
attenuate recording of unwanted sign
Single-Chamber System:
• The pacing lead is implanted in the atrium or
ventricle, depending on the chamber to be
paced and sensed
Advantages:
• Implantation of a single lead.
Disadvantages:
• Single ventricular lead does not provide AV
synchrony.
• Single atrial lead does not provide ventricular
backup if A-to-V conduction is lost
Dual-Chamber System:
• Have Two Leads
• One lead implanted in the atrium
• One lead implanted in the ventricle
Benefits of Dual Chamber Pacing:
• Provides AV synchrony
• Lower incidence of atrial fibrillation
• Lower risk of systemic embolism and stroke
• Lower incidence of new congestive heart failure
• Lower mortality and higher survival rates
Pacemaker Components
A basic pacing system is made up of:
• Implantable pulse generator that contains:
– A power source—the battery within the pulse
generator that generates the impulse
– Circuitry—controls pacemaker operations
–
• Leads—Insulated wires that deliver electrical
impulses from the pulse generator to the heart.
Leads also transmit electrical signals from the
heart to the pulse generator.
• Electrode—a conductor located at the end of the
lead; delivers the impulse to the heart
Schematic of the typical circuits found in modern pacemakers
The Pulse Generator
• Contains a battery that provides
the energy for sending electrical
impulses to the heart
• Houses the circuitry that controls
pacemaker operations
Battery
• A battery converts chemical energy into electrical energy.
• The source of this energy is the electrochemical reactions that occur within the
battery.
• During a spontaneous chemical reaction, substances react to form more stable
products.
• Redox reactions: types of reactions in which oxidation and reduction occur. They
are called redox reactions because electrons are transferred from one reactant to
another.
2Li + I2  2LiI
Battery
• The major parts of a battery are the
anode, the cathode, and the electrolyte.
• The anode and cathode must be
physically separated and both must be in
contact with the electrolyte
Battery
• In a spontaneous galvanic cell such as a battery, the anode is electrically negative
and the cathode positive.
• In nonspontaneous electrochemical reactions, such as those that occur at pacing
electrodes, electrochemical reactions are driven by an externally applied voltage.
• In this case oxidation(loss of electrons) will occur at the positive electrode
and reduction (gain of electrons) at the negative electrode.
• Thus for pacing electrodes, the anode is positive and the cathode is negative.
• The polarities are reversed, but the underlying electrochemical processes defining
anode and cathode, namely, oxidation and reduction, respectively, are the same
Electrolyte
• The anode and the cathode must be separated from each other so they cannot
directly react with one another, but they also need to be connected by an ionically
conducting medium called the electrolyte.
• As the battery discharges, it furnishes electrons at one terminal, pushes them
through an external circuit, and receives them back at a second terminal.
• The electrolyte allows electrical charge, as ions, to flow within the battery,
completing the circuit.
• The electrolyte must conduct ions but not electrons.
Separator
• The separator is a structural member of the battery that keeps the anode and
cathode materials physically apart, thus preventing shorting of the battery.
• In batteries with liquid electrolytes, the separator is usually a porous polymer
film that is immersed in and permeated by the electrolyte.
• In the case of lithium/iodine batteries, traditionally used for pacemakers, the
separator and electrolyte are one and the same, namely, the growing layer of
solid, ionically conductive lithium iodide discharge product.
Current Collector
• The current collector makes the connection between the positive or negative
terminal of the battery and its respective active electrode material inside the
cell.
• It is usually a wire connected to a screen, or grid, which is embedded in the
anode or cathode material.
• The current collector may also serve as a structural member of the battery to
provide physical integrity and strength to that electrode.
Sealing of Batteries
• Batteries are well sealed, most often in hermetically welded containers with glass
electrical feedthroughs to make the electrical connections between the inside and
the outside of the battery.
• Sealing is necessary to prevent any interchanges of materials between the battery
and its surroundings.
• Medical batteries are typically considered hermetically sealed if the leak rate out
of the battery for a test gas, usually helium, is less than 1 × 10−7 cm3/sec at one
atmosphere pressure difference between the inside and the outside of the battery.
Types of Batteries
Primary Batteries
• can only be used once.
• not designed to be recharged
• Most batteries used to power
modern implantable medical devices
are primary batteries that use
lithium anodes because such
batteries
• have very high energy densities.
Secondary Batteries
• Designed to be repetitively discharged
and recharged.
Eg: Lithium-ion batteries – use in LVADs, and
implantable neurologic stimulators for
control of chronic pain.
• Although they are not currently in
development for pacemakers and ICDs,
features such as more frequent telemetry
could cause manufacturers to reconsider
lithium-ion battery use in cardiac
applications in the future.
Implantable Battery Design Requirements
• The most important requirement in battery selection for implantable devices is high
reliability.
• Other significant factors include the desired longevity of the device (directly related
to battery energy density, circuit design, and overall device size) and an appropriate
indication of impending battery depletion (end-of-service warning).
• The basic considerations when designing a battery for an implantable medical device
include the current variations that can be expected from the circuit as the device
provides its service for individual patients.
• Once the range of these application requirements are defined, the current, voltage,
and capacity requirements of the battery can be determined.
Elective Replacement Indicator(ERI)
• All modern implantable pulse generators have an ERI that alerts the clinician to
impending battery depletion and allows adequate time for replacement of the
device.
• Typically, this indicator is designed to occur at least 3 months before the battery
voltage drops to a level that would result in erratic pacing, loss of capture, or loss
of other critical features. Many newer devices conform to CENELEC standards
(European Committee for Electrotechnical Standardization).
• For pacemakers, the standard requires a recommended replacement time (RRT)
such that at least 95% of the devices will have a prolonged service period (PSP) of
at least 180 days.
• For ICD and CRT devices, the requirement is that at least 95% of the devices have a
PSP of at least 90 days
Battery Chemistries Used in Pacemakers
The Lithium/Iodine Battery
• In general, lithium/iodine batteries have a
single, central lithium anode that is
surrounded by cathode material which is at
least 96% iodine and has been thermally
reacted with a polymer material to render
a conductive mixture.
• Central anode with an embedded current
collector wire and the iodine cathode that
fills much of the volume inside the battery.
The Lithium/Carbon Monofluoride Battery:
• CFx is a cathode material with high capacity and moderate power capability.
• The CFxmaterial is a powder that is mixed with carbon as a conductivity enhancer plus a
polymer binder and then pressed into a porous pellet.
• The battery is composed of a lithium anode, a porous polymer separator material, and
the porous pressed-powder cathode pellet that are all inserted into a case which is
welded closed.
• In contrast to the lithium/iodine battery, the lithium/CFxbattery uses a liquid electrolyte
that consists of a lithium salt dissolved in an organic solvent
• Voltage decline near end of battery life tends to be fairly abrupt, making it challenging
for device manufacturers to engineer means of maintaining adequate replacement time
warning.
The Lithium/Hybrid Cathode Battery:
• A new lithium battery chemistry with a cathode comprised of silver vanadium
oxide (SVO), plus CFx, has been developed to meet the needs of implantable
devices with higher rate therapies and features.
• SVO has high power capability and is the same cathode material used to power
ICDs.
• The blend of the two cathode materials, called a hybrid cathode, yields a primary
battery that has an energy density equal to that of a lithium/iodine battery with
approximately 100 times more power capability.
The Lithium/Manganese Dioxide Battery
• Lithium/manganese dioxide batteries are being used in a new generation of
pacemakers commercialized by one manufacturer.
• Manganese dioxide possesses good energy and power density characteristics and
shows low self-discharge.
• As lithium/manganese dioxide batteries demonstrate greater power capability in
comparison to lithium/iodine batteries, devices that require higher power, such as
wireless telemetry capable of communicating over longer distances, can be
successfully powered
Miniature Batteries for Leadless Devices
• Leadless pacemakers are powered by a cylindrical, pin-style, lithium metal based
primary battery that typically contains a monolithic, thick cathode with adequate
electrode area.
• Provide sufficient power for background electronics, stimulation, and low-power
telemetry, while maintaining high energy density to achieve longevity on par with
traditional pacemakers.
• Another distinguishing feature of the miniature batteries used in leadless pacemakers is
that the battery may not be fully enclosed in a titanium (device) case as in traditional
pacemakers;
• Instead, the battery may be welded to the end of the container housing the electronics
and the fixation mechanism, such that the battery enclosure may be in direct contact
with the body fluids.
Microprocessors
• Microprocessors have become the standard control circuits of implantable
pacemakers and ICDs.
• Microprocessors have several advantages over older integrated circuits, including
a far greater circuit density and greatly reduced current drain.
• Microprocessors also allow very sophisticated algorithms, requiring multiple
calculations, to be incorporated into implantable devices, and have vastly
increased data storage.
• The microprocessor can respond to changes in programming instructions that
allow functions to be added or changed after implantation.
• The integrated circuit of pulse generators may contain both read-only memory
(ROM) and random access memory (RAM).
Microprocessors
• ROM (typically 256 KB to 1 MB) is used to guide the sensing and output circuits.
• In addition, RAM is used to store diagnostic information regarding pacing rate, intrinsic
heart rates, and sensor output.
• The amount of RAM (1–16 MB) included in the pulse generator varies between models
and manufacturers, but has rapidly increased in modern pulse generators, allowing for a
far greater amount of diagnostic information to be stored.
• The rapidly expanding diagnostic capabilities of pacemakers has allowed for improved
assessment of the physiological condition of the patient, including stored information
about heart rate variability, respiration, intracardiac pressure, patient activity, lung
water, and arrhythmia logs.
Output circuit
• The output circuit contains the output section and voltage multipliers.
• Pacing outputs of greater than the specified voltage of a cell are achieved by a
variety of methods.
• One approach is to use capacitors located on the output circuits. These capacitors
are charged by the battery in parallel, but then discharged in series.
• Multiple capacitors can be used to achieve the desired output voltages.
• An alternative approach to capacitor-based voltage multipliers is to use the
electromagnetic principle of inductance.
Sensing circuit
• The intracardiac EGM is conducted from the electrodes to the sensing circuit of the PG,
where it is amplified and filtered.
• The intracardiac EGM is filtered to remove unwanted frequencies, a process that
markedly affects the amplitude of the processed signal.
• Following filtering of the intracardiac signal, the processed signal is compared with a
reference voltage to determine if the signal exceeds a threshold detection level
(programmed sensitivity).
• Signals with amplitudes greater than the sensitivity threshold level are sensed as
intracardiac events, whereas signals of lower amplitude are discarded as noise.
• Signals that exceed the threshold level are sent to the timing circuit and logic circuits.
Sensing circuit
• Most permanent pacemakers also contain noise reversion circuits that change the
pulse generator to an asynchronous pacing mode when the sensing threshold level
is exceeded at a rate faster than the noise reversion rate.
• The noise reversion mode prevents inhibition of pacing in the presence of electro
magnetic interference.
• The sensing amplifier must prevent the detection of unwanted intracardiac signals,
such as far-field R waves in the atrial EGM, afterpotentials, T waves, and
retrogradely conducted P waves.
Timing circuit
• The pacing cycle length, sensing refractory and alert periods, pulse duration and AV
interval are precisely regulated by the timing circuit of the pulse generator.
• The timing circuit of a pulse generator is a crystal oscillator that generates a very
accurate signal with a frequency in the kilo Hertz range.
• The output of the crystal oscillator is sent to a digital timing and logic control
circuit that operates internally generated clocks at divisions of the oscillator
frequency.
• The output of the logic control circuit is a logic pulse that triggers the output
pacing pulse, the blanking and refractory intervals, and the AV delay.
Timing circuit
• The timing circuit also receives input from the sense amplifier to reset the escape
intervals of an inhibited pacing system or trigger initiation of an AV delay for
triggered pacing modes.
• The pulse generator also contains a rate limiting circuit that prevents the pacing
rate from exceeding an upper limit in the case of a random component failure.
• This runaway protection rate is typically in the range of 180–200 ppm.
Magnet mode
• Pacemakers and some ICDs have a reed switch which normally is open until a
magnet or magnetic field [e.g. a magnetic resonance imaging (MRI) scanner]
comes into close contact with the device.
• The magnet will close the reed switch and the device will be switched to the
“magnet mode.”
• The magnet mode typically causes asynchronous pacing for a pacemaker (e.g.
VOO, DOO, AOO) and deactivation of tachycardia therapy for most defibrillators,
but bradycardia pacing and sensing functions are typically not affected.
• The magnet may be unable to close the reed switch if it is not powerful enough
or the magnet function is programmed “off.”
• In contemporary devices (some pacemakers and most ICDs), other technology,
such as the Hall effect sensor, integrated solid-state detection, or GMR sensor, is
used in place of the reed switch to optimize response to the magnetic
Rate Responsive Pacing Sensors:
• When the need for oxygenated blood increases, the pacemaker ensures that the heart
rate increases to provide additional cardiac output
• There are several different types of sensors that may be employed in pacemakers. Most
commonly, sensors work to augment heart rate in response to physical activity, whether
by detecting patient movement or changes in respiratory rate.
• Those most accepted in the market place are:
– Activity sensors that detect physical movement and increase the rate according to
the level of activity
– Minute ventilation sensors that measure the change in respiration rate and tidal
volume via transthoracic impedance readings
Types of sensors
• Using technology that detects changes in vibration, acceleration, or minute
ventilation, pacemakers may be pre-programmed to induce concomitant changes
in the paced rate if the physiological heart rate does not meet those parameters.
• Other sensors may use sensitivity to temperature, QT interval, or local myocardial
contractility to respond to conditions under which increases in heart rate are
desired, but for which sensors that detect only changes in physical activity would
be inadequate.
Leads
Leads Are Insulated Wires That:
• Deliver electrical impulses from the
pulse generator to the heart.
• Sense cardiac depolarization.
During Pacing, the Impulse:
• Begins in the pulse generator
• Flows through the lead and the
cathode (–)
• Stimulates the heart
• Returns to the anode (+)
Components of pacemakers leads
1) Electrodes
2) Conductors
3) Insulation
4) Connector pin
5) Fixation mechanism
• A Unipolar Pacing System Contains a
Lead with Only One Electrode Within the
Heart.
• In This System, the Impulse:
– Flows through the tip electrode
(cathode)
– Stimulates the heart
– Returns through body fluid and tissue
to the IPG (anode)
Unipolar leads
• Simplest lead design of all the leads have only one conductor surrounded by
insulation.
• Tip of the lead is the cathode and the PG is the anode.
• Largely replaced by bipolar lead designs for most of the endocardial
applications.
• Still commonly used in CS branches for CRT
• Demonstated impressive longevity and many are still in service today
• A Bipolar Pacing System Contains a
Lead with Two Electrodes Within
the Heart.
• In This System, the Impulse:
– Flows through the tip electrode
located at the end of the lead
wire
– Stimulates the heart
– Returns to the ring electrode
above the lead tip
Bipolar lead
• PG is not the part of pace/sense circuit.
• Both the ring electrode(anode) and the tip electrode(cathode) are in
contact with the myocardium
• Can be used for endocardial, CS and epicardial applications.
• Two main bipolar lead designs – Co-axial
Co-radial
Types of lead design
A) Coaxial lead body design consists of two nested coil
conductors to the tip and ring electrodes, each wrapped in a
separate layer of insulation. The inner coil connects to the
tip cathode and contains a central lumen for stylet passage;
the outer coil connects to the ring anode.can be attached to
endocardium via active/passive fixing mechanisms but
limited by large diameters
B) In coradial design, a single coil contains the parallel
conductor strands to the tip and ring electrodes, with each
strand covered by a fluoropolymer(ETFE) insulator. A
surrounding layer of outer insulation by polyurethane
covers the lead body.Design can allow significantly smaller
(<6F) diameters but limits the fixation mechanism.
ICD leads
• Defibrillator leads have greater complexity, with two, three, or four conductors
that connect to the pacing/sensing electrodes as well as to the high-voltage
shocking coil(s), depending on the specific type of lead.
• Because of the prohibitive bulk that would result from a coaxial design with more
than two conductors, ICD leads generally have a different type of structure, known
as a multilumen design.
• This type of ICD lead consists of a long cylinder of insulating material, with
separate internal channels running down its length. Each conductor runs down an
individual channel, usually with its own additional covering tube of insulation.
ICD leads
Multilumen ICD lead design
consists of a single body of
insulation with several internal
channels that contain the
conductor elements (central
pace/sense coil and outer high-
voltage cables), each covered by
its own layer of fluoropolymer
insulation. A surrounding layer
of outer polyurethane-based
insulation is present. HP, High-
performance.
ICD lead - Structural Overview
• Right ventricular (RV) defibrillation leads comprise a distal tip electrode with a
fixation mechanism that anchors the lead to the heart, proximal terminals that
connect to the generator, and a lead body that connects the two.
• The lead body consists of a flexible insulating cylinder with 3 to 6 parallel
longitudinal lumens through which conductors run from the proximal terminals to
small pace-sense electrodes and larger shock coil electrodes.
• This multilumen design permits more conductors in smaller-diameter leads than
older coaxial designs.
• The subcutaneous ICD uses a parasternal electrode in which the larger shock coil
is straddled by 2 small sensing electrodes.
• ICD leads differ in number of shock coils, number of sensing electrodes, and type
of connector terminals.
Dual-Coil Versus Single-Coil Leads
• All transvenous leads have a distal shock coil in the RV. Dual coil leads have an
additional proximal shock coil, which usually lies in the superior vena cava.
• Dual-coil leads were required for the first transvenous ICD systems, which
preceded those in which the housing (can) of the generator served as a
defibrillation electrode (active can).
• With contemporary biphasic waveforms and active cans, differences in
defibrillation efficacy between dual-coil and single-coil shock pathways are
rarely clinically significant for left-pectoral implants.
Dual-Coil Versus Single-Coil Leads
• The principal disadvantage of dual-coil leads is greater risk of extraction because
of fibrotic tissue ingrowth into the proximal coil.
• An additional disadvantage is the possibility of a short between high-voltage
elements of opposite polarity.
• However, dual-coil leads retain advantages, including more reliable atrial
cardioversion, coil-coil electrograms for morphology algorithms that discriminate
VT from SVT, and proximal coil-can EGMs that display atrial signals to improve
interpretation of single-chamber EGMs.
Integrated Versus True Bipolar Sensing
• Sensing can either be true bipolar between the tip electrode and a small ring
electrode or integrated bipolar between the tip electrode and RV coil.
• Modern integrated bipolar and true bipolar leads have similar performance for
sensing ventricular fibrillation (VF) but differ in susceptibility to oversensing.
• The larger integrated bipolar antenna is more susceptible to external
electromagnetic interference, R-wave double counting, and diaphragmatic
myopotentials.
• True bipolar leads are more susceptible to T-wave oversensing, probably because
of greater variations in R-wave amplitude.
• There is no consensus regarding which is preferred.
Multipolar leads
• The complex challenges of CRT system have resulted in development of multipolar
leads.
• A central coil conductor extends to the tip, allowing the stylet or guidewire
insertion.
• Conductors for the more proximal electrodes are cables arranged in parallel that
terminate at their respective electodes.
• Unlike ICD leads the individual elements may be smaller in diameter as they need
not conduct high voltages needed for defibrillation.
Multipolar leads
• Overall diameter of the leads can be
small allowing stable positioning in the
branches of CS.
• Since multiple pacing vectors are
available, it may be possible to
overcome phrenic nerve stimulation,
pace from electrically more
advantageous sites or pace from sites
with better acing thresholds.
CRT leads
• Leads used in CRT deserve special consideration with respect to fixation and shape.
• CS branch anatomy is highly variable. Manufacturers offer a range of CS leads and it is up to
the implanting physician to decide which lead is appropriate for a given patient.
• The implanter has to balance the frequently competing objectives of lead stability, electrical
performance, avoidance of phrenic nerve capture, and resynchronization performance.
• Familiarity with a range of sizes, shapes, and fixation mechanisms of available leads is
necessary for successful implantation.
• Lead diameters vary from 4 to 6 Fr, some leads do not have isodiametric lead bodies. Most
leads use friction and tension to maintain their position within the CS vasculature.
CRT leads
• Most CS leads that are transvenously placed do not have a fixation helix or screw
that embeds the lead to the myocardium.
• Most leads use friction and tension to maintain their position within the CS
vasculature.
• Most leads are bipolar, but some of the smaller diameter leads are unipolar.
• Some leads (Boston Scientific Easy Trak2®) are straight with tines that allow them
to be wedged into a distal target vessel and help provide additional stability
through frictional force.
• Other leads have multiple cants or curves that apply pressure to the vessel wall,
again providing stability by frictional force.
CRT leads
(A)Bipolar lead with tines; (B) helical biased bipolar pacing lead; (C) bipolar pacing lead with can fixation; (D) helical biased,
unipolar pacing lead; (E) bipolar pacing lead with curve fixation; (F) S-biased bipolar pacing lead. (Sources: (A, B, D, E)
Boston Scientific Corporation. (C) Medtronic, Inc.. (F) St.Jude Medical,
CRT leads
• Alternatives to canted leads are leads with sigmoidal biasing in two dimensions (S
shaped) or helical biasing in three dimensions (spiral).
• These leads also rely on the friction created both from wedging the distal tip in a
branch and from the multiple contact points on the vessel walls.
• Despite the variety of available passive fixation CS leads, current passive leads
often do not remain stable in the proximalportion of the CS tributaries.
• If a distal branch cannot be engaged or is prohibited due to phrenic nerve or
extracardiac stimulation, a proximal position can be targeted with a lead such as
the Medtronic Attain 4195 StarFix® lead.
CRT leads
The Medtronic StarFixR lead with its fixation lobes deployed
(top) and retracted (bottom).
This has a unipolar design
with three lobes that are
deployed by advancing the
push tubing around the
lead.
CRT leads
• Modern CRT devices allow for separate pacing circuits for
the RV and LV leads, and allow LV leads with unipolar,
bipolar, and multipolar configurations.
• The ability to separately assess the RV and LV leads allows
the clinician to program parameters separately, and
simplifies lead assessment while simultaneously reducing
current drain on the battery.
• separate programming of the LV and RV channels allows
pacing with multiple LV configurations and with optional
different timing between the two channels.
• Multiple LV configurations allow non-invasive programming
options to avoid phrenic nerve capture or minimize current
drain during LV pacing.
Quadripolar leads
Fixation mechanisms
• The chronic performance of permanent pacing leads is critically dependent on
stable positioning of the electrode(s).
• Proper fixation to the endocardial surface is essential to lead performance.
• two types of fixation mechanisms:
Passive
Active
Passive fixation leads
• Passive fixation leads have tines (fins) near their tip, which are made of the same
material as the insulation.
• The number (two, three or four), length, and stiffness of the tines are variable.
• Passive fixation leads become entrapped within the trabeculae of the RA
appendage or RV apex immediately upon correct positioning of the lead.
• Effective fixation of the lead can be confirmed at the time of implantation by its
gentle traction or rotation.
• Tines generally add minimal technical difficulty to the implantation procedure,
although they may occasionally become entrapped in the tricuspid valve
apparatus.
Passive fixation leads
• These leads are not suitable for placement in non-traditional locations such as the
high RV septum or RV outflow tract.
• The passive fixation leads are rapidly covered by fibrous tissue, making removal of
the lead by simple traction difficult or impossible in as short a time as 6 months.
• In general, passive fixation leads are more difficult to extract than active fixation
leads.
Active fixation leads
• In the present generation of active fixation, endocardial leads have a fixed helix or
extendable–retractable helix that in most cases also serves as a pacing electrode.
• In leads with a fixed helix design, the helix is coated with mannitol or polyethylene
to facilitate introduction of the lead through the vasculature into the desired
chamber. The coating dissolves within a few minutes, allowing the lead to be
positioned with a stylet.
• Fixation is performed by rotation of the entire lead body and transmission of the
torque to the distal tip. Repositioning is accomplished by reverse
(counterclockwise) rotation of the lead.
Active fixation leads
• The extendable–retractable design has become the most widely used active
fixation mechanism because of its ease of implantation and because the fixation
mechanism allows retraction long after implantation, thereby facilitating
repositioning or extraction.
• Active fixation leads allow for stable positioning of the lead at many sites in either
the atrium or the ventricle.
• Although active fixation leads are easier to extract than passive fixation leads, the
risk of myocardial perforation during and after implantation is higher.
• Historically, the chronic pacing thresholds of active fixation leads were higher than
those of passive fixation leads. This difference has largely been eliminated with the
routine use of corticosteroids in active fixation leads.
Types of fixation
Epimyocardial leads
• Permanent epicardial pacing leads are used when tricuspid valve anormalités,
central venous obstruction, congenital heart disease, or technical issues preclude
transvenous lead placement.
• Epicardial leads may also be used when recurrent bloodstream infection is a
concern, or when there is a need for lead implantation co-incident with another
intrathoracic surgical procedure.
• Epicardial pacing leads are either sutured to the epicardial surface or screwed into
the epimyocardium of the atrium or ventricle.
• Fixation mechanisms include a fishhook-shaped electrode that is stabbed into the
atrial myocardium or a large screw helix that is rotated into the ventricular
myocardium.
Epimyocardial leads
• These are typically unipolar and are frequently
non-steroid eluting, with higher stimulation
thresholds than modern endocardial electrodes.
• A newer epicardial lead design has a steroid-
eluting electrode, rests against the epicardium,
and is sutured in place available in unipolar and
bipolar models with lower chronic pacing
thresholds and better long-term performance.
• In the era of CRT, epicardial leads are also being
used when a functional LV lead cannot be placed
transvenously due to limitations imposed by the
coronary venous anatomy
From top to bottom:
fish hook type,
screw-inhelix type; and
sew-on button type.
Conductors
• The primary conductors used in most pacing and ICD leads are MP35N and silver.
• MP35N is a superalloy that is double melted to remove impurities. It is composed
of cobalt, nickel, chromium, and molybdenum.
• It is characterized by biocompatibility, high tensile strength, and resistance to
corrosion.
• Because of its relatively high electrical resistivity (1033 μΩ), MP35N is combined
with efficient conductors such as silver.
• Single conducting filaments (wires) are made in two designs: a drawn brazed
strand (DBS) and a drawn filled tube (DFT).
Conductors
Cross-sections of the two types of conducting filaments (wires) used in pacemaker and ICD leads:
Drawn brazed strand (DBS, left) and drawn filled tube (DFT, right).
Conductors
• The DBS design consists of the MP35N® alloy stranded around a core of softer,
highly conductive silver. The resulting wire combines increased fatigue life, high
conductivity, and solderability with the ability to withstand greater mechanical
stresses.
• The DFT design is coated with platinum or platinum–iridium. Extreme compressive
forces are applied in order to form a sound mechanical bond between surfaces.
• Single wires are combined together into strands and then wound into cables for
use as cable conductors. Larger cables used for high voltage applications and
smaller cables for pacing applications.
• The number of filaments and their
arrangement within a given cable is variable
from manufacturer to manufacturer, even
for similar applications.
• Compared individually, a 7 × 7 (49 wire)
cable has relatively greater flexibility but
relatively lower torquability than a 1 × 19
(19 wire) design.
• Neither design has been shown to be
clinically superior.
Insulation
• The materials used in pacemaker leads for insulation play a critical role in their
longevity and reliability, as well as in handling and implant characteristics.
• The ideal insulation material should be biologically inert, and exhibit no surface
erosions, no molecular chain disruptions, no uptake of low-molecular-weight
biological materials, and no tendency to calcifications, while retaining stable
mechanical properties.
• There is no single ideal insulation and some leads are constructed using multiple
materials.
• Pacemaker leads are generally made of one insulation type. ICD leads are more
complex, utilizing combinations of insulation material.
Insulation
• Fluoropolymers are fluorocarbon-based polymers characterized by high resistance to
solvents, acids, and bases. They have maximum biocompatibility and tensile
strength, but their stiffness limits their use to thin coating (<0.076 mm) applications.
Eg: PTFE (polytetrafluoroethylene, i.e. Teflon®, DuPont)
ETFE (ethylene tetrafluoroethylene).
• Silicone rubber is a polymer both biostable and biocompatible, also has high
resistance to extreme temperatures but has a low tensile strength, making it prone
to tearing and abrasion wear.
• Polyurethanes are characterized by high tear strength, high elasticity, and a low co-
efficient of friction. These qualities allow for smaller lead diameters.
Insulation
Electrodes
• The pacing electrode is the final interface between the lead and myocardium. Its design
and composition greatly influence the overall electrical performance of the pacing
system.
• In a bipolar pacing lead the tip electrode is the cathode and the ring electrode is the
anode.
• The stimulation threshold is a function of the radius of the electrode. A smaller radius is
associated with a higher current density, lower pacing threshold, and higher resistance at
the electrode–myocardium interface.
• In contrast, smaller radius electrodes result in high sensing impedance and electrode
polarization impairing myocardial EGM sensing.
Electrodes
• The materials currently used for electrodes of permanent pacing leads include
platinum–iridium, Elgiloy (cobalt–chromium–nickel), platinum coated with
platinized titanium, vitreous or pyrolytic carbon coating a titanium or graphite
core, platinum, iridium oxide, or titanium–nitride.
• The use of corticosteroid-eluting electrodes has been a major advance in pacing
lead technology. Dexamethasone sodium phosphate and/or dexamethasone
acetate is impregnated on a silicone core or collar that surrounds the tip electrode.
• Corticosteriods have dramatically reduced the risk of exit block for pacing leads,
ICD leads, and epimyocardial leads.
Steroid Eluting Leads
• Steroid eluting leads reduce
the inflammatory process
and thus exhibit little to no
acute stimulation threshold
peaking and low chronic
thresholds
Porous, platinized tip
for steroid elution
Silicone rubber plug
containing steroid
Tines for
stable
fixation
Connector pins
• Connector pins are made of stainless steel or titanium. The current standard is IS-1
(Industry Standard) for pacing leads.
• The lead diameter at the head block is 3.2 mm and sealing rings are integrated to
the lead. With this industry standard, all current bipolar pacing leads are
compatible with all current manufacturer header designs.
• It is important to visualize the connector pin extending beyond the distal set screw
in the header block. If the connector pin is not properly seated in the header
block, sensing artifact or failure to pace may result.
• Lead impedance measurements alone are not enough to detect this problem.
Connector pins
Adapter for leads terminated with 5/6-mm pins that need
to be connected to a pacemaker with an IS-1 header block.
Photograph and schematic of an IS-1 pin terminated lead.
Sealing rings are integrated into the lead
Radiograph demonstrating proper extension of the connector
pin beyond the set-screw (top; arrow indicates fluoroscopic
image of a pacemaker header with properly implanted lead
pin) and inadequate extension beyond the set-screw (bottom;
circle indicates improper
lead pin position).
Connector pins
• Multipolar pacing leads such as those used in the CS will not be IS-1 compatible.
They are connected to specialized header blocks using an IS-4 LLLL type pin similar
to the pin assembly found on a DF-4 ICD lead.
• Currently, there are two ISO standards for the ICD yoke (connector pin assembly):
IS-1/DF-1 yokes and DF-4 (or IS-4) pin assembly.
• Until recently, the IS-1/DF-1 yoke design was the industry standard. The pace/
sense portion of the lead is terminated with an IS-1 pin and the high voltage coils
are terminated with their respective DF-1 (ISO-11318) pins.
Connector pins
• The DF-4 design eliminates the possibility
of lead connection errors (i.e.
interchanging the proximal and distal
coils) due to a uniquely compatible port at
the header.
• This design also has advantage with
respect to pocket bulk as well as for lead-
to-lead and lead to- can interactions.
• A potential disadvantage is the possibility
of multiple component fracture from one
critically located stress point in the pocket.
Comparison of the current DF-1 connector (bottom)
and the proposed DF-4 connector (top).
Lead anchoring sleeve
• The anchoring sleeve allows the lead to be secured to the fascia or muscle, and its
proper use can help reduce the risk of lead damage.
• Currently, most sleeves are made of a biocompatible, peroxide catalyzed (MDX) silicon
rubber and may contain a radiographically dense material to allow visualization under
fluoroscopy.
• These sleeves have one, two, or three circumferential grooves to which non-
absorbable sutures are secured.
• Inadequate tie down force or an inadequate number of sutures is especially a problem
with lead bodies coated with a lubricious material such as polyvinyl pyrrolidone (PVP).
• Balancing the risk of lead damage with lead dislodgement requires the implanter’s
familiarity with the specific lead and sleeve.
MRI compatible lead system
Micra Transcatheter Pacing System
• The miniaturized Micra™ transcatheter pacing system (TPS) is the world’s smallest
pacemaker, delivered percutaneously via a minimally invasive approach, directly
into the right ventricle and does not require the use of leads.
• 99% implant success rate in 726-patient global trial
• 48% fewer major complications than traditional pacemakers
• 93% smaller than conventional pacemakers5
• Ultra low-power circuit design delivers an estimated average 12-year battery
longevity.
• MRI SureScan™ technology allows the patient to be safely scanned using either a
1.5T or 3T full body MRI.
Linear one-step deployment facilitates consistent capsule
placement, no torque required.
THANK YOU

More Related Content

What's hot

Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
drskd6
 

What's hot (20)

Blood pumps
Blood pumpsBlood pumps
Blood pumps
 
Basics of cpb
Basics of cpbBasics of cpb
Basics of cpb
 
Oxygenators
Oxygenators   Oxygenators
Oxygenators
 
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANBMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
 
Lvad
LvadLvad
Lvad
 
Coronary anatomy and angiographic views
Coronary anatomy and angiographic viewsCoronary anatomy and angiographic views
Coronary anatomy and angiographic views
 
Catheter ablation of ventricular tachycardia
Catheter ablation of ventricular tachycardiaCatheter ablation of ventricular tachycardia
Catheter ablation of ventricular tachycardia
 
Coronary guide wires
Coronary guide wires  Coronary guide wires
Coronary guide wires
 
Pace maker professor md toufiqur rahman
Pace maker professor md  toufiqur rahman Pace maker professor md  toufiqur rahman
Pace maker professor md toufiqur rahman
 
Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography
 
PROSTHETIC VALVES
PROSTHETIC VALVESPROSTHETIC VALVES
PROSTHETIC VALVES
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
 
Left ventricular angiogram (1)
Left ventricular angiogram (1)Left ventricular angiogram (1)
Left ventricular angiogram (1)
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
Basic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacingBasic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacing
 
PTMC/PBMC
PTMC/PBMCPTMC/PBMC
PTMC/PBMC
 
Evaluation of prosthetic valve function and clinical utility.
Evaluation of prosthetic valve function and clinical utility.Evaluation of prosthetic valve function and clinical utility.
Evaluation of prosthetic valve function and clinical utility.
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
BMV balloons- FINAL.pptx
BMV balloons- FINAL.pptxBMV balloons- FINAL.pptx
BMV balloons- FINAL.pptx
 

Similar to Components of Pacemaker and ICDs - understanding the hardware

ECG complete lecture presentation, ECG waveform and leads placement
ECG complete lecture presentation, ECG waveform and leads placementECG complete lecture presentation, ECG waveform and leads placement
ECG complete lecture presentation, ECG waveform and leads placement
DrSUVANATH
 

Similar to Components of Pacemaker and ICDs - understanding the hardware (20)

Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
pacemaker.pptx
pacemaker.pptxpacemaker.pptx
pacemaker.pptx
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
 
ECG Fundamentals.pptx
ECG Fundamentals.pptxECG Fundamentals.pptx
ECG Fundamentals.pptx
 
6a electrophysiology of heart
6a electrophysiology of heart6a electrophysiology of heart
6a electrophysiology of heart
 
Pacemaker and anaesthesia
Pacemaker and anaesthesiaPacemaker and anaesthesia
Pacemaker and anaesthesia
 
PACEMAKER
PACEMAKERPACEMAKER
PACEMAKER
 
Electrophysiology of heart
Electrophysiology of heartElectrophysiology of heart
Electrophysiology of heart
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Present5
Present5Present5
Present5
 
ECG Basics 1.pptx
ECG Basics 1.pptxECG Basics 1.pptx
ECG Basics 1.pptx
 
The pacemaker
The pacemakerThe pacemaker
The pacemaker
 
Defibrillation
DefibrillationDefibrillation
Defibrillation
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemaker
 
pacemaker
pacemakerpacemaker
pacemaker
 
Cardiac monitoring.pdf
Cardiac monitoring.pdfCardiac monitoring.pdf
Cardiac monitoring.pdf
 
Bioelectronics devices
Bioelectronics devicesBioelectronics devices
Bioelectronics devices
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
ECG complete lecture presentation, ECG waveform and leads placement
ECG complete lecture presentation, ECG waveform and leads placementECG complete lecture presentation, ECG waveform and leads placement
ECG complete lecture presentation, ECG waveform and leads placement
 
Electrocardiogram pptx
Electrocardiogram pptxElectrocardiogram pptx
Electrocardiogram pptx
 

More from Raghu Kishore Galla

More from Raghu Kishore Galla (19)

Trans septal Puncture in Cardiology
Trans septal Puncture in CardiologyTrans septal Puncture in Cardiology
Trans septal Puncture in Cardiology
 
CORONARY ARTERY DISEASE(CAD) in WOMEN
CORONARY ARTERY DISEASE(CAD) in WOMENCORONARY ARTERY DISEASE(CAD) in WOMEN
CORONARY ARTERY DISEASE(CAD) in WOMEN
 
Ibutilide
Ibutilide Ibutilide
Ibutilide
 
Tachycardia discriminating algorithms and trouble shooting of ICDs
Tachycardia discriminating algorithms and trouble shooting of ICDsTachycardia discriminating algorithms and trouble shooting of ICDs
Tachycardia discriminating algorithms and trouble shooting of ICDs
 
Fetal Cardiac Interventions
Fetal Cardiac InterventionsFetal Cardiac Interventions
Fetal Cardiac Interventions
 
Localization of WPW( accessory Pathway) by surface ECG
Localization of WPW( accessory Pathway) by surface ECGLocalization of WPW( accessory Pathway) by surface ECG
Localization of WPW( accessory Pathway) by surface ECG
 
Transcatheter therapy for Mitral Regurgitation (MR)
Transcatheter therapy for Mitral Regurgitation (MR)Transcatheter therapy for Mitral Regurgitation (MR)
Transcatheter therapy for Mitral Regurgitation (MR)
 
Sinus Node Dysfunction
Sinus Node DysfunctionSinus Node Dysfunction
Sinus Node Dysfunction
 
Ductus dependent circulation
Ductus dependent circulationDuctus dependent circulation
Ductus dependent circulation
 
Conduction system of the heart
Conduction system of the heartConduction system of the heart
Conduction system of the heart
 
History of echocardiography
History of echocardiographyHistory of echocardiography
History of echocardiography
 
Congenital LV and RV inflow anomalies by Echocardiography
Congenital LV and RV inflow anomalies by EchocardiographyCongenital LV and RV inflow anomalies by Echocardiography
Congenital LV and RV inflow anomalies by Echocardiography
 
PET scan in Cardilogy
PET scan in CardilogyPET scan in Cardilogy
PET scan in Cardilogy
 
Bioprosthetic valve thrombosis
Bioprosthetic valve thrombosisBioprosthetic valve thrombosis
Bioprosthetic valve thrombosis
 
Echocardiographic evaluation of of coronary arteries
Echocardiographic evaluation of  of coronary arteriesEchocardiographic evaluation of  of coronary arteries
Echocardiographic evaluation of of coronary arteries
 
CRT Non Responders - A practical guide
CRT Non Responders - A practical guideCRT Non Responders - A practical guide
CRT Non Responders - A practical guide
 
Inotropic therapy for heart failure
Inotropic therapy for heart failureInotropic therapy for heart failure
Inotropic therapy for heart failure
 
Echo of repaired tof
Echo of repaired tofEcho of repaired tof
Echo of repaired tof
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 

Recently uploaded

Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 

Recently uploaded (20)

Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 

Components of Pacemaker and ICDs - understanding the hardware

  • 1. Components of cardiac pacing and ICD- Understand the hardware Dr Raghu Kishore Galla
  • 2. History: 1700’s Cardiac electrostimulation began in the mid-18th century with the use of currents from the Leyden jar or Voltaic Pile (Allessandra Volta, 1799) to stimulate cardiac nerves and muscles in animals and to attempt resuscitation of intact dead animals.
  • 3. History: 1800’s Dr.DeSanctis used “Re-animation Chair.” as described by Richmond Reece in his The Medical Guide (1820). It had 3 pertinent features: a bellows to give forced ventilation, a metallic tube to be inserted into the esophagus and a voltaic pile attached at one pole to the esophageal tube and at the other to an electrode. The electrode was to be successively touched to “the regions of the heart, the diaphragm and the stomach….
  • 4. History: 1900’s In 1899, Prevost and Battelli demonstrated that electrical currents could cause ventricular fibrillation that often could be reversed by another powerful stimulus of either alternating or direct current. Robinvitch in a series of reports from 1907 through 1909 confirmed this work and designed the first portable electrical resuscitative apparatus for ambulances.
  • 5. 1905 – Einthoven Published first human AV block using string galvanometer
  • 6. History: 1930’s In 1932, Albert Hyman developed a machine for controlled repetitive electrostimulation of heart and named his device the “artificial cardiac pacemaker”
  • 7. • 1950 - John Hopps - Canadian electrical engineer paired with Dr. Wilfred Bigelow to create a pacemaker that could be used for long periods and treat many of the new dysarhythmias that had been discovered. • The device that sat on a tabletop was large and heavy.
  • 8. • 1951 - Dr. Paul Zoll a cardiologist from Boston created pacemaker using modern transistors but still was large, heavy and relied on AC power. • Despite these shortfalls the Zoll pacemaker remained a standard for pacing patients for several years.
  • 9. History: 1950’s • The first clinical implantation into a human of a fully implantable pacemaker was in 1958 at the Karolinska Institute in Solna, Sweden, using a pacemaker designed by Rune Elmqvist and surgeon Åke Senning. • It connected to electrodes attached to the myocardium of the heart by thoracotomy. The device failed after three hours. A second device was then implanted which lasted for two days.
  • 10. • The world's first implantable pacemaker patient, Arne Larsson, went on to receive 5 different lead systems and 22 pulse generators of 11 different models of pacemakers during his lifetime. • He died in 2001 with cancer, at the age of 86, outliving the inventor as well as the surgeon
  • 11. • 1981 – Rate responsive pacing by QT interval, respiration, and movement • 1994 – Cardiac resynchronization pacing • 1998 – Automatic capture detection • 2016 - Leadless pacemaker • Now – Approximately 3 million with pacemakers – Approximately 1 million with ICD device
  • 12. Introduction • Modern day pacemakers serve the primary function of either maintaining a minimum heart rate to avoid symptomatic or potentially life-threatening brady or tachyarrhythmias or offering resynchronization between the left and right ventricles in the setting of heart failure. • Multiple advances in device design, programming, sensor technology, and materials science have afforded the ability to offer an ever-widening range of devices with a variety of specialized features.
  • 13. Pacemaker Functions 1. Stimulate cardiac depolarization 2. Sense intrinsic cardiac function 3. Respond to increased metabolic demand by providing rate responsive pacing 4. Provide diagnostic information stored by the pacemaker
  • 14. Basic Concepts of Pacing • Cardiac myocytes may be “activated” by the delivery of an electrical pacing stimulus. • This stimulus creates an electrical field that allows for the generation of a self- propagating wave front of action potentials that may then advance from the stimulation site. • The minimum amplitude and duration required to generate the self-propagating wave front that results in cardiac activation is referred to as the threshold.
  • 15. • In order to provide for consistent myocardial stimulation, there needs to be a constant source of energy for pulse delivery, which is provided by the pulse generator, a conductor that will deliver the stimulus from the source, which is the lead itself, an electrode at the end of the conductor that delivers the pulse, and underlying myocardium that is excitable. • Failure at any of these points (e.g. due to fracture in the lead or loss of contact between the electrode and the underlying myocardium) will result in failure of myocardial stimulation. • Currently available pacing systems comprise a pulse generator and one or more pacemaker leads with lead tips positioned in the cardiac chamber of interest. Basic Concepts of Pacing
  • 16. Basic Concepts of Pacing • Sensing of local electrograms (EGMs) is essential to the proper function of permanent pacemakers. • Whether sensing takes place in a unipolar or bipolar system and despite the differences between the two, the principles of sensing remain the same. • Two electrodes, a cathode and an anode, are required to complete the electrical circuit between the body and the pacemaker. • In a bipolar system, both anode and cathode are located in the heart, whereas in a unipolar system only the cathode is located in the heart and the pacemaker generator can serve as the anode.
  • 17. Basic Concepts of Pacing • As a wavefront of depolarization travels toward an endocardial electrode in contact with resting myocardium, the electrode becomes positively charged relative to the depolarized region. This is manifest in the intracardiac EGM as a positive deflection. • As the wave front of depolarization passes under the recording electrode, the outside of the cell suddenly becomes negatively charged relative to resting myocardium, and a brisk negative deflection is seen in the intracardiac EGM. • The positive and negative deflections that precede and follow the intrinsic deflection represent activation in neighboring regions of myocardium relative to the recording electrode.
  • 18. Basic Concepts of Pacing A typical bipolar ventricular electrogram (EGM) in a normal individual. The sharp downward deflection in the EGM represents the intrinsic deflection and indicates the moment of activation under the recording electrode. The slope of the intrinsic deflection (dV/dt) is expressed in volts per second and is referred to as the slew rate. For an EGM to be sensed by a sensing amplifier, the amplitude and slew rate must exceed the sensing thresholds.
  • 19. Bipolar Versus Unipolar Stimulation • Modern lead design has largely eliminated any significant clinical difference between bipolar and unipolar stimulation, but some inherent differences remain. • Historically, the unipolar threshold was lower than the bipolar threshold in the same system due to a markedly larger anode and resulting reduced impedance in the unipolar system. • With the contemporary lead tip design, this difference is attenuated as impedance is mainly determined by the lead tip properties.
  • 20. Bipolar Versus Unipolar Stimulation • The clinically important implication of unipolar pacing is that the pacemaker generator may cause pectoral muscle stimulation (muscle twitching) if implanted adjacent to the pectoral muscle. • Because the distance between the cathode and anode is so much greater during unipolar pacing, the pacing stimulus on the surface ECG is also much larger than during bipolar pacing. • The increased size of the “pacer spike” may be visually helpful to determine proper pacemaker function when interpreting cardiac telemetry recordings or trans- telephonic ECG tracings during remote pacemaker monitoring.
  • 21. Bipolar Versus Unipolar Stimulation • The larger amplitude signal from unipolar pacing is far more likely to interfere with appropriate sensing of the cardiac rhythm by an ICD in the rare instance that a patient has a separate pacemaker and ICD implanted. • Specifically, a sensed large-amplitude unipolar pacing stimulus will result in subsequent under-detection of low-amplitude intracardiac signals, such as those that occur during ventricular fibrillation. • It is for this reason that the use of bipolar pacing is critical when a separate pacemaker is present in an ICD patient. • Bipolar pacing leads are also necessary for some rate-adaptive sensors, such as the minute-ventilation sensor, as well as for some types of automatic threshold algorithms, where the ring electrode is used to sense myocardial capture.
  • 22. Bipolar Versus Unipolar Stimulation • Due to the much smaller field of view, bipolar electrodes tend to be minimally influenced by electrical signals that originate outside the heart, while unipolar leads may detect electrical signals that originate near the pulse generator pocket. • Thus, unipolar sensing is more susceptible to interference from electrical signals originating in skeletal muscle (myopotentials), which can result in inappropriate inhibition of pacing (if sensed in the ventricular channel) or triggering of pacing output (if sensed in the atrial channel). • Thus, bipolar sensing is less prone to oversensing myopotentials, far-field cardiac signals, or electromagnetic interference from environmental sources. • Shielding of the device and signal processing in the pacemaker circuitry, such as rectification and filtering of lower or higher frequency components, also help to attenuate recording of unwanted sign
  • 23. Single-Chamber System: • The pacing lead is implanted in the atrium or ventricle, depending on the chamber to be paced and sensed Advantages: • Implantation of a single lead. Disadvantages: • Single ventricular lead does not provide AV synchrony. • Single atrial lead does not provide ventricular backup if A-to-V conduction is lost
  • 24. Dual-Chamber System: • Have Two Leads • One lead implanted in the atrium • One lead implanted in the ventricle
  • 25. Benefits of Dual Chamber Pacing: • Provides AV synchrony • Lower incidence of atrial fibrillation • Lower risk of systemic embolism and stroke • Lower incidence of new congestive heart failure • Lower mortality and higher survival rates
  • 26.
  • 27. Pacemaker Components A basic pacing system is made up of: • Implantable pulse generator that contains: – A power source—the battery within the pulse generator that generates the impulse – Circuitry—controls pacemaker operations – • Leads—Insulated wires that deliver electrical impulses from the pulse generator to the heart. Leads also transmit electrical signals from the heart to the pulse generator. • Electrode—a conductor located at the end of the lead; delivers the impulse to the heart
  • 28.
  • 29. Schematic of the typical circuits found in modern pacemakers
  • 30. The Pulse Generator • Contains a battery that provides the energy for sending electrical impulses to the heart • Houses the circuitry that controls pacemaker operations
  • 31. Battery • A battery converts chemical energy into electrical energy. • The source of this energy is the electrochemical reactions that occur within the battery. • During a spontaneous chemical reaction, substances react to form more stable products. • Redox reactions: types of reactions in which oxidation and reduction occur. They are called redox reactions because electrons are transferred from one reactant to another. 2Li + I2  2LiI
  • 32. Battery • The major parts of a battery are the anode, the cathode, and the electrolyte. • The anode and cathode must be physically separated and both must be in contact with the electrolyte
  • 33. Battery • In a spontaneous galvanic cell such as a battery, the anode is electrically negative and the cathode positive. • In nonspontaneous electrochemical reactions, such as those that occur at pacing electrodes, electrochemical reactions are driven by an externally applied voltage. • In this case oxidation(loss of electrons) will occur at the positive electrode and reduction (gain of electrons) at the negative electrode. • Thus for pacing electrodes, the anode is positive and the cathode is negative. • The polarities are reversed, but the underlying electrochemical processes defining anode and cathode, namely, oxidation and reduction, respectively, are the same
  • 34. Electrolyte • The anode and the cathode must be separated from each other so they cannot directly react with one another, but they also need to be connected by an ionically conducting medium called the electrolyte. • As the battery discharges, it furnishes electrons at one terminal, pushes them through an external circuit, and receives them back at a second terminal. • The electrolyte allows electrical charge, as ions, to flow within the battery, completing the circuit. • The electrolyte must conduct ions but not electrons.
  • 35. Separator • The separator is a structural member of the battery that keeps the anode and cathode materials physically apart, thus preventing shorting of the battery. • In batteries with liquid electrolytes, the separator is usually a porous polymer film that is immersed in and permeated by the electrolyte. • In the case of lithium/iodine batteries, traditionally used for pacemakers, the separator and electrolyte are one and the same, namely, the growing layer of solid, ionically conductive lithium iodide discharge product.
  • 36. Current Collector • The current collector makes the connection between the positive or negative terminal of the battery and its respective active electrode material inside the cell. • It is usually a wire connected to a screen, or grid, which is embedded in the anode or cathode material. • The current collector may also serve as a structural member of the battery to provide physical integrity and strength to that electrode.
  • 37. Sealing of Batteries • Batteries are well sealed, most often in hermetically welded containers with glass electrical feedthroughs to make the electrical connections between the inside and the outside of the battery. • Sealing is necessary to prevent any interchanges of materials between the battery and its surroundings. • Medical batteries are typically considered hermetically sealed if the leak rate out of the battery for a test gas, usually helium, is less than 1 × 10−7 cm3/sec at one atmosphere pressure difference between the inside and the outside of the battery.
  • 38. Types of Batteries Primary Batteries • can only be used once. • not designed to be recharged • Most batteries used to power modern implantable medical devices are primary batteries that use lithium anodes because such batteries • have very high energy densities. Secondary Batteries • Designed to be repetitively discharged and recharged. Eg: Lithium-ion batteries – use in LVADs, and implantable neurologic stimulators for control of chronic pain. • Although they are not currently in development for pacemakers and ICDs, features such as more frequent telemetry could cause manufacturers to reconsider lithium-ion battery use in cardiac applications in the future.
  • 39. Implantable Battery Design Requirements • The most important requirement in battery selection for implantable devices is high reliability. • Other significant factors include the desired longevity of the device (directly related to battery energy density, circuit design, and overall device size) and an appropriate indication of impending battery depletion (end-of-service warning). • The basic considerations when designing a battery for an implantable medical device include the current variations that can be expected from the circuit as the device provides its service for individual patients. • Once the range of these application requirements are defined, the current, voltage, and capacity requirements of the battery can be determined.
  • 40. Elective Replacement Indicator(ERI) • All modern implantable pulse generators have an ERI that alerts the clinician to impending battery depletion and allows adequate time for replacement of the device. • Typically, this indicator is designed to occur at least 3 months before the battery voltage drops to a level that would result in erratic pacing, loss of capture, or loss of other critical features. Many newer devices conform to CENELEC standards (European Committee for Electrotechnical Standardization). • For pacemakers, the standard requires a recommended replacement time (RRT) such that at least 95% of the devices will have a prolonged service period (PSP) of at least 180 days. • For ICD and CRT devices, the requirement is that at least 95% of the devices have a PSP of at least 90 days
  • 41. Battery Chemistries Used in Pacemakers The Lithium/Iodine Battery • In general, lithium/iodine batteries have a single, central lithium anode that is surrounded by cathode material which is at least 96% iodine and has been thermally reacted with a polymer material to render a conductive mixture. • Central anode with an embedded current collector wire and the iodine cathode that fills much of the volume inside the battery.
  • 42. The Lithium/Carbon Monofluoride Battery: • CFx is a cathode material with high capacity and moderate power capability. • The CFxmaterial is a powder that is mixed with carbon as a conductivity enhancer plus a polymer binder and then pressed into a porous pellet. • The battery is composed of a lithium anode, a porous polymer separator material, and the porous pressed-powder cathode pellet that are all inserted into a case which is welded closed. • In contrast to the lithium/iodine battery, the lithium/CFxbattery uses a liquid electrolyte that consists of a lithium salt dissolved in an organic solvent • Voltage decline near end of battery life tends to be fairly abrupt, making it challenging for device manufacturers to engineer means of maintaining adequate replacement time warning.
  • 43. The Lithium/Hybrid Cathode Battery: • A new lithium battery chemistry with a cathode comprised of silver vanadium oxide (SVO), plus CFx, has been developed to meet the needs of implantable devices with higher rate therapies and features. • SVO has high power capability and is the same cathode material used to power ICDs. • The blend of the two cathode materials, called a hybrid cathode, yields a primary battery that has an energy density equal to that of a lithium/iodine battery with approximately 100 times more power capability.
  • 44. The Lithium/Manganese Dioxide Battery • Lithium/manganese dioxide batteries are being used in a new generation of pacemakers commercialized by one manufacturer. • Manganese dioxide possesses good energy and power density characteristics and shows low self-discharge. • As lithium/manganese dioxide batteries demonstrate greater power capability in comparison to lithium/iodine batteries, devices that require higher power, such as wireless telemetry capable of communicating over longer distances, can be successfully powered
  • 45. Miniature Batteries for Leadless Devices • Leadless pacemakers are powered by a cylindrical, pin-style, lithium metal based primary battery that typically contains a monolithic, thick cathode with adequate electrode area. • Provide sufficient power for background electronics, stimulation, and low-power telemetry, while maintaining high energy density to achieve longevity on par with traditional pacemakers. • Another distinguishing feature of the miniature batteries used in leadless pacemakers is that the battery may not be fully enclosed in a titanium (device) case as in traditional pacemakers; • Instead, the battery may be welded to the end of the container housing the electronics and the fixation mechanism, such that the battery enclosure may be in direct contact with the body fluids.
  • 46. Microprocessors • Microprocessors have become the standard control circuits of implantable pacemakers and ICDs. • Microprocessors have several advantages over older integrated circuits, including a far greater circuit density and greatly reduced current drain. • Microprocessors also allow very sophisticated algorithms, requiring multiple calculations, to be incorporated into implantable devices, and have vastly increased data storage. • The microprocessor can respond to changes in programming instructions that allow functions to be added or changed after implantation. • The integrated circuit of pulse generators may contain both read-only memory (ROM) and random access memory (RAM).
  • 47. Microprocessors • ROM (typically 256 KB to 1 MB) is used to guide the sensing and output circuits. • In addition, RAM is used to store diagnostic information regarding pacing rate, intrinsic heart rates, and sensor output. • The amount of RAM (1–16 MB) included in the pulse generator varies between models and manufacturers, but has rapidly increased in modern pulse generators, allowing for a far greater amount of diagnostic information to be stored. • The rapidly expanding diagnostic capabilities of pacemakers has allowed for improved assessment of the physiological condition of the patient, including stored information about heart rate variability, respiration, intracardiac pressure, patient activity, lung water, and arrhythmia logs.
  • 48. Output circuit • The output circuit contains the output section and voltage multipliers. • Pacing outputs of greater than the specified voltage of a cell are achieved by a variety of methods. • One approach is to use capacitors located on the output circuits. These capacitors are charged by the battery in parallel, but then discharged in series. • Multiple capacitors can be used to achieve the desired output voltages. • An alternative approach to capacitor-based voltage multipliers is to use the electromagnetic principle of inductance.
  • 49. Sensing circuit • The intracardiac EGM is conducted from the electrodes to the sensing circuit of the PG, where it is amplified and filtered. • The intracardiac EGM is filtered to remove unwanted frequencies, a process that markedly affects the amplitude of the processed signal. • Following filtering of the intracardiac signal, the processed signal is compared with a reference voltage to determine if the signal exceeds a threshold detection level (programmed sensitivity). • Signals with amplitudes greater than the sensitivity threshold level are sensed as intracardiac events, whereas signals of lower amplitude are discarded as noise. • Signals that exceed the threshold level are sent to the timing circuit and logic circuits.
  • 50. Sensing circuit • Most permanent pacemakers also contain noise reversion circuits that change the pulse generator to an asynchronous pacing mode when the sensing threshold level is exceeded at a rate faster than the noise reversion rate. • The noise reversion mode prevents inhibition of pacing in the presence of electro magnetic interference. • The sensing amplifier must prevent the detection of unwanted intracardiac signals, such as far-field R waves in the atrial EGM, afterpotentials, T waves, and retrogradely conducted P waves.
  • 51. Timing circuit • The pacing cycle length, sensing refractory and alert periods, pulse duration and AV interval are precisely regulated by the timing circuit of the pulse generator. • The timing circuit of a pulse generator is a crystal oscillator that generates a very accurate signal with a frequency in the kilo Hertz range. • The output of the crystal oscillator is sent to a digital timing and logic control circuit that operates internally generated clocks at divisions of the oscillator frequency. • The output of the logic control circuit is a logic pulse that triggers the output pacing pulse, the blanking and refractory intervals, and the AV delay.
  • 52. Timing circuit • The timing circuit also receives input from the sense amplifier to reset the escape intervals of an inhibited pacing system or trigger initiation of an AV delay for triggered pacing modes. • The pulse generator also contains a rate limiting circuit that prevents the pacing rate from exceeding an upper limit in the case of a random component failure. • This runaway protection rate is typically in the range of 180–200 ppm.
  • 53. Magnet mode • Pacemakers and some ICDs have a reed switch which normally is open until a magnet or magnetic field [e.g. a magnetic resonance imaging (MRI) scanner] comes into close contact with the device. • The magnet will close the reed switch and the device will be switched to the “magnet mode.” • The magnet mode typically causes asynchronous pacing for a pacemaker (e.g. VOO, DOO, AOO) and deactivation of tachycardia therapy for most defibrillators, but bradycardia pacing and sensing functions are typically not affected. • The magnet may be unable to close the reed switch if it is not powerful enough or the magnet function is programmed “off.” • In contemporary devices (some pacemakers and most ICDs), other technology, such as the Hall effect sensor, integrated solid-state detection, or GMR sensor, is used in place of the reed switch to optimize response to the magnetic
  • 54. Rate Responsive Pacing Sensors: • When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output • There are several different types of sensors that may be employed in pacemakers. Most commonly, sensors work to augment heart rate in response to physical activity, whether by detecting patient movement or changes in respiratory rate. • Those most accepted in the market place are: – Activity sensors that detect physical movement and increase the rate according to the level of activity – Minute ventilation sensors that measure the change in respiration rate and tidal volume via transthoracic impedance readings
  • 55. Types of sensors • Using technology that detects changes in vibration, acceleration, or minute ventilation, pacemakers may be pre-programmed to induce concomitant changes in the paced rate if the physiological heart rate does not meet those parameters. • Other sensors may use sensitivity to temperature, QT interval, or local myocardial contractility to respond to conditions under which increases in heart rate are desired, but for which sensors that detect only changes in physical activity would be inadequate.
  • 56.
  • 57.
  • 58. Leads Leads Are Insulated Wires That: • Deliver electrical impulses from the pulse generator to the heart. • Sense cardiac depolarization.
  • 59. During Pacing, the Impulse: • Begins in the pulse generator • Flows through the lead and the cathode (–) • Stimulates the heart • Returns to the anode (+)
  • 60. Components of pacemakers leads 1) Electrodes 2) Conductors 3) Insulation 4) Connector pin 5) Fixation mechanism
  • 61.
  • 62. • A Unipolar Pacing System Contains a Lead with Only One Electrode Within the Heart. • In This System, the Impulse: – Flows through the tip electrode (cathode) – Stimulates the heart – Returns through body fluid and tissue to the IPG (anode)
  • 63. Unipolar leads • Simplest lead design of all the leads have only one conductor surrounded by insulation. • Tip of the lead is the cathode and the PG is the anode. • Largely replaced by bipolar lead designs for most of the endocardial applications. • Still commonly used in CS branches for CRT • Demonstated impressive longevity and many are still in service today
  • 64. • A Bipolar Pacing System Contains a Lead with Two Electrodes Within the Heart. • In This System, the Impulse: – Flows through the tip electrode located at the end of the lead wire – Stimulates the heart – Returns to the ring electrode above the lead tip
  • 65. Bipolar lead • PG is not the part of pace/sense circuit. • Both the ring electrode(anode) and the tip electrode(cathode) are in contact with the myocardium • Can be used for endocardial, CS and epicardial applications. • Two main bipolar lead designs – Co-axial Co-radial
  • 66. Types of lead design A) Coaxial lead body design consists of two nested coil conductors to the tip and ring electrodes, each wrapped in a separate layer of insulation. The inner coil connects to the tip cathode and contains a central lumen for stylet passage; the outer coil connects to the ring anode.can be attached to endocardium via active/passive fixing mechanisms but limited by large diameters B) In coradial design, a single coil contains the parallel conductor strands to the tip and ring electrodes, with each strand covered by a fluoropolymer(ETFE) insulator. A surrounding layer of outer insulation by polyurethane covers the lead body.Design can allow significantly smaller (<6F) diameters but limits the fixation mechanism.
  • 67. ICD leads • Defibrillator leads have greater complexity, with two, three, or four conductors that connect to the pacing/sensing electrodes as well as to the high-voltage shocking coil(s), depending on the specific type of lead. • Because of the prohibitive bulk that would result from a coaxial design with more than two conductors, ICD leads generally have a different type of structure, known as a multilumen design. • This type of ICD lead consists of a long cylinder of insulating material, with separate internal channels running down its length. Each conductor runs down an individual channel, usually with its own additional covering tube of insulation.
  • 68. ICD leads Multilumen ICD lead design consists of a single body of insulation with several internal channels that contain the conductor elements (central pace/sense coil and outer high- voltage cables), each covered by its own layer of fluoropolymer insulation. A surrounding layer of outer polyurethane-based insulation is present. HP, High- performance.
  • 69. ICD lead - Structural Overview • Right ventricular (RV) defibrillation leads comprise a distal tip electrode with a fixation mechanism that anchors the lead to the heart, proximal terminals that connect to the generator, and a lead body that connects the two. • The lead body consists of a flexible insulating cylinder with 3 to 6 parallel longitudinal lumens through which conductors run from the proximal terminals to small pace-sense electrodes and larger shock coil electrodes. • This multilumen design permits more conductors in smaller-diameter leads than older coaxial designs. • The subcutaneous ICD uses a parasternal electrode in which the larger shock coil is straddled by 2 small sensing electrodes. • ICD leads differ in number of shock coils, number of sensing electrodes, and type of connector terminals.
  • 70. Dual-Coil Versus Single-Coil Leads • All transvenous leads have a distal shock coil in the RV. Dual coil leads have an additional proximal shock coil, which usually lies in the superior vena cava. • Dual-coil leads were required for the first transvenous ICD systems, which preceded those in which the housing (can) of the generator served as a defibrillation electrode (active can). • With contemporary biphasic waveforms and active cans, differences in defibrillation efficacy between dual-coil and single-coil shock pathways are rarely clinically significant for left-pectoral implants.
  • 71. Dual-Coil Versus Single-Coil Leads • The principal disadvantage of dual-coil leads is greater risk of extraction because of fibrotic tissue ingrowth into the proximal coil. • An additional disadvantage is the possibility of a short between high-voltage elements of opposite polarity. • However, dual-coil leads retain advantages, including more reliable atrial cardioversion, coil-coil electrograms for morphology algorithms that discriminate VT from SVT, and proximal coil-can EGMs that display atrial signals to improve interpretation of single-chamber EGMs.
  • 72. Integrated Versus True Bipolar Sensing • Sensing can either be true bipolar between the tip electrode and a small ring electrode or integrated bipolar between the tip electrode and RV coil. • Modern integrated bipolar and true bipolar leads have similar performance for sensing ventricular fibrillation (VF) but differ in susceptibility to oversensing. • The larger integrated bipolar antenna is more susceptible to external electromagnetic interference, R-wave double counting, and diaphragmatic myopotentials. • True bipolar leads are more susceptible to T-wave oversensing, probably because of greater variations in R-wave amplitude. • There is no consensus regarding which is preferred.
  • 73.
  • 74. Multipolar leads • The complex challenges of CRT system have resulted in development of multipolar leads. • A central coil conductor extends to the tip, allowing the stylet or guidewire insertion. • Conductors for the more proximal electrodes are cables arranged in parallel that terminate at their respective electodes. • Unlike ICD leads the individual elements may be smaller in diameter as they need not conduct high voltages needed for defibrillation.
  • 75. Multipolar leads • Overall diameter of the leads can be small allowing stable positioning in the branches of CS. • Since multiple pacing vectors are available, it may be possible to overcome phrenic nerve stimulation, pace from electrically more advantageous sites or pace from sites with better acing thresholds.
  • 76. CRT leads • Leads used in CRT deserve special consideration with respect to fixation and shape. • CS branch anatomy is highly variable. Manufacturers offer a range of CS leads and it is up to the implanting physician to decide which lead is appropriate for a given patient. • The implanter has to balance the frequently competing objectives of lead stability, electrical performance, avoidance of phrenic nerve capture, and resynchronization performance. • Familiarity with a range of sizes, shapes, and fixation mechanisms of available leads is necessary for successful implantation. • Lead diameters vary from 4 to 6 Fr, some leads do not have isodiametric lead bodies. Most leads use friction and tension to maintain their position within the CS vasculature.
  • 77. CRT leads • Most CS leads that are transvenously placed do not have a fixation helix or screw that embeds the lead to the myocardium. • Most leads use friction and tension to maintain their position within the CS vasculature. • Most leads are bipolar, but some of the smaller diameter leads are unipolar. • Some leads (Boston Scientific Easy Trak2®) are straight with tines that allow them to be wedged into a distal target vessel and help provide additional stability through frictional force. • Other leads have multiple cants or curves that apply pressure to the vessel wall, again providing stability by frictional force.
  • 78. CRT leads (A)Bipolar lead with tines; (B) helical biased bipolar pacing lead; (C) bipolar pacing lead with can fixation; (D) helical biased, unipolar pacing lead; (E) bipolar pacing lead with curve fixation; (F) S-biased bipolar pacing lead. (Sources: (A, B, D, E) Boston Scientific Corporation. (C) Medtronic, Inc.. (F) St.Jude Medical,
  • 79. CRT leads • Alternatives to canted leads are leads with sigmoidal biasing in two dimensions (S shaped) or helical biasing in three dimensions (spiral). • These leads also rely on the friction created both from wedging the distal tip in a branch and from the multiple contact points on the vessel walls. • Despite the variety of available passive fixation CS leads, current passive leads often do not remain stable in the proximalportion of the CS tributaries. • If a distal branch cannot be engaged or is prohibited due to phrenic nerve or extracardiac stimulation, a proximal position can be targeted with a lead such as the Medtronic Attain 4195 StarFix® lead.
  • 80. CRT leads The Medtronic StarFixR lead with its fixation lobes deployed (top) and retracted (bottom). This has a unipolar design with three lobes that are deployed by advancing the push tubing around the lead.
  • 81. CRT leads • Modern CRT devices allow for separate pacing circuits for the RV and LV leads, and allow LV leads with unipolar, bipolar, and multipolar configurations. • The ability to separately assess the RV and LV leads allows the clinician to program parameters separately, and simplifies lead assessment while simultaneously reducing current drain on the battery. • separate programming of the LV and RV channels allows pacing with multiple LV configurations and with optional different timing between the two channels. • Multiple LV configurations allow non-invasive programming options to avoid phrenic nerve capture or minimize current drain during LV pacing.
  • 83. Fixation mechanisms • The chronic performance of permanent pacing leads is critically dependent on stable positioning of the electrode(s). • Proper fixation to the endocardial surface is essential to lead performance. • two types of fixation mechanisms: Passive Active
  • 84. Passive fixation leads • Passive fixation leads have tines (fins) near their tip, which are made of the same material as the insulation. • The number (two, three or four), length, and stiffness of the tines are variable. • Passive fixation leads become entrapped within the trabeculae of the RA appendage or RV apex immediately upon correct positioning of the lead. • Effective fixation of the lead can be confirmed at the time of implantation by its gentle traction or rotation. • Tines generally add minimal technical difficulty to the implantation procedure, although they may occasionally become entrapped in the tricuspid valve apparatus.
  • 85. Passive fixation leads • These leads are not suitable for placement in non-traditional locations such as the high RV septum or RV outflow tract. • The passive fixation leads are rapidly covered by fibrous tissue, making removal of the lead by simple traction difficult or impossible in as short a time as 6 months. • In general, passive fixation leads are more difficult to extract than active fixation leads.
  • 86. Active fixation leads • In the present generation of active fixation, endocardial leads have a fixed helix or extendable–retractable helix that in most cases also serves as a pacing electrode. • In leads with a fixed helix design, the helix is coated with mannitol or polyethylene to facilitate introduction of the lead through the vasculature into the desired chamber. The coating dissolves within a few minutes, allowing the lead to be positioned with a stylet. • Fixation is performed by rotation of the entire lead body and transmission of the torque to the distal tip. Repositioning is accomplished by reverse (counterclockwise) rotation of the lead.
  • 87. Active fixation leads • The extendable–retractable design has become the most widely used active fixation mechanism because of its ease of implantation and because the fixation mechanism allows retraction long after implantation, thereby facilitating repositioning or extraction. • Active fixation leads allow for stable positioning of the lead at many sites in either the atrium or the ventricle. • Although active fixation leads are easier to extract than passive fixation leads, the risk of myocardial perforation during and after implantation is higher. • Historically, the chronic pacing thresholds of active fixation leads were higher than those of passive fixation leads. This difference has largely been eliminated with the routine use of corticosteroids in active fixation leads.
  • 89. Epimyocardial leads • Permanent epicardial pacing leads are used when tricuspid valve anormalités, central venous obstruction, congenital heart disease, or technical issues preclude transvenous lead placement. • Epicardial leads may also be used when recurrent bloodstream infection is a concern, or when there is a need for lead implantation co-incident with another intrathoracic surgical procedure. • Epicardial pacing leads are either sutured to the epicardial surface or screwed into the epimyocardium of the atrium or ventricle. • Fixation mechanisms include a fishhook-shaped electrode that is stabbed into the atrial myocardium or a large screw helix that is rotated into the ventricular myocardium.
  • 90. Epimyocardial leads • These are typically unipolar and are frequently non-steroid eluting, with higher stimulation thresholds than modern endocardial electrodes. • A newer epicardial lead design has a steroid- eluting electrode, rests against the epicardium, and is sutured in place available in unipolar and bipolar models with lower chronic pacing thresholds and better long-term performance. • In the era of CRT, epicardial leads are also being used when a functional LV lead cannot be placed transvenously due to limitations imposed by the coronary venous anatomy From top to bottom: fish hook type, screw-inhelix type; and sew-on button type.
  • 91. Conductors • The primary conductors used in most pacing and ICD leads are MP35N and silver. • MP35N is a superalloy that is double melted to remove impurities. It is composed of cobalt, nickel, chromium, and molybdenum. • It is characterized by biocompatibility, high tensile strength, and resistance to corrosion. • Because of its relatively high electrical resistivity (1033 μΩ), MP35N is combined with efficient conductors such as silver. • Single conducting filaments (wires) are made in two designs: a drawn brazed strand (DBS) and a drawn filled tube (DFT).
  • 92. Conductors Cross-sections of the two types of conducting filaments (wires) used in pacemaker and ICD leads: Drawn brazed strand (DBS, left) and drawn filled tube (DFT, right).
  • 93. Conductors • The DBS design consists of the MP35N® alloy stranded around a core of softer, highly conductive silver. The resulting wire combines increased fatigue life, high conductivity, and solderability with the ability to withstand greater mechanical stresses. • The DFT design is coated with platinum or platinum–iridium. Extreme compressive forces are applied in order to form a sound mechanical bond between surfaces. • Single wires are combined together into strands and then wound into cables for use as cable conductors. Larger cables used for high voltage applications and smaller cables for pacing applications.
  • 94. • The number of filaments and their arrangement within a given cable is variable from manufacturer to manufacturer, even for similar applications. • Compared individually, a 7 × 7 (49 wire) cable has relatively greater flexibility but relatively lower torquability than a 1 × 19 (19 wire) design. • Neither design has been shown to be clinically superior.
  • 95. Insulation • The materials used in pacemaker leads for insulation play a critical role in their longevity and reliability, as well as in handling and implant characteristics. • The ideal insulation material should be biologically inert, and exhibit no surface erosions, no molecular chain disruptions, no uptake of low-molecular-weight biological materials, and no tendency to calcifications, while retaining stable mechanical properties. • There is no single ideal insulation and some leads are constructed using multiple materials. • Pacemaker leads are generally made of one insulation type. ICD leads are more complex, utilizing combinations of insulation material.
  • 96. Insulation • Fluoropolymers are fluorocarbon-based polymers characterized by high resistance to solvents, acids, and bases. They have maximum biocompatibility and tensile strength, but their stiffness limits their use to thin coating (<0.076 mm) applications. Eg: PTFE (polytetrafluoroethylene, i.e. Teflon®, DuPont) ETFE (ethylene tetrafluoroethylene). • Silicone rubber is a polymer both biostable and biocompatible, also has high resistance to extreme temperatures but has a low tensile strength, making it prone to tearing and abrasion wear. • Polyurethanes are characterized by high tear strength, high elasticity, and a low co- efficient of friction. These qualities allow for smaller lead diameters.
  • 98. Electrodes • The pacing electrode is the final interface between the lead and myocardium. Its design and composition greatly influence the overall electrical performance of the pacing system. • In a bipolar pacing lead the tip electrode is the cathode and the ring electrode is the anode. • The stimulation threshold is a function of the radius of the electrode. A smaller radius is associated with a higher current density, lower pacing threshold, and higher resistance at the electrode–myocardium interface. • In contrast, smaller radius electrodes result in high sensing impedance and electrode polarization impairing myocardial EGM sensing.
  • 99. Electrodes • The materials currently used for electrodes of permanent pacing leads include platinum–iridium, Elgiloy (cobalt–chromium–nickel), platinum coated with platinized titanium, vitreous or pyrolytic carbon coating a titanium or graphite core, platinum, iridium oxide, or titanium–nitride. • The use of corticosteroid-eluting electrodes has been a major advance in pacing lead technology. Dexamethasone sodium phosphate and/or dexamethasone acetate is impregnated on a silicone core or collar that surrounds the tip electrode. • Corticosteriods have dramatically reduced the risk of exit block for pacing leads, ICD leads, and epimyocardial leads.
  • 100. Steroid Eluting Leads • Steroid eluting leads reduce the inflammatory process and thus exhibit little to no acute stimulation threshold peaking and low chronic thresholds Porous, platinized tip for steroid elution Silicone rubber plug containing steroid Tines for stable fixation
  • 101. Connector pins • Connector pins are made of stainless steel or titanium. The current standard is IS-1 (Industry Standard) for pacing leads. • The lead diameter at the head block is 3.2 mm and sealing rings are integrated to the lead. With this industry standard, all current bipolar pacing leads are compatible with all current manufacturer header designs. • It is important to visualize the connector pin extending beyond the distal set screw in the header block. If the connector pin is not properly seated in the header block, sensing artifact or failure to pace may result. • Lead impedance measurements alone are not enough to detect this problem.
  • 102. Connector pins Adapter for leads terminated with 5/6-mm pins that need to be connected to a pacemaker with an IS-1 header block. Photograph and schematic of an IS-1 pin terminated lead. Sealing rings are integrated into the lead Radiograph demonstrating proper extension of the connector pin beyond the set-screw (top; arrow indicates fluoroscopic image of a pacemaker header with properly implanted lead pin) and inadequate extension beyond the set-screw (bottom; circle indicates improper lead pin position).
  • 103. Connector pins • Multipolar pacing leads such as those used in the CS will not be IS-1 compatible. They are connected to specialized header blocks using an IS-4 LLLL type pin similar to the pin assembly found on a DF-4 ICD lead. • Currently, there are two ISO standards for the ICD yoke (connector pin assembly): IS-1/DF-1 yokes and DF-4 (or IS-4) pin assembly. • Until recently, the IS-1/DF-1 yoke design was the industry standard. The pace/ sense portion of the lead is terminated with an IS-1 pin and the high voltage coils are terminated with their respective DF-1 (ISO-11318) pins.
  • 104. Connector pins • The DF-4 design eliminates the possibility of lead connection errors (i.e. interchanging the proximal and distal coils) due to a uniquely compatible port at the header. • This design also has advantage with respect to pocket bulk as well as for lead- to-lead and lead to- can interactions. • A potential disadvantage is the possibility of multiple component fracture from one critically located stress point in the pocket. Comparison of the current DF-1 connector (bottom) and the proposed DF-4 connector (top).
  • 105. Lead anchoring sleeve • The anchoring sleeve allows the lead to be secured to the fascia or muscle, and its proper use can help reduce the risk of lead damage. • Currently, most sleeves are made of a biocompatible, peroxide catalyzed (MDX) silicon rubber and may contain a radiographically dense material to allow visualization under fluoroscopy. • These sleeves have one, two, or three circumferential grooves to which non- absorbable sutures are secured. • Inadequate tie down force or an inadequate number of sutures is especially a problem with lead bodies coated with a lubricious material such as polyvinyl pyrrolidone (PVP). • Balancing the risk of lead damage with lead dislodgement requires the implanter’s familiarity with the specific lead and sleeve.
  • 107.
  • 108. Micra Transcatheter Pacing System • The miniaturized Micra™ transcatheter pacing system (TPS) is the world’s smallest pacemaker, delivered percutaneously via a minimally invasive approach, directly into the right ventricle and does not require the use of leads. • 99% implant success rate in 726-patient global trial • 48% fewer major complications than traditional pacemakers • 93% smaller than conventional pacemakers5 • Ultra low-power circuit design delivers an estimated average 12-year battery longevity. • MRI SureScan™ technology allows the patient to be safely scanned using either a 1.5T or 3T full body MRI.
  • 109. Linear one-step deployment facilitates consistent capsule placement, no torque required.