FPGA BASED HEARTBEATS MONITOR WITH
FINGERTIP OPTICAL SENSOR
Wahyu Kusuma R.1
, Ridha I.2
, Yasman Rianto3
, Swelandiah E.P4
1,3
Departement of Electrcal Engineering, Gunadarma University, Jakarta, Indonesia
2,4
Departement of Computer Science, Gunadarma University, , Jakarta, Indonesia
ABSTRACT
The heart is an organ of human body which has a vital function, small abnormalities can have a big impact
on the performance of the body. Heart disease is the number one cause of death in the world. Examination
of the heart can be detected from blood flow in the fingertips, in order to obtain information about the
number and rhythm of the heartbeat. This research aims to design and implement the FPGA board to
monitor the heart rate with optical sensors.The results of this study are expected to facilitate the patient's
medical team or independently in detecting heart health. The series is composed of blocks of sensors,
signal conditioning block, the block pulse counter, block timer 10 seconds and blocks the viewer. Based on
the test results of the 10 respondents with a variety of age and gender, has built a tool that the percentage
error of 3.94%.
KEYWORDS
Heartbeat monitor, Xilinx ISE Webpack, FPGA, optical sensor
1. INTRODUCTION
The heart is an organ of human body which has a vital function, small abnormalities can have a
big impact on the performance of the heart kita.Penyakit body is the number one cause of death
in the world. Based on data from the World Health Organization (WHO), cardiovascular disease
has reached 29% in the percentage of deaths in the world and 17 million people die every year
due to heart and blood vessel disease throughout the world [3].
The development of medical instrumentation systems is growing rapidly along with the need for
medical personnel to diagnose a patient and a medical examination. One medical instrumentation
used for the examination of the heart is Electrocardiograph (ECG). ECG is a medical instrument
that is commonly used by the medical team to detect heart rate and rhythm [6]. EKG can not be
used independently by patients to detect a patient's pulse. In addition to the expensive costs for
the procurement of ECG, ECG devices also require special skills to operate.
Along with the requirement in the design and manufacture of medical devices, digital electronics
design technology is developing very rapidly, both in terms of hardware and software. Xilinx is
one manufacturer that produces equipment or tools for modeling the design of digital systems.
One product is in the form of a kit module board FPGA (Field Programmable Gate Array).
FPGA is a programmable device that is composed of large modules independent logic that can be
configured by the user who is connecting through the canals of programmable routing [4].
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
2
Field Programmable Gate Array (FPGA) has advantages of which can be configured by the End
User, does not require the fabrication process, and the available solutions that support
customized VLSI chip, so the logic circuit capable of implementing, instant manufacturring,
very-low cost prototype. While FPGA product is also supported by the availability of the Xilinx
ISE WebPack permograman language that is used as a GUI for the digital system design
problems is desired.
Based on the above issues, this study will design a device that can detect heart rate by detecting
the frequency of blood flow to the fingers automatically processed electronically using the
module FPGA (Field Programable Gate Array). This research is expected to generate a heartbeat
detector chip implementation can provide facilities for the medical and can be used
independently by the user (patient) without the help of a doctor or paramedic.
2. ARCHITECTURE AND ITS IMPLEMENTATION
This research conducted several phases of hardware design, software design pelangkat, testing
tools, and analysis to draw conclusions.
2.1 Hardware Design
The design of this tool consists of several circuit blocks are arranged into a single system
heartbeats monitoring based on optical sensors, as shown in Figure 1.
Figure 1. Block Diagram Tool Heartbeats Detection Based on Optical Sensors
2.1.1. Optical Sensor Unit
The sensor consists of a transmitter that emits infrared light (IR) and IR photo detector
(photodiode) to act as a receiver. Construction sensor block shown in Figure 2. The use of optical
sensors for process monitoring heart rate with fingertip illustrated as a figure 3.
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
3
Figure 2. Optical sensor circuit Figure 3. Ilustration fingertip to optical sensor
2.1.2. Conditional Signal Unit
The output signal of the sensor is very small and still a lot of noise. This signal before entering
the pulse counter block should be done in conditioning, so signal that result this block is ready to
be read or processed by the block counter. The signal conditioner block consists of a series of low
pass filter and amplifier structure is shown as figure 4.
Figure 4. Diagram Block of Conditional Signal Unit
Figure 5. Optical Sensor and Conditional Signal Circuits
2.1.3. Heart Counter Unit
This block is used to calculate the output signal generated by the signal conditioning block.
Signals that represent a high rate, while the low signal represents the absence of heartbeat. So this
block will calculate the amount of high signal.
The basic principle of this block as a series of counters. This block has three inputs namely are
Clk signal, the signal In is connected to the output signal conditioning block and Enable signal
(En) which is connected to the output signal at the timer block. The output of the pulse counter
block form as 6-bit binary sequence (D5, D4, D3, D2, D1, D0). In this research, the pulse
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
4
counter block is designed with programming languages Xlinx WebPack ISE. The design of the
pulse counter block is shown as figure 6.
Figurer 6. Design of Heart Counter Unit
2.1.4. Timer 10s Unit
This block is designed for pulse counting process is done by calculating the input signal to the
timer block rate for 10 seconds. The basic principle of the timer unit is a clock signal frequency
divider circuit. The clock signal is taken from the internal oscillator signal frequency of 50 MHz
on the FPGA module. In order to get a signal with a period of 20 seconds, then beat signal as a
function of the timer must have a frequency of 50 mHz, by taking half the period of the high
level. The output signal timer 10s unit to be input to the Enable signal (En) at the pulse counter
block. The timer 10s unit is shown in Figure 7.
Figure 7. Diagram Block of Timer 10s
2.1.5. Display Unit
The Display unit that is used is the 6 Led (D5, D4, D3, D2, D1, D0) which represents the value of
the output of the pulse counter block. In addition, the results of the calculation rate is also
displayed with an LCD unit. This block already exists in the module of Spartan 3E FPGA Starter
Kit [4], as shown in the figure 8.
Figure 8. Display of LED and LCD at FPGA Spartan 3E Starter Kit [4]
2.2. Designing of Programming Xilinx ISE Webpack
In this research, design of Xilinx ISE Webpack program done on the pulse counter block and
block the timer 10s. The programming result, then applied to the module Field Programable Gate
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
5
Array (FPGA), through the implementation process. Output of this system is the number of
heartbeat information in 1 minute, which known as beat permenit (bpm). Using of this system,
in the process of beating the count is done for 10 seconds as the sample to calculate the 1 minute,
so to show the heartbeat detection for 1 minute, required multiple of 6 constant number.
Flowchart program of heartbeat monitoring device shown in figure 9.
Results from the VHDL programming of heartbeat monitorr has been simulated correctly, then
the program is implemented on Spartan 3E FPGA board.
Figure 9. Flowchart of Heart Rate Monitoring VHDL Programming
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
6
2.3. RTL Viewer
Figure 10. RTL Viewer Heartrate Monitor
After the process of synthesis and design implementation, next the process is VIEW RTL
Schematic, as shown in the figure 10. This figure describes the gates configuration that represent
the pulse counter and the timer 10s devices.
3. EXPERIMENTAL RESULT
3.1 Conditional Signal Output Testing
Based on Figure 11, can be explained that the heart rate signal measurement results using the
oscilloscope at the output of the signal conditioner has a value of amplitude of 8.4 V, with a value
of 800 ms period or frequency of 1.25 Hz.
Figure 11. Display of output the signal conditioning block on the oscilloscope output
.
3.2. Simulation Results Heartbeat Counters Block
Testing in this section is to observe the results display on the programming VHDL simulation
with Xilinx Isim, as shown as Figure 12.
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
7
Figure 12. Display of simulation result
Figure 12 shows an example of simulation result Heartbeat Counters Block with a insensor input
signal with a period of 450ms (1.33Hz). At the output outcounter generate data 010111b, which
represents the beating of the time there are 23 counters for 10 seconds. So that the calculation
results heartbeat (heart rate) for 23 x 6 = 138 bpm (beats perminute).
3.3 Testing Results of The Heartbeats Monitor Device
Testing is done by calculating the heart rate to some respondents. Based on the output pulse
counter block built instrument (ha) compared with the measurement results using a fingertip pulse
sensor oxymeter (ho) in units of bpm. Difference in outcome differences between the two
devices, then calculated the percentage error (% Er) with the equation:
………………..……………………..(1)
Table 1: Testing results of the heartbeats monitor device
Responden
LED Condition
FPGA
Heart Rate
(bpm)
LCD
Heart Rate
(bpm)
Oxymeter
Error (%)
D5 D4 D3 D2 D1 DL0
1 0 0 1 1 0 1 78 76 2,6
2 0 0 1 0 1 1 66 70 5,7
3 0 0 1 1 1 0 84 82 2,4
4 0 0 1 1 0 1 78 77 1,3
5 0 0 1 1 1 0 84 81 3,7
6 0 0 1 1 0 0 72 75 4
7 0 0 1 1 1 1 90 84 7,1
8 0 0 1 1 0 1 78 82 7,3
9 0 1 0 0 0 0 96 94 2,1
10 0 0 1 1 1 1 90 93 3,2
Table 1 shows the trial against six respondents. At trial respondent 1, the output of the gain block
rate pengitung 001101b value as indicated by the LED flash is worth 1, and if the flame is not
worth 001101b value 0 is equivalent to 13d. Furthermore, to get the heart rate measurement with
units of beats per minute (bpm) needs to be multiplied by a constant 6 Thus, respondent 1 has a
value of 13 x 6 = 78 bpm. The same was done for the other respondents. Measurement accuracy
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
8
in terms of percentage error is calculated using equation (1). Based on the test results in Table 1,
it has been tested tool designed to measure the number of heartbeats to 10 respondents, has an
error (Error) an average of 3.94%.
4. CONCLUSION
This research has successfully designed a heartbeat monitor circuit using the Xilinx ISE program
WebPack 13.1. The series is composed of sensors block, signal conditioning block, pulse counter
block, timer 10s unit and the display unit. Based on the test results of the 15 respondents, a tool
designed to have a percentage error of 3.28%.
5. SUGESTION
a. In the testing process, the accuracy of the optical sensor is strongly influenced on laying /
finger placement ujuag. So to overcome this required the manufacture of containers or the
media, so the detection of blood flow in the fingertips can be precision quickly.
b. In the signal conditioning block in this research, design and manufacturing are still using
electronic components. In further studies need to be developed to block the signal conditioner
made with vhdl programming, thus forming sauté specific integrated circuits or ASICs known
(Application-Specific Integrated Circuits)
Figure 13. FPGA based heartbeats monitor circuit
REFERENCES
[1] Dogan., I., Kadri., B, 2012, Hear Rate Measurement from the Finger using a low cost
Microcontroller, http://www.emo.org.tr/ekler/ a568a2aa8c19a31_ek. pdf,
[2] Machriz, E., Sony S., Achmad R., 2008, Perancangan Perangkat Monitoring Denyut Jantung (Heart-
Beat Monitoring) dengan Visualisasi LCD Grafik Berbasis ATMEL AT89C51, proseding
Konferensi Nasional Sistem dan Informatika 2008; Bali, November 15, 2008.
[3] Hadiyani N., , 2012, Penyakit Jantung Koroner, http://www.dokterku-
online.com/index.php/article/54-penyakit-jantung-koroner, 14 Februari 2014
[4] Pong P. C., 2008, FPGA Prototyping by VHDL Examples: Xilinx Spartan-3 Version, Publisher:
Wiley-Interscience
[5] Sharief F. B., Liena E., Abdel K., Samah M. E., 2011, Microcontroller Based Heart Rate Monitor
using Fingertip Sensors, UofKEJ Vol. 1 Issue 2 pp. 47-51,October 2011
[6] Webster, EDS, 1981, Design of Microcomputer-Base Medical Instrumentation, Prentice Hall
International, New Jersey
Optical Sensor
Conditional Signal
Circuit
Xilinx Spartan 3E
FPGA
International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014
9
Authors
Wahyu Kusuma R received the Dr. in information technology from Gunadarma University, in
2008. Currently, he is a Electrical and Computer Laboratory staff at Gunadarma University.
His research interests include Voice Recognition Systems, Music Information Retrieval and
Embedded Systems with FPGA
Ridha Iskandar received the Dr. in information technology from Gunadarma University, in
2009. Currently, he is a computer Science laboratory staff at Gunadarma University. His
research interests include Biomedical Signal Processing and Embedded Systems with FPGA.
Yasman Rianto received as Student the Doctoral in information technology from
Gunadarma University, in 2013. Currently, he is a Physic laboratory staff at Gunadarma
University. His research interests include Sensoring and Microcontroller Application
Swelandiah E. Pratiwi graduated form Master of Electrical Enginering Gunadarma
University, in 2006. Currently, he is a Information Technology laboratory staff at
Gunadarma University. His research interests include Embedded System and Microcontroller
Application.

More Related Content

What's hot

An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...
An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...
An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...IJTET Journal
 
The research of_portable_ecg_monitoring_system_with_usb_host_interface
The research of_portable_ecg_monitoring_system_with_usb_host_interfaceThe research of_portable_ecg_monitoring_system_with_usb_host_interface
The research of_portable_ecg_monitoring_system_with_usb_host_interfaceArhamSheikh1
 
Ijaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xIjaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xsalborunda
 
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...ijtsrd
 
Instant elelectrocardiogram monitoring in android smart phones
Instant elelectrocardiogram monitoring in android smart phonesInstant elelectrocardiogram monitoring in android smart phones
Instant elelectrocardiogram monitoring in android smart phonesIjrdt Journal
 
Embedded System Real Time Operating System (ERTS) I unit by vijay
Embedded System Real Time Operating System (ERTS) I unit by vijayEmbedded System Real Time Operating System (ERTS) I unit by vijay
Embedded System Real Time Operating System (ERTS) I unit by vijayVijay Kumar
 
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoard
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoardPortable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoard
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoardIJSRD
 
Microcontroller (8051) by K. Vijay Kumar
Microcontroller (8051) by K. Vijay KumarMicrocontroller (8051) by K. Vijay Kumar
Microcontroller (8051) by K. Vijay KumarVijay Kumar
 
A portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacA portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacArhamSheikh1
 
Towards development of a low cost and
Towards development of a low cost andTowards development of a low cost and
Towards development of a low cost andArhamSheikh1
 
Cardionics ECG Presentation mobile cartouch pharma
Cardionics ECG Presentation mobile cartouch pharmaCardionics ECG Presentation mobile cartouch pharma
Cardionics ECG Presentation mobile cartouch pharmaCARDIONICS BELGIUM
 
A novel efficient human computer interface using an electrooculogram
A novel efficient human computer interface using an electrooculogramA novel efficient human computer interface using an electrooculogram
A novel efficient human computer interface using an electrooculogrameSAT Publishing House
 
Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)IJERD Editor
 
683 690,tesma412,ijeast
683 690,tesma412,ijeast683 690,tesma412,ijeast
683 690,tesma412,ijeastArhamSheikh1
 
Report on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino UnoReport on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino UnoAshfaqul Haque John
 
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...TELKOMNIKA JOURNAL
 
Office automation system using arduino
Office automation system using arduinoOffice automation system using arduino
Office automation system using arduinoAshfaqul Haque John
 
A simple portable ecg monitor with iot
A simple portable ecg monitor with iotA simple portable ecg monitor with iot
A simple portable ecg monitor with iotArhamSheikh1
 
Smart hospital technology
Smart hospital technologySmart hospital technology
Smart hospital technologyhiij
 

What's hot (20)

An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...
An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...
An Efficient System Of Electrocardiogram Data Acquisition And Analysis Using ...
 
The research of_portable_ecg_monitoring_system_with_usb_host_interface
The research of_portable_ecg_monitoring_system_with_usb_host_interfaceThe research of_portable_ecg_monitoring_system_with_usb_host_interface
The research of_portable_ecg_monitoring_system_with_usb_host_interface
 
Ijaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xIjaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195x
 
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
 
Instant elelectrocardiogram monitoring in android smart phones
Instant elelectrocardiogram monitoring in android smart phonesInstant elelectrocardiogram monitoring in android smart phones
Instant elelectrocardiogram monitoring in android smart phones
 
Embedded System Real Time Operating System (ERTS) I unit by vijay
Embedded System Real Time Operating System (ERTS) I unit by vijayEmbedded System Real Time Operating System (ERTS) I unit by vijay
Embedded System Real Time Operating System (ERTS) I unit by vijay
 
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoard
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoardPortable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoard
Portable ECG Monitoring System using Lilypad And Mobile Platform-PandaBoard
 
Microcontroller (8051) by K. Vijay Kumar
Microcontroller (8051) by K. Vijay KumarMicrocontroller (8051) by K. Vijay Kumar
Microcontroller (8051) by K. Vijay Kumar
 
A portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacA portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiac
 
Towards development of a low cost and
Towards development of a low cost andTowards development of a low cost and
Towards development of a low cost and
 
Ijigsp v10-n5-3
Ijigsp v10-n5-3Ijigsp v10-n5-3
Ijigsp v10-n5-3
 
Cardionics ECG Presentation mobile cartouch pharma
Cardionics ECG Presentation mobile cartouch pharmaCardionics ECG Presentation mobile cartouch pharma
Cardionics ECG Presentation mobile cartouch pharma
 
A novel efficient human computer interface using an electrooculogram
A novel efficient human computer interface using an electrooculogramA novel efficient human computer interface using an electrooculogram
A novel efficient human computer interface using an electrooculogram
 
Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)
 
683 690,tesma412,ijeast
683 690,tesma412,ijeast683 690,tesma412,ijeast
683 690,tesma412,ijeast
 
Report on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino UnoReport on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino Uno
 
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
 
Office automation system using arduino
Office automation system using arduinoOffice automation system using arduino
Office automation system using arduino
 
A simple portable ecg monitor with iot
A simple portable ecg monitor with iotA simple portable ecg monitor with iot
A simple portable ecg monitor with iot
 
Smart hospital technology
Smart hospital technologySmart hospital technology
Smart hospital technology
 

Similar to FPGA BASED HEARTBEATS MONITOR WITH FINGERTIP OPTICAL SENSOR

Design and implementation of portable electrocardiogram recorder with field ...
Design and implementation of portable electrocardiogram  recorder with field ...Design and implementation of portable electrocardiogram  recorder with field ...
Design and implementation of portable electrocardiogram recorder with field ...IJECEIAES
 
Iaetsd vlsi based implementation of a digital
Iaetsd vlsi based implementation of a digitalIaetsd vlsi based implementation of a digital
Iaetsd vlsi based implementation of a digitalIaetsd Iaetsd
 
Fpga based computer aided diagnosis of cardiac murmurs and sounds
Fpga based computer aided diagnosis of cardiac murmurs and soundsFpga based computer aided diagnosis of cardiac murmurs and sounds
Fpga based computer aided diagnosis of cardiac murmurs and soundseSAT Publishing House
 
-1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid
 -1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid -1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid
-1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaidsairamreddy siddu
 
Pulse Plus: Engineered Heart Monitoring Device
Pulse Plus: Engineered Heart Monitoring  DevicePulse Plus: Engineered Heart Monitoring  Device
Pulse Plus: Engineered Heart Monitoring DevicePalash Sukla Das
 
Introduction to Bedside Monitor .pptx
Introduction to  Bedside   Monitor .pptxIntroduction to  Bedside   Monitor .pptx
Introduction to Bedside Monitor .pptxRevathiJ10
 
Angina Pectoris Predicting System using Embedded Systems and Big Data
Angina Pectoris Predicting System using Embedded Systems and Big DataAngina Pectoris Predicting System using Embedded Systems and Big Data
Angina Pectoris Predicting System using Embedded Systems and Big DataIRJET Journal
 
Arduino Based Abnormal Heart Rate Detection and Wireless Communication
Arduino Based Abnormal Heart Rate Detection and Wireless CommunicationArduino Based Abnormal Heart Rate Detection and Wireless Communication
Arduino Based Abnormal Heart Rate Detection and Wireless Communicationijcisjournal
 
Heart monitoring
Heart monitoringHeart monitoring
Heart monitoringAmit Sheth
 
Electronic stethoscope academic white paper
Electronic stethoscope academic white paperElectronic stethoscope academic white paper
Electronic stethoscope academic white paperclauworlf55
 
ECG monitoring on android smartphone
ECG monitoring on android smartphone ECG monitoring on android smartphone
ECG monitoring on android smartphone Rushikesh Yeldari
 
Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.Arkadeep Dey
 
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded System
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded SystemFPGA Design and Implementation of Electrocardiogram Biomedical Embedded System
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded Systemns risman
 
Vivek_Presentation1.pptx
Vivek_Presentation1.pptxVivek_Presentation1.pptx
Vivek_Presentation1.pptxVishalLabde
 

Similar to FPGA BASED HEARTBEATS MONITOR WITH FINGERTIP OPTICAL SENSOR (20)

Design and implementation of portable electrocardiogram recorder with field ...
Design and implementation of portable electrocardiogram  recorder with field ...Design and implementation of portable electrocardiogram  recorder with field ...
Design and implementation of portable electrocardiogram recorder with field ...
 
Iaetsd vlsi based implementation of a digital
Iaetsd vlsi based implementation of a digitalIaetsd vlsi based implementation of a digital
Iaetsd vlsi based implementation of a digital
 
Fpga based computer aided diagnosis of cardiac murmurs and sounds
Fpga based computer aided diagnosis of cardiac murmurs and soundsFpga based computer aided diagnosis of cardiac murmurs and sounds
Fpga based computer aided diagnosis of cardiac murmurs and sounds
 
C04611318
C04611318C04611318
C04611318
 
-1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid
 -1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid -1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid
-1348064572-13. electronics - ijeceierd - design and - sapna katiyar - unpaid
 
Pulse Plus: Engineered Heart Monitoring Device
Pulse Plus: Engineered Heart Monitoring  DevicePulse Plus: Engineered Heart Monitoring  Device
Pulse Plus: Engineered Heart Monitoring Device
 
Introduction to Bedside Monitor .pptx
Introduction to  Bedside   Monitor .pptxIntroduction to  Bedside   Monitor .pptx
Introduction to Bedside Monitor .pptx
 
Angina Pectoris Predicting System using Embedded Systems and Big Data
Angina Pectoris Predicting System using Embedded Systems and Big DataAngina Pectoris Predicting System using Embedded Systems and Big Data
Angina Pectoris Predicting System using Embedded Systems and Big Data
 
Ijetcas14 323
Ijetcas14 323Ijetcas14 323
Ijetcas14 323
 
Arduino Based Abnormal Heart Rate Detection and Wireless Communication
Arduino Based Abnormal Heart Rate Detection and Wireless CommunicationArduino Based Abnormal Heart Rate Detection and Wireless Communication
Arduino Based Abnormal Heart Rate Detection and Wireless Communication
 
Heartm~1
Heartm~1Heartm~1
Heartm~1
 
Heart monitoring
Heart monitoringHeart monitoring
Heart monitoring
 
Electronic stethoscope academic white paper
Electronic stethoscope academic white paperElectronic stethoscope academic white paper
Electronic stethoscope academic white paper
 
ECG monitoring on android smartphone
ECG monitoring on android smartphone ECG monitoring on android smartphone
ECG monitoring on android smartphone
 
heartbeatsensor
heartbeatsensorheartbeatsensor
heartbeatsensor
 
Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.
 
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded System
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded SystemFPGA Design and Implementation of Electrocardiogram Biomedical Embedded System
FPGA Design and Implementation of Electrocardiogram Biomedical Embedded System
 
Bidirect visitor counter
Bidirect visitor counterBidirect visitor counter
Bidirect visitor counter
 
Vivek_Presentation1.pptx
Vivek_Presentation1.pptxVivek_Presentation1.pptx
Vivek_Presentation1.pptx
 
Report
ReportReport
Report
 

Recently uploaded

Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Lokesh Kothari
 
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Sérgio Sacani
 
Botany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsBotany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsSumit Kumar yadav
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000Sapana Sha
 
Chromatin Structure | EUCHROMATIN | HETEROCHROMATIN
Chromatin Structure | EUCHROMATIN | HETEROCHROMATINChromatin Structure | EUCHROMATIN | HETEROCHROMATIN
Chromatin Structure | EUCHROMATIN | HETEROCHROMATINsankalpkumarsahoo174
 
Disentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTDisentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTSérgio Sacani
 
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service 🪡
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service  🪡CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service  🪡
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service 🪡anilsa9823
 
Green chemistry and Sustainable development.pptx
Green chemistry  and Sustainable development.pptxGreen chemistry  and Sustainable development.pptx
Green chemistry and Sustainable development.pptxRajatChauhan518211
 
Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )aarthirajkumar25
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsSérgio Sacani
 
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisRaman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisDiwakar Mishra
 
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdf
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdfPests of cotton_Borer_Pests_Binomics_Dr.UPR.pdf
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdfPirithiRaju
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​kaibalyasahoo82800
 
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.Nitya salvi
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticssakshisoni2385
 
Animal Communication- Auditory and Visual.pptx
Animal Communication- Auditory and Visual.pptxAnimal Communication- Auditory and Visual.pptx
Animal Communication- Auditory and Visual.pptxUmerFayaz5
 
Natural Polymer Based Nanomaterials
Natural Polymer Based NanomaterialsNatural Polymer Based Nanomaterials
Natural Polymer Based NanomaterialsAArockiyaNisha
 
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencyHire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencySheetal Arora
 
Biological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfBiological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfmuntazimhurra
 

Recently uploaded (20)

Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
 
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
 
Botany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questionsBotany krishna series 2nd semester Only Mcq type questions
Botany krishna series 2nd semester Only Mcq type questions
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
 
Chromatin Structure | EUCHROMATIN | HETEROCHROMATIN
Chromatin Structure | EUCHROMATIN | HETEROCHROMATINChromatin Structure | EUCHROMATIN | HETEROCHROMATIN
Chromatin Structure | EUCHROMATIN | HETEROCHROMATIN
 
Disentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTDisentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOST
 
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service 🪡
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service  🪡CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service  🪡
CALL ON ➥8923113531 🔝Call Girls Kesar Bagh Lucknow best Night Fun service 🪡
 
Green chemistry and Sustainable development.pptx
Green chemistry  and Sustainable development.pptxGreen chemistry  and Sustainable development.pptx
Green chemistry and Sustainable development.pptx
 
The Philosophy of Science
The Philosophy of ScienceThe Philosophy of Science
The Philosophy of Science
 
Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
 
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral AnalysisRaman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
Raman spectroscopy.pptx M Pharm, M Sc, Advanced Spectral Analysis
 
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdf
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdfPests of cotton_Borer_Pests_Binomics_Dr.UPR.pdf
Pests of cotton_Borer_Pests_Binomics_Dr.UPR.pdf
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​
 
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
Animal Communication- Auditory and Visual.pptx
Animal Communication- Auditory and Visual.pptxAnimal Communication- Auditory and Visual.pptx
Animal Communication- Auditory and Visual.pptx
 
Natural Polymer Based Nanomaterials
Natural Polymer Based NanomaterialsNatural Polymer Based Nanomaterials
Natural Polymer Based Nanomaterials
 
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls AgencyHire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
Hire 💕 9907093804 Hooghly Call Girls Service Call Girls Agency
 
Biological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdfBiological Classification BioHack (3).pdf
Biological Classification BioHack (3).pdf
 

FPGA BASED HEARTBEATS MONITOR WITH FINGERTIP OPTICAL SENSOR

  • 1. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 DOI : 10.5121/ijcseit.2014.4501 1 FPGA BASED HEARTBEATS MONITOR WITH FINGERTIP OPTICAL SENSOR Wahyu Kusuma R.1 , Ridha I.2 , Yasman Rianto3 , Swelandiah E.P4 1,3 Departement of Electrcal Engineering, Gunadarma University, Jakarta, Indonesia 2,4 Departement of Computer Science, Gunadarma University, , Jakarta, Indonesia ABSTRACT The heart is an organ of human body which has a vital function, small abnormalities can have a big impact on the performance of the body. Heart disease is the number one cause of death in the world. Examination of the heart can be detected from blood flow in the fingertips, in order to obtain information about the number and rhythm of the heartbeat. This research aims to design and implement the FPGA board to monitor the heart rate with optical sensors.The results of this study are expected to facilitate the patient's medical team or independently in detecting heart health. The series is composed of blocks of sensors, signal conditioning block, the block pulse counter, block timer 10 seconds and blocks the viewer. Based on the test results of the 10 respondents with a variety of age and gender, has built a tool that the percentage error of 3.94%. KEYWORDS Heartbeat monitor, Xilinx ISE Webpack, FPGA, optical sensor 1. INTRODUCTION The heart is an organ of human body which has a vital function, small abnormalities can have a big impact on the performance of the heart kita.Penyakit body is the number one cause of death in the world. Based on data from the World Health Organization (WHO), cardiovascular disease has reached 29% in the percentage of deaths in the world and 17 million people die every year due to heart and blood vessel disease throughout the world [3]. The development of medical instrumentation systems is growing rapidly along with the need for medical personnel to diagnose a patient and a medical examination. One medical instrumentation used for the examination of the heart is Electrocardiograph (ECG). ECG is a medical instrument that is commonly used by the medical team to detect heart rate and rhythm [6]. EKG can not be used independently by patients to detect a patient's pulse. In addition to the expensive costs for the procurement of ECG, ECG devices also require special skills to operate. Along with the requirement in the design and manufacture of medical devices, digital electronics design technology is developing very rapidly, both in terms of hardware and software. Xilinx is one manufacturer that produces equipment or tools for modeling the design of digital systems. One product is in the form of a kit module board FPGA (Field Programmable Gate Array). FPGA is a programmable device that is composed of large modules independent logic that can be configured by the user who is connecting through the canals of programmable routing [4].
  • 2. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 2 Field Programmable Gate Array (FPGA) has advantages of which can be configured by the End User, does not require the fabrication process, and the available solutions that support customized VLSI chip, so the logic circuit capable of implementing, instant manufacturring, very-low cost prototype. While FPGA product is also supported by the availability of the Xilinx ISE WebPack permograman language that is used as a GUI for the digital system design problems is desired. Based on the above issues, this study will design a device that can detect heart rate by detecting the frequency of blood flow to the fingers automatically processed electronically using the module FPGA (Field Programable Gate Array). This research is expected to generate a heartbeat detector chip implementation can provide facilities for the medical and can be used independently by the user (patient) without the help of a doctor or paramedic. 2. ARCHITECTURE AND ITS IMPLEMENTATION This research conducted several phases of hardware design, software design pelangkat, testing tools, and analysis to draw conclusions. 2.1 Hardware Design The design of this tool consists of several circuit blocks are arranged into a single system heartbeats monitoring based on optical sensors, as shown in Figure 1. Figure 1. Block Diagram Tool Heartbeats Detection Based on Optical Sensors 2.1.1. Optical Sensor Unit The sensor consists of a transmitter that emits infrared light (IR) and IR photo detector (photodiode) to act as a receiver. Construction sensor block shown in Figure 2. The use of optical sensors for process monitoring heart rate with fingertip illustrated as a figure 3.
  • 3. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 3 Figure 2. Optical sensor circuit Figure 3. Ilustration fingertip to optical sensor 2.1.2. Conditional Signal Unit The output signal of the sensor is very small and still a lot of noise. This signal before entering the pulse counter block should be done in conditioning, so signal that result this block is ready to be read or processed by the block counter. The signal conditioner block consists of a series of low pass filter and amplifier structure is shown as figure 4. Figure 4. Diagram Block of Conditional Signal Unit Figure 5. Optical Sensor and Conditional Signal Circuits 2.1.3. Heart Counter Unit This block is used to calculate the output signal generated by the signal conditioning block. Signals that represent a high rate, while the low signal represents the absence of heartbeat. So this block will calculate the amount of high signal. The basic principle of this block as a series of counters. This block has three inputs namely are Clk signal, the signal In is connected to the output signal conditioning block and Enable signal (En) which is connected to the output signal at the timer block. The output of the pulse counter block form as 6-bit binary sequence (D5, D4, D3, D2, D1, D0). In this research, the pulse
  • 4. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 4 counter block is designed with programming languages Xlinx WebPack ISE. The design of the pulse counter block is shown as figure 6. Figurer 6. Design of Heart Counter Unit 2.1.4. Timer 10s Unit This block is designed for pulse counting process is done by calculating the input signal to the timer block rate for 10 seconds. The basic principle of the timer unit is a clock signal frequency divider circuit. The clock signal is taken from the internal oscillator signal frequency of 50 MHz on the FPGA module. In order to get a signal with a period of 20 seconds, then beat signal as a function of the timer must have a frequency of 50 mHz, by taking half the period of the high level. The output signal timer 10s unit to be input to the Enable signal (En) at the pulse counter block. The timer 10s unit is shown in Figure 7. Figure 7. Diagram Block of Timer 10s 2.1.5. Display Unit The Display unit that is used is the 6 Led (D5, D4, D3, D2, D1, D0) which represents the value of the output of the pulse counter block. In addition, the results of the calculation rate is also displayed with an LCD unit. This block already exists in the module of Spartan 3E FPGA Starter Kit [4], as shown in the figure 8. Figure 8. Display of LED and LCD at FPGA Spartan 3E Starter Kit [4] 2.2. Designing of Programming Xilinx ISE Webpack In this research, design of Xilinx ISE Webpack program done on the pulse counter block and block the timer 10s. The programming result, then applied to the module Field Programable Gate
  • 5. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 5 Array (FPGA), through the implementation process. Output of this system is the number of heartbeat information in 1 minute, which known as beat permenit (bpm). Using of this system, in the process of beating the count is done for 10 seconds as the sample to calculate the 1 minute, so to show the heartbeat detection for 1 minute, required multiple of 6 constant number. Flowchart program of heartbeat monitoring device shown in figure 9. Results from the VHDL programming of heartbeat monitorr has been simulated correctly, then the program is implemented on Spartan 3E FPGA board. Figure 9. Flowchart of Heart Rate Monitoring VHDL Programming
  • 6. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 6 2.3. RTL Viewer Figure 10. RTL Viewer Heartrate Monitor After the process of synthesis and design implementation, next the process is VIEW RTL Schematic, as shown in the figure 10. This figure describes the gates configuration that represent the pulse counter and the timer 10s devices. 3. EXPERIMENTAL RESULT 3.1 Conditional Signal Output Testing Based on Figure 11, can be explained that the heart rate signal measurement results using the oscilloscope at the output of the signal conditioner has a value of amplitude of 8.4 V, with a value of 800 ms period or frequency of 1.25 Hz. Figure 11. Display of output the signal conditioning block on the oscilloscope output . 3.2. Simulation Results Heartbeat Counters Block Testing in this section is to observe the results display on the programming VHDL simulation with Xilinx Isim, as shown as Figure 12.
  • 7. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 7 Figure 12. Display of simulation result Figure 12 shows an example of simulation result Heartbeat Counters Block with a insensor input signal with a period of 450ms (1.33Hz). At the output outcounter generate data 010111b, which represents the beating of the time there are 23 counters for 10 seconds. So that the calculation results heartbeat (heart rate) for 23 x 6 = 138 bpm (beats perminute). 3.3 Testing Results of The Heartbeats Monitor Device Testing is done by calculating the heart rate to some respondents. Based on the output pulse counter block built instrument (ha) compared with the measurement results using a fingertip pulse sensor oxymeter (ho) in units of bpm. Difference in outcome differences between the two devices, then calculated the percentage error (% Er) with the equation: ………………..……………………..(1) Table 1: Testing results of the heartbeats monitor device Responden LED Condition FPGA Heart Rate (bpm) LCD Heart Rate (bpm) Oxymeter Error (%) D5 D4 D3 D2 D1 DL0 1 0 0 1 1 0 1 78 76 2,6 2 0 0 1 0 1 1 66 70 5,7 3 0 0 1 1 1 0 84 82 2,4 4 0 0 1 1 0 1 78 77 1,3 5 0 0 1 1 1 0 84 81 3,7 6 0 0 1 1 0 0 72 75 4 7 0 0 1 1 1 1 90 84 7,1 8 0 0 1 1 0 1 78 82 7,3 9 0 1 0 0 0 0 96 94 2,1 10 0 0 1 1 1 1 90 93 3,2 Table 1 shows the trial against six respondents. At trial respondent 1, the output of the gain block rate pengitung 001101b value as indicated by the LED flash is worth 1, and if the flame is not worth 001101b value 0 is equivalent to 13d. Furthermore, to get the heart rate measurement with units of beats per minute (bpm) needs to be multiplied by a constant 6 Thus, respondent 1 has a value of 13 x 6 = 78 bpm. The same was done for the other respondents. Measurement accuracy
  • 8. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 8 in terms of percentage error is calculated using equation (1). Based on the test results in Table 1, it has been tested tool designed to measure the number of heartbeats to 10 respondents, has an error (Error) an average of 3.94%. 4. CONCLUSION This research has successfully designed a heartbeat monitor circuit using the Xilinx ISE program WebPack 13.1. The series is composed of sensors block, signal conditioning block, pulse counter block, timer 10s unit and the display unit. Based on the test results of the 15 respondents, a tool designed to have a percentage error of 3.28%. 5. SUGESTION a. In the testing process, the accuracy of the optical sensor is strongly influenced on laying / finger placement ujuag. So to overcome this required the manufacture of containers or the media, so the detection of blood flow in the fingertips can be precision quickly. b. In the signal conditioning block in this research, design and manufacturing are still using electronic components. In further studies need to be developed to block the signal conditioner made with vhdl programming, thus forming sauté specific integrated circuits or ASICs known (Application-Specific Integrated Circuits) Figure 13. FPGA based heartbeats monitor circuit REFERENCES [1] Dogan., I., Kadri., B, 2012, Hear Rate Measurement from the Finger using a low cost Microcontroller, http://www.emo.org.tr/ekler/ a568a2aa8c19a31_ek. pdf, [2] Machriz, E., Sony S., Achmad R., 2008, Perancangan Perangkat Monitoring Denyut Jantung (Heart- Beat Monitoring) dengan Visualisasi LCD Grafik Berbasis ATMEL AT89C51, proseding Konferensi Nasional Sistem dan Informatika 2008; Bali, November 15, 2008. [3] Hadiyani N., , 2012, Penyakit Jantung Koroner, http://www.dokterku- online.com/index.php/article/54-penyakit-jantung-koroner, 14 Februari 2014 [4] Pong P. C., 2008, FPGA Prototyping by VHDL Examples: Xilinx Spartan-3 Version, Publisher: Wiley-Interscience [5] Sharief F. B., Liena E., Abdel K., Samah M. E., 2011, Microcontroller Based Heart Rate Monitor using Fingertip Sensors, UofKEJ Vol. 1 Issue 2 pp. 47-51,October 2011 [6] Webster, EDS, 1981, Design of Microcomputer-Base Medical Instrumentation, Prentice Hall International, New Jersey Optical Sensor Conditional Signal Circuit Xilinx Spartan 3E FPGA
  • 9. International Journal of Computer Science, Engineering and Information Technology (IJCSEIT), Vol. 4, No.5, October 2014 9 Authors Wahyu Kusuma R received the Dr. in information technology from Gunadarma University, in 2008. Currently, he is a Electrical and Computer Laboratory staff at Gunadarma University. His research interests include Voice Recognition Systems, Music Information Retrieval and Embedded Systems with FPGA Ridha Iskandar received the Dr. in information technology from Gunadarma University, in 2009. Currently, he is a computer Science laboratory staff at Gunadarma University. His research interests include Biomedical Signal Processing and Embedded Systems with FPGA. Yasman Rianto received as Student the Doctoral in information technology from Gunadarma University, in 2013. Currently, he is a Physic laboratory staff at Gunadarma University. His research interests include Sensoring and Microcontroller Application Swelandiah E. Pratiwi graduated form Master of Electrical Enginering Gunadarma University, in 2006. Currently, he is a Information Technology laboratory staff at Gunadarma University. His research interests include Embedded System and Microcontroller Application.