AN EMBEDDED SYSTEM FOR NON-INVASIVE BLOOD GLUCOSE
MEASUREMENT IN GLUCOSE TOLERANCE TEST
Indhulekshmi M C1
, Liju Philip2
1
PG Scholar, Department of Electronics Engineering, College of Engineering Chengannur, Kerala, India
2
Assistant Professor, Department of Electronics Engineering, College of Engineering Chengannur, Kerala, India
Abstract
Diabetic patients need to monitor their blood glucose levels frequently to prevent complications related to the disease. Commonly
used glucose measurement methods are invasive, uncomfortable, and usually require frequent finger pricking. This paper deals
with the development of a non-invasive system for continuous measurement of blood glucose concentration levels, using near
infrared light of suitable wavelength. Glucose determination using near infrared can be done by taking measurement of
transmitted light through human body. This device may become very useful while conducting glucose tolerance test (GTT). In
glucose tolerance test, patient should be given a measured dose of glucose solution to drink and blood is drawn at intervals of half
an hour or one hour for two to four hours. This requires five to nine finger pricking for measurement of glucose concentration. By
using this proposed device pricking can be reduced to one. First the glucose concentration is measured using conventional
method for calibration. Then by comparing with the first sample measurement subsequent values can be determined without
further pricking.
Keywords: Non invasive measurement, Near Infrared, Blood Glucose, Transmittance, Diabetes
--------------------------------------------------------------------***----------------------------------------------------------------------
1. INTRODUCTION
Diabetes is metabolic disorder in which blood glucose
fluctuates from its normal range (70-110mg/dl). Insulin is a
hormone produced in body to regulate blood glucose level
naturally. Under some pathological failure, body is not able
to produce insulin or body cells become unable to use
insulin. Thus amount of glucose present in blood exceeds
normal range and patient may become diabetic. Nowadays
number of diabetic people is increasing across the world due
to population growth, unhealthy diet, obesity and lack of
physical activity. Diabetic patients need to monitor their
blood glucose levels frequently to prevent complications
related to the disease.
Most of commercially available glucose measurement
devices are invasive. The invasive methods are painful, have
high recurring cost and danger of spreading infectious
diseases. Non-invasive methods are more desirable and
excellent alternatives to these devices. Non-invasive glucose
monitoring could make millions of people more relaxed and
comfortable during blood glucose testing. Thus it is
necessary to develop a non-invasive blood glucose
measurement system which is painless, convenient and cost
effective. Non-invasive monitoring system will be a major
breakthrough in the area of treating diabetes patients.
Optical methods are painless and promising methods that
can be used for non-invasive blood glucose measurement.
Near-infrared (NIR) is one of the most widely explored
optical techniques because of its high penetration in skin.
This technique has been applied on various body parts:
finger, palm, arm, forearm, earlobe etc.
This work deals with the development of a device for
measuring glucose concentration using near infrared light.
The proposed device may become very useful while
conducting glucose tolerance test (GTT).Glucose tolerance
test (GTT) is a medical test which measures how well our
body is able to break down glucose. In this test patient is
given a measured dose of glucose solution to drink. Then
blood samples are taken at specified intervals for the
measurement of glucose concentration to determine how
quickly it is cleared from the blood. Usually Glucose
tolerance test takes about two to four hours. The patient is
instructed to fast (water is allowed) for eight to twelve hours
prior to the test. First a fasting blood sample is taken and
then patient is given a measured dose of glucose solution to
drink. Blood is drawn at intervals of half an hour or one
hour for two to four hours to determine the glucose
concentration. This requires five to nine finger pricking
which cause pain and discomfort to patients.
By using this proposed device pricking can be reduced to
one. First the glucose concentration is measured using
conventional method for calibration. Then by comparing
with the first sample measurement subsequent values can be
determined without further pricking.
The goal of this work is to develop a reliable, simple,
painless and portable device for glucose measurement that
does not cause repetitive puncturing of skin.
2. PROPOSED SYSTEM
Near infrared based optical technique has been selected
from a number of methods because it is easier to implement
and needs less complex hardware setup.The near infrared
IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308
_______________________________________________________________________________________
Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 64
(NIR) light in the range of 700nm-2500nm is the most
suitable for detecting glucose concentration.
The proposed system is based on the principle of
transmittance measurement. The value of transmission of
light energy is dependent on the number of molecules
present in absorbing material. Thus, intensity of light energy
transmitted out from the absorbing substance is used as a
sign of concentration of that particular substance. Hence by
measuring intensity of transmitted light through the
measurement site, concentration of glucose in blood can be
calculated.
Fig-1: Non invasive glucose concentration measurement
system
This system allows continuous measurement of variation in
blood glucose concentration. The instrumentation for optical
measurements (sensing unit) consists of an NIR emitter and
receiver of 950nm wavelength which is suitable for glucose
concentration measurement.NIR rays emitted by the NIR-
LED passes through the measurement site and the
transmitted rays are received by NIR receiver. After
amplification output signal from the sensing unit is given to
the microcontroller. The microcontroller has an inbuilt
analog to digital converter that converts the analog electrical
signal into digital values. The digitized values are then
processed in microcontroller to determine the corresponding
blood glucose concentration.
For proper measurement site selection two factors should be
considered: i) The measurement site should be easily
accessible ii) NIR light should easily transmit through the
measurement site. Hence earlobe or finger web is selected as
suitable measurement site.
The sensing unit must be fixed in proper position during
entire measurement period. It is designed in such a way that
the detector is placed exactly opposite and in line to the NIR
light source for detecting maximum amount of transmitted
light. The light source and detector should be placed as close
as possible to the measurement site in physical contact with
the skin.
2.1 Operation of the System
First glucose concentration is measured using conventional
invasive method. After calibrating the proposed device with
the first sample measurement, subsequent glucose
concentration values are determined without further finger
pricking. Thus pricking can be reduced to one while
conducting glucose tolerance test.
Once the system is turned on, the infrared light propagates
through the measurement site. During the propagating
phase, NIR light interacts with the molecules present in the
medium (glucose in blood). A part of light gets absorbed
depending on the glucose concentration of blood under test
and remaining part is transmitted through the measurement
site. The transmitted signal is then detected by the infrared
sensitive detector. After amplification the signal is digitized
and is processed in the microcontroller to determine the
blood glucose concentration. The measured blood glucose
concentration is then displayed on the LCD. The time of
measurements and the corresponding value of glucose
concentration are stored in memory for further analysis
3. RESULTS AND DISCUSSION
The design was tested on seven non-diabetic individuals
including both genders. The measurement sites chosen were
earlobe and part of skin between thumb & index finger. The
near infrared emitter and receiver were positioned exactly
opposite to each other to obtain transmitted signal through
measurement site.
First the proposed device is calibrated with fasting glucose
concentration obtained by the invasive method. After meal
the signal from the detector is taken every half an hour for
two to three hours to determine the variation in glucose
concentration. The persons were not allowed to move
throughout the measurement period to avoid change in
sensor unit positioning in the measurement site.
Table-1: Fasting glucose value obtained from invasive
laboratory test
Person A B C D E F G
Fasting
glucose(mg/dl)
89 85 90 87 91 100 85
Table-2: Variation in glucose concentration obtained from
proposed non invasive device after meal
IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308
_______________________________________________________________________________________
Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 65
Tables-1 and 2 given below shows the data collected from
each person, during the experiment. Table-1 shows the
fasting glucose value obtained from invasive laboratory
test.Table-2 shows the variation in glucose concentration
obtained from proposed non invasive device after meal.
For validation of the experiment, a value obtained from the
proposed device is compared with invasive laboratory test
result. Two non-diabetic persons were considered for this
experiment. At first the device is calibrated using the fasting
glucose value which is measured invasively. Then glucose
concentration readings of the above two persons after meal,
using the proposed device and invasive laboratory test, is
taken at an interval of one hour.
Data obtained from Person 1
Invasively measured fasting glucose concentration of
person 1: 83 mg/dl.
Measurement site: finger web
Table-3: Data obtained from sensor and invasive laboratory
test of person 1
Sl.no
Time
(min)
Non invasive
glucose value from
sensor (mg/dl)
Invasive
laboratory test
result
1 60 114 98
2 120 143 138
Data obtained from Person 2
Invasively measured fasting glucose concentration of
person 2: 85 mg/dl.
Measurement site: finger web
Table -4: Data obtained from sensor and invasive laboratory
test of person 2
Sl.no
Time
(min)
Non invasive glucose
value from sensor
(mg/dl)
Invasive
laboratory test
result
1 60 91 89
2 120 103 93
Fig-2: Noninvasive and invasive glucose measurement of
person 1
Fig-3: Non invasive and invasive glucose measurement of
person 2
It was found from the figures 2 and 3, that the values
acquired from this proposed device shows an incremental
nature as compared with the blood glucose level obtained
with invasive laboratory test.
Fig-4: Emitter and detector setup
Fig-5: Final hardware
IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308
_______________________________________________________________________________________
Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 66
4. CONCLUSION
In the present work an embedded system for noninvasive
procedure based blood glucose measurement for conducting
Glucose Tolerance Test is developed. By using this
proposed device pricking can be reduced to one while
conducting glucose tolerance test.Experiments are carried
out using the proposed non invasive device and the results
obtained are comparable with invasive measurement. This
device is also useful for continuous glucose measurement of
any bedridden or ICU patients.
Challenges faced in this work are:
i) The skin transmittance is different for different persons
depending on the colour, texture of the skin and thickness of
measurement site. So the proposed device needs to be
calibrated for each person before conducting the test.
ii) The sensing unit must be properly positioned and not
allowed to move during the entire measurement period.
In future the performance of the proposed system can be
improved by adopting differential measurement in which
the difference between incident light and transmitted light
from skin is used for calculating glucose concentration.
REFERENCES
[1]. Omar S. Khalil, “Spectroscopic and Clinical Aspects of
Noninvasive Glucose Measurements”, Clinical Chemistry
45:2 165–177 (1999)
[2]. JyotiYadav, Asha Rani, Vijander Singh, Bhaskar Mohan
Murari, “Near-infrared LED based Non-invasive Blood
Glucose Sensor”, 2014 international conference on Signal
Processing and Integrated Networks (SPIN).
[3]. Ola S. Abdalsalam, Al-khanssa M. Osman, Roua M.
Abd-Alhadi, Saad D. Alshmaa, “ Design of Simple
Noninvasive Glucose Measuring Device”,2013 International
Conference on Computing, Electrical and Electronic
Engineering (ICCEEE)
[4]. A.G.Pavithra, D.MeneshaKaran,D. Ajith Kumar, P .S.
AnuShalin ,”Non invasive technique to measure glucose and
haemoglobin level in blood using NIR
spectroscopy”,International Journal of scientific research
and management.Volume2,Issue4,2013

More Related Content

Similar to An embedded system for non invasive blood glucose measurement in glucose tolerance test

Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurementijceronline
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...TELKOMNIKA JOURNAL
 
A gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring systemA gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring systemeSAT Journals
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method IJECEIAES
 
IJBET-4-Narasimham et al
IJBET-4-Narasimham et alIJBET-4-Narasimham et al
IJBET-4-Narasimham et alGaurav Kaila
 
Wearable_Technolgies and medical devices.pdf
Wearable_Technolgies and medical devices.pdfWearable_Technolgies and medical devices.pdf
Wearable_Technolgies and medical devices.pdfSaketBansal9
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...IRJET Journal
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET Journal
 
Non invasive anemia screening- Mobile apps
Non invasive anemia screening- Mobile appsNon invasive anemia screening- Mobile apps
Non invasive anemia screening- Mobile appsSaiLakshmi128
 
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptx
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptxIOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptx
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptxaseelahkhazi13
 
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Method
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive MethodMeasurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Method
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Methodijtsrd
 
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTS
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTSAUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTS
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTShiij
 
Imrankhan.N-Occular Biosensors.pptx
Imrankhan.N-Occular Biosensors.pptxImrankhan.N-Occular Biosensors.pptx
Imrankhan.N-Occular Biosensors.pptxImran Khan
 
Real Time ECG and Saline Level Monitoring System Using Arduino UNO Processor
Real Time ECG and Saline Level Monitoring System Using Arduino UNO ProcessorReal Time ECG and Saline Level Monitoring System Using Arduino UNO Processor
Real Time ECG and Saline Level Monitoring System Using Arduino UNO ProcessorAssociate Professor in VSB Coimbatore
 

Similar to An embedded system for non invasive blood glucose measurement in glucose tolerance test (20)

Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurement
 
U34129131
U34129131U34129131
U34129131
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
 
A gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring systemA gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring system
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
 
IJBET-4-Narasimham et al
IJBET-4-Narasimham et alIJBET-4-Narasimham et al
IJBET-4-Narasimham et al
 
Wearable_Technolgies and medical devices.pdf
Wearable_Technolgies and medical devices.pdfWearable_Technolgies and medical devices.pdf
Wearable_Technolgies and medical devices.pdf
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
 
Non invasive anemia screening- Mobile apps
Non invasive anemia screening- Mobile appsNon invasive anemia screening- Mobile apps
Non invasive anemia screening- Mobile apps
 
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptx
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptxIOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptx
IOT BASED NON-INVASIVE BLOOD GLUCOSE LEVEL DETECTION.pptx
 
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Method
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive MethodMeasurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Method
Measurement of GHT Glucose, Heart Rate, Temperature Using Non Invasive Method
 
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTS
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTSAUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTS
AUTOMATIC AND NON-INVASIVE CONTINUOUS GLUCOSE MONITORING IN PAEDIATRIC PATIENTS
 
Imrankhan.N-Occular Biosensors.pptx
Imrankhan.N-Occular Biosensors.pptxImrankhan.N-Occular Biosensors.pptx
Imrankhan.N-Occular Biosensors.pptx
 
Glucose Monitoring Smart Watch
Glucose Monitoring Smart WatchGlucose Monitoring Smart Watch
Glucose Monitoring Smart Watch
 
Materials for glucose measurement
Materials for glucose measurementMaterials for glucose measurement
Materials for glucose measurement
 
IRJET- Smart Saline
IRJET-  	  Smart SalineIRJET-  	  Smart Saline
IRJET- Smart Saline
 
Real Time ECG and Saline Level Monitoring System Using Arduino UNO Processor
Real Time ECG and Saline Level Monitoring System Using Arduino UNO ProcessorReal Time ECG and Saline Level Monitoring System Using Arduino UNO Processor
Real Time ECG and Saline Level Monitoring System Using Arduino UNO Processor
 
40120140507005
4012014050700540120140507005
40120140507005
 
40120140507005
4012014050700540120140507005
40120140507005
 

More from eSAT Journals

Mechanical properties of hybrid fiber reinforced concrete for pavements
Mechanical properties of hybrid fiber reinforced concrete for pavementsMechanical properties of hybrid fiber reinforced concrete for pavements
Mechanical properties of hybrid fiber reinforced concrete for pavementseSAT Journals
 
Material management in construction – a case study
Material management in construction – a case studyMaterial management in construction – a case study
Material management in construction – a case studyeSAT Journals
 
Managing drought short term strategies in semi arid regions a case study
Managing drought    short term strategies in semi arid regions  a case studyManaging drought    short term strategies in semi arid regions  a case study
Managing drought short term strategies in semi arid regions a case studyeSAT Journals
 
Life cycle cost analysis of overlay for an urban road in bangalore
Life cycle cost analysis of overlay for an urban road in bangaloreLife cycle cost analysis of overlay for an urban road in bangalore
Life cycle cost analysis of overlay for an urban road in bangaloreeSAT Journals
 
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materials
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materialsLaboratory studies of dense bituminous mixes ii with reclaimed asphalt materials
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materialseSAT Journals
 
Laboratory investigation of expansive soil stabilized with natural inorganic ...
Laboratory investigation of expansive soil stabilized with natural inorganic ...Laboratory investigation of expansive soil stabilized with natural inorganic ...
Laboratory investigation of expansive soil stabilized with natural inorganic ...eSAT Journals
 
Influence of reinforcement on the behavior of hollow concrete block masonry p...
Influence of reinforcement on the behavior of hollow concrete block masonry p...Influence of reinforcement on the behavior of hollow concrete block masonry p...
Influence of reinforcement on the behavior of hollow concrete block masonry p...eSAT Journals
 
Influence of compaction energy on soil stabilized with chemical stabilizer
Influence of compaction energy on soil stabilized with chemical stabilizerInfluence of compaction energy on soil stabilized with chemical stabilizer
Influence of compaction energy on soil stabilized with chemical stabilizereSAT Journals
 
Geographical information system (gis) for water resources management
Geographical information system (gis) for water resources managementGeographical information system (gis) for water resources management
Geographical information system (gis) for water resources managementeSAT Journals
 
Forest type mapping of bidar forest division, karnataka using geoinformatics ...
Forest type mapping of bidar forest division, karnataka using geoinformatics ...Forest type mapping of bidar forest division, karnataka using geoinformatics ...
Forest type mapping of bidar forest division, karnataka using geoinformatics ...eSAT Journals
 
Factors influencing compressive strength of geopolymer concrete
Factors influencing compressive strength of geopolymer concreteFactors influencing compressive strength of geopolymer concrete
Factors influencing compressive strength of geopolymer concreteeSAT Journals
 
Experimental investigation on circular hollow steel columns in filled with li...
Experimental investigation on circular hollow steel columns in filled with li...Experimental investigation on circular hollow steel columns in filled with li...
Experimental investigation on circular hollow steel columns in filled with li...eSAT Journals
 
Experimental behavior of circular hsscfrc filled steel tubular columns under ...
Experimental behavior of circular hsscfrc filled steel tubular columns under ...Experimental behavior of circular hsscfrc filled steel tubular columns under ...
Experimental behavior of circular hsscfrc filled steel tubular columns under ...eSAT Journals
 
Evaluation of punching shear in flat slabs
Evaluation of punching shear in flat slabsEvaluation of punching shear in flat slabs
Evaluation of punching shear in flat slabseSAT Journals
 
Evaluation of performance of intake tower dam for recent earthquake in india
Evaluation of performance of intake tower dam for recent earthquake in indiaEvaluation of performance of intake tower dam for recent earthquake in india
Evaluation of performance of intake tower dam for recent earthquake in indiaeSAT Journals
 
Evaluation of operational efficiency of urban road network using travel time ...
Evaluation of operational efficiency of urban road network using travel time ...Evaluation of operational efficiency of urban road network using travel time ...
Evaluation of operational efficiency of urban road network using travel time ...eSAT Journals
 
Estimation of surface runoff in nallur amanikere watershed using scs cn method
Estimation of surface runoff in nallur amanikere watershed using scs cn methodEstimation of surface runoff in nallur amanikere watershed using scs cn method
Estimation of surface runoff in nallur amanikere watershed using scs cn methodeSAT Journals
 
Estimation of morphometric parameters and runoff using rs & gis techniques
Estimation of morphometric parameters and runoff using rs & gis techniquesEstimation of morphometric parameters and runoff using rs & gis techniques
Estimation of morphometric parameters and runoff using rs & gis techniqueseSAT Journals
 
Effect of variation of plastic hinge length on the results of non linear anal...
Effect of variation of plastic hinge length on the results of non linear anal...Effect of variation of plastic hinge length on the results of non linear anal...
Effect of variation of plastic hinge length on the results of non linear anal...eSAT Journals
 
Effect of use of recycled materials on indirect tensile strength of asphalt c...
Effect of use of recycled materials on indirect tensile strength of asphalt c...Effect of use of recycled materials on indirect tensile strength of asphalt c...
Effect of use of recycled materials on indirect tensile strength of asphalt c...eSAT Journals
 

More from eSAT Journals (20)

Mechanical properties of hybrid fiber reinforced concrete for pavements
Mechanical properties of hybrid fiber reinforced concrete for pavementsMechanical properties of hybrid fiber reinforced concrete for pavements
Mechanical properties of hybrid fiber reinforced concrete for pavements
 
Material management in construction – a case study
Material management in construction – a case studyMaterial management in construction – a case study
Material management in construction – a case study
 
Managing drought short term strategies in semi arid regions a case study
Managing drought    short term strategies in semi arid regions  a case studyManaging drought    short term strategies in semi arid regions  a case study
Managing drought short term strategies in semi arid regions a case study
 
Life cycle cost analysis of overlay for an urban road in bangalore
Life cycle cost analysis of overlay for an urban road in bangaloreLife cycle cost analysis of overlay for an urban road in bangalore
Life cycle cost analysis of overlay for an urban road in bangalore
 
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materials
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materialsLaboratory studies of dense bituminous mixes ii with reclaimed asphalt materials
Laboratory studies of dense bituminous mixes ii with reclaimed asphalt materials
 
Laboratory investigation of expansive soil stabilized with natural inorganic ...
Laboratory investigation of expansive soil stabilized with natural inorganic ...Laboratory investigation of expansive soil stabilized with natural inorganic ...
Laboratory investigation of expansive soil stabilized with natural inorganic ...
 
Influence of reinforcement on the behavior of hollow concrete block masonry p...
Influence of reinforcement on the behavior of hollow concrete block masonry p...Influence of reinforcement on the behavior of hollow concrete block masonry p...
Influence of reinforcement on the behavior of hollow concrete block masonry p...
 
Influence of compaction energy on soil stabilized with chemical stabilizer
Influence of compaction energy on soil stabilized with chemical stabilizerInfluence of compaction energy on soil stabilized with chemical stabilizer
Influence of compaction energy on soil stabilized with chemical stabilizer
 
Geographical information system (gis) for water resources management
Geographical information system (gis) for water resources managementGeographical information system (gis) for water resources management
Geographical information system (gis) for water resources management
 
Forest type mapping of bidar forest division, karnataka using geoinformatics ...
Forest type mapping of bidar forest division, karnataka using geoinformatics ...Forest type mapping of bidar forest division, karnataka using geoinformatics ...
Forest type mapping of bidar forest division, karnataka using geoinformatics ...
 
Factors influencing compressive strength of geopolymer concrete
Factors influencing compressive strength of geopolymer concreteFactors influencing compressive strength of geopolymer concrete
Factors influencing compressive strength of geopolymer concrete
 
Experimental investigation on circular hollow steel columns in filled with li...
Experimental investigation on circular hollow steel columns in filled with li...Experimental investigation on circular hollow steel columns in filled with li...
Experimental investigation on circular hollow steel columns in filled with li...
 
Experimental behavior of circular hsscfrc filled steel tubular columns under ...
Experimental behavior of circular hsscfrc filled steel tubular columns under ...Experimental behavior of circular hsscfrc filled steel tubular columns under ...
Experimental behavior of circular hsscfrc filled steel tubular columns under ...
 
Evaluation of punching shear in flat slabs
Evaluation of punching shear in flat slabsEvaluation of punching shear in flat slabs
Evaluation of punching shear in flat slabs
 
Evaluation of performance of intake tower dam for recent earthquake in india
Evaluation of performance of intake tower dam for recent earthquake in indiaEvaluation of performance of intake tower dam for recent earthquake in india
Evaluation of performance of intake tower dam for recent earthquake in india
 
Evaluation of operational efficiency of urban road network using travel time ...
Evaluation of operational efficiency of urban road network using travel time ...Evaluation of operational efficiency of urban road network using travel time ...
Evaluation of operational efficiency of urban road network using travel time ...
 
Estimation of surface runoff in nallur amanikere watershed using scs cn method
Estimation of surface runoff in nallur amanikere watershed using scs cn methodEstimation of surface runoff in nallur amanikere watershed using scs cn method
Estimation of surface runoff in nallur amanikere watershed using scs cn method
 
Estimation of morphometric parameters and runoff using rs & gis techniques
Estimation of morphometric parameters and runoff using rs & gis techniquesEstimation of morphometric parameters and runoff using rs & gis techniques
Estimation of morphometric parameters and runoff using rs & gis techniques
 
Effect of variation of plastic hinge length on the results of non linear anal...
Effect of variation of plastic hinge length on the results of non linear anal...Effect of variation of plastic hinge length on the results of non linear anal...
Effect of variation of plastic hinge length on the results of non linear anal...
 
Effect of use of recycled materials on indirect tensile strength of asphalt c...
Effect of use of recycled materials on indirect tensile strength of asphalt c...Effect of use of recycled materials on indirect tensile strength of asphalt c...
Effect of use of recycled materials on indirect tensile strength of asphalt c...
 

Recently uploaded

MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlysanyuktamishra911
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...Call Girls in Nagpur High Profile
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordAsst.prof M.Gokilavani
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Bookingdharasingh5698
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxpranjaldaimarysona
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...ranjana rawat
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSISrknatarajan
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)simmis5
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...roncy bisnoi
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsRussian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 

Recently uploaded (20)

MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptx
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsRussian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
 

An embedded system for non invasive blood glucose measurement in glucose tolerance test

  • 1. IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308 _______________________________________________________________________________________ Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 63 AN EMBEDDED SYSTEM FOR NON-INVASIVE BLOOD GLUCOSE MEASUREMENT IN GLUCOSE TOLERANCE TEST Indhulekshmi M C1 , Liju Philip2 1 PG Scholar, Department of Electronics Engineering, College of Engineering Chengannur, Kerala, India 2 Assistant Professor, Department of Electronics Engineering, College of Engineering Chengannur, Kerala, India Abstract Diabetic patients need to monitor their blood glucose levels frequently to prevent complications related to the disease. Commonly used glucose measurement methods are invasive, uncomfortable, and usually require frequent finger pricking. This paper deals with the development of a non-invasive system for continuous measurement of blood glucose concentration levels, using near infrared light of suitable wavelength. Glucose determination using near infrared can be done by taking measurement of transmitted light through human body. This device may become very useful while conducting glucose tolerance test (GTT). In glucose tolerance test, patient should be given a measured dose of glucose solution to drink and blood is drawn at intervals of half an hour or one hour for two to four hours. This requires five to nine finger pricking for measurement of glucose concentration. By using this proposed device pricking can be reduced to one. First the glucose concentration is measured using conventional method for calibration. Then by comparing with the first sample measurement subsequent values can be determined without further pricking. Keywords: Non invasive measurement, Near Infrared, Blood Glucose, Transmittance, Diabetes --------------------------------------------------------------------***---------------------------------------------------------------------- 1. INTRODUCTION Diabetes is metabolic disorder in which blood glucose fluctuates from its normal range (70-110mg/dl). Insulin is a hormone produced in body to regulate blood glucose level naturally. Under some pathological failure, body is not able to produce insulin or body cells become unable to use insulin. Thus amount of glucose present in blood exceeds normal range and patient may become diabetic. Nowadays number of diabetic people is increasing across the world due to population growth, unhealthy diet, obesity and lack of physical activity. Diabetic patients need to monitor their blood glucose levels frequently to prevent complications related to the disease. Most of commercially available glucose measurement devices are invasive. The invasive methods are painful, have high recurring cost and danger of spreading infectious diseases. Non-invasive methods are more desirable and excellent alternatives to these devices. Non-invasive glucose monitoring could make millions of people more relaxed and comfortable during blood glucose testing. Thus it is necessary to develop a non-invasive blood glucose measurement system which is painless, convenient and cost effective. Non-invasive monitoring system will be a major breakthrough in the area of treating diabetes patients. Optical methods are painless and promising methods that can be used for non-invasive blood glucose measurement. Near-infrared (NIR) is one of the most widely explored optical techniques because of its high penetration in skin. This technique has been applied on various body parts: finger, palm, arm, forearm, earlobe etc. This work deals with the development of a device for measuring glucose concentration using near infrared light. The proposed device may become very useful while conducting glucose tolerance test (GTT).Glucose tolerance test (GTT) is a medical test which measures how well our body is able to break down glucose. In this test patient is given a measured dose of glucose solution to drink. Then blood samples are taken at specified intervals for the measurement of glucose concentration to determine how quickly it is cleared from the blood. Usually Glucose tolerance test takes about two to four hours. The patient is instructed to fast (water is allowed) for eight to twelve hours prior to the test. First a fasting blood sample is taken and then patient is given a measured dose of glucose solution to drink. Blood is drawn at intervals of half an hour or one hour for two to four hours to determine the glucose concentration. This requires five to nine finger pricking which cause pain and discomfort to patients. By using this proposed device pricking can be reduced to one. First the glucose concentration is measured using conventional method for calibration. Then by comparing with the first sample measurement subsequent values can be determined without further pricking. The goal of this work is to develop a reliable, simple, painless and portable device for glucose measurement that does not cause repetitive puncturing of skin. 2. PROPOSED SYSTEM Near infrared based optical technique has been selected from a number of methods because it is easier to implement and needs less complex hardware setup.The near infrared
  • 2. IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308 _______________________________________________________________________________________ Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 64 (NIR) light in the range of 700nm-2500nm is the most suitable for detecting glucose concentration. The proposed system is based on the principle of transmittance measurement. The value of transmission of light energy is dependent on the number of molecules present in absorbing material. Thus, intensity of light energy transmitted out from the absorbing substance is used as a sign of concentration of that particular substance. Hence by measuring intensity of transmitted light through the measurement site, concentration of glucose in blood can be calculated. Fig-1: Non invasive glucose concentration measurement system This system allows continuous measurement of variation in blood glucose concentration. The instrumentation for optical measurements (sensing unit) consists of an NIR emitter and receiver of 950nm wavelength which is suitable for glucose concentration measurement.NIR rays emitted by the NIR- LED passes through the measurement site and the transmitted rays are received by NIR receiver. After amplification output signal from the sensing unit is given to the microcontroller. The microcontroller has an inbuilt analog to digital converter that converts the analog electrical signal into digital values. The digitized values are then processed in microcontroller to determine the corresponding blood glucose concentration. For proper measurement site selection two factors should be considered: i) The measurement site should be easily accessible ii) NIR light should easily transmit through the measurement site. Hence earlobe or finger web is selected as suitable measurement site. The sensing unit must be fixed in proper position during entire measurement period. It is designed in such a way that the detector is placed exactly opposite and in line to the NIR light source for detecting maximum amount of transmitted light. The light source and detector should be placed as close as possible to the measurement site in physical contact with the skin. 2.1 Operation of the System First glucose concentration is measured using conventional invasive method. After calibrating the proposed device with the first sample measurement, subsequent glucose concentration values are determined without further finger pricking. Thus pricking can be reduced to one while conducting glucose tolerance test. Once the system is turned on, the infrared light propagates through the measurement site. During the propagating phase, NIR light interacts with the molecules present in the medium (glucose in blood). A part of light gets absorbed depending on the glucose concentration of blood under test and remaining part is transmitted through the measurement site. The transmitted signal is then detected by the infrared sensitive detector. After amplification the signal is digitized and is processed in the microcontroller to determine the blood glucose concentration. The measured blood glucose concentration is then displayed on the LCD. The time of measurements and the corresponding value of glucose concentration are stored in memory for further analysis 3. RESULTS AND DISCUSSION The design was tested on seven non-diabetic individuals including both genders. The measurement sites chosen were earlobe and part of skin between thumb & index finger. The near infrared emitter and receiver were positioned exactly opposite to each other to obtain transmitted signal through measurement site. First the proposed device is calibrated with fasting glucose concentration obtained by the invasive method. After meal the signal from the detector is taken every half an hour for two to three hours to determine the variation in glucose concentration. The persons were not allowed to move throughout the measurement period to avoid change in sensor unit positioning in the measurement site. Table-1: Fasting glucose value obtained from invasive laboratory test Person A B C D E F G Fasting glucose(mg/dl) 89 85 90 87 91 100 85 Table-2: Variation in glucose concentration obtained from proposed non invasive device after meal
  • 3. IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308 _______________________________________________________________________________________ Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 65 Tables-1 and 2 given below shows the data collected from each person, during the experiment. Table-1 shows the fasting glucose value obtained from invasive laboratory test.Table-2 shows the variation in glucose concentration obtained from proposed non invasive device after meal. For validation of the experiment, a value obtained from the proposed device is compared with invasive laboratory test result. Two non-diabetic persons were considered for this experiment. At first the device is calibrated using the fasting glucose value which is measured invasively. Then glucose concentration readings of the above two persons after meal, using the proposed device and invasive laboratory test, is taken at an interval of one hour. Data obtained from Person 1 Invasively measured fasting glucose concentration of person 1: 83 mg/dl. Measurement site: finger web Table-3: Data obtained from sensor and invasive laboratory test of person 1 Sl.no Time (min) Non invasive glucose value from sensor (mg/dl) Invasive laboratory test result 1 60 114 98 2 120 143 138 Data obtained from Person 2 Invasively measured fasting glucose concentration of person 2: 85 mg/dl. Measurement site: finger web Table -4: Data obtained from sensor and invasive laboratory test of person 2 Sl.no Time (min) Non invasive glucose value from sensor (mg/dl) Invasive laboratory test result 1 60 91 89 2 120 103 93 Fig-2: Noninvasive and invasive glucose measurement of person 1 Fig-3: Non invasive and invasive glucose measurement of person 2 It was found from the figures 2 and 3, that the values acquired from this proposed device shows an incremental nature as compared with the blood glucose level obtained with invasive laboratory test. Fig-4: Emitter and detector setup Fig-5: Final hardware
  • 4. IJRET: International Journal of Research in Engineering and Technology eISSN: 2319-1163 | pISSN: 2321-7308 _______________________________________________________________________________________ Volume: 04 Issue: 08 | August-2015, Available @ http://www.ijret.org 66 4. CONCLUSION In the present work an embedded system for noninvasive procedure based blood glucose measurement for conducting Glucose Tolerance Test is developed. By using this proposed device pricking can be reduced to one while conducting glucose tolerance test.Experiments are carried out using the proposed non invasive device and the results obtained are comparable with invasive measurement. This device is also useful for continuous glucose measurement of any bedridden or ICU patients. Challenges faced in this work are: i) The skin transmittance is different for different persons depending on the colour, texture of the skin and thickness of measurement site. So the proposed device needs to be calibrated for each person before conducting the test. ii) The sensing unit must be properly positioned and not allowed to move during the entire measurement period. In future the performance of the proposed system can be improved by adopting differential measurement in which the difference between incident light and transmitted light from skin is used for calculating glucose concentration. REFERENCES [1]. Omar S. Khalil, “Spectroscopic and Clinical Aspects of Noninvasive Glucose Measurements”, Clinical Chemistry 45:2 165–177 (1999) [2]. JyotiYadav, Asha Rani, Vijander Singh, Bhaskar Mohan Murari, “Near-infrared LED based Non-invasive Blood Glucose Sensor”, 2014 international conference on Signal Processing and Integrated Networks (SPIN). [3]. Ola S. Abdalsalam, Al-khanssa M. Osman, Roua M. Abd-Alhadi, Saad D. Alshmaa, “ Design of Simple Noninvasive Glucose Measuring Device”,2013 International Conference on Computing, Electrical and Electronic Engineering (ICCEEE) [4]. A.G.Pavithra, D.MeneshaKaran,D. Ajith Kumar, P .S. AnuShalin ,”Non invasive technique to measure glucose and haemoglobin level in blood using NIR spectroscopy”,International Journal of scientific research and management.Volume2,Issue4,2013