Wind Tunnel High Speed Test of
Gill Windobserver Ultrasonic Anemometer
Test Item: Gill Ultrasonic Anemometer
Type: Windobserver 1390-90-B-121
SN 152199
Contracted by:
Gill Instruments Ltd
Saltmarsh Park
67 Gosport Street
Lymington
Hampshire
SO41 9EG UK
Deutsche WindGuard
Wind Tunnel Services GmbH
Oldenburger Str. 65
26316 Varel
Project No.: VT150551
Report No.: AK150551.02
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 2 of 14
Wind tunnel high speed test of Gill Windobserver 1390-90-B-121; SN 152199
at 5, 12, 20, 32, 45, 60, 75, 85, 90 m/s for sensor yaw orientation of 45°
plus yaw seep test run at 45 m/s
Deutsche WindGuard Wind Tunnel Services GmbH
Oldenburger Str. 65
26316 Varel
Tel: 04451-9515-0
Email: d.westermann@windguard.de
Contracted by:
Gill Instruments Ltd
Saltmarsh Park
67 Gosport Street
Lymington
Hampshire
SO41 9EG UK
Contact: Richard McKay
Product Manager
Project No.: VT150551
Report No.: AK150551.02
Version: 2
Location, Date: Varel, Jun.22nd 2015
Author: Dipl.- Ing. (FH) P. Busche
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 3 of 14
Results of high speed anemometer test performed at the wind tunnel of Deutsche WindGuard
Engineering located at Überseering 7, D-27580 Bremerhaven . Test section size 1.25 X 2.7 m
across
Test date: May 29th .2015
Test run number: 4859
Type Gill Windobserver 1390-90-B-121
Serial number 152199
Orientation 45° + yaw sweep test at 45 m/s
Test result tunnel speed 5 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 5m/s tunnel speed.
Note: Due to the nature of the test facility (high speed tunnel) the relative tunnel uncertainty at
“low” wind speeds will be more pronounced compared to rel. uncertainties at higher tunnel
speeds!
Ambient conditions throughout 5 m/s test run:
Baro. pressure test section min. 1006.9 hPa max. 1007.3 hPa avg. 1007.1 hPa
Temp. test section min. 24.15 °C max. 24.21 °C avg. 24.18 °C
rel. Humidity test section min. 46.1 % max. 46.3 % avg. 46.2 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 4 of 14
Test result tunnel speed 12 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 12m/s tunnel speed.
Note: Due to the nature of the test facility (high speed tunnel) the relative tunnel uncertainty at
“low” wind speeds will be more pronounced compared to rel. uncertainties at higher tunnel
speeds!
Ambient conditions throughout 12 m/s test run:
Baro. pressure test section min. 1006.5 hPa max. 1007.0 hPa avg. 1006.7 hPa
Temp. test section min. 24.37 °C max. 24.46 °C avg. 24.41 °C
rel. Humidity test section min. 45.5 % max. 45.7 % avg. 45.7 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 5 of 14
Test result tunnel speed 20 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 20 m/s tunnel speed.
Ambient conditions throughout 20 m/s test run:
Baro. pressure test section min. 1005.3 hPa max. 1006.0 hPa avg. 1005.7 hPa
Temp. test section min. 24.57 °C max. 24.70 °C avg. 24.63 °C
rel. Humidity test section min. 44.9 % max. 45.1% avg. 45.0 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 6 of 14
Test result tunnel speed 32 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 32 m/s tunnel speed.
Ambient conditions throughout 32 m/s test run:
Baro. pressure test section min. 1002.9 hPa max. 1003.8 hPa avg. 1003.3 hPa
Temp. test section min. 24.86 °C max. 25.01°C avg. 24.93°C
rel. Humidity test section min. 43.9 % max. 44.1% avg. 44.0 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 7 of 14
Test result tunnel speed 45 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 45 m/s tunnel speed.
Ambient conditions throughout 45 m/s test run:
Baro. pressure test section min. 998.8 hPa max. 999.9 hPa avg. 999.3 hPa
Temp. test section min. 26.54 °C max. 26.83 °C avg. 26.67 °C
rel. Humidity test section min. 39.9 % max. 40.3 % avg. 40.1 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 8 of 14
Test result tunnel speed 60 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 60 m/s tunnel speed.
Ambient conditions throughout 60 m/s test run:
Baro. pressure test section min. 992.8 hPa max. 994.3 hPa avg. 993.6 hPa
Temp. test section min. 26.66 °C max. 26.84 °C avg. 26.75 °C
rel. Humidity test section min. 38.6 % max. 39.0 % avg. 38.8 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 9 of 14
Test result tunnel speed 75 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 75 m/s tunnel speed.
Ambient conditions throughout 75 m/s test run:
Baro. pressure test section min. 983.9 hPa max. 987.0 hPa avg. 985.6 hPa
Temp. test section min. 25.81 °C max. 26.07 °C avg. 25.95 °C
rel. Humidity test section min. 38.3 % max. 38.4 % avg. 38.4 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 10 of 14
Test result tunnel speed 85 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 85 m/s tunnel speed.
Ambient conditions throughout 85 m/s test run:
Baro. pressure test section min. 977.4 hPa max. 981.7 hPa avg. 979.7 hPa
Temp. test section min. 24.34 °C max. 24.83 °C avg. 24.65 °C
rel. Humidity test section min. 38.9 % max. 39.3% avg. 39.1 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 11 of 14
Test result tunnel speed 90 m/s:
Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill
Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 90 m/s tunnel speed.
Ambient conditions throughout 90 m/s test run:
Baro. pressure test section min. 973.7 hPa max. 977.8 hPa avg. 976.2 hPa
Temp. test section min. 17.05 °C max. 17.93 °C avg. 17.51 °C
rel. Humidity test section min. 48.9 % max. 50.4% avg. 49.7 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 12 of 14
Test result of yaw sweep at a tunnel speed of 45 m/s:
Graph showing the flow speed deviation and directional deviation between anemometer wind
signal of Gill Windobserver 1390-90-B-121; SN 152199 and tunnel signal during yaw sweep at
45 m/s tunnel speed; yaw rate aprox. 1.3 °/s
Ambient conditions throughout yaw sweep run at 45 m/s :
Baro. pressure test section min. 997.4 hPa max. 999.9 hPa avg. 998.5 hPa
Temp. test section min. 25.64 °C max. 26.28 °C avg. 25.94 °C
rel. Humidity test section min. 40.8 % max. 41.8 % avg. 41.3 %
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 13 of 14
Picture of a Gill Windobserver 1390 of similar type to 1390-90-B-121; SN 152199, installed at
the turn table of wind tunnel Bremerhaven ; yaw orientation 0°. View into the wind tunnel
nozzle (view into the flow).
High Speed Anemometer Wind Tunnel Test
Deutsche WindGuard
Wind Tunnel Services GmbH, Varel 14 of 14
Lateral view of the Gill Windobserver 1390-90-B-121; SN 152199 installed at the turn table of
wind tunnel Bremerhaven; pictured at a yaw orientation of 0°(flow from right to left)
Close-up of the Gill Windobserver 1390-90-B-121; SN 152199 and mounting bracket installed at
wind tunnel Bremerhaven

More Related Content

What's hot

Overcoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration WebinarOvercoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration Webinarrmelnyk1
 
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015Transcat
 
How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2Transcat
 
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Transcat
 
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow MeterEmerson Exchange
 
Notes on validation
Notes on validationNotes on validation
Notes on validationDhruv Desai
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationTranscat
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationTranscat
 
Labthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink
 
Frequency of calibration
Frequency of calibrationFrequency of calibration
Frequency of calibrationECRD IN
 
Oil burner servicing, combustion efficiency testing cbt 2
Oil burner servicing, combustion efficiency testing cbt 2Oil burner servicing, combustion efficiency testing cbt 2
Oil burner servicing, combustion efficiency testing cbt 2dbartley123
 
Adev triple range hydrogen on analysis hydrogen cooled generators
Adev triple range hydrogen on analysis hydrogen cooled generatorsAdev triple range hydrogen on analysis hydrogen cooled generators
Adev triple range hydrogen on analysis hydrogen cooled generatorsLorenzo Novati
 
Fluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure CalibrationFluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure CalibrationTranscat
 
An overview of diagnostic tools used in RESNET testing
An overview of diagnostic tools used in RESNET testingAn overview of diagnostic tools used in RESNET testing
An overview of diagnostic tools used in RESNET testingBill Spohn
 

What's hot (20)

Thermal Mass Flowmeter (ABB N.V.)
Thermal Mass Flowmeter (ABB N.V.)Thermal Mass Flowmeter (ABB N.V.)
Thermal Mass Flowmeter (ABB N.V.)
 
IDEAL GAS LAB REPORT
IDEAL GAS LAB REPORTIDEAL GAS LAB REPORT
IDEAL GAS LAB REPORT
 
Overcoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration WebinarOvercoming Challenges in Pressure Calibration Webinar
Overcoming Challenges in Pressure Calibration Webinar
 
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
 
How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2How to Calibrate an RTD Using a Dry Block Calibrator v2
How to Calibrate an RTD Using a Dry Block Calibrator v2
 
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
Fluke Calibration on How to Calibrate an RTD Using a Dry block Calibrator Web...
 
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter
40 CFR 98 Tier 4 Greenhouse Gas Reporting with Stack Emissions Flow Meter
 
Notes on validation
Notes on validationNotes on validation
Notes on validation
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke Calibration
 
Throttling calorimeter
Throttling calorimeterThrottling calorimeter
Throttling calorimeter
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer Calibration
 
Labthink Gas Permeation Tester
Labthink Gas Permeation TesterLabthink Gas Permeation Tester
Labthink Gas Permeation Tester
 
Frequency of calibration
Frequency of calibrationFrequency of calibration
Frequency of calibration
 
Oil burner servicing, combustion efficiency testing cbt 2
Oil burner servicing, combustion efficiency testing cbt 2Oil burner servicing, combustion efficiency testing cbt 2
Oil burner servicing, combustion efficiency testing cbt 2
 
Adev triple range hydrogen on analysis hydrogen cooled generators
Adev triple range hydrogen on analysis hydrogen cooled generatorsAdev triple range hydrogen on analysis hydrogen cooled generators
Adev triple range hydrogen on analysis hydrogen cooled generators
 
Fluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure CalibrationFluke Calibration Tips for High Pressure Calibration
Fluke Calibration Tips for High Pressure Calibration
 
Air Duct Leakage
Air Duct LeakageAir Duct Leakage
Air Duct Leakage
 
Draught and chimney
Draught and chimneyDraught and chimney
Draught and chimney
 
Final presentation
Final presentationFinal presentation
Final presentation
 
An overview of diagnostic tools used in RESNET testing
An overview of diagnostic tools used in RESNET testingAn overview of diagnostic tools used in RESNET testing
An overview of diagnostic tools used in RESNET testing
 

Similar to Wind observer 90 windguard test report

S4 oman wind energy potential anemometers 2016
S4 oman wind energy potential   anemometers 2016S4 oman wind energy potential   anemometers 2016
S4 oman wind energy potential anemometers 2016CETN
 
S4 oman wind energy potential - anemometers 2016
S4 oman wind energy potential - anemometers 2016S4 oman wind energy potential - anemometers 2016
S4 oman wind energy potential - anemometers 2016CETN
 
Hydrotest REPORT FOR 4'' 158m Line.
Hydrotest REPORT FOR 4'' 158m Line.Hydrotest REPORT FOR 4'' 158m Line.
Hydrotest REPORT FOR 4'' 158m Line.Emeka Ngwobia
 
fanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdffanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdfBlentBulut5
 
Handheld-Anemometer
Handheld-AnemometerHandheld-Anemometer
Handheld-AnemometerAlizaMorris
 
Edwards Signaling SD2W Installation Manual
Edwards Signaling SD2W Installation ManualEdwards Signaling SD2W Installation Manual
Edwards Signaling SD2W Installation ManualJMAC Supply
 
Beebe steam turbines cm apr2016 r1
Beebe steam turbines cm apr2016 r1Beebe steam turbines cm apr2016 r1
Beebe steam turbines cm apr2016 r1Ray Beebe
 
Final Group Design Presentation 03 06 10
Final Group Design Presentation 03 06 10Final Group Design Presentation 03 06 10
Final Group Design Presentation 03 06 10m_phull
 
Performance Evaluation of Synthetic Jet Cooling For CPU
Performance Evaluation of Synthetic Jet Cooling For CPUPerformance Evaluation of Synthetic Jet Cooling For CPU
Performance Evaluation of Synthetic Jet Cooling For CPUIRJET Journal
 
Bulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMBulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMTiago Morais
 
Airflow measurement
Airflow measurement Airflow measurement
Airflow measurement BSRIA
 

Similar to Wind observer 90 windguard test report (20)

S4 oman wind energy potential anemometers 2016
S4 oman wind energy potential   anemometers 2016S4 oman wind energy potential   anemometers 2016
S4 oman wind energy potential anemometers 2016
 
S4 oman wind energy potential - anemometers 2016
S4 oman wind energy potential - anemometers 2016S4 oman wind energy potential - anemometers 2016
S4 oman wind energy potential - anemometers 2016
 
Hydrotest REPORT FOR 4'' 158m Line.
Hydrotest REPORT FOR 4'' 158m Line.Hydrotest REPORT FOR 4'' 158m Line.
Hydrotest REPORT FOR 4'' 158m Line.
 
S-TU221-TA-ENG
S-TU221-TA-ENGS-TU221-TA-ENG
S-TU221-TA-ENG
 
Conductivity Sensor for High-Temperature Applications W CLS 12
Conductivity Sensor for High-Temperature Applications W CLS 12Conductivity Sensor for High-Temperature Applications W CLS 12
Conductivity Sensor for High-Temperature Applications W CLS 12
 
fanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdffanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdf
 
FAN MODULE.pdf
FAN MODULE.pdfFAN MODULE.pdf
FAN MODULE.pdf
 
Handheld-Anemometer
Handheld-AnemometerHandheld-Anemometer
Handheld-Anemometer
 
Edwards Signaling SD2W Installation Manual
Edwards Signaling SD2W Installation ManualEdwards Signaling SD2W Installation Manual
Edwards Signaling SD2W Installation Manual
 
Beebe steam turbines cm apr2016 r1
Beebe steam turbines cm apr2016 r1Beebe steam turbines cm apr2016 r1
Beebe steam turbines cm apr2016 r1
 
pipe friction for turbulent
pipe friction for turbulentpipe friction for turbulent
pipe friction for turbulent
 
pipe friction for turbulent
pipe friction for turbulentpipe friction for turbulent
pipe friction for turbulent
 
Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13
 
Final Group Design Presentation 03 06 10
Final Group Design Presentation 03 06 10Final Group Design Presentation 03 06 10
Final Group Design Presentation 03 06 10
 
Performance Evaluation of Synthetic Jet Cooling For CPU
Performance Evaluation of Synthetic Jet Cooling For CPUPerformance Evaluation of Synthetic Jet Cooling For CPU
Performance Evaluation of Synthetic Jet Cooling For CPU
 
Bulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMBulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAM
 
Airflow measurement
Airflow measurement Airflow measurement
Airflow measurement
 
Siemens Air velocity sensor
Siemens Air velocity sensorSiemens Air velocity sensor
Siemens Air velocity sensor
 
PPT-on-false-air.ppt
PPT-on-false-air.pptPPT-on-false-air.ppt
PPT-on-false-air.ppt
 
2392-3449
2392-34492392-3449
2392-3449
 

Recently uploaded

GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationMichael W. Hawkins
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationRidwan Fadjar
 
Next-generation AAM aircraft unveiled by Supernal, S-A2
Next-generation AAM aircraft unveiled by Supernal, S-A2Next-generation AAM aircraft unveiled by Supernal, S-A2
Next-generation AAM aircraft unveiled by Supernal, S-A2Hyundai Motor Group
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsMemoori
 
How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonetsnaman860154
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024BookNet Canada
 
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure service
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure serviceWhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure service
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure servicePooja Nehwal
 
The Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxThe Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxMalak Abu Hammad
 
Artificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraArtificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraDeakin University
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions
 
Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountPuma Security, LLC
 
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxMaking_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxnull - The Open Security Community
 
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your Budget
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your BudgetHyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your Budget
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your BudgetEnjoy Anytime
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticscarlostorres15106
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking MenDelhi Call girls
 
Slack Application Development 101 Slides
Slack Application Development 101 SlidesSlack Application Development 101 Slides
Slack Application Development 101 Slidespraypatel2
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsMark Billinghurst
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j
 

Recently uploaded (20)

GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day Presentation
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 Presentation
 
Next-generation AAM aircraft unveiled by Supernal, S-A2
Next-generation AAM aircraft unveiled by Supernal, S-A2Next-generation AAM aircraft unveiled by Supernal, S-A2
Next-generation AAM aircraft unveiled by Supernal, S-A2
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial Buildings
 
How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonets
 
Vulnerability_Management_GRC_by Sohang Sengupta.pptx
Vulnerability_Management_GRC_by Sohang Sengupta.pptxVulnerability_Management_GRC_by Sohang Sengupta.pptx
Vulnerability_Management_GRC_by Sohang Sengupta.pptx
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
 
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure service
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure serviceWhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure service
WhatsApp 9892124323 ✓Call Girls In Kalyan ( Mumbai ) secure service
 
The Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxThe Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptx
 
Artificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning eraArtificial intelligence in the post-deep learning era
Artificial intelligence in the post-deep learning era
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping Elbows
 
Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path Mount
 
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptxMaking_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
Making_way_through_DLL_hollowing_inspite_of_CFG_by_Debjeet Banerjee.pptx
 
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your Budget
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your BudgetHyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your Budget
Hyderabad Call Girls Khairatabad ✨ 7001305949 ✨ Cheap Price Your Budget
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men
 
Slack Application Development 101 Slides
Slack Application Development 101 SlidesSlack Application Development 101 Slides
Slack Application Development 101 Slides
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR Systems
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
 

Wind observer 90 windguard test report

  • 1. Wind Tunnel High Speed Test of Gill Windobserver Ultrasonic Anemometer Test Item: Gill Ultrasonic Anemometer Type: Windobserver 1390-90-B-121 SN 152199 Contracted by: Gill Instruments Ltd Saltmarsh Park 67 Gosport Street Lymington Hampshire SO41 9EG UK Deutsche WindGuard Wind Tunnel Services GmbH Oldenburger Str. 65 26316 Varel Project No.: VT150551 Report No.: AK150551.02
  • 2. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 2 of 14 Wind tunnel high speed test of Gill Windobserver 1390-90-B-121; SN 152199 at 5, 12, 20, 32, 45, 60, 75, 85, 90 m/s for sensor yaw orientation of 45° plus yaw seep test run at 45 m/s Deutsche WindGuard Wind Tunnel Services GmbH Oldenburger Str. 65 26316 Varel Tel: 04451-9515-0 Email: d.westermann@windguard.de Contracted by: Gill Instruments Ltd Saltmarsh Park 67 Gosport Street Lymington Hampshire SO41 9EG UK Contact: Richard McKay Product Manager Project No.: VT150551 Report No.: AK150551.02 Version: 2 Location, Date: Varel, Jun.22nd 2015 Author: Dipl.- Ing. (FH) P. Busche
  • 3. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 3 of 14 Results of high speed anemometer test performed at the wind tunnel of Deutsche WindGuard Engineering located at Überseering 7, D-27580 Bremerhaven . Test section size 1.25 X 2.7 m across Test date: May 29th .2015 Test run number: 4859 Type Gill Windobserver 1390-90-B-121 Serial number 152199 Orientation 45° + yaw sweep test at 45 m/s Test result tunnel speed 5 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 5m/s tunnel speed. Note: Due to the nature of the test facility (high speed tunnel) the relative tunnel uncertainty at “low” wind speeds will be more pronounced compared to rel. uncertainties at higher tunnel speeds! Ambient conditions throughout 5 m/s test run: Baro. pressure test section min. 1006.9 hPa max. 1007.3 hPa avg. 1007.1 hPa Temp. test section min. 24.15 °C max. 24.21 °C avg. 24.18 °C rel. Humidity test section min. 46.1 % max. 46.3 % avg. 46.2 %
  • 4. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 4 of 14 Test result tunnel speed 12 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 12m/s tunnel speed. Note: Due to the nature of the test facility (high speed tunnel) the relative tunnel uncertainty at “low” wind speeds will be more pronounced compared to rel. uncertainties at higher tunnel speeds! Ambient conditions throughout 12 m/s test run: Baro. pressure test section min. 1006.5 hPa max. 1007.0 hPa avg. 1006.7 hPa Temp. test section min. 24.37 °C max. 24.46 °C avg. 24.41 °C rel. Humidity test section min. 45.5 % max. 45.7 % avg. 45.7 %
  • 5. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 5 of 14 Test result tunnel speed 20 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 20 m/s tunnel speed. Ambient conditions throughout 20 m/s test run: Baro. pressure test section min. 1005.3 hPa max. 1006.0 hPa avg. 1005.7 hPa Temp. test section min. 24.57 °C max. 24.70 °C avg. 24.63 °C rel. Humidity test section min. 44.9 % max. 45.1% avg. 45.0 %
  • 6. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 6 of 14 Test result tunnel speed 32 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 32 m/s tunnel speed. Ambient conditions throughout 32 m/s test run: Baro. pressure test section min. 1002.9 hPa max. 1003.8 hPa avg. 1003.3 hPa Temp. test section min. 24.86 °C max. 25.01°C avg. 24.93°C rel. Humidity test section min. 43.9 % max. 44.1% avg. 44.0 %
  • 7. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 7 of 14 Test result tunnel speed 45 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 45 m/s tunnel speed. Ambient conditions throughout 45 m/s test run: Baro. pressure test section min. 998.8 hPa max. 999.9 hPa avg. 999.3 hPa Temp. test section min. 26.54 °C max. 26.83 °C avg. 26.67 °C rel. Humidity test section min. 39.9 % max. 40.3 % avg. 40.1 %
  • 8. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 8 of 14 Test result tunnel speed 60 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 60 m/s tunnel speed. Ambient conditions throughout 60 m/s test run: Baro. pressure test section min. 992.8 hPa max. 994.3 hPa avg. 993.6 hPa Temp. test section min. 26.66 °C max. 26.84 °C avg. 26.75 °C rel. Humidity test section min. 38.6 % max. 39.0 % avg. 38.8 %
  • 9. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 9 of 14 Test result tunnel speed 75 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 75 m/s tunnel speed. Ambient conditions throughout 75 m/s test run: Baro. pressure test section min. 983.9 hPa max. 987.0 hPa avg. 985.6 hPa Temp. test section min. 25.81 °C max. 26.07 °C avg. 25.95 °C rel. Humidity test section min. 38.3 % max. 38.4 % avg. 38.4 %
  • 10. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 10 of 14 Test result tunnel speed 85 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 85 m/s tunnel speed. Ambient conditions throughout 85 m/s test run: Baro. pressure test section min. 977.4 hPa max. 981.7 hPa avg. 979.7 hPa Temp. test section min. 24.34 °C max. 24.83 °C avg. 24.65 °C rel. Humidity test section min. 38.9 % max. 39.3% avg. 39.1 %
  • 11. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 11 of 14 Test result tunnel speed 90 m/s: Graph showing the wind tunnel flow speed and the anemometer wind speed signal of Gill Windobserver 1390-90-B-121; SN 152199; yaw orientation 45° at 90 m/s tunnel speed. Ambient conditions throughout 90 m/s test run: Baro. pressure test section min. 973.7 hPa max. 977.8 hPa avg. 976.2 hPa Temp. test section min. 17.05 °C max. 17.93 °C avg. 17.51 °C rel. Humidity test section min. 48.9 % max. 50.4% avg. 49.7 %
  • 12. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 12 of 14 Test result of yaw sweep at a tunnel speed of 45 m/s: Graph showing the flow speed deviation and directional deviation between anemometer wind signal of Gill Windobserver 1390-90-B-121; SN 152199 and tunnel signal during yaw sweep at 45 m/s tunnel speed; yaw rate aprox. 1.3 °/s Ambient conditions throughout yaw sweep run at 45 m/s : Baro. pressure test section min. 997.4 hPa max. 999.9 hPa avg. 998.5 hPa Temp. test section min. 25.64 °C max. 26.28 °C avg. 25.94 °C rel. Humidity test section min. 40.8 % max. 41.8 % avg. 41.3 %
  • 13. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 13 of 14 Picture of a Gill Windobserver 1390 of similar type to 1390-90-B-121; SN 152199, installed at the turn table of wind tunnel Bremerhaven ; yaw orientation 0°. View into the wind tunnel nozzle (view into the flow).
  • 14. High Speed Anemometer Wind Tunnel Test Deutsche WindGuard Wind Tunnel Services GmbH, Varel 14 of 14 Lateral view of the Gill Windobserver 1390-90-B-121; SN 152199 installed at the turn table of wind tunnel Bremerhaven; pictured at a yaw orientation of 0°(flow from right to left) Close-up of the Gill Windobserver 1390-90-B-121; SN 152199 and mounting bracket installed at wind tunnel Bremerhaven