In the Field or in the Shop: Overcoming
the Challenges in Pressure Calibration
Presenter: Tim Francis, Fluke Calibration
Principles and practical tips about electrical, flow, pressure,
RF and temperature calibration
Agenda
• The Vocabulary – Definition of some basic terms
– Reference Mode – Gauge versus absolute versus
differential
– Pressure Ranges – How high is high and how low is
low?
• Challenges and how to overcome them
– Safety Considerations
– Environmental Impact
– Contamination
– The right process and equipment for the job
• Question and Answer
©2016 Fluke Corporation.
Agenda
• The Vocabulary – Definition of some basic terms
– Reference Mode – Gauge versus absolute versus
differential
– Pressure Ranges – How high is high and how low is
low?
• Challenges and how to overcome them
– Safety Considerations
– Environmental Impact
– Contamination
– The right process and equipment for the job
• Question and Answer
©2016 Fluke Corporation.
Reference Modes
• Pressure values are sometimes followed by the words
“gauge” or “absolute” (or possibly “psig” or “psia”)
• What do these words mean and how do they affect the
calibration process?
• Measurements are relative
– You don’t say, “I live 5 miles.” You say, “I live 5 miles from
here” or “I live “5 miles from my work.” Everything has a
starting point.
– Pressure is no different. Saying “The pressure is 5 psi”
doesn’t say the whole story. More proper to say “The
pressure is 5 psi above atmosphere.”
– Gauge, absolute, and differential are simply shorthand for
this.
©2016 Fluke Corporation.
Gauge Mode
• Referenced to atmospheric
pressure
– 0 means that your test
pressure is the same as
atmosphere
– Most common reference mode
– Can be both positive and
negative (partial vacuum)
– Usually the easiest to measure
• Examples
– Tire Pressure
– Many Process Measurements
Gauge (+ or -)
Barometric
Zero pressure, no molecules
©2016 Fluke Corporation.
Absolute Mode
• Referenced to a perfect vacuum
– 0 means that it is a perfect
vacuum (absolutely no
molecules)
– Can only be positive numbers –
You can’t have negative
molecules in a vessel
– More difficult measurement –
How do you re-zero the
reference?
• Examples
– Barometric pressure
measurement
– Airplane altitude measurement
– Downhole tools
Gauge (+ or -)
Absolute
Barometric
Zero pressure, no molecules
©2016 Fluke Corporation.
Differential Mode
• Referenced to another pressure (line
pressure)
– 0 means that there is no difference in the
two pressures
– Gauge mode can be thought of as a
special case of differential mode.
– Can be a difficult measurement to make.
Few calibrators support high line pressure.
– Can be both positive or negative (above or
below line pressure)
• Examples
– Flow measurements in a pipeline
– Draft Range measurements – difference in
pressure between two rooms
Gauge (+ or -)
Absolute
Differential
Barometric
Zero pressure, no molecules
©2016 Fluke Corporation.
How high is high and low is low?
• Pressure measurement covers a wide range, from
vacuum to > 60,000 psi
– Vacuum – Less than atmosphere
– Draft Range – low differential/gauge pressures (< 2 psi)
– Low Pressure – Normally pneumatic pressures (<= 300 psi)
– Medium Pressure – Could be pneumatic or hydraulic (<=
2000 psi)
– High Pressure – Almost always hydraulic (2000 psi to
20,000 psi)
– Very High Pressure - > 20,000 psi
• General statements – no official rules
©2016 Fluke Corporation.
Challenges created by Reference Mode
• I only have a gauge standard. Can I use it to
calibrate an absolute device?
• The spec sheet says that the range is -15 to 30
psi, but I can’t get down to -15. Why not?
• Is it okay to calibrate a differential device as
though it was a gauge mode device?
• What is sealed gauge and when can I use it?
©2016 Fluke Corporation.
Using a gauge standard to measure absolute mode
• There will be an offset equal to barometric pressure
(~14.7 psi)
• The offset will change with barometric pressure – if
there is a storm that causes barometric pressure to
drop 0.4 psi, then your error will change by 0.4 psi.
• If working with high pressures, you may be able to
estimate barometric pressure and add it to your
standard reading.
• If working with lower pressures, you will need to
measure barometric pressure and continuously adjust
your standard reading to compensate.
©2016 Fluke Corporation.
-15 psi isn’t reachable
• At barometric pressure, 0 psi gauge = ~14.7 psi
absolute.
• -15 psi gauge = -0.3 psi absolute – NOT POSSIBLE
• The lowest you can possibly go is -1 x barometric
pressure
– Denver, Colorado = ~12.1 psi absolute
– Honolulu, Hawaii (at the beach) = ~14.7 psi absolute
– Mauna Kea, Hawaii = ~8.7 psi absolute
• Even at sea level, it may not be possible to get all
the way to -14.7 psi – you have to get all the
molecules out.
©2016 Fluke Corporation.
Calibrating High Line Differential
• Example: Pressure transmitter is used on a
pipeline that has a pressure of ~1500 psi
absolute. It’s measuring the drop in pressure
across an orifice, ~30 psi.
• To calibrate it like it is used, we need to apply
1500 psi to one side of the transmitter and 1530
psi to the other side.
• Difficult to do, so we normally leave one side
open to atmosphere and apply 30 psi to the other
side.
• Is this acceptable? Depends on who you ask.
©2016 Fluke Corporation.
Sealed Gauge
• Special case that is a hybrid of gauge mode and
absolute mode
• If the test port is open to atmosphere, it reads 0 (like
gauge mode)
• If test port is open to atmosphere and barometric
pressure changes, then the measurement will change
with it (like absolute mode)
• Similar to using an absolute mode device and
subtracting a default barometric pressure value from it.
• Useful for high pressure applications where variations
in barometric pressure
©2016 Fluke Corporation.
Agenda
• The Vocabulary – Definition of some basic terms
– Reference Mode – Gauge versus absolute versus
differential
– Pressure Ranges – How high is high and how low is
low?
• Challenges and how to overcome them
– Safety Considerations
– Environmental Impact
– Contamination
– The right process and equipment for the job
• Question and Answer
©2016 Fluke Corporation.
Safety Considerations
• A pressurized vessel has some inherent safety risks
• Make sure all equipment is rated to the proper
pressures for the work being performed.
– Some shops only stock pressure lines and fittings that are
rated to the highest pressure they work to. This works if
the highest pressure is relatively low. Doesn’t work if the
highest pressure is 10,000 psi or higher.
• Use the proper Personal Protection Equipment (PPE)
• When working in the field, be cognizant of other
dangers around you.
©2016 Fluke Corporation.
Impacts from the Environment
• A pressure calibration can be impacted by the environment around it.
• Temperature – Almost everything that measures pressure is also
impacted by temperature. Pressure Calibrator manufacturers will
characterize and compensate, but they still perform best over a given
temperature range.
• Gravity – Acceleration of gravity varies by location. A traditional
deadweight tester is impacted by this.
• Head height – A large difference in elevation between the reference
and the device under test will create an error
– 3 feet in gas = ~0.01% reading
– 3 feet in oil = ~ 10 psi
• Large drafts, temperature swings, or changes in barometric pressure
can make it difficult to get pressure to stabilize
©2016 Fluke Corporation.
Contamination
• Dirt, debris, and liquids from the
Device Under Test can cause problems:
– Damage your calibration equipment
– Cross contaminate other Devices Under
Test
• How do we combat it?
– Clean the devices before we connect to
them
– Make use of filters, traps, or separators
In the field or in the lab?
• Need to consider a few questions
– Can the work be done safely in the field?
– Does the environment allow for the calibration to
be done properly?
• Is the temperature stable and within the allowed for
operating range of the standard?
• If you are using a deadweight tester, do you know the
local gravity sufficiently?
• Can you easily connect to the device? Or is it better to
calibrate a similar device in the shop and rotate it into
the field?
©2016 Fluke Corporation.
The Right Equipment
• Portable versus Bench
– Portable type instruments, like the Fluke
718 can be used on the bench, but…
– They are designed for the field, so they
aren’t always ergonomic for the bench.
©2016 Fluke Corporation.
• Make sure you have devices that cover
the proper range and mode
– Add pressure modules for absolute mode
– Not always appropriate to calibrate low
pressure with a higher range instrument
– Example: Fluke 3130 with 750P Pressure
Modules
Conclusion
Pressure calibrations can be challenging, but
these challenges can be overcome with the right
processes and the right equipment.
©2016 Fluke Corporation.
Special limited-time offer!
• Gift with purchase – up to $1200 value
• +Go to www.transcat.com/deals for more
details
Questions or Comments?
Email the moderator, Nicole VanWert:
nvanwert@Transcat.com
Transcat: 800-828-1470
www.Transcat.com
For related product information, go to:
www.Transcat.com/Fluke

More Related Content

What's hot

How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2Transcat
 
Practical Power Quality Measurements and Fixes Presented by Fluke and Transcat
Practical Power Quality Measurements and Fixes Presented by Fluke and TranscatPractical Power Quality Measurements and Fixes Presented by Fluke and Transcat
Practical Power Quality Measurements and Fixes Presented by Fluke and TranscatTranscat
 
A.introduction tech.
A.introduction   tech.A.introduction   tech.
A.introduction tech.girlie22
 
How to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment WebinarHow to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment WebinarTranscat
 
Transcat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure CalibrationTranscat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure CalibrationTranscat
 
Pressure Measurement
Pressure MeasurementPressure Measurement
Pressure MeasurementLiving Online
 
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Transcat
 
Instrumentation Interview Questions Download
Instrumentation Interview Questions DownloadInstrumentation Interview Questions Download
Instrumentation Interview Questions DownloadInstrumentation Tools
 
Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationFluke Calibration
 
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Transcat
 
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...Transcat
 
Basic concepts of instrumentation
Basic concepts of instrumentationBasic concepts of instrumentation
Basic concepts of instrumentationG. Udhaya Sankar
 
Instrument lesson
Instrument lessonInstrument lesson
Instrument lessonHung Thai
 
Technical interview-questions
Technical interview-questionsTechnical interview-questions
Technical interview-questionsmojahid zaighum
 
Basic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingBasic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingTheunis Venter
 
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...Transcat
 
Pressure Overview Of Measurement And Calibration
Pressure Overview Of Measurement And CalibrationPressure Overview Of Measurement And Calibration
Pressure Overview Of Measurement And CalibrationAsep Herman
 
Yokogawa EJX110A Differential Pressure Tranismitter
Yokogawa EJX110A Differential Pressure TranismitterYokogawa EJX110A Differential Pressure Tranismitter
Yokogawa EJX110A Differential Pressure TranismitterPower Specialties, Inc.
 
Temperature calibration
Temperature calibrationTemperature calibration
Temperature calibrationChetan Yadav
 
Basics Of Instrumentation
Basics Of InstrumentationBasics Of Instrumentation
Basics Of InstrumentationVinoth Ganesh
 

What's hot (20)

How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2
 
Practical Power Quality Measurements and Fixes Presented by Fluke and Transcat
Practical Power Quality Measurements and Fixes Presented by Fluke and TranscatPractical Power Quality Measurements and Fixes Presented by Fluke and Transcat
Practical Power Quality Measurements and Fixes Presented by Fluke and Transcat
 
A.introduction tech.
A.introduction   tech.A.introduction   tech.
A.introduction tech.
 
How to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment WebinarHow to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment Webinar
 
Transcat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure CalibrationTranscat and Fluke Calibration Webinar: Automating Pressure Calibration
Transcat and Fluke Calibration Webinar: Automating Pressure Calibration
 
Pressure Measurement
Pressure MeasurementPressure Measurement
Pressure Measurement
 
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
 
Instrumentation Interview Questions Download
Instrumentation Interview Questions DownloadInstrumentation Interview Questions Download
Instrumentation Interview Questions Download
 
Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke Calibration
 
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
 
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
 
Basic concepts of instrumentation
Basic concepts of instrumentationBasic concepts of instrumentation
Basic concepts of instrumentation
 
Instrument lesson
Instrument lessonInstrument lesson
Instrument lesson
 
Technical interview-questions
Technical interview-questionsTechnical interview-questions
Technical interview-questions
 
Basic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingBasic to Site Specific Instrument Training
Basic to Site Specific Instrument Training
 
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
 
Pressure Overview Of Measurement And Calibration
Pressure Overview Of Measurement And CalibrationPressure Overview Of Measurement And Calibration
Pressure Overview Of Measurement And Calibration
 
Yokogawa EJX110A Differential Pressure Tranismitter
Yokogawa EJX110A Differential Pressure TranismitterYokogawa EJX110A Differential Pressure Tranismitter
Yokogawa EJX110A Differential Pressure Tranismitter
 
Temperature calibration
Temperature calibrationTemperature calibration
Temperature calibration
 
Basics Of Instrumentation
Basics Of InstrumentationBasics Of Instrumentation
Basics Of Instrumentation
 

Similar to Overcoming Challenges in Pressure Calibration Webinar

Basic Instrument Presentation ( pressure )
Basic Instrument Presentation ( pressure  )Basic Instrument Presentation ( pressure  )
Basic Instrument Presentation ( pressure )Aziz u rehman Jamali
 
Pressure Measurement Instruments 1.pptx
Pressure Measurement Instruments 1.pptxPressure Measurement Instruments 1.pptx
Pressure Measurement Instruments 1.pptxHanif61
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxsanjayudps2016
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationTranscat
 
Fulham Reach Block G Risk Review Presentation For review
Fulham Reach Block G Risk Review Presentation For reviewFulham Reach Block G Risk Review Presentation For review
Fulham Reach Block G Risk Review Presentation For reviewMark Platten
 
psv calculation ppt jhjghgh jghjgh jhjgj
psv calculation ppt jhjghgh jghjgh jhjgjpsv calculation ppt jhjghgh jghjgh jhjgj
psv calculation ppt jhjghgh jghjgh jhjgjHashemAli23
 
psv and pvrvdesign calculation.ppt
psv and pvrvdesign calculation.pptpsv and pvrvdesign calculation.ppt
psv and pvrvdesign calculation.pptssuser73bb74
 
psv-calculation_psv-calculation_psv-calc
psv-calculation_psv-calculation_psv-calcpsv-calculation_psv-calculation_psv-calc
psv-calculation_psv-calculation_psv-calcAmit Chopra
 
Control theory basics for process control engineers
Control theory basics for process control engineersControl theory basics for process control engineers
Control theory basics for process control engineerssarkmank1
 
Pressure Thermometer
Pressure ThermometerPressure Thermometer
Pressure ThermometerDeepak Kumar
 
Pressure measurement id Mitesh Kuamr
Pressure measurement id Mitesh KuamrPressure measurement id Mitesh Kuamr
Pressure measurement id Mitesh KuamrMitesh Kumar
 
Labthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink
 
Testo 312 2 3-4 datasheet 2012
Testo 312 2 3-4 datasheet 2012Testo 312 2 3-4 datasheet 2012
Testo 312 2 3-4 datasheet 2012Testo Limited
 
Pressure measurement
Pressure measurementPressure measurement
Pressure measurementBRIJ10PATEL
 

Similar to Overcoming Challenges in Pressure Calibration Webinar (20)

Pressure measurement
Pressure measurement Pressure measurement
Pressure measurement
 
Basic Instrument Presentation ( pressure )
Basic Instrument Presentation ( pressure  )Basic Instrument Presentation ( pressure  )
Basic Instrument Presentation ( pressure )
 
Pressure Measurement Instruments 1.pptx
Pressure Measurement Instruments 1.pptxPressure Measurement Instruments 1.pptx
Pressure Measurement Instruments 1.pptx
 
CALIBRATIONS.pptx
CALIBRATIONS.pptxCALIBRATIONS.pptx
CALIBRATIONS.pptx
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptx
 
Module 3.pdf
Module 3.pdfModule 3.pdf
Module 3.pdf
 
Instrumentation and control
Instrumentation and controlInstrumentation and control
Instrumentation and control
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer Calibration
 
Fulham Reach Block G Risk Review Presentation For review
Fulham Reach Block G Risk Review Presentation For reviewFulham Reach Block G Risk Review Presentation For review
Fulham Reach Block G Risk Review Presentation For review
 
psv calculation ppt jhjghgh jghjgh jhjgj
psv calculation ppt jhjghgh jghjgh jhjgjpsv calculation ppt jhjghgh jghjgh jhjgj
psv calculation ppt jhjghgh jghjgh jhjgj
 
psv and pvrvdesign calculation.ppt
psv and pvrvdesign calculation.pptpsv and pvrvdesign calculation.ppt
psv and pvrvdesign calculation.ppt
 
psv-calculation_psv-calculation_psv-calc
psv-calculation_psv-calculation_psv-calcpsv-calculation_psv-calculation_psv-calc
psv-calculation_psv-calculation_psv-calc
 
Control theory basics for process control engineers
Control theory basics for process control engineersControl theory basics for process control engineers
Control theory basics for process control engineers
 
Leak Tests.pptx
Leak Tests.pptxLeak Tests.pptx
Leak Tests.pptx
 
Pressure Thermometer
Pressure ThermometerPressure Thermometer
Pressure Thermometer
 
Presentation10-9.ppt
Presentation10-9.pptPresentation10-9.ppt
Presentation10-9.ppt
 
Pressure measurement id Mitesh Kuamr
Pressure measurement id Mitesh KuamrPressure measurement id Mitesh Kuamr
Pressure measurement id Mitesh Kuamr
 
Labthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink Gas Permeation Tester
Labthink Gas Permeation Tester
 
Testo 312 2 3-4 datasheet 2012
Testo 312 2 3-4 datasheet 2012Testo 312 2 3-4 datasheet 2012
Testo 312 2 3-4 datasheet 2012
 
Pressure measurement
Pressure measurementPressure measurement
Pressure measurement
 

Recently uploaded

Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSRajkumarAkumalla
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...RajaP95
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingrakeshbaidya232001
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024Mark Billinghurst
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝soniya singh
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineeringmalavadedarshan25
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZTE
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 

Recently uploaded (20)

Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writing
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineering
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 

Overcoming Challenges in Pressure Calibration Webinar

  • 1. In the Field or in the Shop: Overcoming the Challenges in Pressure Calibration Presenter: Tim Francis, Fluke Calibration Principles and practical tips about electrical, flow, pressure, RF and temperature calibration
  • 2. Agenda • The Vocabulary – Definition of some basic terms – Reference Mode – Gauge versus absolute versus differential – Pressure Ranges – How high is high and how low is low? • Challenges and how to overcome them – Safety Considerations – Environmental Impact – Contamination – The right process and equipment for the job • Question and Answer ©2016 Fluke Corporation.
  • 3. Agenda • The Vocabulary – Definition of some basic terms – Reference Mode – Gauge versus absolute versus differential – Pressure Ranges – How high is high and how low is low? • Challenges and how to overcome them – Safety Considerations – Environmental Impact – Contamination – The right process and equipment for the job • Question and Answer ©2016 Fluke Corporation.
  • 4. Reference Modes • Pressure values are sometimes followed by the words “gauge” or “absolute” (or possibly “psig” or “psia”) • What do these words mean and how do they affect the calibration process? • Measurements are relative – You don’t say, “I live 5 miles.” You say, “I live 5 miles from here” or “I live “5 miles from my work.” Everything has a starting point. – Pressure is no different. Saying “The pressure is 5 psi” doesn’t say the whole story. More proper to say “The pressure is 5 psi above atmosphere.” – Gauge, absolute, and differential are simply shorthand for this. ©2016 Fluke Corporation.
  • 5. Gauge Mode • Referenced to atmospheric pressure – 0 means that your test pressure is the same as atmosphere – Most common reference mode – Can be both positive and negative (partial vacuum) – Usually the easiest to measure • Examples – Tire Pressure – Many Process Measurements Gauge (+ or -) Barometric Zero pressure, no molecules ©2016 Fluke Corporation.
  • 6. Absolute Mode • Referenced to a perfect vacuum – 0 means that it is a perfect vacuum (absolutely no molecules) – Can only be positive numbers – You can’t have negative molecules in a vessel – More difficult measurement – How do you re-zero the reference? • Examples – Barometric pressure measurement – Airplane altitude measurement – Downhole tools Gauge (+ or -) Absolute Barometric Zero pressure, no molecules ©2016 Fluke Corporation.
  • 7. Differential Mode • Referenced to another pressure (line pressure) – 0 means that there is no difference in the two pressures – Gauge mode can be thought of as a special case of differential mode. – Can be a difficult measurement to make. Few calibrators support high line pressure. – Can be both positive or negative (above or below line pressure) • Examples – Flow measurements in a pipeline – Draft Range measurements – difference in pressure between two rooms Gauge (+ or -) Absolute Differential Barometric Zero pressure, no molecules ©2016 Fluke Corporation.
  • 8. How high is high and low is low? • Pressure measurement covers a wide range, from vacuum to > 60,000 psi – Vacuum – Less than atmosphere – Draft Range – low differential/gauge pressures (< 2 psi) – Low Pressure – Normally pneumatic pressures (<= 300 psi) – Medium Pressure – Could be pneumatic or hydraulic (<= 2000 psi) – High Pressure – Almost always hydraulic (2000 psi to 20,000 psi) – Very High Pressure - > 20,000 psi • General statements – no official rules ©2016 Fluke Corporation.
  • 9. Challenges created by Reference Mode • I only have a gauge standard. Can I use it to calibrate an absolute device? • The spec sheet says that the range is -15 to 30 psi, but I can’t get down to -15. Why not? • Is it okay to calibrate a differential device as though it was a gauge mode device? • What is sealed gauge and when can I use it? ©2016 Fluke Corporation.
  • 10. Using a gauge standard to measure absolute mode • There will be an offset equal to barometric pressure (~14.7 psi) • The offset will change with barometric pressure – if there is a storm that causes barometric pressure to drop 0.4 psi, then your error will change by 0.4 psi. • If working with high pressures, you may be able to estimate barometric pressure and add it to your standard reading. • If working with lower pressures, you will need to measure barometric pressure and continuously adjust your standard reading to compensate. ©2016 Fluke Corporation.
  • 11. -15 psi isn’t reachable • At barometric pressure, 0 psi gauge = ~14.7 psi absolute. • -15 psi gauge = -0.3 psi absolute – NOT POSSIBLE • The lowest you can possibly go is -1 x barometric pressure – Denver, Colorado = ~12.1 psi absolute – Honolulu, Hawaii (at the beach) = ~14.7 psi absolute – Mauna Kea, Hawaii = ~8.7 psi absolute • Even at sea level, it may not be possible to get all the way to -14.7 psi – you have to get all the molecules out. ©2016 Fluke Corporation.
  • 12. Calibrating High Line Differential • Example: Pressure transmitter is used on a pipeline that has a pressure of ~1500 psi absolute. It’s measuring the drop in pressure across an orifice, ~30 psi. • To calibrate it like it is used, we need to apply 1500 psi to one side of the transmitter and 1530 psi to the other side. • Difficult to do, so we normally leave one side open to atmosphere and apply 30 psi to the other side. • Is this acceptable? Depends on who you ask. ©2016 Fluke Corporation.
  • 13. Sealed Gauge • Special case that is a hybrid of gauge mode and absolute mode • If the test port is open to atmosphere, it reads 0 (like gauge mode) • If test port is open to atmosphere and barometric pressure changes, then the measurement will change with it (like absolute mode) • Similar to using an absolute mode device and subtracting a default barometric pressure value from it. • Useful for high pressure applications where variations in barometric pressure ©2016 Fluke Corporation.
  • 14. Agenda • The Vocabulary – Definition of some basic terms – Reference Mode – Gauge versus absolute versus differential – Pressure Ranges – How high is high and how low is low? • Challenges and how to overcome them – Safety Considerations – Environmental Impact – Contamination – The right process and equipment for the job • Question and Answer ©2016 Fluke Corporation.
  • 15. Safety Considerations • A pressurized vessel has some inherent safety risks • Make sure all equipment is rated to the proper pressures for the work being performed. – Some shops only stock pressure lines and fittings that are rated to the highest pressure they work to. This works if the highest pressure is relatively low. Doesn’t work if the highest pressure is 10,000 psi or higher. • Use the proper Personal Protection Equipment (PPE) • When working in the field, be cognizant of other dangers around you. ©2016 Fluke Corporation.
  • 16. Impacts from the Environment • A pressure calibration can be impacted by the environment around it. • Temperature – Almost everything that measures pressure is also impacted by temperature. Pressure Calibrator manufacturers will characterize and compensate, but they still perform best over a given temperature range. • Gravity – Acceleration of gravity varies by location. A traditional deadweight tester is impacted by this. • Head height – A large difference in elevation between the reference and the device under test will create an error – 3 feet in gas = ~0.01% reading – 3 feet in oil = ~ 10 psi • Large drafts, temperature swings, or changes in barometric pressure can make it difficult to get pressure to stabilize ©2016 Fluke Corporation.
  • 17. Contamination • Dirt, debris, and liquids from the Device Under Test can cause problems: – Damage your calibration equipment – Cross contaminate other Devices Under Test • How do we combat it? – Clean the devices before we connect to them – Make use of filters, traps, or separators
  • 18. In the field or in the lab? • Need to consider a few questions – Can the work be done safely in the field? – Does the environment allow for the calibration to be done properly? • Is the temperature stable and within the allowed for operating range of the standard? • If you are using a deadweight tester, do you know the local gravity sufficiently? • Can you easily connect to the device? Or is it better to calibrate a similar device in the shop and rotate it into the field? ©2016 Fluke Corporation.
  • 19. The Right Equipment • Portable versus Bench – Portable type instruments, like the Fluke 718 can be used on the bench, but… – They are designed for the field, so they aren’t always ergonomic for the bench. ©2016 Fluke Corporation. • Make sure you have devices that cover the proper range and mode – Add pressure modules for absolute mode – Not always appropriate to calibrate low pressure with a higher range instrument – Example: Fluke 3130 with 750P Pressure Modules
  • 20. Conclusion Pressure calibrations can be challenging, but these challenges can be overcome with the right processes and the right equipment. ©2016 Fluke Corporation.
  • 21. Special limited-time offer! • Gift with purchase – up to $1200 value • +Go to www.transcat.com/deals for more details
  • 22. Questions or Comments? Email the moderator, Nicole VanWert: nvanwert@Transcat.com Transcat: 800-828-1470 www.Transcat.com For related product information, go to: www.Transcat.com/Fluke