Page 1 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 System
Training
Liliana Escobar Diab
Support Engineer
Effective Date: marzo 2015
Page 2 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Course Overview
Este curso tiene por objeto definir las diferencias de software y hardware entre los
sistemas de gases RAPIDPoint 405 y el nuevo RAPIDPoint 500.
RAPIDPoint 405 RAPIDPoint 500
Page 3 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Product Overview
Installation
Components
Routine Operation
Screen Navigation
 Status Screen
 Instrument Setup
 Recall Screen
Hardware
Routine System Procedures
Network and Connectivity
Patient Sample Analysis
QC Analysis
Calibrations
Troubleshooting
Cartridge Overview
Q&A
Agenda
Page 4 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Our Blood Gas History…..
The 60’s
& 70’s
175
165
161
pH
The
80’s
278/270
178/2500 HP85
The 90’s
and to date
Rapidlab 800
Rapidlab
238/248/348
Rapidlab 1200
Rapidpoint 400 Rapidlink &
RapidComm
Protection notice / Copyright notice
RAPIDPoint 500
Product Overview
Page 6 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 System Intended Use Statement
El sistema de Diagnostico Siemens RAPIDPoint
500 esta diseñado para uso profesional en un
entorno de Cuidado Crítico o de Laboratorio.
Este sistema realiza pruebas de: gases en
sangre, electrolitos , metabolitos,
hemoglobina total y derivados de la
hemoglobina en muestras de sangre total
arterial, venosa y capilar.
Al proporcionar resultados oportunos, el
sistema permite al personal medico tomar
decisiones mas rápidas sobre el tratamiento
del paciente y mejorar la calidad de la
atención del paciente en:
 Quirófanos y recuperación critica.
 Unidad de cuidados Intensivos.
 Salas de Emergencia.
Page 7 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Selling Points
• Resultados rápidos y disponibles a través de
un sistema de POC
• Facilidad de uso
• Control de Calidad Automático
• Amplio menú de analitos críticos.
• Soluciones de conectividad Únicos.
• Máximo control del sistema incluyendo
bloqueo.
• Diseñado para ser utilizado en zona cercana
al paciente.
Page 8 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Specifications
Page 9 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Specifications
Blood Gases
Electrolytes / Metabolites
Co-Ox Parameters
Page 10 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Unparalleled Ease of Use
• Muestreo automático con detección de
coágulos y burbujas.
• Lector de código de barras Integrado
• Pantalla touch screen a color
• Entrenamiento On-line con videos y
audios de ayuda.
• Impresora a bordo y escáner de código
de barras opcional
• Automatic QC
• Conectividad a través RAPIDComm
Page 11 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 System - Simple to live with!
• Sistemas RAPIDPoint 500 son libres de
mantenimiento completamente.
• Cartuchos de larga duración
• Solo deben mantener en stock 4
números de parte vs. ~30 en los
sistemas tradicionales.
• Ningún procedimiento programado de
limpieza o procedimientos de
mantenimiento de sensores.
Page 12 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 System
• Sistema basado en Cartuchos
Multiuso.
• Menú
– pH, pCO2, pO2
– Na+
, K+
, Ca++
, Cl-
– Glucosa, Lactato
– tHb, CO-ox & nBili**
• Barcode Scanner Integrado
• CO-oximetria
• Control de Calidad Automático
• Guías en Audio/Video
**Not available for sale in US.
Broad Menu of Analytes
Page 13 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Maximum System Control
• Provee seguridad y confiabilidad.
• Analitos seleccionables dependiendo del
muestreo.
• Protección del sistema por password
• Control de seguridad Multi- nivel
• Bloqueo opcional de analitos si falla el control
de calidad.
• Estado remoto de monitoreo y control
Remoto vía RAPIDComm (full remote control)
500
405
Page 14 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Connectivity via RAPIDComm version 4.0
Diseñado para el manejo completo de información de pacientes y el
sistema:
• Monitoreo de los analizadores en tiempo real a
través de la red hospitalaria.
• Control centralizado para máxima eficiencia
• Conocimiento de todos los requerimientos
regulatorios.
• Conexión al LIS y HIS.
Page 15 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ventajas
Caracteristicas
 Cartuchos de medida sin
mantenimiento.
 Exposición por bioseguridad se
reduce
 Los cartuchos son remplazados en
sencillos pasos.
Los sistemas usan cartuchos de
medida que contienen todos los
reactivos y sensores y cartuchos de
lavado /desechos que son
reemplazados cuando se llenan.
 Los sistemas pueden ser ubicados
en un ambiente con limitaciones de
espacio y pueden ser reubicados
fácilmente.
Los sistemas usan un cartucho
compacto, basado en un diseño sin
botellas de reactivo ni tanques de
gas externos.
Features & Benefits
Page 16 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ventajas
Características
 El operador introduce la muestra
consistentemente con el volumen
adecuado, eliminando variabilidad
en los resultados por la técnica de
muestreo.
El sistema automáticamente aspira
las muestras.
 El cliente puede maximizar la
cantidad de tiempo con los
pacientes y no con el sistema.
El sistema calibra automáticamente
los sensores de medida.
Features & Benefits
Page 17 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ventajas
Características
 El operador puede analizar muestras
y moverse por el software con
facilidad.
 Un operador con bajo o cero
conocimiento en manejo de
computadores puede fácilmente
aprender como usar los sistemas.
El sistema presenta la información
esencial de una forma clara,
incluyendo videos y diagramas que
muestran paso a paso las acciones
para tareas clave.
 Operador puede ver la información
y seleccionar fácil y rápidamente.
El sistema usa una pantalla touch
screen a color que permite
visualizar claramente.
Features & Benefits
Page 18 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ventajas
Características
 No hay perdida de cartuchos a causa de coágulos.
 Menos riesgo de exposición por bioseguridad.
Manejo de
Coágulos
 No hay tiempo muerto relacionado con
mantenimiento de electrodos , limpieza manual y
automática, o procedimientos o técnicas que
requieran mantenimiento.
 No requiere calibraciones ni controles de calidad
adicionales posteriores a una rutina de
mantenimiento.
Libre de
Mantenimiento
Features & Benefits
Page 19 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ventajas
Características
 Requisitos reglamentarios
 Proporciona un medio para adminsitrar de forma remota
los sitemas desde una ubicacion central
Información fácil de
comprender y
manejo de la
información vía
RAPIDComm
 Minimiza el tiempo lejos del paciente.
 Incrementa el rendimiento durante procesamientos con
cargas de alto volumen.
Resultados rápidos
500 = ~60 sec.
405 = ~60 sec.
Features & Benefits
Page 20 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
 Mas de una muestra
 Proporciona un mayor conocimiento clínico
Amplio menú
Ventajas
Características
 Asegura la integridad del sistema
 Elimina el tiempo gastado en el manejo manual del
material de control.
 Reduce el tiempo muerto
Control de calidad
automático
Features & Benefits
Page 21 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Installation
Page 23 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Inspect the box for shipping damage
Installation
Page 24 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Open the box and remove the accessories pack
Installation
Page 25 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Grasp the instrument by the handle. While tilting the box toward you, lift the
instrument out. Remove the packing material and plastic bag.
Installation
Page 26 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ubique el sistema sobre un meson u otra
superficie a nivel donde vaya a ser instalado.
Remueva las cintas blancas del analizador.
• Pantalla (4 lugares)
• Puerta (2 lugares)
• Tapa del filtro
Remueva el plastico de burbujas detras del display
Instale un rollo de papel en la impresora.
Installation
Page 27 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Connect barcode scanner, RJ-45 network cable, and power cord and
turn on the power switch.
Installation
Page 28 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Durante la inicialización, cuando la pantalla de Siemens que se muestra
arriba aparezca, toque la pantalla una vez para iniciar la calibración de
emergencia de la pantalla ( cuando la barra de tareas haya progresado
al rededor de 1 cm de la barra).
Installation
Page 29 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
• When the screen calibration screen is displayed, touch the targets when they
appear. Note:These targets only stay on screen for ~12 seconds so you must
be available when they appear.
Installation
Page 30 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
When prompted to replace the Measurement Cartridge, touch Cancel.
Installation
Page 31 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Press system Info to check the software version and serial number
Installation
Page 32 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
When prompted to replace the Measurement Cartridge, touch Cancel.
Installation
Page 33 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Touch Setup and set the options to the customer’s preferences.
Page 34 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Create a Setup Disk by touching Secured Options then Save Setup.
Installation
Page 35 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Return to the Status screen and replace the
Measurement Cartridge by touching Replace.
Installation
Page 36 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
The system door will open and the
cartridge replacement video will begin to play.
Installation
Page 37 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Install a new Measurement Cartridge.
Installation
Page 38 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Install a new Wash / Waste Cartridge.
Installation
Page 39 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Close the system door.
Installation
Page 40 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
The video will stop and the system will begin a cartridge initialization.
~25 minutes for a RAPIDPoint 500.
Installation
Page 41 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Completing the Installation
• Corra los tres niveles de Rapid QC y
verifique que todos los resultados de los
parametros caigan en los rangos de
inserto.
• Corra CV si es requerido.
• Complete el reporte de condiciones del
sistema.
Page 42 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Components
Page 44 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Active Matrix Display Colour Touch-screen
Component Identification
Page 45 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
AQC Cartridge
System Door
Component Identification
Integrated Bar
Code Reader
Page 46 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Built in Roll
Printer
Paper
Advance
Knob
Component Identification
Page 47 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Measurement
Cartridge
Wash/Waste
Cartridge
Measurement Cartridge
Release Lever
Component Identification
AQC
Cartridge
Page 48 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
USB Ports
Carrying
Handle
Component Identification
Page 49 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Screen Tilt/
Release Lever
Printer Paper
Cradle
Component Identification
Page 50 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
AC Power
Connection
Power
Switch
RJ-45 Ethernet
Network Port
RS-232
Serial Port
Barcode Scanner Port
Component Identification
Page 51 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Air Filter (black)
Part Number - SMN 10322638
Lamp
Part Number - SMN 10310472
Component Identification
Page 52 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Routine Operation
Page 54 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ease of Use - User Interface
Sample Type
Test Parameters
Customizable
Panels
System Icons
Page 55 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
1 Scan barcode
2 Insert sample
3 Press Start
Ease of Use - User Interface
Page 56 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ease of Use - User Interface
Page 57 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ease of Use - User Interface
Page 58 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ease of Use - User Interface
Page 59 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Results still in ~60 seconds
Ease of Use - User Interface
Page 60 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Results still in ~60 seconds
Ease of Use - User Interface
Protection notice / Copyright notice
RAPIDPoint 500
Screen Navigation
Page 62 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Home Screen
Banner icons slightly different.
Note START Button replaces ANALYZE
button.
Barcode scanner icon
refers to new on board
scanner
Page 63 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Page 64 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 65 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Help Screen
Page 66 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Extensive Menu
nBilirubin
pH
pCO2
pO2
Co-oximetry
Na+
K+
Ca++
Cl-
Glucose
Lactate
250 Tests
400 Tests
750 Tests
pH
pCO2
pO2
nBilirubin
Co-oximetry
pH
pCO2
pO2
Na+
K+
Ca++
Cl-
Glucose
nBilirubin
Co-oximetry
pH
pCO2
pO2
Na+
K+
Ca++
Cl-
Glucose
Lactate
nBilirubin
Co-oximetry
Page 67 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Status Screen
Page 69 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Access to Status
Touch Here
to Access the
Status Screen
Page 70 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Page 71 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Información del cartucho
Analizadas - numero de muestras y QC los cuales han sido analizados utilizando los
cartuchos actuales.
Días restantes – numero de días que le quedan a cada cartucho a bordo.
Fecha y hora de expiración – Expiración estimada, basada en:
28 días desde la fecha/hora de instalación del cartucho de medida.
10 días desde la fecha/hora de instalación del cartucho de lavado/desecho.
28 días desde la fecha/hora de instalación del cartucho de AQC.
Registro de Eventos
Permite ver los mensajes o eventos actuales del sistema.
Page 72 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Reemplazar.
Luego de la expiración de un cartucho, toque Reemplazar para inicializar las
secuencias de retiro e instalación.
Page 73 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Cartridges
Page 74 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Cartridges
Page 75 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the Measurement Cartridge
Page 76 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
• En la pantalla de Status, toque el icono cartucho de medida y luego toque
reemplazar.
• La puerta se abrirá y el video comenzara a correr.
• Remueva el cartucho de lavado/desechos y descártelo.
• Levante la palanca hasta que el cartucho de medida sea expulsado.
• Alinee las guías de los lados de el cartucho de medida con las guías del
sistema.
• Posicione el cartucho en el sistema empujando el cartucho firmemente hasta
que se escuche que el cartucho caso en el sitio.
• Baje la palanca.
• Inserte un nuevo cartucho de lavado /desecho y cierre la puerta.
• El CARTUCHO DE MEDIDA NO PUEDE VOLVER A SER RE INSTALADO
NI REUBICADO.
Replacing the Measurement Cartridge
Page 77 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Cartridges
Page 78 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the Waste/Wash Cartridge
Page 79 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
• En la pantalla de Status, toque el icono del cartucho lavado/desechos y toque
reemplazar.
• La puerta se abrirá y el video empezara a correr.
• Remueva el cartucho de lavado /desecho y descártelo.
• Instale un nuevo cartucho de lavado /desecho en el sistema, empújelo firmemente
hasta que el cartucho rompa y quede en el lugar.
• Cierre la puerta del sistema.
• El cartucho de lavado /desecho no puede ser reinstalado una vez se ha
puesto en el sistema, ya que tiene un dispositivo de bloqueo que no
permite reinstalarlo una vez haya sido usado.
Replacing the Waste/Wash Cartridge
Page 80 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Cartridges
Page 81 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the AQC Cartridge
Page 82 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the AQC Cartridge
• En la pantalla de Status, toque el icono de AQC y toque reemplazar.
• Presione y deslice el conector del cartucho AQC hacia la derecha y espere a que el
cartucho AQC sea expulsado del sistema.
• Remueva el cartucho de AQC y descártelo.
• Remueva la tarjeta amarilla que esta debajo de la tapa color gris del nuevo cartucho.
• Presione firmemente la pieza gris del AQC con el fin de perforar las bolsas que
contienen el material de control.
• Coloque el cartucho en el sistema y presione firmemente deslizándolo hacia atrás
hasta que se escuche que el cartucho quedo en el lugar.
• Deslice el conector del cartucho AQC hacia la izquierda para cerrarlo.
• Una pantalla de espera aparece en la pantalla mientras el sistema prepara el
cartucho.
• El cartucho de AQC puede ser reubicado en el mismo sistema hasta 6 horas
después que haya sido removido.
Page 83 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Port
Permite al operador reemplazar el puerto de muestras.
Page 84 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Replace Sample Port
Page 85 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Shutdown
Forma correcta de apagar el sistema ya que cierra el sistema operativo e indica al
usuario el momento de apagar el interruptor del equipo.
Page 86 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Shutdown Screen
Page 87 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Shutdown Screen
Page 88 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Clean Screen
Touch to disable the touch screen for 20 seconds to clean.
Page 89 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Clean Screen
20 Seconds to clean the screen with lint free cloth and 0.5% sodium
hypochlorite solution
Page 90 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Calibrate
Permite al operador iniciar manualmente una calibración de un punto, dos puntos, o
una calibración completa de dos puntos.
El icono de calibración en la pantalla del sistema cambia por Reiniciar Cartucho si algún
parámetro falla durante la inicialización o instalación de un nuevo cartucho.
Page 91 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Calibrate
Solo operadores con nivel 1 o 2 de seguridad pueden tener acceso a la calibración del sistema
Page 92 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > Calibrate
Page 93 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > System Info..
Page 94 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen > System Info
Permite ver el numero de teléfono de servicio, el ID del sistema, versión del Software,
contaje de ciclos y seriales de los cartuchos instalados actualmente.
Page 95 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Acceso a las opciones de configuración del
sistema.
Page 96 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Instrument Set Up
Page 98 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Touch Here to Access
the Instrument Setup
Status Screen
Page 99 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC
Page 100 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC > QC Options
Page 101 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > QC > QC Options
Unscheduled
Required QC
Automatic QC (Repeat)
QC Syringe
QC Range Methodology
Page 102 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC
Page 103 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Schedule
Setup Schedule – selecciona días, horas y niveles a ser corridos con un nuevo cartucho de
medida.
Page 104 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > QC > QC Options >
Required QC Schedule > Intervals
Page 105 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Schedule
Setup Schedule – selecciona días, horas y niveles a ser corridos
Page 106 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Schedule
New M-Cart Setup. Puede programar para que el sistema automáticamente corra los
controles de calidad cuando un nuevo cartucho de medida ha sido instalado en el sistema.
Page 107 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC
Page 108 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Ranges
2D
Page 109 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Ranges
Page 110 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Ranges
Page 111 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Ranges
Page 112 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC
Page 113 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Automatic QC Schedule
Page 114 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Required QC Schedule
New M-Cart Setup. Usted puede programas el nivel que desee correr cuando se isntale un
nuevo cartucho de medida en el sistema.
Page 115 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status > Setup > QC
Page 116 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Automatic QC Ranges
Page 117 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Automatic QC Ranges
Page 118 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Automatic QC Ranges
Page 119 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDPoint 500 Setup > QC >
QC Options > Automatic QC Ranges
Page 120 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 121 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 122 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 123 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Patient Ranges
Ingrese los rangos para los resultados de pacientes.
Page 124 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 125 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Patient Demographics
Selecciones los demográficos para paciente, puede activar
la rápida identificación de la muestra.
Page 126 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 127 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Sample Demographics - Screen 1
Seleccione los demográficos de la muestra.
Page 128 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Sample Demographics - Screen 2
Page 129 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 130 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Parameter Selection
Permite al operador seleccionar parámetros de medida para ser medidos en la
pantalla de proceso.
Paneles personalizados , permite configurar hasta 6 paneles.
Un panel puede ser seleccionado como predeterminado.
Page 131 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 132 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 133 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample
Page 134 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Sample > Sample Type
Seleccione el tipo de muestra por defecto.
Page 135 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters
Page 136 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup >Parameters > Parameters On/Off Screen 1
Seleccione los parámetros a ser reportados por el sistema.
Page 137 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup >Parameters > Parameters On/Off Screen 2
Page 138 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup >Parameters > Parameters On/Off Screen 3
Page 139 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters
Page 140 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters > Parameter Units Screen 1
Seleccione las unidades de medida de los parámetros.
Page 141 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters > Parameter Units Screen 2
Page 142 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters
Page 143 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters > Demographic Units screen 2
Page 144 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters
Page 145 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Parameters > Values
Ingrese el valor de la presión atmosférica.
Page 146 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 147 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 148 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options > Country Options
Seleccione el lenguaje del sistema y el formato de la fecha.
Page 149 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 150 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options > Date & Time
Ajuste la fecha y hora del sistema.
Page 151 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 152 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options > Sound
Ajuste el volumen del sistema y el volumen del sonido de video.
Page 153 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 154 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Display Options
Ajuste el birllo de la pantalla.
Page 155 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options
Page 156 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > System Options > Other Options
Ingrese el nombre del sistema , teléfono de contacto para asistencia, y
encienda el aviso de calibración pendiente.
Page 157 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 158 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 159 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Printer Options
Active o desactive la impresora, seleccione las características de impresión.
Page 160 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 161 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices> Bar Code Setup
Active o desactive el lector de código de barras y seleccione el formato de ID del paciente
Page 162 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup> Printer & Devices >Patient ID Bar Code
Page 163 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices >Accession No. Bar Code
Page 164 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices >Password Bar Code
Page 165 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 166 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Communications
Configure las comunicaciones.
Page 167 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices >Configure RS-232
Page 168 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 169 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Network Setup
Configure las opciones de red.
Page 170 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Printer Options
Page 171 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 172 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Email Setup
Configure la funcionalidad de envío automático de Email.
Page 173 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices
Page 174 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Printer & Devices > Remote Viewer
Configure la funcionalidad de acceso remoto.
Page 175 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 176 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 177 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >System Access
Define las restricciones de acceso a las opciones del sistema- No restringido, limitado y restringido.
Acceso Restringido: todos los operadores deben ingresar un password
para usar el sistema.
Acceso limitado: Todos los operadores pueden realizar tareas de
rutina, como analizar muestras y reemplazar cartuchos sin ingresar un
password.
Acceso no Restringido: Todos los operadores pueden realizar tareas de
rutina sin entrar un password, y todos los operadores pueden acceder
a las opciones de configuración.
Page 178 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 179 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >Operator Security
Define la ID del operador, el password tienen 4 niveles de acceso.
NB: Dos operadores IDs o dos passwords pueden ser idénticos.
Page 180 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >System Access
Page 181 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >System Access
Page 182 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >System Access
Page 183 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 184 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >Analysis Options
Seleccione Editar Demográficos, Salvar Demográficos, Mostrar resultados con marca ?, Lista
de pacientes y Rangos analíticos.
Page 185 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 186 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >Interference Correction
Seleccione la opción de corrección de muestra venosa mixta. Se activa solo para reportar PO2 y tHB .
Page 187 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 188 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options >Screen Calibration
Calibre el touch screen
Page 189 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options
Page 190 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Salvar la configuracion en una usb.
Secured Options > Save Setup
Page 191 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options
Page 192 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options > Restore Setup
Restaura los datos de calibracion desde una USB
Page 193 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options
Page 194 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options > Software Installation
Instala una nueva versión de software del sistema.
Page 195 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options > Software Installation
Page 196 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options > Software Installation
Page 197 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Setup > Secured Options
Page 198 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Secured Options > Correlation Coefficients
Ajusta los valores de pendiente y offset
Page 199 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Recall Screen
Page 201 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Status Screen
Page 202 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 203 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Patients
Muestra un archivo que almacena hasta 250 muestras de paciente que han sido corridas.
Page 204 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Patients
Page 205 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Patients
Page 206 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Patients
Page 207 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 208 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 209 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> QC
Muestra un archivo que contiene los últimos 250 QC resultados que hayan sido corridos.
Page 210 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> QC
Page 211 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 212 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 213 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 214 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 215 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Calibrations
Muestra y grafica la fecha y hora de las ultimas 250 calibraciones realizadas.
Page 216 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 217 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Events Log
Muestra un archivo que contiene los ultimos 250 eventos que han ocurrido en el sistema.
Page 218 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 219 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Sample Totals
Muestra el total de muestras procesadas por cada analito.
Page 220 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Sample Totals
Page 221 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
No es informacion para el cliente.
Ingrese usando el Service Password, XXXXX52637, donde XXXXX son los 5
digitos del serial del instrumento. – ESTE PASSWORD NUNGA DEBE SER DADO A
LOS CLIENTES.
Recall Screen > Sample Totals > Service Data
FTP Server - Enter IP address of
RAPIDComm with dedicated line for data
collection
Page 222 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen> Sample Totals
Page 223 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen > Sample Totals > Diagnostics
No es para los usuarios.
Ingrese usando el Service Password
FS diagnósticos pueden ser realizados acá.
Page 224 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Simulator
 Measurement Test
 Nota: El sistema es probado 3 veces y por cada canal que es seleccionado por un
switch de posición en la parte frontal del cartucho simulador de sensores. El
sistema debe ser probado rotando el switch para seleccionar las posiciones HI,
Norm y LOW.
 Instale el cartucho simulador de sensores.
 Low
 Coloque el Cartucho de simulación en LOW.
 Presiones los resultados serán impresos.
 Verifique los datos y asegúrese que el numero serial del cartucho de simulación
este impreso.
 Fíjese en la impresión en la columna ‘L’ Chequee los valores actuales para toda
la columna:
 ‘0’ = Pass
 ‘F’ = Fail
 Para los canales 6 & 7 chequee si los valores están dentro de los limites que
se encuentran en la tabla. Rechace si cualquier unidad esta por fuera de los
rangos.
Page 225 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Simulator
 Test de medicion
 Normal
 Posicione el Cartucho de simulación de sensores en Norm.
 Presione e los resultados serán impresos.
 Fíjese en la impresión en la columna ‘N’ Chequee los valores actuales para toda la
columna:
 ‘0’ = Pass
 ‘F’ = Fail
 Para los canales 6 & 7 chequee si los valores esta en los limites de la tabla.
Rechace si cualquier unidad esta por fuera de los rangos.
Page 226 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Simulator
test de medicion
 High
 Posicione el cartucho de simulación de sensores en la posición High.
 Presione los resultados serán impresos.
 Fíjese en la impresión en la columna ‘H’ Chequee que los valores actuales para
toda la columna:
 ‘0’ = Pass
 ‘F’ = Fail
 Para los canales 6 & 7 chequee que los valores estén en los rangos de la tabla.
Rechace si alguna unidad esta fuera del rango.
Page 227 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Simulator
Diagnostic
Measurement
Test Printout
Name Mux channel
-LOW
limit
+LOW
limit
-MED
limit
+MED
limit
-HIGH
limit
+HIGH
limit
1 GND 1 -0.25 0.25 -0.25 0.25 -0.25 0.25
2 PO2 2 -25.00 25.00 790.00 970.00 2900.00 3600.00
3 HCO3 3 -1382.00 -968.00 -119.00 347.00 1145.00 1662.00
4 GLUA 4 -50.00 50.00 -460.00 -340.00 -3700.00 -3000.00
5 GLUB 5 -50.00 50.00 -460.00 -340.00 -3700.00 -3000.00
6 LACA 6 -50.00 50.00 -345.00 -255.00 -2750.00 -2250.00
7 LACB 7 -50.00 50.00 -345.00 -255.00 -2750.00 -2250.00
8 PH1 8 -1229.00 -809.00 -246.00 214.00 732.00 1232.00
9 K 9 -787.00 -349.00 39.00 511.00 870.00 1376.00
10 CL 10 -79.00 388.00 422.00 910.00 924.00 1432.00
11 NA 11 -649.00 -200.00 132.00 608.00 904.00 1411.00
12 CA 12 -173.00 291.00 594.00 1089.00 1051.00 1882.00
13 REF 13 -200.00 200.00 -200.00 200.00 -200.00 200.00
14 DAC 14 0.00 0.00 0.00 0.00 0.00 0.00
15 PH2 15 -1382.00 -968.00 -119.00 347.00 1145.00 1662.00
16 MG 16 NA NA NA NA NA NA
17 17 0.00 0.00 0.00 0.00 0.00 0.00
18 FD1 FD1 -350.00 650.00 750.00 1818.00 1850.00 3148.00
19 FD2 FD2 -350.00 650.00 750.00 1818.00 1850.00 3148.00
20 PRE NA NA NA NA NA NA
21 BLK NA NA NA NA NA NA
22 SAMP NA NA NA NA NA NA
Page 228 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen > Sample Totals > Diagnostics
Page 229 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Valve Alignment
Page 230 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Valve Alignment
•Asegúrese que la herramienta de alineación de la válvula este asegurada
sobre la base con el tornillo.
•Asegúrese que el tornillo manual este ajustado usando un
destornillador plano.
•Verifique que el indicador de la válvula marque cero ( 0 ±
0.0005)
•Si la lectura de la válvula esta fuera de lo especificado,
suelte el tornillo y ajuste el dial en cero.
•Re-Ajuste el tornillo . Asegúrese que la medida continua en
cero ± 0.0005
•Repita el proceso (para verificar repetibilidad.
•Asegúrese que la medición sea repetible cero ± 0.0005.
Page 231 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
M Valve Test
Instale la herramienta de alineación de la válvula en la interface del
cartucho del sistema.
Presione el embolo para ver que no golpee con la válvula M cundo este
ajustada.
Asegúrese que la herramienta de ajuste este asegurada.
Verifique que la herramienta esta ajustada usando un destornillador .Sea
cuidadoso de no apretar demasiado porque puede dañar el tornillo.
Presione
Presione la válvula se mover a la primera posición.
Presione la válvula se moverá a la posición 2
Verifique que la medición este 0 + 5.
Continúe, presionando hasta que la válvula retorne a la posición
2.
Confirme que la lectura se mantenga en 0 + 5.
Presione espere a que el contador baje.
Regrese a la pantalla de diagnósticos.
Remueva la herramienta de alineación de la interface del cartucho.
Regrese la herramienta a su base.
Troubleshooting > Valve Alignment
Page 232 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen > Sample Totals > Diagnostics
Page 233 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Temperature
Heater Test
Ingrese al modo de diagnósticos.
Presione los resultados serán impresos.
Verifique que el campo 2 esta en : MAIN.
Asegúrese que el campo 3 es mayor a 300
NOTE:Desde el arranque (frio), la temperatura deberá estabilizarse por 25 minutos.
Desde un inicio a temperatura, la temperatura deberá estabilizarse durante 15
minutos antes de que el test sea iniciado. Luego de este tiempo presione
TEMPERATURE. El STEADY conteo, campo 3, deberá estar por encima de 300
antes de chequear las lecturas. Es aceptable tener que repetir la medición para
asegurarse que el campo 3 es mayor a 300.
Page 234 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Temperature
Field Description Lower Limit Upper limit
1 NA NA NA
2 MAIN (If COV printed – close the door) MAIN MAIN
3 Steady State control 300 NA
4 DAC MAIN BLOCK NA NA
5 BLOCK TEMP 37.4 39.6
6 DAC Setting for Pre-Heater NA 700
7 PREHEATER 37.0 38.5
8 SAMPLE TEMP 36.85 37.15
9 CO-ox Sample Temp 36.65 37.35
Compare the temperatures with the following table.
Page 235 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen > Sample Totals > Diagnostics
Page 236 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > Wavelength
Wavelength
Desde la pantalla de diagnósticos presione.
Asegúrese de lo siguiente:
Que el campo 1 es mayor a 1000
Que el campo 5 es menor que ± 1.0
Que el campo 6 es menor de 50000
Que el reporte de impresión reporte PASS
FIEL
D
Description Low
er
Limit
Upper
Limit
1 Mean Peak Counts (Average of Multiple Scans) 1000 NA
2 Standard Deviation of Peak Counts NA NA
3 Centroid Measured from Instrument on-board Neon NA NA
4 Centroid Stored from Factory Alignment Polychromator NA NA
5 Measured Wavelength shift in Pixels fields 3 - 4 -1.0 1.0
6 CMB Integration Time from last Zero Calibration NA 50000
7 Polychromator Serial Number NA NA
Page 237 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 238 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Software
Page 239 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 240 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Software
Copia 3 meses de datos de diagnostico a una USB. Envíe por Email a
siemens para un M-Cart. Remplazo y troubleshooting.
Page 241 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 242 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Software
3 Types of Sensor Data
Type 1 – Gives Sensor Performance Data
Type 2 – Gives CO-Ox Performance Data
Type 0 – Gives Software Performance Data
Page 243 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Recall Screen
Page 244 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Hardware
Page 246 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Enhanced
Communication
Options
Fast Time to
Patient Results
Extended Test Menu to
Include: Lactate &
Neonatal Total Bilirubin
Integrated Barcode
Reader
RAPIDComm and LIS
Connectivity Options
Proven AQC
Application
Page 247 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Ease of Use – Integrated Barcode Reader
Page 248 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Provides Positive Identification
Offers Multiple Configurations
Easily Accessible
Ensures Data Integrity
Meets Hospitals Needs Now and the Future
Offers Convenience
Ease of Use – Integrated Barcode Reader
Page 249 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
2D
1D
Positive Patient ID
Positive Operator ID
2D - Enhanced Information Possibilities
Compatible with common formats
Ease of Use – Integrated Barcode Reader
Page 250 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
On Board Barcode Scanner
Page 251 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Remove the Allen hinge pin securing the front door at the bottom, and remove the front door.
Lay the door on its front side
Page 252 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Remove the 4 screws securing the barcode reader housing. The end of the housing arm
hooks onto a rib on the door
Page 253 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Gently disconnect the flex cable from the barcode reader interface board, then remove the 2
screws securing the barcode reader interface board to its housing and disconnect the USB
cable from the interface board.
Page 254 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Remove the 2 screws securing the cartridge interface assembly and pull it slightly forward.
Page 255 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Note the routing of the BCR cable through the notch in the Cartridge Interface Assembly.
Page 256 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
USB Board with On-Board Barcode Scanner
connected
Page 257 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
3 Available USB Ports
Page 258 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Remove the rear cover and power supply module
Remove the 2 screws securing the USB board.
Page 259 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Replacing the Printer
Rotate the paper cover down.2. Loosen the one captive screw at the top of the printer cover,
and rotate the printer cover down.
Page 260 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Loosen 2 captive printer screws at the top of the printer.
Page 261 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Remove 2 screws at the bottom
Use a flat-bladed screwdriver to pry off the paper advance knob
Disconnect the printer ribbon cable and remove the printer.
Page 262 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Main Board Assembly
(New to RP500)
Page 263 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Display Connection to Main Board
Page 264 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
New Style Hard Drive
(RP500 Only)
Page 265 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Hardware
Page 266 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Routine System Procedures
Page 268 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
• Reemplace el Wash/Waste Cartridge
• Reemplace el Measurement Cartridge
• Reemplace el Automatic QC Cartridge
• Reemplace el Sample Port
• Limpie y desinfecte la pantalla.
• Limpiar y desinfectar las superficies exteriores
• Vacíe el recipiente de ampolletas.
• Reemplace el papel de la impresora.
• Reemplace los filtros de aire.
Routine System Procedures
Page 269 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Cleaning & Disinfecting the Screen
and Exterior Surfaces
• Humedezca un limpión que no suelte hilos con Solución de hipoclorito de sodio al
0.5%
• Escoja el icono Clean Screen de la pantalla de Status.
• Mientras se limpia la pantalla son mostrados (20 segundos), Limpie la pantalla con el
limpión húmedo y séquela.
• Limpie la superficie exterior del sistema con un limpión húmedo y seque las superficies.
Page 270 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Emptying the Ampule Breaker
• Remueva el contenedor de ampolletas
del sistema.
• Vacíe el contenedor de ampolletas en
un sitio adecuado.(guardián)
• Si es necesario , limpie el interior del
contenedor de ampolletas con una
solución blanqueadora al 10%
• Reinstale el contenedor de ampolleta
Page 271 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the Printer Paper
Page 272 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Replacing the Air Filter
•Hale el seguro del filtro del aire del
sistema.
•Remueva el filtro de aire
•Instale un nuevo filtro de aire.
•Reinstale el seguro del filtro del aire al
sistema.
Page 273 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Networking and Connectivity
Page 275 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDComm Information Management System
Page 276 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RAPIDSystems Product Portfolio
Page 277 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Connectivity Real Time View
Page 278 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Compatible with all current Siemens Diagnostics Critical Care Systems
Remote monitoring and control functions
Built on Standard Technology
 Microsoft® XP Operating System
 Microsoft® SQL Server Database
Key Features
Page 279 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Management
 Provides powerful management tools to improve laboratory operations and
lower costs.
Clinical
 Provides clinicians with critical care results, associated reference ranges,
hospital and patient demographics, a clinical interpretation of results, and the
last four patient samples from the database -- All in a single concise report.
Regulatory
 All Quality Control, Maintenance, and Calibration records are stored and
readily accessible for inspection purposes. The program also schedules
maintenance, monitors compliance, and logs activity.
User
 Intuitive point-and-click design, and online training module for easy operation.
Key Benefits
Page 280 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
Patient Sample Analysis
Page 282 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Collecting Patient Samples > Fill Volumes
Sample Type
Collection
Collection
Device
Minimum Fill
Volume
Preparation
Arterial blood, venous
blood, or mixed venous
blood
syringe
• Expel air from the syringe and cap it
immediately after obtaining the sample
1 mL 200 μL
3 mL 800 μL • Do not use cork to cap the syringe
5 mL 1.5 mL
• Mix the sample thoroughly by rolling the
syringe between your palms and gently
inverting it several times.
capillary blood capillary tube 100 μL
• Fill the tube completely and cap it
securely
• Do not use clay or cork to cap the tube
• Do not use capillary tubes containing
mixing fleas
Page 283 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Patient Samples
•Mezcle la muestra.
•Seleccione el tipo de muestra a analizar
en el sistema.
•Inserte la jeringa o capilar en el puerto
de muestra y seleccione Start.
•Remueva la muestra del dispositivo
cundo se le indique y seleccione continuar.
(botón a la derecha de la pantalla).
•Ingrese los demográficos del paciente o
use la ID del paciente (host Query) y
luego seleccione Continuar.
•View results.
Page 284 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis
Syringe, Capillary, Venous
Page 285 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis
Syringe, Capillary, Venous
Page 286 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis
Syringe, Capillary, Venous
Page 287 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Mixed Venous Sample Analysis
Page 288 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis > Remove Sample Device
Page 289 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis > Enter Demographics
Page 290 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis > Keyboard
Page 291 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis > Analyzing
Page 292 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Sample Analysis > Results
Page 293 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A
Protection notice / Copyright notice
RAPIDPoint 500
QC Analysis
Page 295 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
RapidQC Complete
An aqueous control for pH/blood gas, electrolyte,
CO-oximetry, and metabolite analysis
• Storage
• 24 months from date of manufacture, at room temperature
• Catalog Numbers
• Level 130ampules x 2.5mL/box SMN 10309925
• Level 230ampules x 2.5mL/box SMN 10309926
• Level 330ampules x 2.5mL/box SMN 10309927
Page 296 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
•Touch the button on Analysis Screen.
•Scan the barcode on the QC ampule.
•Open the ampule and attach the Quick adapter
•Insert the Quick adapter with attached ampule into the sample port and touch
Continue.
•Remove when prompted and touch Continue.
•View results.
•Parameters that fail Required QC will not be available for sample analysis.
Analyzing Required QC Samples
Page 297 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Required QC Samples
Page 298 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Required QC Samples > Failed
Page 299 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled QC
•Touch the QC ampule Button on the Analysis Screen
•Scan the barcode on the QC ampule
•Open the ampule and attach the Quick adapter
•Insert the Quick adapter with attached ampule into the sample port and touch Analyze
•Remove when prompted and touch Continue
•View results
•Parameters that fail Unscheduled QC will be available for sample analysis
Page 300 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled QC
Page 301 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled QC
Page 302 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled QC
Page 303 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled AQC
Touch the AQC Button on the Analysis Screen
Touch Analyze
Touch the level(s) to be run and touch Start
View results
Parameters that fail AQC will not be available for sample analysis
Page 304 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled AQC
Page 305 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled AQC
Page 306 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled AQC
Page 307 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Analyzing Unscheduled AQC
Protection notice / Copyright notice
RAPIDPoint 500
Calibrations
Page 309 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
One Point Calibration
• Occurs every 30 minutes
• 2 minutes to complete
• pH, pO2, pCO2, Na+, K+, Cl-, Ca++, and Glu.
• The system measures the 200 Cal calibrator on all sensors and adjusts for
cal. drift of all analytes.
Calibrations
Page 310 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Two Point Calibration
• Occurs every 2 hours
• 3 minutes 45 seconds to complete
• pH, pO2, pCO2, Na+, K+, Cl-, Ca++, Glu, and CO-ox Zero
• The system measures a combination of two calibrators, 200 Cal and Zero Cal,
and adjusts for cal. and slope drifts of all analytes.
• If the zero calibration for the CO-ox exceeds drift limits, CO-ox zero occurs
during the 1 pt. Cal until the drift error is cleared.
Calibrations
Page 311 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Full Two Point Calibration
• Occurs every 8 hours
• 5 minutes 30 seconds to complete
• All functions of a One and Two Point Cal and CO-ox Slope.
• If the zero calibration for the CO-ox exceeds drift limits, CO-ox zero occurs
during the 1 pt. Cal until the drift error is cleared.
Calibrations
Page 312 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Interruptible Calibrations
Touch the STAT button to abort the calibration. The system immediately
performs a wash cycle.
Page 313 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Interruptible Calibrations
System will decide if it’s quicker to stop the calibration
and wash or finish the calibration.
Page 314 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
The dialogue box will NOT contain a STAT button next to the
countdown clock.
Un-Interruptible Calibrations
Page 315 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Cartridge Initialization Timeline
Calibration # Sequence Time elapse
(min)
Time since last
cal (min)
Interruptible /
Uninterruptible
1 Full Two Point - - Uninterruptible
2 Two Point - - Uninterruptible
3 Full Two Point 20 Uninterruptible
4 Two Point 44 20 Uninterruptible
5 One Point 66 20 Interruptible
6 One Point 88 20 Interruptible
7 One Point 110 20 Interruptible
8 Two Point 133 20 Uninterruptible
9 One Point 155 20 Interruptible
10 One Point 177 20 Interruptible
11 One Point 200 20 Interruptible
12 Full Two Point 225 20 Uninterruptible
Page 316 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
The initialization and retrospective calibration (Retrocal) period is not complete until either
7.5 hours have passed or all of the following are true:
• Two scheduled calibrations have occurred after cal #12
• Both calibrations passed the first time with no need to repeat
• No samples were run between the two scheduled calibrations
So the Retrospective calibration period can be between 3.5 hours and 7.5 hours depending
upon system performance
Cartridge Initialization Timeline
Page 317 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Retrospective Calibration Sample
During the initialization period, if a patient sample is run the following
happens:
• The sample is analyzed
• The system washes out the sample with 200 cal and loads 200 cal over
the sensors
• The system performs a one point calibration
• Sample results are adjusted based on the calibration drift
The system will also perform this sequence under certain conditions like
detection of interfering substances.
Page 318 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Normal Sample
If a patient sample is run outside of the retrocal period the following
happens:
 Loads patient sample
 Performs electronic measurement of sample
 Calculates patient data and sends it to UIP
 Runs normal wash The sample is analyzed
Protection notice / Copyright notice
RAPIDPoint 500
Troubleshooting
Page 320 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Troubleshooting > D Codes
D2 - Excessive Drift
• System will recalibrate 2X
• Options are to manually initiate a cal, wait for the next cal, or remove M/C
D3 - Slope Error
• System will recalibrate 2X
• Options are to manually initiate a cal , wait for the next cal, or remove M/C
D4 - Offset Drift
• System will recalibrate 2X
• Options are to manually initiate a cal , wait for the next cal, or remove M/C
Page 321 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D21 - Processing Error
• Reboot
D23:2 or 10 - No Wash detected at FD1 or FD2, respectively
• System will reload the Wash up to 2 more times
• If error not cleared, system prompts for W/W replace
D23:3 or 11 - No 200 Cal detected at FD1 or FD2, respectively
• System will reload the 200 Cal up to 2 more times
• If error not cleared, system prompts for M/C replace
Troubleshooting > D Codes
Page 322 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D23:4 or 12 - No Zero Cal detected at FD1 or FD2, respectively
• System will reload the Zero Cal up to 2 more times
• If error not cleared, system prompts for M/C replace
D23:5 or 13 - No RCx detected at FD1 or FD2, respectively
• System will reload the RCx up to 2 more times
• If error not cleared, system prompts for M/C replace
Troubleshooting > D Codes
Page 323 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D23:7 or 8 - No Wash segment detected at either FD1 or FD2
• Repeat Wash up to 2 more times
• If error does not clear, system prompts for W/W replace
D23:9 - No foam Wash detected at FD1
• System initiates aggressive wash to clear clot and checks FD1 up to 2 more times
• If error does not clear after 3rd try, system prompts for sample port replacement
• Error common on new M/C - with more blood less common
Troubleshooting > D Codes
Page 324 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D33:1 Valve Error
• System must be restarted to clear
D35 - Electronics Error
• System must be restarted to clear
• D35:14 Door Error - check for something blocking the door
D38 - Temperature Error
• Qualifiers 1, 2, 3, 6, & 12 - Fatal error
• Qualifiers 4 & 5 - Non-fatal
• System may clear error and will need to be restarted
• Check temperature of environment
• Qualifier 3 & 6 out of box - leave system running for 1 hour and then restart -
error may clear
Troubleshooting > D Codes
Page 325 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D39 - Obstruction
• System prompts for sample port change
Troubleshooting > D Codes
Page 326 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
D40 - Wash not Detected
• Detected during the prime of a W/W cartridge
• Prompts for W/W change up to 2X
• If error not cleared after 3rd attempt, system prompts for M/C replace
D60 - Communication Error
• Check cable connections
• Check IP addresses
Troubleshooting > D Codes
Page 327 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Bubble in Sample
• Will prompt for sample port change
Insufficient Sample
• Will prompt for sample port change
Refer to “System Messages” in the Troubleshooting and Calibrations section of the Service
Manual for a complete list of all Diagnostic Codes and Corrective Actions
Troubleshooting > Error Messages
Protection notice / Copyright notice
RAPIDPoint 500
Cartridge Overview
Page 329 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Extensive Menu
Full Complement of Tests in a Single
Cartridge
Long Measurement Cartridge Use Life
Fully Customizable Patient Panels
Lactate and Neonatal Total Bilirubin
Total Bilirubin
Methodology Spectrophotometric
Lactate
Methodology Lactate Oxidase
Page 330 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Selection of reagents
AQC
zero
RCX
200
cal
wash
Foam
wash
Syringe
Cap
Vent
Page 331 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Bag Contents – M-Cart and Wash
Page 332 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Bag Contents - AQC
Page 333 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Calibration Points
Page 334 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Page 335 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Inside The Sensor Module
Sample
Serpentine flow
preheater sensors Ref
Page 336 Siemens Healthcare Diagnostics
Siemens Confidential – Internal Use Only
Q&A

More Related Content

What's hot

Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS
Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS
Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS Amazon Web Services
 
CNIT 152: 6. Scope & 7. Live Data Collection
CNIT 152: 6. Scope & 7. Live Data CollectionCNIT 152: 6. Scope & 7. Live Data Collection
CNIT 152: 6. Scope & 7. Live Data CollectionSam Bowne
 
Combining Logs, Metrics, and Traces for Unified Observability
Combining Logs, Metrics, and Traces for Unified ObservabilityCombining Logs, Metrics, and Traces for Unified Observability
Combining Logs, Metrics, and Traces for Unified ObservabilityElasticsearch
 
HFM Extended Analytics
HFM Extended AnalyticsHFM Extended Analytics
HFM Extended Analyticsaa026593
 
D53155 hfm 9.3.1_rules_student_guide
D53155 hfm 9.3.1_rules_student_guideD53155 hfm 9.3.1_rules_student_guide
D53155 hfm 9.3.1_rules_student_guideariffsharif
 
Strategic Choices in SAP S/4 HANA Deployment
Strategic Choices in SAP S/4 HANA DeploymentStrategic Choices in SAP S/4 HANA Deployment
Strategic Choices in SAP S/4 HANA DeploymentDirk Oppenkowski
 
How to Easily Build SCADA & HMI HTML5 Web Applications
How to Easily Build SCADA & HMI HTML5 Web ApplicationsHow to Easily Build SCADA & HMI HTML5 Web Applications
How to Easily Build SCADA & HMI HTML5 Web ApplicationsInductive Automation
 
Simplifying EBS 12.2 ADOP - Collaborate 2019
Simplifying EBS 12.2 ADOP - Collaborate 2019   Simplifying EBS 12.2 ADOP - Collaborate 2019
Simplifying EBS 12.2 ADOP - Collaborate 2019 Alfredo Krieg
 
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors Rebekah Rodriguez
 
Knime (Konstanz Information Miner)
Knime (Konstanz Information Miner)Knime (Konstanz Information Miner)
Knime (Konstanz Information Miner)Kiran Buriro
 
Ray: Enterprise-Grade, Distributed Python
Ray: Enterprise-Grade, Distributed PythonRay: Enterprise-Grade, Distributed Python
Ray: Enterprise-Grade, Distributed PythonDatabricks
 
Currency Translation in HFM
Currency Translation in HFMCurrency Translation in HFM
Currency Translation in HFMaa026593
 

What's hot (12)

Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS
Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS
Best Practices for Genomic and Bioinformatics Analysis Pipelines on AWS
 
CNIT 152: 6. Scope & 7. Live Data Collection
CNIT 152: 6. Scope & 7. Live Data CollectionCNIT 152: 6. Scope & 7. Live Data Collection
CNIT 152: 6. Scope & 7. Live Data Collection
 
Combining Logs, Metrics, and Traces for Unified Observability
Combining Logs, Metrics, and Traces for Unified ObservabilityCombining Logs, Metrics, and Traces for Unified Observability
Combining Logs, Metrics, and Traces for Unified Observability
 
HFM Extended Analytics
HFM Extended AnalyticsHFM Extended Analytics
HFM Extended Analytics
 
D53155 hfm 9.3.1_rules_student_guide
D53155 hfm 9.3.1_rules_student_guideD53155 hfm 9.3.1_rules_student_guide
D53155 hfm 9.3.1_rules_student_guide
 
Strategic Choices in SAP S/4 HANA Deployment
Strategic Choices in SAP S/4 HANA DeploymentStrategic Choices in SAP S/4 HANA Deployment
Strategic Choices in SAP S/4 HANA Deployment
 
How to Easily Build SCADA & HMI HTML5 Web Applications
How to Easily Build SCADA & HMI HTML5 Web ApplicationsHow to Easily Build SCADA & HMI HTML5 Web Applications
How to Easily Build SCADA & HMI HTML5 Web Applications
 
Simplifying EBS 12.2 ADOP - Collaborate 2019
Simplifying EBS 12.2 ADOP - Collaborate 2019   Simplifying EBS 12.2 ADOP - Collaborate 2019
Simplifying EBS 12.2 ADOP - Collaborate 2019
 
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors
X13 Pre-Release Update featuring 4th Gen Intel® Xeon® Scalable Processors
 
Knime (Konstanz Information Miner)
Knime (Konstanz Information Miner)Knime (Konstanz Information Miner)
Knime (Konstanz Information Miner)
 
Ray: Enterprise-Grade, Distributed Python
Ray: Enterprise-Grade, Distributed PythonRay: Enterprise-Grade, Distributed Python
Ray: Enterprise-Grade, Distributed Python
 
Currency Translation in HFM
Currency Translation in HFMCurrency Translation in HFM
Currency Translation in HFM
 

Similar to RAPIDPoint 500 System Service Training Deck.ppt

RNA Automation presentation (2014)
RNA Automation presentation (2014)RNA Automation presentation (2014)
RNA Automation presentation (2014)RNA Automation Ltd
 
Why Predictive Maintenance in Dairy Industry is a necessity? Know here
Why Predictive Maintenance in Dairy Industry is a necessity? Know hereWhy Predictive Maintenance in Dairy Industry is a necessity? Know here
Why Predictive Maintenance in Dairy Industry is a necessity? Know hereInfinite Uptime
 
VEGA Pressure & Level Measurement - Pharmaceuticals Industry Applications
VEGA Pressure & Level Measurement - Pharmaceuticals Industry ApplicationsVEGA Pressure & Level Measurement - Pharmaceuticals Industry Applications
VEGA Pressure & Level Measurement - Pharmaceuticals Industry ApplicationsThorne & Derrick UK
 
Siemens win cc manual pcs 7 wincc basic process control
Siemens win cc manual pcs 7 wincc basic process controlSiemens win cc manual pcs 7 wincc basic process control
Siemens win cc manual pcs 7 wincc basic process controlDien Ha The
 
VEGA Pressure & Level Measurement - Food Industry Applications
VEGA Pressure & Level Measurement  - Food Industry ApplicationsVEGA Pressure & Level Measurement  - Food Industry Applications
VEGA Pressure & Level Measurement - Food Industry ApplicationsThorne & Derrick UK
 
Siemens win cc manual pcs 7 runtime
Siemens win cc manual pcs 7 runtimeSiemens win cc manual pcs 7 runtime
Siemens win cc manual pcs 7 runtimeDien Ha The
 
2014_Pantec_Metrology_BU_PPT_revVERJA
2014_Pantec_Metrology_BU_PPT_revVERJA2014_Pantec_Metrology_BU_PPT_revVERJA
2014_Pantec_Metrology_BU_PPT_revVERJAJavier O. Vera
 
Automated vision inspection systems
Automated vision inspection systemsAutomated vision inspection systems
Automated vision inspection systemsRNA Automation Ltd
 
Veracel andritz - i io-t - online wirelless vibration project implementa...
Veracel    andritz - i io-t -  online wirelless vibration project  implementa...Veracel    andritz - i io-t -  online wirelless vibration project  implementa...
Veracel andritz - i io-t - online wirelless vibration project implementa...Jaques Valle
 
Fl lab collection_05_2011_en
Fl lab collection_05_2011_enFl lab collection_05_2011_en
Fl lab collection_05_2011_enVahid RG-zadeh
 
PolygenPPresentation300dpi.pdf
PolygenPPresentation300dpi.pdfPolygenPPresentation300dpi.pdf
PolygenPPresentation300dpi.pdfNurAkbar16
 
Energy Efficient automation solutions for Pharma Industry
Energy Efficient automation solutions for Pharma Industry Energy Efficient automation solutions for Pharma Industry
Energy Efficient automation solutions for Pharma Industry Sneha Scindia
 
Innovations for Asset Integrity Management
Innovations for Asset Integrity ManagementInnovations for Asset Integrity Management
Innovations for Asset Integrity ManagementSGS
 
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdf
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdfDCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdf
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdfOkeke Livinus
 
10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA SystemInductive Automation
 
10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA SystemInductive Automation
 
Advanced APM .NET Hands-On Lab - AppSphere16
Advanced APM .NET Hands-On Lab - AppSphere16Advanced APM .NET Hands-On Lab - AppSphere16
Advanced APM .NET Hands-On Lab - AppSphere16AppDynamics
 

Similar to RAPIDPoint 500 System Service Training Deck.ppt (20)

RNA presentation (2014)
RNA presentation (2014)RNA presentation (2014)
RNA presentation (2014)
 
RNA Automation presentation (2014)
RNA Automation presentation (2014)RNA Automation presentation (2014)
RNA Automation presentation (2014)
 
Why Predictive Maintenance in Dairy Industry is a necessity? Know here
Why Predictive Maintenance in Dairy Industry is a necessity? Know hereWhy Predictive Maintenance in Dairy Industry is a necessity? Know here
Why Predictive Maintenance in Dairy Industry is a necessity? Know here
 
VEGA Pressure & Level Measurement - Pharmaceuticals Industry Applications
VEGA Pressure & Level Measurement - Pharmaceuticals Industry ApplicationsVEGA Pressure & Level Measurement - Pharmaceuticals Industry Applications
VEGA Pressure & Level Measurement - Pharmaceuticals Industry Applications
 
Siemens win cc manual pcs 7 wincc basic process control
Siemens win cc manual pcs 7 wincc basic process controlSiemens win cc manual pcs 7 wincc basic process control
Siemens win cc manual pcs 7 wincc basic process control
 
Programmable logic controller
Programmable logic controllerProgrammable logic controller
Programmable logic controller
 
VEGA Pressure & Level Measurement - Food Industry Applications
VEGA Pressure & Level Measurement  - Food Industry ApplicationsVEGA Pressure & Level Measurement  - Food Industry Applications
VEGA Pressure & Level Measurement - Food Industry Applications
 
Siemens win cc manual pcs 7 runtime
Siemens win cc manual pcs 7 runtimeSiemens win cc manual pcs 7 runtime
Siemens win cc manual pcs 7 runtime
 
2014_Pantec_Metrology_BU_PPT_revVERJA
2014_Pantec_Metrology_BU_PPT_revVERJA2014_Pantec_Metrology_BU_PPT_revVERJA
2014_Pantec_Metrology_BU_PPT_revVERJA
 
Automated vision inspection systems
Automated vision inspection systemsAutomated vision inspection systems
Automated vision inspection systems
 
Veracel andritz - i io-t - online wirelless vibration project implementa...
Veracel    andritz - i io-t -  online wirelless vibration project  implementa...Veracel    andritz - i io-t -  online wirelless vibration project  implementa...
Veracel andritz - i io-t - online wirelless vibration project implementa...
 
Fl lab collection_05_2011_en
Fl lab collection_05_2011_enFl lab collection_05_2011_en
Fl lab collection_05_2011_en
 
PolygenPPresentation300dpi.pdf
PolygenPPresentation300dpi.pdfPolygenPPresentation300dpi.pdf
PolygenPPresentation300dpi.pdf
 
Energy Efficient automation solutions for Pharma Industry
Energy Efficient automation solutions for Pharma Industry Energy Efficient automation solutions for Pharma Industry
Energy Efficient automation solutions for Pharma Industry
 
Innovations for Asset Integrity Management
Innovations for Asset Integrity ManagementInnovations for Asset Integrity Management
Innovations for Asset Integrity Management
 
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdf
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdfDCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdf
DCS_Check-Out_and_Operator_Training_with_HYSYS_Dynamics_White_v1.3.pdf
 
10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System
 
10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System10 Steps to Architecting a Sustainable SCADA System
10 Steps to Architecting a Sustainable SCADA System
 
RNA Presentation 2015
RNA Presentation 2015RNA Presentation 2015
RNA Presentation 2015
 
Advanced APM .NET Hands-On Lab - AppSphere16
Advanced APM .NET Hands-On Lab - AppSphere16Advanced APM .NET Hands-On Lab - AppSphere16
Advanced APM .NET Hands-On Lab - AppSphere16
 

Recently uploaded

Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordAsst.prof M.Gokilavani
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performancesivaprakash250
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...RajaP95
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...Call Girls in Nagpur High Profile
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 

Recently uploaded (20)

Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 

RAPIDPoint 500 System Service Training Deck.ppt

  • 1. Page 1 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 System Training Liliana Escobar Diab Support Engineer Effective Date: marzo 2015
  • 2. Page 2 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Course Overview Este curso tiene por objeto definir las diferencias de software y hardware entre los sistemas de gases RAPIDPoint 405 y el nuevo RAPIDPoint 500. RAPIDPoint 405 RAPIDPoint 500
  • 3. Page 3 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Product Overview Installation Components Routine Operation Screen Navigation  Status Screen  Instrument Setup  Recall Screen Hardware Routine System Procedures Network and Connectivity Patient Sample Analysis QC Analysis Calibrations Troubleshooting Cartridge Overview Q&A Agenda
  • 4. Page 4 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Our Blood Gas History….. The 60’s & 70’s 175 165 161 pH The 80’s 278/270 178/2500 HP85 The 90’s and to date Rapidlab 800 Rapidlab 238/248/348 Rapidlab 1200 Rapidpoint 400 Rapidlink & RapidComm
  • 5. Protection notice / Copyright notice RAPIDPoint 500 Product Overview
  • 6. Page 6 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 System Intended Use Statement El sistema de Diagnostico Siemens RAPIDPoint 500 esta diseñado para uso profesional en un entorno de Cuidado Crítico o de Laboratorio. Este sistema realiza pruebas de: gases en sangre, electrolitos , metabolitos, hemoglobina total y derivados de la hemoglobina en muestras de sangre total arterial, venosa y capilar. Al proporcionar resultados oportunos, el sistema permite al personal medico tomar decisiones mas rápidas sobre el tratamiento del paciente y mejorar la calidad de la atención del paciente en:  Quirófanos y recuperación critica.  Unidad de cuidados Intensivos.  Salas de Emergencia.
  • 7. Page 7 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Selling Points • Resultados rápidos y disponibles a través de un sistema de POC • Facilidad de uso • Control de Calidad Automático • Amplio menú de analitos críticos. • Soluciones de conectividad Únicos. • Máximo control del sistema incluyendo bloqueo. • Diseñado para ser utilizado en zona cercana al paciente.
  • 8. Page 8 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Specifications
  • 9. Page 9 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Specifications Blood Gases Electrolytes / Metabolites Co-Ox Parameters
  • 10. Page 10 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Unparalleled Ease of Use • Muestreo automático con detección de coágulos y burbujas. • Lector de código de barras Integrado • Pantalla touch screen a color • Entrenamiento On-line con videos y audios de ayuda. • Impresora a bordo y escáner de código de barras opcional • Automatic QC • Conectividad a través RAPIDComm
  • 11. Page 11 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 System - Simple to live with! • Sistemas RAPIDPoint 500 son libres de mantenimiento completamente. • Cartuchos de larga duración • Solo deben mantener en stock 4 números de parte vs. ~30 en los sistemas tradicionales. • Ningún procedimiento programado de limpieza o procedimientos de mantenimiento de sensores.
  • 12. Page 12 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 System • Sistema basado en Cartuchos Multiuso. • Menú – pH, pCO2, pO2 – Na+ , K+ , Ca++ , Cl- – Glucosa, Lactato – tHb, CO-ox & nBili** • Barcode Scanner Integrado • CO-oximetria • Control de Calidad Automático • Guías en Audio/Video **Not available for sale in US. Broad Menu of Analytes
  • 13. Page 13 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Maximum System Control • Provee seguridad y confiabilidad. • Analitos seleccionables dependiendo del muestreo. • Protección del sistema por password • Control de seguridad Multi- nivel • Bloqueo opcional de analitos si falla el control de calidad. • Estado remoto de monitoreo y control Remoto vía RAPIDComm (full remote control) 500 405
  • 14. Page 14 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Connectivity via RAPIDComm version 4.0 Diseñado para el manejo completo de información de pacientes y el sistema: • Monitoreo de los analizadores en tiempo real a través de la red hospitalaria. • Control centralizado para máxima eficiencia • Conocimiento de todos los requerimientos regulatorios. • Conexión al LIS y HIS.
  • 15. Page 15 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ventajas Caracteristicas  Cartuchos de medida sin mantenimiento.  Exposición por bioseguridad se reduce  Los cartuchos son remplazados en sencillos pasos. Los sistemas usan cartuchos de medida que contienen todos los reactivos y sensores y cartuchos de lavado /desechos que son reemplazados cuando se llenan.  Los sistemas pueden ser ubicados en un ambiente con limitaciones de espacio y pueden ser reubicados fácilmente. Los sistemas usan un cartucho compacto, basado en un diseño sin botellas de reactivo ni tanques de gas externos. Features & Benefits
  • 16. Page 16 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ventajas Características  El operador introduce la muestra consistentemente con el volumen adecuado, eliminando variabilidad en los resultados por la técnica de muestreo. El sistema automáticamente aspira las muestras.  El cliente puede maximizar la cantidad de tiempo con los pacientes y no con el sistema. El sistema calibra automáticamente los sensores de medida. Features & Benefits
  • 17. Page 17 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ventajas Características  El operador puede analizar muestras y moverse por el software con facilidad.  Un operador con bajo o cero conocimiento en manejo de computadores puede fácilmente aprender como usar los sistemas. El sistema presenta la información esencial de una forma clara, incluyendo videos y diagramas que muestran paso a paso las acciones para tareas clave.  Operador puede ver la información y seleccionar fácil y rápidamente. El sistema usa una pantalla touch screen a color que permite visualizar claramente. Features & Benefits
  • 18. Page 18 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ventajas Características  No hay perdida de cartuchos a causa de coágulos.  Menos riesgo de exposición por bioseguridad. Manejo de Coágulos  No hay tiempo muerto relacionado con mantenimiento de electrodos , limpieza manual y automática, o procedimientos o técnicas que requieran mantenimiento.  No requiere calibraciones ni controles de calidad adicionales posteriores a una rutina de mantenimiento. Libre de Mantenimiento Features & Benefits
  • 19. Page 19 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ventajas Características  Requisitos reglamentarios  Proporciona un medio para adminsitrar de forma remota los sitemas desde una ubicacion central Información fácil de comprender y manejo de la información vía RAPIDComm  Minimiza el tiempo lejos del paciente.  Incrementa el rendimiento durante procesamientos con cargas de alto volumen. Resultados rápidos 500 = ~60 sec. 405 = ~60 sec. Features & Benefits
  • 20. Page 20 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only  Mas de una muestra  Proporciona un mayor conocimiento clínico Amplio menú Ventajas Características  Asegura la integridad del sistema  Elimina el tiempo gastado en el manejo manual del material de control.  Reduce el tiempo muerto Control de calidad automático Features & Benefits
  • 21. Page 21 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 22. Protection notice / Copyright notice RAPIDPoint 500 Installation
  • 23. Page 23 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Inspect the box for shipping damage Installation
  • 24. Page 24 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Open the box and remove the accessories pack Installation
  • 25. Page 25 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Grasp the instrument by the handle. While tilting the box toward you, lift the instrument out. Remove the packing material and plastic bag. Installation
  • 26. Page 26 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ubique el sistema sobre un meson u otra superficie a nivel donde vaya a ser instalado. Remueva las cintas blancas del analizador. • Pantalla (4 lugares) • Puerta (2 lugares) • Tapa del filtro Remueva el plastico de burbujas detras del display Instale un rollo de papel en la impresora. Installation
  • 27. Page 27 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Connect barcode scanner, RJ-45 network cable, and power cord and turn on the power switch. Installation
  • 28. Page 28 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Durante la inicialización, cuando la pantalla de Siemens que se muestra arriba aparezca, toque la pantalla una vez para iniciar la calibración de emergencia de la pantalla ( cuando la barra de tareas haya progresado al rededor de 1 cm de la barra). Installation
  • 29. Page 29 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only • When the screen calibration screen is displayed, touch the targets when they appear. Note:These targets only stay on screen for ~12 seconds so you must be available when they appear. Installation
  • 30. Page 30 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only When prompted to replace the Measurement Cartridge, touch Cancel. Installation
  • 31. Page 31 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Press system Info to check the software version and serial number Installation
  • 32. Page 32 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only When prompted to replace the Measurement Cartridge, touch Cancel. Installation
  • 33. Page 33 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen Touch Setup and set the options to the customer’s preferences.
  • 34. Page 34 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Create a Setup Disk by touching Secured Options then Save Setup. Installation
  • 35. Page 35 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Return to the Status screen and replace the Measurement Cartridge by touching Replace. Installation
  • 36. Page 36 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only The system door will open and the cartridge replacement video will begin to play. Installation
  • 37. Page 37 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Install a new Measurement Cartridge. Installation
  • 38. Page 38 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Install a new Wash / Waste Cartridge. Installation
  • 39. Page 39 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Close the system door. Installation
  • 40. Page 40 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only The video will stop and the system will begin a cartridge initialization. ~25 minutes for a RAPIDPoint 500. Installation
  • 41. Page 41 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Completing the Installation • Corra los tres niveles de Rapid QC y verifique que todos los resultados de los parametros caigan en los rangos de inserto. • Corra CV si es requerido. • Complete el reporte de condiciones del sistema.
  • 42. Page 42 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 43. Protection notice / Copyright notice RAPIDPoint 500 Components
  • 44. Page 44 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Active Matrix Display Colour Touch-screen Component Identification
  • 45. Page 45 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only AQC Cartridge System Door Component Identification Integrated Bar Code Reader
  • 46. Page 46 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Built in Roll Printer Paper Advance Knob Component Identification
  • 47. Page 47 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Measurement Cartridge Wash/Waste Cartridge Measurement Cartridge Release Lever Component Identification AQC Cartridge
  • 48. Page 48 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only USB Ports Carrying Handle Component Identification
  • 49. Page 49 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Screen Tilt/ Release Lever Printer Paper Cradle Component Identification
  • 50. Page 50 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only AC Power Connection Power Switch RJ-45 Ethernet Network Port RS-232 Serial Port Barcode Scanner Port Component Identification
  • 51. Page 51 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Air Filter (black) Part Number - SMN 10322638 Lamp Part Number - SMN 10310472 Component Identification
  • 52. Page 52 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 53. Protection notice / Copyright notice RAPIDPoint 500 Routine Operation
  • 54. Page 54 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ease of Use - User Interface Sample Type Test Parameters Customizable Panels System Icons
  • 55. Page 55 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only 1 Scan barcode 2 Insert sample 3 Press Start Ease of Use - User Interface
  • 56. Page 56 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ease of Use - User Interface
  • 57. Page 57 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ease of Use - User Interface
  • 58. Page 58 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ease of Use - User Interface
  • 59. Page 59 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Results still in ~60 seconds Ease of Use - User Interface
  • 60. Page 60 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Results still in ~60 seconds Ease of Use - User Interface
  • 61. Protection notice / Copyright notice RAPIDPoint 500 Screen Navigation
  • 62. Page 62 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Home Screen Banner icons slightly different. Note START Button replaces ANALYZE button. Barcode scanner icon refers to new on board scanner
  • 63. Page 63 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen
  • 64. Page 64 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 65. Page 65 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Help Screen
  • 66. Page 66 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Extensive Menu nBilirubin pH pCO2 pO2 Co-oximetry Na+ K+ Ca++ Cl- Glucose Lactate 250 Tests 400 Tests 750 Tests pH pCO2 pO2 nBilirubin Co-oximetry pH pCO2 pO2 Na+ K+ Ca++ Cl- Glucose nBilirubin Co-oximetry pH pCO2 pO2 Na+ K+ Ca++ Cl- Glucose Lactate nBilirubin Co-oximetry
  • 67. Page 67 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 68. Protection notice / Copyright notice RAPIDPoint 500 Status Screen
  • 69. Page 69 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Access to Status Touch Here to Access the Status Screen
  • 70. Page 70 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen
  • 71. Page 71 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen Información del cartucho Analizadas - numero de muestras y QC los cuales han sido analizados utilizando los cartuchos actuales. Días restantes – numero de días que le quedan a cada cartucho a bordo. Fecha y hora de expiración – Expiración estimada, basada en: 28 días desde la fecha/hora de instalación del cartucho de medida. 10 días desde la fecha/hora de instalación del cartucho de lavado/desecho. 28 días desde la fecha/hora de instalación del cartucho de AQC. Registro de Eventos Permite ver los mensajes o eventos actuales del sistema.
  • 72. Page 72 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen Reemplazar. Luego de la expiración de un cartucho, toque Reemplazar para inicializar las secuencias de retiro e instalación.
  • 73. Page 73 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Cartridges
  • 74. Page 74 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Cartridges
  • 75. Page 75 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the Measurement Cartridge
  • 76. Page 76 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only • En la pantalla de Status, toque el icono cartucho de medida y luego toque reemplazar. • La puerta se abrirá y el video comenzara a correr. • Remueva el cartucho de lavado/desechos y descártelo. • Levante la palanca hasta que el cartucho de medida sea expulsado. • Alinee las guías de los lados de el cartucho de medida con las guías del sistema. • Posicione el cartucho en el sistema empujando el cartucho firmemente hasta que se escuche que el cartucho caso en el sitio. • Baje la palanca. • Inserte un nuevo cartucho de lavado /desecho y cierre la puerta. • El CARTUCHO DE MEDIDA NO PUEDE VOLVER A SER RE INSTALADO NI REUBICADO. Replacing the Measurement Cartridge
  • 77. Page 77 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Cartridges
  • 78. Page 78 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the Waste/Wash Cartridge
  • 79. Page 79 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only • En la pantalla de Status, toque el icono del cartucho lavado/desechos y toque reemplazar. • La puerta se abrirá y el video empezara a correr. • Remueva el cartucho de lavado /desecho y descártelo. • Instale un nuevo cartucho de lavado /desecho en el sistema, empújelo firmemente hasta que el cartucho rompa y quede en el lugar. • Cierre la puerta del sistema. • El cartucho de lavado /desecho no puede ser reinstalado una vez se ha puesto en el sistema, ya que tiene un dispositivo de bloqueo que no permite reinstalarlo una vez haya sido usado. Replacing the Waste/Wash Cartridge
  • 80. Page 80 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Cartridges
  • 81. Page 81 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the AQC Cartridge
  • 82. Page 82 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the AQC Cartridge • En la pantalla de Status, toque el icono de AQC y toque reemplazar. • Presione y deslice el conector del cartucho AQC hacia la derecha y espere a que el cartucho AQC sea expulsado del sistema. • Remueva el cartucho de AQC y descártelo. • Remueva la tarjeta amarilla que esta debajo de la tapa color gris del nuevo cartucho. • Presione firmemente la pieza gris del AQC con el fin de perforar las bolsas que contienen el material de control. • Coloque el cartucho en el sistema y presione firmemente deslizándolo hacia atrás hasta que se escuche que el cartucho quedo en el lugar. • Deslice el conector del cartucho AQC hacia la izquierda para cerrarlo. • Una pantalla de espera aparece en la pantalla mientras el sistema prepara el cartucho. • El cartucho de AQC puede ser reubicado en el mismo sistema hasta 6 horas después que haya sido removido.
  • 83. Page 83 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Port Permite al operador reemplazar el puerto de muestras.
  • 84. Page 84 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Replace Sample Port
  • 85. Page 85 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Shutdown Forma correcta de apagar el sistema ya que cierra el sistema operativo e indica al usuario el momento de apagar el interruptor del equipo.
  • 86. Page 86 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Shutdown Screen
  • 87. Page 87 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Shutdown Screen
  • 88. Page 88 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Clean Screen Touch to disable the touch screen for 20 seconds to clean.
  • 89. Page 89 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Clean Screen 20 Seconds to clean the screen with lint free cloth and 0.5% sodium hypochlorite solution
  • 90. Page 90 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Calibrate Permite al operador iniciar manualmente una calibración de un punto, dos puntos, o una calibración completa de dos puntos. El icono de calibración en la pantalla del sistema cambia por Reiniciar Cartucho si algún parámetro falla durante la inicialización o instalación de un nuevo cartucho.
  • 91. Page 91 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Calibrate Solo operadores con nivel 1 o 2 de seguridad pueden tener acceso a la calibración del sistema
  • 92. Page 92 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > Calibrate
  • 93. Page 93 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > System Info..
  • 94. Page 94 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen > System Info Permite ver el numero de teléfono de servicio, el ID del sistema, versión del Software, contaje de ciclos y seriales de los cartuchos instalados actualmente.
  • 95. Page 95 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen Acceso a las opciones de configuración del sistema.
  • 96. Page 96 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 97. Protection notice / Copyright notice RAPIDPoint 500 Instrument Set Up
  • 98. Page 98 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Touch Here to Access the Instrument Setup Status Screen
  • 99. Page 99 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC
  • 100. Page 100 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC > QC Options
  • 101. Page 101 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > QC > QC Options Unscheduled Required QC Automatic QC (Repeat) QC Syringe QC Range Methodology
  • 102. Page 102 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC
  • 103. Page 103 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Schedule Setup Schedule – selecciona días, horas y niveles a ser corridos con un nuevo cartucho de medida.
  • 104. Page 104 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > QC > QC Options > Required QC Schedule > Intervals
  • 105. Page 105 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Schedule Setup Schedule – selecciona días, horas y niveles a ser corridos
  • 106. Page 106 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Schedule New M-Cart Setup. Puede programar para que el sistema automáticamente corra los controles de calidad cuando un nuevo cartucho de medida ha sido instalado en el sistema.
  • 107. Page 107 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC
  • 108. Page 108 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Ranges 2D
  • 109. Page 109 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Ranges
  • 110. Page 110 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Ranges
  • 111. Page 111 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Ranges
  • 112. Page 112 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC
  • 113. Page 113 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Automatic QC Schedule
  • 114. Page 114 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Required QC Schedule New M-Cart Setup. Usted puede programas el nivel que desee correr cuando se isntale un nuevo cartucho de medida en el sistema.
  • 115. Page 115 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status > Setup > QC
  • 116. Page 116 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Automatic QC Ranges
  • 117. Page 117 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Automatic QC Ranges
  • 118. Page 118 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Automatic QC Ranges
  • 119. Page 119 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDPoint 500 Setup > QC > QC Options > Automatic QC Ranges
  • 120. Page 120 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 121. Page 121 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 122. Page 122 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 123. Page 123 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Patient Ranges Ingrese los rangos para los resultados de pacientes.
  • 124. Page 124 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 125. Page 125 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Patient Demographics Selecciones los demográficos para paciente, puede activar la rápida identificación de la muestra.
  • 126. Page 126 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 127. Page 127 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Sample Demographics - Screen 1 Seleccione los demográficos de la muestra.
  • 128. Page 128 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Sample Demographics - Screen 2
  • 129. Page 129 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 130. Page 130 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Parameter Selection Permite al operador seleccionar parámetros de medida para ser medidos en la pantalla de proceso. Paneles personalizados , permite configurar hasta 6 paneles. Un panel puede ser seleccionado como predeterminado.
  • 131. Page 131 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 132. Page 132 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 133. Page 133 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample
  • 134. Page 134 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Sample > Sample Type Seleccione el tipo de muestra por defecto.
  • 135. Page 135 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters
  • 136. Page 136 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup >Parameters > Parameters On/Off Screen 1 Seleccione los parámetros a ser reportados por el sistema.
  • 137. Page 137 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup >Parameters > Parameters On/Off Screen 2
  • 138. Page 138 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup >Parameters > Parameters On/Off Screen 3
  • 139. Page 139 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters
  • 140. Page 140 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters > Parameter Units Screen 1 Seleccione las unidades de medida de los parámetros.
  • 141. Page 141 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters > Parameter Units Screen 2
  • 142. Page 142 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters
  • 143. Page 143 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters > Demographic Units screen 2
  • 144. Page 144 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters
  • 145. Page 145 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Parameters > Values Ingrese el valor de la presión atmosférica.
  • 146. Page 146 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 147. Page 147 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 148. Page 148 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options > Country Options Seleccione el lenguaje del sistema y el formato de la fecha.
  • 149. Page 149 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 150. Page 150 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options > Date & Time Ajuste la fecha y hora del sistema.
  • 151. Page 151 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 152. Page 152 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options > Sound Ajuste el volumen del sistema y el volumen del sonido de video.
  • 153. Page 153 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 154. Page 154 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Display Options Ajuste el birllo de la pantalla.
  • 155. Page 155 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options
  • 156. Page 156 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > System Options > Other Options Ingrese el nombre del sistema , teléfono de contacto para asistencia, y encienda el aviso de calibración pendiente.
  • 157. Page 157 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 158. Page 158 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 159. Page 159 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Printer Options Active o desactive la impresora, seleccione las características de impresión.
  • 160. Page 160 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 161. Page 161 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices> Bar Code Setup Active o desactive el lector de código de barras y seleccione el formato de ID del paciente
  • 162. Page 162 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup> Printer & Devices >Patient ID Bar Code
  • 163. Page 163 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices >Accession No. Bar Code
  • 164. Page 164 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices >Password Bar Code
  • 165. Page 165 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 166. Page 166 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Communications Configure las comunicaciones.
  • 167. Page 167 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices >Configure RS-232
  • 168. Page 168 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 169. Page 169 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Network Setup Configure las opciones de red.
  • 170. Page 170 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Printer Options
  • 171. Page 171 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 172. Page 172 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Email Setup Configure la funcionalidad de envío automático de Email.
  • 173. Page 173 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices
  • 174. Page 174 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Printer & Devices > Remote Viewer Configure la funcionalidad de acceso remoto.
  • 175. Page 175 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 176. Page 176 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 177. Page 177 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >System Access Define las restricciones de acceso a las opciones del sistema- No restringido, limitado y restringido. Acceso Restringido: todos los operadores deben ingresar un password para usar el sistema. Acceso limitado: Todos los operadores pueden realizar tareas de rutina, como analizar muestras y reemplazar cartuchos sin ingresar un password. Acceso no Restringido: Todos los operadores pueden realizar tareas de rutina sin entrar un password, y todos los operadores pueden acceder a las opciones de configuración.
  • 178. Page 178 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 179. Page 179 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >Operator Security Define la ID del operador, el password tienen 4 niveles de acceso. NB: Dos operadores IDs o dos passwords pueden ser idénticos.
  • 180. Page 180 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >System Access
  • 181. Page 181 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >System Access
  • 182. Page 182 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >System Access
  • 183. Page 183 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 184. Page 184 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >Analysis Options Seleccione Editar Demográficos, Salvar Demográficos, Mostrar resultados con marca ?, Lista de pacientes y Rangos analíticos.
  • 185. Page 185 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 186. Page 186 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >Interference Correction Seleccione la opción de corrección de muestra venosa mixta. Se activa solo para reportar PO2 y tHB .
  • 187. Page 187 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 188. Page 188 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options >Screen Calibration Calibre el touch screen
  • 189. Page 189 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options
  • 190. Page 190 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Salvar la configuracion en una usb. Secured Options > Save Setup
  • 191. Page 191 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options
  • 192. Page 192 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options > Restore Setup Restaura los datos de calibracion desde una USB
  • 193. Page 193 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options
  • 194. Page 194 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options > Software Installation Instala una nueva versión de software del sistema.
  • 195. Page 195 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options > Software Installation
  • 196. Page 196 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options > Software Installation
  • 197. Page 197 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Setup > Secured Options
  • 198. Page 198 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Secured Options > Correlation Coefficients Ajusta los valores de pendiente y offset
  • 199. Page 199 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 200. Protection notice / Copyright notice RAPIDPoint 500 Recall Screen
  • 201. Page 201 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Status Screen
  • 202. Page 202 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 203. Page 203 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Patients Muestra un archivo que almacena hasta 250 muestras de paciente que han sido corridas.
  • 204. Page 204 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Patients
  • 205. Page 205 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Patients
  • 206. Page 206 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Patients
  • 207. Page 207 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 208. Page 208 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 209. Page 209 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> QC Muestra un archivo que contiene los últimos 250 QC resultados que hayan sido corridos.
  • 210. Page 210 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> QC
  • 211. Page 211 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 212. Page 212 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 213. Page 213 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 214. Page 214 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 215. Page 215 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Calibrations Muestra y grafica la fecha y hora de las ultimas 250 calibraciones realizadas.
  • 216. Page 216 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 217. Page 217 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Events Log Muestra un archivo que contiene los ultimos 250 eventos que han ocurrido en el sistema.
  • 218. Page 218 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 219. Page 219 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Sample Totals Muestra el total de muestras procesadas por cada analito.
  • 220. Page 220 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Sample Totals
  • 221. Page 221 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only No es informacion para el cliente. Ingrese usando el Service Password, XXXXX52637, donde XXXXX son los 5 digitos del serial del instrumento. – ESTE PASSWORD NUNGA DEBE SER DADO A LOS CLIENTES. Recall Screen > Sample Totals > Service Data FTP Server - Enter IP address of RAPIDComm with dedicated line for data collection
  • 222. Page 222 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen> Sample Totals
  • 223. Page 223 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen > Sample Totals > Diagnostics No es para los usuarios. Ingrese usando el Service Password FS diagnósticos pueden ser realizados acá.
  • 224. Page 224 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Simulator  Measurement Test  Nota: El sistema es probado 3 veces y por cada canal que es seleccionado por un switch de posición en la parte frontal del cartucho simulador de sensores. El sistema debe ser probado rotando el switch para seleccionar las posiciones HI, Norm y LOW.  Instale el cartucho simulador de sensores.  Low  Coloque el Cartucho de simulación en LOW.  Presiones los resultados serán impresos.  Verifique los datos y asegúrese que el numero serial del cartucho de simulación este impreso.  Fíjese en la impresión en la columna ‘L’ Chequee los valores actuales para toda la columna:  ‘0’ = Pass  ‘F’ = Fail  Para los canales 6 & 7 chequee si los valores están dentro de los limites que se encuentran en la tabla. Rechace si cualquier unidad esta por fuera de los rangos.
  • 225. Page 225 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Simulator  Test de medicion  Normal  Posicione el Cartucho de simulación de sensores en Norm.  Presione e los resultados serán impresos.  Fíjese en la impresión en la columna ‘N’ Chequee los valores actuales para toda la columna:  ‘0’ = Pass  ‘F’ = Fail  Para los canales 6 & 7 chequee si los valores esta en los limites de la tabla. Rechace si cualquier unidad esta por fuera de los rangos.
  • 226. Page 226 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Simulator test de medicion  High  Posicione el cartucho de simulación de sensores en la posición High.  Presione los resultados serán impresos.  Fíjese en la impresión en la columna ‘H’ Chequee que los valores actuales para toda la columna:  ‘0’ = Pass  ‘F’ = Fail  Para los canales 6 & 7 chequee que los valores estén en los rangos de la tabla. Rechace si alguna unidad esta fuera del rango.
  • 227. Page 227 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Simulator Diagnostic Measurement Test Printout Name Mux channel -LOW limit +LOW limit -MED limit +MED limit -HIGH limit +HIGH limit 1 GND 1 -0.25 0.25 -0.25 0.25 -0.25 0.25 2 PO2 2 -25.00 25.00 790.00 970.00 2900.00 3600.00 3 HCO3 3 -1382.00 -968.00 -119.00 347.00 1145.00 1662.00 4 GLUA 4 -50.00 50.00 -460.00 -340.00 -3700.00 -3000.00 5 GLUB 5 -50.00 50.00 -460.00 -340.00 -3700.00 -3000.00 6 LACA 6 -50.00 50.00 -345.00 -255.00 -2750.00 -2250.00 7 LACB 7 -50.00 50.00 -345.00 -255.00 -2750.00 -2250.00 8 PH1 8 -1229.00 -809.00 -246.00 214.00 732.00 1232.00 9 K 9 -787.00 -349.00 39.00 511.00 870.00 1376.00 10 CL 10 -79.00 388.00 422.00 910.00 924.00 1432.00 11 NA 11 -649.00 -200.00 132.00 608.00 904.00 1411.00 12 CA 12 -173.00 291.00 594.00 1089.00 1051.00 1882.00 13 REF 13 -200.00 200.00 -200.00 200.00 -200.00 200.00 14 DAC 14 0.00 0.00 0.00 0.00 0.00 0.00 15 PH2 15 -1382.00 -968.00 -119.00 347.00 1145.00 1662.00 16 MG 16 NA NA NA NA NA NA 17 17 0.00 0.00 0.00 0.00 0.00 0.00 18 FD1 FD1 -350.00 650.00 750.00 1818.00 1850.00 3148.00 19 FD2 FD2 -350.00 650.00 750.00 1818.00 1850.00 3148.00 20 PRE NA NA NA NA NA NA 21 BLK NA NA NA NA NA NA 22 SAMP NA NA NA NA NA NA
  • 228. Page 228 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen > Sample Totals > Diagnostics
  • 229. Page 229 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Valve Alignment
  • 230. Page 230 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Valve Alignment •Asegúrese que la herramienta de alineación de la válvula este asegurada sobre la base con el tornillo. •Asegúrese que el tornillo manual este ajustado usando un destornillador plano. •Verifique que el indicador de la válvula marque cero ( 0 ± 0.0005) •Si la lectura de la válvula esta fuera de lo especificado, suelte el tornillo y ajuste el dial en cero. •Re-Ajuste el tornillo . Asegúrese que la medida continua en cero ± 0.0005 •Repita el proceso (para verificar repetibilidad. •Asegúrese que la medición sea repetible cero ± 0.0005.
  • 231. Page 231 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only M Valve Test Instale la herramienta de alineación de la válvula en la interface del cartucho del sistema. Presione el embolo para ver que no golpee con la válvula M cundo este ajustada. Asegúrese que la herramienta de ajuste este asegurada. Verifique que la herramienta esta ajustada usando un destornillador .Sea cuidadoso de no apretar demasiado porque puede dañar el tornillo. Presione Presione la válvula se mover a la primera posición. Presione la válvula se moverá a la posición 2 Verifique que la medición este 0 + 5. Continúe, presionando hasta que la válvula retorne a la posición 2. Confirme que la lectura se mantenga en 0 + 5. Presione espere a que el contador baje. Regrese a la pantalla de diagnósticos. Remueva la herramienta de alineación de la interface del cartucho. Regrese la herramienta a su base. Troubleshooting > Valve Alignment
  • 232. Page 232 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen > Sample Totals > Diagnostics
  • 233. Page 233 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Temperature Heater Test Ingrese al modo de diagnósticos. Presione los resultados serán impresos. Verifique que el campo 2 esta en : MAIN. Asegúrese que el campo 3 es mayor a 300 NOTE:Desde el arranque (frio), la temperatura deberá estabilizarse por 25 minutos. Desde un inicio a temperatura, la temperatura deberá estabilizarse durante 15 minutos antes de que el test sea iniciado. Luego de este tiempo presione TEMPERATURE. El STEADY conteo, campo 3, deberá estar por encima de 300 antes de chequear las lecturas. Es aceptable tener que repetir la medición para asegurarse que el campo 3 es mayor a 300.
  • 234. Page 234 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Temperature Field Description Lower Limit Upper limit 1 NA NA NA 2 MAIN (If COV printed – close the door) MAIN MAIN 3 Steady State control 300 NA 4 DAC MAIN BLOCK NA NA 5 BLOCK TEMP 37.4 39.6 6 DAC Setting for Pre-Heater NA 700 7 PREHEATER 37.0 38.5 8 SAMPLE TEMP 36.85 37.15 9 CO-ox Sample Temp 36.65 37.35 Compare the temperatures with the following table.
  • 235. Page 235 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen > Sample Totals > Diagnostics
  • 236. Page 236 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > Wavelength Wavelength Desde la pantalla de diagnósticos presione. Asegúrese de lo siguiente: Que el campo 1 es mayor a 1000 Que el campo 5 es menor que ± 1.0 Que el campo 6 es menor de 50000 Que el reporte de impresión reporte PASS FIEL D Description Low er Limit Upper Limit 1 Mean Peak Counts (Average of Multiple Scans) 1000 NA 2 Standard Deviation of Peak Counts NA NA 3 Centroid Measured from Instrument on-board Neon NA NA 4 Centroid Stored from Factory Alignment Polychromator NA NA 5 Measured Wavelength shift in Pixels fields 3 - 4 -1.0 1.0 6 CMB Integration Time from last Zero Calibration NA 50000 7 Polychromator Serial Number NA NA
  • 237. Page 237 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 238. Page 238 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Software
  • 239. Page 239 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 240. Page 240 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Software Copia 3 meses de datos de diagnostico a una USB. Envíe por Email a siemens para un M-Cart. Remplazo y troubleshooting.
  • 241. Page 241 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 242. Page 242 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Software 3 Types of Sensor Data Type 1 – Gives Sensor Performance Data Type 2 – Gives CO-Ox Performance Data Type 0 – Gives Software Performance Data
  • 243. Page 243 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Recall Screen
  • 244. Page 244 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 245. Protection notice / Copyright notice RAPIDPoint 500 Hardware
  • 246. Page 246 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Enhanced Communication Options Fast Time to Patient Results Extended Test Menu to Include: Lactate & Neonatal Total Bilirubin Integrated Barcode Reader RAPIDComm and LIS Connectivity Options Proven AQC Application
  • 247. Page 247 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Ease of Use – Integrated Barcode Reader
  • 248. Page 248 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Provides Positive Identification Offers Multiple Configurations Easily Accessible Ensures Data Integrity Meets Hospitals Needs Now and the Future Offers Convenience Ease of Use – Integrated Barcode Reader
  • 249. Page 249 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only 2D 1D Positive Patient ID Positive Operator ID 2D - Enhanced Information Possibilities Compatible with common formats Ease of Use – Integrated Barcode Reader
  • 250. Page 250 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware On Board Barcode Scanner
  • 251. Page 251 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Remove the Allen hinge pin securing the front door at the bottom, and remove the front door. Lay the door on its front side
  • 252. Page 252 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Remove the 4 screws securing the barcode reader housing. The end of the housing arm hooks onto a rib on the door
  • 253. Page 253 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Gently disconnect the flex cable from the barcode reader interface board, then remove the 2 screws securing the barcode reader interface board to its housing and disconnect the USB cable from the interface board.
  • 254. Page 254 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Remove the 2 screws securing the cartridge interface assembly and pull it slightly forward.
  • 255. Page 255 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Note the routing of the BCR cable through the notch in the Cartridge Interface Assembly.
  • 256. Page 256 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware USB Board with On-Board Barcode Scanner connected
  • 257. Page 257 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware 3 Available USB Ports
  • 258. Page 258 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Remove the rear cover and power supply module Remove the 2 screws securing the USB board.
  • 259. Page 259 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Replacing the Printer Rotate the paper cover down.2. Loosen the one captive screw at the top of the printer cover, and rotate the printer cover down.
  • 260. Page 260 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Loosen 2 captive printer screws at the top of the printer.
  • 261. Page 261 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Remove 2 screws at the bottom Use a flat-bladed screwdriver to pry off the paper advance knob Disconnect the printer ribbon cable and remove the printer.
  • 262. Page 262 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Main Board Assembly (New to RP500)
  • 263. Page 263 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware Display Connection to Main Board
  • 264. Page 264 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware New Style Hard Drive (RP500 Only)
  • 265. Page 265 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Hardware
  • 266. Page 266 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 267. Protection notice / Copyright notice RAPIDPoint 500 Routine System Procedures
  • 268. Page 268 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only • Reemplace el Wash/Waste Cartridge • Reemplace el Measurement Cartridge • Reemplace el Automatic QC Cartridge • Reemplace el Sample Port • Limpie y desinfecte la pantalla. • Limpiar y desinfectar las superficies exteriores • Vacíe el recipiente de ampolletas. • Reemplace el papel de la impresora. • Reemplace los filtros de aire. Routine System Procedures
  • 269. Page 269 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Cleaning & Disinfecting the Screen and Exterior Surfaces • Humedezca un limpión que no suelte hilos con Solución de hipoclorito de sodio al 0.5% • Escoja el icono Clean Screen de la pantalla de Status. • Mientras se limpia la pantalla son mostrados (20 segundos), Limpie la pantalla con el limpión húmedo y séquela. • Limpie la superficie exterior del sistema con un limpión húmedo y seque las superficies.
  • 270. Page 270 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Emptying the Ampule Breaker • Remueva el contenedor de ampolletas del sistema. • Vacíe el contenedor de ampolletas en un sitio adecuado.(guardián) • Si es necesario , limpie el interior del contenedor de ampolletas con una solución blanqueadora al 10% • Reinstale el contenedor de ampolleta
  • 271. Page 271 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the Printer Paper
  • 272. Page 272 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Replacing the Air Filter •Hale el seguro del filtro del aire del sistema. •Remueva el filtro de aire •Instale un nuevo filtro de aire. •Reinstale el seguro del filtro del aire al sistema.
  • 273. Page 273 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 274. Protection notice / Copyright notice RAPIDPoint 500 Networking and Connectivity
  • 275. Page 275 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDComm Information Management System
  • 276. Page 276 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RAPIDSystems Product Portfolio
  • 277. Page 277 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Connectivity Real Time View
  • 278. Page 278 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Compatible with all current Siemens Diagnostics Critical Care Systems Remote monitoring and control functions Built on Standard Technology  Microsoft® XP Operating System  Microsoft® SQL Server Database Key Features
  • 279. Page 279 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Management  Provides powerful management tools to improve laboratory operations and lower costs. Clinical  Provides clinicians with critical care results, associated reference ranges, hospital and patient demographics, a clinical interpretation of results, and the last four patient samples from the database -- All in a single concise report. Regulatory  All Quality Control, Maintenance, and Calibration records are stored and readily accessible for inspection purposes. The program also schedules maintenance, monitors compliance, and logs activity. User  Intuitive point-and-click design, and online training module for easy operation. Key Benefits
  • 280. Page 280 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 281. Protection notice / Copyright notice RAPIDPoint 500 Patient Sample Analysis
  • 282. Page 282 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Collecting Patient Samples > Fill Volumes Sample Type Collection Collection Device Minimum Fill Volume Preparation Arterial blood, venous blood, or mixed venous blood syringe • Expel air from the syringe and cap it immediately after obtaining the sample 1 mL 200 μL 3 mL 800 μL • Do not use cork to cap the syringe 5 mL 1.5 mL • Mix the sample thoroughly by rolling the syringe between your palms and gently inverting it several times. capillary blood capillary tube 100 μL • Fill the tube completely and cap it securely • Do not use clay or cork to cap the tube • Do not use capillary tubes containing mixing fleas
  • 283. Page 283 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Patient Samples •Mezcle la muestra. •Seleccione el tipo de muestra a analizar en el sistema. •Inserte la jeringa o capilar en el puerto de muestra y seleccione Start. •Remueva la muestra del dispositivo cundo se le indique y seleccione continuar. (botón a la derecha de la pantalla). •Ingrese los demográficos del paciente o use la ID del paciente (host Query) y luego seleccione Continuar. •View results.
  • 284. Page 284 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis Syringe, Capillary, Venous
  • 285. Page 285 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis Syringe, Capillary, Venous
  • 286. Page 286 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis Syringe, Capillary, Venous
  • 287. Page 287 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Mixed Venous Sample Analysis
  • 288. Page 288 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis > Remove Sample Device
  • 289. Page 289 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis > Enter Demographics
  • 290. Page 290 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis > Keyboard
  • 291. Page 291 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis > Analyzing
  • 292. Page 292 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Sample Analysis > Results
  • 293. Page 293 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A
  • 294. Protection notice / Copyright notice RAPIDPoint 500 QC Analysis
  • 295. Page 295 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only RapidQC Complete An aqueous control for pH/blood gas, electrolyte, CO-oximetry, and metabolite analysis • Storage • 24 months from date of manufacture, at room temperature • Catalog Numbers • Level 130ampules x 2.5mL/box SMN 10309925 • Level 230ampules x 2.5mL/box SMN 10309926 • Level 330ampules x 2.5mL/box SMN 10309927
  • 296. Page 296 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only •Touch the button on Analysis Screen. •Scan the barcode on the QC ampule. •Open the ampule and attach the Quick adapter •Insert the Quick adapter with attached ampule into the sample port and touch Continue. •Remove when prompted and touch Continue. •View results. •Parameters that fail Required QC will not be available for sample analysis. Analyzing Required QC Samples
  • 297. Page 297 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Required QC Samples
  • 298. Page 298 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Required QC Samples > Failed
  • 299. Page 299 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled QC •Touch the QC ampule Button on the Analysis Screen •Scan the barcode on the QC ampule •Open the ampule and attach the Quick adapter •Insert the Quick adapter with attached ampule into the sample port and touch Analyze •Remove when prompted and touch Continue •View results •Parameters that fail Unscheduled QC will be available for sample analysis
  • 300. Page 300 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled QC
  • 301. Page 301 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled QC
  • 302. Page 302 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled QC
  • 303. Page 303 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled AQC Touch the AQC Button on the Analysis Screen Touch Analyze Touch the level(s) to be run and touch Start View results Parameters that fail AQC will not be available for sample analysis
  • 304. Page 304 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled AQC
  • 305. Page 305 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled AQC
  • 306. Page 306 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled AQC
  • 307. Page 307 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Analyzing Unscheduled AQC
  • 308. Protection notice / Copyright notice RAPIDPoint 500 Calibrations
  • 309. Page 309 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only One Point Calibration • Occurs every 30 minutes • 2 minutes to complete • pH, pO2, pCO2, Na+, K+, Cl-, Ca++, and Glu. • The system measures the 200 Cal calibrator on all sensors and adjusts for cal. drift of all analytes. Calibrations
  • 310. Page 310 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Two Point Calibration • Occurs every 2 hours • 3 minutes 45 seconds to complete • pH, pO2, pCO2, Na+, K+, Cl-, Ca++, Glu, and CO-ox Zero • The system measures a combination of two calibrators, 200 Cal and Zero Cal, and adjusts for cal. and slope drifts of all analytes. • If the zero calibration for the CO-ox exceeds drift limits, CO-ox zero occurs during the 1 pt. Cal until the drift error is cleared. Calibrations
  • 311. Page 311 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Full Two Point Calibration • Occurs every 8 hours • 5 minutes 30 seconds to complete • All functions of a One and Two Point Cal and CO-ox Slope. • If the zero calibration for the CO-ox exceeds drift limits, CO-ox zero occurs during the 1 pt. Cal until the drift error is cleared. Calibrations
  • 312. Page 312 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Interruptible Calibrations Touch the STAT button to abort the calibration. The system immediately performs a wash cycle.
  • 313. Page 313 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Interruptible Calibrations System will decide if it’s quicker to stop the calibration and wash or finish the calibration.
  • 314. Page 314 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only The dialogue box will NOT contain a STAT button next to the countdown clock. Un-Interruptible Calibrations
  • 315. Page 315 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Cartridge Initialization Timeline Calibration # Sequence Time elapse (min) Time since last cal (min) Interruptible / Uninterruptible 1 Full Two Point - - Uninterruptible 2 Two Point - - Uninterruptible 3 Full Two Point 20 Uninterruptible 4 Two Point 44 20 Uninterruptible 5 One Point 66 20 Interruptible 6 One Point 88 20 Interruptible 7 One Point 110 20 Interruptible 8 Two Point 133 20 Uninterruptible 9 One Point 155 20 Interruptible 10 One Point 177 20 Interruptible 11 One Point 200 20 Interruptible 12 Full Two Point 225 20 Uninterruptible
  • 316. Page 316 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only The initialization and retrospective calibration (Retrocal) period is not complete until either 7.5 hours have passed or all of the following are true: • Two scheduled calibrations have occurred after cal #12 • Both calibrations passed the first time with no need to repeat • No samples were run between the two scheduled calibrations So the Retrospective calibration period can be between 3.5 hours and 7.5 hours depending upon system performance Cartridge Initialization Timeline
  • 317. Page 317 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Retrospective Calibration Sample During the initialization period, if a patient sample is run the following happens: • The sample is analyzed • The system washes out the sample with 200 cal and loads 200 cal over the sensors • The system performs a one point calibration • Sample results are adjusted based on the calibration drift The system will also perform this sequence under certain conditions like detection of interfering substances.
  • 318. Page 318 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Normal Sample If a patient sample is run outside of the retrocal period the following happens:  Loads patient sample  Performs electronic measurement of sample  Calculates patient data and sends it to UIP  Runs normal wash The sample is analyzed
  • 319. Protection notice / Copyright notice RAPIDPoint 500 Troubleshooting
  • 320. Page 320 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Troubleshooting > D Codes D2 - Excessive Drift • System will recalibrate 2X • Options are to manually initiate a cal, wait for the next cal, or remove M/C D3 - Slope Error • System will recalibrate 2X • Options are to manually initiate a cal , wait for the next cal, or remove M/C D4 - Offset Drift • System will recalibrate 2X • Options are to manually initiate a cal , wait for the next cal, or remove M/C
  • 321. Page 321 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D21 - Processing Error • Reboot D23:2 or 10 - No Wash detected at FD1 or FD2, respectively • System will reload the Wash up to 2 more times • If error not cleared, system prompts for W/W replace D23:3 or 11 - No 200 Cal detected at FD1 or FD2, respectively • System will reload the 200 Cal up to 2 more times • If error not cleared, system prompts for M/C replace Troubleshooting > D Codes
  • 322. Page 322 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D23:4 or 12 - No Zero Cal detected at FD1 or FD2, respectively • System will reload the Zero Cal up to 2 more times • If error not cleared, system prompts for M/C replace D23:5 or 13 - No RCx detected at FD1 or FD2, respectively • System will reload the RCx up to 2 more times • If error not cleared, system prompts for M/C replace Troubleshooting > D Codes
  • 323. Page 323 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D23:7 or 8 - No Wash segment detected at either FD1 or FD2 • Repeat Wash up to 2 more times • If error does not clear, system prompts for W/W replace D23:9 - No foam Wash detected at FD1 • System initiates aggressive wash to clear clot and checks FD1 up to 2 more times • If error does not clear after 3rd try, system prompts for sample port replacement • Error common on new M/C - with more blood less common Troubleshooting > D Codes
  • 324. Page 324 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D33:1 Valve Error • System must be restarted to clear D35 - Electronics Error • System must be restarted to clear • D35:14 Door Error - check for something blocking the door D38 - Temperature Error • Qualifiers 1, 2, 3, 6, & 12 - Fatal error • Qualifiers 4 & 5 - Non-fatal • System may clear error and will need to be restarted • Check temperature of environment • Qualifier 3 & 6 out of box - leave system running for 1 hour and then restart - error may clear Troubleshooting > D Codes
  • 325. Page 325 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D39 - Obstruction • System prompts for sample port change Troubleshooting > D Codes
  • 326. Page 326 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only D40 - Wash not Detected • Detected during the prime of a W/W cartridge • Prompts for W/W change up to 2X • If error not cleared after 3rd attempt, system prompts for M/C replace D60 - Communication Error • Check cable connections • Check IP addresses Troubleshooting > D Codes
  • 327. Page 327 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Bubble in Sample • Will prompt for sample port change Insufficient Sample • Will prompt for sample port change Refer to “System Messages” in the Troubleshooting and Calibrations section of the Service Manual for a complete list of all Diagnostic Codes and Corrective Actions Troubleshooting > Error Messages
  • 328. Protection notice / Copyright notice RAPIDPoint 500 Cartridge Overview
  • 329. Page 329 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Extensive Menu Full Complement of Tests in a Single Cartridge Long Measurement Cartridge Use Life Fully Customizable Patient Panels Lactate and Neonatal Total Bilirubin Total Bilirubin Methodology Spectrophotometric Lactate Methodology Lactate Oxidase
  • 330. Page 330 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Selection of reagents AQC zero RCX 200 cal wash Foam wash Syringe Cap Vent
  • 331. Page 331 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Bag Contents – M-Cart and Wash
  • 332. Page 332 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Bag Contents - AQC
  • 333. Page 333 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Calibration Points
  • 334. Page 334 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only
  • 335. Page 335 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Inside The Sensor Module Sample Serpentine flow preheater sensors Ref
  • 336. Page 336 Siemens Healthcare Diagnostics Siemens Confidential – Internal Use Only Q&A