1. 1 2 3 4 5 6 7 8 9 10 11 12
Cost of Other 15-Oct-16 15-Jan-17 Feb 2017 March 2017
2
PURCHASE OF MEDICAL & LAB.
EQUIPMENTS
A09404 SDA 75,000,000
Single Stage
Two Envelope
Tender 15-Oct-16 15-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
3
PURCHASE OF PLANT &
MACHINERY
A09601 SDA 33,000,000
Single Stage
Two Envelope
Tender 15-Oct-16 15-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
4
REPAIR & MAINTENANCE OF
MEDICAL & LAB EQUIPMENT
A13102 SDA 21,000,000
Single Stage
Two Envelope
Tender 15-Oct-16 15-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
5
REPAIR & MAINTENANCE OF
PLANT & MACHINERY
A13101 SDA 10,500,000
Single Stage
Two Envelope
Tender 15-Oct-16 15-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
6 FURNITURE & FIXTURE A09701
SDA /
PLA
4,000,000
Single Stage
Two Envelope
Tender 15-Oct-16 15-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
7 PURCHASE OF IT EQUIPMENTS A9203 SDA 30,000,000
Single Stage
Two Envelope
Tender 22-Oct-16 22-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
8
REPAIR & MAINTENANCE OF IT
EQUIPMENTS
A13703 SDA 6,400,000
Single Stage
Two Envelope
Tender 22-Oct-16 22-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
9 BOOKS & PERIODICALS A03905 SDA 3,000,000
Single Stage
Two Envelope
Tender 22-Oct-16 22-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
10
SERVICES
PEST Control Services
A03919 SDA 3,500,000
Single Stage
Two Envelope
Tender 22-Oct-16 22-Jan-17 Feb 2017 March 2017
For smooth running
of hospital
RAWALPINDI INSTITUTE OF CARDIOLOGY, RAWAL ROAD, RAWALPINDI
2nd ANNUAL PROCUREMENT PLAN FOR THE YEAR 2016-17
(Under Rule 8 of the Punjab Procurement Rules 2014)
Name of the procuring agency: Rawalpindi Institute of Cardiology, Rawalpindi.
Sr. # Name of Procurement (Description)
Head of
Account
Alloction
Estimated
Cost
Procurement
Procedures /
Process
Procurement
Methods
Tentative Date of
Procurement
Notice Publication
Tentative Date of
Award of Contract
Tentative Delivery
Schedule
Tentative Date
of Completion
Benefits Likely to
Attain
For smooth running
of hospital
1 A03942 SDA 46,750,000
Single Stage
Two Envelope
Tender
2.
3. DEMANDS OF BOOKS & PERIODICALS FOR THE YEAR 2016-17
S.# NAME OF BOOKS QTY YEAR
BOOKS OF IT / BIO-MEDICAL
1
Medical Technology for the IT professional (An Essential Guide for
Working Today’s Healthcare setting)
01 2016
2 Medical Equipment Management Manual 01 2016
3 Journal of Clinical Engineering 01 2016
4
Dictionary of health Information Technology and Security (Kindle
Edition)
01 2016
5 The Clinical Engineering Handbook 01 2016
6
BMET Study Guide (Preparing for Certification and Sharpening Your
Skills, 2016 Edition
01 2016
7
HIMSS Dictionary of Healthcare Information Technology Terms,
Acronym and Organization,
01
Latest
Edition
8
Preparing for Success in Healthcare Information and Management System,
The CPHIMS Review Guide.
01
Latest
Edition
BOOKS OF GYNE
1 High Risk Management Option, 6th
/ 7th
Edition 01 2012
2 Textbook of Prenatal Medicine, 2nd
Edition 01 2006
3
Heart Disease in Pregnancy (Complications & Management) by Akmal –
El - Mazny
01 2014
BOOKS OF NURSING
1 Psychiatric nursing Lippincott William 4th
Edition 01 2011
2 Biological & behavioral concept by Thoms Delmar Learning 2003 01 2007
3 Mental Health Nursing & Evidence-Based Introduction 01 2011
BOOKS OF PEADS
1 BNF for Children 2012-13 01 2012
2 Forfar & arneil’s Textbook of Pediatrics with Online Access 7th
Edition. 01 2009
3 Medical Emergencies in Children 5th
Edition. 01 2012
4
Nelson Textbook of Pediatrics (3 Vol Set) With Online Access 20th
Edition.
01 -
5 Park’s Pediatric Cardiology for Practitioners 6th
Edition. 01 2014
6 Pediatric & Neonatal Mechanical Ventilation With CD 2nd
Edition. 01 2011
7 Rogers’ Handbook of Pediatric Intensive Care 4th
Edition. 01 2012
8 The Harriet Lane Handbook 20th
Edition. 01 2014
9
Association of Iron Deficiency Anemia With Febrile Seizures Latest
Edition.
01 -
10 Pediatric Cardiology by Robert Anderson Latest Edition. 01 -
11
Comprehensive Surgical Management of Congenital Heart Disease, by
Richard A. Jonas Latest Edition.
01 -
12 Pediatric Cardiac Surgery, Constantine Mavroudis ET AL. 3rd
Edition. 01 -
13 Congenital Cardiac Surgery, Bruce A. Reitz ET. Latest Edition. 01 -
14
Pediatric & Congenital Cardiology, Cardiac Surgery & Intensive Care.
Eduardo M, Da Craz, Ivy, Dunbar, James & Jaggers.
01 2014
15
Nadas Pediatric Cardiology, John F. Keane, James E.Lock, Donald
C.Flyer by Elsevier.
01
Latest
Edition
16
Critical Heart disease in infants & children-Darid G. Nicholas, Ross M
Ungerleider, Philip J. Spevak, William J. Greclay, Duke E Comeron.
01
Latest
Edition
17 Pediatric Cardiology, The Essential Pocket Guide 3rd
Edition, Walter 01 2016
4. DEMANDS OF BOOKS & PERIODICALS FOR THE YEAR 2016-17
Johnson James Moller Publisher: Willey Blackwell.
18
Manual of Perioperative Care in Adult Cardiac Surgery. Robert M. Bojar,
Willey Blackwell.
01
Latest
Edition
BOOKS OF NUCLEAR CARDIOLOGY
1
Radiation Protection in Medical Imaging and Radiation Oncology (Series
in Medical Physics and Biomedical Engineering) by Richard J. Vetter
(Editor), Magdalena S. Stoeva (Editor) (With CD)
01 2016
2
Nuclear Cardiology, an Issue of Cardiology Clinics By (author) Sharmila
Dorbala , By (author) Piotr Slomka (With CD)
01 2016
3
The Practice of Internal Dosimetry in Nuclear Medicine By (author)
Michael G. Stabin (With CD)
01 2016
4
Radiation Protection in Diagnostic X-Ray ImagingBy (author) Euclid
Seeram , By (author) Patrick C. Brennan (With CD)
01 2016
5
Radioisotope Studies in Cardiology by H J Biersack , by P H Cox (With
CD)
01 2012
6
Nuclear Cardiology Study Guide : A Technologist's Review for Passing
Specialty Certification Exams By (author) Andrzej Moniuszko , By
(author) B. Adrian Kesala (With CD)
01 2015
7
Nuclear Imaging in Clinical Cardiology Edited by M L Simoons , Edited
by Johan H C Reiber (With CD)
01 2014
8
Practical Mathematics In Nuclear Medicine Technology 2nd
Edition,
Patricia Wells, MAE, CNMT (With CD)
01 2014
9 Quick Reference protocol manual for NM Technologist (With CD) 01 2014
10
Nuclear Cardiology: The Basics. (How to maintain a Laboratory) By,
Frans J TH. Wackers MD, PhD, Wendy Bruni, BS, CNMT, Barry L Zaret.
MD(With CD)
01 2014
11
Nuclear and Radiation Physics in Medicine : A Primer By Tony
Key(With CD)
01 2014
12 Atlas of Nuclear Cardiology By (author) Narula Jagat Dilsizian Vasken 01 2014
13
Basic Science of Nuclear Medicine: The Bare Bone Essentials by K H
Lee, PHD (With CD)
01 2015
14 Health Physics, 4th
Edition, Herman Camber ((With CD)) 01
Latest
edition
15
Essentials of Nuclear Medicine Physics and Instrumentation By (author)
Rachel A. Powsner , By (author) Matthew R. Palmer , By (author) Edward
R. Powsner (With CD)
01 2013
BOOKS OF ANESTHESIA
1 Anesthesia Secrets Questions You Will Be Asked 4th
Ed 01 2013
2 Cardiac Anesthesia Practical Aspects 01 2015
3 Clinical Anesthesia Practice 2nd
Ed 01 2002
4 Duke’s Anesthesia Secrets 5th
Ed 01 2016
5
Foundations of Anesthesia Basic Sciences for Clinical Practice With CD-
ROM 2nd
Ed
01 2006
6 Handbook For Stoelting’s Anesthesia & Co-Existing Disease 3rd
Ed 01 2010
7 Handbook of Pediatric Anesthesia 01 2015
8 Handbook of Veterinary Anesthesia 5th
Ed 01 2013
9 Manual of Anesthesia For Operation Theater Technicians 01 2013
10 Manual of Local Anesthesia In Dentistry 01 2016
11 Miller’s Anesthesia With Online Access 2 Vols Set 8th
Ed 01 2015
5. DEMANDS OF BOOKS & PERIODICALS FOR THE YEAR 2016-17
12 Miller’s Anesthesia Review 2nd
Ed with Online Access 01 2013
13 Obstetric Anesthesia Clinical Anesthesiology Guide 01 2015
14
Shnider and Levinson’s Anesthesia For Obstetrics With Online Access 5th
Ed
01 2013
15 Stoelting’s Anesthesia & Co-Existing Disease 2nd
South Asia Ed 01 2015
16 The 5-Minute Anesthesia Consult With Online Access 01 2013
17 The Anesthesia Guide 01 2013
18 Understanding Anesthesia Equipment With Online Access 5th
Ed 01 2011
BOOKS OF RADIOLOGY
1 The Complete Guide to Cardiac CT. 01 2011
2 Practical Textbook of Cardiac CT & MRI (Tae-Hwanlim) Springer 01
Latest
Addition
3
Cardiac CT Imaging-Diagnosis of Cardiovascular Disease (Mathew J.)
01
Latest
Addition
BOOKS OF PATHOLOGY
1 Manual of Laboratory Medicine by AFIP (Latest Edition) 02 -
2 Robbins Basic Pathology (Latest Edition). 01 -
3 Standards for Blood Bank & Transfusion Services AABB (Latest Edition). 01 -
4 Microbiology by Prescott (Latest Edition). 01 -
5 Bailey & Scoh’s Diagnostic Microbiology (Latest Edition) 01 -
6
Clinical Chemistry by Bishop (Techniques, Principles, Correlation) Latest
Edition
01 -
7
Text Book of Blood Banking and Transfusion Medicine by Sally V.
Rudmann (Latest Edition)
01 -
8
Blood Transfusion in Clinical Medicine by Mollison & Marcela Contreras
(Latest Edition).
01 -
9 Practical Haematology by Dacie & Lewis (Latest Edition). 01 -
10
Lynch’s Medical Laboratory Technology by Stanley S Raphael, Methew J
Lynch (Latest Edition).
01 -
11
Clinical Haematology and Fundamentals of Hemostasis by Denise M
(Latest Edition).
01 -
12
Clinical Laboratory Medicine by Richard Ravel (Latest Edition).
01 -
BOOKS OF PERFUSIONIST / CARDIOPULMONARY BYPASS
1 Cardiopulmonary Bypass (Combridge Clinical Guides). 01 2009
2
System theory. A Modified Extra Corporeal Circuit to Attenuate Systemic
Inflammatory Response Syndrome by Kevin Mccuska PHD CCP).
01 -
3 Cardiac Perfusion and Pumping Engineering Echo by Dhajoo and Ghista. 01 -
4 Extra Corporeal CP Support in Critical Care 4th
Ed, by Gail Annich MD. 01 -
5 ECMO Specialist Training Manual 3rd
Ed, (congenital Books). 01 -
6
Comprehensive Surgical Management of Congenital Heart Disease by
Hodder Arnold Publication).
01 -
BOOKS OF PHARMACY
1 Stockley’s Drug Interaction 10th
Edition. 01 -
2 A Manual of Adverse Drug Interaction 5th
Edition. 01 -
3 Medical Toxicology by Richard 01 -
4 Clinical Pharmacy & Therapeutics by Herfindal. 01 -
5 BNF Peads 01 -
6. DEMANDS OF BOOKS & PERIODICALS FOR THE YEAR 2016-17
6 Quimp Latest Edtion 02 -
7 Community Pharmacy by Paul Rutter 01 -
8 Adverse Drug Reactions by Lee, Anne, 2nd
Edition. 01 -
BOOKS OF PHYSIOTHERAPY
1
Improving Functional Outcomes in Physical Rehabilitation 1st
Ed by
Susan B. O’Sullivan.
01 -
2
Orthotics and Prosthetics In Rehabilitation, 3e 3rd
Ed by Michelle M.
Lusardi PHD PT.
01 -
3
Geriatric Physical Therapy, 3rd
Ed by Andrew A. Guccione PT and Rita
Wong.
01 -
4
Physical Rehabilitation of the Injured Athlete: Expert Consult – Online
and Print, 4e by James R. Andrews MD and Gary L. Harrelson
01 -
8. 32 Curved arteries long 20
33 Curved arteries ST 20
34 Nerve hooks fine 10
35 Castro needle holder 8 inches ST 15
36 Castro needle holder 8 inches
curved
15
37 Ring tip forceps 8 inches 10
38 Metzenbaum scissors 8 inches
curved
20
39 ST scissors 8 inches curved 20
40 Mosquito artery forceps 100
41 Rongour 05
42 Bone nebler small 01
43 Bone nebler medium 01
44 Bone cutter 01
45 Cornery scissor forward 45o
10
46 Cornery scissor reverse 60o
10
47 O, Shaugnnesy 5 inches 05
48 O, Shaugnnesy 7 inches 05
49 O, Shaugnnesy 9 inches 05
50 Duct clamps (set of five) 01
51 Deaversretector (set of five) 01
52 Sponge holder 10
53 Arteries curved 7 inches 30
54 Mosqito arteries ST 50
55 Sim forceps 05
56 Potts smith (fine forceps) CA-16-
604
05
57 Waugh (Forceps) C-16-186 18cm 05
58 Tylor forceps C-16-098 05
59 Mixter (Rt angel) C-17-252 22cm 05
60 Baby mixter C-17-246 05
61 Needle holder fine 7 inches 10
62 Needle holder fine 9 inches 10
63 Needle holder fine 8 inches 10
64 Tenotmy scissor 8 inches 20
65 Codwan’s forceps fine 05
66 Seller’s chest spreader 01
67 Pen aortic clamp 05
B. Radiation Protection Devices
1. Liquid for cleaning of lead aprons 30
bottles
Used for cleaning of
lead aprons.
0.2M
2. Ovarian shield 25
As per sample
approval.
Approved any
international
certification. (CE
standard : EU directive
89/686/EU or
0.2M
3. Thyroid collar 15 0.15M
4. Head Cover (Visor) 25 0.25M
5. Apron sleeves 30 0.1M
6. Visors shields 25 0.25M
7. Patient half apron 25 0.5M
8. Breast shield (mammary gland
protective)
25
0.5M
9. 9. Pediatric gonad shield (male) 4 equivalent)
Light weight lead or
equivalent material
Warranty : one year
(start from at the time
of delivery)
As per standard
protection values
0.06M
10. Pediatric mittens (gloves) 4 0.06M
11. Pediatric gonad shield (females) 4 0.06M
12. Lead aprons (adults) 5 0.175M
13. Half Aprons (Cover Gonad
protection)
12
0.18M
C. CSSD
1. ETO kits 36 Used for sterilization
of soft items.
4M
D. Laundry Items
1. De-scalar for Washer drum 600 L Used for Descaling of
washer drums.
0.375M
2. Note: Firm should visit before quoting and take sample before
treatment and after treatment. At the time of services firm qualified
engineer should be there for guideline of chemical dosing.
E. Chemical For Boiler
1. Anti-
corrosion/anti
scale
900 L Used to maintain PH in boiler
and corrosion preventive
(Dynaflok or Equal)
0.56 M
2. De-scalar for
boiler
600 L used to deposit scale
sediment’s from boiler
0.37M
Note: Firm should visit before quoting and take sample before
treatment and after treatment. Also monthly water quality test will be
totally responsibility of the firm.
At the time of annual services firm qualified engineer should be there
for guideline of chemical dosing.
F. Chemicals For HVAC System
1. Chemical for cooling tower regular
dose
170 kg Use for daily cleaning
of cooling tower
0.068M
2. Chemical for cooling tower
descaling
100 kg Use for descaling at the
time of annual services
in cooling tower
0.023M
3. Chemical for close circuit (anti
corrosion anti scale.
1500kg Use for descaling of
close circuit and chiller.
0.75M
4. Chemical for FCUS cleaning 150kg Use for evaporator
cleaning of FCUS
0.067M
Note: Firm should visit before quoting and take sample before
treatment and after treatment. Also monthly water quality test will
betotally responsibility of the firm.
At the time of annual services firm qualified engineer should be there
for guideline of chemical dosing.
Total Estimated coast: RS: 14.898M
10. Annexure “A”
SPECIFICATION OF “HOSPITAL FURNITURE” FOR TENDER 2016-17
S/N
o
Item Required Qty Justification / Purpose
Estimated
coast
1 Crash Cart
10 Used in wards for medicine
and equipment’s
3M
2
Pediatric Patient
Trolley
02
Used for mobile purpose of
medicine and biomedical
equipment in Peads wards.
6M
3 Medicine Trolley 10
Used for mobile of
medicine in wards.
1.5M
4
Hydraulic
Patient Shifting
Trolley With
Matters
04
Used for shifting of un
concision cardiac patients
0.6M
5
Height Stool /
Chair
05
Used for doctors in catch
lab (control room).
0.05M
6 Patient Couch 20
Used for cardiac patients
when space is not
available for beds.
0.6M
Total: 11.75M
11. Sr. No 01
Clinical specialty Hospital furniture
Generic Name Crash cart
Clinical purpose Used in wards for medicine and equipment’s
Quantity 10
Technical specification
The emergency trolley must support different equipment for any type of
intervention.
Drawer always slide smoothly with double side wall construction.
All the accessories are easily attainable to reduce the maneuvers and the
times of intervention.
Accessories
6 drawers unit. Drawer size 3” x 3, 6” x 2 & 9” x 1.
Recessed top storage area.
Side shelf.
Accessory panels.
Defibrillator shelf.
4 TENTE Casters 2 lockable.
Breakaway locking bar.
Stabilizing frame with bumper.
Durable aluminum mounting tracks.
Light System.
Stainless Steel I.V. pole.
Double push handle.
1 x Cylinder holder for D or E size cylinder I.
Cardiac board with bracket.
Universal rail system fitted to width of trolley.
Operating manual, Services manual, error code book, part list and
software if any.
Optional: Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
12. Sr. No 02
Clinical specialty Hospital furniture
Generic Name Pediatric Station Trolley
Clinical purpose Used for mobile purpose of medicine and biomedical
equipment in Peads wards.
Quantity 02
Technical specification
(for use of Pediatric Resuscitation Emergencies)
Drawers are color Coded using the BroselowTMmethod for emergency
pediatric care.
Include:
9 – drawer, A-SMART premier Breakaway Lock Cart with Divider for Small Drawers.
(30” total drawer space).
Tuck-away two-way side shelf, 30” locking bar.(approx.)
Stabilizing frame / bumper system, seamless plastic top, soft grip push handles,
all-swivel TENTE casters (Two locking), and one swivel / fixed tracking caster.
configuration of drawers: 3” (8 No), 6” (1 No) (approx.)
Accessories
1 x Oxygen Tank Bracket.
1 x Suction Unit Holder.
1 x two hook I.V. pole.
Breakaway locking Bar.
Operating manual, Services manual, error code book, part list and
software if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
13. Sr. No 03
Clinical specialty Medical and Lab Equipment
Generic Name Medicine trolley
Clinical purpose Used for mobile of medicine in wards.
Quantity 10
Specification
For the storage and dispensing of medication by ward nursing staff.
Frame and carcass shall be constructed from mild steel powder coated/ Polymer.
Mounted on approx 125mm castors.
The approximate internal dimensions of the medicine container shall be 80 w x50 d x30 h
cm.
The hinged lid of the unit shall slope down to the front.
Lockable with heavy-duty dead lock.
Shall be fitted with the following accessories:
o Document support on under lid surface.
o 4 tier rack for holding medications.
o Waste bins qty 2 off mounted under the medicine container.
o Ampoule holding rack.
o Dispensing tray to collect spillages when pouring liquid oral medication.
o Side mounted fold-down writing flap.
o Different compartments for medicines with labeling.
o Lower shelf mounted under medicine container.
Unit shall be supplied with a security clamp for wall fixing at the appropriate location, thus
ensuring that the trolley is secure when not in use.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
14. Sr. No 04
Clinical specialty Medical and Lab Equipment
Generic Name Hydraulic Patient shifting trolley with matters
Clinical purpose Used for shifting of un concision cardiac patients
Quantity 04
Technical specification
Hydraulic Patient’s trolley for use in patient’s recovery and Emergency area transportation.
Upholstered top, with adjustable head full upholstery
Side rails
Bumper bar/push handle, castors
Safe Working Load 180kg
Head section adjustable.
Accessories
Matters fine quality medical dedicated and autoclave able.
IV pole
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
15. Sr. No 05
Clinical specialty Medical and Lab Equipment
Generic Name Height stool / chair
Clinical purpose Used for doctors in catch lab (control room)
Quantity 05
Technical specification
Height quality stains less steel with lather cover.
As per sample approval.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
16. DEMAND FOR “MEDICAL AND LABORATORY
EQUIPMENT” FOR TENDER 2016-17
S/No Equipment Name With Detail Specifications
Qty In
Hand
Qty
Required
Estimated
Cost
GYNAE DEPARTMENT
1 Anesthesia Work Station 0 1 6 M
2 OT Table for Gynae Surgery 0 1 2 M
3 TVS Probe USG 0 2 1 M
Operation theatre
4 Head Light 03 02 1 M
5 Cardio Blade Ablation Machine 0 01 5 M
6 Crisis Manikin 0 01 0.8 M
7 Bronchial Tree 0 01 0.3 M
8 Human Skeleton With Nerve System 0 01 0.3 M
9 Microplegia Delivery System Machine 0 01 1 M
10 Cerebral Ox Meter 0 01 1.5 M
11 Endoscopic Vein Harvesting System 0 03 30 M
12 Portable Ultrasound Machine 0 01 1.5 M
13 Live 3-D Echo Machine 0 1 20 M
LABORATORY
14 Water Quality Test Equipment 0 1 0.5 M
15 CO2 Incubator 0 01 0.8 M
16 Refrigerator Centrifuge (Cryofuge) 0 01 4.5 M
17 Plasma Thawing System 0 01 0.7 M
PHYSIOTHERAPY
18 Arm Ergo Meter 0 01 0.12 M
19 Multi Pulley System 0 01 1.5 M
20 Storage Rack With Material 0 01 0.3 M
21 Ultrasound Therapy Unit 0 01 0.5 M
22 Hydrocolator 0 01 0.4 M
23 Recumbent Bike 0 01 1.0 M
24 Telemetry Monitoring System 0 01 5 M
25 Rechargeable / Electric Therapeutic Portable /
handy Ultrasound 0 01 1 M
26 Compression Device 0 01 0.3 M
GENERAL MEDICAL EQUIPMENTS FOR ALL DEPARTMENTS
27 Cardiac Monitors (Vital sign) 186 150 16 M
28 Online Blood Pressure Monitor 0 10 8 M
29 Syringe Pump 174 30 4 M
30 Vain Finder 0 03 1.6 M
31 Pulse Oximeter (Portable) 14 10 1.2 M
32 Blood Warmer / Plasma Therm 05 02 1.5 M
33 Defibrillator (Pacing) 6 06 2.4 M
34 Mobile BP Apparatus 0 10 0.18 M
35 AED 0 04 1.5 M
PAEDS ECHO DEPARTMENT
36 Echocardiography Machine 0 1 4.5 M
17. HDU-II
37 Dialysis Machine 1 2 2.5 M
CATH LAB
38 Angio Jet and Jet Stream 0 01 14 M
39 Open Irrigation System 0 01 5.5 M
NUCLEAR MEDICINE
40
Contamination Probe for Survey Meter RADOS
RDS-31 Used To Detect Contamination (PNRA
Requirement)
0 1 0.3M
41
Syringe Pump Digital Injector Used For Patients
To Inject Adenosine
0 1 0.2 M
42
Printer For Dose Calibrator (Hot Lab) To Be
Used With VDC-405 Dose Calibrator To Print
The Daily QC Reports
0 1 0.1 M
43
Printer For Spectra Analyzer System, To Be
Used For The Analysis Of Thallium On Weekly
Basis
0 1 0.1 M
44 Dose Calibrator 0 1 0.8 M
45 Co-57 Flood Source 0 1 0.5 M
46 Co-57 point Source 0 1 0.5 M
47 Pocket Dosimeter 0 3 0.1 M
Radiology
48 Work Station for CT 02 01 5 M
49 Portable Ultrasound 0 01 1M
Infection control department
50 Fogging Machine 0 01 6M
Total Estimated Cost: 167.4 M
SPECIFICATION FOR “MEDICAL AND LABORATORY
EQUIPMENT” FOR TENDER 2016-17
Sr. No 01
Clinical specialty Medical and lab Equipment
Generic Name Anesthesia Work Station
Quantity 01
Clinical purpose Delivery of anesthesia during Operation.
Technical specification
3 Gas Anesthesia Workstation, 02, N20 and Air Up to 1 x O2 and 1 x N20
Cylinder can be connected
Pipelines/fittings/gauges for Central Piping 1 x 02 and 1 x N20 Yokes &
Gauges for Cylinders 1 x 02, 1 x N2O and 1 x Air Pipeline, Pipeline Fitting
and Pressure Gauges Main Working Surface
1 x Monitoring Shelf 1 x Side Folding Shelf
Single 02 High Pressure Outlet 02 Flush
Common Gas Outlet
Side Rails and Centrally Braked Castors with Large Brake Paddle 2 x
Drawers
Anesthesia Machine should be capable of working at Low Flow Anesthesia
18. Scavenging System
Venture Suction
1KG Circle/CO2 Absorber, fully Autoclave able, including Bag/Vent switch,
CO2 by pass and manometer gauge, complete with mounting bracket,
disposable patient circuit, absorber detachable including Y-piece circuit, bag
and fresh gas assembly 1 x Bag Arm 2 Station Selecta Tec Back bar to
accommodate up to two vaporizers at a time.
Electronic / digital flow-meter:
Three Gas Configurations (02, N20 and Air)
Flow Range:
02 and Air: 0 to 12 LPM
N2O: 0 to 12 LPM
Total Flow Meter 0 to 10 LPM Calibrated to 02
Thermal Mass Flow Sensor Type
Mechanical, Hypoxic Guard for minimum of 25% 02 / Auxiliary Flow Meter
Low flow oxygen at 200ml flow meter
Anaesthesia Ventilator
Microprocessor Controlled Anesthesia Ventilator
Full Color Minimum 6 - 8" or more LCD/TFT Touch Screen Display
Automatic Leak and Compliance Test
Fresh Gas and Compensation
Ventilation Modes: Volume Control, Pressure Control. SIMV+PSV, Pressure
Support, CPAP, Spontaneous manual
Ventilator Monitoring:
Oxygen (F102), Inspired & Expired Volumes, Paw, Pressure, Real Time
Waveforms for Flow, Volume,& Pressure
Tidal Volume: 20 — 1400 ml
IE Ratio: 2.0:1 — 1:5
Frequency: 4 — 60 bpm
Minute Volume: 0.3 — 25 LPM
Pressure Limit: 10 - 70 cmH2O (Adult) & 10-50 cmH2O (Paeds) PEEP: 3 —
20 cmH2O
Patient Trigger: 1 — 10 LPM
Battery Back Up of 60 minutes Minimum / charging time.
User Alarms: High/Low Pressure; High/Low Minute Volume; High/Low
Breath Per Minute (BPM); High/Low Oxygen Static Alarms: Apnea, Settings
Error, Low Supply Gas,
Inspiratory Flow Transducer Error, No Charge, Fresh Gas Too High,
Ventilator Inoperative, Under Pressure, Peep Error/Sensor Error, Power Fail,
Battery Low, Apnea Alarm in Bag Mode
User service manuals with error code book..
Anesthesia agent monitoring
FDA, CE or at least two international standards approved.
Accessories :
1 x Vaporizer, Selecta Tec Type Tec 7 or equivalent, Isoflurane
1 x Vaporizer, Selecta Tec Type Tec 7 or equivalent, Sevoflurane
1 Pin Index Cylinders for 02
1 Pin Index Cylinders for N2O
Reuse able/ auto clave able breathing circuit one for adults and one for
paeds.
All allied accessories.
19. Operating manual, Services manual, error code book, part list and software
if any.
Optional: nil
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
20. Sr. No 03
Clinical specialty Medical and lab Equipment
Generic Name TVS probe USG
Quantity 02
Clinical purpose Used for diagnostic purpose with Ultra sound
Technical specification:
Compactable with Toshiba nimo MX.
CE/ISO approved
Accessories: nil
Optional: Nil
Warranty :
Two year
Sr. No 02
Clinical specialty Medical and ward Equipment
Generic Name OT table for Gynae
Clinical purpose Used in Gynae OT for surgery
Quantity 01
Technical specification:
Hydraulic operation table for Gynae procedure
Table Top (Radiolucent) with antistatic mattress 4 section and equipped with
X-Ray cassette holder.
Table top arranged as Head plate, Back plate, Seat Plate, Two separate leg
plates.
All control of table movement toward head side X-Ray cassette tunnel full
length of table top.
Up and down movement height range 810-1250mm
Trendelenburg / Anti-trendelenburg + 300, Lateral tilt + 200
Back rest angle adjustment -150/+750 and Leg rest detachable; angle
adjustment – 850 / +150
Outward angle 900
Safe working load capacity 170-200Kg
FDA, CE or at least two international standards approved.
Accessories:
Complete with Anesthetic screen, body rest-holder, shoulder rest-holder,
Arm rest-holder, Leg sport, Knee rest-holder, Kidney bolster, Restraining
straps, Tray, Foot plate.
Operating manual, Services manual, error code book, part list and software
if any.
Optional : Nil
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
21. Sr. No 04
Clinical specialty Medical and lab Equipment
Generic Name Head light
Clinical purpose Used in OT department during surgery.
Quantity 02
Technical specification:
Operating prism loupes to be supplied on a spectacle frame.
The frames to be supplied in three sizes with five available nose bridges to
allow maximum comfort and security for the surgeon. 300/400 Watt
To be easily identifiable by color coding and simple to clip on when needed.
To have a pocket size light source with Brightness Control and a compatible
plug in charger. Complete system with xenon light source
Accessories :
Complete system with xenon light source.
Operating manual, Services manual, error code book, part list and software
if any.
Optional: Nil
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 05
Clinical specialty Medical and ward Equipment
Generic Name Cardio blade Ablation
Clinical purpose Used in Cardiac OT during surgery
Quantity 01
Technical specification:
Argon grade North America
Pre-purified Argon≥ 99.998% pure (grade 4.8)
Tank pressure 6000psi
Tank volume ~ 50 liters
Tank diameter 10 inches
Tank Height 56 inches
Tank valve fitting CGA77 or CGA680
Cryo time per cylinder
Accessories :
Cylinder cap and valve.
Operating manual, Services manual, error code book, part list and software
if any.
Optional : Nil
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
22. part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 06
Clinical specialty Medical and ward Equipment
Generic Name Crisis Manikin
Clinical purpose Used for training purpose of Cardiac Anesthesia
Quantity 01
Technical specification
Ambu cardiac care trainer with Manual Monitor and CPR software (ECH
Box Built in having 26 ECG Rhythms) and Windows compatible Laptop core
i7, Third generation (HP, Dell, Acer)
Full body, IV arm and Heart core
At least two international standards meet.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 07
Clinical specialty Medical and Lab Equipment
Generic Name Bronchial tree
Clinical purpose Used for training purpose of Anesthesia department
Quantity 01
Technical specification
As per sample approval
US, Europe, Japan
Accessories
Operating manual, Services manual,
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 08
Clinical specialty Medical and Lab Equipment
Generic Name Human skeleton
Clinical purpose Used for training purpose of Anesthesia department
23. Quantity 01
Technical specification
As per sample approval
US, Europe, Japan
Accessories
Nil
Optional
Nil
Warranty
Two year with all spare parts,
Sr. No 09
Clinical specialty Medical and Lab Equipment
Generic Name Microplegia delivery system machine
Clinical purpose Used for patient during surgery.
Quantity 01
Technical specification
(Microplegia Delivery System Machine)
Microplegia Delivery System Shall be optimal for (Myocardial Protection for Cardiac
Surgery Procedure)
General Feature
System shall feature an intelligent operating modes which consists of:
Pressure Must be Auto-regulated
Temperature Required Warm, Cold, Tepid
Arrest Faction Must Be Control OR set by Systems (As Required)
Flow control Required / Continuous Precise, and Cyclic Delivery of Microplegia.
System Delivery Route / Antegrade, retrograde and simulgrade.
Composition Required / Oxygenated Blood, natural buffers.
Technical Specification for System
Flow rate: 10 – 999ml / min.
Accuracy: 1ml / min for 10-100 ml / min.
Resolution: 5 ml / min for 100-999 ml / min.
Air Bubble detection: 100 mul cohesive bubble.
Prime Volume: 190 ml & 45 ml heat exchanger.
Arrest & additive capacity: 50 ml.
Temperature Range: Cold: Ice water controlled.
Temperature Range: Warm: 4-42 Degree Centigrade.
Temperature Accuracy: +- 1 Degree Centigrade.
Blood to Crystalloid Ratio Control: 100% Crystalloid / 100% Blood.
Ration Accuracy: 3% of each component required proportion.
Arrest agent concentration rang: 0 – 40 mEq/L.
Additive concentration rang: 0 – 50 ml/L.
System shall be portable (Easily Moveable).
System shall be fit on HLM (Heart Lungs Machine) if required.
Rated voltage (mains operation) 220 – 240 VAC, 60 – 50 HZ.
Manufacture Requirement
Manufacturer should have ISO Certification for Quality Standards.
FDA / CE Approval.
Country of Origin / USA, Europe.
24. Accessories: Operating manual, Services manual, error code book, part list
and software if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 10
Clinical specialty Medical and Lab Equipment
Generic Name Cerebral Oximeter
Clinical purpose Used for training purpose of Anesthesia department
Quantity 01
Technical specification:
The device should be FDA cleared for measuring absolute cerebral tissue
oxygen latest model.
Saturation (Sto2) for all patient population; neonatal (<8kg) and
pediatric/adult in their cleared weight and sensor size.
Adult sensor with an interrogation depth of 2.5cm.
4 different sensor sizes with 3 different penetration depths in order to
Accommodate for the anatomical differences of the different patient
population.
Compensate for the interference of stool (meconium) when measuring
abdominal
StO2 in neonates <4kg.
Compensate for the effects of melanin and patient specific tissue properties
when
Measuring cerebral Sto2.
Allow for age and weight entry to compensate for anatomical and human
growth
differences.
The device’s data should refresh at every 2 seconds or less.
The device should utilize at least 4 laser wavelengths to achieve greater
precision
and accuracy in Sto2 measurements.
The device’s Co-Oximeter Reference value can be obtained from the formula
StO2=SaO2 X 0.3 + SjvO2 X 0.7.
Accessories: Operating manual, Services manual, error code book, part list
and software if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
25. Sr. No 11
Clinical specialty Medical and Lab Equipment
Generic Name Endoscopic vein Harvesting system
Clinical purpose Used for vein harvesting during cardiac surgery.
Quantity 01
Technical specification:
Full HD Camera Control Unit CMOS Type Compatible with harvesting device
for Cardiac bypass vein and artery harvesting Latest and advanced kit.
Full HD Camera system with Latest Chip or Sensor technology for brilliant
image display.
Full HD image quality (1920 x 1080 pixel provides a crystal clear image)
Scan Mode of system should be Progressive Scan 50/60Hz
Video Signal Outputs: 2 x DVI - D 1080p (50/60Hz) one for teaching
purpose, one for monitor.
2 x HD - SDI 1080i (50/60Hz) one for teaching purpose, one for recording
device, 1 x S - Video PAL for printer.
Camera Head with Zoom coupler.
Camera Resolution: 1920x1080 Pixel
Buttons on Camera Head: at least 4 button with 5 Functions or more.
Optical Zoom:2x or more
Digital zoom: 2.5x or better
Length Camera Cable 4 meter minimum
Harvesting Tool: Should consist with following parts for better
activity.
Harvesting Cannula
Tool Adapter Port
C-Ring Slider
Hemo pro Harvesting Tool place.
Hemo pro Jaws tool space.
C-Ring must be part of it.
Harvesting Tool Extension Cable Connector
Scope Washer Connector (blue)
Distal Insufflations Connector
Activation Toggle for bisector.
Harvester Endoscope: 7mm extended length Telescope with Dissection Tip
or more.
It should have built tip nose for Target focus on vein. Should be 0
Degree Straight forward.
Endoscopic Harvester Power Generator: DC Type. It should have
Extension Cable Connector with Power Cord Connector and low Power
Setting Knob with LED Power-On Indicator and Hanger.
Cutting tool should not have thermal expansion for the better quality of Vein
or artery.
System should not have any bipolar cutting instrument or equipment along
with to avoid Thermal expansion.
System must be equipped with Harvester Extension cable and Harvester
Adapter
LED Light Source:-
Xenon 300 watt or more / LED.
Life span 500 hours on continuous use
26. Colour temperature : 6000 K
Light Guide Cable :1 pcs.
light intensity should be continuously adjustable
HIGH FLOW INSUFFLATOR:-
Maximum Gas Flow 20 L / min or more. With Pressure Range 5 - 25 mm Hg
Fully automatic insufflators, High pressure tube 1m.
Reusable silicon tube and Insufflations balloon with trocar.
System should be closed tunnel for the maximum safety of patient
and better affectivity.
Ultra-High Definition Medical Monitor 26'' with Bright LED
Backlight :-
Panel Quality Active Matrix TFT
Brightness (cd/m2) 800
Resolution Pixel 1920 * 1080 (Full HD) or better
Viewing Angle 1780 / 1780,
Aspect Ratio 16:9
Classification acc. To directive 93/42/EEC : I
Video Signal Inputs:
2 * DVI (DVI - D & DVI - I),
2 * SDI / HD – SDI (BNC),
S – Video (DIN – 4 or 2 * BNC)
System should be FDA and CE with full safety and features.
Country of Origin: UK, USA and Western Europe.
Accessories :
Disposable Kit Harvester: 25 X quantity Included: Harvester
Cannula with Traction device, Harvesting Bisector (Three in one
Cutting, coagulation, and dissection), Trocar Cannulae, syringe.
Disposable kit should have clinical studies more than 100,000
patients for better outcomes and trust on therapy.
Video Endoscopic Cart, Imported from same manufacturer.
Operating manual, Services manual, error code book, part list and
software if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 12
Clinical specialty Medical and Lab Equipment
Generic Name Portable ultra sound machine
Clinical purpose Used for ultra sound in different departments.
Quantity 01
Technical specification
A complete dedicated high end digital Color Doppler Ultrasound unit for wide range of
premium performance application of Abdominal, OBS/GYN, Urology, and vascular
imaging in pediatrics and adult. Mobile Portable system with built in workstation / data
management system for digital acquisition, storage and review of complete ultrasound
27. studies including static and dynamic clips in DICOM format, read/write zoom. Studies can
be reviewed and output to CD / DVD/USB. System Should have built-in battery. The
machine must have sharp and high quality image reproduction with heavy duty
performance. It should have minimum following specification :
DISPLAY:
High resolution non interlaced, flicker free 15” LCD/TFT display tilt able.
OPERATING MODES:
B, 2D M-Mode, Power Doppler, HPRF, Spectral Doppler, Color Doppler, Pw Doppler,
Duplex And Triplex Doppler, IMT imaging, Contrast Imaging, Panoramic Imaging,
Compound Imaging, 3D/4D imaging.
CONTROL PANEL:
Alphanumeric keyboard with built-in trackball.
Direct access to system functions through dedicated keys.
Indicator lights identify activated keys.
Audio volume control with bidirectional / stereo speakers and foot switch.
User selectable image magnification control.
Adjustable transmit focusing control.
Total and Lateral Gain Compensation controls (6 or more).
2 Active Probe connectors
CALIPER / MEASUREMENTS :
4 to 8 calipers for measurement per screen trace length measurements for:
Distance, angle, distance depth from skin line, area, circumferences, compound / volume,
OBS/GYN measurements, Urological measurements, Vascular Measurements with IMT.
APPLICATION:
Abdominal, OBS/GYN, Urology, vascular, Small Parts, Peripheral for pediatric, adult with
all required software for measurements.
DISPLAY MODES:
Live and stored display format: full size and split screen. Review image format: for still and
cine, simultaneous capability B+PW, B+ CFM +PW, B+ or triplex mode, B+ color split screen
display. 2D mode, , M-mode, color Doppler imaging, color Doppler imaging, color m-mode,
PW / HPRF Doppler, IMT, Panoramic Imaging, Vascular imaging , vascular calculations.
CINE MEMORY
Min. Cine Memory for 8000 frames or 250mb min.
IMAGE VIEWING DEPTH:
30 – 300 mm or more for cardiac application
IMAGING MODES / TECHNIQUES:
Tissue harmonic Imaging, Power Doppler Imaging,
Vascular imaging software for carotids
STORAGE DEVICE
Built-in CD / DVD/ USB Drive
SYSTEM DYNAMIC RANGE
Dynamic range minimum 200 dB or more
COMMUNICATION SOFTWARE
System should upgradable for DICOM 3 communication software for:
Image Storage, print, Query / Retrieve, Network Communication.
PORTS:
Video Output, USB / RS 232, Networking
POWER REQUIREMENT
220-240VAC 50 Hz
Battery Back up
3-4 hour built in battery backup.
FDA (510 k) and CE approved.
Accessories:
Multi frequency Convex Probe to cover 2.5 -6.0MHz.
28. Multi frequency Linear Probe to cover 7.5-12.0 MHz.
Imported Trolley
Thermal Printer 256-Gray scale
50 High Density / High Glossy thermal paper Rolls
Gel: 20 liters
Operating manual, Services manual, error code book, part list and software
if any.
Optional:
Multi frequency Transvaginal Probe to cover 5 -7.0MHz
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 13
Clinical specialty Medical and Lab Equipment
Generic Name 3-D echo machine
Clinical purpose Used for live 3-D echo cardiograph
Quantity 01
TECHNICAL SPECIFICATION:
At latest version Digital Echocardiography Unit for the clinical routine and
research and cardiovascular applications. All vendors should quote the top
of the line latest system.
System should have capability for high processing to achieve high frame rate
up to 2500 are better.
System Channels at least 5,500,000 channels. verified by OEM printed TDS
System frequency range 2-20 MHz
21” or more flat panel flicker free monitor LCD.
Tilt and swivel stand for LCD Monitor.
12” or more touch screen control panels.
PW Doppler with independent HPRF.
Image memory kit (4000 frames or more).
Automatic adjustment of B-gain and TGC with one touch button for image
optimization I Scan.
Keyboard with built in trackball and mouse.
Up to 2700 fps in 2D should be verified in the technical brochure of the
equipment.
Dual focus.
Contrast imaging using high or low MI (Mechanical Index). Optional
Contrast pulse sequencing technology (real time low mechanical index for
contrast agent examinations, Soft ware should have capability to show
contrast agent only, tissue only or contrast and tissue only. Optional
Contrast harmonic imaging capability.
Scanning Option:
Electronic Convex.
Electronic Sector.
Electronic Linear with independent PV steering.
B mode specification:
29. Viewing Depth: 2-30 cm.
Display Size: up to 10 x more with PAN / Zoom.
M. Mode Specification:
Scrolling with 4 speeds.
D Mode Specifications:
PWD, HPRF and CWD Doppler employing FET with window function
contrast control and base shift facilities, auto PRF facility.
Color Doppler Specifications:
CWD and PWD Doppler are continuously steerable in the color blood flow
image in real time.
Tissue Velocity.
Complete Cardiac Calculation in 2D and 3D imaging.
Cine Memory 2200 frames.
An integrated patient data management workstation with DCD/CD writer.
All required software in standard.
Tissue Harmonic imaging for cardiac and vascular imaging.
Display mode:
B/M Scan Mode.
o B single / dual.
o B/M.
o M.
Doppler Mode.
o B.
o B/M/D.
o B/D.
o M/D.
o D.
Color Doppler Mode.
Tissue Doppler Imaging.
Edge Assisted Ejection Fraction for assisted boundary detection technique.
Integrated multi stage stress echo system for advanced and flexible stress
echo Examinations.
Communication software Dicom 3.0 complete image storage, point, queries
retrieved, network communication.
2D Spectral Tracking.
Probes
Phased array sector probe for adults cover 2-3.5 MHz
Phased array Sector Probe for pediatrics to cover 3.5-7 MHz or better.
Live 3D/4D probe electronic switchable for adult – that can be used for
routine 2D/B mode as well
Optional Probe
Live 3D/4D probe electronic switchable for peads – that can be used for
routine 2D/B mode as well
3D/4D acquisition rate to be at least 400 mega voxel/sec
Multiplan TEE Probe (3.5-7 MHz or better) matrix quality for 3D (4D)
imaging for Adult and Peads echocardiography cases in mandatory.
3-D (4D) imaging software with complete on-cart and off-cart volumetric
quantification package including 2 Dimensional Quantification-Region of
interest, intimate media Thickness, Strain, Border, Detection, Parametric
Quantification, semi-automated border detection for cardiac chambers and
vessel cavities, LV systolic / diastolic functions including fraction area
change, peak ejection change, 3 dimensional quantification for distance
30. measurement, left ventricle volume, Mass Ejection / Fraction and MPR. LV
Segment based global volume wave form and 3 DCRT Evaluation.
3-D/4D Acquisition with minimum 400 mega voxels per second. Should be
verified by technical/sales brochure
International Standard Must be approved:
CE Marked (MDD) 93/42/EEC.
FDA 510K.
Future Upgrades:
Capability for Future up gradation.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 14
Clinical specialty Medical and Lab Equipment
Generic Name Water quality Test equipment
Clinical purpose Used to check quality of water in laboratory.
Quantity 01
Technical specification
CE marked .
Capable to test microbiology of water.
Take minimum time for test.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 15
Clinical specialty Medical and Lab Equipment
Generic Name CO2 Incubator
Clinical purpose Used in microbiology for microorganism growth.
Quantity 01
Technical specification:
Temperature Range: Up to 60 ºC
Control Accuracy: ± 0.5ºC
31. Volume: 50-80 litters
Display: LED
Operating Voltage: 220 Volts
Humidity Range (% RH): 90-95
CO2 Range:0-20%
Auto-sterilization with hot air at 180 °C
Humidification system with condensation protection
CO₂ gas-mixing jet with Venture effect
Builtin CO₂ sensor and measuring technology
Tightly sealed inner door made of tempered safety glass
Seamless, deep-drawn inner stainless steel chamber
Perforated shelves, stainless steel
Tempered glass inner door
with audible and visual alarm
Over temperature monitor
Easy-to-clean interior,
Reinforced by deep drawn ribbing with integrated and protected
large-area heating on four sides
ISO/FDA/CE marked
Accessories: Operating manual, Services manual, error code book, part list
and software if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 16
Clinical specialty Medical and Lab Equipment
Generic Name Refrigerator Centrifuge (Cryofuge)
Clinical purpose Used for the preparation of blood components (FFP,
platelets )
Quantity 01
Technical specification:
Max Capacity: 6 x 1000 ML bottle
12 quin or quad blood bags
Max RCF (*g): 5735 to 7,500
Max Speed (rpm): 3500 to 5, 00
Speed Range (rpm/g): 500 to 4900 rpm/ 80 to 7300 * g
Acceleration Rates: 08-12
Deceleration Rates: 08-12
Temperature Range: -8°C to +40°C
Control System: Microprocessor control, self diagnostics for servicing
Display: VIDEO set graphics (includes run graph), Info/Error messages
32. in two languages
Motor: High-performance Induction (brushless)
Max Power: 4500 VA
ISO/FDA/CE marked
Accessories: Operating manual, Services manual, error code book, part list
and software if any.
Optional : Nil
Warranty :
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 17
Clinical specialty Medical and Lab Equipment
Generic Name Plasma Thawing System:
Clinical purpose Used for thawing of FFP.
Quantity 01
Technical specification
Thaw Capacity: 06-08 Bags
Thaw time: Short thaw time 06-10 Minutes
Chamber Material: Stainless Steel
Specialized basket-unit for swing
Basket for easily loading
Basket Material: Stainless Steel
With over-temperature alarming system to protect FFP
Digital temperature monitor
Microprocessor temperature controller with audible and visual high
temperature alarm
Overwrap Protection
Large opening for easy cleaning.
ISO/FDA/CE marked
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional: Nil
Warranty:
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 18
Clinical specialty Medical and Lab Equipment
Generic Name Arm Ergometer
33. Clinical purpose Used for upper limbs exercise
Quantity 01
Technical specification
Approved and authorized for medical use
Arm ergometer ideal for Physiotherapy and cardio vascular rehabilitation
User interface must be user friendly
Speed self-generating
High level of safety
Resistance: Isokinetic (speed control): 15 speed settings (increments of 5
deg/sec) or more
Work Rate Range: Up to 600 watts (120 rpm) or more
Speed Control Range: 50 to 120 deg/sec or more adjustable
Readouts: Time, RPM, watts, calories, METs, heart rate, total work, total
cycles ormore.
Telemetry compatible
Accessories
Complete with standard accessories.
Operating manual, Services manual, error code book, part list and
software if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 19
Clinical specialty Medical and Lab Equipment
Generic Name Multi pulley system
Clinical purpose General/ multipurpose body exercise
Quantity 01
Technical specification
Approved and authorized for medical use.
Automated and computerized adjustment of weight.
High level of safety and adjust able seat height.
Pre-stretched nylon coated steel cables,wear free, 4 inch pulleys (or more
)withquality ball bearings.
200 lbs plastic weight (adjustable) stack, Steel safety shrouds around the
weight stack or more.
Offers over 25 (at least) or more upper and lower body exercises.
Dimension 1660mm (L) x 1160mm (W) x 2020mm (H) with in or near
about theses limits.
Back rest must be comfortable and provides good support.
AB strap station, curl/row bar, lat bar and adjustable velcro ankle strap
for leg workouts or more.
Accessories
Operating manual, Services manual, error code book, part list and software
34. if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 20
Clinical specialty Medical and Lab Equipment
Generic Name Storage rack with material
Clinical purpose Used for frozen shoulder and basic weight training.
Quantity 01
Technical specification
Storage rack of wooden
Wooden block of 10*6*4 and 20*10*5 (03*each)
Club original
Exercise stick of wooden 100cm (*6)
Medicine ball leather (1.5kg, 2kg, and 3kg)
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 21
Clinical specialty Medical and Lab Equipment
Generic Name Ultrasound Therapy unit
Clinical purpose Used for pain management at physiotherapy
department
Quantity 01
Technical specification
Ultrasound, Frequencies 1 and 3 MHz
Continuous and Pulse Frequency / Duty Cycle 100 Hz / 5, 10, 20, 50% or
more.
Multi-Frequency Treatment head 1 and 3 MHz Large
Contact Control
Automatic Power Switch-off and treatment time
Interruption in case of insufficient contact
35. The visual indicator of the treatment head switches on when in proper
contact
Number of connections : 1 or more
Intensity : 0-2 W/com2 continuous, 0-3 W/cm2 pulsed
Programmable memory
Mains voltage: 100-240 V + 10% (50/60 Hz) or other
Anatomical based treatment modes.
Dimensions: 290x280x105 mm or more
Complete with standard accessories
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 22
Clinical specialty Medical and Lab Equipment
Generic Name Hydrocolator
Clinical purpose Used for pain and spasm management at
physiotherapy department.
Quantity 01
Technical specification
Approved and authorized for medical used with slandered 12 moist pack
Pack of 25 x30 cm.
Easley moved on rubber wheels
Tank capacity: 69 litter
Temperature range: 71- 74 c or more
Power : 220-240v, 50/60 hz, 1000 w
Accessories
Standard accessories.
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
36. Sr. No 23
Clinical specialty Medical and Lab Equipment
Generic Name Recumbent bike
Clinical purpose Used in exercise for DCM patients.
Quantity 01
Technical specification
Approved and authorized for medical used.
Bicycle ideal for physiotherapy and cardio vascular rehabilitation
User interference must be user friendly multiple level of resistance
Speed self-generating adjustable seat arm support and back support
Compatible with telemetry
Smooth and noise free
High level of safety
Complete with stranded
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 24
Clinical specialty Medical and Lab Equipment
Generic Name Telemetry Monitoring System
Clinical purpose Monitoring system used during cardiac rehabilitation
Exercise.
Quantity 01
Technical specification
8 patients telemetry monitoring system upgrade able upto 32 patients
Dedicated for cardiac rehabilitation
12 lead ECG, HR, spo2, NIBP and temperature PR or more
PC based system architecture with windows operating system.
Pre-configured branded PC hardware from manufacturer/system with
LCD.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
37. equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 25
Clinical specialty Medical and Lab Equipment
Generic Name Rechargeable / Electric Therapeutic Portable / handy Ultrasound
Clinical purpose Used for pain management for different wards
(mobile purpose).
Quantity 01
Technical specification
1 MHz Ultrasound with Pulsed and Continuous Output
Rechargeable ultrasound either in a docking station or via a mains adaptor or USB
cable to your computer
High quality therapeutic ultrasound delivers the same output as used by
physiotherapists and healthcare professional in sports therapy
Very simple and easy to use
Top quality polished stainless steel transducer head delivers exceptional smooth
ultrasound transmissions for the best results in ultrasound therapy treatments.
30 Treatments from one single charge
Relive Pain
Helps reduce inflammation fast
Ideal for reducing scar tissue
Sports injuries - Neck & shoulder pains, ankle injuries, groin injuries hamstring
strains and much more
Sciatica pain, arthritis pains and joint pain
Cellulite reduction
Rechargeable
Comes Complete with:
1 x Ultrasound unit ,1 x Charging Station, 1 x Mains adaptor ,1 x Ultrasound Manual /
User Guide,1 x 250ml Ultrasound Gel, 1 x Professional Hard Carry Case ,Therapeutic
ultrasound in pulses at 3 power settings (low, med & high), Targeted pain relief .
Speed healing - ultrasound is used by physiotherapist in physical therapy to treat pain,
spasms & inflammation. Ultrasound promotes tissue heating & blood flow to speed
the recovery process & injury rehab.
Convenient & easy
FDA & CE Mark approved.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
38. Firm must send annual PPM schedule with installation report.
Sr. No 26
Clinical specialty Medical and Lab Equipment
Generic Name Pneumatic compression device
Clinical purpose Used for preferable edema and bed riddle patients
in physiotherapy department
Quantity 01
Technical specification
Delivers compression to leg or foot, or both simultaneously.
Clinically-proven, sequential, circumferential, gradient leg compression.
LCD Display
Inclusive Power Indicator displays power source.
One Touch Operation
Port A& B indicator.
Audio and visual alarms, e.g. Low battery, single port use etc.
Voltage: 220 VAC
A flexible and transportable design lets you decide when to plug in for therapy or to take
advantage of battery-powered freedom.
Battery Backup
Easy to transport with a small footprint and light- weight
Ergonomic handle and integrated versatile bed mount.
Compression Parameters
Sequential, Gradient, Circumferential
Vascular Refill Detection
Automatic pressure adjustment
Compact & Portable, patient bed hanging system.
Accessories
SCD Sleeve Thigh Length Small INT, SCD Sleeve Thigh Length Medium INT
SCD Sleeve Thigh Length Large INT, SCD Tubing Set, SCD Sleeve Knee Length Medium INT
SCD Sleeve Knee Extra Large, SCD Foot Cuffreg INT
Operating manual, Services manual, error code book, part list and software if
any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
39. Gen ward, CCU, HDU
Sr. No 27
Clinical specialty Medical and Lab Equipment
Generic Name Cardiac monitors(Vital sign monitor)
Clinical purpose Used to check vital signs of the patients.
Quantity 150
Technical specification
For monitoring of patients vital signs
Operating Features and Characteristics:
Non fade TFT color display
Electro-surgical interference suppression/protection
Defibrillator protection
Freeze and cascade facility
Waveform trace speed: 25 to 50 mm/sec
Screen Size: min. 10” or more TFT/LCD color Touch Screen display
PARAMETERS:
ECG:
Numeric: heart rate
Waveform: 12 Wave forms minimum, real time and freeze
ECG trace
Non-Invasive Blood Pressure (NIBP):
Method: Oscillometric principle
Numeric: Systolic, diastolic and mean pressures
Selectable auto inflate interval settings
Rising cuff / continuous pressure display
Temperature:
Numeric: Temperature selectable in oC / oF
Channels: 2 channels
Pulse Oximetry:
Numeric: 0-100% oxygen saturation measuring range.
Waveform: plethysmograph pulse with pulse strength indication.
Reusable sensor electrode.
Reusable cuff.
Arrhythmia Analysis:
Arrhythmia analysis and st analysis.
Minimum 10 arrhythmia items.
Respiration:
Breath rate display and alarms.
Sweep speed; 12.5 to 50 mm/sec.
Other Features:
Trend Data; up to 72 hours trend graphical and tabular
Alarms:
High & low (settable) on all parameters
Visual and audible indication of alarms.
40. Operating Requirements:
Ac 22ov/50hz
Built-in rechargeable battery for at least 1-2 hours AC power failure
at full
parameter
Wall Mount (Original / Local)
Capability of central monitoring system.
CE, FDA, JIS or at least two international standards approved.
Accessories
Note-1: The system must be complete with all sensors, probes, cables or
any other accessories required for measuring all the above selected
parameters. Monitor should be compatible with central monitoring
system.
Note-2: All allied accessories should be original and from the same
manufacturer.
Operating manual, Services manual, error code book, part list and
software if any.
Optional
IABP, ETCO
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 28
Clinical specialty Medical and Lab Equipment
Generic Name Automatic Blood pressure monitor with stand
Clinical purpose Used for blood pressure monitoring in different
department.
Quantity 10
41. Technical specification
Measurement method: Oscillometric
NIBP Measurement range
Systolic:
Diastolic:
Pulse:
40-270 mmHf or better
20-200 mmHg or better
30-240 bpm or better
Display range: 0.-300 mmHg or better
Measurement accuracy
Pressure:
Pulse
+ 3mmHg or better
+ 5mmHg or better
Cuff size:
Thermal printer:
125x300 mm or better
Should allows easy paper roll replacement
and avoid possible paper jam.
Display type: LED Digital
Should have mobile card, adjustable height stool and foot switch.
Should have clock function for date & time
Should have convenient for heart patients, pregnant women’s, patient with back
problem & patient in wheel chair.
Should have safety feature fast stop button, one button operation for stop /
start etc.
Country of origin: US, Europe, Japan or Equivalent
Standards should have CE or other international standard approved.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 29
Clinical specialty Medical and Lab Equipment
Generic Name Syringe Pump
Clinical purpose Used for delivery of calculated medicine to patients
42. Quantity 30
Technical specification
Syringe pump for fluid administration.
Flow Rates: 0.1 - 300 ml/hr.
Digital display of set parameters.
Universal Syringe acceptance capability for disposable, Plastic, Size, 10,
20, 50 ml.
Drive Accuracy. ±3%
Display of drug name, Infusion rate, infused volume and volume to be
infused.
Automatic adaptation of controls according to syringe /infusion set.
Quick freed/rapid infusion facility.
Rechargeable battery and mains operated 220V, 50Hz.
Safety alarm audible and acoustic for occlusion end of infusion, low
battery.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Docking Station:
Mobile Docking Station for simple and quick pump installation.
Features:
Designed to save valuable space around the patient bed.
Allow fast and easy pump mounting.
Infusion Stand.
Technical Specification:
Power 100-240 VAC, 50/60 Hz
Electrical Safety: EN60601-1 or equivalent FDA, CE of other two
international slandered approved.
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 30
Clinical specialty Medical and Lab Equipment
Generic Name Vain finder
Clinical purpose Used for visualization of veins
Quantity 01
Technical specification
Should detect veins to a depth of 10mm.
Sterilize able and medical grade design
Easy to clean
Handheld or hands-free operation.
Battery operated
43. Accessories
Mobile cart
Charger
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 31
Clinical specialty Medical and Lab Equipment
Generic Name Pulse oximeter (Portable)
Clinical purpose Used for check the oxygen saturation of patients.
Quantity 10
Technical specification
Approved and authorized for medical use
Portable Finger Ox meter For emergency Use
Provide SPO2 and pulse rate, pulse strength or more
User friendly with LED display on probe
Auto power shut down after 8-15 seconds when not in use
Low battery indicator flash light
Monitor patient range from pediatrics to adult
Combined monitor and sensor in one, light weight, pocket size.
Battery backup 4hrs or more.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 32
Clinical specialty Medical and Lab Equipment
Generic Name Blood Warmer / Plasmatherm
Clinical purpose Used to warm and shake blood / plasma.
Quantity 02
44. Technical specification:
FOR WARMING OF BLOOD BAGS
MINIMUM THREE TO FOUR BAGS CAPACITY
Temperature setting: +37ºC to+44ºC
Microprocessor controlled unit
Blood warmer
Voltage: 220 V, 50 Hz
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 33
Clinical specialty Medical and Lab Equipment
Generic Name Defibrillator (Pacing)
Clinical purpose Used for defibrillation at different departments.
Quantity 06
Technical specification
Biphasic transthoracic (external) defibrillator with TFT/LCD colour display
Synchronized output with ECG.
Control of energy charging/ delivering on main panel & apex paddle.
Charging Indicators.
The energy range should be adjustable for paeds and adults up to 200Joules.
Charging Time for full energy will be less than 5 sec.
Screen Size of approx. inch coloured or more.
Display of HR, ECG through paddles and Lead I.II & III patient cable.
Built in recorder for printing of full summery on standard 50mm paper.
Alarms for High and low Heart rate, low battery warning.
AC 220V / 50Hz operated.
Built-in Rechargeable battery with charger for minimum 50 shocks at max. energy.
Auto tester/self check.
External pediatric and adults Paddles, ECG cable.
Pacing facility.
AED facility with cable.
Accessories
All standard accessories.
2x Pacing cables
Operating manual, Services manual, error code book, part list and software
if any.
Optional
45. Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 34
Clinical specialty Medical and Lab Equipment
Generic Name Mobile BP Apparatus.
Clinical purpose For BP monitoring in different department.
Quantity 10
Technical specification
CE, ISO /FDA approved.
As per sample approval.
Accessories
Imported trolley .
Optional
Nil
Warranty
Two year with all spare parts.
Sr. No 35
Clinical specialty Medical and Lab Equipment
Generic Name AED
Clinical purpose Used in Ambulance for defibrillation.
Quantity 04
Technical specification
Synchronized output with ECG.
The energy range should be adjustable for paeds and adults up to 200Joules.
Charging Time for full energy will be less than 5 sec..
Display of HR, ECG .
Alarms for High and low Heart rate, low battery warning.
AC 220V / 50Hz and 12 v Dc operated.
Built-in Rechargeable battery with charger for minimum 50 shocks at max energy.
Auto tester/self check.
Pacing facility.
Accessories
10x Disposable Electrodes
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
46. Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 36
Clinical specialty Medical and Lab Equipment
Generic Name Echocardiography machine
Clinical purpose Use for echocardiography for OPD patients.
Quantity 01
SPECIFICATION:
ECHOCARDIOGRAPHY MACHINE.
A complete dedicated digital Echocardiography unit for wide range of premium
performance application of cardiovascular imaging in pediatrics and adult. Mobile trolley
mounted system with built in workstation / data management system for digital
acquisition, storage and review of complete ultrasound studies including static and dynamic
clips in DICOM format, read/write zoom. Studies can be reviewed and output to CD /
DVD/MOD. The machine must have sharp and high quality image reproduction with heavy
duty performance. It should have minimum following specification :
DISPLAY:
High resolution 1280x 1024 non interlaced, flicker free.
Display size Min. 17” Flat/TFT, tilt able and swive able type.
OPERATING MODES:
B, 2D M-Mode, Power Doppler, HPRF, Spectral Doppler, Color Doppler, Velocity Mode, Pw
Doppler, Duplex And Triplex Doppler, CW Doppler Steerable and ECG Gating,
CONTROL PANEL:
Alphanumeric keyboard with built-in trackball.
Direct access to system functions through dedicated keys.
Indicator lights identify activated keys.
Audio volume control with bidirectional / stereo speakers and foot switch
User selectable image magnification control.
Adjustable transmit focusing control.
Total and Lateral Gran Compensation controls (6 or more).
CALIPER / MEASUREMENTS :
6 to 8 calipers for measurement per screen trace length measurements for:
Distance, angle, distance depth from skin line, area, circumferences, compound / volume,
slope, time, heart rate and acceleration.
APPLICATION:
Cardiac, Peripheral, pediatric, adult cephalic and transesophageal with all required software
for measurements.
OPERATING MODES:
2D tissue, 2D angio flow, color M-Mode, tissue velocity M-mode, tissue strain imaging,
disynchrony imaging, continuous wave Doppler, tissue m-mode, pulse wave Doppler,
47. tissue velocity imaging, tissue tracking, tissue synchronization, blood flow imaging, blood
flow angio flow imaging.
DISPLAY MODES:
Live and stored display format: full size and split screen. Review image format: for still and
cine, simultaneous capability B+PW, B+ CFM/TVI+PW, CW, B+ or triplex mode, , B+ color
split screen display.
Tissue Imaging, 2D mode, , M-mode, color Doppler imaging, color flow imaging, color
Doppler imaging, color angio, color m-mode, blood flow imaging, blood flow angio
imaging, tissue velocity imaging, tissue velocity imaging mode, tissue synchronization
imaging mode, PW / HPRF Doppler, CW Doppler, contrast imaging, LVO Contrast,
Vascular / abdominal contrast, vascular calculations, cardiac measurements.
FRAME RATE
Min. 200fps in B-Mode and 100fps in Doppler mode.
CINE MEMORY
Min. Cine Memory for 1000 frames or 250mb min.
IMAGE VIEWING DEPTH:
20 – 280 mm or more for cardiac application
IMAGING MODES / TECHNIQUES:
Tissue harmonic Imaging, Tissue Doppler Imaging, Color Angio, Tissue Velocity Imaging
Tissue Imaging (Display real time Doppler shift information from moving tissue to better
visualize and quantity myocardial function). Capability to display time difference in
myocardial motion in color for CRT (Cardiac resynchronization therapy.
Quantitative strain rate imaging; An advanced quantitative technique of Tissue
Doppler Velocity. Strain rate is a measure of the contractile motion of myocardium.
Auto-Tracking contrast quantification: quantitative technique for on-line assessment of
contrast agent images.
Contrast plus sequencing technology: a real time, low mechanical index, non-linear
imaging technique for contrast agent examinations. The software should have the capability
to show contrast agent only, tissue only or contrast and tissue displays. Contrast Harmonic
Imaging capability.
Vascular imaging software for carotids with IMT measurement.
STRESS ECHO
Integrated multi stage stress echo system for advance and flexible stress echo Acquisition
and measurement for LV B-Mode imaging.
Quantitative analysis for contrast during stress. Examinations
Used with TDI protocols.
STORAGE DEVICE
Built-in MOD/CD / DVD Drive WITH 10 DISKETTES.
SYSTEM DYNAMIC RANGE
Dynamic range minimum 160 dB or more
COMMUNICATION SOFTWARE
System should conform to DICOM 3 communication software for:
Image Storage, print, Query / Retrieve, Network Communication.
48. STANDARD TRANSDUCERS:
Linear Probe multi frequency to cover frequency of 6.0-8.0 MHz.
Multi frequency Phased array sector probe to cover 2.0/2.5 – 4.0MHz.
Multi frequency Phased array sector probe to cover 5.0 – 8.0MHz.
CW Pencil Probe
Multiplane TEE Transducer (3 – 6 MHz) for adults.(Only Two with all machines)
POWER REQUIREMENT
220-240VAC 50 Hz
Accessories
Digital Color Thermal Printer with 10 Packs of 100..
Online UPS for 10 min. backup time for complete unit including
Printer.(Emerson,Liebert,Chloride,MGE)
Digital B/W Thermal Printer with 50 rolls of papers.
Jelly 20 L in bottles.
FUTURE UPGRADES:
The unit should have the capability for up gradation of future software (shall be provided
free of cost within the warranty period).
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Multiplane TEE Transducer (4 – 6 MHz) for peads(Only One with all machines)
Probes:
Should be light weight, capable of multiple centre frequencies on transmit for 2D, color
Doppler PW/CW (Steerable) Imaging and to perform Harmonics.
PORTS:
Video Output
USB / RS 232
Networking
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 37
Clinical specialty Medical and Lab Equipment
Generic Name Dialysis machine
Clinical purpose Used for acute and chronic kidney failures patient
dialysis
Quantity 01
TECHNICAL SPECFICATION:
Touch Panel Color Display 10.4".
Syringe pump allows simultaneous infusion of heparin
or other prescription drug.
Purage infusion, bolus infusion and programmed
Infusions are possible.
49. Air bubble/Blood detector
Blood flow is stabilized by 2 roller arms.
Blood flow rate is easily adjustable.
Parameters display on monitor
Blood detect sensor
Auto off feature (After disinfection)
Fast preparation
Reduced disinfection time
Display for service diagnostic and calibration.
Detects air bubbles in the venous and blood presence to secure patient
infusion.
Universal Tubing set adult/Peads
Automatic Rinse & Disinfection, Thermal / hot 86oc or more method are
available as standard.
Various rinse program can be set daily & weekly
Ultra filtration program & UF profile are available as standard
Electronic Control dial sate flow & blood flow rate with display.
Variable bicarb acetate concentration & Automatic control bicarbonate
proporation.
Computerized Numerical & Graphical Screen.
Acetate and Bicarbonate Dialysis
Sequential Dialysis
Double Needle
UF, Bicarbonate & Sodium Profile
Advanced accurate KT/V Calculation - dose finder
Start up test
Back-up Battery 30 min.
Volumetric Control
UF rate: 0, 0.10 to 5.00L/hr
Accuracy: ±30g/hr
Dialysis Flow Range (Qd) 300 to 800ml/min
Dialysate Temperature: 32.0°C to 39.0°C
Blood Pump Rate (Qb): 0, 10 to 600 ml/min
Syringe Pump
Delivery Range: 0.0 to 9.9 ml/hr
Ultrasonic Sensor
Venous Pressure Sensor
-500 to 500mmHg
Arterial Pressure Sensor
-500 to 500mmHg
Water Rinse
Chemical Rinse (disinfection)
Acid Rinse
Hot Rinse / Hot Citric disinfection
Auto Rinse programs
Blood Leak Detector
Optical Sensor
Optionals:
Blood pressure monitor.
50. Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 38
Clinical specialty Medical and Lab Equipment
Generic Name Angio jet system
Clinical purpose Used in cath lab for cardiac patients
Quantity 01
Technical specification
LCD Display and Keypad feature Console with Rheolytic
thrombectomy catheter. For breaking apart and removing
thrombus with a feature of delivering thrombolytic agents on sight
by utilizing the Bernoulli principle.
Full Range of Catheters for Coronary, Peripheral and AV Access
Applications
System Self Configures for each catheter model
Catheters are of 4,5 & 6 Fr
Catheter Lengths are of 50 - 140cm can treat vessels from 2 to
20mm Internal Diameter
Dual Lumen Catheter with option of Rapid Exchange or over the
Wire Delivery System
Integrated Pump , Catheter and Collection bag are into a single
unit .
System Preparation time is <2 < omits with Isovoilumetic fluid
delivery and removal.
Catheters generates 700 bar internal pressure with audiovisual
alarm for identification malfunction
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 39
Clinical specialty Medical and Lab Equipment
51. Generic Name Open – Irrigation System
Clinical purpose Use in electrophysiology lab for cardiac patients.
Quantity 01
Technical specification
Intelligent User Interface
• Comprehensive, real-time diagnostic information on one screen
• Large, easy-to-read display can be viewed from a distance
• Quick, intuitive menu navigation.
IntellaView™ Infusion Monitoring
• Provides real-time feedback on five different saline assessments
- Volume Remaining - Low Fluid Warning
- Volume Infused -
New Saline Bag
- Volume Dispensed.
Customizable Memory Buttons
• Programmable flow and menu settings can be quickly accessed for future procedures
• Store and retrieve procedure parameters for up to three different scenarios.
Bubble and Occlusion Detection
• Reliable sensor technology designed to
prevent air infusion and occlusion
• Simple placement process to ensure
accurate tubing alignment
• Smaller tubing designed to effectively
clear bubbles and increase flow.
Automatic Titration
• Intelligent, automatic titration ensures
optimal power-to-fluid control
• Instant, clear display of titration status.
Maximum Operating Back Pressure
35 psi (standby and ablation)
60 psi (purge)
Flow Rate Accuracy
-5%/+15% (6-30mL/min)
-10%/+20% (2-5mL/min)
+/- 20% (60mL/min).
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
52. part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 40
Clinical specialty Medical and Lab Equipment
Generic Name Contamination probe:
Clinical purpose Used in hot lab with survey meter RDS 31.
Quantity 01
Technical specification
radiation detection: gamma-, beta- and alpha-rays
• e e g a ge: ga a > keV, eta a > keV, a d alpha > eV
• dete to t pe: haloge ue hed GM tu e
• App o i ate easu e e t a ge: - 10 000 cps
• e d i do a ti e a ea: . . i ; i do thi k ess: . - 2 mg/cm2
• se siti it : . ps fo u ifo “ / Y sou e of . B /
• et a ta le o d, le gth - 2500 mm (31.50 in - 98.42 in)
• te pe atu e a ge: ope atio : -25°C to +55°C (-13°F to +131°F); storage: -40°C to
+70°C (-40°F to +158°F)
• hu idity: 0 - 95% relative humidity, non-condensing
• pa ake di e sio s: . . . i total p o e le gth
mm (11.6 in
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 41
Clinical specialty Medical and Lab Equipment
Generic Name Syringe Pump digital injector
Clinical purpose Used for patients to inject adenosine
Quantity 01
Technical Specification:
Free flow protection.
Automatic calculation of rate based on does increase in mg, microgram,
mmol, weight and time related e.g microgram /kg/min.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
53. Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 42
Clinical specialty Medical and Lab Equipment
Generic Name Printer for dose calibrator (Hot Lab)
Clinical purpose Used to print reports of dose calibrator.
Quantity 01
Technical specification
Dose Calibrator (VDC-405 Version 3.31, Veenstra Instruments) or
equivalent.
HP LaserJetP3015 or equivalent.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 43
Clinical specialty Medical and Lab Equipment
Generic Name Printer for spectro analyzer system,
Clinical purpose To be used for the analysis of thallium on weekly
basis.
Quantity 01
Technical specification
Spectro AnalyzerTM SpectroWipeTM, Wipe Test Counting system,
1024-Channel MCA with Nuclide Identification, SW1000-ID Version
3.1.0 (AccuSync Medical Research Corporation) or equivalent.
Following mentioned printers are recommended and compatible with
above mentioned equipment
Lexmark E260D, Lexmark E260DN, Brother HL-2270DW, Brother
HL-2270 to be used for the analysis of thallium on weekly basis.
Accessories:
54. Operating manual, Services manual, error code book, part list and software
if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 44
Clinical specialty Medical and Lab Equipment
Generic Name Dose Calibrator
Clinical purpose To measure radioactive medicine.
Quantity 01
Technical specification
Ionization Chamber
• T pe: Thi all, deep well, high pressure
• Di e sio s: deep . dia.
• Ca li g: . ’ i te o e ti g a le
Measurement Range
• T pe: Auto Ra gi g
• A ti it : GB Ci , a .
• Resolutio : . MB . µCi , a .
Display Screen
• T pe: Dot Matrix Liquid Crystal Display
• Fo at: Di e t eadi g i B o Ci
• B /Ci Readi g: Use sele ta le o fi ed
• Values Displa ed: Nu lide a e Ato i s ol, Mass u e , ali atio u e
Electrometer
• A u a : Bette tha ± %
• Li ea it : Withi ± 2%
• Respo se Ti e: Withi se ., to se . fo e lo a ti it sa ples use selectable
average period)
Repeatability of Measurement
• Repeata ilit : Withi ± % ithi hou s, du i g hi h ti e the ali ato is on all the
time.
Molybdenum-99 Assay
• Methods: Ca iste o CAPMAC
• Measu ed Values: Mo-99 elution, Tc-99m, Tc-99m/Mo-99 RatioTests
• Diag osti s: Full test of p og a , s ste e o ies
• Qua te l : Dail , A u a a d Li ea it
• Dail : Auto )e o, Auto Ba kg ou d Adjust, Data Che k, A u acy and Constancy Standard
Source Data
• “ ste Me o : Co-57, Co-60, Ba-133, Cs-137 Standard Sources
Nuclear Data
• Nu lea “etti g Ke s: P e-set, 5 User
55. • “ ste Me o : O e u lides / al u e a d half-life)
• A p e-set key measures up to 2.0 Ci (74.0 GBq) of F-18
• Co pati ilit : “ta da d Nu lea Medi i e Ma age e t “ ste s
Power Requirements
• -240 VAC (50/60Hz) 90 MA
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 45
Clinical specialty Medical and Lab Equipment
Generic Name Co-57 flood source
Clinical purpose Used for daily QC of Gamma camera.
Quantity 01
Technical specification
(Eckert & Ziegler Germany product cat no. CTRF10017). Co-57, 10mCi, Rectangular Flood Src
w/ProKem Tech
Transparent Flood Source.
Active dimensions:410 mm x 260
Overall dimensions: 445 x 375 x 8 mm
CE-marked
Activity tolerance: +/- 15%
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 46
Clinical specialty Medical and Lab Equipment
Generic Name Co-57 point source
Clinical purpose Used for daily QC of Gamma camera.
56. Quantity 01
Technical specification
(Eckert & Ziegler Germany product cat no. 3807) Single-encapsulated stainless steel
point source. ISO rating: C66444. 1mCi = 37MBq. Part#CO738070001M.
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Sr. No 47
Clinical specialty Medical and Lab Equipment
Generic Name Pocket dosimeter
Clinical purpose Used to measure instant radiation of individuals.
Quantity 03
Technical specification
Radiological Characteristics
Radiation detected: - gamma and X-rays
• Dete to s:
- energy compensated si-Diode
• Measu e e t a ge:
dose: μs - 9.99 sv or 0.1 mrem - 999 rem –
dose ate: μs /h - 3 sv/h or 0,5 mrem/h - 300 rem/h
• Cali atio :
- better than ±5% (cs-137, 662 keV at 2 msv/h), Hp(10)
• E e g espo se:
- Hp(10), 55 keV - 3 MeV, better than ±25%, up tp 6 MeV, better than ±35%
Dose rate linearity:
- better than ±15%, up to 3 sv/h (300rem/h)
Functional Characteristics
• Ala th esholds:
- six preset values each for integrated dose and dose rate, manually selectable by push-
button
• F o t pa el push
-button functions:
- toggle between dose and dose rate display
- switch On/OFF
- chirp On/OFF
- reset integrated dose
- change alarm thresholds
- activate battery test
• Audi le ala s:
57. - seven separate alarms, sound level typically better than 85 dBA at 30 cm
- integrated dose, dose rate, dose overflow, dose rate overflow at 3 sv/h or 300 rem/h
- low battery 1 and 2
- defect
Power Requirement
-100-240VDC
Reader communication:
- by infrared through bottom part; by using ADR-1 Reader Head in combination with RADOs
Pc software
Reader and software for dosimeter configuration calibration or access control (Qty 1)
Accessories:
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
Radiology
Sr. No 48
Clinical specialty Medical and Lab Equipment
Generic Name Workstation for CT
Clinical purpose Used to warm and shake blood / plasma.
Quantity 02
Technical specification
FDA / CE approved, Independent, multimodality, fully Functional.
All the companies will give undertaking that the offered workstation has been supplied by
them worldwide.
High speed link to operator console on DICOM network.
System architecture and operating system:
a: Dual processor Xeon.
b: 2.66 GHz or more.
c: 512 Cache or more.
d: Graphic card & Network card.
Licensed Software’s: latest Windows systems or Linux; MS Office, Norton antivirus (current
and upgradeable).
Should have one high resolution TFT monitor of 18 inches or more.
DVD writer / CDR.
DICOM -3 Viewer with universal PC display capability (Licensed).
Laser black & white / Color printer, A4 size, 2400 dpi or higher, Two paper trays for A4 /
letter size media (HP, Lexmark, Xerox) Network ready.
UPS, 3 kV (x 03) (Branded, Dry batteries) .capable of providing 20 minutes of backup for
workstations (MG, APC, MC, Chloride).
58. Workstation Software
a: All the latest available cardiac and whole body software will be provided by the
vendor as standard, including reporting packages.
b: Software up gradation of all existing applications for at least next 5 years.
c: Following software should be provided at ALL workstations:
3-D reconstruction display.
a: Maximum and minimum intensity projections.
b: Multi-planner and curved planer reconstruction.
c: 3-D Shaded Surface Display.
d: 3-D Volume Rendering software.
e: 3-D Virtual endoscopy, colonoscopy, bronchoscopy.
CT Angiography
Advanced coronary vessel analysis.
Calcium scoring with ECG gating and prospective & retrospective reconstruction.
Cardiac function analysis, complete package.
Advanced Pheripheral / General vessel analysis.
Lung nodule detection & analysis.
Brain Perfusion analysis.
Fusion software for correlative imaging.
DICOM 3.0
DICOM 3 ready (multi-vendor and multimodality compatible) for Send, Receive, Archive,
Retrieve and Print, on main consol and workstations.
UPGRADEABILITY
a: All vendors will quote their latest systems.
b: The System should have an upgrade route to higher version.
QUALITY AND SAFETY STANDARDS
a: MDD (CE) compliance.
b: FDA 510 K approval.
POWER REQUIREMENT
Three-phase with line voltage of 220 V, 50 Hz.
Fluoroscopy with real time imaging and display of at least 8 frames / sec with required
hardware & software. One high resolution in-room TFT monitor of at least 15 inches or more
on mobile base / ceiling mounted.
Accessories
Operating manual, Services manual, error code book, part list and software
if any.
Optional
Nil
Warranty
Two year with all spare parts, during warranty period firm should maintain
equipments and done PPM as per principle recommendation. If any spare
part required during installation or PPM will be responsibility of the firm.
Firm must send annual PPM schedule with installation report.
59. Sr. No 49
Clinical specialty Medical and Lab Equipment
Generic Name Portable ultra sound
Clinical purpose Used to warm and shake blood / plasma.
Quantity 02
Technical specification
A complete dedicated high end digital Color Doppler Ultrasound unit for wide range of
premium performance application of Abdominal, OBS/GYN, Urology, and vascular
imaging in pediatrics and adult. Mobile Portable system with built in workstation / data
management system for digital acquisition, storage and review of complete ultrasound
studies including static and dynamic clips in DICOM format, read/write zoom. Studies can
be reviewed and output to CD / DVD/USB. System Should have built-in battery. The
machine must have sharp and high quality image reproduction with heavy duty
performance. It should have minimum following specification :
DISPLAY:
High resolution non interlaced, flicker free 15” LCD/TFT display tilt able.
OPERATING MODES:
B, 2D M-Mode, Power Doppler, HPRF, Spectral Doppler, Color Doppler, Pw Doppler,
Duplex And Triplex Doppler, IMT imaging, Contrast Imaging, Panoramic Imaging,
Compound Imaging, 3D/4D imaging.
CONTROL PANEL:
Alphanumeric keyboard with built-in trackball.
Direct access to system functions through dedicated keys.
Indicator lights identify activated keys.
Audio volume control with bidirectional / stereo speakers and foot switch.
User selectable image magnification control.
Adjustable transmit focusing control.
Total and Lateral Gain Compensation controls (6 or more).
2 Active Probe connectors
CALIPER / MEASUREMENTS :
4 to 8 calipers for measurement per screen trace length measurements for:
Distance, angle, distance depth from skin line, area, circumferences, compound / volume,
OBS/GYN measurements, Urological measurements, Vascular Measurements with IMT.
APPLICATION:
Abdominal, OBS/GYN, Urology, vascular, Small Parts, Peripheral for pediatric, adult with
all required software for measurements.
DISPLAY MODES:
Live and stored display format: full size and split screen. Review image format: for still and
cine, simultaneous capability B+PW, B+ CFM +PW, B+ or triplex mode, B+ color split screen
display. 2D mode, , M-mode, color Doppler imaging, color Doppler imaging, color m-mode,
PW / HPRF Doppler, IMT, Panoramic Imaging, Vascular imaging , vascular calculations.
CINE MEMORY
Min. Cine Memory for 8000 frames or 250mb min.
IMAGE VIEWING DEPTH: