New Fifth Generation Tunable Multi-Channel LED light source for sensor testing and calibration based on over 16 years of scientific light emitting diode engineering application experience
2. Gamma Scientific has been
producing LED tunable light
sources for over 16 years
Digital Tunable LED Calibration Light
Source product generations:
First stable optical feedback controlled,
intensity tunable LED source with
Patented technology, RS-5, 1998
The first integrating sphere 4 color LED
tunable source in 1999
8 separate color channel tunable LED
calibration source, RS-5M in 2002
10 separate color channel RS-5B
developed in 2005
3. Gamma Scientific has been
producing LED tunable light
sources for over 16 years
Digital Tunable LED Calibration Light
Source product generations:
RS-5-93K for Agilent tester in 2005
Four RS-5B coupled to a 1 meter
sphere for vision research, 2008
RS-5H high output 8 channel CMOS
image sensor test source in 2007
Current generation RS-7 containing 35
channels and over 500 watts of LED
power.
4. Uniform Flat-Field Source for
Camera and Image Sensor
Test and Calibration
Applications:
Camera and Image Sensor
Calibration (CCD, CMOS, etc.)
Ambient Light Sensor Calibration
OEM Camera Manufacturing
Spectrum/Illuminant Simulation
Diagnostic Medical Imaging
Technical and Industrial
Photography
Cinematography
Copier/Scanner Calibration
5. Powerful Tool for
Production or R&D
Measurement Examples:
White Balance
Quantum Efficiency
Spatial Non-Uniformity
Pixel Defects
Cross Talk
Vignetting Correction
Sensitivity
Responsivity
Signal-to-Noise
Linearity
ISO Speed
Saturation Exposure
Dynamic Range
6. The SpectralLED Advantage
Simplify the calibration process
Improve accuracy
Reduce cycle time
Quickly simulate a huge variety of
traditional light sources
Select a preset CIE Illuminant or
Macbeth™ / X-RITE™ color patch in
milliseconds
Produce a Macbeth™ / X-RITE™
color patch under any CIE illuminant
Import your own spectrum
Let the internal RMS spectral fitting
find the best match instantly.
7. Highest Resolution
Tunable Light Source
35 Discrete Wavelength Channels
380 to 1000 nm range
Nearly continuous simulation of
broadband spectra
Monochromatic outputs
Linear 16-bit digital control on each
channel
Closed-loop feedback for stability
8. Extremely High Power With
Wide Dynamic Range
Over 500 Watts of LED Power
Max Radiance levels (D65) > 30,000 uW/cm2/sr
Max Luminance levels (D65) > 60,000 cd/m2
More than 3 decades of dynamic range
(spectrum dependent)
ND filter and computer controlled IRIS options
available
Standard 75mm exit port provides >97%
uniformity
9. Accuracy and Stability
ISO/IEC 17025 NVLAP Calibration Laboratory (Lab Code 200823-0)
- Ensures Superior Wavelength and Color Accuracy
Full Calibration of Each LED Channel at Multiple Drive Levels
NIST Equivalent Calibration ensures Superior Wavelength and
Color Accuracy
Pure DC Constant Current Drivers and Built-in Optical Feedback
Photodiode System – Ensure Accurate Output in Real Time
On-board Thermal Electric Cooler and Feedback for LED
Temperature Control
10. SpectralLED Synthesis Engine
Built-in RMS Spectral Fitting for Simulation of Any User Imported Target
Spectrum - blackbody, daylight, fluorescent, LED
Quickly Simulate any CIE Illuminant or Macbeth™ / X-RITE™ Color Patch
Convolve standard illuminants with color patches
Flexible Interface – GUI or serial control
Configurable – minimize RMS or ∆E
12. Standard Configurations
RS-7-1 Integrating Sphere Configuration for Radiance Geometries
RS-7-2 Baffle Tube Configuration for Irradiance Geometries
RS-7-1 Can Be Used as Irradiance Source for Small Areas
RS-7-2 Includes Modular Baffle Tube (f/1 to f/8 adjustment)
Additional Baffle Tube Segments Available
RS-7-1 RS-7-2
13. Optional Wavelength Monitor
Optional Hardware System to Measure Long
Term Wavelength Drift of Individual LED
Channels
Eliminates Need for External
Spectroradiometer at Lower Cost
Multi-Channel Photodiode Feedback for
Amplitude and Wavelength Data
Uses Internal Cal Data for Extreme Accuracy
and Repeatability (<0.1nm)
Extends Calibration Interval
Monitor Data Available to External Systems
14. Custom Configurations
Available
Fiber Optic Coupled Outputs (Easily
Adaptable for Test Systems)
Custom LED wavelengths
ND Filters or Computer Controlled IRIS
Larger Port Diameters
Higher f-number Systems
Multiple LED Engines for Higher Output